1. or 0 025 0 05 in of span span psi span bar Performance specifications are based on reference conditions of 25 C 77 F zero 0 static pressure 10 to 55 RH and 316L Stainless Steel barrier diaphragm URL limit or maximum seal pressure rating whichever is lower Performance Under Rated Conditions Model STR17G 0 30 to 0 3 000 psig Parameter Description Upper Range Limit psig 3 000 bar 210 Minimum Span psig 30 bar 2 1 Turndown Ratio 100 to 1 Zero Elevation and Suppression No limit except minimum span from absolute zero to 100 of URL Specifications valid over this range Accuracy Reference Includes combined effects of linearity hysteresis and repeatability e Accuracy includes residual error after averaging successive readings For FOUNDATION Fieldbus use Digital Mode specifications For HART use Analog Mode specifications In Analog Mode 0 15 of calibrated span or upper range value URV whichever is greater For URV below reference point 300 psi accuracy equals 300 psi 1 bar 2 0 10 0 05 or 0 10 0 05 in of span span psi span bar In Digital Mode 0 125 of calibrated span or upper range value URV whichever is greater For URV below reference point 300 psi accuracy equals 300 psi 21 bar 0 075 0 05 OF 0 075 0 05
2. in of span span psi span bar Performance specifications are based on reference conditions of 25 C 77 F zero 0 static pressure 10 to 55 RH and 316L Stainless Steel barrier diaphragm Transmitter URL limit or maximum seal pressure rating whichever is lower ST 3000 Smart Pressure Transmitter Performance Under Rated Conditions STR14A 0 5 to 0 500 psia Parameter Description Upper Range Limit psia 500 bar absolute 35 Minimum Span psia 5 bar absolute 0 35 Turndown Ratio 100 to 1 Zero Elevation and Suppression No limit except minimum span from 0 to 100 URL Accuracy Reference Includes combined effects of linearity hysteresis and repeatability e Accuracy includes residual error after averaging successive readings For FOUNDATION Fieldbus use Digital Mode specifications For HART use Analog Mode specifications In Analog Mode 0 1 of calibrated span or upper range value URV whichever is greater For URV below reference point 20 psi accuracy equals 20 psi 1 4 bar 0 05 0 05 or 0 05 0 05 2 in of span span psi span bar In Digital Mode 0 075 of calibrated span or upper range value URV whichever is greater For URV below reference point 20 psi accuracy equals 20 psi 1 4 bar 0 025 0 05 or 0 025 0 05 in of span span psi span bar Perfo
3. ANSI 150 1 2 ANSI 300 ANSI 150 ANSI 300 ANSI 150 ANSI 300 1 Standard facina 125 250 AARH RF raised face serrated surface finish Plastic Pluas are TEMPORARY ONLY to protect threads and MUST be REMOVED before installation Note Remote seal system pressure rating is body rating or seal rating whichever is less Table ll continued next page 34 ST 3000 Smart Pressure Transmitter 34 ST 16U 32 Issue 50 Page 4 of 12 STR14A STR12D amp 13D 14G 8 17G TABLE ll SEALS continued I Diaphragm Lower Selection 316L SS 316L SS Hastelloy C 276 316L SS Hastelloy C 276 Hastelloy C 276 Monel 400 Monel 400 Tantalum 316L SS Tantalum Hastelloy C 276 Tantalum Clad operen sion EIS A Fel Mpp Carbon Steel 316L SS Flush Flanged No Selection AE Seal Flushing Connections One 1 4 with plastic plug and Plugs One 1 4 with metal plug Metal plug material Two 1 4 with plastic plugs will be the same as Two 1 4 with metal plugs Lower material if One 1 2 with plastic plug metal plug is chosen One 1 2 with metal plug SS Plug for CS Lower Two 1 2 with plastic plugs and Tantalum Clad Two 1 2 with metal plugs Klinger C 4401 non asbestos Gasket Grafoil Teflon Gylon 3510 Wetted Material with Lower Table Il continued below STR14A TABLE ll SEALS Jabra STR12D 8 13D 14G amp 17G ANSI Class 150 ae OD ANSI Class 300 extension DIN DN80 PN40 4 ANSI Class 150
4. 10 Physical and Approval Bodies ST 3000 Smart Pressure Transmitter Parameter Description Process Interface See Model Selection Guide for Material Options for desired seal type Seal Barrier Diaphragm 316L Stainless Steel Monel Hastelloy C Tantalum Seal Gasket Materials Klinger C 4401 non asbestos Grafoil Teflon Gylon 3510 Mounting Bracket Carbon Steel Zinc Chromate plated or Stainless Steel Fill Fluid Meter Body Silicone DC 200 CTFE Chlorotrifluoroethylene S G 25 C 0 94 S G 25 C 1 89 Fill Fluid Secondary Silicone DC 200 CTFE Chlorotrifluoroethylene Silicone Oil 704 Syltherm 800 NEOBEE M 20 S G 25 C 0 94 S G 25 C 1 89 S G 25 C 1 07 S G 25 C 0 90 S G 25 C 0 93 Electronic Housing Epoxy Polyester hybrid paint Low copper aluminum alloy Meets NEMA 4X watertight and NEMA 7 explosion proof Stainless steel optional Capillary Tubing Armored Stainless Steel or PVC Coated Armored Stainless Steel Length 5 10 15 20 25 and 35 feet 1 5 3 4 6 6 1 7 5 and 10 7 meters A 2 inch 51 millimeter S S close coupled nipple is also available See Model Selection Guide Refer to Figure 14 for guide to maximum capillary length vs diaphragm diameter Wiring Accepts up to 16 AWG 1 5 mm diameter Mounting See Figure 17 Dimensions Transmitter See Figures 2
5. 3 70 OD ANSI Class 300 extension DIN DN100 PN40 Flange Seal 316L SS 316L SS with Extended adc Hastelloy C 276 316L 22 Diaphragm Hastelloy C 276 Non Wetted CS Nickel Plated Material flange 316L 2 gt E TEE A lele Extension Length qu 6_ No Selection No Selection No Seeds Ff of ole Table Il continued next page 1 Standard facina 125 250 AARH RF raised face serrated surface finish 4 Plastic Pluas are TEMPORARY ONLY to protect threads and MUST be REMOVED before installation Bolt material will be same as Upper Material However if Table III bolt nut option is chosen seal bolt material will be the same Note Remote seal system pressure rating is body rating or seal rating whichever is less ST 3000 Smart Pressure Transmitter 34 ST 16U 32 Issue 50 Page 5 of 12 STR14A TABLE II SEALS continued STR12D amp 13D 3 14G amp 17G 316L SS 316L SS Hastelloy C 276 316L SS Hastelloy C 276 Hastelloy C 276 Monel 400 Monel 400 Wetted Material Tantalum Tantalum RonWetedWetefal No Sesion 8 fo Po fw a A A ctm PM CIC Calibration Rings None 316L SS Pancake Seal Hastelloy C 276 Monel 400 Flushing Connections One 1 4 with plastic plug and Plugs One 1 4 with metal plug Metal plug material Two 1 4 with plastic plugs will be the same as Two 1 4 with metal plugs Cal Ring material if One 1 2 with plastic plug metal plug is chosen One 1 2 with metal plug SS Plug for CS
6. 50 to 93 C acd NE 7 o Ex d IIC Enclosure IP 66 67 T6 Ta 50 to 78 C T4 Ta 50 to 93 C 4 20 mA DE T5 Ta 50 to 85 C IECEx T6 Ta 50 to 70 C International Electrotechnical E Commission Intrinsically Safe Zone 0 1 T4 Ta 50 to 93 C LCIE 4 20 mA HART T5 Ta 50 to 63 C Ex ia IIC Enclosure IP 66 67 T6 Ta 50 to 48 C Fieldbus T3 Ta 50 to 93 C Not FISCO T4 Ta 50 to 40 C Li 0 except Li 1504H when Option ME Analog Meter is selected DN Type of Protection Comm Option Field Parameters Temp Codes Flameproof Zone 1 All All T5 Ta 50 to 93 C Ex d IIC Enclosure IP 66 67 T6 Ta 50 to 78 C Ui 30V li 100mA T4 Ta 50 to 93 C 4 20 mA DE Ci 4 2nF T5 Ta 50 to 85 C Li T6 Ta 50 to 70 C Pi 1 2W s T4 Ta 50 to 93 C dll td aoe eae e 4 20 mA HART T5 Ta 50 to 63 C x la IIC Enclosure T6 Ta 50 to 48 C Fieldbus T3 Ta 50 to 93 C Not FISCO T4 Ta 50 to 40 C tier Multiple Marki Ui 30V ultiple Marking i D li 100mA T4 Ta 50 to 93 C Flameproof Zone 1 4 20 mA DE Ci 4 2nF T5 Ta 50 to 85 C f pm L1 O Ex d IIC Enclosure IP 66 67 T6 Ta 50 to 70 C Pi 1 2W Intrinsically Safe Zone 0 1 gd Ex ia IIC Enclosure IP 66 67 I 100m a 50 to 4 20 mA HART Ci 4 2nF T5 Ta 50 to 63 C C gt _ O The user must determine the type of 3 DA T6 Ta 50 to 48
7. Lower Housing i Ya Wetted Gasket Diaphragm Wetted Wetted 3 Saddle Seal Instr Conn 489 fi j Upper Body Non wetted Lower Housing if Wetted Gasket Diaphragm Wetted Wetted 4 Saddle Seal Calibration Ring Type Size Rating Dimension 1 ANPT 1 2NPT A 5 00 5 00 Calibration ARS 150 800 B 1 00 1 50 B ris C 3 00 3 00 ZA Calibration Ring 30 Options e High Accuracy Option HA Extends applicable S100 models to 0 025 analog reference accuracy e Angle Mounting Bracket Options MB MX SB SX FB The angle mounting bracket is available in either zinc plated carbon steel or stainless steel and is suitable for horizontal or vertical mounting on a two inch 50 millimeter pipe as well as wall mounting An optional flat mounting bracket is also available in carbon steel for two inch 50 millimeter pipe mounting An option also exists for Marine approved mounting brackets used with Marine certification options e Indicating Meter Options ME and SM Two integral meter options are available An analog meter option ME is available with a dual O to 10 square root and O to 100 linear scale The Smart Meter option SM provides an LCD display for both analog and digital output and can be configured to display pressure in selected engineering units e HART Output Protocol Options HC and H6 Optional electronic modules for the ST 3000 provi
8. C protection required for installation of i the equipment The user shall then check the box V adjacent to the type of protection used on the equipment certification nameplate Once a type of protection has been checked on the nameplate subsequently the equipment shall not be reinstalled using any of the other certification types Fieldbus T3 Ta 50 to 93 C Not FISCO T4 Ta 50 to 40 C Li 0 except Li 150uH when Option ME Analog Meter is selected 14 ST 3000 Smart Pressure Transmitter Type of Protection Comm Option Field Parameters Flameproof Zone 0 111 D Ex tD All Al Enclosure IP 66 67 Temp Codes A20 IP6X T95 C Ta 93 C or T80 C Ta 78 C T5 Ta 50 to 93 C Flameproof Zone 1 T6 Ta 50 to 78 C 91I 2 GD Ex d IIC Ex tD All All OW DEX OBOE 4 ake Enclosure IP 66 67 or T80 C Ta 78 C Ui 30V li 100mA T4 Ta 50 to 93 C 4 20 mA DE Ci 4 2nF T5 Ta 50 to 85 C Li T6 Ta 50 to 70 C Pi 1 2W Intrinsically Safe Zone 0 1 Ui 30V T li 100mA 9II 1 G Ex ia lic 4 20 mA HART Ci 4 2nF Li Pi 1 2W Enclosure IP 66 67 Fieldbus T4 Ta 50 to 93 C T5 Ta 50 to 63 C T6 Ta 50 to 48 C T3 Ta 50 to 93 C T4 Ta 50 to 40 C Ui 30V li 100mA T4 Ta 50 to 93 C 4 20 mA DE Ci 4 2nF T5 Ta 50 to 85 C Li T6 Ta 50 to 70 C Pi 1 2W Not FISCO Non Sparking Zone 2 ATEX
9. Il IIl Division 1 Groups A B C D E F G locations Enclosure Type 4X Imax 225mA 4 20 mA HART Ci 4 2nF T4 Ta 93 C Li Pi 1 2W Vmax 32V Imax 120mA Ci 4 2nF Li 0 Fieldbus Entity Not FISCO T4 Ta 40 C T3 Ta 93 C Pi 0 84W Intrinsically Safe Vmax 24V Imax 250mA Ci 4 2nF Li 0 Pi 1 2W Vmax 17 5V Imax 380mA FISCO Ci 4 2nF Li 0 Pi 5 32W Vmax 42 4V Imax 225mA 4 20 mA DE Ci 4 2nF T4 Ta 93 C Nonincendive Li Pi 1 2W Class Division 2 Groups A B C D Vmax 30V locations Enclosure Type 4X Imax 225mA 4 20 mA HART Ci 4 2nF T4 Ta 93 C Li Pi 1 2W Class Il IIl Division 1 Groups A B C D E F G locations Fieldbus Entity Not FISCO T4 Ta 40 C T3 Ta 93 C FM Class 1 Zone 0 AEx ia Group IIC Approvals Enclosure Type 4X IP 66 67 T4 Ta 40 C T3 Ta 93 C Vmax 32V Nonincendive Fieldbus Entity las oe T4 Ta 40 C Not FNICO Yu T3 Ta 93 C Class l Division 2 Groups A B C Pi 0 84W D Vmax 24V Suitable for Fieldbus Entity alles oo TA Ta 40 C Class II Division 2 Groups F amp G Not FNICO eo T3 Ta 93 C Class III Division 2 Pi 1 2W Class Zone 2 Group IIC Enclosure Type 4X IP 66 67 FNICO T4 Ta 40 C T3 Ta 93 C Li 0 except Li 1504H when Option ME Analog Meter is selected FM Approvals is a service mark of FM Global 12 S
10. control systems provides local measurement accuracy to the system level without adding typical A D and D A converter inaccuracies e Unique piezoresistive sensor automatically compensates input for real world temperature and static pressure variations e Added smart features include configuring lower and upper range values simulating accurate analog output and selecting preprogrammed engineering units for display e Smart transmitter capabilities with local or remote interfacing means significant manpower efficiency improvements in commissioning start up and ongoing maintenance functions e ST 3000 transmitters feature full Dual Seal certification based on ANSI NFPA 70 202 and ANSI ISA 12 27 01 requirements without the use of additional seal protection elements e ST 3000 transmitters are available fully compliant to SIL 2 3 requirements as a standard option Advanced Diagnostics ST 3000 is now available for both HART 6 and Foundation Fieldbus with advanced diagnostics that minimize unplanned plant outages minimize maintenance costs and by providing the industry s most reliable transmitter e Provide advanced warning of possible failure events and avoid costly shutdowns e Three levels of failure reporting e Comprehensive list of on board diagnostics Ref ST 3000 User manual with HART 6 34 ST 25 17 Rev June 09 and Foundation Fieldbus option manual 34 ST 25 15 Rev June 09 ST 3000 Smart Pressure Tran
11. i il l i 18 plus receiver resistance y Operating rea o 10 6 16 25 20653 25 28 3 53570 424 Operating Voltage Vac Figure 16 Supply voltage loop resistance chart LP Side Maximum lev el H2 Fixed ref leg Minimum lev el HP Side 24249 NOTE Lower flange seal should not be mounted over 22 feet below or above the transmitter for silicone fill fluid 11 feet for CTFE fill fluid with tank at one atmosphere The combination of tank vacuum and high pressure capillary head effect should not exceed 9 psi vacuum 300 mmHg absolute Consult Honey well for installation of STR13D Figure 17 The ST 3000 transmitter with remote diaphragm seals shown mounted on a tank 19 Variable Head H 1 20 Application Data Liquid Level Closed Tank Determine the minimum and maximum pressure differentials to be measured Figure 18 PMin SGp x a SGf x d LRV when HP at bottom of tank URV when LP at bottom of tank PMax SGp x b SGf x d URV when HP at bottom of tank LRV when LP at bottom of tank Where minimum level at 4mA maximum level at 20 mA a distance between bottom tap and minimum level b distance between bottom tap and maximum level d distance between taps SGf Specific Gravity of capillary fill fluid See Page 11 for values SGp Specific Gravity of process fluid Contact STC Phoenix concerning applications for model STR13D ST 3000 Smart Pressure Tra
12. li 100mA T4 Ta 50 to 93 C 4 20 mA DE Ci 4 2nF T5 Ta 50 to 85 C INMETRO Li T6 Ta 50 to 70 C CERTUSP Pi 1 2W Brazil Ui 30V Intrinsically Safe Zone 0 1 y li 100mA T4 Ta 50 to 93 C 4 20 mA HART Ci 4 2nF T5 Ta 50 to 63 C BR Ex ia IIC Li T6 Ta 50 to 48 C Enclosure IP 66 67 P 1 2W j Ui 24V Fieldbus ds E T3 Ta 50 to 93 C Not FISCO en T4 Ta 50 to 40 C Pi 1 2W Li 0 except Li 150UH when Option ME Analog Meter is selected This certificate defines the certifications covered for the ST 3000 Pressure Transmitter family of products including the SMV 3000 Smart Multivariable Transmitter It represents the compilation of the five certificates Honeywell currently has covering the certification of these products into marine applications For ST 3000 Smart Pressure Transmitter and SMV 3000 Smart Multivarible Transmitter American Bureau of Shipping ABS 2009 Steel Vessel Rules 1 1 4 3 7 4 6 2 5 15 4 8 3 13 ST 3000 Pressure 4 13 5 4 8 4 27 5 1 4 9 7 13 Certificate number 04 HS417416 PDA Transmitter Marine Certificate Bureau Veritas BV Product Code 389 1H Certificate number 12660 B0 BV MT Option Det Norske Veritas DNV Location Classes Temperature D Humidity B Vibration A EMC B Enclosure C For salt spray exposure enclosure of 316 SST or 2 part epoxy protection with 316 SST bolts to be applied Certificate number A 11476 Ko
13. 3 000 psi from 26 C to 40 C Use of graphite o rings de rates transmitter to 3 625 psi Use of adaptor with graphite o rings de rates transmitter to 3 000 psi Consult factory for MAWP of ST 3000 transmitters with CSA approval ST 3000 Smart Pressure Transmitter Performance Under Rated Conditions Model STR12D 4 to 0 400 inH O Parameter Description Upper Range Limit inH O 400 39 2 F 4 C is standard reference temperature for inH20 range mbar 1 000 Minimum Span inH O 4 Note Recommended minimum span in square root mode is 20 inH O 50 mbar mbar 10 Turndown Ratio 100 to 1 Zero Elevation and Suppression No limit except minimum span within 100 URL Accuracy Reference Includes combined effects of linearity hysteresis and repeatability e Accuracy includes residual error after averaging successive readings For FOUNDATION Fieldbus use Digital Mode specifications For HART use Analog Mode specifications In Analog Mode 0 20 of calibrated span or upper range value URV whichever is greater terminal based For URV below reference point 50 inH2O accuracy equals 90 inH O 125 mbar TER 0 10 0 10 or 0 10 0 10 in of span span inH O span mbar In Digital Mode 0 175 of calibrated span or upper range value URV whichever is greater terminal based For URV below reference point 50 inH20 accuracy
14. Female 316 SS Certified Conduit Adapter ATEX CSA amp IECEx 1 2 NPT Male to 3 4 NPT Female 316 SS Certified Conduit Adapter ATEX CSA amp IECEx M20 Male to 1 2 NPT Female 316 SS Certified Conduit Adaptor ATEX CSA amp IECEx 1 2 NPT Zinc plated Certified Conduit Plug ATEX CSA amp IECEx Marine Approved Angle Mounting Bracket 304 SS Flat Mounting Bracket Carbon Steel Services Certificates Marine Type Approvals Options E 09 Users Manual Paper Copy Standard HC H6 or FF ships accordingly Clean Transmitter for Oxygen or Chlorine Service with Certificate 50039190 Over Pressure Leak Test with Certificate F3392 Calibration Test Report and Certificate of Conformance F3399 Certificate of Conformance F3391 Certificate of Origin F0195 SIL Certificate SIL 2 3 FC33337 NACE Certificate Process Wetted amp Non Process Wetted FC33339 NACE Certificate for all welded meter bodies only F0198 NACE Certificate Process Wetted only FC33338 Material Traceability Certification per EN 10204 3 1 FC33341 Supplied as 316 SS or as Grade CF8M the castina eauivalent of 316 SS ST 3000 Smart Pressure Transmitter 34 ST 16U 32 Issue 50 Page 9 of 12 TABLE lll OPTIONS continued Warranty Options Additional Warranty 1 year Additional Warranty 2 years Additional Warranty 3 years 39 STR14A STR12D amp 13D Em 14G amp 17G Additional Warranty 4 years us Approv
15. LCIE Ell 3 G Ex nA IIC Honeywell T4 Ta 50 to 93 C T5 Ta 50 to 63 C T6 Ta 50 to 48 C Enclosure IP 66 67 Ui 30V li 100mA 4 20 mA HART Ci 4 2nF Li Pi 1 2W Fieldbus Not FNICO Flameproof Zone 1 Ui 30V li 100mA 4 20 mA DE Ci 4 2nF Li E ll 2 G Ex d IIC Pi 1 2W Intrinsically Safe Zone 0 1 Ui 30V s IG li 100mA T4 Ta 50 to 93 C Ell 1 G 4 20 mA HART Ci 4 2nF T5 Ta 50 to 63 C piss T6 Ta 50 to 48 C Non Sparking Zone 2 Pi 1 2W Ell 3 G Ex nA IIC T3 Ta 50 to 93 C T4 Ta 50 to 40 C Multiple Marking TA T 50 to 93 C a 50 to T5 Ta 50 to 85 C T6 Ta 50 to 70 C NOTE The user must determine the type of protection required for installation of the equipment The user shall then check the box Y adjacent to the type of protection used on the equipment certification nameplate Once a type of protection has been checked on the nameplate subsequently the equipment shall not be reinstalled using any of the other certification types Fieldbus T3 Ta 50 to 93 C T4 Ta 50 to 40 C Not FISCO FNICO Li O except Li 1504H when Option ME Analog Meter is selected ST 3000 Smart Pressure Transmitter 15 A Type of Protection Comm Option Field Parameters Temp Codes Flameproof Zone 1 T5 Ta 50 to 93 C BR Ex d IIC T6 Ta 50 to 78 C Enclosure IP 66 67 Ui 30V
16. Lower Two 1 2 with plastic plugs Two 1 2 with metal plugs N DO UZZC I Table Il continued below STR14A STR12D 13D 14G 17G TABLE Il SEALS continued B ECC 99 EIE EAM p 9 9 Wedge 750 psi V 5 O D Chemical Tee Diaphragm Body Selection Taylor Wedge 316L SS 316L SS Wetted Material Hastelloy C 276 316L SS d Hastelloy C 276 Hastelloy C 276 Ron Weted Materai No Selection o e lts O NoSelection o f es Mo Selection 0 Je Wo Selection No Selection _0 e Table Il continued next page 1 Standard facing 125 250 AARH RF raised face serrated surface finish Plastic Pluas are TEMPORARY ONLY to protect threads and MUST be REMOVED before installation Tantalum Bodv has Tantalum wetted parts and 316 SS non wetted parts Note Remote seal system pressure rating is body rating or seal rating whichever is less 36 ST 3000 Smart Pressure Transmitter 34 ST 16U 32 Issue 50 Page 6 of 12 TABLE II SEALS continued STR14A Threaded Process Pressure Rating Pressure Rating STR12D amp 13D 7 i 14G amp 17G Diaphragm I Connection Size 304 SS AND 2 500 1 250 2 4 3 4 NPT s a psi psi 1 NPT ee 2 500 1 250 2 9 3 4 NPT 7 a psi psi 1 NPT cents 1 500 750 3 4 NPT a psi psi 1 Mal 316L SS Carbon Steel 316L SS 316L SS Hastelloy C 276 316L SS Hastelloy C 276 Hastelloy C 276 Monel 400 Monel 400 Tantalum 316L
17. Material upper 316L SS Seal Capillary Center Seal Connection Side Seal Calibration Rings None 316L SS Hastelloy C 276 Monel 400 Table Il continued next page 1 Standard facina 125 250 AARH RF raised face serrated surface finish 2 Limited vacuum availabilitv 3 Minimum static pressure reauirement No vacuum allowed See Specifications Fiaure 15 Tantalum Upper insert has Tantalum wetted parts and 316 SS or CS non wetted parts Note Remote seal system pressure rating is body rating or seal rating whichever is less O OO ST 3000 Smart Pressure Transmitter 34 ST 16U 32 Issue 50 STR14A Page 3 of 12 STR12D amp 13D 14G amp 17G TABLE ll SEALS continued Selection Flushing Flush Flanged Connections One 1 4 with plastic plug and Plugs One 1 4 with metal plug Metal plug material Two 1 4 with plastic plugs Seal will be the same as Two 1 4 with metal plugs Cal ring material if One 1 2 with plastic plug metal plug is chosen One 1 2 with metal plug SS Plug for CS Lower Two 1 2 with plastic plugs Two 1 2 with metal plugs NO TUTZZ lt CITI Table Il continued below STR12D 8 13D STR14A 14G 8 17G TABLE ll SEALS continued Selection Diaphragm Flange Flange Pressure const aes Construction See Spec Diameter Size a uw SES DUI Figure 34 ST 03 64 ST 03 64 ANSI 300 e 2 4 ANSI 300 ANSI 150 pm EM EZ ANSI 300 1 2 ANSI 300 with Lower ANSI 300 ANSI 150
18. SS Seal with Tantalum Hastelloy C 276 Process l Conneccion upper Stainless Steel 7 Carbon Steel Flushing Connections One 1 4 with plastic plug and Plugs One 1 4 with metal plug Metal plug material Two 1 4 with plastic plugs will be the same as Two 1 4 with metal plugs Lower material if One 1 2 with plastic plug metal plug is chosen One 1 2 with metal plug SS Plug for CS Lower Two 1 2 with plastic plugs and Tantalum Clad Two 1 2 with metal plugs Klinger C 4401 non asbestos Gasket Grafoil Teflon Wetted Material Gylon 3510 Table Il continued next page Plastic Plugs are TEMPORARY ONLY to protect threads and MUST be REMOVED before installation If Table III Bolt Nut option is chosen Seal bolts will ship as same material and MAWP may change Note Remote seal system pressure rating is body rating or seal rating whichever is less ST 3000 Smart Pressure Transmitter 14G amp 17G 34 ST 16U 32 Issue 50 Page 7 of 12 STR14A TABLE Il SEALS icontmusa STR12D amp 13D je lea meme MAT Diameter E Customer clamp rating or 600 psi whichever is less 24 24 24 ates Matera D Bghagm oe RHEINE sess sss NA Telefe Non Wetted Materia No Selection its AS No Selection EUER 0 ejeje Styles Tri Clover Tri Clamp __________8_ Lo fel No Selection Size and Seal Pressure Rating Diaphragm seme mem mm Pattern pu for 3 Pipe 8 Bolt gt 4 pipe
19. for clarity TE 135 3 6 5 32 0 14 Plug Optional meters Rotational lock Rotational lock 24262 Figure 20b Approximate vertical mounting dimensions for Remote Seal Transmitter 24 Dimensions and Drawings Flush Flanged Seal 3 Class 150 3 Class 200 3 Class 600 Wetted Materials Diaphragm oS Hastelley C Hastelloy C Mane Tantalum 32 Hastelloy C Hastellcy C Mane Tantalurn 32 Hastelley C Rastelloy C Mane Tantalum 3o Rastelloy C Hastelloy C Mone Tantalum 32 Hastelloy C hastelloy C Mone Tantalum 32 Rastelloy C Hastielley C Mane Tantalum 55 Hasielloy C Hasielloy C Mane Tantalum SS Rasielloy C Hastelley C Mone Tantalum Hastellay C Manel Hastelloy E Hastellay C Monel 3 m 35 3 Hastellay C Manel c NA 55 Hastelloy C Monel Hastelloy E 32 Hastelloy C Manel mit on a Construction See figure 00000000 OA 00 0 600 0 0 00 STD I 0 06 0006 010 00 0 06 00 6 0 ST 3000 Smart Pressure Transmitter ST 3000 Smart Pressure Transmitter Flange V Instr Conn Non wetted A Diaphragm extends over the entire Raised face gasket surface wetted i lix effective Erw diaphragm diaphragm Configuration HS Figure A Instr Conn Flange s Non Wetted a S A _ effective Body diaphragm Diaphragm extends over non wetted diaphragm the entire Raised fac
20. provide one of When digitally integrated with Honeywell s Process Knowledge System EXPERION PKS ST 3000 instruments provide a more accurate process variable as well the following output communication options Communications options as advanced diagnostics Honeywell s high performance ST 4 20 mA 3000 S100 transmitters lead the industry in H Il Digitally Enh d DE oneywell Digitally Enhanced DE Accuracy HART versions 5 x or 6 x e Stability FOUNDATION Fieldbus e Reliability e Rangeability e Warranty ST3000 S900 Transmitter Benefits Total Accuracy 0 0375 Stability 0 01 per year Reliability 470 years MTBF Rangeability 400 to 1 Lifetime Warranty 15 years The devices provide comprehensive self diagnostics to help users maintain high uptime meet regulatory requirements and attain high quality standards S100 transmitters are ideal for critical applications such as custody transfer of natural gas and energy and material balances where accuracy and stability are of the utmost importance Description of Diaphragm Seals Diaphragm seals are traditionally used when a standard pressure transmitter should not be exposed to the process pressure directly Diaphragm seals typically protect the pressure transmitter from one or more damaging aspects of the process media Consideration for using a diaphragm seal should be made in the following circumstances e High Process Temperature e Pro
21. 00 and Experion are registered trademarks of Honeywell International Inc ST 3000 Smart Pressure Transmitter 200 185 180 160 Maximum Ambient 140 Temperature F 120 100 80 60 40 20 165 140 200 350 450 Process Interface Temperature F Figure 13 Ambient temperature and process interface chart Transmitter Minimum Span and Maximum Capillary Length Minimum recommended span for STR12D and STR13D DP Transmitter with two Remote Seals Sze 5 1 i 2 30 38 maximum SY SE RICO EEC TE ET 29 7100 m0 12500 15000 175wo a5 i6iwe 20wc 24iwc 28 iwc 360 awe 3 aa t2iwc we 18iwc 2tiwe 27 we 30me 38 Minimum recommended span for STR12D and STR13D DP Transmitter with one Remote Seal Size Mou 10 1 20 30 35 maximum 24 209 9 2 2 2 5 7 29 10psg tS psig 20psg 25psig 30psg 277 35 50iwe 80iwe 100iwc 120wc 140wc 18060 200we 3 AA 40iwc 60iwc 80iwc 7100 iwc 7120 iwc 160 iwc 7180 iwc 35 Minimum recommended span for STR14G STR14A STR17G Transmitter with Remote Seal Diaphragm Direct Capillary Capillary Length Size Mount 5 10 15 20 30 35 o maximum 20 25psi 30psi 40psi 50psi 1 Minimum span is the higher of the value from the table above or the value defined under Performance Co
22. 0a and 20b Seal See Figures 21 through 31 Net Weight Transmitter 15 4 pounds 7 Kg Total weight is dependent on seal type and capillary length NOTE Pressure transmitters that are part of safety equipment for the protection of piping systems or vessel s from exceeding allowable pressure limits equipment with safety functions in accordance with Pressure Equipment Directive 97 23 EC article 1 2 1 3 require separate examination Vent Drains are sealed with Teflon or PTFE Hastelloy C 276 or UNS N10276 Monel 400 or UNS N04400 Supplied as 316 SS or as Grade CF8M the casting equivalent of 316 SS Carbon Steel heads are zinc plated and not recommended for water service due to hydrogen migration For that service use 316 stainless steel wetted Process Heads Hastelloy C 276 or UNS N10276 Supplied as indicated or as Grade CW12MW the casting equivalent of Hastelloy C 276 Monel 400 or UNS N04400 Supplied as indicated or as Grade M30C the casting equivalent of Monel 400 ST 3000 Smart Pressure Transmitter 11 Certifications Ss Type of Protection Comm Option Field Parameters perborate Explosionproof Class Division 1 Groups A B C D locations Dust Ignition Proof m de Tonic Class Il Ill Division 1 Groups E F G locations Enclosure Type 4X Vmax 42 4V Imax 225mA REN 4 20 mA DE Ci 4 2nF T4 Ta 93 C Intrinsically Safe Liz Pi 21 2W Vmax 30V Class
23. 1 500 psi 1 500 psi Design ud for 3 Pipe 6 Bolt gt 4 pipe 1 250 psi 1 250 psi Design 316L SS Carbon Steel 316L SS 316L SS Wetted Material Hastelloy C 276 316L SS Sanitary Seal Saddle Seal Non Wetted Material Carbon Steel Carbon Steel 316L SS e SS AA aes es NeSeedio 0 iie Klinger C 4401 non asbestos Gasket Grafoil Teflon Gylon 3510 8 If Table III Bolt Nut option is chosen Seal bolts will shin as same material and MAWP mav chanae All sanitarv seals have dairv arade 3A approval 10 Bolts are not included with bodv onlv selection Note Remote seal system pressure rating is body rating or seal rating whichever is less 38 ST 3000 Smart Pressure Transmitter 34 ST 16U 32 Issue 50 Page 8 of 12 i 4A o 351 eee TABLE lll OPTIONS Selection Analog only can be configured using appropriate Honeywell DE tool AN DE Protocol communications DE HART 5 x Protocol compatible electronics HC da X MEE AUS ues electronics de Indicating Meter Options LS eL aa necat Meter 0 100 Even 0 10 Square Root gi Smart Meter Custom Configuration of Smart Meter Local Zero amp Span Local Zero No ERE 1 2 NPT 316 SS Certified Conduit Plug ATEX CSA amp IECEx M20 316 SS Certified Conduit Plug ATEX CSA amp IECEx 1 2 NPT Non certified Conduit plug Zinc plated carbon steel general use NAMUR Failsafe Software SIL 2 TUV Certified transmitter requires HC or H6 an
24. 374 With Analog meter 53 1 65 1 Removal od ri 2 09 T7 2 8C Clearance yas Gin ut meter meter for All Caps 1 8 O ptional external ground Seal omitted for clarity STR14G STR17G LGP I Remote Seal with Horizontal Pipe Mount ST 3000 Smart Pressure Transmitter 3 6 0 14 Plug Optional meters Rotational lock LP seal omitted for clarity Rotational lock 24261 Figure 20a Approximate horizontal mounting dimensions for Remote Seal Transmitter ST 3000 Smart Pressure Transmitter Reference Dimensions With Smart mete ES 352 With Analog meter 23 1 65 1 Removal 45 7 2 09 2 56 Clearance 1 8 Without Without for All Caps meter meter Reference Dimensions millimeters Inches STR12D STR13D DP I Remote Seal with Vertical pp Pipe Mount LJ l With Smart mete 828 29 a 53 1 65 1 Removal 45 7 orn Clearance 1 8 UE Ao STR 14A APII Remote Seal with Vertical for All Caps h Pipe Mount Optional external ground Seal omitted for clarity Removal 45 7 Clearance 1 8 for All Caps With Smart mete 82 9 94 9 With Analog 3 26 3 74 meter 93 1 56 i STR14G STR17G LGP I Remote Seal 23 5 with Vertical Pipe 925 Optional external ground Seal r1 omitted for clarity i i 7 LI Mount Flat Bracket 23 135 3 6 5 32 0 14 Plug IE 2 i Ni Optional q E meters LP seal Rotational omitted lock
25. 77 F zero 0 static pressure 10 to 55 RH and 316L Stainless Steel barrier diaphragm Transmitter URL limit or maximum seal pressure rating whichever is lower Specification applies to transmitters with 2 seals only Apply 1 5 times factor to temperature effect for capillary lengths greater than 10 feet ST 3000 Smart Pressure Transmitter Performance Under Rated Conditions Model STR14G 0 5 to 0 500 psig Parameter Description Upper Range Limit psig 500 bar 35 Minimum Span psig 5 bar 0 35 Turndown Ratio 100 to 1 Zero Elevation and Suppression No limit except minimum span from absolute zero to 100 of URL Specifications valid over this range Accuracy Reference Includes combined effects of linearity hysteresis and repeatability e Accuracy includes residual error after averaging successive readings For FOUNDATION Fieldbus use Digital Mode specifications For HART use Analog Mode specifications In Analog Mode 0 1 of calibrated span or upper range value URV whichever is greater For URV below reference point 20 psi accuracy equals O psi 1 4 bar 2 0 05 0 05 or 0 05 0 05 in of span span psi span bar In Digital Mode 0 075 of calibrated span or upper range value URV whichever is greater For URV below reference point 20 psi accuracy equals O psi 1 4 bar 2 0 025 0 05
26. T 3000 Smart Pressure Transmitter umm Type of Protection Comm Option Field Parameters Temp Codes Explosion Proof Class l Division Canadian Standards Association CSA 1 Groups B C D locations Dust Ignition Proof Class Il Ill Division 1 Groups E F G locations Enclosure Type 4X Intrinsically Safe Class Il IIl Division 1 Groups A B C D E F G locations Enclosure Type 4X Nonincendive Class Division 2 Groups A B C D locations Enclosure Type 4X Canadian Registration Number CRN 4 20 mA DE 4 20 mA HART Fieldbus Entity Not FISCO 4 20 mA DE 4 20 mA HART Fieldbus Entity Not FNICO Vmax 42V Imax 225mA Ci 4 2nF Li Pi 1 2W Vmax 42V Imax 225mA Ci 4 2nF Li Pi 1 2W Vmax 24V Imax 250mA Ci 4 2nF Li 0 Pi 1 2W Vmax 42 4V Imax 225mA Ci 4 2nF Li Pi 1 2W Vmax 30V Imax 225mA Ci 4 2nF Li Pi 1 2W Vmax 24V Imax 250mA Ci 4 2nF Li 0 Pi 1 2W T4 Ta 93 C T4 Ta 93 C T4 Ta 93 C T4 Ta 40 C T3 Ta 93 C T4 Ta 93 C T4 Ta 93 C T4 Ta 40 C T3 Ta 93 C All ST 3000 models except STG19L STG99L STG170 and STG180 have been registered in all provinces and territories in Canada and are marked CRN 0F8914 5C ST 3000 Smart Pressure Transmitter 13 nmm Type of Protection Comm Option Field Parameters Temp Codes Flameproof Zone 1 T5 Ta
27. Technical Information Honeywell ST 3000 Smart Transmitter Series 100 Remote Diaphragm Seals Models gt Specifications gt 34 ST 03 64 March 2013 a Introduction In 1983 Honeywell introduced the first Smart Pressure Transmitter the ST 3000 In 1989 Honeywell launched the first all digital bi directional protocol for smart field devices Today its ST 3000 Series 100 Remote Seal Transmitters continue to bring proven smart technology to a wide spectrum of measurement applications Typical applications include high accuracy level measurement in pressurized vessels in the chemical and hydrocarbon processing industries A second application consists of accurate flow measurement for slurries and high viscosity fluids in the chemical industry Honeywell remote seal transmitters demonstrate proven reliability in hundreds on installations in a wide variety of industries and applications with a wide variety of secondary fill fluids for corrosive or high temperature process fluids STR12D 0 4 to 0 400 inH20 0 10 to 0 1 000 mbar STR13D 0 1 to 0 100 psid 0 0 07 to 0 7 bar STR14G 0 5 to 0 500 psig 0 0 35 to 0 35 bar Figure 1 Series 100 Remote Seal Pressure Transmitters STR17G 0 30 to 0 3 000 psig 2 1 to 0 210 bar feature proven piezoresistive sensors and advanced seal l technology with standard weld connections STR14A 0 5 to 0 500 psia 0 0 35 to 0 35 bara All ST 3000 transmitters can be ordered to
28. al Type Location or Classification Selection No hazardous location approvals Intrinsically Safe Zone 0 Intrinsically Safe Zone 1 Dust tight Enclosure Zone 0 Flameproof and Dust tight Enclosure Zone 1 Non Sparking 40 Zone 2 ATEX LCIE Multiple Marking Int Safe Zone 0 1 and Dust tight Enclosure or Flameproof Zone 1 and and Dust tight Enclosure or Non Sparking Zone 2 RA ee s Ex ia IIC T4 Ta 50 C to 93 C T5 Ta 50 C to 85 C T6 Ta 50 C to 70 C Enclosure IP 66 67 3S Ex ia IIC T4 Ta 50 C to 93 C T5 Ta 50 C to 85 C T6 Ta 50 C to 70 C Enclosure IP 66 67 Ex tD A20 IP6X T95 C at Ta 93 C or T80 C at Ta 78 C Enclosure IP 66 67 Exd IIC T5 Ta 40 C to 93 C 33 T6 Ta 40 C to 78 C Supply 11 42Vdc Ex tD A21 IP6X T95 C at Ta 93 C or T80 C at Ta 78 C Enclosure IP 66 67 b Ex nA IIC T5 Ta 40 C to 93 C T6 Ta 40 C to 78 C Zone 2 Supply lt 42Vdc 23mA Ex tD A22 IP6X 3N T95 C at Ta 93 C or T80 C at Ta 78 C Honeywell Enclosure IP 66 67 Ex ia IIC T4 Ta 50 C to 93 C T5 Ta 50 C to 85 C T6 Ta 50 C to 70 C Ui 30V li 100mA Ex tD A20 IP6X T95 C at Ta 93 C or T80 C at Ta 78 C Ex d IIC T5 Ta 40 C to 93 C T6 Ta 40 C to 78 C Supply 11 42Vdc Ex tD A21 IP6X 3C T95 C at Ta 93 C
29. ccuracy Reference Includes In Analog Mode 0 1 of calibrated span or upper range value URV whichever combined effects of linearity is greater terminal based hysteresis and repeatability For URV below reference point 30 psi accuracy equals e Stated accuracy does not apply for 30 psi 2b l models with 2 9 inch diameter 0 05 0 05 or 0 05 0 05 in of span remote seal diaphragms erage Spanner In Digital Mode 0 075 of calibrated span or upper range value URV whichever is greater terminal based For URV below reference point 30 psi accuracy equals e Accuracy includes residual error after averaging successive readings e TM f 30 psi 2b OF FOUNDATION FIRIODUS USG 0 025 0 05 gt or 0 025 0 05 2 in of span Digital Mode specifications span psi span bar For HART use Analog Mode specifications Combined Zero and Span In Analog Mode 0 33 of span Temperature Effect per 28 C For URV below reference point 60 psi effect equals 50 F 60 psi 4 bar bc 0 05 0 20 or 0 05 0 28 in of span span psi span bar In Digital Mode 0 305 of span For URV below reference point 60 psi effect equals 60 psi 4 bar 0 025 40 25 or 0 025 0 28 in of span span psi span bar Performance specifications are based on reference conditions of 25 C
30. cess Media is Viscous or Contains Suspended Solids e Process Media is Subject to Solidifying e Process Media is Corrosive e Process Application Requires Sanitary Connections e Process Application Subjects the Measuring Instrument to Hydrogen Permeation e Tank Level Applications with Maintenance Intensive Wet Legs e Tank Application with Density or Interface Measurements e Measuring Instrument Requires Remote Mounting The following diaphragm seals are standard from Honeywell please call your local salesperson if you do not see the product you need for your application ST 3000 Smart Pressure Transmitter Figure 2 Flush Flange Seals Flush Flange Seals can be used with differential gauge and absolute pressure transmitters and are available with 3 ANSI Class 150 ANSI Class 300 and DIN DN80 PN40 process connections Flush flange seals can also be provided with Lowers Lowers are essentially calibration rings which allow flushing connections if needed Figure 3 Flange Seal with Extended Diaphragm Flange Seal with Extended Diaphragm can be used with differential gauge and absolute pressure transmitters and are available with 3 and 4 ANSI Class 150 ANSI Class 300 DIN DN80 PN40 and DIN DN100 PN40 process connections 2 4 and 6 extension lengths are available Figure 4 Pancake Seals Pancake Seals can be used with differential gauge and absolute pressure transmitters and are availab
31. d WP options Lightning Protection Custom Calibration and D in Memory Transmitter Configuration non Fieldbus Transmitter Configuration Fieldbus Write Protection Delivered in the enabled position Write Protection Delivered in the disabled position Stainless Steel Customer Wired On Tag 4 lines 26 characters per line customer supplied information Stainless Steel Customer Wired On Tag Meter Body Options Seal bolt material depends on Transmitter bolt material A286 SS NACE Bolts and 304 SS NACE Nuts for Heads CR 316 SS Bolts and 316 SS Nuts for Process Heads SS HH b B7M Bolts and Nuts for Process Heads Remote Seal Options Gold Plated Seal Diaphragm 1 Seal G1 Gold Plated Seal Diaphragm 2 Seals G2 Teflon Coated Seal Diaphragm 1 Seal only for anti sticking N1 b Teflon Coated Seal Diaphragms 2 Seals only for anti sticking N2 Transmitter Mounting Bracket Options EE EE ERR E A Angle Mounting Bracket Carbon Steel MB Marine Approved Angle Mounting Bracket Carbon Steel MX Angle Mounting Bracket 304 SS SB b No housing conduit plugs or adaptors come standard with the ST 3000 For certain approval codes you must select a certified conduit plug from below and it will come packaged in the box with your transmitter 316 SS Electronics Housina with M20 Conduit Connections 316 SS Electronics Housina with M20 to 1 2 NPT 316 SS Conduit Adapter for use with FM and CSA Approval codes 1 2 NPT Male to M20
32. de HART Protocol compatibility in either HART 5 x or 6 x formats Transmitters with a HART Option are compatible with any HART enabled system that provides 5 x or 6 x format support e Foundation Fieldbus Output Option FF Equips transmitter with FF protocol for use in 31 25 kbit s FF networks See document 34 ST 03 72 for additional information on ST 3000 Fieldbus transmitters e SIL2 SIL3 Certification Option SL This ST 3000 product is available for use with safety systems With the SL option we are fully certified to SIL 2 capability for single transmitters and SIL 3 capability for multiple transmitter use through T V Nord Sys Tec GmbH amp Co KG We are in compliance with the following SIL standards IEC 61508 1 1998 IEC 61508 2 2000 IEC 61508 3 1998 ST 3000 Smart Pressure Transmitter Lightning Protection Option LP A terminal block is available with circuitry that protects the transmitter from transient surges induced by nearby lightning strikes NAMUR NE43 Compliance Option NE This option provides software the meets the NAMUR NE43 requirements for failsafe software Transmitter failure information is generated when the measuring information is no longer valid Transmitter failure values are x 3 6 mA and 2 21 0 mA The normal ST 3000 ranges are lt 3 8 mA and 2 20 8 mA Write Protection Options WP and WX Provides the capability to hardwire write protect installed transmitter conf
33. e gasket surface wetted Configuration IS Figure C 25 Flange t Instr Conn Wetted W 5 B me TE ZE p i wo d L od A diaphragm Raised face Gasket surface wetted Configuration HT Figure B Instr Conn Flange Ring Non Wetted A y Bp AN diaphragm he Body __ Raised face Gasket wetted surface wetted Configuration IT Figure D 26 Dimensions and Drawings Flush Flanged Seal with Lower ANSUDIN 247 Diaph a BO a 12 Class 150 Flush Flanged Seal with Lower Class 3002 Class 400 BD iihout Flush B1 B Dimension with 14 NPT Flushing Connection B2 B dimension wih 1 2 NPT Flushing Connection 2 9 Diaph Dia in ST 3000 Smart Pressure Transmitter 4 1 Diaph Dia in ST 3000 Smart Pressure Transmitter Dimensions and Drawings A Instr Conti UpperAmsert Mon Ifletted Flange Rind Mon llletted Fill Screw 0 62 Diaphregm Wetted Flush Flanged Seal with Lower Flanged Seal with Extended Diaphragm 3 Class 150 3 Class OOF ed CIN ONSC Flang Seal with Suse 4 Class Diaphragm 1508 4 Class i 300 DIN DNBD PNA4O E Designed ta meet with schedule 40 pipe I C ul qr Pancake Seal Class 150 200 600 ONEaD PH40 10 00 1 25 3 70 8 25 0 24 3 70 2 e c Fill _ Instr Screw Conn Unp
34. egistered trademark of DuPont Neobee is a registered trademark of Stepan Company Syltherm 800 is a Trademark of Dow Corning Corporation Klinger C 4401 is a registered trademark of THERMOSEAL INC GRAFOIL is a registered trademarks of GrafTech International Holdings Inc Gylon 3510 is registered trademark of Garlock Sealing Technologies Tri Clover Tri Clamp is a registered trademark of Alfa Laval DC 200 and DC 704 are registered trademarks of Dow Corning ST 3000 Smart Pressure Transmitter 43 Sales and Service For application assistance current specifications pricing or name of the nearest Authorized Distributor contact one of the offices below ASIA PACIFIC EMEA NORTH AMERICA SOUTH AMERICA TAC hfs tac Phone 80012026455 or 44 Honeywell Process Solutions Honeywell do Brazil amp Cia support honeywell com 0 1202645583 Phone 1 800 423 9883 Phone 55 11 7266 1900 l FAX 44 0 1344 655554 Or 1 800 343 0228 FAX 55 11 7266 1905 idle Tw Email Sales sc cp apps Email Sales ask Email Sales ask ulis Abk diis salespa62 honeywell com ssc honeywell com or ssc honeywell com or Phone 61 7 3846 1255 or FAX 61 7 3840 6481 TAC hfs TAC BENE AO AB well NTS tac support honeywell com support honeywell com Toll Free 1300 36 39 36 support honeywell com Toll Free Fax 1300 36 04 70 China PRC Shanghai Honeywell China Inc Phone 86 21 5257 4568 Fax 86 21 6237 2826 Singapore Honey
35. equals 50 inH O 125 mbar T 0 075 0 10 or 0 075 0 10 in of span span inH O span mbar Combined Zero and Span Temperature Effect per 28 C 50 F In Analog Mode 1 2 of span For URV below reference point 200 inH O effect equals 200 inH O 900 mbar _ 0 20 0 10 Or 0 20 0 10 _ in of span span inH O span mbar In Digital Mode 1 175 of span For URV below reference point 200 inH O effect equals 200 inH O 500 mbar np 0 175 1 0 _ Or 0 175 1 0 in of span span inH O span mbar Performance specifications are based on reference conditions of 25 C 77 F zero 0 static pressure 10 to 55 RH and 316L Stainless Steel barrier diaphragm Transmitter URL limit or maximum seal pressure rating whichever is lower Specification applies to transmitters with 2 seals only Apply 1 5 times factor to temperature effect for capillary lengths greater than 10 feet ST 3000 Smart Pressure Transmitter 7 Performance Under Rated Conditions Model STR13D 1 to 0 100 psid Parameter Description Upper Range Limit psid 100 bar 7 Minimum Span psid 1 bar 0 07 Turndown Ratio 100 to 1 Zero Elevation and Suppression No limit except minimum span within 18 and 100 of URL Specifications valid from 5 to 100 of URL A
36. er Mon WMetted 0 90 Flush Conn 4 Optional V Gasket Lower Wetted Wetted Diaphregm Iletted Stud Flush Flanged Seal with Lower Note 0 90 dimension is 0 70 for 4 1 Dia Diaphragm A J Instr Conn Fill Screw Flange Non Wetted Raised Face Wetted Extension Length Extension Wetted A C Nu NEN Wetted Flange Extended Seal Instrument Connection A A Pancake Flange Seal 28 Dimensions and Drawings Chemical Tee Taylor Wedge Seal 3 5 Diaph Chemical Seal with Threaded Process Connection Conn Seal Without Flush B1 B Dimension with 1 4 NPT Flushing Connection B2 B dimension with 1 2 NPT Flushing Connection Sanitary Seal B B EE ST 3000 Smart Pressure Transmitter Diaphragm Wetted Wetted Chemical Tee Taylor Wedge Seal Upper Body Non Wetted Instr Conn Bolt Fill Screw f 3 1 Flush Conn Optional Gasket Wetted Lower Diaphragm Wetted Wetted A Threaded Process Connection Seal Instr Conn Body Wetted Fill Screw A ama Diaphragm A m Wetted Sanitary Seal ST 3000 Smart Pressure Transmitter 29 Saddle Seal instr Conn in A Bolt 3 5n Upper Body Non wetted Saddle B 2 80 5eal 3 50 Fill 3 04 Weld Mote Specify 6 or 3 bolt pattern el as
37. igurations Stainless Steel Tagging Option TG Up to 30 characters can be added on the stainless steel nameplate mounted on the transmitter s electronics housing at no extra cost A stainless steel wired on tag with additional data of up to 4 lines of 28 characters is also available The number of characters for tagging includes spaces Transmitter Configuration Options TC and FC With Option TC the factory can configure the analog DE or HART transmitter s linear square root extraction damping time LRV URV and mode analog digital and enter an ID tag of up to eight characters and scratchpad information as specified With Option FC the Device ID Transmitter Tag Unit Level Node Address Output Mode and Damping Time Constants can be specified Custom Calibration and ID in Memory Option CC The factory can calibrate any range within the scope of the transmitter s range and enter an ID tag of up to eight characters in the transmitter s memory Indicator Configuration Option CI Provides custom configuration of Smart Meters Lifetime Warranty Option WL Extends limited 1 year warranty policy to 15 years for ST 3000 S100 pressure transmitters See Honeywell Terms and Conditions ST 3000 Smart Pressure Transmitter 31 Model Selection Guides are subject to change and are inserted into the specifications as guidance only Prior to specifying or ordering a model check for the latest revision Model Selection Guides which a
38. inless Steel Nipples are available for Close Coupled remote seal solutions Figure 12 Welded Meter Body for All Welded Remote Seal Solution Welded Meter Body for All Welded Remote Seal Solution The welded ST 3000 meter body is an important part of an All Welded Remote Seal Solution which is commonly used in Vacuum applications Description The ST 3000 transmitter can replace any 4 to 20 mA output transmitter in use today and operates over a standard two wire system The measuring means is a piezoresistive sensor which actually contains three sensors in one It uses a differential pressure sensor a temperature sensor and a static pressure sensor in delivering the most comprehensive compensated output signal available today Microprocessor based electronics provide higher span turndown ratio improved temperature and pressure compensation and improved accuracy The transmitter s meter body and electronics housing resist shock vibration corrosion and moisture The electronics housing contains a compartment for the single board electronics which is isolated from an integral junction box The single board electronics is replaceable and ST 3000 Smart Pressure Transmitter interchangeable with any other ST 3000 Series 100 or Series 900 model transmitters Features e Choice of linear or square root output conformity is a simple configuration selection e Direct digital integration with Experion PKS and other
39. le with 3 ANSI Class 150 300 and 600 process connections ST 3000 Smart Pressure Transmitter e Figure 5 Chemical Tee Taylor Wedge Figure 8 Saddle Seals Chemical Tee Taylor Wedge can be used with differential pressure transmitters and are available with Taylor Wedge Saddle Seals can be used with differential gauge and 5 O D process connection absolute pressure transmitters and are available with 3 and 4 6 bolt or 8 bolt designs process connections Figure 6 Seals with Threaded Process Connections Figure 9 Calibration Rings Seals with Threaded Process Connections can be used with differential gauge and absolute pressure transmitters and are available with 72 and 1 NPT Female process Calibration Rings are available with Flush Flange Seals and connections Pancake Seals Flushing ports 1 4 or 72 are available with calibration rings gt is n gt Sc 1 7 SO rO LU f Figure 10 Stainless Steel Armor and PVC Coated Stainless Figure 7 Sanitary Seals Steel Armor Capillaries Sanitary Seals can be used with differential gauge and Stainless Steel Armor and PVC Coated Stainless Steel absolute pressure transmitters and are available with 3 and Armor Capillaries are available with Honeywell Remote 4 Tri Clover Tri Clamp process connections Seal Solutions Figure 11 2 Stainless Steel Nipples 2 Sta
40. n accordance with Pressure Equipment Directive 97 23 EC article 1 2 1 3 require separate examination A formal statement from T V Industry Service Group of TUV America Inc a division of T V S ddeutschland a Notified Body regarding the Pressure Equipment Directive can be found at www honeywell com A hard copy may be obtained by contacting a Honeywell representative Electro Magnetic Compatibility EMC 2004 108 EC CE Mark All Models EN 50081 1 1992 EN 50082 2 1995 EN 61326 1 1997 A1 A2 and A3 Industrial Locations Dual Seal Dual Seal Certification based on ANSI NFPA 70 202 and ANSI ISA 12 27 01 requirements without the use of Certification additional seal protection elements Recommended Frequency of Honeywell recommends verifying the calibration of these devices once every four years Calibration Approved Honeywell Process Solutions York PA USA Manufacturing Honeywell Tianjin Limited Tianjin P R China Locations Honeywell Automation India Ltd Pune 411013 India Foundation Fieldbus is a trademark of the Fieldbus Foundation Viton is a registered trademark of DuPont HART is a registered trademark of HART Communications Foundation Teflon is a registered trademark of DuPont Hastelloy C 276 is a registered trademark of Haynes International DC 200 is a registered trademark of Dow Corning Monel 400 is a registered trademark of Special Metals Corporation FM Approvals is a service mark of FM Global ST 30
41. nditions for the range transmitter Figure 14 Typical Maximum capillary length and diaphragm size chart Maximum Pressure Limited T nge Rang of Seal 10000 183 000 psig e Oi 5000 30 to 640 4000 000 cone DC200 e 40 to 400 F 1000 AN tAm i i a oo _CTFE E WS e P 5to 300 F S Bis SS Wx f rse SP E c Special High Vacuum S NN Construction required NNUS 7 Consult STC l Pressure immbHas d a 4 35 50 175 350 450 psig 100 100 200 300 400 a Process Temperature Degrees F Maximum Pressure Limited 1 lange Rating Seal 10000 193 3000 psg Spon Syttherm 800 _ 40 to 600 F i E psig sea gU oO 7 Fl NE 1000 Eo A 5 ooo Ui a os qq dd d ELI gt aie IEEE SA pa E ET od ee 4d le dq 100 o LL a a el Co I E OE oo ee Ls A SSS ee 5 E go E p kak v e dd Pp e a ST 3000 Smart Pressure Transmitter A 1 1 e EE MA A 1f Pressure O PI l 1 a 175 390 450 550 pag 100 Co 500 600 40 5 25 30 1 200 00 Process Temperature Degrees F are Figure 15 ST 3000 Remote Seals operable limits for pressure vs temperature ST 3000 Smart Pressure Transmitter Loop S00 650 Mote Aminimum of 250 ohms of loop resistance is necessary to support communications Loop resistance equals barrier resistance plus wire resistance 450 250 I E i l I i E I E I I
42. nsmitter Max Level Min Level Transmitter above datum Max Level Min Level Transmitter at datum Max Level Min Level Transmitter below datum 24253 Figure 18 Closed tank liquid level measurement distance ST 3000 Smart Pressure Transmitter Density or Interface Calculate the minimum and maximum pressure differentials to be measured Figure 18 Pmin SGmin SGf x d minimum density 4mA output Pmax SGmax SGf x d maximum density 20mA output Where d distance between the taps SGmax maximum Specific Gravity SGmin minimum Specific Gravity SGf Specific Gravity of capillary fill fluid See Page 11 for values Contact STC Phoenix concerning applications for model STR13D 21 Figure 19 Density direct acting transmitter configuration 22 Reference Dimensions s2 T 94 9 With Smart meter gt With Analog FAL 14 meter Lo 4 Removal 45 7 Clearance 1 8 Wied With out for All Caps E meter 5 925 With Smart mete 3 74 65 1 Removal 45 7 2 56 Clearance 1 Without for All Caps meter C2 CO a E Oxo END 2 Ol go Optional external ground Seal omitted for clarity Reference Dimensions millimeters inches STR12D STR13D DP I Remote Seal with Horizontal Pipe Mount With Analog meter STR14A AP I Remote Seal with Horizontal Pipe Mount 82 9 94 9 With Smart mete 3 06 6 7
43. or T80 C at Ta 78 C Ex nA IIC T5 Ta 40 C to 93 C T6 Ta 40 C to 78 C Zone 2 Supply lt 42Vdc 23mA Ex tD A22 IP6X T95 C at Ta 93 C or T80 C at Ta 78 C Honeywell Enclosure IP 66 67 Table Ill Approvals continued next page 40 ST 3000 Smart Pressure Transmitter 34 ST 16U 32 Issue 50 Page 10 of 12 STR14A STR12D amp 13D 14G amp 17G TABLE Ill Approvals Options continued y Ve Jee NR Mal sm 1C Approvals A B C D E F G Canadian Standards A a fefefe Association asi mars Class I Il Ill Div 1 Groups CSA 7 A B C D E F G Flameproof oe is n T5 Ta 40 to 93 C T6 Ta 40 to 78 C IECEx ee CA Intrinsically Safe Ex ia IIC T3 Li T9 T6 See Spec for detailed temperature codes by Zone 0 1 aoe Communications option Intrinsically Safe Ex iallC T4 T5 T6 b Zone 0 1 pa Flameproof cas 1 ExdlIC T5 T6 Enclosure IP 66 67 5 e Africa Multiple Marking Int Safe Zone 0 1 or Ex ia IIC T4 T5 T6 ZA Flameproof Zone 1 Ex d IIC T5 T6 Enclosure IP 66 67 cERTUSP Elameproof Zone 1 BR Ex d IIC T5 T6 6D je e e INMETRO Intrinsically Safe BR Ex ia NC T4 T5 T6 Brazil Zone 0 1 See CERTUSP certificate for detailed temperature 6S codes by Communications option 10 See ATEX installation requirements in the ST 3000 User s Manual 1 The user must determine the type of protection required for installation of the equipment The user shall
44. or FF TABLE METER BODY Selection 1 Remote Seal High Side 2 Remote Seals 1 Remote Seal Low Side Value Added Model VAM unit Fill Fluid DC 200 Silicone EL M Meter bod Non Wetted Adan s Head Material Number of Seals EIN OLIT 316 SS Bonnet for Close Couple D 316 SS bolt on bid __A 316 SS with Almeida meter body C Single Head 316 SS Adapter Head _A SS T ST 3000 Smart Pressure Transmitter 34 ST 16U 32 Issue 50 Page 2 of 12 TABLE ll SEALS Format for Seal Selection Specify 12 characters 14A Common Required Seal 12D amp 13D Note The first 3 characters are common to all seals When selecting required seal you must specify only the 9 selections within the required seal No Fill Fluid Silicone DC 200 CTFE Silicone DC 704 Neobee M20 Syltherm 800 No Capillary No Nipple 5 feet 10 feet 15 feet 20 feet 25 feet Capillary 35 feet Length 5 feet 10 feet 15 feet PVC Coated SS 20 feet 25 feet 35 feet 2 inch long SS nipple close coupled 14G amp 17G Secondary Fill SS Armor Connection of Remote Seal to Meter Body No Selection No Seal Attached to Core Transmitter 000000000 3 3 3 Diameter Size Rating i l HHH ANSI Class 300 NE 80mm 316L SS 316L SS Hastelloy C 276 316L SS Hastelloy C 276 Hastelloy C 276 Flush Flanged Monel 400 Monel 400 Seal Tantalum 316L SS Non Wetted Material CS Nickel Plated Wetted
45. re published at http hpsweb honeywell com Cultures en US Products Instrumentation ProductModelSelectionGuides default htm Model Selection Guide ST 3000 Smart Transmitter Model DP GP amp AP Remote Seals Selection Series 100 Guide 34 ST 16U 32 Issue 50 Page 1 of 12 Instructions Select the desired Key Number The arrow to the right marks the selection available Make selections from each Table I II and IV using the column below the proper arrow Select as many Table III options as desired plus a communications option selection A e denotes unrestricted availability A letter denotes restricted availability Restrictions follow Table IV Key Number Ill Optional KEY NUMBER AAA Description Selection Availability 0 4 to 0 400 HO 0 10 to 0 1 000 mbar STR12D l Body Rating 2 500 psi 172 bar Compound Characterized 0 1 to 0 100 psi 0 0 07 ni bar STR13D Body Rating 2 500 psi 172 bar 0 5 to 0 500 psia ae to 0 35 barA STR44A Body Rating 500 psia 35 bar A 0 5 to 0 500 psi 0 0 34 bar to 0 35 bar STR14G Body Rating 500 psi 35 bar sme i 0 30 to 959 008 psi 0 2 1 oar to 210 bar STR17G l Body Rating 3 000 psi 210 bar Remote seal system pressure rating is body rating or seal rating whichever is less Base STR models no longer include a default communications option All units now require the Important Note selection of a communication option from Table Ill AN DE HC H6
46. rean Register of Shipping KR Certificate number LOX17743 AE001 Lloyd s Register LR Certificate number 02 60001 E1 amp E2 16 European Pressure Equipment Directive PED 97 23 EC ST 3000 Smart Pressure Transmitter The ST 3000 Smart Pressure Transmitters are in conformity with the essential requirements of the Pressure Equipment Directive Honeywell ST 3000 Smart Pressure Transmitters are designed and manufactured in accordance with the applicable portions of Annex Essential Safety Requirements and sound engineering practices These transmitters have no pressurized internal volume or have a pressurized internal volume rated less than 200 bar 2 900 psig and or have a maximum volume of less than 0 1 liter Article 3 1 1 a first indent Group 1 fluids Therefore these transmitters are not subject to the essential requirements of the directive 97 23 EC PED Annex and shall not have the CE mark applied For transmitters rated gt 200 bar 2 900 psig lt 1 000 bar 14 500 psig Honeywell maintains a technical file in accordance with Annex Ill Module A internal production control when the CE mark is required Transmitter Attachments Diaphragm Seals Process Flanges and Manifolds comply with Sound Engineering Practice NOTE Pressure transmitters that are part of safety equipment for the protection of piping systems or vessel s from exceeding allowable pressure limits equipment with safety functions i
47. rmance specifications are based on reference conditions of 25 C 77 F zero 0 static pressure 10 to 55 RH and 316L Stainless Steel barrier diaphragm Transmitter URL limit or maximum seal pressure rating whichever is lower Performance Under Rated Conditions General for all Models Parameter Output two wire Description Analog 4 to 20 mA or digital communications DE mode Options available for Foundation Fieldbus and HART protocol Supply Voltage Effect 0 005 of span per volt Damping Time Constant Adjustable from 0 to 32 seconds digital damping RFI Protection Standard Negligible 20 to 1 000 MHz at 30 volts per meter CE Conformity Europe 89 336 EEC Electromagnetic Compatibility EMC Directive NAMUR NE 43 Compliance Option NE SIL 2 3 Compliance Transmitter failure information is generated when the measuring information is invalid or no longer present Failure information is transmitted as a current signal but outside the normal 4 20 mA measurement signal level Transmitter failure values are lt 3 6 mA and 2 21 0 mA The normal signal range is 2 3 8 mA and s 20 5 mA SIL certified to IEC 61508 for non redundant use in SIL 2 related Safety Systems single use and for redundant multiple use in SIL 3 Safety Systems through TUV Nord Sys Tec GmbH amp Co KG under the following standards IEC61508 1 1998 IEC 61508 2 2000 IEC61508 3 1998
48. smitter Operating Conditions All Models Reference Rated Condition Operative Limits Transportation and Parameter Condition at Storage zero Static PE OF OF C OF C OF Ambient Temperature 25 1 7712 55 to 90 67 to 194 Humidity RH 10 to 55 O to 100 0 to 100 O to 100 Vacuum Region Minimum Pressure mmHg absolute Atmospheric See Figure 15 for vacuum limitation Supply Voltage Current and Load Resistance Voltage Range 10 8 to 42 4 Vdc at terminals Current Range 3 0 to 21 8 mA Load Resistance 0 to 1 440 ohms as shown in Figure 16 Maximum Allowable Working Pressure MAWP MAWP ST 3000 products are rated to Body Maximum Allowable Working Pressure MAWP depends on STR12D Approval Agency and transmitter STR13D materials of construction STR12D STR13D STR14G STR17G STR14A MAWP is minimum of Body Rating or Seal Rating See Model Selection Guide for Seal MAWP 2 500 psig 172 bar 2 500 psig 172 bar 1 450 psig 100 bar 1 450 psig 100 bar 500 psig 35 bar 3 000 psig 207 bar 500 psia 35 bara Bolted Process Heads Tablel__A Bolted Process Heads Tablel__A All Welded Process Heads Table C All Welded Process Heads Table C a e uu a Ambient Temperature Limit is a function of Process Interface Temperature See Figure 13 MAWP applies for temperature range 40 to 125 C However Static Pressure Limit is de rated to
49. then check the box V adjacent to the type of protection used on the equipment certification nameplate Once a type of protection has been checked on the nameplate subsequently the equipment shall not be reinstalled using any of the other certification types TABLE IV Selection Factory Identification XXXX le e e Ordering Example STR12D 11A OAQAFAAA11AH HC LP 2J XXXX ST 3000 Smart Pressure Transmitter 41 34 ST 16U 32 Issue 50 Page 11 of 12 RESTRICTIONS Restriction Available Only With Not Available With Letter Table _______Selection_______ _Table_ ____ _Selection______ Select only one option from this group l TC ME or FISCO FNICO compliance not FISCO FNICO compliance available available with 3C 3N 33 3S 2J CA Z2 ZD ZA 6D amp 6S only with 1C RESTRICTIONS continued ST 3000 Smart Pressure Transmitter 34 ST 16U 32 Issue 50 Page 12 of 12 Restriction Available Only With Not Available With Letter Table Selection Table _ ______Selection________ This approval code requires the selection of a certified conduit plug A5 A6 or A7 FM Approvals is a service mark of FM Global Hastelloy is a registered trademark of Haynes International Monel 400 is a registered trademark of Special Metals Corporation HART is a registered trademark of HART Communication Foundation FOUNDATION Fieldbus is a registered trademark of Fieldbus Foundation Teflon is a r
50. well Pte Ltd Phone 65 6580 3278 Fax 65 6445 3033 South Korea Honeywell Korea Co Ltd Phone 822 799 6114 Fax 822 792 9015 Specifications are subject to change without notice 44 ST 3000 Smart Pressure Transmitter For More Information Learn more about how Honeywell s ST 3000 Series 100 Remote Seal Transmitters can increase performance reduce downtime and decrease configuration costs visit our website www honeywellprocess com or contact your Honeywell account manager Honeywell Honeywell Process Solutions 1860 West Rose Garden Lane 34 ST 03 64 Phoenix Arizona 85027 March 2013 Tel 1 800 423 9883 or 1 800 343 0228 2013 Honeywell International Inc www honeywellprocess com
Download Pdf Manuals
Related Search
Related Contents
Provider Online Authorization Request Portal User Manual Tyco Electronics 19R Laptop User Manual 成田市廃棄物減量等推進員について(平成27年度版) Memorex MPH2430 Cordless Telephone User Manual USB Attitude Indicator Software Manual (English, Pdf) Zanussi ZF62/26 Use & Care Manual C 310 ECO - De Dietrich Copyright © All rights reserved.