1. Comfort range setup The user can adjust the mattress pressure level by adjusting the comfort range from 1 to 6 Please consult your physician for an appropriate setting When the Low Pressure Indicator de illuminates the mattress has reached its preset pressure level Alternating function cycle time setup The user can adjust the alternating cycle time mode as follows Please consult your physician for the appropriate setting AltaDyne XT 773400A cycles between 5 10 15 20 min AltaDyne XD 774200A cycles between 10 15 20 25 min Seat Inflation Function The seat inflation function provides extra pressure compensation around the buttocks area for better surface support when the patient sits on gets on or gets off the mattress Disable this function when user is lying down to prevent excessive pressure around the buttocks area 773400A INS LAB RevE14 11 LOW PRESSURE AND ALARM FUNCTION When air pressure is below the preset pressure for a prolonged period of time the control panel Low Pressure Indicator yellow LED will flash and the alarm will sound Ensure that all connections are secure and correctly fastened per preceding installation instructions To silence the alarm press the Alarm Mute button it will be silent for three minutes but the Low Pressure Indicator will still flash After three minutes if the pressure has not yet returned to normal the alarm will sound again Info If the pressure level is con
2. Atmosphere 10 15 psi 15 37 psi 70 kPa 106 kPa Classification Class Il Type BF Applied Part Air Mattress Not suitable for use in the presence of a flammable anesthetic mixture IPXO enclosed equipment without protection against ingress of water Continuous operation 773400A INS LAB RevE14 17 Mattress Specification Dimensions AltaDyne XT 78 7 x 35 4 x 8 LxWxH 773400A 200 cm x 90 cm x 20 3 cm AltaDyne XD 78 7 x 42 x 10 774200A 200 cm x 106 7 cm x 25 4 cm Arrangement AltaDyne XT 20 x 8 cells 773400A AltaDyne XD 774200A Material Cover Quilted top sheet Slip resistant bottom mat Top cover fabric Nylon PU vapor permeable fluid resistant Nylon PU 18 773400A INS LAB RevE14 LIMITED WARRANTY SCOPE OF WARRANTY GF Health Products Inc GF warrants to the original purchaser only that it will replace or repair components at GF s sole discretion that are defective in material or workmanship under normal use and service All warranties are conditioned upon the proper use of the products strictly in accordance with good commercial practice and applicable GF instructions and manuals including proper use and maintenance To the extent that a component is warranted by a third party GF conveys all of its rights under that warranty to the original purchaser to the extent permitted This limited warranty shall only apply to defects that are reported to GF s customer service tea
3. Z DANGER Do not use this product while bathing DANGER Do not place or store this product where it can fall or be pulled into a tub or sink A DANGER Do not place this product in or drop into water or other liquid NDANGER Do not reach for a product that has fallen into water Unplug it immediately WARNING To reduce the risk of burns electrocution fire or personal injury WARNING Do not leave this product unattended when plugged in Z WARNING Always use close supervision when this product is used by on or near children or those who require close supervision Z WARNING Use this product only as intended and described in this manual Do not use attachments or accessories not recommended by Graham Field 4 773400A INS LAB RevE14 A WARNING Never operate this product if a it has a damaged cord or plug b it is not working properly c it has been dropped or damaged or d it has been dropped into water Return the product to your Graham Field authorized distributor Z WARNING Keep the cord away from heated surfaces A WARNING Never block the air openings of this product or place it on a soft surface such as a bed or couch where the openings may be blocked Keep the air opening free of lint hair and other similar particles A WARNING Never drop or insert any object into any opening or hose A WARNING Connect this product only to a properly grounded power outlet A WARNING Maximum weight cap
4. the mattress While the mattress is filling the low pressure indicator non flashing yellow LED will illuminate When the selected pressure level is reached the low pressure indicator will de illuminate 6 Replace all sheets blankets and pillows on the bed 10 The patient can now lie on the bed The pressure may now be adjusted to the physician prescribed setting using the control panel Comfort Range button The alternating function may now be adjusted to the physician prescribed setting using the control panel Cycle Time button 773400A INS LAB RevE14 PRESSURE ADJUSTMENT Generally a lighter patient will need a lower softer setting while a heavier patient will need a higher firmer setting but pressure adjustment must ultimately be based on the patient s weight distribution A short stocky person may be the same weight as a tall thin person but their settings will not be the same The mattress should be firm enough to support the patient yet soft enough so the patient will sink down to eliminate pressure points Test the pressure by placing four fingers below the mattress under the patient s buttock area your fingers should just fit If the fit is too loose decrease the setting If your fingers do not fit well increase the setting two notches firmer Wait ten minutes and then check again You should only have to adjust this setting one time per patient Please consult your physician for an appropriate setting
5. ATION UNPACKING 1 Before unpacking check for obvious damage to the carton s or its contents If damage is evident please notify the carrier and your Graham Field authorized distributor 2 Remove all loose packing from the carton s 3 Carefully remove all components from the carton s 4 Inspect all components If damage is evident please notify the carrier and your Graham Field authorized distributor PUMP AND MATTRESS INSTALLATION 1 Place the mattress on top of the bed frame ensure that the foot end symbol faces up at foot end as shown at right Secure the mattress to the bed by affixing the elastic straps to the bottom bed frame 2 Fold open the wire hangers on the back of the pump as shown at right 3 Either hang the pump on the bed s a foot end as shown at right or place t the pump on a flat surface easily accessible to the user caregiver and or doctor A NOTICE Place the device ina position where the user caregiver and or doctor can access it easily 4 773400A INS LAB RevE14 4 Connect the mattress pump air hose quick connector as shown at right f ensure that the connector has clicked Ab securely into place before continuing HIN A NOTICE Ensure that the air hoses are not kinked or tucked under the mattress Z WARNING Locate all cords so that they will not be stepped on tripped over or otherwise subjected to damage or stress A WARNING Do not use a household exten
6. LUNEX AltaDyne XT Model 773400A AltaDyne XD Model 774200A Alternating Pressure Low Air Loss Replacement Mattress System Important Do not operate the Mattress System without first reading and understanding this manual Save this manual for future use The latest version of this manual can be found online at www grahamfield com 773400A INS LAB RevE14 CONTENTS INTRODUCTION aaa NA AA AA EEOSE RAEN 3 INTENDED USE OF THIS DEVICE u 1 11 1n aaanaanasaaasaanansana 3 CONTRAINDICATION 1 2 nananana 3 IMPORTANT SAFETY PRECAUTIONS 1111111110 namana 3 DANGER aa GANERN GURL AA 4 WARNING maa al Na a 4 NOTICE AA AA AA 5 EQUIPMENT SYMBOLS ATA AA 6 PRODUCT DESCRIPTION a AA TARARA ARNAN 6 INSTALLATION AN BA BANA RA 7 UNPACKING paaa a ANA NAA Ala 7 PUMP AND MATTRESS INSTALLATION ccsecsesseesseeseseneseeeesnnes 7 OPERATION aaa Aa AA AA 8 CONTROL PANEL FUNCTIONS 1 011 1 nnnnananannananaasaa nananana 9 OPERATION asam NANANA NAA niiae 10 PRESSURE ADJUSTMENT u s 11 COMFORT RANGE SETUP cccccsecccseeeseeeseeeeseeeeeseeeeseeeesenenees 11 ALTERNATING FUNCTION CYCLE TIME SETUP n se 11 SEAT INFLATION FUNCTION cccccssesssesseeeeseeeesseeeeneeensenenees 11 LOW PRESSURE AND ALARM FUNCTION cccccssessseeessteneereeeeees 12 QUICK DISCONNECT i iiscctcccccsccecesscceesssvescscecscceasccsesczerseccessbeverstessszeess 12 CPR FUNCTION aaa NAAN SANA 12 MAINTENANCE RAANG 13 CLEANING sacs secccceccccecscesccncteac
7. OR WAGES PAYMENTS AND BENEFITS The warranties contained herein contain all the representations and warranties with respect to the subject matter of this document and supersede all prior negotiations agreements and understandings with respect thereto The recipient of this document hereby acknowledges and represents that it has not relied on any representation assertion guarantee warranty collateral contract or other assurance except those set out in this document For additional information on this product or this warranty please contact a GF Customer Service Representative NOTES 1 Additional terms and conditions may apply 2 Freight claims must be notated on the Bill of Lading and must be made with immediacy The ICC regulations govern specific requirements for freight claims Failure to abide by those regulations may result in a denial of the freight claim GF will assist you in filing the freight claim 3 Claims for any short shipment must be made within thirty 30 days of the invoice date GF Health Products Inc 2935 Northeast Parkway Atlanta GA 30360 Tel 770 368 4700 Fax 770 368 2386 www grahamfield com 773400A INS LAB RevE14 19 Manufactured for GF Health Products Inc 2935 Northeast Parkway Atlanta Georgia 30360 tel 770 368 4700 fax 770 368 2386 GRAHAM FIELD www grahamfield com 2005 GF Health Products Inc Made in Taiwan
8. acity for this product is AltaDyne XT 773400A 400 Ib 182 kg EVENLY DISTRIBUTED AltaDyne XD 774200A 800 Ib 364 kg EVENLY DISTRIBUTED A WARNING Notice for California Customers California Proposition 65 WARNING This product contains a chemical known to the State of California to cause cancer and reproductive or developmental harm NOTICE A NOTICE The pump can be used only with the accompanying mattress Do not use it for any other purpose 773400A INS LAB RevE14 5 EQUIPMENT SYMBOLS A Attention Consult accompanying documents this manual BF symbol Indicates that this product is in accordance with the degree of protection against electric shock for type BF equipment Class Il equipment 2 concentric squares indicating double 0 insulation IPXO This equipment has no protection against ingress of water PRODUCT DESCRIPTION The AltaDyne XT AltaDyne XD offers both low air loss and alternating pressure therapy The pressure and alternating cycle time are adjustable in order to maximize therapy and comfort for each individual The patented cell in cell feature is designed to prevent bottoming out in the event of a power failure The easy to operate digital control unit features a lock out setting designed to prevent changes by unauthorized personnel P PO mattress NN s ti i lg s77 panel e connector CPR valves pump and mattress system 6 773400A INS LAB RevE14 INSTALL
9. cessccressecteczsessccesanteccoecesscesssnessancestecresse 13 GENERAL MAINTENANCE 1101 111m 13 FUSE REPLACEMENT a aaah aaa 14 AIR FILTER CLEANING REPLACEMENT cccccsseseseeeseeenees 14 STORAGE eicccsceccveccesscscecsseccvsscresescsstccessacsectecactessssecscctteaseesesteesstersaczeeszs 15 TROUBLESHOOTING Nama aaa 16 SPECIFICATIONS RAK LA AA 17 LIMITED WARRANTY ca TAGAN 19 GF Health Products Inc is not responsible for typographical errors All illustrations specifications packaging and warranties contained in this document are based on the latest product information available at the time of printing The most current product information including the most current revision of this manual can be found online at www grahamfield com Graham Field Lumex and AltaDyne are registered trademarks of GF Health Products Inc 2 773400A INS LAB RevE14 INTRODUCTION Use this manual for set up and operation of the Alternating Pressure Low Air Loss Mattress System Read all instructions before using the mattress system Save this manual for future reference INTENDED USE OF THIS DEVICE The AltaDyne XT AltaDyne XD Alternating Pressure Low Air Loss Mattress System is an advanced alternating pressure relief replacement system and low air loss system suitable for the prevention and treatment of up to and including stage 4 pressure ulcers Contraindication A WARNING DO NOT use this product in the presence of flammable ane
10. d or mattress straps are securely fastened Air tube connector Press handles on both sides of air tube connector won t fit into pump Ensure the connector is tightly secured by listening for click while connecting No air produced This is normal since this system has alternating mode from some air tube Air outlets alternate to produce air during their preset connector air outlets cycle time Contact your Graham Field authorized distributor if you are unable to solve your problem after using the above steps 16 773400A INS LAB RevE14 SPECIFICATIONS Standards This pump and mattress UL60601 1 IEC60601 1 2 system has been independently tested and successfully approved to the following standards For U S and Canada only Medical Equipment Air Pump with respect to electrical shock fire and mechanical hazards only in accordance with UL60601 1 and CAN CSA C22 2 No 601 1 Pump Specification Power Supply 120V AC 50 60 Hz Fuse Rating A Cycle Time AltaDyne XT 5 10 15 or 20 minutes 773400A AltaDyne XD 10 15 20 or 25 minutes 774200A Air Output gt 16 LPM Dimensions L x W x H 14 4 x 10 5 x 5 37 x 27 x 13 cm Weight BBA Temperature Operation 50 F to 104 F 10 C to 40 C Storage 14 F to 122 F 10 C to 50 C Shipping 14 F to 158 F 10 C to 70 C Humidity Operation 30 to 75 non condensing Storage 10 to 90 non condensing Shipping 10 to 90 non condensing
11. excessive wear 2 Check mattress cover for signs of wear or damage Ensure mattress cover and tubes are connected correctly 3 Check airflow from the air hose connector 4 Check the air hose connectors for damage 5 Check the air hoses for any kink or break For replacement please contact your Graham Field authorized distributor 6 Contact your Graham Field authorized distributor with any problems 773400A INS LAB RevE14 13 Fuse replacement tool needed small screwdriver A WARNING Ensure that the main power plug is 4 unplugged before fuse replacement Tf you suspect a blown fuse disconnect the plug from the wall outlet immediately Use a small screwdriver to remove the cover of the fuse holder fuse locations are shown in picture at right Insert a new fuse of the same rating Replace the fuse holder cover Air filter cleaning replacement A WARNING Ensure that the main power plug is unplugged before air filter cleaning or replacement The air filter is located on the back of the pump Clean the air filter once per month l 14 Snap out the air filter plate filter location is shown in picture at right Remove the air filter The filter is reusable if not torn and can be washed gently with a mild detergent and water Air dry the filter completely before replacing in pump Replace filter and plate after filter has dried completely Replace the air filter more often than o
12. m within the applicable warranty period and which upon examination by GF or its authorized representative prove to be a warranty item This limited warranty is not transferable The warranted components and time period are set forth below Control unit two years Mattress one year The applicable warranty period shall commence from date of shipment to the original customer unless there is an expiration date on the component in which case the warranty shall expire on the earlier of warranty period or the expiration date OBTAINING WARRANTY SERVICE A GF Customer Service Representative must authorize warranty service Please contact the GF Customer Service department by calling 678 291 3207 sending a fax request to 770 368 2386 or by e mailing a request to cs grahamfield com Specific directions will be provided by the Customer Service Representative Failure to abide by the specific directions will result in denial of the warranty claim EXCLUSIONS The warranty does not cover and GF shall not be liable for the following 1 Defects damage or other conditions caused in whole or in part by misuse abuse negligence alteration accident freight damage tampering or failure to seek and obtain repair or replacement in a timely manner 2 Products which are not installed used or properly cleaned and maintained as required in the official manual for the applicable product 3 Products considered to be of a non durable nature i
13. nce per month if it is gray torn or the environment is dirty Contact your Graham Field authorized distributor for replacement filters 773400A INS LAB RevE14 STORAGE l Disconnect the air hose connectors and CPR valves Ensure that all cells are deflated Lay the mattress out flat and upside down Lay the air hose connector on the inner side of the mattress to protect it Roll from the head end toward the foot end Stretch the foot end strap around the rolled mattress to prevent unrolling Store mattress and pump in a dust free environment with no exposure to direct sunlight A NOTICE Do not fold crease or stack the mattress 773400A INS LAB RevE14 15 TROUBLESHOOTING CU sowon ooy S Power is not ON Ensure that plug is connected to power Check if fuse is blown Alarm is on Check if the connection between pump and hose quick audible amp visible connector is secure Check if the CPR connection is secure Check if any air cells have leakage Check if all tubing connections along mattress are secure Patient is bottoming Pressure setting may be too low for patient adjust out comfort range 1 to 2 levels higher wait a few minutes recheck Mattress becomes too AutoFirm may have been activated for too long firm and pressure can t Release some air by disconnecting the air tube be reduced connector change to preferred comfort level reconnect Air mattress is loose Ensure snap buttons an
14. ncluding but not limited to casters filters fuses gaskets lubricants and charts 4 Accessories or parts not provided by GF 5 Charges by anyone for adjustments repairs replacement parts installation or other work performed upon or in connection with such products which are not expressly authorized in writing in advance by GF 6 Any labor or shipping charges incurred in the replacement part installation or repair 7 Costs and expenses of regular maintenance and cleaning and 8 Representations and warranties made by any person or entity other than GF ENTIRE WARRANTY EXCLUSIVE REMEDY AND CONSEQUENTIAL DAMAGES DISCLAIMER THIS WARRANTY IS GF S ONLY WARRANTY AND IS IN LIEU OF ALL OTHER WARRANTIES EXPRESS OR IMPLIED GF MAKES NO IMPLIED WARRANTIES OF ANY KIND INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE IF ANY MODEL OR SAMPLE WAS SHOWN TO THE CUSTOMER SUCH MODEL OR SAMPLE WAS USED MERELY TO ILLUSTRATE THE GENERAL TYPE AND QUALITY OF THE PRODUCT AND NOT TO REPRESENT THAT THE PRODUCT WOULD NECESSARILY CONFORM TO THE MODEL OR SAMPLE IN ALL RESPECTS THIS WARRANTY IS LIMITED TO THE REPAIR OR REPLACEMENT OF THE DEFECTIVE PARTS GF SHALL NOT BE LIABLE FOR AND HEREBY DISCLAIMS ANY DIRECT SPECIAL INDIRECT INCIDENTAL EXEMPLARY OR CONSEQUENTIAL DAMAGES INCLUDING BUT NOT LIMITED TO DAMAGES FOR LOSS OF PROFITS OR INCOME LOSS OF USE DOWNTIME COVER OR EMPLOYEE OR INDEPENDENT CONTRACT
15. similar to a regular air mattress To go back to alternating mode press this button again Increases the body zone s pressure above the preset setting ensuring adequate support for sitting position and getting on or off the mattress Disable this function when the user is lying down to prevent excessive pressure at body zone Adjusts the comfort level Use button to increase or decrease level increases level decreases level The mattress has 6 comfort ranges 1 is the softest 6 is the firmest Please consult your physician for the appropriate setting Adjusts the cycle time Use button to increase or decrease cycle time to that desired increases cycle time decreases cycle time Please consult your physician for the appropriate setting AltaDyne XT 773400A cycles between 5 10 15 20 min AltaDyne XD 774200A cycles between 10 15 20 25 min 773400A INS LAB RevE14 9 OPERATION Before turning power ON ensure that air tube connectors and CPR valves are securely fastened ba Turn the power switch shown at right to the ON position Press the control panel Power button to start the system A GU power Press the control panel AutoFirm switch button to activate the AutoFirm function Info Use AutoFirm mode for the quickest inflation at set up AutoFirm mode will last for 20 minutes then automatically return to the default setting 4 The pump will begin delivering air into
16. sion cord if the electrical cord does not reach the power outlet Use of an improper extension cord could result in fire and electric shock If an extension cord must be used use a three conductor cord with ground properly wired in good electrical condition and keep it as short as possible Z WARNING Ensure that the local power voltage is appropriate for the pump unit 5 Plug the power cord into a properly grounded electrical outlet OPERATION A WARNING Always read operating instructions before use 8 773400A INS LAB RevE14 CONTROL PANEL FUNCTIONS Pressure Indicator Static Inflation Comfort Range Cycle Time Turns on the control unit It takes approximately 15 minutes to inflate the mattress Enables and disables the audible alarm feature Press it to silence the audible alarm The alarm buzzer will sound again after three minutes if the pressure has not yet returned to normal Locks the current setting to prevent others from changing it To unlock the panel press this button and hold it for three seconds 1 No light normal condition mattress is operating properly 2 Non flashing yellow light pressure has not reached the preset setting 3 Flashing yellow light low pressure For maximum inflation of mattress All other functions are disabled until this button is pressed again This mode lasts for 20 minutes then automatically returns to default setting Selects non alternating mode
17. sistently low check for leakage at air cells or connecting tubes If necessary contact your Graham Field authorized distributor to replace damaged cells or tubes QUICK DISCONNECT To disconnect the device quickly pull the power plug from the wall connector CPR FUNCTION If CPR must be performed on the patient while the mattress is in use to immediately deflate the mattress either 1 Pull the CPR valves from the mattress in the location of the arrow shown at right or 2 Disconnect the quick connector from the pump in the location of the arrow shown at right 12 773400A INS LAB RevE14 MAINTENANCE CLEANING A WARNING Ensure that the main power plug is unplugged during cleaning Perform the following cleaning procedures before use Pump Wipe the pump with a clean damp cloth and mild detergent and keep it away from dust Air dry A NOTICE Do not use phenolic products or corrosive or powdered cleansers to clean the pump A NOTICE Do not immerse or soak the pump Mattress Wipe the mattress unit with a clean damp cloth and mild detergent The mattress may also be cleaned using a 10 solution of sodium hypochlorite bleach diluted in water Air dry all parts thoroughly before use A NOTICE Do not use phenolic products to clean the mattress A NOTICE After cleaning air dry the mattress without direct exposure to sunlight GENERAL MAINTENANCE 1 Check main power cord and plug for abrasion or
18. sthesia There is a possible fire hazard when this product is used with certain oxygen delivery equipment Use nasal cannula face mask or 1 2 length oxygen tent to deliver oxygen in the presence of this product DO NOT use a full length oxygen tent that extends past the top surface of the mattress with this product IMPORTANT SAFETY PRECAUTIONS The safety statements presented in this chapter refer to the basic safety information that should be observed by those using the Alternating Pressure Low Air Loss Mattress System There are additional safety statements in other chapters or sections which may be the same as or similar to the following or specific to the operations Z DANGER Indicates an imminent hazard situation that if not avoided will result in death or serious injury A WARNING Indicates a potential hazard situation or unsafe practice that if not avoided could result in death or serious injury 773400A INS LAB RevE14 3 CAUTION Indicates a potential hazard or unsafe practice that if not avoided could result in minor or moderate personal injury A NOTICE Indicates a potential hazard or unsafe practice that if not avoided could result in product or property damage Info Provides application recommendations or other useful information to ensure that you get the most from your product DANGER To reduce the risk of electrocution A DANGER Always unplug this product immediately after using
Download Pdf Manuals
Related Search
Related Contents
PA-T300 Sommaire Mode d`emploi Electrophoresis Units Bottle Top Dispenser Semiconductor Characterization System Electrolux TJ1053 User's Manual Introduction to the JUROPA3 Experimental Partition FT55-RLAP Manual V.1 Wiley Social Computing with Microsoft SharePoint 2007: Implementing Applications for SharePoint to Enable Collaboration and Interaction in the Enterprise Copyright © All rights reserved.