Archimede User Manual rev.2_EN
Contents
1. UTILITIES The utilities menu features commands to set the DL clock to run a self test and to change the user interface language Click the Utility button in the main menu Utilities 1 Set date amp time 2 Unit status 3 Set language SETTING DATE AND TIME Click the Set date amp time button in the Utilities menu to display the current date and time Set date amp time _Press_an arrow key to edit Press an arrow key to edit To change the current date and time click one of the arrow keys a cursor will be displayed on the first digit of the hours field Set date amp time E Press an arrow key to edit Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual You may move the cursor using the arrow keys or change the current value by typing the new value with the numeric keys When you are finished editing you may either commit the new value by pressing ENTER or discard with ESC CHECKING DL STATUS Click the Unit status button in the Utilities menu to check the current status The DL will display a list of the tests being performed and mark each test for success or failure Unit status y Checking x New Leonardo ver 0 15 S Disk 1928736 kB V Keyboard X Time NOT SET Battery 12 0 AD converter 28 9C AD calibration 06 07 10 E No key should be pressed during the DL self test so if you keep a key pressed the keyboard test will fail The tem
2. To send data to a remote BlueTooth device select it in the list and press the ENTER key Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual Select site ii A TEST The DL will show a list of sites INC 1 01 SPIR 00 GRAF 00 GRAF 01 GRAF 02 When using a PC you should now select the Receive a File option in the BlueTooth menu Select the desired reading file for Archimede DL or site for other DL and press ENTER the DL will start sending the acquisitions file Sending through Bluetooth TO sending file content a The PC or SmartPhone that is receiving the data may ask you to authorize the reception or storage of data Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual There are some precautions to take with this Operative System OS CINDROID e The user should use an application that allows data transfer through Bluetooth because this feature is not natively supported by the OS In the Android Market there are some free application SISGEO suggest a free application called Bluetooth File Transfer very appreciated by users recommended on many sites and forums e This application should be initiated and maintained in the background if you close it the SmartPhone does not receive the file e Once received the file is saved in the default directory of the SD inside the smartphone You should find the file and then you sho
3. Code Unit Decimals Rev 2 31 01 2013 The name used to identify the probe this is initially set the same as the probe category but should be changed by the user before saving the probe User notes related to the instrument The instrument model this field cannot be edited by the user The unit of measure the user may either select it from the drop down list or edit a new unit of measure Can t be changed for digital probes The number of digits after the decimal separator used when displaying readings for example 3 means that the number will be formatted as 20 000 We suggest to set o Digital probes require no calibration since all the parameters are already inside the inside Go to paragraph INCLINOMETER TUBES CONFIGURATION if you are using a digital probe Mji Page Archimede Portable Dataloggers User Manual After editing data in this screen press ENTER the DL will ask you to select the calibration type Select calibration type BI Probe Polynomial You should select the calibration type according to the calibration report of the probe Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual PROBE CALIBRATION FOR ANALOG PROBE Probe calibration is normally used for biaxial inclinometers this calibration requires two sets for parameters one for each axis these parameters are labeled Sensitivity Electrical offset Mechanical offset Cross axis Multiplicative
4. Mini USB Connector for PC connection record key Jack Plug for battery charger SISGEO MILANO ITALY MIL Connector for instrument connection INTRODUCTION Depending on the type of your portable datalogger DL it is possible read different instrument types Analog Digital Probe Vibrating wire Leonardo Y Y New Leonardo Y Y Y Galileo Y Archimede Y Y For further details on instruments type please see Appendix A iPage Rev 2 31 01 2013 Archimede Portable Dataloggers User Manual The Leonardo DL has been developed to make and store readings obtained from various instruments analog and digital except for vibrating wire instruments Leonardo DL can not perform inclinometer tests The Leonardo DL is mainly composed by e Shoulder and shipment bag OCDLOBAG927 e Battery charger for NiMH batteries OECAB12VNMB e Flying cable 1 MIL connector and 6 alligator clips ground OECAV7P6A00 e P68 USB cable for PC connection e CD ROM Manuals Smart Manager Suite Shipment bag Shoulder bag Flying cable USB cable Battery charger Rev 2 31 01 2013 iPage Archimede Portable Dataloggers User Manual The new Leonardo DL has been developed to make and store readings obtained from various instruments analog digital and vibrating wire instruments The newLeonardo DL is mainly composed by e Shoulder and shipment bag OCDLOBAG927 e Battery charger for NiM
5. firmware upgrade scroll NO E 5 5 j m 4 6 JO off Read Smart Manager manual for detail Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual EARLIER FIRMWARE VERSION In case some instruments have been configured with an earlier version of the firmware any changes will not be applied to the instruments already configured To apply the changes introduced by a new firmware you should remove the instrument from the datalogger and re insert it If for example with a new firmware it has been changed the power supply of a instrument this one will be powered with the power supply set in the old firmware If you want to use the new power supply it is necessary to remove the instrument from the datalogger and re insert it Rev 2 31 01 2013 Page Portable Dataloggers User Manual TECHNICAL DATASHEET New Leonardo Leonardo Archimede Galileo Technical Specifications m mV V mV V C Hz ys digit C pE n 2 24bit A D converter with A D converter autocalibration 19 true bit 4 20mA 10mV 400mV 5V Measuring range 10000 Pt100 100000 NTC from 400 to 6000Hz 1114 with F S 20mA 1pV with FS 10mV 10pV with F S 400mV 100uV with F S 5V 0 1 C with F S 10000 Pt100 0 1 C with F S 100000 NTC 0 1 Hz with FS from 400 to 6000Hz 0 001mW V with F S 10mV V Accuracy 0 01 F S Type of measurement Hz u5 digit C pe 100009 NTC from 400
6. materials for a standard and appropriate use Repairs and or recalibrations should always be carried out at SISGEO factory or by personnel trained and authorized by SISGEO In case of repair and or recalibration please send us the device and every accessory in the container provided in order to prevent damage from shocks during transport This procedure involves the filling out of the form on our website www sisgeo com to the link assistance repairs once inserted the request a confirmation with the related authorization code will be sent to the e mail address indicated The material has to be packed properly and sent in door to door basis freight to client charge to SISGEO S r l Via F Serpero 4 F1 20060 MASATE MI Attn to Technical Assistance Office Tel 39 02 95764130 Fax 39 02 95762011 e mail assistance sisgeo com PHONE ASSISTANCE Telephone support for installation configuration of equipment measurement systems and Software TECHNICAL ASSISTANCE ON SITE FOR THE SUPPLIED INSTRUMENTATIONS Technical support for the installation of monitoring systems with its commissioning and start up through the add of specialized technician s teams TRAINING Possibility of training at our Factory or at site related to the installation procedures the system configuration and the use of measurement systems Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual APPENDIX A PROBES DIGITAL PROB
7. A1 A3 Az A4 B1 B3 B2 B64 2 i 2 2 2 The readings in the matching runs should be about opposite so that when reading the second of the matching runs the check values should be close to zero these values are only available when the matching runs has already been performed On the right of this box the current groove is shown The bottom blue box shows the Reference values that is the values read in the same run and at the same depth during the reference test and on the right the Current level in the tube At the center the current values are shown The color of the unit of measurement shown in the left part of the screen alternates from black to blue every time reading values are updated 20 09 12 20 09 12 17 13 29 sin a sin a 17 10 51 Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual When are used digital probe it is possible to read the temperature taken inside the probe 20 09 12 20 09 12 17 26 48 17 26 45 226 C 226 C To display the temperature press the Right Arrow Key to return to the unit of measurements press Left Arrow key Click Measure to record the values and proceed to the next step If the DL finds an incompatibility with the digital instrument i e has been selected a mono axial digital probe while the connected probe is bi axial will be shown a dialog box Error reading instrument Gil Instrument is not compatible If the DL can t read the conne
8. DATACAL Pag 48 USE WITH SMART REMOTE Pag 49 USING THE BLUETOOTH INTERFACE Pag 51 Android 2 2 Froyo Pag 53 Symbian S60 3 Edition Pag 53 RIM ver 5 0 0 822 Pag 53 Windows Mobile ver 5 6 Pag 54 UTILITIES Pag 55 SETTING DATE and TIME Pag 55 CHECKING DL STATUS Pag 56 CHECKING USER INTERFACE LANGUAGE Pag 57 TROUBLESHOOTING Pag 58 ASSISTANCE Pag 59 Repairs and maintenances Pag 59 Phone assistance Pag 59 Technical assistance on site for the supplied instumentations Pag 59 Training Pag 59 APPENDIX A Pag 60 Probes Pag 60 SPECIAL KEYBOARD COMBINATIONS Pag 61 Switch ON Pag 61 Switch OFF Pag 61 Emergency reset Pag 61 Firmware upgrade Pag 61 EARLIER FIRMWARE VERSION Pag 62 TECHNICAL DATASHEET Pag 63 Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual NOTE ON SYSTEM UTILIZATION To ensure a safe and efficient functioning please read the following instructions carefully before use Any unauthorized modification is considered a total responsibility of the user In addition to the standards listed above the user must comply with the existing legislation on safety and health of employees in the workplace Check the product for damages caused by transport Verify that your package includes all standard accessories and any accessories required In case contact SISGEO or the distributor The use of the portable datalogger shall be done only after the proper connection and following the config
9. New button The DL will display a screen asking the name for the tube Name Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual Type the name you wish to use for the new tube and press ENTER Type ESET Ru AIBUASBS Unit e mt Oft Length P Step Azimuth Notes The DL will ask for data for the new tube these data include Type The type of tube this may be either Vertical or Horizontal Run The type of measure to be performed on the tube this may be e A1B1 A3B3 e A2B2 A4B4 e A1B1 A2B2 A3B3 A4B4 e Spiralometer e Trex Unit The unit used to measure the steps either meters or feet Length The useful length of the tube in the specified unit Step The useful length of the step in the specified unit Azimuth The azimuth of the tube this parameter may be left blank Notes User notes Enter the required data and press ENTER to save the tube definition Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual READING INCLINOMETER TUBES Once a tube has been defined and you have already loaded the definitions for the probes you wish to use in your DL you may proceed to perform tests on tubes In case the unit switches off during an inclinometer test for example because of a low battery readings made so far are saved Note however that the inclinometer test file is closed and it is no more possible to add further measures to it Anyway it will be pos
10. both channels of the TREX probe when a ring is detected both values will be in the 1000mV to 9000mMV range On the right the Current Direction Up in our case is shown The bottom blue box displays the Current Level in the tube and the Reference value read in the same run and at the same depth during the reference test At the center the current value is shown this is only displayed when both rings have been detected a single value is calculated from the two channels of the T REX the values read on the two channels are converted in millimeters and their difference is shown le 05 ImV B 3834 pi aa 10 78 Back 4 0 Ring found Ref Level Rev 2 31 01 2013 Mi Page Archimede Portable Dataloggers User Manual When a valid value is displayed you may click the Read button to record the current value REX45100S TR wl 02 08 10 mV A 4838 Dir 17 33 56 mV B 3835 Up mm HHH Ht Save a stored alue store Ref Level The DL will store the value and get ready for the next level BROWSING INCLINOMETER READINGS You may browse data in an inclinometer read while you are performing the test by clicking the Data button in the window From the Main menu you may browse acquired data for any test select Show readings in the Main menu The DL will first ask you to select the site then the inclinometer tube after this it will show a list of tests performed on the tube JO 02 0
11. system stops the power circuit Repeat the startup procedure see paragraph Special combination keys in the reference DL manual Connect the battery charger and try to turn the DL on If the DL turns on contact SISGEO s assistance Wait at least 30 minutes to allow the battery to become colder and go back in the appropriate temperature range Check DL battery level available in menu Utilities gt Unit Status directly on DL The battery charger could have not correctly detected the battery Battery s temperature overcomes 70 C internal security system stops the power circuit Display backlight could be turned off or set on minimum value DL software generates an unexpected error Sensor s connections are wrong A D converter does not understand signals correctly Instrument is connected to DL in a wrong way A D converter does not understand signals correctly USB connection could be wrong Repeat the charge procedure checking the LED colour sequence Insert always first the jack in the DL and then the plug in the network plug Wait at least 30 minutes to allow the battery to become colder and go back in the appropriate temperature range Turn on or set the back light through the special keys see paragraph special combination keys in the reference DL manuals If this procedure will not solve the proble contact SISGEO assistance Reset through keys combination see special combination keys in th
12. test any configured probe this may be useful to check that a probe operates correctly before starting an inclinometer read iPage Archimede Portable Dataloggers User Manual CONFIGURING SITES DLs have a permanent memory used to store configuration data such as probes and Inclinometer Tubes as well as acquired data that is Readings and Inclinometer Tests Also if the memory s maximum capacity is remarkable 2GB DL export and synchronization speed with the PC depends on the inside data quantity We advise not to charge exceedingly DL s memory and to use it only to save the Dl s files All data in a DL is divided in sites how to define sites is up to the user but it is expected that they are named after the physical sites in which measurements are performed To view the site s list select Configuration in the main menu then select Sites The list of sites is initially empty To add a new site make sure that the edit box at the bottom of the screen is active you should see a blinking cursor in it and type the name for the new site then press ENTER Configuration sites LLO A MYSITE Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual The new site will be displayed in the list on the screen Configuration sites C MYSITE A ad The name of a site as well as the name of a Multiplexer Switch Panel Worklist or Inclinometer Tube should be composed
13. the USB cable cannot be used and connecting an USB cable when the DL expects a push button will cause the PC to issue a warning that the USB detection failed To start using the USB select Communication in the main menu Communication Ei 1t USB Disk 2 USB Calibration 3 Bluetooth 4 Remote management SMART MANAGER SUITE must already be installed once you begin PC communication see the manual for installation and software use USE WITH SMART EXPORT AND SMART MANAGER To share data with SmartExport and SmartManager the DL must simulate an external disk for this click the USB Disk button and make sure that an USB cable connects the DL to the PC The DL will appear as an external USB disk If the SmartMonitor feature is enabled SmartManager will start automatically Please refer to the SmartExport and SmartManager manuals for details Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual USE WITH SMART DATACAL NOTE This function is reserved for SISGEO technicians Rev 2 31 01 2013 Mii Page Archimede Portable Dataloggers User Manual USE WITH SMART REMOTE SmartRemote is a software that replicates the DL screen and keyboard on a PC this may be useful for using the DL with a projector in a class or to provide support to a remote user controlling the PC through Internet using such programs as VNC Click the Remote management button the DL will display the init
14. will be used Press ENTER again to view the same data in tabular form again Rev 2 31 01 2013 Page Archimede Portable Dataloggers User Manual T REX Graph When performing T REX tests there is a distinction between a reference test usually the first test performed whose measures are absolute and other tests in which what matters is usually the difference between the current values and the values recorded in the reference test Graphs for these two types of tests are slightly different ALZA 1 o A 0 6 AG a 4 ales a 2 Send a E UF mn 5 E DOWN ca 5 a A A e Ref RA AA AA The graph for the reference test on the left is labeled ABSOLUTE while the graph for other tests report a label indicating the test being used as reference the reference test or the fact that a test is absolute is selected when starting the test and cannot be changed later In the left absolute graph the blue line is the difference between the distance of reference rings in each step and the ideal value of one meter In the relative graph the offset of the reference test is subtracted before drawing the graph and the line represents the elongation of the tube at each depth from the original state If both an UP and a DOWN runs have been performed they will be drawn as different lines blue for UP and green for DOWN For relative graphs a line is drawn only if the relative data is available both in the current and in the r
15. 8 10 17 54 S242SV30 01 02 08 10 17 56 S242SV30 You may delete tests from this window by pressing the DEL key the DL will ask you to confirm the operation Select the test you wish to browse and click ENTER the DL will display all of the data acquired for one run Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual INCL Hiili To view data for other runs press the PREVIOUS scroll and NEXT scroll keys Inclinometer test graphs When performing Inclinometer tests there is a distinction between a reference test usually the first test performed whose measures are absolute and other tests in which what matters is usually the difference between the current values and the values recorded in the reference test Graphs for these two types of tests are slightly different GRAF BU GRAF BU ABSOLUTE Ref 00 400 6 og 0 200 6 a GIG Y WS NORTH NORTH Oy A pe 30 2 an 270 ages EAST EAST moe 180 180 The graph for the reference test on the left is labeled ABSOLUTE whereas the graph for the other tests displays a label indicating the test being used as reference the reference test or the fact that a test is absolute is selected when starting the test and cannot be changed later In the absolute graph on the left the blue line represents the inclinometer tube as seen from the top the center being the top end of the tube In the relative graph the offset of the reference test is su
16. ES REX451XX T REX removable extensometer 12V 10000mV 5241HH30 Horizontal inclinometer probe 30 12V 5000mV SinA S241SHXX Horizontal probe 12V 5000mV SinA S242HVXX Inclinometer probe 12V 5000mV SinA S242SHXX Horizontal probe 12V 5000mV SinA S242SVXX Servo inclinometer probe 12V 5000mV SinA S30PR12X Spiralmeter probe 2 5V 400mV Degrees S 310PV2X Tilt probe for cased borehole 2 5V 400mV SinA SCLIN45X Tili portable tiltmeter 12V 5000mV Degrees Rev 2 31 01 2013 MiPace Archimede Portable Dataloggers User Manual SPECIAL KEYBOARD COMBINATIONS SWITCH ON To switch the DL on keep the INFO key pressed and a press the ON OFF key till the display switches on ua Eb 12 3 4151 scroll 6 7 8 WXYZ AECIJE SWITCH OFF To switch the DL off keep the ON OFF key pressed arm a and press the ENTER key till the DL switches off e A D 1 KIE ABC 23 4 GHI E JKL la 678 9 EMERGENCY RESET Should the DL become unresponsive press the keys esc del p OFF and DEL keys at the same time the DL should g 2 reset immediately ae ae 1 j 2 ABC DEF 3 9 z JKL teal scroll S gt E 9 ay FIRMWARE UPGRADE To upgrade firmware switch the DL off then press arm p the following keys at the same time 1 3 9 and Ta gt ENTER While keeping them pressed press the leo a gt ON OFF key till the DL switches on 1 3 45 The DL will display a blue screen and will be ready for epa 2 DEF GHI JKL enter
17. H batteries OECAB12VNMB e Flying cable 1 MIL connector and 6 alligator clips ground OECAV7P6A00 e P68 USB cable for PC connection e CDROM Manuals Smart Manager Suite Shipment bag Shoulder bag Flying cable USB cable Battery charger Rev 2 31 01 2013 JiPage Archimede Portable Dataloggers User Manual The Galileo DL has been developed to make and store readings obtained from vibrating wire instruments and the NTC thermistor The Galileo DL is mainly composed by e Shoulder and shipment bag OCDLOBAG927 e Battery charger for NiMH batteries OECAB12VNMB e Flying cable 1 MIL connector and 4 alligator clips ground DECAV7P4A00 e IP68 USB cable for PC connection e CD ROM Manuals Smart Manager Suite Shipment bag Shoulder bag Flying cable USB cable Battery charger Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual The Archimede DL has been developed to make and store readings obtained from inclinometer test both analog and digital inclinometers The Archimede DL is mainly composed by e Shoulder and shipment bag OCDLOBAG927 e Battery charger for NiMH batteries OECAB12VNMB e P68 external button for data storage Archimede DL OCDL300REMC e Flying cable 2 MIL connectors M F 7 pins OECAVO7V200 e P68 USB cable for PC connection e CD ROM Manuals Smart Manager Suite Shipment bag Shoulder bag Flying cable USB cable Battery charger IP68 external but
18. TUNING BACKLIGHT Backlight may be switched on and off by pressing the LIGHT key You may tune backlight level to adapt to field conditions Keep the LIGHT key pressed and press the PREVIOUS scroll key to increase light or the NEXT scroll key to decrease it lese ae g lA Se 12 3 45 epa ABC DEF GHI JKL enter MNO gt TUV WXYZ SHEHE ONLINE HELP In order to display online help press the INFO key JT 2 ABC E DEF 3 4 GHI E JKL E ml scroll G gt 8 9 ON of f NOTE actually no help on line is implemented Rev 2 31 01 2013 Page Archimede Portable Dataloggers User Manual THE MAIN MENU The main menu gives access to all the functions of the DL Read inclinometer Show readings Configuration Communication Utility Manual reading Rev 2 31 01 2013 Main menu 1 Manual read 2 Batch read 3 Configuration 4 Communication 5 Utility del In this section you may select an inclinometer tube and start a new reading session on it In this section you may browse readings performed on tubes In this section you may create the sites configure inclinometer tubes and probes used to perform readings on them In this section you may connect to a PC using the USB port some DLs support a Bluetooth module which is also managed in this section In this section you may set date and time select the language for the user interface and run a checkup for the DL In this section you may
19. ation just press the ESC key on the keyboard As an alternative most buttons contain a number at the left of their labels just click the matching number to trigger the button action When no button is selected the ENTER key is normally used to confirm the most obvious operation for example when a list has focus clicking the ENTER key will usually bring you to the next screen with the currently selected item LISTS D232TXXX al D313FVXX ELI420MA L2CTOXXX S412HV 10 REX451XX TMPTIO Lists are used to select one of several items when a list has focus you may change the selected item clicking the UP or DOWN keys The LEFT key will move to the beginning of the list and the RIGHT key will move to the end By clicking ENTER the current item is selected Only in configuration windows when a list box is displayed items defined by the user for example sites or instruments it is usually possible to delete the selected item pressing the DEL key when the desired item is selected and when you are in configuration menus The DL will require a confirmation to avoid accidental deleting it doesn t happen if the item has no data inside Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual RADIO BUTTONS Radio buttons are used to select one of a few possible options To select an item move focus to it clicking NEXT scroll or PREVIOUS scroll and press ENTER key OlStandard O Wire O Biaxial ____
20. btracted before drawing the graph and the line represents the movement of the tube from the original path as seen from the bottom the center being the bottom end of the tube Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual Push ENTER again to view a different representation of the same data This time the line represents the displacement of the tube from the vertical line for the first absolute graph and from the reference measurement for the second graph Again the reference zero is the top of the tube for the first graph and the bottom for the second GRAF WMA GRAF BU SHG El 400 d l SHG H 205 El iag El O ri E M 0 oO M Ed E ppop J a ap E E Ol al El a8 10 0 11 p iz i 13 0 14 0 15 0 10 0 LA iz i 13 0 14 0 15 0 al al A ec ABSOLUTE cia Ref OG eon tino oo sd Please note that only the absolute value of the displacement is represented look at the first graph to view the direction of the displacement Please note that in each graph a single line is represented in case more runs were performed their values are used to obtain a single value represented in the graph In case of relative graphs the common set of runs available both in the reference test and in the current test are used for example if runs A1B1 A2B2 A3B3 A4B4 were performed for the reference test and runs A1B1 A3B3 where performed for the current graph only runs A1B1 A3B3 of both graphs
21. chapter Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual The user interface is composed of a series of screens containing control items similar to those available in Windows Configuration sites SITE NAME 1 SITE NAME 2 ree Screens contain different items such as lists buttons edit boxes and check or radio buttons Only one of the controls on the screen is selected the currently focused control is highlighted in blue and normally indicates the action performed by pressing the ENTER key Configuration sites MO SITE NAME 1 a SITE NAME 3 Active control lt If the selected control is an edit control characters typed on the keyboard will edit its content Configuration sites Text control Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual You may change the focused control by clicking the PREVIOUS scroll and scroll NEXT scroll keys on the left side of the keyboard As all keys they have an auto repeat behavior when held pressed All controls in a page are in circular list so that clicking NEXT scroll when the last control is selected moves the selection to the first control of the screen Similarly clicking PREVIOUS scroll on the first control moves focus to the last control Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual EDIT BOXES An edit box is used to display and edit text when t
22. cted digital probe will be shown a dialog box Error reading instrument Instrument is not connected In this case check the connection cable Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual SPIRALOMETER TESTS When performing a spiralometer test the screen looks as follows TEST NAME PROBE NAME S3OPR120 SPIR DATE AND TIME E 02 08 10 Chk 18 30 50 MEASURE UNIT E 7 A 4 CURRENT GROOVE CHECK SUM VALUES CURRENT LEVEL The top blue box displays the Check sum value it is the halved sum of the current run and the matching run up for down and down for up at the current level the readings in the matching runs should be exactly opposite so that the check value should be close to zero this value is only available when the matching run has already been performed On the right of this box the current Direction is shown The bottom blue box shows the Current level in the tube At the center the current value is shown Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual T REX TEST When performing a TREX test the screen looks as follows TEST NAME PROBE NAME REX45100S TR ATT DATE AND TIME 02 08 10 mv A 0 Dir 171758 mV B O Up CURRENT GROOVE MEASURE UNIT CURRENT VALUES IN mV REFERENCE VALUES BAF Level CURRENT LEVEL 4 0 The top blue box displays the Current values read in mV on
23. e oe Archimede Datalogger Archimede Datalogger gt sc del E 11 23 4812 e7eso sisoea ad User Manual Archimede Portable Dataloggers User Manual INDEX NOTE ON SYSTEM UTILIZATION WASTE DISPOSAL SYMBOLOGY IDENTIFICATION PLATE DL OVERVIEW INTRODUCTION LEONARDO OCDL200UNO0 new LEONARDO 0CDL400N000 GALILEO OCDL100VWO0 ARCHIMEDE OCDL300INCL ACCESSORIES and SPARE PARTS DL STORAGE AND TRANSPORT BATTERIES Battery charger Internal temperature How to use the battery charge Charge cycle and LED indications THE USER INTERFACE Edit boxes Buttons Lists Radio buttons Tuning backlight On line help THE MAIN MANU CONFIGURING SITES CONFIGURING PROBES PROBE CALIBRATION ONLY FOR ANALOG PROBE POLYNOMIAL CALIBRATION ONLY FOR ANALOG PROBE CONFIGURING INCLINOMETER TUBES READING INCLINOMETER TUBES INCLINOMETER TESTS SPIRALOMETER TESTS Rev 2 31 01 2013 Pag 4 Pag 5 Pag 5 Pag 6 Pag 7 Pag 7 Pag 8 Pag 9 Pag 10 Pag 11 Pag 12 Pag 12 Pag 13 Pag 13 Pag 14 Pag 14 Pag 14 Pag 16 Pag 20 Pag 21 Pag 21 Pag 22 Pag 22 Pag 22 Pag 23 Pag 24 Pag 26 Pag 29 Pag 30 Pag 31 Pag 33 Pag 36 Pag 39 W iPage Archimede Portable Dataloggers User Manual T REX TEST Pag 40 BROWSING INCLINOMETER READINGS Pag 41 Inclinometric Test Graph Pag 42 T Rex Graph Pag 44 MANUAL READING Pag 45 USING THE USB INTERFACE Pag 47 USE WITH SMART EXPORT and SMART MANAGER Pag 47 USE WITH SMART
24. e DL manual Reconnect sensors in a correct way verifiying connection scheme Perform automatic diagnostic directly on DL Utilities gt Unit Status menu if one or more red check marks pop up contact SISGEO assistance Verify the instrument conncections and click OK Turn off and turn on the DL and reset Check that USB cable is correctly inserted both in the PC and in DL port DL showed Calibration file is not valid The calibration file is lost or damaged Call SISGEO for assistance DL showed Instruments is not compatible DL showed Instruments is not connected The selected instrument is not compatible with the instrument connected to the DL The energy saving function is enabled The connections to the instruments are incorrect Select the correct instrument and repeat the reading Disable the energy saving function only for Sisgeo digital inclinometer or wait for the time necessary for the awakening of the sensors and eventually check the configuration only for Sisgeo digital inclinometer chains Reconnect the instrument properly and repeat the reading Archimede Portable Dataloggers User Manual ASSISTANCE Sisgeo S r l is aware of the importance covered by the after sales assistant service therefore Sisgeo Technical Support is structured to offer the following services REPAIRS AND MAINTENANCES During the warranty period we assure that the equipment supplied is free from defects in
25. eference test Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual MANUAL READING You may read probes without starting a new test in order to check that they are functional select Manual Reading in the Main menu then select the probe you wish to test the DL will show the values being read Reading instrument S2425V30 Sin During the reading of a digital probe pressing the RIGHT key it is possible to display the temperature read from the probe S242HDXX C Attention this function is possible only if it used a digital probe Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual If the DL finds an incompatibility with the digital instrument i e has been selected a mono axial digital probe while the connected probe is bi axial will be shown a dialog box Error reading instrument Instrument is not compatible ox If the DL can t read the connected digital probe will be shown a dialog box Error reading instrument Instrument is not connected In this case check the connection cable Rev 2 31 01 2013 Mj Pace Archimede Portable Dataloggers User Manual USING THE USB INTERFACE The DL features an mini USB interface connector for connecting to a PC This same connector is used for an external push button used for inclinometer test Archimede DL only These two uses are not compatible when an external push button is connected
26. evisions Rev 2 31 01 2013 MiPage Archimede Portable Dataloggers User Manual WASTE DISPOSAL According to the European standard 2002 96 CE the disused instrument must be disposed in a correct way The recyclable materials are recovered in order to avoid the environmental damage For more information please contact your waste disposal service or the A instrument retailer SYMBOLOGY Here are the symbols used in this manual to alert the reader Attention This operation must be executed only by specialized personnel W Pay particular attention to the following information Rev 2 31 01 2013 Page Archimede Portable Dataloggers User Manual IDENTIFICATION PLATE Each portable datalogger have two types of identification plates One of these reports the product registration Serial Number and is on the back of DL SISGEO llano italy Number gt gt s Y r E 5 i f ya PI ge MA A i SN A r y d x y y w k hd MAA NAO NE It is important to communicate this information when requesting information concerning the product The second plate is always located on the back of the DL Follow the requirements on the plates Failure to comply with these requirements could result in damage to the device and danger to the operator Rev 2 31 01 2013 iPage Archimede Portable Dataloggers User Manual DL OVERVIEW Display TFT DL model Archimede Datalogger Keyboard
27. ext to be entered is a number either an integer or a number with decimals the numeric keys and the decimal point are used to directly type the digits and decimals separator The key is used to enter a or sign while the e for the exponent part Just press the key the DL will show a list of characters Press the key repeatedly till the character you wish to insert is selected then wait till the list of characters disappears and the selected character will be inserted in the edit box When you need to enter text all the numeric keys and the and the keys are valid press any of these keys repeatedly to enter the desired character Press the DEL key to delete the character at the left of the cursor If you keep the DEL key pressed for about a second the content of the edit box is deleted quick delete The characters available for each key are E o 1 epa ABC2abc DEF3def E GHI4ghi JKL5jkl 6 MNO6mno PORS7pqrs E TUVStuyv E WXYZ9wx yz 0 0 _ space Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual BUTTONS Buttons are used to select an action which often means navigating to a new screen Just move focus to the desired button and click the ENTER key 2 ft Manual read 2 Batch read Generally no BACK or ABORT button is on the screen to return to the previous screen or abort the current oper
28. f they are repeatedly and partially discharged In order to prevent this it is recommended to use exclusively the battery charger supplied with the DL and to cycle charge and fully discharge the batteries at least once every two three weeks The batteries should be charged only when they are completely discharged and not before charge it before only when you have to use the DL Furthermore if the DL is not going to be used for an extended period it is preferred to store the DL with the batteries fully charged Well charged batteries provide at least 8 hours of continuous use this means that the batteries provide charge for more than a day s work If the DL is not used for months the batteries lose their capacity due to the auto discharge The first time a battery charger is connected this could signal green also if the batteries are not fully charged it is advised to repeat the charge cycle twice Using a good battery a full battery charge allows a non stop use of at least 8 hours this means that with a full charge it is possible to work on site for more than a day BATTERY CHARGER The battery charger furnished is a microprocessor controlled with rapid charge in less than 3 hours The input voltage range is universal 90 264 V This battery charger utilises a method called dv detection this method is based on the fact that the voltage drops over the NiMH cells when the batteries are fully charged this voltage drop
29. factor O Probe calibration ET Ch A Ch B 9 939e03 9923803 001414 001522 _ 0 01691 0 01609 006414 0 05322 10000 Save The parameter K should be consistent with the previously selected unit If for example it has been selected 20kSen a as the unit of measure it should be inserted 20000 as a parameter K These parameters are used to convert from the electrical measurement to the physical value Sin a The conversion formulas are reading sin arcsin rawDatag Sa Ea Ma readings sin arcsin rawDatag Sg Eg Ms final reading reading readings Xa final readings readings readings Xg where final reading accounts for the cross axis effect NOTE usually the probe s calibration parameters are inserted by SISGEO and the probe is supplied with the DL Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual POLYNOMIAL CALIBRATION ONLY FOR ANALOG PROBE Polynomial calibration is normally used by T REX spiralometers and monoaxial inclinometers Instrument calibration Type NEEDS Ch A Ch B A B a 1 Oe Save The list of available polynomials degrees is listed at the top of the screen Select the polynomial you wish to use with the UP and DOWN keys it is possible to select one of the following families of polynomials e Ax B e AX Bx C e AX BxX Cx D Then click the NEXT button to edit the coefficients for the
30. fter a few seconds about 5 sec the DL will switch on and the display will show a start up page This page will display the type of DL and the version of the firmware At the bottom of the display the name of the DL will show up this is an user assigned name that can be used to distinguish between different DLs Keeping the name of different DLs distinct is important when importing data to a PC so that the source of the data can be traced DLt uii Archimede ver 0 3 0 Battery Status Firmware version 20120000 Unit Name At the upper right corner the display shows the current status of the battery the remaining charge is shown by the number of bars When the battery voltage drops below a certain level the bars will turn red At about 10 5V the DL will stop all activities then it will beep and switch off Rev 2 31 01 2013 Page Archimede Portable Dataloggers User Manual To save battery consumption after three minutes without user activity the display will be switched off after ten minutes the DL will switch off Press ENTER or wait a few seconds to display the Main menu To switch the DL off press the ON OFF key while keeping the ENTER key pressed the DL will display a window on top of the current page with a bar at the bottom When the bar is completely filled the DL will ask to release all keys and then it will switch off Other keys combinations are described in the Special keyboard combinations
31. he next the probe is extracted and rotated 902 Spiralometer A head to bottom run is performed normally a second bottom to head run is performed TREX A bottom to head run is performed first sometimes a second head to bottom run is performed next Although the above sequences are the ones normally used you may choose to perform runs in any order you prefer Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual The DL will suggest you the first run to perform At the end of each run the DL will suggest the next run It is possible to interrupt the test at the end of a run by selecting Save and exit and pressing ENTER A1 B1 A3 B3 Save and exit INCLINOMETER TESTS When performing a run what is displayed in the center of the screen depends on the type of test the upper part of the screen and the left nonetheless contain items that are common to all tests At the top of the screen the Name of the probe being used and the Name of the tube are shown On the left the Current date and time and the Measure unit are shown Four buttons are also present Data Click this button to view acquired data in tabular form Save Click this button to save acquired data and exit Read Click this button to read the value currently displayed and proceed to the next step Back Click this button to revert to last step It is possible to stop the test anytime pushing the ESC key will pop up a confirmation request If y
32. ial solash screen On the PC run the SmartRemote program When it starts click the Connect button at the top of the displayed window on the PC esc del gt 1 2 3 45 The refresh button at the top may be SR used in case the display fails to update Main menu Software s virtual keys in the lower part of the window have the same function 1 read i of the real ones on DL keyboard they 2 Batch read can be pushed with the left mouse 3 Configuration button 5 Utility To press more keys at once for example to obtain the effects described in the Special keyboard combination chapter you may lock a key by clicking on it with the right mouse button Right click it again to unlock it All the time the physical keyboard on the DL is fully functional so you may find more comfortable to use it instead of the simulated keyboard on the PC You will notice that when using Rev 2 31 01 2013 M Page Archimede Portable Dataloggers User Manual SmartRemote the DL is slightly slower than usual this is normal and it is due to the fact that the screen must be transmitted to the PC when it changes The DL is fully functional when being used with SmartRemote except that the USB Disk and USB Calibration functions in the Communication menu are disabled Communication TO t USB Disk 2 USB Calibration 3 Bluetooth 5 Remote management off Also when reading instrume
33. is detected when the voltage has dropped a certain percentage from the highest value and the charge will be automatically completed If this voltage drop does not occur the battery charger has a safety timer which will terminate charging after a given period of time to avoid batteries overcharging For battery cells which have been idle for a long period of time the voltage could drop in the first part of the charge cycle For this reason a start timer is built into the battery charger to prevent dv detection in the first minutes of the charge cycle Don t take instrument readings with battery charger connected to the DL Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual Don t use a different type of battery charger than the one supplied OECAB12VNMB INTERNAL TEMPERATURE If for any reason the internal temperature in the DL is more than 70 C there is a safety system that disconnects the battery until the temperature decrease under 70 C So if after the battery charge the DL won t switch on wait for the battery cooling until it returns within an adequate temperature range HOW TO USE THE BATTERY CHARGER The battery charger is started by connecting the jack to the readout and then connecting the battery charger at the voltage line NOTE To avoid damages always insert the battery charger jack in the DL before connecting the battery charger at voltage line 90 264 VAC 50 60Hz likewise before disc
34. nts it will not be possible to use the external pushbutton Click Remote management off button in the Communication menu to terminate SmartRemote mode Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual USING THE BLUETOOTH INTERFACE The BlueTooth interface allows transferring data to a PC or a SmartPhone equipped with a BlueTooth interface From the main menu select Communication then click Bluetooth the datalogger will beep and in a few seconds the BlueTooth interface will be activated and a page will be displayed for selecting the remote device to which data will be transferred Select BT device MO Check for BT devices This list is initially empty to add one or more remote unit to it make sure that they are visible How a BlueTooth device is made visible depends on the specific device for example when using Windows Seven you should open the BlueTooth settings windows in the lower right of the Windows bar and after selecting Open Settings make sure that the option Allow Bluetooth devices to find this computer is selected Your PC or SmartPhone should now be able to detect the DL The code required for pairing is 574832 Once the DL has been added to the list of devices of the PC or SmartPhone click the Check for BT devices button on the DL the PC or SmartPhone will be added to the list of BlueTooth devices Select BT device TT CMAGPC J wt Check for BT devices
35. of the same characters used for Microsoft Windows valid file names that is all characters except the following lt gt To delete a site select it in the sites list and make sure that the site list has focus then press the DEL key if the site is not empty the DL will ask to confirm the operation Configuration sites TT M ore A Warning Delete site MYSITE SIT Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual CONFIGURING PROBES Probes are configured globally and may be used in any site the user will normally carry his own probes and use them on different tubes in different sites To browse or edit probes select Configuration in the Main menu then select Probes Edit instruments A The list of probes is initially empty later you will be able to edit any of the probes defined in it by selecting it and pressing ENTER to define a new probe click the New button New instrument REX451XX sa S241HH30 S241SHXX S242HV XX S242SHXX S242SVXX S30PR12X Select the probe type in the list according to your probe probes type are listed in Appendix A and press ENTER Rev 2 31 01 2013 Page Archimede Portable Dataloggers User Manual The DL will show configuration data for your probe Instrument identification ML Name 242HVXX Notes Code S242HVXX Unit Sna Decimals 0 These data include Name Notes
36. onnecting the battery charger jack from the readout disconnect the battery charger from the voltage line The LED on the battery charger will be yellow before the fast charge starts and then the LED changes to orange When the batteries are fully charged the LED will be green The trickle charge does not exceed 24 hours If the battery voltage is less than the normal one the LED will then indicate error by flickering green and orange light CHARGE CYCLE AND LED INDICATIONS ED MODE Battery not connected Battery initialisation and analysis Fast charge Gren Tickle charge O Connecting the battery charger at the voltage the LED will become orange for the first 5 7 seconds then it turns to yellow when the initialisation and the analysis starts If the battery is connected the actual charging will start a few seconds later when the LED changes to orange When the battery is charged the LED will be green in approx 4 seconds This is only a signal to test Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual the service When dv has been detected the start of the top off charge is indicated with a green LED with intermittent yellow flashes The LED will be green during trickle charge Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual USER INTERFACE To switch the DL on press the ON OFF key while keeping the INFO key pressed Sehr delle gt ME HS scroll 7 A
37. ou confirm the test will be deleted instead keeping the ESC key pusher for a few seconds the data will be saved and the test will be incomplete You may also abort the test at any time by pressing the ESC key in this case no data will be saved During inclinometer tests you may trigger a read by using an external pushbutton OCDL300REMC this has the same function as the Read button To prevent the user from inadvertently triggering multiple readings the external button is disabled for two seconds after each push We advise to wait at least three seconds between each pushing Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual You may also view data in graphical form for inclinometer and T REX tests just press ENTER when the table is displayed When the last read for the current run is performed the DL will display again the list of runs with the next run preselected When performing an inclinometer test the screen looks as follows PROBE NAME S242SV30_ a INCL TEST NAME DATE AND TIME 02 08 10 Chk A F oe CURRENT GROOVE 17 56 34 Chk B AWVB1 MEASURE UNIT Sna CHECK SUM VALUES roe At 4 EA e B 57 IZ A 4 CURRENT LEVEL Ref B 57 REFERENCE VALUES The top blue box displays the Check Sum values for both channels these are the halved sums of the current run and the matching run A1B1 versus A3B3 and A2B2 versus A4B4 at the current level that is
38. perature displayed on the AD converter line is the measured inside the DL at the AD converter chip so it will normally be a little higher than the room temperature CHANGING USER INTERFACE LANGUAGE Click the Set language button in the Utilities menu to change the language used by the DL Set language Ci Italan The DL will display a list of available languages select your language and press ENTER The DL will revert to the previous screen with the newly selected language Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual Utilita 1 Imposta data e ora 2 Stato unita 3 Imposta lingua If you find a DL set in language that you cannot understand from any screen of the user interface press ESC for a few seconds till the DL reverts to the main menu then press the 5 button followed by the 3 button then select your language Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual TROUBLE SHOOTING DL does not switch on The battery is not correctly recharged Battery recharger LED does not perform the foreseen colorur sequence as in DL manual DL display has no back light DL crashes DL reads incoherent values DL showed Check connections before continuing DL does not update firmware It hasn t been done the correct startup keys combination A connector from the DL internal battery disconnected Battery s temperature overcomes 70 C internal security
39. polynomial When you have finished entering parameters click the Save button the newly created instrument will be displayed in the instruments list for the site NOTE Normally the calibration parameters of the probe is set directly from SISGEO if the probe is supplied with the DL Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual CONFIGURING INCLINOMETER TUBES Tubes are arranged per site to browse or edit tubes select Configuration in the Main menu then select Tubes Tubes as defined in the DL include parameters describing how the physical tube should be read SO you may need to define more tubes on the DL for the same physical tube For example you may wish to define a spiralometer and an inclinometer tube for the same physical tube naming them for example MYTUBE SP and MYTUBE First the DL will display a screen to select the site in which the inclinometer tube is or should be defined after a site is selected it will display a list of tubes defined in the site Select tube iil A When one or more tubes are listed it is possible to delete one of them by selecting it in the list and pressing the DEL key the DL will ask the user to confirm this operation It is also possible to make a copy of an existing tube by selecting it in the list and clicking the Duplicate button the DL will ask the name for the new tube The list is initially empty so that the only possibility is to click the
40. sible to delete the partial test and repeat it in future Click Read inclinometer in the Main menu then select the site in which the inclinometer tube is located the DL will show a list of the tubes defined in the site Select the tube you wish to use and press ENTER to proceed Select test number Tube Number TR bo The DL will ask you to enter the number for the test prompting you with the first available number Enter a new number or simply press ENTER to proceed you may perform up to 99 test on the same tube The DL will ask next for the instrument to use REX45100S 2 S242SV30 S30PR12X Rev 2 31 01 2013 RR Page Archimede Portable Dataloggers User Manual Select the instrument and press ENTER Then the DL will ask you to specify some options on how the test should be performed This screen is not displayed for T REX tests because none of the options that can be selected in it makes sense for such tests You should first select if the readings should be triggered manually or automatically after a delay Delay Direction 30 seconds 1 minute 2 minutes 5 minutes lt We suggest to select the default Delay option that is Manual Then you should select if the runs should be performed from head to end or from end to head Delay Manual v Direction Hi ca Te Y From end to head From head to end This question is not asked for spiralometer
41. tests because in these tests you have to perform a run From head to end run and then perform a run From end to head Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual Next the DL will ask you for the test to be used as Reference this will normally be the first test in the list labeled 00 but you may choose any test in the list or click the No reference button to perform an absolute test Reference test 02 08 10 17 54 S242SV30 Y No reference The values done when performing the test later will actually be the difference between the values actually read and the values read for the same run and the same step during the reference test which will be used a zero reference The above mentioned page will not be shown in the Spiralometer test because it is not expected the reference test choice The reference test is not used for Spiralometer tests so in this case the above window will not be displayed At this point all data has been collected and the DL will ask which run should be performed next the number and type of runs depends on the inclinometer test type A1B1 A3B3 An A1B1 run is performed then the probe is extracted and rotated by 1802 and an A3B3 run is performed A2B2 A4B4 An A2B2 run is performed then the probe is extracted and rotated by 1802 and an A4B4 run is performed A1B1 A2B2 A3B3 A4B4 Runs A1B1 A2B2 A3B3 and A4B4 are performed in sequence between each run and t
42. to 6000Hz Measuring resolution 0 1 Hz with F S from 400 to 6000Hz Temperature drift 0 001 F S C gt 10MO for voltage lt 2 5V mput impedance gt 1M0 for voltage gt 2 5V Storage memory 2GB MH iPace Rev 2 31 01 2013 Portable Dataloggers User Manual Operating time Battery charger Display Rev 2 31 01 2013 IPage e ia Portable Dataloggers User Manual SISG EO 4 New Leonardo Leonardo Archimede Galileo Rev 2 31 01 2013 IPage Se Portable Dataloggers User Manual SISG EO Y New Leonardo Leonardo Archimede Galileo Weight Lx Wx H Rev 2 31 01 2013 iPage
43. ton Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual ACCESSORIES AND SPARE PARTS Flying cables e 1MlLconnector and 2 alligator clips ground OECAV7P2A00 e 1MlLconnector and 4 alligator clips ground OECAV7P4A00 e 1MlLconnector and 6 alligator clips ground OECAV7P6A00 e 2 MlLconnectors M F 7 pins OECAVO7V200 e P68 external button for data storage Archimede DL only OCDL30OREMC e Multiplexer measurement OECAV1OMUXO External 12V battery pack OCDLO12EXBP Internal Bluetooth module OCDLOBTOOTH this module shall be requested together with DL purchase order can be added later only by SISGEO Shoulder and shipment bag OCDLOBAG927 Battery charger for NiMH batteries OECAB12VNMB External 12V battery pack Multiplexer measurement cable DL STORAGE AND TRANSPORT To store correctly the DL it is advised to use the supplied bag Put carefully the DL in the suitable compartment with the display turned inside During the transport it is advised to use the shoulder bag and to avoid bumps Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual BATTERIES Each DL is powered by an internal rechargeable NiMh battery If properly used the NiMH batteries are rated for at least 500 discharge cycles An effect known as the memory effect which will degrade the readout s performance during use may occur if the batteries are left under charge for some days or i
44. uld press and hold a few seconds it will be open a menu selecting Sharing you can send the file as an email attachment SYMBIAN S60 3 EDITION There are some precautions to take with this Operative System OS symbian OS e The user should use an application File Manager that allows to save the file SGX because the OS tries to interpret the content and not allow saving the file The file manager is available for free in Internet SISGEO suggest a free application called Active File ver 1 44 e After receiving the message BlueTooth you have to open the File Manager through it you have to copy and paste the file in the desired folder e By creating a new mail you can attach the file saved as generic RIM ver 5 0 0 822 E e TPL There are some precautions to take with this Operative System OS FT 22 BlackBerry e To receive the file is necessary to entry in the Multimedia folder and thought the menu button you should select Receive using Bluetooth item The smartphone will be available to receive file Now you can send the file from the datalogger Rev 2 31 01 2013 IPage Archimede Portable Dataloggers User Manual WINDOWS MOBILE ver 5 amp 6 ay Windows Mobile There are no special precautions to take with this Operative System OS The file are managed as under Microsoft Windows for PC Rev 2 31 01 2013 Page Archimede Portable Dataloggers User Manual
45. uration needed so the user must carefully perform all the operations described herein SISGEO will not be liable for inconveniences damages accidents etc due to lack of knowledge or non application of the requirements contained in this manual The use any possible maintenance or repair of the equipment is allowed only to skilled and authorized operators These operators must be physically professionally and intellectually capable For information or orders of spare parts always specify the Serial Number on the label on the back of the portable datalogger reported on the identification plate showing the serial number inside each device When replacing parts use only ORIGINAL SPARE PARTS The use of unoriginal SISGEO spare parts may cause irregular functioning and dangers for people and objects The product is not fit for installation in small and or unventilated environments with a high humidity range or in presence of inflammable or explosive liquids All the instruments and the equipment furnished are not suitable for installation in potential dangerous area or when the use of explosion proof components is foreseen The electrical connections must be executed by a qualified installer observing the effective regulations The manufacturer reserves the right to make changes in case of any manufacturing or commercial need without notice It will attempt to ensure that user manuals are updated in order to reflect functional product r