5781 A4 cover template in-house prod manuals
Contents
1. Setup Test Sequences Test Codes Asset Trace Variables System Config Bluetooth Favourites User Admin gt Preferences Chance User Change Password User Profile Memory Options Restore Factory Settings About 67 Page Rigel Medical BP SIM user Manual V3 1 Appendix B Setting up printers 68 Page Rigel Medical BP SIM user Manual V3 1
2. This will display the Test Details Screen 50 Page Rigel Medical BP SIM user Manual V3 1 Asset ID 123455 Tested by Tested on Mode Status Pass Test Period months Use the button F2 to return to the main View Data screen Use the gt button F3 to view the Trace Variables Asset Trace Variables Service Code Site Location Make Model Description Serial Number Client Use the K button F2 to return to the main View Data screen In the Test Details screen press the button F4 to view the following options Asset ID 123456 Tested by Tested on 08 Oct 2008 12 05 PM View Results View Sequence User Comment Print Result Print Label View Results will display the results of the test see section 6 1 View Sequence will display the test Sequence User Comment will allow the user to enter comments if required 51 Page Rigel Medical BP SIM user Manual V3 1 Print Result will print the result on the thermal printer 50 mm roll see Appendix B Print Label will print a pass fail label on the thermal printer BO mm roll See Appendix B 6 1 View Results i From the Test Details screen select button F4 Asset ID 123455 Tested by Tested on 08 Oct 2008 1205 PM View Results View Sequence User Comment Print Result Print Label Use the up and down keys to select the view results option Press the button F4 to display the view text result
3. Press the Date Time button F2 The following menu will be displayed 24 Page Rigel Medical BP SIM user Manual V3 1 Day Enter current day Month Use the up down left and right keys to select the current month Year Enter current year Time Enter current time Use the F1 to separate hours from minutes to ensure the correct time is entered and saved Confirm the settings by pressing the v button F4 Ones all System Configurations have been completed press the v button F4 Changes are saved automatically 3 4 Blue Tooth Favourites In order for the Rigel BP SIM to automatically connect to the correct Bluetooth accessory details of that accessory must be entered into the Bluetooth Favourites list The Bluetooth Favourites list comprises of three sub lists device types each of which can store up to 3 accessories of the same type The three sub lists are shown below e Barcode Scanner e Printer e Computer To select Bluetooth Favourites press the button F4 followed by Setup in the Menu and Bluetooth Favourites from the list as shown below Press V F4 to select 25 Page Rigel Medical BP SIM user Manual V3 1 Heart Rate 80 bpm Default Systolil Test Sequences Asset Trace Variables System Config View O SIE Gi ON EVI UE Data T User Admin gt Siem Memory Options Restore Factory Settings The following shows how to add a Computer to Bluetooth Favourites Bluetooth Favourites
4. T logo and press the v button F4 Heart Rate 80 bpm Default Systolic 1 20 mmHg a View D Download to PC BEIC Upload from PC Setup 60 Page Rigel Medical BP SIM user Manual V3 1 The Rigel BP SIM will now attempt to connect to the PC setup in Bluetooth Favourites When a connection has been established the display will be as shown below The Rigel BP SIM is now ready to receive data The amount of data received is shown numerically on the top left of the display 0 in the screen shot above When using the BlueTooth Downloader application select Send File from the File menu in the task bar This will open a window allowing you to select the Bitmap file Select the required Bitmap and click on Open The Bluetooth Downloader will now transfer the file to the Rigel BP SIM When the transfer is complete the Rigel BP SIM screen will display a message as shown below V4 Press the OK button F4 to save the logo The logo is now available for printing using the Seaward Test nTag printer Note The file transfer can be made using any suitable PC application such as the Seaward Bluetooth Downloader data transfer exe or Hyperterminal 61 Page Rigel Medical BP SIM user Manual V3 1 8 About This screen provides details on the Rigel BP SIM s Firmware version Hardware version Calibration status Last Calibration Date Serial Number Blue Tooth ID Select ABOUT from the menu below ek 80 bpm Defaul
5. 10 Technical Specifications Simulation Waveform Envelopes O curves Simulation via Pulse Volume Heart rate Integrated Pump Leak test Chronometer Overpressure Digital Manometer Pressure Accuracy Pressure Units General Operation Battery charger Battery life Memory Capacity Communication Display Keypad Weight Size Lx W x D Operating conditions Storage environment Environmental Protection 66 Page Oscillometric Tbc updates available through www rigelmedical com T connection into Cuff and monitor High Medium Low paediatric 1 300 bpm 0 to 350 mmHg user configurable user configurable between 0 350mmHg configurable upto 999 secs automatic max 410 mmHg 0 410 mmHg 0 5 FS mmHg inHg kg cm cmH20 mBar PSI inH2O and kPa Battery cell insitu charge 110 230VAC 50 60 Hz standby appr 8 hours or 100 NIBP simulations Appr 10 000 records Via Blue Tooth Monochorme 1 4 VGA full graphics Alpha numeric 1 kg 270 x 110 x 75 mm 10 5 x 4 x 3 10 30 C 0 90 RH NC 15 60 C IP 40 Rigel Medical BP SIM user Manual V3 1 Appendix A Firmware Route Map Manual Mode NIBP Test Settings NIBP Leak Test NIBP Pop off Test NIBP Manometer Mode Set Zero Auto Mode View Data Data Transfer Download to PC Upload fro PC Configuration Data Load T nT Logo
6. Barcode lt gt Printer Computer OS Use the up down arrow keys to highlight the device type in this case Computer then i press C F3 Make sure that the device you want to add to your new favourites is switched on Press a F1 The Rigel BP SIM will search for all discoverable Bluetooth devices within a 10m range The time taken to perform the search will vary depending on the number of Bluetooth devices in range A progress bar will show during the search When the search is completed the progress bar will disappear Setu Bluetooth ID PIN number Use the arrow keys to highlight the field labelled New and press the left or right keys to open the list of devices which have been found during the search 26 Page Rigel Medical BP SIM user Manual V3 1 Note f more than 10 Bluetooth devices have been found the list will show the last 10 devices found In the following example the search has identified Bluetooth devices named SELLTO23 SELPC045 and Elite 36 5011 Setup m Bluetooth ID Elite 36 5011 Nev SELLTO23 eT Te Note f the required device is not shown check that it is powered on with the Bluetooth function active then repeat the search Note Some Bluetooth devices do not report a recognisable name and may cause confusion when setting up Favourites Turn off or disable unused Bluetooth devices when configuring Favourites Use the up down keys t
7. Once a variable has been activated the user is given the ability to enter Trace Variables prior to a safety test by selecting from or adding to the list of default items 22 Page Rigel Medical BP SIM user Manual V3 1 Each variable entered during the automatic test will automatically be added to the list for future use Trace Variables that are de activated set to NO will not appear during the automatic tests Note that a maximum of 40 individual entries are available for each Trace Variable Additional entries will overwrite on a first in first out basis 3 3 System Config This feature can setup the Rigel BP SIM to automatically perform certain features during or after a test as well as allowing the user to set Time and Date preference To select System Config press the button F4 followed by Setup in the menu and System Config from the list as shown below Heart Rate Default Systolil Test Sequences Jg Memory Options Restore Factory Settings The System Config menu is displayed below System Config Asset ID Blank After TesiNew Test o gt Comments Always gt Languege Engieh fe Date MAESA The System Config menu provides the following features 3 3 1 Asset ID Provides automatic Asset ID configuration Use the Left amp Right Keys to select between 23 Page Rigel Medical BP SIM user Manual V3 1 Increment Automatically increments the next Asset ID s least significant number by 1 Bla
8. RIGEL MEDICAL lt _ RIGEL BP SIM NIBP simulator Copyright 2015 SEAWARD GROUP Instruction Manual d Last Update 7th August 2015 370A550 Revision 3 1 Rigel Medical 24 Month Warranty Statement Rigel Medical provides a standard 12 month manufacturer s warranty against breakdown during normal use This warranty can be upgraded to a 24 month warranty terms and conditions apply Problems caused through misuse damage fair wear amp tear consumables and accessories are excluded from standard warranty Such components found to be being used in excess of their manufacturer s operating recommendations are also excluded Shipping to an authorised service centre is the responsibility of the sender Terms and Conditions of 24 Month Warranty The Rigel product must be registered with Rigel Medical within 30 days of purchase to be eligible for the extended 24 month warranty Instruments must be returned to an authorised service centre complete with proof of purchase within 13 months of purchase for calibration at the current rate Any items returned for calibration outside of the 13 month period stated above may not be eligible for the second 12 month section of warranty The second 12 month section of the warranty begins at the expiry of the initial 12 month period not when the unit is calibrated Details correct at time of going to print The manufacturer retains the right to make amendments to the above terms and conditi
9. a number of After Test options are possible New Test Automatically brings up the next test screen No further options or viewing of results available Download Automatically down loads the test results to the PC see 7 1 No further options or viewing of results available The tester will return to the home screen Print Label Automatically prints the test results to a thermal printer No further options or viewing of results available The tester will return to the home screen Test n Tag Automatically prints a Test n Tag label see 4 2 No further options or viewing of results available The tester will return to the home screen Options Menu Provides a menu to allow further choices after a test see 4 3 4 2 Test n Tag The Rigel BP SIM is compatible with the Seaward Test n Tag and Test n tag Elite system To use the Test n Tag printer add your T nT printer to the BlueTooth Favorites See 3 4 A detailed application note is available from the Rigel website on the used of Test n Tag printers with the Rigel BP SIM Please visit http www rigelmedical com knowledgebase 40 Page Rigel Medical BP SIM user Manual V3 1 4 3 Options Menu When Options has been selected in the Setup System Config menu the Rigel BP SIM will show a Test Details screen see below Asset ID 123455 Tested by Tested on Made Status Test Period months The test screen provides the following featur
10. 120 mmHg Diastolic NIBP Test Settings NIBP Leak Test MAP NIBP Pop off Test NIBP Manometer Mode Pulse volume Set ZERO NIBP Test Settings see 5 1 NIBP Leak Test See 5 2 NIBP Pop off Test see 5 3 NIBP Manometer Mode see 5 4 Set Zero see 5 5 Rigel Medical BP SIM user Manual V3 1 5 1 NIBP Test Settings Use the NIBP Test Settings to setup a selection of Dynamic Pressure simulations Manufacturer Default lt gt Adult Infant Systolic Diastolic Heart Rate D Pulse volume Medium lt gt Navigate using the Up amp Down Keys and change each setting by using the left key Manufacturer Select between a default O curve or one specific to a manufacturer for example Manufacturer Adult Infant Systolic Diastolic Heart Rate Pulse volume Medium DeIe 9Ie Additional manufactures will be developed and can be uploaded through new firmware revisions from our website http www rigelmedical com downloads as Adult Infant Toggle between Adult or Infant settings Default settings such as heart rate and pulse volume will be adjusted automatically Systolic Diastolic Select a preset range of NIBP setting from the drop down list 44 Page Rigel Medical BP SIM user Manual V3 1 Manufacturer Adult Infant Systolic Diastolic 180 140 200 150 Heart Rate 220 180 Pulse volume Heart Rate Enter the required simulated heart rate Pulse Volume Select the required pulse
11. Blue Tooth Favourites Setup your Blue Tooth devices see 3 4 User admin Setup users and preferences see 3 5 Memory Options Manage the tester s memory see 3 6 Restore Factory Settings Defaults the tester to factory settings see 3 7 3 1 Test Sequences The Rigel BP SIM can be set up to create new test sequences to meet monitor specific requirements or to modify existing test sequences to meet personal preference Preset test sequences cannot be changed by the user however alterations are possible by producing a copy of the default test sequence setting The total number of possible preset test sequences is 50 including 12 default test sequences 11 Page Rigel Medical BP SIM user Manual V3 1 3 1 1 User definable tests This feature gives the Rigel BP SIM the capability of recording user defined visual inspections checks or tests including measurements from SPO2 ECG NIBP Defib Infusion Ventilation Pressure etc The input is text only as no measurements are performed by the Rigel BP SIM during these tests The user can enter questions or instructions followed by either a PASS FAIL result or alpha numeric input Preset engineering units e g Joules mmHg PSI CmH20 etc are available 3 1 2 View Delete or Copy an existing Test Routine Preset test sequences cannot be deleted or changed by the user however alterations are possible by producing a copy of the default test sequence To enter the Test Sequences
12. Run Mode Semi Automatic lt gt Test Period months B Select the right criteria for the Test from the Asset Details screen Enter the Asset ID number using the keypad or barcode scanner Enter the Test Code if this has been created see 3 2 Use the up down left and right keys to select the required Test Sequence Enter the required test period in months BoON Asset Details Admin Default Adult lt gt Semi Automatic lt gt User Name Asset ID Test Sequence Run Made 5 Returning to the Asset Details screen the O button F2 provides the user with the option to modify or create a new Test Sequence from this menu This feature is identical to that of the Test Sequence feature in the Setup Test Sequence menu See 3 1 for more information V4 When all parameters are set press the button 4 to enter the Asset Trace Variables screen when applicable In this example all Asset Variables are activated 37 Page Rigel Medical BP SIM user Manual V3 1 Service Code Site Peterlee Location Building BAG Make DG Medical Model JB4000 Description Patient Monitor Serial Number XYZ1234557890 Client North East NHS 6 Use the Up amp Down Keys to navigate the menu and the Left amp Right Keys to open the drop down boxes Select the required input and confirm with the button F4 Each variable entered during the automatic test will automatically be added
13. The file transfer can be made using any suitable PC application such as the Seaward Bluetooth Downloader Data Transfer exe or Hyperterminal capable of sending and receiving a text file Note The maximum number of entries for each field is SiteName 40 LocationName 50 UserName 40 and Comment 80 7 3 1 Receive Config Data Follow the steps above in Send Config Data to run the Bluetooth Downloader and connect the Rigel BP SIM to your PC Click on File on the Bluetooth Downloader menu and select Send File This will open a window allowing you to select the text file containing your Trace Variables User Profiles 59 Page Rigel Medical BP SIM user Manual V3 1 and Comments configuration data Select the require file and click on Open The Bluetooth Downloader will now transfer the configuration file to the Rigel BP SIM When the transfer is complete the Rigel BP SIM screen will display a message as shown below Valid Data Received If the transfer is not successful check that the format of the configuration file exactly matches that specified Note The Config file must be ASCII text and the last entry must be the END statement 7 4 Load TnT Logo A bitmap logo can be transferred to the Rigel BP SIM for use with the Seaward Test n Tag print system The logo must be in BMP format with a maximum size of 320 pixels by 240 pixels Press button F4 select Data Transfer using the up down keys select load T n
14. inserting using the up and down keys and press the Insert button F1 from the menu below Note Tests will be inserted before the highlighted position not after NIBP Leak Test Set Level 275 mmHg Duration 180s NIBP Pop off Test Once the Insert button F1 has been pressed a drop down box will appear with all available tests and inspection available Activate the drop down box using the left key and select the Custom Test option Visual Test NIBP Leak Test NIBP Settings o curve NIBP Simulation Custom Test Confirm using the v button F4 The following menu will define the nature of the test eg test Instruction prior to leak test 18 Page Rigel Medical BP SIM user Manual V3 1 Use the drop down box to choose a preset instruction and engineering unit or simply type in the boxes provided Newly entered data will be added to the drop down box ona first come first serve basis Edit Custom Test To insert the new Custom Test press the button F4 or exit using the x button F3 and return to the previous screen without changes being made NIBP Settings o curve P Systolic 30720mmHg Diastolic 80mmHg Custom Test urn into service mode OOGA Further tests can be inserted or deleted by following the steps described herein To exit and save the changes for future press the e button followed by pressing the ni button If the Escape button F3 is pressed the Rigel BP SIM will return to th
15. provides the user with the ability to run a preset test sequence and store the data afterwards Press F4 select Auto Mode using the up down keys Select by pressing the press V F4 to accept Heart Rate 80 bpm Default Systolic 1 20 mmHg View Data Data Transfer gt User Name Admin Asset ID Test Sequence Default Adult m Run Mode Semi Automatic Asset ID Each test record is stored using the Asset ID number 25 character max and date time of test Multiple entries can be applied using the same Asset ID number In this case the memory will hold all entries and differentiate using the test date Test Sequence The Rigel BP SIM comes standard with preset test routines to meet the requirements of e Default Adult Sequence e Default Pediatric Sequence In addition the user can create a total of 50 preset test sequences to meet specific requirements See 3 1 more information on creating new test sequences 35 Page Rigel Medical BP SIM user Manual V3 1 Run Mode e Semi Automatic allows the user the enter data after each individual test that forms a test sequence e Full Automatic will run through the test sequence and provides an opportunity to enter data at the end of the complete sequence Test Period The re test period allows the user to include the next test date on printed labels and PC downloads that make it easier to schedule future work More details on printing labels see 4 2 Trace Variabl
16. the correct value of 57600 In the application select Start Capture from the Capture menu in the task bar Configure COM Port Time 00 00 00 Estimated time to completion 00 00 00 Ready Data received from Tester Capture OFF Fie Name Enter a valid file name for the text file containing the download information see below Save in e Config Data 4 My Recent Documents Desktop My Network File name Places Save as type J tat Once the text file has been created press the Send button F4 to transfer the Asset Trace variables and User names from the Rigel BP SIM to the PC 58 Page Rigel Medical BP SIM user Manual V3 1 The downloaded text will appear in the Bluetooth Downloader screen Select Stop Capture from the Capture menu in the task bar See below w start Capture Fy EUM zWHIOEdFIOSIBSE Tire C0023 20 Estir aed lime u cuinzleli n 20 00 0C Ready Zala reseivac rcr seter Ccrmen depHodueNene AP Modue 1 P Modus z P tdi hare 7 SpUe NIBP CCl Destin Peds Tren piahi i an ji 4 Caidiac Zuiput EMD Capture Os D 1Doz rnc ks and SH rasiohnZosktop bp sim txt Close down the Application and open the text file using word pad or equivalent The downloaded information can be used to clone other Rigel BP SIM s or modified to add additional trace variables Again these can be uploaded to any Rigel BP SIM for convenience Note
17. to the list for future use Trace Variables that are de activated set to NO in the Setup menu section 3 2 will not appear during the automatic tests Note that a maximum of 40 individual entries are available for each Trace Variable Additional entries will overwrite on a first in first out basis When all information is provided press the button F4 to start the safety test After an initial message The test about to start requires the unit to be in service mode switch to service mode now The first test of the Default test setting is the the NIBP Leak Test 38 Page Rigel Medical BP SIM user Manual V3 1 NIBP Leak Test P Drop Start P mmHg 156 6 Timer Following the Leak test the overpressure valve is tested NIBP Pop off Test Set Level mmHg 272 7 Following the Overpressure valve test the BP SIM is ready to make NIBP simulations as per the test routine Pen 60 bpm Default Systolic 1 2 0 mmHg Diastolic 80 MAP 93 Pulse volume Medium Pressing the x STOP button F3 during any test will bring up the failure menu see 4 1 1 At the end of the test the user has the ability to enter comments See 3 3 3 for more information on how to activiate comments after a test 39 Page Rigel Medical BP SIM user Manual V3 1 Enter User Comments Example test INE2 INES Press the button F4 to store the comments Depending on the settings in Menu Setup System Config see 3 3 2
18. Eite 36 6071 Te Printer Type Test n Tag Elite lt gt Bluetooth ID 00037417E631 PIN number New Elite 36 5011 lt gt Ensure when selecting the printer that the Printer type is selected corresponding with the print media Printer used BT ID Printer type Test and Tag Elite Elite 36 xxxx Test n Tag Elite Citizen list printer CITIZEN SYSTEMS Citizen CMP 10 Test and Tag SW T nT 6oo xxxx Test n Tag Press v F4 to store the changes and return to the previous screen 28 Page Rigel Medical BP SIM user Manual V3 1 Bluetooth Favourites Barcode Printer Elite 36 5011 lt gt Computer SELLTO23 lt gt SELLT023 and Elite 36 5011 are now a Bluetooth Favourite and the Rigel BP SIM will automatically establish a connection with SELLT023 when using functions that communicate with a computer e g downloading records to a PC database software package or connect to Elite 36 5011 when requiring a print out Note Press x F2 to exit without saving changes Press C F3 if necessary to select another device V4 Press F4 button to save changes When all your favourite devices have been set up press F4 and they will be saved 3 5 User Admin The User Admin feature allows the user to personalise the way in which the tester behaves during normal use and include features such as Preferences see 3 5 1 Change User see 3 5 2 Change Password see 3 5 3 User
19. Mode The BP SIM has a built in manometer function which allows the user to measure the pressure in a number of engineering units Select the Manometer mode from the NIBP menu 48 Page Rigel Medical BP SIM user Manual V3 1 NIBP Manometer Mode 0 0 Pressure Units mmHg Ensure the manometer is set to zero by pressing the SET ZERO F3 button whilst the pressure connection at the BP SIM panel is open Select the engineering unit from the drop down box using the left and right keys NIBP Manometer Mode mmHg InchH20 mBar cmH20 InchHg Psi eI Ie The manometer will automatically convert the pressure to the desired pressure unit Pressure Units m F2 Press to return to start menu 5 5 Set Zero The Set Zero function is required to ensure the manometer is pressurised to the ambient pressure 49 Page Rigel Medical BP SIM user Manual V3 1 6 View Data To view the stored data press the button F4 and select View Data from the menu Heart Rate 80 bpm Default Systolic 1 20 mmHg Auto Mode Data Transfer gt Setup gt The following screen provides an overview of all available records in the memory identified by Asset ID Site and Pass or Fail View Data Search Result Peterlee Peterlee Use the up and down arrow keys to scroll through the database When the desired record is highlighted press the V button F4 to enter the record and view the data
20. P performance Testing 5 Page Rigel Medical BP SIM user Manual V3 1 1 2 Key Features e Portable and Lightweight Ideal for field service technicians and in situ testing of NIBP performance e Battery Operated For convenient and fast testing without mains power e Manufacturer O curves To ensure accurate simulations across a wide range of e different brands and models e Internal Pump Easy testing of pneumatic leaks and over pressure valves e Prepared for PPM Protocols Configured for automatic performance testing of a variety of parameters e Full Keyboard Easy data capturing and asset management e Intuitive User Interface Fast and easy navigation using graphics display e Graphics Display Allowing real time display of pressure curves and simulation data e BlueTooth Communication PC download and future field upgrades e Web Updates Enjoy easy updates through our website to benefit from the latest features 6 Page Rigel Medical BP SIM user Manual V3 1 1 3 Interfaces Full graphic Monochrone LCD blue with white backlit 1 4 VGA Larger Red key Stop Off Key Larger Green Start On key 4 Programmable soft keys below display Alpha Numeric keyboard Up Down and Left Right cursor control Power Supply inlet Pneumatic interface connector Input Output connector futu
21. Profile see 3 5 4 To enter the User Admin feature press the button F1 from the home screen select Setup followed be User Admin The following menu will appear 29 Page Rigel Medical BP SIM user Manual V3 1 Test Sequences Systolil Asset Trace Variables Data T Memor Change User EU Restord Change Password Navigate through this menu using the Up amp Down Keys and select by pressing the V4 button F4 or the right arrow key 3 5 1 Preferences Provides means of configuring the default settings of the tester s behaviour during normal use All settings are user specific and stored against the logged in user Preferences User Name Admin Set Contrast Auto Off Time mins Data Entry gt lt lt Backlight Mode User Rights e Set Contrast Use the up down left and right keys to set the contrast e Auto Off Time mins Use the up down left and right keys to set the power off time on the tester Settings between 0 10 minutes Note 0 minutes will deactivate the AutoSwitch Off feature e Data Entry Sets the tester up to take data entry from Barcode or Keypad only Use the up down left and right keys to select Barcode or Keypad only Note n data entry mode the unit will automatically try to connect to a barcode scanner via the internal Bluetooth module if the scanner is available this feature will drain the batteries u necessarily In this instance it is recommended that the unit
22. TE NM 52 6 2 Search the database aneen aa n nn neun nn aseene 53 6 3 View Data Options ne d teste die na aane eo tee aa on auae aon aano Edna o 54 2 Page Rigel Medical BP SIM user Manual V3 1 7 Bata Transfer hentai 55 7 1 P Tepic 55 7 2 EIDERRICINGEN 56 7 3 Configuration Data sise eh FR RE eoe o ou ees Phe EN ea coceusedeuecsstecsenss 57 7 4 Load TNT LOB 60 Six WIDOUEL Loire ED OTRO as PARTO dard Ptit 62 9 Maintaining the Rigel BP SIM 1 c eese eese eene ee eene n nennen nn nita eene aa nens nu nan 64 9 1 Cleaning the BP SIM 1 oo ie disi ona e pepe sa ce e Fa Foo eoe be es Poe sana aoo Thus n oae daa To arb aae de add 64 9 2 User Maintenance ses M 64 9 3 Return INStrUctiOnsS 65 10 Technical Specifications sistema 66 3 Page Rigel Medical BP SIM user Manual V3 1 User Notes These operating instructions are intended for the use of adequately trained personnel The following symbols are used in these operating instructions and on the Rigel BP SIM Warning of electrical danger A Indicates instructions must be followed to avoid danger to persons Important follow the documentation This symbol indicates that the AN operating instructions must be adhered to in order to avoid danger If the BP SIM is used in a manner not
23. be set to Keypad only e Backlight Mode Chose between Always Off Always On Power Save Back light on for 30 secs after each key press Use the up down left and right keys to select 30 Page Rigel Medical BP SIM user Manual V3 1 e User Rights For information only To change the user rights see 3 5 4 3 The F2 button allows the user to set the audible warnings and beep on key presses See menu below one preferences Beep on key press Beep on Heartbeat Beep on Barcode Beep on warning Beep on data entry Beep on Tests Navigate using the Up amp Down Keys and change between Yes No using the left key Confirm the settings using o V button F4 this will return to the Home screen o button F2 will return to previous menu and save data x o button F3 will return to the home screen without changes being saved 3 5 2 Change User This feature allows an operator to default to a different existing user or setup a new user Admin rights are required for this feature Alternatively new users can be created in the User Profile menu see below Note A brand new Rigel BP SIM will automatically default to admin rights so new users can be created 3 5 3 Change Password This feature allows a new password to be assigned or changes to an existing password To create a new password highlight the New Password field using the Up amp Down Keys and enter a new password followed by confirmation o
24. blished and ready to communicate the Bluetooth status icon will change to show the Bluetooth icon as shown on the right 55 Page Rigel Medical BP SIM user Manual V3 1 Download ta PC Computer Baud No of assets sELLTO23 57600 Format Rigel SSS aiting to connect Connected The Rigel BP SIM is able to download in a number of formats to suit individual requirements These are e CSV Full Comma Separate Value download only CSV Summary Comma Separate Value download only e Rigel SSS Up and download format Toggle between the options by using the left and right arrow keys The Rigel BP SIM is now ready to transfer data The No of assets field shows the number of records that will be downloaded Ensure that the PC application is ready to receive data then press the Send button F4 A progress bar as shown below will now follow the data transfer until download is complete Download to PC Computer Baud No of assets sELLTO23 57600 Format Rigel SSS aiting to connect Connected E Note To download a selection of the database refer to View Data section 6 Search on the required criteria then select download to pc from the options menu 7 2 Upload from PC This feature is only available when used in combination with the new Med eBase PC software and when the SSS format is activated on your device 56 Page Rigel Medical BP SIM user Manual V3 1 Select the Upload from PC option f
25. download all records within the search criteria to the Test n Tag printer setup in Bluetooth favourites e Download to PC Will download all records within the search criteria to the PC setup in Bluetooth favourites Delete Single Asset Will delete the single selected record e Delete All Assets Will delete all records from the memory this action is irreversible 54 Page Rigel Medical BP SIM user Manual V3 1 7 Data Transfer Data transfer is used to transfer records and configuration data to and from a PC load a logo in bitmap format for use with the Seaward Test n Tag print system and clone data The Data Transfer functions are accessed by pressing the button F4 selecting Data Transfer using the up and down keys and pressing the button F4 Heart Rate 80 bpm Default Systolic 1 20 mmHg ada a E C pload from PC Configuration Data Load TnT Logo 7 1 Download to PC This feature is used to download records to a PC Select Download to PC and press the v button F4 to accept The Rigel BP SIM will now attempt to make a Bluetooth connection with the Computer that has been configured in the Bluetooth Favourites Download ta PC Computer Baud No of assets Fu sELLTO23 57600 Format Rigel SSS aiting to connect Initially the Bluetooth status icon will display the egg timer for a few seconds as shown on the left while the connection is established When the connection is esta
26. e previous meny without changes being made Save changes 19 Page Rigel Medical BP SIM user Manual V3 1 3 1 6 Create a new Test Sequence To create a new Test Sequence enter the Menu button F4 from the home screen select Setup followed by Test Sequences from the menu Heart Rate 80 bpm Default Asset Trace Variables System Config The following screen displays the list of default and customised tests available Copied Test Sequences will appear in the list with an pre fix From this menu the user is able to create a new Test Sequence by pressing the ur CA aa 45 button F1 20 Page Rigel Medical BP SIM user Manual V3 1 Use the keyboard to activate the text box When activated function keys F1 tT and F2 uy appear automatically In this instance Customer test 1 has been created as Test Name New Test Sequence Customer test 1 To change to upper or lower case or use language specific charaters press the tT button F1 and select the required entry To Delete a character press the button F2 To return to the previous menu without changes press the x button F3 To confirm the entry press the button F4 Next step is to save the name of the newly create Test Name by pressing the OK button F4 Pressing the x key F3 would return to the Test Sequence menu without changes The newly created Test Name now appears in the Test Sequence menu and can be L modified to i
27. e shown below the password field and the serial number of the tester A new password will then be created For security reasons this new password can only be provided to the email address of the original purchaser of the device The Rigel BP SIM is designed to allow a user to customise test sequences and allow default settings This customisation can speed up testing by providing the user with default manufacturer lists model numbers user text protocols automatic printing post text and fault menus 10 Page Rigel Medical BP SIM user Manual V3 1 3 Setting up your tester The Rigel BP SIM is designed to allow the user to customise the device to allow default setting in order to speed up the testing such as default lists of manufacturers model numbers user test protocols automatic printing following test fault menu s and so on All custom facilities can be found under the SETUP menu Simply press the F4 from the home screen and select Setup from the menu Hint The underlined character acts as a short key to allow swift navigation through the menu structure est Sequences Asset Trace Variables System Config Bluetooth Favourites User Admin gt Memory Options Restore Factory Settings Restore Factory Sequences Options available Test Sequences Modify or create test sequences see 3 1 Asset Trace Variables Generate default list of variables see 3 2 Systems Config Configure default test options see 3 3
28. e requirements of Council Directive 2004 108 EC and 2006 95 EC This Conformity is indicated by the symbol i e Conformit Europ enne Seaward Electronic Ltd is registered under BS EN I809001 20068 Certificate No Q05356 1 Page Rigel Medical BP SIM user Manual V3 1 Index User NOTES cccessccnsesinecersocdssacasssccedeceanesteusseetostieosssnsatsecadsceceacesansereensecnaesdeseabesasececesssee 4 1 Introduction ea seteainaaedaneion oa dtinaandsouvndeduduapamiesgnds cssatenabdunusaeevedes 5 1 1 Rigel BP SIM Design Philosophy ccsssssssccssssscccesssccccsssssccesssscceenssssceaasseseeaes 5 1 2 CRIT EPHTOLOTIDOTDIDU CLTC D CQ IDDLILTIDIDLICILLDOITL LLT 6 1 3 Interfaces riii niente eer HE HERE nee bete Pea ba Eae PEERS E RESP RUE Ek P PEE BERE etes nt ne ne 7 1 4 Rigel BP SIM includes 1 n1 re itedia soci eno sedo eo aon so tera ou D rao SEPE EANO Ineas go poaae ea dt 8 1 5 Optional Accessories 5 ree iee torpe sa ee pn uera psa a ee nn sa Eee uy ga aa eo vn sossseedessssdeesessexs 8 1 6 Test n Tag Compatible un era SD Ra e Pee ERR ONE Rage 8 1 7 Unique use of ICONS cciccs sccecccceseseccasctececsssedeccaeseneccsessscccsscdacccecedecssacsscsesesssesees 9 PAESE Forrlirguup ER 10 Setting up your tester 4 ioee ee error or pa eoe iren in e aeos aee p reo en Pao ke s uana ones te eisai eaaa 11 3 1 I EITIuIB 11 3 2 Asse
29. erials and components which can be recycled and reused When this symbol is attached to a product it means the product is covered by the European Directive 2012 19 EC Please familiarise yourself with the appropriate local separate collection system for electrical and electronic products or contact your local supplier for further information Please dispose of this product according to local regulations Do not dispose of this product along with normal waste material By offering your old products for recycling you will help prevent potential negative consequences for the environment and human health Certificate of Conformity This product is manufactured by Seaward Electronic Ltd Bracken Hill South West Industrial Estate Peterlee County Durham SR8 2SW UK Statement of Conformity As the manufacturer of the apparatus listed declare under our sole responsibility that the product Rigel BP SIM NIBP Simulator To which the declaration relates are in conformity with the relevant clauses of the following standard BS EN 61010 1 Safety requirements for electrical equipment for measurement control and laboratory use Part 1 General requirements BS EN 61326 Electrical equipment for measurement control and laboratory use EMC requirements Performance The instrument operates within specification when used under the conditions in the above standards EMC and Safety Standards The product identified above conforms to th
30. es e Goto the home screen F1 e Continue with the next test F2 e Options F3 To view the options menu select the button F4 from the test details screen The options menu provides the following sub menu Asset ID 123456 Tested by Tested on 08 Oct 2008 12 12 PM Print Label Print Result View Results will display the results of the test see section 4 4 User Comment will allow the user to enter comments if required Print Label will print a pass fail label on the thermal printer 50 mm roll Print Result will print the result on the thermal printer 50 mm roll 41 Page Rigel Medical BP SIM user Manual V3 1 4 4 View Results Press the View result Button to review the data Use the Up and Down keys to scroll through the screen Put in service mode Use the e button F2 to return to the Test Details screen The button F4 provides extra option to print results label or Test n tag 42 Page Rigel Medical BP SIM user Manual V3 1 5 Manual Mode The Manual mode provides the user with the features of testing a specific individual test and or test condition for example to test a selection of Static pressure calibration Pop off valve test Manometer function Press NIBP F3 from the start screen Rate 80 bpm Default Systolic 120 mmHg Diastolic 80 MAP 93 Pulse volume Medium When NIBP is selected a number of options will be available 43 Page ae 80 bpm Default Systolic
31. es To increase the traceability of the safety test the user can include valuable data with the test record such as Re test period and Asset variables see 3 2 for more information Trace variables can be added once the above data has been entered and confirmed See next chapters for more details 4 1 Performing an Automatic Test The automatic mode provides an option to test all available tests NIBP simulations Leak Test Overpressure test in an automatic sequence 4 1 1 Failure Menu When a specific test fails the user will be provided with a number of options depending on the nature of the fault The Failure Menu will enable the user to Restart Test Individual test e Restart Test Sequence Whole Test sequence Resume Test Skip the failed test e End Test Sequence store result Restart Test Sequence and follow options See 4 3 prie Sequence e Abort Test Sequence no cn es equence F Rhon Test Sequence information stored return to home screen The following procedures describe the setup and completion of an automatic test sequence The steps are applicable to all available test sequences As example we describe a default test sequence 36 Page Rigel Medical BP SIM user Manual V3 1 Press the button F4 from the home screen and select Auto Mode This will show the Asset Details screen as detailed below Asset Details User Name Admin asam Test Sequence Default Adult m
32. f the new password Use the key F4 to finish and press the V button F4 to confirm and save the new password To change an existing password enter current password and repeat the steps above See menu below for an overview of the Change Password screen 31 Page Rigel Medical BP SIM user Manual V3 1 Change Password User Name Admin Current Password New Password XXXXX Confirm New Passw XXXXXI 3 5 4 User Profile This feature allows the administrator to setup new and existing users and inhibit or assign certain feautures to individual users Note that the User Profile is only valid when the ADMIN user has been setup with a PASSWORD See item above If no ADMIN password has been set all users have FULL User Rights Select the applicable user name use the Up amp Down Keys to navigate through the menu use the Left amp Right Keys to activiate yes or de activate no certain features See menu below User Name System Config Download to PC Upload from PC Erase Restore Factory Settings Edit Test Sequences Press the v button F4 to save user profile 3 6 Memory options This feature is used to view information about the Rigel BP SIM memory status Press F4 select setup using the up down keys and select Memory Options by pressing the press F4 to accept 32 Page Rigel Medical BP SIM user Manual V3 1 Memory No of assets o Space available 100 0 Assets Deleted
33. h the life of the instrument and its reliability If the Rigel BP SIM is subject to condensation allow the Analyser to completely dry before use e Always check the Rigel BP SIM and supplied accessories for signs of damage and wear before use e Do not open the Rigel BP SIM under any circumstances e Keep the instrument clean and dry e Avoid testing in conditions of high electrostatic or electromagnetic fields e Maintenance should only be performed by authorised personnel e There are no user replaceable parts in the Rigel BP SIM e The unit should be regularly calibrated at least annually 64 Page Rigel Medical BP SIM user Manual V3 1 9 3 Return Instructions For repair or calibration return the instrument to Rigel Medical Service Dept p a Seaward Electronic Limited Bracken Hill South West Industrial Estate Peterlee Co Durham SR8 2SW England Tel 44 0 191 5878739 Fax 44 0 191 586 0227 sales rigelmedical com calibration seaward co uk www seaward co uk www rigelmedical com Prior to returning your unit for service please contact our service department to obtain a Returns Number By obtaining a Returns Number your service request can be booked in advance thus reducing the down time of your equipment When asking for a Returns Number please quote e Instruments name and model e Serial number see section 8 e Firmware version see section 8 65 Page Rigel Medical BP SIM user Manual V3 1
34. however alterations are possible by producing a copy of the default test setting see 3 1 2 Non default Test Sequences can be modified from the Test Sequences menu When G non default Test Sequences are highlighted the screen F4 will appear in the menu Development only Four Development only One 13 Page Rigel L Pessing the below button F4 will open the Test Sequence for Editing as shown To change the Name and or Class of the Test Sequence press the G Button F4 and enter the Name and or Class required Edit Sequence Name Custom test sequence Class NONE lt gt To change to upper or lower case or use language specific characters press the tT button F1 and select the required entry To Delete a character press the button F2 To return to the previous menu without changes press the x button F3 To confirm the entry press the button F4 14 Page Rigel Medical BP SIM user Manual V3 1 Edit Sequence Name Confirm the changes with the button F3 or cancel using the x button F4 To Insert a new test highlight the test using the Up amp Down arrow keys and press the Insert button F1 To Delete an individual test highlight the test using the Up amp Down arrow keys and press the button F2 To go back to the previous menu highlight the test using the Up amp Down arrow keys and press the x button F3 To modify an individual test highlight the
35. in built pump to generate user specific pressure settings for comprehensive system leak tests and over pressure tests An integral digital manometer also enables highly accurate static pressure measurements to be taken using a wide range of selectable pressure units The new Rigel BP SIM forms part of a comprehensive range of high performance specialist biomedical test equipment supplied by Rigel Medical part of the Seaward Group 1 1 Rigel BP SIM Design Philosophy The BP SIM has been designed to address the increasing demand for smaller more comprehensive test equipment within the Healthcare Industry The challenge was to combine the benefits of the size and weight of a smaller hand held NIBP Simulator with the test versatility and convenience of a larger bench mounted simulator Today s industry demands test equipment that can save time and cost thus offering greater flexibility These factors have all been taken into account during the development of the BP SIM The test capabilities and functionality exceed that of most common NIBP Simulator yet the instrument is only a fraction of the size and weight No other NIBP simulator on the market offers a hand held enclosure with the test capabilities equal to most bench marked version interface large internal memory Blue Tooth communication asset management facilities user configurable Performance Tests and more We believe that the new BP SIM is set to become the new standard in NIB
36. menu click on El followed by Setup Select the Test N Sequences from the list and press F4 el 80 bpm Default Asset Trace Variables System Config M Memory Options Restore Factory Settings The following overview is provided and shows the list of default and customised test sequences availble Copied Test Sequences will appear in the list with an pre fix SMELL p View os From this menu the user is able to View a test sequence by pressing the View button F4 Default test settings cannot be altered at any stage 12 Page Rigel Medical BP SIM user Manual V3 1 To Copy Delete or Print an existing or default test sequence use the Up amp Down arrow keys to highlight the test sequence and press the button F2 Use the Up amp Down Arrow keys to select the required action and press the F4 to confirm Press Escape F3 to cancel and return to previous screen Copied Test Sequences will appear in the list with an pre fix All Copied test Sequences can be edited See 3 1 3 for more on editing test sequences Print Test Sequence will provide an overview of the Test Sequence on the Favourite Blue Tooth Printer See 3 4 for help on setting up Blue Tooth devices Delete Sequence will remove the highlighted test sequence from the memory of the Rigel BP SIM Note Default Test Sequences cannot be Deleted 3 1 3 Modifying Edit Existing Test Sequences Preset test sequences cannot be modified by the user
37. nclude the required individual tests Press the the sequence for editing button F4 to open Customer test 1 21 Page Rigel Medical BP SIM user Manual V3 1 Once the Test Sequence has been opened individual tests can be inserted by highlighting the place where a new test needs inserting using the up and down keys Note Tests will be inserted before the highlighted position not after To modify a test sequence please refer to 3 1 3 3 2 Asset Trace Variables The Asset Trace Variables allow the user to include valuable data to the test results to enhance the traceability and to provide increased search criteria when using database software Asset trace variables may consist of a maximum of 25 characters and can be selected prior to each test when using the Rigel BP SIM in automatic mode Heart Rate 80 bpm Default Systolif Test Sequences Asset Trace Variables System Config View Bluetooth Favourites Data T User Admin gt SU Memory Options Restore Factory Settings The following variables can be added to the test results Asset Trace Variables Service Code Site Location Make Model Description Serial Number Client The default settings on the Rigel BP SIM include the Site and Location variables Both are activated The drop down box is set to Yes To activate or de activate each variable use the up and down keys to select the variable and the left and right keys to select yes or no
38. ng the left key and or scroll through the available tests using the up and down keys Visual Test NIBP Leak Test NIBP Settings o curve NIBP Simulation Custom Test CETTE Once the desired test has been highlighted confirm using the OK button F4 and finally confirm to insert into the test Sequence Pressing the x button F3 will return to the previous screen without changes being made The individual test has now been inserted and can be edited as described in 3 1 4 NIBP Leak Test Set Level 275 mmHg Duration 180s Further tests can be inserted or deleted by following the steps described herein 17 Page Rigel Medical BP SIM user Manual V3 1 3 1 5 Insert Custom Tests or Instructions Custom tests or Instructions are intended to allow the user to capture additional information either prior to or after the main NIBP performance tests Such information could indicate the condition such as Labels software version certain damage or upgrades To create customised inspections use the insert Custom Test function using the instructions below and set the engineering units to blank Hint Create a unique range of visual tests or instructions by creating a new test sequence see 3 1 6 and select Custom Test as the nature of the test This will create a unique test sequence The maximum number of characters in the test description or instruction is 255 To insert a Custom Test highlight the place where a new test needs
39. nk Will leave the next Asset ID field blank Repeat Last Will copy the previous Asset ID into the next Asset ID field 3 3 2 After Test Provides automatic action after a test is completed Use the left and right keys to select from the following New Test Automatically brings up the next test screen Download Automatically down loads the test results to the PC see 7 1 Print Label Automatically prints the test results or label to a printer see Appendix B Options Menu Provides a menu to allow further choices after a test see 4 3 3 3 3 Comments Provides the ability to enter further comments after successful or failed tests Use the left and right keys to select from the following Always Comment field is displayed after each passed or failed test On Pass Comment field is displayed after a passed test only On Failure Comment field is displayed after failed test only Never Comment field is not displayed 3 3 4 Language Provides the ability to change the default language in the Rigel BP SIM A number of languages are available and will be updated which ach firmware revision Use the up down left and right keys to select from the available languages Additional languages firmware upgrades can be downloaded from www rigelmedical com or contact your local Rigel sales office UK 44 0 191 5878701 3 3 5 Date Time Allows the user to set the current date time and preferred formats
40. o Assets Uploaded o The display shows the number of asset records currently stored the remaining memory space number of assets deleted and number of assets in upload memory The nature of Flash memory is such that when records are deleted only the reference to the data is removed The data still remains and will use memory space The memory must be erased in order to release memory used by deleted records If the Erase F2 key is pressed a prompt is shown below To erase the memory press 4 button F4 Clear Results Memory Are you sure 33 Page Rigel Medical BP SIM user Manual V3 1 3 7 Restore Factory Settings The factory settings can be restored at any time using this function Press button F4 Select Setup followed by Restore Factory Settings Heart Rate Default Systolil Test Sequences Asset Trace Variables System Config All the settings provided within the Setup menu will default to Factory Settings including All Asset Trace Variables Test Sequences User Admin Test Codes and System Config Warning Restore Factory Settings cannot be undone and will remove all the customised items described above Hint All customised settings within Setup can be cloned to a PC We advise that this is done on a regular basis to ensure custom settings are saved for back up See 7 4 for further details on Cloning 34 Page Rigel Medical BP SIM user Manual V3 1 4 Automatic Mode The Automatic mode
41. o highlight the required device in this case SELLTO23i and press the v button F4 SELLTO23 is now placed in the New field as shown below TFC ANANI Setup Computer SELLTO23 lt gt Bluetooth ID OOOA3A76A88B PIN number _ New SELLTO23 m Press F3 to add SELLT023 to the Favourite Computer list To add further devices to the list of available computers select the new field again use the up down left and right keys to select the required device and press F3 to add the chosen device SELLT023 has been added to the Computer field with Bluetooth ID 4 Press F4 to store the changes and return to the previous screen 27 Page Rigel Medical BP SIM user Manual V3 1 A PIN number can be entered if required for the device to be added to the Favourites list Note Seaward and Rigel accessories do not require a PIN For other Bluetooth devices refer to the manufacturer s documentation Note Devices that are not required can easily be removed by selecting the Computer field pressing the left of right keys to open the list highlighting a device using the up down keys and pressing F2 Confirm the deletion by pressing the V button F4 Press the v button F4 to save Bluetooth Favourites settings All Bluetooth devices have a unique ID The Bluetooth ID is shown when a device is added to the Favourites list Repeat the actions above to select a printer for the BP SIM Printer
42. ons without prior notice Calibration Statement The Rigel BP SIM NIBP Simulator is fully calibrated and found to be within the specified performance and accuracy at the time of production The Seaward Group provides its products through a variety of channels therefore it may be possible that the calibration date on the provided certificate may not represent the actual date of first use Experience has indicated that the calibration of this instrument in not effected by storage prior to receipt by the user We therefore recommend that the recalibration period be based on a 12 month interval from the first date the unit is placed in to service Date received into service i 1 Copyright 2015 All rights reserved Nothing from this edition may be multiplied or made public in any form or manner either electronically mechanically by photocopying recording or in any manner without prior written consent from the SEAWARD GROUP This also applies to accompanying drawings and diagrams Due to a policy of continuous development the SEAWARD GROUP reserves the right to alter the equipment specification and description outlined in this publication without prior notice and no part of this publication shall be deemed to be part of any contract for the equipment unless specifically referred to as an inclusion within such contract Disposal of old product The Rigel BP SIM NIBP Simulator has been designed and manufactured with high quality mat
43. re use only SO m DIT e 7 Page Rigel Medical BP SIM user Manual V3 1 1 4 Rigel BP SIM includes Battery charger Carry case 5 pc tubing adaptor set Instruction manual Application disc BlueTooth USB adaptor Calibration certificate 1 5 Optional Accessories Med eBase PC Download software p n 301A920 BlueTooth Barcode scanner p n 339A923 BlueTooth Test N Tag System p n 331A950 BlueTooth Test N Tag Elite System p n 339A970 BlueTooth result printer 560mm p n 339A930 1 6 Test n Tag Compatible The Rigel BP SIM is compatible with the Seaward Test n Tag and Elite system Our Test n Tag systems allows you to print customised Thermal PASS FAIL labels The benefits of using the Test n Tag printers are e Robust and durable labels Resistant to most solvents used in the medical sector Free opportunity to advertise your logo and company details or emergency telephone number with every item you Test n Tag e Automatic barcode generation to provide easy use of the barcode scanner thus speeding up your test time The unique Test n Tag label provides test status and retest due date barcode and asset ID number and person conducting the test For more information on using the Test n Tag printer see section 4 8 Page Rigel Medical BP SIM user Manual V3 1 1 7 Unique use of ICONS The Rigel BP SIM features a hi resolution graphic back lit display which not only provides highly
44. rom the Data Transfer menu The Rigel BP SIM will now attempt to establish a Bluetooth connection to the PC as previously described in Download to PC The Rigel BP SIM is now ready to receive upload data from a PC Refer to the Instruction manual of MediGuard plus on how to transmit data from the PC To activate the SSS format on your Rigel BP SIM tester please download the SSS Application note from our knowledge base Please visit www rigelmedical com knowledgebase 7 3 Configuration Data This feature can be used to transfer store or modify the current Trace Variables User Profiles and Comment lists to from a PC It can be used to configure a number of Rigel BP SIM units with identical Trace Variable User Profile and Comment lists Heart Rate 80 bpm Def ault Systolic 1 mmHg Once the BP SIM has connected to the PC the BlueTooth icon will appear Capture the data using the datatransfer exe program or Windows hyperterminal aiting to connect Connected Ready to Send And Receive Data Press the SEND F4 button to send data to the PC The BP SIM will automatically receive data when the ready message is shown as above Press the e F3 button to go back to the start screen 57 Page Rigel Medical BP SIM user Manual V3 1 Open the Bluetooth Downloader Data Transfer exe application on your PC and check that the Com Port is correct for the Bluetooth USB adaptor being used with the P C The baud rate will default to
45. s screen see below View Test Results Visual Test Use the e button F2 to return to main menu Press the button F4 to provide the options to print the results 52 Page Rigel Medical BP SIM user Manual V3 1 View Test Results Visual Test 6 2 Search the database From the main View Data screen press the a button F1 View Data Search Result Asset ID Test Status Date from Date to Service Code Site Location Make Use the up down left and right keys to navigate this screen The default setting is which includes all records that Passed To search for more specific criteria fill in the required fields by either using the keyboard or drop down boxes 53 Page Rigel Medical BP SIM user Manual V3 1 Note The entered data must be identical to the data stored with the test This includes visible and invisible SPACE characters For Date entry use format DDMMYYYY or DD MM YYYY For example for 31 October 2008 use 31102008 Test Status Allows the user to select search criteria based on PASSED FAILED or UPLOADED items To start the search press the v button F4 or the button F3 to leave this screen 6 3 View Data Options te E From the main View Data screen press the button F2 View Data Search Result Download to PC f Delete Single Asset Delete All Assets This will present an options menu to allow the user to e Print All Labels Will
46. specified by the manufacturer the protection provided by the equipment may be impaired The BP SIM is supplied with a mains power supply unit The use of any other power supply unit may cause damage to the equipment and may result in a hazard to the user 4 Page Rigel Medical BP SIM user Manual V3 1 1 Introduction The Rigel BP SIM is a new hand held non invasive blood pressure simulator for the fast and accurate testing of all tyoes of NIBP monitors The new lightweight and battery powered Rigel BP SIM utilises patented technology to simulate a wide range of blood pressures to verify the accuracy and safe operation of all leading NIBP monitors currently used in healthcare and medical facilities The new simulator incorporates a range of custom settings that include a variety of paediatric and adult NIBP pressures pulse volume adjustments heart rate and manufacturer specific O curves The Rigel BP SIM is the first hand held NIBP simulator combining full performance features with a large capacity internal memory for the data capture storage and subsequent downloading of test results for record keeping purposes In addition the Bluetooth enabled technology allows the wireless connectivity of the tester with PCs and other accessory equipment for fast and convenient downloading of performance data and the uploading of manufacturers O curves and manufacturer specific test programs Other special features of the Rigel BP SIM include an
47. strength to simulate different volume strength Manufacturer WA ProPaq Adult Infant Systolic Diastolic guy Heart Rate Pulse volume High Neo natal Note that the settings Low Medium High are not directly linked to any physiological condition 4 Once the selections are complete select the F4 to confirm the settings 45 Page Rigel Medical BP SIM user Manual V3 1 Manufacturer Default lt gt Adult Infant Adult lt gt Systolic Diastolic 120 80 lt gt Heart Rate DSC Pulse volume lt gt Press the Green Start key to start the simulation and close the internal valve within the BP SIM Note when the valve is not closed the pressure within the cuff will not be contained and the monitor might go into alarm When the simulation is present ie the internal BP SIM valve is closed a pulse symbol will be shown ps 80 bpm Default Systolic 1 2 0 mmHg Diastolic 80 MAP 93 Pulse volume Medium 5 2 NIBP Leak Test In the NIBP leak test the user is able to test the pneumatic leakage within the system The settings available are e Start Pressure e Timer seconds Note Prior to initiating the test some NIBP monitors might have to be set to a service mode to ensure that the internal monitor valves are shut for incoming pressures If the monitor s internal valve is open the leak test will not be able to pressurise the system 46 Page Rigel Medical BP SIM user Man
48. t Systolic 120 mmHg Auto Mode View Data Data Transfer gt MEDICAL Rigel BP SiM Tech Support 44 191 587 8718 Firmware Version 0 032 Use the scoll keys F3 to toggle between the different information screens 62 Page Rigel Medical BP SIM user Manual V3 1 About Serial Number YOO 0000 Processor info 2 Bluetooth ID 0012F3066687 Bluetooth ID Status Pass Non Invasive Blood Pressure Firmware Version 0 21 Hardware Version 1 Calibration Status OK Last calibrated 26 Sep 2008 AA Ensure you have this information available when contacting Rigel Medical for Technical Support or Service See chapter 9 for more information on Service and Calibration Procedures 63 Page Rigel Medical BP SIM user Manual V3 1 9 Maintaining the Rigel BP SIM 9 1 Cleaning the BP SIM The Rigel BP SIM case can be cleaned with a damp cloth with if necessary a small amount of mild detergent Prevent excessive moisture around the socket panel or in the lead storage area Do not allow liquid inside the Rigel BP SIM or near the sockets Do not use abrasives solvents or alcohol If any liquid is spilt into the Rigel BP SIM case the simulator should be returned for repair stating the cause of the defect 9 2 User Maintenance The Rigel BP SIM is a rugged quality instrument However care should always be taken when using transporting and storing this type of equipment Failure to treat the product with care will reduce bot
49. t Trace Variables E 22 3 3 System Cohflg ee riris rie cscccesetsscacsscedececsusecscceusccessebsscddsaveccoscssvedccsas ccensodcts 23 3 4 Blue Tooth Favourites ss ee seen n nennen nn nh annuus tnnt nn nnns nnn 25 3 5 BE IN CnIIDESCT ER 29 3 6 Memiory OptiORhs rs eros Ira 22e esae ee so green naga GU 2b donate ob edeedcen ERR SUO Sepa A Oc eo ERR AEE ENAA 32 3 7 Restore Factory Settings ciccss icccsesscscccssessccsssssiccesieseccssssasccossessccosssascconsess Ee Papa ain 34 4 Automatic Mode osi heit Flare eR PRE Va ERE RERR YER EAM MERE RR NR Mn a Ea inst 35 4 1 Performing an Automatic Test ee eese eese eese ee ee eese eene eene n nonet nannte sanant nn a 36 4 2 Test n Tae ccm 40 4 3 Options EID 41 4 4 MIU CIIIEeEC notes te 42 5 Manual Modem denis 43 5 1 NIBP Test Settings cccisiccescisvicesstsiacvedessivcedssiesesdbsssacesssinansdshinucedsssannsdubsencees Tanase 44 5 2 NIBP Leak ee 46 5 3 NIBP Pop off TeStiiis ccc cadesvcdedicsecebedessccessgsedoaddsencdeisesaccsscaseccescesnccedde sncceasesesebedssacd 47 5 4 NIBP Manometer Mode cccccccssssssseeseceeececcececeseeeeeeseecececssseseesesseesaceseeseeeeseees 48 5 5 mln t esienanseseenss 49 6 View Datus eo phe cese a poop Re DRM Vli E nee terne Rae VEPASEAT TUR 50 6 1 MISI Cm E
50. test using the Up amp Down arrow keys and press the e Button F4 Visual Test fin OTOL A 3 1 3 Edit individual Test Press the G button F4 as shown above The Leak Test sub menu will allow you to alter the settings for Test duration and Pressure set point 15 Page Rigel Medical BP SIM user Manual V3 1 est Settings Pressure IEEE nmi Highlight the required field using the up and down keys To change the content use the keypad to enter the correct data To Delete the field content press the U shown below button as est Settings Pressure 275 mmHg When the right settings have been achieved press the V button F4 to save the changes or press X F3 to return to the previous menu without changes Repeat this action for every test that requires editing Once all the requires tests are programmed press x F3 to exit 3 1 4 Insert an individual test To insert an individual test we use the up and down keys to highlight the position where a new test needs inserting and press the Insert button F1 from the menu below Note Tests will be inserted before the highlighted position not after To modify a test sequence please refer to 3 1 3 16 Page Rigel Medical BP SIM user Manual V3 1 NIBP Leak Test Set Level 275 mmHg Duration 180s NIBP Pop off Test OOGA Once the Insert button F1 has been pressed a drop down box will appear with all available tests Activate the drop down box usi
51. ual V3 1 NIBP Leak Test 0 0 Start P 0 0 Timer The BP SIM will inflate the system to around the set point Once the system is stabilised at the set pressure the timer will count up for the set duration This is highlighted below NIBP Leak Test P Drop Start P mmHg 149 8 Timer At the end of the test the BP SIM will show e the start pressure ie 157 1 mmHg e the final pressure ie 149 9 mmHg e the pressure drop over 60 seconds ie 7 2 mmHg Press START F4 to repeat the test Press fat F2 to return to start menu 5 3 NIBP Pop off Test The Pop off test allows the user to verify the correct operation of the protection valve within the NIBP monitor Note Prior to initiating the test some NIBP monitors might have to be set to a service mode to ensure that the internal monitor valves are shut for incoming pressures If the monitor s internal valve is open the leak test will not be able to pressurise the system 47 Page Rigel Medical BP SIM user Manual V3 1 Select the test from the NIBP menu NIBP Pop off Test Set Level mmHg To initiate the test press START F4 The BP SIM will inflate the system until the protection valve is activated NIBP Pop off Test Set Level mmHg 272 7 At the end of the test the BP SIM will show the set point of the protection valve ie 331 mmHg Press START F4 to repeat the test fat F2 Press to return to start menu 5 4 NIBP Manometer
52. visible and easy to follow menu structures but also allows the user to operate the tester using intuitive icons to speed up their test routines Below are of some of the icons used in the Rigel BP SIM Escape OK PASS Print Add Less Copy Iu im Repeat 4 Delete Save Edit a Search Help ei Ter Settings Menu List fa tt Shift Home New Mute 4 Sound 9 Page Rigel Medical BP SIM user Manual V3 1 2 Logging on When switching on the Rigel BP SIM the user will be able to log in to allow for specific default user preferences and to provide traceability of test results To increase the security and protection of the user a password can be set prior to using the Rigel BP SIM See 3 5 3 for further details The default password setting is OFF and the login screen will not require a password In case a user and password are set the tester will default to the last user password User Name Enter Password XXXXXI 95672179 To change the user use the drop down menu provided and select the new user In case a password is set for the new user please enter the password case sensitive In case a password is lost and the unit is blocked please call our helpdesk on 44 0 191 5878701 or contact your local distributor and quote the 8 digit cod