Home

User`s Manual

image

Contents

1. e Purifier Logic Biological Safety Cabinets and PuriCare Lab Animal Research Stations carry a three year warranty from date of installation or four years from date of shipment from Labconco whichever is sooner SteamScrubber amp FlaskScrubber Glassware Washers carry a two year warranty from date of installation or three years from date of shipment from Labconco whichever is sooner Blood Drawing Chairs carry a ten year warranty Carts carry a lifetime warranty Glassware is not warranted from breakage when dropped or mishandled This limited warranty covers parts and labor but not transportation and insurance charges In the event of a warranty claim contact Labconco Corporation or the dealer who sold you the product If the cause is determined to be a manufacturing fault the dealer or Labconco Corporation will repair or replace all defective parts to restore the unit to operation Under no circumstances shall Labconco Corporation be liable for indirect consequential or special damages of any kind This statement may be altered by a specific published amendment No individual has authorization to alter the provisions of this warranty policy or its amendments Lamps and filters are not covered by this warranty Damage due to corrosion or accidental breakage is not covered Returned or Damaged Goods Do not return goods without the prior authorization from Labconco Unauthorized returns will not be accepted If your shipment
2. Carefully follow the instructions that follow and refer to Figure 3 1 and Figure 3 8 that depict views of the rear of the Gas Purifier where all the plumbing connections are made Furthermore refer to Figure 3 2 and Figure 3 8 that shows a typical plumbing connection of the rear of the glove box for the circulation lines 10 Product Service 1 800 522 7658 Chapter 3 Getting Started C A COM Port 6 pin on back of Gas Purifier for shared vacuum pump operation with glove boxes configured with Auto Pressure Control feature 3 Vacuum Pump Outlet for Auto Mode i glove boxes without auto Moisture Oxygen f pressure control Used for a 9 pin connector for dedicated vacuum pump connection to oxygen i or moisture monitor Only used for auto blower control 3 4 Inlet Outlet Compression Fittings for Inert Gas Circulation from the Glove Box 3 8 compression Exhaust Vent used during regeneration Vent Solenoid 3 8 compression 4 Hydrogen Inert Gas Mixture Inlet Flow Control Valve to set the 496 Hydrogen Inert Gas Mixture between 10 25 CFH 3 8 compression y Flow Direction Out d A Vacuum Pump Solenoid Valve Figure 3 1 Rear View of AtmosPure Re Gen Gas Purifier Product Service 1 800 522 7658 11 Chapter 3 Getting Started 3 4 Inlet Outlet Compression Fittings for Connection to Gas Purifier See Chapter 7 Accessories to order AtmosPure 3 4 to 3 8 Adapter Kits
3. 150 200 250 300 350 400 Time Minutes This chart shows oxygen level decrease from 980 2 ppm to 4 9 ppm after 400 minutes Two separate calibrated oxygen monitors confirmed results 52 Product Service 1 800 522 7658 450 Moisture PPM Appendix D Oxygen amp Moisture Removal Rates Precise CA polyethylene Moisture Removal with AtmosPure Re Gen Gas Purifier 1000 100 1 10 100 1000 Time Minutes 10000 This chart shows moisture level decrease from 660 2 ppm to 5 0 ppm after 8475 minutes Two separate calibrated moisture monitors confirmed results Product Service 1 800 522 7658 53 Appendix E Conversions Pressure Conversions e em e m oer nr nr From amis 07 3853 uisi 30152 00954 006 1385 000 Cin 254 1 1360 LIS 0342 4812 03485 3387 4339 Lk rs 5 soe ur osr ast G2 p i Lam 78 8 pre SRI T TON Conversions lit sec gal min ft sec ft min bbl hr bbl day From lit see 8 85 00035827 03532 2 29 22 2266 5438 8 CO 1 p ses imr nas ss fe a 43 1 4 41 SHRI 54 Product Service 1 800 522 7658 Appendix F References Many excellent reference texts and booklets are currently available The following is a brief listing Pharmaceutical Isolators A Guide to their application design and control Pharmaceutical Press 2004 Editors Midcalf Phillips Neiger and Coles Isolation Technology A Practical Guide 2004
4. 5000 ppm is established Loading Materials and Gas Purifier Set Up e Load only the materials and equipment required for the procedure Do not overload the glove box When loading ensure the inner door is closed before opening the outer door Then close the outer door Product Service 1 800 522 7658 31 Chapter 5 Using Your Gas Purifier Perform evacuation and fill cycle on the transfer chamber of the glove box using the vacuum outlet and gas inlet valves Depending on the vacuum level achieved it takes 1 5 cycles to reduce oxygen and moisture levels below 196 Porous materials being transferred may require longer sustained vacuum evacuations to remove the oxygen and moisture trapped in the materials In order to achieve the lowest oxygen and moisture levels 1n a glove box perform the required evacuation and fill cycles on the main chamber using the vacuum outlet and gas inlet valves To automate the evacuation and fill cycles 1 499 for both glove box chambers the addition of the Automatic Pressure Controller 5238600 or 5238601 1s recommended see Chapter 7 or use a Protector Controlled Atmosphere Glove Box with Auto Pressure Control Normally 75 100 evacuation and fill cycles will reduce moisture and oxygen levels below 196 An additional 100 cycles will reduce moisture and oxygen levels below 0 3 or 3 000 ppm The AtmosPure Gas Purifier blower can now be turned on The use of external glove port covers will expedite the evacuatio
5. CRC Press ud Edition www crcpress com International Standard ISO 10648 2 Containment Enclosures e Classification according to leak tightness and associated checking methods NIOSH Pocket Guide to Chemical Hazards U S Dept of Health and Human Resources Centers for Disease Control and Prevention National Institute for Occupational Safety and Health The Manipulation of Air Sensitive Compounds by D R Shriver and M A Drezdzon published by John Wiley and Sons Product Service 1 800 522 7658 55 Protecting your laboratory environment LABCONCO CORPORATION 816 333 8811 e 800 821 5525 8811 Prospect Avenue FAX 816 363 0130 Kansas City MO 64132 2696 www labconco com CE Declaration of Conformity Labconco Corporation 8811 Prospect Avenue Kansas City MO 64132 USA Ph 816 333 8811 Fax 816 363 0130 labconco labconco com being the representative party in the EU solely responsible for this certification hereby declare that the following product Equipment Laboratory Equipment AtmosPure Re Gen Gas Purifier Type Model 5121831 208 230V 50 60 Hz 5121835 208 230V 50 60 Hz 5121840 208 230V 50 60 Hz 5121845 208 230V 60 Hz only Manufacturer Labconco Corporation 8811 Prospect Avenue Kansas City MO 64132 USA Ph 816 333 8811 Fax 816 363 0130 labconco labconco com is in conformity with the standards listed below EN61010 1 EN61326 1 EN55022 EN61000 3 2 3 following th
6. Chapter 3 for materials that reduce Gas Purifier effectiveness DANGER The addition of oxygen inside the glove box must be avoided to prevent explosion death or serious injury Perchloric acid use in this Gas Purifier 1s prohibited Radioisotope material usage in the glove box and Gas Purifier should be approved by your Safety Officer A qualified certification technician should test the Gas Purifier with the glove box before it 1s initially used The Gas Purifier should be recertified at least annually after installation See Chapter 6 The use of safety goggles protective clothing and any other personal protective equipment recommended by your safety officer should be observed Exhaust vacuum pumps should be filtered and or ducted to the outside when using hazardous materials Consult your Safety Officer Manipulations that generate gases from toxic chemicals or radionuclides must be evaluated carefully from the standpoint of buildup to dangerous levels The decontamination and compliance with applicable regulations is the responsibility of the operator and safety officer Venting or filtering of vacuum pump exhaust is the operator s responsibility Ensure that the Gas Purifier and glove box 1s connected to electrical service in accordance with local and national electrical codes Failure to do so may create a fire or electrical hazard Do not remove or service any electrical components without first disconnecting the glove box
7. Gas Purifier 5 Replace the large nut and seal with pipe sealant compound 6 Install the outer service panel 7 Follow 13 hour Regeneration procedure from Chapter 3 8 Once regenerated the Gas Purifier blower can be run as soon as the glove box inert atmosphere is reduced to 3000 5000 ppm through 100 200 evacuation fill cycles 36 Product Service 1 800 522 7658 Chapter 7 Accessorizing Your Gas Purifier There are many ways to accessorize the Gas Purifier for your use with the glove box These include many different accessories listed in this chapter 1 Adjustable Height Base Stands with Work Surface Part 5235500 w casters Part 5235501 w adjustable feet These stands 60 x 31 with attached black laminate work surfaces have an adjustable height range from 33 to 40 838mm 1 016mm and include a lower shelf for supporting accessory equipment See Figure 1 1 2 AtmosPure Roughing Pre Filter Kit Part 5441500 Provides two roughing particulate pre filters and clamp plates to mount inside Protector Glove Boxes only It is used to filter the inlet and outlet of the glove box as the inert gas 1s purified 3 AtmosPure P100 Particulate Filter Kit Part 5441600 Provides two P100 particulate filters and mount housings that are similar to respirator cartridges It is used to filter the inlet and outlet of the glove box as the inert gas 1s purified 4 AtmosPure Regeneration Plumbing Kit Part 5441700 The regeneration plu
8. Product Service 1 800 522 7658 Chapter 3 Getting Started Start Stop Regeneration button to start and stop the Vacuum release button used momentarily for less than one second after the regeneration process is complete regeneration process Main Power ON OFF Switch Indicator Lights show each step of the 13 hour regeneration process Two 3 4 manually operated Isolation Ball Valves to isolate Gas Purifier from the glove box during regeneration Indicator Lights flash indicating proper position of the 3 4 manually operated isolation valves Flow Control Valve to set 10 25 CFH flow rate of 4 hydrogen inert gas mixture during regeneration Blower button switches Gas Purifier blower between on off and auto ON Blower on continuously OFF Blower off AUTO Blower controlled by oxygen or moisture monitors with high and low range limits enabled Accessory Oxygen Moisture Monitor Cable 9 pin is required Figure 3 7 AtmosPure Re Gen Control Panel Features Product Service 1 800 522 7658 19 Chapter 3 Getting Started Regenerating the Gas Purifier Catalysts The AtmosPure Re Gen Gas Purifier 1s initially shipped with the copper catalyst and molecular sieve fully regenerated and ready to use The copper catalyst can last 1 3 months with normal use The performance levels and frequency of regeneration are
9. can further reduce moisture below 10 ppm in 6 hours below 5 ppm in 16 hours below 2 3 ppm in 2 days and as low as 1 ppm in 3 6 days See Appendix D Internal stainless steel tubing has low permeation rates below 0 10 ppm min for both oxygen and moisture The sealed blower and circulation system assures leak tight operation for optimized performance The Gas Purifier has been tested on all three Labconco glove box liner materials stainless steel fiberglass FRP and polyethylene All three liners meet ISO 10648 2 test methods for oxygen permeation rates below 1 67 ppm min when tested at 4 inches of water gauge The lowest attainable oxygen levels are 0 2 ppm for stainless steel 0 4 ppm for FRP and 3 5 ppm for polyethylene The lowest attainable moisture Product Service 1 800 522 7658 25 Chapter 4 Features and Safety Precautions 7 8 9 levels are 1 ppm for stainless steel 2 ppm for FRP and 5 ppm for moisture The vacuum pump outlet on the back 1s rated at 115V 8A or 230V 4 5A Three solenoid valves control the regeneration of the copper catalyst and molecular sieve All are located on the back of the Gas Purifier for easy connection to the 4 hydrogen inert gas mixture vacuum pump and vent exhaust A flow control valve regulates the 4 hydrogen inert gas mixture between 10 25 CFH Two 3 4 diameter 1solation valves are closed during the regeneration process 1solating the catalysts from the glove box envir
10. ferrules properly and install the tube insert into the flexible line on both ends per Figure 3 3 Tighten the compression fittings on both ends 1 1 4 turns past finger tight second locate the vacuum pump and route the vacuum line from the vacuum pump to the vacuum solenoid valve located on the back of the AtmosPure Re Gen Gas Purifier per Figure 3 1 The vacuum line connection tubing is included in the Gas Purifier Regeneration Tubing Kit Again be sure to orient the ferrules properly install the tube inserts and tighten the compression fittings 1 1 4 turns past finger tight An alternate view for connection to the vacuum solenoid valve 1s provided in Figure 3 5 If one shared vacuum pump will be used and shared between the controlled atmosphere glove box and the Gas Purifier then the AtmosPure C A COM cable 6 pin must be connected between the Gas Purifier and the back of the glove box Auto Pressure Control refer to Figures 3 1 3 6 and 3 8 If shared with the controlled atmosphere glove box only with auto pressure control then plug the vacuum pump into the back of the glove box refer to Figure 3 6 If a dedicated vacuum pump will be used for the Gas Purifier plug the vacuum pump into the electrical outlet on the back of the AtmosPure Re Gen Gas Purifier Figure 3 1 Third the vent valve 1s exhausted through the 10 25 CFH flow control valve as viewed in Figure 3 1 The vent line must be exhausted to the outside or to a fume hood
11. for attachment to glove boxes with 3 8 inlet outlet The adapter kits allow for a tube size transition so 3 4 tubing can be run to the Gas Purifier Figure 3 2 Typical Plumbing Connection for the Circulation Lines 12 Product Service 1 800 522 7658 Chapter 3 Getting Started Connecting the Circulation Plumbing The circulation plumbing continuously circulates the inert gas atmosphere from the controlled atmosphere glove box and purifies it by removing oxygen and moisture The connections from the 3 4 inlet outlet compression fittings must be made to the glove box Depending on the neoprene lined tubing kit or stainless steel tubing kit selected route one line from the inlet of the Gas Purifier to the outlet of the glove box Refer to Figures 3 1 3 2 and 3 8 Next route the outlet of the Gas Purifier to the inlet of the glove box A continuous tubing loop for Gas Purifier circulation 1s now established Be sure to properly tighten all the 3 4 compression fittings at least 1 1 4 turns past finger tight as leaks will affect Gas Purifier performance The orientation of all the compression fittings must be properly set per Figure 3 3 Should your glove box have 3 8 compression fittings such as Labconco polyethylene glove boxes use the 3 4 to 3 8 adapter kits to route the circulated air through the neoprene lined hose kit Refer to Chapter 7 to order all circulation kits and adapter kits Note The Gas Purifier ships fu
12. from electrical service Product Service 1 800 522 7658 29 Chapter 4 Features and Safety Precautions 11 Ensure only trained operators use the Gas Purifier with the glove box New users must review the User s Manual and should be trained by operators familiar with glove box operation 12 Tag the Gas Purifier and glove box with appropriate warnings if work inside the glove box 1s hazardous or needs to remain untouched Tag the Gas Purifier and glove box if it requires servicing 13 Consult your Safety Officer before removing any hoses to vacuum pumps gas lines pressure gauges or exhaust hoses they may be contaminated with chemicals 30 Product Service 1 800 522 7658 Chapter 5 Using Your Gas Purifier Now that the installation of your Gas Purifier and glove box 1s completed you are ready to use it Read this chapter to learn about e Routine Daily Work Procedures e Hazardous Use with Chemicals e Prohibited Materials Use Routine Daily Work Procedures Planning e Thoroughly understand procedures and equipment required before beginning work Work from a written plan e Consult your Safety Officer for personal protective equipment recommendations and operating pressure confirmation Start up e Basic start up procedure includes turning on the vacuum pump opening the inert gas valve on the gas supply tank turning on the glove box light and turning on the blower to the Gas Purifier once an inert atmosphere below
13. the transfer chamber and locking the internal door the transfer chamber must be evacuated and filled approximately 1 5 times depending on vacuum level achieved This ensures the removal of any remaining airborne contaminants before the outer door is opened Never apply a high vacuum purge below 2 mercury pressure on sensitive electronic moisture and oxygen monitors The exact number of cycles should be determined by your Safety Officer Shutdown e To help preserve the controlled atmosphere leave the glove box main chamber under 1 to 1 water gauge pressure and continuously run the Gas Purifier blower The use of the accessory external glove port covers can assist 1n preserving the inert atmosphere by reducing permeation through the gloves e Keep the outer door closed when not in use Hazardous Use with Chemicals Hazardous chemicals may be used safely only when the glove box vacuum pump is properly vented and or filtered to the outside Consult your Safety Officer for proper venting and filtering of the vacuum pump exhaust Prohibited Materials Use Due to stainless steel components the Gas Purifier and glove box interior must not be exposed to mineral acid The Gas Purifier and glove box life spans will be limited with use of mineral acids Additionally materials referenced in Chapter 3 cannot be used such as alcohols halides halogens hydrazine heavy metals mercury arsine phosphene and sulfur containing compounds beca
14. used in AUTO mode Allows for automatic control of the AtmosPure blower from an accessory Oxygen or Moisture Monitor sold separately 9 Electrical 9 pin Sealed Pass Through Kit Part 5075619 Provides a leak tight connection for data transfer via the 5442100 Cable sold separately connected to the AtmosPure Re Gen Gas Purifier to an interior mounted Oxygen or Moisture Monitor Includes sealed 9 pin connector D subminiature with male pins on one side and female pins on the other and 10 cord Note For Precise polyethylene glove boxes or Pre 2001 Protector glove boxes order the 9 pin installation kit separately which includes all the drill bits and hole saws for field installation Part 5241100 10 Gas Inlet Vacuum Outlet Tubing Kits and Direct Drive Vacuum Pumps These kits provide the necessary flexible tubing and fittings to make connections to an inert gas supply tank and a rotary vane vacuum pump See Figure 3 8 Direct Drive Vacuum Pump 230V 50 60Hz NEMA 6 15P plug Direct Drive Vacuum Pump 230V 50 60Hz with reverse IEC plug 38 Product Service 1 800 522 7658 Chapter 7 Accessorizing Your Gas Purifier 11 Oxygen Monitors Oxygen monitors allow the user to continuously monitor oxygen levels inside the glove box Monitors are normally placed inside the main chamber to provide the best performance Never operate a transfer chamber vacuum evacuation on sensitive oxygen monitors as damage can occur 5244500 0 10 00
15. was damaged in transit you must file a claim directly with the freight carrier Labconco Corporation and its dealers are not responsible for shipping damages The United States Interstate Commerce Commission rules require that claims be filed with the delivery carrier within fifteen 15 days of delivery Limitation of Liability The disposal and or emission of substances used in connection with this equipment may be governed by various federal state or local regulations All users of this equipment are required to become familiar with any regulations that apply in the user s area concerning the dumping of waste materials in or upon water land or air and to comply with such regulations Labconco Corporation is held harmless with respect to user s compliance with such regulations Contacting Labconco Corporation If you have questions that are not addressed in this manual or if you need technical assistance contact Labconco s Customer Service Department or Labconco s Product Service Department at 1 800 821 5525 or 1 816 333 8811 between the hours of 7 00 a m and 6 00 p m Central Standard Time Part 45432400 Rev ECO F944 TABLE Or CONTENTS CHAPTER 1 INTRODUCTION General Description Circulation System Purifier System Regeneration Control CHAPTER 2 PREREQUISITES support Requirements Exhaust Requirements for the Regeneration Vent Valve Electrical Requirements Plumbing Requirements for Circulation 4 Hydrogen Iner
16. 0 ppm Oxygen Monitor with three ranges of 0 100 ppm 100 1 000 ppm and 1 000 10 000 ppm 100V 115V 50 60 Hz 5244501 0 10 000 ppm Oxygen Monitor with three ranges of 0 100 ppm 100 1 000 ppm and 1 000 10 000 ppm 230V 50 60 Hz International 5244400 0 1 100 Oxygen Monitor 1 000 ppm lowest 100V 115V 230V 50 60 Hz 12 Moisture Monitors Hygrometers Moisture monitors or hygrometers allow the user to continuously monitor moisture levels inside the glove box Moisture monitors are normally placed inside the main chamber to provide the best performance Never operate a transfer chamber vacuum evacuation on these sensitive hygrometers as damage can occur 5244900 1 1000 ppm Moisture Monitor Hygrometer Factory Set at 1 1 000 ppm with three settings of 1 10 ppm 1 100 ppm and 1 1 000 ppm 100V 115V 50 60 Hz 5244901 1 1000 ppm Moisture Monitor Hygrometer Factory Set at 1 1 000 ppm with three settings of 1 10 ppm 1 100 ppm and 1 1 000 ppm 230V 50 60 Hz International 5244800 0 100 Hand Held Digital Hygrometer with Temperature ME y meren oerda indy andtompetie 5442200 Moisture Meter Probe and Calibration Stand 80 to 0 C dew CH gohrrangeor sooo ppm Display snow penton 13 Exterior Monitor Plumbing Kits Part s below The exterior monitor plumbing kits provide attachment plumbing to oxygen or moisture monitors outside the glove box Note The control and accuracy of the oxygen and moisture mo
17. 2 1 e where normally only non conductive atmospheres are present Occasionally however a temporary conductivity caused by condensation must be expected in accordance with IEC 664 Product Service 1 800 522 7658 45 Appendix D AtmosPure Re Gen Gas Purifier Oxygen and Moisture Removal Rates Protector CA Stainless Steel Oxygen Removal with AtmosPure Re Gen Gas Purifier 1000000 210000 100000 10000 653 6 220 cycles stop and AtmosPure circulation blower starts and isolation valves are 1000 opened 100 Oxygen PPM 10 0 1 Time Minutes This chart shows oxygen level decrease from 210 000 ppm at atmospheric conditions to 0 7 ppm after 210 minutes Two separate calibrated oxygen monitors confirmed results 46 Product Service 1 800 522 7658 Moisture PPM Appendix D Oxygen amp Moisture Removal Rates Protector CA Stainless Steel Moisture Removal with AtmosPure Re Gen Gas Purifier 100000 10000 1000 100 100 Time Minutes 1000 10000 This chart shows moisture level decrease from 11 253 ppm 409 relative humidity to 1 0 ppm after 8965 minutes Two separate calibrated moisture monitors confirmed results Product Service 1 800 522 7658 47 Oxygen PPM 1000 100 0 1 Appendix D Oxygen amp Moisture Removal Rates Protector CA Stainless Steel Oxygen Removal with AtmosPure Re Gen Gas Purifier Time Minutes This chart shows oxygen level decrea
18. Auto Mode Moisture Oxygen Connector 9 pin for connection to the oxygen or moisture monitor relays Refer to Labconco accessories for oxygen monitors moisture monitors and the Oxygen Moisture Monitor Cable 9 pin The Gas Purifier inlet outlet connections are 3 4 compression fittings Connection accessory kits are available with either neoprene lined flexible hose or hard plumbed stainless steel tubing All glove boxes were tested with the neoprene lined hose An accessory 3 4 to 3 8 adapter kit 1s also available for connection to polyethylene glove boxes and pre 2011 fiberglass stainless steel glove boxes The Gas Purifier s compact design easily rests beneath the glove box on Labconco s base stands Dimensions are 14 9 W x 21 9 H x 31 8 D The Gas Purifier can be connected to one vacuum pump shared with the glove box and powered by the glove box vacuum pump outlet receptacle A controlled atmosphere communications port C A COM 6 pin on the back of the Gas Purifier allows for shared access of the vacuum pump when used with Controlled Atmosphere glove boxes with Auto Pressure Control A separate accessory C A COM Cable with 6 pin connectors is available for communication between the Gas Purifier and the Auto Pressure Controller of the glove box The cable must be used to share the one vacuum pump Product Service 1 800 522 7658 21 Chapter 4 Features and Safety Precautions Easily removed service panel 4 Hydrogen flow r
19. D 5235500 WCASTERS OR 5235501 WFEET ORDERED SEPARATE ROTARY VANE VACUUM PUMP 1472100 7739400 OR 7739402 ORDERED SEPARATE 40 Product Service 1 800 522 7658 Chapter 8 Troubleshooting Refer to the following table 1f your Gas Purifier box fails to operate properly If the suggested corrective actions do not solve your problem contact Labconco for additional assistance PROBLEM CAUSE CORRECTIVE ACTION AA ESA AAA Control panel lights won t Unit not plugged into outlet Plug unit into appropriate electrical service operate and turn main power switch on NURSES ANN A A NESPIMIHIMIMEEEMMMMZMZMME IIA REN Regeneration process not working A EEE Replace heater Catalysts poisoned Replace catalysts No power to exterior Unit not plugged in Plug unit into electrical service vacuum pump electrical receptacle Circuit breaker s tripped Reset or replace circuit breaker Gas Purifier blower won t Isolation valves closed Open manually operated isolation valves so operate blower can circulate inert gas and purify the gas Blower not running Check electrical connection and make sure terminals are connected Circulation blower not Replace blower For 230V replace blower working and transformer Product Service 1 800 522 7658 41 Appendix A Replacement Parts The components that are available for your Gas Purifier are listed The parts shown are the most commonly requested If other parts are requir
20. From the Regeneration Tubing Kit install the exhaust line to the flow control valve and route the tubing to the outside or to a fume hood Again properly install the tubing per Figure 3 3 Product Service 1 800 522 7658 15 Chapter 3 Getting Started Hose Connector Small Ferrule Compression Nut Note Hose Clamp not supplied Figure 3 5 Alternate View for Vacuum Connection 16 Product Service 1 800 522 7658 Chapter 3 Getting Started C A COM Port 6 pin on Back of Auto Pressure Control Glove Box Used for connection to AtmosPure Re Gen for single shared vacuum pump operation Vacuum Pump electrical plug on Auto Pressure Controller for single shared vacuum pump use Figure 3 6 Rear View of Auto Pressure Control on Glove Box Product Service 1 800 522 7658 17 Chapter 3 Getting Started Initial Start up of the Gas Purifier for Removal of Oxygen and Moisture Do not open the isolation valves Figure 3 4 until the inert atmosphere in the controlled atmosphere glove box main chamber has been reduced below 5000 ppm for both oxygen and moisture The Gas Purifier arrives fully regenerated and ready for use Usually 5000 ppm can be attained through 200 evacuation and fill cycles of inert gas into the glove box Once 5000 ppm or less 1s achieved the Gas Purifier 1solation valves may be opened and the blower turned on Simply turn the blower ON as identified on the control panel per F
21. M User s Manual AtmosPure Re Gen Gas Purifier Models Voltage Frequency Electrical Receptacle Plug Type 5121810 100 115V 50 60 Hz North America 5121831 208 230V 50 60 Hz British UK 5121835 208 230V 50 60 Hz Schuko 5121840 208 230V 50 60 Hz China Australia 5121845 208 230V 60 Hz only North America To receive important product updates complete your product registration card online at register labconco com Protecting your MN laboratory environment Labconco Corporation 8811 Prospect Avenue Kansas City MO 64132 2696 800 821 5525 816 333 8811 FAX 816 363 0130 E MAIL labconco g labconco com HOME PAGE www labconco com Please read the User s Manual before operating the equipment Copyright O 2010 Labconco Corporation All rights reserved The information contained in this manual and the accompanying products are copyrighted and all rights reserved by Labconco Corporation Labconco Corporation reserves the right to make periodic design changes without obligation to notify any person or entity of such change Warranty Labconco provides a warranty on all parts and factory workmanship The warranty includes areas of defective material and workmanship provided such defect results from normal and proper use of the equipment The warranty for all Labconco products will expire one year from date of installation or two years from date of shipment from Labconco whichever is sooner except the following
22. URGE 2 HRS 0N OPEN OPEN CLOSED Continues until manually VACUUM Suae CLOSED CLOSED OPEN operator pushing Start Stop button 8 hr min VACUUM Momentary by Pra ay NOTE The Gas Purifier has optimized programming of each function as shown in the above chart Electrical Power Outage The regeneration system automatically compensates for power outages If the outage 1s longer than 10 minutes the regeneration system will reset and start over 22 Product Service 1 800 522 7658 Chapter 3 Getting Started luide 3 09 JO UOW eunsiowjueB Axo UONDINIIID JONA S09 H E 18407 Buiqn uunno2bA UIDA QO L Buiqn uunn2DA JO 8 3 plOus OS JVA a piouajos juaA TX Jeuj Yeu uonp noJr5 Jang _ ZH 9epL s09 p seddn ybnoJy sspg Uld 6 palpas panpana PIOUS OS ZH isd 1 01 8JnjXlN S09 119U ueDoJp H 9er B E asn peJDus JOJ uonoeuuo 091119313 duung LUNNIDA uaDoup H 9egpjuoBuy H42 Sz ueboJpAH 9epjuaDOoJIIN H3O OZ USDOJP H 9eg uuni aH H42 0I H49 62 01 1D jas Jojo nbay MO y sdp 0 JMXIH SOS Lau 8 asn dund uunn2DA paJjoys Jo J uidg 31909 WOD Y 2 9eungsouujv Sjaju DJOD Jamog ajoJodas aaJu Figure 3 8 Connection Diagram for AtmosPure Gas Purifier and Controlled Atmosphere Glove Box Product Service 1 800 522 7658 23 Chapter 3 Getting Started Requirements Note Refer to Chapter 7 for proper accessories required for set up At a minimum
23. at and H2 purge cycles Step 6 After a minimum 8 hour vacuum cycle push the START STOP button to end regeneration Step 7 Turn off the vacuum pump then push the VAC REL button for one second or less Note An optional method releases the vacuum through the inert gas inside the glove box and enhances oxygen removal capacity Step 8 Optional Do not push the VAC REL button Instead partially open both isolation valves The isolation valves must be opened partially to slowly replace the vacuum inside the Gas Purifier with inert gas from the glove box As the glove box pressure drops due to the Gas Purifier vacuum refill the glove box with inert gas from the inert gas tank directly or use the Auto Pressure Control of the glove box Hold the glove box pressure between 1 w g to 4 w g Once the vacuum is fully released open the isolation valves 100 This method fully preserves the regenerated copper catalyst and molecular sieve Product Service 1 800 522 7658 21 Chapter 3 Getting Started Step 9 The regeneration procedure has now been completed The blower button needs to be turned on to begin circulating the inert gas in the glove box through the Gas Purifier AUTOMATIC REGENERATION PROGRAMMED FUNCTIONS SOLENOID VALVES o FUNCTION TIME HEATER 4 a VENT VACUUM H Set EA Purge manual adjustment of 4 2 MIN OPEN OPEN CLOSED hydrogen inert gas flow rate HEAT 3 HRS 0N CLOSED CLOSED CLOSED H P
24. conco also offers a 3 4 to 3 8 adapter elbow used to connect to 3 8 fittings found on polyethylene and older style glove boxes An AtmosPure Regeneration Tubing Kit should be selected from Chapter 7 Accessories as 1t provides all the plumbing to connect the 4 hydrogen inert gas mixture vacuum pump and exhaust vent A rotary vane vacuum pump must also be selected from Chapter 7 Accessories Rotary vane vacuum pumps establish a very deep vacuum below 10 microns that allows the copper catalyst and molecular sieve to be properly regenerated Chapter 3 the detailed Figures 1n Chapter 3 and Figure 3 8 show typical connections for all plumbing Space Requirements The dimensions are shown in Appendix B Dimensions Adequate space 1s provided to support the Gas Purifier on Labconco Adjustable Base Stands as listed in Chapter 7 Accessories The lower shelf of the adjustable base stand provides a convenient location to place the Gas Purifier on the right side below the glove box transfer chamber and out of the way of the operator Refer to Figure 1 1 in Chapter 1 Introduction Product Service 1 800 522 7658 7 Chapter 3 Getting Started Once the site for your Gas Purifier 1s properly prepared you are ready to unpack inspect and install your system Read this chapter to learn how to Unpack and move the Gas Purifier Set up the Gas Purifier with the proper supporting structure Connect the electrical supply Connect the plumbing lines
25. d perform one of the following tests Both tests validate the system Leak Test of the Gas Purifier connected to the glove box e Oxygen Method Leak Test per ISO 10648 2 test methods See glove box manual e Pressure Change Leak Test per ISO 10648 2 test methods See glove box manual Solenoid Valve Replacement l 2 Decontaminate the glove box interior and Gas Purifier 1f required Wear appropriate protective equipment as determined by your Safety Officer Turn the Gas Purifier and vacuum pump OFF Close the manually operated isolator valves Remove the solenoid valve Refer to Figure 3 1 that shows the solenoid valves Install the new solenoid valve Regenerate the catalysts and follow the 13 hour Regeneration listed in Chapter 3 Turn the glove box and Gas Purifier ON and re certify Catalyst Replacement le The catalysts can become ineffective for a number of reasons as described in Chapters 3 5 Turn the unit off and remove the outer service panel Loosen the large nut near the heater terminal connections on top of the Gas Purifier cylinder Refer to Figure 4 1 and Figure A 1 Remove the catalysts using a vacuum cleaner hose Catalysts are hazardous and need to be disposed of properly Fill the bottom with 4 pounds of molecular sieve followed by 5 pounds of copper catalyst and then followed by 1 pound of molecular sieve on top Product Service 1 800 522 7658 35 Chapter 6 Maintaining Your
26. dependent on several influencing factors including e Purity of inert gas being used and the amount of O2 H2O removal by evacuation and filling during initial glove box operation with the inert gas e Permeation rate of the glove box rubber gloves e Oxygen and moisture content of materials introduced into the glove box e Glove box liner material and its permeation rate e Frequency of use and applications that introduce oxygen and moisture into the inert atmosphere The regeneration procedure requires 13 hours and may be done overnight Refer to Figure 3 7 for control panel features make sure that the Gas Purifier exhaust vent is connected to safely discharge vapors from the regeneration process to the outside or to a fume hood The exhaust vent is attached to the flow control valve per Figure 3 1 qi WARNING Before starting the regeneration procedure Reseneration Mode isolation valves must be closed The regeneration process consists of a Hz Set cycle a Heat cycle a Hz Purge cycle using 4 hydrogen inert gas mixture and a Vacuum cycle that takes 13 hours and can be done overnight Moisture is driven from the molecular sieve by the heat Oxygen 1s removed from the heated copper catalyst by the hydrogen combining with the copper oxide which forms H20 After the heat and purge cycles the vacuum cycle removes the residual gas and moisture from the purifier The Hz set cycle allows the operator to set the 4 hydrogen iner
27. e 2 hour 4 hydrogen inert gas mixture flows through purifier heat also remains on e Vacuum cycle 8 hours minimum continues until manually shut off and red LED flashes The control panel contains push button switches for automatic cycling Ho SET HEAT H2 PURGE VACUUM The lighted display shows which stage the regeneration process is 1n NOTE Refer to Chapter 3 Chapter 4 and Chapter 5 for more detailed descriptions of performance features control functions operating instructions and safety features 4 Product Service 1 800 522 7658 Chapter 2 Prerequisites Before you install the Gas Purifier to the controlled atmosphere glove box you need to prepare your site for installation You must be certain that the area 1s level and of solid construction In addition a dedicated source of electrical power should be located near the installation site to power the glove box and the Gas Purifier Additionally the glove box and accessory Gas Purifier should be strategically placed in the lab to provide efficient workflow and prevent operator interference from normal traffic flow Carefully read this chapter to learn the requirements for your installation site e The support and location requirements e The exhaust requirements for the regeneration vent valve e The electrical power requirements e The plumbing requirements for the circulation system of the Gas Purifier the regeneration 4 hydrogen inert gas mixture
28. e STP Depending on the application glove box liner material chosen and frequency of use regeneration may be necessary only once every 1 3 months and take only 13 hours to complete the process The AtmosPure Re Gen Gas Purifier offers many unique features to enhance performance with use of controlled atmosphere glove boxes To take full advantage of all the features you must read and understand this manual and keep it handy for future reference during regeneration AtmosPure Re Gen Gas Purifier shown with Labconco Protector Controlled Atmosphere Glove Box with Auto Pressure Control mounted on an Adjustable Height Base Stand Figure 1 1 2 Product Service 1 800 522 7658 Chapter 1 Introduction General Description The AtmosPure Re Gen Gas Purifier consists of three distinct systems A Circulation System B Purifier System C Regeneration Control Circulation System The circulation blower continually circulates the inert gas inside the glove box through the Gas Purifier to remove oxygen and moisture As the blower circulates the glove box inert atmosphere through the purifier column water molecules are adsorbed by the molecular sieve Molecular sieve 1s a synthetic zeolite that adsorbs all molecules less than 10 angstroms in size Oxygen molecules are removed from the inert gas by reacting with the copper catalyst and forming copper oxide Purifier System A six inch diameter stainless steel column contains 5 lbs of
29. e for the vacuum with inert gas Set the glove box main chamber pressure between 1 w g to 4 w g for this preferred method This method is preferred because it maximizes the capacity of the Gas Purifier and does not allow atmospheric air to deplete the Gas Purifier column Note Atmospheric air contains 21 oxygen 210 000 ppm 40 relative humidity air contains 11 250 ppm moisture e Main power switch turns the inlet power to the Gas Purifier on and off 10 The optimized regeneration process is completed in 13 hours and 26 may be completed overnight while unattended Product Service 1 800 522 7658 11 12 13 14 15 16 175 Chapter 4 Features and Safety Precautions The Gas Purifier automatically compensates for power outages and restarts the regeneration process 1f interrupted for 10 minutes or longer This feature assures complete regeneration of the copper catalyst and molecular sieve as the system would cool and not regenerate properly Leak tight Each Gas Purifier has been factory tested with a helium mass spectrometer and found to have no detectable leaks greater than 1 x 10 cc sec Each Gas Purifier ships fully regenerated and 1s ready to use The Gas Purifier has the capacity to remove 13 500 ppm oxygen and 18 million ppm moisture The Gas Purifier may be programmed by selecting auto mode and connecting it to an accessory oxygen or moisture monitor The back of the Gas Purifier has an
30. e provisions of the following directives 89 336 EEC 2004 108 EC 2006 95 EC when installed and operated in accordance with the manufacturers installation and operating instructions Mark Schmitz Place and date of issue Vice President Research and Engineering Labconco Corporation Labconco P N 3696076 Rev 56 Product Service 1 800 522 7658
31. ed please contact Product Service Item Qty Part Description Number po Electrical Components l 2 A 1 130700 Main Power Switch DPST 2 POS _ 3B 2 132704 CircuitBreaker SA S 4A 1 f 1305800 Power Cord 100 115V N America 4B 1 f 1332600 PowerCord 208 230V British UK 4C 1 1336100 Power Cord 208 230V Schuko 4D 1 133270 PowerCord 208 230V China Australia 4E 1 133800 Power Cord 208 230V N America 5A 1 1289101 Relay Panel Mount 115V Coil EN A Gas Purifier Components l O Ring Receptacle an O NUM AA Regeneration Components _______________ 14A 1 5431500 Re Gen Sub Assy w heater and catalysts complete 115V 14B A41 9 5431501 Re Gen Sub Assy w heater and catalysts complete 230V 15A 1 5118300 Heater Top Weld Assy UISV ISB 1 f 5118301 Heater Top Weld Assy 230V 16 1 J 5117300 Gasket Cylinder Tp gt VA l 141470 Plug Cylinder Top access to load catalysts 21A 3 5417602 _ Solenoid Valve Assy w Back Pressure 115V 2IB 3 5417603 Solenoid Valve Assy w Back Pressure 230V 22 1953100 Flow Regulator Valve from vent solenoid to control 4 hydrogen inert gas mixture flow rate Plumbing Components 23 2 5432200 Bulkhead Union 3 4 Compression w ferrules and nu
32. egulator valve set between 10 25 CFH Isolation Ball Valves Manually operated for control CONTROL PANEL INDICATOR LIGHTS AND TOUCH CONTROL BUTTONS 13 HEATER TERMINAL CONNECTIONS CIRCUIT BOARD W MICROPROCESSOR GLOVE BOX INERT GAS IN TRANSFORMER 230V ONLY GAS PURIFIER INLET GAS PURIFIER OUTLET 4 HYDROGEN IN SOLENOID 12326 BLOWER POWER INLET VENT OUT SOLENOID 8 SK VACUUM OUT SOLENOID 12 PURIFIED INERT GAS OUT Figure 4 1 Features 28 Product Service 1 800 522 7658 Chapter 4 Features and Safety Precautions Safety Precautions l 10 It is the responsibility of the user to determine the suitability of this product for the intended applications Consult your Safety Officer for application review This product is not designed or intended to be explosion proof It is the responsibility of the user to determine and avoid the lower explosive limits and flammability of the enclosed gases used within the controlled atmosphere glove box The user is responsible for taking proper precautions to prevent equipment damage or injury due to an explosion or combustion of chemicals The use of flammable gases or solvents in the glove box should be limited as these gases are exposed to the Gas Purifier Care must be taken to ensure against the concentration buildup of flammable or explosive gases or vapors These vapors can potentially poison the Gas Purifier catalysts and reduce their efficiency See
33. for circulation 4 hydrogen inert gas mixture vacuum and exhaust vent Start up the Gas Purifier for removal of oxygen and moisture Regenerate the Gas Purifier catalysts after 1 3 months of use shipping container allows for lifting with a mechanical lift qi Each Gas Purifier model weighs 75 Ibs each 34 kg The truck or floor jack If you must lift the Gas Purifier manually follow safe lifting guidelines Do not lift by the upper sheet metal facade and front panel as damage can occur Unpacking the Gas Purifier We recommend that you do not remove the Gas Purifier from its shipping container until it is ready to be placed into its final location Carefully remove the shrink wrap on the Gas Purifier and inspect it for damage that may have occurred in transit If damaged notify the delivery carrier immediately and retain the entire shipment intact for inspection by the carrier Product Service 1 800 522 7658 Chapter 3 Getting Started THE UNITED STATES INTERSTATE COMMERCE COMMISSION RULES REQUIRE THAT CLAIMS BE FILED WITH THE DELIVERY CARRIER WITHIN FIFTEEN 15 DAYS OF DELIVERY DO NOT RETURN GOODS WITHOUT THE PRIOR AUTHORIZATION OF LABCONCO UNAUTHORIZED RETURNS WILL NOT BE ACCEPTED IF GAS PURIFIER WAS DAMAGED IN TRANSIT YOU MUST FILE A CLAIM DIRECTLY WITH THE FREIGHT CARRIER LABCONCO CORPORATION AND ITS DEALERS ARE NOT RESPONSIBLE FOR SHIPPING DAMAGES i8 8 18 Do not discard the packing material un
34. igure 3 7 Once turned ON the blower circulates inert gas air from the glove box through the Gas Purifier column oxygen and moisture will start to drop immediately The button switches the Gas Purifier blower between ON OFF and AUTO ON operates the blower continuously OFF shuts the blower down and AUTO turns the blower on and off if an oxygen or moisture monitor control range is desired If AUTO 1s chosen the oxygen or moisture monitor available from Chapter 7 must be programmed and the monitor automatically controls the blower function The oxygen or moisture monitor should be placed inside the glove box for best results The accessory Oxygen Moisture Monitor Cable 9 pin and Electrical 9 pin Sealed Pass Through must be installed between the monitor oxygen or moisture and the Gas Purifier Please refer to Figure 3 1 Figure 3 8 and Chapter 7 for accessories and connections for AUTO mode control of the sealed circulation blower inside the Gas Purifier CAUTION Some chemicals used in a glove box will poison the copper catalyst inside the Gas Purifier and reduce the oxygen removal capability These chemicals should not be used in the glove box in order to avoid replacing the copper catalyst and molecular sieve 1 Sulfur containing compounds hydrogen sulfide carbonyl sulfide disulfides sulfur dioxide sulfur trioxide Halides and halogens Freon Alchohols Hydrazine Heavy Metals Mercury Arsine Phospene wd I ES 18
35. lly regenerated and ready qi to use Keep the isolation valves manually operated closed until ready to use The isolation valves Figure 3 4 must be closed until the glove box inert vas atmosphere is below 5000 ppm Opening the isolation valves before the Gas Purifier and glove box are ready for operation will cause a loss in oxygen and moisture removal capacity Tube Insert only used with flexible lines for support during ferrule compression on the tube Large Ferrule A Small Ferrule Gey Ox Compression Nut Figure 3 3 Typical Orientation of Ferrules Product Service 1 800 522 7658 13 Chapter 3 Getting Started Isolation Valves manually operated e Closed initially and upon regeneration only Opened for normal mode circulation when inert gas atmosphere is below 5000 ppm Figure 3 4 Isolation Valves 14 Product Service 1 800 522 7658 Chapter 3 Getting Started Connecting the Regeneration Plumbing The Regeneration Plumbing allows the copper catalyst and molecular sieve to be regenerated as needed The regeneration process takes 13 hours and may be completed overnight Refer to Figures 3 1 3 3 and 3 8 First locate the tank of 4 hydrogen inert gas mixture Using the AtmosPure Regeneration Tubing Kit route the tubing from the 4 hydrogen gas mixture pressure regulator to the valve inlet per Figures 3 1 and 3 8 The 4 hydrogen mixture provides the gas to restore the copper catalyst Be sure to orient the
36. mbing kit includes flexible tubing and fittings to make all the regeneration tubing connections only These are the three connections to the 4 hydrogen inert gas mixture vacuum and vent 5 AtmosPure Circulation Kits see Part s below The circulation kits include all the plumbing to make connection from the Gas Purifier to the glove box There are options for neoprene lined hose and hard ducted stainless steel 5441800 3 4 I D Neoprene lined hose All Glove Boxes single width 5441801 3 4 I D Neoprene lined hose All Glove Boxes double width 5441802 3 4 Rigid Stainless Steel Protector Glove Boxes single width 5441803 3 4 Rigid Stainless Steel Protector Glove Boxes double width Product Service 1 800 522 7658 37 Chapter 7 Accessorizing Your Gas Purifier 6 AtmosPure 3 4 to 3 8 Adapter Kit See Part s below Fittings permit 3 4 ID neoprene hose included with 5441800 and 5441801 Kits sold separately to attach to glove box s 3 8 inlet outlets Two fittings included with each kit Not required on current Protector Glove Boxes manufactured after November 2010 5441900 Precise Glove Boxes 5441901 Pre 2011 Protector Glove Boxes 7 AtmosPure 6 pin Communication Cable Part 5442000 Required for use when AtmosPure and glove box configured with Auto Pressure Controller are sharing one vacuum pump 8 AtmosPure 9 pin Oxygen Moisture Monitor Cable Part 5442100 Required for use when the AtmosPure is
37. molecular sieve for H20 removal and 5 lbs of a reactant copper catalyst material for O2 removal The purifier 1s equipped with components necessary for regeneration of the molecular sieve and copper scavenging media Attached to the column are e Two isolation valves manually operated These valves are kept open during circulation of the glove box gas through the purifier and MUST BE CLOSED during the regeneration procedure e Three solenoid valves for connecting 4 hydrogen inert gas mixture exhaust vent and vacuum These are located at the back of unit e A flow meter for 10 25 CFH regulation of the 4 hydrogen inert gas mixture during regeneration e A 180 watt electric heating coil imbedded in the purifier media The heating coil 1s used in regeneration of the catalysts e Ceramic fiber high temperature insulation e Anexterior metal case to protect all the components Product Service 1 800 522 7658 3 Chapter 1 Introduction Regeneration Control Regeneration is controlled by a microprocessor which automatically sequences the solenoid valves during the regeneration cycle The regeneration cycle restores the molecular sieve and copper catalyst It is programmed for the following time functions e Initial 2 minute purge to allow adjustment of 4 hydrogen inert gas mixture flow rate to 10 25 CFH 10 CFH Helium 4 Hydrogen 20 CFH Nitrogen 4 Hydrogen 25 CFH Argon 4 Hydrogen e Heat cycle 3 hours e Purge cycl
38. n and fill cycle time and help maintain low moisture and oxygen levels when the glove box is not in use Maintaining a 1 to 1 water gauge pressure difference can help to preserve the glove box atmosphere The AtmosPure Re Gen Gas Purifier is required to maintain oxygen and moisture levels below 0 3 0 5 Turn the Gas Purifier blower on to start circulation and reduce oxygen and moisture Labconco AtmosPure Re Gen Gas Purifiers can maintain oxygen levels below 1 ppm and moisture levels below 1 ppm depending on liner material and applications Monitoring moisture and oxygen levels inside the glove box 1s recommended by using accessory monitors see Chapter 7 The Gas Purifier catalysts will last for 1 3 months before requiring regeneration See Chapter 3 for instructions to start regeneration Work Techniques 32 Maintain work at a slightly negative pressure 1f working with hazardous compounds Continuously run the Gas Purifier blower to preserve the inert atmosphere 1f so desired Product Service 1 800 522 7658 Chapter 5 Using Your Gas Purifier Transferring Materials to the Outside e All open chemical weigh vessels or containers must be sealed before removing materials from the glove box main chamber e Objects in contact with previously contaminated internal atmospheres must be surface decontaminated or bagged before removal from the glove box through the transfer chamber e After placing materials to be removed in
39. nitors will be reduced due to the surrounding ambient air rather than inert gas inside the glove box Descriptions Used With 5442400 Plumbing Kit for Oxygen Monitor 0 to 10 000 ppm 5244500 01 5442500 Plumbing Kit for Monitor Hygrometer with pump 0 to 10 000 ppm 5244900 01 5442300 Plumbing Kit for Moisture Meter Probe 80 to 0 C or 0 to 6000 ppm 5442200 14 Automatic Pressure Controller for use with Precise polyethylene and manual control Protector Glove Boxes The Pressure Controller provides the operator the convenience of automatically regulating pressure and vacuum within the glove box main chamber and transfer chamber The end user can also automate Product Service 1 800 522 7658 39 Chapter 7 Accessorizing Your Gas Purifier the evacuation fill cycles up to 499 cycles for the main chamber and up to 199 cycles on the transfer chamber The vacuum and pressure can be automatically regulated between 5 to 5 inches of water gauge in the main chamber and between 0 to 29 inches of mercury gauge in the transfer chamber See illustration below and figures in Chapter 3 Electrical Configuration 5238600 100 115 Volts 50 60 Hz operation 5238601 230 Volts 50 60 Hz operation 5245200 TUBING KIT FOR AUTOMATIC OPERATION ORDERED SEPARATE INERT GAS TANK AUTOMATIC PRESSURE CONTROLLER AND REGULATOR 5238600 OR 5238601 D IDE CUSTOMER SUPPLIED ORDERED SEPARATE RE ADJUSTABLE HEIGHT BASE STAN
40. oisture inside the controlled atmosphere glove box The Gas Purifier enables the user to work in an inert gas environment with materials that are sensitive to oxygen and moisture The Gas Purifier comes fully regenerated and ready to use after the main chamber oxygen and moisture levels are reduced to 3000 5000ppm using evacuation and re fills of an inert gas The Gas Purifier has been tested with all three controlled atmosphere glove box liner materials manufactured by Labconco stainless steel fiberglass FRP and polyethylene All three Labconco controlled atmosphere glove boxes meet or exceed ISO 10648 2 standards when used with the Gas Purifier The test results are as follows Glove Gas Port Permeant Covers Precise polyethylene Protector FRP Attainable O Installed 3 5 0 4 Purity H O Installed 5 2 ppm O5 None 4 5 0 9 H50 None 5 3 Oxygen Permeation Oz None 0 60 0 30 at 4 w g per ISO 10648 2 Test Method ppm min ISO 10648 2 O5 None 1 67 1 67 Acceptance Level for Class 1 Controlled Atmosphere Glove Box ppm min Normal Permeation O None 0 33 0 16 with pressure from H O None 0 21 0 17 1 w g to 1 w g ppm min Product Service 1 800 522 7658 Protector stainless 0 2 1 0 8 2 0 16 1 67 0 14 0 10 Chapter 1 Introduction The Gas Purifier has the capacity to remove up to 5 liters of oxygen 13 500ppm and 660 grams of moisture 18 million ppm at standard temperature and pressur
41. onment After the 13 hour regeneration process is complete the 1solation valves are slowly opened up to relieve the vacuum in the column and return the Gas Purifier back to normal circulation mode Control Panel features include a Blower button controls the Gas Purifier blower between on off and auto In the ON position the blower runs continuously OFF shuts the blower off and AUTO controls blower with pre programmed auto mode which can be adjusted using control settings on the Labconco oxygen moisture monitors An accessory Oxygen Moisture Monitor Cable 9 pin must be used to control the Gas Purifier to high or low limit settings of the oxygen moisture monitors b Flashing indicator lights show proper positioning of the 3 4 manually operated isolation valves c Start Stop Regeneration button starts the programmed 13 hour regeneration process The isolation valves must be closed Indicator lights show each step of the regeneration process Once the regeneration process 1s complete the red indicator light will flash indicating successful program completion The Start Stop button is pushed to end the process d The Vacuum Release button can be used to momentarily release the vacuum in the purifier column after regeneration 1s complete However Labconco recommends operators should release the column vacuum through the glove box chamber by slowly opening the isolation valves and allowing the auto pressure controller to compensat
42. se from 753 6 ppm to 0 7 ppm after 150 minutes Two separate calibrated oxygen monitors confirmed results 48 Product Service 1 800 522 7658 Moisture PPM Appendix D Oxygen amp Moisture Removal Rates Protector CA Stainless Steel Moisture Removal with AtmosPure Re Gen Gas Purifier 1000 100 1 10 100 1000 Time Minutes 10000 This chart shows moisture level decrease from 853 2 ppm to 1 0 ppm after 8905 minutes Two separate calibrated moisture monitors confirmed results Product Service 1 800 522 7658 49 Oxygen PPM 1000 100 10 0 1 Appendix D Oxygen amp Moisture Removal Rates Protector CA Fiberglass Oxygen Removal with AtmosPure Re Gen Gas Purifier Time Minutes This chart shows oxygen level decrease from 969 2 ppm to 0 77 ppm after 150 minutes Two separate calibrated oxygen monitors confirmed results 50 Product Service 1 800 522 7658 Moisture PPM Appendix D Oxygen amp Moisture Removal Rates Protector CA Fiberglass Moisture Removal with AtmosPure Re Gen Gas Purifier 1000 100 1 10 100 1000 Time Minutes 10000 This chart shows moisture level decrease from 600 ppm to 2 3 ppm after 9873 minutes Two separate calibrated moisture monitors confirmed results Product Service 1 800 522 7658 51 Oxygen PPM 1000 100 Appendix D Oxygen amp Moisture Removal Rates Precise CA polyethylene Oxygen Removal with AtmosPure Re Gen Gas Purifier 50 100
43. sphere glove box The Gas Purifier 1s rated 100 115V 10A or 230V SA One exterior outlet for a vacuum pump connection is located on the rear of the Gas Purifier and is rated for 100 115V 8A or 230V 4 5A A single shared vacuum pump can be used for glove boxes configured with auto pressure control please see Chapter 7 Accessories to order the AtmosPure C A COM Cable 6 pin Accessory oxygen or moisture monitors can be programmed to provide automatic control of the Gas Purifier s blower Please see Chapter 7 Accessories to order the AtmosPure Oxygen Moisture Monitor Cable 9 pin Oxygen or Moisture Monitor and Sealed Electrical 9 pin Pass Through Chapter 3 details all the connections for proper start up Figure 3 8 shows a typical connection diagram for electrical connections 6 Product Service 1 800 522 7658 Chapter 2 Prerequisites Plumbing Requirements for Circulation 4 Hydrogen Inert Gas Mixture Vacuum and Exhaust Vent The Gas Purifier has two 3 4 compression fittings for connecting the inlet outlet of the circulation system of the Gas Purifier to the glove box inert gas and three solenoid valves with 3 8 compression fittings for connection to the regeneration system Regeneration Control Refer to Chapter 3 Getting Started and Chapter 7 Accessories to select the proper Circulation Tubing Kit for attachment of the AtmosPure Re Gen Gas Purifier Labconco offers neoprene lined tubing kits and stainless steel tubing kits Lab
44. t used to route circulation to glove box Can Do E E T NE NE 42 Product Service 1 800 522 7658 Appendix A Replacement Parts 12 13 17 18 amp 19 Figure A 1 Replacement Parts for AtmosPure Re Gen Gas Purifier 43 Product Service 1 800 522 7658 Appendix B Dimensions The following dimensions are for all the Gas Purifiers All dimensions in inches and metric 1306 33l 7mm 12 00 304 8mml 318 18077 mml 75 190 5rmmi 90 228 6mml Figure B 1 44 Product Service 1 800 522 7658 Appendix C Specifications This Appendix contains technical information about all the Gas Purifiers including electrical specifications and environmental operating conditions e 10 Amps 100 115V or 5 Amps 230V 50 60 Hz AtmosPure Re Gen Gas Purifier Environmental Conditions e Indoor use only e Maximum altitude 12 000 feet 3 657 meters e Ambient temperature range 41 to 104 F 5 to 40 C e Maximum relative humidity 80 for temperatures up to 88 F 31 C decreasing linearly to 50 relative humidity at 104 F 40 C e Main supply voltage fluctuations not to exceed 10 of the nominal voltage e Transient over voltages according to Installation Categories II Over voltage Categories per IEC 1010 Temporary voltage spikes on the AC input line that may be as high as 1500V for 100 115V models and 2500V for 230V models are allowed e Used in an environment of Pollution degrees
45. t Gas Mixture Vacuum and Exhaust Vent Space Requirements CHAPTER 3 GETTING STARTED Unpacking the Gas Purifier Installing the Gas Purifier on a Supporting Structure amp Work Surface Connecting the Electrical Supply Source to the Gas Purifier Connect the Plumbing Lines for Circulation 4 Hydrogen Inert Gas Mixture Vacuum and Exhaust Vent Connecting the Circulation Plumbing Connecting the Regeneration Plumbing Initial Start up of the Gas Purifier for Removal of Oxygen and Moisture Regenerating the Gas Purifier Catalysts CHAPTER 4 FEATURES AND SAFETY PRECAUTIONS Gas Purifier Performance Features safety Precautions CHAPTER 5 USING YOUR GAS PURIFIER Routine Daily Work Procedures Hazardous Use with Chemicals Prohibited Materials Use BR UU Y me NNN t 10 15 18 20 25 25 29 31 33 33 CHAPTER 6 MAINTAINING YOUR GAS PURIFIER Routine Maintenance Schedule Initial Certification Recertification Solenoid Valve Replacement Catalyst Replacement CHAPTER 7 ACCESSORIZING YOUR GAS PURIFIER CHAPTER 8 TROUBLESHOOTING APPENDIX A REPLACEMENT PARTS APPENDIX B DIMENSIONS APPENDIX C SPECIFICATIONS APPENDIX D ATMOSPURE RE GEN GAS PURIFIER OXYGEN amp MOISTURE REMOVAL RATES APPENDIX E CONVERSIONS APPENDIX F REFERENCES DECLARATION OF CONFORMITY 4 42 44 45 46 54 55 56 Chapter 1 Introduction The AtmosPure Re Gen Gas Purifier reduces the levels of oxygen and m
46. t gas mixture flow rate with the flow control valve Figures 3 1 and 3 7 as follows 4 hydrogen 96 helium 10 CFH 4 hydrogen 96 nitrogen 20 CFH 4 hydrogen 96 argon 25 CFH 20 Product Service 1 800 522 7658 Chapter 3 Getting Started Regeneration Steps Step 1 Check to see that the 4 hydrogen inert gas mixture tank valve 1s open and its pressure regulator is set between 10 14 psi At least Y a tank of 4 hydrogen inert gas must be available Also be sure that the vacuum pump electrical power cord is plugged in Step 2 Close both isolation valves CAUTION Both isolation valves must remain closed until the entire regeneration cycle is completed and vacuum is released Step 3 Push the automatic START STOP button on the AtmosPure control panel see Figure 3 7 to begin automatic regeneration Step 4 Adjust the Hydrogen Purge Flow Rate Knob to provide CFH reading as specified on the back of the Gas Purifier and listed in this instruction manual on the previous page If proper flow cannot be set within the two minute period push Start Stop switch to cancel program Return to Step 3 when you are ready Regeneration will begin with the heat cycle after the two minute flow adjustment period After the 3 hour initial heat cycle the 2 hour H purge cycle will begin automatically Step 5 With the vacuum pump on and properly connected the vacuum pump will run for at least eight hours after completion of the he
47. til you have checked all of the components and tested the Gas Purifier with the controlled atmosphere glove box Installing the Gas Purifier on a Supporting Structure and Work Surface Exercise caution when lifting or moving the Gas Purifier Do not lift the Gas Purifier using the isolation valve handles When installing the Gas Purifier onto a surface ensure that the structure can safely support the weight of the Gas Purifier and any related equipment such as the vacuum pump The Labconco base stands from Chapter 7 offer a lower shelf with optimized placement of the Gas Purifier on the right side under the transfer chamber Refer to Figure 1 1 in Chapter 1 and Figure 3 8 in Chapter 3 The lower shelf can also support the rotary vane vacuum pump accessory as shown in Figure 3 8 Connecting the Electrical Supply Source to the Gas Purifier Connect the power cord supplied to the IEC electrical supply plug located on the back of the Gas Purifier 100 115V models are rated at 10 amps total 230V models are rated at 5 amps total If a dedicated vacuum pump is used for the Gas Purifier one electrical outlet 1s located on the rear of the Gas Purifier for vacuum pump use and is rated at 100 115V 8A or 230V 4 5A Product Service 1 800 522 7658 9 Chapter 3 Getting Started Connect the Plumbing Lines for Circulation 49 Hydrogen Inert Gas Mixture Vacuum and Exhaust Vent All the plumbing lines to the Gas Purifier must be connected properly
48. use they will poison the Gas Purifier catalysts and reduce their effectivity and lifespan NOTE Should the catalysts inside the Gas Purifier become ineffective after two regeneration cycles the two catalysts can be replaced See Chapter 6 Maintenance and Appendix A Replacement Parts for replacement Product Service 1 800 522 7658 33 Chapter 6 Maintaining Your Gas Purifier Review this chapter on maintenance for the following Routine maintenance Initial certification Recertification Solenoid valve replacement Catalyst replacement ic decem Routine Maintenance Schedule Weekly e Monitor and record oxygen and moisture levels inside the glove box with accessory monitors see Chapter 7 1 3 Months e Regenerate the catalysts every 1 3 months depending on your application glove box liner material and frequency of use Initial Certification Each Gas Purifier has been leak tested at the factory with helium gas at positive 5 w g pressure no leaks greater than 1 x 10 cc sec are allowable The Gas Purifier and glove box should be certified by your Safety Officer before use with one of the two tests listed below under Recertification 34 Product Service 1 800 522 7658 Chapter 6 Maintaining Your Gas Purifier Recertification Under normal operating conditions the Gas Purifier and glove box system should be recertified at the time of installation and at least annually thereafter The certifier should choose an
49. vacuum and exhaust vent e The space requirements Refer to Appendix B Gas Purifier Dimensions Refer to Appendix C Gas Purifier Specifications for complete electrical and environmental conditions specifications and requirements Product Service 1 800 522 7658 5 Chapter 2 Prerequisites Support Requirements In the preparation of installing the Gas Purifier at the glove box site please consider the following e A bench or stand that rests on the floor or fixed to the wall but not both may be appropriate 35 to 40 889mm to 1 016mm is typical for standing height Labconco stands offered in Chapter 7 vary from 33 to 40 or 838mm to 1 016mm The stands Labconco offers include a convenient place on the right side shelf to support the AtmosPure Re Gen Gas Purifier See Figure 1 1 in Chapter 1 Exhaust Requirements for the Regeneration Vent Valve The vent valve exhausts 4 hydrogen mert gas during the regeneration process and must be vented safely to the outside or exhausted to a fume hood The regeneration vent valve 1s exhausted through the gas flow regulator An open compression fitting 1s attached to the gas flow regulator and can be used to connect the exhaust hose from the Regeneration Plumbing Kit Refer to Chapter 7 to purchase the Regeneration Plumbing Kit and Chapter 3 for start up Electrical Requirements Electrical receptacles should be located nearby for connecting the Gas Purifier and the controlled atmo
50. you will require the following from Chapter 7 24 Adjustable height base stand Regeneration Tubing kit Circulation Tubing kit Gas Vacuum Vent Tubing kit Direct drive rotary vane vacuum pump Oxygen or Moisture monitor or both Recommended A glove box configured with auto pressure controller facilitates best performance of the AtmosPure Re Gen Gas Purifier Product Service 1 800 522 7658 Chapter 4 Features and Safety Precautions Labconco has engineered the Gas Purifier to meet the ISO 10648 2 Class 1 atmosphere leak specification when used with Labconco controlled atmosphere glove boxes The information that follows details performance features so you can get the most out of your Gas Purifier Refer to Figure 4 1 Gas Purifier Performance Features 1 2 3 4 5 6 Capacity to remove up to 5 liters of oxygen 13 500 ppm and 660 grams of moisture 18 million ppm with normal oxygen and moisture permeation rates below 0 16 ppm min The adsorption column contains 5 lbs of copper catalyst to remove oxygen and 5 lbs of molecular sieve to remove moisture Depending on the application and use the Gas Purifier may operate for 1 3 months before initiating the required 13 hour regeneration process to restore the catalysts Ability to reduce oxygen to 1 ppm in less than 2 hours starting from 1000 ppm See Appendix D Ability to reduce moisture below 15 ppm in 2 hours starting from 1000 ppm The Gas Purifier