Thermo Scientific™ SampleSeal™ 384

image

Contents

1. 1 6 1 ici 6 1 Opem COSC settee ad ee eee 6 1 6 2 7 aes 3 E E E E TO 6 3 CVE TMC use tts Owe ai a dhe ca 6 3 Chapter 7 Maintenance FA cee eee eee eee A AEE aU 7 1 USEF MAINTENANCE ue 7 1 MW Dll nae SR aa Geek aa 7 1 Cleaning and de contamination YY FE FF FF FF Fad 7 1 RACKINGSE cos iat ane A CF 7 2 Thermo Scientific Thermo Scientific SampleSeal 384 tube Heat Sealer 3 Table of Contents CAG DIAN vr A A a AA WAA NF y 1 2 GUN ar an seaman eens ea woh wee ern HEFY EA 7 3 Touchscreen User 000 dida E E ddw 7 3 Non User Maintenance ccdx ts FR Ddw o DD a eee a ad 4 Cutting tool replacement YY FF FF FFF FF Fad 4 Guillotine replacement YF FF tenet need 7 4 Chapter 8 Troubleshooting 00c ccc sce c cence eect ence AA A A AE ARR 8 1 MESSAGES ssc ac a ad o ae atic Gee 8 1 8 2 Chapter 9 SampleSeal Remote Control o0ooococcccccccc 9 1 General Command 9 1 Command Set a WED DO a OLE ba 9 2 O a 9 2 Rescata UN NEU FN NT 9 3 E ES A 9 3 A Ae eee he arden nate Wats oe 9 3 O DDED 9 3 CUT TOI ck tents od di NR 9 4 Roques Esta CANCR DN aides wee op 9 4 Chapter 10 DISPOSAL ooo c nica iia a la ee Pelee ca 10 1 1 4 Thermo Scientific SampleSeal 384 tube Heat Sealer Thermo Scientif
2. CAUTION If the waste bin is overfilled waste foil may not be ejected from the instrument causing the guillotine to fail The waste bin at the rear of the instrument has sufficient capacity for 200 racks to be processed before emptying Cleaning and de contamination The outer surfaces of the SampleSeal unit can be cleaned with typical non abrasive laboratory cleaning agents or 70 alcohol solutions The system also tolerates decontamination regimes based on hydrogen peroxide e g that offered by Bioquell Inc CAUTION Do not use Formaldehyde as a decontamination agent as this may cause permanent damage to components and invalidate the SampleSeal warranty Do not use acetone as a Cleaning agent as damage to components especially plastics or painted surfaces may occur Internal surfaces may be wiped clean using the same agents a vacuum cleaner may be used to remove any debris within the instrument The system may be used with samples containing DMSO as a cryopreservative but care must be taken to prevent contact with painted surfaces and plastic components Thermo Scientific SampleSeal 384 tube Heat Sealer 1 Thermo Scientific Chapter 7 Maintenance Rack Nest Seal quality may be impaired if the base of the 384 tube rack does not sit directly on the base of the rack nest The rack nest should be regularly inspected and wiped clean or a vacuum cleaner used to remove any debris Sealing platen The sealing
3. cutter and guillotine which are wear parts Cutting tool replacement The cutting tool is a wear item Contact the service team for details if replacement is required Guillotine replacement The guillotine is also a wear item Contact the service team if replacement is required 7 4 Thermo Scientific SampleSeal 384 tube Heat Sealer Thermo Scientific Troubleshooting Messages Notifications and error messages are displayed on the messages screen accessed by touching the Msgs button on the main screen Selecting an individual message by touching the message text displays further information in the Details section of the screen To acknowledge and clear a message select the message and press OK Press Back to return to main screen Thermo Scientific Thermo Scientific SampleSeal 384 tube Heat Sealer 1 Chapter 8 Troubleshooting Manual Recovery WARNING Do not attempt to override the loading or service door safety interlocks Failure to comply may result in serious injury WARNING Do not attempt to access the interior of the device through any other openings and do not remove panels there are no user serviceable components inside Failure to comply may result in serious injury or death WARNING Do not attempt to access the interior of the SampleSeal unit when the sealing platen is hot Contact with the hot platen may result in burns To recover failed labware from the interior of the unit e Wai
4. WARNING Do not attempt to access the interior of SampleSeal when the sealing platen is hot Contact with the hot platen may result in burns WARNING Do not attempt to use foil other than the specified material Failure to comply may result in the device being damaged and invalidate the warranty CAUTION The edges of the sealing foil are sharp and unless handled carefully may result in minor cuts Wear latex or nitrile laboratory gloves when handling foil To load a roll of foil e Remove the service door to disconnect power to heating pins allowing safe access to internal components e Allow unit to reach a safe temperature lt 70 C e Remove the foil cover from the top of the instrument e Remove the waste bin from the rear of the instrument e From the function screen disable the guillotine e From the function screen go to the foil screen on the user interface e Wait until the sealing platen is at a safe temperature as indicated at the touch screen then remove the service door by turning the quarter turn fasteners and lifting the panel from the location tabs at its base Opening the service door will trigger the safety circuit and remove air from all actuators which will now be raised e Load the foil roll and thread the foil through the instrument following the instructions on the inside face of the service panel Thermo Scientific Thermo Scientific SampleSeal 384 tube Heat Sealer 5 1 Chapter5 Loadin
5. platen is PTFE coated to prevent the foil sticking but may need wiping clean periodically when unit is cool to ensure optimum seal quality WARNING Do not attempt to access the interior of the SampleSeal unit when the sealing platen is hot Contact with the hot platen may result in burns CAUTION Do not use abrasive cleaning agents or scouring pads on the sealing platen as the non stick coating will be damaged CAUTION Do not spray cleaning agents directly onto the sealing platen or into other parts of the instrument This may result in damage to the unit To clean the sealing platen e Wait until the platen is at 40 C as indicated at the touchscreen then remove the service door by turning the quarter turn fasteners and lifting the panel from the location tabs at its base Opening the service door will trigger the safety circuit and remove air from all actuators which will now be raised e Wipe the platen with a soft wet cloth DO NOT use abrasives cleaning agents or scouring pads as the non stick coating will be damaged e For stubborn deposits a solvent may be required acetone may be used but take extreme care to prevent contact with painted parts e Wipe the platen dry with a soft dry cloth and inspect for any residue if there is none replace the service door and reset the unit e Reset the sealing temperature 7 2 Thermo Scientific SampleSeal 384 tube Heat Sealer Thermo Scientific Maintenance Chapt
6. 19sf Thermo Scientific SampleSeal 384 tube Heat Sealer Version 01 April 2014 Table of Contents Chapter 1 011005 cs inau ud a a a iaa 1 5 Sale ma A YAR 1 5 oale MOS air ar Soni ow tad a ack Goad Aare Ue BL 1 5 DALEY LADEIS cotos aia ed 1 5 WI a aia 1 6 Technical Support Contact Details 1 6 Ghapter2 IntrOdUCtON woes eee er eae a ij 2 1 Pesci SL LS cata eaten anna Ae 2 1 QVEIVIGW coso ir der Gh bones A ld DWN I A Fd Gh 2 2 Chapter 3 Unpacking and 1 0 51 31 3 1 Unpacking the SampleSeal units ee aeb brinda is aid 3 1 Installing the SampleSeal Heat Sealer 3 2 Operating Conditions 0 08 00 Fad 3 2 POWE CFE debe neta and a i ia eai eS phases dad hae 3 2 IN so mice d oe Cade MoS Geko eed FFO E avec teeta alee 3 3 Ebro 3 3 PUIG Chapter 4 Stat UD cocina a a il 4 1 Sale Ne raea 4 1 Pon st A E A A A E 4 2 Sealing Partially Populated Racks 4 4 FUNCIONS iena arpa O TF grada NU NHU S84 aa FFAN a aa 4 5 Accessing the function SCTEEn 4 5 Gas aos ca ld Damen 4 5 A cdc tats wean tia EN HN HY NN 4 5 6 7 onan te teh a hance 4 5 Model INMOMMAUON oho dae acta eek oe oe e bee eae eR a eee ee 4 6 Chapter Loading TOM saa sais cosa car aaa 5 1 Chapter6 Sealing 384 format 2D tube racks
7. 6mm pushfit connector at the rear of the instrument The installer must ensure that there is a means of safely isolating the air supply for service or in the event of an emergency CAUTION Oil or water in the air supply and or a supply pressure greater than 10bar 140psi will damage the instrument and invalidate the warranty Quality Pressure Flow Clean dry oil free 5 9 to 8 3 bar 85 to 120 psi 50 slpm 1 77scfm Purge Gas The purge gas connection is via a 4mm pushfit connector at the rear of the instrument The installer must ensure that there is a means of safely isolating the purge gas supply for service or in the event of an emergency WARNING Argon and Nitrogen are odorless colorless gases that can cause serious injury or death by suffocation Use only in well ventilated areas and install an oxygen monitor in an appropriate location WARNING DO NOT connect flammable or reactive gases to the purge supply Failure to comply may result in serious injury or death CAUTION Oil or water in the purge gas supply and or a supply pressure greater than 10bar 140psi will damage the instrument and invalidate the warranty Quality Pressure Clean dry inert gas or nitrogen 3 5 to 8 3 bar 50 to 120 psi 5 sipm 0 18 scfm NOTE Purge gas must always be connected for correct operation of SampleSeal If inert gas is not available tee off the air supply line via suitable Gmm 4mm reducing connector Thermo Scientific Sampl
8. al 384 tube Heat Sealer 3 1 Chapter 3 Unpacking and installation Installing the SampleSeal Heat Sealer SampleSeal is designed for laboratory use Dimensions are given below 85 cm 33 5 in 36 2 cm 14 3 in ollo Pic 61 cm 24 in The instrument should be installed on a flat surface in close proximity to electricity a compressed air supply and a purge gas supply if gas purge is required NOTE Ensure that there is sufficient room behind the instrument to access the ON OFF switch Do not cover the air vents at the rear of the machine as this may cause overheating Operating Conditions Operating Temperature Humidity Maximum Duty Cycle 15 C to 30 C lt 75 RH non condensing 30 racks hour Power The SampleSeal instrument requires an AC main supply The connection is made at the switched IEC inlet at the rear of the unit Voltage Frequency 85 to 264V AC 47 to 63 Hz 3 2 Thermo Scientific SampleSeal 384 tube Heat Sealer Thermo Scientific Unpacking and installation Chapter 3 Air The SampleSeal instrument requires a clean dry oil free compressed air supply The air connection is via a
9. d not be used for any other purpose SampleSeal can be used as a standalone device or integrated into robotic platforms SampleSeal 384 format Sealer Catalog Number Description 4220 384 SampleSeal 384 format Foil Heat Sealing and Cutting Instrument 4221 025 SampleSeal 25um Easy pierce Foil Compatible Labware 3815 Matrix 384 format 70ul Open Top 2D Barcoded tube NS Thermo Scientific Thermo Scientific SampleSeal 384 tube Heat Sealer 2 1 Chapter 2 Introduction Overview Removable foil cover AOS Waste bin Operator push buttons x4 Rack loading Service door door Electrical panel cover Identification label Services connection point 2 2 Thermo Scientific SampleSeal 384 tube Heat Sealer Thermo Scientific Unpacking and installation Unpacking the SampleSeal unit WARNING The weight of the SampleSeal instrument is approximately 47 6kg 105Ib Please ensure that the instrument is handled safely by at least two people and that appropriate lifting methods are employed Always lift the unit using the recessed lifting handles in the base of the unit Never lift this equipment by the extended carriage Failure to follow this instruction may cause the carriage to become detached which may result in serious injury Check that all tables safety cabinets and other equipment are capable of safely supporting the SampleSeal unit Thermo Scientific Thermo Scientific SampleSe
10. e return CR character ASCII 13 or a line feed LF character ASCII 10 e commands are case insensitive Responses from the instrument are formatted as follows e responses from the board start with a lt character ASCII 60 e this is followed by a single character digit indicating the type of response 0 Successful end of a command 1 Successful start of a command only generated if the command does not end immediately 2 Data from an executing command there will be 0 many of these between the lt 1 start and gt 0 end 3 Failed command was corrupted probably a bad checksum 4 Failed unrecognised Command was correctly formatted but does not correspond to a known command or it has bad parameters 5 Failed generic failure then some optional response specific text checksum represented by characters xx where xx is the hexadecimal sum of all characters before the 5 responses are terminated with a carriage return CR character ASCII 13 Command Set Definition The following sections describe the commands used to remotely operate the instrument and the responses expected for each command Stop e stops any currently underway operation equivalent to pressing Stop button e command stop e normally no response 9 2 Thermo Scientific SampleSeal 384 tube Heat Sealer Thermo Scientific SampleSeal Remote Control Chapter 9 Reset Instrument e resets the instrument equivalent t
11. eSeal 384 tube Heat Sealer 3 3 Thermo Scientific Start up Safety The SampleSeal heat sealer contains moving parts and high temperature components that can cause serious injury The SampleSeal unit is designed to protect users from these hazards under normal operating conditions WARNING Do not attempt to override the loading or service door safety interlocks Failure to comply may result in serious injury Do not attempt to access the interior of the device through any other openings and do not remove panels there are no user serviceable components inside Failure to comply may result in serious injury or death In an emergency turn off the SampleSeal unit The power switch is located at the back of the instrument Thermo Scientific Thermo Scientific SampleSeal 384 tube Heat Sealer 1 Chapter 4 Startup Power Up To power up the instrument connect the electrical and compressed air supplies and turn on the power switch at the back of the instrument After a short delay the touchscreen will be illuminated Reset Required Press the reset button Press the RESET pushbutton The instrument will initialize and if foil is loaded and the compressed air supply is connected the sealer will start to warm up 4 2 Thermo Scientific SampleSeal 384 tube Heat Sealer Thermo Scientific Startup Chapter 4 When the temperature is close to set point there is a short d
12. elay to allow the temperature of the sealing platen to stabilize The SampleSeal heat sealer is now ready for use Thermo Scientific SampleSeal 384 tube Heat Sealer 3 Thermo Scientific Chapter 4 Startup Sealing Partially Populated Racks CAUTION It is important that there are no tubes present in columns that have been designated as empty The SampleSeal operating system assumes that racks are placed on the rack nest with column 1 closest to the instrument Failure to comply with these conventions will result in a failure to process the rack and manual recovery of the labware will be required is possible to configure the SampleSeal unit to seal racks with empty columns at each or either end On the main screen touch the Columns tab The 384 way SampleSeal unit is capable of leaving columns 1 2 and 23 24 unsealed Remove tubes from these columns for this procedure Touch OK or if the selection has been changed Save to accept the column selection NOTE To retain the settings after a power cycle it is necessary to seal a rack 4 4 Thermo Scientific SampleSeal 384 tube Heat Sealer Thermo Scientific Startup Chapter 4 Functions The SampleSeal heat sealer has a number of functions that can be enabled disabled by the user according to requirements Accessing the function screen Press Func on the main screen to access the function screen Gas Purge purge gas is connec
13. er 7 Cutting tool The 384 format cutter has a life of at least 1250 racks but may need cleaning periodically before replacement to ensure optimum seal quality WARNING Take extreme care when accessing the interior of the SampleSeal unit when the sealing platen is hot Contact with the hot platen may result in burns CAUTION The 384 format cutter is sharp wear laboratory gloves and take care to avoid injury To clean the cutter e Remove the service door by turning the quarter turn fasteners and lifting the panel from the location tabs at its base Opening the service door will trigger the safety circuit and remove air from all actuators which will now be raised e Using a soft bristle brush remove any debris from the cutting faces For stubborn debris an air duster may be required Do not use any hard or abrasive item to clean the cutting tool as this may cause permanent damage to it e Replace the service door and reset the unit Touchscreen User Interface CAUTION Do not use fluid cleaners on the touchscreen Failure to comply may result in the device being damaged and invalidate the warranty To clean the touch screen gently wipe with a microfiber lens cleaner or similar lint free cloth Thermo Scientific SampleSeal 384 tube Heat Sealer 3 Thermo Scientific Chapter 7 Maintenance Non User Maintenance The SampleSeal unit is designed to be maintenance free with the exception of the 384 format
14. f not avoided may result in injury Failure to follow CAUTION notices may result in personal injury or damage to labware or the instrument Warranty Standard warranty is 12 months or 1250 sealing cycles as recorded by the Cycle Counter for the cutting tool and guillotine both of which can be replaced in the field by a qualified technician Enhanced preventative maintenance and service packages are available For further details contact Technical Support Technical Support Contact Details North America and ROW Technical Support Europe Technical Support Tel 49 0 6184 906 000 email orders labequipment de thermofisher com Thermo Scientific Thermo Fisher Scientific Thermo Fisher Scientific 75 Panorama Creek Drive Robert Bosch Str 1 Rochester NY 14625 USA 63505 Langenselbold Tel 585 586 8800 Germany 1 800 625 4327 email technicalsupport thermofisher com Chapter 1 Notices 1 6 Thermo Scientific SampleSeal 384 tube Heat Sealer Introduction Description The Thermo Scientific SampleSeal Heat Sealer is a laboratory instrument designed for the foil sealing of racks of 2D and non 2D barcoded storage tubes This manual describes the function of the 384 format version of SampleSeal only and due to a policy of continuous improvement contents are subject to change without notice The SampleSeal Heat Sealer is intended for research use only and for use as a sealer cutter It shoul
15. g foil e Index the foil by pressing the gt gt button on the foil screen Step 12 on the foil loading instruction label Approximately 2 meters of foil will be unwound from the reel to ensure the foil used to seal subsequent racks has not been handled by the operator The foil index may be stopped at any time by pressing the STOP button e From the function screen enable the guillotine e Replace the foil cover e Replace the waste bin e Replace the service door and reset the SampleSeal instrument e Wait for unit to return to sealing temperature 5 2 Thermo Scientific SampleSeal 384 tube Heat Sealer Thermo Scientific Sealing 384 format 2D tube racks Reset Press the RESET button to initialize the SampleSeal unit after it has been switched on and to return the unit to a known state after a STOP or to recover from a fault or to initialize after manual recovery from a fault Open Close By default the instrument leaves the carriage extended at the end of the cycle ready for the next rack The rack carriage may be retracted or extended between cycles by pressing the OPEN CLOSE push button Thermo Scientific Thermo Scientific SampleSeal 384 tube Heat Sealer 1 Loading Labware WARNING Do not place your hand inside the SampleSeal unit while loading labware there are hot components inside the unit WARNING Use only specified labware Failure to comply may result in the device being damaged and invalidate the war
16. ic Notices Safety Before using the Thermo Scientific SampleSeal heat sealer you should ensure that you are properly trained in gt The correct and safe operation of the SampleSeal instrument gt The correct and safe maintenance procedures if you are involved in the servicing or repair of the instrument Safety Standards The SampleSeal heat sealer is CE compliant Product has been Tested according to CAN CSA 022 2 12 UL 61010 1 2012 EN 61010 1 2010 The product was voluntarily tested according to the relevant safety requirements noted above lt can be marked with the certification mark above The mark must not be altered in any way This product certification system operated by TUV SUD Testing and Certification Regulations TUV SUD America Inc is an OSHA recognized NRTL and a Standards Council of Canada accredited certification body NOTE All references to TUV and SUD include the two dots over the U Safety Labels The SampleSeal unit has a number of safety labels intended to protect the operator from injury pay attention to these labels at all times Thermo Scientific SampleSeal 384 tube Heat Sealer 1 5 Thermo Scientific A WARNING notice indicates a potentially hazardous situation which if not avoided could result in serious injury or death Failure to follow WARNING notices may result in serious injury or death A CAUTION notice indicates a potentially hazardous situation which i
17. ires FTDI VCP drivers on the computer The serial communications parameters required for remote control of the instrument are e 115200 baud e 8 data bits e 1 stop bit e no parity e no flow control General Command Syntax All commands and responses are expressed in plain ASCII Any terminal emulator program can conveniently be used to communicate with the instrument e g Hyperterminal or PuTTY For ease of use it is likely that the user would wish to configure their terminal emulation software with the following features enabled e local echo enable so you can see what you type e send carriage return and line feed for enter add carriage return to newline so your typed commands end on a new line e add Linefeed to received carriage return so responses from the instrument are displayed correctly Thermo Scientific Thermo Scientific SampleSeal 384 tube Heat Sealer 9 1 Chapter 9 SampleSeal Remote Control All commands sent to the instrument must be formatted as follows gt begin with a gt character ASCII 62 e a gt is regarded as the beginning of a command even if it occurs in the middle of another command e the body of the command follows the gt character e commands may contain an optional checksum after the body of the command This is represented by characters xx where xx is the hexadecimal sum of all characters before the 5 commands are terminated with a carriag
18. is returned as a textual description equivalent to what is displayed on the touch screen panel e the status description may contain internal delimiters represented as I characters ASCII 124 gt command st e response lt 0ST status text 9 4 Thermo Scientific SampleSeal 384 tube Heat Sealer Thermo Scientific Disposal SampleSeal is required to comply with the European Union s Waste Electrical amp Electronic Equipment WEEE Directive 2002 96 EC and is marked with the following symbol Thermo Fisher Scientific has contracted with one or more recycling or disposal companies in the European Union EU and these companies will dispose of or recycle this product Please contact Thermo Fisher Scientific at the address given in the Technical Support Contacts section of this User Manual for further information on compliance with these Directives and the recyclers in your country Thermo Scientific Thermo Scientific SampleSeal 384 tube Heat Sealer 10 1 Thermo Fisher Scientific 75 Panorama Creek Drive Rochester NY 14625 USA Technical Support North America Tel 585 586 8800 1800 625 4327 Email technicalsupport thermofisher com Europe Tel 49 0 6184 906 000 Email orders labequipment de thermofisher com thermoscientific com O 2014 Thermo Fisher Scientific Inc All rights reserved All trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries Specifications
19. o pressing Reset button command srm e normal response is series of messages similar to e lt T o lt 2 Reset e lt 0 Go e Start a cycle equivalent to pressing the Go button gt command sgo e normal response is series of messages similar to e lt T e lt 2 Seal and Punch e lt 0 Open Close e equivalent to using the Open Close button gt command soc opens the hatch if presently closed closes it if presently open toggle action like button e command sop opens the hatch command scl closes the hatch e normal response is series of messages similar to e lt I e lt 2 Open e lt 0 Feed foil gt equivalent to selecting one of the feed foil actions on the touch screen panel gt command sin feeds small amount of foil e command sfd feeds longer length of foil e normal response is series of messages similar to e lt T e lt 2 Inch Foil e lt 0 Thermo Scientific SampleSeal 384 tube Heat Sealer 9 3 Thermo Scientific Chapter 9 SampleSeal Remote Control Cut foil e equivalent to selecting the cut foil action on the touch screen panel command set e normal response is series of messages similar to e lt T gt lt 2 Cut Foil e lt 0 Request status e this command allows the controlling computer to request the status of the instrument e the status
20. ranty CAUTION The SampleSeal operating system assumes that racks are placed on the rack nest with column 1 closest to the instrument It is important that this convention is observed when processing partially populated racks Check that the rack nest is free from debris Seal quality may be impaired if the base of the 384W tubes do not sit directly on the base of the rack nest Place the rack to be sealed in the rack nest The rack should be oriented so that A1 is located at the back right corner of the rack nest A red screw indicates the A1 position Chapter6 Sealing 384 format 2D tube racks 6 2 Thermo Scientific SampleSeal 384 tube Heat Sealer Thermo Scientific Sealing 384 format 2D tube racks Chapter 6 Check that the rack is level and within the raised tabs at the corners of the rack nest as shown in the picture below GO The GO button initiates a sealing cycle and flashes throughout the sealing cycle STOP The STOP button is illuminated throughout a sealing cycle and other operations Pressing STOP halts the SampleSeal unit immediately If STOP is pressed mid cycle then manual recovery of labware may be required and the instrument must be RESET Cycle Time The SampleSeal unit processes a fully populated rack of 384 format tubes in approximately 100 seconds Thermo Scientific Thermo Scientific SampleSeal 384 tube Heat Sealer 6 3 Maintenance User maintenance Waste Bin
21. t for the unit to reach a safe temperature lt 0 C e Remove the service door by turning the quarter turn fasteners and lifting the panel from the location tabs at its base e The actuators will all be raised and it should be possible to move the rack carriage by hand when the power is off Similarly the foil drive may also be driven by hand if necessary e Take care not to damage components inside the SampleSeal unit during recovery and remove any debris that may be present gt If the cutting tool fails to retract from the rack nest turn off power to the instrument and manually free the tool from the rack nest Check that there is no debris in the holes or on the cutting tool datum pins If necessary apply a small amount of silicone grease with a cotton swab to the inside of the rack nest datum holes e Reload the foil if required e Replace the service door and reset the instrument e Wait for the sealing temperature to return to the original value e Ensure the drive roller shown below is free of debris 8 2 Thermo Scientific SampleSeal 384 tube Heat Sealer Thermo Scientific SampleSeal Remote Control The SampleSeal heat sealer can be controlled remotely from a connected computer using simple serial commands The instrument is provided with a USB connector for remote control operation USB A female When connected to a USB port on a computer a virtual serial port will be created on the computer requ
22. ted to the unit tubes are automatically purged as the rack moves into the instrument The gas purge can be disabled at the function screen by pressing the circular button to the left of the text Guillotine Similarly the guillotine can be disabled By default SampleSeal automatically cuts the waste foil into thin ribbons for optimum compaction in the waste bin and easy disposal The user may wish to disable this function and handle the waste using an alternative method contact Technical Support for advice on suitable waste handling methods Carriage Extend By default SampleSeal extends the rack loading carriage to its full extent see section 3 2 for dimensions for optimum accessibility especially when integrated with a robotic picker For manual operation on a laboratory bench the user may prefer that the rack is only partially extended to save space This is achieved by disabling the Full extend Enable function Manual access is unimpaired Thermo Scientific SampleSeal 384 tube Heat Sealer 4 5 Thermo Scientific Chapter 4 Startup Model Information The info screen is accessed from the main screen and displays build information and the number of racks the SampleSeal has processed Due to quality control checks during manufacture the info screen the cycle count will be greater than zero on delivery 4 6 Thermo Scientific SampleSeal 384 tube Heat Sealer Thermo Scientific Loading foil
23. terms and pricing are subject to change Not all products are available in all countries Please consult your local sales representative for details hermo SCIENTIFIC Part of Thermo Fisher Scientific MNSAMPLESEAL384 0414

Download Pdf Manuals

image

Related Search

Related Contents

Operating Instructions Manual  User Manual -- Channel Creation  AmpFlSTR® Identifiler® Plus PCR Amplification Kit User`s Guide  Mode d`emploi de l`outil Test Compta Demat  HP LeftHand Storage with VMware vSphere: Design considerations  WARNING: AVERTISSEMENT: Pb  System User Manual  ST-C530 SM-BC05  Ewent EW7038 storage enclosure  Mode d`emploi des générations d`appareils précédentes  