Home
R&S FSC Spectrum Analyzer
Contents
1. without input signal RBW lt 30 kHz RF attenuation 0 dB tracking generator OFF fin 2 x 4881 MHz 60 dBc Intermediate frequencies 21 4 MHz 831 4 MHz 4881 4 MHz 60 dBc typ 80 dBc 8931 4 MHz 50 dBc Other interfering signals f lt 3 6 GHz signal level RF attenuation lt 20 dBm spurious at fin 2440 7 MHz lt 60 dBc 3 6 GHz lt f lt 6 GHz spurious at fin 4465 7 MHz lt 60 dBc Other interfering signals f lt 3 6 GHz related to local oscillators Af lt 300 kHz 60 dBc Af 2 300 kHz lt 60 dBc f gt 3 6 GHz Af lt 300 kHz 54 dBc Af 2 300 kHz lt 54 dBc f receive frequency Residual spurious response input matched with 50 Q lt 90 dBm Level display Logarithmic level axis 1 2 5 10 20 50 100 dB 10 divisions Linear level axis 0 to 100 10 divisions Number of traces 2 Trace detectors max peak min peak auto peak sample RMS Trace functions clear write max hold min hold average view Setting range of reference level 80 dBm to 30 dBm Units of level axis dBm dBmV dByV V W Level measurement uncertainty Absolute level uncertainty at 100 MHz 20 C to 30 C 0 3 dB o 0 1 dB Frequency response 20 C to 30 C 9 kHz lt f lt 10 MHz 1 5 dB nominal 10 MHz lt f lt 3 6 GHz 1 dB o 0 33 dB 3 6 GHz lt
2. Level Display range displayed noise floor to 30 dBm Maximum rated input level with RF attenuation gt 10 dB DC voltage 50 V CW RF power 30 dBm 1 W Peak RF power lt 3 s duration 33 dBm 2 W Max pulse voltage 150 V Max pulse energy pulse width 10 us 10 mWs Maximum rated input level with RF attenuation lt 10 dB DC voltage 50 V CW RF power 20 dBm 100 mW Peak RF power lt 3 s duration 23 dBm 200 mW Max pulse voltage 50 V Max pulse energy pulse width 10 us 1mWs Intermodulation nominal values Third order intermodulation TOI intermodulation free dynamic range signal level 2 x 20 dBm RF attenuation 0 dB without RF preamplifier R amp S FSC B22 option or RF preamplifier OFF fin lt 300 MHz gt 54 dBc TOI gt 7 dBm typ 11 dBm 300 MHz lt fin lt 3 6 GHz gt 60 dBc TOI gt 10 dBm typ 15 dBm 3 6 GHz lt fn lt 6 GHz gt 46 dBc TOI gt 3 dBm typ 10 dBm signal level 2 x 40 dBm RF attenuation 0 dB RF preamplifier R amp S FSC B22 option ON fin lt 300 MHz gt 50 dBc TOI 15 dBm 300 MHz lt fin lt 6 GHz gt 56 dBc TOI 12 dBm nominal values Second harmonic intercept SHI RF attenuation 0 dB without RF preamplifier R amp S FSC B22 option or RF preamplifier OFF fin 20 MHz to 1 5 GHz 40 dBm fin 1 5 GHz to 3 GHz 30 dBm RF attenuation 0 dB RF preamplifier R amp S FSC B22 option
3. 100 kHz lt f lt 300 kHz gt 60 dB typ 80 dB 300 kHz lt f lt 3 GHz gt 70 dB typ 90 dB 3 GHz lt f lt 6 GHz model 16 only gt 70 dB typ 90 dB Reverse power DC voltage 50 V CW RF power 20 dBm 0 1 W Max pulse voltage 50 V Max pulse energy 10 us 1 mWs 6 Rohde amp Schwarz R amp S FSC Spectrum Analyzer Inputs and outputs Version 01 03 August 2010 RF input Impedance 50 Q Connector N female VSWR 100 kHz lt f lt 1 GHz lt 1 5 nominal 1 GHz lt f lt 6 GHz lt 2 nominal Setting range of input attenuator 0 dB to 40 dB in 5 dB steps RF preamplifier gain with R amp S FSC B22 option 20 dB nominal AF output AF demodulation types AM and FM Connector 3 5 mm mini jack Output impedance 32 Q nominal Voltage open circuit Vrms adjustable from O V to gt 100 mV USB interface Front panel USB host interface version 1 1 Connector USB type A plug version 1 1 Memory sticks supported lt 4 Gbyte USB version 1 1 or 2 0 Rear panel USB device interface version 1 1 Connector USB type B plug version 1 1 External reference external trigger Connector BNC female 50 Q Mode selectable external reference external trigger External reference input required level 0 dBm frequency 10 MHz External trigger threshold low high transition 2 4 V
4. R amp S FSC Spectrum Analyzer Specifications ROHDE amp SCHWARZ Version 01 03 August 2010 CONTENTS T en ica 3 ETA QUE Ya a At ita ia dida 3 Sweep TIME sees ee eee eee eee eee 3 Bandwidth S scr dd AA AAA AAA AAA A 3 A OOO NOTE 4 aiee L d aeie THH 5 Tracking generator model 13 160MlV 5 occicaesk cccttscandesecesathssbssoapcuaetieciaeouassuecuipesuctvaceudicadedsubanadncecustvasteteneptisoishetynesatocnsudsbusoasctenuteniecdas 6 INPUES AMC QUIE ati ii A A a tica 7 General data id daa eine 8 Ordering information ia 9 Options sivsat nrneieen east hemi ce AAA ci ee De te a ee ees es 9 A ca 10 Specifications apply under the following conditions 15 minutes warm up time at ambient temperature specified environmental conditions met calibration cycle adhered to Data without tolerances typical values only Data designated as nominal applies to design parameters and is not tested 2 Rohde amp Schwarz R amp S FSC Spectrum Analyzer Version 01 03 August 2010 Base unit Frequency Frequency range model 03 13 9 kHz to 3 GHz model 06 16 9 kHz to 6 GHz Frequency resolution 1 Hz Reference frequency internal nominal Aging per year 1x 10 Temperature drift 0 CT to 30 C 1x 10 30 C to 50 C 3 x 107 Achievable initial adjustment accuracy 5x10 Total reference uncertainty time since last adjustment x aging rate temperature drift calibration accuracy
5. 0 C to 55 C storage temperature range 40 C to 70 C Climatic loading relative humidity 25 40 C at 85 relative humidity IEC 60068 2 30 Mechanical resistance Vibration sinusoidal IEC 60068 2 6 random IEC 60068 2 64 Shock 40 g shock spectrum in line with MIL STD 810E method 516 4 procedure 1 IEC 60068 2 27 EMC in line with EMC Directive 2004 108 EC including IEC EN 61326 1 IEC EN 61326 2 1 CISPR 11 EN 55011 IEC EN 61000 3 2 IEC EN 61000 3 3 Weight and dimensions Dimensions WxHxD 233 mm x 158 1 mm x 350 mm 9 2 in x 6 2 in x 13 8 in Weight 4 5 kg 9 9 Ib Recommended calibration interval 1 year 1 RF emission in line with EN 55011 class A operation in residential commercial and business areas or in small size companies is not covered Thus the instrument may not be operated in residential commercial and business areas or in small size companies unless additional measures are taken to ensure that EN 55011 class B is complied with 2 Immunity test requirement for industrial environment EN 61326 table 2 8 Rohde amp Schwarz R amp S FSC Spectrum Analyzer Version 01 03 August 2010 Ordering information Designation Type Order No Spectrum Analyzer 9 kHz to 3 GHz R amp S FSC3 1314 3006 03 Spectrum Analyzer 9 kHz to 3 GHz R
6. ON fin 100 MHz to 3 GHz 0 dBm Displayed average noise level RF attenuation 0 dB termination 50 Q RBW 100 Hz VBW 10 Hz sample detector log scaling tracking generator OFF normalized to 1 Hz without RF preamplifier R amp S FSC B22 option or RF preamplifier OFF frequency 9 kHz to 100 kHz lt 108 dBm typ 118 dBm 100 kHz to 1 MHz lt 115 dBm typ 125 dBm 1 MHz to 10 MHz lt 136 dBm typ 144 dBm 10 MHz to 2 GHz lt 141 dBm typ 146 dBm 2 GHz to 3 6 GHz lt 138 dBm typ 143 dBm 3 6 GHz to 5 GHz lt 142 dBm typ 146 dBm 5 GHz to 6 GHz lt 140 dBm typ 144 dBm RF attenuation O dB termination 50 Q RBW 100 Hz VBW 10 Hz sample detector log scaling tracking generator OFF normalized to 1 Hz RF preamplifier R amp S FSC B22 option ON frequency 100 kHz to 1 MHz lt 133 dBm typ 143 dBm 1 MHz to 10 MHz lt 157 dBm typ 161 dBm 10 MHz to 1 GHz lt 161 dBm typ 165 dBm 1 GHz to 2 GHz lt 159 dBm typ 163 dBm 2 GHz to 5 GHz lt 155 dBm typ 159 dBm 5 GHz to 6 GHz lt 151 dBm typ 155 dBm 4 Rohde amp Schwarz R amp S FSC Spectrum Analyzer Version 01 03 August 2010 Immunity to interference nominal values Image frequencies fin 2 x 21 4 MHz lt 70 dBc typ 80 dBc fin 2 x 831 4 MHz lt 70 dBc typ 90 dBc
7. Frequency readout Marker resolution 0 1 Hz Uncertainty marker frequency x reference uncertainty 10 x resolution bandwidth Ye span sweep points 1 1 Hz Number of sweep trace points 631 Marker tuning frequency step size span 630 Frequency counter resolution 0 1 Hz Count uncertainty S N gt 25 dB frequency x reference uncertainty Ye last digit Frequency span 0 Hz 10 Hz to 3 GHz 6 GHz Span setting uncertainty span 630 Spectral purity SSB phase noise f 500 MHz carrier offset 30 kHz lt 95 dBc 1 Hz typ 105 dBc 1 Hz 100 kHz lt 100 dBc 1 Hz typ 110 dBc 1 Hz 1 MHz lt 120 dBc 1 Hz typ 127 dBc 1 Hz Sweep time Sweep time span 0 Hz 200 us to 100 s 10 Hz lt span lt 600 MHz 20 ms to 1000 s span gt 600 MHz 20 ms x span 600 MHz to 1000 s Uncertainty span 0 Hz 1 nominal span gt 10 Hz 3 nominal Bandwidths Resolution bandwidths Range 3 dB bandwidth 10 Hz to 3 MHz in 1 3 sequence Bandwidth accuracy 10 Hz lt RBW lt 300 kHz lt 5 nominal RBW gt 300 kHz lt 10 nominal Selectivity 60 dB 3 dB lt 5 Gaussian type filters nominal Video filters Range 3 dB bandwidth 10 Hz to 3 MHz in 1 3 sequence Rohde amp Schwarz R amp S FSC Spectrum Analyzer 3 Version 01 03 August 2010
8. amp S FSC3 1314 3006 13 with tracking generator Spectrum Analyzer 9 kHz to 6 GHz R amp S FSC6 1314 3006 06 Spectrum Analyzer 9 kHz to 6 GHz R amp S FSC6 1314 3006 16 with tracking generator Accessories supplied Power cable USB cable for connection to PC quick start guide and CD ROM with operating manual and service manual Options Designation Type Order No Preamplifier 100 kHz to 3 GHz 6 GHz R amp S FSC B22 1314 3535 02 for the R amp S FSC3 6 Rohde amp Schwarz R amp S FSC Spectrum Analyzer 9 Version 01 03 August 2010 Recommended extras Designation Type Order No Ethernet Cable R amp S HA Z210 1309 6152 00 Headphones R amp S FSH Z36 1145 5838 02 19 Rack Adapter R amp S ZZA T33 1109 4458 00 for installing two R amp S FSC 19 Rack Adapter R amp S ZZA T34 1109 4464 00 for installing one R amp S FSC Matching pad 50 75 Q R amp S RAM 0358 5414 02 0 Hz to 2700 MHz matching at both ends N connectors Matching pad 50 75 Q R amp S RAZ 0358 5714 02 0 Hz to 2700 MHz matching at one end N connectors 75 ohm matching pad N to BNC female R amp S FSH Z38 1300 7740 02 Near Field Probe Set R amp S HZ 15 1147 2736 02 Preamplifier for R amp S HZ 15 R amp S HZ 16 1147 2720 02 Supported Power Sensors Designation Type Order No Universal Power Sensor R amp S NRP Z11 1138 3004 02 10 MHz to
9. f lt 6 GHz 1 5 dB o 0 5 dB Attenuator uncertainty 0 3 dB o 0 1 dB Uncertainty of reference level setting 0 1 dB nominal Display nonlinearity S N gt 16 dB 0 dB to 50 dB logarithmic level display 0 2 dB o 0 067 dB Bandwidth switching uncertainty reference RBW 10 kHz 0 1 dB nominal Total measurement uncertainty 95 confidence level 20 C to 30 C S N gt 16 dB 0 dB to 50 dB below reference level RF attenuation auto 10 MHz lt f lt 3 6 GHz 1 dB typ 0 5 dB 3 6 GHz lt f lt 6 GHz 1 5 dB typ 1 dB Trigger functions Trigger Trigger source free run video external External trigger level low high transition 2 4 V nominal high low transition 0 7 V nominal Rohde amp Schwarz R amp S FSC Spectrum Analyzer 5 Version 01 03 August 2010 Tracking generator model 13 16 only Frequency range model 13 100 kHz to 3 GHz model 16 100 kHz to 6 GHz Connector N female 50 Q VSWR 100 kHz lt f lt 1 GHz lt 1 5 nominal 1 GHz lt f lt 3 GHz lt 2 nominal 3 GHz lt f lt 6 GHz model 16 only lt 2 nominal Output level tracking generator attenuation 0 dB 0 dBm nominal Tracking generator attenuator 0 dB to 40 dB in 1 dB steps Dynamic range RF attenuation 0 dB tracking generator attenuation 10 dB RBW 1 kHz
10. nominal high low transition 0 7 V nominal IF out Connector BNC female 50 Q Frequency 21 4 MHz DC supply input Connector 5 mm DIN 45323 female Input voltage range 14 V to 16 V nominal Input current 0 9 A to 0 7 A Rohde amp Schwarz R amp S FSC Spectrum Analyzer 7 Version 01 03 August 2010 General data Power supply AC supply input specifications 100 V AC to 240 V AC 50 Hz to 60 Hz 400 Hz 130 VA DC supply input specifications 14 V to 16 V 0 9 A to 0 7 A nominal Power consumption 12 W nominal Safety in line with IEC 61010 1 EN 61010 1 CAN CSA C22 2 No 61010 1 04 UL61010 1 Test mark VDE GS cCSAus Manual operation Languages Chinese English French German Italian Hungarian Japanese Korean Portuguese Russian Spanish Remote control Command set SCPI 1997 0 LAN interface 10 100BaseT RJ 45 USB interface rear panel USB device type B Display Type 14 5 cm 5 7 LCD TFT color Resolution 640 x 480 pixel Audio Speaker internal Mass memory Mass memory flash memory internal USB memory stick not supplied Data storage internal gt 256 instrument settings and traces external on 1 Gbyte USB memory stick gt 5000 instrument settings and traces Temperature operating temperature range 0 C to 50 C permissible temperature range
11. 8 GHz 200 mW Universal Power Sensor R amp S NRP Z21 1137 6000 02 10 MHz to 18 GHz 200 mW Universal Power Sensor R amp S NRP Z22 1137 7506 02 10 MHz to 18 GHz 2 W Universal Power Sensor R amp S NRP Z23 1137 8002 02 10 MHz to 18 GHz 15 W Universal Power Sensor R amp S NRP Z24 1137 8502 02 10 MHz to 18 GHz 30 W Universal Power Sensor R amp S NRP Z31 1169 2400 02 10 MHz to 33GHz 200mW Thermal Power Sensor 0 Hz to 18 GHz R amp S NRP Z51 1138 0005 02 100 mW Thermal Power Sensor 0 Hz to 40 GHz R amp S NRP Z55 1138 2008 02 100 mW Thermal Power Sensor 0 Hz to 50 GHz 100 mW R amp S NRP Z56 1171 8201 02 Thermal Power Sensor 0 Hz to 67 GHz 100 mW R amp S NRP Z57 1171 8401 02 200 mW Wideband Power Sensor R amp S NRP Z81 1137 9009 02 50 MHz to 18 GHz 100 mW Average Power Sensor 9 kHz to 6 GHz R amp S NRP Z91 1168 8004 02 Average Power Sensor 9 kHz to 6 GHz 2W R amp S NRP Z92 1171 7005 02 All power sensors require the following adapt er cable for operation on the FSC Passive USB adapter to connect NRP sensors on R amp S FSC R amp S NRP Z4 1146 8001 02 For product brochure see PD 5214 3330 12 and at www rohde schwarz com 3 For average power measurement only 10 Rohde amp Schwarz R amp S FSC Spectrum Analyzer Version 01 03 August 2010 Rohde amp Schwarz R amp S FSC Spectrum Analyzer 11 Service you can r
12. bH amp Co KG 81671 Munchen Germany
13. ely on 1 Worldwide 1 Local and personalized 1 Customized and flexible 1 Uncompromising quality 1 Long term dependability About Rohde amp Schwarz Rohde amp Schwarz is an independent group of companies specializing in electronics It is a leading supplier of solu tions in the fields of test and measurement broadcasting radiomonitoring and radiolocation as well as secure communications Established more than 75 years ago Rohde amp Schwarz has a global presence and a dedicated service network in over 70 countries Company headquar ters are in Munich Germany Environmental commitment 1 Energy efficient products 1 Continuous improvement in environmental sustainability 1 ISO 14001 certified environmental management system Certified Quality System 150 9001 Rohde amp Schwarz GmbH amp Co KG www rohde schwarz com Regional contact 1 Europe Africa Middle East 49 89 4129 137 74 customersupport rohde schwarz com 1 North America 1 888 TEST RSA 1 888 837 87 72 customer support rsa rohde schwarz com 1 Latin America 1 41091079 88 customersupport la rohde schwarz com 1 Asia Pacific 65 65 13 04 88 customersupport asia rohde schwarz com R amp S is a registered trademark of Rohde 8 Schwarz GmbH amp Co KG Trade names are trademarks of the owners Printed in Germany as we PD 5214 3330 22 Version 01 03 August 2010 R amp S FSC Subject to change 2009 2010 Rohde amp Schwarz Gm