image

Contents

1. 9 3 2 1 Disassembly and assembly of 9 3 2 2 Furiction test so ex UAE WE ES 9 3 3 Pressure reducing valve unit item 0180 9 3 4 Gontrol PGB iteimi 0210 c ieri ev eR in Len ge in qued 9 3 4 1 Disassembly and assembly of 9 3 4 2 Functionitests or npe EP mb bnc bct eu 9 3 5 Control PCB item 0220 10 3 5 1 Measuring points and supply voltages 10 3 5 2 Disassembly and assembly of 10 3 5 3 Function tests iss PR RE VASA HER Meee EE EE a Ha 10 4 Anhang Annex Ee ten ede ee 11 4 1 Reparaturteile Repair parts 11 4 2 Explosionszeichnung Exploded view 12 4 3 Lageplan Schaltnetzteil Component drawing of Switching Power Supply Pos Item 0160 oan ke apre o APR rU Altea aes gained geb Rh 13 4 4 Lageplan Bedien E Karte Component drawing Operation PCB Postem 0210 z aa aypa u Q Uu tas SN ub OS asus bua 14 4 5 Lageplan Steuer E Karte Component drawing of Control PCB Pos Itemi0270 a Fait ee onte aa un eh Mass 15 4 6 Blockschaltbild Block diagram 64 157 108 16 4 7 Verdrahtungsplan Wiring diagram 64 156 251
2. 17 4 8 Pneumatikplan Pneumatic diagram 64 280 021 18 5 Protokolle Reports eee Rl e Ix an 19 5 1 Pr fprotokoll Test Report 19 5 2 Wartungsprotokoll Maintenance report 20 SM 2233 WOLF 1 General information 1 1 Ordering spare parts The part number required for ordering a spare part is listed in the repair parts list under the item number used in the service manual CG IMPORTANT Specify the following numbers when ordering spare parts e Part number of the spare part e Model type number of the device e Serial number of the device Example Repair parts list 0030 72311 112 Kaltger teeinbaubuchse 0160 64 291 022 Geh usedeckel beschichtet Top cover epoxy coated Exploded view 1 SM 2233 WOLF 1 2 Disassembly of housing cover Required tools Phillips screwdriver Unscrew the 4 housing feet 1 2 fig 1 Slide trim plates 1 3 from the rails 1 4 fig 2 Remove the 4 bottom Phillips screws 1 5 fig 2 Pull the housing cover 1 0 as indicated by the arrow a until the tongues 1 8 are disengaged from the corresponding grooves in the front plate 1 1 fig 3 Lift off the housing cover 1 0 as indicated by the arrow b fig 3 Disconnect the earthing ground cable 1 9 fig 3 SM 2233 2 WOLF 2 Maintenance 2 1 Important notes This
3. Important general instructions for use Ensure that this product is only used as intended and described in the instruction manual by ade quately trained and qualified personnel and that maintenance and repair is only carried out by au thorized specialized technicians Operate this product only in the combinations and with the accessories and spare parts listed in the instruction manual Use other combinations accessories and spare parts only if they are ex pressly intended for this use and if the performance and safety requirements are met Reprocess the products before every application and before returning them for repair as required by the instruction manual in order to protect the patient user or third parties Subject to technical changes Due to continuous development of our products illustrations and technical data may deviate slightly from the data in this manual CAUTION USA only Federal law restricts this unit to be used or sold except under the supervision of a medical doctor Safety instructions and levels of danger Symbol Level of danger WARNING Failure to observe can result in death or severe injury CAUTION Failure to observe can result in slight injury or damage to the product IMPORTANT Failure to observe can result in damage to the product or surrounding NOTE Tips for optimum use and other useful information SM 2233 0 DEUTSCH Inhalt 1 22 22 2 EES
4. 0910 0220 5 5 A ddns jewod 8 qod jojjuoo E id 16 ui S L WALZLANLTWHOS I H x x 9x x 02 LO T 5 eI EIL Blei xj E TS MOOTETILNAA Er zr aunssaud SQ uols a nssaud 2279 1909 NaaogasnyH395 17 WOLF Pneumatikplan Pneumatic diagram 64 280 021 4 8 8JDMDJOH r gt 13 01110 gt gt isd 77 JDQ om N E Sd A P d t nl A 0 jualjod ndino m 2 u lu liDd 1 E T l unz Bunbsny Ls doJd A A ja la eanssejg 4 bunjnyu EA IL A SM 2233 WOLF 5 Protokolle Reports 5 1 Pr fprotokoll Test Report Durchgef hrte Pr fung Serien Nr Name Unterschrift Test carried out Serial No Name Signature 19 SM 2233 5 2 Wartungsprotokoll Maintenance report Wartungsma nahmen Serien Nr Servicing check up Serial No WOLF Name Unterschrift Name Signature SM 2233 20 21 GERMANY RICHARD WOLF GmbH D 75438 Knittlingen Pforzheimerstr 32 Tel 49 0 7043 35 0 Fax 49 0 7043 35300 MANUFACTURER E mail info richard wolf com In
5. Every 10 minutes a cyclic test is carried out releasing CO2 gas through the relief valve Flow control Connect hygiene filter with insufflation tube to the patient output Pre select a nominal pressure of 25 mm Hg on the LAPARO PNEU automatic insufflator Actuate the INSUFFLATION button the button must light up Read the current flow on the digital display b Current flow 1 1 l min to 1 5 l min Connect the flow tube CO 1 10 l min to the insufflation tube Actuate the high flow button the button must light up Pre select a nominal flow of 5 liters per minute b Current flow in accordance with digital display 4 l min to 6 l min b Current flow in accordance with flow tube 4 l min to 7 l min Pre select a nominal flow of 8 liters per minute Read the current flow on the digital display and on the flow tube b Current flow in accordance with digital display 7 l min to 9 l min b Current flow in accordance with flow tube 7 l min to 10 l min Switch off insufflation Consumption indicator Actuate the RESET button The digital display for the gas volume consumed reads 0 0 SM 2233 6 WOLF Designation Required checks Pressure control Pre select a nominal pressure of 15 mm Hg and a nominal flow of 10 liters per minute Connect the pressure gauge and the insufflation tube via the Y fitting to the abdomen model Actuate the INSUFFLATION button and then the HIGH FLOW button After a
6. gt 2 l Pin9 GND Pin 10 SDA Pin 11 GND Pin 12 SCL Pin 13 GND 3 5 2 Demontage und Montage der Baugruppe Frontrahmen komplett ausbauen und E Karte wechseln 3 5 3 Funktionspr fungen Voreingestellte Karte einbauen und Funktion vollst ndig berpr fen entsprechend Abschnitt 2 6 und GA A 223 SM 2233 10 ENGLISH Contents 1 General information 2 55 e 12 i uu ee nie 1 1 1 Ordering spare parts Br ETIN be ee eee ae 1 1 2 Disassembly of housing 2 2 Maintenance ann 3 2 1 Important Le csse a heuer EE 3 2 2 Maintenance of device and accessories 3 2 3 Y s al Check EE 4 2 4 Electrical safety test to EN 60601 1 4 2 5 Measuring Instrumentation and other Equipment 5 2 5 1 Measuring set up for overpressure simulation 5 2 6 Eunction Check censet ep een 6 2 6 1 Alarm safety functions 6 2 6 2 EPO Oe HR Ee KEE Res M S eU PRAE YER usss 8 3 Modules aus E 9 3 1 Switching power supply item 0160 9 3 1 1 Disassembly and assembly of 9 3 2 Valve block item 0170
7. 1 1 Bestellung von Ersatzteilen eee eben an ee bg es 1 2 Demontage der Geh useabdeckung 2 Wartung ul a ee ah 2 1 Wichtige Ma EE 2 2 Wartung von Ger t und Zubeh r 2 3 Sichtpr fung s ect aree stets nm 2 4 Elektrische Sicherheitspr fung nach EN IEC 60601 1 2 5 Mess und Hilfsmittel Rm 2 5 1 Messaufbau berdrucksimulation 000 000 cece eee e 2 6 Funktionskontrolle e Eed tei eee een erkenne 2 6 1 Alarm Sicherheitsfunktionen 2 6 2 5 ui resonet RT Cus Ls pa Re 3 Baugruppen 3 1 Schaltnetzteil Pos 0160 n 3 1 1 Demontage und Montage der Baugruppe 3 2 Ventilblock Pos 0170 eet repe a pet recepti Muted eR ER CIE EIN Mate 3 2 1 Demontage und Montage der Baugruppe 3 2 2 Funktionspr fungen 3 3 Druckminderer Einheit Pos 0180 3 4 Bedien E Karte Pos 0210 3 4 1 Demontage und Montage der Baugruppe 3 4 2 Funktionspr fungen 3 5 Steuer E Karte Pos 0220 3 5 1 Messpunkte und Versorgungsspannungen 3 5 2 Demontage u
8. WOLF Service Manual LAPARO PNEU Automatic Insufflator 2233 WOLF Wichtige allgemeine Anwendungshinweise Das Produkt nur bestimmungsgem und unter Beachtung der Gebrauchsanweisung durch ent sprechend ausgebildetes und qualifiziertes Fachpersonal einsetzen Wartung und Reparatur nur durch autorisierte Fachkr fte Das Produkt nur in den Kombinationen und mit dem Zubeh r und den Ersatzteilen betreiben die in der Gebrauchsanweisung angegeben sind Andere Kombinationen Zubeh r und Verschlei teile nur dann verwenden wenn diese ausdr cklich f r die vorgesehene Anwendung bestimmt sind und Leistungsmerkmale sowie Sicherheitsanforderungen nicht beeintr chtigen Die Produkte vor jeder Anwendung und R cksendung zum Schutz von Patient Anwender und Drit ten entsprechend der Gebrauchsanweisung aufbereiten Technische nderungen vorbehalten Durch Weiterentwicklungen k nnen Abbildungen und Technische Daten geringf gig abweichen Struktur der Sicherheitshinweise Bildzeichen Klassifizierung der Gef hrdung WARNUNG Das Nichtbeachten kann zum Tod oder zu schwersten Verletzungen f hren VORSICHT Das Nichtbeachten kann zu leichten Verletzungen oder zu Sch den am Produkt f hren r gt WICHTIG Das Nichtbeachten kann zu Sch den am Produkt oder in der Umgebung f hren HINWEIS Anwendertips f r eine optimale Ger tenutzung und sonstige n tzliche Informationen
9. anschluss lt 0 1 Ohm Ohne Netzanschlu leitung lt 0 1 Ohm Schutzleiteranschluss lt 0 2 Ohm lt 0 2 Ohm Mit Netzanschlussleitung Erdableitstrom Patientenableitstrom Zus tzliche Hinweise b lt 170uA lt 104A Messpunkte ER Messung am Insufflationsanschluss Pr fbedingung Pr fbedingung Imess 25 10 Vo lt 6 V tpr f 5 s bis 10s 50 Hz 60 Hz Gepr ft werden Der Widerstand zwischen Schutzleiterkontakt bzw Schutzleiter stift im Netzstecker und jedem anderen schutzleiterverbundenen ber hrbaren metallischen Teil lt 500 uA N C Normalzustand Typ BF lt 100 uA b N C Normalzustand Typ CF lt 10 uA Messanordnung MD und Messaufbau nach EN IEC 60601 1 Gepr ft werden Der Ableitstrom der durch den Schutzleiter bzw vom Anwen dungsteil ber den Patienten zur Erde flieBen kann Den Insufflationsschlauch vom Ger t trennen SM 2233 4 2 5 2 5 1 LAPARO CO PNEU 2233 Mess und Hilfsmittel WOLF F r die Durchf hrung der in dieser Service Manual beschriebenen Wartungs Reparatur und Testroutinen sind nachfolgend aufgef hrte Hilfs und Messmittel unerl lich Vergewissern Sie sich vor einer Reparatur dass die Messmittel kalibriert und in einem einwandfreien Zustand sind Zu Beachten Angegebene Testkonfigurationen sind ausschlie lich f r Testzwecke definiert und keine Anwendungsempfehlung Bezeichnung Multimeter Flow
10. ble to withstand 30 mm Hg e g an empty 3 liter infusion bag can be used 4171 111 8170 101 64 202 147 50 ml approx 1m 4x 2mm for silicon tubes GA A 223 Measuring set up for overpressure simulation Hygiene filter Syringe Connector djs Silicon tube 5 o o a Y fitting 5 SM 2233 WOLF 2 6 Function check 2 6 1 Alarm safety functions Designation Required checks Preparation Connect the gas cylinder with pressure reducing valve and open the stop valve During the test the cylinder pressure should be in the normal range The green LED of the cylinder pressure indicator is lit The test must be stopped and the cylinder changed if the cylinder pressure indicator changes from continuous light green to blinking light during the test Switch on condition self test While the self test is in process The cylinder pressure indicator is lit and the flow indicator counts up The self test takes may 1 minute The end of the self test is acknowledged by a brief sound signal The digital actual pressure indicator displays the current software version 2 digits Upon completion of the self test After successful completion of the self test the digital displays show the follo wing readings gas volume consumed 0 0 flow 0 to 0 5 I min and intraabdomi nal pressure 0 1 mmHg There must be no error message EO1 to E14 on the gas volume consumed indicator
11. ce The valves cannot be replaced individually Function test Check the full range of functions after replacing the modules in accordance with Section 2 6 and GA A 223 Pressure reducing valve unit item 0180 The unit is supplied in pre set condition Control PCB item 0210 Disassembly and assembly of module Remove the front frame assembly and replace the PCB Function tests Install pre set card PCB and check full range of functions in accordance with Section 2 6 and GA A 223 9 SM 2233 3 5 Control PCB item 0220 H 2 HWH H m H H H H H M WOLF PFD 12V 0 4V GND 24V GND 12V 0 4V GND 5V 5V DS 2 low DS 1 high 2V 0 4V GND DM GND 5 GND Heating GND Flow gt 21 min GND SDA GND SCL GND Remove the front frame assembly and replace the PCB Install the pre set card PCB and check the full range of functions in accordance with Section 2 6 and GA A 223 3 5 1 Measuring points and supply voltages Plug X1 Stecker X3 Stecker X5 3 5 2 Disassembly and assembly of modules 3 5 3 Function tests SM 2233 4 1 Anhang Annex Reparaturteile Repair parts Pos 0010 0030 0040 Type Model 72 311 144 64 073 072 72 316 229 Bezeichnung Kombi Filter mit 2 Sicherungen Schlauchanschluss f r COs Pneu Membrandruckschalter 1 10 bar WOLF Designation Combi filter with 2 fuses Hose coupling for CO gt Pn
12. e Positionsnummern siehe Explosionszeichnung For position numbers see exploded view SM 2233 4 2 Explosionszeichnung Exploded view RICHARD WOLF 0030 0060 0010 0040 0050 wW D e amp H 8 0270 0420 0180 O 0250 0160 0230 0240 0260 5 alll 0100 0090 0080 0220 0210 0200 0190 SM 2233 12 4 3 Lageplan Schaltnetzteil Component drawing of Switching Power Supply Pos Item 0160 Voltage Adjust 13 SM 2233 WOLFA 4 4 Lageplan Bedien E Karte Component drawing Operation PCB Pos Item 0210 SM 2233 14 WOLF 4 5 Lageplan Steuer E Karte Component drawing of Control PCB Pos Item 0270 15 SM 2233 WOLF Blockschaltbild Block diagram 64 157 108 4 6 jsneuxe Bunynnug juened jueneg WNZ eui sure BunuuedszieN isd 9 1 4eq 8 0100 Jo eur suren Si an en Isd 9 1eq 8 IjueAsyeuleuoiS UT Isd 6 isd s osuas J
13. endem Selbsttest Die Flaschendruck Anzeige leuchtet und die Flow Anzeige wird hochge z hlt Der Selbsttest dauert maximal 1 min Am Ende des Selbsttest erfolgt ein kurzer Ton Auf der Digitalanzeige f r den Istdruck wird der aktuelle Software Stand 2 Stellen angezeigt Nach abgeschlossenem Selbsttest Nach bestandenem Selbsttest erscheint auf den Digitalanzeigen f r ver brauchte Gasmenge 0 0 f r Flow bis 0 5 l min und f r intraabdominalen Druck 0 1 mmHg Es darf an der Anzeige f r verbrauchte Gasmenge keine Fehlermeldung E01 bis E14 angezeigt werden Alle 10 min wird der zyklische Test durchgef hrt wobei CO gt Gas durch das Entl ftungsventil abgeblasen wird Flowregelung Hygienefilter mit Insufflationsschlauch am Patientenausgang anschlieBen Am LAPARO CO PNEU einen Solldruck von 25 mmHg vorw hlen Taste INSUFFLATION bet tigen die Taste muss leuchten Auf der Digitalanzeige aktuellen Flow ablesen b Aktueller Flow 1 1 l min bis 1 5 l min Die Flowr hre 1 10 l min an den Insufflationsschlauch anschlie en Die Taste High Flow bet tigen die Taste muss leuchten Einen Sollflow von 5 l min vorw hlen Aktueller Flow laut Digitalanzeige 4 l min bis 6 l min b Aktueller Flow laut Flowr hre 4 l min bis 7 l min Einen Soll Flow von 8 l min vorw hlen Aktuellen Flow von der Digitalan zeige und von der Flowr hre ablesen Aktueller Flow laut Digitalanzeige 7 l min bis 9 l mi
14. equ 26 Jeq E a nssaud Josues LA Ina ZA LA USJOSU9S MO 4 yoniq 9 Je Je d d n n n Bel sd py 1 d sa Josues 55 Jonuo2 e1eMpueH ueJosues 09L0 S n u Bunysemusqn ee 0120 yoed jewod Buryoums Ireyzjeuyeu s Jejouo2 qrt 0220 19 god jeneis x3 uoneiedo god ueipeg 16 SM 2233 WOLF 099 02 9 c0 0LOvV9 40109 do 3d N 1 0 LOO eent Je ull N39N H3HOIS LIA H3 LTIJZL3N 0120 mee CES DEEN SM 2233 Verdrahtungsplan Wiring diagram 64 156 251 4 7 F 1875 31HVMN3lIO38 i bls us x LHe REI
15. eration Safety relevant labeling Lettering symbols e g warnings Device fuses Cables PCBs Lamp plate light sources only 2 4 Completeness and legibility Completeness and legibility The values specified by the manufacturer on the identification plate Nominal current and melting characteristics For perfect condition position insulation and brittleness Corrosion or other damage In place and easy to read Replace damaged parts immediately Electrical safety test to EN IEC 60601 1 Item to be checked Protective ground earth connection Without power cord Protective ground earth connection With power cord Ground earth leakage current Patient leakage current Additional information Maximum admissible measurement values lt 0 1 Ohm lt 0 2 Ohm b lt 170 uA lt 10 Measuring point On the insufflation connection Disconnect the insufflation tube Limit values to EN IEC 60601 1 lt 0 1 Ohm lt 0 2 Ohm Test conditions Imeas 25 A 1096 Vo lt 6 V ttest 5 sto 10 s 50 Hz 60 Hz To be tested The resistance between protective ground earth contact or protective grounding earth pin in the power plug and other exposed bare metal parts connected to the protective ground earth conductor lt 500 uA N C Normal condition for BF type lt 100 uA N C Normal condition for CF
16. eu Diaphragm pressure switch 0050 74 002 141 Dichtung G 1 4 PA6 Seal G 1 4 PA6 0060 74 007 057 Mini Sicherheitsventil G1 4 Mini safety valve G1 4 0080 64 074 074 Filteranschluss Filter connector 0090 0100 0110 0120 0130 0140 0150 64 113 113 64 064 280 64 010 023 64 070 604 64 204 165 64 127 065 64 010 024 Druckst ck Arretierh lse Geh usedeckel kpl Ger tefu kpl Halteschiene Blende Deckel kpl rot Pressure fitting Retaining sleeve Top cover cpl Equipment foot cpl Support bar Shield Cover cpl red 0160 72 321 394 Schaltnetzteil Power Supply 0170 72 321 395 Ventilblock Valve unit 0170 72 321 395A Ventilblock Valve unit 0180 64 110 364 Druckminderer nachgearbeitet Pressure regulator reworked 0190 0200 0210 0220 0230 0240 64 211 126 64 159 882 64 352 130 64 352 132 71 620 000 Frontrahmen kpl Frontplattenfolie Bedien E Karte Steuer E Karte Silikonschlauch 1 0m 4 x 2 mm Front frame cpl Front plate foil Operation PCB Control PCB Silicon tube 1 0m 4 x 2 mm 0250 71 620 034 Silikonschlauch 1 0 m 2 x 1 5 mm Silicon tube 1 0 m 2 x 1 5 mm 0260 71 620 035 Silikonschlauch 1 0 m 8 x 3 mm Silicon tube 1 0 m 8 x 3 mm 0270 64 226 307 Schlauch druckfest 140 mm Tube pressure proof 140 mm A Reparaturaustausch nur gegen Einsendung der defekten Baugruppe Repair exchange only available by sending in the defective modul
17. flation tube closed Error 06 Pressure measurement defective Error 07 Flow measurement defective Error 08 V3 defective V2 defective pressure measurement defective flow measurement defective Error 09 Invalid nominal value Error 10 Not applicable Error 11 Current pressure gt 30 mm Hg for more than 5 seconds Error 12 General hardware error pressure measurement defective V1 defective flow measurement defective Error 13 Input pressure pressure switch defective Error 14 Electronic pressure switch defective Error 15 Wrong DIL switch coding electric pressure monitoring fast slow or Pressure switch defective SM 2233 8 3 3 1 3 2 3 2 1 3 2 2 3 3 3 4 3 4 1 3 4 2 WOLF Modules WARNING Danger of electric shock Danger of touching live parts when working on open devices which are connected to the power supply mains Any work on the open device while the device is connected to power may only be carried out provided the valid safety regulations and protective measures are observed Switching power supply item 0160 Disassembly and assembly of modules Install pre set card PCB and check complete function in accordance with Section 2 6 and GA A 223 Valve block item 0170 Disassembly and assembly of module Open the union nut locking collar and pull off the outlet tube Unscrew the additional screw connection to the housing bottom plate from un derneath the devi
18. g Doku mentation von erstgemessenen Werten ist der Betreiber selbst verant wortlich Nach einer Reparatur m ssen alle Messwerte Ausgangswerte des Ge r ts anhand der Angaben im Service Manual gepr ft und bei Abweichun gen neu eingestellt werden VORSICHT Das Produkt darf nicht betrieben werden wenn die vorgegebenen Messwerte nicht erreicht oder die Funktionen nicht erf llt werden HINWEIS Alle Wartungs und Pr farbeiten an Ger t oder Zubeh r m ssen doku mentiert werden 3 SM 2233 2 3 Sichtpr fung Benennung WOLF Durchzuf hrende Kontrollen Ger t und Zubeh r Bedienelemente Beschriftung Symbolik Sicherheitsgef hrdende Verschmutzung und allgemeine Sauberkeit Mechanische Besch digung Lose oder fehlende Teile Mechanische Funktion und Freigangigkeit Vollst ndig und gut lesbar Sicherheitsrelevante Aufschriften z B Warnhinweise Sicherungseins tze Verkabelung E Karten Vollst ndig und gut lesbar Auf die vom Hersteller auf dem Typenschild angegebenen Werte Nennstrom und Abschmelzcharakteristik Auf einwandfreien Zustand Sitz Isolation und Br chigkeit Korrosion oder andere Besch digungen Lampenschild Nur Lichtquellen Vorhanden und gut lesbar Besch digte Teile sofort austauschen 2 4 Elektrische Sicherheitspr fung nach EN IEC 60601 1 Benennung Maximal zul ssige Messwerte Grenzwerte nach EN IEC 60601 1 Schutzleiter
19. l Reparaturteileliste 0030 72311 112 Kaltger teeinbaubuchse 0160 64 291 022 Geh usedeckel beschichtet Top cover epoxy coated Explosionszeichnung 1 SM 2233 WOLF 1 2 Demontage der Geh useabdeckung Ben tigtes Werkzeug Kreuzschlitz Schraubendreher 4 Ger tef e 1 2 abschrauben Abb 1 Blenden 1 3 von der Halteschiene 1 4 schieben Abb 2 Die 4 untersten Kreuzschrauben 1 5 entfernen Abb 2 Geh useabdeckung 1 0 in Pfeilrichtung a ziehen bis die Federn 1 8 frei aus der Frontplatte 1 1 ragen Abb 3 Geh useabdeckung 1 0 in Pfeilrichtung b abheben Abb 3 Erdungskabel 1 9 l sen Abb 3 SM 2233 2 2 2 1 2 2 Wartung Wichtige Hinweise Fr WOLF Das Service Manual beschreibt die f r das Produkt festgelegten externen Servicema nahmen WICHTIG Zur Durchf hrung der Servicema nahmen ist die Gebrauchsanweisung des Produktes unbedingt zu beachten Wartung von Ger t und Zubeh r Die Wartung und die Pr fung am Ger t soll zum Schutz des Pr fenden in der angegebenen Reihenfolge durchgef hrt werden Sichtpr fung Elektrische Sicherheitspr fung Funktionskontrolle Die im Abschnitt Elektrische Sicherheitspr fung angegebenen Pr f und Messverfahren beziehen sich auf die Pr fung nach EN IEC 60601 1 Alternativ kann nach DIN VDE 751 gepr ft werden F r die Beachtung der aktuellen Grenzwerte und die Erfassun
20. n b Aktueller Flow laut Flowr hre 7 l min bis 10 l min Insufflation ausschalten Verbrauchsanzeige Die Taste RESET bet tigen Die Digitalanzeige f r die verbrauchte Gasmenge zeigt 0 0 an SM 2233 6 WOLF Benennung Durchzuf hrende Kontrollen Druckregelung Einen Solldruck von 15 mmHg und einen Soll Flow von 10 l min vorw hlen Das Druckmessger t und den Insufflationsschlauch ber das Y St ck am Bauchmodell anschlie en Die Taste INSUFFLATION und danach HIGH FLOW bet tigen Nach kurzer Zeit stellt sich der vorgew hlte Druck von 15 mm Hg im Bauch modell ein Ein kurzes Uberschreiten des Druckes mit anschlieBender Ent l ftung ist zul ssig Toleranz der Digitalanzeige 3 mmHg Abweichung zur Anzeige des Digital Druckmessger tes 3 mmHg Insufflation ausschalten Einen Soll Druck von 20 mmHg vorw hlen Die Taste INSUFFLATION bet tigen Nach kurzer Zeit stellt sich der vorgew hlte Druck von 20 mmHg im Bauch modell ein Ein kurzes Uberschreiten des Druckes mit anschlieBender Ent l ftung ist zul ssig Toleranz der Digitalanzeige 3 mmHg Abweichung zur Anzeige des Digital Druckmessger tes 3 mmHg Gasvorrats berwachung RESET bet tigen Das Absperrventil an der Gasflasche schlieBen Einen Solldruck von 20 mmHg und einen Sollflow von 5 l min vorw hlen Insufflation High Flow einschalten Nach einer kurzen Zeit wechselt die Gasvorratsanzeige von gr
21. n Dauerlicht auf Blinklicht Der Alarmton Medium Alarm wird ausgegeben Entl ftungsfunktion in der Druckquelle z B Spritze an Insufflationsschlauch anschlieBen Betriebsart HIGH FLOW Der LAPARO CO2 PNEU muss den anstehenden berdruck entl ften ohne dass der Alarm Uberdruck Medium Alarm auftritt Insufflation aus Alarm Uberdruck Druckquelle z B Spritze an Insufflationsschlauch anschlieBen Uberdruckschalter Einen Solldruck von 15 mmHg vorw hlen Die Entl ftungs ffnung an dem Geh useboden verschlie en Die Taste Insufflation bet tigen Wird der Solldruck um 4 mmHg berschritten ert nt der Alarm berdruck Alarmton Medium Alarm wird ausgegeben Uberschreitet der Druck im Insufflationsschlauch 30 mmHg so tritt nach 5 s Verz gerungszeit der Alarm Uberdruck 30 mmHg auf Alarmton High Alarm wird ausgegeben Die Alarm LED leuchtet permanent Auf der Digitalanzeige f r die verbrauchte Gasmenge erscheint die Fehlermeldung E11 und die Digitalanzeige f r Druck blinkt Der Druck im Insufflationsschlauch steigt auf ca 50 mmHg an Uberschreitet der Druck im Insufflationsschlauch 45 mmHg so spricht nach einer Verz gerungszeit der Uberdruckschalter an Alarmton High Alarm geht in einen Dauerton ber Die Alarm LED leuchtet permanent Ventil 2 wird abgeschaltet Schaltger usch Druckquelle abnehmen Insufflation ein 7 SM 2233 WOLF Benennung Durchzuf hrende Kontr
22. nd Montage der Baugruppe 3 5 3 Funktionspr fungen 4 Anhang Anrex ug dee dee ese bene 4 1 Reparaturteile Repair parts 4 2 Explosionszeichnung Exploded view 4 3 Lageplan Schaltnetzteil Component drawing of Switching Power Supply Pos Item 0160 ke apre a Mr ae geb Rh 4 4 Lageplan Bedien E Karte Component drawing Operation PCB Pos Item 02 10 zur o ised eile LIES pepe elei uit eques Re a RR 4 5 Lageplan Steuer E Karte Component drawing of Control PCB Pos Itemi0270 a rre eub ee onte UO y eh Mass 4 6 Blockschaltbild Block diagram 64 157 108 4 7 Verdrahtungsplan Wiring diagram 64 156 251 4 8 Pneumatikplan Pneumatic diagram 64 280 021 5 Protokolle Reports eee Rl e Ix an 5 1 Pr fprotokoll Test Report 5 2 Wartungsprotokoll Maintenance report SM 2233 WOLF 1 Generelles 1 1 Bestellung von Ersatzteilen Die zur Bestellung erforderliche Artikelnummer von Ersatzteilen ist in der Reparaturteileliste unter der im Service Manual verwendeten Positions Nummer aufgelistet WICHTIG Folgende Angaben sind bei der Bestellung von Ersatzteilen anzugeben e Artikel Nr des Ersatzteiles e Modell Nr des Ger tes e Serien Nr des Ger tes Beispie
23. ollen Kurzzeitunterbrechung Ger t f r ca 10 Sekunden ausschalten Nach dem Wiedereinschalten m ssen alle zuvor angew hlten Funktionen und Einstellungen noch vorhanden sein Werksseitige Grundeinstellung Einstellung am Ger t Soll Flow 5 l min Soll Druck 15 mm Hg 2 6 2 Fehlercodes Fehlercodes werden im Display Verbrauchte Gasmenge angezeigt Error 01 COs Flasche geschlossen bzw leer Error 02 Tastendruck beim Einschalten Error 03 Spannungsversorgung au erhalb der Toleranz Error 04 V1 defekt Flowmessung defekt V3 defekt Error 05 V2 defekt Insufflationsschlauch angeschlossen Error 06 Druckmessung defekt Error 07 Flowmessung defekt Error 08 V3 defekt V2 defekt Druckmessung defekt Flowmessung defekt Error 09 Sollwert ung ltig Error 10 Entf llt Error 11 Aktueller Druck gt 30 mm Hg f r mehr als 5 Sekunden Error 12 Hardwarefehler allgemein Druckmessung defekt V1 defekt Flowmessung defekt Error 13 Druckschalter Eingangsdruck defekt Error 14 Elektronischer Druckschalter defekt Error 15 Falsche DIL Schalter Codierung elektrische Druck berwachung schnell langsam oder Druckschalter defekt SM 2233 8 WOLF 3 Baugruppen WARNUNG Gefahr durch elektrischen Schlag Bei Arbeiten am ge ffneten und am Netz angeschlossenen Ger t besteht Gefahr durch das Ber hren spannungsf hrender Teile Arbeiten am ge ffeten und unter Spannung stehendem Ger t d rfen nur un
24. r hre Digital Druckmessger t CO Flasche CO Druckschlauch Druckminderer Bauchmodell Hygienefilter Insufflationsschlauch Connector Spritze Silikonschlauch Y St ck Gebrauchsanweisung Messaufbau berdrucksimulation Hygienefilter Zus tzliche Angaben Echteffektiv 1 101 min z B von Platon Flow Management Jays Close Viables internet www platon co uk Erforderliche Aufl sung 1 mm Hg im Druckbereich unterhalb von 50 mmHg z B VDM 100 von Fa ASF M nchen oder GDH 14 AN von Fa Greisinger electronic D 93128 Regenstauf 0 75 kg 2046 961 74 021 029 74 007 054 muss 30 mm Hg standhalten geeignet ist z B ein geleerter 3 Infusionsbeutel 4171 111 8170 101 64 202 147 50 ml 1m 4 2 mm f r Silikonschlauche GA A 223 Connector jm Spritze Silikonschlauch 2 E c gt Y St ck SM 2233 WOLF 2 6 Funktionskontrolle 2 6 1 Alarm Sicherheitsfunktionen Benennung Durchzuf hrende Kontrollen Vorbereitung Die Gasflasche mit Druckminderer anschlie en und das Absperrventil ff nen Der Flaschendruck muss bei der Pr fung im normal Bereich sein Die gr ne LED der Flaschendruck Anzeige leuchtet Wechselt w hrend der Pr fung die Flaschendruck Anzeige von Dauerlicht gr n auf blinkend muss die Pr fung abgebrochen und die Flasche ge wechselt werden Einschaltzustand Selbsttest W hrend lauf
25. service manual describes the external service measures defined for the product IMPORTANT For carrying out the service measures the product instruction manual must be observed 2 2 Maintenance of device and accessories To protect the technician carrying out the inspection the maintenance jobs and inspections must be carried out in the order specified below Visual check Electrical safety test Function check The inspection and measuring procedures listed in the section entitled Electrical safety test relate to the test in accordance with EN IEC 60601 1 Alternatively the test can be carried out in accordance with DIN VDE 751 The responsibility for complying with the current limit values and the doc umentation of any values measured for the first time lies in the hands of the operator After repair all measurement values specified output values of the de vice must be checked as described in the service manual and readjusted if deviations are observed CAUTION Do not use this product if the specified measurement values or functions are not fulfilled CG NOTE Any maintenance or inspection job on the device or its accessories must be documented 3 SM 2233 2 3 Visual check Item to be checked WOLF Check for Device and accessories Controls Safety relevant soiling contamination and general cleanliness Mechanical damage Loose or missing parts Mechanical function and easy op
26. short time the pre selected pressure of 15 mm Hg prevails in the abdomen model It is permissible that the pressure is briefly exceeded and then released Tolerance of digital display 3 mmHg Deviation from reading of digital pressure gauge 3 mmHg Switch off insufflation Pre select a nominal pressure of 20 mmHg Actuate the INSUFFLATION button After a short time the pre selected pressure of 20 mmHg prevails in the abdomen model It is permissible that the pressure is briefly exceeded and then released Tolerance of the digital display 3 mmHg Deviation from reading of digital pressure gauge 3 mmHg Gas supply monitoring Actuate RESET Close the stop valve on the gas cylinder Pre select a nominal pressure of 20 mmHg and a nominal flow of 5 liters per minute Switch insufflation to High Flow After a short time the gas supply indicator switches from green continuous light to blinking light Medium alarm is sounded Relief function in the Connect a pressure source e g syringe to the insufflation tube HIGH FLOW mode The LAPARO COs PNEU automatic insufflator must release the excessive pressure without triggering the overpressure alarm medium alarm Switch off insufflation Alarm overpressure Connect pressure source e g syringe to insufflation tube Overpressure switch Pre select a nominal pressure of 15 mmHg Connect the release opening in the housing bottom panel Actuate the In
27. sufflation button If the nominal pressure is exceeded by 4 mmHg the overpressure alarm is sounded Medium alarm is sounded If the pressure in the insufflation tube exceeds 30 mmHg the overpressure alarm 30 mmHg is sounded with a delay of 5 seconds High alarm is sounded The alarm LED is permanently lit The digital display for the gas volume consumed shows error message E11 and the digital display for pressure blinks The pressure in the insufflation tube increases to approx 50 mmHg If the pressure in the insufflation tube exceeds 45 mmHg the overpressure Switch is triggered with some delay The high alarm changes to a continuous alarm The alarm LED is permanently lit Valve 2 switches off switching sound Remove the pressure source Switch insufflation on 7 SM 2233 WOLF Designation Required checks Brief interruption Switch off the device for approx 10 seconds When switching on again all the pre selected functions and settings must still be active Basic factory setting Settings on the device Nominal flow 5 l min Nominal pressure 15 mm Hg 2 6 2 Error codes Error codes are displayed on the gas volume consumed display Error 01 cylinder closed or empty Error 02 Button is pressed when switching on Error 03 Power supply out of tolerance Error 04 V1 defective Flow measurement defective defective Error 05 V2 defective insuf
28. ter Beachtung der g ltigen Sicherheitsvorschriften und Schutzma nahmen erfolgen 3 1 Schaltnetzteil Pos 0160 3 1 1 Demontage und Montage der Baugruppe Voreingestellte Karte einbauen und Funktion vollst ndig berpr fen entsprechend Abschnitt 2 6 und GA A 223 3 2 Ventilblock Pos 0170 3 2 1 Demontage und Montage der Baugruppe berwurfmutter ffnen und Ausgangsschlauch abziehen Zus tzliche Verschraubung mit dem Geh useboden von der Ger teunterseite aus l sen Ventile k nnen nicht einzeln getauscht werden 3 2 2 Funktionspr fungen Funktion nach Baugruppentausch vollst ndig berpr fen entsprechend Abschnitt 2 6 und GA A 223 3 3 Druckminderer Einheit Pos 0180 O wird voreingestellt geliefert 3 4 Bedien E Karte Pos 0210 3 4 1 Demontage und Montage der Baugruppe Frontrahmen komplett ausbauen und E Karte wechseln 3 4 2 Funktionspr fungen Voreingestellte Karte einbauen und Funktion vollst ndig berpr fen entsprechend Abschnitt 2 6 und GA A 223 9 SM 2233 WOLF 3 5 Steuer E Karte Pos 0220 3 5 1 Messpunkte und Versorgungsspannungen Stecker X1 Pin 1 PFD Pin2 12V 0 4V Pin3 GND Pin4 24V Pin5 GND 6 12V 0 4V Pin7 GND Pin8 5V Pin9 NC Pin10 5V Stecker X3 Pin1 DS 2 low Pin 2 DS 1 high Pin3 2V 0 4V Stecker X5 Pin 1 GND Pin2 5V Pin3 GND Pin4 5V 5 GND Pin 6 Heizung Pin 7 GND 8 Flow
29. ternet www richard wolf com USA RICHARD WOLF Medical Instruments Corp 353 Corporate Woods Parkway Vernon Hills Illinois 60061 Tel 847 913 1113 Fax 847 913 1488 E mail sales amp marketing richardwolfusa com Internet www richardwolfusa com UK RICHARD WOLF UK Ltd Waterside Way Wimbledon SW 17 OHB Tel 020 8944 7447 Fax 020 8944 1311 E mail admin richardwolf uk com Internet www richardwolf uk com BELGIUM N V Endoscopie RICHARD WOLF Belgium S A Industriezone Drongen Landegemstraat 6 B 9031 Gent Drongen Tel 32 9 280 81 00 Fax 32 9 282 92 16 E mail endoscopy richard wolf be FRANCE RICHARD WOLF France S A R L Rue Daniel Berger Z A C La Neuvillette F 51100 Reims Tel 33 3 26 87 02 89 Fax 33 3 26 87 60 33 E mail endoscopes richardwolf fr AUSTRIA RICHARD WOLF Austria Ges m b H Wilhelminenstra e 93 a A 1160 Wien Tel 43 1 405 51 51 Fax 43 1 405 51 51 45 E mail info richard wolf at Internet www richard wolf at WOLF SM 2233
30. type lt 10 uA Test condition Measuring device MD and measuring setup to EN IEC 60601 1 To be tested The leakage current which can flow through the protective ground earth conductor or from the applied part via the pati ent to ground earth SM 2233 4 2 5 2 5 1 LAPARO CO PNEU 2233 automatic insufflator WOLF Measuring Instrumentation and other Equipment In order to carry out the maintenance repair and test routines described in this Service Manual the measuring instrumentation and other equipment listed in the following are compulsory Before you start a repair job make sure that the measuring equipment is properly calibrated and in perfect working condition Please observe the following The test configurations indicated have been defined for testing purposes exclusively and do not represent a recommendation for use Designation Multimeter Additional data True RMS Flow tube Digital pressure meter cylinder CO pressure hose 1 10 min e g by Platon Flow Management Jays Close Viables Internet www platon co uk Required resolution 1 mm Hg in pressure range below 50 mmHg e g VDM 100 by ASF Munich or GDH 14 AN by Greisinger electronic D 93128 Regenstauf 0 75 kg 2046 961 74 021 029 74 007 054 Pressure reducing valve Abdomen model Hygiene filter Insufflation tube Connector Syringe Silicon tube Y fitting Manual must be a

Download Pdf Manuals

image

Related Search

Related Contents

Ford 2004 Escape Automobile User Manual  Untitled - saginomiya global  1 Biblioteca BOLETIN DE ADQUISICIONES Enero  C.A 6030 - Chauvin Arnoux  User`s Manual  BCM50 Telephony Device Installation Guide  Manual de instrucciones Instruções de Utilização Istruzioni d`uso  Global Crossing Ready-Access User Guide    Mitsubishi Electronics DLP WD-57734 User's Manual  