Keysight E5080A ENA Series Network Analyzer

image

Contents

1. Safety Notice Supplement Network Analyzer This equipment complies with EN IEC61010 1 2010 This equipment is of MEASUREMENT CATEGORY CAT I Do not use for CAT Il Ill or IV This equipment is a POLLUTION DEGREE 2 INDOOR USE product This equipment is tested in stand alone condition and in combination with the accessories supplied by Keysight Technologies against the requirement of the standards described in the Declaration of Conformity If it is used as a system component compliance of related regulations and safety requirements are to be confirmed by the builder of the system Do not connect the measuring terminals to mains General Safety Precautions The following general safety precautions must be observed during all phases of operation service and repair of this instrument Failure to comply with these precautions or with specific WARNINGS elsewhere in this manual may impair the protection provided by the equipment Such noncompliance would also violate safety standards of design manufacture and intended use of the instrument Keysight Technologies assumes no liability for the customer s failure to comply with these precautions The E5080A complies with INSTALLATION CATEGORY Il as well as POLLUTION DEGREE 2 in IEC61010 1 The E5080A is an INDOOR USE product The LEDs in the E5080A are Class 1 in accordance with IEC60825 1 CLASS 1 LED PRODUCT Ground the Instrument To avoid electric
2. b 2 November 1995 as applicable in any technical data Safety Notices CAUTION A CAUTION notice denotes a haz ard It calls attention to an operat ing procedure practice or the like that if not correctly performed or adhered to could result in dam age to the product or loss of important data Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met WARNING A WARNING notice denotes a hazard It calls attention to an operating procedure practice or the like that if not correctly per formed or adhered to could result in personal injury or death Do not proceed beyond a WARNING notice until the indicated condi tions are fully understood and met This Note sign denotes important information It calls attention to a procedure practice or condition that is essential for the user to understand Caution Safety Summary N Do not exceed the operating input power voltage and current level and signal type appropriate for the instrument being used refer to your instrument s Function Reference Electrostatic discharge ESD can damage the highly sensitive microcircuits in your instrument ESD damage is most likely to occur as the test fixtures are being connected or disconnected Protect them from ESD damage by wearing a grounding strap that provides a high resistance path to ground Alternatively ground yourself to discharge any static charge built up by
3. E5080A When you start up the E5080A for the first time you need to perform the initial registration of the Windows operating system of the E5080A When the Windows agreement message box appears click Agree The rest of the procedure runs automatically The Touch Screen does not require any calibration and it works from the very first time you use the E5080A 20 Network Analyzer Keysight E5080A Network Analyzer Installation Guide 2 Troubleshooting Troubleshooting during Startup 22 System Recovery 23 This chapter describes the troubleshooting process during start up and the procedure of the operating system OS recovery when the Windows OS has been damaged 21 2 Troubleshooting Troubleshooting during Startup When you encounter problems during start up see Table 1 System recovery saves most of problems Table 1 Troubleshooting during startup Symptom Turning ON the standby switch does not start up the system Solu Conf prop Confi rear ion irm that the power cable is erly plugged in rm that the line switch at the panel is turned ON The system starts up but it automatically shuts down immediately The system starts up but it enters the service mode The status bar at the lower right of the screen displays SVC in red start up but the with Error Messa background in th the screen The measurement screen appears after power on test fails ge 241 prompt
4. Figure 5 System recovery confirmation screen Step 8 The progress of the system configuration is displayed on the screen The recovery takes a few minutes depending on the amount of data Figure 6 System configuration progress screen 25 2 26 Troubleshooting CAUTION Step 9 Once the recovery process is completed completion message prompts Click OK The E5080A restarts automatically Step 10 Software End User License Agreement appears Press Agree I have reviewed and agree to comply with the torms in thes software kcorse agrooment pra Step 11 After the restart execute initial registration For information on the execution procedure refer to Initial Registration of E5080A on page 20 Never turn OFF the power during the system recovery because doing so may cause serious damage to the E5080A Network Analyzer Index A assistance 6 D documentation warranty 6 E environmental requirements operating environments 13 ventilation requirement 13 error correction temperature range 13 exclusive remedies 6 inspection checking the shipment 12 M manual contents 8 P power disconnection from supply source 18 power cable options 15 verification and connection of power cable 15 verification of the power supply 15 S safety symbols 5 shutdown turning the power OFF 17 Network Analyzer 27 Index 28 Network Analyzer www keysight com Key
5. touching the outer shell of any grounded instrument chassis before touching the test port connectors When you notice any of the unusual conditions listed below immediately terminate operation and disconnect the power cable Contact your local Keysight Technologies sales representative or authorized service company for repair of the instrument If you continue to operate without repairing the instrument there is a potential fire or shock hazard to the operator Instrument operates abnormally Instrument emits abnormal noise smell smoke or a spark like light during operation Instrument generates high temperature or electrical shock during operation Power cable plug or receptacle on instrument is damaged Foreign substance or liquid has fallen into the instrument Network Analyzer Manufacturer s Declaration Herstellerbescheinigung GERA USCHEMISSION LpA lt 70 dB am Arbeitsplatz normaler Betrieb nach DIN 45635 T 19 Manufacturer s Declaration ACOUSTIC NOISE EMISSION LpA lt 70 dB operator position normal operation per ISO 7779 Regulatory Compliance Information This product complies with the essential requirements of the following applicable European Directives and carries the CE marking accordingly The Low Voltage Directive 2006 95 EC The EMC Directive 2004 108 EEC To obtain Declaration of Conformity please contact your local Keysight Technologies sales office agent or distributor
6. 1 Environmental Requirements Set up the E5080A where the following environmental requirements are met Operating Environments Ensure that the operating environment meets the following requirements Temperature 0 C to 40 C Temperature range at 23 C 3 C lt 1 C deviation from the temperature the error correction when performing the error correction Humidity 20 to 80 at wet bulb temperature lt 29 C non condensation Altitude 0 to 2 000 m 0 to 6 561 feet Vibration 0 21 G maximum 5 Hz to 500 Hz Above environmental requirements are NOT for the specifications and measurement accuracy of the analyzer but for the operating environment of the analyzer Ventilation Requirement Network Analyzer To ensure safety requirements the specifications and measurement accuracy of the analyzer you must keep the environmental temperature within the specified range by providing appropriate cooling clearance around the analyzer or for the rack mount type by forcefully air cooling inside the rack housing For more information on environmental temperature to satisfy the specifications and measurement accuracy of the analyzer see the specification in the E5080A Datasheet When the environmental temperature around the analyzer is kept within the temperature range of the operating environment specification See Section Operating Environments on page 13 the analyzer conforms to the r
7. Keysight Es080A ENA Series Network Analyzer Installation Guide KEYSIGHT TECHNOLOGIES Notices Keysight Technologies Inc 2015 No part of this manual may be repro duced in any form or by any means including electronic storage and retrieval or translation into a foreign language without prior agreement and written consent from Keysight Technologies Inc as governed by United States and international copy right laws Manual Printing History The manual s printing date and part number indicate its current edition The printing date changes when a new edition is printed minor correc tions and updates that are incorpo rated at reprint do not cause the date to change The manual part number changes when extensive technical changes are incorporated January 2015 First Edition Manual Part Number E5080 90000 Edition First Edition January 2015 Printed by Keysight Technologies International Japan G K 1 3 3 Higa shikawasaki cho Chuo ku Kobe shi Hyogo Japan Warranty The material contained in this document is provided as is and is subject to being changed with out notice in future editions Fur ther to the maximum extent permitted by applicable law Key sight disclaims all warranties either express or implied with regard to this manual and any information contained herein including but not limited to the implied warranties of merchant ability and fitness for a particular
8. and troubleshooting information when the Windows cannot be boot up See this manual first when you use the E5080A for the first time Online Help The online help provides the information about the quick start measurement operation programming and I O interface This is pre installed in the E5080A Press Help hard key on the front panel to open Quick Start helps in understanding the E5080A operation quickly The latest version of online help is available at http ena support keysight com E5080A manuals webhelp eng The online help has context sensitive help which is a great feature of the E5080A help It allows you to get information about the selected softkey by pressing the Help key in the E5080A or by pressing F1 in a keyboard attached to the E5080A or by clicking the help button in a dialog box With context sensitive help users can receive information quickly about the area the user is working in the firmware of the E5080A It provides information relevant to the task that needs to be accomplished and reduces the time to search relevant information required to complete a task Service Guide The service manual provides information about the parts troubleshooting performance test adjustment and service menu In This Guide The following shows the contents of this manual Chapter 1 Installation This chapter provides information about how to set up the Keysight E5080A Chapter 2 Troubleshooting This ch
9. apter describes the troubleshooting during start up and the procedure of the operating system OS recovery when the Windows OS has been damaged 8 Network Analyzer Contents 1 Installation Checking the Shipment 12 Environmental Requirements 13 Operating Environments 13 Ventilation Requirement 13 Protection Against Electrostatic Discharge ESD 14 Ensuring Adequate Free Space around Analyzer for Immediate Disconnection of Power Cable in Case of Emergency 14 Power Supply 15 Verification of the Power Supply 15 Verification and Connection of Power Cable 15 Blown Fuses 16 Starting the E5080A 17 Turning the Power ON and OFF 17 Disconnection from Supply Source 18 Initial Registration of E5080A 20 2 Troubleshooting Troubleshooting during Startup 22 System Recovery 23 Notes on executing the factory recovery function 23 Procedure to execute the factory recovery 23 Index Network Analyzer 9 10 Network Analyzer Keysight E5080A Network Analyzer Installation Guide 1 Installation Checking the Shipment 12 Environmental Requirements 13 Power Supply 15 Blown Fuses 16 Starting the E5080A 17 Initial Registration of E5080A 20 This chapter provides information about how to set up the Keysight E5080A 11 1 Installation Checking the Shipment 12 After you receive the analyzer carry out checks during unpacking according to the following procedure When unpacking the analyzer if the exterior of the analyz
10. aximum power 300 VA consumption Verification and Connection of Power Cable The three wire power cable attached to the E5080A has one wire serving as the ground Using this power cable allows the E5080A to be grounded thereby protecting you against any potential electrical shock from the power outlet Step 1 Confirm that the power cable is not damaged NEVER use a power cable showing any sign of damage Faulty cables can cause electrical shock Step 2 Use the supplied cable to connect between the power cable receptacle Figure 2 on page 18 on the rear panel of the E5080A and a three wire power outlet with the grounding prong firmly connected in the ground slot Use the supplied power cable with grounding wire to securely ground the E5080A Network Analyzer Power cord list 16000 99101 shows the power cable options 15 1 Installation Blown Fuses If the fuse appears to have blown during operation this instrument may be subject to failure and must be repaired For any assistance contact Keysight Technologies Customer contact center listed at the end of this guide The E5080A uses the following fuse types UL CSA Type Slow Blo 10 A 250 Vac WARNING DO NOT replace the fuse yourself as doing this may expose yourself to electrical shock 16 Network Analyzer Starting the E5080A Installation 1 This section describes how to turn ON OFF the E5080A power and how to cut off the power suppl
11. equirements of the safety standard Furthermore under that temperature requirement the analyzer still conforms to the requirements of the safety standard even when placing the analyzer with cooling clearance as follows 13 1 Installation Requirements Back lt 180mm Sides lt 60mm both right and left Protection Against Electrostatic Discharge ESD Set up a static safe work station to protect the electronic components from the damage by the electrostatic discharge ESD as shown in Figure 1 Figure 1 Building Ground 1 MegOhm Resistor Example of the static safe work station Ensuring Adequate Free Space around Analyzer for Immediate Disconnection of Power Cable in Case of Emergency As described in Disconnection from Supply Source on page 18 the power supply is disconnected by removing the power cable s connector plug from either the AC outlet or the E5080A unit When installing the E5080A ensure that there is sufficient free space around the unit to permit quick disconnection of the plug from AC outlet or E5080A unit in case of emergency 14 Network Analyzer Power Supply Installation 1 Before turning on the E5080A power check the following Verification of the Power Supply Confirm that the power supplied to the E5080A meets the following requirements Requirements Voltage 90 to 264 VAC Frequency 47 to 63 Hz M
12. er such as the cover front rear panel LCD screen power switch and port connectors appear to be damaged during transport do not turn on the power switch Otherwise you may get an electric shock Step 1 Check that the packing box or shock absorbing material used to package the analyzer has not been damaged If the packing box or shock absorbing material has been damaged leave the packing box and shock absorbing material as is until other inspection items are checked as follows Step 2 Check the packaged items supplied with the analyzer for any damage or defects Step 3 By referring to the furnished contents list check that all packaged items supplied with the analyzer have been received as per the specified options Step 4 After checking if one of the following applies contact your nearest Keysight Technologies sales and service office 1 The packing box or shock absorbing material used to package the analyzer has been damaged or the shock absorbing material has traces where extreme force has been applied 2 A packaged item supplied with the analyzer has mechanical damage or defects 3 A packaged item supplied with the analyzer is missing 4 A fault has been detected in the subsequent operation check of the analyzer If an abnormality is detected in step 1 contact the company that transported the analyzer as well as your nearest Keysight Technologies sales and service office Network Analyzer Installation
13. his manual Instructions contained in the warnings must be followed WARNING Dangerous voltage levels capable of causing death are present in this instrument Use extreme caution when handling testing and adjusting this instrument Safety Symbols General definitions of safety symbols used on the instrument or in manuals are listed below Instruction Manual symbol the product is marked with AN this symbol when it is necessary for the user to refer to the instrument manual Alternating current Direct current On Supply Off Supply A chassis terminal a connection to the instrument s chassis which includes all exposed metal structure Standby G o Network Analyzer Certification Keysight Technologies certifies that this product met its published specifications at the time of shipment from the factory Keysight Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology to the extent allowed by the Institution s calibration facility or by the calibration facilities of other International Standards Organization members Documentation Warranty The material contained in this document is provided as is and is subject to being changed without notice in future editions Further to the maximum extent permitted by applicable law Keysight disclaims all warranties either express or implied with regard to this man
14. ine Switch ON and never turn it If you directly interrupt the power supply to the power cable receptacle when the power supply is turned ON or turn OFF the Line Switch the shutdown process will not work This could damage the software and hardware of the E5080A and lead to device failure Turning ON the power supply after a faulty shutdown may cause the system to start up in a condition called safe mode If this occurs first shut down the system to set it to the standby state and then turn ON the power supply again to start up the system in normal mode Power Cable Receptacle Line Switch Always ON I Figure 2 Line switch Always ON and power cable receptacle from Supply Source The power supply of the E5080A is cut off by disconnecting the plug of the power cable on either AC outlet side or E5080A side When it is necessary to disconnect the power supply in order to avoid shock hazards etc pull out the power cable plug from either the AC outlet side or the E5080A side Network Analyzer Installation 1 To allow this operation to be performed smoothly be sure to follow the guidelines in Ensuring Adequate Free Space around Analyzer for Immediate Disconnection of Power Cable in Case of Emergency on page 14 When turning the power OFF under normal circumstances always follow the methods described in Turning the Power OFF on page 17 Network Analyzer 19 1 Installation Initial Registration of
15. ing on a red e instrument s message warning area at the lower left of Exec ute the system recovery 22 Network Analyzer System Recovery CAUTION Troubleshooting 2 By executing system recovery you can return the E5080A s system the Windows operating system and the firmware to the factory state at the time of purchase Strictly follow the steps described below If you do any operation other than the following steps the system may not be recovered Notes on executing the factory recovery function Executing the factory recovery function causes the following conditions In addition to the Windows operating system and the firmware the following settings of the E5080A are returned to the factory state Network setting GPIB setting Printer setting The driver for the supported printer which is installed after purchase is deleted You need to execute initial registration again Files that you have created using the save function files in the D drive are not affected However Keysight recommends backing them up before executing system recovery for precautionary purposes For more information on backup refer to Backing Up the Data as described in E5080A Online Help Procedure to execute the factory recovery Network Analyzer This section describes how to return the contents of the C drive to the factory state You need a keyboard for this operation 1 The default
16. purpose Keysight shall not be liable for errors or for incidental or consequential damages in con nection with the furnishing use or performance of this document or of any information contained herein Should Keysight and the user have a separate written agreement with warranty terms covering the material in this doc ument that conflict with these terms the warranty terms in the separate agreement shall control Technology Licenses The hardware and or software described in this document are fur nished under a license and may be used or copied only in accordance with the terms of such license Restricted Rights Legend If software is for use in the perfor mance of a U S Government prime contract or subcontract Software is delivered and licensed as Commer cial computer software as defined in DFAR 252 227 7014 June 1995 or as a commercial item as defined in FAR 2 101 a or as Restricted com puter software as defined in FAR 52 227 19 June 1987 or any equiva lent agency regulation or contract clause Use duplication or disclosure of Software is subject to Keysight Technologies standard commercial license terms and non DOD Depart ments and Agencies of the U S Gov ernment will receive no greater than Restricted Rights as defined in FAR 52 227 19 c 1 2 June 1987 U S Government users will receive no greater than Limited Rights as defined in FAR 52 227 14 June 1987 or DFAR 252 227 7015
17. setting of the hard disk is recovered during the system recovery For the hard disk that had failed earlier and had been replaced after purchase its default setting may not be the same as the factory state 23 2 Troubleshooting Step 1 Shut down the E5080A Step 2 Disconnect all of the USB device from the USB ports Step 3 Connect the keyboard to the E5080A Step4 Press the standby switch of the E5080A to turn it ON Step 5 When the screen as shown in the figure below appears select Instrument Image Recovery System and press Enter windows Boot Manager choose an operating system to start To specify an advanced option for this choice ENTER Choose ESC Exit Figure 3 Instrument Image Recovery System selection screen After several seconds the next screen appears automatically even if you do not press any key 24 Network Analyzer Network Analyzer Troubleshooting 2 Step 6 Type 2 to select Recover the original factory system image option and press Enter coverySystem vbs Figure 4 System recovery selection screen Step 7 A confirmation message appears Click OK to continue the process Restoring the factory image will overwrite all the data on the C partition Other partitions will not be affected Do NOT interrupt power before the restoration process has completed Do you wish to continue the restore process OK Cance Press the Enter key for OK and lt tab gt then Enter for Cancel
18. shock the instrument chassis and cabinet must be grounded with the supplied power cable s grounding prong DO NOT Operate in an Explosive Atmosphere Do not operate the instrument in the presence of inflammable gasses or fumes Operation of any electrical instrument in such an environment clearly constitutes a safety hazard Keep Away from Live Circuits Operators must not remove instrument covers Component replacement and internal adjustments must be made by qualified maintenance personnel Do not replace components with the power cable connected Under certain conditions dangerous voltage levels may remain even after the power cable has been disconnected To avoid injuries always disconnect the power and discharge circuits before touching them DO NOT Service or Adjust the Instrument Alone Do not attempt internal service or adjustment unless another person capable of rendering first aid and resuscitation is present DO NOT Substitute Parts or Modify the Instrument 4 Network Analyzer To avoid the danger of introducing additional hazards do not install substitute parts or perform unauthorized modifications to the instrument Return the instrument to a Keysight Technologies Sales and Services Office for service and repair to ensure that safety features are maintained in operational condition Dangerous Procedure Warnings Warnings such as the example below precede potentially dangerous procedures throughout t
19. sight Technologies Inc 2015 Printed in Malaysia First edition January 2015 E5080 90000 KEYSIGHT TECHNOLOGIES
20. ual and any information contained herein including but not limited to the implied warranties of merchantability and fitness for a particular purpose Keysight shall not be liable for errors or for incidental or consequential damages in connection with the furnishing use or performance of this document or any information contained herein Should Keysight and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms the warranty terms in the separate agreement will control Exclusive Remedies The remedies provided herein are Buyer s sole and exclusive remedies Keysight Technologies shall not be liable for any direct indirect special incidental or consequential damages whether based on contract tort or any other legal theory Assistance Product maintenance agreements and other customer assistance agreements are available for Keysight Technologies products For any assistance contact your nearest Keysight Technologies Sales and Service Office Addresses are provided at the back of this manual Network Analyzer Manuals for E5080A Network Analyzer Keysight provides the following three manuals for E5080A The latest version of all documentations can be downloaded from http www keysight com find E5080A manual Installation Guide The installation guide this manual provides start up setup information when you use the E5080A for the first time
21. y in the case of emergency Turning the Power ON and OFF Network Analyzer The standby switch can turn ON OFF the E5080A The color on the button shows the status as indicated below Indicator Color Description Green Normal power on status Orange Standby status Red Illegal power on status Perform the following steps to turn the power ON or OFF Turning the Power ON Step 1 Confirm if the Line Switch on the rear panel is turned ON The switch should always be turned ON Step 2 Press the standby switch once It changes to green color This operation turns ON the power and the E5080A starts the self test Step 3 Confirm that the self test indicates normal operation Normal operation is confirmed by the self test if no error message appears Turning the Power OFF Step 1 To turn OFF the power of the E5080A first press this standby switch or send a shutdown command from the external controller to activate the shutdown process the process of software and hardware necessary to turn OFF the power supply This puts the E5080A into the standby state and the button changes to orange color Step 2 Next if necessary turn OFF the power supply to the power cable receptacle Line Switch on the rear panel 17 1 18 Installation Disconnection Under normal use never directly interrupt the power supply to the power cable receptacle on the rear panel when the power supply is turned ON Always keep the L