Appendix Quick Installation Guide for TXR Generators

image

Contents

1. Table Generator PROGENCY TRUE SPEED BUCKY pre wired Terminal Strip TB3 on board in center Wire Terminal Strip Wire Color of cabinet Color Left Side of Strip marked table B1 Blue 1 Blue B2 Orange 2 Orange B3 White 3 White B4 Yellow N Yellow B8 Brown L Brown Any labeled earth ground in cabinet Green Ground Green Terminal Strip TXR 96 mounted Table Lock Printed Circuit Board right side of control cabinet Left Table Tab Red 24 VAC Red Right Table Tab Black 24 VAC Black Note Red and Black continue through bucky junction box Both terminate at table lock printed circuit board Spare Purple Spare Purple Wall Stand Generator PROGENCY TRUE SPEED BUCKY pre wired Terminal Strip TB3 Wire Terminal Strip Wire Color Right Side marked wall Color B1 Blue 1 Blue B2 Orange 2 Orange B3 White 3 White B4 Yellow N Yellow B8 Brown L Brown Any labeled earth ground in cabinet Green Ground Green Terminal Strip TXR 96 mounted on Wall Stand Lock Printed Circuit Board right side of control cabinet Left Table Tab Red 24 VAC Red Right Table Tab Black 24 VAC Black Note Red and Black continue through bucky junction box Both terminate at wall stand lock printed circuit board Spare Purple Spare Purple Technical Pamphlet W 2009 7 20 31 Tingle X Ray Products 1 Technical Pamphlet TXR Standard Frequency Generators Pictur
2. 2 tube 2 Bucky 2 no lon Chamber 1 0 0 tube 1 No Bucky no lon Chamber No Bucky AP 0015R2 11 HF Series Generators Appendix Quick Installation Guide for TXR Generators 6 Continue Configuration for the remaining Workstations T Exit from configuration by pressing the 200 and 800 pushbuttons down simultaneously for at least two seconds until you hear an audible double beep This sound signals that you have successfully saved the workstations you had selected Configuration of the Workstations is now completed 2 3 CONFIGURING THE TUBE TYPE T To configure the X ray Tube type it is necessary to enter into Calibration Mode by pressing the 200 and 800 push buttons simultaneously The light of the selected workstation blinks Bucky 1 Bucky 2 or No Bucky kVp mAs mA EG Time off On Bu AEC Film Speed g FocusS FocusL Ready X Ray AE PREP EXPOSE Density faa WARNING T ACA A MEET RE CLTC NC REDE 2 Press the mA Increase push button until you reach the maximum mA capacity Press the mA Increase push button twice more to reach the Extended Memory Code E02 that will appear on the mA Display Density 3 In order to configure the X ray Tube type and model press the or Density push buttons until one of the Tube Type numbers listed bel
3. REGION1 VIEW3 PROJECT1 MINIMUM THICKNESS 3 cm DEFAULT THICKNESS 5 cm THICKNESS STEP cm ENTER TECHNIQUE 28 AP 0015R2 Series Generators Appendix Quick Installation Guide for TXR Generators 10 Modify the Radiographic Technique Values In order to modify the Radiographic Technique values for a specific Region View Projection press the Display Selector close to ENTER TECHNIQUE d g REGIONI VIEW3 PROJECT1 m g MINIMUM THICKNESS 3 cm DEFAULT THICKNESS 5 cm Enter Selection and go to next screen THICKNESS STEP cm Go to previous screen 6 ENTER TECHNIQUE lt The Display will show the Region View Projection and the Default Thickness highlighted for the selected Region View Projection Press Enter to edit the APR technique for the displayed Thickness or before pressing Enter modify the Thickness with the Up Down Edit Arrows ifthe APR technique to be modified corresponds to another Thickness value REGIONI VIEW3 PROJECT1 Chahge TECK I Thickness value CANCEL Cancel and go to previous screen After pressing Enter the Console automatically displays all the radiographic parameters kVp mA mAs Time selections Workstation and AEC and SID indication related to that Region View Projection Now the operator can modify and store the Radiographic Parameters Workstation AEC Selections and SID indication 36 40 60 or 72 selected w
4. Mn CLR PORE mh f SW1 Jumper T1 Terminal Strip 2 AP 0015R2 Series Generators Appendix Quick Installation Guide for TXR Generators 4 TXR products require 24 VAC for locks At top of the Generator cabinet a 24 VAC Distribution Board has been added to connect these locks rte us s E S Rn E BELA A Em 4 5 6 7 8 24VAC OVAC 24VDC OVDC 24VDC OVDC Terminals 1 and 2 are in coming 24VAC Collimator is connected to terminals 3 and 4 5 Remove Allen screws that secure hinged door on MODS side AP 0015R2 3 HF Series Generators Appendix Quick Installation Guide for TXR Generators 6 The Terminal Strip 4TS2 is located on Module 4 back of Module 3 of the hinged door Connect the X ray Tube Rotor wiring as follows WIRE FROM X RAY TUBE CONNECTION IN 4TS2 BLACK RED WHITE THERMAL THERMAL GROUND f X ray Tube have not a shielded Ground wire terminal 4TS2 6 is not used 4TS2 Terminal Strip T Carefully connect the J5 side of the Serial cable to the Console and then route the Handswitch wire Connect the J1 side of the Serial cable from the Console to the Generator Cabinet Connection of M A Connection of Serial Cable Serial Cable du TOREM UT Ics at Generator Cabinet at Console P t located on upper back 4 AP 0015R2 Series Generators Appendix Quick Installation Guide for TXR Generators 8
5. 3 contains eight Dip Switches for factory use Swicth 2 Swicth 3 Swicth 1 Configuration of Switch 1 For use with 60 Hertz Line Supply All four Dip Switches must be placed in the down OFF position For use with 50 Hertz Line Supply The first Dip Switch must be placed in the up ON position The second third and fourth Dip Switches must be placed in the down OFF position Configuration of Switch 2 The first third and fourth Dip Switches must be placed in the up ON position The second Dip Switch must be placed in the down OFF position Configuration of Switch 3 All eight Dip Switches must be placed in the down OFF position AP 0015R2 HF Series Generators Appendix Quick Installation Guide for TXR Generators 2 2 CONFIGURING THE WORKSTATIONS With the Generator OFF make sure Dip 3 of Switch 2 on the ATP Console CPU Board Module 4 side is placed in the closed ON position Service Mode enabled Turn ON the Line Power and then the Console power Wait for the self testing to finish then enter in configuration mode by maintaining pressed Power ON 200 and 800 simultaneously for at least two seconds until all of the workstations push buttons are illuminated FocusS FocusL Ready X Ray Density Si S PREP EXPOSE WARNING m P ATENCION cs Results you should see A display of numbers will be visible on the kVp mAs mA and
6. Section Page 1 INSTALLATION rrr RR rev eee 1 2 CONFIGURATION 21 l i kg RR E Ruhm ee 9 2 1 Proper Dip Switch Selections on the ATP Console CPU Board 9 2 2 Configuring the Workstations 10 2 3 Configuring the Tube Type 12 3 CALIBRATION ee t eed eee eee 15 3 4 Previous TaSKS oss odere SEDI ae S oe Ros See aeos Sae eR s 15 3 2 Digital kV Loop 16 3 3 Digital mA Loop Closed sssssesssseseee eh 17 3 4 Auto calibration of Digital mA Loop Open 17 3 5 FinalChecks ee cee GR EXC ROS UE A pde 22 4 APR EDITING obs eus 23 5 BUCKY INTERFACE CONNECTIONS 31 AP 0015R2 i HF Series Generators Appendix Quick Installation Guide for TXR Generators ii AP 0015R2 Series Generators Appendix Quick Installation Guide for TXR Generators SECTION 1 INSTALLATION THIS DOCUMENT PROVIDES A STEP BY STEP GUIDE AND IS LT WARNING NOT INTENDED TO TAKE THE PLACE OF THE SERVICE MANUAL AS STEPS ARE COMPLETED CHECK OFF THE LIST TO THE LEFT OF EACH NUMBER OR PARAGRAPH 1 Place the Power Module close to its permanent position For easy interconnection and calibration place the
7. Series Generators SECTION 3 3 1 PREVIOUS TASKS Appendix Quick Installation Guide for TXR Generators CALIBRATION Before calibration keep in mind that To calibrate and measure the kVp output a Non Invasive kVp Meter is needed Place and center a Non Invasive kVp Meter on the X ray Tube output at the required SID refer to the Non Invasive kVp Meter documentation To calibrate and measure the mA or mAs a mAs Meter needs to be plugged to the banana plug connection at the top of the HV Transformer The cross bar link plate should not be connecting the banana plug Turn the Generator OFF to connect the mAs meter if you did not connect one at the beginning Remember to re install the cross bar between the banana plug connections after removing the mAs meter Remove the cross bar link and plug the mAs meter here AP 0015R2 15 HF Series Generators Appendix Quick Installation Guide for TXR Generators 3 Before calibration procedure verify that the Dip Switches of Switch 2 of the HT Controller Board A3000 xx are in the OFF position This board is located on Module 3 at the hinged door Switch 2 on the HT Controller Board 3 2 DIGITAL KV LOOP OPEN Extended Memory Location E06 contains the calibration factor for Digital kV Loop Open Note gt The Value in E06 Memory Location is only related to the Generator performance it is not related to the X ray Tube s or another
8. Spot second group EACH TIME THAT AUTO CALIBRATION IS ACTIVATED IN A n WARNING mA STATION OR IN E01 MEMORY LOCATION ALL THE FILAMENT CURRENT NUMBERS OF THE SELECTED FOCAL SPOT ARE AUTOMATICALLY SET TO 344 SO A NEW COMPLETE CALIBRATION OF THE FILAMENT CURRENT NUMBERS FOR THIS FOCAL SPOT WILL BE REQUIRED AP 0015R2 17 HF Series Generators Appendix Quick Installation Guide for TXR Generators Auto calibration starts with the minimum available mA station for the selected Focal Spot at 50 kV and follows with the other combinations of mA stations for the selected Focal Spot at 80 kV 120 kV and 40 kV 1 For this procedure set Dip Switch 4 of Switch 2 on the HT Controller Board back in the ON position Digital mA Loop Open Filament Current Constant 1 Select the No Bucky workstation Direct 2 Enter calibration mode by pressing the 200 and 800 push buttons simultaneously and select one of the configured mA stations for the Small Focal Spot Light will be blinking for workstation selected if you are in Service Mode Off On g Bucky ij Bucky No Film Speed 5 X Ray Density PREP EXPOSE 7x WARNING CEA ATENCION con 18 AP 0015R2 HF Series Generators Appendix Quick Installation Guide for TXR Generators 3 Enter Auto calibration mode by pressing the Power On and kV increase
9. or Exposure Controls to stop momentarily the Auto calibration procedure The alarm will stop when the X ray Tube begins to cool and recovers the Heat Units capacity Exposures can be performed again even though code 777 is shown on the Console At this point it is recommended to wait until the Heat Units available are close to 8096 of the X ray Tube capacity without making any exposure The Generator will try to calibrate each kV mA combination for ten 10 attempts maximum If calibration is aborted after ten attempts code 888 will flash on the Console until the Power On push button is pressed to exit Calibration Calibration might also abort due to space charge during calibration of the lowest kV at the highest mA stations for the Focal Spot selected so code 777 will be flashing on the Console until the Power On push button is pressed IF AUTO CALIBRATION IS ABORTED CODE 888 OR 777 CONTINUE AUTO CALIBRATION FOR THE OTHER FOCAL SPOT CHECK AT THE END OF AUTO CALIBRATION WHICH KV MA COMBINATIONS HAVE NOT BEEN AUTO CALIBRATED FOR EACH FOCAL SPOT THESE COMBINATIONS HAVE THE FILAMENT CURRENT NUMBER SET TO 344 MANUALLY CALIBRATE THESE KV MA COMBINATIONS AS EXPLAINED IN THE SERVICE MANUAL CALIBRATION CHAPTER MANUAL CALIBRATION OF DIGITAL mA LOOP OPEN When Auto calibration is successfully performed code 999 will be flashing on the Console until press the Power On push button for 2 seconds
10. previous screen or press the Display Selector close to CANCEL to go back to previous screen without saving the changes d MINIMUM THICKNESS mes Edi DEFAULT THICKNESS Thickness CANCEL lt a Cancel and go to previous screen THICKNESS STEP AP 0015R2 27 HF Series Generators Appendix Quick Installation Guide for TXR Generators Note gt Minimum Thickness MT setsthe minimum value in centimeters used for the selected Region View Projection Default Thickness DT J setsthe default value in centimeters according to the operator preferences Thickness Step TS sets the number of centimeters between each thickness value Each Region View Projection may have a maximum of twenty 20 Steps The maximum range for a Thickness value is 99 cm which means that a Minimum Thickness value too high or a Thickness Step too large will not be allowed if the resulting value is over 99 cm Calculate if the Thickness values are allowed by applying the formula MT 19 x TS where MT Minimum Thickness 19 Maximum number of Steps minus one TS Thickness Step value For example in the figure below the formula is 3 19 x 2 44 cm In which 3 is the Minimum Thickness 19 is the number of Steps 20 1 and 2 is the Thickness Step value 44 is the resulting maximum Thickness but if the Thickness Step value is 8 it would not be allowed as the range is over 99 cm 84 19 x 8 168
11. rename other Views g REGION1 VIEW VIEW1 VIEW2 VIEW3 VIEW4 Edit name VIEW5 VIEW6 VIEW7 VIEW8 Cancel and go to previous screen ENTER VIEW cime Press the Display Selector close to CANCEL to go back to previous screen without saving the changes or press the Display Selector close to ENTER VIEW to go to step 6 AP 0015R2 25 HF Series Generators Appendix Quick Installation Guide for TXR Generators Enter Selection and go to next screen 71 Enter Selection and go to next screen 71 Projection Once the desired Projection highlighted is chosen with the Edit Arrow buttons if the operator requires to modify the Projection go to step 7 If not press the Display Selector close to ENTER PROJECTION to go to step 8 or BACK to return to the previous d d d PROJECT1 PROJECT2 PROJECT3 screen PROJECT4 PROJECT5 Go to previous screen ENTER PROJECTION Modify the Projection name Press Enter to edit the name of the selected Projection The first letter of the name is highlighted the Up Down Edit Arrow buttons will change the letter in alohanumerical order plus underscore and blank the Right Left Edit Arrow buttons will scroll in the word Once the name is written press Enter to save the new Projection name If necessary repeat this process to rename other Projections REGION1 VIEW3 PROJECT PROJECT1 PROJECT2 PROJECT3 Edit name Cancel and go
12. the PC APR Editor Refer to Section 4 of the Operating Manual Proceed in the following way APR 1 With the APR deactivated press simultaneously APR ON OFF and Ll m Enter buttons All the LEDs of the Edition Module buttons lights up Enter On Off Edition Module with Edit Arrow buttons and Enter button d REGIONI Messen oi REGION3 REGION4 m d REGIONS REGION6 REGION7 REGIONS Display Selector D ENTER REGION The Edition Module is operated with the Edit Arrow buttons the Enter button and the Display Selectors Press the Edit Arrow buttons to scroll in the APR Display When pressing the Display Selector on the left lower corner the APR Display will show sequentially the selected Region View Projection atthe upper line AP 0015R2 23 HF Series Generators Appendix Quick Installation Guide for TXR Generators 2 Body Region Once the desired Body Region highlighted is chosen with the Edit Arrow buttons if the operator requires to modify the Region name go to step 3 If not press the Display Selector close to ENTER REGION to go to step 4 g g g REGION1 EKAL REGION3 REGION4 Enter Selection and go to next screen vL ENTER REGION REGION5 REGION6 REGION7 REGION8 3 Modify the Region name Press Enter to edit the name of the selected Region The first letter of the name is highlighted the Up Down Edit Arrow buttons will change the letter
13. time displays The visible numbers will either be O 1 2 etc The Console is equipped to operate in five different modes workstations The three push buttons atthe upper left provide a way to program for different operations See the following tables for proper selections 10 AP 0015R2 Series Generators 4 Appendix Quick Installation Guide for TXR Generators Selectthe first workstation to be configured Bucky 1 Bucky 2 No Bucky by pressing the respective push button only this push button blinks and the console shows one of the following values DISPLAY FUNCTION VALUE DESCRIPTION No configured workstation 15 Value SELECT THE TUBE Tube 1 Tube 2 Direct No Bucky Bucky 1 SELECT THE DEVICE WORKING MODE 2nd Value Bucky 2 Standard Tomo Standard RF No AEC lon Chamber 1 SELECT THE ION 3rd Value CHAMBERS BEING lon Chamber 2 USED ONLY WITH AEC lon Chamber 3 lon Chamber 4 Notes Some values may not be configurable depending on Generator model 1 Only when the Tomo is controlled by the Generator Set the new value by pressing the Density or buttons located at the bottom left corner ofthe Operator Console Whenthe first value is O the second and third value are 0 The following table is an example of a 2 tube System with 2 Buckys and no lon Chambers Bucky 1 1 Quy 1 no lon NE 2 2 0
14. to revious screen PROJECT4 PROJECT5 p ENTER PROJECTION CANCEL ET Press the Display Selector close to CANCEL to go back to previous screen without saving the changes or press the Display Selector close to ENTER PROJECTION to go to step 8 26 AP 0015R2 Series Generators Appendix Quick Installation Guide for TXR Generators 8 Once the Display Selector close to ENTER PROJECTION has been pressed the APR Display shows The name of the Region View Projection upper line The Thickness values for Minimum Thickness Default Thickness and Thickness Step that might be applied to the selected Region View Projection If required these Thickness values can be edited as described below g g REGION1 VIEW3 PROJECT1 m g MINIMUM THICKNESS 3 cm DEFAULT THICKNESS 5 cm Enter Selection and go to next screen THICKNESS STEP cm Go to previous screen ENTER TECHNIQUE lt 9 Modify the Minimum Thickness the Default Thickness or the Thickness Step If the operator decides to modify a Thickness value follow this step if not go to next step Press the Edit Arrow button to highlight the corresponding line and press Enter button to modify the selected Thickness MT minimum DT default or TS thickness step appears atthe right corner of the Display then select the desired number with the Edit Arrow buttons and press Enter to save the changes and go back to
15. to exit from Auto calibration mode The 999 code will disappear when you have successfully exited Calibration To calibrate the Large Focal Spot select one of the configured mA stations for the Large Focal Spot by pressing the mA push button Repeat steps 2 5 Before starting the exposures it is recommended to wait until the Heat Units available are closed to the 8096 of the X ray Tube capacity 20 AP 0015R2 Series Generators 10 Table 3 1 mA Calibration Numbers Appendix Quick Installation Guide for TXR Generators After performing both procedures for Small and Large Focal Spots select in calibration mode each combination of the available mA stations at the kV break points 40 50 80 and 120 kV Press the Reset push button to read on the kV Display the new value of the Filament Current Number stored for each combination and write down the new values in the Data Book Note that the highest mA station for Small Focal Spot may have numbers larger than the lowest mA station for Large Focal Spot This is normal Repeat the above calibration process for the second tube Exit calibration mode Set Dip Switch 4 of Switch 2 on the HT Controller Board back inthe OFF position Digital mA Loop Closed FILAMENT CURRENT NUMBERS AT kV BREAK POINT mA STATION 50 80 Note Some Generator models do not contain all the mA statio
16. Add insulating compound or insulating oil to both High Voltage Transformer receptacles Connect both High Voltage cables Anode and Cathode from the X ray Tube to the HV Transformer receptacles 9 The Terminal Strip 3TS1 is located on Module3 out side of hinged door Verify or connect a jumper between terminals 3TS1 22 and 3TS1 23 to bypass the Door Interlock Terminal Strip 3TS1 10 THIS POINT ONLY APPLIES TO LINE POWERED GENERATORS Remove the plastic cover over the Line Fuses and connect the Line Power cables to the right side of the Fuses at L1 amp L2 for Single Phase Generators and also at L3 for Three Phase Generators Plastic Cover over Line Fuses 1 Input Transf 6T2 PREQQ UCION Esto GND Stud L1 Input Line L2 Input Line L3 Input Line Single Phase Neutral Stud optional AP 0015R2 5 HF Series Generators Appendix Quick Installation Guide for TXR Generators 11 THIS POINT ONLY APPLIES TO BATTERY POWERED GENERATORS a Connector 6J10 on the right side of Generator Cabinet Module 6 is unplugged for safety during transport Plug in it Connector 6J10 b Connect the Line Power cable to Line Connector wire 1 to left wire 2 to middle ground wire to right These connections are color coded Line Connector Circuit Breaker 6 AP 0015R2 Series Generators Appendix Quick Installation Guide for TXR Generators c TurnOnthe C
17. Appendix Technical Publication AP 0015R2 Quick Installation Guide for TXR Generators HF Series Generators Series Generators Appendix Quick Installation Guide for TXR Generators REVISION HISTORY REVISION DATE REASON FOR CHANGE 0 JAN 7 2004 First edition 1 OCT 16 2007 General Update amp Improvement 2 AGO 06 2009 Dip Switch configuration of HT Controller This Document is the English original version edited and supplied by the manufacturer The Revision state of this Document is indicated in the code number shown at the bottom of this page ADVISORY SYMBOLS The following advisory symbols will be used throughout this manual Their application and meaning are described below DANGERS ADVISE OF CONDITIONS OR SITUATIONS THAT 7 ANGER IF NOT HEEDED OR AVOIDED WILL CAUSE SERIOUS PERSONAL INJURY OR DEATH DATA Advise of conditions or situations that if not heeded or avoided could cause personal injury or damage to equipment or data ADVISE OF CONDITIONS OR SITUATIONS THAT IF NOT 0 WARNING HEEDED OR AVOIDED COULD CAUSE SERIOUS PERSONAL INJURY OR CATASTROPHIC DAMAGE OF EQUIPMENT OR CAUTION Note gt Alert readers on pertinent facts and conditions Notes represent information that is important to know but which do not necessarily relate to possible injury or damage to equipment AP 0015R2 Series Generators Appendix Quick Installation Guide for TXR Generators TABLE OF CONTENTS
18. Generator Cabinet on a table or pedestal or mount it to the wall using the Wall Support provided 2 Verify that wire marked on Transformer 6T2 terminal strip is matched to the incoming power Identify the Generator type Single Phase Three Phase or Battery Powered POWER SUPPLY LINE FOR SINGLE PHASE OR BATTERY POWERED GENERATORS 110 VAC or Stand Alone option 208 VAC 220 230 VAC 240 VAC Cable gt in Transformer 6T2 TB 3 or TB 8 TB 21 TB 4 or TB 5 TB 6 or TB 7 POWER SUPPLY LINE FOR THREE PHASE GENERATORS 220 230 240 VAC 400 VAC 415 VAC 440 VAC 480 VAC Cable gt in Transformer 6T2 TB 4 or TB 5 TB 8 TB 7 TB 9 TB 10 6T2 Terminal Strip AP 0015R2 1 HF Series Generators Appendix Quick Installation Guide for TXR Generators 3 THIS POINT ONLY APPLIES TO BATTERY POWERED GENERATORS a Open hinged door to access to Charger Module Mod 1 by removing the Allen screws b Verify that wire marked gt is connected to Transformer 2T1 terminal strip 110 for 110 VAC c Remove plastic covers over Charger Board and Line Monitor Board d Plug J1 into the Charger Board it is unplugged at the factory for safety e Verify Jumper SW1 connection between SW1 5 and SW1 4 110 VAC f Re install plastic covers Plastic Covers Charger Board Line Monitor Board j CAUTION F Gv gt Y 7112211 FEEFEERETR EI d
19. components installed this Memory Location is factory adjusted Only perform this procedure if the HT Controller Board and or HV Transformer have been replaced in the unit For that refer to the Service Manual Calibration Chapter 16 AP 0015R2 Series Generators Appendix Quick Installation Guide for TXR Generators 3 3 DIGITAL mA LOOP CLOSED Extended Memory Locations E03 and E05 contain the calibration factor for Digital mA Loop Closed Note gt Values in E03 and E05 Memory Locations are only related to the Generator performance they are not related to the X ray Tube s installed These Memory Locations are factory adjusted Only perform this procedure if the HT Controller Board and or HV Transformer have been replaced in the unit For that refer to the Service Manual Calibration Chapter 3 4 AUTO CALIBRATION OF DIGITAL mA LOOP OPEN To achieve the most accurate calibration this procedure has to be automatically performed by the Generator Auto calibration The Calibration procedure should be manually performed if Auto calibration is not possible In this case refer to the Service Manual Calibration Chapter Section Manual Calibration of Digital mA Loop Open Auto calibration of the Filament Current Numbers is divided in two separated procedures related to the mA stations configured for the Small or Large Focal Spots Start with the Small Focal Spot first group and continue with the Large Focal
20. es y j Terminal Strip TB3 P e Terminal Strip TXR 96 ale t Rear View Rear Panel Removed 32 Technical Pamphlet W 2009 7 20 Tingle X Ray Products LLC Technical Pamphlet High Frequency Generators SHF Series Table Generator PROGENCY TRUE SPEED BUCKY pre wired Terminal Strip 3TS1 Wire Terminal Strip Wire Color located on right side of swing out door Color 3TS1 5 Blue 1 Blue 3TS1 6 Orange 2 Orange 3TS1 4 White 3 White 3TS1 3 Brown L Brown 3TS1 1 Yellow N Yellow Any labeled earth ground in cabinet Green Ground Green Terminal Strip TXR 96 mounted on Table Lock Printed Circuit Board top of control unit Left Table Tab Red 24 VAC Red Right Table Tab Black 24 VAC Black Note Red and Black continue through bucky junction box Both terminate at table lock printed circuit board Spare Purple Spare Purple NOTE Add following jumpers in generator 1 Jump 3TS1 2 to 3TS1 3 2 Jump 3TS1 1 to 3TS1 27 Wall Stand Generator PROGENCY TRUE SPEED BUCKY pre wired Terminal Strip 3TS1 Wire Terminal Strip Wire Color located on right side of swing out door Color 3TS1 12 Blue 1 Blue 3TS1 13 Orange 2 Orange 3TS1 11 White 3 White 3TS1 10 Brown L Brown 3TS1 8 Yellow N Yellow Any labeled earth ground in cabinet Green Ground Green Terminal Stri
21. in alohanumerical order plus underscore and blank the Right Left Edit Arrow buttons will scroll in the word Once the name is written press Enter to save the new Region name If necessary repeat this process to rename other Regions g R g REGION1 Eixe Koy REGION3 REGION4 Enter Selection and REGIONS REGION6 REGION7 REGIONS go to next screen 610 ENTER REGION Press the Display Selector close to ENTER REGION to go to step 4 Edit name 24 AP 0015R2 Series Generators Enter Selection and go to next screen 103 Enter Selection go to next screen u Appendix Quick Installation Guide for TXR Generators Anatomical View Once the desired Anatomical View highlighted is chosen with the Edit Arrow buttons if the operator requires to modify the View name go to step 5 If not press the Display Selector close to ENTER VIEW to go to step 6 or BACK to return to the previous Screen g g g VIEW2 VIEW5 VIEW6 Go to previous screen ENTER VIEW Modify the View name Press Enter to edit the name of the selected View The first letter ofthe name is highlighted the Up Down Edit Arrow buttons will change the letter in alohanumerical order plus underscore and blank the Right Left Edit Arrow buttons will scroll in the word Once the name is written press Enter to save the new View name If necessary repeat this process to
22. ircuit Breaker Visually check that all green lights and LED 051 are lit on the Charger Board 1 LED DS1 Connector J1 Row of Green LEDs Charger Board veec000000000606 d Visually verify that LED DL1 yellow is flashing on the Line Monitor Board 2A3 The MAX and MIN LEDs red on the Line Monitor Board check the line coming in neither should be on Potentiometer POT1 is factory adjusted If MAX or MIN LED is ON first check that Mains Voltage is according to specifications on page 1 ofthis guide If Mains Voltage is correct adjust POT1 CW up CCW lower measuring between TP 1 and TP 2 on Line Monitor Board Potentiometer adjustment should be Mains Voltage Potentiometer x 2 5 Nominal Voltage Potentiometer Max and Min Lights Line Monitor Board AP 0015R2 7 HF Series Generators Appendix Quick Installation Guide for TXR Generators This page intentionally left blank 8 AP 0015R2 Series Generators SECTION 2 Appendix Quick Installation Guide for TXR Generators CONFIGURATION 2 1 PROPER DIP SWITCH SELECTIONS ON THE ATP CONSOLE CPU BOARD Note gt ATP Console Board The ATP Console CPU Board is located on Module 4 side of the Hinged Door Locate Switches 1 42 and 3 onthe ATP Console CPU Board from the illustration n Switch 1 contains four Dip Switches Switch 2 contains four Dip Switches Switch
23. ith the Up down Edit Arrows AP 0015R2 29 HF Series Generators Appendix Quick Installation Guide for TXR Generators Keep in mind thatall Thickness values of a Projection will share the same AEC values Workstations and SID that means that if one of these Parameters is modified it will affect to all Thickness values for that specific Region View Projection AEC Workstations and SID are common values for all Thickness in a Projection mAs Time FocusS FocusL REGION1 VIEW3 PROJECT1 THCK 3 eo Ready X Ray onc PREP Note F If an APR technique is to be stored with parameters suitable back up time and or mAs MUST be stored in the program by the operator 11 Press Enter to save in memory the new parameters and selections assigned to the selected Region View Project and Thickness value 12 Press simultaneously APR ON OFF and Enter buttons to exit from the APR edition mode 30 AP 0015R2 Tingle X Ray Products LLC Technical Pamphlet INTERFACE CABLE Bucky to X Ray Control A nine conductor 18 AWG cable is provided to supply control and power wiring for both the bucky and the table or wall stand electric locks All bucky s are pre wired at the table or wall stand Standard Frequency Generators Models 425 amp 525
24. ns listed above AP 0015R2 21 HF Series Generators Appendix Quick Installation Guide for TXR Generators 3 5 FINAL CHECKS Verify that all Configuration and Calibration data have been properly stored in memory Enter in calibration mode and check that the values noted for the Filament Current Numbers and Extended Memory Locations of the Data Book are the same values displayed and stored in memory Press the Reset button to read the stored values Exit from calibration mode and Service mode Turn the Generator OFF and verify position of Dip Switches of Switch 2 on the HT Controller Board are Dip Switch 2 in Off position enables Filament and Rotor Interlocks Dip Switch 4 in Off position Digital mA Loop Closed Set Dip Switch 3 of Switch 2 on the ATP Console CPU Board in Off position to place the Generator in normal operating mode 22 AP 0015R2 Series Generators Appendix Quick Installation Guide for TXR Generators SECTION 4 APR EDITING Any APR Technique may be modified to fit the operator requirements It is possible to change and store in memory the names Body Region Anatomical View Projection SID Thickness value range and default value selections Workstation and AEC and radiographic parameters related to selected Thickness value Note gt Creating or Deleting an Body Region Anatomical View Projection and Thickness can only be performed from
25. ow is shown on the kVp Display The entire list of Tube Data is found in your Service Manual on the last page of the Configuration Section Normally one of the following three Tube capacities is chosen TUBE NUMBER TUBE CODE ID FOCAL ROWER RATINGS 5 MODEL kVp Display mAs Display SPOT LS kW HS kW 001 139 TOSHIBA E7239X 133 003 140 TOSHIBA E7242X 187 004 090 TOSHIBA E7252X 26 73 300 12 AP 0015R2 Series Generators Appendix Quick Installation Guide for TXR Generators Verify that the Tube code ID shown on the mAs Display is the same as the tube code listed in the table above or in the Tube Table of the Service Manual Press the Rese Push button to save and permanently store the X ray Tube configuration You will once again hear an audible beep to confirm that you have successfully saved your selection AEC Film Speed we Density EA WARNING ATENCION FocusS Focus L Ready X Ray PREP EXPOSE Note F At this point if your selected Bucky is still blinking you are already in Service Mode and ready for Autocalibration If your selected Bucky is not blinking press 200 and 800 simultaneously for at least two seconds to enter Service Mode AP 0015R2 13 HF Series Generators Appendix Quick Installation Guide for TXR Generators This page intentionally left blank 14 AP 0015R2
26. p TXR 96 mounted on Wall Stand Lock Printed Circuit Board top of control unit Left Table Tab Red 24 VAC Red Right Table Tab Black 24 VAC Black Note Red and Black continue through bucky junction box Both terminate at wall stand lock printed circuit board Spare Purple Spare Purple NOTE Add following jumpers in generator 1 Jump 3751 9 to 3751 10 2 Jump 3TS1 8 to 3TS1 27 Technical Pamphlet W 2009 7 20 33 Tingle X Ray Products 1 Technical Pamphlet High Frequency Generators Pictures amp KS M S E REXU gt Terminal Strip TXR 96 A Terminal Strip 3TS1 Cover Removed 34 Technical Pamphlet W 2009 7 20
27. push buttons simultaneously for 2 seconds After releasing both push buttons code 222 will be flashing on the Console waiting for confirmation to then enter Auto calibration mode Press the kV increase and Power push buttons again until code 222 disappears on the Console to confirm Auto calibration mode Auto calibration is activated after releasing both push buttons At this moment the Generator will check the mA stations available for the Small Focal Spot When 50 kV at 10 mA is displayed in the display windows the Unit is ready for Auto calibration mA BB Focus S FocusL Tm PREP EXPOSE Density Z T WARNING v ATENCION ren 4 Check that the Heat Units available for the X ray Tube are 100 kV Display shows H on the Console 5 Keep the Handswitch push button fully pressed or use the Exposure Controls on the Console to perform continuous exposures Auto calibration can be stopped momentarily by releasing the Handswitch push button or the Exposure Controls Do not exit from Auto calibration before the procedure has been completed AP 0015R2 19 HF Series Generators Appendix Quick Installation Guide for TXR Generators If the Heat Units available for the X ray Tube are less than 40 exposures are inhibited momentarily and code 177 will be flashed on the Console accompanied by an alarm In this case release the Handswitch