1. 2 BY al ae Jal 9 YY Y Y Medical Equipment Fault Detection MASH 209 A Eshra Spring 2011 12 2 Semester Course Instructor Ayman Elsayed Eshra Assistant Prof Dr Eng Tel 9664 352 401 Room 24A 7 3 Office hours see time table E mail aeshra ksu edu sa Textbooks 1 Carr J J amp Brown J M Introduction to biomedical equipment technology 4 ed ch 16 pp 427 2 Khandpur R S Handbook of Biomedical Instrumen tation 250 ed McGraw Hill 2008 3 User and Service Manuals of Medical Devices 4 www ecri org 5 www who org 6 www youtube com Grading amp Assessment Assignments 10 Reports 10 Quizzes 10 Midterm 20 Experimental Test 10 Final 40 Be able to read and understand the Service s and User s Manuals Be able to use biomedical hand tools Be able to use multimeter Voltmeter Ammeter Ohmmeter 4 Be able to use Oscilloscope To make sure that the medical equipment is safe in proper working order Equipment should be inspected to ensure that it is calibra
2. P den fete hem P po 0 03 sec 6 YY Typical faults fall into the following categories Patient electrode connection problems high impedance connections to the scalp or broken electrode wires Cable connection problems broken wires and bent connector pins Incorrect switch position operator error or broken knob indicators Broken switches faulty switch contacts Graphic recorder malfunctions drive roller slipping or ink pens clogged or unseated Electronic malfunctions circuit faults in individual channels system control or power supply 1 Md The EEG technician should perform a routine inspection procedure before daily use as follows Machine turned on to warm up Calibration set usually at 100 uV and rectangular pulse observed on all channels pen recorder Sensitivity set system and individual channel for proper deflection corresponding to 100 UV Pressing and holding calibration switch set to observe time constant decay All inputs grounded to observe zero signal on all channels 1 2 Symptom machine runs but j Solutions the tracing on one or more 1 For dry ink reservoirs fill to level suggested by manufacturer usually below top rim To overfill causes messy operation and can damage circuitry and mechanisms if allowed to drip into the machine For clogged ink tubes remove the tube and pen and soak in
3. an internal calibration control can be used to check the instrument electronics and to distinguish problems with it from those related to the electrodes Problems with the EEG are compounded by the fact that the signal levels are a thousand times smaller than the ECG potentials Furthermore there are multiple channels A problem with one of the channels can be isolated to eliminate the possibility of problems with elements common to all channels such as the power supply Electrical or electronic and mechanical faults rarely occur in clinical EEG machines Y To isolate a problem in the diff amp of particular channel and its corresponding electrode the calibration switch can be used If the channel under test can be calibrated with the switch this would indicate that the electronics beyond it is working and that the problem would probably be in a lead 4 The frequency response of filters can be checked by measuring the output of the filter due to a variable frequency input The most vulnerable element in active filters is the diff amp chip Failure in the chip would cause a radical change in the frequency response However if the changes in frequency response are small the problem may be a leaky capacitor or aging or thermal damage in the components d i ea pui T Veni Ar fe Mix maj OE epar 3 o w y im SS nume E ee qe Poe ge j ii i 0 1sec pirati
4. that a component is faulty is that changing the part either changes the symptoms of the problem or causes it t disappear Y Strain gauge gt Troubles in this transducers are affected by mechanical contact with the patient Elimination of these problems is usually the responsibility of the equipment operator Problems with the balancing bridge and the excitation voltage are similar to those affecting the thermistor Because the operation of pressure transducer often depends on the measuring pressure transmitted through a fluid column especial in INBP any air bubbles in the line seriously degrade both the transducer sensitivity and its frequency response Bubbles often can be seen visually and flushed and that must be done when the transducer is removed from the patient to remove any hazard of transferring bubbles into the bloodstream Calibration switch t provides a handy troubleshooting aid fit works properly it is clear that the electronics beyond it to output is working and that a failure if present has occurred before it in the signal path gt The following Box gives troubleshooting tips for an ECG Possible Artifacts Check the following Base line with no Wave Trace switch ON and gain control set high enough Readjust as required Form _ Select appropriate lead lead wires and patient cable fully inserted into proper receptacle Cable or lead wires damaged check
5. warm water Use a fine wire to gently push the clog through Be certain not to punch a hole in the tube For bent pens remove the pen in question and gently bend the pen downward Be careful not to bend at right angles as these pens are delicate and will crack channels Is missing Possible causes 1 Ink reservoirs for pens are dry on missing channels 2 Ink tubes are clogged 3 Pen is not touching Troubleshooting machine off 1 Check ink reservoirs 2 Check ink tubes for clogging Check for upwardly bent pens gently push pen onto paper with finger or pencil to observe any touching Symptom Spotty recordings light or dark Possible causes 1 Worn pens or incorrectly loaded paper Troubleshooting 1 Check paper loading 2 And if proper then check pen for worn tip ink not feeding properly solutions 1 For paper loading perform manufacturer s procedure 2 For worn pen tip replace with manufacturer s part or equivalent Grounded all EEG leads and check for straight line tracing noiseless and if good connect an EEG simulator if available Check for good tracings If noise appears on the trace the problem is properly inside the machine Refer to the service manual for troubleshooting Solutions 1 For patient connection physically inspect all electrodes and connectors to the machine For machine problem internal repair will be necessary Symptom No
6. blem Check to see if it had reoccurred before looking for other problems gt Logical analysis of given evidence Data relating to the problem is gathered and used to isolate the case analytically Because circuit theory is basic to design of medical equipment it could be used to deduce every problem with the hardware Systematic approach to troubleshooting uses both methods Repair procedure will involve systematic disassembling and reassembling of the equipment gt To disassemble the equipment number each part as you remove it Then to reassemble replace the parts in the reverse order in order to be sure you are putting all the parts back together correctly Make test point measurements Analyze data for board level problem isolation Make measurements using the block diagram Analyze data for board level and ancillary component isolation Make measurements using module schematic diagrams Analyze data for component level problem isolation When problem is cleared assess the need for preventive maintenance 11 Open equipment casing 12 Make a module level assessment 13 Check cable interconnections 14 Analyze symptoms for board level problem isolation Use specialized diagnostic aids maintenance manuals and diagnostic software Check shop records for equipment history Analyze data for problem isolation Interview the manu
7. facturer s technical representative by phone Assess the equipment environment with visual auditory and Look for operator error E 4 E 5 ESSA 6 i 7 lu 8 Em Look for operator error Make an assessment at the system level interconnections Check equipment Use equipment troubleshooting aids Check diagnostic error messages Execute equipment lt A flow diagram outlining logical steps to troubleshooting 4 Follow the steps indicated until the problem is identified A general troubleshooting flowchart 4 Step 1 all available resources should be used emergency or hazard are involved Step 2 interview the person who reported the problem 4 Step 3 system consists of units Unit might b e different types of equipment ECG pressure Unit consists of modules particular functions Modules consist of transducer and another the signal processing section of the unit The modules consists of circuit boards and discrete elements and circuit boards themselves consist of discrete elements he troubleshooting process is the effort to find the discrete element or elements that have failed The logical process is to start at the system level and work down to the discrete element 4 Step 4 interconnecting cable breakage loosening and corrosion Early checking 4 Step 9 many medical instruments display error messages that aid in troubleshooting 4 Step 13
8. he monopolar active output and the patient return are located just above the center The keying switches on the electrodes feed into the upper right corner of the schematic Optical coupling isolates the electrodes as do transformers on both the monopolar and bipolar electrodes The troubleshooting guide is given in form of common symptoms and suggested remedies stated in terms of the block schematic modules An outline of some symptoms and possible remedies follows gt Symptom there are no RF output from the monopolar ESU electrode gt Possible causes High voltage power supply Fuse blown Surgical pencil open circuit v Symptom there are RF output in all modes except one v Possible causes If the bipolar is at fault the bipolar control module should be checked If a monopolar mode has failed the clock control module should be checked Symptom the power output in the monopolar modes is more than 30 below normal Possible causes Some components in the power amplifier stripline and output modules may have failed Symptom The display is dead Possible causes The low voltage power supplies may be faulty vr
9. isy or poor recording Possible causes 1 Lead connection or electronic or mechanical problems Troubleshooting 1 Place selector switches to Standard calibration position and check for noise and improper operation If calibration operation is normal the problem is properly the patient connection Problems show up during routine testing procedures Procedures fro its use must be fail safe Safety of defibrillator is ensured by frequent testing and by training of the operator Test procedures for the following defibrillator block diagram are done with defibrillator analyzer gt Provides a 50 Q resistance discharge path for the defibrillator paddles that stimulate the torso resistance The defibrillator electrodes are placed on the metal test pads Energy levels of discharges into the test pads are measured up to 1000 J The discharge of this discharge may be displayed on an oscilloscope gt The defibrillator analyzer generates an R wave of an ECG pattern at approximately 60 bom and it then displays the time between the leading edge of the R wave and the cardioverter discharge gt This measurement detects failures in the QRS detector and the 30 ms delay circuit in the block diagram gt n the defibrillator mode the R wave would be used along with the energize switch to ensure that the defibrillator properly inhibits unnecessary defibrillator pulses gt The tester may a
10. lso simulate a fibrillation waveform to ensure that a defibrillator discharge can be activated by the attendant switch as necessary Any error in the waveform or energy level would indicate problems in the defibrillator itself gt The wearing elements that would yield low energy output is the battery pack The battery voltage should be tested under loaded conditions The battery should be periodically fully discharged After recharging the battery will be capable of delivering its rated energy Otherwise polarization on the battery terminals could limit its performance A logical first step to better define the symptom would be to interview the operator if possible An alternative to test the ESU with the electrosurgical analyzer This test equipment can be used to measure the output power in all modes of ESU operation It can be used also to measure the RF leakage to ground as well as low frequency current It is important for the troubleshooter to detect 60 Hz ac leakage currents because they can cause microshock Module level troubleshooting can be done by following block schematics or diagrams provided in the manual 4 Troubleshooting tip turn the power off before removing or inserting any circuit board Otherwise transients induced may damage the board n YY There are two outputs to the ESU electrodes The bipolar outputs are on the lower right side of the schematic T
11. ocedures M Since they are the contact point between the patient and the instrument transducers are often vulnerable to damage Troubleshooting techniques apt to be effective with transducers are visual inspection interview of the operator and voltage or resistance measurement Since transducers are often moved about and have delicate parts they are subject to wear and abuse Surface electrode gt Check its attachment to the skin to make sure the electrode gel is adequate and has not dried out gt Be sure from the adhesive connection gt No excessive hair under the electrode or a scarred or bony surface gt Visual inspection of cable connections for frays breaks or corrosion may reveal a problem hermistors Poor thermal conductivity between the body tissue monitored and the thermistor Be sure that the mechanical connection free of air gaps In case that the thermistor is a branch of a Wheatstone bridge a voltage check of the bridge output and resistance measurements on its components could reveal thermistor related faults Nonlinearity of the thermistor for example could be caused by a failure in a linearizing resistor attached across the thermistor The sensitivity of the thermistor can be effected by v Faults in the bridge branch components v Drop in the excitation voltage of the bridge To isolate the problem it can be used circuit analysis voltage or resistance checks The final proof
12. one good test of a cable is to invert it in its connector if its possible If the symptoms of the problem change the cable is probably bad 4 Step 14 15 troubleshooting at the circuit board level is facilitated if the equipment block diagram provided in the service manual is used as a guide Data for problem analysis may be gathered from designated test points 4 Step 18 broad perspective on the kinds of problems that the equipment may be susceptible to 4 Step 22 data can be used to isolate the problem either to a particular board or to an ancillary component that is often too large or that gets too hot to be attached to a board of heat sink requirements iy 4 When a circuit board has been found to be faulty component level troubleshooting should be done This involves Detailed signal tracing Voltage and resistance measurements 4 Use of the equipment schematic showing interconnection between the individual components If a particular circuit board should be replaced by a another one the following consideration should be taken A visual inspection should be performed to look for any evidence of short circuits or overheating Checking the power supply over voltage which could be damage the new board Using of antistatic spray to prevent damage due to static charge buildup Because the circuit boards are expensive all precautions should be taken not to damage them during troubleshooting pr
13. ted accurately PM can protect the patient by reducing the likelihood of mistakes when the equipment gives inaccurate data PM procedures are recommended by the manufacturers of equipment and are usually given in the equipment service manual The frequency of PM depends on how vital the instrumentation is and on the observed failures blood circulation or breathing devices more than other Wearing parts of equipment must be changed periodically A ray tubes chemical electrodes air filters Inoperative equipment should be troubleshooted and repaired 4 Systematic approach to locating the cause of a fault in an electronic circuit or system gt If your keyboard will not type check to ensure that the cable is securely fastened to the keyboard port Determining which part of a system is responsible for a problem sensing electrode reference electrode instrument solution measuring technique and operator Unexpected solution chemistry incorrectly prepared standardizing solutions improper plotting of data unsuitable reference electrodes operator error and poor choice of method account for many more problems than do instrument or electrode failure A logical way of testing hardware or software in order to determine how to fix a problem Troubleshooting is done by one of the following methods Case study approach is used if a piece of equipment were known to have a chronic or repetitive pro
14. with a lead continuity tester Possible Artifacts Check the following Base line Wander Patient moving excessively Secure lead wires and cable to patient Caused by patient s respiration Reposition electrodes Electrodes dry Re prep skin and apply fresh moist electrodes Static buildup around patient Check with Engineering i ue ho a c Noise Gain set too high Readjust as required Unit in diagnostic mode Select monitoring mode Electrodes dry Re prep skin and apply fresh moist electrodes Patient cable entwined with cables of other electrical devices Separate patient cable from all other cables Possible Artifacts Check the following Connections not tight and properly Intermittent Signal secured electrode to lead lead to cable cable to monitor Ensure proper connection Electrodes dry Re prep skin and apply fresh moist electrodes Cable or lead wires damaged Check with continuity tester Low battery in telemetry transmitter Replace with fresh battery Gain set too low Readjust as required Low mn ECG Gain set too low Readjust ired Signal Skin properly prepared Abrade skin els the patients normal complex Check with 12 lead electrocardiogram 4 Troubleshooting an EEG is similar in many ways to troubleshooting of the ECG They both have common mode interference problems They both use multiple surface electrodes and leads As with ECG also

Download Pdf Manuals

image

Related Search

troubleshooting troubleshooting troubleshooting meaning troubleshooting hardware and devices troubleshooting printer troubleshooting settings troubleshooting windows 11 troubleshooting synonym troubleshooting ac unit troubleshooting guide troubleshooting keyboard troubleshooting control panel troubleshooting steps troubleshooting hp printer problems troubleshooting maytag washer troubleshooting ge dishwasher troubleshooting camera troubleshooting bluetooth troubleshooting printer not printing troubleshooting methodology troubleshooting mouse troubleshooting microphone troubleshooting 1202 events troubleshooting tool troubleshooting att troubleshooting windows 10

Related Contents

Allied Telesis Multi Rate SFP, 15km  user manual in pdf    Verhandlungsleiter Ulrich Marten: ........................................................  LG GSA-4120B User's Manual  User manual  WELL T18 User Manual-V0.2  GB F NL N S SF DK I E P 5500 & 7700 D  ココセコム・Gマネージャーご利用ガイド(操作編) 2683KB  Contents  