Home
HEARTSTART MRx
Contents
1. HEARTSTART MRx MONITOR DEFIBRILLATOR PRODUCT DESCRIPTION No compromises That s how we created the HeartStart MRx a combination transport monitor cardiograph manual defibrillator and AED HeartStart MRx unites our industry leading patient monitoring capabilities with our patented biphasic defibrillation therapy in one comprehensive easy to use rugged and lightweight device optimized to meet a variety of user needs With carefully arranged controls connections and menus HeartStart MRx efficiently provides comprehensive diagnostic measurements without compromising critical lifesaving therapy Monitoring and resuscitation functions are clearly delineated with monitoring located on one side of the device and defibrillation pacing and cardioversion on the other Patient status and care data including numeric vital signs and up to four waves as well as text prompts alarms an event timer event markers and battery status indicators are logically organized on the device s large backlit color display PHILIPS HEARTSTART MRx MONITOR DEFIBRILLATOR It is our mission to produce the finest quality medical equipment enabling you to deliver effective care and ultimately save lives We strive to meet the evolving needs of the healthcare community with outstanding technology and innovation while always focusing on what s most important positive patient outcomes So whether it s the long life
2. Long Soft Single Patient Disposable Cuff Thigh Intubated Circuits Filter Line Set Adult Pediatric 25 sets case Filter H Set Humidified Adult Pediatric 25 sets case Filter H Set Humidified Infant Neonatal yellow 25 sets case Non Intubated Dual Purpose Circuits CO O3 Smart CapnoLine Pediatric Smart CapnoLine Intermediate Smart CapnoLine Adult EtCO Non Intubated Single Purpose Circuits CO3 M2524A Smart CapnoLine Pediatric M2525A Smart CapnoLine Intermediate M2526A Smart CapnoLine Adult Power M3538A Lithium lon Battery with fuel gauge M3539A AC Power Module 989803135301 2 Bay Battery Support System for Lithium lon Batteries 989803135331 4 Bay Battery Support System for Lithium lon Batteries 989803135341 4 Bay Battery Support System for Sealed Lead Acid and Lithium lon Batteries Paper 40457C 50 mm Chemical Thermal Gray Grid 10 rolls 40457D 50 mm Chemical Thermal Gray Grid 80 rolls 989803138171 75 mm Chemical Thermal Gray Grid 10 rolls 989803138181 75 mm Chemical Thermal Gray Grid 80 rolls Color Handle M5521A Green and color label set M5522A Blue and color label set M5523A Yellow and color label set M5524A Rose and color label set M5525A Gray and color label set Miscellaneous M1781A Test Load for use with M3507A Pad Cable M3725A Test Load for use with M3508A Pad Cable M3541A Carrying Case includes 3 accessory pouches and shoulder strap M3544A Data
3. Adult Pediatric High Limit 20 to 95 mmHg Adult Pediatric Calibration Schedule yearly or every 4 000 hours Sample Size 50 ml per min PHILIPS HEARTSTART MRx MONITOR DEFIBRILLATOR Airway Respiration Rate Range 0 to 150 rpm Resolution 1 rpm Accuracy O to 40 rpm 1 rpm 41 to 70 rpm 2 rpm 71 to 100rpm 3 101 to 150rpm 5 Alarm Range Low Limit 0 to 99 rpm Adult Pediatric High Limit 10 to 100 rpm Adult Pediatric Apnea Alarm Time 10 40 seconds in increments of 5 Calibration Gas for COs Measurement System Ingredients 5 Carbon Dioxide 21 Oxygen 74 Nitrogen Cylinder Size BD Method of Preparation Gravimetric Blend Tolerance 0 03 Accuracy 0 03 absolute Moisture 10 PPM Maximum Expiration Period 2 years Pressure 144 PSIG Volume 10L 12 Lead ECG Inputs With a 10 Lead cable leads Il III aVR aVL aVF V C 1 V C6 can be obtained All 12 Lead ECG waves can be viewed on the display simultaneously All 12 leads can be printed on the strip chart printer in 3x4 format ECG Bandwidth Filters 0 15 40 Hz 0 05 40 Hz 0 05 150 Hz 12 Lead ECG Transmission Data card to a PC Receiver Type HeartStart Event Review Pro TraceMaster ECG Management System Version A 04 02 and TraceMaster Vue Patient Data Storage Internal Event Summary Up to 8 hours of continuous ECG waveforms and events and up to 62 megabytes of patient data Data Card Event Summary Up to 8 hours of continuou
4. Card Tray M3545A Data Card and Tray M3537A Bedrail Hook mount PHILIPS HEARTSTART MRx MONITOR DEFIBRILLATOR Specifications General Dimensions with pads 12 4 in W x 8 3 in D x 11 7 in H 31 5 cm x 21 0 cm x 29 5 cm Dimensions with paddles 13 4 in W x 8 3 in D x 13 6 in H 34 0 cm x 21 0 cm x 34 5 cm Weight 13 2 Ibs 5 99 kg including pads pads cable full roll of paper and battery Incremental weight of external standard paddles and paddle tray is 2 5 Ibs 1 1 kg Additional battery weighs 1 6 Ibs 0 73 kg Defibrillator Waveform Biphasic Truncated Exponential Waveform parameters adjusted as a function of patient impedance Shock Delivery Via multifunction defib electrode pads or paddles Delivered Energy Accuracy Selected Nominal Delivered Energy vs Patient Impedance Accuracy Energy Load Impedance 25 50 75 100 125 150 175 1J 1 2 1 3 1 2 1 1 1 0 0 9 0 8 2J 2J 1 8 2 0 2 0 1 9 1 7 1 6 1 5 2J 3J 2 8 3 0 3 0 3 1 3 0 2 9 2 7 2J 4J 3 7 4 0 40 4 1 4 2 4 2 40 2J 5J 4 6 5 0 5 1 5 1 5 2 5 2 5 0 2J 6J 5 5 6 0 6 1 6 2 6 3 6 3 6 1 2J 7J 6 4 7 0 7 1 7 2 7 3 7 3 71 2J 8J 74 8 0 8 1 8 2 8 4 8 3 8 1 2J 9J 8 3 9 0 9 1 9 3 9 4 9 4 9 1 2J 10J 9 2 10 10 10 10 10 10 2J 15J 14 15 15 15 16 16 15 15 20J 18 20 20 21 21 21 20 15 30J 28 30
5. Standard Pad Optional Features Cable SpOpo with Fourier Artifact Suppression Technology FAST C15 5 Lead ECG Cable gt Noninvasive Blood Pressure LPT Insiruchons tor Use LP2 User Training Video English only Microstream EtCOo LP3 User Training DVD English only Noninvasive Pacing SM1 Service Manual English only gt 12 Lead ECG with Philips interpretation algorithm SM3 Service Training Video English only gt 75 mm Printer wo1 One Year On Site Warranty W24 Five Year Bench Warranty with Loaner U S amp Canada Standard Accessories only gt AC power module WA1 Five Year Biomed Warranty U S Canada amp Australia only Y Lithiumion batteryithituel gauge wos Three Year Bench Warranty with Loaner Australia amp gt Hands free multifunction electrode cable New Zealand only 3 Lead ECG cable Disposable monitoring electrodes Upgrades Supplies Accessories Upgrades M3530A SpOo M3531A NBP M3532A EtCOo M3533A Pacing M3534A 12 Lead ECG Option BO2 Acquisition Option BOS Transmission Option BO4 Wider Printer M4760A Handle and Cap Plate for Pads M5527A External Paddles with Paddle Tray Option CO1 Standard Paddles Option C02 Water Resistant Paddles External Paddles M3542A Standard External Paddles M3543A Water Resistant External Paddles M4745A Sterilizable External Paddles Internal Paddles M1741A 7 5 cm Switchless M1742A 6 0 cm Switchless M1743A 4 5 cm Switchless M1744A 2 8 cm S
6. batteries automated operational checks or a data management solution HeartStart MRx has the capabilities you need and the performance you demand no compromises M3535A PHILIPS HEARTSTART MRx MONITOR DEFIBRILLATOR Features gt Strip chart recorder paper gt Defibrillator test load Standard Features Documentation CD containing Instructions for Use User s guide User gt ST AR Basic algorithm for arrhythmia detection training workbook and Application notes ECG monitoring through monitoring electrodes and defibrillation pads gt Quick reference cards Synchronized cardioversion Adjustable ECG size and autogain Package M3535A HeartStart MRx Monitor Defibrillator gt Manual and AED operation SMART Biphasic waveform for defibrillation therapy Ordering Option Information A01 SpO gt Large 4 wave color display A02 SpO and NBP Strip chart printer A03 SpO9 NBP and EtCO9 gt Individual adjustable volume of QRS beeper voice prompts and alerts A04 EtCO5 gt Event summary B01 External Pacing k Coritiguration mode B02 12 Lead ECG Acquisition Service mode B04 Wider Printer gt Operational checks c01 Standard External Paddles gt Automated self tests with ready for use indicator 02 Waler Resistant External Paddles c03 Data Card Lithium ion battery with fuel gauge C04 Carrying Case ECG out capability co5 Additional Battery C07 Barrel style Pad Cable Replacement for
7. for CO9 Wave Viewing Time 5 seconds ECG Battery Type 6 0 Ah 14 8 V rechargeable Lithium lon Dimensions 6 5 in H x 3 8 in W x 1 6 in D 165 mm x 95 mm x 42 mm Weight 1 6 Ib 0 73 kg Charge Time Approximately 3 hours to 100 Approximately 2 hours to 80 indicated by battery gauge indicator Charging the battery at temperatures above 45 C may degrade battery life Capacity At least 5 hours of monitoring with ECG SpO and COs monitored continuously and NBP measured every 15 minutes or at least 50 200 joule shocks on a new fully charged battery at room temperature 25 C At least 3 5 hours while pacing at 180 ppm at 160 mA and monitoring as described above Battery Indicators Battery gauge on battery capacity indicator on display flashing RFU indicator chirp and LOW BATTERY message appears on display for low battery condition Low battery condition triggered with at least 10 minutes of monitoring time and 6 maximum energy discharges remain with a new battery at room temperature 25 C Storage Storing the battery for extended periods at temperatures above 40 C will reduce battery capacity and degrade battery life PHILIPS HEARTSTART MRx MONITOR DEFIBRILLATOR Thermal Array Printer Continuous ECG Strip The Print key starts and stops the strip The printer can be configured to run real time or with a 10 second delay The strip prints the primary ECG lead with event annotat
8. 30 31 31 31 30 15 50J 46 70J 64 70 71 72 73 73 71 15 50 51 51 52 52 50 15 100J 92 100 101 103 104 104 101 15 120J 110 120 121 123 125 125 121 15 150J 138 150 152 154 157 156 151 15 170J 156 170 172 175 177 177 172 15 200J 184 200 202 206 209 209 202 15 Charge Time Less than 5 seconds to 200 Joules with a new fully charged Lithium lon battery pack at 25 C Less than 15 seconds when operating without a battery using the M3539A AC power module alone at 90 rated mains voltage Patient Impedance Range Minimum 15 Ohm internal defibrillation 25 Ohm external defibrillation Maximum 180 Ohm Note Actual functional range may exceed the above values Manual Mode Manual Output Energy Selected 1 10 15 20 30 50 70 100 120 150 170 200J maximum energy limited to 50J for internal defibrillation Controls On Off Therapy Knob Charge Shock Sync Print Mark Event ECG Lead Select Alarm Pause Event Review Energy Selection Control Front panel therapy knob Charge Control Front panel button button on external paddles Shock Control Front panel button buttons on external or switched internal paddles Synchronized Control Front panel SYNC button Indicators Text prompts audio alerts QRS beeper battery status Ready For Use external power battery charging Sync mode A
9. 8A M1949A 10 lead ECG Patient Trunk Cable 12 pin ECG Input Connector for 5 Lead and 12 Lead use M1968A 10 electrode Cable Set Extremities Grabber use with M1976A AAMI M1976A 10 electrode Cable Set Chest Grabber use with M1968A AAMI M1971A 10 electrode Cable Set Extremities Grabber use with M1978A IEC M1978A 10 electrode Cable Set Chest Grabber use with M1971A IEC ECG Cables M1500A M1605A M1510A M1615A M1520A M1625A M1530A M1635A Sync Cables M1783A M5526A 3 Lead Cable Set 3 lead ECG Trunk Cable AAMI 3 lead ECG Lead Set with Snaps AAMI 3 lead ECG Trunk Cable IEC 3 lead ECG Lead Set with Snaps IEC 5 Lead Cable Set 5 lead ECG Trunk Cable AAMI 5 lead ECG Lead Set with Snaps AAMI 5 lead ECG Trunk Cable IEC 5 lead ECG Lead Set with Snaps IEC Sync Cable 8 ft 2 4 m Sync Cable 25 ft 7 6 m SpOs Sensors Cables M1191A M1192A M1193A M1194A M1195A M1941A M1943A NBP M1598B M1599B 40400A 40400B 40401A 40401B Reusable SpO Sensor Adult Finger 6 5 ft 2 m Reusable SpOg Sensor Pediatric Small Adult Finger Reusable SpOg Sensor Adult Thumb Reusable SpOg Sensor Adult Ear Clip Reusable SpOg Sensor Infant SpO Extension Cable 2 m 6 5 ft Nellcor SpOo Sensor Adapter Cable 1 m use with M1903 4B Interconnect Cables Adult Pressure 5 ft 1 5 m Adult Pressure 10 ft 3 m Reusable Blood Pressure Cuffs Reus
10. Pad Fault Dashed line appears on the display if a pad becomes disconnected Heart Rate Display Digital readout on display from 15 to 300 bpm with an accuracy of 10 Heart Rate Arrhythmia Alarms HR Asystole VFIB VTACH VTACH Extreme Tachy Extreme Brady PVC rate Pacer not capture Pacer not pacing Hands Free Defibrillation Cable Length M3507A 7 ft 2 2 m M3508A 7 ft 2 2 m ECG Cable Length 9 ft 2 7 m Common Mode Rejection Greater than 90 dB measured per AAMI standard for cardiac monitors EC 13 ECG Size 2 5 5 10 20 40 mm mV autogain Frequency Response AC Line Filter 60 Hz or 50 Hz 3 lead 5 lead and Pads Pads ECG for Display Monitor 0 15 40 Hz or EMS 1 30 Hz Pads ECG for Printer Monitor 0 15 40 Hz or EMS 1 30 Hz Leads ECG for Display Monitor 0 15 40 Hz or EMS 1 30 Hz Leads ECG for Printer Diagnostic 0 05 150 Hz or Monitor 0 15 40 Hz or EMS 1 30 Hz 12 lead ECG for Display 0 05 150 Hz 0 05 40 Hz 0 15 40 Hz ECG for Report 0 05 150 Hz 0 05 40 Hz 0 15 40 Hz 0 05 150 Hz Patient Isolation defibrillation proof ECG Type CF SpOo Type CF EtCOo Type CF NBP Type CF External Defib Type BF Internal Defib Type CF Display Size 8 4 in diagonal 128 mm x 171 mm Type TFT Color LCD Resolution 640 x 480 pixels VGA Sweep Speed 25 mm s nominal stationary trace sweeping erase bar for ECG and SpOo 6 25 mm s
11. able NBP Cuff Kit 3 sizes pediatric adult large adult Reusable NBP Cuff Kit 5 sizes infant pediatric adult large adult thigh Traditional Reusable NBP Cuff Infant Traditional Reusable NBP Cuff Pediatric NBP 40401C 40401D 40401E M4552A M4553A M4554A M4555A M4557A M4559A M1572A M1573A M1574A M1575A M4572A M4573A M4574A M4575A M4576A M4577A M4578A M4579A EtCO M1920A M1921A M1923A M2520A M2521A M2522A PHILIPS HEARTSTART MRx MONITOR DEFIBRILLATOR Traditional Reusable NBP Cuff Adult Traditional Reusable NBP Cuff Large Adult Traditional Reusable NBP Cuff Thigh Antimicrobial Reusable NBP Cuff Infant Antimicrobial Reusable NBP Cuff Pediatric Antimicrobial Reusable NBP Cuff Small Adult Antimicrobial Reusable NBP Cuff Adult Antimicrobial Reusable NBP Cuff Large Adult Antimicrobial Reusable NBP Cuff Thigh Multi Patient Comfort Cuffs Pediatric Multi Patient Comfort Cuffs Small Adult Multi Patient Comfort Cuffs Adult Multi Patient Comfort Cuffs Large Adult Disposable Blood Pressure Cuffs Soft Single Patient Disposable Cuff Infant Soft Single Patient Disposable Cuff Pediatric Soft Single Patient Disposable Cuff Small Adult Soft Single Patient Disposable Cuff Adult Soft Single Patient Disposable Cuff Adult X Long Soft Single Patient Disposable Cuff Large Adult Soft Single Patient Disposable Cuff Large Adult X
12. ions and measurements Auto Printing The printer can be configured to automatically print on Marked Events Charge Shock and Alarm When an alarm condition occurs the unit prints the primary ECG wave and the alarming wave if configured Reports The following can be printed Event Summary 12 Lead Operational Check Configuration Status Log and Device Information Speed 25 or 50 mm s with an accuracy of 5 Amplitude Accuracy 5 or 40 uV whichever is greater Paper Size 50 mm W by 30 m 100 ft L 75 mm W by 30 m 100 ft L Noninvasive Pacing Waveform Monophasic Truncated Exponential Current Pulse Amplitude 10 mA to 175 mA 5 mA increments accuracy 10 or 5 mA whichever is greater Pulse Width 40 ms with 10 accuracy Rate 30 ppm to 180 ppm 10 ppm increments accuracy 1 5 Modes Demand or Fixed Rate Refractory Period 340 msec 30 to 80 ppm 240 msec 90 to 180 ppm SpO Pulse Oximetry Range SpO2 0 100 pulse rate 30 to 300 bpm Resolution 1 Display Update Period 1 sec typical numeric update rate Accuracy with M1191A sensor 1 standard deviation 70 to 100 2 5 M1192A sensor 1 standard deviation 70 to 100 2 5 M1193A sensor 1 standard deviation 70 to 100 2 5 M1194A sensor 1 standard deviation 70 to 100 4 0 M1195A sensor 1 standard deviation 70 to 100 4 0 NELLCOR sensors 1 standard deviation 80 to 100 3 0 Pulse Rate Accuracy 2 or 1 b
13. m lon PHILIPS HEARTSTART MRx MONITOR DEFIBRILLATOR Philips Medical Systems is part of Royal Philips Electronics PHILIPS HEARTSTART MRx MONITOR DEFIBRILLATOR UNITED STATES Philips Medical Systems Cardiac and Monitoring Systems 3000 Minuteman Road Andover MA 01810 Tel 800 934 7372 CANADA Philips Medical Systems 2660 Matheson Blvd E Mississauga Ontario L4W 5M2 Tel 800 291 6743 EUROPE MIDDLE EAST AND AFRICA Philips Medizinsysteme B blingen GmbH Cardiac and Monitoring Systems Hewlett Packard Str 2 71034 B blingen Germany Fax 49 7031 463 1552 LATIN AMERICA HEADQUARTERS Philips Medical Systems 5200 Blue Lagoon Drive 9th Floor Miami FL 33126 Tel 805 267 4220 ASIA PACIFIC HEADQUARTERS Philips Medical Systems 26 F Hopewell Centre 17 Kennedy Road Wanchai Hong Kong Tel 852 2821 5888 Fax 852 2527 7136 www medical philips com 2004 Koninklijke Philips Electronics N V All rights are reserved Re production in whole or in part is prohibited without the prior written consent of the copyright holder Koninklijke Philips Electronics N V reserves the right to make changes in specifications or to discontinue any product at any time without notice or obligation and will not be li able for any consequences resulting from the use of this publication Published December 2004 4522 981 98961
14. pm whichever is greater Alarm Range Low Limit 50 to 99 Adult Pediatric High Limit 51 to 100 Adult Pediatric Alarm Delay 10 seconds Noninvasive Blood Pressure Pressure Range Systolic 40 260 mmHg Diastolic 20 200 mmHg Initial Pressure 160 mmHg Adult 120 mmHg Pediatric Maximum Pressure 280 mmHg Overpressure Safety Limits Maximum of 300 mmHg Cuff Inflation Time 50 second maximum pediatric or adult Accuracy 3 mmHg Alarm Range Systolic high limit 35 270 Adult 35 180 Pediatric Systolic low limit 30 265 Adult 30 175 Pediatric Diastolic high limit 15 245 Adult 15 150 Pediatric Diastolic low limit 10 240 Adult 10 145 Pediatric Mean high limit 25 255 Adult 25 160 Pediatric Mean low limit 20 250 Adult 20 155 Pediatric Calibration Schedule yearly or every 10 000 cycles Auto Mode Repetition Time 1 2 5 5 10 15 30 60 or 120 minutes Measurement Time Auto manual mode 30 seconds average HR gt 60 bpm 170 seconds maximum Interconnect Tube Length M1598B Connect tubing 5 ft 1 5 m M1599B Connect tubing 10 ft 3 m End Tidal Carbon Dioxide Range 0 to 99 mmHg Resolution 1 mmHg 0 1kPa Accuracy For values between O and 38 mmHg 2 mmHg For values between 39 and 99 mmHg 5 of reading 0 08 for every 1 mmHg above 40 mmHg Values read at sea level after gt 20 min warm up Alarm Range Low Limit 10 to 94 mmHg
15. rmed Indicators Charging tone charged tone flashing shock button and energy level indicated on display AED Mode AED Energy Profile 150 Joules nominal into a 50 Ohm test load Text and Voice Prompts Extensive text audible messages guide user through configured protocol AED Controls On Off Shock Indicators Monitor display messages and prompts voice prompts battery status Ready For Use external power battery charging Armed Indicators Flashing shock button charged tone and voice prompts ECG Analysis Evaluates patient ECG and signal quality to determine if a shock is appropriate and evaluates connection impedance for proper defibrillation pad contact Shockable Rhythms Ventricular fibrillation and certain ventricular tachycardias including ventricular flutter and polymorphic ventricular tachycardia Shock Advisory Algorithm Sensitivity and Specificity Meets AAMI DF 39 PHILIPS HEARTSTART MRx MONITOR DEFIBRILLATOR ECG and Arrhythmia Monitoring Inputs Up to four 4 ECG waves may be viewed on display and up to 2 waves printed simultaneously Lead Il or Ill is obtained through the 3 lead ECG cable With a 5 lead cable leads aVR aVL aVF and V can also be obtained Pads ECG is obtained through 2 multifunction defibrillation electrode pads Lead Fault Leads Off message and dashed line appear on the display if an electrode or lead wire becomes disconnected Lead Off indicator in wave sector
16. s ECG waveforms and events and up to 62 megabytes of patient data Environmental Temperature 0 C to 45 C operating 20 to 70 C storage Humidity Up to 95 relative humidity Altitude Operating O to 15 000 ft O to 4 500 m Storage O to 15 000 ft O to 4 500 m Shock Operating Half sine waveform duration lt 3 ms acceleration gt 145 g 1 time on all six faces Storage Trapezoidal waveform acceleration gt 30 g velocity change 742 cm s 10 on all six faces Vibration Operating Random vibration 0 30 grms 5 500 Hz for gt 10 minutes axis PSD 0 0002 g Hz from 5 to 350 Hz 6 dB octave slope from 350 to 500 Hz Storage Random vibration 2 41 grms 5 500 Hz for gt 10 minutes axis PSD 0 02 g Hz from 5 to 100 Hz 6 dB octave slope from 100 to 137 Hz O dB octave slope from 137 to 350 Hz 6 dB octave slope from 350 to 500 Hz Solids Resistance P2x Water Resistance Meets EN 60601 2 4 Water testing performed with cables connected to the device EMC Complies with the requirements of standard EN 60601 1 2 2001 Safety Complies with the requirements of applicable safety standards Other Considerations Device not suitable for use in the presence of concentrated oxygen or a flammable anesthetic mixture with air oxygen or nitrous oxide Mode of Operation Continuous AC Power Module Input 100 240 VAC 50 60 Hz 1 0 46 A Class 1 Output 18V 5A 90W Battery 14 8 V Rechargeable Lithiu
17. witchless M4741A 75 cm Switched M4742A 6 0 cm Switched M4743A 4 5 cm Switched M4744A 2 8 cm Switched 989803127121 Large Disposable Switched 989803127131 Medium Disposable Switched 989803127141 Small Disposable Switched 989803127151 Large Disposable Switchless 989803127161 Medium Disposable Switchless 989803127171 Small Disposable Switchless M4740A Internal Paddles Adapter Cable PHILIPS HEARTSTART MRx MONITOR DEFIBRILLATOR Multifunction Electrode Pads White barrel connector M3501A Defib Adult AAMI M3502A Defib Adult IEC M3503A Defib Pediatric IEC M3504A Defib Pediatric AAMI Grey plug connector M3713A Adult Plus M3716A Adult Radiolucent M3717A Pediatric Plus M3718A Adult Radiotransparent Reduced Skin M3719A Pediatric Radiotransparent Reduced Skin DP2 Self adhesive adult electrode pads DPG Self adhesive adult electrode pads Pads Cable M3507A Defib Hands free Pads Cable barrel style M3508A Defib Hands free plug style 05 10200 Pads Adapter use with M3507A ECG Monitoring Electrodes M2202A High Tack Foam 5 electrodes pack 60 packs case ECG Cables 12 Lead Cable Set M3525A 10 lead ECG Trunk Cable 12 pin Connector for 3 Lead and 12 Lead use M3526A 3 lead ECG Set and Plug with Snap AAMI M3527A Add 7 lead ECG Set for 12 Lead use AAMI M3528A 3 lead ECG Set and Plug with Snap IEC M3529A Add 7 lead ECG Set for 12 Lead use IEC M5530A Combiner Plug for 3 wire Lead Set for use with M3526A M352
Download Pdf Manuals
Related Search
Related Contents
CTY-WS - Saunacore "取扱説明書" Canton Movie 1005 Leica DNA03/DNA10 User Manual DEPOT EXPRESS - mode d`emploi - E User Manual - Smartphone Bedienungsanleitung E-Deluxe Manuale Modalità Operative Copyright © All rights reserved.