1. Electronics Corp Rev D_07 02 12 Auto Therm 390 Shortwave Diathermy Rov D 07 02 12 2 Symbol Glossary and List of Abbreviations 24 Symbol Glossary e EE Om One Os Onone Onone ESO ER Omoa GE Cem of 8 o Start Time display symbol Output display symbol Treatment frequency selection indicators Treatment pulse width selection indicators Drum applicator selection indicator Soft rubber plate applicators selection indicator Go back one step in the setup procedure Go forward one step in the setup procedure Increase value for time watts and beeper Go up for selection of Frequency Pulse Width and applicators Decrease value for time watts and beeper Go down for selection of Frequency Pulse Width and applicators Press Enter to select a value or selection Press Start to begin a treatment Melior Electronics Corp Rev D 07 02 12 Hold Cal Volume 2001 c Press Hold to stop a treatment Parameters that have been chosen will be retained Press this selector to change the volume of the beeper Main Power On Main Power Oft Not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide Attention consult instruction manual Non ionizing radiation Type BF Equipment Ciass Year of manufacture EF WM 22 Abbreviations cw Hz LED us MHz Pour S
2. coil applicator 3980 3982 Four jointarm 3983 Lanyard for Patient Safety Switch 3984 Strap tor securing 14 cm coil applicator to the patient 7293 Detachable hospital grade line cord 96 Cart for use with arm and coil applicator EZ Vote Electronics Corp Rev D_07 02 12
3. may cause burning In addition remove jewelry and watches during treatment 5 Remove hearing aids and watches during treatment to prevent interference with or damage to these devices amp Do not treat over the pelvic or low back area when an IUD is present 7 Shortwave diathermy should not be applied over the pregnant or potentially pregnant uterus Therefore shortwave diathermy should not be applied over the uterus unless specific assurance can be attained from the patient that she is not pregnant 8 Shortwave diathermy should not be applied to the eye 9 Neoplastic tissues or space occupying lesions should not be exposed to shortwave diathermy 10 Shortwave diathermy should not be applied to the testes to avoid increases in temperature 11 Do not treat ischemic tissues in individuals with vascular disease where the blood supply would be unable to follow the increase in metabolic demand and tissue necrosis might result 12 Shortwave diathermy should not be applied over the epiphyseal areas bone growth centers of the bones of growing children E Melior Electronics Corp Rev D 07 02 12 53 Precautions 1 Shortwave diathermy should not be applied in areas of reduced sensation or circulation Patients having reduced sensation will nat be able to notify the practitioner of discomfort i the intensities are too high Patients with compromised circulation may have an excessive heat buildup in the treatment area 2 A
4. part of arm 2426 27 __im m BL Xo Fixing bolt opposite side 21 23 e F Covering 25 Siotted nut Arm Figure 3 4 Auto Therm 390 optional applicator arm assembly 34 Flexible condenser applicators These applicators are available as one pair The permanently attached highly flexible cables ensure maximum flexibility at the point of applicator connection and high adaptability to bent or curved parts of the patient s body The optimum distance between the applicator and the skin is achieved by using thick felt spacers that have to be placed in a cloth bag together with the applicator One to three of the 5 mm thick felt spacers are used for tuning between electrode and skin The number of spacers depends on the application and characteristics of the tissue Velcro straps are used for fastening these applicators 3 5 Inductive drum applicator An optional inductive drum applicator is available This drum applicator uses the high frequency magnetic field generated by the inductive coil of an electrical resonant circuit for therapy The AAuto Therm 390 inductive drum applicator is electrically shielded and connected to the unit via a removable shielded coaxial cable Use only the cable ME 1881 designed for this application Failure to do so may result in damage to the unit or abnormal output 3 Vote Electronics Corp Rev D_07 02 12 Auto Therm 390 Shortwave Diathermy
5. AutoxTherm 390 Instruction Manual Meriter SLectRonics corp 1338 South Claudina Street Anaheim CA 92805 U S Toll free 800 854 9305 Telephone 714 533 2221 FAX 714 635 7539 Web Site htip www mettlerolectronics com Email mail mettlerelectronics com 1R9 36 Rev D 07 02 12 Copyright 2005 by Mettler Electronics Corp Anaheim CA Matter Electronics Corp Rev D 07 02 12 FCC Frequency Interference Statement Warning This equipment generates and uses radio frequency energy and if not installed and operated in strict accordance with the manufacturer s instructions may cause radio frequency interference Notice 1 This equipment has been verifed to comply with the specifications in Part 18 of FCC Rules which are designed to provide reasonable protection against radio frequency interference However there is no guarantee that interference will not occur in a particular installation Notice 2 If this equipment is found to be the source of radio frequency interference which can be determined by turning the equipment off and on the user should try to correct the interference by one or more of the following measures Reorient the receiving antenna as applicable Relocate the Auto Therm with respect to the receiver Move the Auto Therm away from the receiver Plug the Aulo Thorm into a different outlet than the receiver It necessary the user should consult with the dealer
6. Read the contents of this manual before treating patients with the Auto Therm 390 This manual has been written to assist you with the safe operation of the Auto Therm 390 It is intended for use by the owners and operators of the Auto Therm 390 The goal of this manual is to direct the correct operation and maintenance of this unit The specifications and instructions presented in this manual are in effect at the time of its publication These instructions may be updated at any time at the discretion of the manufacturer To ensure that you have the most up to date information about your product please go online ta register it at hitp jwaw metlerelectronics comproductregistration 13 Safety Precautions The Auto Therm 390 operates with high voltages Qualified biomedical technicians with raining in service of shortwave diathermy equipment should perform servicing of the Auto Therm 390 or it Should be returned directly to the factory To maximize safety during use the unit should be plugged into a grounded wall outlet General safety guidelines for medical electronic equipment should be followed Service may be obtained from the manufacturer by sending the Auto Therm 390 in its original shipping container to Mettler Electronics Corp 1333 South Claudina Street Anaheim CA 92805 ATTN Service Department Telephone toll ree 800 854 9305 Alternate telephone number 1 714 533 2221 NOTE All warranty repairs must be performed by Mettler Elec
7. Rev 0_ 07 02 12 4 Operating Instructions 4 1 Shortwave diathermy applications The Auto Thorm 390 is a shortwave diathermy device that operates at 27 12 MHz It provides traditional shortwave diathermy therapy using condenser and electromagnetic inductive coil fields in both continuous and pulsed modes of operation Therefore itis suited for al diathermy treatments in both the clinic and the medical practice The use of shortwave diathermy for heat therapy has the advantage of penetrating deeper than other conventional methods such as hot packs infrared lights and heating pads and even microwave diathermy The heat generated by shortwave diathermy induces a whole range of physiological effects It relaxes muscles tendons and other connective tissues and increases blood circulation in the treatment area When shortwave diathermy is applied in short high energy pulses the depth of penetration is increased having a particularly positive effect on the blood circulation while the temperature sensitive skin hardly feels the heat Shortwave diathermy can be applied over a large area of the skin s surface Avoid allowing the flexible cables or soft rubber applicators to touch each other during treatment 42 Preparing the patient For optimum therapy ensure that the patient is in a comfortable and relaxed position To accomplish this either have the patient in a seated or supine position Patients shall never be treated on
8. e displayed in the output window and there will be an audible alarm The LED to the left of CW will be blinking Select either CW or one of the pulse frequencies by pressing the up and down arrows g Auto Therm 390 Shortwave Diathermy Rov D 07 02 12 CE Cie GER italy M Once the LED of the desired frequency is lit press the next arrow to go to the Pulse Width selection Press the up and down arrows to select the Pulse Width that you would like Please note if CH is selected then the unit will automatically skip this selection Once the LED of the desired pulse width is lit press the next arrow to go to the Applicator selection Press the up and down arrows to select the applicator that is plugged into the Auto Therm 390 for use during this treatment Once the LED of the installed applicator is lit press the next arrow to set the Timer Press the up and down arrows to select the treatment time that you would like Please note time can be set in whole minutes only Press the Volume button to increase or decrease the loudness of the beeper Press the up and down arrows to adjust the volume to your surroundings This is an optional step during the operation of this device Make sure that the applicator s is are plugged in and coupled to the patient Press start to begin treatment The Auto Therm 390 wil fist tune to the patient During this time litle dashes will move around the perimeter of
9. e to allow the inician to cary this shortwave diathermy around the clinic Figure 1 1 The Auto Therm 390 An optional accessory package upgrades your Auto Therm 390 to do inductive coil diathermy using the same device This package includes a multi jointed arm a 14 cm diameter drum applicator special coaxial cable and a cart with large locking wheels that allows the Auto Therm 390 to be moved easily around the clinic and supports the arm and drum applicator during treatment The Auto Therm 390 has been certified by Intertek Testing Services to meet the requirements for ETL Listing per the following standards UL60601 1 UL Standard for Safety Medical Electrical Equipment Part 1 General Requirements for Safety Second Edition CSA C222 NO 601 1 M90 Medical Electrical Equipment Part 1 General Requirements for Safety General Instruction No 1 Supplement 1 1994 R 1997 IEC60601 2 3 Safety of Shortwave Therapy Equipment In addition the Auto Therm 390 meets the following standards for radio frequency emissions FOO Part 15 A IEC EN 60601 1 2 Mettler Electronics Corp has been certified by VTT Expert Services LTD to be compliant with EN ISO 13485 2003 and MDD 93 42 EEC Annex Il requirements In addition Mettler is certified by DOS Medizinprodukte GMBH to be compliant with ISO 13485 2003 CMDCAS Canadian Medical Device requirements Meter Electronics Corp Rev D 07 02 12 1 2 Introduction to This Manual
10. erie Continuous wave 100 duty cycle Hertz pulses per second Light Emitting Diode Microsecond 1 x 10 second Megaherz 1 x 10 pulses per second Minutes Output power Seconds Serial Number Watts Volts EJ Auto Therm 390 Shortwave Diathermy Rev D_07 02 12 3 Installa n 3 1 Installation Instructions 1 Connect the line cord to the back of the Auto Therm 390 shortwave diathermy See Figure 3 1 2 Plug the line cord ME 7293 into a grounded wall outlet that is rated at 100 to 240 Volt AC 50 60 Hz Your power supply must match the voltage requirements listed on the serial number label of your device Do not connect the Auto Therm 390 to a power supply rated differently than that described above 3 Theline cord comes equipped with a standard 3 prong plug This plug provides grounding for the Auto Therm 380 Do not defeat its purpose by using 3 to 2 prong adapters or any other means of attaching to a wall outlet 4 Plug the soft rubber electrode cables into the appropriate receptacles as seen in Figure 3 1 or plug the optional coaxial drum applicator cable into its receptacle Do not plug both applicators in simultaneously as this may damage the Auto Therm 390 5 The Patient Safety Pul Cord is permanently attached to the back of the unit as seen in Figure EG 6 The Auto Therm 390 may emit radio interference Avoid operating other electrical devices adjacent to and simultaneously with the A
11. essed above or if additional troubleshooting guidance is desired call 800 854 8305 or 1 714 533 2221 outside the USA or email our service department at service mettiorelectronics com a Auto Therm 390 Shortwave Diathermy Rev D 07 02 12 7 Specific input ETL and C ETL Listed Fuse Weight Unit Cart Dimensions Unit Cart Frequency HF output Continuous mode Pulsed Mode Pulse frequency Pulse duration range Treatment ime Operating Temperature Humidity Storage Shipping Temperature Storage Humidity Storage Pressure Environmental Disposal Standard Accessories Optional Accessories ns 100 240 VAC 50 60Hz Model ME 390 9801427 For continuous protection replace fuse with the same type and rating 6 3A 250V GDC S506 5x 20 mm time delay 2xT63 AL250V 15 lbs 45 lbs Tin H x 14in W x 16 5 in D 30 25 in H x 19 in W x 19 in D 27 12 MHz 100 W Average Power 200 W Peak Power 10 Hz 20Hz 50Hz 100Hz 400Hz 65 us 100 js 200 us 300 ps and 400 us 1 30 minutes 450 F to 1042 F Operating 30 to 75 Relative Humidity at 104 F Non operating 5 to 95 Relative Humidity Non condensing AD F to 1672 F Storage 30 to 90 Relative Humidity at 104 F Non condensing 700 900 mB The device contains lead in the form of solder used to produce electrical contact between components To avoid adverse environmental impact utiize a disposal faci
12. g NOTE All warranty repairs are to be performed by Mettler Electronics Corp or an authorized Metier Electronics warranty repair center 1 6 Package Contents Your new Auto Therm 390 comes complete with all the necessary components to perform shortwave diathermy Below is a list of items that are included in the shipping carton There are also other optional accessories available for this unit They are listed at the back of this book 1 Auto Therm 390 2 Two each 12 x 18 cm soft rubber plate applicator complete with flexible cable 3963 3 Six each 14 x 20 cm felt spacer for soft rubber plate applicators 3965 Auto Therm 390 Shortwave Diathermy Rov D 07 02 12 Two each 18 x 26 cm cloth cover for softrubber plate applicators 3967 Two each 48 inch Velcro receptive elastic straps 2009 One detachable hospital grade line cord 7293 Instruction manual and warranty card 17 Limited Warranty The Auto Therm 390 shortwave diathermy unit is warranted against defects in materials and workmanship for a period of two years from date of purchase one year on applicators and accessories During the applicable warranty period Mettler Electronics Corp will at is discretion either repair or replace the Product without charge for these types of defects For service under this warranty the Product must be returned by the buyer within the applicable warranty period to Mettler Electronics Corp Shipping charges to Mettler Electronic
13. he patient can lay or sit directly on the pads Always try to maximize contact with the entire area of he electrode The better the connection the better the treatment Pads can be positioned in any manner on the patient They do not have ta be opposite each other or even on the same plane of the body Greater distance between the pads will increase the treatment area De 45 Inductive method of application The shortwave magnetic field of an inductive drum coll applicator generates eddy currents that are transformed to heat in the tissues These currents increase with increasing electric conductivity of the corresponding tissue region issues with good blood circulation e g muscle tissue and inner organs 16 Auto Therm 390 Shortwave Diathermy Rev D_ 07 02 12 To reach these deeper tissues ho inductive coil applicators of the Auto Therm 390 are provided with an electrostatic shielding that prevents the electric feld of the inductive coil applicator from heating the upper skin fat issue Therefore heat sensation by the patient is basically delayed when using the inductive coil applicators is recommended to start the treatment with an intensity level below the desired one and increase it later rather than be guided by tho patient s heat sensation in the treated area For maximum deep effecs apply the inductive coil applicator directiy o the body through a single layer of toweling Note To avoid damaging the inductive dr
14. lity that performs complete incineration of the device at a temperature in excess of 1000 C The shipping materials are fabricated of cardboard and may be disposed of with other paper products 2 12 x 18 cm soft rubber plate applicators with flexible cables 6 fet spacers 2 cloth covers 2 48 inch elastic Velcro receptive straps 1 Instrument cart 1 four joint folding arm 1 14 cm diameter coil applicator 1 special feld coil cable Vote Electronics Corp Rev D_07 02 12 ES Auto Therm 390 Shortwave Diathermy Rov D 07 02 12 8 Accessories 8 1 Ordering Information Therapy products and accessories are available from Mettler Electronics authorized Distributors For information regarding either Mettler products or a distributor near you please call oll free 800 854 2305 or phone 714 533 2221 in areas outside the continental United States Ask for Customer Service Mettler Electronics is open from 7 AM until 5 PM Pacific Time for your convenience The email address for Customer Service is mail mettlerelectranics com 8 2 Auto Therm 390 Accessories Part Description 2009 4 each 48 inch Velcro receplive straps 3063 12 x 18 cm soft rubber plate applicator complete with flexible cable 3065 14 x 20 cm felt spacer for soft rubber plate applicators 3067 18 x 26 cm cloth cover for soft rubber plate applicators 3980 14 cm diameter coil applicator without cable 3981 Special coaxial cable for
15. m 3 Increasing range of motion of contracted joints using heat and stretch techniques 4 Increasing blood flow to tissues in the treatment area 2 Contraindications Diathermy must not be applied over areas of the body which may contain metal implants surgical staples etc for heat will become concentrated in that area increasing the possibility of tissue damage and deep burns 2 Any patient with an implanted electronic device such as a cardiac pacemaker bladder stimulator spinal cord stimulator or electrodes for a my electr prosthesis or implanted metalic leads must not bo treated with shortwave diathermy and should not be in the vicinity ofthe Auto Therm 390 when itis in operation Do not use on patients who have had an implant inthe past unless you are absolutely certain that the implant and all leads in their entirely have been removed Note that leads are often left implanted after the implant is removed 3 Do not treat on a metal treatment table mattress with metal springs wheel chair or a metal stool Make sure that the patient cannot come into any contact with metal during treatment The patient should not be allowed to come into contact with conductive parts which are grounded or have an appreciable capacitance to ground and which may provide unwanted pathways for the radi frequency current 4 Remove clothing from the treatment area Any metal contained in the clothing such as zippers bra hooks or rivets
16. metal chairs tables or beds For safety reasons remove all hearing aids watches rings chains bracelets and other metal objects before starting the treatment The clinician should also take care to remove metalic objects while operating the Auto Therm shortwave diathermy Do not treat patients through their clothing Remove clothing from the treatment area Clothes made of synthetic materials should especially be removed because they insufficiently absorb moisture allowing moisture to pool on the skin causing local overheating Use a single layer of absorbent toweling between the patient and the applicator to absorb any perspiration produced during the treatment and to prevent the development of hot spots due to pooling of perspiration on the skin s surface Hand the Patient Safety Pull Cord to the patient before starting the treatment so that the patient can shut off the Auto Therm 390 if any discomfort is felt such as overheating or nausea Do not leave children or patients who cannot pull the cord unattended while operating this device 43 Heat effect of the applicators Evaluating the heat effect of various applicators is based on the heat felt by the patient This is strongly influenced by a number of factors e g the thickness of fat layers treatment through clothes or bandages blood circulation temperature of the skin etc Therefore the following sections generally explain the operation of the applicators used with the Au
17. nt possible material damage avoid the use of products based on halogen spliting compounds strong organic acids and oxygen spliting compounds solvents benzene and similar agents ATTENTION Do not allow any liquids to penetrate the unit or its accessories while cleaning and disinfecting Dry all sockets and connectors that have become wet before any further use 6 2 Protecting Other Electronic Devices from the Output of the Auto Therm 390 The Auto Therm 880 generates high frequency electric and magnetic fields that can penetrate walls ceilings and floors It cannot be prevented that components of these fields exist in the vicinity of the device Sensitive electronic instruments that are in the immediate vicinity of the Auto Therm 390 can be adversely affected by these fields This danger largely depends on the distance between the devices Therefore the Auto Therm 390 must not be installed any closer than 5 meters from other sensitive devices if possible farther The applicators should never be applied to sensitive devices e g neuromuscular electrical stimulators or their electrodes or cables This problem can be completely eliminated when the Auto Therm 390 is installed in a shielded room Le one containing a Faraday cage A Faraday cage is enclosed by a metal housing or grid that prevents the penetration of electric fields Use of absorbing curtains is enough for most treatment rooms We recommend that operators and other pers
18. ny bleeding tendency is increased by heating because of the increase in blood flow and iscularty of the heated tissues Care therefore should be used in treating patients with verapeutic shortwave diathermy who have bleeding disorders 3 Heating of the joint capsule in acute or subacute arthritis should be avoided 4 Use a single layer of toweling to absorb moisture during treatment with the inductive drum applicators 5 Shortwave diathermy may interfere with other electronic therapeutic devices such as neuromuscular stimulators and therapeutic ultrasound units Never use another electronic device on the same patient when shortwave diathermy is being applied 8 Use caution when treating obese patients with capacitive plate electrodes since this method of application may heat fat excessively E Auto Therm 390 Shortwave Diathermy Rov D 07 02 12 6 Maintenance and Troubleshooting 6 1 Cleaning and disinfecting Turn off the unit and unplug it from the main power supply before cleaning or disinfecting it Clean and disinfect the unit and its accessories except for the felt spacers with commercially available surface disinfectants To prevent damage to the surface materials of the Auto Therm 390 use only surface disinfectants based on agents Ike aldehydes alcohol or ammonium compounds that are suitable for wipe and spray disinfection Use them according to their instructions for use and duration of action To preve
19. ons keep a distance of at least 2 meters to applicators and cables while they are in use In case of doubt itis recommended to measure the field strength 63 _ Routine Maintenance 1 Standard medical electrical safety checks should be performed annually by qualified biomedical engineers or technicians trained to perform these procedures 2 Inspect cables and associated connectors for damage 64 Troubleshooting the Auto Therm 390 Symptom Action T Nothing ighis when main power Switch is Ts line cord connected to outlet tumed on Does tho outlet have power 2 E232 displays in the Output Display Window Patient pulled Patient Safety Pul Cord to stop the units output Find out why the cord was pulled and check the patient s skin under the applicators it necessary 3 E233 displays in the Output Display Window Coupling error Can occur during RFM treatment Melior Electronics Corp Rev D 07 02 12 4 E234 displays in the Output Display Window 5 E235 displays in the Output Display Window 6 E20 21 100 201 209 210 216 219 or 224 226 displays in the Output Display Window Inductive drum applicator is connected with capacitive soft rubber pad selected on membrane panel Inductive drum applicator not connected when the inductive drum is selected on the membrane panel Try turning the unit on and off again to see ifthe error code is resolved If not the unit requires servicing If problem is not addr
20. or manufacturer for additional suggestions The user may find FCC s Interference Handbook helpful It is available from the U S Government Printing Office Washington D C 20402 Stock No 004 000 00450 7 Notice 3 The manufacturer is not responsible for any interference caused by unauthorized modification to this equipment Mettler Electronics Corp 1333 S Claudina St Anaheim CA 92805 Toll Free 800 854 9305 Or 714 583 2221 Auto Therm 390 Shortwave Diathermy Rov D 07 02 12 TABLE OF CONTENTS FCC Frequency Interference Statement Domestic Only 2 1 Introduction 5 14 Introduction to the Auto Therm 390 5 12 Introduction to this manual 6 13 Safety precautions 6 14 Caution 6 15 Shipping damage 6 16 Package contents 6 17 Limited Warranty 7 2 Symbol Glossary and List of Abbreviations 9 2 1 Operational controls displays receptacles and symbols a 22 Abbreviations 10 3 Installation 3 1 Installation Instructions 32 Attaching the Auto Therm 390 to the Optional Cart 3 3 Attaching the Arm to the Cart 12 3 4 Flexible condenser applicators 13 3 5 Inductive drum applicator 13 4 Operating Instructions 15 A1 Shortwave diathermy applications 15 42 Preparing the patient 15 43 Heat effect of the applicators 15 44 Capacitive method of application with the soft rubber applicator 15 4 5 Inductive method of application 16 48 Choosing the applicator 17 4 7 General instructions 17 4 8 Instructons for o
21. ord to a grounded wall receptacle and the Auto Therm 390 d Do not plug in more than one applicator type at a time d Plug in either both soft rubber plate applicators or the inductive drum before starting treatment Couple the soft rubber plate applicators or the inductive drum to the patient before starting a trealment g Hand the Patient Safety Pul Cord to the patient H Note Descriptions of the symbols used on controls are in Section 2 7 Vote Electronics Corp Rev D_07 02 12 AutoxTherm 390 0000000 Intensity W Frequency OW Osus Oz Oops Os Oops mm Oops aan Oops DS OCW pusewiam CDI og OS AMerruea S8tectaonics A Figure amp 1 Autor Therm 390 Membrane Control Panel and LED Indicators Gen On ot os EE Owore tions for operation Turn on the main power switch located on the back of the unit by pressing T icon on switch The timer will display 0000 the intensity will display 0 and the LED indicators to the left of CW and will be blinking Plug in the applicator s that you wil be using for the treatment Apply them to the treatment area on the patient taking care to securely couple the applicator to the patient Hand the Patient Safety Pull Cord to the patient Instruct the patient to pull on the cord if the diathermy treatment becomes uncomfortable If the patient pulls the cord ll output will immediately stop E232 willb
22. peration 18 5 Indications Contraindications and Precautions for Shortwave Diathermy 21 S1 Indications 21 52 Contraindications 21 53 Precautions 2 6 Maintenance and Troubleshooting 23 6 1 Cleaning and disinfecting 23 6 2 Protecting other electronic devices from the output of the Auto Therm 390 23 6 3 Routine Maintenance 23 6 4 Troubleshooting 28 Melior Electronics Corp Rev D 07 02 12 7 Specifications 8 Accessories 8 1 Ordering Information 82 Auto Therm 390 Accessories LIST OF FIGURES 1 1 The Auto Therm 390 31 Auto Therm 890 Controls and Receptacles Located on the Back of the Unit 32 Oplional Auto Therm 390 Cart with unit arm and drum applicator attached 33 Mounting the Arm to the Cart 3 4 Auto Therm 390 Optional Applicator Arm Assembly A3 Auto Therm 390 Membrane Control Panel and LED Indicators 25 27 27 Auto Therm 390 Shortwave Diathermy Rev D 07 02 12 1 Introduction 1 1 Introduction to the Auto Therm 390 Thank you for purchasing the Auto Therm 390 shortwave diathermy This unit features soft rubber electrodes for the application of shortwave diathermy energy The unit may be used for continuous or pulsed shortwave diathermy with a variety of frequencies and pulse widths to choose from An easy to use membrane panel guides the user through setup A patient pullcord allows the patient to stop all output if the treatment becomes uncomfortable There is a handle molded into the cas
23. rm assembly Figure 3 4 for all the following instructions With the exception of the horizontal swing of arm and hinge 20 of the applicator holder 19 all hinges can be adjusted by the user as the treatment demands 3 The lock knob 26 on the lower rotating hinge 25 locks the whole arm in position even when the arm is maximally extended The other main hinges 21 23 that control the length and height of the arm can be adjusted after the appropriate knobs have been loosened To do this it is practical to support the arm above the hinge that has been loosened 20 or 21 4 The shaft of the inductive coll drum applicator is inserted into the aperture of applicator holder 19 until the slide 18 snaps shut 5 Remove the drum applicator by pushing back the slide 18 and pulling the applicator out of the holder The central joint of the arm 23 is factory adjusted to a basic braking force After long term application it may be necessary to readjust the basic braking force due to wearing d plastic parts and brake disks To adjust this joint remove the blue cover using a screwdriver and tighten the hexagonal bolt located below the cover by about a half turn 6 Adjust the breaking of the horizontal motion of the arm at the joint 25 by tightening the lower slotted nut Auto Therm 390 Shortwave Diathermy Rev D_07 02 12 Figure 3 3 Mounting The Arm To The Cart Hexagon bolt 18 18 20 Middle part of arm Lower
24. s Corp under this warranty must be paid by the buyer The buyer must also include a copy of the sales receipt or other proof of the date of purchase If the Product is returned without proof of the date of purchase it will be serviced as an out of warranty product at Mettler Electronics Corp s prevailing service rates Alteration misuse or neglect of the Product voids this warranty Except as specifically set forth above Mettler Electronics Corp makes no warranties express or implied including without limitation any implied warranty of merchantability or fitness for a particular purpose with respect to the Product if any implied warranties apply as a matter of law they are limited in duration to one year Mettler Electronics Corp shall not be liable for any indirect special consequential or incidental damages resulting from any defect in or use of the Product Any legal action brought by the buyer relating to this warranty must be commenced within one year from the date any claim arises and must be brought only in the state or federal courts located in Orange County California Some states do not allow limitations on how long an implied warranty lasts or the exclusion or limitation of incidental or consequential damages so the above limitations or exclusions may not apply to the buyer This warranty gives the buyer specific legal rights and the buyer may also have other rights which vary from state to state Vote
25. the output display At the end of tuning a will be displayed in the output window it tuning is successful If any problems are encountered during this time an error code willbe displayed inthis window 3 Meter Electronics Corp Rev D 07 02 12 14 16 Use the Up arrow to adjust the output in Watts In the CW mode the maximum is 100 Watts In any of the pulsed modes the maximum is 200 Watts Press Enter to stop blinking the intensity display Press the Hold button to momentarily stop a treatment The Auto Therm 390 will remember all the treatment parameter settings However the output display will reset to 0 watts You will need to repeat steps 12 through 14 to resume the session While you are in the hold mode you can also adjust any of the other parameters before restarting the treatment session At the end of treatment an alarm will sound and the time and output displays will return to zero EJ Auto Therm 390 Shortwave Diathermy Rev D 07 02 12 5 Indications Contraindications and Precautions for Shortwave Diathermy 51 Indications Shortwave diathermy delivers energy in the radio band of 27 12 MHz to provide deep heating therapeutic effects to body tissues When shortwave diathermy is delivered to the body at intensities capable of generating a deep tissue temperature increase it can be used to treat selected medical conditions such as 1 Relieving pain 2 Reducing muscle spas
26. to Therm 390 The heat effect of the electrodes difers fundamentally between the inductive and capacitive method of application 44 Capacitive method of application with the soft rubber applicators The capacitive method of application transforms the shortwave energy into heat in tissues of low blood circulation e g fat connective tissues Thus tissues near the skin s surface are heated so that 5 Mettler Electronics Corp Rev D 07 02 12 not only the transformed energy but also the distance between applicator and skin is important for the subjective sensation of heat For best results place soft rubber electrodes parallel to the skin s surface This will spread out the energy over the entre treatment area and prevent hat spots Do not allow the soft ubber applicator cables to come into contact withthe patient If necessary insulate Ihe patient rom the cables using several layers of toweling 1 Place the rubber plate electrodes inside the cloth cover Place one felt spacer on each side of the electrode inside the bag The electrode should be sandwiched between the two felt spacers Press the two Velcro squares to close each bag 2 Plug the cables into the back of the Auto Therm 390 The cables plug into the grey receptacles and should be plugged in as far as possible 3 Place the covered rubber plate electrodes directly on the skin not over clothing Strap the electrodes to the patient using the Velcro straps or t
27. tranics Corp or by a service facility authorized by Mettler Electronics to perform warranty repair work A service manual for the Auto Therm 390 is available fram Meter Electronics Corp for a nominal charge 14 Caution Federal law restricts the sale of this device to or on the order of a physician dentist veterinarian or any other practitioner licensed by law of the state in which he practices Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous exposure to radio frequency energy Treatment should be administered only under the direct supervision of a health care professional 15 Shipping Damage Your new Auto Therm 390 is shipped complete in one or mare cartons Upon receipt please inspect the cartons and the unit and its accessories for visible and hidden damage If you discover any damage hold al shipping materials including the carton and call the shipping agent who delivered the unit They are responsible for all damage in transit therefore all claims should be filed directly with them The factory will not be responsible for any damage in shipment nor allow any adjustments unless proper formal claim has been filed by the receiver against the carrier The carton in which your new Auto Therm 390 was received is specially designed to protect the unit during shipping Please retain all shipping materials in the event that you will need to return your unit tor servicin
28. um always keep the surface of the drum in contact with the patient during treatment It the patient responds to excessive output power intensity by moving away from the drum applicator rather than pulling the ripcord insufficient coupling to the body may occur If his is allowed to happen over a period of time the adjusted electric output power will be converted into heat not only in the patient but also in the drum applicator This overheating may cause damage to the drum applicator It is imperative for this reason to frequently check the patient when using the drum applicator 46 Choosing the applicator Choose the appropriate applicator dependent upon the area and volume of the body parts to be treated and whether inductive coil or condenser feld application is preferred Press the buttons on the membrane panel o indicate the applicator that you intend to use for this treatment The following selectors are assigned to the corresponding applicators PIN Applicator Description Applicator lustration Symbol patos 3980 14 cm diameter inductive coil applicator of Soft rubber 12 x 18 om condenser e 3909 applicators oc 4 7 General Instructions Before you start 2 Review precautions contraindications and side effects adverse reactions listed in Section 5 H Ask patient about possible implants before deciding to administer shortwave diathermy If patient has an implant do not use this device c Verify connection of the line c
29. uto Therm 390 shortwave diathermy unit 7 Once you have verified proper functioning of your Auto Therm 390 using the instructions in Section 4 please fil in the enclosed self addressed Warranty Registration Card and mail it to Mattler Electronics Figure 3 1 Auto Therm 390 Controls and Receptacles Located on the Back of the Unit 3 2 Attaching the Auto Therm 390 to the Optional Cart 1 Remove the model 96 car from the packing material See Figure 3 2 2 Remove the five rubber feet from the bottom of the unit 3 Place the Auto Therm 390 on top of the cart Attach it to the top of the cart by screwing four of the rubber feet to the bottom of the unit from the inside of the cart It necessary use a Philip head screwdriver to tighten the screws in the feet Save the fifth rubber foot for use without tho car m Matter Electronics Corp Rev D 07 02 12 e Fae 32 Opa Ao Tm 390 an win unt um and soc 3 3 Attaching the Arm to the Cart 1 The optional arm is mounted on the side of the cart by means of keyhole type connectors See Figure 3 3 for positioning After slideing the arm into postion tighten the nuts using the wrench provided with the cart to Secure the arm to the cart The inductive drum applicator requires the arm to be installed The clinician can safely position the applicator in many differing treatmer positions because the arm has five locking joints 2 Refer to the illustration of the a