Home

SRV949R Manual

image

Contents

1. Ensure that nothing is in contact with the scale system during this procedure and that all six 6 feet are touching a level surface Remove hands from the system when noting the displayed calibration results Use a known calibrated weight traceable to NIST STEP 1 First place scale platform on a flat surface and adjust scale feet so that scale is level and supported by the feet STEP 2 Perform a Two Point Calibration with 400 pounds full scale according to Steps 3 8 Continued next page SR Instruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2011 SRgR Instruments Inc Model SRV949R 24 X 60 Remote Display Platform Scale Operating and Service Manual S N 1000 Part No MAN949R 110502 Page 11 of 16 CALIBRATION con d STEP 3 To access all settings below press the CAL button S2 through the access hole in the back of the display housing Figure 5 Calibration Button STEP 4 The display will read CAL The Right Button WEIGH on the front of the display housing allows you to toggle through the menus without saving any parameters The Left Button ZERO on the front of the display housing permits a menu item to be selected or a parameter saved CAL MENU data Displays saved calibration and model information Unit Set display units Lb Kg Both SEC Set Auto Power Down Timer betwe
2. S N 1000 Part No MAN949R 110502 Page 8 of 16 BUTTON FUNCTIONS Sos oe SR 23533 by SR Instruments Inc 8 WEIGH ZERO MODE Operating Insructions Press WEIGH to wake up scale Press and hold ZERO for 3 sec Press MODE to select Ib or kg Figure 3 Button Display ZERO The ZERO button is used to turn the scale on or off Press and hold for 3 seconds to zero the display MODE The MODE button is used to select pounds or kilograms WEIGH The WEIGH button is used to wake up the scale prior to weighing the patient BASIC SYSTEM OPERATION TURNING SYSTEM ON Make sure the scale is free and clear of any obstructions and all six 6 feet are touching a level surface Press the ZERO button and hold for 3 seconds The display will read ZERO then and display 0 0 when system is ready for weighing ec Continued next page SR Instruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com Copyright 2011 SR R Instruments Inc email sri Qsrinstruments com Model SRV949R 24 X 60 Remote Display Platform Scale Operating and Service Manual S N 1000 Part No MAN949R 110502 Page 9 of 16 BASIC SYSTEM OPERATION cont d UNIT Press the MODE button to toggle between pounds or kilograms The system will remain in the same mode as last used unless manually changed Example If the last
3. 2011 SRgR Instruments Inc Model SRV949R 24 X 60 Remote Display Platform Scale Operating and Service Manual S N 1000 Part No MAN949R 110502 Page 3 of 16 PACKING CHECKLIST Model SRV949R Portable Platform Scale 4 DESCRIPTION QUANTITY PORTABLE PLATFORM SCALE lea REMOTE DISPLAY lea DISPLAY EXTENSION CABLE lea AA 1 5 VOLT ALKALINE BATTERIES 6 ea CALIBRATION CERTIFICATE lea WARRANTY CARD lea MANUAL lea SR Instruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2011 SR R Instruments Inc Model SRV949R 24 X 60 Remote Display Platform Scale Operating and Service Manual S N 1000 Part No MAN949R 110502 Page 4 of 16 ASSEMBLY Figure 1 Assembled SRV949R To assemble follow steps below STEP 1 Turn scale over and connect extension cable to connector on frame assembly STEP 2 Connect other end of extension cable to display coil cable To install included six 6 AA 1 5 volt alkaline batteries follow steps below STEP 1 Unlatch tab on the Battery Compartment cover on bottom of scale figure 2 STEP 2 Insert batteries according to instructions on the Battery Compartment holder STEP 3 Replace Battery Compartment cover E Figure 2 Battery Compartment Continued next page SR Instruments Inc 600 Young Str
4. weight displayed was in kilograms and the scale shuts down the weight mode will be in kilograms when the scale is turned back on WEIGH The WEIGH button is used to wake up the scale prior to weighing the patient BATTERY REPLACEMENT The Portable Platform Scale contains six 6 AA 1 5 volt alkaline batteries The low battery indicator is displayed when the battery power is insufficient to power the scale Refer to Assembly Instructions Figure 2 SR Instruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2011 SRgR Instruments Inc Model SRV949R 24 X 60 Remote Display Platform Scale Operating and Service Manual S N 1000 Part No MAN949R 110502 Page 10 of 16 CALIBRATION IMPORTANT CALIBRATION CHECK Qualified service personnel only should perform this procedure Load cells have no user serviceable components and should not be tampered with for any reason Re calibration is generally not required but should be verified periodically to ensure accuracy The recommendation for calibration check is at least once every 12 months or as individual maintenance policy requires SR 3535 by SR Instruments Inc bog 5 ZERO MODE WEIGH Operating Insructions Press WEIGH to wake up scale Press and hold ZERO for 3 sec Press MODE to select Ib or kg Figure 4 Button Display NOTE
5. Model SRV949R 24 X 60 Remote Display Platform Scale Operating and Service Manual S N 1000 Part No MAN949R 110502 Page 1 of 16 SRScaleg by SR Instruments Inc Model SRV949R Portable Platform Scale Operating and Service Manual Serial Numbers 1000 SR Instruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2011 SRgR Instruments Inc Model SRV949R 24 X 60 Remote Display Platform Scale Operating and Service Manual S N 1000 Part No MAN949R 110502 Page 2 of 16 TABLE OF CONTENTS TABLE OF FIGURED m 2 PACKING CHECKLIST MODEL SRV949R scscssscssssccsccscscscscsscscscssecssscscssesssesencssessoes 3 FS Mf 4 SYSTEM DESCRIPTION AND INTENDED USE eese eese eee eese testen seen sess n stas senses stone 5 MAINTENANCE AND CLEANING i cccscossissccsnscessenscassonteccsoasssosssaoeceenseonscesseoscassodstcasecssonssesseessies 6 STORAGE AND TRANSPORTATION sisiscsccssncacsescovtscscsseccesesascneeanencsovsyeisnnessessentensensecdpsucaneasexes 6 SPECIFICATIONS iiiccioitsphstevi irati PROS Y SEE HP OPEP REN YE PLA MEER ENSE E Ae bo siiis osso 00s 455500 7 BUTTON FUNCTIONS iicsceoniwxvukeUe ne eh oH XVRCHT NC PN Y Y UERNE FEN ORQLLHETYNCE C00 ssas SE ERR isos QUAE ER URN UA E RLSK UE YER HUE 8 BASIC SYSTEM OPTRA TION sisscccsssccc
6. anda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2011 SRgR Instruments Inc Model SRV949R 24 X 60 Remote Display Platform Scale Operating and Service Manual S N 1000 Part No MAN949R 110502 Page 13 of 16 CALIBRATION con d STEP 8 Save the parameter changes while in the CAL mode by pressing the Left Button when the display shows SAVE To exit the CAL mode without saving the changes press the Left Button when the display shows quit CALIBRATION TOLERANCE FOR SRV949R Low Limit Applied Load High Limit Ib Ib Ib 0 399 400 401 CAUTION The integrated circuits and semiconductors on the printed circuit boards may be damaged by electrostatic discharge ESD Be sure to use proper handling precautions at all times TROUBLESHOOTING SYMPTOM REASON CORRECTIVE ACTION Display reads LOW Battery needs to be replaced Display reads LOW Ensure scale is on a level surface Check to make sure all six feet are touching ground and ensure that nothing is touching bottom or sides of platform For additional information or assistance phone our Service Hotline 1 800 654 6360 or e mail sri srinstruments com SR Instruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2011 SRgR Instruments Inc M
7. cesascnssesnscaunntscuiscetssstexsasocentcoassenswasauns HR QUARE I Foe uiu Ie E E 8 BATTERY REPLACEMDBNTC i iipooen orn nir ien ER Pis bera Hi DROP Dr TEES I I FEVER 9 LB TLLA V CO Pe M 10 CALIBRATION COND c csnsa s cis oQubn Nena FEE nU Rue EDEN tERS RN QE SENT RE EXRS CEPR SERE eHMFE IE QEKEN QUK INE tassie satus 11 CALIBRATION CON DLsceiseesckhtut eh ce vh ee ei IN eo PEH VIR EKe VERRE EON Ue bd XR NREE E ERUNT INL UE TE OH REGE RCEOU TEE s sitos UR QR 12 CALIBRATION CON LE rianesbPetebtu Hee oHHEFEREQUUS SEDI sedisse FLEERMFLUQRRREER ITE Y svs PESE NEQU PREREM RUE 13 TROUBLESHOOTING ierkxnset exe i ERES PROUVER ERE EERVENVE EXE HER MPEIE QE VUE UHR NE PITEEN SE EEXE mana duane 13 WARRANTY 14 LO d Wr m T 15 TABLE OF FIGURES Figure 1 Assembled SEV pri Mc 4 Figure 2 Battery Compartment sesssesssocesooessocesoeessecssocsscossoocssscessocesocssoosesoessseessocssoosssossssesssese 4 Figure 3 Button Display eee 8 Fig r 4 Button Display m 10 Figure 5 Calibration Button eosssoessscessecesocesocessocesseessocssocsscosssoessoesssocesocssoosssocesseessocssocsssosssse 11 SR Instruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright
8. eet Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2011 SRgR Instruments Inc Model SRV949R 24 X 60 Remote Display Platform Scale Operating and Service Manual S N 1000 Part No MAN949R 110502 Page 5 of 16 REPLACEMENT PARTS AND ACCESSORIES Part Description CA8467 Remote Display Label SE8507 Extension Cable MF2284T52 Replacement Foot SYSTEM DESCRIPTION AND INTENDED USE SYSTEM DESCRIPTION The SRV949R Portable Platform Scale is a low profile design which allows the animal easy access The SRV949R Scale can be used in a fixed location or as a portable scale The Scale employs the latest in microprocessor and load cell technology to provide accurate and repeatable weight data Six 6 identically matched transducers are strategically placed to ensure an accurate representation of the patient s weight The patient s weight is displayed on an LCD screen With a push of a button weight data may be viewed in either pounds or kilograms INTENDED USE The SRV949R Portable Platform Scale is a lightweight portable scale designed for use as a weighing system for animals not exceeding 400 pounds or 182 kilograms gross weight Please read the following warnings to avoid injury to the animal or attendant WARNING DO NOT USE FOR ANIMAL TRANSPORT DO NOT EXCEED MAXIMUM LIMIT OF 400 LBS 182 KG SR Instruments Inc 600 Young St
9. en 15 and 300 seconds orAv Set Gravitational Acceleration in m s FULL Applied calibrated mass at full scale HALF Used with three point calibration curve Applied calibrated mass at a second point between zero and full scale usually at half ZERO No calibrated weight applied 2 Pt Two point calibration curve zero and full scale 43 Pt Three point calibration curve zero half and full scale SAVE Saves the calibration data qUIt Quits without saving the calibration data Continued next page SR Instruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2011 SRgR Instruments Inc Model SRV949R 24 X 60 Remote Display Platform Scale Operating and Service Manual S N 1000 Part No MAN949R 110502 Page 12 of 16 STEP 5 CALIBRATION con d Set Auto Power Down Timer NOTE Set time to 60 seconds prior to QC inspection a b C STEP 6 Select the Auto Power Down Timer from the CAL menu by pressing the Left Button when the display shows SEC The display will show four digits 0123 Set the desired time in seconds by selecting the first digit and toggle through the digits until desired value is shown Press the Left Button and continue the same with the remaining digits The value can not be less than 15 or greater than 300 Press the Left Button to save or Right Butto
10. n to toggle to quit Select quit by pressing the Left Button to quit without saving changes Set the Full Scale value NOTE Full scale value is 400 pounds a b STEP 7 Select the Full Scale from the CAL menu by pressing the Left Button when the display shows FULL The display will show only the five digits lt XXXX X gt with the thousand place flashing 1 Toggle to the next digit by pressing the Right Button Set the full scale value maximum load in pounds by toggling to the hundred place by pressing the Right Button Select the hundred place digit and toggle through the digits until desired value is shown Press the Left Button and continue the same with the remaining digits Press the Left Button to save or Right Button to toggle to quit Select quit by pressing the Left Button to quit without saving changes Calibrate the scale with Two point calibration curve zero and full scale Select Two point calibration curve by pressing the Left Button when the display shows 2Pt Zero the scale by pressing the Left Button when the display shows zero Set the full scale point by placing the calibrated mass equal to the predetermined full scale value on the scale and pressing the Left Button Press the Left Button to save or Right Button to toggle to quit Select quit by pressing the Left Button to quit without saving changes Continued next page SR Instruments Inc 600 Young Street Tonaw
11. nts com Copyright 2011 SRgR Instruments Inc Model SRV949R 24 X 60 Remote Display Platform Scale Operating and Service Manual S N 1000 Part No MAN949R 110502 Page 15 of 16 NOTES SR Instruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2011 SRgR Instruments Inc Model SRV949R 24 X 60 Remote Display Platform Scale Operating and Service Manual S N 1000 Part No MAN949R 110502 Page 16 of 16 SRScaleg By SR Instruments Inc Precision amp Technology in Perfect Balance SR Instruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2011 SRgR Instruments Inc
12. ny misuse improper installation or tampering shall void this warranty DAMAGED SHIPMENTS Title passes to purchaser upon delivery to Transportation Company Any claims for shortage or damage should be filed with the delivery carrier by purchaser RETURN POLICY All products being returned to SR Instruments Inc require a Return Goods Authorization number RGA To receive an RGA call our Technical Service Team at 716 693 5977 or toll free in the USA and Canada at 800 654 6360 When inquiry is made please supply model and serial numbers purchase order if the scale was bought on contract and reason for return Generally deleted damaged and outdated merchandise will not be accepted for credit A minimum restocking charge of 15 will be assessed on return of current merchandise All returns are to be shipped FREIGHT PREPAID to SR Instruments Inc 600 Young Street Tonawanda NY 14150 RESTOCKING FEE e 15 fee for any scale that has been opened and used e 10 fee for any scale returned that has been ordered incorrectly or refused delivery with no model change e 5 fee if an error in ordering has been made and a different model exchanged e No fees will be charged if the scale is returned because of an error on the part of SR Instruments Inc e Noreturns accepted after 60 days SR Instruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstrume
13. odel SRV949R 24 X 60 Remote Display Platform Scale Operating and Service Manual S N 1000 Part No MAN949R 110502 Page 14 of 16 WARRANTY TWO 2 YEAR LIMITED WARRANTY Each SRSecales system is manufactured with high quality components SR Instruments Inc warrants that all new equipment will be free from defects in material or workmanship under normal use and service for a period of two 2 years from the date of purchase by the original purchaser Normal wear and tear injury by natural forces user neglect and purposeful destruction are not covered by this warranty Warranty service must be performed by the factory or an authorized repair station Service provided on equipment returned to the factory or authorized repair station includes labor to replace defective parts Goods returned must be shipped with transportation and or broker charges prepaid SR Instruments Inc s obligation is limited to replacement of parts that have been so returned and are disclosed to SR Instruments Inc s satisfaction to be defective The provisions of this warranty clause are in lieu of all other warranties expressed or implied and of all other obligations or liabilities on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no event shall SR Instruments Inc be liable for any subsequent or special damages A
14. reet Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2011 SRgR Instruments Inc Model SRV949R 24 X 60 Remote Display Platform Scale Operating and Service Manual S N 1000 Part No MAN949R 110502 Page 6 of 16 MAINTENANCE and CLEANING The SRV949R Portable Platform Scale is made with a powder coated aluminum platform Exercise caution when cleaning the display window as it is made of a clear plastic and can be scratched by abrasive cleaners Mild soap and water is recommended for general cleaning and disinfecting WARNING DO NOT use pressurized water or steam The scale system contains microprocessor circuitry and strain gauge sensors that may be adversely affected by exposure to such an environment STORAGE and TRANSPORTATION STORAGE If storing this equipment for periods longer than three 3 months remove the batteries To maintain proper operation of this instrumentation storage and transport conditions should not vary outside the following conditions Relative Humidity 0 to 85 Ambient Temperature 14 F to 122 F 10 C to 50 C TRANSPORTATION The SRV949R Portable Platform Scale has four 4 recessed handles on the underside of the scale for easy portability SR Instruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2011 SRgR Instr
15. uments Inc Model SRV949R 24 X 60 Remote Display Platform Scale Operating and Service Manual S N 1000 Part No MAN949R 110502 Page 7 of 16 SPECIFICATIONS MAXIMUM WEIGHT CAPACITY 400 Ib 182 kg 24 in x 60 in x 2 in PLATFORM p SIAN 61 cm x 153 cm x 5 cm DISPLAY TYPE 75 LCD display with 4 5 active digits DISPLAY RESOLUTION 0 21b 0 1 kg ACCURACY 0 2 2 digits of displayed resolution for calibrated range AUTO ZERO One button operation Hold for zero AUTO POWER DOWN Approximately 120 seconds AVERAGING Automatic digital filter POWER SUPPLY Six 6 AA 1 5 volt alkaline batteries CALIBRATION Calibration is traceable to NIST standards Normal operating conditions for this product OPERATING CONDITIONS Ambient Temperature Range 68 F to 85 F 20 C to 30 C Relative Humidity Range 0 85 Avoid exposure to high pressure water or steam Storage and transport conditions should not vary outside the following conditions Relative Humidity TRANSPORT and STORAGE 0 to 85 Ambient Temperature 14 F to122 F 10 C to 50 C Remove batteries if storing longer than three 3 months SR Instruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri Qsrinstruments com Copyright 2011 SR R Instruments Inc Model SRV949R 24 X 60 Remote Display Platform Scale Operating and Service Manual

Download Pdf Manuals

image

Related Search

Related Contents

User Manual PDF - Lemos International  MIL-COTS製品は  平成17年度医療機器自主回収一覧  Yamaha Drums BS-751 User's Manual  WMIX-D5250  User manual    Display Shield AI - Protective Enclosures Company  Airlink101 AWLL6077V2 Network Card User Manual  DE VINGT ANS  

Copyright © All rights reserved.
Failed to retrieve file