Home
Chapter 3 Operation - Spectrum Medical X
Contents
1. The HF Series X ray Generator is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the HF Series X ray Generator can help pre vent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the HF Series X ray Generator as recommended below according to the maximum output power of the communications equipment Rated maximum output Separation distance according to frequency of transmitter power of transmitter m K 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz d 1 2VP d 1 2VP d 2 3VP 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation dis tance d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people ZZE SHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HEHEHE ZZ ZZ ZZ ZZ
2. 1mS NOTE AEC TECHNIQUES SHOULD HAVE EXPOSURE TIMES EXCEEDING 8 MILLISECONDS mAs is tube dependent the generator may not reach maxi mum mAs due to tube type SHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HEHEHE HHH HHH HEHEHE ZZ HHH HHH HHH HHH HHH SD 2 V SC HF Series X ray Generators Operator s Manual Revision Q b Quantum Medical Imaging LLC 2 5 Refer to Table 2 2 for performance specifications of single phase stored energy SE HF Series X ray Generators Table 2 2 HF Series Single Phase Stored Energy Generator Performance Specifications Maximum kW N A 32 40 50 mA Stations 258 50S 758 258 50S 75S 25S 50S 75S Small Focus S 1508 1508 150S Large Focus L MA 100L 160L 200L 100L 200L 250L e Sie Ge 250L 320L 400L 320L 400L 500L i i 600L mA Accuracy N A 5 1mA 5 1mA 5 1mA kVp Range kVp N A 40 125 40 125 40 125 kVp increments kVp N A 1 0 1 0 1 0 step kV Accuracy N A 4 4 4 150 kVp optional N A No No No Time Range sec N A 0 001 6 3 0 001 6 3 0 001 6 3 Time Accuracy N A 1 MS 0 5 1 MS 0 5 1 MS 0 5 Minimum Pestle N A 0 001 0 001 0 001 Time seconds mAs Range N A 0 025 400 0 025 400 0 025 400 High Speed Starter N A No No No a Voltage out N A 5 5 5 Maximum exposure time is a function of kV and mAs setting
3. QG 32 SE QG 3200 SE 320 mA 100 kV 400 mA 80 kV QG 40 SE QG 4000 SE 320 mA 125 kV 400 mA 100 kV 500 mA 83 kV QG 50 QG 50 2 QG 5000 QG 5000 2 400 mA 125 kV 500 mA 100 kV 650 mA 76 kV QG 50 SE QG 5000 SE 400 mA 125 kV 500 mA 100 kV 600 mA 83 kV QG 65 QG 6500 500 mA 130 kV 650 mA 123 kV 800 mA 81 kV QG 80 QG 8000 650 mA 123 kV 800 mA 100 kV HF Series X ray Generators Operator s Manual Quantum Medical Imaging LLC PHO HHH HOE HHF OOH HOF EEE HFEF HOF HOF HEE HEF EHH HOE HHH OOH TOOTH HHH EHH HEH EHH EE EEE EE III IDEE Highest constant output power at 100 kV 0 1 second s Table 2 4 Electric Output Data N A N A QG 32 QG 32 2 QG 32 3 QG 32 5 QG 3200 QG 3200 2 QG 3200 3 QG 3200 5 32 kW 320 mA 100 kV 0 1 5 QG 40 QG 40 2 QG 40 3 QG 40 5 QG 4000 QG 4000 2 QG 4000 3 QG 4000 5 40 kW 400 mA 100 kV 0 1 5 N A N A QG 32 SE QG 3200 SE 32 kW 320 mA 100 kV 0 1 s QG 40 SE QG 4000 SE 40 kW 400 mA 100 kV 0 1 s QG 50 QG 50 2 QG 5000 2 QG 50 SE QG 5000 SE QG 65 QG 6500 50 kW 500 mA 100 kV 0 1 s 65 kW 650 mA 100 kV 0 1 s QG 80 QG 8000 50 kW 500 mA 100 kV 0 1 5 80 kW 800 mA 100 kV 0 1 s Nominal shortest irradiation All models 0 008 second time AEC exposures Lowest mAs All models 0 02
4. 320 mA 150 kV 250 mA with QG 150 option N A N A QG 32 SE QG 3200 SE 125 kV 250 mA QG 40 SE QG 4000 SE 125 kV 320 mA QG 50 QG 50 2 QG 5000 QG 5000 2 125 kV 400 mA 150 kV 320 mA with QG 150 option QG 50 SE QG 5000 SE 125 kV 400 mA QG 65 QG 6500 150 kV 800 mA QG 80 QG 8000 150 kV 800 mA Maximum X ray tube current N A N A and highest X ray tube volt 96 37 0G 32 2 QG 32 3 500 mA 64 kV age obtainable at that cur QG 32 5 QG 3200 QG 3200 2 SR QG 3200 3 QG 3200 5 QG 40 QG 40 2 QG 40 3 500 mA 80 kV QG 40 5 QG 4000 QG 4000 2 QG 4000 3 QG 4000 5 N A N A QG 32 SE QG 3200 SE 400 mA 80 kV QG 40 SE QG 4000 SE 500 mA 80 kV QG 50 QG 50 2 QG 5000 2 650 mA 76 kV QG 50 SE QG 5000 SE 600 mA 83 kV QG 65 QG 6500 800 mA 81 kV QG 80 QG 8000 800 mA 100 kV Revision Q HF Series X ray Generators Operator s Manual 2 8 Quantum Medical Imaging LLC Combination of X ray tube current and X ray tube volt age resulting in highest out put power Note All mA stations below those listed Table 2 4 Electric Output Data N A N A QG 32 QG 32 2 320 mA 100 kV QG 32 3 QG 32 5 400 mA 80 kV QG 3200 QG 3200 2 500 mA 64 kV QG 3200 3 QG 3200 5 QG 40 QG 40 2 400 mA 100 kV QG 40 3 QG 40 5 500 mA 80 kV QG 4000 QG 4000 2 QG 4000 3 QG 4000 5 N A N A
5. OOOOH HOOD HF Series X ray Generators Operator s Manual Revision Q Quantum Medical Imaging LLC i THIS PAGE INTENTIONALLY LEFT BLANK THO HOE HEHEHE HHH FOE HHH HEH II III HEF OHHH HHH HHH HHH HOH HHH HHH HHH HHH HHH HHH HHH HEH HHH HHH HHH HHH HHH HHH HHH HEH HHH EHH HHH EHH HHH HOH HOOD Revision Q HF Series X ray Generators Operator s Manual ii Quantum Medical Imaging LLC CHAPTER 1 SAFETY NOTICES GENERAL SAFETY INFORMATION c ccecceeseseseeeetseeeeeeetetssueesestansanees 1 3 WARNINGS CAUTIONS NOTES 1 3 REGULATORY COMPLIANCE ccccecceceesseeeeseesesseeeeeersesseeeetensaaaeeetentea 1 5 CLASSIFICATION cccccesccccecceseseeseeteesenseseetsnseeeeeeessaaeeseeetansesseteas 1 5 INTENDED USE AE 1 6 COMPATIBILITY cccceccccceeeeceeceesesenseeeeeseesesesteesenseeeetseaensestanseness 1 6 INTENDED OPERATOR aeren dese 1 6 TRAINING aaa da aa eege ENEE Ee aa iiaei 1 6 ACCOMPANYING DOCUMENTATION ees 1 6 APPLICABLE STANDARDS ccccccccssseeeeeeseseeeeeeetanseseeeeessneeeeteensaees 1 7 DISPOSAL OF BATTERIES AND ACCUMULATORS DIRECTIVE 2006 66 EC ccccceecccccesseeeseeteesssseseeetetsnaeeseseeenanees 1 8 ELECTROMAGNETIC COMPATIBILITY EN 60601 1 2 2007 IEC e0o6e0t 1L 23 2001 1 8 ABBREVIATION DEFINITION cccccccceseseeseeeeesnseseeeeetsnsesseetensneesees 1 14 CHAPTER 2 GENERAL I NFORMATION OVERVIEW Ze ege 2 3 INTENDED USE ccccscissecesissisvanndesavsvectaenvaiscncedteass vd teats tiea
6. expressed in thousands of Joules as the kJ value Built in tube protection alerts the operator prior to a subsequent exposure until sufficient heat has dissipated from the tube to allow the next exposure without exceeding tube limits If the heat units exceed the maximum heat limit the READY LED will blink on and off The READY LED will only stop blinking when the tube has cooled down sufficiently such that the heat produced by taking an exposure at the current kV mA and time mAs setting will not exceed tube maximum heat unit limit FAULT MESSAGES Under certain conditions the HF Series X ray Generator shows a dialog box in the top center of the screen containing a message This message informs the user that either a problem has occurred or that a requested action cannot be per formed The user must read these messages carefully They will provide information on what to do from then on This will be either performing an action to resolve the problem or to contact the technical service support organization Many minor problems can be corrected without a service technician To clear an error mes sage press any key on the operator control panel except for the PREP and EXPOSE keys Refer to the HF Series X ray Generators Service Manual for sys tem faults that require the attention of a technical service support representative NOTE If the READY light on the operator control panel ch is off or flashing this indicates a fault has occurred
7. it means the system shut itself down automatically due to inactivity refer to Automatic Power Stand By Mode para graph in this chapter In this case set the power on standby switch to the standby position wait five seconds then set switch to the on position SHH SHH HHH HHH HHH FHF HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH ZZ ZZ ZZ HHH HHH HHH HHH HHH HEHEHE HHH HHH HHH HEH HOS OHHH HOOD 2 V SE HF Series X ray Generators Operator s Manual Revision Q b Quantum Medical Imaging LLC 3 3 DT Revision Q 3 4 3 The system automatically runs a series of self checks for approxi mately three to six seconds to ensure proper operation Quantum Medical Imaging and the OCP software revision level appears on the display 4 Upon successful completion of self tests the Manual or APR mode menu is displayed depending on the operating mode used when the system was last shut down Power Off Procedure For Short Periods Of Time 1 Set the power on standby switch on the Operator Control Panel OCP to the standby OO position 2 Ensure OCP is dark System is now shut down For Long Periods Of Time 1 Set the power on standby switch on the Operator Control Panel OCP to the standby position 2 Locate the main circuit breaker on the generator cabinet see Fig ure 2 2 and set to the OFF position System is now shut down 3 If shut down period is anticipated to be longer than 15 days
8. should assure that it is used in such an environment Immunity test IEC 60601 Compliance Electromagnetic environment guid test level level ance Portable and mobile RF communications equipment should be used no closer to any part of the HF Series X ray Generator including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance Conducted RF 3 Vrms 3 Vrms JP IEC 61000 4 6 150 kHz to 80MHz d 1 2VP Radiated RF 3 V m 3 V m JP IEC 61000 4 3 80 MHz to 2 5 GHz d 1 2VP op MHz to 800 MHz d 23VP ong MHz to 2 5 GHz where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation dis tance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compli ance level in each frequency range Interference may occur in the vicinity of equipment marked with the following sym bol Uer SHH SHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HEHEHE HHH HHH HEHEHE HHH HHH HHH HHH HHH SD S j SC HF Series X ray Generators Operator s Manual Revision Q b Quantum Medical Imaging LLC 1 11 Guidance and manufacturer s declaration electromagnetic immunity NOTE 1 At 80 MHz
9. HEH HHH EEE EE HOH EE HOH HEH HHH HHH HHH HHH HHH HHH HEH EHH HHH HHH HHH HHH HOH HEHEHE HEH EHH HEH HHH HEH HHH Revision Q HF Series X ray Generators Operator s Manual 6 4 Quantum Medical Imaging LLC BUYER S REMEDIES If Quantum Medical Imaging LLC determines that any Product fails to meet any war ranty during the applicable warranty periods Quantum Medical Imaging LLC shall correct any such failure as follows A By repairing adjusting or replacing any defective or non conforming Parts or Products B By making available any necessary repaired or replacement parts or assemblies Quantum Medical Imaging shall have the option to furnish either new or exchange replacement parts or assemblies All returned parts shall become the property of Quantum Medical Imaging if said parts have been determined by QMI to be defective by reason of manufacture The preceding Paragraphs set forth Buyer s Remedies and Quantum Medical Imag ing s sole liability for claims based upon failure of the product to meet any warranty whether the claim is on contract warranty Tort including negligence and strict liabil ity or otherwise and however instituted And upon the expiration of the applicable warranty period all such liability shall terminate In no event shall Quantum Medical Imaging be liable for special or consequential damages arising out of the use or ability to use its equipment whatsoever The warranties and remedies available to the
10. HHH D S SC HF Series X ray Generators Operator s Manual Revision Q ka Quantum Medical Imaging LLC 1 7 DISPOSAL OF BATTERIES AND ACCUMULATORS DIRECTIVE 2006 66 EC In accordance with the European Directive 2006 66 EC batteries and accumulators are labeled to indicate that they are to be collected separately and recycled at end of life The label on the battery may also include a chemical sym bol for the metal concerned in the battery Pb for lead Hg for mercury and Cd for cadmium Users of batteries and accumulators must not dispose of batteries and accumula tors as unsorted municipal waste but use the collection framework available to customers for the return recycling and treatment of bat teries and accumulators Participation is important to minimize any potential effects of batteries and accumulators on the environment and human health due to the potential presence of hazardous substances ELECTROMAGNETIC COMPATIBILITY EN 60601 1 2 2007 IEC 60601 1 2 2007 The HF Series X ray Generator is intended for use in the electromagnetic environment specified below As such the generator must be installed and put into service accord ing to the information provided in the accompanying Service Manual The HF Series X ray Generator complies with the requirements of applicable EMC standards Portable and mobile RF communications equipment can affect medical electrical equipment It is therefore recommended that the operation o
11. HHH HHH HHH HHH HHH HHH HHH HHH HFEF HHH HHH HHH HHH HFEF HHH HHH HHH HEHE HHEH HHH HHH EHH HHH HEHEHE ZZ HEHEHE HHH HHH HHH ZZ 3 S SC HF Series X ray Generators Operator s Manual Revision Q a Quantum Medical Imaging LLC 6 3 Fill in and mail Warranty Card promptly to Quantum Medical Imaging LLC 2002 B Orville Drive North Ronkonkoma NY 11779 7661 USA Any component furnished without charge to Buyer Dealer during the warranty period to correct a warranty failure shall be warranted only to the extent of the unexpired term of the warranty of the original product This warranty extends only to the origi nal purchase and is not transferable unless authorized in writing by Quantum Medical Imaging LLC Products manufactured by parties other than QMI where QMI acts solely as distribu tor or reseller will carry their respective manufacturers warranties including each of their independent terms and conditions Warranty consideration will be given only for defective QMI products properly returned to the factory in accordance with QMI s Returned Materials Procedure refer to Dealer Price Book or contact QMI customer service WARRANTY EXCLUSIONS The foregoing warranties are exclusive and in lieu of all other warranties whether written oral express implied or statutory NO IMPLIED WARRANTY OF MERCHANT ABILITY OR FITNESS FOR PARTICULAR PURPOSE SHALL APPLY Quantum Medical Imaging LLC OMI Warranty is exclusive of 1
12. HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH EHF HEF HHH HHH HHH HHH HHH HHH HHH HHH HHH HEHEHE HEHEHE ZZ ZZ ZZ ZZ ZZ ZZ ZZ HEHEHE HEHEHE HHH HHH HHH SD e BR HF Series X ray Generators Operator s Manual Revision Q a e Quantum Medical Imaging LLC 4 5 Editing a Region Name In the APR Edit Screen select a region to edit by pressing the corre sponding field select key For example to edit the SKULL region see Figure 4 5 press key adja cent to the SKULL field The APR Region Edit screen appears Change View SELECT THIS Settings KEY TO EDIT d APR VIEW Name Ee SELECT aai EDIT SELECT THIS REGION ACTION KEY TO EDIT APR REGION NAME BEING Previous EDITED Screen SKULL Figure 4 3 APR Region Edit Screen 1 To edit the name of the view SKULL in the example shown in Fig ure 4 6 press the key adjacent to the Region Name field if you do not want to edit the region name skip to step 4 The APR Region or View Name Edit screen appears PRESS THIS KEY TO RESTORE THE NAME TO FACTORY USE THESE KEYS DEFAULT TO SELECT VALUE CHARACTER eng USE THESE INSERT KEYS TO SELECTED SELECT CHARACTER CHARACTER POSITION USE THIS KEY TO DELETE SELECTED CHARACTER CENTER DISPLAY BOX SHOWS CURRENT REGION OR VIEW NAME Figure 4 4 APR Region or View Name Edit Screen SHH HHH HHH HHH HHH HHHHHEH HHH HHH HHH
13. II EXPOSUFES geg g eegente 3 25 Instantaneous EXPoSUres rrnrrrnennnnennnenen annan 3 25 HEAT UNIT COMPUTER nirrernenenennnnnnennnnnn na enannna nae 3 25 FAULT MESSAGES isi sects cine 3 26 HELP SCREEN FEATURES une ans 3 31 CHAPTER 4 APR EDITOR OVERVIEW Aert sen iii re 4 3 ACCESSING THE APR EDIT MODE nennen 4 3 APR EDIT MODE nen 4 6 Editing a View Name u u us een 4 6 Editing Programmable Exposure Settings serrer 4 10 Making a Back Up Copy of the APR Program File 4 20 Restoring APR Program from Back Up Copy 1 4 21 CHAPTER 5 USER MAINTENANCE OVERVIEW ed giess egenen SE Ee Ee 5 3 USER MAINTENANCE una Bst 5 3 SYSTEM SETTINGS 42 0 een 5 4 ACCESSING SERVICE MODE mmmmininnnenenennamanaanannananna 5 4 CHAPTER 6 WARRANTY I NFORMATION WARRANTY STATEMENT sssessnneneenseneenneenneennennennnnnnn nenn 6 3 WARRANTY EXCLUSIONS nennen nn 6 4 BUYER S REMEDIES 54 isistsinisescedavedcctatedoathoda cavaiedassaleamidackunacendtecguveia sis 6 5 WARRANTY RETURN PROCEDURE cescceceeeceeeeeeeeeeeeeeeeeteneesneseanes 6 5 EQUIPMENT IN TRANSIT eeren 6 5 VOIDING WARRANTY usage 6 6 Revision Q HF Series X ray Generators Operator s Manual iv Quantum Medical Imaging LLC Chapter SAFETY NOTICES SHSHSHSHSSHSHSHSHSHSHSHHSHSHHSHSHHSHHSHSHHSHHSHHSHHHHHHHHHHHHHHHOHHHHHHOOH OSH 1 1 GENERAL SAFETY INFORMATION Quantum produc
14. Overcharged Calibration ERROR Generator kV MA and or fila Call Technical Service Support ment calibration settings are outside of acceptable limits for normal system operation Cathode Volt ERROR The filtered cathode volts are Call Technical Service Support outside of preset limits at any time Collimator ERROR Automatic collimator not Call Technical Service Support responding only appears on systems using an automatic collimator Door WARNING Entrance door to x ray room is Verify x ray room door is closed and open re take exposure Early Term WARNING Exposure terminated prema Press any key on Operator Control turely because EXPOSE key Panel to clear warning message Re was released before exposure take exposure was completed Expose WDT ERROR Exposure sequence disabled Call Technical Service Support due to loss of communication with EXPOSE watchdog timer Filament Curr ERROR Indicates a problem with the Call Technical Service Support filament control regulator cir cuit voltage SHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HEHEHE FHF HHH HFEF HHH HHH HHH HHH HHH HHH HHH HHH HHH HEHEHE HEH HHH HHH HHH HHH HHH SD 2 V SC HF Series X ray Generators Operator s Manual Revision Q b Quantum Medical Imaging LLC 3 23 Table 3 3 Fault Messages Filament Prog ERROR The filtered Filament Amps are Call Technical Service Support ou
15. Press any key on the operator control panel to display a message describing the cause of the fault When the system will not initiate an exposure an audible alert will sound and a fault message will appear on the Operator Control Panel s LCD screen SHH SHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH FHF HHH HHH HHH HHH HHH HHH HHH HHH HEHEHE HHH HEHEHE HEHEHE HEHEHE HHH HHH HHH HHH HHH SD S K SR HF Series X ray Generators Operator s Manual Revision Q b Quantum Medical Imaging LLC 3 21 Refer to the following descriptions for assistance in interpreting the fault message Table 3 3 Fault Messages AEC No ERROR AEC mode has been selected 1 Verify that WALL or TABLE Bucky without selecting a receptor receptor is selected at OCP WALL or TABLE or no ion 2 Verify that at least one ion cham chamber detectors selected ber detector is selected AEC Not ERROR AEC mode was selected on Do not select AEC mode Note If Installed system not configured with this message appears on a system AEC equipped with AEC option AEC must be enabled in Service mode contact service Anode WARNING Indicates that the tube maxi Reduce technique factor s and or Heat or mum heat load capacity will be wait until tube cools sufficiently Re Housing exceeded by next exposure take exposure Heat READY lamp flashing Anode Volt ERROR Indicates that there is a prob Call Technical Ser
16. ZZ ZZ FHF HHH HHH HHH HHH HHH HHH HEHEHE HHH HHH HHH HEHEHE HHH HHH HHH HHH HHH SD 2 V SC HF Series X ray Generators Operator s Manual Revision Q b Quantum Medical Imaging LLC 1 13 ABBREVIATION DEFINITION The following abbreviations and acronyms may be found in this document Their def inition is explained below ADC AEC APR cm HF HSS kVp LCD mA mAs MHz OCP RF SE sec Revision Q 1 14 Analog to digital converter Automatic Exposure Control Anatomical Programmed Region Centimeters thickness High Frequency High speed starter Tube voltage kilovolts peak Liquid crystal display meters Tube current milliampere Time current product MegaHertz Operator Control Panel Radiofrequency Stored Energy Expose on time seconds Watts HF Series X ray Generators Operator s Manual Quantum Medical Imaging LLC Chapter 2 GENERAL INFORMATION LE 2 2 OVERVIEW NOTE The user should read this manual in its e entirety prior to using this equipment It should be kept in a location near the equipment and be readily accessible to those who operate it This document is intended to assist users in the safe and effective operation of the equipment described herein Pay special attention to all the information described in the Safety section refer to Chapter 1 SAFETY NOTICES This manual is written for trained users of the ODYSSEY High Frequency HF Series X ray Generator hereinafter re
17. any other occurrence that takes the tomography system out of tomographic mode such as selecting a receptor that is not enabled for Tomo Figure 3 7 when tomography mode was de selected the Manual mode screen appears see Figure 3 3 Conversely if the generator was in APR Tomo mode shown in Figure 3 8 when tomography mode was de selected the APR Man ual mode screen appears see Figure 3 4 The time for the last sweep selected will remain dis played on the screen d NOTE If the generator was in Manual Tomo mode SHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HEHEHE HEHEHE HEHE HHH ZZ ZZ ZZ HHH HEHHEHEHEH HEH ZZ EZ ZZ HOOD A Revisi HF Series X ray Generators Operator s Manual evision Q h e Quantum Medical Imaging LLC 3 19 TAKING AN EXPOSURE WARNING Ionizing radiation can lead to radiation PA injuries if handled incorrectiy When radiation is applied the required protective measures must be complied with These procedures are identical for all operating modes After verifying or adjusting techniques and exposure factors ensure that all other X ray exposure precautions are met RECEPTOR AEC FIELD 1 WALL NONE TABLE AUXILLIARY PREP EXPOSE 1 23 E K E amp N e KOED Pie vay unt ny ta go nat an ere 7 Ae ina C no errors are present system is SET LED Indicates self checks are complete prepared for exposu
18. as an example the CM STEP value indicated is 4 Therefore if the minimum or Base CM value is set to 10 CM this con figuration establishes the following series of patient size thickness set points at which the exposure values can be programmed e 10 CM e 14 CM e 18 CM e 22 CM e 26 CM CM STEP VALUE 4 30 CM 46 CM 10 CM 36 CM RANGE e 34 CM e 38 CM e 42 CM e 46 CM As shown in the example above starting at 10 CM there are nine 4 CM steps The CM STEP value and number of steps 9 determine the total thickness CM range for the procedure 9 x 4 36 HF Series X ray Generators Operator s Manual Quantum Medical Imaging LLC NOTE The CM Step value you set in the a APR Edit procedure is only used to estab lish the number of CM steps between pro grammed set points When the system is in actual APR mode of operation pres sing the CM increase decrease keys will change the CM value in increments of one The kVp mA and mAs or time values between each of the ten thickness set points will have interpolated values based on the lower set points exposure settings For example suppose at 10 CM s the kVp value is set to 50 the CM increment value is set to 4 and the next 14 CM thickness set point s kVp value is set to 54 When the operator chooses this view and selects a thickness of 11 CM s the system automatically adjusts the kVp for 51 and at 12 CM s it will be 52 and so forth up to
19. buyer are conditioned upon all claims under this warranty being made in accordance with the aforementioned warranty statement WARRANTY RETURN PROCEDURE A fully completed Field Returned Material Evaluation Form must be returned with any defective product or any returned item All returns must have the Serial Number of the Equipment and or the Specific Part written on the Field Returned Material Evalu ation Form All freight charges resulting from Warranty Returns are the responsibility of the Buyer or Dealer EQUIPMENT IN TRANSIT QMI cannot assume responsibility for any equipment damaged in transit To protect the buyer dealer the receiver of any equipment should examine all cartons and crates carefully at time of delivery If damage is apparent make a notation on the delivery receipt request an inspection by the freight carrier and if applicable file appropriate carrier claim Should concealed damage be detected immediately notify the freight carrier and request an inspection The purchaser dealer customer is fully responsi ble for the filing of freight damage claims to the freight carrier Quantum Medical Imaging LLC is not responsible for any loss or damage to products once they have been shipped from our factory The dealer or customer is responsible for full payment to Quantum Medical Imaging LLC for all invoices as per our stan dard payment terms regardless of freight damage or processing of an insurance claim by the dea
20. con trol panel Ensure that the appropriate RECEPTOR i e WALL 40 WALL 72 or TABLE 40 is selected prior to AEC FIELD selection A configuration of three boxes representing AEC detectors is depicted on the LCD A shaded box indicates an enabled AEC detector To select a different AEC detector press the corresponding AEC FIELD select key until the desired detector s is enabled There are seven possible combinations of AEC detector configurations OOOH OOH HOO OOD FOOTE HOH HHH HEHEHE HHH EEE HEH HEH HHH EH HEH EHH HHH HHH HHH HHH HOH HHH HHH HHH HHH HHH HHH HOH HEH HHH HHH HHH HHH HHHHEH HHH OOH HOHE Revision Q HF Series X ray Generators Operator s Manual 3 16 Quantum Medical Imaging LLC OPERATION WITH TOMOGRAPHY SYSTEM WHEN INTERFACED The HF Series X ray Generator has been designed for seamless operation with either Siemens MULTIX TOP Tomographic System or the Pausch CS 2000 Micro tome Tomography System The tomographic exposure settings are set and con trolled using the tomography system s controls with the generator providing a display of angle time setting The kV mA etc are made using the generator s operator control panel in the same manner as for any radiographic procedure NOTE On Siemens Tomography Systems If 30 0 8s a is selected on the Siemens system the generator display will indicate 8 0 8s Disregard the 8 angle indication it is not the value used to determine sweep angle How ever the
21. fields are toggle fields i e there are only two choices such as Yes No Other fields may contain multiple selections the choices are seguenced scrolled each time the button is pressed 6 N A kVp increment Control Push Adjusts the tube voltage kVp value decrement button from minimum of 40 kVp to maximum of 125 kVp in 1 kVp increments 150 kVp maximum available in systems equipped with 150 kVp option 7 N A mA increment Control Push Sets the tube current mA value mA decrement button stations that are available for selec tion are model dependent 8 N A mAs Time incre Control Adjusts the exposure time or current ment decrement Push button time mAs value SHH SHH HHH HHH HHH FHF HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH FHF HHH HHH HHH HHH HHH HHH HEHE HHH HEHEHE HHH HEHEHE HHH HHH HHH HHH HHH SD 2 V SE HF Series X ray Generators Operator s Manual Revision Q b Quantum Medical Imaging LLC 2 15 9 E Selects wall Control Used for examinations where wall receptor push button receptor and AEC are selected AEC feature is optional patient is posi tioned in front of wall receptor Press ing this button will toggle between WALL 40 and WALL 72 settings 10 o Disables wall and Control Push Used for examinations where neither table receptors button table nor wall image receptor is re
22. 001 6 3 Time Accuracy 1 mS 0 5 1 mS 0 5 1 mS 0 5 1 mS 0 5 1 mS 0 5 wimimi Papa 0 001 0 001 0 001 0 001 0 001 sure Time sec mAs Range 0 025 600 0 025 600 0 025 800 0 025 800 0 025 800 High Speed e Starter No Option Option Yes Yes Ripple Voltage 5 5 5 5 5 output Time measured at 75 of the peak kVp waveform For exposure times from 1 0 mS to 49 mS time accuracy is 2 1 mS for exposure times from 50 mS to 100 MS time accuracy is 1 1 mS NOTE AEC TECHNIQUES SHOULD HAVE EXPOSURE TIMES EXCEEDING 8 MS mAs is tube dependent the generator may not reach maximum mAs due to tube type In AEC mode the mAs is limited to 600 one minute period High Speed Starter Duty Cycle are not to exceed two activations within any SHH SHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HFEF HHH HHH HEF HHH HHH HHH HHH EHH HEHEHE HHH HHH HHH HEHEHE HHH HOOD 4949494444944464444 H 7 HF Series X ray Generators Operator s Manual S e Quantum Medical Imaging LLC Revision Q 2 7 ELECTRIC OUTPUT DATA Nominal X ray tube voltage Table 2 4 Electric Output Data N A N A and highest X ray tube cur rent obtainable at that volt age QG 32 QG 32 2 QG 32 3 QG 32 5 QG 3200 QG 3200 2 QG 3200 3 QG 3200 5 125 kV 250 mA 150 kV 200 mA with QG 150 option QG 40 QG 40 2 QG 40 3 QG 40 5 QG 4000 QG 4000 2 QG 4000 3 QG 4000 5 125 kV
23. 5 mAs 25 mA 1 0 mS Revision Q HF Series X ray Generators Operator s Manual 2 10 Quantum Medical Imaging LLC MAIN COMPONENTS See Figure 2 1 The HF Series X ray Generator is comprised of 1 Generator Cabinet Contains the electronics for high voltage generation and control 2 Operator Console Contains the operator control panel which has all generator operator controls and indicators all aspects of X ray techniques are entered at the operator control console The operator console may be mounted on a shelf QG WM option table top or ona 3 pedestal QG PDL option as shown in Figure 2 1 d aan ODYSSEY IF xis a i Figure 2 1 HF Series X ray Generator Main Components HF Series X ray Generators Operator s Manual Revision Q Quantum Medical Imaging LLC CONTROL PANEL SERIAL LABEL Model OGV 80 Serial No fourm oer IS KKT eis GENERATOR REGULATORY MARKINGS LABEL REFER TO CHAPTER 1 FOR MEANINGS OF SYMBOLS p X Au A EDUGA ELBE GA aam menen Eat EX ce GENERATOR SYSTEM LABEL GENERATOR CABINET LABEL Complies with FDA radiat Jards set forth in CFR 21 Subchapter Te Ee Zu 380 480 vc 100 A 3PHASE Model QG 80G Serial No QG80G XXX XXXX Manufactured Month Year MAX MA 800 MAX KVP 150 AMP MOM 200 0440 VAC AMPSLT lt 10 Manufactured by Q
24. 54 h Select the appropriate film screen combination to be used for this procedure Ensure the selected film screen combination has been previously calibrated Pressing the key adjacent to the Film Speed Combination field see Figure 4 10 item g sequences through the available film screen combination choices The factory default choices are WALL 40 WALL 72 TABLE 40 100 Speed 200 Speed 800 Speed and Film Screen7 If the Digital Receptor setting in Service Con figuration Menu is enabled an eighth option entitled Digital is also provided Note that selection of film screen combina tion is the only method to select a film speed for APR proce dures Only select film screens that have been calibrated i The FILM SPEED field see Figure 4 10 item h displays the base film speed setting The base film speed is the speed used to formulate the APR procedure j Select the appropriate kVp mA and mAs or time values for Procedure Step 1 see Figure 4 10 item i These values are the settings that will appear automatically when the operator accesses this APR anatomical view at minimum thickness selection Note During normal APR operation any pre pro grammed kVp mA or mAs time value can be manually modi fied i e overridden by the operator without returning to APR edit mode SHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HEHEHE HHH HHH HHH EHH HEHE HH
25. ANUAL AEC APR MANUAL AEC APR PRESSED WALL WALL WALL40 WALL40 WALL WALL40 WALL40 WALL72 WALL72 WALL72 WALL72 NONE NONE NONE NONE NONE NONE NONE TABLE TABLE TABLE40 TABLE40 TABLE TABLE40 TABLE40 AUX NOT NOT NOT WALL2 WALL2 WALL2 ENABLED ENABLED ENABLED AUX AUX AUX Table 3 2 shows the various receptor selection options The receptors shown with an asterisk are selected by toggling the indicated receptor key eg there is a choice of two SID s for the wall receptor WALL40 or WALL72 when in AEC or APR mode SHH SHH HHH HHH HHH HHH HHH HHH ZZ HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH ZZ ZZ ZZ ZZ HHH HHH HHH HHH HEHEHE HHH HHH HEHE HHH HEH HOD OHHH HHH HHO D 2 k oi HF Series X ray Generators Operator s Manual Revision Q a b Quantum Medical Imaging LLC 3 7 MODE SELECTION The generator has three operational modes Manual AEC optional eAnatomical Programmed Regions APR NOTE AEC mode can be used either as a stand alone mode or in conjunction with APR mode READY CN WALL d as AUXILLIARY ie PREP WEISSE o wel EE MODE This key enables the Anatomical MODE Programmed Regions mode This mode assure proper exposure times SELECTION provides pre programmed technique settings for up KEYS to ten 10 available anatomical regions each con taining up to ten programs for the selected region E the optional Automatic Exposure Control e AEC aes f
26. E We O 0 PS uue SETTING TOMO Figure 3 8 APR Tomo Mode Screen The tomo sweep time appears in the APR Tomo Mode screen depending upon whether TIME or MAS display mode is selected If the sweep time is changed by selecting a different angle time key on the tomogra phy system control panel and the generator is in tomography mode the exposure time updates instantly and the generator is properly set up for the changed selection but the angle and time displayed in the circle will not automatically update and remains with the original sweep time selec tion This indication can be updated by a exiting out and then back in to tomo mode or b selecting a different receptor or NONE then re select ing the correct tomo receptor the generator OCP after the tomography system sets the time the sweep time will not match what the tomography system expects and therefore the exposure will not be correct d NOTE If a change in the time setting is made on PREP key on the operator control panel to reset d NOTE On Pausch tomography systems press the tomo back to its initial position Returning to Radiographic Mode The OCP display will automatically switch from tomography mode to radi ography mode when e the radiographic mode select button on the tomography system s control panel is pressed Revision Q HF Series X ray Generators Operator s Manual 3 18 Quantum Medical Imaging LLC e
27. EMITIES EXTREMITIES PEDIATRICS SPECIAL Lele CHEST arrmove SKULL AP PA SKULL LATERAL THORAX CERVICAL TOWNES WATERS SELECT ABDOMEN E LUMBAR CALDWELL THE VIEW LR UPPER LOWER MANDIBLE TMJ EXTREMITIES EXTREMITIES PEDIATRICS SPECIAL mas AP PA SKULL LATERAL mm TOWNES WATERS SELECT d N CALDWELL THE VIEW En Pal MANDIBLE 70 150S 00 0 r 16CM 5 MA ES 0 APR TECHNIQUE SCREEN WITH AEC ENABLED BEER LLULLU APR PATIENT TYPE MODE SCREEN PRESS THE AEC MODE KEY TO TOGGLE WITHOUT AEC BETWEEN AEC ENABLED DISABLED Figure 3 5 Accessing APR Technique Screen THEO HOE HHH HEF HHH FEE HHH HOH HOH EE HEE EE EE EE HOH HHH EE HHH III EEE III III III III III III III EE EE REEFF EEE Revision Q HF Series X ray Generators Operator s Manual 3 14 Quantum Medical Imaging LLC Ap P SKULL LATERAL KVP MA MAS PATIENT 70 1508 00 0 THICKNESS 6 __ ES ER UP DOWN ARROW INDICATES VALUE WAS INCREASED DECREASED FROM ORIGINAL APR VALUE Ap P SKULL o LATERAL AEC PATIENT KVP MA MAS DENSITY TYPE l 70 1508 00 0 KEYS KEYS l em e lt FILM I Ei SCREEN 1 SELECT i KEY U _ EEE EEE M Ap 25 0 SKULL AEC FIELD A LATERAL Qa o KVP MA MAS 70 150S 00 0 Figure 3 6 Making APR Mode Exposure Selections 6 AEC enabled exposures only Verify AEC film density or Patient Type non AEC exposures only film speed anatomical thick nes
28. ENANCE The system user is responsible for the basic cleanliness of the eguipment Ona regular basis at least every six months painted metal surfaces should be cleaned using a clean cloth slightly moistened in warm soapy water use mild soap Wipe the operator control panel with a clean wet cloth then dry Never use strong cleaners solvents or abrasive polish on this eguipment WARNING Always disconnect the equipment A from the main power supply prior to any cleaning WARNING Electric shock hazard The HF Series A X ray Generator contains no user serviceable components Do not attempt to disable these components or remove any trim covers Refer service to gualified service personnel SHH SHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HEHEHE HHH HHH HHH HHH HHH HHH HEHEHE HHH ZZ HHH HHH HHH HHH HHH SD S K SC HF Series X ray Generators Operator s Manual Revision Q b Quantum Medical Imaging LLC 5 3 SYSTEM SETTINGS There are several operational parameters that can be set by the user These set tings include setting the date and time date time format language AEC back up time MAS generator automatic shut off delay time setting and default film screen designations for AEC mode The following paragraphs describe access to and selection of the various system settings ACCESSING SERVICE MODE The system settings screens are accessed by entering the X ray generator s Service
29. Failure of Buyer Dealer to prepare the site or provide power requirements or operating environmental conditions in compliance with any applicable instructions or recommendations of OMI 2 Failure of Buyer Dealer to provide the proper incoming power required to support the equipment in accordance with the recommendation of QMI 3 Any modification of product performed by a party other than QMI 4 Combining products deemed by QMI to be incompatible 5 Improper or extraordinary use of the Product improper maintenance of the Prod uct or failure to comply with any applicable instructions or recommendations of Quantum Medical Imaging LLC 6 Misuse tampering or negligent storage handling of the Product by Buyer its employees agents or contractors 7 Fuses glassware high voltage cables and other items deemed by QMI to be expendable 8 Acts of God acts of civil or military authority fires floods power failure or electri cal power surges strikes or other labor disturbances war riots or other causes beyond the reasonable control of Quantum Medical Imaging LLC 9 Installation troubleshooting or repair service are not included in this warranty Technical service and maintenance is the responsibility of the dealer selling the equipment 10 The Manufacturer is relieved of any responsibility for damage during shipment after the freight carrier picks up and begins transport of the unit for delivery SOHO HOE HHH HHH HHH EE HOH HOH EE
30. H HHO SD 44944444 4444444 S V SC HF Series X ray Generators Operator s Manual Revision Q b Quantum Medical Imaging LLC 4 13 k When all exposure factors for Procedure Step 1 are correct press the PATIENT SI ZE up key see Figure 4 10 item b to advance to the next procedure step i e Procedure Step 2 Modify exposure factors as required for this thickness set point and continue until all ten procedure steps have been configured m Using the PATIENT SIZE up down keys see Figure 4 10 item b select the procedure step as the default setting that will appear when the operator initially accesses this pro cedure using APR mode When the desired procedure step is selected press the Set Start Step key see Figure 4 10 item j to set this step as the default Note that the Set Start Step field should now match the Procedure Step number as a confirmation the setting was made n Press the Previous Save key see Figure 4 10 item k to save the modifications and return to the APR View Edit Screen see Figure 4 9 or press the Previous No Save key to ignore the changes and return to the APR View Edit screen 1 Access the APR View Edit screen see Figure 4 9 as described in the Editing a View Name procedure 2 Press the key adjacent to the Procedure Value 2 field The APR Procedure Value 2 screen is displayed SELECT EDIT ACTION Figure 4 8 APR Procedure Value 2 Screen a The setting in the MAS SEC fiel
31. HE HHH SD e s HF Series X ray Generators Operator s Manual Revision Q e Quantum Medical Imaging LLC 4 3 Figure 4 1 provides an overview for accessing the various APR edit mode screenss Each screen is described in detail in the paragraphs that follow ACCESSING THE APR EDIT MODE APR Edit mode is accessed through the Service Mode To gain access to the Ser vice Mode password entry is required This is to prevent unintended and or unauthorized changes to system and or APR screen settings To access the APR Edit mode proceed as follows 1 Set the ON OFF circuit breaker CB1 on the generator cabinet to ON position 2 Set the power on standby switch on the Operatorback of the E ZVET Control Panel OCP to the standby position then to the on position 3 The system begins the start up sequence the Welcome Screen is displayed and Testing Standby is flashing as shown in Figure 4 2 PRESS TO ACCESS PASSWORD MENU General Hospital Medical Center Quantu 1 Dealer Name EE XXX XXX XXXX Model QG 250 1 Version 1 00 Testing Stand by 4 Welcome ScreenWithin five seconds of start up press the upper left field select key as shown in Figure 4 3 The Password menu displays an eo 4 P I 4 I E b 4 I I I wee s MEE I I I E 4 FIELD SELECT KEYS Sb TT Figure 4 2 Password Menu THEO OHH HOE HHH HEF HHH EE H
32. HEHEHE HHEH HHH HHH HEHEHE HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HEHEHE HEH ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ HHH SD ae e Revision Q HF Series X ray Generators Operator s Manual k E 4 6 Quantum Medical Imaging LLC 2 The APR Region or View Name Edit screen allows you to edit the name of the selected Region or View Name as it appears on the APR screen or if no existing Region or View Name has been created it allows one to be created The text string being modified if one exists appears in the center display box as shown in Figure 4 7 Names of regions or Views are limited to 49 char acters The APR Region or View Name Edit screen provides three ways to edit text e Insert a new character e Delete an existing character e Change an existing character The following steps explain each method Inserting a New Character a Use the up and down keys next to the Sentence Position field to select the location to insert the new character in the text string the new character is always inserted to the left of the selected character pushing existing text to the right For example if the text string SKULL is displayed in the center display box choosing character position number 0 will place a character before the letter S posi tion 1 will place a character before the K and so on Note that the character in the selected position appears in the Character field b Use the up and down keys next to the Cha
33. HH HEHEHE HHH HHH HHH HEHEHE HHH HHH HHH HHH HOOD e bes HF Series X ray Generators Operator s Manual Revision Q e Quantum Medical Imaging LLC 3 11 1 Press the AEC mode button on the operator control panel The system displays the AEC MODE screen 2 Press MAS TIME key to choose between display of mAs or time value on AEC mode screen at end of exposure 3 Verify the technique selections are correct Modify technique set tings using the KV and MA increment decrement keys as required see Figure 3 6 4 Verify AEC film density and film screen combination selections are correct If necessary modify film density and or film screen combination settings using the increment up and decrement down keys adjacent to each field Available choices are e AEC Density Each step increases decreases mAs 15 above and below nominal N There are a total of eleven available steps 75 60 45 30 15 nominal 15 30 45 60 75 Film Screen choices are WALL 72 WALL 40 TABLE 40 100 Speed 200 Speed 800 Speed and FILM Screen7 If the Digital Receptor setting in Service Configuration Menu is enabled an eighth option entitled Canon Digital will be also be selectable Note These are the system default film screen combination names however each can be edited by user as necessary NOTE Ensure that the appropriate RECEPTOR amp i e WALL 40 WALL 72 or TABLE 40 is selected prior to AEC FIELD selectio
34. HHSHSHSSHHSHHSHSHSHSHHSHSHSHHHHSHSHHHHSHSHHHHHSHHHHHH OSH OVERVIEW This chapter is designed to assist the system user in maintaining the proper operation of the generator Also included in this chapter are instructions for making changes to the system s user configurable settings such as setting the date and time and lan guage and automatic power shut off settings This product has been factory tested to assure its required performance in an X ray system The user is responsible for performing routine maintenance and inspection procedures to ensure proper equipment operation Aside from routine maintenance any abnormal noise vibration or unusual performance should be investigated by a qualified service representative Preventive maintenance or any repair service should be performed only by qualified service personnel WARNING Failure to follow manufacturer s or service personnel s recommendations may result in serious injury on this equipment In this context qualified means those legally permitted to work on the equipment and authorized means those specifi cally authorized by local management d Only qualified and authorized persons shall work Changes additions or maintenance to the equip AN ment carried out by persons without appropriate gualifications and training and or using un approved spare parts may lead to serious risk of injury and damage to the eguipment as well as making the warranty void USER MAINT
35. HSHSHHHHHSHHHHHH OS WARRANTY STATEMENT Quantum Medical Imaging LLC herein known as QMI warrants to buyer that any new product manufactured by QMI will be free from defects in material and manufac turing and conform substantially to applicable specifications in effect on the date of shipment when subjected to normal proper and intended usage by properly trained personnel QMI will act as the sole judge in determining whether equipment or part is defective by reason of manufacture All QMI products shall be warranted for a period of 12 months from the original instal lation the date of which will be determined by a completed returned warranty card which must be returned to OMI headquarters within 30 days of system installation In no case shall the warranty exceed 15 months from the date of shipment If the warranty card is not returned to QMI then the warranty period will begin immediately on the date of shipment invoice date and last for twelve months Buyers should complete only one 1 form per system or component WARRANTY CARD Cut along dashed line Q Name of Owner Name of Facility Address 1 Address 2 City State Country Zip Phone e mail Name of Distributor Installation Date Check Type of Equipment and Provide ID No s Model No Serial No Hi Freq Generator Table Collimator Hi Tension Cable Tube Tube Stand Wall Stand Other OO Oo OD O0 0 00 SHH SHH HHH HHH HHH HHH HHH HHH HHH HHH HHH
36. IE EI DIE HF Series X ray Generators Operator s Manual Quantum Medical Imaging LLC Revision Q 4 8 Removing an Existing Character a Use the up and down keys next to the Sentence Position field to select the character to remove from the text string For example if the text string SKULL is displayed in the center display box choos ing character position number 0 selects the letter S for removal position 1 selects the letter K for removal and so on Note that the character in the selected position appears in the Character field Pressing the key next to the Delete Character field deletes the selected character Note that when the character is deleted any characters to the right of the deleted character will shift to the left one space thereby closing up the deleted character s space Replacing an Existing Character a Use the up and down keys next to the Sentence Position field to select the character to replace For example if the text string SKULL is displayed in the center display box choosing character position number 0 selects the letter S for replacement position 1 selects the letter K for replacement and so on Note that the character in the selected position appears in the Character field Use the up and down keys next to the Character field to scroll through all available alphanumeric characters beginning with a no character or blank space until the desired character appears in t
37. L 72 NONE or TABLE 40 See Figure 4 10 item d For AEC enabled exposures either the WALL or TABLE receptor must be selected SHH SHH HHH HHH HHH FHF HHH HHH ZZ HHH HHH HHH HHH HHH HHH HHH HHH HHH ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ HHH HHH ZZ ZZ ZZ HEH HHH ZZ HEHEHE HEH HHH OHHH HHH HHH SD 2 V SC HF Series X ray Generators Operator s Manual Revision Q b Quantum Medical Imaging LLC 4 11 Revision Q 4 12 6 f NOTE The AUX receptor is functional on systems equipped with the Universal AEC option it is usually reserved for Tomo receptor selection For AEC enabled exposures select the desired ion chamber field combination using the three AEC field keys on the oper ator control panel A configuration of three boxes repre senting AEC chamber detectors is depicted on the LCD see Figure 4 10 item e A solid box indicates that a field is selected for use IMPORTANT Do not change the CM STEP value setting if you are only reviewing procedure values or if you only want to change one or more of the procedure s exposure values Doing so will change the relationship between patient thick ness and corresponding exposure factors The CM STEP field Figure 4 10 item f sets the thickness step size value this value determines the size in centime ters CM between each of the ten patient size thickness set points for this procedure The displayed value is the cur rent thickness step size setting Using Figure 4 10
38. ME ege A 10 17 am 10 20 am Time Z n KVP MA MAS KP ma SS le i Jul 16 01 10 17 am we pm mmm mm e Figure 3 3 Making Manual Mode Exposure Settings 1 Press the MANUAL mode key on the operator control panel The system displays the MANUAL MODE screen 2 Press the MAS TIME key to toggle display between mAs and time settings 3 Verify exposure technique settings are correct Modify tech nique using the KV MA or MAS TIME increment decrement keys as required SOHO EHH HOE HHH HEF HHH EE HOH HOF HHH HEE EE EEE EE HOH RER HOH HHH HHH HHH HHH HOH HHH HHH HEH HHH HHH HHH HHH HEH HOH HHH HHH HHH ERR III HHH OOH HEHE Revision Q HF Series X ray Generators Operator s Manual 3 10 Quantum Medical Imaging LLC AEC Mode When operating in AEC mode the system automatically sets required mAs or time for achieving proper optical density with compensation for programmable film screen speed To make AEC exposure settings see Figure 3 4 and proceed as follows lt MAS TIME gt 70 150S 00 0 KEY El KVP MA MAS genen ki CD 70 150S 00 0 j elle SCH i CD EB N wE Ne _ D AEC DENSITY KVP MA MAS _ SELECTION KEYS 70 150S 00 0 FILM SCREEN ea ea fe SELECTION KEYS AEC FIELD ka lal AI Figure 3 4 Making AEC Mode Exposure Settings SHH SHH HHH HHH HHH FHF HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HFEF HHH HHH HHH HHH HHH H
39. Mode To gain access to the Service Mode password entry is required This is to prevent unintended and or unauthorized changes to sys tem settings To access the Service Menu proceed as follows 1 Verify the ON OFF circuit breaker CB1 on the generator cabinet is set to the ON position 2 Set the power on standby switch on the Control Panel OCP to the standby position then to the on position 3 The system begins the start up sequence the Welcome Screen is dis played and Testing Standby is flashing See Figure 5 1 PRESS TO ACCESS General Hospital PASSWORD MENU Medical Center Quantum Dealer Name Medical Imagine XXX XXX XXXX Model QG 250 1 Version 1 00 OO AE EAL EL Testing Stand by Figure 5 1 Welcome Screen THHHHHHHH HOE HEHE HEH HHH EE HOH HOFF HOE HHH HEE EE HOH HHH HHH HHH HHH HHH HHH HOH HHH HHH HEH HHH HHH HHH HHH HEH HHH HHH HHH HHH HHHHEH HHH OOH HOHE Revision Q HF Series X ray Generators Operator s Manual 5 4 Quantum Medical Imaging LLC 4 Within five seconds of start up press the upper left field select key as shown in Figure 5 1 5 The Password menu displays Use Ke To Enter FIELD SELECT KEYS 7 Figure 5 2 Password Menu 6 Inthe Password menu enter the default X Ray Technician password 3497 using the field select keys next to the corresponding numbers This password enables access to Service Mode for modification
40. OH HOH HHH HEE HOH EEE EE HOH HHH HOH HHH HHH HHH HHH HHH HHH HHH HHH EE EFFE HHH HHH HEH HEH HHH HHHHHH HHH HEH HHH EHH HOHE Revision Q 4 4 HF Series X ray Generators Operator s Manual Quantum Medical Imaging LLC 5 Inthe Password menu enter the default system password 3497 using the field select keys next to the corresponding numbers 6 Upon validation of the password the Service Menu displays Dec 13 01 2 23 pm SERVICE Configuration D Calibrati te APR EDIT kid FIELD SELECT KEY Figure 4 1 Service Menu 7 In the Service Menu press the key next to the APR Edit field The APR Edit mode screen appears see Figure 4 5 The system accommodates up to 10 user configurable preset region names as shown inthe sample screen below Each region can have up to 10 selectable views which appear upon selection of the region Each view contains pre programmable technigue factors that are configurable according to the facility s particular reguirements and or Bahai preferences CHEST SKULL REGION THORAX CERVICAL FIELD SELECT ABDOMEN kanton LUMBAR KEYS To Edit UPPER LOWER EXTREMITIES EXTREMITIES PEDIATRICS SPECIAL Figure 4 2 APR Edit Mode Screen Region APR EDIT MODE APR Edit mode provides editing utilities for adding modifying duplicating or deleting the name of a view and for modifying the programmable exposure set tings associated with a particular view SHH HHH HHH HHH
41. Revision Q HF Series X ray Generators Operator s Manual 1 6 Quantum Medical Imaging LLC ACCOMPANYING DOCUMENTATION The documentation consists of a User manual this document and related docu mentation Service Manual P N DC30 011 Contains technical and service documentation for this product including installation and configuration instructions to be performed by qualified persons The documentation shall be kept with the system for easy reference APPLICABLE STANDARDS HF Series X ray Generators comply with the following regulatory standards FDA Center for Devices and Radiological Health CDRH Title 21 CFR Subchapter J EN 60601 1 1990 A1 1993 A2 1995 A13 1996 IEC 60601 2 7 1998 E CAN CSA C22 2 No 601 1 M90 2005 Medical Electrical Equipment part 1 General Requirements for Safety UL 60601 1 1st Edition 2006 04 26 Medical Electrical Equipment part 1 General Requirements for Safety IEC 60601 1 Medical electrical equipment Part 1 General requirements for safety IEC 60601 1 2 2007 EC Directive 93 42 EEC for Medical Devices EU Authorized Representative Medizintechnik Berlin GmbH Altentreptower Strasse 59 12683 Berlin Germany Phone 49 302 82 4726 0473 Fax 49 302 82 6382 E mail Medizintechnik Berlin GmbH t online de STHHHHH HHH HHH HHH FHF HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HEHE HHH EHH HHH OD SHHHHHH HHH HHH HHH HHH HEHEHE HHH HEHEHE HHH HEH HEHEHE HEHEHE
42. S VIEWS EXPOSURE SETTINGS De PRESS THIS KEY TO EDIT FOR ADDITIONAL SELECT EXPOSURE EDITS THE VIEW EDIT NAME SEEN PRESS THIS KEY TO RESTORE ns PROCEDURE SETTINGS TO FACTORY DEFAULT VALUES Figure 4 6 APR View Edit Screen d Press the key adjacent to the View Name field The APR Region or View Name Edit screen appears see Figure 4 7 with the View name displayed in the center display box The method for editing the View name is the same as for editing a Region name as described previously in the Editing a Region Name procedure HF Series X ray Generators Operator s Manual Revision Q Quantum Medical Imaging LLC 4 9 e To modify programmed exposure settings for the selected view e g view SKULL A P P A as shown in Figure 4 9 select the Procedure Value key or Procedure Value 2 key for additional programmable exposure parameters Pro ceed with instructions in the next paragraph Editing Programmable Exposure Settings Each procedure view has programmable exposure factors settings To accommodate the operator in providing maximum system flexibility two separate screens are provided containing all of the configurable exposure settings for each programmable procedure The APR Procedure Value Screen see Figure 4 10 and the APR Procedure Value 2 Screen see Fig ure 4 11 APR Procedure Value Screen Includes the kV mA time or mAs anatom ical thickness range and initial thickness setting image receptor sele
43. al Revision Q Quantum Medical Imaging LLC 2 13 RECEPTOR EISE WALL NONE TABLE AUXILLIARY PREP OEXPOSE eo eo O Figure 2 5 OperatorControl Panel Controls and Switches Table 2 5 Operator Control Panel Controls and Indicators 1 S Selects Manual Control Enables manual selection of kVp mA R mode of operation Push button mAs and time settings for exposure 2 Selects Automatic Control Push System automatically sets required 45 Exposure Control button mAs for achieving proper optical den mode of opera sity with compensation for program tion Only func mable film screen speeds tional if AEC option is installed Revision Q HF Series X ray Generators Operator s Manual 2 14 Quantum Medical Imaging LLC 3 Selects Anatomi Control Push Enables technique selection from pre e cal Programmed button programmed anatomical regions and Regions APR radiographic projections mode of operation Note When in APR Mode pressing the AEC key on the Operator Control Panel toggles between APR AEC mode and APR Patient Type mode 4 N A LCD screen Indicator Displays various system screens menus and messages The LCD screen viewing angle is adjustable for maximum visibility 5 N A Selects the field Control The functions activated by the Display adjacent to the Push button Function buttons depend on the infor Display Function mation displayed on the LCD Some button pressed
44. and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the HF Series X ray Generator is used exceeds the applicable RF compliance level above the HF Series of X ray generators including TechVision option should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as re orienting or relocating the HF Series X ray Generator b Over the frequency range 150 kHz to 80 kHz field strengths should be less than 3 V m SOHO HOE HHH HEH HHH EE III EI III III III EI ZI HOH HHH III HHH HHH HHH HHH HOH HHH HHH HHH HHH HHH HHH HHH HEH HHH HHH HHH HHH ERR EEE HHH OOH HEHE Revision Q HF Series X ray Generators Operator s Manual 1 12 Quantum Medical Imaging LLC Recommended separation distances between portable and mobile RF communications equipment and the HF Series X ray generators including TechVision option
45. c tion AEC on off film screen combination film speed only functional with non AEC exposures and active ion chamber detector settings APR Procedure Value 2 Screen Includes the time mAs default AEC Den sity Proc procedure Enable On Off and backup mAs settings To edit a view s programmed exposure settings proceed as follows 1 Access the APR View Edit screen see Figure 4 9 as described in the Editing a View Name procedure 2 Press the key adjacent to the Procedure Value field The APR Procedure Value screen is displayed kire MA kv P MA MAS 73 1505 028 Procedure Step 1 Figure 4 7 APR Procedure Value Screen THEO EHH HOE HHH HEH HHH FEE HHH HOH HOHE HEE EE HEE HHH HOH HHH EE HHH HHH HHH HHH HOH III III III II III EE HHH EE EE DIDI III ERR EEE III Revision Q HF Series X ray Generators Operator s Manual 4 10 Quantum Medical Imaging LLC a Procedure Step 1 is displayed directly below the center dis play box Figure 4 10 item a The system provides 10 Pro cedure Steps or patient size thickness set points for a procedure each set point can be programmed with particular exposure factors i e KV mA time mAs When pro grammed appropriately the ten steps or set points allow you to vary the exposure factors to correspond with increasing patient size thickness settings across the range of select able size settings The Procedure Step 1 set point is used to set the minimum thicknes
46. cumentation Si Earth ground h Dangerous Voltage Attention consult accompanying amp Protective Earth ground AN documents Power ON Power OFF C part of equipment O part of equipment Stand By 6 Non ionizing radiation ey lonizing radiation Nj Alternating current TEA COMPATIBILITY The eguipment described in this manual must only be used in combination with other equipment or components if these are expressly recognized by Quantum Medical Imaging LLC as compatible INTENDED OPERATOR The HF Series X ray Generator is intended to be installed used and operated only in accordance with the safety procedures given within this manual for the purpose for which it was designed Before attempting to work with this equipment read understand note and strictly observe all warnings cautions and safety markings on the equipment Users include those persons who actually handle the equipment and those who have authority over the equipment TRAINING Users of HF Series X ray Generator shall have received adequate training on its safe and effective use before attempting to work with the equipment Training requirements may vary from country to country The User shall make sure that training is received in accordance with local laws or regulations that have the force of law SOHO HOE HHH HEH HHH EE HOH EI III III III III III III III III III III III III HOH HHH HHH HEH HHH EFFE HHH HHH HEH HOH EHH HHH HHH ERR EEE HHH OOH HEHE
47. d an eighth option entitled Digital will be also be selectable Note that the factory default film screen combination names may be changed using the film screen combination name edit utility as described in Service Manual SOHO EEH HOE HHH HEH HHH EE HOH HOH EE HEE HHH HEE EE HOH HHH EHH HHH HHH HHH HHH HOH HHH HHH HEH HHH HHH HHH HHH HEH EE HHH HHH HHH ERR EEE HHH OOH HEHE Revision Q HF Series X ray Generators Operator s Manual 5 10 Quantum Medical Imaging LLC The Film Screen Defaults screen shown below displays the current receptor film screen combination settings To change the default film screen settings proceed as follows W e F Ge Film Screen Previous 40 Defaults Screen Wall 72 FS Wall 72 H Table 40 FS Table 40 Wall2 FS 100 Speed USE THESE KEYS TO SET FILM SCREEN COMBINATION THESE KEYS ONLY USED ON UNIVERSAL TYPE AEC SYSTEMS Auxilliary FS 200 Speed Figure 5 7 Film Screen Defaults Screen NOTE The default film screen combination set A tings automatically link specific calibrated film screen combinations that will be used in non APR and APR exposures that use AEC However the operator may choose any available calibrated film screen combination using the Film Screen selection keys provided in the AEC and APR mode screens 1 Wall 40 FS setting Press the key next to the Wall 40 FS field until the desired film screen combination name appears The selected film screen co
48. d incorrect checksum OCP Micro ERROR Operator Control Panel micro Call Technical Service Support processor Al6U3 returned incorrect checksum Phase Missing ERROR Hardware has detected that at Call Technical Service Support least one phase of a three phase generator is not present Please Wait WARNING Operator attempted to take Allow more time for system to reach exposure but system not yet its ready state before taking next ready exposure Note The system will not allow more than two high speed tube starts within a one minute time period SOHO HEHE HEHEHE HHH FEE HEHEHE HOE FEOF EEE HEHEHE HHH EEE HEHEHE HEE HHH FEE HHH EEE HEHEHE HEF EHH EEE HHH EHH HHH EEE RRE EE EEE EHH EEE HHH EEE Eat Revision Q HF Series X ray Generators Operator s Manual 3 26 Quantum Medical Imaging LLC Table 3 3 Fault Messages Plus 15V ERROR Missing or incorrect 15 VDC Call Technical Service Support supply voltage Plus 24V ERROR Missing or incorrect 24 VDC Call Technical Service Support supply voltage Plus 5V ERROR Missing or incorrect 5 VDC Call Technical Service Support supply voltage Prep WDT ERROR Exposure seguence disabled Call Technical Service Support due to loss of communication with PREP watchdog timer Receptor Not ERROR The signal on the Receptor Call Technical Service Support Ready Ready terminal of the Digital Interface Board A9 AY40 034T is not ready within 500 ms fr
49. d see Figure 4 9 item a determines the unit of measure of the exposure duration in terms of either time SEC or mA seconds mAs Pressing the key next to this field toggles between these two settings SHEE HOE HHH HEH HHH EE HOH HOFF HEE EE HEE EE HOH HHH HOH HHH HHH HHH HHH HOH HHH HHH HHH EE EFFE HHH HHH HEH HHH HHH HHH HHH ERR III HHH OOH HEHE Revision Q HF Series X ray Generators Operator s Manual 4 14 Quantum Medical Imaging LLC b The AEC DENSITY field see Figure 4 10 item b sets the default AEC Density for the procedure Select the AEC Den sity setting using the increment up and decrement down keys adjacent to the AEC DENSITY field Each step of this setting adjusts the optical density 15 above and below nominal N for a total of eleven available settings 75 60 45 30 15 nominal 15 30 45 60 75 c The Proc Enable field see Figure 4 10 item c setting enables or disables the selection of this particular pro grammed procedure view in the APR View Selection Screen see Figure 4 8 In the example shown in Figure 4 8 the system s default programmed procedures such as P A GRID are selectable since the Proc Enable setting for each of these views has been set to On at the factory Note however the blank fields that appear in the APR View Selection screen see Figure 4 8 are views that have been disabled because their Proc Enable setting is set to Off and the view names have been ed
50. d with Universal type AEC Te AEC Board A11 AY40 0275 is installed Press the key next to the Auxiliary FS field until the desired film screen combination name appears The selected film screen combination is now the film screen combination used when AUX image receptor is selected for exposures using AEC and APR AEC mode exposures Language Setting The Service Settings Menu shown below displays the current language set ting The default language setting is English To change the language setting proceed as follows PRESS Set Date SERVICE THIS KEY Apr 15 05 SETTINGS TO SELECT Set Time Language LANGUAGE 2 49 pm English gt 4 Film Screen Back Up Defaults Time MAS 4 gt APR Format Generator All Timeout Windows 60 A Date Format US Figure 5 8 Choosing Language and Date Format Settings Press the key next to the Language field to scroll through available lan guages English French Spanish German Italian etc until desired lan guage is displayed AEC Backup Setting The AEC Backup Setting screen shown below displays the current AEC backup mAs setting Note thatthis is the system wide backup mAs setting and as such will be effective on all AEC exposures The procedure specific back up time is set using the APR Edit function refer to APR Edit chapter in the Operator Manual for instructions on setting individual procedure backup mAs Note The lower of the two backup settings will take pr
51. dow at the bottom center of the operator control panel display screen SS AEC FIELD m b j WALL G az Wf S OPREP EXPOSE 0000 WK vi MESSID i ry tray se argos pit eed apr reste ptt fr nd pr butions a ott WALL ID Selects the Wall Stand Film Cabinet as the image receptor IMAGE NONE f Selects the cassette i e neither Table nor RECEPTOR D Wall Stand Film Cabinet as the image SELECTION receptor Use with Special exams such as KEYS cassette table top or cross table exams TABLE N Selects the Table Film Cabinet as the image receptor AUXILLARY f Selects additional image receptor s such as a second wall stand or separate digital AON receptor etc Press once select the receptor assigned as WALL2 press second time to select receptor assigned as AUXILLIARY Figure 3 1 Receptor Selection Revision Q HF Series X ray Generators Operator s Manual 3 6 Quantum Medical Imaging LLC The available receptor choices are dependent on the mode of operation Manual AEC or APR whether or not the system is equipped with AEC and if so the type of AEC standard or universal AEC Standard AEC provides for the selection of up to two separate receptors WALL and TABLE Universal AEC provides for the selection of up to four separate receptors WALL TABLE WALL2 and or AUX Table 3 2 Receptor Selection Options STANDARD AEC OPTION UNIVERSAL AEC OPTION RECEPTOR KEY M
52. e HF Series X ray Generator must only be used in rooms that comply with all applicable laws or regulations that have the force of law concerning electrical safety for this type of equipment Scheduled maintenance is essential to the assurance of continued integrity of this equipment with respect to regulatory compliance The continuance of certified perfor mance to the regulatory standard is incumbent upon the user s diligent conformance to recommended maintenance instructions Do not use this equipment until you are sure that the planned maintenance program is up to date CLASSIFICATION This product has been classified as Class by Underwriters Laboratories Inc Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or with nitrous oxide Protection against Harmful Ingress of Water Ordinary enclosed equipment without protection against ingress of liquids NSS Ry MEDICAL ELECTRICAL EOUIPMENT RY gt WITH RESPECT TO ELECTRIC SHOCK FIRE U MECHANICAL HAZARDS ONLY GAS L US IN ACCORDANCE WITH UL 60601 1 AND CAN CSA C22 2 NO 601 1 98UA SHH SHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HEHEHE HHH HHH HFEF HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HEHEHE HHH HHH HHH HHH SD S SC HF Series X ray Generators Operator s Manual Revision Q ka Quantum Medical Imaging LLC 1 5 The following symbols may be used for marking on this equipment or equipment do
53. ecedence when set to unequal values SOHO HHE HHH HEF HHH EE HOH HOH HHH HEE HHH EEE EE HOH HHH HHH HHH HHH HHH HHH HOH HHH HHH HHH HHH EFFE HHH HHH HEH HHH HHH HHHHHHHHHHEH HHH OOH HHH Revision Q 5 12 HF Series X ray Generators Operator s Manual Quantum Medical Imaging LLC To change the system wide backup mAs setting proceed as follows Previous AEC Backup Screen Setting ackup v t MAS Set gt E AEC Backup MAS KEYS TO SET BACK UP MAS Eu AR an Figure 5 9 Choosing a Back Up Time MAS Setting 1 Adjust the backup mAs setting using the appropriate keys as shown in Figure 5 9 above until desired back up mAs is displayed 2 To save the indicated back up mAs setting and return to the Service Set tings Menu press the key next to the Save field or the Previous Screen field Generator Timeout Setting The Service Settings Menu shown below displays the current automatic gen erator power shutdown setting The time period selected represents the amount of delay time after which if no user activity is detected the system automatically enters Automatic Power Stand By mode This mode is entered to help reduce power consumption while protecting system electronics To set the generator timeout setting proceed as follows Set Date SERVICE Apr 15 05 SETTINGS 4 Set Time Language 2 49 pm English PRESS THIS KEY TO SELECT GENERATOR TIMEOUT SETTING 4 gt Film Screen Back Up Defa
54. em Quantum gt ME IMAGING Division of Carestream ODYSSEY HF Series X ray Generators Operator s Manual This manual is copyrighted and all rights are reserved No portion of this document may be copied photocopied reproduced translated or reduced to any electronic medium or machine readable form without prior consent in writing from Quantum Medical Imaging LLC QMI Copyright 2010 QMI Quantum Medical I maging LLC Made in U S A 2002 B Orville Drive North Ronkonkoma NY 11779 7661 USA Phone 631 567 5800 Fax 631 567 5074 E mail info qmiteam com www quantummedical net REVISION DATE TYPE OF MODIFICATION A 7 13 01 Initial Release B 7 30 01 Incorporated UL comments added Model QG 50 C 9 25 01 Added Model AG 40 3 D 12 10 01 Added Model QG 65 incorporated ECO 0619 E 2 26 02 Incorporated ECO 0714 0742 F 10 9 02 Incorporated ECO 0787 0876 G 4 23 03 Incorporated ECO 0923 1028 H 7 11 03 Incorporated ECO 1256 J 7 31 03 Incorporated ECO 1269 1272 K 2 5 04 Incorporated ECO 1144 1359 L 5 4 05 Incorporated ECO 1607 M 5 12 08 Incorporated ECO 1985 N 3 2 09 Added protection against harmful ingress of water EU battery directive P 9 21 09 Incorporated ECO 2199 Q 9 13 10 Safety Notices enhancements i iv 1 1 to 1 14 2 1 to 2 18 3 1 to 3 28 4 1 to 4 16 5 1 to 5 14 6 1 to 6 6 0 0 0 0 0 0 0 error
55. f equipment of this type such as mobile telephones cordless microphones and other similar mobile radio equipment be restricted from the vicinity of this device Use of accessories transducers and cables other than those specified in the accom panying documents may result in increased emissions or decreased immunity of the equipment Guidance and manufacturer s declaration electromagnetic emissions in such an environment The HF Series X ray Generator is intended for use in the electromagnetic environment specified below The customer or the user of the HF Series X ray Generator should assure that it is used Emissions test Compliance Electromagnetic environment guidance RF emissions Group 1 The HF Series X ray Generator uses CISPR 11 RF energy only for their internal functions Therefore the RF emis sions are very low and are not likely to cause any interference in nearby electronic equipment SOHO HEHEHE HHH FEE HEHEHE HEHE HHH EEE HEHEHE HEHEHE SE EHH HEE HHH EEE HHH EHH SE EHH HEHEHE EEE HHH HEHE HEH HOH RRE EHH EEE EHH EEE HHH EEE Eat Revision Q 1 8 HF Series X ray Generators Operator s Manual Quantum Medical Imaging LLC Guidance and manufacturer s declaration electromagnetic emissions RF emissions Class A The HF Series X ray Generator is CISPR 11 suitable for use in all establishments i KE other than domestic and those Harmonic emissions Class A di
56. ferred to as the HF Series X ray Generator and for authorized field service personnel Quantum Medical Imaging LLC assumes no liabil ity for use of this document if any unauthorized changes to the content or format have been made The HF Series X ray Generator includes the following models Model QG 32 32 kW 208 260 VAC single phase input configuration Model QG 32 SE 32 kW STORED ENERGY 115 230 VAC single phase input configuration Model QG 32 2 32 kW 208 240 VAC three phase input configuration Model QG 32 3 32 kW 380 480 VAC three phase input configuration e Model QG 32 5 32 kW 380 480 VAC single phase input configuration e Model QG 40 40 kW 208 260 VAC single phase input configuration e Model QG 40 SE 40 kW STORED ENERGY 115 230 VAC single phase input configuration e Model QG 40 2 40 kW 208 240 VAC three phase input configuration e Model QG 40 3 40 kW 380 480 VAC three phase input configuration e Model QG 40 5 40 kW 380 480 VAC single phase input configuration Model QG 50 SE 50 kW STORED ENERGY 115 230 VAC single phase input configuration Model QG 50 3 50 kW 380 480 VAC three phase input configuration Model QG 50 2 50 kW 208 240 VAC three phase input configuration Model QG 65 65 kW 380 480 VAC three phase input configuration Model QG 80 80 kW 380 480 VAC three phase input configuration This product is intended to be used and operated only in accordance
57. he Character field window To replace the character in the position chosen in the previous step press the key next to the Replace Character field The new char acter will now appear in the selected position of the text string When done editing the Region or View name you may select Previous Save to save the new settings and return to the previous screen Region Edit or View Edit screen or select Previous No Save to ignore the changes and return to the previous screen Region Edit or View Edit screen To restore region or view name to the factory default settings press the key next to the Factory Reset field Editing a View Name Subordinate to each region there may exist up to 10 individual view names To change the name of a view or create a new view where there is currently a blank field proceed as follows a Access the APR Region Edit screen as described previously see Figure 4 6 b Inthe APR Region Edit screen press the key adjacent to the Change View Settings field The APR View Selection screen appears AP A PA SKULL LATERAL TOWNES WATERS SELECT SUBMENTAL THE VIEW VERTEX Figure 4 5 APR View Selection Screen c Select a view to edit by pressing the corresponding field select key For example to edit the A P P A view see Figure 4 8 press key adjacent to the A P P A field The APR View Edit screen appears PRESS THIS KEY TO EDIT THI
58. heavy etc or exposure parameters are adjusted by selecting cm or setting individual exposure factors NOTE On systems equipped with AEC option all a techniques may be toggled between AEC and non AEC modes using the AEC mode button AEC MODE Automatic Exposure Control optional uses an ion chamber to ensure proper optical density with compensation for programmable film screen speeds If AEC option is not installed AEC mode is inaccessible to operator e MANUAL MODE enables manual selection of kVp mA and mAs or time settings for exposures For detailed operating instructions for each of the above modes proceed to the required procedure MANUAL AEC or APR with or without AEC on the follow ing pages NOTE In the sample screens that follow dis a played values are for illustrative purposes only and do not necessarily reflect valid techniques SHH SHH HHH HHH HHH FHF HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HFEF HHH HHH HHH HHH HHH HHH HEHEHE HHH HHH HHH HEHEHE HEHEHE HHH HHH HHH HHH SD S K SE HF Series X ray Generators Operator s Manual Revision Q b Quantum Medical Imaging LLC 3 9 Manual Mode When operating in Manual mode all X ray techniques and exposure fac tors must be set by the operator To make manual exposure settings see Figure 3 3 and proceed as follows 1 16 01 MAS 17 am MANUAL MODE TIME Jul 10 AUER EEE PRESS TO DISPLAY EITHER MAS OR TI
59. ility Access limited to authorized service technician only Performs various system self tests Last Calibration Date Access imited to authorized service technician only Provides the date on which calibration was last performed Settings utility Provides access to system settings such as the language date time date format international or U S wall and table film screen designations generator automatic shut off time delay setting and system wide AEC backup time mAs setting APR Edit utility Used to edit existing APR screens or use to create new screens refer to Chapter 4 APR Editor for details HHH HOE HEHE HEF HHH Ett HOH HOH HHH HEE EE EEE EE HOH HHH HEH HHH HHH HHH HHH HHH HHH HHH HEH HHH EFFE HHH HHH HEH HHH HHH HHH HHH HEHEHE HHH OOH HEHE Revision Q 5 6 HF Series X ray Generators Operator s Manual Quantum Medical Imaging LLC SYSTEM SETTINGS GENERAL DESCRIPTIONS In the Service Menu see Figure 5 4 press the key next to the Settings field to access the Service Settings Menu shown below Set Date SERVICE Apr 15 05 SETTINGS 4 Set Time Language 2 49 pm English K d Film Screen Back Up Defaults Time MAS gt 4 APR Format Generator All Timeout Windows 60 d Date Format US BERGE Figure 5 4 Service Settings Menu The Service Settings Menu provides access to various system settings The fol lowing briefly describes the various Service Settings Menu selections e Set Date setti
60. ing between 10 and 600 mAs in 10 mAs increments e Generator Timeout Sets the delay time for generator to automatically shut down if no keys are pressed on operator control panel selectable in 30 minute increments between 0 minutes automatic shutdown disabled up to 180 minutes Date Format setting Sets date time format options are US or I nternational HHH HOE HHH HEH HHH EE HOH HOH HOH HEE EE EOE EE HOH HHH HEH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HEH HHH HHH HHH HHH HHH HEH HHH OOH HEHE Revision Q HF Series X ray Generators Operator s Manual 5 8 Quantum Medical Imaging LLC SYSTEM SETTINGS DETAILED DESCRIPTIONS To change a setting press the key next to the system setting you want to change After all system settings are set appropriately press the key next to the Main field The system settings are saved in non volatile read only mem ory NVRAM and the system returns to the main Service Menu see Figure 5 3 The following paragraphs provide specific instructions for configuring each of the above system settings Set Date Screen The Set Date screen shown below displays the current date setting To change the date setting proceed as follows USE THESE KEYS TO SET DAY Jul 17 01 USE 10 34 am THESE USE THESE KEYS TO KEYS TO SET DAY SET YEAR Figure 5 5 Setting the Date 1 Adjust the month date or year setting using the appropriate keys as shown in Fig
61. ited to all blank spaces refer to Editing a View Name procedure on the preceding pages of this chapter The result is that when the user presses the key adjacent to any of the blank fields there is no response by the system In the event that you need to activate a currently inactive procedure change the Proc Enable setting to On the view is now enabled for selection by the user in the APR View Selection Screen d The Backup MAS field see Figure 4 11 item d sets the procedure specific AEC backup mAs limit The mAs limit range is configurable from 10 to 600 mAs in10 mAs incre ments Note The procedure specific backup mAs setting is independent of the system wide system backup mAs setting described in Chapter 5 User Maintenance The lower of the two backup settings will take precedence when set to unequal values SHH SHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HEHE HHH HHS SHH HHH HHH HHH HEHEHE HEHEHE HEHEHE HHH HEHEHE HEHEHE HEHEHE HHH EHH SH 2 V SC HF Series X ray Generators Operator s Manual Revision Q b Quantum Medical Imaging LLC 4 15 THIS PAGE INTENTIONALLY LEFT BLANK THOOHH HOE HHH HEE HHH FEE HHH HOH HHE HEH HEF HHH HHH HHH HHH HHEH HOH HHH EHH HHH ER EE HHH HHH HHH HHH HHH HHH HHH HHH HHH HEH HHH EE HHH EHH HHH HOH HOOD Revision Q HF Series X ray Generators Operator s Manual 4 16 Quantum Medical Imaging LLC Chapter 5 USER MAINTENANCE SPHHFHSSHSSSHSHSHSSHS
62. ler or customer SHH SHH HHH HHH HHH FHF HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HFEF HHH HHH HHH HHH HHH HHH HFEF HHH HHH HEHEHE HHH HHH HHH HEHE HHH HHH HEHEHE ZZ HHH HHH HEHE HHH HHH SD S Y SC HF Series X ray Generators Operator s Manual Revision Q ka Quantum Medical Imaging LLC 6 5 VOIDING WARRANTY Any installation maintenance repair service relocation or alteration to or of or any other tampering with the product performed by any person or entity other than Quantum Medical Imaging or a certified Quantum Medical Imaging dealer without the written approval of an authorized person at Quantum Medical Imaging shall immedi ately void and cancel all warranties with respect to the affected product THHHHH HOE III III IE HHH HOH HHE EE HEF HHH HEH EE HHH HEH HOH HHH EHH HHH HHH HHH HHH HHH HHH HHH HHH EHH HHH HHH HHH HHH HHH HHH EE HHH EHH HHH HOH HOOD Revision Q HF Series X ray Generators Operator s Manual 6 6 Quantum Medical Imaging LLC
63. mability APR pre defined and customized technique selection e Wall table none image receptor selection e Multi language display capability e Optional pedestal wall mount control console Compact generator cabinet design THEO EI ZZ IE III III HOH HOF EHH HEE EE HEE EE HOH HHH HEH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HEH HEH HHH HHH HHH ERR EEE HHH OOH HOHE Revision Q HF Series X ray Generators Operator s Manual 2 4 Quantum Medical Imaging LLC PERFORMANCE SPECIFICATIONS Refer to Table 2 1 for performance specifications of single phase non stored energy HF Series X ray Generators Table 2 1 HF Series Single Phase Generator Performance Specifications Maximum kW N A 32 40 mA Stations 25S 75S 1505 25S 755 1505 Small Focus S N A 100L 200L 250L 100L 200L 250L Large Focus L 320L 400L 500L 320L 400L 500L mA Accuracy N A 5 1mA 5 1mA kVp Range kVp N A 40 125 40 125 kVp increments kVp N A 1 0 1 0 step kV Accuracy N A 4 4 150 kVp optional N A Yes Yes Time Range sec N A 0 001 6 3 0 001 6 3 Time Accuracy N A 1 MS 0 5 1 MS 0 5 ai Ge N A 0 001 0 001 mAs Range N A 0 025 600 0 025 600 High SpeedStarter N A No No Ripple Voltage output N A 5 5 Time measured at 75 of the peak kVp waveform For exposure times from 1 0 mS to 49 mS time accuracy is 2 1mS for exposure times from 50 mS to 100 mS time accuracy is 1
64. mbination is now the film screen combination used when WALL 40 image receptor is selected for exposures using AEC and APR AEC mode exposures 2 Wall 72 FS setting Press the key next to the Wall 72 FS field until the desired film screen combination name appears The selected film screen combination is now the film screen combination used when WALL 72 image receptor is selected for exposures using AEC and APR AEC mode exposures 3 Table 40 FS setting Press the key next to the Table 40 FS field until the desired film screen combination name appears The selected film screen combination is now the film screen combination used when TABLE image receptor is selected for exposures using AEC and APR AEC mode exposures 4 Wall2 FS setting enabled only generators equipped with Universal type AEC Te AEC Board A11 AY40 0275 is installed Press the key SHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZZZ ZZ Z ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ HEHEHE ZZ EZ ZZ ZZ E A SC HF Series X ray Generators Operator s Manual Revision Q e Quantum Medical Imaging LLC 5 11 next to the Wall2 FS field until the desired film screen combination name appears The selected film screen combination is now the film screen combination used when WALL2 AUX image receptor is selected for exposures using AEC and APR AEC mode exposures 5 Auxiliary FS setting enabled only generators equippe
65. n 5 Select the desired ion chamber field combination using the three AEC field keys on the operator control panel A configuration of three boxes representing AEC chamber detectors is depicted on the display screen A shaded box indicates that a field is selected for use To select a different AEC field configuration press the left center and right AEC FIELD select keys until the desired con figuration is displayed There are a total of eight possible configu rations M KAKKAKKAKKA KI TOOTH HEH HHH HOH HEE HHH HEE III HHH HOH HHH EE EEE EE HHH HHH EE HHH EE HHH EE EFFE EE HOH ERR HHH HHH EE EE E EI E EE ERR HOH HHH Revision Q HF Series X ray Generators Operator s Manual 3 12 Quantum Medical Imaging LLC APR Mode When APR mode is selected X ray techniques and exposure factors are automatically set according to your specific program 1 Press the APR mode key on the operator control panel The sys tem displays the APR MODE ANATOMICAL REGION screen dis plays up to 10 anatomical regions 2 Select an anatomical region by pressing the corresponding key The system displays the APR MODE ANATOMICAL VIEW screen displays up to 10 anatomical views 3 Select an Anatomical View by pressing the corresponding key The system displays theAPR screen for the selected anatomical view see Figure 3 5 a On systems equipped with optional AEC IF AEC exposure control is enabled for this procedure but the operator wants to switch t
66. nce of high voltage A WARNING Warning hot surface SHH SHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HFEF HHH HHH HHH HHH HHH HHH HFEF HHH HHH HHH HEH HHH HHH HEHEHE HHH HHH HHH HEH HHH HHH HHH HHH HHH SD S Y SC HF Series X ray Generators Operator s Manual Revision Q ka Quantum Medical Imaging LLC 1 3 A WARNING Quantum Medical Imaging LLC disclaims all responsibility from any injury resulting from improper application of this equipment This equipment is sold to be used exclusively under the prescribed direc tion of a person who is licensed by law to operate equipment of this nature This equipment must be used in accordance with all safety procedures described in this manual and must not be used for purposes other than those described herein In the United States Federal law restricts this device to sale distribution and use by or on order of a licensed physician Quantum Medical Imaging LLC cannot assume responsibility for any mal functioning of this equipment resulting from improper operation mainte nance or repair or from damage or modification of its components Failure to observe these warnings may cause serious injuries PA WARNING X rays are hazardous to both patient and operator unless established safe exposure factors and operating instructions are observed Only qualified and authorized personnel shall operate this system In this context qualified means those legally permitted to o
67. ng Accesses the Set Date Screen for setting current date Set Time setting Accesses the Set Time Screen for setting current time e Film Screen Defaults Accesses the Film Screen Defaults Screen which is used to configure the default film screen combination for the various image receptors used with the generator e APR Format Press the key next to the APR Format field to scroll through available APR Procedure Screen format options All Windows APR Procedure Screens will display both Film Speed and CM thickness screen fields No CM APR Procedure Screens will not display the CM thickness screen field No FS APR Procedure Screens will not display the film speed screen field No FS No CM APR Procedure Screens will not display the film speed nor the CM thickness screen fields e Main returns the system to the Service Menu Language setting Sets language used for operator control panel display screens menus i e English French Spanish German or Italian 99 999 49496 9949 4994 999 99 4999 99 4 94996 9994 99 9 99094 9944994 994 9494 494 99 9 4994 9494 9944 49 94 9994 494 494494 944444 444 444 44449444494 49444444494 4494444494444944444 e Se HF Series X ray Generators Operator s Manual Revision Q e Quantum Medical Imaging LLC 5 7 Back Up Time MAS setting Displays current AEC exposure back up mAs setting only functions on systems equipped with QG AEC option The system allows for selection of a mAs sett
68. ns 21 CFR Sub chapter J and the performance standards attendant thereto Upon installation all certified products require the filing of Form FD 2579 Report of Assembly of a Diag nostic X ray System by the assembler i e the installer with the appropriate agen cies the Installation Quality Assurance Checklist must also be completed and properly distributed upon installation A copy of each form pink copy is provided to the user The installation report is completed by the installer and returned to Quan tum Medical Imaging LLC Those responsible for the planning of X ray equipment installations must be thor oughly familiar and comply completely with NCRP Report No 49 Structural Shielding Design and Evaluation for Medical Use of X rays and Gamma Rays of Energies up to 10 MeV as revised or replaced in the future Those authorized to operate test par ticipate in or supervise the operation of the equipment must be thoroughly familiar and comply completely with the currently established safe exposure factors and pro cedures described in publications such as Subchapter J of Title 21 of the Code of Fed eral Regulations Diagnostic X ray Systems and Their Major Components and NCRP Report No 102 Medical X ray Electron Beam and Gamma Ray Protection for Ener gies Up to 50 MeV Equipment Design and Use as revised or replaced in the future The ODYSSEY High Frequency HF Series X ray Generator hereinafter referred to as th
69. o manual non AEC exposure mode press the AEC button on the operator control panel b To switch back to AEC mode press the AEC button on the operator control panel once again 4 Verify the patient size thickness is correct To increase or decrease this value press the keys next to the PATIENT SIZE up down fields to adjust the thickness value the system automati cally adjusts all techniques according to pre programmed tech niques for the selected patient thickness value 5 Verify Select Exposure Factors In APR mode exposure factors kVp mA and mAs time are pre programmed Displayed expo sure factor values can be modified using the KV MA or MAS TIME increment decrement keys to select the required value The system displays an up or down arrow indicator up to indi cate the value was increased from the initially programmed set ting down to indicate the value was decreased from the initially programmed setting above any technique that is modified from the original APR setting The MAS TIME key allows you to select either mAs or time for display and adjustment SHH SHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HEHE HHH HHS SHHHHH HHH HHH HEHEHE HEHEHE HHEH HEHEHE HEHEHE HEHEHE HEHEHE EHH HHH SH 2 V be HF Series X ray Generators Operator s Manual Revision Q b Quantum Medical Imaging LLC 3 13 CHEST apnmove SKULL CERVICAL Si ABDOMEN fr LUMBAR d UPPER LOWER EXTR
70. of system settings and for access to the APR Edit utility 7 Upon validation of the password the Service Mode Menu displays Sa ep 4 st Calibrati b 4 ACCESSIBLE TO 3 u TECHNICIAN Self Test Figure 5 3 Service Menu LAIIZIZZZZZZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZZZ ZZ ZZ Z ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ EZ ZZ ZZ e s HF Series X ray Generators Operator s Manual Revision Q e Quantum Medical Imaging LLC 5 5 The Service Menu offers a selection of service oriented system utilities each of which is described briefly below Configuration utility Access limited to authorized service technician only Provides access to system configuration settings such as AEC enable disable rotor drive and tube type settings etc In addition it provides the Edit Welcome utility that enables the user to change the name of the Facility e g General Hospital and Dealer contact information as it appears on the Welcome Screen see Figure 5 1 Calibration utility Access limited to authorized service technician only Provides access to system calibration functions such as filament mA kVp calibration and analog to digital A D conversion calibrations Reports utility Access limited to authorized service technician only Provides access to Service Reports menu which provides Generator Tube Error Warnings and Diagnostic reports Self Test ut
71. om the time PREP was ini tiated Rotor Current ERROR Tube rotor is not getting cor Call Technical Service Support rect current to start Rotor Overrun ERROR Tube has been rotating for Call Technical Service Support longer than 30 continuous sec onds Rotor Temp WARNING Tube rotor is too hot Allow tube to cool for thirty minutes and then re take exposure Table ERROR The table bucky has been Verify that TABLE receptor was not Bucky selected but did not return the selected at OCP with system not ready signal to start exposure having table receptor This may take up to five sec onds after exposure is reguested Tube WARNING Backup time was reached Reduce technigue factor s Re take Limits based on current AEC tech exposure nique factors exceeding tube exposure limits SHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH ZZ ZZ ZZ ZZ HEHHE HHH HHH HHH HHH HEHE HHH HEHEHE HHH EHH HEHE HHH HOOD LEE S V See HF Series X ray Generators Operator s Manual Revision Q b Quantum Medical Imaging LLC 3 27 vr Table 3 3 Fault Messages V24 Brake ERROR Missing or incorrect 24 VDC Call Technical Service Support supply voltage supplies volt age for operation of external equipment e g tube stand wall stand etc Vref ERROR Missing or incorrect positive Call Technical Service Support voltage reference Vref value returned Vref ERROR Missing or incorrec
72. perate this equipment in the jurisdiction in which the equipment is being used and authorized means those authorized by the authority controlling the use of the equip ment Full use must be made of all radiation protection features devices systems procedures and accessories It is important that everyone having anything to do with x radiation be properly trained and fully acquainted with the recommendations of the National Council on Radiation Protection and Measurements as published in NCRP Reports available from NCRP Publications 7910 Woodmont Ave nue Suite 800 Bethesda Maryland 20814 3095 www ncrp com and of the International Commission on Radiological Protection www icrp org and take adequate steps to protect against injury N WARNING X ray equipment may cause injury if used improperly The instructions con tained in this manual must be read and followed when operating this unit Personal radiation monitoring and protective devices are available You are urged to use them to protect against unnecessary X ray exposure SOHO HEE HEE ERR FEE HEHEHE HOE EHH FEE HEH EEE HHH OEE HHF HEHEHE EE EEE EES EHH HEHEHE HEF EE HEHEHE EHH RE HOH HEHE EE EEE EHH EEE HOHE Eat Revision Q HF Series X ray Generators Operator s Manual 1 4 Quantum Medical Imaging LLC REGULATORY COMPLIANCE This certified Quantum Medical Imaging LLC medical device has been designed man ufactured and calibrated to comply with governing Federal Regulatio
73. quired such as table top and off table techniques 11 o Selects table Control Push Used for examinations where table Ld receptor button receptor and AEC are selected AEC feature is optional patient is posi tioned on tabletop 12 Selects auxiliary Control On systems equipped with more than receptor s Push button two image receptors press this key once to select third receptor default assignment is WALL2 press again to select a fourth receptor default assignment is AUX Used for exami nations when a typical wallstand or table bucky is not desired such as portable digital receptor applications 13 i Selects top leftion Control Push Turns on off ion chamber s top left chamber detector button detector for AEC exposures location 14 Selects middle ion Control Bush Turns on off ion chamber s middle 8 chamber detector button detector for AEC exposures location 15 u Selects top right Control Push Turns on off ion chamber s top right ion chamber button detector for AEC exposures detector location THEO HOE HEHE HEF HHH EE HOH HOH EE HHH EE EEE EE HOH HHH HHH HHH HHH HHH HHH HOH HHH HHH HEH HHH HHH HHH HHH HEH HHH HHH HHH HHH HHH HEH HHH OOH HOHE Revision Q HF Series X ray Generators Operator s Manual 2 16 Quantum Medical Imaging LLC 16 N A 17 N A 18 N A 19 N A 20 N A Ready Prepped PREP button prepares genera tor for exposure release to inhibit exposure Exposure On EXPOSE button initia
74. racter field to scroll through all available alphanumeric characters begins with a no character or blank space until the character you want to insert appears in the Character field window c To insert the selected character in the position chosen in step a press the key next to the Insert Character field The selected character will now appear in the designated position of the text string NOTE THIS CHARACTER NOTE THIS CHARACTER IS ALINE RETURN IS ABLANK SPACE USE IT TO START NEW LINE BELOW PA THE CURRENT LINE O space 4 4 0 0 1 2 3 4 5 6 7 819 3 lt lt gt A B C D E F G H I J K L M N O P Q R S T U V W X Y Z _ a b c d e f g h i j k l m n O p q T t u V W X y Z ET y i 2 p 0 4 Chal a Yo Yay Yay in A A A A A E C E E 1 1T D 0 0 Y P B i Eh 3 SHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HFEF HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH ZZ HEHEHE HHH HHH HHH HHH SD S K Weg HF Series X ray Generators Operator s Manual Revision Q b Quantum Medical Imaging LLC 4 7 HHH HHH HOE HEHEHE HHH FEE HHH HEH HHH EE HEH EE HEE EE HHH EE EE HHH EE EEE EE III II EEE III III III III EE EE HOH II III III
75. re LED Indicates tube rotor and filament OPREP are prepared for exposure EXPOSURE PREP Key Starts the tube rotor pre heats CONTROLS AND tube filament and performs generator self test INDICATORS O EXPOSE LED Indicates exposure in process EXPOSE Key Starts the exposure NOTE AN AUDIBLE TONE IS EMITTED DURING EXPOSURE IT STOPS TO INDICATE END OF EXPOSURE EXPOSE KEY MUST REMAIN PRESSED THOUGHOUT EXPOSURE PERIOD OTHERWISE EXPOSURE WILL BE TERMINATED Figure 3 9 Exposure Controls PHHHHH HOE III III ZEIT EI ZEIT HEH HHH EHH HHH HHH EE HOH HHH EHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH EE HHH EHH HHH HOH HOOD Revision Q HF Series X ray Generators Operator s Manual 3 20 Quantum Medical Imaging LLC Exposures Press and hold the EXPOSE key switch until the exposure is complete The amber EXPOSE indicator will light X ray beam is on and a single audible tone will sound during the exposure Instantaneous Exposures 1 Press and hold PREP until the green PREP indicator lights 2 Press and hold EXPOSE until the exposure is complete The amber EXPOSE indicator will light X ray beam is on and a single audible tone will be emitted during the exposure HEAT UNIT COMPUTER Anode heat units are computed and displayed continuously on screen thereby monitoring the dissipation of heat between exposures The amount of tube anode heating expressed in thousands of heat units is the kHU value or
76. rectly connected to the public low IEC 61000 3 2 voltage power supply network that Voltage fluctuations Complies supplies buildings used for domestic flicker emissions purposes IEC 61000 3 3 HF Series X ray Generators Operator s Manual Revision Q Quantum Medical Imaging LLC 1 9 Guidance and manufacturer s declaration electromagnetic immunity The HF Series X ray Generator is intended for use in the electromagnetic environment specified below The customer or the user of the HF Series X ray Generator should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environ ment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are cov ered with synthetic material the relative humidity should be at least 30 Electrical fast transient burst 2 kV for power supply lines 2 kV for power supply lines Mains power quality should be that of a typical commercial or IEC 61000 4 4 1 kV for input output 1 kV for input output hospital environment lines lines Surge 1 kV differential mode 1 kV differential mode Mains power quality should be IEC 61000 4 5 2 kV common mode 2 kV common mode that of a typical commercial or hospital environment Voltage dips lt 5 Ur lt 5 Ur Mains power guality sho
77. renieavivecusees 2 4 KEY FEATURES eieiei deeg 2 4 PERFORMANCE SPECIFICATIONS cccecccceeseseeceeeeeseceseetsnaesseseetanaes 2 5 ELECTRIC OUTPUT DATA 2 9 MAIN COMPONENTS iiiiinnenennsnnan aa ennem 2 12 ACCESSORIES u nen en 2 13 CONTROLS AND INDICATORS uesssssssesseeseesnnnsneeenennnnenenennnenne nen 2 13 HELP SERVICES SCREEN ccccescecceeseeeseeeeeesseeeeeteessnsesteeseaaneeeteeaaa 2 19 CHAPTER 3 OPERATION OVERVIEW aa en anne 3 3 POWER ON OFF PROCEDURES cccccsseeseceeeseeeseeteetsnseeseetense nesses 3 3 Power On Procedure 3 3 Power Off Procedure ueessessesennseenenennnnnnnnnen nennen nenn nenn nennen 3 4 Automatic Power Standby Mode eessesiinreerereeeeee nn 3 5 DAILY TUBE WARM UP PROCEDURE ccccccccssesseecerseeeeseeeensenesses 3 6 SETTING UP TO TAKE EXPOSURES cccccccsseeeseeteesseseeseetensnaeesesteessnees 3 7 RECEPTOR SELECTION cccccssseseeeceeseseeeeteeseneeseetsasenseeeetsnsessetes 3 7 MODE SELECTION sicisadececidcnbictsdiacitundoatiaieaieesindietudadbeawaitanaaandavadaees 3 9 Manual Mode ccccccccccccsceccceceeseseeaceeueeeseceeeceeeeuevaeeuueeeuaeenanes 3 11 IEG IM OGG a Satan Ee 3 12 APR MOG Gs i antava vas sania vanaaja ama maas a a ajalt il va Saaja 3 18 TAKING AN EXPOSURE ccccccecsseceeceeeeseeseseetseeenesteessneeseeteassnens 3 24 HF Series X ray Generators Operator s Manual Revision Q Quantum Medical Imaging LLC
78. rt due to missing PREP hardware line or PREP signal interruption during PREP cycle SHH SHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HFEF HHH HHH HHH HHH HHH HHH HHH HHH HEHEHE HHH HHH HHH EHH HHH HHH HOD OHHH HHH HHH HHS a HF Series X ray Generators Operator s Manual Revision Q b Quantum Medical Imaging LLC 3 25 Table 3 3 Fault Messages Lost Expose WARNING Exposure terminated prema Press any key on Operator Control turely because EXPOSE key Panel to clear warning message was released before exposure Re take exposure was completed MA Current ERROR Indicates the exposure mA cur Call Technical Service Support rent is out of range Minus 15V ERROR Missing or incorrect 15 VDC Call Technical Service Support supply voltage Model Limits ERROR PREP was pressed while a Call Technical Service Support combination of the selected parameters is outside of the limits of the generator model or tube type No HSS ERROR No High Speed Starter Call Technical Service Support Connected detected when selected No HV ERROR High Voltage transformer or Call Technical Service Support Power Module not functioning properly Not Rotating ERROR Tube rotor or generator rotor Call Technical Service Support drive high speed starter HSS unit not functioning properly NVRam ERROR Non volatile random access Call Technical Service Support memory NVRAM returne
79. s and ion chamber field combinations selections are correct If required these settings may be adjusted as follows SHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH ZZ ZZ ZZ ZZ ZZZZ ZZ ZZ ZZ Z ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ HEHEHE HEHEHE HEHEHE HHH SD e See HF Series X ray Generators Operator s Manual Revision Q e Quantum Medical Imaging LLC 3 15 a The default AEC density setting may be modified using the AEC DENSITY increase up arrow and decrease down arrow keys adjacent to the Density field Each step adjusts density up down by 15 relative to the average setting For APR manual non AEC exposures the PATIENT TYPE set ting changes the time mAs value in 25 increments Modify the patient type setting using the increase up arrow and decrease down arrow keys adjacent to the PATIENT TYPE field b Film Screen choices are WALL 72 WALL 40 TABLE 40 100 Speed 200 Speed 800 Speed and FILM Screen If the Digital Receptor setting in Service Configuration Menu is enabled an eighth option entitled Digital will also be select able Note that these are the factory default film screen combination names However each can be edited as neces sary by user NOTE In the event light films are produced cA verify the correct film screen combination is selected c For AEC exposures select the desired ion chamber field com bination using the three AEC field keys on the operator
80. s and the age of the batteries Time measured at 75 of the peak kVp waveform For exposure times xK BEER SC Revision Q 2 6 from 1 0 mS to 49 mS time accuracy is 2 1 mS for exposure times from 50 mS to 100 mS time accuracy is 1 1 mS NOTE AEC TECHNIQUES SHOULD HAVE EXPOSURE TIMES EXCEEDING 8 MILLI SECONDS mAs is tube dependent the generator may not reach maximum mAs due to tube type TEE HHH HEHEHE EHH HEE HHH EEE HHH EEE HEE EHH HEH EE EEE ERR EEE EE HEHEHE EE EHH HEH HEH ESE EHH EEE EHH EEE HHH EHH Eat HF Series X ray Generators Operator s Manual Quantum Medical Imaging LLC Refer to Table 2 3 for performance specifications of three phase HF Series X ray Generators Table 2 3 HF Series Three Phase Generator Performance Specifications Maximum kW 32 40 50 65 80 a e 295 755 1505 258 758 1508 258 758 150S 258 758 150S D Large Focus L 100L 200L 100L 200L 100L 200L 100L 200L 3201 AOOL 9 250L 320L 250L 320L 320L 400L 320L 400L 500L 650L 400L 500L AOOL 500L 500L 650L 500L 650L 800 800 mA Accuracy 5 1mA 5 1mA 5 1mA 5 1mA 5 1mA kVp Range 40 125 40 125 kVp 40 125 40 150 with 40 150 with 40 150 40 150 QG 150 option QG 150 option kVp increments kVp step 1 0 1 0 1 0 1 0 1 0 kV Accuracy 4 4 4 4 4 Ge kvp Yes Yes Yes Standard Standard optional o 0 001 6 3 0 001 6 3 0 001 6 3 0 001 6 3 0
81. s value i e the lowest limit of the thickness range that the technologist will be able to achieve when using this particular procedure IMPORTANT The following step is only a required if you are configuring the proce dure for the first time or to change the thickness range it is not required when changing exposure values for a particu lar thickness set point s To edit expo sure values for a particular set point press the PATIENT SI ZE up or down keys until the desired set point e g 32 CM is displayed then edit kVp mA and or time mAs values as necessary b Set the CM STEP field Figure 4 10 item f to Edit Base CM c Using the up down arrow keys next to the PATIENT SI ZE field Figure 4 10 item b set the CM value to the minimum thickness setting reguired for this procedure referred to as the Base CM value d Select the AEC On Off setting by pressing the key next to the AEC On Off field Figure 4 10 item c Note AEC is only functional on generators equipped with AEC option When this field is set to AEC On automatic exposure control will be enabled for this procedure When set to Off the proce dure is configured for non AEC or manual exposure mode of operation Note When set to AEC On the operator dur ing normal operation can switch between AEC and non AEC modes using the AEC mode button e Select image receptor setting by pressing the appropriate image receptor keys WALL 40 WAL
82. set the facility main power disconnect switch to the OFF position Automatic Power Standby Mode To help save power and protect system electronics when left unattended the generator is equipped with an automatic power off feature that places the system in stand by mode i e system low voltage state After a user defined period of inactivity selectable from 0 30 60 90 120 150 or 180 minute delay settings the system will automatically enter a low power state through software control leaving all power switches in the on position during generator configuration by the service technician Setting the Generator Timeout to 0 will disable the system from entering Automatic Power Stand By Mode d NOTE The power standby period must be set Refer to the paragraph entitled Generator Timeout Setting in Chapter 5 USER MAINTENANCE for instructions on setting the automatic shut off time To avoid entry into Automatic Power Stand By mode press any key on the operator control panel before the delay setting expires or set the Auto Shut off time to 0 To re start after Automatic Shut Down refer to Power On Procedure HF Series X ray Generators Operator s Manual Quantum Medical Imaging LLC DAILY TUBE WARM UP PROCEDURE All tube manufacturers recommend X ray tube seasoning following installation and daily prior to use Systems used infrequently should have the tube seasoned on a daily basis It is important that these proced
83. t positive Call Technical Service Support voltage reference Vref value returned Wall ERROR The wall bucky has been Verify that WALL receptor was not Bucky selected but did not return the selected at OCP with system not READY signal to start expo having wall receptor sure This may take up to five seconds after exposure is requested SHH HOE HHH HEH HHH EE HOH HEHEHE HHH EEE EE HOH HHH HHH HHH HHH HHH HHH HHH HHH HHH HEH HHH HHH HHH HHH HEH HHH HHH HHH HHH HHH HEH HHH OOH HEHE Revision Q HF Series X ray Generators Operator s Manual 3 28 Quantum Medical Imaging LLC Chapter APR EDITOR SPHSHSHSSHSHSHSHSHHSHSHHSHHHHSHHSHHHHHHHHHHHHHSHHHHHHHHHHHHOHHHHODH 4 1 OVERVIEW This chapter provides instructions for creating and editing Anatomical Programmed Region APR technique screens The HF Series X ray generator has the capability of recalling up to 5000 pre programmed APR screens by storing up to 10 views tech niques in each of the 10 anatomical regions with each view having up to 50 thick ness cm selections Any of the existing X ray techniques can be edited to suit the individual needs of your practice SKULL AP DA uc KVP MA MASIC u 73 150S 028 1 SCH Du enen Figure 4 1 APR Edit Mode Screens SHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HEHEHE HHH HHH ZZ HEF HHH HHH HHH HHH HHH ZZ ZZ ZZ ZZ ZZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ HEHEHE HEHEHE HEHE
84. tes exposure Indicator Indicator Control Push button Indicator Control Push button When lit indicates generator is ready for exposure When blinking indicates one of the following conditions When the PREP LED is illuminated green indicates tube rotor and fila ment are prepared for exposure When pressed initiates rotor acceler ation and X ray tube filament pre heating During the production of X rays the EXPOSE LED is illuminated yellow Dead man type switch i e requires continuous switch activation throughout entire exposure cycle Initiates the exposure If preparation cycle PREP is not complete initiates preparation cycle tube heat limit will be exceeded by the next exposure On stored energy systems the bat teries are re charging blinking stops when fully charged On non stored energy systems 400V capacitors are re charging SHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HEF HHH HHH HHH HHH HHH HEHEHE HHH HHH HHH HEHEHE HEF HEHEHE ZZ ZZ ZZ ZZ ZZ ZZ ZZ HEHEHE HEHEHE HEHEHE HEH HHH SD H 7 HF Series X ray Generators Operator s Manual ka Quantum Medical Imaging LLC Revision Q 2 17 THIS PAGE INTENTIONALLY LEFT BLANK THHHHH HOE HHH HEE HHH FFE HHH HOH HHE EHH HEF HEHEHE HHH HEHE HHH HOH HHH EHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HEH HHH EE HHH EHH HHH HOH HOOD Revision Q HF Series X ray Generators Operator s Manual 2 18 Q
85. time setting 0 8s is the actual value used to determine the exposure time Refer to the operator s manual provided with the tomography system for complete tomography system operation instructions Tomographic Mode The HF Series X ray Generator Operator Control Panel OCP display will automatically switch between radiographic exposure mode to tomo graphic exposure mode when the following operator actions are taken e A valid tomographic combination selection is made using the tomography system s angle time keys and The tomography system table receptor is selected usually this receptor is selected using the TABLE receptor key on the HF Series X ray Generator operator control panel If the generator is in either Manual or AEC mode and tomo mode is acti vated the Manual Tomo Mode screen will appear as shown below 7 03 TOMO ANGLE ea te EXPOSURE TI TIME SETTING SETTING TOMO BE kvp MA mas SEN 70 1508 15 0 kHU me m Figure 3 7 Manual Tomo Mode Screen 6er TOOT HHH HHH HHH HEH III III III III III III II HF Series X ray Generators Operator s Manual Revision Q Quantum Medical Imaging LLC 3 17 If the generator is in APR mode and tomo mode is activated the APR Tomo Screen will appear as shown below El Apr 7 03 THORAX MAS 6 13 pm EEN TIME Patient ANGLE KvP MA Che Ge TIME ER l SETTING 77 400L 600 a FROM K AT e TOMO 30CM SYSTEM Tomo kHU 400 EXPOSURE
86. ts are designed to meet stringent safety standards All medical elec trical equipment requires proper installation operation and maintenance particularly with regard to safety It is vital that the user read understand note and where applicable strictly observe all Warnings Cautions Notes and Safety markings within this document and on the equipment and that the user strictly follow all safety directions in this manual to help ensure the safety of users and patients Every reasonable precaution has been taken during manufacture to safeguard the health and safety of persons who will operate this equipment The following precau tions must be observed at all times WARNINGS CAUTIONS NOTES The following samples show how warnings cautions and notes appear in this docu ment The text explains their intended use WARNING Indicates injury or death is possible if the instruc tions are not obeyed Instructs users to refer to documentation if displayed without warning text AN CAUTION Indicates that damage to equipment is possible if the instructions are not obeyed Gq NOTE Notes provide advice and highlight unusual points A note is not intended as an instruction The purpose of safety icons such as those shown below is to indicate at a glance the type of caution warning or danger WARNING lonizing radiation indicates the possibility of increased levels of radiation AN WARNING Dangerous voltage indicates the prese
87. tside of preset limits at any time Filament Too ERROR The OCP sent a request to the Call Technical Service Support Low Generator to set a PREP value that is too low Flash ERROR Flash memory returned incor Call Technical Service Support rect checksum Flash Not ERROR Flash Memory U2 on OCP Call Technical Service Support Loaded Control Board A16 AY40 004S1 not loaded with pro gram that corresponds with the version of the Microcontroller program in the OCP Flash Memory is experiencing a read problem Generator ERROR Received communication mes Press any key on Operator Control Comms sage not valid Panel to clear warning message Re take exposure Generator Not ERROR 1 No operator activity for over Call Technical Service Support Ready 60 minutes 2 Failed communication between generator cabinet and operator control panel 3 Wrong program loaded into OCP Control Board A16U3 Hold Button WARNING Exposure terminated prema Re take exposure making sure turely because the EXPOSE key EXPOSE key is continuously pressed was released before the expo throughout entire duration of expo sure was completed sure HSS Detected ERROR A High Speed Starter has been Call Technical Service Support detected when none is selected Keybrd ERROR Indicates a key on the Front 1 Verify a switch on the keyboard Panel is stuck on was not accidentally depressed during power up 2 If not the above call Technical Ser
88. uantum Medical Imaging LLC Chapter OPERATION SPHHSHSSHSSHSHSHSHSSHSHHSHSHSSHSHHHSHSHSHSHHHSHSHHHSHSHHHHSHSHHHHHSHHHHHH OS OVERVIEW This chapter provides the information necessary to operate the generator The following operating procedures are outlined and described in detail Power on off procedures e X ray tube seasoning procedure e Operating instructions with reasonable care Manufacturer s equipment recommendations described in this manual must be observed N WARNING This equipment must be operated WARNING Do not operate the generator if water A has leaked into or around the generator cabinet Call service before applying power to the system POWER ON OFF PROCEDURES NOTE On STORED ENERGY SE models after og long periods of non usage allow system to re charge batteries for approximately four hours prior to using disable AUTO SHUTOFF for contin uous charge and turn on system refer to Gener ator Timeout Setting paragraph in this chapter for instructions It is recommended the system be turned on at least once a week to maintain battery charge The following procedures describe the steps necessary to perform system power on and power off shut down Power On Procedure 1 Verify the main circuit breaker on the generator cabinet is set to the ON position 2 Set the power on standby switch on Operator Control Panel OCP to the on position If the switch is already in the on position
89. uantum Medical imaging LLC 2002 Orville Drive Nortt Ronkonkoma NY 11779 USA LABEL LOCATIONS CDR Figure 2 2 Generator Label Locations ACCESSORIES Equipment described in this manual shall only be used in combination with other equipment or components if these are expressly recognized by Quantum Medical Imaging LLC as compatible A list of such equipment and components is available from Quantum Medical Imaging LLC on request CONTROLS AND INDICATORS The operator controls and indicators on the HF Series X ray Generator are divided into two basic groups SHH SHH HHH HHH HHH HHH HHH HHH HHH HHEH HHH HHH HEHEHE HHH HEHEHE HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HEH EHH HHH HEHE HEH HHH HHH HEHE ZZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ CRT 7 Revision Q HF Series X ray Generators Operator s Manual k E 2 12 Quantum Medical Imaging LLC Generator Cabinet and Operator Console Power Switches see Figures 2 3 and 2 4 Operator Control Panel Operator Controls and Indicators see Figure 2 5 and refer to Table 2 5 for descriptions Power ON OFF Switch ON up OFF down Figure 2 3 Generator Cabinet Power On Off Switch POWER ON STANDBY SWITCH Figure 2 4 Operator Console Power ON OFF Switch PHO OHHHOH HHH OHH HOF EHH FHF EZ FEF EE OOH III EHH HOE H HH HOF HHH HEH HHH EHH HHH HHH HEE HHH OEE ERR HHH HHH OHH HEH HHH HHH EE HHH RER HOH HEH HHH HEH HOOD HF Series X ray Generators Operator s Manu
90. uld be short interruption gt 95 dip in UT gt 95 dip in UT that of a typical commercial or and voltage for 0 5 cycle for 0 5 cycle hospital environment If the variations on power user of the HF Series X ray Gen supply 40 Ur 40 Ur erator reguires continued opera input lines of Ars a A tion during power mains IEC 60601 4 11 a S nu nn 2 lfi Ur interruptions it is recommended or 5 cycles or 5 cycles that the HF Series X ray Genera tor be powered from an uninter 70 UT 70 UT ruptible power supply or battery 30 dip in UT 30 dip in UT lt 5 UT lt 5 UT gt 95 dip in Ur for 5 s gt 95 dip in UT for 5 s Power freguency 50 60 Hz IEC 61000 4 8 3 A m 3 A m Power freguency magnetic fields should be atlevels characteristic of a typical location in a typical commercial or hospital environ ment NOTE Ur is the A C mains voltage prior to application of the test level SHEE HEHEHE HEHE HEH FEE HEHEHE HOE HOH EEE HEHEHE HHH EEE HHH HEHEHE EE EEE HHH EHH SE EHH HEHEHE EEE HHH EHH HHH OH RRE EHH EEE EHH EEE HHH EEE Eat Revision Q HF Series X ray Generators Operator s Manual 1 10 Quantum Medical Imaging LLC Guidance and manufacturer s declaration electromagnetic immunity The HF Series X ray Generator is intended for use in the electromagnetic environment specified below The customer or the user of the HF Series of X ray generators including TechVision option
91. ults Time MAS kb 4 APR Format Generator Windows 60 d Date Format ED m Figure 5 10 Choosing Language and Date Format Settings ES ES Press the key next to the Generator Timeout field to scroll through available time period settings 0 timeout disabled 30 60 90 120 150 180 minutes until desired setting is displayed SHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HEHEHE HEHEHE HEHHE HEHEHE ZZ ZZ ZZ ZZ ZZ ZZ ZZ HEHEHE HEHEHE HHH HHH HHH SD Revisi HF Series X ray Generators Operator s Manual evision Q 5 e Quantum Medical Imaging LLC 5 13 Datei Time Format Setting The Date Format field see Figure 5 10 displays the current date time format setting To change the date time format setting proceed as follows 1 Press the key next to the Date Format field to select either US e g JUN 11 01 2 00 pm or Int e g 11 JUN 01 14 00 date time format used for display of date and time on all applicable system screens 2 To exit the screen and return to the Service Menu press the key next to the Main field THEE HOE HHH HEE HHH FEE HHH HOH II III HOF HHH HHH EE HHH HHH HOH HHH EHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HEH HHH HHH EE HHH EHH HOE HOH HOOD Revision Q HF Series X ray Generators Operator s Manual 5 14 Quantum Medical Imaging LLC Chapter 6 WARRANTY INFORMATION SPHHSHSSHSSHSHSHSSSHSSHSHHSHSHSSHSHHSHSHSHSHHHHSHSHHHSHSHHH
92. unction AEC uses an ion chamber to fcr for manual selection of all technique E 1 UAL ee without automatic system intervention Figure 3 2 Mode Selection e ANATOMICAL PROGRAMMED REGIONS APR MODE After a thickness measurement is entered prior to exposure the system PHO OOH HOE III III ZEIT ZI EHH HEF HHH HHH III III II EI HHH EHH HHH HHH HHH HHH HHH HEH HHH HHH HHH HHH HHH HHH HHH HHH HHH EE HHH EHH HEE HOH HOOD Revision Q HF Series X ray Generators Operator s Manual 3 8 Quantum Medical Imaging LLC automatically sets programmed X ray techniques and exposure factors including AEC mode if available bucky selection kVp mA time initial anatomical thickness and AEC density Pre programmed APR settings can be modified at any time through the operator control panel either in the actual stored program or for the current exam i e temporarily During exposure AEC optional works in conjunction with APR to provide uniform repeatable high quality images with the smallest variation of optical density between patients Additionally a Patient Type sub mode is provided for APR selections that are configured for manual exposure control i e non AEC When this sub mode is accessed the patient type is initially set for average i e typical cm setting The PATIENT TYPE up down keys provide quick time mAs compensation 25 and 50 settings in accordance with physical characteristics of patient e muscular thin
93. ure 5 5 above until desired date is displayed 2 To exit the screen and return to the Start Up Screen press the key next to the Main field HF Series X ray Generators Operator s Manual Revision Q Quantum Medical Imaging LLC 5 9 Set Time Screen The Set Time screen shown below displays the current time setting To change the time setting proceed as follows USE THESE moon EA HOURS 10 35 am Za E SS a MINUTES Figure 5 6 Setting the Time 1 Adjust the hours or minutes setting using the appropriate keys as shown in Figure 5 6 above until desired time is displayed 2 To exit the screen and return to the Start Up Screen press the key next to the Main field Film Screen Defaults Screen The Film Screen Default Screen see Figure 5 7 is used to configure the default film screen combination for the various image receptors used with the generator This setting applies to AEC and APR AEC mode exposures only If the selected film screen combination has not already been calibrated it must be calibrated according to the AEC Calibration procedures later in this chapter using the selected film screen combination setting The generator has been pre configured with seven different film screen com bination selections designated Wall 40 Wall 72 Table 40 100 Speed 200 Speed 800 Speed and Film Screen7 If the Digital Recep tor setting in Service Configuration Menu is enable
94. ures be performed to maintain both the tube manufacturer s and Quantum Medical Imaging LLC s warranties A typical daily seasoning procedure is provided below If the system has not been used for several days or upon installation refer to the tube manufacturer s instructions Proceed as follows 1 Select MANUAL mode from the Operator Control Panel 2 Take exposures listed in Table 3 1 using a 200L mA focal spot and allow 30 seconds between exposures 3 The system is ready for operation Table 3 1 Daily X ray Tube Seasoning Exposures 1 50 0 1 2 60 0 1 3 70 0 1 4 80 0 1 5 90 0 1 6 100 0 1 7 110 0 1 8 120 0 1 NOTE Exposure 8 X ray Tube seasoning set ch tings applies only to generators capable of 150kV SHH SHH HHH HHH HHH FHF HHH HHH ZZ ZI ZZ ZZ ZZ ZZ ZZ ZI ZZ ZZ EZ ZZ ZZ ZZ ZZ ZZ ZZ ZZ HEHEHE HHH HHS SHHHHH HHH HHH HHH HEHEHE HEHEHE HHHE HEHEHE HEHEHE EHH HH HH HH HH HH S K SE HF Series X ray Generators Operator s Manual Revision Q b Quantum Medical Imaging LLC 3 5 SETTING UP TO TAKE EXPOSURES Before taking an exposure the operator must prepare the system as follows e Select Receptor Select Mode Select verify exposure factors RECEPTOR SELECTION In any operating mode use the RECEPTOR selection keys see Figure 3 1 to select the appropriate image receptor Select NONE if not using a table or wall receptor The currently selected receptor is indicated in the win
95. vice Support Revision Q 3 24 HF Series X ray Generators Operator s Manual Quantum Medical Imaging LLC Table 3 3 Fault Messages KVP Error ERROR 1 The anode and cathode KV Call Technical Service Support are sufficiently different to cause an error 2 The anode or cathode volt age sense is exceeding 15 of program voltage 3 Communication between OCP Control Board A16 AY40 00451 and Logic Board Al AY40 006S was interrupted 4 Arc in system H V Trans former x ray tube or high voltage cables 5 Faulty missing J 1 connection on KVP Control Board A2 AY40 003S KVP Volts ERROR Indicates the exposure kVp Call Technical Service Support voltage is out of range LBMastMicro ERROR Logic Board master micropro Call Technical Service Support cessor A1U9 returned incor rect checksum LBSlaveMicro ERROR Logic Board slave microproces Call Technical Service Support sor A1U10 returned incorrect checksum Line 400V ERROR Missing or incorrect anode Call Technical Service Support Anode sensing voltage Vsense A value returned Line 400V ERROR Missing or incorrect cathode Call Technical Service Support Cathode sensing voltage Vsense C value returned Long Prep WARNING PREP time exceeded the eigh Press any key on Operator Control teen seconds limit Panel to clear warning message Re take exposure Lost Prep ERROR Exposure sequence disabled Call Technical Service Suppo
96. vice Support lem with the anode kVp circuit high voltage transformer or Power Driver Module Backup WARNING Applicable in AEC mode only 1 Reduce technique factor s Re Backup time was reached take exposure based on current technique 2 Verify the preset backup setting factors This includes exceed displayed on APR screen is suffi ing the maximum output cient This setting is configurable capacity of either the x ray using APR Edit function refer to tube or the x ray generator Chapter 4 for instructions model 3 Verify appropriate receptor is selected Battery ERROR Faulty stored energy cell in top Call Technical Service Support Discharged battery tray Cathode on two tray systems or Line Monitor Charger board not adjusted correctly HF Series X ray Generators Operator s Manual Quantum Medical Imaging LLC Revision Q 3 22 Table 3 3 Fault Messages Battery 2 ERROR Faulty stored energy cell in Call Technical Service Support Discharged bottom battery tray Anode on two tray systems or Line Mon itor Charger board not adjusted correctly Battery NotRdy WARNING Battery voltage is too low Wait until READY light on Operator This warning message applies Control Panel is steady before taking only to stored energy genera exposure tors Battery ERROR Line Monitor Charger board Call Technical Service Support Overcharged not adjusted correctly Battery 2
97. with the safety procedures given within this manual for the purpose for which it was designed The intended use is given below Nothing stated in this manual reduces user s profes sional responsibilities for sound judgment and best practice SHH SHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HHH HEHE HHH HHH HHH HHH HEHE HHH HEHEHE HHEH EHH HEHEHE HHH HOOD LEE S V SC HF Series X ray Generators Operator s Manual Revision Q b Quantum Medical Imaging LLC 2 3 INTENDED USE The HF Series X ray Generator is a diagnostic X ray high voltage generator It is intended to supply and control the electrical energy applied to a diagnostic X ray tube for medical veterinary radiographic examinations Use of the equipment for purposes other than those intended and expressly stated by the manufacturer as well as incorrect use or operation may relieve the manufacturer or his agent from all or some of the responsibility for resultant non compliance damage or injury KEY FEATURES It is imperative that all safety procedures described in this manual be strictly adhered to in order to ensure the safety of both patient and user The significant physical and performance characteristics of the HF Series X ray Generators are as follows e Ultra high frequency up to 120 kHz voltage waveform for highly efficient X ray production e Manual AEC optional and automatic APR modes of operation e Anatomic Program
Download Pdf Manuals
Related Search
Related Contents
取扱説明書等 SVED® Wound Treatment System User Manual コルクタイルの施工方法 P25 Medium – Heavy Duty Side Mount Pan/Tilt Unit Extractos vegetales control Ojo de gallo en café MAINTENANCE INTERVALS - Safety HI 3841 Estojo de Testes de Dureza Gama Média Instrucciones para el acceso de los padres a navega Kombi Feedback Switch bearbeitet.pages Copyright © All rights reserved.
Failed to retrieve file