1. ADAM Adam Equipment MCB SERIES MEDICAL SCALE MCB 300 and MCB 400 Wheelchair and Bed Scales Software Per 4 ADAM EQUIPMENT CO LTD pn 3822 Rev B2 July 2002 ALYS Labware Lausanne Tel 021 312 42 60 Fax 021 312 42 61 labware alys technologies com INTRODUCTION The MCB 300 and MCB 400 scales are designed for weighing wheel chairs and beds They have a capacity of 300kg 400kg and readability of 0 1kg They are fast accurate and simple to use Main features of the scale are The scale is very simple to use It has the ability to run from internal AAA alkaline batteries or an external power supply HOLD function to lock the display until the operator releases it Automatic turn off to maximise battery life Filters optimised for weighing people ASSEMBLY INSTRUCTIONS The MCB 300 400 comes with the 2 beams that the wheelchair or bed can be rolled up and a separate indicator module Cables connect the beams to the indicator The indicator could be mounted to the wall or used on a table top Brackets are available for wall mounting the indicator Connect the cables to the indicator The cables must be secure for proper operation POWER SUPPLY lf the external power supply is used plug it into the jack on the indicator If batteries are used install them before mounting the indicator in the pillar Adam Eguipment 2002 1 OPERATOR S INSTRUCTIONS Key description ON OFF Turns the scale on or of
2. The equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense Shielded interconnect cables must be employed with this equipment to insure compliance with the pertinent RF emission limits governing this device Changes or modifications not expressly approved by Adam Equipment could void the user s authority to operate the equipment Adam Equipment 2002 8 Adam Equipment 2002 ADAM EQUIPMENT is a global supplier with more than 30 years experience in the production and sale of electronic weighing equipment Products are supplied and serviced from our company locations in the UK USA and SOUTH AFRICA and distributed through a world wide dealer network ADAM EQUIPMENT s products are predominantly designed for the laboratory educational medical and industrial markets The product range can be summarised as follows Analytical and Precision laboratory balances Basic range of top loading balances for schools and educational establishments Counting scales for general industrial and warehouse applications Digital weighing check weighing scales High performance platform scales with extensive software fac
3. ones lf an external 6V lead acid battery is used it will be necessary to use a charger compatible with the battery Contact your supplier for more information Typical Battery Life is 12 hours Adam Equipment 2002 5 CALIBRATION To prevent unauthorised calibration the Calibration Procedure can be performed after entering a code number It is the last function displayed at the end of the service parameters code 1948 PROCEDURE Zero the balance with no weight on the pan Press TARE then within 2 seconds press TARE and ZERO together The display will show code then 000000 Press TARE to move the flashing digit to the right and PRINT HOLD to increment the digit until the display reads 001948 then press F The first service parameter will now be displayed To change any parameter press PRINT HOLD to step to the next parameter press F Con Stability time Min 1 Max 9 Typical value Con 3 dlS Display update rate Min 1 Max 9 Typical value dlS 3 Zon Zero tracking range Typical value is 0 13kg Uni Weight display Units Kg 1 Ib 2 Normally set to kilograms Now the raw A D counts are displayed Press TARE display shows cALibr Press F to skip calibration or Press PRINT HOLD to enter the calibration program no CAL will be displayed initially when LOAD is displayed place the calibration weight on the centre of the pan The balance will tell you the value of the calibration wei
4. c switch off is disabled Press the F key to return to weighing Uni WEIGHING UNITS SELECTION This function allows the weighing units to be changed from Kilos to Pounds and back even when to scale is loaded To set the parameter press the F key to show uni Press PRINT HOLD to see the parameter Press PRINT HOLD to select either uni 1 or uni 2 Uni 1 Weighing in Kgs Uni 2 Weighing in Lbs Press the F key to return to weighing bL AUTOMATIC BACKLIGHT CONTROL This function may not be shown if the scale does not have a backlight lf the scale includes a backlight it can be either enabled or disabled To set the parameter press the F key to show bL Press PRINT HOLD to see the parameter Press PRINT HOLD to select either oL 0 or bL 1 bL 0 Backlight is enabled bL 1 Backlight is disabled Press the F key to return to weighing Adam Equipment 2002 4 POWER SUPPLY Power can be supplied by mains adapters delivering a regulated 6 to 9VDC 100 mA or using batteries inside of scale s head 6 x AA Size or by an external 6V lead acid battery Internal batteries and external lead acid batteries are not standard equipment CAUTION DO NOT use external power supplies and internal batteries together damage to the batteries may result If the batteries are low on charge BAT LO will be displayed The scale will be automatically switch off after 5 min The batteries should be replaced by new
5. elect the desired value Press the F key to return to weighing AVE 1 Fastest display rate AVE 4 Slowest display rate for uses such as an unsteady weighing or unstable environment HOLd SET METHOD OF DISPLAY HOLD The scale has 3 options for the HOLD function Press the PRINT HOLD key to change form HOLd 0 HOLd 1 to HOLd 2 Press The F key to set this option and return to weighing HOLd 0 Hold function is disabled HOLa 1 The display will be locked when the PRINT HOLD key is pressed The display will return to normal weighing when the PRINT HOLD key is pressed a second time or when the display is zeroed or tared with no weight on the scale HOLd 2 The display will be locked automatically when the weight is stable The display will return to normal weighing only when the display is zeroed or tared with no weight on the scale Adam Equipment 2002 3 T1 AUTOMATIC POWER SWITCH OFF The scale includes an internal battery The typical operation time using only the battery is 12 hours The scale includes the T1 function to turn the power off after 5 minutes if it is not being used This function can be disabled if the balance is powered from the mains power supply or if the disruption of the power might effect the weighing procedure To set the parameter press the F key to show T1 Press PRINT HOLD to see the parameter Press PRINT HOLD to select either T1 0 or T1 1 T1 0 Automatic switch off is enabled T1 1 Automati
6. f F set functions PRINT HOLD To hold display value or initiate the RS 232 transmission if included TARE Set the display to zero by storing the current weight in the tare memory ZERO Sets a zero condition when the platform is empty To weigh on the scale turn the scale on using the ON OFF key After the display is at zero place an item to be weighed on the platform The display will show the weight The unit will be shown in kilograms To zero the display press the TARE key In practice the beams are placed on the floor and an empty wheelchair or bed is rolled onto the beams The TARE key is pressed to zero the display lf necessary remove the wheelchair used to zero the display and then roll the wheelchair with the person to be weighed onto the beams The display will show the weight of the person in the wheelchair without the added weight of the wheelchair Likewise the bed weight will be tared without the person in the bed and then when the weight of the person is added only that weight will be shown DISPLAY SYMBOLS Symbol Description ZERO scale in auto zero range indication accurate zero STAB result is stable kg g scale in weighing mode BAT LO battery is weak after 5 minutes the scale will turn off p gt Scale in HOLD mode HOLD FUNCTION If the HOLD function has been enabled see below the display will hold the value unitl the user presses the PRINT HOLD key or ZERO key or TARE key The HOLD functio
7. ght to be used MCB 300 use 300kg MCB 400 use 400kg Even if the scale is set to display in pounds the weights shown above should be used for calibration Press the PRINT HOLD key CAL will be displayed initially then when UNLOAD is displayed remove the calibration weight The display will now show cALu 0 press the F then the display will show bL u 0 lf the scale has a backlight set bL u 0 otherwise set bL u 1 by pressing the PRINT HOLD key Press the F key to return to normal weighing Note The only permissible calibration weight is the value shown on the display immediately after LOAD To calibrate using any other weight the initial parameters will need to be reprogrammed using the AE106 service manual Adam Equipment 2002 6 Adam Equipment 2002 Manufacturer s Declaration of Conformity This product has been manufactured in accordance with the harmonised European standards following the provisions of the below stated directives Electro Magnetic Compatibility Directive 89 336 EEC Low Voltage Directive 73 23 EEC Adam Equipment Co Ltd Bond Avenue Denbigh East Estate Milton Keynes MK1 1SW United Kingdom FCC COMPLIANCE This equipment has been tested and found to comply with the limits for a Class A digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment
8. ilities including parts counting percent weighing etc Digital electronic scales for medical use Retail price computing scales ADAM EQUIPMENT and their distributors offer a full range of technical services such as on site and workshop repair preventative maintenance and calibration facilities ADAM EQUIPMENT operates an approved quality management system and is certified to ISO9002 Visit our Website at www adameguipment com Adam Eguipment Co Ltd Adam Eguipment Inc Adam Eguipment S A Pty Ltd Bond Avenue 26 Commerce Drive P O Box 1422 Milton Keynes Danbury CT Kempton Park 1620 MK1 1SW 06810 Johannesburg UK USA Republic of South Africa Phone 44 0 1908 274545 Phone 1 203 790 4774 Phone 27 0 11 974 9745 Fax 44 0 1908 641339 Fax 1 203 792 3406 Fax 27 0 11 392 2587 e mail sales adamequipment co uk e mail sales adamequipment com e mail sales adamequipment co za Copyright by Adam Equipment Co Ltd All rights reserved No part of this publication may be reprinted or translated in any form or by any means without the prior permission of Adam Equipment Adam Equipment reserves the right to make changes to the technology features specifications and design of the equipment without notice All information contained within this publication was to the best of our knowledge timely complete and accurate when issued However we are not responsible for misimpressions which may result form the reading of
9. n operates in 2 ways when enabled 1 gt The hold function will lock the displayed value when the user presses HOLD The value will continue to be shown untill HOLD is pressed a second time or the scale is zerod or tared 2 gt The HOLD funciton will operate automatically whe the weight is stable The user can unlock the display by pressing the PRINT HOLD key or zeroing or taring the scale Adam Equipment 2002 o FUNCTIONS To change functions press the F key The functions that are available can be cycled through by pressing the F key Not all functions may be enabled To use any ofthe functions see the corresponding section of the manual If you continue to press the F key you will eventually be returned to weighing Functions Available Function Deseipion 11 AUE Change Averaging of weighing result 3 tt Enable disable the automatic turn off after a period of not being used 4 Uni Changes weighing units from kgs to Lbs 5 bL Controls the LCD backlight if it is installed 0 When the function is a simple enable or disable then it is only necessary to select the function using the F key then press PRINT HOLD to view the current setting Pressing PRINT HOLD again will change the setting to the next value Pressing F again will return to normal weighing AVE DISPLAY AVERAGING RATE Select the required averaging for the display udpdate Press the F key until AVE is displayed Press PRINT HOLD to s
10. this material The latest version of this publication can be found on our Website ALYS Labware Lausanne Adam Eguipment 2002 Tel 021 312 42 60 Fax 021 312 42 61 labware alys technologies com