Corix PRO 70 Digital - Dentalaire Products

image

Contents

1. 29 7 Maintenance gt css rara 31 7 1 General Features 31 7 2 Arms Regulation a 31 8 System Calibration 33 9 Drawings uut eder eee a leale ascolta 35 10 Electrical SCHEMES 38 2 Corix PRO 70 Installation Manual 1 INTRODUCTION Corix PRO 70 manufactured by Coramex S A performs high quality intra oral radiographs ensured by the repeatability of examination combined with reduced exposure times and with the small focal spot Corix PRO 70 has a powerful and user friendly hand held controller that fits into the palm of your hand This manual is intended to assist the user and installer in the safe and efficient operation and installation of the equipment described 2 SAFETY INFORMATION This manual provides all the necessary information for the correct handling of the equipment as well as warnings related to risks associated to X ray generators Coramex S A shall not be responsible for use of the equipment different from what it has been designed for Any damage to the equipment the operator or the patient caused by incorrect installation and maintenance not compliant with the procedures contained in the relevant user s and installation manuals or by incorrect operation techniques Any mechan
2. HOLO3NNOO 2 PRO 70 Installation Manual 39 Coramex S A A Division of Corix Medical Systems 94 B Lauro Villar Mexico City D F 02440 Mexico Phone 011 52 555 394 1192 Fax 011 52 555 394 8120 Cat Part P141PRO R1 0 40 Corix PRO 70 Installation Manual
3. Z 140 N ul 120 amp 100 5 80 lt 60 T 40 20 0 50 100 150 200 250 300 30 400 450 500 TIME 18 PRO 70 Installation Manual EMISSION AND FILAMENT FEATURES COMPAGNIA ELECTTRONICA ITALIANA S R L TUBE MOD OX 70P RATING CHARTS 25 gt 20 4 8 15 EE E 4H H 72 70KV B ll 0 411 i 10 0 01 0 1 1 10 100 EXPOSURE TIME SEC EMISSION CHARACTERISTICS FILAMENT CARACTERISTICS 12 24 10 9 2 0 E 2 6 5 19 4 2 1 8 24 26 28 3 0 3 2 24 26 28 3 0 32 FILAMENT VOLTAGE FILAMENT VOLTAGE 4 2 DPM Timer Preset Exposure Times The following table of pre set exposure times in Seconds shows the rated exposure time for a nominal line voltage of 120V and the final corrected exposure time as a function of the line voltage correction factor and patient size for the minimum 109V and maximum 132V line voltage operating range 19 Corix PRO 70 Installation Manual Notes Suggested exposure times in Seconds for E type films Film speed Factory pre set for E type Films If set for Digital x ray C C D Sensor pre set exposure times are reduced by 1 3 e Corre
4. co Woo 110 pd 9 LX T amp q L NN T E 2050 178 2 8 S NN 9L SOW 350 Jn VW si VOW e o 99 6 MN l A p vi A q e p AA el 1 VW a HOW A r pou SSA arvo n vos 8 So 8 6 e EIN N 31 wa 288 ST San a Woo M 7 988 SWH NL noop s gg Li 9 3 woo sz VAN 91 ZAZ 2 8 AMI 99 ag 5 nos ot zz 3 logy 1 89 8717 NV T in s s MN 9 MT ec n 4 L ZH Wild LMS 9u noo OP Tar i o _ AAA xo P Ul ug 199 9H Sad SH 6 9031 u o A XOL CET Me cani DEAN 6H d di al 19 neo o b ii ppi i 109135 A PP vam eg ui ord D P oz om o AA 6 Ul eam eu j IL aa AS 29A AAA 4 S n AYH X OL QHOO 1109 A n da 6 or df lad 1031 t A 9 eem P LOIN SLOWLNOO PON GespueH Woo ino et
5. The timer supply plate hand remote control and connection cable assembly supplied by the manufacturer must be kept unadulterated Using different parts requires new system calibration Judgment of wall sturdiness is left to the installer Fixing bosses to be used for each type of wall are the following Concrete walls expansion bosses Wooden studs self threading screws Hollow bricks chemical bosses 5 1 Electrical Features The supply line must meet the following requirements Single phase mains voltage ground 120V 10 Line frequency 60Hz Absorbed current 10A 120V Apparent line resistance 0 20 max 120V The equipment must be wired to an electrical panel whose characteristics comply with the electrical norms in force in the country where it is installed A dedicated line protected by a 10A circuit breaker is recommended The general ground connection must be performed according to the norms in force Inadequate ground connection of the equipment may represent a hazard for the operator and or cause the electrical equipment to malfunction Maximum distance between electrical panel and supply terminal block varies according to the section of supply wires and is reported in following table 120 V 60 Hz Minimum Required Wire Size Wire Run Distance 12 AWG 4mm 25 Feet 7 5 Meters 50 Feet 15 Meters 10 AWG 6 3mm kkkkkkkkkkkkk kkkkkkkkkkkkk 75 Feet 22 5
6. 1 LABEL 3 MANUFACTURER CORAMEX S A LAURO VILLAR No 94 B 02440 MEXICO D F MEXICO MOD CORIX PRO 70 PART P103USV X RAY TUBE HOUSING ASSEMBLY WARNING THIS X RAY UNIT MAY BE DANGEROUS TO THE PATIENTAND OPERATOR UNLESS SAFE EXPOSURE FACTORS AND OPERATING INSTRUCTIONS ARE OBSERVED RATED PEAK TUBE POTENTIAL 70 79 ELECTRICAL SHOCK HAZARD DO NOT 3 RATED TUBE CURRENT 8mA 15 TOTAL FILTRATION 2mm AI SERIAL No XXXXX FOCAL SPOT 0 8mm MANUFATURED MMMMMMMM YYYY MEXICO X RAY TUBE C E l srl ITALY MOD OX 70 P S N XXXXX COMPLIES WITH DHHS PERFORMANCE STANDARD 21 CFR SUBCHAPTER J REMOVE PANELS RISK OF EXPLOSION DO NOT USE IN PRESENCE OF FLAMMABLE ANESTHETICS FOR CONTINUED PROTECTION AGAINST RISK OF FIRE REPLACE ONLY WITH SAME TYPE AND RATING OF FUSE LABEL 4 MANUFACTURER CORAMEX S A LAURO VILLAR 94 02440 MEXICO D F MEXICO LABEL 2 MANUFACTURER CORAMEX S A LAURO VILLAR No 94 B 02440 MEXICO D F MEXICO MOD CORIX PRO 70 X RAY CONTROL PART P101USV LINE VOLTAGE 120Vac 10 109 132Vac CURRENT MAX 12A 132Vac 50 60Hz CLASS 1 TYPE B SHORT TIME OPERATION SERIAL No XXXXX MANUFATURED MMMMMMMM YYYY MEXICO COMPLIES WITH DHHS PERFORMANCE STANDARD 21 CFR SUBCHAPTER J COMPLIES WITH UL 2606 1 MOD CORIX PRO 70 BEAM LIMITING DEVICE PART P104USV SOURCE TO SKIN DISTANCE FFD 20cm X RAY FIELD O AT MINIMUM FFD 6cm SERIAL No XXXXXX MANUFATUR
7. Radiation leakage at 1 m lt 50 mR h technical Factors 70Kvp 8mA 15 Corix PRO 70 Installation Manual Exposure Interval Duty Cycle 01 30 The minimum Exposure Interval between exposures 30 time units of cooling time for every time unit of exposure is a Preset value in the microprocessor controlled digital timer Model DPM X Ray Tube Part of the Tube Housing Assembly Manufacturer C E I S R L Model Designation OX 70 P Focal spot 0 8 mm IEC 336 Inherent Filtration 0 5 mm AI eq Beam Limiting Device Manufacturer Coramex S A Part Designation P104USV Minimum Focal Spot to Skin Distance 200 mm FFD X Ray Field at Minimum FFD 60mm Note Pre heating time is the time required by the tubehead to enable the correct radiation output When testing the timer the correct time is the displayed time plus the pre heating time For example it display reads 0 40s the correct time is 0 40s 0 15s 0 55s 17 Corix PRO 70 Installation Manual 4 1 Tube Characteristics X RAY TUBEHEAD CURVES THERMAL CURVES OF TUBE 200w 1 sow 110 80W N A Uu HEAT STORAGE KJ 0 1 2 3 4 5 COMPAGNIA ELECTTRONICA ITALIANA S R L TIME min TUBE MOD OX 70P COOLING CURVES OF TUBEHEAD 200 180 160
8. has been carried out the scissors arms holding pack can be removed 5 Lower cover on wedge and screw safety screw back NOTE The function of the cover is to avoid that wedge leaves the relevant seat Therefore cover must be held in the right position by means of the relevant holding screw All operations for extension arm friction regulation must be performed 26 Corix PRO 70 Installation Manual 6 5 Installation of Options Warning Coramex S A is not responsible for any damage to the equipment the operator or the patient caused by incorrect installation and maintenance not compliant with the procedures contained in the relevant user s and installation manuals or by incorrect operation techniques Single stud mount This mount is available for those cases where an installation requires the use of a limited surface The term single stud means that the stud must be at least two 2 x 4 or two 2 x 6 inch studs sandwiched together The installer is reminded that this feature must be carefully used since the manufacturer makes no claim whatsoever as to the fitness of this installation Most of the times this type of installation is used to mount onto a cabinet type structure or subdividing walls Installation to brick or concrete walls Use lead shields or chemical bosses to secure the device Follow proper shielding utilization procedures Always secure the device so as to make a solid fastening Sandy and hollow bricks may be danger
9. the UP and DOWN keys and the factor will be 1 00 once again 34 Corix PRO 70 Installation Manual 9 DRAWINGS CORIX PRO 70 WALL MOUNT EXTENSION ARM A TOTAL REACH x B TOTAL INSIDE REACH z C 4 4 EXTENCION ARM A TOTAL TOTAL INSIDE DISTANCE D REACH C 13 3 4 35 cm 25 3 4 65 4 cm 53 3 8 135 6 cm 27 3 8 69 4 cm ST 31 1 2 80 cm 43 1 2 110 5 cm 71 1 8 180 6 cm 45 114 4 cm 35 3 8 90 cm 47 1 2 120 6 cm 75 190 6 cm 49 124 4 cm 86 5 219 7 cm 57 1 8 145 cm DISTANCE D FLOOR LEVEL 35 Corix PRO 70 Installation Manual 70KVp 8mA Q CORIX MEDICAL SYSTEMS WARNING THIS X RAY UNIT MAY BE DANGEROUS TO PATIENT AND OPERATOR UNLESS SAFE EXPOSURE FACTORS AND OPERATING INSTRUCTIONS ARE OBSERVED BRACKET REMOTE TIMER KIT CORIX PRO 70 OPTIONAL P N P116USV 36 Corix PRO 70 Installation Manual Z 6Sd S NIAYEIO w SS exeo 2002 2018 L ILNG vos 04 XIHOO GOW 1N3WdInO3 AvH X V S X3aWwvHuoo 1 zi WWZESP 9 uuutp 90t yug Aa OL TULIDIO ASN 5078 02 gX1109 SWALSAS XIHOI 9 LE 4 101 2mm 7 13 16 199 2mm 4 3 4 120 2mm Corix PRO 70 Installation
10. Coramex S A Corix PRO 70 Digital VETERINARY X RAY EQUIPMENT Installation Manual Release 1 0 March 2005 CONTENTS 1 introduction Lisa 3 2 Safety Information 3 2 1 Warnings cool aa 3 3 DESerplon LDL 5 3 1 Identification Labels 5 3 2 Equipment Parts 11 3 3 Configurations 12 a Standard Configuration 12 b Remote Timer Configuration 13 c Mobile Stand Configuration 14 d Portable Stand Configuration 15 4 Technical Features lia 16 4 1 Tube Characteristics 18 4 2 DPM Timer Preset Exposure Times 19 5 Pre Installation nl 20 5 1 Electrical Features 21 6 Installation aiar 22 6 1 Wall Mobile and Portable Mounting Installation 22 6 2 Scissors Arm and Extension Arm Mounting 24 6 3 Arms Mounting on Support 25 6 4 Tubehead Mounting 26 6 5 Installation of Options 27 6 6 Electrical Connections 27 6 7 Final Functioning Tests
11. ED MMMMMMMM YYYY MEXICO COMPLIES WITH DHHS PERFORMANCE STANDARD 21 CFR SUBCHAPTER J LABEL LABEL 8b LABEL 5 MANUFACTURER CORAMEX S A LAURO VILLAR No 94 B 02440 MEXICO D F MEXICO EXTENSION ARM PART P105USV Serial XXXXX MMMMMMMM YYYY 70KVp 8mA A TO OPEN PUSH THE COVER EXPOSURE TIME IN SECONDS LABEL 6 v MANUFACTURER CORAMEX S A LAURO VILLAR No 94 B 02440 MEXICO D F MEXICO SCISSORS ARM PART P106USV Serial XXXXX LABEL 7 MANUFACTURER CORAMEX S A LAURO VILLAR No 94 B 02440 MEXICO D F MEXICO CORIX MEDICAL SYSTEMS THIS X RAY UNIT MAY BE DANGEROUS TO PATIENTAND Sy E MANUFACTURER CORAMEX S A LAURO VILLAR No 94 B 02440 MEXICO D F MEXICO CONTROL UNIT DPM Mod P102USV Serial MOBILE ARM PART P131USV Serial XXXXX MMMMMMMM Y Y Y Y REAR 8 Corix PRO 70 Installation Manual LABEL 9 Q CORIX MEDICAL SYSTEMS Corix Pro 70 WM DIGITAL VETERINARY X RAY SYSTEM 70 KVp 8mA WARNING THIS UNIT MAY DANGEROUS TO PATIENT AND OPERATOR UNLESS SAFE EXPOSURE FACTORS AND OPERATING INSTRUCTIONS ARE OBSERVE WHEN NOT IN USE TURN OFF THE POWER MAIN CONTROL UNIT MERE SWITCH POWER CONTROL HPA O SUPPORT FRONT PANEL LABEL FOR WALL MOUNT 9 Corix PRO 70 Installation Ma
12. ED s will light up with its corresponding timer numerical value visualized on the display 29 Corix PRO 70 Installation Manual 3 Select manual function by pressing the relevant keys and check that display visualizes the different manual exposure times The following test implies emission conditions Follow all applicable regulations and safety precautions 4 Position fluorescent screen for radiation visualization at extreme end of collimator press relevant X ray Switch key and check simultaneous switching on of LED X ray signaling and generation of acoustic signal accompanying x ray emission Keeping X ray Switch pressed the display will visualize the real exposure time Error Messages Error message LO on display means that mains input voltage is too low and not within the accepted limit of 10 Error message HI on display means that network input voltage is too high and exceeds the accepted limit of 10 IN case calculated exposure time is higher than 4s no exposure will be performed If these error messages are frequently displayed adjust input mains voltage as indicated in Section 8 Once these checks have been completed Equipment is ready for use 30 Corix PRO 70 Installation Manual 7 MAINTENANCE 7 1 General Features Corix PRO 70 requires correct use and also regular maintenance and checks Such measures will guarantee safe and effective equipment functioning and will prevent any risk for both patien
13. Manual 37 10 Electrical Schemes ScSd N ONIMVHG 2002 90 91 394 04 XIHOO GOW LNAWdINOA AM0Z A18IN3SSY SNISNOH Jan 9 anyo MI 042 Met Oo 91 120 god enpo LINA 10 1 00 OL 8809 1109 Y00PNI y p i MOL SLOWLNOO eu Qespuey LO3NNOO BY InGOW 92 20208 39IA3Q SIYNINWIL NIYIN NHO ANI TweLNAN o MOTISA N3389 vw 091 AcV02L 21088 2018 ny Yum jas puey Bnid 9 1 5 su og e HOLIMS LINN TOHLNO9 OL QHOO 100 Corix PRO 70 Installation Manual 38 928d N SNIMYNA 3 3 a or Z00Z 0 s d vO N W A3H 294 1300N TvLISIA SunSodxdAVUJcEE Ed Fe AQVYH 21 C31 l V S X3NVUOOD 1O313S 037 01 097 3091 dro 193739 1V1N3Q 6 C31 1 037 248491214 461
14. Meters For 120 V supply is recommended to use wires whose section is not lower than 4 mm 12 AWG 21 Corix PRO 70 Installation Manual For standard configuration and for mobile stand configuration the supply terminal block is the same as that of the Timer For remote configuration the Timer s supply terminal block is only a link between the electrical panel and the supply terminal block of the arms support For proper functioning the equipment must be installed in air conditioned environments having the following characteristics Relative humidity 50 75 not condensing Temperature 18 28C 6 INSTALLATION Warning Coramex S A is not responsible for any damage to the equipment the operator or the patient caused by incorrect installation and maintenance not compliant with the procedures contained in the relevant user s and installation manuals or by incorrect operation techniques 6 1 Wall mobile and portable mounting Installation a Wall mounting installation Wall Plate plus Timer Standard Configuration The installer must verify the consistency of the wall and must keep in mind that each set pin can carry a load of 200kg 440 pounds If the wall can support this weight expansion metal cast pins can be used If the fixing position has wooden studs screws of 8 x 40 mm can be used 22 Corix PRO 70 Installation Manual If the wall is not strong enough to support the weight of t
15. cted exposure times rounded to the nearest 1 100 of second This table will not show the added 0 15 seconds of preheating time for the x ray tube LINE VOLTAGE 120V 109V 132V LINE VOLTAGE 1 0 1 9 0 55 CORRECTION FACTOR PATIENTS SIZE MEDIUM SMALL MEDIUM SMALL MEDIUM SMALL UPPER JAW INCISOR 0 24 0 16 0 47 0 31 0 13 0 09 CUSPID 0 24 0 16 0 47 0 31 0 13 0 09 BICUSPID 0 30 0 20 0 57 0 38 0 16 0 11 MOLAR 0 33 0 22 0 63 0 42 0 18 0 12 LOWER JAW INCISOR 0 18 0 12 0 33 0 22 0 09 0 06 CUSPID 0 18 0 12 0 33 0 22 0 09 0 06 BICUSPID 0 21 0 14 0 41 0 21 0 12 0 08 MOLAR 0 24 0 16 0 47 0 31 0 13 0 09 PAW 0 24 0 16 0 47 0 31 0 13 0 09 5 PRE INSTALLATION Proper planning prior to installation is required There are three areas of concern before installation 1 Mounting structure 2 Reach of the tubehead 3 Electrical connections Warning The precision of the installation is the full responsibility of the installer Inappropriate installation of the equipment may cause it to drop from its support resulting in damage to individuals and materials near its range The manufacturer fully disclaims expressed or implied in this regards Different wall structures require different type of fasteners It is the responsibility of the installer to use the appropriate fasteners Always making sure that the installation is properly leveled 20 Corix PRO 70 Installation Manual
16. e value to the voltmeter use by selectively pressing the UP and DOWN keys Once the values correspond press the X Ray key to accept this new value At this point the displayed voltage value might begin to flicker or change according to the true voltage cycling followed by 0 00 Unit is now calibrated to exit this mode simply press any of the patient selection keys Timer Multiplication Factor Occasionally the end user may need to use timer settings other than those pre set at the factory For example when different types of film speeds are required The unit is capable of automatically varying these values to reflect the new time setting required The multiplication factor can be changed from a value of 1 31 to 0 69 At the factory is set for a value of 0 69 Values less than 1 00 will reduce the timer setting while those above will increase the time settings for the automatic keys a Procedure Power up the device while simultaneously pressing the Up and Down keys to enter the calibration mode The display will now show CAL and after 3 seconds the voltage reading Press the Tooth selection key the display will now show FAC To increase time press the UP key until the display indicates 1 31 to decrease time press the DOWN key until the display indicates 0 69 To exit this mode press the Patient selection key the factor will be save in the timer memory To cancel the factor start the procedure without pressing
17. he x ray device it will be necessary to use the optional 4 synthetic set pins 12 mm with bushings Check that all parts are present Check wall consistency and mark holes for wall plate mounting on wall in the selected position at a distance of 57 1450mm from floor Drill holes in wall and mount plate Make sure plate is leveled Secure wall plate to wall by using the appropriate screws Wall Plate Configuration with Remote Timer Follow as indicated in section 6 1 but install the optional remote timer mounting assembly in the area where the hand switch assembly will be located 1 Insert optional remote timer mounting on wall and bring connecting wire through the hole on the plate 2 Make sure wire strain relief is used This will help eliminate undue stress on the connecting cord assembly Control Box Assembly 1 2 Remove control box cover by loosening the relevant captive screws Secure control box to plate by using the screws and check that top plane is leveled Make sure power wires are brought through the internal hole and prepare for electrical installation Mobile stand mounting and Timer Installation 23 Corix PRO 70 Installation Manual 1 Make sure all parts for mobile mounting are present e Cross the two base legs and fix them together to the column Position the articulated arm do not remove safety belt into the column mak
18. ical and or electrical changes caused during or after installation different from those reported in the service manual Any expenses related to the eventual disposal of the equipment or parts 2 1 Warnings The equipment must be used in compliance with the procedures contained in the present manual and shall never be used for purposes different from those envisaged by it Only qualified service personnel are allowed to perform technical interventions on the equipment and to remove the tubehead from its support and access the internal components There is risk of injury if proper procedures are not used The tubehead cover or the relevant cone collimator assembly should not be touch during X ray emission No objects should be hanged on the extension arms 3 Corix PRO 70 Installation Manual The equipment is not designed to be used in the presence of flammable anesthetics oxygen or nitrous oxide Before performing any maintenance intervention the equipment must be disconnected from the input line voltage by means of the relevant magnetic thermal switch 4 Corix PRO 70 Installation Manual 3 DESCRIPTION 3 1 Identification labels LABELING CORIX PRO 70 WM WALL MOUNT 5 Corix PRO 70 Installation Manual LABELING CORIX PRO 70 MM MOBILE STAND 6 Corix PRO 70 Installation Manual LABELING CORIX PRO 70 PS PORTABLE STAND 10 Corix PRO 70 Installation Manual LABEL
19. ing appropriate wire connections Mount tubehead following section 6 4 Portable stand Installation ww sure all parts for portable stand mounting present 1 Position base on table to be used 2 Fix removable legs into the base 3 4 Mount tubehead following section 6 4 Position the arm into the column making appropriate wire connections 6 2 Scissors Arm and Extension Arm Mounting Assembly of extension arm and scissors arm Warning Do not remove safety belt from scissors arm Releasing the arm prior to mounting the tubehead may cause severe damage or injury Check that Extension Arm pivot is fitted with a spacer By means of tape put the scissors arm cable and extension arm traction wire together Pull wire until cable appears then separate cable from traction wire and introduce scissors arm pivot into extension arm 24 Corix PRO 70 Installation Manual Traction Wire Extension Arm Cable Arm Cover ES _ Scissors Friction to Shaft Calibration 6 3 Arms Mounting on Support Wall mounting of arms assembly NOTE To insert extension arm rotation pivot keep arm in orthogonal position with respect to plate Do not free scissors arms from safety belt 1 Mount complete extension arm on wall plate by inserting rotation pi
20. l and connect the timer remote with the PC board through a flat cable type telephonic of 4 threads with a not greater length to 10mts 30 feet with connectors RJ hand set in its extremes assure to do the cable according to the electric diagram 28 Corix PRO 70 Installation Manual 6 7 Final Functioning Tests DPM Timer Description GREEN LED 12 DIP SWITCH SW 1 70KVp 8mA AMBER LED 13 I UV GREEN LED 1 TO LED 9 9 ID 8 al 5 DB 2 E E 3 x W N GREEN LED 10 TOLED 11 M M CAUTION X RAYS D w 0 FRONT REAR CORIX PRO 70 INSTALLATION DPM TIMER HAND HELD CONTROLLER All equipment functions are set at standard values and are tested in the factory during final tests Some of the functions may be regulated by Service engineers only after installation has been completed or according to specific requirements After equipment has been connected to a dedicated line source perform the following functional tests 1 Set circuit breaker line switch to the ON position and verify that the front rocker switch on the control box is illuminated when it is set in the ON position Verify also that the Hand Control display shows a given value and that the relevant keys are illuminated 2 Check correct functioning in automatic mode by depressing relevant function selection keys in sequence Each of the relevant L
21. libration 1 Hand Control replacement Cord replacement Installation of remote control systems Main P C Board replacement a N When display indicates either LO low voltage or HI high voltage Calibration and adjustment of the DPM timer The DPM timer requires calibration and adjustment according to manufacturer s specifications for the nominal voltage of 120 volts This is necessary to enable the timer adjustment by the software algorithm to be made within a specified window or voltage range Any voltage exceeding this range will cause an error code such as LO or HI The timer is calibrated to the power supply at the factory any deviation will result in a new calibration requirement Calibration affects the relationship between power supply the length of the interconnecting cable between the hand control and the control box and the hardware associated with the timer and the main P C Board Qualified technicians must perform calibration a Test equipment Digital voltmeter true RMS preferable b Procedure 33 Corix PRO 70 Installation Manual Connect the leads of the voltmeter to the input terminals and verify the correct voltage 120 10 Power up the device while simultaneously pressing the Up and Down keys to enter the calibration mode The display will now show CAL and after 3 seconds the voltage reading Compare this value to the voltage reading of the voltmeter Calibrate the displayed voltag
22. nual LABEL 10 O CORIX MEDICAL SYSTEMS Corix Pro 70 MM DIGITAL VETERINARY X RAY SYSTEM 70 KVp 8mA WARNING THIS X RAY UNIT MAY BE DANGEROUS TO PATIENT AND OPERATOR UNLESS SAFE EXPOSURE FACTORS AND OPERATING INSTRUCTIONS ARE OBSERVED MAIN EU SWITCH POWER WHEN NOT IN USE TURN OFF THE POWER SWITCH AND LEAVE THE CONTROL UNIT HERE CONTROL 3 O Ho Ww FRONT PANEL LABEL FOR MOBILE STAND AND PORTABLE STAND 10 Corix PRO 70 Installation Manual 3 2 Equipment Parts A set of different models of the complete extraoral dental X Ray device share the certified components listed Model Corix PRO 70 Wall Mount Common Part X Ray Tube Housing Assembly Part P103USV Beam Limiting Device P104USV X Ray Control External Panel Part P101USV Control Unit Part P102USV Particular Part Extension Arm Standard 80cm Part P105USV Extension Arm Large 90cm Part P105LUSV Extension Arm Short 35cm Part P105SUSV Bracket Remote Timer Kit optional Part P116USV Scissors Arm Part P106USV Wall Plate Part P115USV Model Corix PRO 70 Mobile Stand Common Part X Ray Tube Housing Assembly Part P103USV Beam Limiting Device Part P104USV X Ray Control External Panel Part P101USV Control Unit Part P102USV Particular Part Scissors Arm Part P106USV Mobile Base Part P121USV Model Corix PRO 70 Portable Stand Common Part X Ray T
23. on 4 Once friction adjustment is achieved replace cover 31 Corix PRO 70 Installation Manual Balancing of arms 1 Observe arms to determine which one requires adjustment Anterior or Posterior 2 Position the articulated arm as shown in the figure 3 Locate and remove the cover plug of the arm that requires adjustment 4 Insert the hexagonal key and rotate clockwise if the arm tends to go down or rotate counter clockwise if arm tends to go up 5 Once adjustment is finished replace cover plug 1 2 Hexagonal Key Cover Plug Articulated Arm Anterior Cover Pluc Articulated Arm Posterior 32 Corix PRO 70 Installation Manual Tubehead replacement The scissors arm must be held closed together using a heavy duty strip This will avoid serious injury to personnel or assembler and limit the potential arm breakage 1 Remove screw from scissors tubehead arm holder 2 Hold Arm at top height hold the defective tubehead and slide sleeve upwards and hold it with a piece of tape 3 Remove the wedge securing the tubehead assembly and gently work the tubehead pivot post out Care should be exercised not to damage the arm or the connection 4 Proceed to install the replacement tubehead as instructed under section 6 4 8 SYSTEM CALIBRATION Corix PRO 70 with DPM timer is calibrated at the factory prior to shipment for the nominal voltage of 120 VAC However any of the following situations requires a new ca
24. ous Consider using all thread rods or bolts to go through walls and use another external fastener or clamp to hold baking in a secure place 6 6 Electrical Connections The device must be connected to a properly grounded power source Follow all applicable electrical regulations Use dedicated lines with the correct gauge and circuit breakers 27 Corix PRO 70 Installation Manual Electrical connections for standard versions Refer to section 5 1 for appropriate wire size 1 Connect power cables from the wall source to the terminal block of the control box by means of a bipolar cable ground whose minimum section must be 14 AWG crimp the terminals provided to the ends of each wire and insert them into the appropriate terminal on the connection block Labels indicate the following L Line black wire N Neutral white wire Ground symbol ground wire green cable Connect tubehead cable from arms to the main P C Board as follows Crimp the terminals with the appropriate connectors and connect them as indicated to the sliding connectors shown on the P C board The black wire to L live tubehead wire the white wire to N neutral tubehead wire and the ground wire green to a physical ground Electrical connection for versions equipped with remote Timer For this case the control unit timer DPM can be located in a remote place as indicates the section 3 3 b use the bracket remote timer to hold the control to the wal
25. t and operator Maintenance checks that can be performed by the operator and periodical maintenance interventions to be performed by Service engineers only are the following Regularity Test Type Annual Check that tags are intact and correctly secured Check that no oil traces are present on tubehead Check that hand remote control cable does not reveal signs of interruption or abrasion Check that equipment does not revel external signs of damage which could affect safety protection from x ray Check scissors arm balancing Every two years Check x ray beam centering Check tube current absorption 7 2 Arms Regulation Arms regulations do not require removal of tubehead In case this operation is considered useful or necessary before removing tubehead bring scissors arm in closed position and secure it with safety fixing device This operation is necessary to avoid damage to people and to the arm Arms regulation may be necessary in the following cases Scissors arm is not perfectly balanced in this case operate on spring regulation After a certain time arms balancing springs may sag Should this happen tubehead will no longer be balanced in all positions and spring calibration will be required Extension arm friction regulation 1 Locate and remove the plastic cover at the end of the arm 2 Access the friction screw behind this cover 3 Regulate friction by means of a hexagonal wrench and check arm rotati
26. tic X Ray Imaging General Purpose Dental Manufacturer Coramex S A Lauro Villar 94 B Mexico D F 02440 Mexico Model Designation Corix PRO 70 Class Itype B Rated Line voltage 120 V 10 109 132 Vac Line frequency 50 60 Hz Line current 10 A max 120 Vac Power consumption 1 050 KW max 132Vac Apparent line resistance 0 2 Ohms max 120 V Line Voltage regulation lt 3 Main Fuse IOAF X Ray Control Microprocessor Controlled Digital Timer Manufacturer Coramex S A Part Designation P101USV panel with Main Terminals Device TIMER DPM Timer Control Unit Part Designation P102USV Hand Held Controller Exposure Times Manual time selection from 0 03 s to 3 00 5 in steps of 0 01 s Plus 9 pre set exposure times with automatic line voltage compensation A pre heating time of 0 15 s for loading the x ray tube must be added to the time selection shown on the display Timer Accuracy t 10 Tube Housing Assembly Extra oral Diagnostic X ray generator and Beam Limiting Device Manufacturer Coramex S A Part Designation P103USV Rated output voltage 70 kVp 7 Single phase self rectifying Rated output current 8 mA 15 120Vac Maximum deviation of output current 4 5 mA over the voltage range Total filtration 2 mm AI eq Transformer insulation Oil Bath Cooling Thermal Convection Cooling
27. ube Housing Assembly Part P103USV Beam Limiting Device Part P104USV X Ray Control External Panel Part P101USV Control Unit Part P102USV Particular Part Mobile Arm Part P131USV Part Portable Base Part P132USV 11 Corix PRO 70 Installation Manual es X un 9 jouyuoo X p wwe uOISUS X3 Ule SIOSSIOS Z peeueqn LINNON TIVAA WM 02 Od PRO 70 Installation Manual 12 Aey X pun S X wue SJOSSIOS Z peeueqn YAWIL 3 LOIN3H LNNOW TIVM WM 02 O ld jeuoado PRO 70 Installation Manual 13 c Mobile Stand Configuration CORIX PRO 70 MM MOBILE STAND 5 Control unit with X Ray button 1 Tubehead 2 Scissors arm 3 X ray Control 4 Mobile Stand 16 h gt gt S S XN NN lt my Ti TIT N 14 Corix PRO 70 Installation Manual d Mobie Stand Configuration CORIX PRO 70 PS PORTABLE STAND DT a 2 E 2 Mobile arm 3 X ray Control 4 Portable Stand 15 5 Control unit with X Ray button Corix PRO 70 Installation Manual 4 TECHNICAL FEATURES Technical features Equipment Intra oral Diagnos
28. vot in the relevant thimble 2 Check that the extension arm this leveled through a level the plate of wall should be leveled horizontal and vertical if is necessary can wear the plate against the wall to obtain the level desired NOTE In this phase extension arm must not support tubehead weight it is recommended to keep the angle slightly wider than 90 degrees this allowing a full load flexion of about 4 mm with a 900 mm extension arm 25 Corix PRO 70 Installation Manual Stand mounting of arms assembly NOTE To insert extension arm rotation pivot in thimble keep arm in orthogonal position with respect to stand Do not free scissors arms from safety belt 1 Mount scissors arm extension arm is absent in this configuration on stand by inserting rotation pivot in the relevant thimble 2 Check that extension arm is leveled Portable mounting of arms assembly 1 Mount articulated arm on column 6 4 Tubehead Mounting nm 1 Remove safety screw located on joint 2 Keep arm articulation at maximum height and slide safety cover over the connection post of the scissors arm 3 Insert tubehead connection pivot post into the scissors arm post assembly about half way Slide in the wedge into the groove on the pivot post and then push the head upwards until the wedge sits itself into the groove of the pivot post 4 Completely insert rotation pivot so that wedge fits the relevant slots on the pivot Only after this