Service Manual RW Series
Contents
1. METRIC VITON O RING METRIC SILICONE O RING WASHER FACE SEAL NEOPRENE 70 DURO O RING 2115 FLOTEC PART 112 1156 008 810 6490 002 810 6490 102 810 6039 003 810 6428 001 810 6033 021 810 6293 001 810 6714 011 810 6496 001 810 6496 022 810 6672 021 010 6206 031 010 6206 041 010 6207 001 HB12 HBD2 810 6568 003 810 6634 003 241 0400 100 810 6737 003 810 6236 001 210 6022 015 210 6028 006 210 6028 019 810 0562 01 810 0562 001 810 0495 001 210 6022 022 SILICONE 70 DURO O RING 210 6028 010 O RING EPDM 70 DURO 220 6036 006 O RING EPDM 70 DURO 220 6036 015 O RING EPDM 70 DURO 220 6036 019 O RING EPDM 70 DURO 220 6036 022 STAINLESS STEEL DETENT SPRING 151 0001 001 STAINLESS STEEL DETENT BALL 553 0001 003 DOME HANDLE DH209 CHEMETRON STYLE QD TO 1 8 MNPT QDC2 12 OHIO STYLE QD TO 1 8 MNPT QDO2 12 DISS 1040 FITTING 1 8 MNPT DF 1040 DISS 1060 FITTING 1 8 MNPT DF 1060 DISS 1120 FITTING 1 8 MNPT DF 1120 DISS 1160 FITTING 1 8 MNPT DF 1160 DISS 1180 FITTING 1 8 MNPT DF 1180 DISS 1200 FITTING 1 8 MNPT DF 1200 DISS 1220 FITTING 1 8 MNPT DF 1220 DISS 1240 FITTING 1 8 MNPT DF 1240 DISS 1040 CHECK VALVE 1 8 MNPT DCV 1040 DISS 1060 CHECK VALVE 1 8 MNPT DCV 1060 DISS 1120 CHECK VALVE 1 8 MNPT DCV 1120 DISS 1160 CHECK VALVE 1 8 MNPT DCV 1160 DISS 1180 CHECK VALVE 1 8 MNPT DCV 1180 DISS 1220 CHECK VALVE 1 8 MNPT DCV 1220 DISS 1240 CHECK VALVE 1 8 MNPT DCV 1240 GAUG
2. REPAIR OF FLOTEC MEDICAL GAS REGULATORS 1 2 Unscrew the flow meter sub assembly and remove the piston manifold assembly from the base of the regulator Also remove the CGA washer unscrew the retainer and remove the valve seat insert taking care to not scratch the tip of the valve seat insert Take the valve seat insert and place a small green viton o ring part 241 0400 001 on the side of the insert that already contains an o ring Put the valve seat insert back into the hole you removed it from in the same orientation in which you found it Also screw in the retainer and place the CGA washer back onto the retainer making sure to tighten the retainer to somewhere between 12 and 15 ft Ib this can be accomplished by using a model ES70T screwdriver at scale 5 Reassemble the piston manifold assembly and screw in the flowmeter sub assembly making sure to torque the flowmeter sub assembly from 15 to 20 ft Ib 2 DISS AUXILIARY OUTLETS a REMOVAL 1 Using a 5 8 open end wrench remove 13155 auxiliary outlet from regulator 2 Remove old sealant from the male and female threads CAUTION Hold the regulator with the DISS auxiliary outlet ports down so that any particulate matter will fall out of the port b SERVICE 1 Disassemble the DISS auxiliary outlet by depressing the check valve plunger until the opposite end containing the o ring clears the body Remove the o ring to gain access to the plunger and sp
3. knob shaft for tightness Torque rotor assembly using model ES70T electric screwdriver scale 9 Any other service work must be done by the manufacturer there are no user serviceable parts in the flowmeter assembly If the flowmeter needs to be returned to the manufacturer reassemble the flowmeter with the upper body before returning CAUTION The flow control valve is manufactured to exacting tolerances and utilizes number of extremely small orifices To prevent improper valve operation due to contaminants ensure hands tools and worktable are clean when disassembling and servicing the flow control valve There are no user serviceable parts in the flowmeter assembly Clean all flow control valve components with isopropyl alcohol and thoroughly blow dry CAUTION Do not allow a lubricant film applied on the rotor face to accumulate in the rotor orifices WARNING Do not interchange rotors and flow control valve bodies Always assemble rotor with its original flow control valve body Verify that the valve shaft rotates smoothly and stops at each detented position c INSTALLATION 1 2 Replace o ring on flowmeter assembly as needed Thread the flowmeter into the upper body and turn clockwise to tighten Using a vise on the regulator body and a strap wrench on the flowmeter VIII assembly for Yoke style regulators or two strap wrenches on Nut amp Nipple style regulators turn the flowmeter clockwise to tighten to 10 15 f
4. psig 396 Boom Box Leak detector Nupro Snoop or equivalent oxygen compatible C SERVICE MATERIALS DuPont Krytox performance lubricants GPL206 Teflon tape 4 in 6 mm Isopropyl alcohol Cloth lint free Oxygen tubing in 6 mm I D Screw 6 32 UNC Service Instructions Work Instruction WI PRO 010 General Assembly Procedures for RW and Exhibit A SPECIFICATIONS Flotec Medical Gas Regulators are small lightweight devices that are available in a variety of configurations to meet the various requirements for medical gas delivery Flotec Medical Gas Regulators can be equipped with 12 position detented flow control valve or variety of designated inlet flow connectors Units are also available with single or dual high flow DISS outlets Both small cylinder and large cylinder inlet connections are available on all regulators The following instructions provide information to assist qualified personnel in the service and repair of Flotec Medical Gas Regulators It is recommended that the repair technician read these service and repair instructions thoroughly before servicing Flotec Regulators Storage Temperature 30 degrees F to 140 degrees F Operating Temperature 20 degrees F to 150 degrees F Materials Brass Bronze Anodized Aluminum Stainless Steel Teflon Neoprene Silicone Viton EPDM Filtration 20 micron Supply Pressure Range 250 to 3000 PSI Operating Press
5. test gauge stabilizes at 50 or 5 PSI E FLOWRATE TEST NOTE A number of variables including ambient atmospheric conditions and test instrumentation resolution and accuracy will affect the results of flowrate testing Compensation must be made for deviation from ambient conditions of 70 degrees F and 29 3 inches HG at the location of the test 1 Connect regulator to full gas cylinder and SLOWLY open cylinder valve Connect the flow control barbed outlet connector the designated flow connector or the flowmeter outlet to the test flowmeter using a suitable length of oxygen tubing Verify that the flows read on the test flowmeter are within or 10 of the flow set on the regulator For the pediatric RW or FM flow from 0 02 LPM to 0 5 LPM the tolerance of the flow should be or 20 10 IX REPLACEMENT PARTS LIST FOR RR8xxx xxxxxx AND RR5xxx xxxxxx DESCRIPTION PIN DOWEL RETAINER ORIFICE BRASS THREAD INSERT THREAD INSERT Helicoil FILTER DISC new WASHER CGA VALVE SEAT INSERT w seals SPRING 22 PSI SPRING 50 PSI SPRING 70 PSI ALUMINUMPISTON SUB ASSY w seals BRASS PISTON SUB ASSY w seals MANIFLOD SUB ASSY w seals HOSE BARBS TO 1 8 MNPT RESIDENT NIPPLE ADAPTER LOW PROFILE T HANDLE standard LARGE T HANDLE VITON 80 DURO O RING 4mm x 1mm VITON 90 DURO O RING 2 5mm x 1mm SEAL ORIFICE PISTON NEOPRENE 70 DURO O RING SILICONE 70 DURO O RING SILICONE 70 DURO O RING
6. E AND BOOT ASSEMBLY 010 6100 081 GAUGE AND BOOT ASSEMBLY 010 6100 330 to 333 GAUGE AND BOOT ASSEMBLY 010 6100 335 to 341 12 TROUBLESHOOTING GUIDE CONTENTS GAUGE SYMPTOM Pointer does not move after cylinder valve is turned on Reading drops sharply with flow through regulator THE REGULATOR LEAKS AT Vent holes in body Auxiliary DISS outlet Adjustable flow outlet with knob in off position Designated flow connector Threaded NPT connections THERE IS NO FLOW OR INACCURATE FLOW AT THE METERED FLOW OUTLET PROBABLE CAUSE Cylinder is empty Regulator inlet filter is blocked Contents gauge is damaged Cylinder valve cracked open inlet filter partially occluded Particulate matter in regulator Piston O ring leaks Piston seal leaks Relief valve gasket leaks Check valve O ring leaks Flowmeter not functioning properly Connector not properly tightened O ring leaks Loose connections Regulator inlet filter is blocked Flowmeter flow control is turned off Flowmeter not functioning properly Flow control valve is obstructed Flow control knob is incorrectly positioned on shaft 13 ACTION Replace cylinder Replace inlet filter Replace contents gauge Fully open cylinder valve Replace inlet filter Clean regulator internal components Replace piston O rings 2 Replace piston seal Replace relief valve gasket Replace check valve O ring Return to Flotec Tight
7. FLOTEC INC 7625 West New York Street Indianapolis IN 46214 4911 317 273 6960 Fax 317 273 6979 e mail flotec floteco2 com web site http www floteco2 com Service Manual RW Series M Service Manual RW Series Regulator 2014 07 31 docx pdf TABLE OF CONTENTS I TOOLS TEST EQUIPMENT AND SERVICE MATERIAL A TOOLS B TEST EQUIPMENT C SERVICE MATERIALS SPECIFICATIONS 0 THEORY OF OPERATION IV DEFINITION OF STATEMENTS V PRODUCT CLASSIFICATION DATABASE VI INSTALLATION INSTRUCTIONS VII SERVICE AND REPAIR A CLEANING LUBRICATION AND SEALING 1 INLET CONNECTIONS a REMOVABLE CGA OR INTERNATIONAL CONNECTION b INSTALLATION CGA OR INTERNATIONAL CONNECTION 2 DISS AUXILIARY OUTLETS a REMOVAL b SERVICE c INSTALLATION 3 CONTENTS GAUGE a REMOVAL b INSTALLATION 4 FLOWMETER a REMOVAL b SERVICE c INSTALLATION 5 REGULATOR INTERNAL COMPONENTS a REMOVAL b SERVICE c INSTALLATION VIII TESTING CONTENTS GAUGE TEST LEAK TEST STATIC PRESSURE TEST DYNAMIC PRESSURE TEST FLOWRATE TEST oos IX REPLACEMENT PARTS LIST FOR RR8xxx xxxxxx AND RR5xxx xxxxxx X TROUBLESHOOTING GUIDE XI DRAWINGS a RXAXX XXXXX CGA 500 RW REGULATOR W NUT AND NIPPLE b RXCXX XXXXX CGA 320 RW REGULATOR WITH NUT AND NIPPLE c RXDXX XXXXX CGA 880 CGA 930 CGA 960 CGA 965 RW REGULATOR mo EB n e o F B r m RXEXX XXXXX RXGXX XXXXX RXNXX XXXXX RXRXX XXXXX KNURLED NUT RXSXX XXXXX STEEL
8. LTER DISC FLOTEC P N 810 6033 021 D O RING FLOTEC P N 241 0400 001 FLOTEC P N 810 6495 001 1 O RING FLOTEC P N 810 6737 003 CONFIDENTIAL 1999 14 FLOTEC Inc FLOTEC DATE REVISION RECORD ECN 8 08 CHK 6 11 RELEASE amp ISSUE fen o2ej upal 5 gt 7 31 14 WAS 810 6737 002 ADD P N 241 0400 100 E14 061 JDA J 3 PTFE WASHERS FLOTEC P N 810 6495 001 PE D 3 S S DETENT BALLS FLOTEC P N 553 0001 003 2 PTFE WASHERS 3 O RINGS FLOTEC P N 810 6562 001 4 O RING FLOTEC P N 810 6562 V01 3 S S DETENT SPRINGS D O RING FLOTEC FLOTEC P N 151 0001 001 P N 210 6022 022 D O RING FLOTEC P N 210 6028 010 ENLARGED VIEW OF FLOWMETER ASSEMBLY PTFE WASHER P N 810 6495 001 ON O RING P N 810 6562 001 O RING P N 210 6022 022 O RING P N 810 6562 V01 ON TOP O RING P N 810 6562 001 ON BOTTOM D O RING FLOTEC P N 210 6028 019 SECTION D O RING FLOTEC SE P N 210 6028 006 ENLARGED FLOTEC 870 RW SERIES REGULATOR AND FLOWMETER ASSEMBLY WITH YOKE STYLE INLET Lo J SHEET DRAWING NUMBER REV RW SERIES CGA 870 COMPONENT PART NUMBERS lor Rw REG PARTS CGA 870
9. NIPPLE ASSY RX1XX XXXXX RX5XX XXXXX RX7XX XXXXX RX8XX XXXXX RX8XX S XXXXX XXXXX TEST PROJECT MODULATED FLOW RW REG PARTS RW REG PARTS BS 341 3 REGULATOR WITH NUT AND NIPPLE DIN 477 9 RW REGULATOR WITH NUT AND NIPPLE CGA 660 NITRIC OXIDE REGULATOR CGA 626 REGULATOR W CHROME PLATED CGA 626 REGULATOR W BRASS NUT STAINLESS CGA 580 REGULATOR WITH NUT AND NIPPLE CGA 540 REGULATOR WITH NUT AND NIPPLE CGA 940 AND CGA 973 RW REGULATOR YOKE ASSY CGA 870 YOKE UNIBODY OXYGEN REGULATOR RW REGULATOR W OXYSAVRR FLOWMETER SUB ASSEMBLY STATIC AND DYNAMIC TEST PROCEDURE MANIFOLD IS CLOSED AND OPENED CGA 870 COMPONENT PART NUMBERS CGA 540 COMPONENT PART NUMBERS TOOLS TEST EQUIPMENT AND SERVICE MATERIAL A TOOLS Screwdriver 1 4 in 6 mm flat blade Screwdriver No 2 Phillips Open end wrenches 7 16 in 9 16 in 5 8 in 11 16 in Adjustable wrench 10 in 25 cm Hex key wrench 1 16 in Internal retaining ring pliers Punch 1 8 in 3 mm Hammer C Clamp 3 in 8 cm Dental Pick or similar pointed instrument B TEST EQUIPMENT Test flowmeter with an accuracy 0 or 290 if reading 70 degrees 21 degrees C and 14 7 PSI 101 kPa Test pressure gauge with a range of 0 100 psi 0 690 kPa and accuracy in according to ASME 40 1 1998 Grade permissible error 96 of span lower 4 of scale 0 25 psig 2 390 middle 12 of scale 25 75 psig 2 upper 34 of scale 75 100
10. OWMETER installation NOTE References made to the flow control valve body are also applicable to the regulator end cap TESTING Perform the following test procedures after servicing the regulator or to determine if a regulator problem exists WARNING If service to the internal components of the regulator has been performed conduct a preliminary static pressure test per the STATIC PRESSURE TEST instructions with a medical gas cylinder pressure of approximately 500 PSI Conduct the preliminary static pressure test before beginning any functional testing to verify that all regulator internal components have been installed correctly NOTE functional testing should be performed using a full cylinder of gas at 1900 to 2100 PSI A CONTENTS GAUGE TEST 2 3 With the regulator unpressurized verify that contents gauge pointer is at or below the red REFILL area Connect the Medical gas regulator to a full gas cylinder and SLOWLY open the cylinder valve Verify that the contents gauge pointer moves into the green FULL area B LEAK TEST WARNING Do not attempt to repair leaks while the regulator is pressurized 1 3 Connect regulator to an appropriate cylinder or Boom Box Post valve On regulators equipped with a flowmeter set the flow to the off position Apply compatible leak test solution to all outlets fittings and the junction of the contents gauge and regulator body SLOWLY open the cylin
11. The piston assembly then moves toward the seat and the seal blocks the flow path and prevents the pressure from rising above 50 PSI This all occurs in a fraction of a second When gas is let out of the output cavity through the flow controlling device the pressure in the cavity drops The spring then pushes the piston assembly away from the seat to allow the oxygen to restore the pressure to 50 PSI and deliver the flow through the flow controlling device at prescribed rate This too happens in a fraction of a second In the unlikely event that the regulator does not regulate properly and should the internal pressure rise above 75 PSI or 95 PSI the relief valve will unseat and safely release gas through the vent holes in the side of the body NOTE The Flotec Medical Gas adjustable flow regulator is shown in enclosed drawings The internal regulator components are common to all Flotec Oxygen Regulators and Pressure Reducers IV DEFINITION OF STATEMENTS Statements in this manual proceeded by the following words are of special significance WARNING means there is the possibility of injury or death to you or others CAUTION means there is the possibility of damage to the unit or other property NOTE indicates points of particular interest for more efficient and convenient operation Always disassemble inspect clean and repair regulator components in accordance with these instructions Be aware of all the potential hazards associated with han
12. V01 1 O RING FLOTEC P N 210 6022 022 1 O RING FLOTEC O RING P N P N 210 6028 019 210 6022 022 D O RING FLOTEC P N 210 6028 006 FLOTEC OXYGEN CGA 540 RW SERIES REGULATOR AND FLOWMETER ASSEMBLY WITH NUT AND NIPPLE STYLE INLET CONNECTION RW REG CGA 540 COMPONENT PART NUMBERS REVISION RECORD ECN 8 188 RELEASE amp ISSUE E11 026 JDA vas 810 6737 002 ADD P N 241 0400 100 14 06 3 PTFE WASHERS FLOTEC P N 810 6495 001 3 S S DETENT BALLS FLOTEC P N 553 0001 003 3 5 8 DETENT SPRINGS FLOTEC P N 151 0001 001 D O RING FLOTEC 7 210 6028 010 ENLARGED VIEW FLOWMETER ASSEMBLY PTFE WASHER P N 810 6495 001 ON TOP O RING P N 810 6562 001 ON 1 BOTTOM O RING P N 810 6562 V01 ON TOP O RING P N 810 6562 001 ON BOTTOM A A SECTION ENLARGED REFERENCE amp RELATED DWGS CAD 8 E CAD FILE 8 RW REG COMPONENTS CGAS40B ASSY 8 DRAWING NUMBER REV 1 1 RW REG PARTS CGA 540 XD CGA WASHER FLOTEC P N 810 6293 001 4 PTFE SEAL FLDTEC P N 810 6236 001 D O RING FLOTEC P N 210 6022 015 FOR 17 TO 58 PSI REGS ONLY ENLARGED VIEW OF PRESSURE RELIEF ASSEMBLY INDIANAPOLIS IN USA ALL RIGHTS RESERVED D FI
13. der valve or Boom Box Post valve Constant bubbling of the leak test solution denotes the presence of a leak Tighten fittings as required to eliminate all external leaks DO NOT overtighten threaded connections Close cylinder valve and note the pressure indicated on the contents gauge This pressure should not decay more than 100 PSI in 90 seconds or less If the regulator decays more than 100 PSI return to factory for repair C STATIC PRESSURE TEST 1 Connect test pressure gauge to the flow control outlet flowmeter outlet or designated flow connector Turn the flow control knob or flowmeter knob where applicable to the largest flow setting On emergency oxygen regulators connect the test pressure gauge to one of the DISS auxiliary outlets Connect the regulator to a medical gas cylinder and SLOWLY open the cylinder valve Verify that the test gauge pointer stabilizes and becomes steady at between 45 PSI and 65 PSI NOTE Steady is defined as no detectable movement of the test gauge pointer for a minimum of 15 seconds D DYNAMIC PRESSURE TEST 4 Connect test pressure gauge to one of the DISS auxiliary outlets Connect the regulator to an oxygen cylinder and SLOWLY open the cylinder valve Set a flow on the regulator flow control or flowmeter On emergency regulator models slightly depress the second auxiliary outlet check valve plunger to establish a small flow of gas from the outlet Verify that the pressure
14. dling and using high pressure gas equipment Also be aware that the possibility of fire exists when the combination of a combustible material a source of ignition and oxygen is present called the Fire Triangle WARNING Replacement of parts on high pressure regulators should be made only by qualified personnel familiar with their operation Do no remove or install parts with the regulator installed on a cylinder Use only the proper repair tools and parts Always wear eye protection when servicing high pressure regulators WARNING Contaminants or hydrocarbons in the presence of an ignition source and oxygen may combine and burn violently Never permit oil grease or other combustible substances to come in contact with oxygen cylinders regulator parts or repair tools Provide a clean oil free surface on which to place disassembled regulator parts WARNING Always open high pressure oxygen cylinder valves SLOWLY This should always be done in order to allow the heat of gas compression to dissipate Always verify that the oxygen cylinder valve is fully closed clockwise before disconnecting the regulator from the cylinder WARNING Do not obstruct the vent holes in the body of the regulator These holes must be open to atmosphere for proper regulator operation WARNING Flotec Regulators with yoke are equipped with a rubber and metal inlet seal washer Never use a non Flotec inlet seal washer Always use only one inlet seal washer and only this s
15. en connector Replace connector O ring Disassemble clean and reseal threads reassemble and tighten Replace inlet filter Turn flowmeter flow control to desired setting Refer to Flowmeter service Instructions Remove flow control valve from regulator and backflush Align knob set screw with flat on shaft and tighten securely THERE IS NO FLOW THE AUXILIARY DISS OUTLETS Designated flow connector filter screen dirty Cylinder is empty Regulator inlet filter is blocked Hose connector not properly engaged 4 Backflush connector replace connector Replace Cylinder Replace inlet filter Re engage and tighten hose connector MODULATING PISTON IN SHUT OFF POSITION MODULATING PISTON SHUT OFF RELIEF VALVE MANIFOLD OPEN 16 D FILTER DISC FLOTEC P N 810 6033 021 210 6024 011 MEN Rt i EUN K D O RING FLOTEC P N O 4 O RING FLOTEC ENLARGED VIEW OF CONNECTOR ASSEMBLY 1 PTFE SEAL FLOTEC P N 810 6236 001 FOR 17 58 PSI REGS ONLY D O RING FLOTEC P N 210 6022 015 lt ENLARGED VIEW OF PRESSURE RELIEF ASSEMBLY CONFIDENTIAL 1999 14 FLOTEC 22 INDIANAPOLIS IN USA ALL RIGHTS RESERVED 4 6 1 a B 7 31 14 1 O RING FLOTEC P N 810 6737 003 2 PTFE WASHERS FLOTEC P N 810 6495 001 3 O RINGS FLOTEC P N 810 6562 001 4 O RING FLOTEC P N 810 6562
16. move any residual fibers O ring seals that have been cleaned or are new replacements should be lubricated with a film of DuPont Krytox performance lubricants GPL206 Whenever pipe threaded NPT connections must be disassembled reseal the threads with Teflon tape When applying teflon tape remove old sealant from the male and female threads and wrap the male threads with Teflon tape in a clockwise direction starting 1 to 2 threads back from the end 1 INLET CONNECTIONS a REMOVABLE CGA OR INTERNATIONAL CONNECTION 1 Using a strap wrench on the regulator body and an appropriately sized open end wrench on the nipple tailpiece or yoke assembly remove the assembly from 4 female NPT port 2 Remove old sealant from the male and female threads CAUTION Hold the regulator with the inlet port down so that any particulate matter will fall out of the port 3 If the inlet filter replacement is required tap the old filter out of the tailpiece with inserted in the threaded end of the nipple or tailpiece b INSTALLATION CGA OR INTERNATIONAL CONNECTION 1 2 Install a replacement inlet filter in the nipple or retainer or yoke assembly and secure in place using a 5 16 rod and a light tap of a ball peen hammer Reseal the male threads on the nipple retainer or yoke assembly with Teflon tape Always put Teflon tape after the first full thread of the 14 male NPT threads Use wrenches to tighten the assembly c
17. ring 2 Inspect the components for wear or damage and replace as necessary Reassemble in reverse order c INSTALLATION l 2 Reseal the male threads on the 13155 auxiliary outlet with tape Thread the outlet into the regulator body and tighten 3 CONTENTS GAUGE a REMOVAL Using 8 7 16 wrench remove contents gauge from regulator body Remove old sealant from the male and female threads CAUTION Hold the regulator with the contents gauge port down so that any particulate matter will fall out of the port b INSTALLATION 1 Reseal the male threads on the contents gauge with Teflon tape 2 Thread the contents gauge into the regulator body and tighten The word OXYGEN on the gauge dial should be parallel to the regulator body FLOWMETER Before attempting to disassemble the flow control valve to resolve flow related problem perform the following procedures for back flushing the valve to remove any debris from the valve orifices NOTE The flow control valve must be removed from the regulator before back flushing REMOVAL 1 Using 8 vise on regulator body and strap wrench on flowmeter assembly for Yoke style regulators or two strap wrenches on Nut amp Nipple style regulators turn the flowmeter counter clockwise to loosen b SERVICE 1 Backflush 11 flow positions with medical gas Check holding the rotor assembly onto the
18. t Ibs 5 REGULATOR INTERNAL COMPONENTS a REMOVAL 1 Remove either the flowmeter or the end cap by following instructions under FLOWMETER removal NOTE References made to the flow control valve body are also applicable to the regulator end cap 2 Ifthe regulator has a removable inlet connection remove the connection from the regulator per INLET CONNECTIONS removable CGA or international connection b SERVICE 1 Remove the large o ring from the piston for inspection Clean the piston with isopropyl alcohol and thoroughly blow dry Inspect both O rings for wear or damage Replace as necessary with new O rings that have been lubricated with a film of Dupont Krytox performance lubricants GPL206 Inspect the seal on the end of the piston for cuts or abrasions WARNING Failure to install the piston seal will render the regulator inoperative and will result in uncontrolled high flows through the relief valve vent ports 2 Clean the relief valve with isopropyl alcohol and thoroughly blow dry Inspect the relief valve o ring for cuts and wear Replace the o ring if it is damaged 3 Clean the inside of the regulator body with isopropyl alcohol and a lint free cloth and thoroughly blow dry Inspect the seat for damage Apply a film of Dupont Krytox performance lubricants GPL206 approximately 12 wide to the inner surface of the regulator body c INSTALLATION 1 Install either the flowmeter or the end cap by following instructions under FL
19. tyle of washer when attaching the regulator to a cylinder V PRODUCT CLASSIFICATION DATABASE Device Regulator Pressure Gas Cylinder Device Description Pressure regulator Medical Specialty Anesthesiology Product Code CAN Regulation Number FDA 868 2700 Device Class 1 GMP Exempt No 510 k Exempt Yes European Commission Meddev 2 4 1 Rev 8 Part 2 Rule 11 Active devices to administer remove medicines and other substances to or from body or Directive 93 42 Class 2B Pressure regulator for medical gases Health Canada CAN CSA Z305 87M Medical Devices Regulations Classification Rules for Medical devices Part 1 Rule 9 Active devices to administer or withdraw energy to or from the body Class 2 Medical Pressure regulator VI INSTALLATION INSTRUCTIONS See Instructions for use provided with each regulator See Work Instruction WI PRO 010 General Assembly Procedures for RW and FM Exhibit AT SERVICE AND REPAIR Always perform the regulator test procedures in this manual before placing a repaired regulator back in service Note Refer to the Parts List and Parts Illustration for identification of parts referenced with bold numbers A CLEANING LUBRICATION AND SEALING Clean the metal parts of the regulator with isopropyl alcohol Use a mild solution of soap and water to clean the plastic parts Use a clean lint free cloth with isopropyl alcohol to clean the internal parts being certain to re
20. ure 50 PSI 17 20 24 30 40 58 70 90 110 PSI available upon request Relief Valve Pressure 75 PSI or 95 PSI Inlet Connections Any CGA inlet connection is available for specific gases and most International is available upon request Dimensions Variable depending on type and style specified Weight Variable depending on type and style specified Indicators Cylinder Contents Medical Gas Flow Rate Power take off Outlet Connections none 1 or 2 of any combination of the following DISS check valve Chemetron QD Hansen QD Ohio QD Puritan Bennett QD Schrader QD plugged 1 8 FNPT ports or plugged 14 FNPT ports Flowmeter Outlet Connection DISS fitting or Hose Barb HI THEORY OF OPERATION The Flotec Medical Gas Regulator is a single stage piston type regulator When it is initially attached to a high pressure gas cylinder the internal components are in the position shown in enclosed drawings With the cylinder off the spring exerts force against the piston assembly and pushes the piston assembly away from the regulator seat When the cylinder valve is opened high pressure medical gas approximately 2000 PSI oxygen for a full cylinder passes through the inlet filter and flows through the seat The gas then flows through the shaft of the piston and exits to the regulated output cavity The gas continues to flow into this area until the cavity builds enough pressure to overcome the force of the spring which occurs at 50 PSI
Download Pdf Manuals
Related Search
Related Contents
JLI-UV Installation-, adjustment and user guide TracVision Cruiser Owner`s Manual Equipo GPS VT400 Ficha Técnica manuel d`utilisation et d`entretien Craftsman 8' Workbench Frame - Red Service Parts EN SITES ARTISTIQUES TEMPORAIRES ? FR - Rasor 1946 Mode d`emploi Projet collectif FATS, FATAR FATSCOPY User Manual GettinG started Copyright © All rights reserved.