1. Power Grounding 6 of 6 Site Preparation Jun 2001 Agilent 6890 Gas Chromatograph Service Manual
2. 200 V 5 2 950 15 amp dedicated Fast heating 220 V 5 2 950 15 amp dedicated Fast heating 230 V 5 2 950 16 amp dedicated Fast heating 230 V 5 2 250 10 amp dedicated Slow heating Switzerland or Denmark with 10 amp maximum service 240 V 5 2 950 13 or 16 amp dedicated Fast heating Frequency range for all voltages is 48 to 66 Hz The fast heating oven requires at least 200 V Most countries standard voltage meets this requirement GCs for use in the USA Denmark and Switzerland will be equipped with a slow heating oven unless they are ordered with power option 002 which specifies a fast heating oven Although your GC should arrive ready for operation in your country compare its voltage requirements with those listed in Table 530 2 If the voltage option you ordered is not suitable for your installation contact Agilent Technologies USA fast heating oven The fast heating oven requires 240 V 15A power Do not use 208 V power Lower voltage causes slow oven ramps and prevents proper temperature control The power cord supplied with your GC is rated for 250 V 15A and is a two pole three wire cord with grounding type L6 15R L6 15P Receptacle Plug L6 15R L6 15P 2 of 6 Site Preparation Jun 2001 Agilent 6890 Gas Chromatograph Service Manual Power 530 Grounding The fast heating oven requires at least 200 V Most countries standard voltage meets this requirement GCs intended for use in the USA D
3. 530 Grounding Caution Caution Power A proper earth ground is required for GC operations To protect users the metal instrument panels and cabinet are grounded through the three conductor power line cord in accordance with International Electrotechnical Commission IEC requirements The three conductor power line cord when plugged into a properly grounded receptacle grounds the instrument and minimizes shock hazard properly grounded receptacle is one that is connected to a suitable earth ground Proper receptacle grounding should be verified Make sure the GC is connected to a dedicated receptacle Use of a dedicator receptacle reduces interference Any interruption of the grounding conductor or disconnection of the power cord could cause a shock that could result in personal injury Line voltage The GC operates from one of the AC voltage supplies listed in Table 530 1 depending on the standard voltage of the country from which it was ordered GCs are designed to work with a specific voltage make sure your GC voltage option is appropriate for your lab The voltage requirements for your GC are printed near the power cord attachment Jun 2001 Site Preparation 1 of 6 Agilent 6890 Gas Chromatograph Service Manual 530 Power Grounding Table 530 1 Line Voltage Requirements Maximum power Power line Voltage consumption VA requirement Oven type 120 V 5 2 250 20 amp dedicated Slow heating
4. enmark and Switzerland will be equipped with a slow heating oven unless they are ordered with power option 002 which equips the GC with a fast heating oven Canadian installation When installing a GC in Canada make sure your GC s power supply circuit meets the following additional requirements e The circuit breaker for the branch circuit which is dedicated to the instrument must be rated for continuous operation e The service box branch circuit must be marked as a Dedicated Circuit Jun 2001 Site Preparation 3 of 6 Agilent 6890 Gas Chromatograph Service Manual 530 Grounding Table 530 2 Voltages and Line Cord Terminations by Country Country Power cord Voltage Oven type termination Australia 10 amp 240 V Regular C Australia 15 amp 240 V Fast China 10 amp 220 V Regular China 16 amp 220V Fast Continental Europe 3 220V Fast 100 D Continental Europe 230 V Fast Denmark 10 amp 230 V Regular Denmark and Switzerland 230 V Fast 16 amp Switzerland 10 amp 230 V Regular Hong Kong 220 V Fast India South Africa 240 V Fast Israel 230 V Fast Japan 200V Fast 4of 6 Agilent 6890 Gas Chromatograph Service Manual Site Preparation Jun 2001 Power 530 Grounding United Kingdom 240 V Fast et USA gt 120 V Regular USA A 240 V Fast S Site Preparation 5 of 6 Jun 2001 Agilent 6890 Gas Chromatograph Service Manual 530
Download Pdf Manuals
Related Search
Related Contents
VienteVienteViente VienteVienteViente IMPORTANT! SERVICE MANUAL RIVA COMPACT M90E.24S M90E.28S Miele DA 390-6 40" Operating and Installation Instructions 取扱説明書[EW-DJ61] (2.09 MB/PDF) LM6100 Series LCD Monitor Logan Graphic Products 560-1 User's Manual Charlo Avec Charlo, qu`est Samsung RT32CHSW User Manual Behringer X-USB Product Information Copyright © All rights reserved.
Failed to retrieve file