1. Z amanu PIPA dd 3LV9 LUVAS andang 5 ino C 19915 OLIVA Luwis Block Diagram of the Model 566 Time to Amplitude Converter Bin Module Connector Pin Assignments For Standard Nuclear Instrument Modules per DOE ER 0457T Pin Function 1 4 3V 2 3V 3 Spare bus 4 Reserved bus 5 Coaxial 6 Coaxial 7 Coaxial 8 200Vdc 9 Spare 10 6V 11 6V 12 Reserved bus 13 Spare 14 Spare 15 Reserved 16 12V 17 12V 18 Spare bus 19 Reserved bus 20 Spare 21 Spare 22 Reserved Pin 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 Function Reserved Reserved Reserved Spare Spare 24 V 24 V Spare bus Spare Spare 117 V ac hot Power return ground Reset Scaler Gate Reset Auxiliary Coaxial Coaxial Coaxial 117 V ac neutral High quality ground Ground guide pin Pins marked are installed and wired in ORTEC s 4001A and 4001C Modular System
2. 4 Switch on ac power and check the dc power voltage levels at test points on the 4001C BASIC SWITCH SETTINGS Set the 566 controls as follows Range 50 ns Multiplier 1 Logic Input Anti Coin Strobe Int CONVERSION TESTS Use the typical test setup shown in Fig 6 1 and supply a start and stop pair of input signals with known time difference into the 566 Observe the TAC output Then use the following procedures PULSE GENERATOR Be 05 t 2 nie a Neg Output FANOUT Start Input 10 Output OSCILLO SCOPE Stop Input Neg Output Fig 6 1 Test System for Checking Conversion 1 Adjust the delay for the stop input to 50 ns 2 Check to see that the full scale time range is 50 ns x 1 3 Measure the signal through the TAC Output connector It should be about 10 V for a 50 ns delay or 5 V for a 25 ns delay 4 Turn the Range ns switch through its 50 100 and 200 settings and observe the pulse amplitude at each setting each successive switch position should decrease the pulse amplitude to about half of the amplitude for the previous setting 5 Return the Range ns switch to 50 and set the Multiplier switch at 10 The output amplitude should be reduced to about 1 10 of the reading for step 3 RESOLUTION TESTS See Fig 6 2 for the typical test setup used for resolution checks The start and stop pulses used for this test must have fast rise time and be jitter free The
3. current Using the typical setup shown in Fig 6 1 follow the setup procedures outlined in Conversion Tests steps 1 through 3 Adjust potentiometer R128 counterclockwise to minimum If the TAC is functioning properly a pulse output of 8 5 to 9 8 V should be present on the oscilloscope If this voltage is not present on the oscilloscope refer to Section 6 3 Troubleshooting or return the unit to the ORTEC factory for repair as outlined in Section 6 4 If a pulse is present at the TAC output the conversion current can be calibrated by adjusting R128 clockwise slowly until the output pulse obtains an amplitude of 10 V and begins to half fire Adjust R128 counterclockwise slightly until the pulse is solid or full firing OUTPUT WIDTH ADJUSTMENT Potentiometer R163 located on the PWB adjusts the TAC output pulse width The potentiometer should vary the output pulse width from 1 us to 3 us The width can be adjusted to customer requirements factory set at 2 Us OUTPUT OFFSET CALIBRATION Potentiometer R223 located on the PWB adjusts the TAC output offset from 100 mV to 100 mV Remove the start and stop signals from the TAC input Measure the offset voltage at the TAC output connector with a voltmeter Adjust potentiometer R223 until the offset voltage is 0 mV 1 mV Capacitor C93 located on the PWB is used to filter out any linear gate signals from the TAC output pulse This is factory set and should need fu
4. CIRCUIT DESCRIPTION The jumpers locations and settings discussed in this chapter are depicted in Figs 3 and 4 Please refer to these figures as you read the circuit description given below To access the PWB as viewed in these illustrations remove the right side panel as viewed from the front panel The jumpers are located near the bottom edge of the board The identifiers used in the illustrations are Jumper Identifiers Definition SP 5 STG 04 ST J3 STR 01 VST J7 5 1 START CIRCUITRY The start circuitry is used to generate a logic signal which begins the time to amplitude conversion in the 566 Two inputs the START front panel and the GATE IN rear panel are used to control the start circuitry The GATE IN allows START signals to be blocked by a user supplied logic signal The user can select blocking in coincidence or anticoincidence mode via a rear panel switch Also PWB jumpers allow the selection of either positive or negative NIM logic signals for both the START and the GATE IN The input circuitry for the START and the GATE IN provide for the proper termination of the input signals level shifting and buffering The START input circuitry consists of Q1 Q2 1 2 of U1 J3 D4 D6 and associated passive components while the GATE IN input circuitry includes Q3 Q4 half of U1 J4 D1 D3 and associated passive components If the GATE IN signal meets the proper coincidence anticoinci
5. Q41 which opens the linear gate A4 allowing the output signal to appear at the TAC output If the EXT position is selected on the STROBE switch the logic signal supplied by the user on the STROBE input will control the timing of the TAC output Either positive or negative NIM logic can be selected for this input via PWB jumper If the EXT mode is selected the STROBE input triggers one shot U15 13 which opens the linear gate The user may allow a time window for the STROBE signal of 10 us or 100 us If a strobe signal does not occur in this time window time window begins with the STOP signal the TAC output is not enabled and the 566 is reset P N 4274300 23 Hise a fol Fs fot Oy RIO ce a8 5 a 1 SP STG ST STR VST VCV Fig 3 Jumper Locations MODEL 566 JUMPER SETTINGS ALL JUMPERS SET TO POSITIVE oloo VST O O OO OO 5 STG ST STR ALL JUMPERS S
6. Time ranges from 50 ns to 2 ms are provided via the front panel controls The 566 Start input can be inhibited by a pulse ora dc level at the rear panel Gate In input connector Valid Start and Valid Conversion outputs are provided on the rear panel for each accepted start and stop input respectively The duration of the Valid Start output indicates the interval from the accepted start until the end of reset Valid Conversion occurs from the end of the internal delay after stop to the end of reset The selectable TAC output width and variable delay which are easily adjustable further serve to make the 566 a flexible instrument that can be easily adapted into many time spectroscopy systems The output of the TAC may be synchronized with the stop signal or an external strobe signal to further enhance its versatility The 566 is dc coupled and gated so that input count rates will not paralyze or otherwise hinder normal operation The TAC output should be connected to the dc coupled input of a multichannel analyzer MCA for optimum high count rate performance 1 2 OPERATION Start to stop time conversion is accomplished only after a valid start has been identified and after a stop pulse has arrived within the selected time range The start input is disabled during the busy interval to prohibit pileup the stop input is disabled after the first accepted stop signal The input gate for the start circuit can be operated in either an ant
7. allotment for a single NIM width on the 12 V and 12 V bin lines However this position allows the 566 to be used with power supplies not providing 6 V and 6V 2 4 INPUTS All four inputs listed below are dc coupled edge triggered and printed wiring board PWB jumper selectable to accept either negative or positive NIM standard signals Input impedance is 50 in the negative position and gt 1kQ in the positive position The threshold is 400 mV in the negative position and 2 V in the positive position STROBE Front panel BNC connector provides an external means to strobe a valid output signal from the TAC in the Ext Strobe mode The input signal exceeding threshold within the Ext Strobe reset interval after the Stop input initiates the read cycle for the linear gate to the TAC output Factory set in the positive input position Ext Strobe reset interval has a minimum value of 0 5us a maximum value of nominally 10 us START Front panel BNC connector initiates time conversion when Start input signal exceeds threshold Factory set in negative input position STOP Front panel BNC connector terminates time conversion when Stop input signal exceeds threshold Factory set in negative input position GATE IN Rear panel BNC connector provides external means of gating the Start circuitry in either Coincidence or Anticoincidence with the Start input signal Gate input signal must cross threshold gt 10 ns prior to t
8. signal after it has been enabled by the start condition It may be enabled immediately after start When a stop input signal is accepted this indicates that an interval has been measured and its analog equivalent is stored and available A signal is furnished through the Valid Converter output that continues until the leading edge of a subsequent reset If no stop input is accepted before an over range condition is sensed the measurement will be aborted and no output signals for the TAC will be generated The TAC output must be strobed The source of the strobe can be switch selected from the internal signal or from an external signal If internal is selected the strobe occurs after a delay that has been adjusted with the front panel TAC Output delay control 0 5 us to 10 5 us after the leading edge of the signal If the Strobe switch is set at Ext a signal must be furnished through the Strobe Ext BNC connector to strobe the output promptly 2 SPECIFICATIONS 2 1 PERFORMANCE TIME RESOLUTION FWHM lt 0 01 of full scale plus 5 ps for all ranges TEMPERATURE INSTABILITY lt 0 01 C 100 ppm C of full scale or 10 ps C whichever is greater 0 to 50 C DIFFERENTIAL NONLINEARITY lt 2 from 10 ns to 2 of full scale whichever is greater to 100 of full scale INTEGRAL NONLINEARITY lt 0 1 from 10 ns to 2 of full scale whichever is greater to 100 of full scale RESET CYCLE Fixed 1 0 us for X1 and X10 mu
9. 0 0c eects 6 5 4 STROBE CIRCUITRY Scena iene Se Sat an ees ana te vee viata 6 6 MAINTENANGE i resa dee Sethe ee eine raat hada Paar Gale Beats cen 8 6 1 TESTING PERFORMANGE 8 6 2 CORRECTIVE MAINTENANCE 12 6 3 TROUBLESHOOTING 27 2 Seg ede beet Legenda ee beds he 13 6 4 EACTORY REPAIR asso dates i eu eee aa ewe Ya Sage ee eae ee 13 SAFETY INSTRUCTIONS AND SYMBOLS This manual contains up to three levels of safety instructions that must be observed in order to avoid personal injury and or damage to equipment or other property These are DANGER Indicates a hazard that could result in death or serious bodily harm if the safety instruction is not observed WARNING Indicates a hazard that could result in bodily harm if the safety instruction is not observed CAUTION Indicates a hazard that could result in property damage if the safety instruction is not observed Please read all safety instructions carefully and make sure you understand them fully before attempting to use this product In addition the following symbol might appear on the product to Manual DANGER High Voltage Please read all safety instructions carefully and make sure you understand them fully before attempting to use this product SAFETY
10. 00 ns to ensure that it will occur later than the full time range 5 Switch the 566 to its external strobe mode and observe the TAC output pulse It should be identical to the pulse observed in step 3 except for the time at which it occurs 6 Vary the front panel Delay us and observe that there is no change in the TAC output amplitude but that the output delay follows the strobe delay 6 2 CORRECTIVE MAINTENANCE Clean the surfaces of the printed circuits the connectors and all chassis parts periodically to prevent accumulated dust from forming leakage paths between the circuit components If the instrumentis suspected of malfunctioning use the performance tests of Section 6 1 to aid verification When incorrect operation is identified disconnect the 566 from its position in the system and perform routine diagnostic tests with a pulse generator and an oscilloscope Use the timing chart in Fig 6 7 to isolate the problem CALIBRATION There is one critical calibration and three adjustments that may need to be made GATE INPUT 1 anti il START ee 05 13 STOP INPUT mr 10512 Lu S VALID START J T L OUTPUT CONVERTER OUTPUT TL rT Q23 EMITTER VALID CONVERTER L INTERNAL DELAY U11 8 L 11 RESET 1 LJ OVERRANGE qr rT Fig 6 7 Timing Chart CONVERSION CURRENT ADJUST Potentiometer R128 located on the PWB adjusts the conversion
11. 7 Subtract the peak channel number in step 3 from the peak channel number in step 6 This is the number of channels that represents the time variation injected at step 4 8 Using the formula below calculate time resolution effective for the established system calibration At per channel stop delay increase channel shift 9 With the equation below calculate the converter resolution using the FWHM channel width from either step 3 or step 7 These widths should be the same at either peak location Time resolution FWHM FWHM channel width x At per channel This resolution is affected adversely by any jitter that may be present in the discriminator and by the resolution of the amplifier Allowances should be made for these contributions COUNT RATE TESTS In many applications it is important for a time to amplitude converter to handle high count rates both external and internal Since the start input is gated internally and the conversion circuits are all direct coupled the limit for its external count rate capability is determined solely by the input pulse width and there are no pileup effects The limit on the internal count rate is imposed by the conversion and reset process where the start input is disabled through a converter busy interval following each accepted start signal A converter busy interval is the measured time plus 7 us for start stop intervals within the selected time range or is the selected time range pl
12. ET TO NEGATIVE o 0010 VST O olo SP STOP IN SP STG ST STR TATE STARTIN STR EXT STROBE IN VST VALID START OUT VCV VALID CONV OUT Fig 4 Jumper Settings top all positive bottom all negative 6 MAINTENANCE 6 1 TESTING PERFORMANCE The following test procedures are furnished as a guide during installation and checkout of the 566 TAC TEST EQUIPMENT The following test equipment is recommended Each test procedure refers to this list by the unit identification number for the required item s of test equipment An equivalent unit may be substituted for any item in the list providing that it furnishes the function required for each specific application Hewlett Packard 215A Pulse Generator ORTEC 436 100 MHZ Discriminator ORTEC 416A Gate and Delay Generator ORTEC 425A Nanosecond Delay Photomultiplier tube with scintillator and radiation source ORTEC 449 Log Lin Ratemeter Tektronix Type 475 Oscilloscope ORTEC 7100 7150 7450 Multichannel Analyzer or 918 Multichannel Buffer 9 ORTEC 414A Fast Coincidence 10 ORTEC 444 Gated Biased Amplifier OND PRELIMINARY PROCEDURES Take the following preliminary steps when the 566 is installed 1 Check the module visually for possible damage 2 With the power turned off install the 566 into a NIM Standard bin and power supply such as the ORTEC 4001C 4002A 3 Check the installation for proper mechanical alignment
13. Model 566 Time to Amplitude Converter TAC Operating and Service Manual Printed in U S A ORTEC Part No 678950 0411 Manual Revision F Advanced Measurement Technology Inc a k a ORTEC a subsidiary of AMETEK Inc WARRANTY ORTEC warrants that the items will be delivered free from defects in material or workmanship ORTEC makes no other warranties express or implied and specifically NO WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE ORTEC s exclusive liability is limited to repairing or replacing at ORTEC s option items found by ORTEC to be defective in workmanship or materials within one year from the date of delivery ORTEC s liability on any claim of any kind including negligence loss or damages arising out of connected with or from the performance or breach thereof or from the manufacture sale delivery resale repair or use of any item or services covered by this agreement or purchase order shall in no case exceed the price allocable to the item or service furnished or any part thereof that gives rise to the claim In the event ORTEC fails to manufacture or deliver items called for in this agreement or purchase order ORTEC s exclusive liability and buyer s exclusive remedy shall be release of the buyer from the obligation to pay the purchase price In no event shall ORTEC be liable for special or consequential damages Quality Control Before being approved for shipment each ORTEC instrume
14. NS 2 Poe el ede ee 2 2 1 PERFORMANCE ies eases Gaels 2 2 2 FRONT PANEL CONTROLS ae 2 2 3 REAR PANEL CONTROLS casio obes nii pehi bhi 2 24 INPUT S bie gre G18 Vine a a 2 225 OUTPUTS paket 3 2 6 ELECTRICAL AND MECHANICAL 0 0 0 0c cette eee 3 Si INSTALLATION Oe ie Soe eh Se ilo She Ok A a Ad Oe ble Ae 3 Bila GENERALS is mae BAe eae sola Path geal aa Doge a Paks Ved Nae held Pak geod 3 3 2 CONNECTION TO POWER 3 3 3 CONNECTION INTO A SYSTEM 0 00 cect tte ees 3 3 4 LINEAR OUTPUT SIGNAL CONNECTIONS AND TERMINATING IMPEDANCE 4 3 5 LOGIC SIGNAL CONNECTIONS 2 0 0 0 eee 4 4 OPERATINGINSTRUGCTIONS tale eds Pedal yaaa Pha pat ded deeds 5 4 1 TIME TO AMPLITUDE CONVERSION 0 0 0 0 c eects 5 5 CIRGUIT DESCRIPTION pared Sa ok etalon dl Stee a Pee aed net teow Pet 5 START CIRCUITIRY eerie idee eed Geeta eae 6 52 STOP CIRCUITRY Sis ti Seti eda bene ta dete aie Riga ae sae Wee 6 5 3 GATED BASELINE RESTORER
15. ONS NIM standard single width module 3 43 x 22 13 cm 1 35 x 8 714 in per TID 20893 Rev 3 INSTALLATION 3 1 GENERAL The ORTEC 4001 A 4002A 4001 A 402D or 4001 C 402D Bin and Power Supply or equivalent in which the 566 will be installed is intended for rack mounting If vacuum tube equipment is operated in the same rack there must be sufficient cool air circulating to prevent localized heating of the all transistor circuits in the 566 and in the other modules in the bin and power supply Rack mounted equipment subjected to the temperatures in vacuum tube equipment can exceed the maximum for which the transistorized circuits are designed unless this precaution is taken The 566 should not be subjected to temperatures in excess of 120 F 50 C 3 2 CONNECTION TO POWER The 566 is designed per TID 20893 and accepts its operating power requirements through a mating power connector when it is installed in an ORTEC 4001A 4002A Bin and Power Supply As a safety precaution always turn off the power for the bin before inserting or removing any modules Monitor the dc voltages at the test points on the control panel of the bin after all modules have been installed and the power is turned on in order to determine that none of the four power levels have been reduced by an overload 3 3 CONNECTION INTO A SYSTEM The 566 can accept both start and stop pulses from NIM modules that furnish NIM standard positive and fast nega
16. Regulate the random start rate to cause the MCA dead time to be 10 4 Clear the analyzer to zero and operate the system until the average count level stored in each channel is sufficient to ensure statistical accuracy 5 Compare any nonlinearity indications to the specifications listed in Section 2 Some nonlinearity can be expected in channels in the lower 5 of the MCA range as shown in Fig 6 5 because of the stop pulse width and the TAC gating time CHECKING EXTERNAL STROBING MODE The system for checking the external strobing mode is shown in Fig 6 6 This system can be used to verify the principles of operation of the 566 Range 1 Multiplier X1 100 ns Vert Full Scale 5 x 10 counts Horiz Full Scale 105 range Range 1 Multiplier X10 Vert Full Scale 5 x 10 counts Horiz Full Scale 105 range Tus Range 1 Multiple 108 Vert Full Scale 5 x 10 counts Horiz Full Scale 105 range Fig 6 5 Differential Linearity for the Indicated Ranges 1 Setthe delay for the stop signal at about 400 ns 2 Set the 566 time range for 500 ns FAST OSCILLO DISCRIMI 7 SCOPE RATOR NATOR NANOSECOND DELAY Fig 6 6 Test System for Checking External Strobing Mode 3 Use the internal strobe mode for the 566 Adjust the oscilloscope sweep as required to identify the TAC output pulses 4 Adjust the delay for the strobe signal gt 5
17. WARNINGS AND CLEANING INSTRUCTIONS DANGER Opening the cover of this instrument is likely to expose dangerous voltages Disconnect the instrument from all voltage sources while it is being opened WARNING Using this instrument in a manner not specified by the manufacturer could impair the protection provided by the instrument Cleaning Instructions To clean the instrument exterior e Unplug the instrument from the ac power supply Remove loose dust on the outside of the instrument with a lint free cloth e Remove remaining dirt with a lint free cloth dampened in a general purpose detergent and water solution Do not use abrasive cleaners CAUTION prevent moisture inside of the instrument during external cleaning use only enough liquid to dampen the cloth or applicator e Allow the instrument to dry completely before reconnecting it to the power source vi ORTEC MODEL 566 TIME TO AMPLITUDE CONVERTER 1 DESCRIPTION 1 1 PURPOSE AND FEATURES The ORTEC 566 Time to Amplitude Converter TAC is a single width NIM standard module that measures the time interval between pulses to its start and stop inputs and generates an analog output pulse proportional to the measured time Timing experiments requiring time ranges of 50 ns to 2 ms 10 ns to 2 ms usable time range may be performed giving the experimenter flexibility in analyzing random nuclear events that occur within a selected time range
18. an output pulse through the TAC Output connector The output pulse will not be furnished through this connector unless it is strobed The strobe condition is selected by a front panel switch When the output does occur its peak amplitude will be proportional to the ratio of the measured start to stop interval to the selected full scale time in a 0 to 10 V range Internal logic eliminates any pulse ambiguity No output pulse is furnished unless a stop signal is accepted within the selected full range time A stop signal is not effective unless it is preceded by a valid start signal For further logical control either a coincidence or anticoincidence mode can be selected for gating control of the start input circuit To eliminate gating for the start input set the Gate Mod switch at Anti Coin and leave the gate input circuit without any connection When the same switch setting is used and an input signal is furnished start signals are not accepted when the gate signal is 2 V or more For coincidence gating of the start input circuit set the Gate switch at Coin and furnish a signal of 2 V or more through the Gate logic Input connector when start signals are to be accepted The front panel Strobe switch selects the source for the strobe signal for the TAC output When the switch is set at Int the strobe is generated by a delayed valid start signal and the delay is fixed at a time that is longer than the selected full range time 5
19. dence condition selected by the user an ECL logic true level is set on U2 7 the D input of a dual ECL flip flop allowing a START signal to set this flip flop The Q output of this flip flop U2 2 sets the other half of U2 place U2 15 in the ECL true state and U2 14 in the ECL false state These two signals go through level shifting transistors Q10 Q11 and QI5 and turn transistor Q17 off Then the current which was flowing through Q17 discharges C38 and any other capacitor selected by the front panel MULTIPLIER switch This discharging will continue until halted by the stop circuitry The discharge rate of C38 is set by the precision current source comprised of U7 Q29 DI9 and associated passive components Both the RANGE switch and the MULTIPLIER switch control the magnitude of this current source A logic signal is generated on the VALID START output by each start signal which is enabled by the GATE IN This signal extends from the beginning of the start pulse until the TAC is reset 5 2 STOP CIRCUITRY The stop circuitry generates a logic signal which ends the time to amplitude conversion in the 566 This circuitry is controlled by the front panel STOP input and a PWB jumper allows the selection of either a positive or negative NIM logic signal for STOP The STOP input circuitry is identical to the START input circuitry When the STOP input becomes active an EOL flip flop 1 2 of U4 is set The Q output of this
20. e only about 90 of the maximum signal amplitude for each pulse The termination of a 93 0 cable ina 93 0 load will cause 50 loss for the signal 3 5 LOGIC SIGNAL CONNECTIONS The start and stop input circuits accept positive or negative NIM standard signals Each of these input circuits is jumper selectable Input impedance of 50 for negative signals or gt 1 for positive signals is intended as the proper termination for the signals Impedance considerations for each of the remaining logic inputs and output for the 566 are noncritical 93 0 cable is usually used They can be terminated with 100 prevent ringing if the signal is used to drive a high impedance load 4 OPERATING INSTRUCTIONS 4 1 TIME TO AMPLITUDE CONVERSION There are four front panel controls and two rear panel controls on the 566 Of these four are directly associated with the conversion of a start to stop interval into an analog equivalent TAC output pulse These controls are Range ns Multiplier Delay us and Anti Coinc The Range ns and Multiplier switches determine the full scale limit for time conversion Any of 15 combinations can be selected as follows Full Scale Time Range ns Multiplier For example with the Range switch set at 50 and the Multiplier switch at X100 the full scale time range is 5 us Any stop input signal that occurs within 5 us after a valid start signal will initiate the gating of
21. engths are required for transfer of a linear signal the cable should be terminated in a resistive load equal to the cable impedance to prevent reflections and oscillations in the cable Oscillation suppression can be effected by either a series termination at the sending end of the cable or by a shunt termination at the receiving end For convenience a BNC tee can usually accommodate both the cable and a mating terminator at the input of the receiving instrument These units ate available commercially including B C terminators with nominal values of 50 100 and 1000 ORTEC stocks a limited quantity of all but the 1000 terminators for your convenience as listed below B C Tee Connector 29 50 Terminator C 28 100 Terminator C 27 When a shunt termination at the receiving end of the cable is impractical consider series termination at the sending end For a series termination the full signal amplitude span is available at the receiving end only if the input impedance is many times the characteristic impedance of the cable For series termination install the correct resistance between the actual amplifier output on the etched circuit board and the output connector Effectively the terminating resistance is in series with the input impedance of the receiving instrument and may result in some loss in signal amplitude For example if the series terminator is 93 and the driven load is 900 Q the available signal span will b
22. flip flop sets the other 1 of U4 if a valid START signal has previously set U2 2 to the ECL true state causing U4 15 to switch to the ECL true state and U4 14 to switch to the ECL false state These signals propagate through level shifting transistors Q12 Q13 and Q19 to turn off transistor Q20 terminating the current flow out of C38 and ending the time to amplitude conversion cycle 5 3 GATED BASELINE RESTORER The voltage on C38 is buffered by Q21 Q23 The voltage on the emitter of Q23 is held to a quiescent value of 0 V by a gated baseline restorer BLR consisting of U8 C29 Q26 Q31 Q32 U5 and associated passive components In the quiescent state the gated BLR acts as a high gain feedback circuit which maintains the output of Q23 at 0 V However when a valid START signal is received an ECL false signal on U5 13 disables U8 which opens the feedback path The proper dc voltage is maintained throughout the loop though since the output voltage of U8 is held on C29 After the START flip flop is cleared the BLR circuit is again enabled 5 4 STROBE CIRCUITRY When the INT position is selected on the STROBE switch the output signal is gated out by the buffered STOP signal U5 12 after a user selected delay time This delay time is set by the front panel DELAY potentiometer which controls the width of the negative going TTL logic signal on U10 4 The rising edge of this signal triggers one shot U15 13 which fires
23. he Start input signal and must overlap the trigger edge of the Start input signal Factory set in the positive input position 2 5 OUTPUTS TAC OUTPUT Front panel BNC connector provides unipolar pulse e Amplitude 0 V to 10 V proportional to Start Stop input time difference e Time End of delay period Int Strobe mode prompt with Strobe input in Ext Strobe mode Width Adjustable by PWB potentiometer from lt l us to gt 3 Us e Impedance Z lt 10 e Rise Time 250 ns Fall Time 250 ns VAL ST Rear panel BNC connector provides NIM standard slow positive logic level signal e Amplitude Nominally 5 V Complement signal selectable by PWB jumper e Time and Width From accepted Start input to end of reset e Impedance Z lt 100 Rise Time lt 50 Fall Time lt 50 ns VAL Rear panel BNC connector provides NIM standard slow positive logic level signal to indicate a valid conversion e Amplitude Nominally 5 V Complement signal selectable by PWB jumper Time and Width From end of internal delay after Stop to end of reset Impedance Z lt 1 0 e Rise Time 250 ns e Fall Time 250 ns 2 6 ELECTRICAL AND MECHANICAL POWER REQUIRED Log Switch 6V 24 V 45 mA 24 V 50 mA 12 V 95 mA 12 V 140 mA 6 V 140 mA 6 V 300 mA 12V 24 V 45 mA 24 V 50 mA 12 V 210 mA 12 V 405 mA WEIGHT Net 1 5 kg 3 3 Ib Shipping 3 0 kg 7 0 Ib DIMENSI
24. icoincidence or a coincidence mode Time ranges may be switch selected for full scale intervals from 50 ns to 2 ms Each time measurement is analog stored in low loss stretcher amplifier until a linear gate is opened by either an internal or an external strobe The internal strobe can be obtained from either the start or the stop input pulse and in either case occurs automatically at a selected delay following the reference An external strobe can be used for a prompt output at the strobe time provided that atime measurement has been completed and reset has not occurred If reset occurs before a strobe no TAC output signal is available Reset also occurs if the start to stop time interval exceeds the range that is selected 1 3 LOGIC An input can be accepted through the Start input connector on the front panel unless the 566 is busy processing a previous set of information or the response is inhibited by a gate input condition The acceptance of a start input is essential in order to initiate a response in the 566 When a start input is accepted a positive logic signal is available through the rear panel Valid Start output connector and is continued until the leading edge of a subsequent reset The reset can be caused by a TAC output or by the sensing of an over range condition The start signal permits the internal circuits to start measuring a time interval and enables the stop input circuit The Stop input BNC can accept an input
25. ime of the 566 At each start input signal the 566 will analyze the time until its related origin signal is furnished to the stop input The time measured is then delay time minus T and produces a so called inverted time spectrum The purpose of this type of system connection is to reduce the number of conversions and the corresponding dead time during the experiment For each signal accepted through the start input there must be a conversion but for each signal through the stop input there need not be a conversion For each start signal that is not followed by a stop signal within the selected time full range the converter measures a time equal to the total range even though no output pulse is generated 3 4 LINEAR OUTPUT SIGNAL CONNECTIONS AND TERMINATING IMPEDANCE The source of impedance of the standard TAC output with the 0 to 10 V linear range is about 1 through the connector on the front panel For the front panel circuit the interconnection to other modules does not usually require any special considerations especially it the interconnecting cable is shorter than 4 ft in length Paralleling several loads on a single output will still not reduce the 0 to 10 V signal span significantly unless the combined load is lt 100 As with any analog instrument oscillations may be observed occasionally when unterminated lengths of cable are used Short cable lengths up to 4 ft need not be terminated When longer cable l
26. in warranty should follow the same procedure and ORTEC will provide a quotation Damage in Transit Shipments should be examined immediately upon receipt for evidence of external or concealed damage The carrier making delivery should be notified immediately of any such damage since the carrier is normally liable for damage in shipment Packing materials waybills and other such documentation should be preserved in order to establish claims After such notification to the carrier please notify ORTEC of the circumstances so that assistance can be provided in making damage claims and in providing replacement equipment if necessary Copyright 2011 Advanced Measurement Technology Inc All rights reserved ORTEC is a registered trademark of Advanced Measurement Technology Inc All other trademarks used herein are the property of their respective owners CONTENTS WARRANTY saya oaa a eat kas oe Seal ji SAFETY INSTRUCTIONS AND SYMBOLS 0 0 0 cect eee iv SAFETY WARNINGS AND CLEANING INSTRUCTIONS 1 DESCRIPTION 244 00 dee eek 2a pee pe a eee ee ee 1 1 1 PURPOSE AND FEATURES ok a gas VEG Pah a eal a a legs 1 1 2 OPERATION bx hh nce gare de ngs ed tne ete 1 1 3 LOGIC us Gt Bains Per ee een ase ais De eee Rea Ob out kd me Paste Beate 1 2 SPECIFICATIO
27. is used as the source for start signals and a pulse generator with a fixed rate is used for stop signals The measurable time interval between a start and stop is a random value with equal probability that it will be any time difference up to the periods between the regular stop signals For an infinite number of TAC outputs the count levels for each channel of the MCA should be equal After the test has been run long enough to assure statistical accuracy e g gt 25 000 counts channel the spectrum should be similar to those illustrated in Fig 6 4 Any deviation from a straight line represents a differential nonlinearity and the percent of deviation is the difference between this count level and the average divided by the average count level RADIATION SOURCE PHOTO MULTIPLIER DISCRIMI NATOR SCINTILLATOR PULSE GENERATOR Fig 6 4 Test System for Checking Differential Linearity 1 Select the 566 time range to be tested 2 Calculate the maximum stop pulse repetition rate for the selected time range This should be slightly lower than the reciprocal of the time range For example for the 1 us time range the reciprocal is 1 x 10 and a pulse generator rate of 4 to 5 times 10 should be satisfactory A lower rate increases the time required to run the test while a faster rate will reduce the response because of MCA dead time 3 Operate the system and monitor the dead time meter on the MCA
28. ltipliers fixed 5 us for X100 multiplier and fixed 50 us for X1k and X10k multipliers Occurs after Over range Strobe cycle or Ext Strobe Reset cycle START TO STOP CONVERSION TIME Minimum lt 5 ns 2 2 FRONT PANEL CONTROLS RANGE ns Three position rotary switch selects full scale time interval of 50 100 or 200 ns between accepted Start and Stop input signals MULTIPLIER Five position rotary switch extends time range by a multiplying factor of 1 10 100 1k or 10k DELAY us 20 turn screwdriver adjustable potentiometer varies the delay of the TAC output from 0 5 us to 10 5 us relative to an accepted Stop input signal operable in the Int Strobe mode only STROBE MODE Two position locking toggle switch selects either Internal or External source for initiating the strobe cycle to strobe valid information from the TAC output 2 3 REAR PANEL CONTROLS GATE MODE Two position locking toggle switch selects Coincidence or Anticoincidence mode of operation for the Start circuitry Start circuitry is enabled in the Coinc position or inhibited in the Anti position during the interval of a Gate input signal LOG CURR Two position locking toggle switch selects the use of 6 V or 12 V bin lines to provide current for the internal logic circuitry In the 6 V position the 566 is within the current allotment for a single NIM width when using a NIM standard Class V power supply In the 12 V position the 566 exceeds the current
29. minimum delay recommended for the stop pulses is 15 ns The resolution of any scale can be measured with this setup and the main consideration is that each stop signal delay be within the linear region of the selected time range The testing procedure consists of the following FANOUT PULSE GENERATOR Neg Output Start Input GATED BIASED AMPLIFIER Stop Input Fig 6 2 Test System for Checking Converter Resolution 1 Adjust the delay for the stop input to a basic setting of 30 to 80 of the selected time range 2 Operate the system and obtain a timing spectrum Normalize the output amplitude full range for the normally digitized full range of the ADC in the analyzer 3 After you have accumulated an adequate spectrum to assure statistical accuracy of photopeak measurements 1000 counts in the peak channel identify the peak channel number and measure the FWHM channel number limits Log for reference 4 Increase the delay for the stop signal by a fixed and known amount This may be done by switching in a fixed delay line cable ORTEC 425A or by careful adjustment of the delay unit controls The total delay for the stop signal must still be lt 100 of the selected time full range 5 Accumulate a spectrum for this measurement of increased time intervals 6 Observe the relocated photopeak in the timing spectrum and record its peak channel number and its FWHM channel number limits
30. nt must pass a stringent set of quality control tests designed to expose any flaws in materials or workmanship Permanent records of these tests are maintained for use in warranty repair and as a source of statistical information for design improvements Repair Service If it becomes necessary to return this instrument for repair it is essential that Customer Services be contacted in advance of its return so that a Return Authorization Number can be assigned to the unit Also ORTEC must be informed either in writing by telephone 865 482 441 1 or by facsimile transmission 865 483 2133 of the nature of the fault of the instrument being returned and of the model serial and revision Rev on rear panel numbers Failure to do so may cause unnecessary delays in getting the unit repaired The ORTEC standard procedure requires that instruments returned for repair pass the same quality control tests that are used for new production instruments Instruments that are returned should be packed so that they will withstand normal transit handling and must be shipped PREPAID via Air Parcel Post or United Parcel Service to the designated ORTEC repair center The address label and the package should include the Return Authorization Number assigned Instruments being returned that are damaged in transit due to inadequate packing will be repaired at the sender s expense and it will be the sender s responsibility to make claim with the shipper Instruments not
31. rther adjustment 6 3 TROUBLESHOOTING Use the tests in Section 6 1 and the circuit description in Section 5 6 4 FACTORY REPAIR This instrument can be returned to the ORTEC factory for service and repair at a nominal cost Our standard procedure for repair ensures the same quality control and checkout that are used for a new instrument Always contact our Global Service Center at ORTEC 865 482 4411 before sending in an instrument for repair to obtain shipping instructions and so that the required Return Authorization Number can be assigned to the unit Write this number on the address label and on the package to ensure prompt attention when it reaches the ORTEC factory 14 Anant 1 andyng AS vzs andang Ag Edl 1 2 AV1a0 390815 wOISHaANOD 620 6N 214 610 43 1 Asnray 130 ats UILIS 13437 nami dois HO 31V9 UV3NIT 90 911 109530 BLOM 104170 LOHSINO SWL sin oun 7 vesay 123 84 see w AS iy ndin 434400 21 a GE bia 111907 1 31 5 13 21 5 ee 920 120 110 810 910 10 610 11 7 925 wyo nhng YOLVNIWIYISIO 14 193 LOKS3NO 13530 11 610
32. tive logic signals or from the timing output of a photomultiplier tube base Typical ORTEC instruments that provide compatible signals are the 583 584 and 473A discriminators and the 265 269 270 and 271 photomultiplier tube bases The start and stop inputs will properly terminate 50 cable and this type is recommended to ensure proper termination of the signals No input or output connectors need be terminated when they are not in use In any experiment in which it is reasonable to assume that the count rates for start and stop will be equal or nearly equal use the signal furnished from the origin of events into the start input and the signal furnished from the response into the stop input The 566 will then measure the time difference T from origin to response and furnish an output amplitude that is some fraction of the selected full scale amplitude proportional to the ratio of T to the selected full scale time range In any experiment in which the two count rates differ noticeably such as one in which fewer responses than event origins can be expected use the lower count rate as the start input to the 566 This assures that the 566 dead time will be minimized because it analyzes the time difference only after a start signal is accepted When the response is used as a start signal furnish the signals from the origin of events through a delay line into the stop input and adjust the delay to match the selected full scale t
33. us 4 us if no stop signal is furnished within the time range The following test based on the system connection shown in Fig 6 3 permits accumulation of a basic timing spectrum for the start stop input pulses at 60 Hz As the external count rate for start only is increased by regulating the random pulse generator the internal pulse rate in the 566 is increased and a ratemeter will monitor the resulting rate at which the internal capability is impaired 1 The photomultiplier may be used as a random pulse generator triggered by a radiation source Use an initial sensitivity setting above the energy level for a zero output pulse rate 2 Adjust the delay for the stop input to 0 4 us RADIATION SOURCE PHOTO MULTIPLIER and CINTILLATOR 60 Couan s PULSE GENERATOR Fig 6 3 Test System for Checking Count Rate 3 Select the 0 5 us time range with the 566 4 Adjust the system for a timing spectrum accumulated for the 60 Hz input pulses 5 Decrease the threshold of the discriminator to generate random start signals with no corresponding stop signals Monitor the random rate with the ratemeter 6 Observe the timing spectrum as the random input rate is gradually increased Watch for interference in the accumulated spectrum DIFFERENTIAL LINEARITY MEASUREMENTS A system for testing differential linearity of the 566 is shown in the block diagram in Fig 6 4 In this system the random pulse generator