Home

User Guidelines for Mattress & Cushion Systems (dual voltage) 50

image

Contents

1. pressing MUTE button Reset power unit by pressing MUTE button J Check that CPR device if applicable is fully closed i e dial of CPR device is rotated clockwise until clicked into closed position a v If alarm fault persists power unit requires inspection MATTRESS FLAT no alarm Check air cells and tubing are intact AA Reset power unit by pressing MUTE button Check that mains power is connected indicated by illuminated LCD display and allow at least 20 minutes for mattress to inflate If alarm fault persists system requires inspection l Reset power unit by pressing MUTE button If alarm fault persists power unit requires inspection UNCALIBRATED WARNING The system will continue to operate but contact Talley Medical for recalibration as soon as possible Indicates mains power failure Silence alarm by C pressing MUTE button Reconnect to power supply to continue using mattress Caused by external RF fields Fault will clear when interference ceases 10 11 Static Foam Mattresses Installation and User Guidelines MATTRESSES 1 Ensure the bed frame to be used has no existing mattress components and is free from items which could cause damage to the SUPREMA DORMIRA mattress It should be noted that this mattress is inten
2. QUATTRO PLUS and PULSAIR CHOICE mattress replacement no min weight 200kg 31 stone max QUATTRO OVERLAY 26kg 4 stone min 160kg 25 stone max PULSAIR CHOICE mattress overlay 26kg 4 stone min 127kg 20 stone max B A S E SEQUENTIAL and B A S E RECLINER MAT cushions no min weight 127kg 20 stone max PULSAIR CHOICE cushion 26kg 4 stone min 102kg 16 stone max CPR FACILITY Fig 5 ALL ALTERNATING AIR PRESSURE MATTRESSES EXCEPT QUATTRO PRIME The CPR device is situated at the head end on the right hand side of the mattress viewed from foot end as indicated by arrows on the mattress tag For rapid deflation rotate the dial of the CPR device anti clockwise to click into the open position W If re inflating the mattress make sure the dial of the CPR device is rotated clockwise until clicked into the closed position a PATIENT TRANSPORT FACILITY Fig 6 ALL ALTERNATING AIR PRESSURE MATTRESSES EXCEPT QUATTRO PRIME Press MAX INFLATE button if applicable to 2 fully inflate the mattress When maximum pressure is achieved detach the mattress air supply hose from the power unit by rotating the mattress hose connector anti clockwise until the black lines align and 5 Pa pull the mattress hose connector away The mattress will remain inflated so supporting gt the patient Fig 6 All alternating air pressure mattresses can be used
3. To reconnect the mattress to the power unit after using the cushion switch off the power unit and disconnect the cushion adaptor Re connect the mattress air supply tubing to the power unit and switch on the power unit operation will automatically default to medium comfort setting 3 b To use the cushion whilst the mattress is operating simply attach the air supply hose from the cushion to the connection port on the side of the head end of the mattress by matching up the alignment markings and pushing together the CPC connectors until they click into place Fig 4 increasing the comfort control setting to maximum 6 The mattress and cushion will operate simultaneously allowing the simple transfer of the patient between the bed and chair N B It should be noted that the use of an alternating air pressure cushion will increase the patient s seated height by approximately 5cm and care should be taken to ensure the patient s comfort and security regarding height of foot and arm rests OPERATION BUTTONS The operation buttons on the face of the power unit provide the following functions Please refer to your individual model to find those relevant to your mattress system THERAPY MODE Pressing the THERAPY MODE button will toggle between ACTIVE 1 in 4 alternating air pressure cycle and STATIC for continuous low pressure therapy modes The selected therapy mode is shown on the display screen The default mode is ACT
4. power lead entry The system will display STARTING then INITIALISING whilst the mattress inflates this will take between 15 20 minutes 10 11 Note It is important that during the INITIALISING phase the mattress connector is not disconnected from the power unit If this is done the power unit must be switched off MUTE button pressed when the audible alarm is made the mattress connector re engaged and the power unit restarted If UNCALIBRATED is displayed when switched on the system will continue to operate but should be recalibrated as soon as possible When a bottom sheet is added to the mattress ensure that it is left loose to allow the mattress surface to conform with the patient s body as much as possible Allow the mattress to inflate before positioning the patient on the mattress Once inflated the system will automatically default to medium comfort setting 3 The comfort setting and the associated patient weight range guidelines are shown on the display screen The comfort setting can be adjusted using the up and down arrow buttons see COMFORT CONTROL on page 5 12 Place the carry bag and user manual in a safe place for future use INSTALLING CUSHIONS The QUATTRO PLUS QUATTRO OVERLAY AND PULSAIR CHOICE power units can all be used to operate their respective alternating air pressure cushions Place the cushion on a chair ensuring that it is placed the correct way up with the BACK labelling if
5. smoking whilst the system is in use to prevent the accidental secondary ignition of associated items which may be flammable such as bed linen In order for alternating air pressure mattress cushion systems to operate effectively please avoid placing objects on the surface that may obstruct the movement of air between the cells For the same reason please discourage people from sitting on the edge or on the end of the mattress whilst it is in use Do not place heavy objects on the surface of foam mattresses when not in use Do not allow sharp objects to penetrate the cover material Do not store in damp conditions The system is used as part of a pressure ulcer prevention programme not solely relied upon for this purpose Alternating pressure therapy should not be used for patients with unstable fractures gross oedema burns or an intolerance to motion It should be noted that the use of a cushion will increase the patient s seated height by approximately 5cm and care should be taken to ensure the patient s comfort and security regarding height of foot and arm rests Alternating Air Pressure Mattresses amp Cushions Installation and User Guidelines INSTALLING MATTRESSES 1 If using a mattress replacement system which is intended to completely replace the bed mattress remove any existing mattress from the bed frame If using a mattress overlay system ensure bed frame to be used has an existing mattress in place Plac
6. 2W 2 x 500mA QUATTRO ACUTE variable continuous QUATTRO PRIME variable continuous QUATTRO PLUS 15 minutes continuous QUATTRO OVERLAY 15 minutes continuous PULSAIR CHOICE 15 minutes mattress replacement 10 minutes mattress overlay cushion continuous ALTERNATING AIR PRESSURE MATTRESSES Construction Options Type Dimensions UK standard size BASE QUATTRO ACUTE QUATTRO PRIME QUATTRO PLUS QUATTRO OVERLAY Woven nylon 940 DTEX PU coated both sides PULSAIR CHOICE PVC coated 20z nylon 350 micron CELLS PU film COVER PU coated stretch nylon 255g m Standard Long Narrow QUATTRO ACUTE QUATTRO PRIME QUATTRO PLUS PULSAIR CHOICE MATTRESS REPLACEMENT Orthodifferential pleated air cells operating in a 1 in 4 alternating air pressure cycle O QUATTRO OVERLAY Orthodifferential air cells operating in a 1 in 4 alternating air pressure cycle O PULSAIR CHOICE mattress overlay One piece air cell set operating in a 1 in 2 alternating air pressure cycle QUATTRO ACUTE QUATTRO PRIME 1980mm x 880mm x 240mm QUATTRO PLUS PULSAIR CHOICE MATTRESS REPLACEMENT 1950mm x 880mm x 180mm O QUATTRO OVERLAY 1930mm x 855mm x 130mm Weight PULSAIR CHOICE MATTRESS OVERLAY 2000mm x 840mm x 128mm QUATTRO ACUTE 12 5kg O QUATTRO PRIME 10 6kg QUATTRO PLUS 9 4kg O QUATTRO OVERLAY 6 3 kg PULSAIR CHOICE 8 0kg mattress replacement 5 5kg matt
7. 500 er ek GROUP USA re order Talley USA 888 259 9994 rji 07 2011
8. IVE therapy To switch to STATIC MODE press THERAPY MODE button for 2 seconds until an audible tone is heard The system will then display PLEASE WAIT and requires approximately 2 minutes to initialise COMFORT CONTROL Air pressure is regulated within each of the cells throughout the cycle so that support posture and therapy are constantly maintained at optimum levels irrespective of a patient weight movement or position Equalisation of cell pressure automatically takes place at each stage of the 1 in 4 cycle again to ensure precise pressure and therapy is provided The automatic default comfort setting is level 3 However if the patient prefers a firmer or softer mattress increase or decrease the comfort control setting accordingly using the up and down arrow buttons soft 1 firm 6 The comfort setting and the associated patient weight range guidelines are shown on the display screen Check periodically to ensure patient Support and comfort DATA D Pressing the DATA button at any time switches the display into DATA mode Use the up and down arrow buttons to scroll through the product data and user information set Pressing the DATA button again returns the display to the previous mode NB Used for accessing information only does not affect mode of operation MUTE Press to silence alarm and to clear alarm display from display screen MUTE LOCK M Press to silence the alarm and to clear the alarm display from the display s
9. MIRA mattress 1970mm x 880mm x 127mm DORMIRA cushion 432mm x 432mm x 76mm SUPREMA mattress cushion 13 5kg 1 0kg DORMIRA mattress cushion 10 0kg 0 9kg SUPREMA mattress cushion 2 way DORMIRA mattress cushion 4 way MANUFACTURER S GUARANTEE All power units alternating mattresses cushions and foam mattresses cushions are covered by a 24 month manufacturer s guarantee 14 EXPLANATION OF LABEL SYMBOLS AND STATEMENTS A N ATTENTION Consult accompanying documents C Medical Devices Directive 93 42 EEC Electrical Protection Type B Class II Equipment Double Insulated R X Do not dispose of with the normal household waste SS please refer to www talleygroup com for further details This is a statement that alerts the user to the WARNING possibility of serious injury or other adverse reactions with the use or misuse of the device This is a statement that alerts the user to the CAUTION possibility of a problem with the system associated with its use or misuse ple This product is CLASSIFIED by Underwriters Laboratories Inc with respect to electric shock fire cus and mechanical hazards only in accordance with 1BD1 UL60601 1 and CAN CSA C22 2 No 601 1 USER MANUAL PART NUMBER 50 02 07 200 TALLEY GROUP LIMITED z Premier Way Abbey Park Industrial Estate Gi nA 2 Romsey Hampshire S051 9DQ England ALLEY Web www talleygroup com Brith l Haikan Europe re order 0 1794 503
10. TALLEY MEDIZ AL User Guidelines for Pressure Area Care Mattresses and Cushions QUATTRO ACUTE Systems QUATTRO PRIME Systems QUATTRO PLUS Systems QUATTRO OVERLAY Systems PULSAIR CHOICE Systems SUPREMA DORMIRA B A S E Seating Systems Contents INTRODUCTION CAUTIONS AND WARNINGS INFORMATION ALTERNATING AIR PRESSURE MATTRESSES amp CUSHIONS INSTALLATION AND USER GUIDELINES CARE AND MAINTENANCE FAULT FINDING FOAM MATTRESSES amp CUSHIONS INSTALLATION AND USER GUIDELINES CARE AND MAINTENANCE SPECIFICATIONS Introduction Page 11 11 13 Thank you for choosing to use a Talley Medical pressure area care product effective for the prevention and management of pressure ulcers The range of alternating air pressure products is as follows QUATTRO ACUTE mattress replacement system for patients at very high risk of pressure ulcers QUATTRO PRIME mattress replacement system for patients at very high risk of pressure ulcers QUATTRO PLUS mattress replacement system for patients at high risk of pressure ulcers ulcers The range of static foam products is as follows PULSAIR CHOICE mattress overlay for patients at medium risk of pressure ulcers PULSAIR CHOICE cushion seat cushion for patients at high risk of pressure ulcers B A S E SEQUENTIAL seat cushion for patients at very high risk of pressure ulcers B A S E RECLINER MAT seat and back c
11. applicable facing the back of the chair Secure cushion to chair using adjustable straps where available 1 The PULSAIR CHOICE cushion is connected to the PULSAIR CHOICE power unit exactly as described in points 5 to 9 above and will take approx 30 seconds to inflate The B A S E SEQUENTIAL and B A S E RECLINER MAT cushions can be used with the QUATTRO PLUS and QUATTRO OVERLAY mattress systems either operated with the mattress or using a cushion adaptor directly with the power unit as follows a To use the cushion directly with the power unit connect the cushion adaptor to the power unit rotating the adaptor clockwise until the green indicator fills the indicator hole on the power unit connector Connect the air supply hose from the cushion onto the adaptor matching up the alignment markings Fig 3 Ensure that this has been correctly clicked into place otherwise a leak may occur Attach the power lead plug into the mains outlet and switch the power on at the side of power unit adjacent to the power lead entry Once inflated approx 5 minutes increase the comfort control setting to maximum 6 NB If using the power unit from an operating mattress system first press the MAX INFLATE button if applicable to fully inflate the mattress then rotate the air supply hose connector anti clockwise and detach from the power unit to seal the air within the mattress to leave it fully inflated
12. creen The power unit will automatically lock 2 minutes after the last button operation when running to prevent the inadvertent operation of button functions except MUTE as indicated by L on the display screen Press and hold the MUTE UNLOCK button until the power unit beeps if further button operation is needed i e comfort setting The power unit will lock again 2 minutes after the last button operation NB After power failure switching the power off pressing MUTE cancels the system s previous settings When power returns the default setting of ACTIVE mode comfort setting 3 is invoked Note that previous settings are automatically cancelled if the duration between switch off and switch on is greater than 12 seconds If power returns before a period of 12 seconds has passed and the MUTE button has not been pressed the system will return to the previous mode of operation MAX INFLATE Necessary for some nursing procedures the MAX INFLATE mode inflates the mattress to maximum static pressure 50mmHg for a period of 15 minutes After pressing the MAX INFLATE button to inflate mattress the system displays PLEASE WAIT followed by READY and a 5 second audible tone when maximum pressure is achieved and MAX INFLATE is shown on the display screen After 15 minutes the system automatically returns to the active mode of operation USER WEIGHT GUIDELINES QUATTRO ACUTE and QUATTRO PRIME no min weight 250kg 39 stone max
13. ded to completely replace the existing mattresses 2 Place the SUPREMA DORMIRA mattress on the bed frame NB If using the SUPREMA mattress ensure that the mattress is positioned with the printed cover uppermost with the grey non slip base facing downwards 3 Place the user manual in a safe place for future use The SUPREMA DORMIRA mattress can be used on profiling bed frames slatted frames in filled frames and divans A gap of 2 5cm either side of the mattress should not be exceeded when side rails are deployed CUSHIONS Place the SUPREMA DORMIRA cushion onto a chair If using the SUPREMA cushion ensure that it is positioned with the green cover uppermost with the grey non slip base facing downwards N B It should be noted that the use of a foam cushion will increase the patient s seated height by approximately 5cm and care should be taken to ensure the patient s comfort and security regarding height of foot and arm rests Care and Maintenance The SUPREMA DORMIRA mattress cushion will provide years of efficient service if it is cared for in the following way COVER Always keep the mattress cushion cover as clean as is practicable The material is waterproof and vapour permeable Inspect top cover for signs of damage or wear which could result in the contamination of the interior e g tears holes damage to seams or zips underside staining etc The frequency of the
14. e the mattress on the bed frame existing bed mattress If using a powered mattress ensure the mattress hose exits at the foot end on the right hand side and the coloured cover is uppermost Secure the mattress to the bed frame existing mattress using adjustable straps pass straps beneath or around bed frame mattress platform bed mattress and secure with the buckles Fig 1 allowing for moving components on electric bed frames This is important as it stops the mattress becoming detached from the bed frame existing bed mattress and causing possible injury to the patient Suspend the power unit from the foot board of the bed having first adjusted the hanger brackets as instructed at the rear of the unit Alternatively the power unit may be placed on the floor Y ia IK ij A i i Fig 1 Fig 2 Attach the air supply hose to the power unit by aligning the black line on the air supply hose connector with the black line on the power unit connector and pushing together Rotate the mattress hose connector clockwise until the green indicator fills the indicator hole on the power unit connector Fig 2 Plug the smaller end of the power lead into the left hand side of the power unit and the other end into mains outlet in wall Ensure that the mains lead and tubing cannot become trapped in the bed frame Switch the power on at the outlet and at the side of power unit adjacent to the
15. ge INTERIOR Inspect the mattress cushion interior for signs of damage or contamination e g staining or evidence of fluid ingress The frequency of these checks should be at each decontamination process i e between patients or patient occupancy or weekly for longer term patients Contaminated mattresses cushions should be removed from service and disposed of in accordance with local policy Regular turning will prolong the life of the mattress The SUPREMA has a 2 way turning cycle and the DORMIRA has a 4 way turning cycle Please refer to instructions printed on cover SYSTEM DESIGN LIFE 3 years HANDLING AND STORAGE Handle with care Please report instances of damage or impact to Talley Service Department The mattress should be stored flat Temperature Operational 10 C to 40 C Storage 10 C to 55 C 13 TRANSPORTATION OF MATTRESS SYSTEM The mattress should be transported flat and not rolled Mattresses should not be stored more than 6 high as this can potentially be a safety hazard when handling To protect the covers mattresses should be packed in minimum 200 micron clear polythene Specifications ALTERNATING AIR PRESSURE POWER UNITS Construction ABS Plastic Dimensions 335mm 13 2 x 233mm 9 2 x 165mm 6 5 Weight 3 4 kg 7 5 Ibs Mains Cable 5 metres 16 5 Electricity Supply 240V 50Hz Full Load Wattage Fuse Rating Cycle Time USA 110V 6OHz Saudi Arabia 240V GOHz 1
16. nation process i e between patients or patient occupancy or weekly for longer term patients Care should be taken to avoid puncturing cover with objects such as needles scalpels pat slides acrylic nails etc The cover may be removed and cleaned in accordance with The Revised Healthcare Cleaning Manual June 2009 subject to the following action Following the use of a detergent and or disinfectant solution the mattress cover should be rinsed with clean water using a clean cloth and allowed to dry Frequent or prolonged exposure to high concentration disinfectant solution may reduce the useful life of the mattress cushion cover Where high concentration disinfectants e g 10 000ppm chlorine releasing agent CRA or combined chlorine releasing agent CRA and detergent solutions are used to clean and disinfect blood or other body fluids mattresses cushions should be rinsed with clean water using a clean cloth This will help prevent any possible long term compatibility issues associated with disinfectant residues Alternatively disinfection may be achieved by laundering at temperatures not exceeding 65 C for 10 minutes or 73 C for 3 minutes which may include a chlorine rinse Do not use abrasive cleaners solvents or alcohol based cleansers e g Phenicol Hibiscrub Clearsol Stericol Hycoline as these will destroy the cover materials Do not iron Ensure that the mattress cushion is thoroughly dried before remaking the bed or placing in s
17. on profiling bed frames slatted frames in filled frames and divans Back rests or pillows for support should be placed beneath the mattress to allow uninterrupted body contact with the mattress surface Place the bottom sheet loosely on the mattress to allow the mattress surface greater contact with the patient s body Avoid using fitted sheets The use of incontinence sheets excessive bedding beneath the patient may reduce the pressure relieving effect of the mattress To remove air from the mattress when dismantling the system use the CPR facility as described above all mattresses except QUATTRO PRIME to remove air from the QUATTRO PRIME detach mattress air supply hose from power unit by rotating mattress hose connector anti clockwise until black lines align and pull mattress hose connector away Care should be taken when raising and lowering bed safety side rails in order to avoid possible interference with the CPR and cushion connection port where fitted A gap of 2 5cm on either side of the mattress should not be exceeded when side rails are deployed Care and Maintenance COVER Always keep the mattress cushion cover as clean as is practicable The material is waterproof and vapour permeable Inspect top cover for signs of damage or wear which could result in the contamination of the interior e g tears holes damage to seams or zips underside staining etc The frequency of these checks should be at each decontami
18. pped or there is a fault in the system Switch power off press MUTE button then switch power on again If the fault re occurs contact Talley Medical LOW PRESSURE warning will alert if pressure falls below the minimum allowable levels Check that the hose is connected to the power unit correctly Check that the CPR device is fully closed i e the dial of the CPR device is rotated clockwise until clicked into closed position A Check that the internal cells are connected and that no cell is punctured Press the MUTE button to clear alarm Note that the alarm will reoccur if the fault persists if the fault re occurs contact Talley Medical EMI fault indicates that the unit detects the pressure sensor amplifier is adversely affected by external RF fields This will clear when interference ceases OTHER ALARMS PUMP OR TRIAC fault indicates a pump control failure or an open pump coil fault Should this occur contact Talley Medical UNCALIBRATED contact Talley Medical for recalibration If you have any queries relating to this system please contact Talley Medical or your local authorised dealer LOW PRESSURE FAULT ROTOR SYSTEM FAULT PUMP or TRIAC FAULT Check air supply hose is correctly attached to power unit i e mattress hose connector is correctly aligned with power unit connector and is fully rotated clockwise so green indicator fills window on power unit connector Reset power unit by
19. ress overlay ALTERNATING AIR PRESSURE CUSHIONS Construction Type Dimensions UK standard size Weight BASE B A S E SEQUENTIAL RECLINER MAT cushion Woven nylon 940 DTEX PU coated both sides O PULSAIR CHOICE cushion PVC coated 20z nylon 350 micron INNER B A S E SEQUENTIAL RECLINER MAT cushion PVC bellows within punched CMFR foam O PULSAIR CHOICE cushion PU film COVER PU coated stretch nylon 255g m7 B A S E SEQUENTIAL RECLINER MAT cushion 8 20 rows of 6 bellow cell strips operating in a 1 in 4 alternating air pressure cycle PULSAIR CHOICE cushion One piece air cell set operating in a 1 in 2 alternating air pressure cycle B A S E SEQUENTIAL cushion 430mm x 430mm x 70mm B A S E RECLINER MAT cushion 1080mm x 430mm x 70mm PULSAIR CHOICE cushion 430mm x 430mm x 40mm B A S E SEQUENTIAL cushion 1 9kg O B A S E RECLINER MAT cushion 3 9kg O PULSAIR CHOICE cushion 1 3kg STATIC FOAM MATTRESSES amp CUSHIONS Construction Cover Dimensions Weight Turning cycle SUPREMA mattress cushion 57 63kg m density viso elastic foam over a base of 38 42 kg m nominal density 150 190N nominal hardness SAFEGUARD CMHR O DORMIRA mattress cushion Multi layered SAFEGUARD CMHR foam 38 42 kg m nominal density 110 140N 100 130N nominal hardness PU coated stretch nylon 255g m SUPREMA mattress 1970mm x 880mm x 160mm SUPREMA cushion 432mm x 432mm x 76mm DOR
20. se checks should be between patients or patient occupancy weekly for longer term patients Care should be taken to avoid puncturing cover with objects such as needles scalpels pat slides acrylic nails etc The cover may be removed and cleaned in accordance with The Revised Healthcare Cleaning Manual June 2009 subject to the following action Following the use of a detergent and or disinfectant solution the mattress cover should be rinsed with clean water using a clean cloth and allowed to dry Frequent or prolonged exposure to high concentration disinfectant solution may reduce the useful life of the mattress cushion cover Where high concentration disinfectants e g 10 000ppm chlorine releasing agent CRA or combined chlorine releasing agent CRA and detergent solutions are used to clean and disinfect blood or other body fluids mattresses cushions should be rinsed with clean water using a clean cloth This will help prevent any possible long term compatibility issues associated with disinfectant residues Alternatively disinfection may be achieved by laundering at temperatures not exceeding 65 C for 10 minutes or 73 C for 3 minutes which may include a chlorine rinse Do not use abrasive cleaners solvents or alcohol based cleansers e g Phenicol Hibiscrub Clearsol Stericol Hycoline as these will destroy the cover materials Do not iron Ensure that the mattress cushion is thoroughly dried before remaking the bed or placing in stora
21. torage INTERIOR COMPONENTS Check air cells and mattress cushion interior for signs of damage or contamination e g staining or evidence of fluid ingress The frequency of these checks should be at each decontamination process i e between patients or patient occupancy or weekly for longer term patients Care should be taken to avoid puncturing air cells with objects such as needles scalpels pat slides acrylic nails etc The individual cells can also be wiped clean with a mild antiseptic solution All cells are replaceable and can be obtained easily from Talley Medical Do not immerse the PULSAIR CHOICE mattresses cushion in water The QUATTRO range of mattresses are fully launderable but ensure the CPR device is fully closed i e the dial of the CPR device is rotated clockwise until clicked into the closed position 8 prior to laundering to prevent water entering the air cells tubing POWER UNIT Always disconnect the power unit from the electricity supply before cleaning Power units can be wiped down with a detergent or disinfectant solution or wipe Do not use solvents The unit contains no user serviceable parts and should only be serviced by a competent electrical engineer or returned to Talley Medical your local authorised dealer All Talley Medical products should be serviced every two years by Talley Medical or authorised dealer in order to comply with warranty conditions It is the customer s responsibility to ens
22. ure the following prior to collection ithe system is cleaned of any obvious contaminants contamination status is documented assistance is given to Talley Medical personnel to bag the equipment if the mattress has been in a known or suspected infectious environment SYSTEM DESIGN LIFE 5 years HANDLING AND STORAGE Handle with care Please report instances of damage or impact to Talley Service Department Temperature Operational 10 C to 40 C Storage 10 C to 55 C Humidity and other environmental conditions refer to the Service Manual TRANSPORTATION OF MATTRESS SYSTEM The mattress should be loosely rolled lengthwise with the cover innermost taking care not to strain the umbilical It can then be transported and stored in the carry bag with the power unit mains cable and this booklet Do not stack bagged mattresses more than two high to avoid strain on the umbilical Fault Finding All alarms can be reset by pressing the MUTE button once This also cancels the fault displays MUTE also silences the audible alarm All systems have a fault log that records the last 5 faults via the DATA display mode If problems reoccur contact Talley Medical AC FAIL fault indicates a mains power failure an audible alarm will sound if power is interrupted e g power unit switched off power cut disconnection of mains lead Press MUTE or re connect to power supply ROTOR SYSTEM fault indicates the automatic sequential cycle has sto
23. ushion for patients at very high risk of pressure QUATTRO OVERLAY mattress overlay system for patients at high risk of pressure ulcers PULSAIR CHOICE mattress replacement for patients at high risk of pressure ulcers SUPREMA mattress replacement and cushion for patients at medium risk of pressure ulcers DORMIRA mattress replacement and cushion for patients at risk of pressure ulcers Cautions and Warnings Before using alternating air pressure mattress cushion systems ensure that the electricity supply is of the type indicated on the power unit the mains lead is free from damage and is positioned so as not to cause an obstruction the system is not used in the presence of flammable anaesthetics For 110V units only means to isolate the power unit from the electricity supply shall be carried out via disconnecting the plug attached to the non detachable mains cord from the wall socket The equipment conforms to IEC 60601 1 2 for electromagnetic interference however should the equipment be subjected to electromagnetic interference outside this then the unit should be recalibrated by a qualified service technician Do not place mattress cushion or power unit on or near a heat source Do not use with hot water bottles or electric blankets Although the materials used in the manufacture of all components of the mattress cushion system comply to the required fire safety regulations Talley Medical advise against

Download Pdf Manuals

image

Related Search

Related Contents

Extech Instruments Thermometer 445715 User's Manual  Tristar KB-7147 cool box  Removal and Installation  Samsung DVD-HR720 manual de utilizador  校正サービス - アズビル金門製作所  Otoport Screener  JamVOX Owner's manual  Professional Weather Center  

Copyright © All rights reserved.
Failed to retrieve file