Contents
1. P6 33 ST IM84c SES 113 and 225 AUTOCLAVES 2 0 Getting started WARNING Read these instructions before use and keep them in a safe convenient place for future reference The use of this autoclave should be under the control of a person with adequate training the Autoclave Manager The Autoclave Manager is responsible for ensuring that the correct sterilization cycle is used for the items being sterilized ensuring that records are kept correctly and keeping the amended PIN code confidential refer to Appendix 7 Special functions The autoclave should only be used as specified by Eschmann as detailed within these instructions Never tamper with bypass or interfere with any of the safety features There are no user serviceable parts inside the autoclave Eschmann Equipment is not responsible for malfunction or a reduced level of protection provided by the equipment if it is used in a manner not specified in these instructions Only use Eschmann accessories and mains leads Any spare parts required must be supplied by Eschmann They must be fitted in accordance with the Service Manual or the instructions supplied with them by EschmannCare engineers or engineers who have been trained by Eschmann Autoclave Manager The person who is ultimately accountable for the operation of the premises Depending on the nature of the organisation this may be the owner or other person of similar authority In small autonomous installati
2. JAM Sister for use gt Q ESCHMANN LittleSister P2 33 Contents Page INMOOUCHON zart 3 Suitable loads 2s 2cin cease u Hassan ee neueren 3 ESchimannCare 0c cheese heehee KERR Ge eke ikea 3 1 0 Controls and displays ccsc0 vere cen red eve RR an 6 2 0 Getting started erstere 7 3 0 Running a cycle 222intstesserdertetagetadeadeideiees 8 4 0 Direct Data Download instructions 10 5 0 Printer instructions 000 0c eee eee 13 6 0 Technical data achece 0bentesseencdasesebe one te ee ve 15 APPENDIX 1 Installation 2 44 2404 04 2 18 APPENDIX 2 Warnings and cautions 0005 19 APPENDIX 3 Loading the autoclave 44 21 APPENDIX 4 Daily Testing Vacuum B type autoclaves 22 APPENDIX 5 Maintenance 0002 cece eee eee 23 APPENDIX 6 Cycle errors and faults and Display messages 24 APPENDIX 7 Special Functions 000 eae eee 26 APPENDIX 8 Accessories and spares 000 32 Figures BEE Fig Fig Fig Fig Fig Fig Fig Fig ONOarwoNn Identification of the main parts of the autoclave 4 The main information display and control buttons 5 Filling and draining the reservoirs 04 5 DDD Controls active connected 10 DDD Download button symbols 10 Cycle Select and various display screens 11
3. 5 ea button en __ Progress Stop E SESCHMANN bar Fig 4 Controls active Connected Click the disk button Fig 4 to select the type of download required The image shown next to the disk button indicates which files you are about to download refer to Fig 5 For single files enter the cycle number in the box provided af P dl Single file Files not downloaded All files Fig 5 Type of download symbols After selection click the green button to start the download If required click the red button to stop The Progress bar Fig 4 shows the download progress and the following Off Normal Green File retrieval is complete all files retrieved OK Red An error occurred that prevented the retrieval Yellow Retrieved all files but at least one was corrupt To view files browse to the user allocated folder storing these files If the letter E appears by the cycle number this file is corrupt You may try to retrieve this again as a single file If in doubt refer to the instructions within the DDD Application CD P10 33 ST IM84c SES 113 and 225 AUTOCLAVES Normal Cycle Select display prior to running a cycle Door message Indicates cycle l Other door message Cycle selected with drying ee Cycle type Cycle temperature Statement 121 or 134 hi Lie y Ha PILI LIPEJ Time of day clock Other statment or Cycle counter Cycle number Cycle type B or N Normal
4. 318 kPa B121 121 C to 124 C approx 215 kPa Sterilization temperature and pressure held i e at temperature and pressure as above 4 Pressure maintained as above the precondition hold Cycle without drying pressure returned to atmospheric pressure Cycle with drying drying pulses of varying pressure Note For cycles with drying there are several drying pulses the chamber is then returned to atmospheric pressure P18 33 ST IM84c SES 13 and 225 AUTOCLAVES APPENDIX 1 Installation These Autoclaves should be installed by an EschmannCare or Eschmann Trained Engineer For Warranty Registration Free Installation and Free User Training contact EschmannCare on 01903 765040 Programmes and autoclave configuration can be adjusted by an EschmannCare or Eschmann Trained Engineer on request To protect against tampering access for such adjustment can only be gained by use of a PIN code refer to User PIN number below The Autoclave should be lifted or moved by at least two people Observe good manual handling techniques at all times with hands placed under the base each side of the autoclave Unpacking Remove the outer packaging and all packaging pieces carefully and place the autoclave on a flat water and heat resistant surface Ensure there is a working clearance of 50mm around the unit Do not cover the ventilation ports at the rear or side of the autoclave The following items should be checked and placed to
5. Various display screens 20000000 00 ee 12 Printer control panel o s4e 0404 dee 24 ae 13 ST IM84c SES 113 and 225 AUTOCLAVES Introduction The Eschmann Little Sister SES 113 and 225 autoclaves have been designed to meet the needs of the healthcare professional and have been designed in accordance with European Standards e g BS EN 13060 and Department of Health guidelines SES 113 Autoclave The SES 113 is an N type displacement autoclave with an 11 litre chamber It is intended for the sterilization of solid instruments which have not been wrapped SES 113 B Autoclave The SES 113 B vacuum is a B type vacuum autoclave with an 11 litre chamber It is intended for the sterilization of solid and hollow instruments pouched or wrapped instruments and porous loads SES 225 B Autoclave The SES 225 B vacuum is a B type vacuum autoclave with an 22 litre chamber It is intended for the sterilization of solid and hollow instruments pouched or wrapped instruments and porous loads Suitable loads The following table details the types of load that may be successfully processed in each model Please refer to the table in the Technical Data section for details of the cycles available on each model Type of load and cycle type SES 113 SES 113 B Vacuum SES 225 B Vacuum Solid instruments for immediate use N Solid instruments processed for aseptic storage N Hollow cannula instruments B Single wra
6. at all times when the autoclave is not in use To close the door fully With the handle vertical push the door shut firmly and turn the door handle clockwise until horizontal 3 o clock position The door indicator light will stop flashing and will not be illuminated Loading Note Prior to sterilization instruments should be thoroughly cleaned ideally in an Eschmann Instrument Washer Disinfector Open the door fully The door indicator light will flash indicating the door should be closed after loading together with a display message Close Door Following the guidance provided in APPENDIX 3 and local procedures load the autoclave with the items to be sterilized Reservoir warning If a Reservoir Warning Indicator detailed on page 6 is illuminated fill or drain reservoir as indicated referring to page 7 Fill the reservoir or page 9 Draining a reservoir as required Closing door Close the door fully The green door indicator light will stop flashing and will not be illuminated Cycle selection The digital display indicates the default cycle If another cycle is required press the Cycle select button until the required cycle is displayed Ensure the cycle is appropriate for the load refer to section 6 Starting cycle Press the Start cycle button The green door indicator light will illuminate indicating the door is locked and cannot be opened until cycle completion The printer if fitted wi
7. load x Small porous loads Full porous load Hollow load A Dryness solid load Dryness porous load Residual air x KK KK KK XK X in compliance with all applicable clauses of BS EN 13060 ST IM84c P17 33 LittleSister 6 14 Detailed N type cycle N type autoclave information TSS TSE Pressure Typical N TYPE cycle N type autoclave pressure diagram showing stages and printer codes diagram O56 Pressure rises i e heating to sterilization temperature N134 134 C to 137 C approx 318 kPa N121 121 C to 124 C approx 215 kPa 5 Sterilization temperature and pressure held i e at temperature and pressure as above Chamber pressure reduced to atmospheric i e condensing and draining chamber 6 15 Detailed B type cycle information T S So ESE Typical B TYPE cycle a TPP TDS pressure diagram showing R STD or ETD stages and printer codes if no drying TPV u i ETD Cycles with drying only Stage on Stage detail diagram Pressure drops to below 15 kPa Initial vacuum Pressure increase as chamber i e boiler fills with the required volume of water Initial pressure rise during initial heating temperature is pressure related B134 pressure is 175 to 185 kPa B121 pressure is 125 to 135 kPa Pressure drops to below 30 kPa the steam penetration pulse Pressure rises i e heating to sterilization temperature B134 134 C to 137 C approx
8. minute or a fail if greater than 1 3mbar minute T3 Steam penetration MUST NOT be used to process loads This test runs a cycle suitable for the B type Helix test basically a B134 C cycle without drying Check the test device is placed in the usable space in an empty chamber before starting the test Also refer to Appendix 4 User Can Select When this function is enabled the operator will be able to select any of the two options listed i e T1 or T3 as a cycle using the cycle select button To enable or disable this option press the Up or Down buttons to select this option indicated by the cursor then press the Enter Select button An asterisk will indicate when this option is enabled The cursor will return to the T1 option after pressing the Enter Select button To return to the CYCLE menu press the Cancel Exit button or select Exit Note The user should be trained how to use T1 and T3 they are user selectable until disabled by the Autoclave Manager Delayed start For this option do not enter the CYCLE menu until you have completed the following open the autoclave door load the chamber select the required cycle and check reservoirs do not required filling or emptying Press Enter Select the CYCLE screen is displayed Press Enter Select the CYCLE menu is displayed Press Up or Down to move the cursor to the Delayed start
9. mm length by cross section shown Chamfers 20 x 20 mm 197 mm 197 mm 6 6 Standards compliance These autoclaves have been designed in accordance with the following standards BS EN 13060 2004 BS EN 61326 1998 Class A BS EN 980 2003 BS EN 1041 1998 PD 5500 2006 EN 61010 2 041 1997 BS EN 61010 1 1993 BS EN ISO 14971 2007 6 7 Applicability This manual applies to the following autoclave models SES 113 Autoclave N type from Serial Number L1NA or SINA or SIXA REF 87 11 SES 113 B Vacuum Autoclave B type from Serial Number L1BA or S1BA REF 87 12 SES 225 B Vacuum Autoclave B type from Serial Number L2BA or S2BA REF 87 13 indicate model variant numbers 6 8 Environmental considerations During cleaning and disinfection procedures any potentially contaminated waste materials produced during these actions should be handled in accordance with local procedures and National legislation for the disposal of potentially contaminated waste At the end of the working life of the autoclave it should be dismantled and recycled as much as possible in line with the recommended procedure available from Eschmann Equipment and in accordance with local procedures and National legislation 6 9 Technical Lifetime The Eschmann Little Sister SES 113 and 225 autoclaves are designed to have a technical lifetime of 36 000 operating cycles 10 years normal use subject to their being
10. option press Enter Select The display shows Set Start Time with the hour and minutes indicated by hh mm Use the Left or Right buttons to move the cursor to each time element and adjust them with the Up or Down buttons When the start time is correct press Enter Select Press Cancel Exit twice You will be prompted to close the door When the door is closed the Delayed Start display will be shown illustrated below The message Delayed Start and the start time are shown Note The cycle select button cannot be used after setting the delayed start the cycle must be selected first as detailed above On completion of the delayed start cycle the autoclave is ready for normal use To cancel a delayed start press the Cancel Exit button or the Cycle Start button The display will show Delayed Start alternating with Abort and possibly CAUTION HOT Pressing the Cancel Exit button whilst Abort is displayed stops the cancellation Pressing the Enter select button whilst Abort is displayed cancels the delayed start If a delayed start is cancelled the Autoclave will record a User abort for the Cycle Error number 2 and increment the Cycle count ST IM84c P31 33 LittleSister i Press Enter Select to accept I or press Cancel Exit to return 1 to Cycle Select display I Up Down
11. or overridden a Door closed position sensor and door locks position sensors These prevent a cycle starting if the door is not closed and locked Pressure Relief Valve safety valve This valve safely releases excess pressure up to a maximum of 2 85 10 barG Also refer to section 6 11 for autoclaves fitted with an adjustable pressure relief valve Independent cycle control and monitoring There are two microcomputers independently monitoring each other and other autoclave components Either is able to stop a cycle and place the autoclave into a safe condition in the event of a failure In the case of a venting valve failure the autoclave may not be able to eject water and steam i e pressure will remain until cool and the door will remain locked until the pressure temperature has dropped to a safe level Overheat Control The microcomputer operates in conjunction with an independent manual reset thermostat to protect the heating element from overheating Band Heater Thermostat This prevents the band heater exceeding a preset temperature Pressure door switch Prevents the door being opened when the residual chamber pressure is too high by removing power to a door lock so that it locks the door closed Safety Checks The user should ensure that the following safety checks are carried out Daily by User i ii iii iv Check that the door opens and closes easily Check the door seal and chamber face for any signs of d
12. papers recess with the roll unwinding from the back and bottom of the roll when placed in the recess left picture below Note the paper can only be printed on one side the outside of the roll do not place the roll in back to front iv Close the lid by applying equal amounts of pressure on each side ensure the lid is in the locked position 9 P14 33 Do not close the lid with one finger as shown below V Now tear the spare paper away by pulling the printout down towards the tear bar and tear from one side to the other across the serrated edge 5 3 3 Removing a printout Push the paper feed button to advance the paper about 5 mm if required To remove a printout from the printer pull the printout down towards the tear bar and tear from one side to the other across the serrated edge as 5 3 2 v 5 3 4 Printer door During use the printer door can be closed the printout will feed through the recess in the bottom of the door When closing the door hold the printout down and take care not to trap any paper in the door 5 4 Printer maintenance Power On Self Test The self test procedure is initiated by supplying power to the printer i e switch the autoclave on whilst depressing the paper feed button Fig 8 until the printer starts printing the test print Head Thermal Limit After extensive printing the print head temperature may rise to an unacceptable level The Status LED will report when this o
13. the applicable parts of HTM2010 and DB9804 are available from EschmannCare The EschmannCare team is staffed and equipped to provide advice and assistance during normal office hours To avoid delays when making enquiries please quote the Model and Serial Number refer to Fig 1 item 20 and the error number displayed if applicable All correspondence relating to the after sales service of Eschmann Equipment should be addressed to EschmannCare Eschmann Equipment Peter Road Lancing West Sussex BN15 8TJ Tel 44 0 1903 765040 Fax 44 0 1903 875711 CarePlan service cover EschmannCare CarePlan service cover should be taken out for your Autoclave An autoclave will only continue to function correctly if it is properly serviced and maintained CarePlan provides the assurance that the exacting conditions necessary for sterilization are maintained throughout the autoclave s working life Safety Checks Refer to General safety and Safety Checks in Appendix 2 Cycle History When the message Please Download Your Cycle Data For Archiving appears consult the Direct Data Download Software CD package supplied with the autoclave and download the autoclave data to a PC Certification Eschmann can offer certification to PSSR 2006 regulations as part of our CarePlan maintenance packages This certification entails inspecting the pressure vessels and testing a number of safety critical devices to ensure continued safe operat
14. to help initiate instrument traceability within the surgery environment Examination sized accessories enable a one set of instrument per tray per patient policy to be adopted Full details can be found in the accompanying EiMS brochure also refer to Appendix 8 Unwrapped solid items N or B type cycles Place unwrapped solid items loose on trays or into hands free instrument transfer HFiT cassettes Do not overload trays or HFiT cassettes TECHNICAL DATA details maximum tray loads and maximum load Avoid bunching items together and ensure all items are positioned so that they do not touch and drain freely Only use trays or HFiT cassettes supplied by Eschmann Loaded trays and HFiT cassettes should then be placed in the sterilizing chamber All items must be positioned so that they drain freely and do not trap rising air bubbles Unwrapped items are for immediate use only For details on the choice of cycle refer to Section 6 Hollow instruments including items with narrow lumens e g dental handpieces porous loads and single wrapped e g pouched hollow or solid instruments B type cycles ONLY Items in pouches should only be sterilized using the special pouch rack which holds 9 pouches with only one instrument per pouch otherwise instruments touch together in the pouch restricting steam from penetrating all surfaces Avoid packing pouches too tightly and note that paper film pouches must be loaded such t
15. to select Enter Select l to accept Accept Exit to return I I I to CYCLE SCREEN Cycle Options Cycle History Test Cycl Delayed Start Exit 1 Up Down to select 1 Enter Select to accept i Download Ei Cycle History Frimt amp Download Exit T ie hae Enter Fin Code 0 Cycle Options H 134 H 121 H 134 we ala E 134 E 121 Cycle History st Successful 103 Las SE Failed 10 Exi Cycle History Day leek Honth oe Month i Up Down to adjust digit Enter Select 1 to accept digit and move to next digit i Selected digit becomes an E Right then Up tick or Down cross Right again then Enter Select sets default I l indicated by a 3 i I Left moves cursor back to enable Up Down t 1 Cancel Exit to return to CYCLE SCREEN d Cycle Options H 134 H 121 Hoo 124 H 121 E 134 B 121 T U E LH be i Up Down to select 1 Enter Select to accept I J i Press Cancel Exit to return i to CYCLE menu or select and 1 1 accept the Exit option i Up Down to select 1 Enter Select to accept I Note The currently ae option is marked with an Note If a printer is not installed Cycle History i t j Remi displ h t will assume the Download option is required erminder display when se Cycle History Cycle History pS Month Please Download Your Don t Remind me Cycle Data for Ext i
16. used in accordance with these instructions and subject to proper maintenance and repair 6 10 International Symbols zn_ This symbol indicates that the equipment is for use on alternating current This symbol indicates that fuses adjacent to the symbol have a rating and type as detailed P15 33 LittleSister N This symbol warns the user to read the accom panying documents these Instructions for use SN This symbol indicates the unit serial number is as indicated adjacent to the symbol REF This symbol indicates the catalogue number is as indicated adjacent to the symbol This symbol indicates that the date of manufacture is as indicated adjacent to the symbol This symbol indicates the manufacturer is as indicated adjacent to the symbol This symbol warns the user of potentially hot surfaces This symbol indicates that this product was placed on the market after 13th August 2005 At the end of its working life it should be deposited at an appropriate facility to enable recycling For BE information on how to recycle this product responsibly contact Eschmann 6 11 Miscellaneous For Conditions for safe use refer to Appendix 2 The autoclave will be ready for use after initial installation and setup in less than 1 0 minute The pressure relief valve is nonadjustable and set to 2 85 bar Catalogue numbers ending 4 80 or 3 80 have an adjustable pressure relief valve set d
17. 5 33 LittleSister Service Due The service due date will be set in the factory for 10 months after manufacture or 1500 cycles i e whichever comes first When the date or cycle count is reached the digital display will show the message Service Required Please Contact Eschmann This message will be displayed for 5 seconds and appears when the power is switched on and after each cycle Warranty Reminder The warranty reminder will be set in the factory to appear until 30 Cycles have been completed If still active the digital display will show the message Welcome Please contact Eschmann to register for FREE installation and training Whilst still active the message will be displayed for 30 seconds when the power is switched on Please contact Eschmann and you will be advised which buttons to press when this message is displayed to cancel the message irrespective of the cycle count Warranty arrangements will also be advised and initiated Cycle Errors Table Error P26 33 Cause Power failure during cycle User cycle abort Door lock failed Initial vacuum pulse timeout Water in chamber at cycle start Failed to fill boiler Overheat cutout has tripped or a leak Boiler water level low during cycle Temperature pressure error Temperature error Not enough water in fresh water tank Too much water in waste water tank Air detected in chamber Pressure fell too much during leak test Pressur
18. 875711 Overseas Customers Contact your local distributor In case of doubt contact Eschmann Equipment ST IM84c P3 33 LittleSister is es Tem m OL Part rear view Part rear view Autoclave door Serial Label front Door handle Overheat indicator Process display Fig 2 Waste water drain Start button Fig 2 Waste water reservoir Fixed Indicators Fig 2 Fresh water drain Digital display Fig 2 Fresh water reservoir Slides out to fill Cycle select button Fig 2 Drain release clip Control buttons Fig 2 Serial label back Printer Mains fuses Mains on off switch IEC Mains connection Adjustable foot Antibacterial air filter B type units only USB port Basic instructions UK models only OANDOABRWN Fig 1 Identification of the main parts of the autoclave P4 33 ST IM84c SES 113 and 225 AUTOCLAVES Progress indicators inner ring Process display symbols outer ring Start cycle button Indicators Control Button detail and function Left Right a Cycle select button Enter Select Down Cancel Exit Digital display Figs 5 and 6 Fig 2 The main information display and control buttons shown with all indicators illuminated and a typical Cycle select display Filling the fresh water reservoir slide out Draining the waste water reservoir fixed Each reservoir has a capacity of 3 36 litres Drain tube before connection direct end in
19. Archiving Note User Can Select can be enabled or disabled when selected using Enter Select When accepted this is indicated by an Typical display when a test is selected Test Cycles Measuring Se pTi Chamber Integrity 1 Up Down T3 Steam Penetration Ito Select User Can Select f or select and accept 1 Remain 1 Enter Select i to accept the Exit option CUCLE Pressure Fise kPa B S i Press Cancel Exit to I return to CYCLE SCREEN Hins Door Closed Start Time Cycle T Selected I Press Cancel Exit or Start to abort A Vacuum Dry l a delayed start Delayed Start an Fuad alternates on display with fbart 28 Delayed Start I Press Cancel Exit to stop abort Press Enter Select when fort I I shown to abort delayed start Note The cycle must be selected before setting the start time TIME 14 30 l Left Right to select element Up Down to adjust element 1 Enter Select to accept the start time i and initiate the cycle selected I P32 33 ST IM84c SES 113 and 225 AUTOCLAVES APPENDIX 8 Accessories and spares For prices and to order any of these accessories call EschmannCare telephone 01903 775730 Eschmann instrument Management System EiMS The Little Sister SES autoclave range incorporates furniture accessories from the Eschmann instrument Management System EiMS EiMS has been developed to provide effective instrument patie
20. CIAL FUNCTIONS CYCLE MENU also refer to pages 30 and 31 Note To access a menu option shown below move the cursor to the required option by pressing the Up or Down buttons then press the Enter Select button Select the Exit menu option or press the Cancel Exit button to exit the special function options Menu Option Cycle Options Cycle History P30 33 Method of use To use this function requires knowledge of the PIN code 5 numeric digits factory set to 00001 The PIN code is should be restricted to those people responsible for the use and management of the autoclave only Access to this function is not normally available to the operator The display shows Enter PIN Code and a 0 To enter the PIN code press Up or Down to change the first digit press the Enter Select button to accept this and move to the next digit The digit will be transformed into an asterisk to protect the PIN code Continue enter the PIN code until the Cycle Option display is shown If the PIN code entry is incorrect the display will reset to enable PIN code entry again If the PIN code cannot be entered correctly e g forgotten press the Cancel Exit button then contact EschmannCare Tel 01903 765040 The available cycles will be shown and numbered in a table the default cycle will be marked with an asterisk Note The default cycle cannot be disabled The cycle type will be i
21. Tray in bottom clips Tray in top clips HFiT cassette gripped HFiT cassette gripped CORRECT INCORRECT at the side at the end When using a tray lifter to hold pouch racks or the porous load basket ensure the rack or basket is correctly located as illustrated below End bar of pouch rack or basket located in tray lifter s top centre clip Second bar of pouch rack or basket located in the tray lifter s top channel Stopping a cycle In the unlikely event of needing to stop a cycle e g wrong cycle selected press the Cancel Exit button Fig 2 This will induce Cycle Error No 02 Refer to Errors and Faults Appendix 6 to clear this error and start the cycle again Draining a reservoir The waste water reservoir will need emptying when illuminates WARNING TAKE CARE THE WASTE WATER MAY BE VERY HOT Using the drain tube supplied if lost a replacement can be ordered Part No 115036 drain the water into a container of at least 4 litres capacity or a sink Plug the drain tube s connector into the drain socket Fig 1 item 15 this automatically opens the socket s drain valve if the plug cannot be pushed into the drain push the drain release clip Fig 1 item 19 until it clicks and then insert the plug Release the red clip on the end of the drain tube with the drain tube pointing into the container or sink Note that a direct drain kit REF 87 271 55 can be fitted please contac
22. a download via a USB port They may also be bought with a built in printer Installation of a printer can also be arranged at a later date through the EschmannCare service team contact 01903 765040 and order REF 87 270 23 a printer for these autoclaves fitted with 1 roll of paper and 4 spare rolls Note This must be fitted by an EschmannCare Engineer The following instructions only apply to autoclaves fitted with a printer IMPORTANT OPERATOR NOTES Follow the notes and cautions below when changing the paper roll Do not pull on the paper whilst the printer is in operation since it may damage the printer 5 1 Printer Output The printer output refer to example below gives the following information Autoclave type Eschmann Little Sister SES 113B Autoclave Serial number SRN A unique code Cycle type CYC e g B134 followed by Cycle count e g 00345 Time and Date started STD hh mm ss dd mm yy Cycle information Cycle stage codes key to codes is below with time pressure and temperature Sterilization information Time pressure and temperature every 20 seconds during sterilization Maximum sterilization temperature TMX Minimum sterilization temperature TMN Time and Date finished ETD hh mm ss dd mm yy Status STS followed by Cycle Complete or FAIL Error number ERR If applicable If an error occurs during a cycle it is recorded with time and date on the printout and the message FAIL err
23. als or solutions containing chlorine to clean the autoclave Do not attempt to open the chamber door until the message Ready to Open has been displayed NOTE This may be accompanied by the message CAUTION HOT when the chamber is above 55 C f If an error display appears during a cycle do not switch off power refer to Errors and Faults Appendix 6 g Donot place heavy items on top of the autoclave or containers with liquids e g cups etc h Do not operate this autoclave if obviously damaged deformed or if parts are missing P20 33 ST IM84c SES 113 and 225 AUTOCLAVES Environmental conditions for safe use This autoclave has been designed for use in the following conditions only a p290 Indoor use Pressure 700 to 1060 mbar Temperature 5 C to 40 C Relative humidity 20 to 90 non condensing Mains supply voltage fluctuations Not to exceed 10 of the 230 V supply Note If it is required to operate this Autoclave outside of these conditions contact Eschmann Equipment Warning markings The warning markings on this equipment have the following meanings N Caution refer to accompanying documents i e these Instructions for Use A Caution hot surface Safety features Eschmann autoclaves are designed to be safe and effective No changes should be made to them except by an EschmannCare engineer In particular the following safety features must not be interfered with circumvented
24. amage Check for any obvious escape of steam or water during a cycle For type B autoclaves run a steam penetration test Helix or Bowie Dick type If any problems are found please contact an EschmannCare engineer Tel 01903 765040 Do not use the autoclave until repair has been effected Annually by an EschmannCare engineer i ii iii iv Check that the pressure relief valve operates correctly at the set pressure Inspect the pressure system for integrity Check calibration Check door locking mechanism for integrity NOTE Annual inspections should only be undertaken by properly trained engineers EschmannCare comprehensive service contracts cover preventive maintenance and repair to ensure trouble free operation of the autoclave including annual inspections of the pressure system to satisfy the requirements of all relevant legislation and guidelines including the Pressure Systems Safety Regulations 2006 Please contact the EschmannCare team for details Tel 01903 765040 The EschmannCare team can also provide testing and certification against HTM2010 DB9804 and DB2002 06 Refer to Eschmann s Instructions for Testing Vacuum Autoclaves ST IM85 or contact the EschmannCare team ST IM84c P21 33 LittleSister APPENDIX 3 Loading the autoclave CAUTION When loading the autoclave take care that you do not damage the door or front face of the chamber especially the door seal and mating face Damage to the
25. autoclave Close the left hand access door ST IM84c P7 33 LittleSister 3 0 Running a cycle Switching on and off Switch the autoclave on using the mains switch Fig 1 item 10 The digital display and applicable indicators will illuminate to show the autoclave has been switched on After normal checks and a Welcome message shown for 30 seconds for the first few cycles the Cycle select screen will be displayed Fig 6 If the door is closed the green door indicator will flash to indicate the door should be opened together with a display message Open Door To switch the autoclave off press 0 on the mains switch Notes 1 At times the message PLEASE WAIT may be displayed for a few seconds 2 B type autoclave If running test T1 run it directly after switching on a cold and dry chamber For notes on T1 refer to Appendix 7 Special Functions To open the door The autoclave must be switched on Once the normal Cycle select display is shown Fig 6 turn the door handle anticlockwise until vertical 12 o clock position The green door indicator light will flash To close the door onto the secondary catch With the handle vertical push the door closed gently and turn the door handle 45 clockwise 2 o clock position It will meet a stop at this position and the door will only be partially closed The door should be shut onto the secondary catch
26. ave ONLY load capacity If all trays are removed one porous load basket external dimensions 280 x 147 x 87 mm or one 9 slot pouch rack can be accommodated SES 225 B vacuum Autoclave load capacity The tray carrier load configuration has five levels that can accommodate the following 15 x examination trays or HFiT 10 cassettes with instruments or 5 x standard trays or HFiT 20 cassettes and 5 x examination trays or HFiT 10 cassettes with instruments P22 33 ST IM84c SES 13 and 225 AUTOCLAVES APPENDIX 4 Daily Testing Vacuum B type autoclaves IMPORTANT The Steam Penetration Test Device hereafter called Device must only be used and stored as detailed in the Instructions for Use supplied with the Device Failure to do so could lead to failure of the Device and dangerously misleading results In order to sterilize products and devices by steam the steam must be able to penetrate through to and contact all surfaces Items such as packaging pouches and narrow or hollow instruments are not easy to penetrate with steam For this reason the SES 113 B vacuum autoclave has been fitted with a pump to actively remove air and thus allow steam to penetrate into the load The Device order REF 87 272 32 which complies with BS EN 13060 for B type autoclaves hollow load A has been designed to test the steam penetration ability of the autoclave by presenting a defined challenge to the autoclave The Devic
27. be changed but ensure the new code is kept confidential The PIN code should be changed before the autoclave is used NOTE When using the buttons shown in the Figure below note they will auto repeat when required i e when changing items such as date or time simply press and hold button to effect multiple changes Left Up Right Control Button detail CF gt and function Enter Select Down Cancel Exit To gain access to the special functions press the Enter Select button when the normal Cycle Select display is shown the CYCLE screen is displayed To select this option press the Enter Select button or to access the next option SETTINGS press the Down button To return to the normal Cycle Select display without selecting a Special Function press the Cancel Exit button To return to the CYCLE screen from the SETTINGS screen press the Up button The Special Function menus available are shown below they are explained in greater detail on the following pages CYCLE menu options Cycle Options Cycle History Test Cycles Delayed Start SETTINGS menu options Date Time Language Display Set user PIN code Audio Not applicable on UK models NOTE There is a third option SERVICE which will be displayed if the Down button is pressed when the SETTINGS screen is displayed This option is not applicable to the user it requires knowledge of a spe
28. bout its suitability for processing in an autoclave Electrical safety The user should ensure the following electrical safety measures are followed a This equipment must be earthed b The autoclave is switched off and disconnected from the mains power supply before renewing fuses checking and cleaning the autoclave c No attempt is made to service this autoclave internally General safety IMPORTANT The design of the autoclave pressure vessel is certified by a third party accredited test house to International Standards In order to ensure safety and to comply with UK and or International regulations the vessel and fittings should be inspected by a competent person at regular intervals Eschmann recommends this should be carried out at least once every 14 months by an Eschmann trained engineer The operator should follow these general safety measures a Donotcover any of the ventilation grills or attempt to use the autoclave if the covers are damaged loose or missing or if the door seal is damaged or missing b Always refer to the Errors and Faults section of Appendix 6 if the normal cycle select display is not shown on the display or it is blank or corrupted c Ensure the fresh water reservoir lid is in place and returned to its normal rear position before using the autoclave Check the fresh water reservoir inlet filter is cleaned if dirty or replaced if damaged Do not use abrasive powders chemic
29. by pushing a new one into the mount c At weekly intervals remove the chamber furniture and clean out any dirt or debris that has collected around the filter mesh cone in the base at the back of the chamber At monthly intervals remove any dirt or debris around the filter and then gently pull it out of its recess Clean the recess and the filter carefully Replace the filter before using the autoclave again If the filter is damaged or blocked it should be replaced order Part No 391037 d Ona weekly basis rinse out the fresh water reservoir with Sterile Water for Irrigation drain and allow to dry Check the drain tube and the O ring on its connector regularly replace if damaged order Part No 115036 Fuse Renewal Fuses Part No 112474 are at the rear of the autoclave Fig 1 item 23 For fuse ratings refer to TECHNICAL DATA To extract a fuse switch off and disconnect the autoclave from the mains supply Insert a screwdriver or small coin in the slot of the fuse holder and twist it counter clockwise After inspecting or renewing a fuse reverse the above procedure to secure the fuse Internal fuses should not be replaced by the user this should only be performed by an Eschmann trained service engineer Service The Service Manual ST SM48 which also contains an illustrated spare parts list and a Manual reference ST IM85 providing the Manufacturer s recommended instructions for testing vacuum autoclaves to
30. cable has a fused plug the correct fuse will be installed Connect the mains power cable as supplied to the mains supply and to the rear IEC mains socket Fig 1 item 22 DO NOT use equivalent mains leads Switch the mains supply on if controlled by a switch If changing the plug ensure that the plug is earthed Check the fuse if a fused plug is appropriate for the equipment The connections to observe when connecting a mains plug are Yellow Green Earth Brown Live Blue Neutral Switch the autoclave on using the mains switch Fig 1 item 10 The switch is marked 0 for off and I for on it illuminates green when on The digital display and applicable indicators will illuminate to show the autoclave has been switched on After normal checks and a Welcome Please contact Eschmann message shown for 30 seconds for the first few cycles the Cycle select screen will be displayed To switch the autoclave off press 0 on the mains switch Opening Closing the door To open the door The autoclave must be switched on Once the normal Cycle select display is shown Fig 6 turn the door handle anticlockwise until vertical 12 o clock position The green door indicator light will flash To close the door onto the secondary catch only With the handle vertical push the door closed gently and turn the door handle 45 clockwise 2 o clock position It will meet a s
31. ccurs section 5 3 1 Printing will be suspended until the head temperature returns to normal levels once cooled the printout will continue and no information will be lost Spares Pack A spare printer roll will be provided with the autoclave when you require more please contact the EschmannCare team on 01903 765040 and order the following REF 87 270 45 a set of 4 thermal paper rolls for the Eschmann Little Sister SES 113 113B vacuum autoclaves ST IM84c 6 0 Technical data 6 1 Electrical Supply 230 V 410 or 220 V 6 15 see rating plate 50 60 Hz a c v For use with alternating current Loading at220V 1 85kW 10A max Loading at230V 2 0kW 10A max Fuses 2 T12 5A 440 V Part No 112474 Protection Class 6 2 Sterilizing and test data Sterilizing time refer to Note below 3 minutes 20 seconds at 134 137 C or 15 minutes at 121 124 C B type cycle drying time if selected is 15 minutes Maximum time for T1 less than 20 minutes Maximum time for T3 less than 25 minutes Operating pressure 1 03 bar at 121 C 1 23 bar at 124 C 2 03 bar at 134 C 2 32 bar at 137 C For basic cycle details refer to section 6 12 and for more detailed cycle information sections 6 13 to 6 14 Note An Eschmann trained engineer can increase the sterilization times up to a max of 30 minutes on request contact the EschmannCare team on 01903 765040 6 3 Maximum Loads 113 autoclave Maximum load per tra
32. cial service PIN code if this is required please contact Eschmann and is not discussed or detailed within this manual If this is selected in error press the Cancel Exit button to return to the Cycle Select display NOTE When any parameters are changed using the special functions use the table below to record your settings This will ensure your settings can be entered quickly in the unlikely event of them being lost time date and language are obvious do not write down the pin code Speaker tick Cycles enabled tick Default cycle tick Cycle History tick ST IM84c P27 33 LittleSister TABLE 3 SPECIAL FUNCTIONS SETTINGS MENU also refer to page 28 Note To access a menu option shown below move the cursor to the required option by pressing the Up or Down buttons then press the Enter Select button If the Exit option is selected this will return you to the normal SETTINGS screen press the Cancel Exit button to exit the special function options Menu Option Date Time Language Display Set User PIN Code P28 33 Method of use When selected the display will show Set Time followed by the hour and minutes indicated by hh mm below this is Set Date followed by the day month and year indicated by dd mm 20yy The cursor shows which element of the time or date can be changed Using the Left and Right buttons m
33. display during a cycle Cycle stage in progress Actual temperature Actual pressure Animated progress bar NOTE Other information as detailed above Normal display at the end of a cycle When temperature is above 55 C When temperature is below 55 C CAUTION HOT alternates with Ready To Open every second Fig 6 Cycle Select and various display screens ST IM84c P11 33 LittleSister Normal cycle display after pressing Cycle Select during a cycle Note TIME is replaced with CLE and the cycle number Note This message may only appear briefly once the autoclave has emptied either of the displays below will be shown until cleared Press Erber and Carcel Enter PIH Code 5 Display if PIN code is required Display if PIN code is not required See Key Combination in Set User PIN Code Special functions Appendix 7 Special functions selection To access press and hold the Enter Select button when the Cycle Select display is shown The third option SERVICE is not applicable to any user and is not discussed in these instructions If return to the Cycle Select display is required without selection of either special function press the Cancel Exit button To use these Special functions refer to Appendix 7 Fig 7 Various display screens P12 33 ST IM84c 5 0 Printer instructions All Eschmann Little Sister SES 113 and 225 autoclaves incorporate direct dat
34. e consists of a plastic tubular device with an indicator receptacle at the end and is supplied with indicator strips The Device must be replaced after 250 uses This Device satisfies the requirement to perform a daily steam penetration test as given in EN 554 1994 Sterilization of medical devices Validation and routine control of sterilization by moist heat British Standards Institution clause 6 3 4 DB 2002 06 Device Bulletin Benchtop Steam Sterilizers Guidance on purchase Operation and Maintenance Medical Devices Agency clause 5 3 1 SN 2002 24 Safety Notice Steam penetration tests in vacuum benchtop sterilizers Medical Devices Agency DB 2000 05 Device Bulletin Guidance on the Purchase Operation and Maintenance of Vacuum Benchtop Steam Sterilizers Medical Devices Agency clause 5 3 1 DB 9804 Device Bulletin The validation and periodic testing of benchtop vacuum steam sterilizers Medical Devices Agency clause 5 1 Note Some of the above documents are available on the MHRA s website www mhra gov uk Frequency of use To comply with the requirements of the documents listed above Test T1 chamber integrity test and Test T3 steam penetration test must be conducted at the start of each day the autoclave is to be used If the tests are passed it verifies the suitability of the autoclave s air removal stage for the day ahead Eschmann recommends the daily use of the Device and a Test Log E
35. e rose too much during leak test Temperature change during leak test Temperature pressure error Various errors during cycle User setup corrupted Button locked pressed Items to check before running another cycle Check mains supply check mains connections check mains fuses and mains lead fuse Run cycle again Correct reason for Cycle abort Run required cycle Visually check door locks for a cause of failure e g foreign body stopping door lock operating check door is firmly closed Correct visual faults if possible then run cycle again Visually check for obstructions in door seal area clean door seal and its mating face Run cycle again Empty water from chamber check load is dry and run cycle again Ensure fresh water reservoir is full and run cycle again Refer to Overheat cutout section page 24 If cutout has not tripped visually check for obstructions in door seal area clean door seal and its mating face Run cycle again Consider reducing load Note Error may be due to a particularly cold load Check fresh water reservoir level Run cycle again Visually check for obstructions in door seal area clean door seal and its mating face Run cycle again Check chamber and load is dry Run cycle again Fill fresh water reservoir Run cycle again Empty waste water reservoir Run cycle again Check door seal for damage and cleanliness in door seal area Clean door seal and its mating face Run cyc
36. gineer refer to Appendix 7 ST IM84c P9 33 LittleSister 4 0 Direct Data Download instructions Introduction The Eschmann Little Sister SES 113 and 225 autoclaves include Direct Data Download DDD application software as standard All cycle records are stored on the unit IMPORTANT After 20 000 cycles the autoclave s memory may be getting full and data storage could stop please contact Eschmann for details of how the autoclave s memory can be replaced DDD allows cycle records to be downloaded via a USB cable direct to a PC or Laptop Cycle records can then be stored directly onthe PC printed out using a conventional printer several records to a page or pasted directly into the computerised patient record DDD enables the implementation of instrument patient management and traceability DDD application removes the need for an integral printer and the issues relating to accurately managing this type of paper based data Using the DDD application If in doubt refer to the instructions supplied within the Application CD or contact Eschmann The DDD application software should be installed on a PC refer to the instructions provided with the DDD Software To open the DDD application click on the DDD icon on the Desktop of your PC or Laptop Using the cable provided connect the PC to the autoclave s USB port Fig 1 item 12 Switch on the autoclave The application will access the autoclave refer to Fig 4 Disk Start
37. hat the film sides of adjacent pouches are face to face and sterilized using a B type vacuum cycle with drying Only use pouches recommended for steam sterilization i e conform to EN868 details are provided below Ensure all seals have been made correctly to maintain sterility post sterilization All items must be positioned so that they drain freely and do not trap rising air bubbles Wrapped item can be stored for later use For details on the choice of cycle refer to Section 6 The following sterilisation pouches are available from Eschmann Pouch box of 200 small 90 mm x 230 mm REF 87 033 10 Pouch box of 200 medium 133 mm x 254 mm REF 87 033 18 Pouch box of 200 large 190 mm x 330 mm REF 87 033 29 The configuration of the load support system follows the load should be confined to the usable space as defined in section 6 5 Full details on the cassettes etc can be found in the accompanying EiMS brochure also refer to Appendix 8 SES 113 and SES 113 B vacuum Autoclave load capacity The tray carrier load configuration has four levels that can accommodate the following Top 1 x examination tray or HFiT 10 cassette 140 mm x 180 mm Upper Middle 1 x standard tray or HFiT 20 cassette 280 mm x 180 mm or 2 x examination trays or HFiT 10 cassettes Lower Middle 1 x standard tray or HFiT 20 cassette or 2 x examination trays or HFiT 10 cassettes Bottom 1 x examination tray or HFiT 10 cassette SES 113 B vacuum Autocl
38. hes or wrapping should be used CAUTION Take care during the following procedure the water chamber and load could be very hot and potentially contaminated Remove the tray carrier and check for water in the rear of the chamber 350 ml maximum drain any found using for example a suitable syringe and tube Replace the tray carrier The chamber should be allowed to dry before starting another cycle Errors that can be corrected or checked by the operator are detailed in the error table on page 25 ensure these checks if applicable are performed before running the cycle again For errors that are not detailed try running the cycle again checking procedures during loading If the same error persists contact the EschmannCare team Tel 01903 765040 Do not attempt to solve an error using tools or tampering with the autoclave The engineer will need to know the Model Serial Number refer to label in Fig 1 item 20 and the error number displayed Display messages Certification Expired The certification date has been set in the factory to 18 months after manufacture When the date is reached the digital display will show the message Certification Expired Please contact Eschmann This message will be displayed for 5 seconds and appears when the power is switched on and after each cycle It will not be displayed if the service due message has been triggered because service due takes precedence ST IM84c P2
39. ion for the next year P24 33 ST IM84c SES 113 and 225 AUTOCLAVES APPENDIX 6 Cycle errors and faults and Display messages Cycle errors and faults The autoclave is continually checking the parameters required for effective sterilization if a problem is encountered a CYCLE FAILED or SYSTEM ERROR message will be displayed followed by an error number and Please Wait There will be five rapid beeps and also any illuminated progress indicators will flash For future reference record the type and error number before continuing The autoclave returns to a safe condition and displays either Enter PIN Code or Press Enter and Cancel depending on configuration refer to Appendix 7 The factory set PIN code is 00001 this should be changed refer to Appendix 7 Special Functions CYCLE FAILED and SYSTEM ERROR message The cycle will abort and provided power supply to the autoclave is maintained the message CYCLE FAILED Please Wait or SYSTEM FAILED Please Wait and an error number will be displayed Finally the message Enter PIN Code or Press Enter and Cancel will be displayed refer to Enter PIN Code or Press Enter and Cancel below Follow the section below Attending to an error and run the cycle again Overheat cutout Error 8 If the heating element reset thermostat operates e g in a nil water condition indicated by the overheat indicator light illuminating Fig 1
40. item 14 and Error 8 proceed as follows Allow the autoclave to cool and press the reset button on the underside of the autoclave position indicated by the serial label Fig 1 item 13 then clear the error in the normal way DO NOT clear the error first and then press the reset button the unit will try to reset with the thermostat tripped inducing errors No mains power If after switching on power there is no visual display check the power supply and connections both fuses at the back of the autoclave refer to Fuse Renewal and the fuse in the mains plug Enter PIN Code Message Enter the user PIN code as follows refer to Fig 2 for button information Press Up or Down to select the first digit press Enter Select to accept this digit and move to the next The accepted digit changes to an asterisk to protect the PIN Continue to select and enter the PIN If the PIN is incorrect you will be asked for it again Press Enter and Cancel Message Press the Enter select and Cancel Exit buttons together refer to Fig 2 for button detail and hold until error clears Selection of which message is displayed is via the SETTINGS menu refer to Special Functions Appendix 7 Attending to a cycle error Items must be considered non sterile if an error occurs and all items in the load should be reprocessed before use All items should be removed and dried If applicable new pouc
41. l and configuration information when using the Special Functions Indicators Fig 2 Low water indicator When illuminated the fresh water reservoir top one should be refilled before a new cycle is started section 2 It will flash if the reservoir is low and an attempt to start a cycle is made Drain reservoir indicator When illuminated the waste water reservoir bottom one is full and should be emptied before a new cycle is started section 2 It will flash if the reservoir is full and an attempt to start a cycle is made Door indicator This is illuminated to indicate the door is locked or flashing to indicate the door should be opened or closed the digital display provides a related message It will be off after you close the door prior to starting a cycle 000 Cycle complete indicator This is illuminated at the end of a successful completed cycle Progress indicators The display has indicators that illuminate in turn as the cycle progresses The main stages are identified by three symbols The symbols and stages are detailed below Cycle started Water fill completed Heating to sterilization temperature Sterilization started Sterilization stage passed halfway Condensing Drying started vacuum only Less than 1 minute remaining to cycle completion Cycle completed successfully OONOOARWND I Y Pre conditioning jo Sterilizing N Post sterilization cooling drying
42. le again Leave Autoclave to cool for a few minutes then repeat test Visually check for obstructions in door seal area clean door seal and its mating face Ensure chamber is dry Run cycle again Leave Autoclave to cool for a few minutes then repeat test Visually check for obstructions in door seal area clean door seal and its mating face Run cycle again Refer to Overheat cutout section page 24 If cutout has not tripped visually check for obstructions in door seal area clean door seal and its mating face Run cycle again All user settings will need to be reset as the autoclave will revert back to the original factory defaults refer to Appendix 7 and enter your user settings refer to list provided on page 26 Run cycle again A button may have stuck momentarily Run cycle again ST IM84c SES 113 and 225 AUTOCLAVES APPENDIX 7 Special Functions IMPORTANT These special functions should only be used by the Autoclave Manager and the EschmannCare Engineer The autoclave s special functions include Amending display language except UK models Printing or downloading cycle history Setting the date and time Switching the audible beep on or off Enabling additional cycles Programming a delayed start Running special tests PIN Number Some special functions are protected by a PIN code this has been set in the factory to 00001 One of the special functions SETTINGS menu Set user PIN code enables this to
43. ll rights reserved This booklet is protected by copyright No part of it may be reproduced stored in a retrieval system or transmitted in any form or by any means electronic mechanical photocopying recording or otherwise without written permission from Eschmann Holdings Limited The information in this publication was correct at the time of going to print The Company however reserves the right to modify or improve the equipment referred to C The CE marking affixed to the product certifies that it complies with the European ST IM84c 0473 Medical Devices Directive 93 42 EEC as amended and related legislation November 2007 0473 P34 33 ST IM84c DISCOVER NEW INNOVATION NEW TECHNOLOGY DISCOVER QESCHMANN Eschmann Equipment Peter Road Lancing West Sussex BN15 8TJ England Tel 44 0 1903 753322 Fax 44 0 1903 875789 www eschmann co uk 113387 ST IM84c
44. ll start printing a cycle record refer to section 5 Indicators will show the status ofthe cycle The digital display will indicate the chamber temperature pressure and an animated progress bar The Cycle Count can be displayed if the Cycle select button is pressed during a cycle Fig 6 and 7 Note Cycles cannot be started when the green door indicator is flashing and the Open Door message is displayed Open and then close the door to correct this situation should it occur Cycle errors WARNING When a cycle is aborted and or an error display is indicated the load must be treated as non sterile The load must be sterilized by running the cycle again If wrapped or pouched new packaging must be used If the autoclave detects a fault it will display a message e g Door not closed Fig 6 or if it detects an error it will display CYCLE FAILED or SYSTEM ERROR there will be five rapid beeps and illuminated progress indicators will flash Refer to Errors and Faults Appendix 6 Anyone who becomes aware of a fault when using the autoclave should report it immediately to the Autoclave Manager and all necessary remedial action should be taken before the autoclave is used again Should the overheat cutout trip its indicator light Fig 1 item 14 will illuminate refer to Appendix 6 Completed cycles Completion of a successful cycle will be indicated by v illuminating and the display showing Cycle Co
45. mplete and Ready to Open the normal end of cycle display refer to Fig 6 There will be three audible beeps unless disabled Open the door fully take care the load will be hot and remove the load using the tray lifter Note On N type autoclaves ONLY at the end of cycles with drying the beep if enabled will continue intermittently During these beeps open the door and remove the load It should still be hot enough to self dry If the door is not opened during the intermittent beeps the load may not dry when removed Items sterilized using an N type cycle are for immediate use only P8 33 ST IM84c SES 113 and 225 AUTOCLAVES Unloading the chamber WARNING TAKE CARE WHEN UNLOADING THE LOAD AND CHAMBER MAY BE VERY HOT Use the tray lifter as detailed in the following section If running consecutive cycles Eschmann recommends waiting for five minutes between cycles with the door open to allow the chamber conditions to stabilise If the autoclave is not to be used for several minutes close the door onto the secondary catch as detailed opposite Using the tray lifter The tray lifter has one bottom support and two pairs of top clips The top pair is for Hand Free instrument Transfer HFiT cassettes and the bottom pair is for trays Do not use the top clips when lifting trays The spacing of the top clips enables HFiT cassettes to be held at the side or end Refer to illustrations below v
46. nd indicated by an asterisk To return to the CYCLE menu select the Exit option or press the Cancel Exit button Table 2 is continued over the page ST IM84c SES 113 and 225 AUTOCLAVES TABLE 2 continued SPECIAL FUNCTIONS CYCLE MENU also refer to page 30 Menu Option Method of use Test Cycles B type units The three options below will be provided plus Exit Press the Up or Down buttons to select the required option T1 T3 etc indicated by the cursor on the left Pressing the Enter Select button returns you to the cycle select display with the test selected as the cycle To start the test after checking the door is closed and chamber is empty press the Start cycle button To cancel the test before starting it press the Enter Select button this returns you to the CYCLE screen To return to the CYCLE menu press the Enter Select button or the Cancel Exit button to return to the normal cycle select display T1 Chamber Integrity MUST NOT be used to process loads Note The chamber MUST be dry and at ambient temperature before running T1 Once switched on the chamber is heated runT1 directly after switching on The autoclave reduces pressure in accordance with BS EN 13060 and measures the leak rate various display messages detail progress The Chamber Integrity Test is deemed a pass if the rate of pressure rise is less than 1 3mbar
47. ndicated by N for non vacuum or B for vacuum drying will be indicated by the inclusion of a drying symbol and the temperature will be shown in C Cycles available to the operator enabled will be indicated by a v and those not available to the operator disabled will be indicated by a x One cycle will be selected and indicated by a cursor on the left to move the cursor press the Up or Down buttons as required To change the v to a or a to a v for a cycle press Up or Down to select the cycle and then press the Right button to move the cursor in front of the v or x to be changed To select a v press the Up button or to select a x press the Down button Press the Left button to move the cursor left to enable another cycle to be selected by pressing Up or Down To change the default cycle indicated by an x press Up or Down to select the cycle and then press the Right button twice to move the cursor past the tick Note To make a cycle the default the cycle must be enabled Press the Enter Select button to make that cycle the default Press the Left button twice moves cursor left so that another cycle can be selected CAUTION If the default cycle is changed it is of paramount importance all users are informed helping ensure the correct cycle is always selected when proces
48. not between 00001 and 59999 the change will not be successful and the message PIN must be between 1 and 59999 Try Again Enter Yes Cancel No Press Enter Select to try again or Cancel Exit to abort the PIN change When the change of PIN is accepted the message PIN Confirmed appears and you will be returned to the SETTINGS menu The display shows three options Speaker On Speaker Off and Exit The current option is identified by an asterisk Move the cursor up or down to the required option using the Up or Down buttons and press the Enter Select button to confirm selection When correct select the Exit option to return to the main SETTINGS menu or press the Cancel Exit button ST IM84c SES 113 and 225 AUTOCLAVES SETTINGS SCREEN i Press Enter Select to accept f 1 Or press Cancel Exit to I I return to Cycle Select display i 1 Up Down to select I Enter Select to accept l 1 Accept Exit or press i Cancel Exit to return I jt SETTINGS SCREEN 1 1 Left Right to select element l l Up Down to adjust element Enter Select to confirm i setting and return to the A SETTINGS SCREEN Note Press the Up or Down button to select 1 Up Down to select the language required from those listed then Enter Select to accept Enter Select to accept All the languages may Accept Exit or pre
49. nt management allowing the tracking of all instruments through the decontamination cycle Chair gt washer disinfector gt autoclave gt chair EiMS features a choice of colour coded instrument trays and Hands Free instrument Transfer HFiT cassettes in particular Examination size components that allow the cost effective implementation of a One set of instruments per tray per patient regime Details of the full EiMS range together with prices and ordering information can also be found in the brochure accompanying your autoclave Standard Trays 280 mm x 180 mm REF 87 244 51 Aluminium Silver REF 87 244 54 Aluminium Green REF 87 244 52 Aluminium Blue REF 87 244 55 Aluminium Yellow REF 87 244 53 Aluminium Red REF 87 244 56 m Aluminium Turquoise REF 87 254 50 Stainless Steel Standard Covers and Storage 280 mm x 180 mm REF 87 214 51 Storage Lid Silver REF 87 214 52 Instrument Support 16 REF 87 214 55 Storage Rack empty REF 87 214 58 Storage Rack complete with 6 x trays lids amp holders Examination Trays 180 mm x 140 mm REF 87 244 61 Aluminium Silver REF 87 244 64 Aluminium Green REF 87 244 62 Aluminium Blue REF 87 244 65 Aluminium Yellow ae REF 87 244 63 Aluminium Red REF 87 244 66 m Aluminium Turquoise REF 87 254 60 Stainles
50. one side some will be packed within the chamber All packaging should be recycled in accordance with local recycling practices DO NOT connect to the mains until setup has been completed Items included with the autoclave Instructions for Use Mains power cable Levelling bull s eye Spare printer roll if printer fitted Drain tube Tray carrier and tray lifter Furniture model specific DDD Software CD Manual and USB cable Misc Documentation Setup Adjustable feet enable the autoclave to be levelled correctly To adjust the feet slacken the locking nut adjust the height as required and tighten the locking nut The autoclave should be adjusted when the chamber is cold until level in all directions using the bull s eye levelling device supplied Note that on B type units the band heater will warm the chamber when switched on once the door has been opened switch the autoclave off until levelled The levelling device should be positioned directly on top of the flat heater plate inside and at the back of the chamber IMPORTANT Remove levelling device before use and if moved to a new location after installation check the autoclave is levelled again adjusting if required Mains connection WARNING This equipment must be earthed Ensure that the mains plug is easily accessible during use as this is the means of disconnection The autoclave is supplied with an IEC mains input socket at the rear and a mains power cable If the
51. ons the manager might also be the user Autoclave Installation Full installation details are provided in APPENDIX 1 this includes a list of items supplied with the autoclave and a note to observe if the autoclave is relocated Ensure the autoclave has been correctly installed before use by an EschmannCare trained engineer contact the EschmannCare team on 01903 765040 Installation will also include initial training covering the use of these autoclaves It is essential that the autoclave is levelled and connected to the electrical mains supply correctly On request and during installation programmes and autoclave configuration can be adjusted by the EschmannCare engineer To protect against tampering access for such adjustment can only be gained by use of a PIN code Direct Data Download DDD Software Installation Refer to the instructions provided with the Direct Data Download Software and install the DDD application Warnings and Cautions Before using the autoclave ensure you understand the Warnings and Cautions which are set out and summarised in APPENDIX 2 RESERVOIR CARE AND WATER QUALITY CAUTION Eschmann recommends filling the water reservoir with Sterile Water for Irrigation This is low in dissolved solids and has a low microbial count In the U K the Department of Health recommends that Sterile Water for Irrigation is used in bench top Autoclaves NHS Estates document HTM2031 If Sterile Water for Ir
52. ors are designated by error codes refer to Appendix 6 Key to codes on printout if not detailed above are TPV Time of vacuum pulse TPP Time of pressure pulse TSS Start of sterilization TSE End of sterilization TDS Start of drying phase TDE End of drying phase ETD End time and date STS Status at cycle end Eschmann Little Sister SES 113B SRN L1BA6E0056 CYC B134 STD 15 28 20 15 04 07 Time kPa 15 30 28 14 15 40 35 178 15 42 47 29 15 46 17 318 15 46 37 320 15 46 57 320 15 47 17 320 15 47 37 320 15 47 57 320 15 48 17 320 15 48 37 319 15 48 57 319 15 49 15 319 135 8 135 0 15 04 07 00034 TPV TPP TPV TSS TSE TMX TMN ETD STS 15 50 47 Cycle Complete ST IM84c SES 113 and 225 AUTOCLAVES 5 2 Interpretation of printer output The most important part of the printout that should be checked is the final line coded STS this should be followed by Cycle Complete indicating a successful cycle Should any aspect of the cycle not match the correct parameters the indication following STS is FAIL This will be accompanied by an error number refer to Appendix 6 Errors and Faults Other aspects of the printout that can be checked if required are as follows but note that any deviation will result in the FAIL message above 121 C Cycles TSS TSE TMX TMN Temperature gt 121 C lt 124 C Time TSS Time TSE not less than 15 minu
53. ove the cursor to the time or date element to change adjust this element using the Up or Down buttons When both the time and date are correct pressing the Enter Select button confirms this and returns you to the main SETTINGS menu The display shows Language followed by a list of the available languages all options may not be displayed at the same time Move the cursor up or down to the required language using the Up or Down buttons and press the Enter Select button to confirm the language required and return to the main SETTINGS menu Select the Exit option at the bottom of the language list or press the Cancel Exit button to return to the SETTINGS menu if no change of display language is required Not applicable on UK models The display shows Adjust Viewing Angle with an up and down arrow and a row of lines Use the Up or Down buttons to adjust the viewing angle of the display and press the Enter Select button when set as required i e provides the clearest contrast or press the Cancel Exit button if no change of display angle is required or to reject any adjustments made You will be returned to the SETTINGS menu This function requires knowledge of the PIN code 5 digits 00001 59999 factory set to 00001 The PIN code is normally restricted to those people responsible for control and management of the autoclave Acce
54. pped pouched instruments B Single wrapped cassettes B Single wrapped hollow instruments B Double wrapped instruments B Double wrapped hollow instruments B Porous loads B v v x x x x x x x KLKLRK RK KKK Warning markings The warning markings on this equipment have the following meanings AN Caution refer to accompanying documents i e these Instructions for Use A Caution hot surface EschmannCare Eschmann recommends direct installation and commissioning in the UK for all its autoclaves by its own EschmannCare team of professional engineers Eschmann also recommends a programme of after sales service and maintenance for all Eschmann autoclaves using only the correct Eschmann spare parts Service and maintenance programmes can include PSSR 2000 certification and testing to HTM 2010 DB9804 The EschmannCare service department is staffed and equipped to provide advice and assistance during normal office hours To avoid delays when making enquiries please quote the Model and Serial Number of your Autoclave which is shown on the Serial Label the location of which is shown in Fig 1 item 20 please include all alpha and numeric digits For further information visit www eschmann co uk All correspondence relating to the after sales service of Eschmann Equipment to be addressed to UK Customers EschmannCare Peter Road Lancing West Sussex BN15 8TJ England Tel 44 0 1903 765040 Fax 44 0 1903
55. rigation is not being used then Eschmann strongly recommends the use of either distilled water deionized water purified water or water treated by the reverse osmosis process These types of water are low in dissolved solids and can help reduce the effects of tap water detailed below DO NOT USE TAP WATER this is high in dissolved solids and can deposit lime scale block filters and cause damage to the pressure vessel Don t leave water standing in the reservoir for more than a few hours If you are not sure how long the water has been there change it Also contents of part used containers of sterile water for irrigation should be discarded microbiological purity will be compromised once opened At the end of the working day or whenever the sterilizer is to be left unused for several hours drain both the water reservoirs and leave them to dry Fill the reservoir Open the left hand access door Fig 3 and slide the top fresh water reservoir out about 100 mm Press down gently the back of the reservoir cover slide it forward and remove it Fig 3 DO NOT USE TAP WATER Fill the reservoir up to the MAX mark engraved on the internal mesh cover 3 36 litres maximum When the reservoir is filled to the MAX mark the autoclave will run 8 cycles before it needs to be filled again Replace and slide the reservoir cover back until it clicks into position and return the reservoir to its normal rear position before using the
56. s Steel Examination Covers and Storage 180 mm x 140 mm REF 87 214 61 Storage Lid Silver REF 87 214 62 Instrument Support 10 REF 87 214 65 Storage Rack empty REF 87 214 68 Storage Rack complete with 6 x trays lids amp holders Miscellaneous Accessories The following sterilisation pouches are available REF 87 033 10 Box of 200 small 90 mm x 230 mm REF 87 033 18 Box of 200 medium 133 mm x 254 mm REF 87 033 29 Box of 200 large 190 mm x 330 mm Other accessories REF 87 270 23 Printer Kit REF 87 270 45 Printer Rolls 4 REF 87 271 21 Tray Lifter REF 87 271 55 Direct Drain Kit REF 87 272 20 SES 113 Porous Load Basket REF 87 272 25 00 SES 113 Pouch Rack REF 87 272 32 Helix Test Pack 250 Uses REF 87 272 66 Bowie Dick Test Pack Part No 115036 Spare Drain Tube Part No 112782 Direct Data Download USB cable Part No 391037 Filter mesh cone Part No 112474 Mains fuse ST IM84c P33 33 LittleSister Patents and Trade marks The ESCHMANN name and logo Little Sister EschmannCare and SES are trade marks of Eschmann Holdings Limited Eschmann Equipment is a trading name of Eschmann Holdings Limited Patents Pending Copyright 2007 Eschmann Holdings Limited A
57. s of the chamber and door must be avoided Water in the waste reservoir can be very hot take care when draining to avoid contact with the hot water b During certification service and maintenance work it may be necessary to remove covers This increases the risk of exposure to hot surfaces and electrical systems Accordingly all service and maintenance work should be carried out by an Eschmann engineer or at the least by an engineer recently trained and certified by Eschmann c If reservoirs are not emptied and left to dry after use preferably on a daily basis water may become contaminated d If the load is not suitable for the cycle selected the load may remain non sterile It is recommended that the user consults the Instructions for Use of the items to be processed If in doubt contact EschmannCare or an expert on sterilizing processes such as an Authorized Person AP in the UK Limitations on use The operator should be aware of the following limitations during use a Do not process hollow pouched or wrapped items or porous loads using an N type cycle b Never use trays or cassettes without perforations c Do not process fluids in this autoclave d Items in pouches should only be processed as detailed in these instructions e Do not use autoclaves near inflammable materials or gases e g anaesthetics f When sterilizing instruments not of solid metal construction the manufacturer of the instrument must be consulted a
58. schmannCare can provide a Test Log for this purpose please contact EschmannCare Tel 01903 765040 and order Part No 113885 Conducting the tests To run T1 refer to Appendix 7 Special Functions for T3 proceed as follows Full instructions are provided with the device The autoclave should be loaded with the Device ONLY Use the T3 cycle or T3 via the special functions if T3 has been disabled as a cycle to run the test Remember the Device must be removed within 10 minutes of the cycle end Compare the indicator test strip with the device s instructions HTM2010 DB2002 06 DB9804 Testing Where there is a requirement to satisfy the guidance in HTM2010 DB2002 06 and DB9804 then further Daily Weekly Quarterly and Annual Testing may be required Our recommendations on these testing requirements are detailed within our Manufacturer s recommended Instructions for Testing Vacuum Autoclaves reference ST IM85 Nand B type autoclaves For details on the automatic control test for HTM2010 please contact EschmannCare ST IM84c P23 33 LittleSister APPENDIX 5 Maintenance Cleaning and care CAUTION Chlorine even in the concentrations found in tap water can cause stainless steel to crack and could damage the chamber Disconnect from the mains electrical supply before cleaning the Autoclave Do not use flammable liquids abrasive powders chemicals or solutions containing chlorine to clean the autocla
59. se parts can adversely affect performance When loading pouches ensure they do not make paper cuts in the door seal Do not reprocess single use items The autoclave is designed to sterilize instruments it does not wash or clean them Effective cleaning and drying of items prior to sterilization is an essential part of the decontamination cycle Air retained within the load can also prevent steam contact Prior to sterilization instruments should be thoroughly cleaned ideally in an Eschmann Instrument Washer Disinfector contact the EschmannCare team on 01903 765040 for more information When placed in an autoclave items with hinges or ratchets should be open and unlocked Multiple part items should be dismantled or only loosely assembled to allow steam penetration Once loaded the autoclave cycle should be started do not leave items awaiting processing in the autoclave Check the suitability of nonmetallic items for processing in an autoclave and that the temperatures are acceptable Load types and load arrangements should conform to that specified below and detailed in section 6 3 If in doubt about loads and processing details please discuss this directly with EschmannCare Eschmann Instrument Management System EiMS This autoclave uses furniture and accessories from the Eschmann instrument Management System Instrument trays and Hands Free instrument Transfer HFiT cassettes are available in a variety of sizes and colours designed
60. sing items To return to the CYCLE menu press the Cancel Exit button The options available are Print Download and Exit Note that if a printer is not fitted the download option will automatically be selected Press the Up or Down buttons to select the required option indicated by the cursor on the left then press the Enter Select button to access that option To return to the CYCLE menu press the Cancel Exit button or select the Exit option Print This option is only available if a printer is fitted There are three options Last Successful 10 Last Failed 10 and Exit Press the Up or Down buttons to select the required option indicated by the cursor on the left then press the Enter Select button to accept that option The required printout will be executed providing there is paper in the printer you will be returned to the CYCLE menu Download This option alters the frequency of the reminder to Download the cycle history There are seven options Day Week Month 3 Month 6 Month Don t Remind me and Exit the currently selected option will be indicated by an asterisk Press the Up or Down buttons to select the required option indicated by the cursor then press the Enter Select button to accept that option The required frequency will be set a
61. ss not be displayed at the same time The lowest i I Cancel Exit to return option the bottom of the list is Exit select this 1 SETTINGS SCREEN to return to the SETTINGS SCREEN without changing the selected language I 1 Up Down to change the viewing angle I I Enter Select to accept the change or Cancel Exit to reject it and 1 return to the SETTINGS SCREEN i Up Down to adjust digit Enter Select 1 1 Up Down to select I to accept digit and move to next digit I i I Enter Select to accept I i Selected digit becomes an I Accept Exi t to initiate F Note Currently selected option will be identified by an Accepting Change PIN will initiate display below Accept Key Combination to return to the SETTINGS SCREEN 1 Up Down to select digit i Enter Select to accept digit and move to next digit I Selected digit becomes an i Press Cancel Exit to abort Note Confirm New PIN only appears after the first entry of the new PIN If a PIN code is not valid a message will confirm this PIN must be between 1 and Accept Exit or press 59999 Try Again Enter Yes l Cancel Exit to return to the Cancel No If codes vary you 1 SETTINGS SCREEN I will be returned to start again Note The currently selected option is marked with an f Up Down to select 1 Enter Select to accept ST IM84c P29 33 LittleSister TABLE 2 SPE
62. ss to this function is not normally available to the operator The display shows Enter PIN Code and 0 To enter the PIN code press Up or Down to change the first digit press Enter Select to accept this digit and move to the next The accepted digit changes to an asterisk to protect the code Continue to enter the other digits If the wrong code is entered you will be prompted to Enter PIN Code again To escape this option if the PIN has been forgotten press Cancel Exit The options provided are Change PIN and Key Combination Use the Up or Down buttons to move the cursor to the required option and press the Enter Select button the selected option will be marked with an asterisk Move the cursor to the Exit option and press the Enter Select button proceed as detailed below for the option selected If Key Combination is selected you will be returned to the SETTINGS menu and cycle errors can be cleared by the operator using a key combination Enter Select and Cancel Exit buttons pressed together and not the PIN code If Change PIN is selected the display will show Enter New PIN and a 0 Enter the new pin code as above The display will then show Confirm New PIN and a 0 enter the new PIN code again to confirm it If the second entry does not match you will be prompted to Enter New PIN again If the PIN code is
63. t EschmannCare for details Also note that the waste water reservoir is fixed To release the drain tube from the autoclave push the drain release clip Fig 1 item 19 Don t forget to close the red clip To drain the fresh water reservoir pull it forward about 100mm as for filling Fig 2 and drain the water into a 4 litre container as detailed above but plug the drain tube s connector into the fresh water reservoir s drain socket Fig 1 item 17 After use end of day When the autoclave is not in use switch it off at the mains switch Fig 1 item 10 Also leave the door open on the secondary catch with the chamber empty Note Whenever a B type autoclave is switched on the chamber is heated At the end of the day and or week ensure the instructions for care of the autoclave are followed as detailed in Appendix 5 It is recommended that the waste water and the fresh water reservoirs should be drained and allowed to dry at the end of each day Display beep and Special function notes At programmed intervals the autoclave will display special messages e g Service required refer to Appendix 6 The autoclave will sound 5 rapid beeps if there is a fault 3 slower beeps on cycle complete and one beep per key press The beep can be disabled using the Special Functions refer to Appendix 7 Special functions should only be used by the Autoclave Manager or an EschmannCare en
64. tes 134 C Cycles TSS TSE TMX TMN Temperature gt 134 C lt 137 C Time TSS Time TSE not less than 3 min 20sec 5 3 Printer operation Mode button not used Status LED Paper feed button Fig 8 Printer control panel 5 3 1 Status LED The printer incorporates a Status LED Fig 8 to report its condition If there is a fault the LED will flash in sequence The fault can be identified by observing the flash sequence detailed below Printer On Printer Off Paper out fit new roll Thermal head too hot allow head to cool LED On continually LED Off LED flashing LED flashing P13 33 LittleSister 5 3 2 Loading paper CAUTION DO NOT USE ALTERNATIVE PAPER ROLLS The quality and size of paper rolls used in the printer should only be supplied by Eschmann The printer will automatically detect when the printer s paper has run out and report this using the Status LED section 5 3 1 Use the Paper feed button to feed through the last few centimetres of paper and fit a new roll as follows i Pull the top of the opaque centre window at the finger hold forward until the lid is released from its locked position To avoid damage do not use excessive force 7 _ ii Remove any paper remaining in the paper recess iii Reel off a few centimetres from a new roll of paper Hold approximately 50 mm of paper outside the printer as you place the new roll into the
65. to a suitable container of at z er a a least 4 litres capacit Fig 3 Filling and draining the reservoirs peel ST IM84c P5 33 LittleSister 1 0 Controls and displays Mains switch Fig 1 item 10 The mains switch is positioned on the front of the autoclave The symbol I indicates on and the symbol 0 indicates off The switch illuminates green when on Push buttons Cycle select button Fig 1 item 7 Press to select a sterilization cycle The Autoclave Manager may have enabled more than one cycle Cycle details will be shown in the main display After scrolling the available cycles which may include the test cycles T1 and T3 the Cycle Count the number of cycles started will be displayed Start cycle button Fig 1 item 4 Press to start the selected cycle Control Buttons The control buttons shown in Fig 2 are behind the hinged panel below the main display Their use is explained in Appendix 7 Special functions They are not required during normal use of the autoclave Note When a button is pressed a beep will sound unless disabled refer to Appendix 7 Audio Digital display Fig 2 The digital display provides the information required during normal use of the autoclave e g to select the required cycle and to provide information on the cycle status It may display other details instructions and warnings as detailed in this manua
66. top at this position and the door will only be partially closed The door should be shut onto the secondary catch at all times when the autoclave is not in use To close the door fully With the handle vertical push the door shut firmly and turn the door handle clockwise until horizontal 3 o clock position The door indicator light will stop flashing and will not be illuminated User PIN number The autoclave is supplied with the User PIN number 00001 This should be changed before the autoclave is used the new PIN number should ONLY be available to the Autoclave Manager Refer to Appendix 7 Special Functions to change the PIN number Direct drain kit If you require the waste reservoir to drain directly e g into an adjacent sink a direct drain kit can be fitted REF 87 251 55 Please contact EschmannCare Tel 01903 765040 for information ST IM84c P19 33 LittleSister APPENDIX 2 Warnings and cautions The grouped warnings and cautions below should be followed at all times all users operators should be made aware them Potential Hazards The operator should be aware of the following hazards a All autoclaves operate with steam at high temperatures and pressure The autoclave door must only be opened when the machine indicates it is safe to do so Care must be taken to avoid contact with any residual steam or hot water The Eschmann tray lifter must be used to remove trays etc Contact with hot internal surface
67. uring manufacture to 2 85 bar P16 33 ST IM84c SES 13 and 225 AUTOCLAVES 6 12 Basic cycle information CYCLE DETAILS B134 C Dry Solid or Hollow instruments and porous loads vacuum cycle with drying items can be pouched single or double wrapped B121 C Dry Solid or Hollow instruments and porous loads vacuum cycle with drying items can be pouched single or double wrapped Chamber integrity test MUST NOT be used to process loads Steam penetration test MUST NOT be used to process loads This cycle is not normally installed on any autoclave Not installed on SES 225 B vacuum autoclave Note On N type autoclaves ONLY at the end of cycles with drying a beep will continue intermittently Refer to section 3 Completed cycles 6 13 Detailed cycle information The operating cycles are detailed in the diagrams and explanation text provided in 6 13 and 6 14 These sections detail the maximum temperature for each sterilization cycle within the useable space The autoclave can be fitted with a printer producing cycle records interpretation of the printout is detailed in section 5 Note The configuration of the load support system is detailed in Appendix 3 In accordance with BS EN 13060 the relevant type tests for the cycles listed are as below Type tests Sterilization cycle type and cycle number B type cycle N type cycle x Dynamic sterilizer chamber pressure Air Leakage Empty chamber X Solid
68. ve Keep chamber trays door seal chamber and chamber face clean The door seal and chamber face should be checked for damage and cleaned every day using only a lint free cloth moistened ideally with Sterile Water for Irrigation or as a minimum the same water used in the reservoir refer to section 2 Clean the outside of the autoclave by wiping down with a cloth dampened with a 70 solution of industrial methylated spirit IMS and water Allow to dry by evaporation The tray lifter should be cleaned and sterilized regularly e g once a week CAUTION In common with other systems containing static water reservoirs water in this autoclave can become contaminated over a period of time or an aborted cycle and should be treated as a potential risk of infection It is recommended that the reservoirs are drained daily and allowed to dry Do not leave water in either reservoir when the autoclave is not in use To drain the reservoirs refer to section 2 0 Refill the fresh water reservoir refer to section 2 only when required For guidance on clean steam management and HTM2031 contact Eschmann Attention to the following will increase the life of the your Autoclave a Always leave the chamber empty and the door open on its secondary catch when not in use b Replace the antibacterial air filter Fig 1 item 23 on vacuum autoclaves every three months order Part No 115190 To remove the filter pull gently whilst twisting Replace
69. y pouch rack or cassette is 1 5 kg Maximum load for the porous load basket is 0 87 kg Maximum weight per item is 1 5 kg Maximum total load is 3 kg 225 autoclave Maximum load per tray pouch rack or cassette is 1 5 kg Maximum load for the porous load basket is 1 0 kg Maximum weight per item is 1 5 kg Maximum total load is 6 kg 6 4 Water and reservoir Reservoir capacity 3360 ml maximum Minimum fill 750 ml approx Volume used for 1 cycle 350 ml approx 6 5 1 Dimensions and Weight approx 113 autoclave Dimensions 582 mm wide 435 mm high 530 mm deep 435 5 mm depending on adjustment of feet Weight SES 113 35 0 kg SES 113 B vacuum 39 5 kg Shipping includes unit accessories and packing Dimensions WxDxH 720 x 720 x 640 mm Weight SES 113 46 5 kg SES 113 B vac 51 5 kg The useable space is defined as Porous load basket 280 x 147 x 87 mm external Trays cassettes etc 290 mm by cross section shown dimensions in mm e 61 34 20 145 20 ST IM84c SES 113 and 225 AUTOCLAVES 6 5 2 Dimensions and Weight approx 225 autoclave Dimensions 582 mm wide 435 mm high 580 mm deep 435 5 mm depending on adjustment of feet Weight SES 225 B vacuum 48 kg Shipping includes unit accessories and packing Dimensions WxDxH 720 x 720 x 640 mm Weight SES 225 B vac 60 kg The useable space is defined as Trays carrier 370