Home
H igh C ontrol S olution
Contents
1. un agente de ajuste de la viscosidad un conservante y un colorante La concentraci n de glucosa aproximada de la soluci n de control Alto OneTouch Select es de 0 35 En la etiqueta de sus tiras reactivas OneTouch Select est impreso el rango de resultados de glucosa esperado correspondiente a la soluci n de control Informacion de seguridad importante e Utilice nicamente las soluciones de control OneTouch Select con el medidor OneTouch Select Escriba la fecha de descarte fecha en que se abri m s tres meses en el espacio suministrado en la etiqueta del frasco No utilice la soluci n una vez transcurrida la fecha de caducidad o de descarte seg n lo que suceda primero nicamente para uso de diagn stico in vitro AA PRECAUCI N No ingiera la soluci n de control no es para consumo humano No aplique soluci n de control sobre la piel ni los ojos ya que puede causar irritaci n Almacenamiento y manipulaci n Cierre bien el envase y almac nelo a temperatura ambiente inferior a 86 F No refrigerar Procedimiento de prueba Lea detenidamente las instrucciones incluidas en su manual del propietario para realizar la prueba con la soluci n de control Materiales suministrados soluci n de control Alto OneTouch Select Materiales requeridos pero no suministrados su medidor manual del propietario tiras reactivas OneTouch Select Siga estos pasos cuando realice una prueba con la soluci n de con
2. the dispenser tip e Failure to match the result to the correct control range Refer to your owner s booklet for assistance If you are unable to resolve the problem contact Customer Service Used test strips may be considered biohazardous waste in your area Be sure to follow your local regulations for proper disposal A CAUTION If you continue to get control solution test results that fall outside the range printed on the test strip vial Do Not use the meter test strips or control solution Call LifeScan Customer Service at 1 800 227 8862 OUR COMMITMENT TO YOU Our goal is to provide you with quality health care products and dedicated customer service If you are not fully satisfied with this product or if you have questions about the use of any LifeScan product please call us at 1 800 227 8862 E LIFES e A a folinsen fofomon company LifeScan Inc Milpitas CA 95035 USA 2007 LifeScan Inc Milpitas CA 95035 Rev date 09 2007 AW06488001B Uso recomendado Para utilizar con el medidor de glucosa en sangre OneTouch Select para verificar que el medidor y las tiras reactivas estan funcionando de manera adecuada en conjunto y que se esta realizando la prueba correctamente Resumen y explicaci n La soluci n de control Alto OneTouch Select contiene una cantidad conocida de glucosa que reacciona con las tiras reactivas OneTouch Select El resultado que obtenga debe encontrarse dentro del rango de resultados q
3. Intended Use For use with the OneTouch Select Blood Glucose Meter to check that the meter and test strips are working together properly and that you are performing the test correctly Summary and Explanation OneTouch Select High Control Solution contains a known amount of glucose that reacts with OneTouch Select Test Strips The test result you obtain should fall within the range of results that you will find printed on the vial of the test strips that you are using A CAUTION The control solution range printed on the test strip vial is for control solution only It is used to test meter and test strip performance It is not a recommended range for your blood glucose level A WARNING Keep the meter and testing supplies away from young children Small items such as the battery door batteries test strips lancets protective disks on the lancets and control solution vial cap are choking hazards Do a control solution test e To practice the test process instead of using blood e Once a week Whenever you open a new vial of test strips If you suspect that the meter or test strips are not working properly If you have had repeated unexpected blood glucose results If you drop or damage the meter A single control solution test that is within the expected range is enough to tell you that the meter and test strips are working together and that you are doing the test properly The control solution consists of glu
4. cose in water with buffers and stabilizers a viscosity adjusting agent a preservative and a dye The approximate glucose concentration for the OneTouch Select High Control Solution is 0 35 An expected range of glucose results for the control solution is printed on the label of your OneTouch Select Test Strips Important Safety Information e Use only OneTouch Select Control Solutions with the OneTouch Select Meter Write the discard date date opened plus three months in the space provided on the vial label Do Not use after the expiration or discard date whichever comes first e For in vitro diagnostic use A CAUTION Do Not swallow control solution it is not for human consumption Do Not apply control solution to the skin or eyes as it may cause irritation Storage and Handling Close the bottle tightly and store at room temperature below 86 F Do Not refrigerate 810088790 MV Select High Control Solution ONE TOUCH Test Procedure Read the instructions carefully in your owner s booklet to perform the control solution test Materials provided OneTouch Select High Control Solution Materials required but not provided your meter owner s booklet OneTouch Select Test Strips Follow these steps whenever you test with OneTouch Select High Control Solution 1 Control solution tests must be done at room temperature 68 77 F Make sure your meter test strips and control s
5. dedicado Si no est totalmente satisfecho con este producto o si tiene alguna duda acerca del uso de cualquier producto de LifeScan ll menos al 1 800 381 7226 E LIFESCAN e afohmonfohmon company LifeScan Inc Milpitas CA 95035 USA O 2007 LifeScan Inc Milpitas CA 95035 Rev date 09 2007 AW06488001B
6. olution are at room temperature before testing 2 Shake the control solution vial well before testing 3 Discard the first drop of control solution you squeeze out and wipe the dispenser tip 4 Be sure that APPLY CONTROL appears on the meter display before applying control solution to the test strip See your owner s booklet for details 5 Apply the drop of control solution to the narrow channel in the top edge of the test strip 6 Your OneTouch Select Test Strips have two control ranges printed on the vial Be sure to compare the result to the correct control range Tightly close the cap on the control solution vial immediately after use to avoid contamination Each vial contains sufficient volume for a minimum of 75 tests Expected Results At room temperature OneTouch Select High Control Solution test results should fall within the expected range printed on the test strip vial label about 95 of the time If test results fall outside the expected range first repeat the control solution test Results outside the expected range may indicate e An error in the way you are doing the test Expired or contaminated control solution Damaged or expired test strips Use of a test strip or control solution past its discard date A problem with the meter Control solution that is too warm or too cold Failure to shake the control solution vial well Failure to discard the first drop of control solution and to wipe
7. rado primero repita la prueba con la soluci n de control Los resultados que est n fuera del rango esperado pueden indicar e Un error en la forma en que est realizando la prueba La soluci n de control ha caducado o est contaminada Las tiras reactivas est n da adas o han caducado Se est usando una tira reactiva o soluci n de control que ha pasado su fecha de descarte El medidor tiene alg n problema La soluci n de control est demasiado caliente o demasiado fr a El frasco de soluci n de control no se agit lo suficiente No se desech la primera gota de soluci n de control ni se limpi la punta del dispensador No se compar el resultado con el rango de control correcto Consulte su manual del propietario si necesita ayuda Si no puede resolver el problema p ngase en contacto con Servicios al cliente Las tiras reactivas usadas pueden considerarse residuos de riesgo biologico en su rea Aseg rese de seguir las regulaciones locales para su descarte adecuada A PRECAUCI N Si la prueba con soluci n de control contin a arrojando resultados que se encuentran fuera del rango impreso en la etiqueta del frasco de tiras reactivas no utilice el medidor las tiras reactivas ni la soluci n de control Llame a Servicios al cliente de LifeScan al 1 800 381 7226 NUESTRO COMPROMISO CON USTED Nuestro objetivo es ofrecerle productos de calidad para el cuidado de la salud y un servicio de atenci n al cliente
8. trol Alto OneTouch Select 1 Las pruebas con soluci n de control se deben realizar a temperatura ambiente 68 a 77 F Aseg rese de que su medidor las tiras reactivas y la soluci n de control est n a temperatura ambiente antes de la prueba 2 Agite bien el frasco de la soluci n de control antes de realizar la prueba Soluci n de control Alto ONE TOUCH Select 3 Deseche la primera gota de soluci n de control que saque y limpie la punta del dispensador 4 Aseg rese de que aparezca el mensaje APPLY CONTROL aplicar control en la pantalla del medidor antes de aplicar soluci n de control a la tira reactiva Consulte m s detalles en el manual del propietario 5 Aplique la gota de soluci n de control al canal angosto en el borde superior de la tira reactiva 6 Sus tiras reactivas OneTouch Select tienen dos rangos de control impresos en el frasco Aseg rese de comparar el resultado con el rango de control correcto Cierre bien la tapa del frasco de soluci n de control inmediatamente despu s de usarla para evitar que se contamine Cada frasco contiene un volumen suficiente para realizar 75 pruebas como m nimo Resultados esperados A temperatura ambiente los resultados de la soluci n de control Alto OneTouch Select deben encontrarse dentro del rango esperado impreso en la etiqueta del frasco de tiras reactivas aproximadamente el 95 de las veces Si los resultados se encuentran fuera del rango espe
9. ue encontrar impreso en el frasco de tiras reactivas que est utilizando A PRECAUCI N El rango de soluci n de control impreso en el frasco de tiras reactivas es para la soluci n de control nicamente Se utiliza para verificar el funcionamiento del medidor y las tiras reactivas No es un rango recomendado para el nivel de glucosa en su sangre A ADVERTENCIA Mantenga el medidor y los materiales para realizar las pruebas fuera del alcance de ni os peque os Los art culos peque os como la compuerta de la bater a las bater as las tiras reactivas las lancetas los discos protectores de las lancetas y la tapa del frasco de la soluci n de control conllevan riesgos de ahogamiento Realice una prueba con la soluci n de control e Para practicar el proceso de prueba en vez de utilizar sangre e Una vez por semana e Siempre que abra un nuevo frasco de tiras reactivas e Si sospecha que el medidor o las tiras reactivas no est n funcionando correctamente e Si ha obtenido resultados inesperados de glucosa en sangre en repetidas ocasiones e Si se cay o se da el medidor Una sola prueba con soluci n de control que muestre un resultado dentro del rango esperado es suficiente para confirmar que el medidor y las tiras reactivas est n funcionando de manera adecuada en conjunto y que se est realizando la prueba correctamente La soluci n de control consta de glucosa en agua con soluci n amortiguadora y estabilizadores