Programmable DC Power Supply User's Manual
Contents
1. ADDRess lt NR1 I gt set or return GPIB address 1 31 ADDRess lt 1 gt set return IP address CONFig lt STATic 0 DHCP 1 1 gt set or return IP config mode KEY LOCK Boolean 9 set or return key lock state 66 LCD BL Boolean 1 gt set or return LCD backlight state POWer CURRent lt NRf gt set or return user define current level 5 Boolean gt set or return user define output state lt OFF 0 LAST 1 USER 2 1 gt set or return power up mode VOLTage lt NRf gt set or return user define voltage level RECall DEFault recall factory default setting REMote USB 0 GPIB 1 ETHERNET 21 gt set or return remote interface SERies Return series number TIMER Timer Subsystem lt Boolean gt enable or disable timer mode return timer state HOUR lt NRI 1 gt set or return timer hours MINutecNRI gt set or return timer minutes SECond lt NR1 gt set or return timer seconds 67 Examples Q 28 How to cancel all actions ABOR ABORT Q 29 How to fetch current value FETC CURR FETCH CURRENT Q 30 How to fetch voltage value FETC VOLT FETCH VOLTAGE Q 31 How to measure current MEAS CURR MEASURE CURRENT Q 32 How to measure voltage MEAS VOLT MEASURE VOLTAGE Q 33 How to set up and read back the specific memory set MEM 1 MEMORY 3 MEM MEMORY Q 34 How to set up and read back the
2. No 30 fe No 31 EI E ix Terminator 5 12 1 Series Connection Command List The series connection command used by all DS series through ASCII string and must add the ending character 0x0A behind the string The system will respond after every data delivers For example the system will respond with the corresponding value or string when delivering the inquire command or respond when delivering the setting command In case any errors happen the system will respond what kind of error it is Please refer to Sec 5 12 2 ERROR RESPONSE LIST Command table is as below 51 SYSTEM CONTROL COMMAND Command Description CADR followed by address which can be 1 to 31 and is used to access the power supply CCLS Clear status CRST Reset command Brings the power supply to a known state CIDN Returns the power supply model identification CREV Returns the firmware version CSN Returns the serial number CST Returns the device status CCLR Clear protect OUTPUT CONTROL COMMAND Command Description CPV Sets the output voltage value in Volts CPV Reads the output voltage setting CMV Reads the actual output voltage CPC Sets the output current value in Amperes CPC Reads the output current setting CMC Reads the actual output current CDVC Display voltage and current data COUT Turns the output to ON or OFF COUT Returns the output On Off status COV Sets the OVP level COV Returns
3. 6 8 Error Input Message When users enter a voltage or current setting that is beyond the acceptable range system will display RANGE ERROR in response and show users the correct input range Press to re enter the voltage current setting RANGE ERROR MIN RANGE 0 000 MAX RANGE 36 000 57 7 Remote Interface communication protocol The communication protocol includes the MOTECH commands and SCPI commands 7 1 Prefaces The SCPI interface enables users to operate DS3640 DS6024 DS8018 0510014 through a computer or terminal that has IEEE 488 2 GPIB or USB interface equipped And it allows carrying out remote control and monitoring SCPI IEEE 488 2 version supports multi units control capability that allows a user to control up to 32 DS3640 DS6024 DS8018 DS 10014 units 7 2 Parameters Definition Type Valid Arguments boolean gt OFF or 0 NRI gt The data format NR 1 is defined in IEEE 488 2 for integers Zero positive and negative integer numeric values are valid data lt NRf gt The data format lt NRf gt is defined in IEEE 488 2 for flexible Numeric Representation Zero positive and negative floating point numeric values are some examples of valid data lt string gt Characters enclosed by single or double quotes lt NL gt New line Hex code is OxOAh lt Rtn gt Return Hex code is OxXODh END End or identify Note All commands shall be ended with the R
4. CHAIN ON OFF CHAIN ADDRESS On Off Series Connection Mode Setting Address 1 31 For the detailed setting information please refer to Sec 5 12 SERIES CONNECTION FUNCTION 29 4 1 2 Rear Panel 2 00000000000 0 OOOOOOO 00000 Q Q 9 O O get exe Rear Panel of DS3640 9 49 6 e SN on 29 ORLICE CH oooga o mE Ka D E ER ER RR x H H x 0000 OD III OOOCOOCO 554545455552 CER 0000000 Beep le 09 16 17 18 A A Raer Panel of 056024 058018 0510014 Cooling fans Automatically adjust fan s rpm according to the internal tempeture This is a wisdom design Power output terminal Please pay attention to the correct polarities when doing the connections RMT LCL Sense When Remote sense is selected the wire connection shall be as follows shown in the figure below positive sense 5 and positive lead of the DC output are connect
5. will be shown as below String Description OK No error Time out Wait response time out Range error Input value is out of range Multi master There are more than one controller in the whole system 54 6 Protections and Error Messages 6 1 Over voltage Protection OVP When the OVP is activated and voltage measured is over the setting point of protected voltage system will enter the Over Voltage Protect mode that will shut off the output and show OVP on the display Press to reset the protection mode and deactivate buzzer V 36 000 V I 40 000 OFF 0 000 V 0 000 A 6 2 Over current Protection OCP When the OCP is activated and current measured is over the setting point of protected current system will enter the Over Current Protect mode that will shut off the output and show OCP on the display Press to reset the protection mode and deactivate buzzer 36 000 V I 10 000 A OFF 0 000 V 0 000 A 6 3 Overpower Protection OPP When the OPP is activated and power measured is over the setting point of protected power system will enter the Over Power Protect mode that will shut off the output and display OPP on the screen Press to reset the protection mode and deactivate buzzer 36 000 V I 10 000 OFF 0 000 V 0 000 2 55 6 4 Constant Voltage Protection CV TO CC When this function is activated DS3640 DS6024 DS8018 DS 10014 will stay in CV mode If load chan
6. 1 PROG TOTA 8 Set program 1 to have 8 steps in total PROG STEP 1 Following 3 settings are for step 1 PROG STEP CURR 1 Set output current to 1 ampere PROG STEP VOLT 5 Output voltage is set to 5 volts PROG STEP ONT 0 1 Output ON time is set to 0 1 sec PROG STEP 2 Following 3 settings are for step 2 PROG STEP CURR 1 PROG STEP VOLT 10 PROG STEP ONT 0 1 PROG STEP 3 Choose step 3 PROG STEP CURR 1 PROG STEP VOLT 15 PROG STEP ONT 0 1 PROG STEP 4 Choose step 4 46 PROG STEP CURR 1 PROG STEP VOLT 20 PROG STEP ONT 0 1 PROG STEP 5 Choose step 5 PROG STEP CURR 1 PROG STEP VOLT 15 PROG STEP ONT 0 1 PROG STEP 6 Choose step 6 PROG STEP CURR 1 PROG STEP ONT 0 1 PROG STEP 7 Choose step 7 PROG STEP CURR 1 PROG STEP VOLT 5 PROG STEP ONT 0 1 PROG STEP 8 Choose step 8 PROG STEP CURR 1 PROG STEP VOLT 0 PROG STEP ONT 0 1 PROG NEXT 0 Select next program to run after program 1 is complete 0 means stop PROG SAV After edit use Save command to store program 1 in the hardware 1 run the program stored in the PROG RUN ON hardware it shall select program number and then use RUN ON command to execute the program 47 Example 2 Tek 1 CH1 5 00 To output the wave form shown above the following example program can Stop M Pos 2 0005 TRIGGER Type Source
7. Boolean LIMit VOLTage lt NRf 1 gt CURRent lt NRf gt 5 PROTection CLEar PROGram lt 11 gt Abort the Output Action Fetch Subsystem Return the fetched output current Return the fetched output voltage Measure subsystem Return the measured output current Return the measured output voltage Memory Subsystem select or return memory number range from 0 9 set or return voltage 0 36V 0 60V set or return current 0 40A 0 24A store memory subsystem parameters Output Subsystem enable or disable output action return output state set or return voltage limit value set or return current limit value Return output mode CV or CC Reset latched protection Program Subsystem select or return memory number range from 1 10 64 CLEar ALL NEXT lt 1 1 gt REPeat NR11 RUN Boolean gt SAV STEP lt Boolean gt CURRent lt NRf gt ONTimet lt NRf gt VOLTage lt NRf 1 gt lt NR1 gt PROTection PS 9 CCCV Boolean gt CLEar CVCC Boolean gt OCP Boolean gt LEVel lt NRf gt OPP Boolean gt LEVel lt NRf gt OVP Boolean gt LEVel lt NRF gt clear program n parameters clear all program parameters ser or return next program number 1 10 0 for end set or return repeat times max 50000 set or query program on off state
8. CH1 Slope Rising Made Normal Coupling CH1 2 50 Kall M 500 5 5 Jun 11 30 be used PROG 2 Choose program number PROG CLE Clear program 2 data 0 No repeat after running this program PROG TOTA 8 Set program 2 to have 8 steps in total PROG STEP 1 Settings for step 1 PROG STEP CURR 2 Set output current to 2 amperes PROG STEP VOLT 20 Set output voltage to 20 volts PROG STEP ONT 0 5 Set output ON time to 0 5 sec PROG STEP 2 Choose step 2 PROG STEP CURR 2 PROG STEP VOLT 15 PROG STEP ONT 0 5 PROG STEP 3 Settings for step 3 PROG STEP CURR 2 PROG STEP VOLT 20 PROG STEP ONT 0 5 PROG STEP 4 Choose step 4 PROG STEP CURR 2 48 PROG STEP VOLT 10 PROG STEP ONT 0 5 PROG STEP 5 Choose step 5 PROG STEP CURR 1 PROG STEP VOLT 20 PROG STEP ONT 0 5 PROG STEP 6 Choose step 6 PROG STEP CURR 2 PROG STEP VOLT 5 PROG STEP ONT 0 5 PROG STEP 7 Choose step 7 PROG STEP CURR 2 PROG STEP VOLT 20 PROG STEP ONT 0 5 PROG STEP 8 Choose step 8 PROG STEP CURR 2 PROG STEP VOLT 0 PROG STEP ONT 0 5 PROG NEXT 0 Select next program to run after program 2 is complete 0 means stop PROG SAV After edit use Save command to store program 2 in the hardware PROG2 To run the program stored in the PROG RUN ON hardware it shall select program number and th
9. mp See M t DS3640 Programmable DC Power Supply WEB CONTROL Home Configuration ee ON GOFF Web Contro Come Veet 3899 0365 bez 10 4054 D I ert MOTECH IDUSTRES INC A Rs Rasarad 7 7 2 Using Telnet Enter behind the command prompt under MS DOS Telnet DeviceIP 5024 IP address of the system is Device IP 5024 is as Telnet port will into 78 Welcome page shown as below Telnet 192 168 10 208 WELCOME TO DC POWER SUPPLY Input SCPI command behind the command prompt to communicate with the system Telnet 192 168 10 208 MELCOME TO DC POWER SUPPLY INC DS 3640 01234567 1 13 8 7 7 3 Using Sockets DS series using port 5025 to provide the standard function of SCPI socket Users can give SCPI command to achieve the purpose of controlling the systems through users program by using the standard SCPI socket 79 8 Assemble Accessories 8 1 Assemble Rack Mount Brackets The DS3640 DS6024 DS8018 DS10014 is designed to fit in a space of one rack unit 1U and can be mounted in a standard 19 inch rack panel or cabinet Rack mount brackets shall be assembled before mount in a rack Refer to the following figure to assemble the rack mount brackets 8 2 Assembly of Output Protective Shield DS6024 DS8018 DS10014 Safety regulation requests protective shield when the output voltage is h
10. This feature provides an extra safety for burn in room and electroplating applications DS3640 DS6024 DS8018 DS 10014 also provides over voltage protection OVP over current protection OCP and over power protection OPP features used to keep the output voltage and current within the safety level preventing the target from the damage caused by the over current The key lock feature is added to avoid accidentally wrong setting to the DS3640 DS6024 DS8018 DS10014 When the input power and the load change DS3640 DS6024 DS8018 DS10014 provides a stable output with the load and line regulation within 0 05 and the transient time less than 1 ms to maintain a very steady output When in remote mode DS3640 DS6024 DS8018 DS 10014 may output a new setting in 50 ms after receive a command This quick response time may increase the throughput on the mass production lines 1 2 1 2 3 4 Features Output Voltage amp Current Voltage output range 0 36V DS3640 0 60V DS6024 0 80V 058018 0 100V 0510014 Current output range 0 40 053640 0 244 056024 0 184 058018 0 14 44 0810014 Power output range 0 1440W Digital knob controller numerical keys and functions keys The knob controller can be used to rapidly change the output voltage setting and simulate the surge of the voltage output It successfully offers a good solution to the tests of triggering circuit Numerical keys facilitate us
11. output voltage and 0 40A for 053640 or 0 24A for DS6024 or 0 18A for DS8018 or 0 14 4A for DS10014 output current therefore the resolution of voltage and current will be different in response The screen thus will show as the following picture accuracy is 10 V 36 00 V I 40 00 OFF 0 00 V 0 00 5 10 Timer of Current Flow This function offers tests to the cutoff time of a breaker or a fuse First connect the test objective to output terminal as shown below 43 DS 3640 DS 3640 Press 6 and then 6 in the main page to enter to the current flow counter page V 10 00V I 1 00A b 0 00A OFF 00 00 000 0 ms After setting up output voltage current and press to output system will start counting the timer till the breaker or fuse is open The resolution of the timer is 100us 0 1 ms and the maximum counting period is one hour If counter doesn t work after pressing OniOff error occurs herein and screen will display error message after two seconds one of the following three cases may happen A Connector not ready If a fuse is not properly connected to the output connector or a breaker under test has not been switched to ON position the screen will glimmer to show an error message as shown below TEST FAIL CONNECTOR NOT READY 00 00 000 0 ms B Current setting too large The output current is set too big that a breaker is open or a fuse is burnt to ope
12. parts of this product are safety certified that can satisfy wide applications under the specified specification Users should thoroughly read over this Manual and follow the instructions to use them For example when changing side fasteners users should pay attention on the correct information regarding the order screw amount and respective lengths etc to ensure for a correct installation rather than making any mistake in assembly that would jeopardize user s rights and interests of warranty service Editing Version Ver 1 0 82 MOTIS MOTECH INDUSTRIES INC 6 F No 248 Sec 3 Pei Shen Rd Shen Keng Hsiang Taipei Hsien 222 Taiwan Tel 886 2 26625093 886 2 26625194 Fax 886 2 26625097 E mail instrument motech com tw http www motech com tw ZOMG 3640ME 2
13. save program parameters set or return step number set or return step n current setting set or return step n output time 0 050 200005 set or return step n voltage setting set or return program n total step numbers max 150 Protection Subsystem return protect state set or return CC to CV protect state Resets latched protection set or return CV to CC protect state set or return over current protect state set or return over current protect value set or return over power protect state set or return over power protect value set or return over voltage protect state set or return over voltage protect value Parallel Series Subsystem MODE lt OFF 0 PARALLEL 1 SERIES 2I gt 65 set or return parallel series mode TYPE lt MASTER 0 SLAVEA 1 SLAVEB 2 SLAVEC 3I gt set or return master slave setting SOURce Source Subsystem CURRent lt NRf gt set or return current level 0 40 0 24A PROTection Boolean gt set or return over current state LEVel lt NRf gt set or return over current level VOLTage lt NRf 1 gt set or return voltage level 0 36 0 60V PROTection Boolean gt set or return over voltage state LEVel NRf gt set or return over voltage level SYStem System Subsystem Boolean gt set or return state ESV Boolean 1 gt enable or disable extra 5V output ERRor return system error EX TernalcOFF 0 VOLT 1 RES 21 gt set or return external state
14. the CURRENT COUNTER TEST page 20 V 10 00V 1 1 00A Ib 0 00A OFF 00 00 000 0 ms Refer to Sec 5 10 Current Counting for the detailed setting procedure 6 2 PROGRAM MODE Press 2 to enter the PROGRAM MODE page PROGRAM NUMBER 1 PROGRAM OFF Before running the program user needs to input the programs through the USB or GPIB interface into DS3640 DS6024 DS8018 DS10014 Users may save 10 programs program number 1 through 10 in the hardware and recall here in this Program Mode page by selecting the program number and then pressing to execute the program 6 3 MEASURE AVERAGE Press 3 to enter the MEASURE AVERAGE page AVERAGE TIME 2 AVERAGE TIME Set the measure average time 7 TIMER CONTROL Press in the third page of Menu Setting to enter TIMER CONTROL page 21 TIMER OFF TIME 0Min 0 Sec TIMER Turn on off TIMER function TIME Set up OUTPUT ON time Max 999Hr 59Min 59Sec 8 CALIBRATION Press in the third page of Menu Setting to enter CALIBRATION page Users shall enter the password and then access to the calibration procedures PLEASE KEYIN PASSWORD _ 8 1 Equipment Requirments 1 5 1 2 DVM HP34401A 2 Shunt for Current Calibration 100A 10 mQ 22 Voltage Calibration 77 RMTI LCL Sense DS series DC Voltage measurement Figure 1 8 2 Calibration Items and Procedures 8 2 1 VOLTAGE CALIBRATION A Connect DS3640 o
15. the OVP setting level COVP Sets the OVP on off COVP Returns the OVP on off COC Sets the OCP level COC Returns the OCP setting level 52 COCP Sets the OCP on off COCP Returns the OCP on off COP Sets the OPP level COP Returns the OPP setting level COPP Sets the OPP on off COPP Returns the OPP on off CMODE Returns the power supply operation mode SYNCHRONOUS CONTROL COMMAND Command Description GRST Reset command Brings the power supply to a known state GCLS Clear status GCLR Clear protect GPV Sets the output voltage value in Volts GPC Sets the output current value in Amperes GOUT Turns the output to ON or OFF GOV Sets the OVP level GOVP Sets the OVP on off GOC Sets the OCP level GOCP Sets the OCP on off GOP Sets the OPP level GOPP Sets the OPP on off EXAMPLES Q1 How to read back ID for Address 5 on the system CADR 5 gt OK CIDN Motech inc DS 3640 A1234567 1 13 0 Q2 How to set up Voltage for Address 7 on the system CADR 7 gt OK CPV 20 gt OK 53 Q3 How to set up Output for Address 7 on the system CADR 3 gt OK COUT 1 gt OK Q4 How to read back Voltage value for Address 1 on the system CADR 1 gt OK CMV gt 10 001 Q5 How to set up Current for all the systems GPC 5 gt No response Q6 How to set up Output for all the systems GOUT 1 gt No response 5 12 2 Error Response List If PC delivers an error command or connection fails the response string
16. the output voltage and current To release from this operation mode parallel operation mode in this example press LCL to access to the series parallel setting page and turning the knob to select OFF for series parallel selection to leave the series or parallel mode and back to the local operation mode that the DS3640 DS6024 DS8018 DS10014 can be controlled by press the keys again It shall not change the operation mode while DS3640 DS6024 DS8018 DS 10014 is in output state or it will cause the Master s communication failure and display error message 5 8 2 Cascade Setting While connecting four sets of DS3640 DS6024 DS8018 DS 10014 in cascade the wiring is shown in the following figure 38 Power Line Power Line After wiring is complete select one DS3640 DS6024 DS8018 DS10014 as the Master and the other three as the Slave A B and C After the DS3640 DS6024 DS8018 DS10014 is set to the Master it will start searching for all Slaves that are connected to the Master Therefore in order to reach the control status user must set up the Slaves before the Master is set To set a D 3640 DS6024 DS8018 DS10014 to the slave mode press Menu CL and 4 in the main page to enter the series parallel setting option And then select the SERIES mode by turning the knob SCPI command is PS MODE SERIES and then press to confirm It will continue to the next line for the MASTER SLAVE selection T
17. voltage stored in specific memory set MEM VSET 10 MEM VSET MEMORY VSET 20 MEMORY VSET Q 35 How to set up and read back the current stored in specific memory set MEMIISET 15 MEM ISET MEMORY ISET 25 68 MEMORY ISET Q 36 How to save data to memory set MEM SAVE MEMORY S AVE Q 37 How to set up and cancel output OUT ON OUTPUT 0 Q 38 How to set up and read back the voltage limit OUT LIM VOLT 30 OUT LIM VOLT OUTPUT LIMIT VOLTAGE 35 OUTPUT LIMIT VOLTAGE Q 39 How to set up and read back the current limit OUT LIM CURR 30 OUT LIM CURR OUTPUT LIMIT CURRENT 35 OUTPUT LIMIT CURRENT Q 40 How to set up and read back the voltage SLEW RATE OUT SR VOLT 2 4 OUT SR VOLT OUTPUT SR VOLTAGE 0 01 OUTPUT SR VOLTAGE Q 41 How to set up and read back the current SLEW RATE OUT SR CURR 2 5 OUT SR CURR OUTPUT SR CURRENT 0 01 OUTPUT SR CURRENT Q 42 How to read back the output state OUT STAT OUTPUT STATE Q 43 How to read back the protection state PROT 69 PROTECTION Q 44 How to set up and read back the CC to CV protection state PROT CCCV ON PROT CCCV PROTECTION CCCV 0 PROTECTION CCCV Q 45 How to set up and read back the CV to CC protection state PROT CVCC ON PROT CVCC PROTECTION CVCC 0 PROTECTION CVCC Q 46 How to clear the state of protection PROT CLE PROTECTION CLEAR OUT PROT CLE OUTPUT PROTECTION CLEAR Q 47 How to set up and read back the overcur
18. C power source still try one s best to operate it in a background environment with low power noise and ensure that the system earth is made substantially If the power noise is unavoidable please install a power filter 3 4 Storage The storage temperature range of this product is within 10 70 and should be within 80 without moisture condensing If not operating this product for a long time interval pack it with original packaging or similar one and put it in a dry place without hit by direct sunlight 3 5 Power line voltage Rated AC power source connected to this product is within 100V 240V refer to the Product Specification for details Before connecting to external power source be sure that the power switch is in OFF state and look over the suitability of power cable including the extension line being as compatible to the rated voltage current and sufficient circuit capacity and then connect the cable tight Warning The power cable attached with this product is safety certified for this product operating in rated range To change a cable or add a extension cable be sure that it can meet the rated range of this product any misuse of additional cable would jeopardize user s rights and interests of warranty service 3 6 Fuses This product is a switching mode power supply The fuse installed inside is a multi barrier protection design of hardware It is hardly be breakdown under n
19. DS3640 DS6024 DS8018 DS10014 Programmable DC Power Supply User s Manual MOTIS Legal Notices The information in this document is subject to change without notice MOTECH makes no warranty of any kind with regard to this manual including but not limited to the implied warranties of merchantability and fitness for a particular purpose MOTECH shall not be held liable for errors contained herein or direct indirect special incidental or consequential damages in connection with the furnishing performance or use of this material MOTECH INDUSTRIES INC 6F NO 248 SEC 3 PEI SHEN RD SHEN KENG HSIANG TAIPEI HSIEN 222 TAIWAN R O C Copyright Notices Copyright 2008 MOTECH all rights reserved Reproduction adaptation or translation of this document without prior written permission is prohibited except as allowed under the copyright laws Warranty MOTECH persists in the faith of reliable quality guarantees and forever thoughtful service in one year since delivering the products we guarantee for free repairs on the troubles or damages of the products caused while they are under normal operation MOTECH does not take a repair free responsibility to any one of the following situations and we will charge to the respective repairs in accordance with the maintaining situation 1 The product is not sold by MOTECH s formally authorized agency directly 2 Because the irresistible calamity changes or force ma
20. DS6024 DS8018 DS10014 will be set to the last output state when power off If USER user defined is selected it will jump to next line to make setting of output voltage current and output state Then after power on DS3640 DS6024 DS8018 DS10014 will set up the output to ON OFF and the output voltage and current according the settings here HOT KEY OFF A HOT KEY Set the HOT KEY functioin ON OFF if the HOT KEY function is ON User can using 0 9 number keys to recall the voltage and current setting value in the memory that have been saved If entering the HOT KEY mode screen will present RMT symbol as shown in the following picture 17 36 000 V I 40 000 A OFF 0 000 V 0 000 A 2 OUTPUT SETTING Press 2 in the first page of Menu Setting to enter OUTPUT SETTING page VOLT LIMIT MAX _ 60 500 V CURR LIMIT MAX 24 500 A VOLT LIMIT MIN 0 000 V 7 CURR LIMIT MIN 0 010 A a VOLT SLEW RATE 3 0000 5 CURR SLEW RATE 1 2000 A mSY CONNECTOR DROP DISABLE EXT FULL VOLT 10 V VOLT LIMIT Limit of the output voltage setting CURR LIMIT Limit of the output current setting VOLT SLEW RATE Voltage ascending descending slope 3640 0 01 2 4V mS 6024 0 01 3V mS 8018 0 01 3 2V mS 10014 0 01 3 3V mS CURR SLEW RATE Current ascending descending slope 3640 0 01 2 5A mS 18 6024 0 01 1 2A mS 8018 0 01 0 72A mS 10014 0 01 0 48A mS CONNECTOR DROP Turn on off
21. GPIB ADDRESS 1 31 EXTERN CONTROL Set up the external control to voltage control VOLT 0 10 0 5V resistance control RES 0 5 or off OFF IP CONFIG STATIC A IP ADDRESS 255 255 255 255 KEY LOCK OFF Y IP CONFIG STATIC users input IP address DHCP IP address is assigned by the server IPADDRESS If IP CONFIG is set to STATIC users shall enter the IP addresses here If IP CONFIG is set to DHCP the assigned IP address will be shown here KEY LOCK While exit the setting page after enable KEY LOCK all keys except the key that can disable KEY LOCK status will be deactivated Simultanously presses both and keys in the main page can also unlock keys While entering KEY LOCK state screen will present LCK symbol as shown in the following picture 36 000 V 1 15 000 OFF 0 000 V 0 000 A V BEEP ON LCD BACKLIT ALWAYS ON RECALL DEFAULT NO BEEP Turn the Buzzer ON OFF 16 LCD BACKLIT Set the backlight of the LCD to always ON or OFF after 1 5 10 30 minutes RECALL DEFAULT Restore the manufactuer default setting EXT 5V OUTPUT OFF A POWER ON STATE OFF V 0 000 1 0 000 0 Ext 5V OUTPUT Turn the extra 5V power on the rear panel ON OFF POWER ON STATE Users may be able to set the output state after the power is on When OFF is selected DS3640 DS6024 DS8018 DS10014 will do nothing after power on If LAST is selected then after power on DS3640
22. O Operation 0 40 C Temperature Ratings S Storage 10 C 70 C Dimensions W H D 420 43 6 432mm Weight 9kg DS8018 DS10014 Specifications Output Rate DS8018 DS10014 Output Voltage 0 80 0 100 Output Current 0 18 0 14 4 Output Power 1440W 1440W Line Regulation Voltage lt 8mV lt 10 Current lt 4mA lt 4 Load Regulation Voltage lt 10 lt 12 Current lt 6 5mA lt 6mA Ripple and Noise 20Hz 20MHz Normal Mode Volage Load 20 5 Load 205 Normal Mode Current lt 50 lt 40 Resolution Programming 2mV 1mA 2 5mV 1mA Readback 2mV 1mA 2 5mV 1mA Programming Accuracy output offset Voltage 0 05 20mV 0 05 25mV Current 0 05 7mA 0 05 6mA Readback Accuracy output offset Voltage 0 05 20mV 0 05 25mV Current 0 05 7mA 0 05 6mA Transient Response Time lt 1mS lt 1mS Efficiency 280 280 OVP Adjustment Range 4 85V 5 105V OVP Accuracy 400mV 500mV Commond Response Time 50mS 50mS Power Factor ru di d Full load Full load Remote Sense Compensation 2V 2V Rising Time at Full Load lt 25mS lt 30msS Rising Time at No Load lt 25mS lt 30msS Falling Time at Full Load lt 25mS lt 30 5 Falling Time at No Load 1000mS 1000mS Standard Interface USB Standard Power Cord Terminal Block for Rapid
23. Over voltage Protection 55 6 2 Over current Protection OCH 55 6 3 Overpower Protection 55 6 4 Constant Voltage Protection CV TO 56 6 5 Constant Current Protection CV 56 6 6 Over temperature Protection OT 56 6 7 Low Voltage Protection 57 6 8 Error Input Message eee e eee e Re 57 Remote Interface communication protocol eere 58 7 1 Prtetaces i aee seda eas 58 7 2 Parameters Definition 58 7 3 The Error Event 59 7 4 Compatible Protocol of MOTECH LPS amp PPS Product Family 60 7 5 SCPI Conformable 63 7 5 1 SCPI 63 7 5 2 SCPI Command subevstem 64 7 6 State Bit Definition EE 75 77 LAN Communction 4 nennen nennen nennen 75 Using Web Server 76 1 7 1 1 2002020 ee tee tee to cede 76 7 7 1 2 Setting Page TI 7 7 1 3 Status of Page Stats enaena TI Web Control 78 4 70 Wsmg Telnet nee 78 PAS Usina SOCKS 79 TE OT E 80 8 1 Assemble Rack Mount Brackets seen 80 8 2 Assembly of Output Protective Shield DS6024 DS8018 DS 10014 80 8 3 Assembly of Remote S
24. Plug Connector Optional Interface LAN GPIB Optional Accessory RS 485 Cable General AC Line Rated Input Voltage por 1596 1096 Tolerance Variation in Voltage 10 power de rating mode when voltage under 95 Vac Rated Frequency 47Hz 63Hz Maximum Rated Input Power 1700VA Temperature Ratings O Operation 0 C 40 C Temperature Ratings S Storage 10 C 70 C Dimensions W H D 420 43 6 432 mm Weight 9kg Features of DS3640 DS6024 DS8018 DS10014 Timer 1 sec 100 hours Graphical LCD easy to read Compact high efficiency and high power density 40A output connector for fast connecting Numerical amp function keys for friendly setting Store and recall settings up to 10 sets Programmable SCPI command only with up to 10 sets of program and maximum 150 steps in total Extra 5V 1A output suitable for fixture circuit without the need of additional power Precise voltage and current measurement OVP OCP and key lock function Series amp parallel connection modes are applicable Series connection mode Max is up to 31 units using RS485 interface Mean measuring time per measurement is 50mSec Standard USB interface Optional interfaces GPIB LAN 3 3 1 Cautions Before Using Check and Confirm Accessories before Using As to safeguard your rights after receiving this product please confirm the items received in accordance with the ones li
25. RODUCT is designed meeting safety code and has passed the related qualification test To assure user the consistency of safety user should after equipping DC output cable according to field condition follow the Manual s suggestion to install the fixing sheets In case output cable is connected close the protection shield and fasten the screws in so as to protect user from getting an electric shock or other hazards 33 5 Operation Interpretation 5 1 Voltage Setting Press and set the output voltage by press numerical keys directly and then press to confirm it and complete the setting V 15 000 A OFF 0 000 V 0 000 A 5 2 Current Setting Press and set the output current by press numerical keys directly and then press to confirm it and complete the setting V 36 000 V 2 A OFF 0 000 V 0 000 A 5 3 Over voltage Protection OVP Press to enter the Configuration page and press C 3 to enter the PROTECTION setting page Then using the knob to set OVP to ON and press to confirm it Now the cursor will move to the value setting for the OVP on the right hand side Enter the OVP value here by press the numerical keys 38 000 V 42 000 A 1440 000 W Y 5 4 Over current Protection OCP Press to enter the Configuration page and press C 3 to enter the PROTECTION setting page Then using the knob to set OCP to ON and press 34 to confirm it Now the cursor will move to the va
26. T 1 gt output on OUT OFF gt output off Q 21 How to read state value STATUS gt return status value Q 22 How to read the firmware version VER gt return version information VERSION gt return version information 62 7 5 SCPI Conformable Information SCPI is entirely compatible with all IEEE 488 2 and SCPI commands 1995 Commands are included in SCPI commands in Vol 2 1995 7 5 1 Common SCPI commands Command Explanation CLS Clear status include error code IDN Response lt Manufacturer gt lt model gt lt serial number gt lt firmware type amp version gt RCL Recalls settings from memory Memory numbers from 0 to 9 are valid RST Resets the power supply to its power on state SAV 1 Saves defined parameters 2 Saves current settings to memory Memory numbers from 0 to 9 are valid Examples Q 23 How to store the voltage current settings to memory devices SAV 5 gt save current settings to memory location 5 Q 24 How to recall the voltage current settings from memory devices RCL3 gt recall setting from memory location 3 Q 25 How to set up software reset RST Q 26 How to identify the instrument s model number IDN Q 27 How to clear error message CLS 63 7 5 2 SCPI Command subsystem ABORt FETCh CURRent VOLTage MEASure CURRent VOLTage MEMory lt NR11 gt VSET lt 95 15 gt SAVE OUTput
27. To operate this product be sure that the protective earth connection of power is already linked firmly This precaution sustains for the other electrical equipment it collocates in operation to ensure personnel and equipment s safety Suitable user of this equipment is one who can hold basic electrical idea and comprehend this User Manual in details The power cable attached to this product meets with the related safety regulations and can secure operating safety under correct use To pull power plug out of receptacle user should hold the plug and pull it out Directly pulling power cable would result in danger At the instance of turning off power switch the residual electric charges in electric capacitors still might cause electrical shock to user We recommend user to wait for five minute till the inner fan ceases which means that the discharge process it finished Before carrying on general maintenance and safeguarding please shut off power supplied to this product pull out the plug and cutting off the joined load Use dry cloth to whisk or wipe off outer cover and cable for safety concern avoid using suds or organic solvent to clean this product Store Move Maintain Handle 3X 3X Storage When this device is not in use properly package it and store it on environment suitable for storage if in a good preserving environment the packaging process can be waived Freight While moving this product move it by us
28. V VOLTAGE 45 gt set voltage to 45V for DS6024 Q 7 How to read the voltage setting value VSET gt return voltage setting VOLT gt return voltage setting VOLTAGE gt return voltage setting Q 8 How to set up current ISET 1 1 gt set current to 1 1A CURR 4 3022 gt set current to 4 3022A CURRENT 0 250 gt set current to 250mA Q 9 How to read the setting value of current ISET gt return current setting CURR gt return current setting CURRENT gt return current setting Q 10 How to read the voltage outputting value VOUT gt return voltage output 61 Q 11 How to read the current outputting value IOUT gt return current output 12 How to lock buttons and the knob controller LOCK 1 gt lock the keypad and knob LOCK ON gt lock the keypad and knob Q 13 How to read the product model number MODEL gt return machine model name Q 14 How to set up OVP function OVP 1 gt enable OVP protect OVP OFF gt disable OVP protect Q 15 How to set up OVP voltage value OVSET 38 gt set OVP level to 38 V Q 16 How to set up OCP function OCP 1 gt enable OCP protect OCP OFF gt disable OCP protect Q 17 How to set up OCP electric current value OISET 40 gt set OCP level to 40 A Q18 How to set up OPP function OPP 1 gt enable OPP protect OPP OFF gt disable OPP protect Q 19 How to set up OPP power value OPSET 1000 gt set OPP level to 1000 W Q 20 How to set up output OU
29. YS REM GPIB SYS REM SYSTEM REMOTE ETHERNET SYSTEM REMOTE Q 68 How to read back the SERIES NUMBER SYS SER SYSTEM SERIES Q 69 How to set up and read back the output voltage SOUR VOLT 30 73 SOUR VOLT SOURCE VOLTAGE 35 SOURCE VOLTAGE Q 70 How to set up and read back the output current SOUR CURR 40 SOUR CURR SOURCE CURRENT 35 SOURCE CURRENT Q 71 How to set up and read back parallel series mode PS MODE PARALLEL PS MODE PS MODE 2 PS MODE OFF Q 72 How to set up and read back master slave in parallel series mode 5 MASTER PS MODE PS MODE SLAVEB PS MODE 3 Q 73 How to set up Program function Refer to Sec 5 11 Programmable capability 74 7 6 State Bit Definition When STATE command is used the system will return three 3 bytes in the order as shown below Byte 2 Byte 1 Byte 0 bit 7 bit 0 bit 7 bit 0 bit 7 bit 0 The definition of each bit is described in the following table byte 0 bit 7 bit 6 bit 5 bit 4 bit 3 bit 2 bit 1 bit 0 byte 1 bit 7 bit 6 bit 5 bit 4 bit 3 bit 2 bit 1 bit 0 byte 2 bit 7 0 7 7 LAN Communction OVP on off status OCP on off status OPP on off status CC to CV on off status CV to CC on off status output on off status LCD back light on off status External 5V output on off status OVP occur flag OCP occur flag OPP occur flag CC to CV occur flag CV to CC occur flag AC dectect
30. ct while using it For any double or unclear thing in operation please contact MOTECH or our agent s professional service personnel While removing the outer cover to repair by the professional person proper protection is still necessary to prevent people from getting an electric shock Forbid repacking this product or its accessories or it would cause danger to user and make the warranty failure Precautions on Normal Operation The recommended operation temperature of this product is 0 40 degrees C relative humidity 20 80 Do not boot this product when the ambient temperature or humidity is beyond the one specified above locating at a place filled with volatilized explosive gas flammable liquid materials Also appropriately locate the cable that won t stumble on passengers walking along corridor or is exerted under external stress To protect personal safety and warranty rights of product advise the user to visually look over the integrality of the appearance of this product before electrifying and starting this product If parts become flexible strength sunken on outer casing sharp angle or crack or under other flawed or damaged state contact MOTECH or its agent Avoid using the incomplete damaged cable or cable not acquiring the safety permission defined by related enactments To renew the cable or install additional extension cable it should also select the cable material meeting the specification of this product
31. e eue uuu uuu US 11 3 8 End 11 3 9 Cautions Im EEN 11 4 Panel Demonstration 12 4 1 DS3640 DS6024 DS8018 DS 10014 12 414 Aet eee teet 12 4 1 2 30 5 Operation Interpretation 34 5 1 Voltage Settina 52 mee eee te eate 34 KENE ee e te diete e ee 34 5 3 Over voltage Protection ONPR 34 5 4 Over current Protection OCH 34 5 5 Voltage gu UG 35 5 6 Voltage Output Controlled by Knob controller 35 5 7 te PIE else use see 35 5 8 Make Series Cascade Parallel Setting 36 5 8 1 Parallel Connection 5 36 5 8 2 Cascade Setting iiini Pena iones 38 5 8 3 Error Message of Series Parallel Connection 41 5 9 External T ning Setting i c rt e RR EFE 42 5 10 Timer of Current low 43 5 11 Programmable Capability SCPI Command Only 45 5 12 Series Connection Function RS485 sss 51 5 12 1 Series Connection Command List esses 51 512 2 Error Response 54 Protections and Error Messages eese 000 etn ntnnnn 55 6 1
32. e switched on or off in the System Setting menu USB It is the USB interface connector EXT CTL DS3640 DS6024 DS8018 DS10014 offers the capability of setting the output voltage current by adjusting an external input voltage resistance The range of the external input voltage is 0 10Vdc or 0 5V which is corresponding to the output voltage of 0 36V for 053640 and 0 60V 256024 and 0 80V for DS8018 and 0 100V for DS10014 and corresponding to the output current of 0 40A for DS3640 and 0 24A for 056024 and 0 18A for 058018 and 0 14 4A for DS10014 The range of the external resistance is 0 5K ohm which is corresponding to the output voltage of 0 36V for DS3640 and 0 60V for DS6024 and 0 80V for DS8018 and 0 100V for DS10014 and corresponding to the output current of 0 40A for DS3640 and 0 24A for DS6024 and 0 18A DS8018 and 0 14 4A for DS10014 RS485 While at series parallel connection or multi unit series connection RS485 interface can be used for communications and synchronize signals between master and slavers Protection shield of the RMT LCL Sense connector DS6024 DS8018 DS10014 WARNING When the RMT LCL Sense is not activated and or not used the protection cover shall be keeping in shielded state do not dismantle it that might get an electric shock 32 28 Protection shield of the output leads DS6024 DS8018 DS10014 WARNING This P
33. ed to the positive end of the device whereas negative sense S and negative lead of the DC output are connected to the negative end of the device under test This connection will compensate the voltage dropped due to the current flow through the long power wires the maximum compensating voltage is 2V 30 19 Q0 Q1 Q2 LOAD LOAD Local Sense Remote Sense When Local sense is selected the wire connection shall be as follows shown in the figure above positive sense S is connected to the positive lead and negative sense S is connected to the negative lead whereas the positive lead of the DC output is connected to the positive end of the device and the negative lead of the DC output is connected to the negative end of the device under test When this sensing mode is selected the power wires from the DC ouput leads to the device under test shall be as short as possible LAN optional The ETHERNET interface connector GPIB optional The GPIB interface connector AC power input The power receptacle is for power source within 100Vac 240Vac Earthing point Used for earth ground connection 31 Q3 Q4 Q5 26 27 5V 1A Output DS3640 DS6024 DS8018 DS10014 offers an extra output that the constant output voltage is 5V and the maximum output current is 1 amperes This extra power supply can b
34. en use RUN ON command to execute the program 49 Example 3 Tek P Ready M Pos 2 3005 TRIGGER Type Source CH1 r Slope Rising 1 1 1 Coupling 5 00 50075 2 604 5 Jun 08 11 32 lt 10Hz If it needs to execute Program 2 right after Program 1 is executed then program 1 shall be modified to have NEXT 2 command The following steps can be taken for the modification and execute the programs PROG 1 Select program 1 PROG NEXT 2 Change the NEXT command from NEXT 0 to NEXT 2 PROG SAV After edit is complete use Save command to store changes in the hardware 1 To run the program by selecting the PROG RUN ON program number first and then using RUN ON command to execute it 50 5 12 Series Connection Function RS485 DS3640 DS6024 DS8018 DS10014 can use RS485 to provide multi units series connection function for up to 31 units If more than 10 units please add the resistance 120 in the last unit as shown in the below figure Turn on the system after series connection is completed Press gt 9 on the main page and set CHAIN ON OFF as ON Series Parallel connection will be cancelled and set each unit with the distinct Address 1 31 Then by using USB to connect PC to any unit multiple units can be controlled by using the commands in Sec 5 12 1 SERIES CONNECTION COMMAND LIST No l No 2
35. ense Protective Shield 056024 058018 0810014 80 SA CCESSOLIOS oro dome 82 1 Preface 1 1 Products Outline MOTECH DS3640 DS6024 DS8018 DS10014 is programmable DC power supply with single output that offers the maximum power output up to 1440 watts 0 36V 40A or 0 60V 24A or 0 80 18 or 0 100V 14 4A With a 16 bit D A A D converter embedded DS3640 DS6024 DS8018 DS10014 comes with the resolution of 1mV in voltage setting and 1mA in current setting With the cascaded series or parallel connections of up to four 4 devices of DS3640 DS6024 DS8018 DS10014 it may be able to provide a maximum power output up to 5760 watts When four 4 DS10014 are cascade connected the maximum output voltage can reach up to 400V When four 4 DS3640 are put in parallel connections the output current can reach up to 160A The feature of series parallel connections makes DS3640 DS6024 DS8018 DS10014 more convenient and flexible for varies applications DS3640 DS6024 DS8018 DS10014 has provided a knob controller numerical and function keys for users to manipulate this electronic instrument more easily and fast Also DS3640 DS6024 DS8018 DS 10014 provides a memory space for user to store 10 settings that can be recalled directly This feature offers an easy way to restore the application setting In addition users can use a program to control when to cut off the output
36. eration mode in this example press LCL to access to the series parallel setting page and turning the knob to select OFF for series parallel selection to leave the series or parallel operation mode and back to the local operation mode that the DS3640 DS6024 DS8018 DS 10014 can be controlled by press the keys again It shall not change the operation mode while DS3640 DS6024 DS8018 DS10014 40 is in output state or it will cause the Master s communication failure and display error message 5 8 3 Error Message of Series Parallel Connection If RS485 wiring is wrong or the signal is not correct Master will display the following message in the page of searching Slaves SELECT MODE PARALLEL MASTER SLAVE MASTER FOUND SLAVE NONE If more than one Master being set in the series parallel operation mode the following message will be displayed SELECT MODE PARALLEL MASTER SLAVE MASTER MULTI MASTER PLEASE CHECK AGAIN If one of the DS3640 DS6024 DS8018 DS 10014 is set to different series parallel operation mode screen will appear the following message SELECT MODE PARALLEL MASTER SLAVE MASTER ERROR MODE PLEASE CHECK AGAIN After finishing setting procedure if the Master cannot have a communication with one of the three 3 Slaves Master will present the following error message Slave A in this example SLAVE A COMMUNICATION ERROR 41 After finishing setting procedure if a Slave receives only
37. ers making fast setting without withstanding the conventional VR analog simulation Using function keys to switch modes makes the overall operation more humanizing and easy to deal with Precise voltage and current measurement Besides the precise output DS3640 DS6024 DS8018 DS 10014 also offers the capacity of voltage amp current measurement read back accurately that users may be able to save extra expense and space for extra measuring instruments Memory device and timer function DS3640 DS6024 DS8018 DS10014 provides a memory space for users to save 10 settings This feature makes users no need to hold and memorize too many settings in mind They simply need to store the settings in DS3640 DS6024 DS8018 DS 10014 hardware and restore the setting at any time The DS3640 DS6024 DS8018 DS 10014 has one 1 timer with the resolution of 1 second The timers are used to time the outputs When the timer counts down to zero the DS3640 DS6024 DS8018 DS10014 will automatically turn the output off This feature is useful when the DS3640 DS6024 DS8018 DS 10014 is providing power to the test object in burn in room The programmable capability allow this equipment to run at a burn in room that operator needs not to care about time setting points because this equipment will shut off output when the specific set time points reaches It has covered the advantages of both safety and flexibility DS3640 DS6024 DS8018 DS10014 will perform perfect control of
38. ges and forces the system turning from CV to CC constant current mode system will enter the TO CC Protect that will shut off the output and display the CVC message on the screen Press to reset the protection and deactivate buzzer V 36 000 10 000 OFF 0 000 V 0 000 A 6 5 Constant Current Protection CC TO CV When this function is activated DS3640 DS6024 DS8018 DS 10014 will stay in the CC mode If load forces the measurement turn from CC to CV constant voltage mode system will enter the CC TO CV Protect that will shut off output and display CCV message on the screen Press to reset the protection and deactivate buzzer 36 000 V 10 000 OFF 0 000 V 0 000 A 6 6 Over temperature Protection OTP When machine has detected abnormally high temperature system will enter the Over Temperature Protect mode that will shut off output and display the erro message as shown in the following figure Press to reset the protection and deactivate buzzer OTP ERROR TEMPERATURE IS OVER HEATING PLEASE CHECK AND TRY AGAIN 56 6 7 Low Voltage Protection ACD When the machine has detected abnormally low AC power input system will enter the Detect Low Protect mode that will shut off output and display the error message shown in the following figure Press to reset the protection and deactivate buzzer AC DETECT LOW LINE VOLTAGE IS UNDER SPEC OR POWER SHUTDOWN
39. igher than 45Vdc Therefore there are protective shields for the output connectors for the DS6024 DS8018 DS10014 Refer to the following figure for the assembly disassembly of the output protective shield SE Step1 Step 22 8 3 Assembly of Remote Sense Protective Shield 80 DS6024 DS8018 DS10014 Safety regulation requests protective shield when the voltage is higher than 45Vdc Therefore there are protective shields for the remote sense connectors for the DS6024 DS8018 DS10014 Refer to the following figure for the assembly disassembly of the remote sense protective shield 81 9 Accessories Product Name DS3640 DS6024 DS8018 DS 10014 Parts List Item Name amp Specification Qty Remarks 1 Terminal Block for Rapid Plug 1 Connector 2 Fixing sheet slice 2 3 Hand holding bar 2 4 Operation Manual 1 5 Power Cord 1 6 Screw M4x8 10 7 Screw M5x8 4 8 Guarantee certificate WARNING All
40. ing the original wrappings to pack this product in advance If the wrapping material is lost use an equivalent buffer material to replace it in packaging and with external marks indicating fragile amp water prevention This product is sort of precise device please try one s best to use qualified means of transport to transport it And try one s best to avoid dropping or others that would jeopardize this product Maintenance This product is free from any maintainable item for user normal user to maintain about except the one indicated in this Manual When any abnormal condition happens to this product under user s judgment please contact us or our agent immediately Never try to maintain this product by oneself that would cause unnecessary danger in consequence and which might cause even severe damage to this product Disposal When the device in badly condition and can t be used or repaired please discard it according to your company disposal procedures or local legal procedures Don t discard arbitrary to avoid polluting environment Index M VE 1 1 1 Products Outline as ade ite t te 1 1 2 2 2 SPECHIEATION 4 3 Cautions Before Using nn 9 3 1 Check and Confirm Accessories before Uemng 9 3 2 Operation 5 9 3 3 Ambient Environment 9 3 4 Storage HE 10 3 5 Power line voltage oec ee eee ee S 10 3 6 USCS c En 10 3 7 Preheatiuig Tim
41. jeure which may attribute to the user who has not followed the User Manual to operate or caused by user s fault such as malfunction or damage made from improper operation or others 3 User dismantles or repairs or repacks the product or installs additional accessories to it without MOTECH s permission or authorization which cause the trouble or damage of the product s Within guaranteeing service duration user should be responsible and afford respective expense to transport the trouble or damaged product to MOTECH or the place which MOTECH specifies And MOTECH will bear the expense in delivering the fixed product back to user Taiwan area only or user specified place Taiwan area only User shall apply for the transporting insurance from the insurance company alone MOTECH INDUSTRIES INC 6F NO 248 SEC 3 PEI SHEN RD SHEN KENG HSIANG TAIPEI HSIEN 222 TAIWAN R O C Customer Service Hotline 886 2 2662 5093 Fax No 886 2 2662 5097 E mail instrument motech com tw Website http www motech com tw Important Safe Operation Guidelines Properly Preserve This User Manual This Manual contains important graphical amp literal interpretations regarding installations maintenance and safety in operation etc Qualified Professional Maintenance Personnel As stated in this User Manual neither spare parts nor exclusive fine tuning is included inside this product forbid user removing the outer cover of a produ
42. low occur flag OTP occur flag reserved reserved DS series provide three LAN control modes including Web server Telnet and Sockets First enter the first Item System Setting on the menu to set the interface selection Remote Control as Ethernet and select IP setting procedure as Static or use DHCP Dynamic Host Configuration Protocol If using static IP it needs to set up IP address After completed it will be able to use the above LAN control mode to control the DS series 75 7 7 1 Using Web Server DS series have built in Web Server User can control the DS series on the computer by using the Web browser Open Web browser and input IP address to enter Welcome page Kp sito IA e D S 3 6 4 0 Programmable DC Power Supply Welcome to Motech 095 3640 status Web Login Page Please key in the password to get in the web control page Data Sheet or key in GUEST to visit the web page PASSWORD von Enter password presetted as 123456 to login the main page Home and then click to select the Items on the left side to the other settings or control page 7 7 1 1 Main Page Home The main page shows all the basic message and network information 76 pne 5 3 6 4 0 Programmable DC Power Supply IMFORMATION E Programmatte DC Power Supply Data Sheet 0017 28 01 2 0 mortech com tw Fo
43. lue setting for the OCP on the right hand side Enter the OCP value by press the numerical keys ON SET 38 000 V ON SET 42 000 A OFF SET 1440 000 W Y 5 5 Voltage Output After the voltage current and protection point settings press to output voltage User shall be able to recognize the setting values and the actual output values from the LCD V 36 000 V I 15 000 A 35 999 V 0 000 A 5 6 Voltage Output Controlled by Knob controller When the output is ON user can may still increase or decrease the output voltage by turnning the knob controller The procedure is press and cursor appears in response press or to move the cursor to the digit that change will make and turn the knob to increase or decrease the output voltage The changes of the voltage seting and the output voltage can be observed concurrently V 36 000 V I 15 000 A 35 999 V 0 000 A 5 7 Timer Function When the Timer Setting function at item 7 of system parameters is ON it will activate the timer After time setting is made return to the main page After setting up the output current amp voltage and press to output the screen now will show the countdown of Timer that when it counts down to zero 35 system will turn off the output automatically TIMER 000 00 55 Sec 39 999 V 0 000 A 5 8 Series Cascade Parallel Setting The DS3640 DS6024 DS8018 DS 10014 uses the series parallel mode 4 sets in maximum
44. merical setting or jump to the previous page 12 7 8 9 10 Numerical keys 0 9 They are used to fast input the voltage or current value or choose the setting option in Menu page Down Right Store This key is a compound key for the following three functions Down Right Store In Menu Setting status use this Down key to move cursor to the next item Under Output status use this key to move cursor right Under Memory Setting status use this key to store setting to the selected memory set Up Left Recall This key is a compound key for the following three functions Up Left Recall In Menu Setting status use this Down key to move cursor to the up item Under Output status use this key to move cursor left Under Memory Setting status use this key to recall setting from the selected memory set Display Display In Menu Setting page press Display to return to main page or switch the display to show the power and the loading resistance V 36 000 I 15 000 A OFF 0 000 V 0 000 A 36 000 V I 15 000 A OFF 0 00W 0 00 13 11 12 13 14 45 Output On Off Control the On Off of the output power The knob controller Use this tuner to adjust voltage or current press first to let cursor display first It is an option item in Menu Setting Enter Enter This key is the confi
45. n ADDRess set the address of the machine ADDRess return the address of the machine BEEP set beep on 1 or off 0 CLR clear protect status CURRent set current CURRent return current setting ERRor return error message IOUT current readback ISET set current ISET return current setting LOCK set rotary and keypad lock on 1 or off 0 MODEL return model name OCP set current protect to off 0 or on 1 OISET set overcurrent protect level OISET return overcurrent value OPP set power protect to off 0 or on 1 OPSET set overpower protect level OPSET Return overpower value OUT set output on 1 or off 0 OVP set voltage protect to off 0 or on 1 OVSET set overvoltage protect level OVSET return overvoltage value STATUS return status of the machine VERsion return version number VOLTage set voltage VOLTage return voltage setting VOUT voltage readback VSET Set up output voltage VSET return voltage setting 60 Examples Q 1 How to set GPIB address ADDR 10 gt address is 10 Q 2 How to read back GPIB address ADDR gt return GPIB address ADDRESS gt return GPIB address Q 3 How to set up buzzer BEEP 1 gt trigger beep to on BEEP off gt trigger beep to off Q 4 How to clear the protecting state CLR gt clear protect status Q 5 How to read back to error information ERR gt return error code Q 6 How to set up voltage VSET 10 gt set voltage to 10V VOLT 3 3V gt set voltage to 3 3
46. n before output current is reaching the setting value The screen will glimmer to 44 show an error message as shown below TEST FAIL I SETTING TOO LARGE 00 00 000 0 ms C Voltage setting too small The voltage is set too small that output current is unable to reach the setting value The screen will glimmer to show an error message as shown below TEST FAIL V SETTING TOO SMALL 00 00 000 0 ms 5 11 Programmable Capability SCPI Command Only DS3640 DS6024 DS8018 DS10014 provides the capability for users to program it There are memory spaces to store 10 programs that can have 150 steps in total for setting purpose Each program does not restrict the step quantity however the sum of 10 programs are restricted to 150 steps For each program users can set up how many times to repeat the program For each step users may be able to set up the output voltage current and period of time 50 ms minimum to output the settings Please refer to Sec 7 5 2 SCPI command subsystem for detail SCPI commands 45 Example 1 Tek M Pos 332 0ms Print i Stop Ink Saver Layout File Format 7 2 60 lt 10H2 CH1 5 00 100ms 5 08 11 25 To output the wave form shown above users may edit the program as the following orders PROG 1 Choose program number PROG CLE Clear program 1 data PROG REP 0 No repeat repeat one time for
47. n DVM fill in with the corresponding function and press ENTER key 27 FIX PARAMETER 10 000 CALIB CURR Lo 0 604 mV CALIB CURR MIDL 128 981 mV CALIB CURR MIDH 258 246 mV CALIB CURR Hi 377 559 mV C Please inspect the hardware if the current value is inconsistent with the value in the tolerance error table after the value of calibration function is transformed to actual current value DS 3640 Function Setting Range of Transformation Value Lo 0 1A 0 0 5A MIDL 13A 11 14A MIDH 26A 23 27 A Hi 38A 34 39 05 6024 Function Setting Range of Transformation Value Lo 0 06A 0 0 5A MIDL 7 8 6 6 9 MIDH 15 6A 14 17 2 Hi 22 8A 21 24 6 A DS 8018 Function Setting Range of Transformation Value Lo 0 045A 0 0 MIDL 5 9A 5 3 6 5 MIDH 12A 10 8 13 2 Hi 17A 15 3 18 7 DS 10014 Function Setting Range of Transformation Value Lo 0 03625A 0 0 1A MIDL 4 7A 4 2 5 2 MIDH 9 5A 8 55 10 45 A Hi 13 7A 12 33 15 07 A D The calbriation value will be stored in FLASH after pressing ENTER at Hi function 8 2 4 CONNECTOR DROP CALIBRATION Not available 8 2 5 EXTERNAL VOLTAGE CALIBRATION Not available 8 2 6 EXTERNAL CURRENT CALIBRATION Not available 9 Series Connection Control Setting CHAIN Press 9 in Menu setting page to enter CHAIN SETTING page CHAIN ON OFF CHAIN ADDRESS
48. nal adjustment state SYS EXT VOLT SYS EXT SYSTEM EXTERNAL RESISTANCE SYSTEM EXTERNAL Q 57 How to set up and read back the GPIB address SYS GPIB ADDR 5 SYS GPIB ADDR SYSTEM GPIB ADDRESS 6 SYSTEM GPIB ADDRESS 58 How to set up and read back the IP address SYS IP ADDR 192 168 0 208 SYS IP ADDR SYSTEM IP ADDRESS 192 168 10 10 SYSTEM IP ADDRESS Q 59 How to set up and read back the IP mode SYS IP CONF DHCP SYS IP CONF SYSTEM IP CONFIG STATIC SYSTEM IP CONFIG Q 60 How to set up and read back the key locking SYS KEY LOCK ON SYSTEM KEY LOCK Q 61 How to set up and read back the LCD backlighting function SYS LCD BL ON SYSTEM LCD BL Q 62 How to set up and read back the booting mode SYS POW TYPE LAST SYS POW TYPE SYS POWER TYPE USER 72 SYS POWER TYPE SYSTEM GPIB ADDRESS Q 63 How to set up and read back the voltage under user defined booting mode SYS POW VOLT 10 SYS POW VOLT SYSTEM POWER VOLTAGE 20 SYSTEM POWER VOLTAGE Q 64 How to set up and read back the current under user defined booting mode SYS POW CURR 10 SYS POW CURR SYSTEM POWER CURRENT 20 SYSTEM POWER CURRENT Q 65 How to set up and read back the output state under user defined booting mode SYS POW STAT ON SYS POW STAT SYSTEM POWER STATE 0 SYSTEM POWER STATE Q 66 How to recall back default setting SYS REC DEF SYSTEM RECALL DEFAULT Q 67 How to set up and read back the communication interface S
49. ormal operation status In case the fuse is melted to break it shows another malfunction exists that causes the fuse to break In such case it is suggested to send this product back for services Warning To sustain user s rights of warranty service and avoid misconnection that would cause jeopardy from it we don t recommend user to dismantle the casing and change fuse by oneself 10 3 7 Preheating Time All movements of this product will activate at the same time when the power is on However to reach the specified equipment accuracy please preheat it for at least 30 minutes 3 8 End test When testing is over and not to be restarted for a while or when this product isn t under use or user needs to leave during operation be sure to turn the power switch on the panel to OFF position as to turn off the power After power switch is turned to OFF position the inner fans will still run for a few seconds to carry on the inside electric capacitor discharge process per safety code requirement Once the discharge process is complete this product will make automatic shut down in response 3 9 Cautions in Operation A While under cascaded connection mode each DS3640 DS6024 DS8018 DS10014 shall be in power on state and output shall be ON In case there is any one DS3640 DS6024 DS8018 DS 10014 is in power off state or output is OFF the associated output current will flow over the output bypass diode of
50. r help and support please out website Copysight 2008 MOTECHINDUSTRES INC All Righes 7 7 1 2 Setting Page Configuration Input each argument and reset password for the system 1B Nae EC oett D S 3 6 4 0 Programmable DC Power Supply Sm Mv zeo IAD CONFIG Home Configuration OVP Seming Ov van PIE Status OCP Secing Gorr 0200 0 434 Nen conet OPP Setting OPP 1889000 0 taoa w Pata Sheet Extra LCD back ight OFF OFF unis OFF emm tech ao Change Password new CONFORM NEY For heb and support please wait out website Copyright 2008 MOTECH INOUSTRES INC Rights Seet 7 7 1 3 Status of Page Status Display the status of the system In case of any error codes please refer to Sec 7 3 Error Event list T TE Vo DC Rug CS wm D S 3 6 4 0 Programmable DC Power Supply STATUS Home Configuration laa troc Warning mei Status L Web Control Data Sheet 7 7 1 4 Web Control Set up voltage current and output In addition monitoring the output voltage current value on line is also allowed JAVA support is required D amp
51. r the MASTER SLAVE selection Turning the knob to select SLAVE SCPI command is PS TYPE SLAVEA for the DS3640 DS6024 DS8018 DS10014 and press to confirm the setting Using the same procedure to set up SLAVE B and C for another two 2 DS3640 DS6024 DS8018 DS 10014 as graphically demonstrated below SELECT MODE PARALLEL MASTER SLAVE SLAVE A To set the master unit press Menu Cy and 4 in the main page to enter the series parallel setting option And then use the knob to select the parallel mode SCPI command is PS MODE PARALLEL and press to confirm It will continue to the next line for the MASTER SLAVE selection Use knob again to select MASTER SCPI command is PS TYPE MASTER and then press to confirm After it is confirmed as the Master the DS3640 DS6024 DS8018 DS10014 will start searching for all Slaves as graphically demonstrated below SELECT MODE PARALLEL MASTER SLAVE MASTER CHECKING FOR SLAVE If wiring is correct screen will appear 37 SELECT MODE PARALLEL MASTER SLAVE MASTER FOUND SLAVE ABC After receiving the control order from the Master all Slaves will be locked on the SLAVE page that all keys except are disabled so that slaves can not be operated by press keys except controlled by the Master as graphically demonstrated below MODE PARALLEL NONE SLAVE Once the Master and Slave settings are done users can then operate the Master only to set up
52. rent protection state PROT OCP ON PROT OCP PROTECTION OCP 0 PROTECTION OCP SOUR CURR PROT ON SOUR CURR PROT SOURCE CURRENT PROTECTION 0 SOURCE CURRENT PROTECTION Q 48 How to set up and read back the overcurrent protection point PROT OCP LEV 30 PROT OCP LEV PROTECTION OCP LEVEL 40 PROTECTION OCP LEVEL SOUR CURR PROT LEV 25 SOUR CURR PROT LEV SOURCE CURRENT PROTECTION LEVEL 35 SOURCE CURRENT PROTECTION LEVEL 70 Q 49 How to set up and read back the overpower protection state PROT OPP ON PROT OPP PROTECTION OPP 0 PROTECTION OPP Q 50 How to set up and read back the overpower protection point PROT OPP LEV 30 PROT OPP LEV PROTECTION OPP LEVEL 40 PROTECTION OPP LEVEL Q 51 How to set up and read back the overvoltage protection state PROT OVP ON PROT OVP PROTECTION OVP 0 PROTECTION OVP SOUR VOLT PROT ON SOUR VOLT PROT SOURCE VOLTAGE PROTECTION 0 SOURCE VOLTAGE PROTECTION Q 52 How to set up and read back the overvoltage protection point PROT OVP LEV 30 PROT OVP LEV PROTECTION OVP LEVEL 40 PROTECTION OVP LEVEL SOUR VOLT PROT LEV 25 SOUR VOLT PROT LEV SOURCE VOLTAGE PROTECTION LEVEL 35 SOURCE VOLTAGE PROTECTION LEVEL Q 53 How to set up the buzzer SYS BEEP ON SYSTEM BEEP 0 54 How to set up extra 5V voltage output SYS E5V ON SYSTEM ESV 0 71 Q 55 How to read back error code SYS ERR SYSTEM ERROR Q 56 How to set up and read back the exter
53. rming key of current or voltage setting value or press under output status to dynamically adjust voltage at CV mode or current at CC mode Mem Mem Press this key to enter the memory device status page Users can then use the numerical key or knob to select the target memory set to save or recall the configuration by pressing the STORE or RECALL key Ten sets are available in selection MEM 0 0 000 V 0 000 A Menu Use this key to make system parameters setting There are eight 8 major items under operation Users may press and the numerical keys to enter the corresponded setting page Press Or key to jump to one of the following three pages 1 SYSTEM SETTING 2 0UTPUT SETTING 3 PROTECTION 14 4 SERIES PARALLEL 5 INFORMATION 6 SPECIAL TEST FUNC 7 TIMER CONTROL 8 CALIBRATION 9 CHAIN SETTING 1 SYSTEM SETTING Pressing 1 key in the first page of Menu Setting can enter the following SYSTEM SETTING page REMOTE CONTROL GPIB ADDRESS EXTERN CONTROL REMOTE CONTROL Choose the transmission interface USB GPIB ETHERNET USB interface is a virtual COM port baud rate is 57600bps Data bit 8 Parity check none Stop bit 41 Please download USB Driver on http www motech com tw f entering the Remote mode screen will present RMT symbol as shown in the following picture V 36 000 V I 15 000 OFF 0 000 V 0 000A on 15 GPIB ADDRESS Set up
54. sted below 1 The appearance of the products is without scratch or other damages 2 Standard parts as shown in Attachment 1 Parts List of Chapter 9 If there is any situation as stipulated above found please inform us as soon as possible and we would correct it for you immediately 3 2 Operation Instructions This PRODUCT is a precise instrument In order to prevent it from improper operation causing the damage be sure to read and comprehend this User Manual And in order to maintain accuracy send it to perform factory s calibration once per year 3 3 Ambient Environment l Do not locate or operate this product in an environment with dust vibration or corrosive gas and do not explore this product directly to the sunlight Operate it in an environment with temperature 0 40 C amp relative humidity 2090 8090 Pause the operation when ambient temperature is over 40 undo the operation only after the ambient temperatures drops to the acceptable temperature range Operating temperature over the above range would damage the host part This product is equipped with one blow out type cooling fan on the back board and three suck in cooling fans on inner side of front board Note on good ventilation near the cooling fans and keep the boards with a space above 10cm away from wall To maintain a good accuracy operation do not block the ventilation holes spaces Although this product is designed to well block the noise from the A
55. t Rate DS3640 DS6024 Output Voltage 0 36V 0 60V Output Current 0 40A 0 24A Output Power 1440W 1440W Line Regulation Voltage lt 4mV lt 6mV Current lt 4mA lt 4mA Load Regulation Voltage lt 8mV lt 8mV Current lt 8mA lt 7mA Ripple and Noise 20Hz 20MHz Normal Mode Voltage 4 Normal Mode Current lt 90 lt 70 Resolution Programming 1mV 1mA 1 5mV 1mA Readback 1mV 1mA 1 5mV 1mA Programming Accuracy output offset Voltage 0 05 10mV 0 05 15mV Current 0 05 10mA 0 05 8mA Readback Accuracy output offset Voltage 0 05 10mV 0 05 15mV Current 0 05 10mA 0 05 8mA Transient Response Time 1085 1085 Efficiency 8046 80 OVP Adjustment Range 2 38V 3 64 OVP Accuracy 200mV 300mV Command Response Time 50mS 50mS Power Factor Ta Full load Full load Remote Sense Compensation 2 2 Rising Time at Full Load lt 15mS lt 20 5 Rising Time at No Load lt 15mS lt 20msS Falling Time at Full Load lt 15mS lt 20 5 Falling Time at No Load 1000mS 1000mS Standard Interface USB Power Cord Terminal Block for Rapid Plug Connector Optional Interface LAN GPIB Optional Accessory RS 485 Cable General AC Line Rated Input Voltage 1 15 41096 Tolerance Variation in Voltage 10 power de rating mode when voltage under 95 Vac Rated Frequency 47Hz 63Hz Maximum Rated Input Power 1700VA Temperature Ratings
56. that can apparently improve the power supplying capability Under the 4 set parallel mode of DS3640 units the combines unit can offer 36V 160A power output And under the 4 set cascading mode of DS10014 units the combined unit can offer 400V 14 4A output Note that you cannot perform both parallel and series mode at the same time Series connection will be off once Series Cascade Paralled connection function 1 turned on 5 8 1 Parallel Connection Setting While connecting four sets of DS3640 DS6024 DS8018 DS 10014 in parallel the wiring is shown in the following figure Power Line Pdwer Line rx Y EF RJ 45 LOAD RJ 45 RJ 45 df del After wiring is complete select one DS3640 DS6024 DS8018 DS10014 as the Master and the other three are Slave A B and C After the DS3640 DS6024 DS8018 DS10014 is set to the Master it will start searching for all Slaves that are connected to the Master Therefore in order to reach the control status user 36 must set up the Slaves before the Master 1 set To set a D 3640 DS6024 DS8018 DS10014 to the slave mode press Menu AD tid 4 in the main page to enter the series parallel setting option And then select the parallel mode by turning the knob SCPI command is PS MODE PARALLEL and then press to confirm It will continue to the next line fo
57. the connector drop calibration function EXT FULL VOLT External tuning setting with full scale voltage 10V 5V 3 PROTECTION SETTING PROTECTION Press 3 key in the first page of Menu Setting to enter PROTECTION page z OFF SET 38 000 V z OFF SET 42 000 A z OFF SET 1440 000 W Y OVP turn on off the SET set up the overvoltage overvoltage protection protecting point OCP turn on off the SET set up the overcurrent overcurrent protection protecting point OPP turn on off the overpower SET set up the overpower protection protecting point CV TO CC OFF CC TO CV OFF CV TO CC Enable disable the protection of the conver sion from CV to CC mode CC TO CV Enable disable the protection of the conver sion from CC to CV mode 19 4 SERIES PARALLEL SETTING Pres 4 in the second page of Menu Setting to enter SERIES PARALLEL page SELECT MODE OFF MASTER SLAVE MASTER SELECT MODE Choose the Series Parallel mode MASTER SLAVE Refer to Sec 5 8 Series Parallel Setting for the detailed setting procedure of MASTER SLAVE mode 5 INFORMATION Press 5 in the second page of Menu Setting to enter INFORMATION page MOTECH INC DS 3640 PROGRAMMABLE DC SOURCE VERSION 1 00 6 SPECIAL TEST FUNC Press 6 in the second page of Menu Setting to enter SPECIAL TEST FUNCTION page 1 CURRENT COUNTER TEST 2 PROGRAM MODE 3 MEASURE AVERAGE 61 CURRENT COUNTER TEST Press C1 to enter
58. the output order sent by Master but not receives the synchronous signal then it will present the following error messages The error message SYNC ON is shown when the slave is not receiving the synchronous output ON signal The error message SYNC OFF is shown when the slave is not receiving the synchronous output OFF signal MODE PARALLEL ERR SYNC ON SLAVE A MODE PARALLEL ERR SYNC OFF SLAVE A 5 9 External Tuning Setting DS3640 DS6024 DS8018 DS10014 provides the capability to control the output voltages by an external voltage resistance The range of the external voltage is in 0 10 Vdc or 0 5 Vdc and the range of the external variable resistance is between 0 5 Kohm to control the output voltage 0 36V for 053640 and 0 60V for DS6024 and 0 80V for DS8018 and 0 100V for 0510014 and the output current 0 40A for 053640 and 0 24 for 256024 and 0 18A for 058018 and 0 14 4A for DS10014 The wirings of the external control is shown in the following figure 42 External Voltage Control External Resistance Control 0 5 0 5 0 10 0 10V 7 EXT CTL EXT CTL The external voltage control or external resistance control can be set in the system setting page Since the external control circuit uses 12bit D A converter for 0 36V for DS3640 or 0 60V for DS6024 or 0 80V for 058018 or 0 100V for 0510014
59. the power off unit and burn it out B While under parallel connection mode the output voltage of each DS3640 DS6024 DS8018 DS10014 shall be set to equal If the setting value of each unit is not the same the higher output voltage will feed back to the smaller unit and destroy its inner parts C When the AC input voltage is lower than the full load voltage which is 100 Vac DS3640 DS6024 DS8018 DS10014 will activate an inner overtemperature protector and cut off the output in response To ensure that the entire test process can be complete smoothly confirm that the input AC voltage is within the specified range 11 Panel Demonstration DS3640 DS6024 DS8018 DS10014 Panel DL SSS epe 0 000 0 006 9 9 9 d Se 4 6 D amp 2 4 1 1 1 2 3 4 5 6 Front Panel Power switch Please consult the Cautions before use before turnning on power switch Display 192 32 Graphic LCD Module Current setting Iset Press to set up the current limit Voltage setting Vset Press to set up the output voltage Dot Local This button is applied as a decimal point Or push this button after entering REMOTE online state to resume into LOCAL mode unit operation mode Or press this button to release after entering LOCK page ESC CLR Esc Press this button to clean up nu
60. time 2 5 6 and current resolution as entirely meeting customer s demanding OVP over voltage protection OCP over current protection and OPP over power protection and key lock functions The over voltage protection OVP over current protection OCP and over power protection OPP features limit the maximum output current and voltage to avoid damages to the test objectives when doing experimentations in a laboratory The key lock feature disables all keys except the CLR key It prevents DS3640 DS6024 DS8018 DS 10014 from the accidentally wrong setting and then avoids the damages to the target Series amp parallel connection mode The series parallel connection mode of two or more units maximum to 4 units largely increases product s power supplying capability Under the parallel connection mode of four DS3640 the maximum output is 36V 160A and under the cascaded connection mode of four DS10014 the maximum output is 400V 14 4A Power supplied under these two modes reach up to 5760 watts output that can meet most electric experiments demands 7 Multi units series connection mode Computer can connect to system through USB interface RS485 interface can be used to connect 31 units at most It can save GPIB interface card and GPIB wiring expense and simplify the connection control 2 Specification DS3640 DS6024 Specifications Outpu
61. ting Range of Transformation Value Lo 5V 4 5 5 5V MIDL 33V 29 7 36 3 V MIDH 66V 59 72V Hi 95V 85 104 V ENTER 3 2 at Hi function the calibration value will be stored in FLASH and returned to calibration main page Items CALIB OVP Lo CALIB OVP Hi in the calibration main page to enter OVP calibration gt Start ENTER function CALIB OVP Lo CALIB OVP Hi 25 to access OVP calibration procedure at Low gt Calibrating C Jump to Hi function when completed CALIB OVP Lo gt OK CALIB OVP Hi gt Start D Press ENTER to access OVP calibration procedure at High function CALIB OVP Lo OK CALIB OVP Hi gt Calibrating E After completed calibration value will be stored in FLASH and returned to calibration main page Items 2 3 F If calibration is not completed in 10 seconds after starting please inspect OVP circuit 8 2 3 CURRENT CALIBRATION A Connect two output terminals of DS3640 to two ends of current shunt and connect DVM to the sensor of current shunt to measure DC voltage as shown Figure 2 Press 3 to enter current calibration page 26 Current Calibration Sense 4 Sense DC Voltage measurement Figure 2 B First input transformation parameter resistance of current shunt m 2 of current measurement fixture According to voltage value shown o
62. tn and NL And there shall be a space between the command and the parameter For example to set the GPIB address of 10 to a DS3640 DS6024 DS8018 DS 10014 The command line is as follows ADDR 10 lt Rtn gt lt NL gt Note The lt NL gt and lt Rtn gt are not presented in the following examples and command descriptions However users shall add them to the end of each command when doing the coding 58 7 3 The Error Event List SCPI interface can offer an error event list that contains up to 10 errors events We can read the errors events through the error command in a first in first out manner Once an error event is read the read process will clean it from the memory To clean all errors events from the memory the CLS command can be used Error Description 000 No error 001 Command error 002 Execution error 003 Query error 004 Input Range error 005 Parallel Series function Error mode 006 Parallel Series function Multi Master 007 Parallel Series function No Slave found 008 Parallel Series function Communication with Slave A error 009 Parallel Series function Communication with Slave B error 010 Parallel Series function Communication with Slave C error 011 Parallel Series function Sync signal error when output on 012 Parallel Series function Sync signal error when output off 59 7 4 Compatible Protocol of MOTECH LPS amp PPS Product Family Command Descriptio
63. urning the knob to select SLAVE A SCPI command is 5 SLAVEA for the DS3640 DS6024 DS8018 DS10014 and press to confirm the setting Using the same procedure to set up SLAVE B and C for another two DS3640 DS6024 DS8018 DS10014 SELECT MODE SERIES MASTER SLAVE SLAVE A To set the master unit press and 4 in the main page to gt enter the series parallel setting option And then use the knob to select the 39 series mode SCPI command is PS MODE SERIES and press to confirm It will continue to the next line for the MASTER SLAVE selection Use knob again to select MASTER SCPI command is PS TYPE MASTER and then press to confirm After it is confirmed as the Master the DS3640 DS6024 DS8018 DS10014 will start searching for all Slaves as graphically demonstrated below SELECT MODE SERIES MASTER SLAVE MASTER CHECKING FOR SLAVE If wiring is correct screen will appear SELECT MODE SERIES MASTER SLAVE MASTER FOUND SLAVE ABC After receiving the control order from the Master all Slaves will be locked on the SLAVE page that all keys except are disabled so that slaves can not be operated by press keys except controlled by the Master as graphically demonstrated below MODE SERIES ERR NONE SLAVE A Once the Master and Slave settings are done users can then operate the Master only to set up the output voltage and current To release from this operation mode series op
64. utput terminal to DVM as shown in Figure 1 Then select DVM at DC volatage measurement and turn on DS3640 Once entering main page press MENU and select 8 Calibration and key in password 13579 to enter the following calibration main page Items 1 2 3 1 VOLTAGE CALIBRATION 2 0VP CALIBRATION 3 CURRENT CALIBRATION 23 4 CONNECTOR DROP CALIBRATION a 5 EXT VOLT CALIBRATION 6 EXT CURR CALIBRATION B Press 1 to access Voltage calibration procedure CALIB VOLT Lo 1 8640 V CALIB VOLT MIDL 12 5540 V CALIB VOLT MIDH 25 1140 V CALIB VOLT Hi 35 7800 V C According to voltage value displayed on DVM please fill in with the corresponding function and press ENTER If any DVM read back value at each voltage function does not fit with the following table please inspect the hardware DS 3640 Function Setting Range of Transformation Value Lo 1 8V 1 5 2 0 V MIDL 12V 10 13 24V 21 25 Hi 32 4V 31 34 V DS 6024 Function Setting Range of Transformation Value Lo 3V 2 4 3 6 MIDL 20V 18 22V MIDH 40V 36 44 V Hi 57N 53 61 24 D When pressing 129 8 2 2 OVP CALIBRATION A Press page B Press DS 8018 Function Setting Range of Transformation Value Lo 4V 3 6 4 4 V MIDL 26V 23 5 28 5 V MIDH 53V 48 58 V Hi 76 69 83 DS 10014 Function Set