Home

PicoScope 9000 Series User's Guide

image

Contents

1. AnyPass one or more active measurements is good ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 33 Format of Stored Files Header Limit FileFormat Type Selector Arguments Binary Verbose YOnly Action set file format File Name Header Limit FileName Type Data Argument text string Forms command query command with query Action define file name for saving the specified signals to disk Parameter Activity Headers Limitl Activ Limit2 Activ Limit3 Activ Limit4 Activ Type On off Action Enable disable the Limit Test for relevant parameter Parameter Limit Mode Headers Limitl1 Mode Limit2 Mode Limit3 Mode Limit4 Mode Type Selector Arguments Center Limit Action set mode of limits for the relevant parameter 2010 Pico Technology All rights reserved ps9000pg en 34 Full list of commands Upper and Lower Limits of Parameters Headers Limitl UpLimit Limitl LowLimit Limit2 UpLimit Limit2 LowLimit Limit3 UpLimit Limit3 LowLimit Limit4 UpLimit Limit4 LowLimit Type Float Arguments absolute value of limit Action sets the limit s value It is used only for Limit mode of the parameter s limit Parameter Center Mode Headers Limitl CenterMode Limit2 CenterMode Limit3 CenterMode Limit4 CenterMode Type S
2. set the kind of trend for the specified function Used for Trend operator only 4 14 FFT commands Enable Spectra Headers Type Action Spectrl Display Spectr2 Display On off enable disable the calculation and display of the relevant spectrum Spectrum Source Headers Type Arguments Action Window Headers Spectrl Sourcel Spectr2 Sourcel Selector Chl Ch1B2 Ch2 Ch2B2 Fl F2 F3 F4 Ml M2 M3 M4 set the source of the specified spectrum Spectrl Window Spectr2 Window Spectr3 Window Spectr4 Window 2010 Pico Technology All rights reserved ps9000pg en 40 Full list of commands Type Selector Arguments Rectang Hamming Hanning Flattop BlackHarr KaiserBess Action set the window for specified spectrum 4 15 Histogram commands 4 15 1 Setting Histogram Parameters Histogram Axis Header Type Arguments Action Hist Axis Selector Off Vert Horiz set axis of histogram Histogram Source Header Type Arguments Action Hist Source Selector Chi Ch1B2 Ch2 Ch2B2 Fl F2 F3 M3 M4 Sl S2 select specified signal as source of histogram Histogram Visibility Header Type Action Hist Visible On off F4 Ml M2 set visibility of histogram Acquisition of the histogram proceeds independently of this command Histogram Finish Condition Header Type Arguments Action His
3. Header Meas lt src gt XDualPar Type On off group Items DellR1R DellR1F Del1F1R DellF1F Del1RnR DellRnF DellFnR DellFnF PhaseDeg PhaseRad PhasePerc Action define the set of the X axis inter signal measurements for the specified signals List of Y Inter Signal Measurements Header Meas lt src gt YDualPar Type On off group Items Gain DBGain Action SG the set of the Y axis inter signal measurements for the specified signals Delete all Measures for all Sources Header Meas ClearAll Type Execution Action Clear list of all measurements for all signals Measurements of Spectrum Signals Limits Definition Method for Spectrum Header Meas lt src gt FFTMethod Type Selector Arguments Harmonic Peak Action sets the method of the limits definition for the specified signal It used for spectrum signals only Left and Right Spectrums Margin Headers Meas lt src gt FFTLeft 2010 Pico Technology All rights reserved ps9000pg en 30 Full list of commands Meas lt src gt FFTRight Type Float Arguments absolute frequency value Action sets the position of margin for the specified spectrum signals It used for searching for peak 1 of the spectrum for the Harmonic method Peak Level of Spectrum Header Meas lt src gt PeakLevel Type Float Arguments 100 to 80 dBV Action sets the level for the specified spectrum signals It used for searching a peak of the spectrum
4. Headers Type Argument Action Hist VertDBScale Hist HorDBScale Float 6 to 60 dB div set the scale of the histograms Used for Manual mode and Logarith type of scale only Logarithmic Offset of Vertical or Horizontal Histogram Headers Type Argument Action Hist VertDBOffs Hist HorDBOffs Float 60 to 0 dB set the offset of the histograms Used for Manual mode and Logarith type of scale only 2010 Pico Technology All rights reserved ps9000pg en A4 Full list of commands 4 15 2 Getting Results of Histogram Get Histogram Data Headers Type Argument Forms Action Hist Data Vert Hist Data Hor Data none query only return text string with values of the histogram comma separated Get Histogram Measurement Headers Hist Res lt Param gt Parameter lt Paramo gt Type Argument Forms Action InBox Number of Hints in Box Wfm Number of Waveforms Peak Peak Value of Histogram PP Difference between highest and lowest Values of Signal Median Centre between highest and lowest Values of Signal Mean Average of Distribution of Histogram StdDev Standard Deviation of Histogram MeanlS number of hints in Mean StdDev Region Mean2S number of hints in Mean 2StdDev Region Mean3S number of hints in Mean 3StdDev Region Min Min Value of Signal Max
5. e Here are some invalid concatenations DISPlay STYlE DOTS ACQuire NAVG 10 no colon before ACQuire DISPlay STYlE DOTS FORMAT YT extra colon before FORMAT use DISPlay STY1E DOTS FORMAT YT instead Acq Chl Mode Sample Chl RecLen 1024 levels of the mnemonics are different either remove the second use of Ch1 or place Acg in front of Ch1 2010 Pico Technology All rights reserved ps9000pg en Command Classification 3 1 3 2 3 3 Command Classification Most commands can be related to one of a few types For example the execution type commands tell the instrument to perform a specific action the selector type commands modify a specific instrument setting to the one of few fixed values and so on All commands of a given type have similar behavior Execution type commands The execution type commands tell the instrument to perform a specific action For example Run ClrDispl There are no arguments for these commands All execution type commands have a set form only and not a query form On off type commands The on off type commands tell the instrument turn on or turn off a specific function For example Header Off Chl Display 0 There are four fixed arguments possible in these commands On Off 0 1 Arguments On and 1 are equivalent and turn on the corresponding function Arguments Off and O are also equivalent and turn off the specific func
6. Meas lt src gt MidThresh Meas lt src gt LowThresh Type Float Arguments absolute voltage value for Volt threshold units only 4 to 4 for Division threshold units only Action sets the threshold position for the specified signals Percentage of Upper Middle or Lower Threshold Headers Meas lt src gt UpThPerc Meas lt src gt MidThPerc Meas lt src gt LowThPerc Type Integer Arguments 80 to 200 Action sets the threshold percentage for the specified signals It used for Percent threshold units only Argument O correspond to the Base of the signals and argument 100 corresponds to the Top of the signals Margins Definition Mode Header Meas lt src gt MargMode Type Selector Arguments Slope Marker Action sets margins definition mode Slope of Left or Right Margins Headers Meas lt src gt LeftSlope Meas lt src gt RightSlope Type Integer Arguments 0 to 127 Action sets the margin for the specified signals on the specified slope It used for slope margins definition mode only Argument O means the first rise value 1 is first fall 2 second rise 3 second fall and so on 2010 Pico Technology All rights reserved ps9000pg en 28 Full list of commands Thresholds of Left and Right Margin Slopes Headers Type Arguments Action Meas lt src gt LeftTresh Meas lt src gt RightTresh Selector Upper Middle Lower sets the thresholds for defin
7. incremented For example basename_00001 wfm basename_00002 wfm basename_00003 wfm and so on Format of stored files Header Save Disk FileFormat Type Selector Arguments Binary Verbose YOnly 2010 Pico Technology All rights reserved ps9000pg en 22 Full list of commands Action set file format Save to Disk Header Save Disk Save Type Execution Action save selected source to previously specified file Select Memory for loading signal from disk Header Save Disk ToMemo Type Selector Arguments M1 M2 M3 M4 Action selects which of the available memory locations the instrument loads the saved file into Load Header Save Disk Load Type Execution Action load the previously specified disk file into the previously specified Memory Zone 4 9 3 Work with Setups Recall Factory Setup Header Save Setup RecFact Type Execution Action returns instrument to manufacturer s default setting Recall Power Off Setup Header Save Setup RecLast Type Execution Action returns instrument to last setting before power supply was last switched off Save Setup as Default Header Save Setup SvAsDefault Type Execution Action stores present front panel setup as default setup ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 23 Name of Custom Setup File Header Save Setup FileName Type Data Argument text string Forms command
8. Max Value of Signal Max Max Difference between two maxima of histogram Data none query only return text string with value of the specified parameters ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 45 4 16 4 16 1 Mask Test commands Common Mask Commands Mask Erasing Header Mask EraseMask Type Execution Action Clear the current mask from the display Signal for Mask Testing Header Mask CompareWith Type Selector Arguments Chl Ch2 Action select the signal for mask testing Actuate Mask Testing Header Mask Test Type On off Action enable disable mask test execution User Masks File Name Header Mask MaskFile Type Data Argument text string Forms command query command with query Action define the file name for next loading of the user mask from the disk Load User Mask Header Mask LoadUser Type Execution Action load the previously specified user mask Mask Test Finish Condition Header Mask RunUntil Type Selector Arguments StopBtn FailedwWfms FailedSmpls Wfms Samples Action set condition of Mask Test Termination 2010 Pico Technology All rights reserved ps9000pg en 46 Full list of commands Number of Failed Waveforms Header Mask FailWfms Type Integer Argument 1 to 1000000 Action set the number of the failed waveforms for the FailedWfms finish condition Number of Failed Samples
9. DutCycDistP DutCycDistS EyeWidth EyeWidthP FallTime Freg JitterPP JitterRMS Period RiseTime Action define the set of the X axis measurements for NRZ signals List of Y axis NRZ Measurements Header Eye YNRZParam Type On off group 2010 Pico Technology All rights reserved ps9000pg en 52 Full list of commands 4 17 2 Items AcRMS AvgPower AvgPWdBm CrossPerc CrossLevel ExtRatioDB ExtRatioP ExtRatio EyeAmpl EyeHeight EyeHeightDB Max Mean Mid Min NegOver PPNoiseOne PPNoiseZero RMSNoiseOne RMSNoiseZero OneLevel PeakPeak PosOver RMS SNRaio SNRaioDB ZeroLevel Action define the set of Y axis measures for NRZ signals List of X axis RZ Measurements Header Eye XRZParam Type On off group Items Area BitRate BitTime CycleArea EyeWidth EyeWidthP FallTime JittPpFall JittPpRise JittRMSFall JittRMSRise NegCross PosCross PosDutyCyc PulseSymm PulseWidth RiseTime Action define the set of X axis measurements for RZ signals List of Y axis RZ Measurements Header Eye YRZParam Type On off group Items AcRMS AvgPower AvgPWdBm Contrast ContrastBb ContrastP ExtRatioDB ExtRatioP ExtRatio EyeAmpl EyeHeight EyeHeightDB EyeOpenFact Max Mean Mid Min PPNoiseOne PPNoiseZero RmsNoiseOne RMSNoiseZero OneLevel PeakPeak RMS SignToNoise ZeroLevel Action define the set of the Y axis measures for R
10. LabVIEW Base Development System 8 2 1 2010 Pico Technology All rights reserved ps9000pg en 60 Programming Examples 5 3 Visual Basic NET The project is located in the PicoScope9000VBdotNETClient subdirectory of the PicoScope9000 SDK The subdirectory is a standard VB NET project directory with the project file PicoScope9000VBdotNETClient vbproj solution file PicoScope9000VBdotNETClient sln MainForm files and three subdirectories bin My Project obj Assembly obj Interop PicoScope9000 d11 which is referenced in the source code as PicoScope9000 COMRC is a bridge between NET platform and Windows Component Object Model COM It has been created with Visual Basic IDE by executing the command Project Add Reference COM PicoScope9000 File MainForm vb contains the whole source code of the example Other files were created automatically either by IDE itself or by the visual form editor This example has been created and tested with Microsoft Visual Basic 2008 Express Edition ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 61 Index A Acquisition commands Acquisition Mode of Channel 16 Ag ion when Number of Waveforms reached Channel Averaging 16 Channel Envelopes 16 Channel Record Length 16 Number of Waveforms 16 Sampling Mode 16 Termination of Acquisition 16 Type of signal 16 All off mode 6 C Case insensitivity 4 Cha
11. Save Recall commands 4 9 1 Work with Memo Zones M1 M2 M3 M4 Memory Display Header Save Memo On Type On off group Items M1 M2 M3 M4 Action control display of memory zones Source for storing into Memory Header Save Memo Source Type Selector Arguments Chi Ch1B2 Ch2 Ch2B2 F1 F2 F3 F4 M1 M2 M3 M4 S1 S2 Action define signal as source for storing into memory zone Select Memory for Saving Header Type Save Memo ToMemo Selector ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 21 Arguments M1 M2 M3 M4 Action define memory zone for saving Save into Memory Header Save Memo Save Type Execution Action store selected source into selected memory Work with Disk Source for saving to file Header Save Memo Source Type Selector Arguments Chil Ch1B2 Ch2 Ch2B2 F1 F2 F3 F4 M1 M2 M3 M4 S1 S2 Action define signal as source for saving to file File Name Header Save Disk FileName Type Data Argument text string Forms command query command with query Action define file name for saving specified signal to disk File Name Mode Header Save Disk NameMode Type Selector Arguments Manual Auto Action set file name mode In Auto mode the file name consists of a base name followed by a underscore _ and a five digit number Each time you save a waveform the number in the file name is automatically
12. Type 23 Motion of Markers 23 X position of Marker 23 Y position of Marker 23 Mathematics commands Constant Value 37 Enable Mathematical Function 37 Function Operator 37 Operand 1 37 Operand 2 37 Smoothing Parameter 37 Trend Measurement 37 Measure commands 24 Measurement of Spectrum Signals commands for Left and Right Spectrum Peaks 29 Left and Right Spectrums Margin 29 Limits Definition Method for Spectrum 29 List of Spectrum Frequency Measurements 29 List of Spectrum Magnitude Measurements 29 Peak Level of Spectrum 29 Measurements of Time Domain Signals commands Base Value for Marker Method 24 Delete all Measures for all Sources 24 Execute Single Measurement 24 List of X Inter Signal Measurements 24 List of X Measurements 24 List of Y Inter Signal Measurements 24 List of YMeasurements 24 Margins Definition Mode 24 Measurement Mode 24 Measurement Source 24 Measurement Type 24 Percentage of Upper Middle or Lower Threshold 24 Position of Left or Right Margin 24 Position of Upper Middle or Lower Threshold Second Source for Inter Signal Measurements Slope of Left or Right Margins 24 Statistic Measurement Mode 24 Threshold Definition Method 24 Threshold Units 24 Thresholds of Left and Right Margin Slopes Top Value for Marker Method 24 Top Base Definition Method 24 Viewing of Define Parameters 24 Weight Value 24 Windows Value 24 Messages 2 Mnemonics 4 ps9000pg en 2010 Pico Tec
13. Waveform Source 55 White space 4 Work with Disk commands File Name 21 File Name Mode 21 Format of stored files 21 Load 21 Saveto Disk 21 Select Memory for loading signal from disk 21 Source for saving to file 21 Work with Memo Zones M1 M2 M3 M4 commands Memory Display 20 Save into Memory 20 Select Memory for Saving 20 Source for storing into Memory 20 Work with Setups commands Name of Custom Setup File 22 Recall Custom Setup 22 Recall Factory Setup 22 Recall Power Off Setup 22 Save Custom Setup 22 Save Setup as Default 22 ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 65 2010 Pico Technology All rights reserved ps9000pg en Pico Technology James House Colmworth Business Park ST NEOTS Cambridgeshire PE19 8YP United Kingdom Tel 44 0 1480 396 395 Fax 44 0 1480 396 296 www picotech com ps9000pg en 4 8 12 10 2010 Pico Technology All rights reserved
14. en 26 Full list of commands Action sets Top and Base vertical reference thresholds for amplitude measurements of specified signals Top Value for Marker Method Header Meas lt src gt Top Type Integer Argument 2 to 1023 Action sets Top vertical reference threshold for specified signals Argument O corresponds to the bottom of the screen and argument 1023 corresponds to the top of the screen independently of the real screen s height Base Value for Marker Method Header Meas lt src gt Base Type Integer Argument 1 to 1022 Action sets Base vertical reference threshold for specified signals Argument O corresponds to the bottom of the screen and argument 1023 corresponds to the top of the screen independent of the real screen s height Threshold Definition Method Header Meas lt src gt Thresh Type Selector Arguments 10 90 20 80 Custom Action sets lower middle and upper thresholds for measurements of the specified signals May be set to the fixed values 10 50 90 20 50 80 or custom values Threshold Units Header Meas lt src gt Unit Type Selector Arguments Percent Volt Division Action sets units of thresholds for specified signals It used for custom threshold definition method only Position of Upper Middle or Lower Threshold Headers Meas lt src gt UpThresh ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 27
15. has only the query form For example Meas Chl XParam 2010 Pico Technology All rights reserved ps9000pg en Command Classification 3 4 3 5 3 6 The answer may be ClearA11 if all items are turned off or one or more items separated by comma For example the answer Freg Period means two turned on items Selector type commands The selector type commands modify a specific instrument setting to one of a few fixed values For example Trig Source has these possible arguments Direct ExtHF IntClock and Trig Mode has these possible arguments Free Trig Between 2 and 32 custom arguments are available for these commands The full set of arguments is specified for each command in the list of commands 10 The selector type commands have a query form It is possible to use the query form with an argument for example Trig Source Direct This command sets the Direct input as the trigger source and returns DIRECT Integer type commands The integer type commands modify specific integer value functions For example the command Acq Chl RecLen 1024 sets the length of Channel 1 signals to 1024 points The valid range and increment of each value is different and is described in the list of commands 10 The integer type commands have a query form It is possible to use the query form with an argument for example Acg Chl RecLen 24 returns 32 since 32 is the m
16. the following arguments On Off 0 1 For example the next command turns on a frequency measurement for Channel 1 Meas Ch1 XParam Freg 1 Single item mode has a query form similar to the On Off commands So the query Meas Ch1 XParam Period 1 or Meas Ch1 XParam Freq returns either ON or OFF Group on mode Multi item mode is used to simultaneously turn on a custom group of items In this case the Include mnemonic is added to the end of the command This is then followed by a space and a few items separated by commas For example the next command turns on rise time and fall time measurements for Channel 1 Meas Chl XParam Include Rise Fall Group off mode Multi item off mode is used to simultaneously turn off a custom group of items In this case the Exclude mnemonic is added to the end of the command This is then followed by a space and a few items separated by commas For example the next command turns off frequency and period measurements for Channel 1 Meas Ch1l XParam Exclude Freg Period All off mode All off mode is used for simultaneously turning off all items In this case the ClearAl1 mnemonic is added to the end of the command For example the next command turns off all measurements for Channel 1 Meas Ch1 xXParam ClearAll Group on Group off and All off modes do not have a query form Group query mode Group query mode is used find out which items are currently turned on This mode
17. 00pg en 36 Full list of commands Parameter Delta Percentage for User Defined mode Headers Limitl UserPerc Limit2 UserPerc Limit3 UserPerc Limit4 UserPerc Type Float Arguments 0 01 to 90 standard deviations of the parameter Action sets the delta value It used for UserPerc mode of delta of the parameter only Failure When Headers Limitl FailWh Limit2 FailWh Limit3 FailWh Limit4 FailWh Type Selector Arguments Outside Inside en en en en Action set the mode of the quality control for the according parameter If Measurement Undefined Headers Limitl NotFound Limit2 NotFound Limit3 NotFound Limit4 NotFound Type Selector Arguments Ignore Fail Pass Action set limit status when measurement is undefined ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 37 4 13 Mathematics commands Enable Mathematical Function Headers Type Action F1l Display F2 Display F3 Display F4 Display On off enable disable the calculation and display of the relevant functions Function Operator Headers Type Arguments Action Operand 1 Headers Type Arguments Action Operand 2 Fi Operat F2 Operat F3 Operat F4 Operat Selector Add Sub Mult Div Invert Abs Exp e Exp 10 Log e Log 10 Dif al I
18. Header Mask FailSmpls Type Integer Argument 1 to 1000000 Action set the number of failed samples for the FailedSmp1s finish condition Number of Waveforms Header Mask NWfms Type Integer Argument 1 to 1000000 Action set number of waveforms for the Wfms finish condition Number of Samples Header Mask NSamples Type Integer Argument 1 to 1000000 Action set number of samples for the Samples finish condition Mask Test Actions Header Mask Action Type On off group Items Beep Save Action Save every failed signal is stored to disk Beep the beep signal will sound for every failed signal Format of Stored Files Header Mask FileFormat Type Selector ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 47 4 16 2 Arguments Binary Verbose YOnly Action set file format Used when Save action is on Stored File Name Header Mask FileName Type Data Argument text string Forms command query command with query Action define file name for storing failed signals on Disk Used when Save action is on Standard Mask Commands Alignment of Signal with Standard Mask Header StdMask Align Type On off Action enable disable alignment of the tested signal with the standard mask parameters Enable Margins Header StdMask MarginsOn Type On off Action enable disable the margin control of eye type masks Margins Value Header StdMask MarginsV
19. ICO E PicoScope 9000 Series PC Sampling Oscilloscopes Programmer s Guide ps9000pg en 4 2010 Pico Technology All rights reserved 9 Technology PicoScope 9000 Series User s Guide l Contents T Picoscope 9000 API E 1 1 PicoScope9000 COM Server sessersessereessrsessereersereeseessesensersersereereeseeseesersereereereeseereeseeseneereerseneeseeseeseeseeseneenee 1 2 ExecCommand Method WE 1 Helef idelOE NEE 1 EE 2 1 Command and Query Structure EEN 2 1Overview O rnssesasasaseasasnecacacsccucncnecncncnccncncnecnsacascuenenesnsncaesnenenesassessenenenesnsassesasnenesusaesesnsassesassesesneneneensasees 2 2 Messages EE 2 3 ef TIET E DN 3 4 Queries 5 Headers 2 Command Entry 1 Rules 2 Concatenating ee H 3 Command Classification ggecticscacciuctactedweisecesesinsieiencilga acest aliesoandsic ineei iaa i ona Ea aed oair 6 1 Execution type Commands ENEE 6 RTE ses aa jon Ra natalia Ra 6 3 On off group type commands sesessesesersesesereesererseseserseseressesensesesereeseressesesensesesenseneseeseneneeseseneeseneneeneseneenenee 6 4 Selector type Commande NENNEN 8 5 Integer type Commande ENEE 8 6 Float type commands sssessessesssessesssessesssesesssessssssesessssesssssesesssessnessesenesseseneseesenssessesesessesessesenssessenssessensaes 8 7 Data type commande ENEE 9 Full list GEHOL 10 T Header EH STE 10 2GUlI comimands 205505 ss sesh ce cee cad case ihcceeds OE owns tebessdctetsvckebecectbasvedeageveccouse
20. Z signals Getting Eye Measurement Results Get List of Measured Parameters Header Eye Res List Type Data Argument none Forms query only Action return list of active eye measurements with ordinal index Get Current Value of Parameter Header Eye Res lt N gt Parameter lt N gt index of parameter in the list Type Data ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 53 4 18 Argument none Forms query only Action return the result of the specified measured parameter Get Statistic Value of Parameter Header Eye Res lt N gt lt Val gt Parameter lt N gt index of the parameter in the list Parameter lt Val gt Wm Min Max Mean StdDev Type Data Arguments none Forms command with query only Action return the specified statistical parameter of the measured parameter Utilities commands Start Autocalibration of Channels Header Flash Calibr AutocalCh Type Execution Action Start self calibration of channels Start Autocalibration of Timebase Header Flash Calibr AutocalTB Type Execution Action Start self calibration of timebase Get the Autocalibration status query Header Flash Calibr AutocalResult Type Integer Action Command is ignored query returns an integer O Autocalibration finished OK 1 Signal must be disconnected from Chi Input Autocalibration of the Channels is aborted 2 Signal must be disconn
21. a query form The query form of the command differs from the set form by a question mark at the end For example the set command ACQuire Ch1 MODe has a query form ACQuire Ch1 MODe Not all commands have both a set and a query form Some commands have set only and some have query only Messages A command message is a command or query name followed by any information the instrument needs to execute the command or query Command messages may contain five element types as defined in the following table Command message elements Symbol lt Header gt This is the basic command name If the header ends with a question mark the command is query The header may begin with a colon character If the command is concatenated with other commands the beginning colon is required lt Mnemonic gt This is a header of the sub function Some command headers have only one mnemonic If a command header has multiple mnemonics a colon character always separates item from each other lt Argument gt This is a quantity quality restriction or limit associated with the header Some commands have no arguments while others have multiple arguments A space separates arguments from the header A comma separates arguments from each other lt Comma gt A single comma is used between arguments of multiple argument commands Optionally there may be white space characters before and after the comma lt Space gt A white space charac
22. ain by executing PicoScope9000 exe with the UnregServer or RegServer switches ExecCommand Method The COMRC object contains only one method ExecCommand The method has one argument a text string with a command or query The method returns e NULL Nothing in Visual Basic if a command without query has been successfully executed e The text string ERROR if the command was invalid e Another text string with query results if the command was a query or a command with query The syntax of the commands and query as well as the full list of commands is described in the following pages COMRC Object To implement the API the server exposes only one object which is called COMRC The object supports automation so it can be used by high level languages like JavaScript HTML pages or VBA Microsoft Word However low level languages like C are also supported The string defining the system wide name of the object and used for object creation is PicoScope9000 COMRC 2010 Pico Technology All rights reserved ps9000pg en Commands Syntax 2 1 2 1 1 Commands Syntax Command and Query Structure Overview PicoScope 9000 commands consist of set commands and query commands usually called commands and queries Commands modify instrument settings or tell the instrument to perform a specific action Queries cause the instrument to return data and information about its status Most commands have both a set form and
23. al Type Float Arguments 100 to 100 Action set the margins value Used when margins is enabled Get List of Standards Header StdMask StdsList Type Data Argument none Forms query only Action return list of mask standards with ordinal index 2010 Pico Technology All rights reserved ps9000pg en 48 Full list of commands Select Standard Header StdMask StdIndex Type Integer Argument O to number of standards 1 Action select the current standard by its ordinal index Get List of Masks Header StdMask MasksList Type Data Argument none Forms query only Action return list of masks with ordinal index from the selected standard Select Standard Mask Header StdMask MaskIndex Type Integer Argument O to number of masks in the current standard 1 Action load the specified mask by its ordinal index 4 16 3 Getting Mask Results Get Integrated Results of Mask Test Headers Mask Res lt Param gt Parameter lt Param gt Allwfm number of waveforms FailWfm number of failed waveforms AllSmpl number of samples FailSmpl number of failed samples Type Data Argument none Forms query only Action return text string with value of the specified parameter Get Number of Samples in Selected Polygons Headers Mask Res Poly lt N gt ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 49 Parameter lt N gt number of t
24. ayed time base used in intensified or delayed timebase Timebase delta delay Header TB DeltaDel Type Float Argument 0 to 10 Action set delta delay of intensified delayed timebase in divisions used in dual delayed timebase 4 6 Trigger commands Trigger Source Header Trig Source Type Selector Arguments Direct ExtHF IntClock ClockRecov Action set trigger source Direct Input Trigger Level Header Trig ExtLevel Type Float Argument 1 to 1 Action set trigger level for direct input volts Trigger Period for Internal Clock Sources Header Trig IntRate Type Float ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 15 Argument 16e 9 to 0 002 Action set period for internal clock trigger source seconds Trigger Mode Header Trig Mode Type Selector Arguments Free Trig Action set Freerun or Triggered mode of the trigger Direct Trigger Slope Header Trig Slope Type Selector Arguments Pos Neg Action set Positive or Negative slope of trigger Holdoff Time Header Trig Holdoff Type Float Argument 5e 6 to 1 Action set the holdoff time seconds Direct Trigger Hysteresis Header Trig Hister Type Selector Arguments Norm HighSens Action set hysteresis for direct trigger Norm or set off HighSens Attenuator Unit for Direct Input Header Trig Atten Unit Type Selector Arguments Off Ratio DB Action
25. commands are listed below organised by type Header commands Header Header Type On Off Action Enable disable headers as part of the query response GUI commands Header Gui Type Selector type command Arguments RemoteLocal RemoteOnly Invisible Action Set the behavior of the GUI when it controls by COM object Version This command can be used with PicoScope SW v 2 3 2 or later System commands Clear Display Header ClrDispl Type Execution Action Clear Display immediately Start Cycle Acquisition Header Run Type Execution Action Run the instrument Start Single Acquisition Stop Acquisition Header StopSingle Type Selector Arguments Stop Single Action Single Start a single acquisition top Immediately stop the acquisition nN Response Stop the instrument is stopped Single the instrument is in the acquisition state ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 11 4 4 Start Autoscaling Header Autoscale Type Selector Arguments Auto SingleVal NRZ RZ Action set the type of signal and start autoscaling of the instrument Response selected type of signals Recall Default Setup Header DefSetup Type Execution Action Restore the instrument to its Default Setup Channels commands Display a Channel Header Chl Display Ch2 Display Type On Off Action turn on or turn off the di
26. ction Displ TraceSel Selector Chl Ch1B2 Ch2 Ch2B2 Fl F2 F3 F4 Ml M2 M3 M4 Sl 82 XY select active trace for AllLocked trace mode Set Display Style Header Type Arguments Action Displ lt src gt Style Selector Dots Vectors VarPersist InfinPers VarGrayScal InfGrayScal VColorGrad IColorGrad set display style for specified trace in PerTrace mode set display style for all traces in AllLocked mode if lt src gt equal to the active trace or do nothing if lt src gt not equal to the active trace Persistence Time seconds for VarPersist Style Header Type Argument Action Displ lt src gt PersistTime Float 0 1 to 20 set persistence time for specified trace in PerTrace mode set persistence time for all traces in AllLocked mode if lt src gt is equal to the active trace or do nothing if lt src gt not equal to the active trace ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 19 Refresh Time seconds for VarGrayScal or VColorGrade Styles Header Displ lt src gt RefreshTime Type Float Argument 1 to 200 Action set refresh time for specified trace in PerTrace mode set refresh time for all traces in AllLocked mode if lt src gt equal to the active trace or do nothing if lt src gt not equal to the active trace Reset Display Style Header Displ ResetAll Type Execution Action Reset D
27. ctor Arguments Volt Watt Ampere Unknown Action set the units of the converter used with the channel ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 13 4 5 Timebase commands Timebase Units Header TB Units Type Selector Arguments Time Bit Action set units of timebase to s div or bit div Timebase mode Header TB Mode Type Selector Arguments A AB B Action set main intensified or delayed timebase Main timebase scale sec div Header TB ScaleA Type Float Argument 10e 12 to 50e 3 Action set scale of the main timebase when time units are used Delayed timebase scale sec div Header TB ScaleB Type Float Argument 10e 12 to 50e 3 Action set scale of delayed timebase when time units are used Main timebase scale bit div Header TB BitScaleA Type Float Argument depends on actual bit rate Action set scale of main timebase when bit units are used Delayed timebase scale bit div Header TB BitScaleB Type Float Argument depends on actual bit rate 2010 Pico Technology All rights reserved ps9000pg en 14 Full list of commands Action set scale of delayed timebase when bit units are used Timebase delay Header TB Delay Type Float Argument 0 to 10 Action set delay of intensified delayed timebase in divisions Dual delayed timebase Header TB DualDel Type On off Action turn on or off the dual del
28. decessaedecssideneedounebecceteerecs vege dee 10 3 System commands oeei iier ia N EE E A ENE E a E E A E E 10 4 Channels commands i lt eissssscsceed dctescescesceccesvesvevecssscecedvesvecadeseedeesvevsececesesceesvevsessesseeddesdessesseteescdedeveesvcsesereedsess 11 5 Tiniebase COmMMandS cee iccessdccsseseedcesesecessiccvesdccndesessdodoesdacucetecessdveciscesncessdeceatiosousdessonsecaussdecessdocbadiecsaesessaodess 13 EE 14 eege 16 8 Display commands E 18 O Save Recall commajds E 20 1 Work with Memo Zones M1 M2 M3 M4 tetas 20 Work with vIG DR EN 3 Work with Setups EE 22 10 Markers commands 11 Measure commands SE 24 1 Measurements of Time Domain Signals 2 Measurements of Spectrum Signals 3 Getting Measurement Results cece cerceeeereeeeteseteeteeseeeeeeeseeeeeeeeeeneeesseenseenseenseesseeeseseeseseeeseeeseeesseesseessresseeeseeeseegs 12 Limit RT E d ET TE 31 13 Mathematics commands ce eg Zeg eege te Sava dese eae eege 37 2010 Pico Technology All rights reserved ps9000pg en ll Contents 14 FET commands escsssessosmssicsasssseriosssca or detec sismo nosso ses so pice na ipa Li sina c uceduced Seb sap SASS na LE sina ESTEE E ISESE EREE rE OSPS ESA 15 Histogram commands s eseseseeseseseesesesersesereesenersesesersesenersesesersenesersenesensenerersenereeseserersesereesenersesesersesesereene 1 Setting Histogram Parameters 2 Getting Results of Histogram e cubes optesseeeysubspontesi
29. ds Measurements of Time Domain Signals Measurement Type Header Meas Display Type Selector Arguments Off Param Statistic Action set measurement type Measurement Source Header Meas DisplSrc Type Selector Arguments Ch1 Ch1B2 Ch2 Ch2B2 F1 F2 F3 F4 M1 M2 M3 M4 Sl S2 Action set source for Measurement Viewing of Define Parameters Header Meas View Type On off ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 25 Action set visibility of define parameters markers for selected sources Measurement Mode Header Meas Mode Type Selector Arguments Permanent Single Action set measurement mode Execute Single Measurement Header Meas SingleMeas Type Execution Action execute single measurement in Single mode Statistic Measurement Mode Header Meas StatMode Type Selector Arguments Permanent Window Weight Action set Statistic Measurement mode Windows Value Header Meas Window Type Integer Argument 8 to 8192 Action set number of recently acquired waveforms for Window mode of Statistic Measurement Weight Value Header Meas Weight Type Integer Argument 8 to 8192 Action set weight variable for Weight mode of Statistic Measurement Top Base Definition Method Header Meas lt src gt Method Type Selector Arguments Hist MinMax Marker 2010 Pico Technology All rights reserved ps9000pg
30. e X axis physical units Get Y axis Unit Header Wfim Preamb YU Type Data Argument none Forms query only Action return the Y axis physical units 4 20 System commands This group of commands is used to control devices in a multi instrument system Count of Instruments in the system query Header Instr List Count Type Integer Argument none Forms query only Action return the number of devices in the system Version This command can be used with PicoScope 9000 SW v 2 4 0 or later ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 57 Get a description of device number N query Header Instr List Unit N when N O count of instruments 1 Type Data Argument none Forms query only Action return the description of the instrument number N in the format N User Name Sys Name Interface IP where N number of devices in the system User Name user name of the Instrument optional Sys Name system name of the Instrument from the list ABSENT DEFAULT DEMO DEFAULT LAN DEFAULT USB DEMO2 DEMOS UNIT1 UNIT2 UNIT3 UNIT4 UNIT5 UNITO UNIT7 UNIT8 Interface type of the interface from the list NONE DEMO LAN USB IP IP address for LAN interface only Version This command can be used with PicoScope 9000 SW v 2 4 0 or later Get the name of the current device query Header Instr Current Get Type Data Argu
31. e used with PicoScope 9000 SW v 2 4 1 or later ps9000pg en O 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 55 4 19 Waveforms commands This group of commands is designed for receiving acquired waveforms from the oscilloscope Waveform Source Header Wim Source Type Selector Arguments Chil Ch1B2 Ch2 Ch2B2 F1 F2 F3 F4 M1 M2 M3 M4 S1 S2 Action set the signal to be received Spectrum Format Header Wfm Complex Type Selector Arguments Mod Ph Re Im Action select the received component of the complex signal Used for spectrum data Get Waveform Data Header Wfm Data Type Data Argument none Forms query only Action return text string with values of all points of the signal comma separated Get Number of Points in Waveform Header Wfim Preamb Poin Type Data Argument none Forms query only Action return number of points in signal Get X axis Step Header Wfm Preamb XInc Type Data Argument none 2010 Pico Technology All rights reserved ps9000pg en 56 Full list of commands Forms query only Action return the increment on the X axis for one signal point Get X axis Origin Header Wfim Preamb XOrg Type Data Argument none Forms query only Action return the X axis value for the first signal point Get X axis Unit Header Wfim Preamb XU Type Data Argument none Forms query only Action return th
32. echnology All rights reserved ps9000pg en 42 Full list of commands Action set the Y positions of the histogram window Window Visibility Header Type Action Hist Display On off set visibility of the window Calculation Mode Header Type Arguments Action Hist Mode Selector Normal Exponent set mode of histogram calculation Weight for Exponential Calculation Header Type Argument Action Scale Type Header Type Arguments Action Scale Mode Header Type Arguments Action Linear Scale Headers Type Hist Weight Integer 8 16 32 8192 set the number of signals for termination of acquisition Hist ScaleType Selector Linear Logarith set the type of the histogram scale Hist ScaleMode Selector Auto Manual set the mode of the histogram scale of Vertical or Horizontal Histogram Hist VertScale Hist HorScale Float ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 43 Argument Action 10 to 100 div set the scale of the histogram Used for Manual mode and Linear type of scale only Linear Offset of Vertical or Horizontal Histogram Headers Type Argument Action Hist VertOffset Hist HorOffset Float 0 to 100 set the offset of the histograms It used for Manual mode and Linear type of scale only Logarithmic Scale of Vertical or Horizontal Histogram
33. ected from Ch2 Input Autocalibration of the Channels is aborted 3 Signal must be disconnected from Chi and Ch2 Inputs Autocalibration of the Channels is aborted 2010 Pico Technology All rights reserved ps9000pg en 54 Full list of commands 5 Autocalibration failed Version This query can be used with PicoScope 9000 SW v 2 3 2 or later When to Begin Autocalibration Header Util CalibrWhen Type On off group Items PowerOn Period Temperat Action PowerOn autocalibration begins on the next Power On Period autocalibration begins periodically with the specified interval Temperat autocalibration begins when deviation of temperature inside the instrument exceeds the specified value Note for the PicoScope 9000 SW v 2 3 2 or later When the GUI is in Invisible or RemoteOnly state the autocalibration routine can not be performed spontaneously and independently on this command see at the GUI command above Autocalibration Period Header Util CalPeriod Type Float Argument 0 5 to 16 hours Action sets autocalibration period in hours Temperature Deviation Header Util TempChange Type Float Argument 0 5 to 10 C Action sets temperature deviation for autocalibration Get the Temperature of the Instrument query Header Calibr Temperature Type Float Argument none Forms query only Action return the temperature inside the device in degrees Celsius Version This command can b
34. ed Parameters 52 Get Statistic Value of Parameter 52 Getting Mask Results commands for Get Integrated Results of Mask Test 48 Get Number of Samples in Margins of Selected Polygon 48 Get Number of Samples in Selected Polygon with Margins Together 48 Get Number of Samples in Selected Polygons Getting Measurement Results commands for 2010 Pico Technology All rights reserved ps9000pg en 62 Index Getting Measurement Results commands for Get Current Value of Parameter 31 Get List of Measured Parameters 31 Get Statistic Value of Parameter 31 Getting Results of Histogram commands for Get Histogram Data 44 Get Histogram Measurement 44 Group off mode 6 Group on mode 6 Group query mode 6 GUI Ready query 56 H Header commands Header 10 Headers 3 Integer type commands 8 L LabView 59 Limit Tests commands Action 31 Action If 31 Center Value 31 Failure When 31 File Name 31 Format of Stored Files 31 If Measurement Undefined 31 Limit Test On Off 31 Limit Test Termination Condition 31 Number of Failures 31 Number of Waveforms 31 Parameter Activity 31 Parameter Center Mode 31 Parameter Delta Mode 31 Parameter Delta Percentage for User Defined mode 31 Parameter Delta Value for Standard Deviation mode 31 Parameter Delta Value for User Defined Mode Parameter Limit Mode 31 Upper and Lower Limits of Parameters 31 Line feeds 4 List of commands 10 M Markers commands Marker Sources 23 Marker
35. eeessccbssontsesses 16 Mask Test EINEN TEE 1 Common ZE Det IT EI E DE 45 2 Standard Mask Commands s88 eo ae OA A ee 47 3 Getting Mask Results W ee 17 Eye Diagram commands s seseseeeeseesesseseessenessesnsssssessessessosseesnsssssessessessossesseesseseesessessessessoneseeseeseese 1 Setting Eye Parameters EE 49 2 Getting Eye Measurement Results EE 52 18 Utilities Command seis siecceeescocsc sossicas osso cao anss ei peer ires nosso NE o le ada DO CSN toed aces ees Neto ee saga VOCE TESES EN cega Dos 53 19 Waveforms commands EE 55 20 System commands eseeseeseeseeseetseetseetseeseeeseeeseeeseessessceesseeseseseesseenaeeeesseesseesseeeseeeeeseeseseseeeaeenseenseaes 56 5 PROSMAMIMINS EXAIMPIES seio gusto ico ege Ee EEN 59 Us TONN 59 BE ZAIT SE 59 3 Visual Basics d Ta W 60 JE 61 ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 1 1 1 1 1 2 1 3 PicoScope 9000 API Reference PicoScope 9000 provides an API for any third party application or library to control the oscilloscope and collect signals The API is COM based and is provided by the PicoScope 9000 GUI application ps9000pg en 4 PicoScope9000 COM Server The COM server implementing the API is called PicoScope9000 and is implemented by the PicoScope 9000 GUI application PicoScope9000 exe It is registered in the system during the setup process and can be explicitly unregistered and registered ag
36. elector Arguments CurrMean UserDef Action set the mode of the center definition for the relevant parameter It used only for the Center mode of the parameter limit Center Value Headers Limitl CenterVal Limit2 CenterVal Limit3 CenterVal Limit4 CenterVal Type Float Arguments absolute value of center Action set the absolute center value It used for UserDef mode of the center of the parameter ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 35 Parameter Delta Mode Headers Limitl Delta Limit2 Delta Limit3 Delta Limit4 Delta Type Selector Arguments StdDev UserDef UserPerc Action set mode of delta definition for relevant parameter It used for Center mode of parameter limit only Parameter Delta Value for Standard Deviation mode Headers Limitl StdDev Limit2 StdDev Limit3 StdDev Limit4 StdDev Type Float Arguments 0 1 to 100 standard deviations of the parameter Action sets the delta value It used for St dDev mode of parameter delta only Parameter Delta Value for User Defined Mode Headers Limitl UserDef Limit2 UserDef Limit3 UserDef Limit4 UserDef Type Float Arguments absolute value of delta Action sets the delta value It used for UserDef mode of delta of the parameter only 2010 Pico Technology All rights reserved ps90
37. for Vertical or Horizontal Histogram 40 Weight for Exponential Calculation 40 Window Visibility 40 Single item mode 6 Specifications 6 Standard Mask Commands Alignment of Signal with Standard Mask 47 Enable Margins 47 Get List of Masks 47 Get List of Standards 47 Margins Value 47 Select Standard 47 Select Standard Mask 47 System commands 56 T Clear Display 10 Recall Default Setup 10 Start Autoscaling 10 Start Cycle Acquisition 10 Start Single Acquisition Stop Acquisition 10 System commands 10 Timese commands Acquire a Channel 13 Attenuator dB 13 Attenuator linear log 13 Attenuator ratio 13 Attenuator unit 13 Bandwidth of Channel 13 Display a Channel 13 Offset a Channel 13 Scale a Channel 13 Timebase Commandes 13 Trigger Commands Acquire a Channel 14 AttenuatordB 14 Attenuator linear log 14 Attenuator ratio 14 Attenuator unit 14 Bandwidth of Channel 14 Display a Channel 14 Offset a Channel 14 Scale a Channel 14 2010 Pico Technology All rights reserved ps9000pg en 64 Index U Utilities commands V Autocalibration Period 53 Start Autocalibration of Channels 53 Start Autocalibration of Timebase 53 Temperature Deviation 53 When to Begin Autocalibration 53 Visual Basic NET 60 W Waveforms commands Get Number of Points in Waveform 55 Get Waveform Data 55 Get X axis Origin 55 Get X axis Step 55 Get X axis Unit 55 Get Y axis Unit 55 Spectrum Format 55
38. for the Peak method Left and Right Spectrum Peaks Headers Meas lt src gt PeakLeft Meas lt src gt PeakRight Type Integer Arguments 1to41 Action sets the first and second peaks for the specified spectrum signals List of Spectrum Frequency Measurements Header Meas lt src gt XFFTPar Type On off group Items Freq DFreg Action define the set of the frequency measurements for the specified signals List of Spectrum Magnitude Measurements Header Meas lt src gt YFFTPar Type On off group Items Magn DMagn TDH Action define the set of the magnitude measurements for the specified signals ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 31 4 11 3 Getting Measurement Results 4 12 Get List of Measured Parameters Header Meas Res List Type Data Argument none Forms query only Action return text with the list of the active measurements for all signals with ordinal index Get Current Value of Parameter Header Meas Res lt N gt Parameter lt N gt index of the parameter in the list Type Data Argument none Forms query only Action return the last result of the specified measured parameter Get Statistic Value of Parameter Header Meas Res lt N gt lt Val gt Parameter lt N gt index of the parameter in the list Parameter lt Val gt Wim Min Max Mean StdDev Type Data Arguments none Forms command wit
39. h query only Action return the specified statistic parameter of the measured parameter Limit Tests commands Limit Test On Off Header Limit TestOn Type On off Action Enable disable the Limit Test Must be set On after full definition of all other Limit Test parameters 2010 Pico Technology All rights reserved ps9000pg en 32 Full list of commands Limit Test Termination Condition Header Type Arguments Action Selector StopBtn Number of Failures Header Type Argument Action Limit RunUntil Failur Wfm set condition of Limit Test Termination Limit Failures Integer 1 to 10000 Number of Waveforms Header Type Argument Action Action Header Type Items Action Action If Header Type Arguments Action set number of failures for the Failur Condition of the Limit Limit NWfms Integer 1 to 1000000 set the number of waveforms for the F m Condition of the Limit Limit Action On off group Beep Save Stop Save every signal with a limit condition is stored to the disk Beep the beep signal will sound for every limit condition Stop acquisition immediately stops after the first limit condition Limit If Selector AnyFail AllPass AllFail AnyPass define the limit condition AnyFail one or more active measures fails A all 1Pass all active measures are good Fail all active measures fail
40. h1 NEnv Acq Ch2 NEnv Type Integer Argument 1 2 4 8 16 4096 8192 Action set number of signals for envelope mode for specified channel Argument 8192 is used for unlimited number of signals Channel Record Length Header Acq Chl RecLen Acg Ch2 RecLen Type Integer Argument 32 64 128 4096 Action set number of points for specified channel Termination of Acquisition Header Acgq RunUntil Type Selector Arguments StopBtn NAcg Action set condition for terminating acquisition when the Stop Button pressed or after specified number of waveforms is reached Number of Waveforms Header Acgq NAcg Type Integer Argument 1 to 65535 Action set number of signals for terminating acquisition Action when Number of Waveforms reached Header Acg React Type On off group Items Beep Save Action if Save is turned on every signal is stored to disk if Beep is turned on the beep signal will sound after the specified number of waveforms is reached 2010 Pico Technology All rights reserved ps9000pg en 18 Full list of commands 4 8 Display commands Trace mode Header Type Arguments Action Displ TraceMode Selector AllLocked PerTrace in PerTrace mode every waveform may be displayed in its own style in All Locked mode the display style of all waveforms is set as the style of the active trace Select active trace Header Type Arguments A
41. he polygon 1 to 8 Type Data Argument none Forms query only Action return text string with value of failed samples on specified polygon Get Number of Samples in Margins of Selected Polygon Headers Mask Res Poly lt N gt Mar Parameter lt N gt number of the polygon 1 to 4 Type Data Argument none Forms query only Action return text string with value of failed samples on margin of specified polygon Used when Margins enabled Get Number of Samples in Selected Polygon with Margins Together Headers Mask Res Poly lt N gt Al11 Parameter lt N gt number of the polygon 1 to 4 Type Data Argument none Forms query only Action return text string with total number of failed samples on the margin and on the specified polygon Used when Margins enabled 4 17 Eye Diagram commands 4 17 1 Setting Eye Parameters Type of Eye Measurements Header Eye Measure Type Selector Arguments Off NRZ RZ Action set type of eye measurements Source for Eye Measurements Header Eye Source 2010 Pico Technology All rights reserved ps9000pg en 50 Full list of commands Type Selector Arguments Chil Ch1B2 Ch2 Ch2B2 F1 F2 F3 F4 M1 M2 M3 M4 S1 S2 Action set source for eye measurements Number of Waveforms in One Measurement Header Eye WfmsInCycle Type Integer Argument 64 128 256 512 1024 Action set number of waveforms in one measurement Eye Frame Visibi
42. hnology All rights reserved PicoScope 9000 Series User s Guide 63 O On Off Group type commands 6 On Off type commands 6 Overview 2 P PicoScope9000 COM Server 1 Programming Examples 59 Q Queries 3 Query commands 2 S Selector type commands 8 Semicolon 4 Set commands 2 Setting Eye Parameters commands for Eye Frame Visibility 49 Left and Right Boundary for NRZ Top Base Finding 49 List of X axis NRZ Measurements 49 List of X axis RZ Measurements 49 List of Y axis NRZ Measurements 49 List of Y axis RZ Measurements 49 Measurement Statistics 49 Measurement Statistics Mode 49 Nagnber of Waveforms in One Measurement Source for Eye Measurements 49 Threshold Definition Mode 49 Type of Eye Measurements 49 Upper and Lower Threshold 49 Weight Value 49 Window Value 49 Setting Histogram Parameters commands Calculation Mode 40 Histogram Axis 40 Histogram Finish Condition 40 Histogram Source 40 Histogram Visibility 40 Left and Right Window Limits for Vertical or Horizontal Histogram 40 Limit Mode for Histogram Window 40 Limit Units for Histogram Window 40 Linear Offset of Vertical or Horizontal Histogram 40 Li gar Scale of Vertical or Horizontal Histogram Logarithmic Offset of Vertical or Horizontal Histogram 40 Logarithmic Scale of Vertical or Horizontal Histogram 40 Number of Samples for Histogram 40 Number of Waveforms for Histogram 40 Scale Mode 40 Scale Type 40 Top and Bottom Window Limits
43. inimum valid length of a signal Float type commands The float type commands modify specific real value functions For example the command Chl Scale 0 1 ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 3 7 sets the Y scale for Channel 1 to 100 mV div The valid range and increment of each value is different and is described in the list of commands 10 Float type commands have a query form It is also possible to use the query form with an argument for example Chl Scale 0 1 returns 100 mV div when V div is dimension of the scale and the prefix m is milli The commands TB ScaleA 0 0000001 TB ScaleA 100e 9 TB ScaleA O 1lu TB ScaleA 100p are equivalent and set the Scale A of the timebase to the value 100 ns div All of these commands return 100 ns div Data type commands The data type commands are used to send some data to the instrument or to receive some data from the instrument such as an acquired signal s array of points the result of a measurement and so on Some data type commands have a query form only while others have both a command and a query form The structure of the data is different for each command and is specified in the list of commands 10 2010 Pico Technology All rights reserved ps9000pg en 10 Full list of commands 4 1 4 2 4 3 Full list of commands All of the PicoScope 9000 API
44. isplay Styles to initial state variable persistence 2 c Display Format Header Displ Format Type Selector Arguments YT 2YT 4YT XY CombYTXY Comb2YTXY Action select number and kinds of screens Define Trace Screen for 4YT Format Header Displ Screen4 lt trace gt where lt trace gt is Chl Ch1B2 Ch2 Ch2B2 Fl F2 F3 F4 M1 M2 M3 M4 S1 S2 Hist Type Selector Arguments 1 2 3 4 Action move specified trace onto specified screen in 4YT format Define Trace Screen for 2YT Comb2YTXY Formats Header Displ Screen2 lt trace gt where lt trace gt is Chl Ch1B2 Ch2 Ch2B2 Fl F2 F3 F4 M1 M2 M3 M4 S1 S2 Hist Type Selector Arguments 1 2 Action move specified trace onto specified screen in 2YT or Comb2YTXY formats Source of X Axis for XY Screen Header Displ XAxis 2010 Pico Technology All rights reserved ps9000pg en 20 Full list of commands Type Selector Arguments Chi Ch1B2 Ch2 Ch2B2 F1 F2 F3 F4 M1 M2 M3 M4 S1 S2 Action set specified signal as X axis for XY screen Source of Y Axis for XY Screen Header Displ YAxis Type Selector Arguments Chil Ch1B2 Ch2 Ch2B2 F1 F2 F3 F4 M1 M2 M3 M4 S1 S2 Action set specified signal as Y axis for XY screen Graticule Type Header Displ Gratic Type Selector Arguments Grid Frame Axis Off Action define type of graticule for YT and XY screens 4 9
45. itions of the left or right slope It used for slope margins definition mode only Position of Left or Right Margin Headers Type Arguments Action Meas lt src gt LeftMarker Meas lt src gt RightMarker Float absolute time value sets the position of margin for the specified signals It used for marker margins definition mode only List of X Measurements Header Type Items Action Meas lt src gt XParam On off group Period Freq PosWidth NegWidth Rise Fall PosDuty NegDuty PosCross NegCross BurstWidth Cycles TimeOfMax TimeOfMin PosJitterPp PosJitterRMS NegJitterPp NegJitterRMS define the set of the X axis measurements for the specified signals List of Y Measurements Header Type Items Action Meas lt src gt YParam On off group Max Min PP Top Base Ampl Middle Mean dcRMS acRMS Area CycMean CycDCRMS CycAcRMS CycArea PosOver NegOver define the set of the Y axis measurements for the specified signals Second Source for Inter Signal Measurements Header Meas Source2 ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 29 4 11 2 Type Selector Arguments Chi Ch1B2 Ch2 Ch2B2 F1 F2 F3 F4 M1 M2 M3 M4 S1 S2 Action set the second source for the inter signal measurements List of X Inter Signal Measurements
46. lity Header Eye DispaylWind Type On off Action set visibility of eye frame Measurement Statistics Header Eye Statistic Type On off Action enable disable measurement statistics Measurement Statistics Mode Header Eye Mode Type Selector Arguments Permanent Window Weight Action set mode of statistics calculation Used when statistics enabled Window Value Header Eye Window Type Integer Argument 8 16 32 8192 Action set window value Used for Window mode of statistics Weight Value Header Eye Weight Type Integer ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 51 Argument 8 16 32 8192 Action set weight value Used for Weight mode of statistics Left and Right Boundary for NRZ Top Base Finding Headers Eye LeftBound Eye RightBound Type Float Argument 10 to 90 of the NRZ period Action set the zone of the period of the NRZ signal for the top base calculation Threshold Definition Mode Header Eye TreshMode Type Selector Arguments 10 90 20 80 Custom Action set mode of threshold definition Upper and Lower Threshold Headers Eye UpTresh Eye LowTresh Type Float Argument 5 to 95 of amplitude Action set the thresholds for the slopes calculation Used for Custom mode only List of X axis NRZ Measurements Header Eye XNRZParam Type On off group Items Area BitRate BitTime CrossTime CycleArea
47. ment none Forms query only Action return the name of the current instrument in the format User Name Sys Name when User Name user name of the Instrument optional Sys Name system name of the instrument from the list ABSENT DEFAULT DEMO DEFAULT LAN DEFAULT USB DEMO2 DEMOS UNIT1 UNIT2 UNIT3 UNIT4 UNIT5 UNIT6 UNIT7 UNIT8 Version This command can be used with PicoScope 9000 SW v 2 4 0 or later 2010 Pico Technology All rights reserved ps9000pg en 58 Full list of commands Switch to another device command with query Header Type Argument Forms Action Version Instr Current Set Argument Data Specifier of the required instrument May be one of three alternatives a number of devices in the system a user name or a system name of the instruments Note that all space symbols in the names must be changed to the underline symbol in the commands command with query only The GUI changes to the given device If the switch is successful the command returns OK If the new device is already used by another interface the command returns BUSY If a new device is not connected to PC the command returns NOT DETECTED This command can be used with PicoScope 9000 SW v 2 4 0 or later GUI Ready query Header Type Argument Forms Action Instr GUlReady On Off none query only return OFF when the GUI is not finished l
48. nnels commands Acquire a Channel 11 Attenuator dB 11 Attenuator linear log 11 Attenuator ratio 11 Attenuator unit 11 Bandwidth of Channel 11 Channels commande 11 Display a Channel 11 Offset a Channel 11 Scale a Channel 11 Characteristics 6 Command classification 6 Command entry rules 4 Command header 3 Command messages 2 Commande 3 list of 10 Common Mask Commands Actuate Mask Testing 45 Format of Stored Files 45 Load User Mask 45 Mask Erasing 45 Mask Test Actions 45 Mask Test Finish Condition 45 Number of Failed Samples 45 Number of Failed Waveforms 45 Number of Samples 45 Number of Waveforms 45 Signal for Mask Testing 45 Stored File Name 45 User Masks File Name 45 COMRC Object 1 Concatenating 4 D Data type commands 9 Delphi 59 Display commands Define Trace Screen for 2YT Comb2YTXY Formats 18 Define Trace Screen for 4YT Format 18 Display commands 18 Display Format 18 Graticule Type 18 Persistence Time seconds for VarPersist Style 18 Refresh Time seconds for VarGrayScal or VColorGrade Styles 18 Reset Display Style 18 Select active trace 18 Set Display Style 18 Source of X Axis for XY Screen 18 Source of Y Axis for XY Screen 18 Trace mode 18 E ExecCommand Method 1 Execution type commands 6 F FFT commands Enable Spectra 39 Spectrum Source 39 Window 39 Float type commands 8 G Getting Eye Measurement Results commands for Get Current Value of Parameter 52 Get List of Measur
49. nt al IFFT LinInt SinInt Smooth Trend set the operator of the specified function Fi Sourcel F2 Sourcel F3 Sourcel F3 Sourcel Selector Chl Ch1B2 Ch2 Ch2B2 EI F2 F3 F4 MI M2 M3 M4 Sl S2 set the first operand of the specified function 2010 Pico Technology All rights reserved ps9000pg en 38 Full list of commands Headers Fl Source2 F2 Source2 F3 Source2 F4 Source2 Type Selector Arguments Chi Ch1B2 Ch2 Ch2B2 F1 F2 F3 F4 M1 M2 M3 M4 Sl S2 Const Action set the second operand of the specified function Used for Add Sub Mult Div operators only Constant Value Headers F1l Const F2 Const F3 Const F4 Const Type Float Arguments absolute value of constant Action set the constant for the specified function Used when Const is the second operand Smoothing Parameter Headers F1l SmoothLen F2 SmoothLen F3 SmoothLen F4 SmoothLen Type Integer Argument 3 5 7 9 49 51 Action set the length of the smoothing interval in points for specified function Used for Smooth operator only Trend Measurement ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 39 Headers Type Arguments Action Fi Tren F2 Tren F3 Tren F4 Tren Selector Period Freq PosWidth NegWidth RiseTime FallTime PosDuty NegDuty
50. oading and is not ready and return ON when GUI is ready Version This command can be used with PicoScope 9000 SW v 2 4 1 or later ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 59 5 Programming Examples Your PicoScope installation includes programming examples in the following languages and development environments e Delphi e LabVIEW e Visual Basic NET 5 1 Delphi The program PicoScopeDelphiClientExample dproj in the Delphi_Client_Example subdirectory of the PicoScope9000 SDK materials demonstrates how to operate PicoScope 9000 Series PC Oscilloscopes The file PicoScope9000_TLB pas is the description of the PicoScope9000 COMRC object You must include this file in your own programs Other files required for the example are MainClient pas MainClient dfm PicoScopeDelphiClientExample dproj PicoScopeDelphiClientExample dsk PicoScopeDelphiClientExample identcache PicoScopeDelphiClientExample res All these 8 files must be put into the same directory and compiled This has been tested with Delphi 2009 5 2 LabVIEW The program PicoScope_Example vi Test_Get_Data vi in the LabView_Client_Example subdirectory of the PicoScope9000 SDK materials demonstrates how to operate PicoScope 9000 Series PC Oscilloscopes This files must be put into the some directory and compiled It has been tested with
51. query command with query Action define file name for storing Custom Setup Save Custom Setup Header Save Setup Save Type Execution Action stores present front panel setup as previously specified custom setup Recall Custom Setup Header Save Setup Recall Type Execution Action recall setup previously saved to file The name of the setup must first be defined by the command Save Setup FileName 4 10 Markers commands Marker Type Header Mark Type Type Selector Arguments Off MX MY XY Action set marker type Marker Sources Header Mark M1l Source Mark M2 Source Type Selector Arguments Chi Ch1B2 Ch2 Ch2B2 F1 F2 F3 F4 M1 M2 M3 M4 S1 S2 Action attach specified marker to specified signal X position of Marker Header Mark M1 XPos Mark M2 XPos Type Float Argument real value of X axis 2010 Pico Technology All rights reserved ps9000pg en 24 Full list of commands 4 11 4 11 1 Action set the X position of specified marker Y position of Marker Header Mark M1 YPos Mark M2 YPos Type Float Argument real value of Y axis Action set Y position of specified marker Motion of Markers Header Mark Motion Type Selector Arguments Independ Paired Action when Paired motion is selected you can move both markers with the M1 POSITION variable simultaneously while the difference between markers can be moved with the M2 POSITION variable Measure comman
52. returns all the histogram parameters Headers You can control whether the instrument returns headers as part of the query response Use the HEADer command to control this feature If header is on the query response returns command headers and formats itself as a valid set command When the header is off the response includes only the values This may make it easier to parse and extract the information from the response The table below shows the difference in responses Comparison of Header Off and Header On Responses Chi Scale 200 mV div CH1 SCALE 200 mV div Acq Ch1 RecLen ACQ CH1 RECLEN 512 2010 Pico Technology All rights reserved ps9000pg en Commands Syntax 2 2 Command Entry 2 2 1 Rules The following rules apply when entering commands e You can follow a mnemonic by any letters for more easy understanding of the program s text For example the commands Chl ATTEN DIMENS Volt Chl ATTENuator DIMENSion Volt Ch1 ATTENblabla DIMENSblabla Volt are equivalent How ever arguments must be written without any following letter e You can enter commands in upper or lower case e You can precede any command with white space characters White space characters include any combination of the ASCII control characters 00 to 09 and OB to 20 hexadecimal 0 to 9 and 11 to 32 decimal e The instrument ignores commands consisting of any combination of white space characte
53. rs and line feeds 2 2 2 Concatenating You can concatenate any combination of set commands and queries using a semicolon The instrument executes concatenated commands in the order received When concatenating comma nds and queries you must follow these rules e Separate completely different headers by a semicolon and by the beginning colon on all commands except the first one For example the commands TRIGger MO DE FREE ACQuire NUMAVg 10 can be concatenated into the following single command TRIGger MO DE FREE ACQuire NUMAVg 10 If concatenated commands have headers that differ by only the last mnemonic you can abbreviate the second command and eliminate the beginning colon For example you can concatenate the commands ACQuire Ch 1 MODE ENVMINMAX ACQuire Chl NAVG 10 into a single command ACQuire Ch 1 MODE ENVMINMAX NAVG 10 The longer version works equally well ACQuire CH 1 MODE ENVMINMAX ACQuire NAVG 10 ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 5 e Set commands and queries may be concatenated in the same message For example ACQuire CH1 MODE AVGSTAB NAVG is a valid message that sets the acquisition mode to Stable Averaging The message then queries the number of acquisitions for averaging Concatenated commands and queries are executed in the order received
54. set presence and unit of attenuator or converter used with direct trigger input Direct Input Attenuation ratio Header Trig Atten Ratio 2010 Pico Technology All rights reserved ps9000pg en 16 Full list of commands Type Float Argument 0 0001 to 1000000 Action set attenuation ratio This setting is active only when attenuator unit is ratio Direct Input Attenuation dB Header Trig Atten DB Type Float Argument 80 to 120 Action set the attenuation in dB This setting is active only when attenuator unit is decibels 4 7 Acquisition commands Type of signal Header Acg FitTo Type Selector Arguments Multi Single Action prepare the instruments for best acquisition of single valued or multi valued Sampling Mode Header Acq Sampl Type Selector Arguments Simult Altern Action Simult set simultaneous acquisition on Channels 1 and 2 Alternate set alternate acquisitions on Channels 1 and 2 Acquisition Mode of Channel Header Acg Ch1i Mode Acg Ch2 Mode Type Selector Arguments Sample AvgStab AvgMult EnvMinMax EnvMax EnvMin Action set acquisitions mode of specified channel Channel Averaging Header Acq Chl NAvg Acq Ch2 NAvg Type Integer ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 17 Argument 1 2 4 8 16 4096 Action set averaging coefficient for specified channel Channel Envelopes Header Acqg C
55. splay of corresponding channel s signal Acquire a Channel Header Chl Acquire Ch2 Acquire Type On Off Action turn on or off the acquisition of the channel s signal when its display is turned off Scale a Channel Header Chl Scale Ch2 Scale Type Float Argument 0 002 to 0 5 or other when attenuator is used Action set the specified display scale in V div Offset a Channel Header Chi Offset Ch2 0ffset Type Float Argument 1 to 1 or other when attenuator is used Action set the specified compensation voltage of the channel in V 2010 Pico Technology All rights reserved ps9000pg en 12 Full list of commands Bandwidth of Channel Header Chi Band Ch2 Band Type Selector Arguments Full Narrow Action set the bandwidth of the channel Attenuator linear log Header Chil Atten Unit Ch2 Atten Unit Type Selector Arguments Off Ratio DB Action set presence and scale of attenuator or converter used with the channel Attenuator ratio Header Chl Atten Ratio Ch2 Atten Ratio Type Float Argument 0 0001 to 1000000 Action set the attenuation ratio This setting is active only when attenuator unit is ratio Attenuator dB Header Chl Atten DB Ch2 Atten DB Type Float Argument 80 to 120 Action set the attenuation in dB This setting is active only when attenuator units is decibels Attenuator unit Header Chl Atten Dimens Ch2 Atten Dimens Type Sele
56. t RunUntil Selector StopSingle Wfms Samples set finish condition for acuiring histogram Number of Waveforms for Histogram Header Type Argument Hist NwWfm Integer 1 to 1000000 ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 41 Action set number of signals for termination of histogram acquisition Number of Samples for Histogram Header Hist NSample Type Integer Argument 1 to 10000000 Action set number of samples for termination of histogram acquisition Limit Mode for Histogram Window Header Hist Limits Type Selector Arguments Paried Independ Action set the mode of the limits of histogram window Limit Units for Histogram Window Header Hist Units Type Selector Arguments Absolute Percent Action set the units of the limits of histogram window Left and Right Window Limits for Vertical or Horizontal Histogram Headers Hist WVert Left Hist WVert Right Hist WHor Left Hist WHor Right Type Float Argument real value of the X axis for Absolute units 0 to 100 of the X axis for Percent units Action set the X positions of the histogram window Top and Bottom Window Limits for Vertical or Horizontal Histogram Headers Hist WVert Top Hist WVert Bottom Hist WHor Top Hist WHor Bottom Type Float Argument real value of the Y axis for Absolute units 0 to 100 of the Y axis for Percent units 2010 Pico T
57. ter is used between a command header and its argument Optionally a white space may consist of multiple white space characters ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 3 2 1 3 2 1 5 Commands Commands cause the instrument to perform a specific function or change one of its settings Commands have the structure lt Header gt lt Space gt lt Argument gt lt Comma gt lt Argument gt A command header consists of one or more mnemonics arranged in a hierarchical or tree structure The first mnemonic is the base or root of the tree and each subsequent mnemonic is a level or branch off the previous one Commands at a higher level in the tree may affect those at a lower level The leading colon always returns you to the base of the command tree Queries Queries cause the instrument to return information about its status or settings Queries have the structure e lt Header gt e lt Header gt lt Space gt lt Argument gt lt Comma gt lt Argument gt You can specify a query command at any level within the command tree unless otherwise noted These branch queries return information about all the mnemonics below the specified branch or level For example HIStogram STATistics STDdev returns the standard deviation of the histogram while HIStogram STATistics returns all the histogram statistics and HIStogram
58. tion All on off type commands have a query form The queries return one of two fixed values ON or OFF It is also possible to use the query form with an argument For example Chl Display O This command turns off the graphic of Channel 1 and returns OFF On off group type commands Some functions of the instrument have items that may be set independently on or off It is also possible for the items to be either all on or all off An example of this type of command is Meas Chl1 XParam This command has a set of parameters for automatic X axis measurements for Ch1 It is possible to select up to 10 parameters from a list of 18 Period Freg PosWidth NegWidth Rise Fall PosDuty NegDuty PosCross NegCross BurstWidth Cycles TimeOfMax TimeOfMin PosJitterPp PosJitterRMS NegJitterPp NegJitterRMS There are between 2 and 64 custom items in the on off group type commands The full set of items is specified for each commands in the list of commands 0 ps9000pg en 2010 Pico Technology All rights reserved PicoScope 9000 Series User s Guide 7 The on off group type commands can be used in several modes Every such command can be used in every mode Single item mode Single item mode is used to control one commande item without changing the other items In this case the item s mnemonic is added to the end of command after a colon character This must be followed by a space character and then one of

Download Pdf Manuals