LSR II Quick User Guide

image

Contents

1. acquisition does not record any data 5 Set the number of events to record 6 Click Record Data a Recording will stop once the cytometer reaches the number of events to record 7 Remove the tube from the SIP Let two drops fall down before installing the next tube This limits the carry over from one sample to the other If rear cells are of interest run some water in between the samples Attention The fluidic system is controlled by the panel located on the cytometer and not by the software If you stop the acquisition in the software the sample continues to be sucked into the machine and you will loose your cells To stop the aspiration of your sample press the standby button on the control panel of the cytometer and remove your sample from the SIP Creating Plots and Histograms Plots and histograms are created in the Global Worksheet The Global Worksheet is the main working area where you create plots define gates show statistics and population hierarchies and enter custom text Global Worksheet Global Sheet sot rr Be H Buttons to create dot Buttons to create gates plots contour plots and histogram Compensation Set up 1 Select Experiment gt Compensation Setup gt Create Compensation Controls 2 Check that all the parameters that you require are listed Remove the parameter for the live dead dye this can be added again after the compensation a Other option is to kill some of the cells to ensure that ther
2. b Press the PRIME fluid control button to force the fluid out of the flow cell and into the waste container c Repeat the priming procedure d Install a FACS tube clear plastic with 1ml of DI water on the SIP and run it on high for about 2min 4 The cytometer is ready to use
3. e BD FACSDiva Software Administrator1 File Edit Yiew Experiment Populations Worksheet Cytometer HTS Help 4 SeHFee f Browser Experiment_002 x Cytometer LSRII 1 x vr Wenchest Ganal Grea TELELE m WE gt se lA 2 2 a _ Name Date BQ Administrator 1 BS Practice Bes Experiment_002 11 9 06 7 03 34 PM ere 3S Cytometer Settings ERS Global Worksheets ax Specimen_O01 as Calibrite Beads a Tube_o04 12 19 03 10 16 14 a f Tube _003 12 19 03 10 11 00 fa Tube_002 12 19 03 10 07 06 6 fa Tube _001 12 19 03 10 00 24 i ay Compensation Controls i a Shared View Calibrite Beads Tube_001 Pues Cytometer Connected 50 100 150 200 250 FSC A tx 1 000 Inspector Global Sheet1 Tube Tube_001 General Headers and Footers Worksheet Grid Population Events Parent Total Name Global Sheet1 a ma poy pa yes Number of Pages Ls i F2 700 23 2 3 1 Vertical P3 19 684 65 6 65 6 Horizontal Pa 4 188 140 14 0 Show Page Breaks Experiment Name Experiment_o02 Specimen Name Calibrite Beads Tube Name Tube_001 Record Date Dec 19 2003 10 00 24 AM ee SOP Administrator EH Acquisition Dashboard GUID c557f1f0 cbc1 420b 9cca 9hecl be2c942 Current Activity Display Range Linear Scale 0 10 000 Active Tubewell Threshold Rate Stopping Gate Event
4. e is a strong signal in the live dead channel 3 Click OK and a new specimen called Compensation Controls appears in your experiment Click on the to display the individual compensation control tubes 4 Set the current tube pointer to the Unstained Control tube under the Compensation Control Specimen ft B 4 Compensation Samples SS Cytometer Settings H fy Unstained Control WA FITC Control J fy PE Control a fg PerCP Cy55 Control WA APC Control ol S el e 5 Load the unstained control tube press Run on the LSR and click Acquire in the software 6 Adjust the P1 gate to include the population of interest 7 Adjust the voltage for each parameter to ensure appropriate background signal 8 You should see both sides of the histogram usually the negative signal is within the first decade of the histogram FITC Stained Control 9 Click Record Data a The software automatically records 5000 evt 10 Select P1 right click and select Apply to All Compensation controls 11 Run and record each single color compensation tubes 12 Adjust the P2 gates to fit the positive populations 13 Select Experiment gt Compensation Setup gt Calculate Compensation 14 Rename the compensation with the current date and your name Single Stained Setup Compensation calculation has completed successfully Name Experiment A Link amp Save Apply Only 15 Select Link amp Save to apply your com
5. he ECOLE POLYTECHNIQUE LSR Il Quick User Guide FEDERALE DE LAUSANNE Flow Cytometry Core Facility August 2010 v 3 During normal workdays the LSR II will already be up and running and the FCCF staff will have done a performance check The software will be ready for you to log in with your personal username and password Please be aware that every second month we will delete all the experiment in the software and on the computer Please make sure that you take the experiment and FCS files with you You are responsible for your own backup Only use the clear FACS tubes on the LSR II the FACS tubes that you use on the Cyan do not work properly on the LSR II All samples must be filtered just before acquiring P2 cells have to be fixed before running on the LSRII Start up Only required if you are using the LSR II on the weekend or on public holidays 1 Turn on computer and log in 2 Start BD FACSDiva software by double clicking the shortcut on the desktop and log in to the software Use your personal username and password to log in to the FACSDiva software 3 Turn the flow cytometer on Green button in front of cytometer The cytometer connects automatically While connecting the message Cytometer Connecting is displayed in the window footer When connection completes the message changes to Cytometer Connected 5 Allow 30 minutes for lasers to warm up and stabilize 6 During this time run some water on low FACSDiva Workspac
6. our experiment right click and select Export gt FCS file a Select FCS 3 0 b Click OK c Inthe Save Export dialog box verify the file storage location In between users 1 Run 3min of Decontamination solution on high 2 Run 3min of Rinse on high 3 Run 3min of a new tube of Water on high 4 Log out from the software Last User of the day and Shutdown 5 Run 3min of Decontamination solution on high Run 3min of Rinse on high Run 3min a new tube of Water on high Shut down the software Shut down computer 0 Press the green button in front of LSRII to shut down the cytometer Se ase Changing the Sheath and Waste Tank Always ask staff to change the tanks for you If nobody is around Make sure that the cytometer is set to Standby before changing the tanks 1 Sheath Tank a Unscrew the lid and carefully remove sample line and put it in the holder to your left b Press the Alarm button to turn off the Alarm c Geta new full BD FACS Sheath box on the right next to the entrance d Put the lid with the sample line back on the new box and put the box back to original position e Press Restart 2 Waste Tank a Remove the black holder that holds up the opening of the waste tank b Unscrew the lid of the waste tank c Getan empty waste tank from under the sink to the left of the LSR II d Exchange the lids e Put the full tank back under the sink 3 Prime the fluidics after changing the tanks a Install an empty tube on the SIP
7. pensation to your acquisitions 16 If you excluded the Live Dead marker in the compensation setup a Right Click on Cytometer Settings b Select Unlink from compensation c Runandrecord your PI sample 17 Switch back to the global worksheet to start creating graph to view the data Global Worksheet Global Sheet1 BIS 2 k BBL O PAH gE Sen ESHS TAN S ma gt e Buttons to switch between compensation worksheet and the global worksheet Define labels for each parameter 1 Choose Experiment gt Experiment Layout 2 Selecta label and enter the parameter e g CD3 into the label text field under the Quick Entry 3 Select the next label and enter the next parameter etc 4 When all the parameters are assigned to a label are click OK Quick Entry Labels Label Name Administrator we 7 4 CD16 56 Specimen_O01 hdd to List Delete From List U Tube_001 CD3 016 56 FITC PE U Tube_002 cD3 CD16 56 ls PE Assign or Remove Labels Export Experiment and FCS files 1 Select your experiment right click and select Export gt Experiment Export Experiments Delete experiments after export Directory D BCExport Experiment Practice Experiment Date Simple_Analysis 12 20 06 3 24 10 PM A v a Note The experiment can only be opened with the DIVA software 2 Export your experiment to the D drive 3 To export the FCS files select y
8. pgeeeee gt KAARE Es gt gt 5 gt 3 5 gt 5 gt gt lt 4 To change a parameter click on the name and a drop down list with all available parameters appears 5 Click the New Specimen button to add a specimen and a tube to the experiment be Browser Manual aa ee q 6 Click the New Tube button to add a second tube 7 To rename Experiment Specimen or Tube right click and select Rename 8 Click to set the current tube pointer to Tube_001 in the Browse The pointer changes to green and five green B Ld r 001 tabs appear in the Cytometer window The 3 jesse 35 Cytometer Settings current tube pointer indicates the tube for a amp ro Tage TEE ees eae which cytometer settings adjustments will apply and for which acquisition data will be 3 44 specimen _001 shown D T Tube 001 j J Tube_002 9 Click the Parameters tab in the Cytometer window to check the parameters Select H and W parameters for FSC and SSC to be able to exclude doublets Cytometer FACSAria 1 Parameter Voltage Log amp FSC 347 a PE Texas Red PerCP Cy5 5 so MIM PE Cy APC APC Cy A ooopopoar o oO oO al oO Ll oO Running Samples 1 Install your tube onto the SIP 2 Set the current tube pointer to appropriate tube 3 Press Run on the fluid control panel 4 Click Acquire Data in the Acquisition Dashboard in the Diva Software a Be aware that just clicking
9. s Elapsed Time FSC A Population Events Parent Mean Tube_001 0 evtis 0 evt 00 00 00 H All Events 30 000 HE 3 646 Mri 28 879 96 3 3 526 Basic Controls Hr 700 2 3 7 028 EF 19 684 65 6 3 653 l irs 4188 14 0 3 649 of Next Tube i Acquire Data OB Record Data Restart al o Acquisition Setup Stopping Gate o All Events w Events To Record 10000 evt w Stopping Time sec Storage Gate Bb i Events w Events To Display 1000 evt v ii Ready E 00 06 10 Connected A 1 Browser 2 Acquisition Dashboard 3 Inspector 4 Global Worksheet 5 Cytometer Setting up an experiment 1 Use the browser toolbar to add an experiment red border Or in the menu bar go to Experiment gt New Experiment be Browser AE 2 Select Cytometer Settings in the browser 3B LEJ MyExperiment 1 4 07 8 54 15 Cytometer Settings h a Global Worksheets 3 Inthe Inspector window delete the parameters not needed a To delete parameters click the selection button next to each unneeded parameter Hold down the Ctrl key to select more than one When you are finished selecting click Delete Inspector Cytometer Settings Cytometer Settings Parameters Threshold Ratio Compensation Parameter voltage Log nN a gt as oO oO Fic PE PerCP Cy5 5 v El Pacific Blue AmCyan Indo 1 Violet Indo 1 Blue ae Peppi ePe ppe

Download Pdf Manuals

image

Related Search

Related Contents

Dossier Classif mode d`emploi-BiblioVie 071001  WorkAbout Pro 802.11b Compact Flash Radio Installation  Gebrauchsanleitung Manual Mode d`Emploi  ダウンロード - 中川電気工業  User`s manual - Vt.vtp  Page 1 Réfrigérateur-Congélateur C`est bon de pouvoir compter sur    Bedienungsanleitung  Alpine SWR-1542D Application Guide  A3判高遠X-Yレコ一夕  