Contents
1. DA Appendix D System Specifications D Appendix D System Specifications D 1 Power Specifications TF ergeon Console Patient Car 100 120 230 VAC 100 120 230 VAC 100 120 230 VAC Voltage 50 60Hz 50 60Hz 50 60Hz Auto Sense Auto Sense Auto Sense 1000VA Continuous 1000VA Continuous 1500VA Continuous Rating and Typical Current 8 4A at 115V 8 4A at 115V 12A at 115V 4 2A at 230V 4 2A at 230V 6A at 230V D 2 Physical Dimensions OOOO O sencon patient Cart e a ina7sem souchacreenstoned wath inom E E GCM Depth 5ain Gem E 127m 365in 627m Weight 580 bs 264k 12001bs aako 446 Tos 2023ko K9 in 8 mm D 3 Environmental Specifications Environmental Conditions Operating Temperatures 0o30 C0086 The 1S3000 shall function properly in atmospheric pressures ranging from 523mm HG 10 000 ft to 774mm HG 500 ft For every 1000 feet above sea level the 30 C operational temperature limit specified above will be reduced by 1 C For example the maximum operating temperature at 5000 feet will be 25 C and the maximum operating temperature at 10 000 feet will be 20 C Atmospheric Pressure Environmental Conditions Storage and Transport Temperature 101055 cron Humiditv 5 to 95 non condensing for transport 10 to 85 non condensing for storage o Note Storage and transport of the system are not affected by atmospheric pressure da Vinci Si
Download Pdf Manuals
Related Search
Related Contents
PeaPod Plus novembre - Chambre d`Agriculture de la Gironde Baixar Manual 【担当曜日】 第2・第4金曜日 【出身地】 神奈川県 【出身大学 HP Compaq Presario CQ56-200SI Descrizione del mobile - Cabinet description User Manual - Princess Auto La Source opale Mode d`emploi des affouages à Charrey sur Saône Copyright © All rights reserved.