Home
Monarch Thermometer 309 User's Manual
Contents
1. 4 4 13 Digital Output 4 Appendix Thermocouple Probe Specifications 5 5 0 Setup TestLink RS232 Interface Software 5 DECLARATION OF CONFORMITY Monarch Instrument Division of Monarch International Inc 15 Columbia Drive Amherst NH 03031 USA declares that the product to which this declaration relates is in conformity with the following standards and therefore conforms in accordance with 89 336 EEC EMC Directive The testing of this product was performed by GesTek EMC Lab in April of 2001 Ref No 0103105E Name Thermometer Four Channel Type K Model Monarch 309 EMC EN55022 1998 CISPR 22 Class B EN50082 1 1997 EN61000 4 3rd May 2001 Importer Amherst NH Alan Woolfson VP Engineering Authorized Signature MODEL 309 FOUR CHANNEL DATALOGGING THERMOMETER 1 1 0 Introduction This instrument is a four channel recording digital thermometer for use with any Type K thermocouple as temperature sensor Temperature indication complies with the NIST and IEC584 temperature voltage tables for Type K thermocouples Its internal memory can store up to 16312 records per channel in any number of recording intervals It is equipped with a RS232 interface for bi directional communication with a PC 2 0 Specifications Measurement Ran
2. 190 C VR2 0 1 C 1000 C VR3 1 C 1900 F VR4 1 F 4 13 Digital Output The Digital Output is a 9600 bps N 81 serial interface RX is a 5V normally high input port TX is a 5V normally high output port Appendix Thermocouple Probe Specifications Model Range Tolerances Description TP K01 Wire probe 50 C to 200 C 58 F to 392 F 2 2 C or 0 75 3 6 F or 0 75 Teflon insulation Maximum insulation temperature 260 C 500 F TP K01 Probe for general condition measurements especially for complex and hard to reach places A simple single point calibration may be performed by adjusting VR1 to 0 C with thermally stabilized ice water
3. then indicates T1 value on the top left T2 value on the bottom left and the difference T1 T2 on the bottom right 4 6 Clock Setup To set the real time clock 1 Press and hold the T1 T2 button and then power on the meter 2 Press HOLD clock 3 Press REC or C F to increase or decrease the number press HOLD to advance to the next item The adjusting order is year month day hour minute Press HOLD again after the last setting to complete the process To abort during a setup process press the button 4 7 Recording Interval Setup To set the recording interval 1 Press and hold the T1 T2 button and then power on the meter 2 Press MAX MIN interval 3 Press REC or C F to increase or decrease the number press MAX MIN to advance to the next item Press MAX MIN again after the last setting to complete the process To abort during a setup process press the button MODEL 309 FOUR CHANNEL DATALOGGING THERMOMETER 4 TX RX GND 4 8 Recording Data Each momentary press of the REC button will alternately start and stop recording To clear the memory power off the meter press and hold the REC button followed by the button holding both on simultaneously for at least 5 seconds The display will show CLR SURE 5 and releasing all buttons then clears the memory 4 9 MAX MIN Operation Press the MAX MIN button to enter the MAX MIN mode In this mode both the maximum and minimum values are simultaneously ret
4. MONARCH INSTRUMENT 15 Columbia Drive Amherst NH 03031 2334 USA Phone 603 883 3390 Fax 603 886 3300 E mail support monarchinstrument com Website www monarchinstrument com Instruction Manual Monarch 309 Datalogging Four Channel Thermometer Monarch Instrument 2002 all rights reserved 1071 8031 410 Safeguards and Precautions 1 Read and follow all instructions in this manual carefully and retain this manual for future reference 2 Do not use this instrument in any manner inconsistent with these operating instructions or under any conditions that exceed the environmental specifications stated 3 Making measurements of high or low temperature can be dangerous Keep the hand holding the temperature probe well away from the object being measured 4 This instrument is not user serviceable For technical assistance contact the sales organization from which you purchased the product or Monarch Instrument directly LIMITED WARRANTY SELLER warrants hardware products to be free from any defect in materials or workmanship for a period of one 1 year from date of shipment to BUYER SELLER s entire liability and BUYER s sole and exclusive remedy resulting from any defect in workmanship or material in the hardware product covered by this limited warranty shall be limited to and fully discharged by the SELLER s option of replacement or repair of such item without charge The limited warranty provided in thi
5. ained in memory and updated with every new data sample The instrument first enters the MAX mode and the MAX symbol and value are both displayed Pressing MAX MIN again advances the display to the MIN symbol and value The next press of the MAX MIN button will cause both the MAX and MIN symbols to flash This indicates that the maximum and minimum values have been updated in memory and the displayed reading is the present temperature Each successive press of the MAX MIN button circulates the display mode among these options To exit the MAX MIN mode press and hold the MAX MIN button for two seconds In the MAX MIN mode the C F button is disabled as indicated by two consecutive beeps if pressed 4 10 Auto Power Off By default the instrument powers on in the Auto Power Off mode and will automatically shut off 30 minutes after the last key operation or RS232 communication To disable this feature press and hold the HOLD button and then power on the meter Two successive beeps will indicate that Auto Power Off is disabled and the will not be displayed 4 11 Low Battery Condition When the battery voltage is at or below the minimum for proper operation the symbol will show on the display indicating that the battery must be replaced 4 12 Calibration Calibration is conducted by adjusting the following potentiometers to within the tolerances shown Ambient Temperature 23 3 C Input Adjust VR Tolerance 0 C VR1 0 1 C
6. ed into the sockets marked T1 through T4 taking care to observe the polarity of the thermocouple pins If a differential measurement is required this is available only between T1 and T2 which determines which thermocouples should be connected to these locations 4 2 Power and Backlight The button turns both the thermometer and the display backlight ON or OFF Press it once to turn the thermometer ON Every subsequent momentary press toggles the backlight ON or OFF Press and hold this button for 3 seconds to turn the power OFF When first powered on the display will show how much memory space is available for use as shown to the right 4 3 Selecting the Temperature Scale When first turned on the instrument defaults to reading in Celsius C The instrument toggles between Celsius C and Fahrenheit F each time the C F button is pressed The instrument remembers the scale setting when last turned off and powers on in that setting the next time 4 4 Data Hold Operation The present reading is held on the display by pressing the HOLD button Pressing the HOLD button again releases the hold function and returns the instrument to continuous reading In the HOLD mode the T1 T2 MAX MIN and C F buttons are disabled as indicated by two continuous beeps when pressed 4 5 T1 T2 Operation When the T1 T2 button is pressed the display will indicate 1 2 indicating the instrument is in the differential mode of T1 minus T2 The display
7. ge 200 C 1370 C 328 F 2498 F Accuracy At ambient of 23 5 C Range Accuracy 200 C 200 C 0 2 reading 1 C 200 C 400 C 0 5 reading 1 C 400 C 1370 C 0 2 reading 1 C 328 F 200 F 0 5 reading 2 F 200 F 200 F 0 2 reading 2 F 200 F 2498 F 0 3 reading 2 F Note The basic accuracy specification does not include the error of the probe Please refer to the probe accuracy specification for additional details Resolution 200 C 200 C 0 1 C 200 C 1370 C 1 C 200 F 200 F 0 1 F outside this range 1 F Number of Sensor Inputs Four Type K Thermocouples Input Protection at Thermocouple Input 60V DC or 24Vrms AC Display Four 4 Digit Temperature Displays plus Indicators see Section 3 0 Switchable backlight Sample Rate 3 seconds per period Time Function Real time clock see Section 4 6 Datalogging Capacity 16 312 records per channel Recording Interval User adjustable see Section 4 7 Digital Output Bi directional RS232 Software and Cable included Power requirement 9 Volt Battery NEDA 1604 or JIS 006P or IEC6F22 Battery Life Approx 100 hours with alkaline battery Low battery indication Operating Conditions Operating Temperature and Humidity 0 C 50 C 32 F 122 F 0 80 RH Storage Temperature and Humidity 10 C 60 C 14 F 140 F 0 80 RH Altitude Up to 2000 meters 6500 feet Dimensi
8. o BUYER at law or in equity as the result of such breach or default regardless of whether based on contract tort strict liability or other theory of liability IN NO EVENT SHALL SELLER BE LIABLE FOR ANY SPECIAL INDIRECT INCIDENTAL CONSEQUENTIAL OR PUNITIVE LOSSES OR DAMAGES INCLUDING BUT NOT LIMITED TO LOSSES OR DAMAGES FOR ANY LOST PROFITS OR LOST DATA AS THE RESULT OF ANY BREACH OR DEFAULT BY SELLER WITH RESPECT TO THE HARDWARE OR SOFTWARE EVEN IF SELLER HAS BEEN ADVISED OR MADE AWARE OF THE POSSIBILITY OF ANY SUCH LOSSES OR DAMAGES AND REGARDLESS OF WHETHER THE CLAIM IS BASED ON CONTRACT TORT STRICT LIABILITY OR OTHER THEORY OF LIABILITY This limited warranty does not extend or apply to consumables including but not limited to lamps and batteries if applicable or equipment instruments or accessories which are warranted separately by the original manufacturer of these items MODEL 309 FOUR CHANNEL DATALOGGING THERMOMETER Table of Contents Section Page 1 0 Introduction 1 2 0 Specifications 1 3 0 Symbol Definitions and Feature Locations 2 4 0 Operating Instructions 3 4 1 Prepa
9. ons 184 64 30 mm 5 4 2 5 1 2 in Weight Approx 250g 8 7 oz Accessories Two Type K Wire Thermocouple Probes Battery Carrying Case Instruction Manual Software RS232 Cable Option AC Adapter 9VDC 15 100mA Plug Diameter 3 5mm 1 35mm Temperature Coefficient For ambient temperatures from 0 C 18 C and 28 C 50 C for each C ambient below 18 C or above 28 C add the following tolerance into the accuracy spec 0 01 of reading 0 03 C 0 01 of reading 0 06 F Notes MODEL 309 FOUR CHANNEL DATALOGGING THERMOMETER 2 3 0 Symbol Definitions and Feature Locations Indicates Auto Power Off is enabled T1 T3 T2 T4 Indicates the value below is theT1 T2 T3 or T4 temperature sensor C F Centigrade or Fahrenheit indication Negative temperature indication Low battery indication Battery voltage is not sufficient for proper operation Thermocouple type indication Indicates that the instrument is recording When flashing the memory is full The Maximum value is now being displayed The Minimum value is now being displayed Indicates that the display data is being held Indicates the value below is T1 T2 Feature Locations 1 T1 through T4 Type K temperature sensor input connectors 2 LCD display 3 ON OFF amp Backlight button 4 Hold button 5 Record button 6 T1 T2 button 7 MAX MIN function control button 8 C F control button 9 Offse
10. ration for Measurement 3 4 2 Power and Backlight 3 4 3 Selecting the Temperature Scale 3 4 4 Data Hold Operation 3 4 5 T1 T2 Operation 3 4 6 Clock Setup 3 4 7 Recording Interval Setup 3 4 8 Recording Data 4 4 9 MAX MIN Operation 4 4 10 Auto Power Off 4 4 11 Low Battery Condition 4 4 12 Calibration
11. s clause is in lieu of all other warranties expressed or implied arising by law or otherwise ALL IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE EXCLUDED This limited warranty shall not be modified except by an arrangement signed by both parties specifically referencing this clause SELLER warrants that any software supplied will operate in accordance with the documentation or manual supplied therewith in all material respects when used in strict compliance with such documentation or manual Notwithstanding the foregoing BUYER acknowledges that since software is complex and therefore may have defects BUYER s sole and exclusive remedy for any such defects or breach of this warranty shall be to require SELLER within a reasonable period of time to provide all reasonable programming services to correct programming errors in the software Except as provided above SELLER MAKES AND BUYER RECEIVES FROM SELLER NO EXPRESS OR IMPLIED WARRANTIES OF ANY KIND WITH RESPECT TO ALL OR ANY PORTION OF SOFTWARE AND BUYER HEREBY AGREES AND ACKNOWLEDGES THAT IT ACCEPTS THE SOFTWARE IN AS IS CONDITION SELLER HEREBY EXPRESSLY EXCLUDES ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE WITH RESPECT TO THE SOFTWARE BUYER agrees that any specific right or remedy granted to BUYER hereunder with respect to any breach or default by SELLER shall be in lieu of all other rights and remedies otherwise available t
12. t calibration screw 10 Digital output connector RS232 11 AC power adapter connector 12 Tripod connector 13 Battery cabinet cover T2 24V MAX 60V T1 T4 T3 3 4 5 2 1 OUTPUT DC9V 6 7 8 CAL 9 MANUAL FOR SAFETY OPEN NEDA 1604 6F22 006P PLEASE READ 9V BATTERY HOLD C F T1 T2 SETUP INTV CLOCK 2498 1370 RANGE REC 200 328 POWER UP OPTIONS F C F C REC MODEL 309 FOUR CHANNEL DATALOGGING THERMOMETER 5 5 0 Setup TestLink SE 309 RS232 Interface Software The TestLink package contains 1 One 80 mm CD 2 Custom designed RS232 cable for TestLink System Requirements Windows 95 Windows 98 or Windows NT 4 0 Minimum Hardware Required Pentium 90MHz PC 32 MB RAM At least 5 MB byte hard disk space available to install TestLink Recommended display resolution is 800X600 Install TestLink 1 Close all other application before installing TestLink 2 Insert setup CD disk in CD disk drive 3 Choose the Start button on the Taskbar and select Run 4 Type E SETUP and choose OK to copy SE309 exe executable file and Help file to your hard disk default is C program files TestLink SE309 For other operating instructions please refer to the online help while executing TestLink MODEL 309 FOUR CHANNEL DATALOGGING THERMOMETER 3 4 0 Operating Instructions 4 1 Preparation for Measurement Plug as many thermocouples as are to be monitor
Download Pdf Manuals
Related Search
Related Contents
Vulkan TU Engleska.cdr Manuel d`utilisation - Collection Tuscany Jet 2013 Samsung 14MP User's Manual DOC TECHNIQUE ( 848 Ko) Nokia 6210 Navigator Manual do Usuário SV470-600 - Kohler Engines Viessmann Bedienunsganleitung IT PPF 631 PPF 632 Manuale d`istruzioni CS 299 & M XL ADC-9033 User Manual Copyright © All rights reserved.