1. S BASIC AMERICAN MEDICAL PRODUC ZENITH APS SERIES NOTHING ON STAFF CONTROL WORKS BUT PENDANT WORKS a Check the Staff Control connections at the foot end Is the Staff Control Cable plugged securely into the T Cable Also check if the other end of the T Cable is plugged securely into the Control Box or Underbed Light If the STAFF CONTROL STILL DOESN T WORK go to step b Unplug the Power Cord from your wall outlet Remove the Cable Cover Shroud on the inside of the footboard and locate the single terminal end that plugs directly into the back of the Staff Control Panel Make sure it is properly seated into the panel Plug in the power cord and test the staff control If the STAFF CONTROL STILL DOESN T WORK Replace the Staff Control Assembly See page 29 for dis assembly See page 32 for order information TROUBLESHOOTING PENDANT amp DOUBLE T CABLE If orange lock icon lit press key symbol and function to unlock A 1 STAFF CONTROL WORKS BUT PENDANT HAND CONTROLLER FUNCTION DOESN T a Check to see if the functions are locked out on the Staff Control panel if orange lock icons show on panel then that function is locked out Press the Key and Function i e Head button at the same time until the orange lock icon goes out Test the function If the PENDANT STILL DOESN T WORK go to step b Check the connections at the seat pan Pendant Hand Controller to T Cable and Staff Control Cabl
2. TROUBLESHOOTING REPLACING YOUR APS STAFF CONTROL REMOVING AND REPLACING THE STAFF CONTROL ASSEMBLY UNPLUG THE POWER CORD FROM THE WALL OUTLET Staff Control Assembly 2 Unscrew the four Phillips Screws 1 that attach the Shroud cable cover to the footboard and set aside 2 3 At the foot end of the bed unscrew the retaining S cap on the T Cable 3 and unplug the two cables Screws gt 7 NN Foot Board with cutout 4 Unscrew the two remaining Phillips Screws 2 from the inside of the footboard that attach the Staff Control Assembly bevel to the board Set aside screws for later re assembly 2 Staff Control Bevel Screws 5 Remove the OLD Staff Control Assembly and fish the old cable out through the large hole 1 Shroud Screws 6 Feed the cable of the NEW Staff Control Assembly from the outside through the large hole and insert the bevel into the cutout on the outside of the footboard 7 Secure the Staff Control by re inserting the two Phillips Screws 2 you removed in Step 4 IMPORTANT Run a Cable Tie through the end hole in the mounting tube loop the excess cable run the tie around cable and pull tight to secure Cut off excess cable tie 8 Re position the Shroud on VP the inside of the footboard and secure with the four Phillips Screws 1 you removed in Step 2 CO Plug the T Cable 5 into the Staff Control Cable 4
3. making sure the phone jack plug is seated correctly into the Staff Control cable plug Screw on the retaining cap 3 to secure 10 Run a Cable Tie through the large end hole on the right hand footboard mounting tube Female end of Staff Control Cable 11 Loop the excess Staff Control Cable Run the cable tie around the cable and pull tight to secure Cut off any excess Cable Tie End of Double T Cable at foot end Grey N T runs on either side of Seat Pan for attaching your Hand Controller Locking End Cap Basic American Medical Products a division of GF Health Products Inc www grahamfield com 999 0831 190A Service Manual 4 z E BASIC AMERICAN MEDICAL PRODUCTS RECOMMENDED MAINTENANCE AND INSPECTION gt Maintenance Inspection of All Components Receipt of shipment Make sure all parts components are included Please see page 9 Unpacking Your Bed Check all bed components for obvious damage nspect the Power Supply Cord for any cuts and or damage Check to see all actuator cords are routed and connected properly to the control box gt Mechanical Inspection of Assemblies 6 months nspect all welds on the sleeping surface frame and base assemblies for stress fractures nspect all fasteners for wear and looseness IMPORTANT Lubricate all pivot points actuator motor clevis pins and control arm clevis pins as needed White Lithium Grease is reco
4. GF Health Products Inc When operating the High Low Knee or Back Functions of the bed ALWAYS ensure that the confined individual is positioned properly within the confines of the bed DO NOT let any extremities protrude over the side or between the bed rails when performing these functions DO NOT lower the bed when objects are beneath it Failure to inspect under the bed can result in damage to property or personal injury The bed s Pendant Cord MUST BE ROUTED AND SECURED PROPERLY to ensure it does not become entangled and eventually severed during use Also make sure all electrical cords DO NOT get tangled around the bed side rails or legs during transport or normal operation of the bed When using nasal or masked type administering equipment all oxygen or air tubing MUST BE ROUTED AND SECURED PROPERLY to ensure that the tubing does not become entangled and eventually severed during the normal operation of the bed Basic American Medical Products a division of GF Health Products Inc A A A PF The bed should ALWAYS be left in its lowest position when unattended to reduce the risk of injury while getting in or out of the bed Keep all moving parts free of obstructions 1 e blankets sheets heating blankets pads wiring etc DO NOT use the assist devices as push handles for moving the bed Assist devices can be deformed or broken if excessive side pressure is exerted Assist devices are not meant for patie
5. Position 2 mounting tubes on the outside of the headboard with L facing inward Align the top hole of the mounting tubes with the top holes in the headboard Insert a 40mm hex drive bolt through each of the top holes and bottom holes and screw into each insert Tighten with the Hex Allen wrench included in your kit e Slide the L portions of the Mounting tubes into the hollow ends of the main frame rails at the head deck end FOR 80 BEDS Slide the mounting tubes in until the 1ST hole in the tubes lines up with the 1ST hole in the rails See sample below FOR 76 BEDS Slide the mounting tubes in until the 1ST hole in the tubes lines up with the 2ND hole in the rails See sample below HEADBOARD CENTER OF BED Inserts a oninside 40mm Bolts p 70 hole position 80 hole 2nd hole in position 1st hole in Basic American Medical Products a division of GF Health Products Inc 2 FOOTBOARD INSTALLATION e The footboard comes with four pre installed inserts consider this the outside of the board e Position 2 mounting tubes on the inside of the footboard with L facing outward e Align the top hole of the mounting tubes with the top holes in the footboard e Insert a 40mm hex drive bolt through each of the top holes and bottom holes and screw into each insert Tighten with the Hex Allen wrench included in your kit e Slide the L portions of the Mounting tubes
6. FOOT DECKS EDICAL PRODUC 4X 100 4752 002 SS Round Plugs Caps sides of Foot Deck Only 12X 100 4200 004 Square Caps in o EE nm D 2X 100 4700 018 he e Rectangular End Caps ik 4X 100 4715 011 E 4 i a Round End Caps CAPS OR PLUGS ARE NOT COVERED UNDER WARRANTY PACKS OF 12 CAN BE ORDERED IN THE EVENT YOU DAMAGE OR LOSE ANY CAPS PLUGS ON YOUR BED SEE BOTTOM OF PAGE 32 FOR ORDER NUMBERS Basic American Medical Products a division of GF Health Products Inc www grahamfield com 999 0831 190A Service Manual WY E S BASIC AMERICAN ZENITH APS SERIES zm NOTES Basic American Medical Products a division of GF Health Products Inc www grahamfield com 999 0831 190A Service Manual GRAHAM FIELD 336 Trowbridge Dr Fond du Lac WI 54937 Customer Service 1 800 365 2338 e Fax 920 929 8213 www grahamfield com O 2012 GF Health Products Inc All Rights Reserved Basic American Medical Products and Zenith are registered trademarks of GF Health Products Inc GF121699RevA12
7. bed components for obvious damage e Inspect the Power Supply Cord for any cuts and or damage e Check to see all actuator motor cables are routed and connected properly to the control box Cut and discard cable tie Basic American Medical Products a division of GF Health Products Inc DISCARD 1 Cardboard Blocks 2 Large Block 3 Large Cable Tie CUT 4 Notched 2 x 4 Board 5 Foam Wrapping 6 End Caps with Foam 7 Notched Leg Foam 8 Small Cable Tie CUT www grahamfield com NOTE Underbed Light is optional and available on APS981741 Models Underbed Light and Mounted Battery optional combination is available on APS981742 Models ZENITH APS SERIES KEEP 9 Wireform Wallsaver 10 Two Mattress Retainers 11 Service Manual Documents 12 Wallsaver Adaptor Brackets for Trendelenberg and Reverse Trendelenberg Positioning 13 Pendant Holster attached to pendant NOTE END OF POWER CABLE IS COILED FOR SHIPPING AND TIED WITH CABLE TIE TO GRID WIRE ALONG WITH PENDANT PENDANT CABLE amp HOLSTER PLEASE CUT AND DISCARD CABLE TIE AROUND CABLES WHEN YOU UN PACK YOUR BED ALSO DISCARD TOP AND BOTTOM CARTONS eg 999 0831 190A Service Manual WY BASIC AMERICAN m MEDICAL PRODUCTS ZENITH APS SERIES HEAD AND FOOTBOARD ASSEMBLY 1 HEADBOARD INSTALLATION e The headboard comes with four pre installed inserts consider this the inside of the board
8. pull out the black knob on the outside of the pivot assembly with the other hand to release the locking mechanism For ease of patient access it is recommended that you pivot the rail toward the headboard until it stops and rests in place Recommend Pivot Toward poo 2 To set the assist bar back to its upright locked position grab the top of the assist bar with one hand and pivot the assembly upward until the black knob mechanism Black Knob to snaps into place locking the assist bar Lock or Release vertically 17 00 HEAD SLEEP SURFACE DECK END FOR YOUR SAFETY MAKE SURE YOUR FINGERS PLEASE MAKE SURE THE PIVOT ASSIST BAR ARE CLEAR OF THE SIDES OR UNDERNEATH THE IS ALWAYS IN THE FULL VERTICAL LOCKED PIVOTING ASSEMBLY WHEN PIVOTING THE POSITION WHEN RESIDENT PATIENT IS LEFT ASSIST BAR UP OR DOWN UNATTENDED Basic American Medical Products a division of GF Health Products Inc www grahamfield com 999 0831 190A Service Manual WY SE BASIC AMERICAN gt MEDICAL PRODUCT Ay 3 ZENITH APS SERIES OPTIONAL TRAPEZE SUPPORT INSTALLATION The Trapeze Support is attached to the head section of the bed Please Order Kit Number ZA79000 Use third hole from the end of the main frame tubes Begin by removing the headboard panel assembly 2 if previously assembled to the bed by pulling out the two Clevis Pins and two Hair Pin Clips that attach the main frame tubes to the headboard
9. z BASIC AMERICAN EDICAL PRODUC ZENITH APS SERIES APS9000 SERIES ELECTRICAL siio d STANDARD POWER CABLE T CABLE HEAD CABLE FOOT CABLE HI LO CABLE FOOT HI LO CABLE HEAD BATTERY BACKUP CABLE GROUND WIRE ooo 9000 006 Portable Battery Cable Extension Ground Wire Power Cable T Cable HB Hi Lo Ft Port 4 Foot Port 3 Hi Lo Hd Port 2 Head Port 1 Power Cable amp Battery Cable Side of Control Box Basic American Medical Products a division of GF Health Products Inc 0 im E AIN sa AME MEM DENM RENE GENS Gum T 1 d KK 4 a CTL n cm Lr 3 a a E m rey 5 v Ba me ri Fir Un Ola ate gt i HU nA Ci bea n es HT www grahamfield com 999 0831 190A Service Manual NG 2 gt AUT BASIC AMERICAN MEDICAL PRODUCT ZENITH APS SERIES APS9000 SERIES ELECTRICAL CABLING WITH UNDERBED LIGHT oo O O 000006 Basic American Medical Products a division of GF Health Products Inc POWER CABLE T CABLE HEAD CABLE FOOT CABLE HI LO CABLE FOOT HI LO CABLE HEAD BATTERY BACKUP CABLE GROUND WIRE UNDERBED LIGHT CABLE UBL Portable Battery Cable Extension Power Cable O O Q Jc xD o O S Es eo gt T Cable Hi Lo Ft Port 4 in Control Box Foot
10. END LONG T OF T CABLE AND BATTERY CABLE SECURED TO ROD AND MAIN FRAME BRACKET WITH CABLE TIES FOOT MOTOR E ZA Na gt ie SHORTER SIDE OF T CABLE WITH PUSH CABLE TIE IN SMALL HOLE CABLE PROTECTION GROMMET IN SEAT PAN KEY HOLE POWER CABLE gt PLUG AT HEAD END POWER CABLE SECURED TO ROD WITH CABLE TIES 1 TO ORDER A REPLACEMENT 2 TO ORDER A REPLACEMENT 3 TO ORDER A REPLACEMENT HI LO MOTOR USE SERVICE PACK HEAD MOTOR USE SERVICE PACK FOOT MOTOR USE SERVICE PACK 999 0831 051SP Includes motor 999 0822 052SP Includes motor 999 0822 053SP Includes motor mounting hardware amp installation mounting hardware amp installation mounting hardware amp installation instruction sheet instruction sheet instruction sheet 4 TO ORDER A REPLACEMENT 5 TO ORDER A REPLACEMENT 6 TO ORDER A REPLACEMENT CONTROL BOX PLEASE USE DOUBLE T CABLE USE SERVICE BATTERY MOTOR USE SERVICE 999 0831 300 Includes Control PACK 999 0806 200SP Includes PACK 999 0711 001SP Includes Box and mounting clip Power T Cable cable ties grommets amp battery cable long cable ties and Cord must be ordered separately installation instruction sheet installation instruction sheet see page 31 for order numbers WHEN ORDERING REPLACEMENT PARTS WITH CUSTOMER SERVICE PLEASE HAVE YOUR BED S SERIAL NUMBER AVAILABLE TO CONFIRM WHETHER THE PART IS COVERED UNDER WARRANTY SEE PAGE 5 FOR LOCATION OF SERIAL NUMBER ID LABEL Basic Ame
11. IS PINCHED With cord unplugged 3 Prong Power Cable 2 Prong Power Cable try to move the bed part slightly to release the pinched 999 0775 208 999 0775 206 cord If you can release the cord re plug the power cord into the wall outlet and test the bed BED WORKS Make sure the cord is not frayed or exposed If it is OK you should not have to replace POWER CORD IS DAMAGED Cut cable ties and immediately replace Power Cord ALWAYS UNPLUG THE POWER CORD BEFORE DOING IF 3 PRONG OR OPTIONAL 2 PRONG POWER CORD ANY ELECTRICAL WORK ON YOUR BED IS DAMAGED REPLACE IMMEDIATELY TROUBLESHOOTING STAFF CONTROL BLINKING LIGHTS 1 STAFF CONTROL LIGHTS ARE BLINKING C If you have a spare staff control assembly plug the NOTHING ON STAFF CONTROL WORKS cable into the T Cable at the foot end and test BUT PENDANT WORKS WORKS AND LIGHTS NO LONGER BLINK Replace the Staff Control Assembly in your foot board See page 29 for dis assembly Please see page 32 for order number a This could mean that your staff control is not getting enough power from the Control Box b Unplug the power cord from the wall outlet Unplug the staff control cable from the T Cable at the foot DOESN T WORK amp LIGHTS STILL BLINKING end of the bed Replace the Control Box Please see Page 31 for order number Basic American Medical Products a division of GF Health Products Inc www grahamfield com 999 0831 190A Service Manual WY e
12. Inspection LL 30 Replacement Service Part List 31 32 Service Part Diagrams 33 36 LABEL SYMBOL DEFINITIONS Oo P A A Consult Safe Double Accompanying Working Insulated Documents Load Basic American Medical Products a division of GF Health Products Inc Protected Grounded Device www grahamfield com R Type B Equipment amp Applied Parts ZENITH APS SERIES 999 0831 190A Service Manual gt Fr PPP P FP BASIC AMERICAN MEDICAL PRODUCT ZENITH APS SERIES IMPORTANT SAFETY AND WARNING INFORMATION This product is a variable height adjustable mattress platform which will provide comfort and convenience for residents patients and caregivers in long term care settings The MAXIMUM SAFE WORKING LOAD for the APS 9000 Bed with weight evenly distributed including bedding resident patient support surface and all accessories is 600 lbs NEVER operate the bed if a Power Cord or Plug is damaged or not working properly Contact qualified Service Personnel for examination and repair Always unplug the Power Cord when performing any mainte nance on the bed DO NOT open assemblies such as the Actuators Hand Control or Control Box If unauthorized personnel perform work on these components the manufacturer s warranty becomes void DO NOT use unauthorized parts accessories or adaptors other than those specified authorized by
13. K e _ BASIC AMERICAN ZENITH MEDICAL PRODUCTS SERIES INSTRUCTIONS FOR USE SERVICE MANUAL To avoid personal injury or damage to bed please read all sections pertaining to your bed model before use ETL UL CSA APPROVED UL 60601 1 IEC 60601 2 38 CSA 22 2 601 1 Basic American Medical Products a division of GF Health Products Inc www grahamfield com Zenith APS Service Manual 999 083 1 190A This service manual covers the following APS Models 76 amp 80 Full Electric with Pedal Lock Caster System amp Grid Decks APS981 74 76 amp 80 Full Electric with Pedal Lock Caster System Grid Decks and Underbed Light APS981741 76 amp 80 Full Electric with Pedal Lock Caster System Grid Decks Underbed Light and Onboard Battery Backup System APS981742 WY gt dm WT BASIC AMERICAN MEDICAL PRODUCTS Basic American Medical Products 336 Trowbridge Drive Fond du Lac WI 54937 For APS 9000 Bed Service Parts please contact our Customer Service Department at 1 800 365 2338 GF Health Products Inc 2935A Northwest Parkway Atlanta GA 30360 www grahamfield com To order an APS 9000 Bed please contact a Graham Field Sales Representative at 1 800 554 9215 IMPORTANT NOTICE GF Health Products Inc is not responsible for typographical errors All illustrations specifications packaging and warranties contained in this Service Manual are based on the latest product information avai
14. L A EF 8 4 BB A D gu 9 p Press Underbed Light button 14 to turn underbed light on or off Press CPR Heart button 7 to Level Flatten the bed horizontally for CPR position Safe Mode 15 LED light will glow green automatically once power cord is plugged in Green light will blink intermittently for approximately 10 minutes After 10 minutes of non activity Hi Lo Lock icons on the nurse staff control panel will show orange locked LOCK SYMBOLS LOCK SYMBOL LOCK SYMBOLS 5 11 Press Head Knee or Chair and Key at the same time to unlock individual lock out functions Press Hi Lo Locked to lock out Hi Lo Trendelenberg and Reverse Trendelenberg functions o 8 Press Hi Lo Locked and Key at the same time LOCK SYMBOLS to unlock all Hi Lo lock out functions po LOCK SYMBOLS LOCK SYMBOL LOCK SYMBOLS v 4 L k SE EY a S P j HI LO Press All Locked and Key at the same time to unlock all functions LOCKED TLOCKED The LED will be lit green when Safe Mode automatically engages when plugging in the bed Press Safe Mode and Key to disable LED the Safe Mode feature Basic American Medical Products a division of GF Health Products Inc www grahamfield com ALL LOCKED BUTTON HEAD DECK UP BUTTON HEAD DECK DOWN BUTTON KNEE amp FOOT DECK UP BUTTON NEE amp FOOT DECK DOWN BU
15. LABLE TO CONFIRM WHETHER THE PART IS COVERED UNDER WARRANTY SEE PAGE 5 FOR LOCATION OF SERIAL NUMBER ID LABEL Basic American Medical Products a division of GF Health Products Inc www grahamfield com 999 0831 190A Service Manual ZENITH APS SERIES SERVICE REPLACEMENT PART DIAGRAM MAIN FRAME CAPPING EDICAL PRODUCT WY amp BASIC AMERICAN a7 o U j DETAIL A SCALE 1 5 2x 100 4738 005 Round Flug Cap 2X 100 4700 0186 Rectangular Cap EX 1004700017 Rectangular t x HEAD END EL Vw pee ae a AG STE EP F i i 5 it iB Ni IN K i Ka B MM i 3 r a a a 2 i ni 1 i wd n i i p M J i NG z he zs l EX JE 6 3 Ita J Jy 1 m 1 n I y j ES TET Bo 0 E b O a I aoe x Cs E a gh a n zr a et ey Se ERE r a p lt MM mm vr Pr T 2 a Main Frame Rail Foot End CAPS OR PLUGS ARE NOT COVERED UNDER WARRANTY PACKS OF 12 CAN BE ORDERED IN THE EVENT YOU DAMAGE OR LOSE ANY CAPS PLUGS ON YOUR BED SEE BOTTOM OF PAGE 32 FOR ORDER NUMBERS o MEM Basic American Medical Products a division of GF Health Products Inc www grahamfield com 999 0831 190A Service Manual WY as BASIC AMERICAN ZENITH APS SERIES SERVICE REPLACEMENT PART DIAGRAM CAPPING ON HEAD KNEE AND
16. Port 3 in Control Box Hi Lo Hd Port 2 in Control Box Head Port 1 in Control Box Port 2 in Underbed Light Box Power Cable amp Battery Cable Side of Control Box Light NOTE Cabling for Beds with Underbed Light feature is identical to Standard APS beds except that the T Cable plugs into Port 2 in the Underbed Light Box and an extra Extension Cable is plugged into Port 1 on the Underbed Light Box and the HB Port in the Control Box EA l I z mer cE KE m 1 pi dL UNDERBED 8 8 LIGHT UBL E www grahamfield com 999 0831 190A Service Manual PE BASIC AMERICAN ZENITH APS SERIES EDICAL PRODUC APS9000 SERIES ELECTRICAL CABLING WITH UNDERBED LIGHT amp BATTERY POWER CABLE e e T CABLE HEAD CABLE FOOT CABLE HI LO CABLE FOOT HI LO CABLE HEAD BATTERY BACKUP CABLE GROUND WIRE UNDERBED LIGHT CABLE UBL O O OS 0000906 Mounted Battery Cable Power Cable Mi ot IL Ll rv O O 03 L D an s A m C ES I T Underbed Light UBL Attached ee NOTE Cabling for Beds with Underbed Light and Mounted aat abie Battery Backup features are identical to Standard APS beds except that the T Cable plugs into Port 2 in the Underbed Light Box and an extra Extension Cable is plugged into Port 1 on the Underbed Light Box and the HB Port in the Control Box MOUNTED BATTERY A Battery Cable runs from
17. SAME FEATURES BUT WITHOUT UNDERBED LIGHT BUTTON OPERATION 1 To angle the Head Deck upward PRESS 1 2 To angle the Head Deck downward PRESS 2 3 To raise the Bed Up horizontally PRESS 3 4 To lower the Bed Down horizontally PRESS 4 5 To angle the Knee Foot Decks upward PRESS 5 6 To angle the Knee Foot Decks downward PRESS 6 T To put the Bed in the Chair position PRESS 7 c 8 To release the Bed from the Chair position PRESS 8 m 9 To turn the underbed light on or off optional pendant only 8 PRESS THE LIGHT BULB ICON near the bottom of the hand controller not shown on sample at left Back of both Pendants include 9 LAK GB attachment gt hook BACKLIT PENDANT i Non Backlit Pendant meea lg Model H00HEBATA D KAR and includes Part 999 0806 305 Comes Underbed Light Button x without Underbed standard E Light Button Wine button 9 g Eod NOTE THE ZENITH 9000 APS BED COMES STANDARD WITH A FOOTBOARD NURSE STAFF CONTROL PLEASE MAKE SURE TO ORDER WITH APPROPRIATE BOARD STYLE THE BACKLIT PENDANT WITH UNDERBED LIGHT BUTTON AND THE UNDERBED LIGHT CAN BE PURCHASED AS OPTIONS PLEASE SEE NEXT PAGE FOR NURSE STAFF CONTROL OPERATING INSTRUCTIONS Basic American Medical Products a division of GF Health Products Inc www grahamfield com 999 0831 190A Service Manual WY BASIC AMERICAN MEDICAL PRODUCTS AUT ZENITH APS SERIES BED OPERATIONS NURSE STAFF CONTROL PANE
18. TING MOTOR CABLES amp CONTROL BOX 1 HEAD FOOT OR HI LO FUNCTION NOT WORKING a Check plugs to make sure they are firmly seated in all ports on your Control Box and are plugged into the correct ports POWER PLUG See diagram at right b Check all cables to make sure they are not frayed pinched or damaged in any way If any cable is damaged UNPLUG YOUR BED S POWER CORD FROM THE WALL OUTLET and REPLACE THE CABLE AND OR THE ELECTRONIC COMPONENT cu mom BATTERY c HEAD DECK NOT RAISING Unplug the power cord Switch the head and foot plugs ports 1 and 3 Re plug the power cord and press the FOOT button to test HEAD WORKS Control Box head port bad Replace box See page 31 for order information DOESN T WORK Replace the Head Motor See page 31 for order information FOOT HEAD HILO HEAD d FOOT AND KNEE DECKS NOT RAISING Unplug the power cord Switch the foot and head plugs ports 3 and 1 Re plug the power cord and press the HEAD button to test FOOT WORKS Control Box foot port bad Replace box See page 31 for order information T CABLE GE He GRE FOOT HILO amp 2 BED LOOKS UNLEVEL HILO MOTORS ARE NOT IN SYNC a Press the Hi Lo DOWN button repeatedly on either the Staff Control or Pendant Hand Controller until e BED DOES NOT GO UP AND DOWN IN HORIZONTAL the bed goes completely down and rests in its full POSITION If you press the Up or Down Hi Lo button and horiz
19. TTON CHAIR POSITION BUTTON CPR UNDO CHAIR POSITION BUTTON EY LOCK UNLOCK BUTTON HI LO T TR LOCKED BUTTON I LO UP BUTTON ILO DOWN BUTTON TRENDELENBERG REVERSE TRENDELENBERG UNDERBED LIGHT BUTTON SAFE MODE RELEASE BUTTON 1 X KCN KO EC LOCK OUT SINGLE FUNCTIONS To individually lock out the Head Knee Chair and Hi Lo functions press the appropriate top icon 22 3 4 10 or 12 button and the Key button at the same time An orange LED lock symbol will appear under the related icon To Unlock any of the individual functions press the top icon 2 3 4 10 or 12 and the Key Button 8 simultaneously LED lights will not show up LOCK OUT HI LO FUNCTIONS To lockout the functions for raising and lowering the entire bed and tilting the bed for Trendelenberg positions press the Hi Lo Locked 9 button Orange LED lock symbols will appear under the icons To Unlock all Hi Lo functions press the Hi Lo Locked button 9 and the Key button 8 simultaneously Orange LED lights will not show up LOCK OUT EVERYTHING To lock out all functions press the All Locked button 1 Orange LED lock symbols will appear under the Head Knee Chair Hi Lo and Trendelenberg Reverse Trendelenberg icons To Unlock all functions press the All Locked button 1 and the Key button 8 simultaneously Orange LED ligh
20. Up and Full Down positions WARRANTY 15 Years on Frame 5 Years on Motors and Control Box longest motor warranty in the industry Years on Other Electronics The warranties contained herein contain all the representation and warranties with respect to the subject matter of this document and supersede all prior negotiations agreements and understanding with respect thereto The recipient of this document hereby acknowledges and represents that it has not relied on any representation assertion guarantee warranty collateral contract or other assurance except those set out in this document Basic American Medical Products a division of GF Health Products Inc www grahamfield com ZENITH APS SERIES 999 0831 190A Service Manual fx BASIC AMERICAN ZEN ITH APS SERI ES MEDICAL PRODUCT N SERVICE REPLACEMENT PART RECORD Page 1 DECKING Pari Number Description amv orderbate Proe Weossn _ eaBeck eck HaronaroPack wa OOo CU a vem CU aa vararoPack ek a Pad Monupgs LIP 5 16 Retaining Rings Deck Brackets MOTORS ACTUATORS Pari Number Deeiplon amv OrderDate Price Hi Lo Motor Cable not included amp Hardware Pack tMotor fof OR Head Motor Cable not included amp Hardware Pack tMotor fof 999 0822 053SP OR Foot Motor Cable not included amp Hardware Pack 1 Motor Motor Hardware Pack INCLUDES 1 Pkg C2 A 12
21. allsaver 1 The APS 9000 wallsaver features Position w o Trapeze 80 Bed w o Trapeze Position w Trapeze 3 positions for easy reconfiguration 76 80 and Trapeze if you use Trendelenberg Reverse Trendelen berg positions please see optional adaptor brackets on next page 2 For 76 beds position the wallsaver ends in the caster base slots closest to the FOOT end of the bed For 80 beds position the wallsaver ends in the MIDDLE slots on the bases 3 If you have an optional Trapeze unit on a 76 bed position the wallsaver ends in the MIDDLE slots on the caster bases For 80 beds position wallsaver in slots closest to HEAD Basic American Medical Products a division of GF Health Products Inc www grahamfield com 999 0831 190A Service Manual A BASIC AMERICAN ZENITH APS SERIES aan MEDICAL PRODUCT OPTIONAL WALLSAVER ADAPTOR BRACKET INSTALLATION TO USE WITH TRENDELENBERG REVERSE TRENDELENBERG FEATURE 4 Secure brackets to caster bases using short Nylon Lock Nuts 3 100 6744 002 included in your adaptor bracket bag kit STEP 1 BRACKET INSTALLATION 1 Remove the tall Lock Nut 1 from the head end of both the right and left hand caster bases 2 Position the two saddle T Y 3 adaptor brackets 2 mep V Ls 999 0831 002 so that the ends with slots face away from the bases as shown 3 Lower the brackets onto the caster bases aligning holes in brackets w
22. ces the bed in a horizontal position and hold the CPR Heart icon button 2 This illustration shows bed all the way down as a reference only the CPR button will not lower the bed Basic American Medical Products a division of GF Health Products Inc www grahamfield com 999 0831 190A Service Manual WY BASIC AMERICAN m MEDICAL PRODUCTS BED OPERATIONS TRENDELENBERG REVERSE TRENDELENBERG ZENITH APS SERIES 1 Make sure the Hi Lo function on the staff control panel is not locked out orange lock icon To unlock press the Key icon 1 and Hi LO Locked button 2 at the same time 2 To move the bed to the Trendelenberg position foot end up press on button 3 3 To move the APS bed into the Reverse Trendelenberg position head end up press on button 4 TRENDELENBERG NOTE THE TRENDELENBERG AND REVERSE TRENDELENBERG FEATURE CAN ONLY BE ACCESSED USING THE STAFF CONTROL IT CANNOT BE ACCESSED USING THE HAND CONTROLLER PENDANT REVERSE TRENDELENBERG Basic American Medical Products a division of GF Health Products Inc www grahamfield com 999 0831 190A Service Manual z E BASIC AMERICAN MEDICAL PRODUCTS ZENITH APS SERIES APS 9000 SERIES ELECTRICAL COMPONENTS LINAK Control Junction Box Part 999 083 1 300 Quantity 1 Power Cable plugs into side port part listed at bottom Head Motor Actuator Part 999 0822 052 Quantity 1 Foot Motor Actuato
23. devices This gap must be small enough to prevent of an improperly fitted mattress could result in injury resident patient from getting his her head or neck or death caught in this location Basic American Medical Products a division of GF Health Products Inc www grahamfield com 999 0831 190A Service Manual A BASIC AMERICAN ZENITH APS SERIES aan MEDICAL PRODUCT STANDARD WIREFORM WALLSAVER INSTALLATION STEP 1 WALLSAVER ASSEMBLY 1 Position the Wireform Wallsaver Part 999 0844 180 with bent end facing upward and tab ends facing inward as shown at right 2 Determine the position desired See bottom of page 3 Gently squeeze the tab ends of the wallsaver inward toward the center of the wallsaver and holding the tabs parallel with the slots in the caster bases See DETAIL A slide the tabs into the slots while letting the wallsaver gently expand outward 4 Turn the wallsaver downward until it rests on the floor See Wireform Wallsaver Detail B 999 0844 180 Bend faces upward STEP 2 WALLSAVER CO REMOVAL l 1 To remove or move the Wireform Cal Wallsaver to a new position raise l the wallsaver off the floor until LLI the end tabs are horizontal e 2 Squeeze the ends toward the Start with tab held OA center of the wallsaver until the nonzontaly wt end tabs slide out of the caster base slots STEP 2 WALLSAVER POSITIONS 76 Bed Wallsaver 76 Bed Wallsaver Position w Trapeze or 80 Bed W
24. e RU alang nta oe factory you will need to first attach it to the footboard amp Staff Control inside of board With the cutout side of the footboard facing outward ias as shown insert the staff control cable through the large round hole Q N assembly into the cutout slot on the footboard Shroud Y H Cable Cover and attach to the board using the two outside small holes and two screws from your staff control hardware kit p Attach the Shroud Cover over the cable on the Cable Tie inside of the board using the four remaining Vp Paani a NN screws from your kit 3 e ZN cable to mounting tube foot end with phone jack into the round plug making sure the phone jack is seated correctly inside the female plug arrow to arrow see DETAIL B Foot Board with cutout NN the Staff control s female plug to secure see DETAIL B SEE PAGE 10 OF THIS Loop Male Staff Control Cable f KY f MANUAL FOR PROPER A a m maging y fh Y ASSEMBLY OF ELBOW Euge ns ep MOUNTING BRACKETS S TUBES amp FOOTBOARD IMPORTANT Y i O Make sure to tie j y Cia off the staff control iB board mounting tube with a cable tie as shown cable to the foot o sA 4 NN Female end of Staff Control Assembly Cable plugs into T Cable Locking End Cap on Double T Cable DETAIL B MAKE SURE LOCK End of Double T Cable at f d G CT Fine on either side of Seat Pan END CAPS ARE for a
25. e connection to the T Cable at the foot end making sure plugs are fully engaged Also check the T Cable connection at the Control Box If you have an Underbed Light feature make sure all cables are secure If the PENDANT STILL DOESN T WORK move to step c UNPLUG THE POWER CORD FROM THE WALL OUTLET Unplug the Pendant Hand Controller plug from the end of the T Cable on either side of the beds Seat Pan Unplug the Staff Control Cable from the end of the T Cable at the foot end of the bed and unplug the other end from the Control Box Plug the Pendant Hand Controller plug directly into the HB port on the Control Box making sure it is fully seated Plug in the Power Cord and test the Pendant ALL BUTTONS WORK Replace the T Cable See page 32 for order information NOTHING WORKS Replace the Hand Controller See page 31 for order information Basic American Medical Products a division of GF Health Products Inc Make sure plugs are fully seated O ring does not show Make sure plugs at foot end are fully seated and cap lockedon B Unplug Staff Control Cable from T Cable at foot end D Unplug Power Plug before working on bed electonics Plug Pendant Cable directly into HB port on Control Box or into port 2 in Underbed Light Box and test www grahamfield com 999 0831 190A Service Manual A BASIC AMERICAN ZENITH APS SERIES MEDICAL PRODUCT TROUBLESHOO
26. e guidelines set forth by the FDA in order to avoid injury To that end we have provided the FDA s web address at right as a resource for understanding and following these guidelines for the safety of patients residents Basic American Medical Products a division of GF Health Products Inc It is also essential to have the correct bed components acces sories that correspond with the needs of your patient resident and the particular bed you have purchased Matching the correct bed component that correlates with the regulatory guidelines can be a daunting task Our sales team at GF Health Products Inc and our friendly Customer Service Representatives at Basic American Medical Products can help you sift through the wide array of compliance and bed options We will help you determine which bed bed part is best for your patient s resident s particular needs and help you with your compliance issues The APS 9000 bed and accessories listed in this manual are in full compliance with FDA guidelines for reducing the risk of bed entrapment Hospital Bed System Dimensional and Assessment Guidance to Reduce Entrapment Details can found at www fda gov www grahamfield com 999 0831 190A Service Manual WY S BASIC AMERI Ay 3 MEDICAL PRODUCT CAN ZENITH 9000 APS MECHANICS NOTE All dimensions are in a range of 25 inches e Overall Bed Length with brds amp wallsaver 827 86 e Overall Bed Wid
27. into the hollow ends of the main frame rails at the foot deck end e FOR 80 BEDS Slide the mounting tubes in until the 1ST hole in the tubes lines up with the 2ND hole in the rails See sample below e FOR 76 BEDS Slide the mounting tubes in until the 1ST hole in the tubes lines up with the 3RD hole in the rails See sample below Inserts on outside FOOTBOARD 40mm Bolts CENTER OF BED 80 hole position 2nd hole in 76 hole position 3rd hole in NOTE The first hole at the foot end is reserved for attaching the Optional 4 pan extension www grahamfield com 999 0831 190A Service Manual gt Ay 3 E Se ZENITH APS SERIES gt MEDICAL PRODUCT PLUGGING IN THE FOOTBOARD NURSE STAFF CONTROL STEP 1 ATTACHING THE FOOTBOARD The APS 9000 bed features a Nurse Staff Controller in the footboard however the board is ordered separately with your bed because of the variety of board styles available If ordered at the same time as the bed the Staff Control Assembly and Shroud Cover will be pre installed to the Footboard at the manufacturing factory STEP 2 CONNECTING YOUR CABLES Please refer to DETAIL A shown below O a O DETAIL A CLOSEUP OF STAFF CONTROL CABLE CONNECTION With text right side up insert the staff control Insert the T Cable end extending out the Screw on the round lock cap onto If the Staff Control Assembly is not installed at th
28. ith caster stems NOTE MOUNTING OF THE ADAPTOR D BRACKETS IS ONLY NECESSARY IF YOU PLAN ON USING THE TRENDELENBERG REVERSE TRENDELENBERG FEATURE STEP 2 WALLSAVER ASSEMBLY 1 Position the Wireform Wallsaver Part 999 0844 180 with bent end facing upward and tab ends facing inward as shown at right DETAIL B 2 Gently squeeze the tab ends of Finish with the wallsaver inward toward the T Wallsaver center of the wallsaver and hold EE with resting on ab held floor ing the tabs parallel with the horizontally slots in the caster bases See DETAIL A slide the tabs into the slots while letting the wallsaver gently expand outward l 4 Turn the wallsaver downward until it rests on the floor See Detail B STEP 3 WALLSAVER REMOVAL 1 To remove or move the Wireform Wallsaver to a new position raise the wallsaver off the floor until the end tabs are horizontal 2 Squeeze the ends toward the center of the wallsaver until the end tabs slide out of the caster base slots IMPORTANT WALLSAVER END NEAREST THE WALL MUST FACE UPWARD AS SHOWN FOR WALLSAVER TO SEAT PROPERLY IN SLOTS Wireform Wallsaver 999 0844 180 Bend faces upward Basic American Medical Products a division of GF Health Products Inc www grahamfield com 999 0831 190A Service Manual WY BASIC AMERICAN m MEDICAL PRODUCTS ZENITH APS SERIES BED OPERATIONS APS PENDANT HAND CONTROLLER 9 a 9
29. lable at the time of printing The most current product information can be found online at www grahamfield com Please check all parts for shipping damage and test before using In case of damage DO NOT USE WY SE BASIC AMERICAN gt MEDICAL PRODUCT Ay 3 TABLE OF CONTENTS Safety BAG AA 4 ID Warning Labels 39 21 5 Gav cade any AA KANA 5 6 Entrapment Compliance Ls 7 Features amp Specifications 8 Unpacking Your New Bed 0200005 9 Assembly Boards 2 00002 cece eee 10 Plugging in Your Staff Control 11 Standard Mattress Retainers 12 Standard Wireform Wallsaver 13 Wallsaver Adapter for Trendelenberg Positions 14 Operations Controllers amp Staff Control 15 16 Placing the Bed into the Chair Position 17 Using Trendelenberg Reverse Trendelenberg 18 Electronic Components xb E x Rx 19 Electrical Cabhng iud Steves kati aided Bee trae 20 22 Optional Pivot Assist Bar sus 23 Optional Trapeze Support Adaptor 24 Optional Battery Backup Configurations 25 Troubleshooting No Power LL 26 Troubleshooting Staff Control amp Pendant 26 27 Troubleshooting Motors Control Box 28 Replacing the Staff Control Assembly 29 Maintenance amp
30. les in the Trapeze Support screwing them into the inserts in the headboard OPTIONAL FOR TMI TRAPEZE BAR WITH SLEEVE Insert the Sleeve into the vertical support with the slot on the bottom Rotate the sleeve until it falls into place over the carriage bolt you inserted in Step 4 This will lock it into the proper position NOTE The top of the Sleeve should sit flush with the top of the vertical support on the Trapeze Support Adaptor www grahamfield com 999 0831 190A Service Manual wa BASIC AMERICAN ZENITH APS SERIES MEDICAL PRODUC OPTIONAL BATTERY BACKUP REPLACEMENT CONFIGURATIONS Configuration 1 Replacing Single Battery Cable for a Portable Battery Backup Unit la 1 Replacing your existing battery backup cable Part 999 0711 001 is fairly simple O First if your external battery unit is hooked up unplug it from the foot end battery cable plug 1 ie Sed 2 Carefully cut all cable ties 2 eins wit eri olm CETT EN 1 1 1913 CU i O 3 Unplug the existing cable from the Control NATI Nh Box 3 Plug in the new cable and run Hi along same path as previous cable 4 Use cable ties to re secure to tie rod and a BIRD OIS RSEN jen fel LI Configuration 2 Replacing Mounted Battery Backup Unit includes attached battery cable 1 If your APS bed came with a mounted Battery 2 Using a 3 8 socket wrench remove the two Nylon Lock Nu
31. mm Retaining Rings Hair Pin Clips Snap Rings Standard Cable Ties Installation Instruction Sheet Head and Hi Lo Motor Cables 999 0806 200 1 per Motor Foot Motor Cable 1 Cable 070101 1510 E 10 Q Fena R Description Order Date Linak Control Box and Clip Standard 3 Prong Power Cable Ground Wire Optional 2 Prong Power Cable PENDANT HAND CONTROLLER Description QTY Order Date Backlit Pendant with Underbed Light button 999 0806 202 Part Number 999 0831 300 999 0775 208 999 0775 206 Part Number 999 0806 301 1 Pendant 999 0791 000 Pendant Holster 1 Holster 999 0806 305 Non Backlit Pendant no Underbed Light button 1 Pendant Basic American Medical Products a division of GF Health Products Inc www grahamfield com 999 0831 190A Service Manual WY BASIC AMERICAN MEDICAL PRODUCTS AUR N mi lt T I U o U m A m DM N SERVICE REPLACEMENT PART RECORD Page 2 DOUBLE T CABLE Part Number Description Order Date 999 0806 200SP Double T Control Cable 1 Cable o l Deck Hardware Pack INCLUDES 1 Pkg 2 4 1 T Split Stem Cable Retainer Push Cable Ties Standard Cable Ties Installation Instruction Sheet Part Number Description Order Date UNDERBED LIGHT Underbed Light Unit 1 Motor cty Motor Hardware Pack INCLUDES 1 Pkg Adaptor Cable 999 0806 212 Standard Cable Ties Attachment Material Installation Instructi
32. mmended b Mechanical Inspection of Casters amp Pedal Locking Mechanism 3 months Check the pedal locking mechanism to make sure it engages and disengages properly Check the bottom foot pads on the foot end casters making sure they are clean of debris and undamaged Replace if needed Check all casters to ensure that they roll properly and are unobstructed b Electrical Inspection of Control Box 6 months Check the external power cord that plugs into the control box for any chafing cuts or wear Make sure all attaching hardware is securely tightened Check all electrical connections for any wear or fractures Check your external or mounted battery see page 24 Replace if needed gt Electrical Inspection of Pendant and Staff Control 6 months Check the pendant cord for any chafing cuts or wear Check all pendant functions v Head raises and lowers properly V Foot raises and lowers properly v Entire bed raises and lowers properly v Bed moves to Chair position properly v Bed moves to Trendelenberg and Reverse Trendelenberg positions properly Check to make sure each button and associated function work properly i e the head section raises when the Head Up button is activated gt Electrical Inspection of Actuators Motors 6 months Check the actuator motor cords for any chafing cuts or wear Check the range of movement on all motors to ensure they do not bind in the Full
33. nts considered as high risks for entrapment 1 e patients with pre existing conditions such as confusion restlessness lack of muscle control altered mental status either organic or medicinal or a combi nation thereof Additional safety measures should be considered for such high risk patients NEVER permit more than one 1 person on in the bed at any time Body weight should be evenly distributed over the sleeping surface of the bed DO NOT allow the patient to lay sit or lean in such a way that their entire body weight is placed only on the raised head or foot sections of the bed This especially applies when repositioning or transferring a patient in or out of the bed Increased risk may occur when the patient s size and or weight are inappropriate for the bed s dimensions or weight capacity The bed is intended for use storage and transport within a temperature range of 40 C to 60 C It has a water resistance rating of IPX4 and IS NOT to be power washed or submersed Operating Conditions Operation of the bed is based on the following conditions Ambient Temperature of 10 C to 40 C Relative Humidity of 30 to 75 Non condensing Atmospheric Pressure of 700hPa to 1060hPa and a Splash Protection of IEC 60529 Storage Storage of the bed is based on the following conditions Ambient Temperature of 10 C to 70 C Relative Humidity of 10 to 100 and an Atmospheric Pressure of 500hPa to 1060hPa Radio F
34. on Sheet STAFF CONTROL Description Order Date Switch Pad with Bezel amp Cables Staff Control Service Pack INCLUDES Switch Pad with Bezel amp Cables Shroud Cable Cover 12 4 1 1 PLEASE SEE PAGE 11 FOR 6 Phillips Head Truss Screws INSTALLATION INSTRUCTIONS 4 Nylon Cable Tie Installation Instruction Sheet CAPS amp PLUGS Description NOTE CAPS amp PLUGS CAN ONLY BE PURCHASED IN SETS OF 12 100 4200 004PK 1 25 x 1 25 Square Caps Pack 100 4700 017PK 1 00 x 2 00 Rectangular Caps Pack 100 4700 018PK 1 25 x 2 00 Rectangular Caps Pack 100 4715 011PK 1 50 Round End Caps Pack cover arm bearings 999 0775 001PK Half Moon End Caps Pack caster bases 100 4762 002PK 5 8 Round Plugs Pack both sides of foot deck 100 4738 005PK 3 8 Round Plugs on main frame rails foot end FOR GENERAL CAP PLUG DIAGRAMS PLEASE SEE PAGES 36 amp 37 32 Basic American Medical Products a division of GF Health Products Inc www grahamfield com 999 0831 190A Service Manual he miki utin ZENITH APS SERIES WT MEDICAL PRODUCT Ay 2 SERVICE REPLACEMENT PART DIAGRAM ELECTRONICS FOOT END in HI LO MOTOR ap END OF BATTERY CABLE CONTROL BOX BLACK ATTACHED TO MOTOR HEAD END FOOT END HI LO MOTOR OF T CABLE GREY LONGER SIDE OF T CABLE WITH PUSH CABLE TIE IN SMALL HOLE a PA A FOOT KNEE AQ FA MOTOR FOOT
35. ontal position The two HI LO motor shafts you hear a click or buzz but nothing happens Check your should be completely retracted inside the motors Hi Lo Motors by touching them amp pushing the Hi Lo buttons DOESN T WORK Replace the Foot Motor See page 31 for order information YOU FEEL NOTHING possible bad motor go to the QUICK REFERENCE third step below HI LO amp FOOT WORKS BUT HEAD DOES NOT YOU FEEL SLIGHT VIBRATION Motor OK Switch Head 4 Foot Motor Cable at the Control Box and test See Step C in left column Unplug the Power Cord Switch the two Hi Lo plugs ports 4 and 2 in the Control Box Re plug the power cord HI LO amp HEAD WORKS BUT FOOT DOES NOT and press the same Hi Lo button then touch the same Switch Foot amp Head Motor Cable at the Control Box non vibrating motor and test See Step D in left column i MOTOR NOW VIBRATES Bad port in the HEAD amp FOOT WORKS BUT HI LO DOES NOT Control Box Replace box See page 31 for a Switch Head and Foot Hi Lo Motor Cables at the order information Control Box and test See Step E in left column b Test to see if Hi Lo Motors need to be re aligned ii MOTOR DOES NOT VIBRATE Replace See Step F above the bad Hi lo Motor See page 31 Basic American Medical Products a division of GF Health Products Inc www grahamfield com 999 0831 190A Service Manual Pay sini scene ZENITH APS SERIES MEDICAL PRODUCTS
36. r Part 999 0822 053 Quantity 1 Hi Lo Motor Actuator Part 999 0831 051 Quantity 2 STANDARD 3 Prong Power Cable Part 999 0775 208 Quantity 1 OPTIONAL 2 Prong Power Cable pt Part 999 0775 206 Quantity 1 T DO NOT use unauthorized parts accessories or adaptors other than those specified authorized by GF Health Products Inc DO NOT open assemblies such as the Actuators Hand Control or Control Box If unauthorized personnel perform work on these components the manufacturer s STANDARD Pendant Hand Controller with Underbed Light Button Part 999 0806 301 Quantity 1 Bed also includes Pendant Holster Part 999 0791 000 Quantity 1 Double T Cable Part 999 0806 200 Quantity 1 Motor Cable Hd amp Hi Lo Part 999 0806 203 Quantity 3 Bon xp Motor Cable Ft Part 999 0806 202 Quantity 1 Battery Backup Cable Part 999 0831 200 Quantity 1 OPTIONAL UNDERBED LIGHT Underbed Light Adaptor Cable Part 999 0806 212 Quantity 1 Underbed Light Part 999 0806 220 Quantity 1 warranty becomes void NEVER operate the bed if a Power Cord or Plug is damaged or not working properly Contact qualified Service Personnel for examination and repair Always unplug the Power Cord when performing any maintenance on the bed Basic American Medical Products a division of GF Health Products Inc www grahamfield com 999 0831 190A Service Manual a
37. requency Interference RFI RFI influences most electronic equipment Caution should be exercised with regard to the use of portable communications equipment in the area around such equipment If RFI causes erratic behavior shut the bed off immediately Leave it off while the transmission is in progress www grahamfield com 999 0831 190A Service Manual WY SE BASIC AMERICAN gt MEDICAL PRODUCT Ay 3 ZENITHAPS SERIES TYPICAL APS 9000 BED IDENTIFICATION LABELS with Grounded Electrical Cable MODEL NO 806 0000 000000 MFG DATE 12 23 11 VOLT HZ AMP TYPE 110 120 50 60 4 5 AC SERIAL NO 000000000000 GF HEALTH PRODUCTS INC DUTY CYCLE 2 min Continuous 18 min e MODE OF OPERATION Intermittent A 600 Ibs IPX4 A p S Q 8 1 T 4 ENITH 806 SERIES SERIALAPS 9 A SAFE WORKING LOAD Z PATENT PENDING APS9000 MODEL NO 806 0000 000000 MFG DATE 12 23 11 TITULI I 000000000000 SERIAL APS98174 Bed labels are an important part of identifying your bed s make and model when ordering replacement parts The Serial Number is essential if you are claiming parts or service under warranty These helpful labels can be located on the main frame rails immediately below the sleep decks on either side of the bed Please have this IMPORTANT infor mation ready when calling our customer service or technical support staff at 800 365 2338 it will allow us to better assist you and quickl
38. rican Medical Products a division of GF Health Products Inc www grahamfield com 999 0831 190A Service Manual A BASIC AMERICAN ZENITH APS SERIES MEDICAL PRODUCT SERVICE REPLACEMENT PART DIAGRAM DECKING ER Deck Bracket Retaining RI HEAD DECK ZA ox we Aa TO MAIN FRAME seat nA MEN S ui SEAT PAN BRACKETS Pan pe HEP Bracket k Re us or Pe 5 Existing second f Bearing deck Y bracket g X Re use Existing Tam O og Clevis Pin DETAIL OF DECK P T ATTACHMENT roce CEDE Similar for all decks E s TOMAIN TRAME SEAT PAN BRACKETS Re use Short T Clevis Pins Wa Re use Pa Existing Retaining _ Ratchets Ring amp _ Re use Long Clevis Pins GETRETEN 989 0844 930G FOOT DECK Retaining Bracket TO KNEE DECK Ring MOUNTING BRACKETS DETAIL OF EDEMA RATCHET RE ATTACHMENT Use existing ratchets and hardware 1 TO ORDER A REPLACEMENT 2 TO ORDER A REPLACEMENT 3 TO ORDER A REPLACEMENT HEAD DECK USE SERVICE PACK KNEE DECK USE SERVICE PACK FOOT DECK USE SERVICE PACK 999 0844 920SP includes deck 999 0844 940SP includes deck 999 0844 930SP includes deck mounting hardware amp installation mounting hardware amp installation mounting hardware amp installation instruction sheet instruction sheet instruction sheet Re use existing edema ratchets amp ratchet hardware WHEN ORDERING REPLACEMENT PARTS WITH CUSTOMER SERVICE PLEASE HAVE YOUR BED S SERIAL NUMBER AVAI
39. s for reducing the risk of bed entrapment Hospital Bed System Dimensional and Assessment Guidance to Reduce Entrapment The new Guidance identifies potential entrapment areas and those body parts most at risk for entrapment provides design criteria for manufacturers of new hospital convalescent beds recommends particular test methods to assess the conformance of existing hospital conva lescent bed systems and answers frequently asked questions about entrapment issues The new Guidance was a result of a long standing collaboration between the FDA and the Hospital Bed Safety Workgroup HBSW formed in 1999 GF Health Products Inc s Long Term Care Bed division Basic American Medical Products 1s an HBSW charter member As a result of our commitment to product safety all our current long term care beds have been strictly tested and conform to the new FDA Guidance The guidelines set forth by the FDA Guidance layout specific dimensional limitations on potentially injury threatening gaps and spaces that can occur between bed system components such as rails when not properly installed GF Health Products Inc and Basic American Medical Products have conformed to these guidelines from a manufacturing aspect However entrapment issues can often arise when a healthcare provider facility has not correctly assembled the components on a bed It is essential that the provider facility fully understand their responsibility in complying to th
40. s mounting tubes Set all pins aside for later use 2 Detach the headboard panel from the mounting tubes by unscrewing the four 40mm Allen Head Bolts that hold the board to the tubes these will be replaced with the longer Phillips Head Pan Screws from your trapeze adaptor hardware bag Do not remove the four 1 4 20 inserts 3 from the board They will be used again to mount the board to the trapeze support 3 Assemble the Trapeze Support 1 to the main frame tubes using the two Clevis Pins and Hair Pin Clips you removed in Step 1 The first holes in the Trapeze tubes must line up with the third hole from the end in the main frame tubes as shown Basic American Medical Products a division of GF Health Products Inc Trapeze Support vertical support tube TMI Bar Sleeve would be inserted N inside the tube Use first holes in Trapeze Support to mount to main frame tubes USE SAME CLEVIS PINS AND HAIR PIN CLIPS THAT WERE USED TO CONNECT HEAD BOARD TO MOUNTING TUBES 4 Insert the 1 4 20 x 2 50 Carriage Bolt 5 into the vertical support as shown and fasten with a 1 4 20 Hex Head Lock Nut 6 Both parts should be in your hardware bag 5 Assemble the headboard to the inside of the Trapeze Support so that the side of the headboard with the inserts is facing away from the Trapeze Support From the outside insert the four 1 4 20 x 1 75 long Machine Screws 4 from your hardware bag through the ho
41. th with boards 36 00 S High Height aiu ex eres EE EE 30 00 S Ko Am Cl AT 7 00 e Maximum Head Back Deck Angle 70 e Maximum Knee Foot Deck Angle 30 e Maximum Safe Working Load 600 Ibs WITH WEIGHT EVENLY DISTRIBUTED includes bedding resident support surface and all accessories e Mass of bed without assist devices or boards 204 Ibs Bed height calculated from floor surface to top of sleep deck ZENITH 9000 APS ELECTRICAL Power Frequency 120 Volt 50 60 Hz QUIDUT RAUNG Gia ai eee wares dove aa 12 33V IPX4 Overall Movement Draw 4 50 Amps e Classification ER Class I Type B Electrical Cord coii t 18 AWG 3 Conductor Type SJT ClassitGAalOM 3330 Hated ta bee Class Il Type B Electrical Cord eei s 18 AWG 2 Conductor Type SJT Mode of Operations 10 Max Duty Cycle 2 minutes on 18 minutes off Battery Backup Option 1 120V External Battery and Charger Pack and Charger can be purchased separately as an accessory Battery Backup Option 2 120V External Battery Pack is mounted to the bed frame no charger needed Basic American Medical Products a division of GF Health Products Inc www grahamfield com ZENITH APS SERIES ERN 999 0831 190A Service Manual KHA 2 x BASIC AMERICAN MEDICAL PRODUCTS UNPACKING YOUR BED e Make sure all parts components are included e Check all
42. the Control Box directly to an d OPTION external Battery Pack at the foot end Basic American Medical Products a division of GF Health Products Inc www grahamfield com 999 0831 190A Service Manual A BASIC AMERICAN ZENITH APS SERIES aan MEDICAL PRODUCT OPTIONAL PIVOT ASSIST BAR INSTALLATION amp OPERATION INSTALLATION OS 1 At the head end of the bed position the Pivoting Assist Bar 2 Position the Assist Bar ANY bracket over the Head Deck 1 frame and align the two holes Bracket ON TOP of the 3 in the Assist Bar bracket with the two holes in the frame rail on ie el ena M N o either side of the head deck sd 2 From the outside insert the Clevis Pins 3 through the holes in the pivot bracket and deck frame HEAD Gb END 3 From the inside insert a Hair Pin Clip 4 through each of the small holes in the clevis pins PLEASE ORDER KIT ZA85400 KIT INCLUDES ONE PIVOTING ASSIST BAR ASSEMBLY AND TWO LANYARD ASSEMBLIES WITH CLEVIS PINS AND HAIR PIN CLIPS A STATIONARY ASSIST BAR IS ALSO AVAILABLE AS AN OPTION KIT ZA85000 THE ASSIST BAR CAN BE POSITIONED ON EITHER THE RIGHT SIDE OR LEFT SIDE OF THE HEAD DECK SAMPLE SHOWS PIVOTING IT IS NOT OFFERED FOR THE FOOT END OF THE BED ASSIST BAR POSITIONED ON THE LEFT SIDE OF THE BED PIVOTING ASSIST BAR OPERATION 1 To release the pivoting assist bar from its vertical lock position hold onto the top of the bar with one hand and slightly
43. ts Backup Pack Part 999 0711 003 you must 6 Part 100 6725 003 that secure the Battery Mounting first unplug the power cord from the wall unit Bracket 5 Part 999 0831 001 to the welded Main Frame then unplug the attached battery cable from Bracket 4 Remove the bolts 7 Part 100 5225 009 and the Control Box Carefully cut all cable ties set aside along the battery cable 3 Once the Bracket and Battery Pack 9 are free remove the pack from the bracket by unscrewing the four Phillips Head Screws 8 Part 100 5414 001 Place the new battery pack on the bracket and reverse the process to reattach bracket and cable Basic American Medical Products a division of GF Health Products Inc www grahamfield com 999 0831 190A Service Manual A BASIC AMERICAN ZENITH APS SERIES aan MEDICAL PRODUCT TROUBLESHOOTING NOTHING WORKS NO POWER 1 NOTHING WORKS NO POWER a Check to make sure you have power coming from your outlet Unplug the power cable from the outlet and test the outlet by plugging in a lamp or similar portable device OUTLET WORKS move to Step b Test Outlet for Power Unplug Power Cord and check entire cord to make sure it is not pinched frayed or damaged b Replug the power cord into the outlet NOTHING WORKS UNPLUG THE POWER CORD and check the power cord from the head end to the control box making sure it is not pinched frayed or damaged in any way ie POWER CORD
44. ts will not show up SAFE MODE Green light will show automatically when bed is plugged in and will blink intermittently during 10 minute sleep mode After 10 minutes of lapsed time since the last function was used the LED light shows green and Hi Lo Lock icons will show orange To manually disable the function hold the Safe Mode button 15 and Key button 8 at the same time Green LED light will go out 999 0831 190A Service Manual 2G nasi AMERICAN ZENITH APS SERIES gt MEDICAL PRODUCT Ay 2 BED OPERATIONS CHAIR POSITION 3 2 Q 1 Make sure the Chair position on the staff control panel is not locked out orange lock icon To unlock press the Key icon 1 and Chair icon 2 or All Locked button 3 at the same time 2 To move the bed to the Chair position foot amp knee decks angled up head deck angled up and head of bed tilted upward press on the chair icon button 2 HEAD END To release the chair feature and return the bed back into a horizontal position press the CPR icon 4 The Chair feature can also be operated using the Pendant See page 15 1 Make sure the bed positions on the staff control panel are not locked out orange lock icon To unlock press the Key icon 1 and the All Locked button 3 at the same time 2 To smoothly move the bed into a prone CPR POSITION horizontal position for CPR access press The CPR button pla
45. ttaching your Hand Controller SCREWED ON SECURELY Basic American Medical Products a division of GF Health Products Inc www grahamfield com 999 0831 190A Service Manual Pake SUC ne ZENITH APS SERIES gt MEDICAL PRODUCT Ay 3 STANDARD MATTRESS RETAINER INSTALLATION STEP 1 HEAD RETAINER INSTALLATION gt 1 Position retainer with curved end toward headboard 2 Squeeze ends of the retainer toward the inside of the bed and lower between the deck rails and first outside grid wires 3 Align the ends of retainer with small holes on the inside of deck rails and slowly release STEP 2 FOOT END RETAINER INSTALLATION 1 Position retainer with curved end toward footboard 2 Squeeze ends of the retainer toward the inside of the bed and lower between the deck rails and first outside grid wires 3 Align the ends of retainer with small holes on the inside of deck rails and slowly release Use Outside Holes for 80 Bed Use Inside Holes for 76 Bed STEP 3 RETAINER POSITIONS M 9 FOOTEND gp mM e FOR 80 BEDS Use outside small holes See sample at right FOR 76 BEDS Use inside small holes See sample at right A Be sure to use a mattress that is properly sized to fit A Use a properly sized mattress in order to minimize the sleep deck which will remain centered on the the gap between the side of the mattress and assist deck relative to State and Federal Guidelines Use
46. y answer your questions and concerns TYPICAL APS 9000 BED IDENTIFICATION LABELS for Class Il Models MODEL NO 806 0000 000000 MFG DATE 12 23 11 VOLT HZ AMP TYPE 110 120 50 60 45 AC SERIAL NO 000000000000 GF HEALTH PRODUCTS INC DUTY CYCLE 2 min Continuous 18 min MODE OF OPERATION Intermittent QD cons E T IPX4 N SAFE WORKING LOAD ENITH 806 SERIES SERIALAPS 9 Z PATENT PENDING APS9000 MODEL NO 806 0000 000000 MFG DATE 12 23 11 TITULAR TTL 000000000000 SERIAL APS98174 APS98174 If you opted for an electrical system without grounding on your electrical cord your bed is designated as a Class ll The Identification bed labels to the left represent the labels for Class II beds They can be located on the main frame rails immediately below the sleep decks on either side of the bed Please have this IMPORTANT information ready when calling our customer service or technical support staff at 800 365 2338 The labels at left will allow us to better assist you in answering your questions and concerns Basic American Medical Products a division of GF Health Products Inc www grahamfield com 999 0831 190A Service Manual WY SE BASIC AMERICAN gt MEDICAL PRODUCT Ay 3 ZENITH APS SERIES TYPICAL SAFETY WARNING BED LABELS The following warning labels have been placed on your bed to help protect you and your bed from damage Please do not remove an
47. y labels from your bed WARNING DO NOT LOWER BED WHEN OBJECTS ARE BENEATH BED FAILURE TO INSPECT UNDER BED CAN RESULT IN DAMAGE TO PROPERTY OR PERSONAL INJURY ATTENTION S assurer de ne pas faire descendre le lit lorsque des objets se trouvent sous le lit Ne pas inspecter le dessous du lit pourrait entrainer des dommages materiels et des risques de blessures WARNING DO NOT LOWER BED WHEN OBJECTS ARE BENEATH BED FAILURE TO INSPECT UNDER BED CAN RESULT IN DAMAGE TO PROPERTY OR PERSONAL INJURY Basic American Medical Products a division of GF Health Products Inc CAUTION THIS BED IS SUITABLE FOR USE ONLY WITH OXYGEN ADMINISTERING EQUIPMENT OF THE NASAL OR MASK TYPE ORA TENT COVERING ONLY THE UPPER HALF HEAD END OF THE BED OXYGEN TENT CANOPIES SHOULD NOT EXTEND BELOW BED SPRING LEVEL LOCK HAND CONTROL AT FOOT OF BED WHEN USING OXYGEN ADMINISTERING EQUIPMENT ATTENTION CE LIT PEUT ETRE UTILISE UNIQUEMENT AVEC UN EQUIPMENT DESTINE A L ADMINISTRATION D OXYGENE DE TYPE NASAL OU MASQUE OU AVEC UNE TENTE RECOUVRANT SEULEMENT LA MOTTIE AVENT TETE DU LIT LES COTES DE LAS TENTE OXYGENE NE DOIVENT PAS SE PROLONGER PLUS DAS QUE LA SOMMIER DU LIT www grahamfield com 999 0831 190A Service Manual z E BASIC AMERICAN MEDICAL PRODUCTS ZENITH APS SERIES ENTRAPMENT amp COMPLIANCE INFORMATION On March 10 2006 the FDA U S Food and Drug Adminis tration released long awaited guideline