1. Waist Garment ti Is WH Whi A HOW TO OPERATE THE RECOVERY EDGE THERAPY SYSTEM Overview Step 1 Set up the Recovery Edge Therapy System Pump Step 2 Attach the Air Hoses to the Garment Step 3 Dress Appropriately Step 4 Putting on the Leg Garment Step 4 A Putting on the Leg Extender Step 4 B Putting on the Arm amp Shoulder Garment Step 4 C Putting on the Waist Garment Step 5 Connect the Air Hoses to the Recovery Edge Therapy System Pump Step 6 Select the Pressure Level Step 7 Lock the Pressure Knob Step 8 Set the Pause Time Step 9 Continue the Treatment Session Step 10 Ending the Treatment Session Step 1 Set up the Recovery Edge Therapy System Pump A Place the Recovery Edge Therapy System on a flat surface You should be able to reach it easily and operate all the controls from your relaxed treatment position B Make sure the pump is securely positioned so that it will not slip or fall C Connect the power cord to an electrical outlet Step 2 Attach the Air Hoses to the Garment A Attach the air hoses to the garment by inserting the hose into each air inlet See Figure 3 B Make sure each connection is securely attached Figure 3 Attaching Hoses to Garment Step 3 Dress Appropriately A Always wear light clothing underneath the Recovery Edge Therapy System garment for hygienic reasons to avoid irritation and to absorb perspiration B Do not wear the Recovery
2. C AFTER Use Recovery Edge Therapy System for a relaxing treatment at the end of the day or after exercise to relax and soothe tired aching muscles Suggested treatment time is 30 45 minutes after your work out or in the evening WARNING The treatment sensation should be pleasant and comfortable If you experience pain or discomfort during or after the treatment discontinue use and consult your physician 43 Step 10 Ending the Treatment Session A To end treatment shut OFF the Recovery Edge Therapy System using the ON OFF switch B Remove the garments C Unplug the pump from the wall socket D We recommend that you do not disconnect the hoses from the garment If you need to disconnect the hoses from the garment do this by gently working the end fitting free from the air inlet using a twisting pulling motion NOTE If the Recovery Edge Therapy System stops working and the warning signal beeps it means that the time needed to fill up the garment is too long Usually this is due to air leakage because the air hoses are not properly connected or because one of the air outlets on the pump is unplugged If this happens turn Off the pump at the ON OFF switch and then A Check that all the hose ends are correctly inserted into the air inlets B Check that the hose connectors are properly connected to the air outlets on the pump C If using only one garment make sure that the unused outlet is plugged with
3. 20 80 mmHg 120 V 60 Hz 14VA F500mMAH250V ABS Plastic 5 1 130mm 11 8 300mm 8 3 210mm 4 5lbs 2kg SAFETY GUIDE Overview General Usage Guidelines General Health Guidelines Health Safety Warnings Electrical Safety Operating Safety Guidelines Product Safety Features General Usage Guidelines Read the entire instruction manual before using the Recovery Edge Therapy System Proper application and connection to the pump is essential Garments should be removed if the patient experiences tingling numbness or pain Do not use any garment other than official Garments with the Recovery Edge Therapy System pump Do not use if under the influence of alcohol or drugs Do not use the device in high humidity place such as a Bathroom or sauna where fire accidents or electric shocks are possible General Health Guidelines This device is intended for use by people in good health This device is not recommended for people who have heart problems or vascular problems have a condition requiring the use of any medical device or have any condition that might affect their normal well being Do not use this device over insensitive or numb areas or in the presence of poor circulation Do not use if you have been diagnosed with blood clots deep vein thrombosis or phlebitis This device should not be used over swollen or inflamed areas or skin eruptions Do not use in the presence of unexplained calf pain If you are
4. 3 Proper Labeling The RMA number must appear on the outside of the shipping container Return shipments will not be accepted if the RMA number is not clearly visible 4 Written Description Please provide a written statement indicating the model number serial number and a brief description of the reason for return 5 Shipping Address send returns to AUTHORIZED REPAIR CENTER Attn Warranty Department Recovery Edge Therapy Systems 9601 Owensmouth Ave 8 Chatsworth CA 91311 Shipping Instructions and Charges 1 Packaging The unit must be shipped to the Company s Authorized Repair Center in either its original package or similar package affording an equal degree of protection Failure to provide this may result in voiding the warranty 2 Customer Return Address Instructions must be provided indicating an address to which the repaired unit must be returned 3 Shipping Charges Under Warranty The Company is responsible for the freight amp insurance charges for any product that is being repaired or replaced that is under the warranty 4 Shipping Charges Out of Warranty For any product that is not covered under the warranty the customer is responsible for any freight and insurance charges 18
5. or may be pregnant consult with your physician before use If you have questions or concerns consult your physician prior to use Health Safety Warnings The unit should NOT be used in the following conditions e Severe arteriosclerosis or other ischemic vascular diseases e Known or suspected acute deep vein thrombosis e Severe congestive cardiac failure e Any local condition in which garments would interfere for example e Dermatitis e Untreated infected leg wounds e Recent skin graft e Gangrene Do not use this product if you are under the care of a physician or have a contraindication requiring the use of any medical device Do not use this product on sensitive skin or in the presence of poor circulation Do not use this product if you have any of the following conditions e Acute congestive cardiac failure e Acute infections e Acute pulmonary edema e Acute thrombophlebitis e Bone fractures or dislocations at or in the vicinity of application e Deep vein thrombosis dvt or blood clots e Episodes of pulmonary embolism e Wounds lesions or tumors at or in the vicinity of application where increased venous and lymphatic return is undesirable Electrical Safety This product contains electronic components When using an electrical appliance basic safety precautions should always be followed to reduce the risk of burns fire electric shock or injury to persons NOTE This product complies with the re
6. Edge Therapy System garment directly over bare skin C Clothing should be unrestrictive and absorbent and free of zippers buttons or other items that could rub and chafe under the treatment We recommend wearing cotton clothes such as sweat pants or leggings along with cotton socks underneath the Recovery Edge Therapy System leg garment or a long sleeve t shirt under the Recovery Edge Therapy System arm garment _9 Step 4 A Putting on the Leg Garment Figure 4 Putting on the Recovery Edge Therapy System Garment A Zip the leg garment closed at least part of the way up before inserting your leg B Insert your foot into the garment see Figure 4 C Pull the top of the garment towards your torso extending the garment to its maximum length D Close the zipper all the way up E Repeat for other leg if desired The Recovery Edge Therapy System leg garment is designed to fit on either the right or left leg Step 4 B Putting on the Arm amp Shoulder Garment A Zip the arm garment closed at least part of the way up before inserting your arm B Insert your arm into the arm garment pulling it over your shoulder Make sure the zipper is facing up and that the cutout area is under your armpit C Close the zipper all the way up Support your arm on an armrest or a pillow during the treatment Supporting the arm in a horizontal position allows the garment to cover the shoulder area D If the garment
7. be owner s sole remedy and RE s sole liability on contract or warranty of otherwise for the products This warranty shall not be modified or amended without the written approval of an officer of the Company Extended Warranties The Company may make available an extended warranty s for additional years beyond the Standard Warranty Plan Please contact the Company for specific details regarding the costs for Extended Warranty Plans NOTICE In no event shall the company be liable for any indirect special incidental or consequential damages resulting from the failure to perform or use or improper use of any goods or services sold pursuant hereto whether due to breach of contract breach of warranty negligence or otherwise 17 PRODUCT RETURNS AND REPAIRS Return Materials Authorization RMA Required and Issuance of RMA Number RMA number must be obtained from customer service or a sales representative for any and all returns The customer agrees that any product to be serviced repaired or returned to the Company shall maintain the following procedures 1 Notification The customer must notify the Company within 15 days of first noticing the defect and promptly return the defective product upon receipt of RMA number s before expiration of the warranty period 2 Prior Approval Contact the Company directly to obtain a Return Materials Authorization RMA number for shipping purposes Customer Service Toll Free 1 855 525 EDGE
8. the Air Hoses to the Recovery Edge Therapy System Pump A The first time you use the Recovery Edge Accelerated Recovery for Athletes Therapy System set the Pressure Knob to S their minimum settings see Figure 7 Figure 7 General View of the Recovery Edge Therapy System Front Panel B Turn On the Recovery Edge Therapy System by pressing the ON OFF switch at the bottom right side of the pump to the ON position See Figure 8 Figure 8 The ON OFF Switch C Wait for a few cycles to become accustomed to the treatment sensations Note that the first inflation cycle may take a longer time to fill especially if you are using two garments D Set the pressure level see Figure 9 and pause time see Figure 10 to your comfort according to the guidelines below E Guidelines for Selecting Pressure Changing the pressure level will change the intensity of the treatment by increasing the amount of air pushed into each air chamber e To decrease the pressure Turn the pressure controller to the left for decreasing the pressure intensity Minimum 20mmHg e To increase the pressure Turn the pressure controller to the right for increasing the pressure intensity Maximum 200mmHg A lower pressure setting will apply a lighter treatment while a higher pressure setting will apply a more intense treatment see Table 1 as a guide for pressure selection If the treatment is too intense lower the pressure You ma
9. the Prong Plug If all the connections are OK and the problem persists contact the Company D In the case of a power failure or malfunction Turn Off the pump at the ON OFF switch Take off the garment s Remove the electrical cord from the wall outlet If the system has malfunctioned contact the Company For more details see Section 10 Troubleshooting PEN 14 MAINTENANCE AND STORAGE Warnings Only an authorized technician may open the pump Before cleaning the pump disconnect the power cord from the electrical wall outlet Clean only according to instructions Do not hand or machine wash dry clean hand or power wring iron tumble or force heat dry Surface wipe only Do not use bleach O OA WN Cleaning the Recovery Edge Therapy System Pump 1 Gently wipe the outside of the pump using a nonabrasive cloth 2 Do not spill any liquids on the pump Cleaning the Recovery Edge Therapy System Garments 1 Never submerge the garments in liquid 2 They are cleanable by surface wiping only 3 Gently wipe the inner and outer surfaces of the garment using a soft cloth moistened with warm water not exceeding 100 F 40 C and a mild detergent Do not allow water to enter the air inlets of the garment at any time 4 If needed a soft brush can be used to remove stubborn dirt 5 Towel dry and surface wipe again with a soft cloth moistened in water to remove all detergent residue aga
10. 6 6 6 6 fc gt ere AE E ees 9 Step 4 A Putting on the Leg Garment cccccccccccceecceeeeceeecececeeeceueceueceeeesaeesaeesaeeseuesseesaueeseeesaees 10 Step 4 B Putting on the Arm Shoulder Garmentt ccccccccceccceecceeeeeeeeseeeceeeseeeesueseeeeseeeseeesaees 10 Step 4 C Putting on the Waist Garment c cc cccccc cece eeccseeeceeeceeeceeeceueeeeeeseeeseeeseeeseueseeeaeeeaeeesaess 11 Step 5 Connect the Air Hoses to the Recovery Edge Therapy System Pump 0ccseeeeeees 11 Sleep 6 Selecihe Pressure Leve isescwancadnancesarsiascsncasteacedacadins EE E 11 Step Lock the Pressure KNOD assictesscncenpiacistmnmaiainonsactessntnassenteduaistynaianeaed annsantialotamiigemnmdectanbebnteianeaiadeneamdanwes 12 NEDO DCL Ue P USONO eaa E E E RR 13 Step 9 Continue the TreatMent SESSION eesse riarena o aR r ANTEE aT 13 Step 10 Ending the Treatment Session ccccceccesecececeeeseeeceeeceuecueeeeeeseeesaeesaeeseuesueeeeeeseeesaees 14 MAINTENANCE AND ST ORAGB opiisseccesicuscceaehinciasedatsuainsecsseatescuctxbateseneanibinbeantuatnnvesthooieesaueeteseaseseds 15 TROUBLESHOOTING renra o ene eae a eee eee et eee E eee ee cee eee eee ee 16 WARRANT ie ee ae ee ee ee 17 PRODUCT RETURNS AND REPAIRS ciiccicesccecechtnetacecctsuatnsecsseatestuctxateseneanibinioantuadiarestaooieesaneeteseasesed 18 INTRODUCTION Thank you for purchasing our product Our goal is to provide you with a quality manufa
11. RECOVERY EDGE I herap y Ae USER MANUAL TABLE OF CONTENTS INTRODUC MON eere E E EE E EE E EES 2 SPECIFICATIONS AND DIMENSIONS 00000 0 0c ceccc eee ceeece cece eeeee esse esse eeseeeseeeseuesseessaeeaeeesaaes 3 soy gy all eC 0 2 Seen eee ee ee ene eee ene er er eee eee een eee ee eee ere 4 6 General Usage GUIGCHINGS sedaane E EEO E 4 General Heath GUIdeElNNES icc2scsc2cccsececesssazonncenseaastencieeaansensadacosssciataratsasedseeneeaaeadcadansdeeceteceiinebiterszouce 4 eME N S ee E E 4 ELECCION E E E E E TE 5 Operating Safety Guidelines cgi ssc oicesae engaskeedeeussnesincuaeenauseunsesccuemoeanseiescanedvanteanaiates eros eeeoeteeuenedskeuend 6 Product oade FO NUG S eee ee eee eee Ce ene er nee eee eee ee eee eee ee eee ee 6 COMPONENTS OF THE RECOVERY EDGE THERAPY SYSTEM ooo ooo ccceccceeeeeeeeeees 7 8 TFS FOD eaters eae race E E ROSE BUNGI rea a E I E EE E E AE E E E E A T FONI PIU e E E E ee aeesene noceuceeeesnes T BOREIA Ms E E E ie cc alae corto T EEEE EAE EA ERE AT 8 Arm and Shoulder Carme 9 Caen eee ne eet E EE EA E 8 eE EEEE a E E E E O E E EE E E a ee 8 HOW TO OPERATE THE RECOVERY EDGE THERAPY SYSTEM 00a eeceeeeeeeeeeeees 9 14 Step 1 Set up the Recovery Edge Therapy System PUM cccccccsecceeeeeeeeseeseeeeeeeseeeeaeeeaeeees 9 Step 2 Attach the Air Hoses to the Garment cccccceeccseeseeteeeseeeeeeeeeeseeesseeseeteeeeeeeeeteeeteeeaeeteeess 9 Slepp 3 Dress 21
12. ause interval If you want to slow down the treatment pace select a longer pause interval see Table 2 Table 2 Guideline for Pause Time Selection Pause Duration Pause Time seconds Short Pause Interval 10 20 Medium Pause Interval 30 50 Long Pause Interval 60 70 NOTE If you experience tingling or numbness during the treatment anywhere in the treated area increase the pause time to 30 to 45 seconds or more If this sensation persists after increasing the pause time reduce the pressure If this sensation continues stop using the system immediately and consult your physician When using two large garments a longer pause time may be necessary to ensure that all the air deflates from the garments in between treatment cycles The Recovery Edge Therapy System will not inflate the garments if the air from the previous cycle has not adequately emptied If the pause time you selected is too short for adequate deflation the pump will automatically double the pause time If this is bothersome increase the pause time slightly and try again Step 9 Continue the Treatment Session A Continue the treatment session for the length of time you desire for up to 60 minutes per treatment session B BEFORE Using Recovery Edge Therapy System as a relaxing invigorating treatment before you exercise will increase circulation and can help to get you into the workout more quickly Suggested treatment time is 15 to 30 minutes before your workout
13. ctured product designed for regular use and superior operating performance This user s manual contains the information how to use keep and repair this product Please keep this manual with this system and read it before using the system for the first time Please review this manual periodically for best use of this product and to prevent potential malfunctions This product is an Advanced Air Compression Therapy System The Recovery Edge Therapy System takes air from the room environment and sends it through hoses into four individual air chambers inside the Recovery Edge Therapy System garment The Recovery Edge Therapy System fills the four chambers with air one after the other moving the treatment wave up the treated area from the base of the leg or arm towards the torso After all the chambers have been filled with air the Recovery Edge Therapy System deflates the garment for a brief pause Then the treatment wave repeats The air chambers in the Recovery Edge Therapy System garment are specially designed to overlap each other for fluent comfortable treatment without gaps You can adjust both the pressure level of the treatment and the pause time in between inflation cycles Specifications Model Class Pressure Range Supply Voltage Supply Frequency Electrical Rating Fuse Rating Case Material Dimensions Height Length Width Weight SPECIFICATIONS AND DIMENSIONS RETS P1000 T
14. e plug from the wall outlet Operating Safety Guidelines Garments Do not use garments on an undressed body bare skin Use the garments with wearing clothing as the garment s endurance can be weakened due to sweat cosmetics and sebum Do not use other manufacturer air compression pumps to blow air into the garments Do not use oil benzene alcohol gasoline on the garments Do not use garments together with accessories such as watches or jewelry that can physically damage the garment Do not keep or use garments near heating materials such as stove or cigarettes Pump Keep the device horizontally The pump s rear case should only be removed by authorized technical personnel Do not use the pump in the presence of flammable gases such as anesthetic agent Make sure that the main power cable is positioned to avoid causing a trip or other hazard and is clear of moving bed mechanisms or other possible entrapment areas Cleaning Do not use oil benzene alcohol gasoline chemicals when cleaning the device and garments Use dry towels Do not use oil benzene alcohol gasoline chemicals when cleaning garments Use wet dust cloths Be careful to not put clips staples needles metals or food into the device Do not repair the device by yourself Product Safety Features Automatic Deflation In Case of Power Failure lf a power outage occurs during use the Recovery Edge Therapy System pump will deflate the air out the garm
15. ents One way Hose Connection To ensure the air chambers will fill in the correct sequence the Recovery Edge Therapy System has one way connection The connector on the hose can be inserted into the pump in one orientation only Overload Protection Fuse The Recovery Edge Therapy System is equipped with an overload fuse Locking Knob The Lock knob locks the pressure adjustment knob in place preventing the pressure knob from being moved inadvertently COMPONENTS OF THE RECOVERY EDGE THERAPY SYSTEM The Pump The Recovery Edge Therapy System Pump is powered by electricity You will need to plug it into a wall outlet with a grounding connection The pump takes air from the room environment and sends it to the individual air chambers in the garment in a sequence that starts from the base of the garment and continues to the top The pump then vents the air from the garments there is a brief pause and the treatment cycle repeats The hose bundle transfers the air from the pump to the Recovery Edge Therapy System garments It includes four hoses a connector that attaches to the front of the pump on the Recovery Edge Therapy System Prong Plug Insert the plug into the Recovery Edge Therapy System pump to seal it off and prevent air from escaping Leg Garment Each zippered Recovery Edge Therapy System leg garment has four air chambers These garments are available in a range of sizes
16. in taking care not to allow liquid to enter the air inlets Towel dry and then allow the garment to air dry completely before use 6 If desired the inner and outer surfaces of the garment may be wiped down with a cloth or wipe moistened with a small quantity of 50 alcohol 7 Work in a well ventilated area and wear gloves 8 Allow the garment to air dry completely before use Storage 1 Store in a dry shaded place at temperatures between 4 F to 158 F 2 Roll up the power cord neatly before storing the pump Do not bend or kink the cord 3 Since the garment is inflated by air it must remain airtight Therefore avoid contact with pins needles and any other sharp objects or instruments If one of the inflatable chambers is damaged the garment must be replaced 4 The hoses should be kept untwisted and unfolded Roll them up neatly Transport 1 Transport in original packaging or in luggage with padding 15 TROUBLESHOOTING Symptom Possible Cause Corrective Action Symptom Possible Cause Corrective Action Symptom Possible Cause Corrective Action Symptom Possible Cause Corrective Action Symptom Possible Cause Corrective Action Symptom Possible Cause Corrective Action The pump is not working No electricity Check the electrical wall outlet Check the power cord visually for any defects Check the fuses and replace if necessary The pump starts wor
17. king and stops immediately The air cannot move through the hose bundle Check hose bundles for kinks twists and folds One garment inflates but the second one does not The second garment does not receive air Check its hose bundle for kinks twists and folds The pump stops working and buzzer beeps Hose bundle is not connected properly to garment or pump or Prong Plug is not inserted into unused air outlet Check and fasten all air connections If treating only one limb make sure the unused air outlet is plugged with the prong plug provided with the pump If all air connections are OK and problem persists contact the Company The pump works at a very low pressure regardless of the pressure set by the user Defective garment Replace garment and check again An internal problem Contact the Company An irregular noise Pump transferring vibrations to a surface Make sure the pump is standing evenly on all four of its bumpers An internal problem Contact the Company 16 WARRANTY Standard Warranty Terms TIME PERIOD a Pump Devices The Company warrants each Pump device to be free from defects in workmanship and materials for the period of twelve 12 months from the date of shipment and under normal use and service b Accessories Pump accessories are warranted for a period of six 6 months from the date of shipment and include the full leg half leg arm and waist garments AUTHORIZATION The Company
18. quirements of applicable Electromagnetic compatibility EMC Standards The use of accessories not specified by the manufacturer may result in increased emissions by or decreased immunity of the equipment affecting its performance Always unplug this product from the electrical outlet immediately after use Use this product only for its intended use as described in this manual Do not use with any accessories not recommended by the Manufacturer Never use pins or other metallic fasteners with this product Never operate this product if it has a damaged cord or plug if it is not working properly if it has been dropped or damaged in any manner or dropped into water or if the product shows any sign of damage or deterioration such as cracks or worn parts Call Recovery Edge Therapy System for examination and repair Do not operate pump on a soft surface such as a pillow or mattress or under a blanket or other covering Keep this product out of reach of children and pets Keep the equipment dry Do not operate while bathing in a shower in or around water or in a wet or moist condition Do not reach for the product if it has fallen into water Unplug immediately Do not carry this product by the power cord or use the cord as a handle Keep the cord away from heated surfaces Do not use where aerosol Spray products are being used or oxygen is being administered To disconnect turn the ON OFF switch to the OFF position then remove th
19. s obligation under this warranty is limited to providing any and all necessary product repairs upon the Company s examination and final determination as to cause or existence of defect and at its option to repair or replace the products which prove to be defective during the warranty period REPLACED PRODUCTS Products replaced under warranty will be warranted only for the balance of the warranty period from the original supplied equipment ORIGINAL PURCHASER This warranty extends only to the original purchaser of the equipment from the Company or its authorized distributor TRANSFER OF WARRANTY Any transfer of warranty must be approved in writing from the Company Standard Warranty Conditions This Standard Warranty will not apply to those products which have been i repaired or altered other than in accordance with the terms of this Agreement or ii abused misused improper handling in use or storage or used in an unauthorized or improper manner or without following written procedures supplied by the Company or ili original identification markings or labels have been removed defaced or altered or iv any other claims not arising directly from material defects in material or workmanship This Standard Warranty is the exclusive warranty made by Recovery Edge Therapy and is in lieu of all other warranties whether written oral or implied including any warranty of merchantability or fitness for a particular purpose and shall
20. tends to slide down use the optional belt To use the belt attach one end of the belt to the snap closure at the back of the garment then pull on the sleeve zip it closed pull the belt around your body under the opposite arm and fasten it to the other snap closure at the front When correctly fastened the belt should cross from the shoulder of the arm across the chest and under the opposite arm Tighten the belt just enough to keep the sleeve from slipping down IMPORTANT FOR ALL GARMENTS The garment should not fit too tightly you should be able to insert three fingers into the garment when it is zipped closed before inflation If the garment is too tight you will need a larger size Contact Recovery Edge Therapy System Make sure the garment is zipped completely closed before turning on the Recovery Edge Therapy System and leave it closed for the duration of the treatment At the end of treatment the garment will deflate within a short time allowing it to be removed easily You should be able to slip the garment off and pull it on again for each use without completely unzipping the garment 10 Step 4 C Putting on the Waist Garment Step 5 Connect the Air Hoses to the Recovery Edge Therapy System Pump A Securely insert the hose bundle connector with the logo facing up into one of the air sockets on the pump see Figure 5 B Make sure the air hoses are not bent kinked or pinched Figure 5 Connecting
21. y also increase the pause time see Step 8 below Table 1 Guideline for Pressure Selection Treatment Intensity Pressure Setting mmHg Light 20 30 Medium 40 60 Intense 70 80 Step 7 Lock the Pressure Knob After you have selected the pressure lock the Pressure Knob by turning the Lock Knob clockwise until you feel the knob tighten closed see Figure 9 This will prevent an inadvertent change in the pressure settings If you want to change the pressure simply loosen the Lock Knob change the pressure settings and re tighten the Lock Knob You may wait to lock the pressure settings until after you have tried your first session so that you can adjust the pressure easily during your first treatment session Figure 9 Locking the Pressure Knob 12 Step 8 Set the Pause Time celerated Recovery for Athletes A Pause is the time that the garments deflate and remain vanesa T relaxed between each treatment cycle Select this time by turning the Pause Knob see Figure 10 Changing the pause time allows you to modify the treatment by increasing the pause interval in between treatment waves Standard setting for pause time is 10 to 15 seconds Figure 10 Setting the Pause Time B Considerations for Choosing Pause Time Changing the pause time adjusts the length of the time the garments are deflated in between each treatment wave If you would like a more rapid treatment pace select a shorter p
Download Pdf Manuals
Related Search
Related Contents
Philips ChargeOn Charging base DLP2274 Toshiba Satellite S875-S7376 Patton electronic 2155 Network Card User Manual Vivitar ViviCam 3355 Digital Camera Epson 1240U Product Support Bulletin MANUAL DEL PROPIETARIO TM-WB ESPECIFICAÇÃO TÉCNICA - licit - Tribunal de Justiça de Minas Installation Manual Copyright © All rights reserved.