Home
Operating instructions
Contents
1. The DI30 has a number of unit parameters with which the function of the indicator can be adjusted to the relevant measuring tasks Because of the large number of these settings and the limited possibility of displaying them on the 7 segment display the parameters have been given consecutive numbers see chapter 6 7 Program Number Table PNO Measuring input One of the aspects in the basic configuration is the desired measuring input which consists of the terminal selection and the relevant factory or sensor calibration PN1 and PN2 Scaling The two program numbers 1 and 2 are used to scale the indication With these two parameters the end value and start value are parameterised If sensor calibration has been selected via PNO 0 then the current for the relevant sensor signal must be applied during programming Otherwise a simple allocation of the selected input configuration will be made For example at PNO 3 which corresponds to a standard signal input 0 10 V the value saved under PN1 will be indicated at 10 V and the value saved under PN2 will be indicated at 0 V PN3 Decimal point By changing this parameter the number of places shown after the decimal point in the display is changed This parameter has no influence on the scaling of the indication value only on the position of the decimal point in the display PN5 Offset shift With this parameter it is possible to carry out a parallel shift of the parameteris
2. fixed via screws for panel thicknesses from 1 mm to 10 mm Wall mounting enclosure Material ABS black cable gland Ingress protection IP 65 Dimensions 160 x 130 x 60 mm w x h x d Cable gland Cable thickness 4 0 8 0 mm Mounting Fixing holes for screws Permissible ambient conditions Operating temperature 0 60 C Storage temperature 20 80 C Humidity 0 75 relative humidity non condensing CE conformity EMC directive 2004 108 EC EN 61326 1 Emission Group 1 Class B and Immunity industrial locations Low voltage directive 2006 95 EC EN 61010 1 Items in curved brackets are optional extras for additional price For further specifications see WIKA data sheet AC 80 05 and the order documentation 10 WIKA operating instructions digital indicator DI30 4 Design and function Operating elements 1 2 3 4 1 Program key 4 Zero key 2 DOWN key 5 7 segment display 3 UP key 6 7 LEDs for displaying the alarm contacts 4 Design and function 4 1 Short description With the digital indicator DI30 standard signals DC 0 4 20 mA or DC 0 5 10 V can be measured The 4 digit display shows the measurements or the scaled value of the measurement During programming the display is used to indicate the set values and the user prompts Two relays alarm outputs 1 and 2 are available to monitor threshold values the status of which is displayed via two separate LED on the front The integrated DC 2
3. DI30 and gas cylinder scale GCS 1 Adjustment hints for digital indicator DI30 in combination with gas cylinder scale GCS 1 In the following there is an example explaining the setting parameters of DISO With the help of the gas cylinder scale GCS 1 the level of a gas cylinder has to be calculated This value shall be displayed on DI30 At a certain level an alarm will be triggered Please look at the type label for relevant datas Measuring range 0 136 1 kg Output signal 4 20 mA 2 wire Additionally the TARA weight of the empty gas cylinder which is printed on the outside of the cylinder eg TARA 20 0 kg Connection example Measurement of current signal 4 20 mA 2 wire using the integrated transmitter supply sensor 36 WIKA operating instructions digital indicator DI30 10 Abpplication example DI30 and linder scale GCS 1 Adjustment With the program numbers 0 1 2 and 3 the indicator is configured according to the connected scale For program number 5 Offset shift on display value the TARA value has to be entered as negative value The weight of the scale is subtracted automatically only the actual weight of the filling will be shown in the display PN Function Set values 0 Input signal 2 4A 20 mA 1 Scaling final value 1361 value which is shown at 20 mA Corresponds to the full measuring range of GCS 1 2 scaling starting value 0 value which is shown at 4 mA Corresponds to
4. RC spark extinguishing combinations or free wheeling diodes Keep input output and supply lines separate from one another and do not lay them parallel with each other Position go and return lines next to one another Where possible use twisted pair So you receive best measuring results screen off and twist sensor lines Do not lay current carrying lines in the vicinity Connect the screening on one side on a suitable potential equaliser normally signal ground The device is not suitable for installation in areas where there is a risk of explosion Any electrical connection deviating from the connection diagram can endanger human life and or can destroy the equipment The terminal area of the devices is part of the service Here electrostatic discharge needs to be avoided Attention High voltages can cause dangerous body currents Galvanic insulated potentials within one complex need to be placed on a appropriate point normally earth or machines ground So a lower disturbance sensibility against impacted energy can be reached and dangerous potentials that can occur on long lines or due to faulty wiring can be avoided The fuse rating of the supply voltage should not exceed a value of 1A N B fuse The manufacturer shall not be liable for claims of any type based on operation contrary to the intended use WIKA operating instructions digital indicator DI3O 2 2 Personnel qualification WARNING Risk of injury sho
5. lock is activated the key must be pressed for at least 1 second Example Change i to programming mode by pushing key P The first released program number PN P v A appears in this case PNO 22 WIKA operating instructions digital indicator DI30 6 Commissoning operation Changing to other program numbers To change between individual program numbers hold the P key down and press the A key for changing to a higher program number or the V key for changing to a lower number By keeping the keys pushed e g P amp A the display will begin after approx 1 second to automatically run through the program numbers Example Under PNO a 1 is parameterized Hold the P key down and press the A key once P y A PN1 appears in the display Under this parameter the final value of input can be changed Change to the parameter Once the program number appears in the display you can press the V or A key to get to the parameters set for this program number The currently stored parameters are displayed Example By pressing the V or A key the currently 7565H stored value for PN1 appears in the display P v A In this case it is 75 64 Changing a parameter After changing to the parameter the lowest digit of the respective parameter flashes on the display The value can be changed with the A or V key To move to the next digit the P key must be briefly pressed Once the highest digit has been
6. the beginning of the measuring range 3 Number of decimal places after separator 1 0 0 5 Offset shift on display value 20 0 negative TARA value which is printed on the gas cylinder If an alarm should be activated at an certain filling level eg 5 kg the alarm output eg A1 has to be configured accordingly Alarm output 1 60 Alarm output 1 1 activated 61 Threshold 5 Program numbers P62 P65 have to be set accordingly to your technical conditions WIKA operating instructions digital indicator model DI30 37 11 Appendix Declaration of conformit 11 Appendix Declaration of conformity EG Konformit tserkl rung Dokument Nr 11528461 01 Wir erkl ren In alleiniger Verantwortung dass die mit CE gekennzeichneten Produkte Typ D130 Beschreibung Digitalanzeige gem guitigem Datenblatt AC 20 05 die grundlegenden Schutzanforderungen der folgenden Richtliniein erf llen 2006 95 EG NSR A004 T08 EG EMV Die Ger te wurden entsprechend den folgenden Normen gepr ft EN 61010 1 2001 EN 61326 1 2006 Unkerzeichnet f r und im Namen von Signed for and on behalf of WIKA Alexander Wiegand SE amp Co KG Klingenberg 2009 11 25 Gesch ftsbereich Company division MP CT Alfred H fner Tal 4 STs 132 0 WIRA Alene Wiogand BE 6 Ca K j Fou 48 SATZ Pe An Niegend lrbe J0 Kommandigeseista Si irgerbeig Amingericht Aschotenturg HAA 1819 WIKA EC Declaration o
7. 4 V transmitter supply is used to supply connected sensors WIKA operating instructions digital indicator model DI30 11 5 Transport packagino 4 2 Scope of delivery The scope of delivery is Indicator for panel mounting Indicator 2 fixing elements Seal Operating instructions Indicator for wall mounting Indicator Operating instructions Cross check scope of delivery with delivery note 5 Transport packaging and storage 5 1 Transport Check instrument for any damage that may have been caused by transport Obvious damage must be reported immediately 5 2 Packaging Do not remove packaging until just before mounting Keep the packaging as it will provide optimum protection during transport e g change in installation site sending for repair 5 3 Storage Permissible conditions at the place of storage Storage temperature 20 80 C Humidity 0 75 relative humidity no condensation Avoid exposure to the following factors Direct sunlight or proximity to hot objects Mechanical vibration mechanical shock putting it down hard Soot vapour dust and corrosive gases Potentially explosive environments flammable atmospheres 12 WIKA operating instructions digital indicator DI30 6 Commissoning operation Store the instrument in its original packaging in a location that fulfils the conditions listed above If the original packaging is not available pack and store the instrument
8. Operating instructions Digital indicator model DI30 14037419 e V2 3 e 02 2012 Digital indicator model DI30 for panel mounting or wall mounting w Kall in Part of your business GB Operating instructions model DI30 page 1 40 2010 WIKA Alexander Wiegand SE amp Co KG All rights reserved WIKA is a registered trademark in various countries Prior to starting any work read the operating instructions Keep for later use 2 WIKA operating instructions digital indicator DI30 1 General information 42 s 02 00nn0nnnnnnnnnnnnnnnn nn nn nun nn nun nn nun nn nun nn nun nn nn 4 2 Sale mussen ee ge ee een 5 2 1 WALENOCO US Ce een 5 2 2 PELSONNEl GUalill CAV ON nesmierna A EE 7 2 3 Special hazards cccccccccsecccececeeeeceeeeeeeeeeceeeeceuceseueeseeeesseeeeseeeeseeeeees 7 24 Labeling Salely Marks een 8 3 Speeificallons ns 9 4 Desin and TuUneClIoNn enee een 11 A SNOrR ESCchHDlOn essen 11 4 2 Scope of delivery zuuuasssenssennanennnnnnnnennnnnnnnennnnnnnnennnnnnnnennnnennne nennen 12 5 Transport packaging and Storage uuunuuenannnnnnnnnnnnnnnnnnnnnnnnnnennnennnnnnen 12 S MANS OO nee ea E 12 92 FACKA eene ee E A nee 12 32 GO AS 12 6 Commissoning operation uuszanunnanunnanunnnnnnnnnnnnnnnnnnnnunnnnunnnnnnnnnn 13 OT MOUDING 2222er 13 5 2 ElecicalcConne HON rsatan aa 16 6 3 Function and operation GESCriIPtiOn cccecceeeeeeeeeeeeeeeeeeesee
9. as described below 1 Wrap the instrument in an antistatic plastic film 2 Place the instrument along with shock absorbent material in the packaging 3 If stored for a prolonged period of time more than 30 days place a bag containing a desiccant inside the packaging WARNING Before storing the instrument following operation remove any residual media This is of particular importance if the medium is hazardous to health e g caustic toxic carcinogenic radioactive etc 6 Commissoning operation Please read the safety instructions and installation instructions in chapter 2 before installation and keep this user manual for future reference 6 1 Mounting Indicator for panel mounting Before assembly a cut out must be made to accommodate the device The sizes and tolerances are given in the technical data On front of the DI30 are the operating and display elements On the sides are the fixing elements to mount the device in the panel On the back are the terminals for all the electrical connections A gasket is inserted between the contact surface of the front collar and the control panel WIKA operating instructions digital indicator model DI30 13 6 Commissoning operation 1 After removing the fixing elements insert the device 2 Check the seal to make sure it fits securely 3 Click the fixing elements back into place and tighten the clamping screws by hand Then use a screwdriver to
10. ctions not be observed there is a risk of serious or fatal injury WARNING Before installation commissioning and operation ensure that the appropriate instrument has been selected in terms of measuring range design and specific measuring conditions Non observance can result in serious injury and or damage to equipment Further important safety instructions can be found in the individual chapters of these operating instructions 2 1 Intended use The DI30 is designed for the evaluation and display measuring transmitter signals With the alarm outputs it is possible to perform simple control tasks The instrument has been designed and built solely for the intended use described here and may only be used accordingly WIKA operating instructions digital indicator model DI30 5 Please read the following safety advice and the assembly before installation and keep it for future reference If the instrument is transported from a cold into a warm environment the formation of condensation may result in the instrument malfunctioning Before putting it back into operation wait for the instrument temperature and the room temperature to equalise Notes on installation There must be no magnetic or electric fields in the vicinity of the device e g due to transformers mobile phones or electrostatic discharge Do not install inductive consumers relays solenoid valves etc near the device and suppress any interference with the aid of
11. e e g PN103 so that it is taken over by the indicator Otherwise no confirmation will be shown On the other hand the relevant indication values do not need to increase constantly They can also fall or alternate between rising and falling Example Linearisation of a pressure transmitter 0 100 mbar with output 0 20 MA To program e g 5 additional calibration points 5 must be entered under PN100 Subsequently for each of the calibration points the voltage current must be applied to the unit and the respective indication value programmed under the following program numbers PN101 PN105 The indicated value before correction can be either calculated from the known characteristic line of the transmitter or determined empirically The non linear range is between 0 75 mbar For calibration point 101 this means A pressure of 15 mbar the transmitter delivers 3 3 mbar instead of the ideal value of 3 0 mbar Since 20 mA in the display corresponds to 100 0 mbar 3 3 mA in the display corresponds to 16 5 mA before the correction Setpoint Pressure Output Indication Desired PN mbar transmitter before indication mA correction IN OUT 2 0 5 2 5 0 0 101 15 3 3 16 5 15 0 102 30 6 2 31 0 30 0 103 40 9 2 46 0 40 0 104 60 11 4 57 0 60 0 105 75 14 7 73 5 75 0 1 100 20 0 100 0 100 0 WIKA operating instructions digital indicator model DI30 25 6 Commissoning operation 6 6 Description of device parameters
12. ed characteristic line This may be necessary if for example a pressure transmitter ages over the course of time giving rise to a shift in the zero point With the parallel shift the transmitter can be adjusted back to the zero point Another application is to parameterise a certain tank level to zero and have any deviation from this level displayed PN10 Zero point suppression Via the zero point suppression an indication value window can be defined as zero This means for example that at PN10 10 all indication values between 10 10 are shown in the display as zero This function is intended to produce a reliable zero indication at high display resolution and low sensor accuracy around zero This could be for example the rom of an engine for which a zero would be expected in the display when standing still 26 WIKA operating instructions digital indicator DI30 6 Commissoning operation PN13 Indication time The indication time is the interval at which the display is updated The longer the time between two indication cycles the more calm the display As a rule the eye perceives the indication time of 1 sec as very pleasant If the adjusted indication time is longer than the adjusted measuring time PN14 no averaging of the detected measuring values during the display time takes place The display will be always refreshed with the newest detected measuring value PN14 Measuring time The measuring time corresponds t
13. eeeeenees 19 04 AGOGO IK ANNO 22 6 5 Description of the measuring input uus2u04440Hene nenn nenn nenn nenn nennen 24 6 6 Description of device parameters cccccceeeeseeeseeeneeeneeeneeeneeeseeenees 26 6 7 Program number table u2200200200200 nn nnnnnnnnnennnnnnnnnennnnnnnenennn 30 626 DETAUIEVAlNES ee een een 32 7 Maintenance and cleaning 2u02200 200200an00n0nnunnnnnnunnnnnnnn nun nn 32 fat NAMEN AN nneer 32 7 2 eN 32 8 False 33 9 Dismounting return and disposal uu uu00au00an0nan0nanunanunanunnnunnnen 34 9 1 DISMOUNING euere 34 I 2 RUN ee E A E AAEE E E EE 34 LI DIS 6 19 otc Ree ee ee ee een 35 10 Application example DI30 and gas cylinder scale GCS 1 36 11 Appendix Declaration of conformity u u00au00u0nan0nannnnunanennn 38 Declarations of conformity can be found online at www wika com WIKA operating instructions digital indicator model DI30 3 1 General information 1 General information he instrument described in the operating instructions has been designed and manufactured using state of the art technology All components are subject to stringent quality and environmental criteria during production Our management systems are certified to ISO 9001 hese operating instructions contain important information on handling the instrument Working safely requires that all safety instructions and work instruction
14. f Conformity Document No 11528461 01 We declare under our sole responsibility that the CE marked products Model DI30 Description Digital Indicator according to the valid data sheet AC 80 05 ere in conformity with the essential protection requirements of the diractiva s 2006 95 EC LVD 2004108 EC EMC The devices had been tested according to the following standards EN 61010 1 2001 EN 61326 1 2006 Quelit tsmanagernent Quality manager MP CT Wendl Harald Hartl Unterschrifl autorisiert durch das Unternehmen Signature authorized by the companiy Bean ph ran WIKA ninmational SS Si Kingenierg BSS Kingenberg Ed ade Kanplassrnl rn WIA Verwaltungs SE Bo KIG Asiesarchi Aschaffenburg HAB 100 Canney wwe wil do air Hingenberg Arrkgeneht Macha Ren bun Yoani Aecander Wiegand HRA at Worsikzender des Aufsichisrate Dr Wow Egi 38 WIKA operating instructions digital indicator DI30 11 Appendix Declaration of conformit WIKA operating instructions digital indicator model DI30 39 WIKA global WIKA subsidiaries worldwide can be found online at www wika com wi KAI WIKA Alexander Wiegand SE amp Co KG Alexander Wiegand Stra e 30 63911 Klingenberg Germany Tel 49 9372 132 0 Fax 49 9372 132 406 E Mail info wika de www wika de 40 WIKA operating instructions digital indicator DI30
15. h a measuring input for standard signals that enables standardised signals e g 4 20 mA from all kinds of different measuring transmitters on the market to be measured directly Factory calibration For this various sensor values for 0 10 V 0 5V 0 20 mA and 4 20 mA are stored in the unit and can be called up via the parameter PNO They are called factory calibrations because the data were established during production and are saved permanently in the unit It means that a pre adjusted transmitter can be operated directly with the indicator without any need to previously connect the signal to be measured to the indicator The indicator can be scaled freely according to the physical dimension to be measured Sensor calibration If on the other hand the sensor has not been pre calibrated the indicator can be adjusted and calibrated direct via the measurement together with the sensor path This can be selected via the parameter PNO O and is consequenily called sensor calibration 24 WIKA operating instructions digital indicator DI30 6 Commissoning operation Sensor linearisation In addition non linear sensors can be linearised with the aid of a characteristic line that can be saved in the indicator This is described with the following example The sensor signal must be parameterised in a strictly monotonously manner i e every new calibration point e g PN104 must have a higher input signal than the previous on
16. o the conversion time of the A D conversion which determines the response time of the alarm outputs The longer the conversion time the smaller the influence of disturbances and the higher the resolution of the measured signal PN50 to PN52 Security setting user level With the parameters in the security settings access to the program numbers is regulated through the setting of various user levels The user levels divide the access into various levels The user is only given access to the settings authorised by the system operator such as the setting of thresholds The lower the figure for the user level given in PN52 the lower is the security level of the unit parameters against user intervention e available function at the selected userlevel Userlevel PN 52 Accessto PN a N E a E O A A a Measuring input parameters 0 3 5 10 13 74 elele T TT ejeje JTI Linearization parameters for measuring pe eee input Parameters of the alarm outputs Pr 65 70 Hysteresis ofthe alarm outputs jan jeleleleeie Thresholds of the alarm outputs je e elele e ele Accessos o o eebe Serial number o feleleleleteleieie The user levels 1 3 and 7 are reserved user levels at which access to the higher numbered user level is active The parameterised user level PN52 is active as long as the locking code PN51 and the access code PN50 are different On delivery of the DI30 both parameters are set to 0000 so that
17. perating instructions digital indicator model DI30 33 9 Dismounting return and disposal WARNING Residual media in dismounted instruments can result in a risk to persons the environment and equipment Take sufficient precautionary measures 9 1 Dismounting To dismount the instrument do the following steps Indicator for panel mounting 1 Open the clamping screws and remove the fixing elements 2 Remove the instrument and seal from panel cutout Indicator for wall mounting For dismounting the indicator open the four fastening screws of the enclosure and remove it 9 2 Return WARNING Strictly observe when shipping the instrument All instruments delivered to WIKA must be free from any kind of hazardous substances acids bases solutions etc When returning the instrument use the original packaging or a suitable transport package Enclose the completed return form with the instrument The return form is available on the internet www wika de Service Return 34 WIKA operating instructions digital indicator DI30 9 3 Disposal Incorrect disposal can put the environment at risk Dispose of instrument components and packaging materials in an environmentally compatible way and in accordance with the country specific waste disposal regulations WIKA operating instructions digital indicator model DI30 35 10 Abpplication example DI30 and linder scale GCS 1 10 Application example
18. pply as long as their current consumption is not higher than 50 mA Power consumption lt 1 2 VA Transmitter suppl 13 14 Transmitter supply Transmitter supply 6 2 2 Connection Examples This section gives a few examples of practical connections a 4 20 mA signal 2 wire transmitter b 0 20 mA signal 3 wire transmitter The voltage drop on the transmitter supply wire is irrelevant which is why this can also be bridged directly on the indicator sensor WIKA operating instructions digital indicator model DI30 17 6 Commissoning operation c 0 5 V 0 10 V signal 3 wire transmitter The ground is bridged via the sensor which eliminates the measuring error through the drop in voltage on the transmitter supply wires transmitter supply oln a i e sensor 18 WIKA operating instructions digital indicator DI30 6 Commissoning operation 6 3 Function and operation description 6 3 1 Operation The DI30 has 4 keys P A V O with which you can parameterise and call up various functions during operation Additionally it has a 4 digit 7 segment display and two LEDs for displaying the alarm contacts 1 2 3 4 Name Description 1 Program key P With the program key you can call up the programming mode or perform various functions in the programming mode 2 DOWN key V With the DOWN decrease key you can call up the MIN memory or alter pa
19. r memory Data storage gt 20 years Input Number and type 1 input for standard signals Input signal 4 20 mA input resistance lt 100 Q or 0 20 mA input resistance lt 100 Q or DC 0 5 V input resistance 2 150 kQ or DC 0 10 V input resistance 2 150 kO Input configuration selectable via terminal configuration and programming Accuracy 0 1 of the measuring span 1 Digit Temperature drift All measuring inputs 50 ppm K Measuring principle Voltage Frequency converter Resolution approx 20 bit at measuring time Isecond Measuring time 0 1 10 sec Transmitter supply DC 24 V max 50 mA galvanically isolated Power supply AC 230 V 50 60 Hz 10 AC 115 V 50 60 Hz 10 Power consumption Max 8 VA Electrical connection removable screw terminals Line cross section up to 2 5 mm Alarm output Number and type 2 independent switching contacts relays freely adjustable Contact load AC 230 V 5A resistive load DC 30 V 5A resistive load Switching cycles 0 5 10 at max contact rating 5 10 mechanically Separation as per DIN EN 50178 Characteristics as per DIN EN 60255 WIKA operating instructions digital indicator model DI30 9 3 Specifications Material Aluminium black Ingress protection Front IP 54 Rear IP 00 Dimensions 96 x 96 x76 mm w x h x d including terminals Mounting clearance horizontal 120 mm vertical 120 mm recommended Weight Approx 530 g Mounting sliding fasteners
20. rameters in the rogramming mode 3 UP key A With the UP increase key you can call up the MAX memory or alter parameters in the rogramming mode 4 Zero key O Zero key for TARA and HOLD function 5 segment display In the 7 segment display measuring values are displayed respectively during programming the program numbers or parameters are displayed 6 7 Alarm displays The alarm display indicates the state of the relays of the alarm outputs If a relay is switched the LED lights up WIKA operating instructions digital indicator model DI30 19 6 Commissoning operation 6 3 2 Switching on Before switching on you have to check all the electrical connections to make sure they are correct On completion of the installation the device can be switched on by applying the power supply During the switching on process a segment test is performed for approx 1 second whereby all LEDs on the front including alarm LEDs are triggered After this the type of software is indicated for approx 1 second and then also for 1 second the software version After the starting procedure the unit changes to operation display mode 6 3 3 General functions MIN MAX Memory The measured minimum and maximum values are saved in a volatile memory in the unit and get lost when the unit is switched off You can call up the contents of the memory by pushing less than 1 second the A or WV key The relevant value is indicated for approx 7
21. s are observed Observe the relevant local accident prevention regulations and general safety regulations for the instrument s range of use The operating instructions are part of the instrument and must be kept in the immediate vicinity of the instrument and readily accessible to skilled personnel at any time Skilled personnel must have carefully read and understood the operating instructions prior to beginning any work he manufacturer s liability is void in the case of any damage caused by using the product contrary to its intended use non compliance with these operating instructions assignment of insufficiently qualified skilled personnel or unauthorised modifications to the instrument he general terms and conditions contained in the sales documentation shall apply Subject to technical modifications Further information Internet address www wika de www wika com Relevant data sheet AC 80 05 Application consultant Tel 49 9372 132 0 Fax 49 9372 132 406 E Mail info wika de 4 WIKA operating instructions digital indicator DI30 Explanation of symbols 2 Safety WARNING Indicates a potentially dangerous situation that can result in serious injury or death if not avoided Information points out useful tips recommendations and information for efficient and trouble free operation DANGER dentifies hazards caused by electric power Should the safety instru
22. seconds By briefly pressing the same key again you will return immediately to the display mode A Display of the MAX value V Display of the MIN value You can erase the value shown in the display by simultaneously operating the A amp VY keys The erasure is acknowledged by horizontal bars The content of the memory is lost when the unit is switched off Overflow and Underflow An overflow of the display is indicated by horizontal bars at the top of the 7 segment display An underflow of the display i is indicated by horizontal bars at the bottom of the 7 segment display i 20 WIKA operating instructions digital indicator DI30 6 Commissoning operation 6 3 4 Description of the alarm outputs With the aid of the LEDs below the 7 segment display you can view the switching state of the alarm outputs An active relay is indicated by the relevant LED lighting up Working principle The alarm outputs have the following properties with regard to their switching properties Parameter Description Alarm output Deactivated activated Threshold Threshold value for switchover Hysteresis Width of the window between the switch thresholds Operating principle Active above SP value active below SP value Switch on delay Time between reaching the threshold and the resultant switching on of the relay Switch off delay Time between reaching the threshold and the resultant switching off of the relay Ac
23. set and confirmed with P the lowest digit will begin to flash again Example U The 4 is flashing this is the lowest value digit and by flashing it is asking for a figure to be entered In our example the value is to be P y A changed from 75 64 to 75 00 You can change the value by changing the figure from 4 to 0 using the A or V key To move to the next digit the P key must be briefly pressed The 6 begins flashing Change the value from 6 to 0 using the A or V key The 5 and the 7 need no change WIKA operating instructions digital indicator model DI30 23 6 Commissoning operation Saving parameters All parameters must be acknowledged by the user by pressing the P key for one second Ihe changed parameters are then taken over as the current operating parameters and saved in the EEPROM This is confirmed by El bars lighting up in the display Example Save the the parameters by pressing P for one second All the newly entered data are confirmed by the unit If no confirmation is received the relevant parameters have not been saved Example You receive confirmation from the unit that the changes have been saved through the appearance of horizontal bars in the middle Piiviial segments Change to the operating mode If no key is pressed in programming mode for approx 7 seconds the unit automatically returns to operating mode 6 5 Description of the measuring input The DI30 is equipped wit
24. switch on delay 2 switch off delay 3 switch on off dela 1 activated 71 Threshold 999 9999 1000 6 72 Hysteresis 1 9999 1 5 cae an 1 threshold value 1 active above 74 Switch delay 0 1000 seconds 1 4 19 Delay type 0 none 2 switch off delay 3 switch on off dela Linearisation eo Pe setpoints pyem m u 110 0 9999 o j8 gt O A gt WIKA operating instructions digital indicator model DI30 31 7 Maintenance and cleaninco 6 8 Default values Reset to default values To return the unit to a defined basic state a reset can be carried out to the default values The following procedure should be used Switch off the power supply Press button P Switch on the power supply and press P for further approx 2 seconds until is shown in the display With reset the default values of the program table are loaded and used for subsequent operation This puts the unit back to the state in which it was Supplied Caution This is only possible when the programming lock PN50 allows access to all PNs or HELP is shown in the display Caution All application related data are lost 7 Maintenance and cleaning 7 1 Maintenance This instrument is maintenance free Repairs must only be carried out by the manufacturer 7 2 Cleaning CAUTION Before cleaning correctly disconnect the instrument from the mains Clean the instrument
25. terminals are designed as removable screw type terminals with a grid pitch of 5 08 mm This makes it possible to connect wires of up to 2 5 mm 6 2 1 Terminal configuration 9 pole connector strip On the 9 pole connector strip with terminals 1 to 9 the power supply supply voltage of the digital indicator and the two relay outlets of the alarm outputs are connected Power supply 4 5 6 N PE L Relay 2 A2 Relay 1 A1 1 2 3 7 8 9 Com Normally Normally Com Normally Normally open closed open closed 5 pole connector strip On the 5 pole connector strip with terminals 10 to 14 the input signal and the transmitter supply are connected Measuring input for standard signals The input signal is connected to terminals 10 to 12 The DI30 has a voltage input and a current input to which most conventional sensors with a standard signal output can be connected The measuring input is galvanically isolated from the power supply Supply voltage and the transmitter supply Input for standard signals Sensor signal Measuring range 10 11 12 U U en 12 12 V I a ss 24 24 mA 16 WIKA operating instructions digital indicator DI3O 6 Commissoning operation Transmitter supply The DC 24 V transmitter supply is available at terminals 13 and 14 The transmitter supply is galvanic insulated from the measuring input Current loop sensors 2 wire and 3 wire and 4 wire sensors can be operated via the transmitter su
26. the programming lock is deactivated WIKA operating instructions digital indicator model DI30 27 6 Commissoning operation To activate the set user level you must enter a 4 digit number under PN51 as a locking code and confirm it by pressing the P key for approx 1 second On changing to programming mode the unit jumps to the first authorised program number If user level PN52 3 then for example the parameters of the alarm outputs can be changed but changing the parameter of the measuring input PNO is not possible at this user level In order to obtain access to all program numbers later equivalent to user level 0 you have to enter under PN50 the same code you entered before under PN51 You must then acknowledge this by pressing the P key for approx 1 s After this you have access to all program numbers Information lf the locking code becomes lost the unit can be set to the default value 0000 at the manufacturer s without any data loss PN54 Allocation of functions of the zero key By pushing of the zero key O functions like e g TARA and HOLD can be actuated or an offset shift can be preset By allocation of TARA the key needs to be pushed for at least 1 second The device confirms the function by showing 0000 in the display With activated offset shift the value in PN5 will be added to the initial value directly by pushing the key A display hold can only be actuated for approx 20 seconds By p
27. tighten them another half a turn CAUTION The torque should not exceed 0 1 Nm Indicator for wall mounting The wall type housing consists of an upper part and a lower part A suitable seal has been inserted in the lower part to achieve protection IP 66 In the upper part is the DI30 with its operating and display elements On the back are the terminals for all the electrical connections 14 WIKA operating instructions digital indicator DI30 6 Commissoning operation m 139 0 1 The cover flaps have to be opened on the right and left side Loose the four fixing screws which connect the upper and lower part of the enclosure with a screw driver Remove the upper part Fasten the enclosure to the wall Connecting cables have to be fed into the enclosure via the PG screws After connection the upper part of DI30 will be placed on the lower part Using the four fixing screws screw the upper part to the lower part Be sure that the screws are completely tightened in order to attain ingress protection IP66 Finally close the cover flaps again UN Screws for wall mounting are not part of scope of delivery Please use screws according to the material of the wall where you want to install the enclosure WIKA operating instructions digital indicator model DI30 15 6 Commissoning operation 6 2 Electrical connection All the necessary signals for operation are connected to the rear terminals The connecting
28. tive above threshold alarm output E The alarm output is off N D below the threshold and 5 switched on when reaching ON a A the threshold l l OFF Ceapa hysteresis measuring value O alarm output E Active below threshold D The alarm output is on below the threshold and ON PA a switched off on reaching the threshold l OFF m 4 lem hysteresis measuring value WIKA operating instructions digital indicator model DI30 21 6 Commissoning operation Switch on delay The relay is on e g 10 seconds after reaching the threshold Briefly exceeding the threshold does not lead to ar the relay being switched on treshold time s time s The switch off delay functions in a similar manner in other words it keeps the alarm output switched on until the parameterised time has elapsed Optical response display flashing The switching on of one or more alarm outputs can also be set to trigger a flashing of the display to enhance the optical response 6 4 Programming The display shows the program numbers PN right aligned as a 3 digit number with a P at the front 7 71 Display of e g program number 0 l Ll Programming Procedure The entire programming of the DI30 is done by the steps described below Change to the programming mode Pushing the P key changes to programming mode The unit goes to the lowest available program number When the programming
29. uld qualification be insufficient Improper handling can result in considerable injury and damage to equipment The activities described in these operating instructions may only be carried out by skilled personnel who have the qualifications described below Keep unqualified personnel away from hazardous areas Skilled electrical personnel Skilled electrical personnel are understood to be personnel who based on their technical training knowledge of measurement and control technology and on their experience and knowledge of country specific regulations current standards and directives are capable of carrying out work on electrical systems and independently recognising and avoiding potential hazards The skilled electrical personnel have been specifically trained for the work environment they are working in and know the relevant standards and regulations The skilled electrical personnel must comply with current legal accident prevention regulations 2 3 Special hazards DANGER Danger of death caused by electric current Upon contact with live parts there is a direct danger of death Electrical instruments may only be installed and mounted by skilled electrical personnel Operation using a defective power supply unit e g short circuit from the mains voltage to the output voltage can result in life threatening voltages at the instrument WARNING Do NOT use this product as safety or emergency stopping device or in an
30. ushing the zero key again one turns back into the current measurement directly During HOLD phase the display flashes PN59 Display flashing By activating the display flashing various alarm states can be optically reinforced The trigger for the display flashing can be freely assigned to the setpoint outputs PN60 to PN75 Alarm outputs The behaviour of the alarm outputs can be influenced via various program numbers The data refer to the scaled measurement and are updated with the set measuring time A description of the various parameters is given in chapter 6 3 4 Description of alarm outputs 28 WIKA operating instructions digital indicator DI30 6 Commissoning operation PN100 to PN110 Linearisation Through linearisation the user has the possibility to linearise a non linear sensor signal A detailed description can be found in the chapter on Sensor linearisation Under PN100 the desired number of additional set points can be released Only after changing the number are they accessible via the configuration PN101 to max PN110 If PN100 0 then no PN101 will be displayed To monitor the desired linearisation all the set points should first be parameterised otherwise there may be deviations from the desired value in the display PN200 Serial number Under PN200 the 4 digit serial number can be called up that allows allocation to the production process and the manufacturing procedure This parameter can only be
31. viewed WIKA operating instructions digital indicator model DI30 29 6 Commissoning operation 6 7 Program number table The program table lists all the program numbers PN with their function range of values default values and user level ange of value User level R Measuring input Current Voltage 2 2 0 Sensor calibration 1 0 20 mA 2 4 20 mA 3 0 10V 4 0 5V Parameters 1 to 4 make use of the factory calibration Scaling final value 999 9999 1000 Scaling starting value 999 9999 o f2 3 0 0 Number of decimal places 2 1 0 0 2 0 00 3 0 000 Offset shift on display 999 9999 ae value 10 _ Zeropointsuppression 10 99 o 2 General settings Security settings 50 __ Access code 0000 99 10000 8 51 Locking code 10000 999 0000 0 52 Userlevel BB Function of the zero ke Actuator via zero key O deactivated 2 1 TARA 2 Offset shift 3 HOLD O1 Flashing of the LED displa 59 Display flashing approx 4 0 5 seconds No flashing 0 no flashing Flashes at alarm 1 1 flashes at 1 Flashes at alarm 2 2 flashes at 2 Flashes at alarm 1 and 2 3 flashes at 1 and 2 30 WIKA operating instructions digital indicator DI30 6 Commissoning operation Range of value User level NA output 1 0 deactivated 1 O 999 9999 1000 6 62 Hysteresis 1 9999 1 5 threshold value 1 active above 65 Delay type 0 none 4 1
32. with a moist cloth Electrical connections must not come into contact with moisture For information on returning the instrument see chapter 9 2 Return 32 WIKA operating instructions digital indicator DI30 8 Faults Fault Solution The unit permanently indicates The input has a very high overflow measurement check the measuring 4 bars in the upper part of the circuit display The unit permanently indicates The input has a very low underflow lt measurement check the measuring 4 bars in the lower part of the circuit display The device shows HELP in the 7 The unit has found an error in the segment display configuration memory Perform a reset on the default values and reconfigure the unit according to your application Program numbers for parameterising The program lock is set to a user the input are not available level that does not permit access Err1 lights up in the 7 segment This error can only be eliminated by display the manufacturer Please return to the manufacturer CAUTION If faults cannot be eliminated by means of the measures listed above the instrument must be shut down immediately and it must be ensured that pressure and or signal are no longer present and it must be prevented from being inadvertently put back into service In this case contact the manufacturer If a return is needed please follow the instructions given in chapter 9 2 Return WIKA o
33. y other application where failure of the product could result in personal injury or material damage Failure to comply with these instructions could result in death or serious injury and material damage WIKA operating instructions digital indicator model DI30 7 2 4 Labelling Safety marks Product label WIKA Alexander Wiegand SE amp Co KG C lt 4 WIKA 63911 Klingenberg Germany www wika de Digitalanzeige Digital Indicator DI30 5 1 Hilfsenergie Power Supply AC 230 V 50 60 Hz a Bestell Nr Order No 4282686 LL 2 Bestell Order Code DI30 S H25ZB Z 3 Serien Nr Serial No E power S2 supply S1 input D 2 T L PENT Sa 4 6 1j2 s 4 s5 6e 7 8 9 max max u AC 230V 5A AC 230V AC 230V 5A DC 24V DC 30V 2A 50 60Hz DC 30V 2A 50mA Made in Germany 0 4 20mA 1 power supply 4 CE mark 2 order code 5 product name 3 serial no 6 connection diagram Explanation of symbols Before mounting and commissioning the instrument ensure you read the operating instructions CE Communaute Europeenne Instruments bearing this mark comply with the relevant European directives 8 WIKA operating instructions digital indicator DI30 3 Specifications 3 Specifications Specifications Display Principle 7 Segment LED red 4 digit Character size 20 mm Indication range 999 9999 Indication time 0 1 10 sec Memory EEPROM Paramete