1. From the system configuration main menu shown in Figure 1 select ADJUST DISPLAY You will see the Display Options menu shown in Figure 2 Selecting EXIT returns the display to the Main System Configuration menu shown in Figure 1 Select ADJUST CONTRAST from the menu You will see the Progeny logo Use the up and down arrows to increase or decrease the contrast between the menu text and the display background Press the right arrow to save your settings Select REVERSE IMAGE from the menu The text and display background colors will be swapped Press the right arrow to save your settings 23 00 02 1601 Rev F VetVision DC Changing Pre programmed Exposure Settings Displaying the Preset Options Menu Figure 3 Preset Options Menu Changing All Receptor Settings Globally The VetVision DC Dental X Ray System allows the operator to increase or decrease image density for all presets for a receptor simultaneously or to change each of the technique factors for a preset individually You can also restore factory default settings For charts of the factory default settings refer to Factory Default Exposure Settings later in this manual Note Before changing presets use the table on page 14 to write down the presets you are programming 1 From the Main System Configuration menu shown in Figure 1 select CHANGE PRESETS You will see the Preset Options menu shown in Figure 3 2 Selecting EXIT returns the display to the
2. 847 415 9810 www progenyvetimaging com O Progeny Dental 2008 U S Patents D470237 D469182 D470589 6 837 468 and 6 664 853 Table of Contents El it EE Cee ral ale 110 AOR EEE EEE 1 VetVision DC Dental X ray Wall mounted Model 1 VetVision DC Dental X ray Mobile Gvstem 2 Compliance with Applicable Standards eres 3 Cerca elle e 3 EG Declaration o Com aa pvr 4 Authorized Hepresentaiives eee nnne 9 DIC ceara a dM eL ELLE NL 5 Explanation of Symbols on Technical Labels 6 Obtaining Technical SUDDOMN EE 6 Operating the VetVision DC Dental X Ray System 7 Usingitne Operalor Ranes 8 Taking an X Ray Using Preset Techniques eese 9 Pre programmed Exposure Tmes eese 11 Baseline Technique Charts CGanme e ereereeren 12 Feline Technique Geitnge e reeeereeeeeereeeerrreranna 13 Record Your Exposure Settings CLanime eese 14 Record Your Exposure Settings Femme 15 System o ontig balol sacos EE Res matte EE 16 System Configuration Mode cccoocccccccncccccncconcnccnocnnonacnnnoncnononnnononenononens 16 AOUSUIGTIE Re EE 17 Changing Pre programmed Exposure Settings occccoccncccocncccnnncnonos 18 Showing Current System Confguraton erre 20 PENNEN EN 21 Trolble S ie el dn Le eee 21 Penormance oisi SNR nal 22 sp o 22 Recommended Maintenance ccocccccccnccncnncncnnnncnnnnnnonnnnnnannnnannnnnnnonannonann
3. Pho ne 1 847 415 980 Toll Fre e 8 88 924 380 Fax 1 11 all e FP 41 vi FEET XE IV p JE W progeny C Fu ed 00 02 1601 Rev F VetVision DC 847 415 981 0 tec hsu ppo rio pro gen yde ntal CO m Serial Number SN numbers are located on the tube head or on the base of mobile systems Location y 12 P a f r YE a f T F ir L A A F Ya A Y I UJ e 00 02 1601 Rev F VetVision DC Operating the VetVision DC Dental X Ray System Using the Operator Panel ICONS 1 coca 9995 e B e 8 H 9 GO 0 E Qu mn D a a e Large LED screen displays technique settings It also displays menu selections when the system is in menu mode Up and Down arrows are used to change kV mA and time settings Tooth Icon Pressing this button allows the user to select Upper or Lower Canine Incisors Premolars and Molars Receptor Icon Choices are Digital D Speed film E F Speed Film If using phosphor plates select E F Speed Film Setting Patient Size Icon Press to select Large or Small Ready Indicator Circle lights up to indicate that the system is ready to produce x ray Right Arrow Button Use this button to move between kV mA and time selections This button is also used as an Enter key when the system is in menu mode Radiation Indicator This symbol lights up
4. KL ZA S Fa D LY I C eer 00 02 1601 Rev F VetVision DC Ra di ati on Sa fet rts ar nour Troen L SE X 4 E L Only qualifi ed and author ized perso nnel may operat e this equip ment obser ving all laws and regula tions conce rning radiati on protec tion wa s 3 0 00 02 1601 Revit e fo VetVision DC e O e nl ec y tri q U cal ali Sa fi e fet E y a n d a ut h or IZ e d S er vi C e p er S O n n o3Oo0cooo cooovocooo cog3gococoo hi 00 02 1601 RevSF e q VetVision DC Ex pl OS Io n Sa fet Pinto B d Es DIY A 1 ri A 9 This equipm ent must not be used in the presen ce of flamma ble or potenti ally explosi ve gases or vapors which could ignite Causing person al injury and or damag e to the equipm ent If such disinfec tants are used the vapor must be allowed to dispers e before using the equipm ent Ms Cy 00 02 1601 Rev F VetVision DC y progeny 00 02 1601 Rev F VetVision DC Explan ation of Symb ols on Techni cal Labels Typ e B An Prot Con N sult writt AT V TE NTI WA LARNI NE Eart Lr Gro Obtaini ng Techni cal Suppor g progeny 00 02 1601 Rev F VetVision DC Co Pro gen nt y ac iA t Mid mar Co mp any 675 Hea thro Driv Linc olns hire IL 600 69
5. when an x ray is produced Exposure Button Pressing this button will produce an x ray exposure 13 MAPAS TT p PIOSCEIT 00 02 1601 Rev F VetVision DC Using the Operator Panel Power On Settings Display Selecting Techniques Technique Tables Adjusting Exposure Settings Exposure Button and Ready Indicator Ready Indicator Radiation Indicators When the VetVision DC is powered on the Operator Panel will display the selections that were used when the system was last powered off The Operator Panel displays the exposure settings KV mA and exposure time for the currently selected tooth image receptor type and patient size To select the desired technique for a procedure 1 Press the Tooth Selection Icon until the correct tooth anatomy is lit 2 Press the Image Receptor Icon to select the type of receptor you will be using 3 Press the Patient Size Icon to choose large or small The appropriate kV mA and time settings will now appear in the display and the system is ready to take an x ray For a table of the factory programmed exposure settings refer to the Pre programmed Exposure Settings tables on page 11 Preset exposure settings can be adjusted prior to making an exposure Use the right arrow to select the exposure setting to adjust Then use the up and down arrow buttons to adjust the value To save new presets refer to System Configuration mode on page 16 The Exposure button is used to i
6. 65 kilovolts 7 milliamperes and the exposure time of one second Make five exposures at this level observing the normal cooling time Select 70 kilovolts 6 milliamperes and the exposure time of one second Make five exposures at this level observing the normal cooling time ORE SL DE SA RE 32 PP vi TE a f y T FA i L II JS LL I C JP e 00 02 1601 Rev F VetVision DC Specifications VetVision DC Dental X Ray System The following specifications contain information required to be provided to the user per Federal Regulation 21 CFR Line Voltage Line Load Maximum Rated Tube Potential kVp Accuracy Tube Current Exposure Time Timer Accuracy Source to Skin Distance Minimum Half Value Layer Minimum Inherent Filtration Focal Spot Automatic Cooling Time Leakage Technique Factors Target Angle Operating Temperature Storage Temperature Maximum Altitude Cone Focal Length Diameter of X Ray 100 230 VAC 10 50 60 Hz Max current 5 amps 70 KVp 5 selectable 4 7 mA 1 mA 20 ms through 2 00 seconds 5 1 ms 8 inch 20 cm 1 7 mm Aluminum equivalent at 70 kVp 2 mm Aluminum equivalent at 70 kVp 0 4 mm IEC 336 15 times the exposure time wait before the next exposure can begin 1 5 mA at 70 kVp 12 5 degrees 50 F 95 F 10 C 35 C 31F 150 F 35 C 66 C 12 000 ft 8 inch 20 cm 2 inch 6 9 cm at the end of the Cone opecifications for optional c
7. Main System Configuration menu shown in Figure 1 1 Select ALTER DENSITIES from the Preset Options menu The first Image Receptor Type illuminates The display shows the selected Image Receptor Type and current density 2 Using the Image Receptor Type button select the image receptor to adjust 3 Use the up and down arrow buttons to specify a percentage by which densities will be increased or decreased for the selected receptor Densities can be increased or decreased according to values provided on the display 4 Press Enter to save your settings 24 00 02 1601 Rev F Preprogramming to Digital Sensors Changing Presets Individually Recall Presets A VetVision DC Energize the system Press the Tooth Selection and Patient Size Selection buttons for five full seconds Select CHANGE PRESETS from the Menu Options screen Select SELECT RECEPTOR from the Preset Options menu Figure 3 Press the up or down button to highlight the sensor or phosphor plate to change and press Enter Select YES or NO on the Verification screen Exit the Preset Options menu Exit the Menu Options menu A message of Saving Settings will display briefly and then the system will return to the normal operational mode Note When you are working in service mode the green light next to the exposure button will blink Select EDIT PRESETS from the Preset Options menu The display notifies you that you are entering Edit Preset Mod
8. NFIG The display will show e Current software version e Cone size e Diagnostic mode on or off Press any button on the Operator Panel to return to the Configuration menu CONFIGURE UNIT SET CONFIG SHOW MAINT EXIT 26 g progeny DENTAL 00 02 1601 Rev F VetVision DC Diagnostic Mode About Diagnostic Mode Showing the Maintenance Summary Figure 5 Set Configuration Menu The VetVision DC Dental X Ray System has a diagnostic mode in which you can display a summary of maintenance data or display feedback values after each exposure From the Main System Configuration menu shown in Figure 1 select CONFIGURE UNIT 2 You will see the Configuration menu shown in Figure 4 To display a summary of maintenance data highlight select SHOW MAINT The following maintenance data are displayed e Total KJ kilojoules total system heat on x ray tube e Exposure Count e Reboots power up cycles e OT Counts over threshold counts 4 Press any button on the Operator Panel to return to the Configuration menu If you take an x ray while in diagnostic mode the display shows feedback values for that exposure Until you exit diagnostic mode the display will continue to show feedback values after each exposure 1 From the Main System Configuration menu shown in Figure 1 select CONFIGURE UNIT You will see the Configuration menu shown in Figure 4 2 Select SET CONFIG You will see the Set Confi
9. all when returning it to the storage position 15 PTE f V t Br DIR Jh Y A I UJ a 00 02 1601 Rev F VetVision DC If it becomes necessary to increase or decrease the kV mA or time from the preset values follow these steps Changing kV mA or time Press the Enter button P to highlight the current kV mA or time setting Use the up and down buttons to increase or decrease the value Press the Exposure button After the image s has been captured press any of the tooth receptor type or patient size buttons to return the display to its preset values NA Note When exposure settings are being adjusted the Tooth Selection Image Receptor Type and Patient Size buttons are not illuminated 16 PTE f V DIR Jh Y A I UJ a 00 02 1601 Rev F VetVision DC Pre programmed Exposure Times The tables below show the factory default exposure settings for each combination of Tooth Image Receptor Type and Patient Size on the Operator Panel These exposure settings can be modified using the System Configuration mode See the System Configuration section for details 8 Cone Progeny Schick Dexis Kodak Sirona D Speed E F Speed Large Small Large Small Large Small Large Small Large Small Large Small Large Small Upper Canine Lower Canine Incisors Premolars Molars 12 Cone Progeny Schick Dexis Kodak Sirona D Speed E F Speed Large Small Large Small Large Small Large Small Large Small Large Small Lar
10. e and Tooth Size Image Receptor Type and Patient Size are illuminated Use the Tooth Selection Image Receptor Type and Patient Size Selection buttons to select the preset to change The display shows the current values for the preset Use the right arrow button to display the technique factor to change Use the up and down arrow buttons to set the value for the selected technique factor and preset Repeat steps 2 4 to change additional presets When you have completed all changes press the Tooth Selection and Patient Size Selection buttons simultaneously for 5 seconds to record the change To return all presets to factory defaults select RECALL PRESETS from the Preset Options menu The menu will ask you to confirm your choice Select YES using the up arrow button and return all presets to factory default settings Selecting YES will erase any custom preseis that have been set up Select NO using the down arrow button and retain current preseis To retain the preseis select YES 25 P y d ap r YE f y T FA i L A A F Ya A Y I CAT e 00 02 1601 Rev F VetVision DC Showing Current System Configuration Figure 4 Configuration Menu The VetVision DC Dental X Ray System displays the current system configuration This display is informational only 1 From the Main System Configuration menu shown in Figure 1 select CONFIGURE UNIT You will see the Configuration menu shown in Figure 4 Select SHOW CO
11. everal exposures and verify that the Radiation Indicator illuminates and the audible Exposure signal is heard Indicators Select the longest exposure time possible using the up and down arrows Initiate an exposure but release the Exposure button after a brief period of time before the timer terminates the exposure Verify that the display indicates Pre termination Error and returns to normal operating mode Premature Termination When a coil cord hand switch is used inspect the switch housing and coil cord for damage or wear Replace if evidence of damage is present Coil cord Hand Switch User Information make certain that the user of the system has received the User Manual 31 F Cie if Ti T a y i d X Zi F FE Fx X KL M AT ts A Y A A SI e 00 02 1601 Rev F VetVision DC New Tube Seasoning Procedure X ray tubes that sit dormant for several months can become electrically unstable To remedy this condition it is recommended you perform a new tube seasoning procedure This process establishes stable high voltage operation and will ultimately extend the life of the tube Repeat this procedure before returning to normal operation any time the system has been unused for more than two months Verify system operation Energize the system Select 60 kilovolts 7 milliamperes and the exposure time of one second Make five exposures at this level observing the normal cooling time Select
12. g progeny A Midmark Company VetVision DC X Ray System User Manual LE 00 02 1601 Rev G ECN P3250 Attention The equipment must only be installed and operated in accordance with the safety procedures and operating Instructions given in this manual and in the User Manual for the purposes and applications for which it was designed Modifications and or additions to the equipment may only be carried out by Progeny Midmark Company or by third parties expressly authorized to do so Such changes must comply with legal requirements as well as with the generally accepted technical rules It is the responsibility of the user to ensure that existing legal regulations regarding installation of the equipment with respect to the building are observed X RAY PROTECTION X ray equipment may cause injury if used improperly The instructions contained in this manual must be read and followed when operating the Vet Vision Your Dental dealer will assist you in placing the Vet Vision in operation The Vet Vision X Ray System provides a high degree of protection from unnecessary x radiation However no practical design can provide complete protection nor prevent operators from exposing themselves or others to unnecessary radiation VetVision DC y progeny VetVision DC Dental X Ray System User Manual Progeny Midmark Company 675 Heathrow Drive Lincolnshire IL 60069 Phone 1 847 415 9800 Toll Free 888 924 3800 Fax
13. ge Small Upper Canine Lower Canine Incisors Premolars Molars Large over 40 Ibs Small under 40 Ibs 17 00 02 1601 Rev F VetVision DC bar progeny Baseline Technique Charts Canine Baseline Technique Charts D speed film Maxillary incisors J Je 17 108 gt Maxillary canine J Je 7 108 gt Maxillary anterior premolars 60 7 032 gt Maxillary PM4 6 17 108 gt Maxillary PM4 caudal view 60 7 040 gt MaxilaryMIM2 IA Mandibular incisors 60 7 025 f Mandibular canine 160 7 fose Meme Je poesi lg lateral oblique view Mandibular posterior pm amp molars 160 17028 Weieren o jo los Weieen Je je Jos Mailaramiororpendas fo fe Jos fm s Joe feo s e po Dm oT y 60 Mandibular incisors 17 fe 1004 1 od 0 0 o cd sonen Lais fom lateral oblique view pp gt Mandibular posterior pm amp molars 6 16 10082 1 Kr Progeny 00 02 1601 Rev F VetVision DC Feline Technique Settings Maday nases nu Weien feo 7 nu Malay prmolars Mt meer 60 7 01 Malay premolars M1 odraoral 60 7 0125 Mandibular incisors and canine 60 17 104 Mandibular PM3 lateral oblique 60 17 046 poa n Ik ise Weieren feo 7 002 Weien mn 7 002 Weien pre
14. guration menu shown in Figure 5 3 From the Set Configuration menu use the up and down arrows to highlight DIAG MODE ON Press the right arrow button to turn on diagnostic mode 4 Exit System Configuration mode by highlighting and selecting EXIT in the Configuration and Main menus 5 Make an exposure The display will show the following feedback values e kV e mA e Filament current 6 Press any button on the Operator Panel to clear the feedback values from the display 7 To exit diagnostic mode depress the Tooth Selection and Patient Size Selection buttons simultaneously for 5 seconds to display the Main System Configuration menu From the Main System Configuration menu highlight and select CONFIGURE UNIT Then highlight and select SET CONFIG On the Set Configuration menu highlight and select DIAG MODE OFF Go 2 00 02 1601 Rev F Trouble Shooting VetVision DC Performance Issues Light or dark images 2 3 No X Ray 1 2 Pre termination Adjust the selected exposure time kilovoltage or tube current to produce an acceptable image If necessary reprogram the techniques factors as explained in the System Configuration section of this manual Verify the kilovoltage and tube current during an exposure using the diagnostic mode as explained in the System Configuration section of this manual Alternatively you may employ a non invasive meter to evaluate kilovoltage and exposure time Inspect the condition
15. molars Mi nr ral 60 7 0047 Medie premolars Mt exraoral 60 7005 Mandibular incisors and canine 60 7 0 032 Mandibular PMS lateral oblique 0 025 Mandibular PM3 M1 0 025 19 g progeny 00 02 1601 Rev F Record Your Exposure Settings VetVision DC If the pre programmed exposure settings do not produce the density desired adjust the settings using System Configuration mode Record your settings in the table below Canine 8 inch 20 cm Cone Digital Receptor Lud D speed Film B Tooth Selection kV Upper Canine kV Lower Canine seconds kV Large 3 Large e Incisors Upper R and Lower seconds Lower Molar and 5 Premolars seconds u wp gt Molar and ma Large more than 30 Ibs Small 2 30 Ibs and under lt A Ri Palpi III MENU 20 Small E F Speed Film l Large Small MI Im y progeny 00 02 1601 Rev F VetVision DC Record Your Exposure Settings If the pre programmed exposure settings do not produce the density desired adjust the settings using System Configuration mode Record your settings in the table below Feline Tooth Selection kV Upper Canine Lower Canine lt seconds lt Incisors Upper and Lower lt Lower Molar and Premolars lt Upper Molar and Premolars seconds 3 gt Use small settings for feli
16. ne dentistry 21 g progeny 00 02 1601 Rev F VetVision DC System Configuration System Configuration Mode About System Configuration Mode Using System Configuration Mode Figure 1 Main System Configuration Menu The VetVision DC Dental X Ray System has a software driven system configuration mode When the VetVision DC is in system configuration mode you can perform the following procedures e Adjusting the Display e Changing Pre programmed Exposure Settings e Changing the Cone Size e Showing Current System Configuration e Displaying Diagnostic Data 1 To enter system configuration mode depress the Tooth Selection and Patient Size Selection buttons on the Operator Panel simultaneously for 5 seconds The display shows the Main System Configuration menu as shown in Figure 1 and the Ready Indicator blinks 2 To select menu items while in system configuration mode use the up and down arrows to highlight a menu option Then use the right arrow button as an Enter button to select the highlighted option When changing presets the right arrow button is also used to select the technique factor 3 After selecting a menu option use the up and down arrows to increase or decrease values 22 00 02 1601 Rev F VetVision DC Adjusting the Display Adjusting Contrast Reversing the Image Figure 2 Display Options Menu The VetVision DC Dental X Ray System allows the operator to adjust the display image 1
17. ng PP 1 SC EE Ensure that all certified components bear labels that include the model and serial number Labels aspi e date of manufacture and a statement of certification as noted elsewhere in this manual Check for oil leaks or other evidence that could indicate internal damage Replace the Tubehead 3 P Tubehead if necessary Tubehead Rotation Ensure thatthe Tubehead maintains its position around the horizontal axis while remaining easy to rotate and position Also check the vertical pivot of the Tubehead for easy movement while remaining in position after moving Check that all movements are smooth and quiet Verify that the Tubehead is properly counterbalanced for vertical drift and that the Horizontal and Articulating Arms do not drift horizontally Verify that the switch is working properly and that the Ready Indicator is illuminated when the power switch is in the ON position Operator Panel With the power switch located at the upper right of the Control Unit in the ON position verify E that technique factors appear on the Operator Panel Also check the function of the selection Controls buttons for Tooth Selection Image Receptor Type and Patient Size Pressing a selection Verify that the Exposure button on the Operator Panel is functioning properly To make an exposure press and hold the Exposure button until the Radiation Indicator is extinguished and the audible signal is no longer heard Exposure Button Make s
18. nitiate an x ray exposure For a complete exposure the button must be pressed and held until the Radiation Indicator no longer illuminates and the audible signal is no longer heard Releasing the Exposure button immediately terminates the x ray exposure CAUTION Releasing the Exposure button prior to the completion of the x ray exposure will result in an incomplete exposure of the image This may require the operator to re take the radiograph When a premature release of the Exposure button occurs the system will notify the operator by blinking momentarily then it will return to operating mode The Ready Indicator lights up when the system is ready to make an exposure Immediately after an exposure the Ready Indicator flashes until the x ray tube cools down sufficiently to make the next exposure When the Ready Indicator is flashing no exposure can be made The VetVision DC has a visible and an audible Radiation Indicator When an exposure is in progress the Radiation Indicator on the Operator Panel is illuminated and an audible tone is heard The exposure is complete when the Radiation Indicator light is extinguished and the audible tone is no longer heard 14 00 02 1601 Rev F VetVision DC Taking an X Ray Using Preset Techniques 1 The power switch is located at the upper right of the Control Unit or on the control unit on the footed stand of the mobile system Press button to the On position The Ready Indicato
19. nnos 23 Regular Maintenance arennnrnnnnnnnnennnrnnnnnnnnnnnnrnnnnnnnnennnennnnnnnnnnnnennnsnnnnennn 23 Cleaning and DISIMSCUING BE 23 Checking System FUNCION Sins 24 System Function Checklist aaa aa aada anaa aaa 25 New Tube Seasoning Procedure ccooocccccccncccncncccocnconacnncnncnononcnononcnonnnnnnnnns 26 SPEARS 27 VetVision DC Dental X Ray System eres 21 TUBE FEM IA ee 28 VetVision DC VetVision DC General Information VetVision DC Dental X ray Wall mounted Model Articulating Arm Tube Head Figure 1 Wall mounted VetVision DC Single Stud Mount Figure 2 Wall mounted VetVision DC Double Stud Mount g progeny 00 02 1601 Rev F VetVision DC VetVision DC Mobile System GILT Articulating Arm Handle SJON e ORE ee Tube Head 6 b amp amp 6 5 6 oe e Control Panel X ray remote ovos B ER Power center and Locking brake on On Off control both front wheels y progeny 00 02 1601 Rev F VetVision DC Compliance with Applicable Standards The certified components of the VetVision DC X Ray System comply with Radiation Performance Radiation Standards 21 CFR Subchapter J at the time of manufacture Protection The certified components of the VetVision DC X Ray System comply with IEC 60601 1 3 Radiation protection x ray equipment UL 2601 1 File Classified by Underwriters Laboratories Inc with respect to electrical
20. oducts such as Discide leanin Cleaning Ultra Cavicide or the equivalent Compounds Cleanin Between each patient perform the following cleaning and disinfecting steps g 1 Remove gross bio burden from the cone handles and structure with a Methods disposable towel moistened with water 2 Dry the cone handles and structure with disposable towels 3 Wipe the cone handles and structure with the disinfectant product following the disinfectant manufacturer s instructions 4 Clean any remaining disinfectant from the component with water This additional step prevents possible product discoloration or corrosion 5 Dry the cone handles and structure with disposable towels Caution The Progeny Dental X Ray System is not waterproof Use only moistened not saturated towels 29 00 02 1601 Rev F VetVision DC Checking System Functions The following checks must be performed to complete the installation of the VetVision DC Dental X Ray System and as part of the recommended maintenance as indicated in the User Manual CAUTION If the VetVision DC Dental X Ray System does not perform the functions below advise the owner that the system is not to be used See the Troubleshooting section of this manual or contact Progeny s Technical Support 30 00 02 1601 Rev F VetVision DC System Function Checklist Ensure that the wall support is adequate and that the system is properly mounted to the wall Wall Mounti
21. of the remaining imaging chain components such as the film chemistry and processor or the condition of the x ray sensor and computer If no x ray is produced check the following Verify that the line cord if one is in use is properly connected Verify that the power switch is in the ON position Early release of the exposure switch will cause a pre termination error to occur After five seconds the system will return to the normal operating Error Technical Support Contact condition Be advised that this will result in an underexposed image PROGENY DENTAL 675 Heathrow Drive Lincolnshire IL 60069 Phone 1 847 415 9800 Toll Free 888 924 3800 Fax 1 847 415 9810 techsupport progenydental com 28 00 02 1601 Rev F VetVision DC Recommended Maintenance Regular Maintenance In the interest of equipment safety a regular maintenance program must be established This maintenance program should consist of cleaning and disinfecting as well as annual system function checking It is the owner s responsibility to arrange for this service and to assure that the personnel performing this are fully qualified to service Progeny Dental x ray equipment Cleaning and Disinfecting The VetVision DC Dental X Ray System requires disinfection The cleaning and disinfecting methods described here protect operators and patients in a manner that is safe for the equipment Progeny Dental recommends the use of disinfectant pr
22. ones differ 33 00 02 1601 Rev F VetVision DC 4500 4000 3500 3000 VI TT fe RAR cot ae a E a E PEE 2500 nd o o o 1500 HEAT STORAGE J 1000 AS AA CA TL 500 0 50 100 150 200 250 300 TIME s 3500 3 Ear 3000 VETT TT Te 5 2500 NT tt lt 1500 da COOLING TEN EUN IL TI E SES EES 500 i Wis 29 e a 0 50 100 150 200 250 300 TIME s Figure 7 Toshiba Tube Rating Charts 34 g progeny 00 02 1601 Rev F VetVision DC og CO ka DY GO r CO ER zl D O S Rating Chart Thermal Characteristics Chart Figure 8 Kailong Tube Rating Charts 35 f progeny 00 02 1601 Rev F
23. r on the front of the Operator Panel will light Verify that the unit is set for the Tooth to be imaged The icon for the currently selected Tooth is illuminated To change press the Tooth Selection button until the correct tooth type is selected Verify that the unit is set for the correct Image Receptor Type The icon for the currently selected Image Receptor Type will be illuminated To change the Image Receptor type press the Image Receptor button until the correct Image Receptor Type is selected Verify that the system is set for the appropriate Patient Size The icon for the currently selected Patient Size will be illuminated To change Patient Size press the Patient Type button to select Large or Small Make sure the patient is properly immobilized and place the tube head 2 3 inches from the patient and directly centered on the receptor Press and hold the Exposure button until the audible signal is no longer heard and the Radiation Indicator is no longer illuminated Releasing the Exposure button or coil cord hand switch at any time will immediately terminate the exposure Note When using the coil cord hand switch it is recommended that the operator exit the operatory if possible Note In order to comply with regulations and good safety practices the technique factors must be visible to the operator from the remote location Return the Tubehead to the storage position Note Be careful not to strike the Tubehead on the w
24. shock fire and mechanical hazards only in accordance with UL 2601 1 and CAN CSA C22 2 NO 601 1 M90 and to the Number following particular standards IEC60601 2 7 IEC 60601 2 28 E181750 EMI EMC IEC60601 1 2 Certified Components Svstem Component Reference Number d Tubehead 30 A1027 Control Unit 30 A1041 Cone 8 in 30 A2016 3 P y d ap r YE f y T FA i L A A F Ya A Y I CAT e 00 02 1601 Rev F EC Declaration of Conformity Name and Description of Product Reference Numbers to which Conformity is Declared Contact Progeny VetVision DC Catalog P7017V 76 inch reach Model 30 A1041 Control 30 A1028 Articulated Arm Catalog P7016V 66 inch reach Model 30 A1041 Control 30 A1028 Articulated Arm Catalog P7015V 56 inch reach Model 30 A1041 Control 30 A1028 Articulated Arm Catalog P7017VM Mobile Model 30 A1032 Control Class IIb The following regulatory documents apply UL 2601 1 IEC 60601 1 2 IEC 60601 1 3 IEC 60601 2 7 IEC 60601 2 28 IEC 60601 2 32 Technical S upport techsupportQ progenydental com Authorized Representatives VetVision DC 00 02 1601 Rev F North America Europe Midmark Corporation 675 Heathrow Dr Lincolnshire IL 60089 Phone 888 924 3800 Fax 847 415 9810 CE Partner 4U Esdoornlaah 13 3951DB Maarn Netherlands Phone 31 343 442 524 Fax 31 343 442 162 VetVision DC Safety dl Ps Ta Ta Tas ae J d AA Ju A d
Download Pdf Manuals
Related Search
Related Contents
Pregunta a ELIKA responde Huffy M5800151 User's Manual レーザー距離計 LD-420(約1.5MB) Samsung NONE User Manual User manual Interroll Carton Wheel Flow ABI PRISM 7700 - Thermo Fisher Scientific 取扱説明書 HDB848 English UserManual HDB848-MG01-03 提出書類一覧表 (PDF : 200KB) Manuel de cette série Copyright © All rights reserved.