Home

ThoraCool DFU

image

Contents

1. Do NOT perform the procedure with any damaged equipment Kimperty CLark Cooled Radiofrequency Sterile Tube Kit Fig 2 1 Place the burette into the burette holder on the side of the Pump Unit The side of the burette with 2 or 3 ports indicates the top of the burette Fig 3 2 Fill the burette with room temperature sterile water Use sterile handling techniques Fill the burette to the 70 mL mark Burette can be filled by injecting sterile water through a port in the lid or by temporarily removing the lid and pouring sterile water in Warning BE SURE TO FILL THE BURETTE TO THE 70 mL MARK Not filling the burette to the 70 mL mark will result in an inadequate supply of water for circulation Use ONLY sterile room temperature water Ensure the lid is snapped back onto the body of the burette after filling Fig 4 5 Inject sterile water into burette OR remove lid and pour sterile water 3 Place the thick walled tubing coming out of the bottom of the burette into the pumphead of the Pump Unit Place the tubing in the channels of the L shaped bracket to ensure that the tubing is not obstructed while closing the pumphead Close the lid on the pumphead to clamp down on the tubing 4 Remove the caps on the male and female luer locks Connect the appropriate luer lock to the corresponding luer lock on the Probe Do not over tighten the connection Fig 6 Caution Connect one Tube Kit to one THoraCoo Probe Fig 7 5 Atthe end
2. THoraCoo Probes only The THoraCooL Introducer provides a path for the THoraCoo Probe to the nervous tissue Kimperty CLark THoraCooL Cooled Radiofrequency Probe sterile single use It is inserted through a ThoraCool Introducer into or near nervous tissue Sterile water circulates internally to cool the THoraCoo Probe while it delivers radiofrequency energy A thermocouple in the THoraCoo Probe measures cooled electrode temperature throughout the procedure Indications For Use The Kimperty CLark THoraCooL Cooled Radiofrequency Kit in combination with the Kimperty Cark Radiofrequency RF Generator PMG 115 TD PMG 230 TD formerly Baylis Pain Management Generator is indicated for use to create RF lesions in nervous tissue Contraindications For patients with cardiac pacemakers a variety of changes can occur during and after the treatment In sensing mode the pacemaker may interpret the RF signal as a heartbeat and may fail to pace the heart Contact the pacemaker company to determine if the pacemaker should be converted to a fixed rate pacing during the radiofrequency procedure Evaluate the patient s pacing system after the procedure Check the compatibility and safety of combinations of other physiological monitoring and electrical apparatus to be used on the patient in addition to the RF Generator If the patient has a spinal cord deep brain or other stimulator contact the manufacturer to determine if the s
3. CH Kimberly Clark THoraCoo Cooled Radiofrequency Kit Pain Management Instructions for Use CD Kimberly Clark Trusted Clinical Solutions Pump Connector Cable THoraCooL Stylet Generator Tube Kit i THorACooL Probe THoraCooL Introducer Cooled Radiofrequency Connector Cable Grounding Pad Monopolar System Sterile water injection port Luer Locks for connection of Tube Kit to THoraCooL Probe 70 mL Mark Burette Reservoir Cooled Radiofrequency Sterile Tube Kit WS Inject sterile rA water into burette Remove lid and pour sterile water Burette Placement in Pump Unit Fig 5 Placement of Tubing in the Pumphead of the Pump Unit Fig 6 i Do Not Use Attention See Single STERILE If Pack Pie 2 Use Only E0 Is Dano d Instructions for Use Ne Upper Limit Keep Away Dispose of 4 of Temperature AS from Sunlight Properly 0086 KIMBERLY CLARK THoraCoo Cooled Radiofrequency Kit Pain Management Federal Law USA restricts this device to sale by or on the order of a physician Device Description Kimperty CLark Cooled Radiofrequency Sterile Tube Kit sterile single use non body contact It is used for closed loop circulation of sterile water through a Kimperty CLark THoraCooL Cooled Radiofrequency RF Probe It includes a burette and tubing Kimperty CLark THoraCooL Cooled Radiofrequency Introducer sterile single use It is to be used with the
4. aCooL Probe The THoraCoo Probe THoraCooL Introducer and Tube Kit are ethylene oxide sterilized and supplied sterile The devices should be stored in a cool dry environment Note Please contact Kimberly Clark for a list of all model numbers and sizes Inspection Prior To Use The sterile packaging should be visually inspected prior to use to detect any compromise Ensure that the packaging has not been opened or damaged Do not use the equipment if the packaging has been compromised Equipment Required Procedures should be performed in a specialized clinical setting equipped with a fluoroscopy unit The equipment required to perform RF procedures include THoraCooL Cooled Radiofrequency Probe THoraCool Cooled Radiofrequency Introducer s e Cooled Radiofrequency Peristaltic Pump Unit and Cable e Cooled Radiofrequency Sterile Tube Kit Cooled RF Connector Cable e Radiofrequency Dispersive Electrode e Radiofrequency Generator PMG 115 TD PMG 230 TD Instructions For Use Monopolar System Fig 1 Assemble all the equipment required for the procedure Set up the Radiofrequency Generator PMG 115 TD PMG 230 TD and the Pump Unit pump as directed in their Instructions for Use Connect the Cooled RF Connector Cable to the RF Generator as described in its Instructions for Use Open the package in the sterile field using appropriate sterile techniques Inspect the devices visually to make sure there is no damage to them
5. ate for example spinal cord or nerve root irritation If these indications are suspected discontinue energy delivery 7 After treatment remove the Probes and the Introducer and discard as biohazards Remove the Dispersive Electrode from the patient and discard appropriately Disconnect the Cooled RF Connector Cable from the RF Generator Follow standard hospital techniques to handle reusable items Troubleshooting The following table is provided to assist the user in diagnosing potential problems TROUBLESHOOTING Ensure all connections are made Probe s to Cooled RF Connector Cable OR Cooled RF Connector Cable to the RF Generator Inaccurate RF Generator to power outlet erratic or sluggish Check for an error message on the RF Generator temperature aaa Visually inspect the Probe or Cable for damage Ensure g that devices are dry and at room temperature If problem persists discontinue use PROBLEM No temperature measurement Water does not flow through the Probe and Tube Kit Stop the procedure immediately Check the luer lock connections to ensure the Tube Kit is connected to the Probe Check the Pump to ensure the lid is not open Check RF Generator for any error messages Probe Connector does not fit in Probe Plug in Check that the connector s keys are lined up in the proper orientation Ensure that the Connectors are clean and unobstructed Damage to Insulation on Probe or Introducer Do no
6. ate to each individual patient all foreseeable risks of the RF lesion procedure The sterile packaging should be visually inspected prior to use to detect any compromise Ensure that the packaging has not been damaged Do not use the equipment if the packaging has been compromised Proper sterile techniques must be used when assembling and filling the Tube Kit Do not place the lid down on a non sterile surface Kimperty CLark Cooled Radiofrequency Sterile Tube Kit The Tube Kit is for use with a single THoraCoo Probe Care must be taken to ensure all luer fittings are secure to prevent leaking Do not disconnect luer fittings while the pump is running Arrange equipment to minimize tubing tripping hazards Do NOT perform cooled RF lesion procedures if water is not circulating through the Tube Kit water is leaking or air bubbles are seen in the tubing Immediately stop the procedure and correct circulation before restarting the procedure Do NOT pinch the tubing of the Tube Kit Kimperty CLark THoraCoot Cooled Radiofrequency Introducer Be careful while handling the THoraCoo Introducer The sharp tip can cause injury to the operator if handled carelessly Handle the Introducer safely when it is in use due to electric currents Do not move the Introducer without the stylet fully inserted Choose the properly sized Introducer Kimperty CLark THoraCoot Cooled Radiofrequency Probe The Tube Kit should never be disconnected from the Probe when RF del
7. e of purpose Limitation of Liability for Damages In any claim or lawsuit for damages arising from alleged breach of warranty breach of contract negligence product liability or any other legal or equitable theory the buyer specifically agrees that Kimberly Clark shall not be liable for damages for loss of profits or claims of buyer s customers for any such damages Kimberly Clark sole liability for damages shall be limited to the cost to buyer of the specified goods sold by Kimberly Clark to buyer which give rise to the claim for liability The buyer s use of this product shall be deemed acceptance of the terms and conditions of these limited warranties exclusions disclaimers and limitations of liability for money damages
8. emperature readings are observed Use of damaged equipment may cause patient injury Do not modify Kimserty Ciark Equipment Any modifications may compromise safety and efficacy of the device When the RF Generator is activated the conducted and radiated electrical fields may interfere with other electrical medical equipment The RF Generator is capable of delivering significant electrical power Patient or operator injury can result from improper handling of the Probes particularly when operating the device During power delivery the patient should not be allowed to come in contact with grounded metal surfaces Do not remove or withdraw the device while energy is being delivered Precautions Do not attempt to use the THoraCooL Kit before thoroughly reading the accompanying Instructions for Use and the User s Manual for the RF Generator and Dispersive Electrode PMA GP BAY Apparent low power output or failure of the equipment to function properly at normal settings may indicate 1 faulty application of the dispersive electrode or 2 power failure to an electrical lead Do not increase power level before checking for obvious defects or misapplication To prevent the risk of ignition make sure that flammable material is not present in the room during RF power application Only physicians familiar with RF lesion techniques should use the THoraCoo Kit components It is the physician s responsibility to determine assess and communic
9. ernational Customers 1 770 587 7200 Notes In order to return products under limited warranty you must have a return authorization number before shipping the products back to Kimberly Clark Limited Warranties Kimberly Clark warrants that these products are free from defects in original workmanship and materials If these products prove to be defective in original workmanship or original materials Kimberly Clark in its absolute and sole discretion will replace or repair any such product less charges for transportation and labor costs incidental to inspection removal or restocking of product This limited warranty applies only to original factory delivered products that have been used for their normal and intended uses Kimberly Clark s limited warranty shall NOT apply to Kimberly Clark s products which have been repaired altered or modified in any way and shall NOT apply to Kimberly Clark s products which have been improperly stored or improperly installed operated or maintained contrary to Kimberly Clark Instructions The warranty period for Kimeerty CLark RF Probes and RF Connector Cables is 90 days from the date of purchase unless otherwise stated Disclaimer and Exclusion of Other Warranties There are no warranties of any kind which extend beyond the description of the warranties above Kimberly Clark disclaims and excludes all warranties whether expressed or implied of merchantability or fitness for a particular us
10. ivery is in progress The lumen of the Tube Kit should not be obstructed in any way during the procedure as this will stop cooling of the Probe Disconnect the Probe by pulling the connector not the cable Handle the Probe safely when it is in use due to electric currents and the hot tip While inserting the Probe through the THoraCoot Introducer watch the fluoroscope for any buckling Do not attempt to further insert the Probe if any buckling is observed or significant resistance is felt Do not move the THoraCooL Introducer when the Probe is in it If repositioning is needed retract the Probe from the Introducer and then reposition the Introducer with the stylet inserted The Cooled RF Temp displayed on the RF Generator refers to the cooled electrode temperature and not the hottest tissue temperature Adverse Events Potential complications associated with the use of this device include but are not limited to infection nerve damage increased pain visceral injury failure of technique paralysis and death Product Specifications The THoraCooL Probe is comprised of an electrically insulated shaft with an active tip that functions as an electrode for RF energy delivery a handle tubes with luer locks and a cable with a 7 pin connector The THoraCooL Introducer includes an insulated stainless steel cannula and a stylet The Tube Kit is comprised of a burette and flexible tubing fitted with luer locks for connection to the THor
11. of the procedure discard the Tube Kit appropriately Kimperty CLark THoraCooi Cooled Radiofrequency Introducer 1 With the stylet in the Introducer carefully insert the Introducer into the patient using fluoroscopic guidance to place it at the desired lesion location 2 Once the Introducer is in the proper position carefully remove the stylet from the Introducer 3 Repeat steps 1 2 with a second Introducer if necessary Kimperty CLark THoraCooi Cooled Radiofrequency Probe 1 Insert the Probes into the tissue through the Introducer Never force the Probe in if significant resistance is felt 2 Attach the Dispersive Electrode to the RF Generator and place the Dispersive Electrode Pad on the patient as directed in the Instructions for Use accompanying the package 3 Connect the Probe to the Tube Kit 4 Connect the 14 pin connector of the Cooled RF Connector Cable into the RF Generator Connect the Probe to the 7 pin connector on the Cooled RF Connector Cable 5 Select the Treatment mode in the RF Generator Set advanced settings and the parameters for RF delivery in the RF Generator as described in the User s Manual 6 Perform the procedure as described in the RF Generator User s Manual The procedure comprises pre cooling treatment and optional post cooling stages Note Other than reproduction of their usual referred pain or irritation due to probe introduction monitor the patient for unexpected symptoms that may indic
12. t use Discard immediately Water is not circulating through tubing during THoraCoo pre cooling ON and post cooling states Ensure the Tube Kit is correctly connected to the Probe Ensure the Tube Kit has been correctly placed in the pumphead Ensure the burette reservoir has been filled Visually inspect the Tube Kit tubing and joints for leaks and occlusions Ensure that the float ball in the burette is floating and not occluding the outflow of water from the burette Ensure the pump tubing thick walled tubing that is coming directly out of the bottom port of the burette is placed in the pumphead Water is not dripping into the burette Check to see if water is running down the wall of the burette Float is stuck on bottom port of the burette Close the pumphead lid Gently shake the burette to try and loosen the ball from the bottom of the burette The lid of the burette cannot be removed Inject sterile water through the port of the lid rather than removing the lid PROBLEM TROUBLESHOOTING Tube Kit breaks Immediately discard the Tube Kit is leaking or is occluded Customer Service and Product Return Information If you have any problems with or questions about Kimeerty CLark Equipment contact our technical support personnel Kimberly Clark 1400 Holcomb Bridge Road Roswell GA 30076 2199 Email InterventionalPain KCHC KCC COM U S Customers 800 KCHELPS 800 742 1996 Int
13. timulator needs to be in the bipolar stimulation mode or in the OFF position This procedure should be reconsidered in patients with any prior neurological deficit The use of general anesthesia is contraindicated To allow for patient feedback and response during the procedure it should be performed under local anesthesia Systemic infection or local infection in area of the procedure Blood coagulation disorders or anticoagulant use Warnings The TuoraCoot Kit contains single use devices Do not reuse reprocess or resterilize this medical device Reuse reprocessing or resterilization may 1 adversely affect the known biocompatibility of the device 2 compromise the structural integrity of the device 3 lead to the device not performing as intended or 4 create a risk of contamination and cause the transmission of infectious diseases resulting in a patient injury illness or death The THoraCoo Probe must be used with the correct connector cable Attempts to use it with other connector cables can result in electrocution of the patient or operator Laboratory staff and patients can undergo significant x ray exposure during radiofrequency procedures due to the continuous use of fluoroscopic imaging This exposure can result in acute radiation injury as well as increased risk for somatic and genetic effects Therefore adequate measures must be taken to minimize this exposure Discontinue use if inaccurate erratic or sluggish t

Download Pdf Manuals

image

Related Search

Related Contents

CME-965 User Manual - Diamond Systems Corporation  VIZIO VF550M User's Manual  none 306150 Use and Care Manual  Peerless PLP-V4X3  Manual de Instruções Balanças NewClassic Modelos ME  携帯用複合型ガス検知器  User's Guide - SMART TECH INTERNATIONAL  Langue et matières scolaires      

Copyright © All rights reserved.
Failed to retrieve file