ResMed Autoset Respond

image

Contents

1. AutoSet Respond Reorder number 308145 2 03 09 USER S MANUAL English USA www resmed com Waking people up to sleep RESMED AutoSet Respond Reorder number 308145 2 03 09 USER S MAN UAL English USA Manufactured by ResMed Ltd 97 Waterloo Road North Ryde NSW 2113 Australia Tel 61 2 9886 5000 or 1 800 658 189 toll free Fax 61 2 9878 0120 Email reception resmed com au Distributed by ResMed Corp Danielson Street Poway CA 92064 6857 USA Tel 1 858 746 2400 or 1 800 424 0737 toll free Fax 1 858 746 2900 Email reception resmed com ResMed UK Limited 65 Milton Park Abingdon Oxfordshire OX14 4RX UK Tel 44 1235 862 997 Fax 44 1235 831 336 Email reception resmed co uk ResMed Asia Pacific Ltd 97 Waterloo Road North Ryde NSW 2113 Australia Tel 61 2 9886 5000 or 1 800 991 900 toll free Fax 61 2 9889 1471 Email reception resmed com au ResMed Finland Niittykatu 6 FIN ESPOO Suomi Puh 358 9 8676820 Faksi 358 9 86768222 S hk posti reception resmed fi ResMed GmbH amp Co KG RudolfstraBe 10 D 41068 M nchengladbach Deutschland Tel 49 0 2161 3521 0 Reception 49 0 180 22 22 668 Service Telefon 0 06 Anruf 49 0 180 22 66 888 Wartungstelefon 0 06 Anruf Fax 49 0 2161 3521 1499 Email reception resmed de ResMed Japan Nihonbashi Hisamatsu Bldg 4F 2 28 1 Nihonbashi Hamacho Chuo Ku Tokyo 103 0007 Japan Tel 81 3 3662 5056 Fax 81 3 3662 5
2. Component failure Return your AutoSet Respond for servicing Display error message High leak in last session You have experienced excessively high leak levels during the night Check that your air tubing is connected properly and that your mask does not leak excessively Use the mask fitting feature to help you to fit your mask properly If this message appears again contact your clinician Display error message HIGH LEAK Adjust mask You have experienced excessively high leak levels for more than 20 seconds Check that your air tubing is connected properly and that your mask does not leak excessively Use the mask fitting feature to help you to fit your mask properly If this message appears again contact your clinician 24 SYSTEM SPECIFICATIONS Performance Operating pressure range 4 to 20 cm H5O Dimensions H x W x D 5 6in x 9 lin x 10 6in Weight 5 116 Power Supply SYSTEM SPECIFICATIONS Input range 100 240V 50 60Hz lt 120VA maximum power consumption Actual power consumption will vary depending on factors such as the use of accessories height above sea level and ambient temperature Power consumption values for typical treatment conditions using a breathing machine set to 0 5L and 15 breaths min with 6ft6in 2m air tubing and a modular mask are given in the following table Device Pressure Power cm H5O VA AutoSet Respond 5 20 10 26 15 32 20 39
3. Housing Construction Flame retardant engineering thermoplastic Environmental Conditions Operating Temperature 41 F to 104 F Operating Humidity 1026 9526 non condensing Storage and Transport Temperature 4 F to 40 F Storage and Transport Humidity 1026 9526 non condensing Electromagnetic Compatibility Product complies with all applicable electromagnetic compatibility requirements EMC according to IEC6060 1 1 2 for residential commercial and light industry environments Air Filter Two layered powder bonded polyester non woven fiber Air Tubing Flexible plastic 6ft6in 2m Pressure Sensor Adapter Thermoplastic rubber IEC 60601 1 Classifications Class double insulation Type CF Note The manufacturer reserves the right to change these specifications without notice 25 GLOSSARY OF SYMBOLS N Attention consult accompanying documents Type CF applied part Class equipment Drip proof IPX P A eon weer Start Stop or Mask fit MASK SETTINGS FLOW PRESSURE CHARACTERISTICS 70 60 50 40 30 o un Flow L min 10 0 45678 9 10111213 14 15 16 17 18 19 20 Pressure cm H5O STANDARD MIRAGE MIR FULL ULTRA LIMITED WARRANTY ResMed warrants that your ResMed product shall be free fro m defects in material and workmanship for the period speci
4. Mask Fitting 19 Mask Screen 16 Mask Type 16 Mask fit Star Rating definitions 19 Mask fitting Feature 18 Masks 6 Medical Information 1 Menu 15 Menus 15 Mouth Leaks 19 N Nasal Irritation 20 Operating Instructions 17 Oxygen Use 3 P Passover 11 Performance 25 Power Supply 25 Preparingfor Use 9 Pressure Sensor Adapter 25 R Replacing the Air Filter 22 Responsibility user owner 1 Right Key 15 Runny or Blocked Nose 20 S Servicing 22 Setting Up 9 Side Effects 4 Standby Screen 16 Starting Treatment 17 Stopping Treatment 17 SW Screen 16 System Specifications 25 31 32 T Travelling 20 Troubleshooting 23 U Up Down Key 15 Used Screen 16 User Owner Responsibility 1 Using the Mask fitting Feature 18 Using the Menus 15 Warnings Related to Oxygen Use 3 Warnings Related to Treatment 3 Weekly Cleaning 21 Weight 25
5. You can use the standby screen the period during which the AUTOSET RESPOND maintains a low pressure before it starts to adjust pressure in response to respiratory events You can set the settling time at 5 minute intervals from to 45 minutes To change the settling time press the Up Down key until the desired settling time appears THE MASK SCREEN The Mask screen shows the current mask type You can use the Mask screen to change the mask type Mask options are STANDARD Modular Vista ULTRA MIRAGE MIR FULL Mirage Full Face Mask Mirage Full Face Mask Series 2 Figure summarizes the AUTOSET RESPOND menu screens WELCOME gt gt gt gt gt AUTOSET RESPOND gt gt gt gt gt gt gt gt SETTLING 10min menu USED 50 70days change 1 Figure 1 AutoSet Respond Menu ACCESSING THE MENUS SW SX1161234 il exit 400hrs exit MASK ULTRA exit To access the AUTOSET RESPOND menu To exit out of an item within the menu Press the Right key exit Press the Left key menu while the standby screen is displayed OPERATING INSTRUCTIONS STARTING TREATMENT The AUTOSET RESPOND unit should be assembled beside your bed with the air tubing and mask system connected See Setting Up The AutoSet Respond on page 9 Turn the main power switch at the back of the unit to on 0 U When the AutoSet Respond is turned on a welcome message is disp
6. air filter cover protects the device in the event of accidental liquid spillage onto the device Ensure the air filter and air filter cover are fitted at all times Align the pressure sensor adapter with the pressure port Fit the pressure sensor adapter onto the air outlet of the unit PREPARING FOR USE 5 Assemble your mask system according to the mask user instructions 6 Connect your mask system to the free end of the air tubing The AUTOSET RESPOND is now ready for use To start treatment see Operating Instructions on page 17 HUMIDIFIER USE HUMIDAIRE 2i The HUMIDAIRE 2i attaches to the front of the AUTOSET RESPOND to provide heated humidification No other accessories are required for its use Please refer to the HumidAire 2i User s Manual for details HUMIDAIRE 2iC The HUMIDAIRE 21 attaches to the front of an AUTOSET RESPOND unit to provide passover humidification No other accessories are required for its use Please refer to the HumidAire 2iC User s Manual for details 10 HUMIDAIRE AND PASSOVER Short Am 19cm and medium size 21in 52cm air tubing are necessary accessories for connecting the AUTOSET RESPOND unit to a HUMIDAIRE humidifier Only the medium size tubing is needed for connecting to the ResMed PASSOVER To set up the AUTOSET RESPOND with the HUMIDAIRE or ResMed PASSOVER Fill the HUMIDAIRE or PASSOVER with water as described in the humidifier manual HumidAire U
7. low blood pressure dehydration surgery to the brain middle or inner ear pituitary gland or sinuses respiratory distress syndrome middle ear infection or perforated ear drum severe nosebleed WARNINGS CPAP therapy should be used with caution if you have any of the following conditions respiratory failure cavities or cysts in the lung called bullae and usually due to emphysema or previous history of pneumothorax previous history of severe nosebleed sinus infection Tell your doctor if you have any of these conditions Your doctor will advise you whether the likely benefits of CPAP therapy outweigh the expected risks Special care should be exercised if you are dehydrated or may become dehydrated for example as a result of fluid restriction or diuretic therapy including changes in therapy Discontinue therapy and seek medical advice if during therapy or when you start therapy each night you feel faint or light headed The following are general warnings that pertain to your use of an AUTOSET RESPOND unit Specific warnings appear next to the relevant instructions in the manual WARNING This is NOT a life support device It may stop operating with power failure or if a fault occurs in the unit The air flow temperature for breathing produced by this device can be as much as 6 C 11 higher than the temperature of the room Caution should be exercised if the room tempe
8. 040 ResMed Malaysia Sdn Bhd Suite E 10 20 Plaza Mont Kiara No 2 Jalan 1 70C Mont Kiara Kuala Lumpur Malaysia Tel 60 3 6201 7177 Fax 60 3 6201 2177 Email reception resmed com my ResMed NZ Ltd PO Box 51 048 Pakuranga Auckland New Zealand Tel 64 274 737 633 Fax 64 9 239 0193 Email reception resmed co nz ResMed SA Parc de la Bandonni re 2 rue Maurice Audibert 69800 Saint Priest France T l 33 0 4 37 251 251 Fax 33 0 4 37 251 260 Email reception resmed fr ResMed Singapore Pte Ltd 57 Ubi Ave 1 407 09 Ubi Centre Singapore Tel 65 6284 7177 Fax 65 6284 7787 Email reception resmed com sg ResMed Spain SL C Arturo Soria 245 Madrid Tel 34 93 5908154 Fax 34 93 5908153 Email angelo resmed es ResMed Sweden AB Industrigatan 2 S 461 37 Trollhattan Sverige Tel 46 520 420 110 Fax 46 520 397 15 Email reception resmed se Labhardt AG Thannerstrasse 57 CH 4054 Basel Schweiz Tel 41 061 307 9711 Fax 41 061 307 9722 Email info labhardt ch per US DESIGNATED ResMed Corp EU AUTHORIZED REPRESENTATIVE ResMed UK Ltd www resmed com Protected by patents AU AU AU AU AU AU AU AU AU EP 0651971 DE FR GB NZ US 4944310 US 5199424 US 5245995 US 5522382 US 5704345 US 6029665 US 6138675 US 6152129 US 6279569 US 6363933 US 6367474 US 6398739 US 6425395 US 6502572 Other patents pending Protected by design registrations AU AU AU CH 128 709 CH 128 710 CH 128 711 DE
9. 40202007 DE 40202008 DE 40201723 DE 40202020 ES FR 02 1407 GB 3001791 GB 3001819 GB 3001820 GB 3001821 JP 1164087 JP 1164265 JP 1164266 JP 1164267 SE SE SE SE US D467335 US D468011 US D476077 Other designs pending d iutoset T AutoSet AutoSet Respond and HumidAire 2i are trademarks of ResMed Ltd ResMed Ltd 2003 CONTENTS INTRODUCTION DEFINITIONS USERMOWNER RESPONSIBILITY WHAT AN AUTOSET RESPOND UNIT IS FOR TELL YOUR DOCTOR IF WARNINGS CAUTIONS POSSIBLE SIDE EFFECTS THE AUTOSET RESPOND AUTOSET RESPOND COMPONENTS MASKS HUMIDIFIERS PREPARING FOR SETTING UP THE AUTOSET RESPOND FEATURES OF THE AUTOSET RESPOND OPERATING INSTRUCTIONS STARTING TREATMENT STOPPING TREATMENT USING THE MASK FITTING FEATURE HELPFUL HINTS CLEANING AND MAINTENANCE DAILY WEEKLY PERIODICALLY REPLACING THE AIR FILTER SERVICING TROUBLESHOOTING 24 2 OCDE RU SC la Cas SYSTEM SPECIFICATIONS LIMITED WWVARRANTY yest ea eb ra S Ec S US bh BRN sch sch CONTENTS INTRODUCTION DEFINITIONS This manual contains special terms and icons that appear in the margins to draw your attention to specific and important information WARNING Alerts you to possible injury CAUTION Explains speci
10. ER 7 Connect the long air tubing 6ft6in to the air outlet connector PREPARING FOR USE 13 14 8 Assemble your mask system according to the mask user instructions 9 Connect the mask system to the free end of the long air tubing The final assembly should lock like this HUMIDAIRE HumidAire Users Plug the HUMIDAIRE power cord into a power outlet and turn the power on 11 Connect the power cord to the socket at the rear of the AUTOSET RESPOND Plug the other end of the power cord into a power outlet and turn the power on WARNING Make sure that the power cord and plug are in good condition and the equipment is not damaged 12 AUTOSET RESPOND is now ready for use with the HUMIDAIRE or the ResMed PASSOVER FEATURES OF THE AUTOSET RESPOND LCD SCREEN AND KEYPAD The control panel of the AUTOSET RESPOND includes an LCD screen and keypad PREPARING FOR USE AutoSet LCD screen Left key Right key Up down key Front key The AUTOSET RESPOND keypad has the following keys Key Function Front aa FE p e Starts or stops treatment Extended hold for at least 3 seconds starts the mask fitting feature Up Down 8 Allows you to scroll through the AuTOSET RESPOND menus submenus and setting options Allows you to change the settling time from the standby screen Left Si Performs the function indicated by the guidin
11. WARNINGS RELATED TO TREATMENT WARNING f you stop your CPAP treatment your sleep apnea will return Always consult your clinician if you expect to be in a situation where you cannot use your AUTOSET RESPOND unit f you are admitted to a hospital or prescribed any other form of medical treatment always inform the medical staff that you are being treated with CPAP It is also important to contact the clinician who is treating you for sleep apnea f you experience an infection of the upper respiratory tract middle ear or sinuses contact your clinician before continuing your CPAP treatment You may be advised to stop treatment until the infection has cleared If you continue with treatment during an infection be sure to clean your mask and tubing after every use WARNINGS RELATED TO OXYGEN USE WARNING If oxygen is used with this device the oxygen flow must be turned off when the device is not operating Explanation When the CPAP device is not in operation and the oxygen flow is left on oxygen delivered into the air delivery tubing may accumulate within the CPAP machine enclosure and create a risk of fire This applies to most types of CPAP machines Oxygen supports combustion Oxygen should not be used while you are smoking or in the presence of an open flame Always ensure airflow is being generated by the device before the oxygen supply is turned on Always turn the oxygen supply off before stopping the a
12. al measures for the safe and effective use of the device Note 5 an informative or helpful note USER OWNER RESPONSIBILITY The user or owner of this system shall have sole responsibility and liability for any injury to persons or damage to property resulting from operation which is not in accordance with the operating instructions supplied maintenance or modifications carried out unless in accordance with authorized instructions and by authorized persons Please read this manual carefully before use MEDICAL INFORMATION WHAT AN AUTOSET 9 RESPOND UNIT IS FOR The AUTOSET ResPOND CPAP system is indicated for the treatment of obstructive sleep apnea OSA in adult patients The optional integrated HUMIDAIRE 2im humidifier is indicated for the humidification and warming of air from the AUTOSET RESPOND flow generator The optional integrated HUMIDAIRE 2iC humidifier is indicated for the humidification of air from the AutoSet Respond flow generator The AUTOSET RESPOND CPAP flow generator HUMIDAIRE 2i and HUMIDAIRE 2iC are for home and hospital use CAUTION Federal law restricts this device to sale by or on the order of a physician TELL YOUR DOCTOR IF You must tell your doctor and CPAP continuous positive airway pressure therapy must not be used if you have any of the following conditions pneumothorax or pneumomediastinum air in the pleural cavity or mediastinum INTRODUCTION severe heart failure
13. and hang it in a clean dry place until next use Do not hang the air tubing in direct sunlight as it may harden and crack over time Clean the mask according to the mask user instructions f you are using a humidifier clean it according to the instructions in the manual Remove the air tubing from the AUTOSET RESPOND unit and the mask Wash the mask system according to the instructions supplied with it Wash the air tubing in warm water using mild detergent Rinse thoroughly hang and allow to dry Before next use assemble the mask and headgear according to the mask user instructions Reconnect the air tubing to the pressure sensor adapter and mask CAUTION Do not use bleach chlorine alcohol or aromatic based solutions including all scented oils moisturizing or antibacterial soaps to clean the cushion mask air tubing or the AutoSet Respond These solutions may cause hardening and reduce the life of the product Do not wash or dry the mask frame at a temperature above 80 176 F Exposure to higher temperatures may reduce the life of the product Do not hang the air tubing and pressure sensor adapter in direct sunlight as the tubing may harden over time and eventually crack PERIODICALLY 4 The mask and air tubing are subject to normal wear and tear Inspect them regularly for damage Clean the exterior of the flow generator with a damp cloth and mild detergent Inspect the air filter to check if it is blocked
14. by dirt or contains holes See Replacing the Air Filter on page 22 Replace the pressure sensor adapter every 2 months CLEANING AND MAINTENANCE 21 22 WARNING Beware of electric shock Do not immerse the flow generator or power cord in water Always unplug the flow generator before cleaning and be sure that it is dry before reconnecting CAUTION Do not attempt to open the AutoSet Respond There are no user serviceable parts inside Repairs and internal servicing should only be performed by an authorized service agent REPLACING THE AIR FILTER Inspect the air filter every month to check if it is blocked by dirt or contains holes With normal use of an AUTOSET RESPOND unit the air filter needs to be replaced every six months or more often if your unit is in a dusty environment To replace the air filter Remove the air filter cover at the back of the AUTOSET RESPOND Air filter cover 2 Remove and discard the old air filter 3 Insert a new filter with the blue tinted side facing out from the unit 4 Replace the air filter cover WARNING Do not wash the air filter The air filter is not washable or reusable Note The air filter should be inspected once a month SERVICING Your AUTOSET RESPOND flow generator is designed to give you years of trouble free operation The flow generator should not require regular servicing if it is maintained according to the instructions in this manual If you f
15. eel that your unit is not performing properly see Troubleshooting on page 23 CAUTION Inspection and repair should only be performed by an authorized agent Under no circumstances should you attempt to service or repair the flow generator yourself TROUBLESHOOTING If there is a problem try the following suggest contact your equipment supplier or ResMed Problem No display Possible Cause Power not connected or switch at back is not on ions If the problem cannot be solved Do not attempt to open the unit TROUBLESHOOTING Solution Ensure the power cable is connected and that the switch at the back of the unit is in the ON position Insufficient air delivered from AutoSet Respond Settling Time is in use Air filter is dirty Air tubing is kinked or punctured Air tubing not connected properly Mask and headgear not positioned correctly Plugs missing from access port on mask Pressure required for treatment may have changed Wait for air pressure to build up Replace air filter Straighten or replace tubing Check air tubing Adjust position of mask and headgear Replace plugs See your clinician to adjust the pressure AutoSet Respond does not start Power cord not connected properly Power outlet may be faulty AutoSet Respond unit not switched on There is excessive leak Connect power cord firmly at both ends Try another power outlet Swi
16. ent size or style of mask You can also use the mask fitting feature to help you fit your mask properly See Using the Mask Fitting Feature on page 18 Before wearing your mask wash your face to remove excess facial oils This will allow a better fit and prolong the life of the mask cushion OPERATING INSTRUCTIONS 19 20 NASAL IRRITATION DRYNESS You may experience dryness of the nose mouth and or throat during the course of treatment especially during winter In many cases a humidifier may resolve this discomfort Contact your clinician for advice RUNNY OR BLOCKED NOSE You may experience sneezing and or a runny or blocked nose during the first few weeks of treatment In many cases nasal irritation can be resolved with a humidifier Consult your clinician for advice TRAVELLING WITH THE AUTOSET RESPOND INTERNATIONAL USE Your AUTOSET RESPOND flow generator has an internal power adapter that enables it to operate in other countries It will operate on power supplies of 100 240V and 50 60Hz No special adjustment is necessary but you may need a plug adapter for the power outlet CLEANING AND MAINTENANCE You should regularly carry out the cleaning and maintenance described in this section WARNING DAILY 4 3 WEEKLY l 2 3 4 5i Do not wash the pressure sensor adapter Do not wash the pressure sensor line that is used with the HumidAire or the ResMed Passover Disconnect the air tubing
17. fied below from the date of purchase by the initial consumer Product Warranty Period AutoSet CS flow generator ResMed humidifiers ResControl 1 Year ResLink CPAP continuous positive airway pressure APAP automatic 2 Years positive airway pressure bilevel and AutoSet CS 2 flow generators Accessories mask systems including mask frame cushion 90 Days headgear and tubing Excludes single use devices Note Some models are not available in all regions Ifthe product fails under conditions of normal use ResMed will repair or replace at its option the defective prod not cover a any damage uct or any of its components This Limited Warranty does caused as a result of improper use abuse modification or alteration of the product b repairs carried out by any service organization that has not been expressly authorized contamination due to ciga water being spilled on or by ResMed to perform such repairs c any damage or rette pipe cigar or other smoke d any damage caused by into the flow generator Any product repaired or replaced under warranty will be returned freight prepaid to the dealer designated by the consumer The cost of transporting the product to an authorized service organization will be borne by the consumer This warranty is in lieu of al other express or implied warranties including any implied warranty of merchantability or fitness for a par
18. g text displayed above it on the LCD screen Guiding text includes menu enter change and apply Right Performs the function indicated by the guiding text displayed above it on the LCD screen Cuiding text includes exit and cancel To assist you in adjusting the AUTOSET RESPOND the keypad and LCD are equipped with a backlight The LCD backlight comes on when the unit is turned on or when you press a key and turns off after 2 minutes The keypad backlight is on at all times when the AUTOSET RESPOND is on USING THE MENUS The AUTOSET RESPOND unit provides a set of functions which are arranged in menus Via the LCD screen the menus allow you to view and change the settings for a particular function You can access the menus when the AUTOSET RESPOND is in stand by mode or downloading data MENU The menu allows you to view details about the usage time and software version of your AUTOSET RESPOND You can also change the settling time and mask type 15 16 The standby screen allows you to view change the settling time The Used screen displays the total number of days and hours for which the AUTOSET RESPOND has been used The Mask screen allows you to view and change the mask type The SW screen displays the current software version instal THE STANDBY SCREEN The standby screen shows the current sett to change the settling time Settling time is ed in the AUTOSET RESPOND ling time
19. ibed in the mask user instructions Hold down the Front key for at least 3 seconds until air pressure delivery starts The following display will appear on the LCD screen indicating that the mask fitting feature is in operation The flow generator will ramp to the mask fit pressure and will remain at this pressure for 3 minutes A mask fit star rating is also displayed See Definitions of Mask Fit Star Rating on page 19 MaskFit Mask fit star rating excelnt RESMED Note The mask fit star rating display disappears after 3 minutes The mask fitting feature can only be started from the standby screen The mask fit pressure is the highest pressure that you are likely to receive based on the previous night s session or 10 cm H30 whichever is greater Adjust your mask mask cushion and headgear until you have a secure and comfortable fit Once you have a secure and comfortable fit check your mask fit star rating on the LCD screen Definitions of the mask fit star ratings are presented in Table Note If there is another person nearby to check your mask fit star rating you can adjust your mask mask cushion and headgear while lying down After 3 minutes the pressure will drop to the set minimum pressure and treatment will begin Definitions of the mask fit star ratings are presented in Table e If you do not wish to wait 3 minutes hold down the Front key for at least 3 seconds and treatment will begi
20. iliarize yourself with the following components of the AUTOSET RESPOND unit AUTOSET RESPOND Unit Front View LCD outlet Removable front cap CD D Power cord Pressure sensor adapter WARNING Filter cove Power Rear View switch Power socket Carry bag Auxiliary Serial port communications port Air tubing 6ft6in 2m Do not connect any device to the auxiliary port Although your health care provider may connect specially designed devices to the auxiliary port of the AutoSet Respond unit connection of other devices could result in injury or damage of the unit In the clinical environment any personal computer that is used with your CPAP system must be at least 5ft 1 5m away from or at least 8ft 2 5m above the patient It must also comply with the relevant test standard For personal computers the international standard is IEC or equivalent THE AUTOSET RESPOND SYSTEM MASKS You will also need a ResMed mask system supplied separately The following ResMed mask systems are recommended for use with the AUTOSET RESPOND MIRAGE VISTA MASK MIRAGE MASK MODULAR FULL MASK FULL FACE MASK SERIES 2 WARNING Only ResMed mask systems are compatible for use with the AutoSet Respond Please refer to Warnings on page 2 HUMIDIFIERS A humidifier may be required if you are experiencing dryness of the nose throat o
21. irflow from the device Note At a fixed rate of supplemental oxygen flow the inhaled oxygen concentration will vary depending on where the oxygen is introduced the pressure settings patient breathing pattem mask selection and leak rate INTRODUCTION CAUTIONS The following are general cautions Specific cautions appear next to the relevant instructions in the manual Do not open the AUTOSET RESPOND case There are no user serviceable parts inside Repairs and internal servicing should only be performed by an authorized service agent POSSIBLE SIDE EFFECTS The AUTOSET RESPOND flow generator is designed to help you have a good night s sleep However you need to be aware of possible problems that may arise during CPAP treatment WARNING Consult your clinician immediately if you experience any of the following symptoms during your CPAP treatment headache middle ear or sinus discomfort chest pain dryness of the nose mouth or throat feeling bloated due to air swallowing air continually leaking out of the mouth while sleeping recurrence of any sleep apnea symptoms while on CPAP Skin irritation may occur from sensitivity to the mask materials or from excessively tight headgear straps A correctly fitted mask and appropriately adjusted straps will often prevent skin irritation If problems persist contact your clinician for advice AUTOSET RESPOND SYSTEM AUTOSET RESPOND COMPONENTS Please identify and fam
22. layed on the LC screen then the product name is displayed The standby screen then appears Note If you have the HumidAire 2i attached please refer to the HumidAire 2i User s Manual for operating instructions Fit your mask as described in the mask user instructions WARNING Before starting therapy ensure that you are wearing the mask and that the mask and air tubing are fitted correctly Lie down and arrange the air tubing so that it is free to move if you turn in your sleep CAUTION Do not leave long lengths of air tubing around the top of your bed It could twist around your head or neck while you are sleeping Make sure the area around the flow generator is dry and clean It should also be clear of bedding clothes and other potential blockages To start treatment press the Front key After starting treatment the treatment pressure is displayed on the LCD Note Pressure displayed on the LCD during therapy is the intended delivery pressure STOPPING TREATMENT To stop treatment at any time remove your mask and press the Front key OPERATING INSTRUCTIONS 17 18 USING THE MASK FITTING FEATURE The AUTOSET RESPOND mask fitting feature can be used to help you fit your mask properly The mask fitting feature delivers air pressure for a three minute period prior to starting treatment for checking and adjusting your mask fit to minimize leaks To use the mask fitting feature Fit your mask as descr
23. n immediately f you press the Front key for less than 3 seconds the unit will return to standby mode the standby screen is displayed Table 1 Definitions of Mask Fit Star Rating Star rating Definition FEX Excellent A Very good We Good Na Adjust mask Kaes Adjust mask ee Adjust mask HELPFUL HINTS STARTING OUT FIRST USE When you first use your AUTOSET RESPOND unit you may feel uncomfortable breathing against the air flow This is normal You will continue to breathe normally while you sleep Take time to adjust to this new sensation Taking a few deep breaths may help GETTING UP f you need to get up during the night remove your mask and turn the flow generator off Always remember to put your mask on and turn your flow generator on when you return to bed MOUTH LEAKS f using a nasal mask try to keep your mouth closed during treatment Air leaks from your mouth can decrease the effectiveness of your treatment If mouth leaks are a problem a full face mask or chin strap may help Contact your clinician or equipment supplier for further details MASK FITTING The flow generator delivers the most effective treatment when the mask is well fitted and comfortable Treatment can be affected by leaks so it is important to eliminate any leaks that may arise If you have problems trying to get a comfortable mask fit contact your sleep clinic or equipment supplier You may benefit from a differ
24. r mouth The AUTOSET RESPOND is compatible for use with the following humidifiers HUMIDAIRE 2iC humidifier HUMIDAIRE heated humidifier ResMed PASSOVER humidifier WARNING Only the HumidAire 2i HumidAire 2iC HumidAire heated humidifier and the ResMed Passover are compatible for use with the AutoSet Respond Please refer to Warnings on page 2 WARNING Only the 6ft6in 2m air tubing can be used with the AutoSet Respond The following humidifier accessories are available for use with the AUTOSET RESPOND for HUMIDAIRE and ResMed PASSOVER only gt Pressure Short air tubing Medium air tubing sensor line 7 5in 19cm 2lin 52cm HUMIDAIRE and PASSOVER HUMIDAIRE only HUMIDAIRE and PASSOVER THE AUTOSET RESPOND SYSTEM PREPARING FOR USE SETTING UP THE AUTOSET RESPOND 1 Place the AUTOSET RESPOND unit on a table near the head of your bed CAUTION Be careful not to place the device where it can be bumped or where someone is likely to trip over the power cord Note The AutoSet Respond unit may be placed on the floor beside or underneath your bed Ensure that the area is free from dust and clear of bedding clothes or any other objects that could block the air inlet Connect the power cord to the socket at the rear of the flow generator Plug the other end of the power cord into a power outlet WARNING Make sure the power cord and plug are in good condition and the equipment is not damaged The
25. rature is warmer than 32 C 90 F The AUTOSET RESPOND unit should only be used with masks and connectors recommended by ResMed or by your physician or respiratory therapist A mask should not be used unless the AUTOSET RESPOND unit is turned on and operating properly The vent hole or holes associated with the mask should never be blocked Explanation The AUTOSET RESPOND unit is intended to be used with special masks or connectors which have vent holes to allow continuous flow of air out of the mask When the device is turned on and functioning properly new air from the device flushes the exhaled air out through the mask vent holes However when the device is not operating insufficient fresh air will be provided through the Ports may be incorporated in the mask or in connectors that are near the mask mask and the exhaled air may be rebreathed Rebreathing of exhaled air for longer than several minutes can in some circumstances lead to suffocation This applies to most models of CPAP devices The AUTOSET RESPOND unit should only be connected to the components humidifiers or accessories specified in this manual Connection of other items may result in injury or damage to the AUTOSET RESPOND unit At low pressures the flow through the exhalation ports of your mask may not clear all exhaled gas from the tubing Some rebreathing may occur Explosion hazard do not use in the vicinity of flammable anaesthetics
26. sers Place the filled water chamber inside the HUMIDAIRE Connect the medium 2lin 52cm air tubing to the left connector port and the short 7 5in 19cm air tubing to the right connector port on the humidifier Close the HUMIDAIRE lid ResMed Passover Users Connect the medium air tubing to the left connector on the PASSOVER Short 7 5in air tubing Medium 2 lin air tubing Medium 2 lin air tubing HUMIDAIRE ResMed PASSOVER PREPARING FOR USE 11 3 Place the AUTOSET RESPOND on top of the HUMIDAIRE PASSOVER Do not place the AUTOSET RESPOND unit underneath the humidifier This is to avoid water spilling into the AUTOSET RESPOND unit HUMIDAIRE amp ResMed PASSOVER 4 HumidAire Users Connect the free end of the medium air tubing to the air outlet of the AUTOSET RESPOND ResMed Passover Users Connect the free end ofthe air tubing to the air outlet of the AUTOSET RESPOND HUMIDAIRE ResMed PASSOVER 5 HumidAire Users Take the pressure sensor line and press the air outlet connector onto the short air tubing ResMed Passover Users Take the pressure sensor line and press the air outlet connector onto the right connector port on the PASSOVER 12 ResMed PASSOVER CAUTION Do not wash the pressure sensor line It is not washable 6 Connect the free end of the pressure sensor line firmly onto the pressure port of the AUTOSET RESPOND HUMIDAIRE ResMed PASSOV
27. tch power switch at rear of AutoSet Respond to ON Adjust position of mask and headgear Plugs may be missing from ports on mask Replace them Air tubing not connected properly Connect firmly at both ends Air tubing kinked or punctured Straighten or replace 23 Problem Pressure rises inappropriately Possible Cause Talking coughing voluntarily breath holding eg while rolling over in bed or intentionally breathing in an unusual manner Mask cushion is buzzing against the skin Cushion seated incorrectly Cushion seated incorrectly causing excessive leak Air tubing is kinked or punctured Humidifier control dial set too high resulting in accumulation of water in the air tubing Solution Avoid talking with a nasal mask on and breathe as normally as possible Adjust the headgear Adjust headgear or re fit cushion Straighten or replace air tubing Turn humidifier control dial down and empty the water from the air tubing Display error message Check tube Key if done The air tubing is loose or blocked Check that the air tubing is connected securely to your mask and the air outlet on the front of the AutoSet Respond When you have checked the air tubing press the Front key to restart the machine If this does not clear the message turn the power switch off and then turn it back on to restart the machine Display error message SYSTEM ERROR Call service
28. ticular purpose Some regions or states do not allow limitations on how long an implied warranty lasts so the above limitation may not apply to you ResMed shall not be responsible for any incidental or consequential damages claimed to have occurred as a result of the sale installation or use o f any ResMed product Some regions or states do not allow the exclusion or limitation of incidental or consequential damages so the above limitation may not apply to you This warranty gives you specific legal rights and you may also have other r region to region For further information on your warranty rights contact your local ResMed dealer or ResMed office ights which vary from R000 303 6 03 08 LIMITED WARRANTY 27 28 INDEX Accessories 7 Air Filter 22 25 Air Tubing 25 B Back light LCD and Keypad 15 Cleaning Maintenance 21 Cleaning Periodically 21 Components 5 CPAP 1 D Daily Cleaning 21 Definitions 1 Dimensions 25 Dryness 20 E Environmental Conditions 25 Error Messages 24 F Features of the AutoSet Respond 15 Filter 22 First Use 19 Flow Pressure Characteristics 27 Front Key 15 G Getting Up 19 Glossary of Symbols 26 H Helpful Hints 19 Housing Construction 25 HumidAire 2i 10 HumidAire 2iC 10 11 Humidifier Accessories 7 Humidifier Use 10 Humidifiers 6 I International Use 20 K Keys functions 15 INDEX L LCD Screen and Keypad 15 Left Key 15 M

Download Pdf Manuals

image

Related Search

Related Contents

FIXA Sierra de calar  Digitus PCI Serial interface card  Chef'sChoice 834 Use and Care Manual  MELSEC iQ-F FX5-4AD-ADP Hardware Manual  WSG18SFP User Manual  Samsung AS09FCN manual de utilizador  Bijlage 26 Generating Type 1 Fonts from Metafont Sources  UM10848 - NXP Semiconductors  6 - Smartcaisse  Sony KV-XA25M61 25" TV  