1. Note To protect the crosslinker from corrosive buffers lay a piece of plastic wrap on the compartment bottom before placing the blotting paper and sample important safety information Important safety information Always wear protective gloves and goggles when working with the crosslinker The UVC 500 is a powerful source of ultraviolet radiation The door is safety interlocked to prevent accidental user exposure to UV light Do not attempt to override the safety interlocks If the safety interlock is disengaged the UV lamps do not light and the unit will beep 3 times when the START key is pressed If a malfunction occurs and the lamps in the crosslinker remain on when the door is open immediately close the door and turn off the mains power For repair information call the Amersham Biosciences Technical Support The view window blocks UV radiation and allows viewing while the unit is in operation If the window is broken or removed stop using the crosslinker immediately For repair information call Amersham Biosciences Technical Support Always disconnect the mains power cord before cleaning the unit or replacing fuses or lamps Operating instructions Set up and membrane placement 1 2 Place the unit on a stable and level surface Plug the power cord into the receptacle at the back of the unit and into a properly grounded electrical outlet of the proper power rating Flip the mains power switch locate
2. restart close the door and press the START key uvc 500 ultraviolet crosslinker e care and maintenance Caution Always turn the mains power switch off and unplug the instrument before cleaning Important The photofeedback system is calibrated to 254 nm UV light Do not install lamps rated for a different wavelength um 80 6256 32 e p6 Care and maintenance Cleaning Preventative maintenance includes regular cleaning of the compartment surface the sensor and UV lamps Use mild soap and water on a soft cloth or sponge to clean the exterior Gently wipe the photofeedback sensor and UV lamps regularly with a soft cotton cloth dampened with alcohol Always wipe any water and spills from the inside of the unit and clean away all chemicals after each use Never use abrasive cleaners acetone or chloroform on any part Replacing UV lamps Periodically check that all five lamps function by looking through the view window while the unit is on If one or more lamps are burned out it is recommended that all five be replaced at the same time To replace each lamp 1 Turn off the mains power switch and disconnect the power cord 2 Locate the lamps along the top of the compartment Remove each lamp by carefully rotating it 1 4 1 3 turn and then carefully pulling it loose 3 To insert the new lamp hold it so that the metal plugs at the ends align verti cally Place the lamp in the sockets at each end of the uni
3. 90 594 Sarkar Gobinda and Sommer Steve S 1990 Shedding Light on PCR Contamination Nature 343 27 Yamada H and Hieda K 1992 Wavelength Dependence 150 290nm of the Formation of the Cyclobutane Dimer and the 6 4 Photoproduct of Thymine Photochemistry and Photobiology 55 541 548 customer service information Customer service information Technical service and repair Amersham Biosciences offers complete technical support for all our products If you have any questions about how to use this product or would like to arrange to repair it please call or fax your local Amer sham Biosciences representative Important Request a copy of the Amersham Biosciences Conta mination Clearance Certificate before returning the item No items can be accepted for servicing or return unless this form is properly completed Ordering information Quantity Code No UVC Ultraviolet Crosslinker For membranes to 25 x 30 cm 115 VAC 1 80 6222 31 230 VAC 1 80 6222 50 Replacement UV lamp 8 W 254 nm 1 80 6223 64 Replacement fuses 5 80 6109 45 uvc 500 ultraviolet crosslinker e notes Notes um 80 6256 32 e p10 is a trademark of Amersham Biosciences Limited or its subsidiaries Amersham Amersham Biosciencesand is a trademark of Amersham plc All other trademarks and registered trademarks are the property of their respective companies or organizations O 2000 Amersham Biosciences Inc All rights r
4. Qu blotting equipment UVC 500 Ultraviolet Crosslinker transilluminator Amersham um 80 6256 32 Rev DO 02 00 Biosciences Page finder UV Crosslinker function and description cece eee 1 UnpackKillg ore n eem m ett bete A aa 2 Specifications ee ee ee ee ee ee k kkkh kk kkk k k k kk i i LL EEE EEIE EEnEnnn nnn 2 Important safety information eee 3 Operating instructions ee ee kikki kikki k ki kkk i k LLE n rnnr nann n an 3 Energy setting eee eene kk kk a xa 4 Time setting 5 ee ee 5 Care and maintenance keke yekeke 6 Bibliography eee eene ee ee ee ee ee 8 Customer service information ooo ee RR 9 um 80 6256 32 e pii Crosslinker front view uv crosslinker function and description UV Crosslinker function and description The UVC 500 Ultraviolet Crosslinker provides a uniform and intense source of ultraviolet UV light for such uses as e DNA crosslinking e DNA nicking e Rapid Site mapping e Testing RecA function e Ultraviolet sterilization e Elimination of PCR contamination The internal photo feedback system continually measures the energy within the chamber and automatically deactivates the UV light source once the programmed energy level is reached This system also automati cally adjusts to variations in UV intensity that occur as lamps age Pul
5. d under the keypad to on Note The instrument settings are always the values last programmed and are indicated by the LED display and a lit red indicator lamp Open the door by pulling the door handle above the window toward you Note Do not obstruct or damage the photofeedback sensor which is mounted on a small metal tab protruding from the lower right side Moisten do not wet completely a sheet of blotting paper with transfer buffer and place it in the center of the compartment Place the membrane on the paper with the nucleic acid side facing upward Set the energy exposure energy or time as described in the next section e p3 uvc 500 ultraviolet crosslinker energy setting UVC 500 Crosslinker user interface The LED display shows values for energy multiply by 100 for actual value in J cm and time in minutes and tenths of a minute starr e ENERGY e TIME STOP RESET o PRESET fe ENERGY X100 y J CM TIME 999 9 MINUTES MAX um 80 6256 32 e p4 Energy setting The amount of energy recommended for UV crosslinking of DNA and RNA by covalently binding nucleic acids to Hybond membranes after Northern Southern slot or dot blotting is 70 000 nJ cm This value is preset at the factory and displays as 700 on the LED display which only shows 4 digits As the unit operates a countdown of the remaining energy to be emitted is displayed on the LED Note Multiply all displayed ene
6. eserved All goods and services are sold subject to the terms and conditions of sale of the company within the Amersham Biosciences group that supplies them A copy of these terms and conditions is available on request Printed in the USA Amersham Biosciences UK Limited Amersham Place Little Chalfont Buckinghamshire England HP7 9NA Amersham Biosciences AB SE 751 84 Uppsala Sweden Amersham Biosciences Inc 800 Centennial Avenue PO Box 1327 Piscataway NJ 08855 USA Amersham Biosciences Europe GmbH Munzinger Strasse 9 D 79111 Freiburg Germany www amershambiosciences com Amersham e Biosciences
7. l the door open along this handle LED display SEA UV Crosslinker View window Keypad Power switch e pl uvc 500 ultraviolet crosslinker e unpacking um 80 6256 32 e p2 Unpacking Unwrap all packages carefully and compare contents with the packing list making sure all items arrived If any part is missing contact your local Amersham Biosciences sales office Inspect all components for damage that may have occurred while the unit was in transit If any part appears damaged contact the carrier immediately Be sure to keep all packing material for damage claims or to use should it become neces sary to return the unit Specifications Ultraviolet wavelength and sources 254 nm shortwave Five 8 W 254 nm UV dual bi pin discharge type tubes Features Microprocessor controlled UV sensor feedback system e Large LED readout Tactile membrane switch keypad e Laydown type door Multiple set functions pre set UV energy or time exposure user specified UV energy or time exposure Max energy setting 999 900 uJ cm Max time setting 999 9 min Dimensions h x d x w External 22 2 x 34 9 x 40 0 cm 8 75 x 13 75 x 15 75 in Chamber interior 12 7 x 30 5 x 25 4 cm 5 x 12 x 10 in Input fuses e 115 V F 1A 250V 5 x 20 mm 2 fuses e 230 V F 1A 250V 5 x 20 mm 2 fuses Weight e 7 5 kg 15 5 lbs Safety e Safety interlocked door e UV blocking view window e Safety fused and removable power cord
8. ntdown value displays At the end of the exposure the unit automatically stops and beeps 5 times Reprogramming the preset time setting 1 Press and hold the PRESET key until you hear the second beep The lamp next to the PRESET key flashes when this key switches to program mode 2 Press the TIME key Use the numeric keypad to enter the new value The LED display shows the new value 3 Press the ENTER key Programming the TIME key 1 Press the TIME key 2 Enter a value using the numeric keypad The LED will display the new value 3 Press the ENTER key After programming pressing the TIME key retrieves this setting Change the setting by entering a new value and pressing ENTER Stopping and restarting the crosslinker Press the red STOP RESET key to stop the exposure The LED will display the remaining exposure level or time Press the green START key to restart the exposure The program will continue from where it was stopped To enter a different value once the program has started press STOP RESET enter the new value then press ENTER Press the START key to begin the new exposure value To revert to the last used setting after the exposure is interrupted press the STOP RESET key a second time Press START to start exposure from the previous setting f the door is accidentally opened during exposure the UV lamps instantly turn off and the remaining exposure level or time is displayed on the LED display To
9. rgy settings by 100 for the actual energy value Using the preset energy setting 1 First press the PRESET key and then the ENERGY key The red indicator lamp next to each of these keys should light and the LED should display 700 or the value last programmed 2 Press the green START key The UV lamps take a few seconds to energize before the countdown value displays At the end of the exposure the unit automatically stops and beeps 5 times Reprogramming the preset energy setting 1 Press and hold the PRESET key until you hear the second beep The lamp next to the PRESET key flashes when this key switches to program mode 2 Press the ENERGY key Use the numeric keypad to enter the new value The LED display should show the new value 3 Press the ENTER key Programming the ENERGY key 1 Press the ENERGY key 2 Enter a value using the numeric keypad The LED will display the new value 3 Press ENTER After programming pressing the ENERGY key retrieves this setting Change the setting by entering a new value and pressing the ENTER key time setting Time setting The display shows time in minutes and tenths of a minute Using the preset time setting 1 First press the PRESET key and then the TIME key The red indicator lamp next to each of these keys should light and the LED should display 5 0 or the value last programmed 2 Press the green START key The UV lamps take a few seconds to energize before the cou
10. sequencing Proc Natl Acad Sci USA 81 1991 1995 Khandjian E W 1987 Optimized hybridization of DNA blotted and fixed to nitrocellulose and nylon membranes BioTechnology 5 165 167 Khandjian E W 1986 UV crosslinking of RNA to nylon membrane enhances hybridization signals Molec Biol Report 11 107 115 Reed K C and D A Mann 1985 Rapid transfer of DNA from agarose gels to nylon membranes Nuc Acid Res 13 7207 7221 DNA Nicking Vollrath D and R W Davis 1987 Resolution of DNA molecules greater than 5 megabases by contour clamped homogeneous electric fields Nuc Acid Res 15 7865 7876 Rapid Site Mapping Whittaker P A and E M Southern 1986 Ultraviolet irradiation of DNA A way of generating partial digests for rapid restriction site mapping Gene 41 129 134 Testing RecA function Maniatis T 1989 Maintenance and Testing of Lysogens of Bacterio phage Lambda in Molecular Cloning A Laboratory Manual eds Sambrook Fritsch and Maniatis 2 96 Ultraviolet sterilization Reich R R and Anderson H D 1986 Sterilization of Membrane Filters with Ultraviolet Radiation Pharmaceutical Manufacturing Jan Elimination of PCR contamination Wilson K H 1992 UV Absorption Complicates PCR Decontamina tion BioTechniques 13 208 210 Sarkar Gobinda and Sommer Steve S 1991 Parameters Affecting Susceptibility of PCR Contamination to UV Inactivation BioTechniques 10 5
11. t and lightly push it up into place Rotate the tube turn so that it locks into place UVC 500 Crosslinker back view 115 V model care and maintenance Replacing fuses Fuses protect equipment by disconnecting loads too large for the instru ment s circuit design For continued protection only use fuses with the specified voltage and current rating Each of two round modules on the left side of the control panel above the mains power cord receptacle holds one F 1A 250V 5 x 20 mm Fast blo fuse Model ID and power specifications Mains power cord receptacle Fuse modules 2 Caution Set the mains power switch to off and detach the power cord before replacing fuses Insert a small flat blade screwdriver into the slot on the fuse module and turn it 1 4 turn counterclockwise The spring loaded module cap will loosen and you can then pull the cap fuse holder out Pull the fuse out of its holder and inspect If the fuse element is burned or broken replace the fuse with an identical type If the fuse appears to be intact check it with an ohmmeter Insert a good fuse into the holder and then insert this assembly back into the unit Seat the module by inserting the screwdriver into the slot pressing gently and turning the cap 1 4 turn clockwise uvc 500 ultraviolet crosslinker e bibliography um 80 6256 32 e p8 Bibliography DNA Crosslinking Church M C and G Gilbert 1984 Genomic