Contents
1. gt If the power indicator is ON and the pump still doesn t work contact your local dealer immediately gt If the power indicator is OFF there may be a faulty outlet Try to connect the power cord to another outlet If the power indicator is still OFF contact a qualified electrician for main power check 1 For a quick check adjust the pressure to Firm 2 Check to see if the inflation tubes connected to the pump are twisted or disconnected or any leakage occurring Incomplete inflation Low pressure gt Always keep the tubes straight gt Change the tubes if there is any leakage Cleaning In this section we describe the procedures to clean and decontaminate the pump and mattress It is important to follow these procedures before using the system on patients The cleaning task is required at least once a week to maintain personal hygiene Pump gt DO NOT immerse or soak the pump gt Check for external damage and move the pump to the cleaning area gt Place the pump on a work surface and spray or wipe the outside of the case with quaternary ammonium solution DO NOT spray any cleaning solution directly on the surface of the pump DO NOT use a Hypocarbonate or Phenolic based cleaning solution as this may cause damage to the case Allow the solution to incubate for 10 minutes or accordingly as stated by the cleaning product used gt Wipe case with a clean cloth Make sure all areas are cl
2. 70 C gt Humidity Operation 15 93 non condensing Storage lt 93 non condensing Safety Standards UL c UL CE Mattress Mattress Specification Size 78 x 35 4 x 7 8 Material Nylon PVC EN 71 low toxic Weight Capacity 160 kg or 350 Ibs Warranty Your Drive brand product is warranted to be free of defects in materials and workmanship for one year of the original consumer purchaser This device was built to exacting standards and carefully inspected prior to shipment This 1 Year Limited Warranty is an expression of our confidence in the materials and workmanship of our products and our assurance to the consumer of years of dependable service This warranty does not cover device failure due to owner misuse or negligence or normal wear and tear The warranty does not extend to non durable components such as rubber accessories casters and grips which are subject to normal wear and need periodic replacement If you have a question about your Drive device or this warranty please contact an authorized Drive dealer NOTE NOTE V 01 Jan 2014
3. User Manual Alternating Pressure Pump and Mattress System Item AP8000N IMPORTANT SAFEGUARDS When using electrical products especially when children are present basic safety precautions should always be followed including the following READ ALL INSTRUCTIONS BEFORE USING DANGER To reduce the risk of electrocution gt Always unplug this product immediately after use Do not use while bathing gt Do not place or store product where it can fall or be pulled into a tub or sink gt Do not place in or drop into water or other liquids gt Do not reach for a product that has fallen into water Unplug immediately WARNING To reduce the risk of burns electrocution fire or injury to persons gt A product should never be left unattended when plugged in gt Close supervision is necessary when this product is used by on or near children or physically challenged individuals gt Use this product only for its intended use as described in this manual Do not use attachments not recommended by the manufacturer gt Never operate this product if it has a damaged cord or plug if it is not working properly if it has been dropped or damaged or dropped into water Return the product to a service center for examination and repair gt Keep the cord away from heated surfaces gt Never block the air openings of the product or place it on a soft surface such as a bed or couch where the air openings may be bl
4. e switch is on the front faceplate of the pump Turn ON the power the pump will start operation gt Pressure adjust Knob 2 Turn the soft firm knob to set a comfortable pressure level Mattress Lay it flatly on a bed frame Please refer to the Specifications section for details Secure the mattress to the bed frame by properly tightens all straps Do not tighten them on the base bed frame but on the moveable frame Manual Always read this manual before using this product 3 Installation Step 1 Place the mattress flatly on a bed frame The inflation tubes should be towards the foot end so that they can be connected to the inflation nozzles on the pump Step 2 Hang the pump over the frame or board at the foot end of the bed Make sure the pump is secured to the board Step 3 gt Connect the inflation tubes from the mattress to the pump s inflating nozzles Make sure they are properly attached CAUTION Make sure the air hoses are not kinked or tucked under the mattress Also check to ensure the CPR valve is properly attached Step 4 Plug the power cord into an electrical outlet with grounded AC power If the power cord has a grounding wire with a grounding pin then this three wire plug must be plugged properly into an outlet and grounded as shown in the figure in the Grounding Instructions section NOTE Before inserting the plug into the outlet make sure the voltage is compatible 4 CAUTION Make sur
5. e this product is well grounded Step 5 Turn on the power by pressing the power switch at the faceplate of the pump Proceed to the Operation section Step 6 Make sure to disconnect the pump by unplugging the power cord when it is not in use Grounding Instructions This product should be grounded when it is equipped with a cord that contains a grounding wire with a grounding plug In the event of an electrical short circuit grounding reduces the risk of electric shock by providing an escape wire for the electric current DANGER Improper use of the grounding plug can result in a risk of electric shock If repair or replacement of the cord or plug is necessary do not connect the grounding to either flat blade terminal The wire with insulation having an outer surface that is green with or without yellow stripe is the grounding wire Check with a qualified electrician or serviceman if the grounding instructions are not completely understood or if in doubt as to whether the product is properly grounded If it is necessary to use an extension cord use only a 3 wire extension cord that has a three prong grounding plug and a 3 slot receptacle that will accept the plug on the product Replace or repair if damaged Polarized Plug If the product is provided with a 2 wire polarized attachment plug This appliance has a polarized plug one blade is wider than the other To reduce the risk of electric shock this plug is inte
6. ean top and bottom both sides gt Spray cloth with cleaning solution and clean faceplate DO NOT allow excess cleaning solution on faceplate or control panel If solution gets inside damage will occur Allow surface to thoroughly dry after cleaning gt After the pump is thoroughly cleaned and dried proceed to plug in the pump and test to see if it runs normally gt Unplug the pump and store with proper identification tag Avoid long exposure to sunlight Mattress gt Brush off or wipe down all surfaces of the cover sheet with soap and water before wetting with any liquid disinfectant gt Any obvious blood spots should be wet thoroughly with 1 9 Hypochlorite solution 1 part bleach to 9 parts water and allow drying for at least 10 minutes Then blot with a clean damp cloth gt Unzip the top cover from the mattress gt Brush or wipe down all surfaces with soap and water before applying any liquid gt Covers are immersed and soaked in disinfectant for the required incubation time gt After pre soaking the cover is rinsed through a regular cycle in a washer with no soap then laundered with mild detergent wash temperature 93 F rinse temperature 78 F or on the coldest setting gt Covers are aerated until they are fully dry Drying temperature range 90 120 F or on the coldest setting gt The air cells are unsnapped from one side and are sprayed on all sides with a disinfectant Let it sit for the requir
7. ed incubation time and wipe down with a clean cloth Make sure to disconnect all the air cells one by one and spray the disinfectant on all sides including all the connecting tubes and hoses Let it sit for least 10 minutes gt If there is a base after you remove all the air cells the base has to be sprayed down with the disinfectant inside and outside Let it sit for the required incubation time and wipe down with a cloth gt Repeat the process with the tubing set spray incubate and then wipe clean gt The optional carrying bag should be turned inside out and completely wiped down using the disinfectant solution Allow it to thoroughly air dry Once the inside is dry turn it back wipe down the outside of the bag with disinfectant gt Dry the mattress on a SUNLESS area after cleaning 7 Handling and Storage gt Lay the mattress out flat and upside down gt Roll from the foot end towards the head end the foot end strap can then be stretched around the rolled mattress to prevent unrolling gt Do not fold crease or stack the mattress gt Follow the national requirement to dispose the pump Maintenance General gt Check the power cord and plug to see if there are abrasions or excessive wear gt Check the mattress cover for signs of wear or damage Ensure the mattress cover and tubes are connected together correctly gt Plug in the pump and check the airflow from the hose connection port The airflow sh
8. nded to fit into a polarized outlet only one way If the plug does not fit fully into the outlet reverse the plug If it still does not fit contact qualified electrician Do not attempt to modify the plug in any way Operation step by step Always read the operating instructions before use AP8000N Step 1 Turn on the power The green indicator of the power switch will come on The pump starts to pump air into the mattress e Step 2 For time saving in the first inflation turn the pressure knob to the full Firm position NOTE It takes a while for the mattress to be fully inflated for the first time You may want to consider using a quick inflation device for the first inflation Step 3 Turn the pressure adjust knob to set a comfortable pressure level from soft to firm according to patient s weight and comfort Pressure set up It is recommended that the pressure adjust knob being turned to Firm when the mattress is initially inflated Users can then easily adjust the pressure level of the mattress to a desired firmness by themselves or according to the suggestion from a health care professional If the pressure is consistently low check for any leakage occurring in inflation tubes Contact your local dealer for replacement of any damaged tubes Troubleshooting Problems Solutions 1 Check if the plug is inserted firmly into the outlet The pump doesn t work y 2 Turn on the power switch again
9. ocked Keep the air openings free of lint hair and the like gt Never drop or insert any object into any opening or hose gt Do not use outdoors or operate where aerosol spray products are being used or in an oxygen enriched environment gt Connect this product to a properly grounded outlet only See Grounding Instructions Introduction This manual should be used for the initial set up of the system and for reference purposes General The AP8000N is a high quality and affordable air support surface system suitable for pressure ulcer prevention and treatment It has been specifically designed for prevention and treatment of pressure ulcers and offers an effective solution to 24 hour pressure area care The AP8000N has been tested and certified to the following standards gt ANSI AAMI ES60601 1 gt CAN CSA C22 2 NO 60601 1 EN 60601 1 EN 60601 1 2 IEC 60601 1 11 Intended Use The AP8000N is intended to reduce the incidence of pressure ulcers while optimizing patient comfort It also is appropriate for the following gt Individual home care setting and long term care facilities Pain management as prescribed by a physician Contraindication Patient conditions for which the application of pressure relieving therapy on an alternating system is contraindicated are as follows gt Cervical or skeletal traction gt Unstable spinal cord injuries Product Description Pump gt Power Switch 1 Th
10. ould alternate between ports every half cycle time gt Check the air hoses to see if there are kinks or breaks For replacement please contact your local agent or dealer gt Make sure the mattress tube is well connected gt Check the pump and make sure both power and power indicator are off when the switch is turned off Low pressure Examine if there is air leakage between the pump and the mattress connections or from the air mattress tubes gt Check connectors between the air mattress and pump If there is any disconnection please reconnect it gt Check the CPR Valves Ensure their outlets are sealed gt Check the air connecting tubes Ensure each single cell is not broken gt Turn the pressure adjust knob to Firm Keep the tubes fully inflated and inspect for air leakage gt Check if there is any air leakage from cells Ensure no leakage occurs If any leakage occurs please contact your local agent or dealer Specifications Pump Pump Specification Power Supply Class AP APG NO Type BF Air output gt 5 liter min Pressure range 30 mmHg 80 mmHg Cycle time 10 min 60Hz or 12 min 50Hz Power consumption AC 120V 60Hz or 230V 50Hz 12W 1A Fuse Rating 250V T A Size 28 0 x 13 0 x 8 5 cm 11 0 x 5 1 x 3 3 in Weight Pump Q1 02 1 1 kg 2 4 lbs Environment Requirements gt Temperature Operation 41 F 104 F 5 40 C Storage 13 F 158 F 25