Home

ICS669-check - Mettler Toledo

image

Contents

1. 4 2 1 Overview Level 1 Level 2 Level 3 Calibration Display Unit 1 g kg oz lb lb oz t Units Unit 2 g kg oz lb lb oz t Resolution Unit roll On Off Zero AZM Off 0 5 d 1d 2d 5d 10d Tare Auto tare On Off Chain tare On Off A Clear On Off 9 d tare Restart On Off Filter Vibration Low Medium High Process Universal Dosing Stability Fast Standard Precise MinWeigh Function On Off Reset Perform reset 46 METTLER TOLEDO User manual ICS649 check Order number 22021156A 12 10 4 2 2 Description of the analog Scale menu block Analog Scale Calibration This menu item is not available for verified scales Perform calibration 1 Unload scale 2 Start calibration with E gt The scale determines the zero point 0 appears in the display The calibration weight to be placed on the scale flashes in the display 3 If necessary change the weight value displayed with gt T lt 4 Place the calibration weight on the scale and confirm with La The scale calibrates with the calibration weight loaded After calibration is completed Done appears briefly in the display and the calibration protocol is displayed Calibration passed COMMENT 5 You can now enter User name Weight name and Comments if applicable Note e n order to achieve particularly high precision calibrate under full load The calibra
2. Settings in kg g Settings in Ib Model Range Readability Range Readability A3 1 5 kg 3 kg 0 5g 1lg 2 5 lb 5 Ib 0 0005 Ib 0 001 Ib AG QAG 3 kg 6 kg 1g 2g 5 Ib 10 Ib 0 001 Ib 0 002 Ib A15 QB15 6 kg 15 kg 2g bg 10 Ib 25 Ib 0 002 Ib 0 005 Ib BB30 B30 QB30 15 kg 30 kg 5g 10g 25 Ib 50 Ib 0 005 Ib 0 01 Ib BB60 B60 BC60 CC60 QB60 QC60 30 kg 60 kg 10g 20g 50 Ib 100 Ib 0 01 Ib 0 02 Ib BC150 B150 CC150 QC150 60kg 150kg 20g 50g 100 Ib 250 Ib 0 02 Ib 0 05 Ib BC300 CC300 150kg 300kg 50g 100g 250 lb 500 Ib 0 05 Ib 0 1 Ib CC600 300 kg 600 kg 100g 200g 5001b 10001lb 0 1 Ib 0 2 Ib 11 10 Operation limits maximum static safe load Order number 22021161A Model a center load b side load c corner load A 30 kg 60 Ib 20 kg 40 Ib 10 kg 20 Ib BB 100 kg 200 Ib 70 kg 140 Ib 35 kg 70 Ib B 200 kg 400 Ib 140 kg 280 Ib 75 kg 150 Ib BC 400 kg 800 Ib 300 kg 600 Ib 150 kg 300 Ib cc 700 kg 1400 Ib 400 kg 800 Ib 200 kg 400 Ib QA 15 kg 30 Ib 10 kg 20 Ib 5 kg 10 Ib op 100 kg 200 Ib 70 kg 140 Ib 35 kg 70 Ib ac 200 kg 400 Ib 140 kg 280 Ib 75 kg 150 Ib METTLER TOLEDO User manual ICS669 check 85 86 METTLER TOLEDO User manual ICS669 check Weights approx values Model Weight in kg Weight in Ib A 5 2 11 5 BB 7 4 16 3 B 12 7 28 0 BC
3. Order number 22021161A 12 10 1 2 4 Default layout Weight values in 3 line mode 12 10 Order number 22021161A Display If you specify a target weight for Over Under Checkweighing the display automatically switches from the straight weighing display to the Over Under Checkweighing display To meet your special requirements three different display layouts are selectable in the terminal menu Straight weighing display 2 3 110 Date 85 16 2618 Time 13 44 32 Gemen 10 Dev Id AAATI2 Te s UserName SUPERVISOR Metrological data for details see below Weight value with star sign and stability monitor for details see below Spanner icon service needed for details see Event and error messages Battery symbol Net Gross Unit Calender can be activated in the menu Navigation hint use cursor keys gt for scrolling soft key pages Soft keys factory setting page 1 10 Auxiliary data 3 lines can be defined in the menu 11 Symbol and info line for details see below o d OO OO P Ga MM ca e 116 METTLER TOLEDO User manual ICS669 check 9 10 Default layout 3 line mode Colour mode METTLER TOLEDO User manual ICS669 check Over Under Checkweighing display In Over Under Checkweighing the display uses colours fo indicate the Over Under Checkweighing status LOW 1 64kg HIGH 2 38kg Instead of the weight display a bargraph is
4. Checkweighing pieces Posi tion Tolerance Checkweighing result METTLER TOLEDO Tel 49 7431 140 Header Germany Dote 087012010 Tine 00 02 53 Date amp time 102 Identifications Gross 2 090 kg Weight Target 90 PCS Tol 1 PCS Tol 1 PCS Checkweighing parameters Tol Tupe mo Piece counting Date 08 01 2010 Tine 00106131 1 Date amp time ae city 0 700 K9 Net weight quantity APH 23 96766 g Average piece weight METTLER TOLEDO User manual ICS669 check 95 8 6 Index A Accessories sssssssss 89 Alibi log sssse M 76 Average weighing 27 B Barcode Destination 67 Identification 28 Battery Handling 19 Specification 82 Deene 61 C Calendar 60 Calibration og 18 Cleaning eese 30 Commissioning 18 CONNECTIONS sisisi 16 D po 60 Display Backlighting 59 Brightness 59 Coniroet eeeceeeeeeeeees 59 Metrological data 11 Over Under Checkweighing 9 10 Presentation in the menu 44 Symbols and info line 12 Weight disploy 11 Disposal 94 Dynamic weighing 21 E Environment 30 82 ENOS non
5. HETTLER TOLEDO Tel 49 7431 140 Germany HHH DL Copn Dote 27 04 2010 Time 22 21 14 Net 0 37 ko Tore 0 82 ko Dev Id 4591 22 A Dev Loc Building B9 Header 4 New line 1 Date amp time 1 Net tare weights New line Device information 4 New line Checkweighing standard printout Posi tion Good METTLER TOLEDO Tel 49 7431 140 Germany ron ot com Date 07 01 2010 Tine 23154156 101 102 67195 Town Net 0 710 kg Tare 0 025 kg Gross 0 735 kg Target 0 700 kg Tol 0 050 kg Tol 0 090 kg Tol Type Relative Dev 0 010 kg Checkweighing result 1 Header Date amp time Identifications 1 Net tare gross weights Checkweighing parameters Checkweighing minimum printout Checkweighing result 1 Net weight Posi tion gt Tolerance Net 0 925 kg 12 10 Order number 22021161A Average weighing 3900000000000000000000060 Gross 1 19 kg Netfiverage 0 37 kg Tare 0 82 kg Starline Gross net tare weights 1 New line Printout with header and identification data HETTLER TOLEDO Tel 49 7431 140 Germany HHH At com Date 27 04 2010 Tine 21 50 48 101 Company ABC 102 67195 Tonn Net 0 57 kg Tare 0 82 kg Gross 1 39 kg T Header T Date amp time Identifications Net tare gross weights New line
6. The scale checks the zero point 0 appears in the display 2 The test weight value flashes in the display If necessary change the weight value displayed using T 3 Putthe test weight on the scale and confirm with gt The scale checks the test weight 4 After the test is completed the deviation from the last calibration briefly appears in the display ideally d 0 0g after which the device changes to the next menu item Keyboard Test Keyboard test Start 1 Press LZ to start the keyboard test 2 Press the keys in the displayed order If the key works the device switches to the next key Display Test Display test Start 1 Press LZ to start the display test A checkerboard pattern is displayed in all colours 2 Press LZ to leave the display fest The display works properly if the dark and bright fields are displayed without missing pixels Serial number Display of the serial numbers Start 1 Press E gt The serial numbers of the weighing terminal and the active weighing platform are displayed 2 Press to leave the item Print Setup Printout of a list of all menu settings Start 1 Press E gt A safety prompt is displayed 2 Press gt again to start printing 12 10 Order number 22021156A METTLER TOLEDO User manual ICS649 check 71 Print Records Printout of a list of the target value records stored in the database Start 1 Press Ca A safety p
7. 33 Lord co 41 P Piece counting 27 Power oft 58 Power supply 19 82 UL DT 24 Printout Examples 95 Templates 70 a Quick Select Alibi log file 76 Enter main menu 73 LEOGOUE iiis cete tis 73 Routine test 74 75 76 R Routine test 55 74 75 17 S Safety instructions 5 31 Getpointe suse 68 Order number 22021161A 11 10 11 10 SICS Z Command set 62 Zero Identifications 28 Automatic zero update 48 50 Simple weighing 22 Zeroing Sleep mode eee 58 Zero point correction Smart weighing counter 81 SO KOY S 5 ets 14 Spanner icon 9 81 Switching on and off 21 T Tare Automatic clearing 23 48 Automatic taring 23 48 Chain tare 23 48 Tore preset 24 leit OH 22 Technical data Weighing platforms 85 Weighing terminal 82 Terminal and platform combinations 6 85 Test EI eil VE 71 Keyboard 0c eeeeeeeeeees 71 Spleen gege ege 71 Verification 32 UE 60 Totalising sssssee 25 U User management 17 V Verified instruments in EC COUNTIES ienas 92 W
8. Display at Over Under Checkweighing to zero With display type Colour mode and default colours the following is displayed for too light good too heavy ZA Ke2 40 3 7 Terminating Over Under Checkweighing Press C Cleared appears in the display The target values are cleared and the straight weighing display appears The device operates in straight weighing mode 42 METTLER TOLEDO User manual ICS669 check Order number 22021161A 12 10 d Settings in the menu In the menu settings can be changed and functions can be activated This enables adaptation to individual weighing requirements The menu consists of the following 5 main blocks containing various submenus on several levels Scale see section 4 2 analog scales or 4 3 IDNet scales Application see section 4 4 Terminal see section 4 5 Communication see section 4 6 Maintenance see section 4 7 4 1 Operating the menu 4 1 1 Calling up the menu and entering the password The menu differentiates between 2 operating levels Operator and Supervisor The Supervisor level can be protected by a password When the device is delivered both levels are accessible without a password Operator menu 1 Press E The Quick Select menu opens Menu is highlighted 2 Press EP Enter code is displayed 3 Press LZ again The menu item Terminal is displayed Only parts of the submenu Device are accessible Supervisor menu 1 Press E The Quick Select menu opens M
9. Minimum weighing in quantity Before you can use this function the METTLER TOLEDO service technician has to determine and to enter a minimum weight value Function Switching minimum weight function on off If the weight on the scale drops below the stored minimum weight B appears in the symbols and info line Analog Scale Reset Resetting scale settings to factory settings Perform reset Confirmation inquiry e Reset the analog scale settings to factory settings with YES Do not reset the analog scale settings with NO 4 3 Scale menu block IDNet scales Factory settings are printed in bold in the following overview 4 3 1 Overview Level 1 Level 2 Level 3 Display Unit 2 g kg oz lb t Units Unit roll On Off Zero AZM On Off Tare Auto tare On Off Chain tare On Off A Clear On Off 9 d tare Restart On Off Filter Vibration Stable Normal Unstable Process Finefill Universal Absolut Stability ASD 0 ASD 1 ASDz2 ASD 3 ASD 4 Update The possible settings depend on the connected scale MinWeigh Function On Off Reset Perform reset 12 10 Order number 22021156A METTLER TOLEDO User manual ICS649 check 49 4 3 2 Description of the IDNet Scale menu block IDNet Scale Display Weighing unit Unit 2 Select weighing unit 2 g kg oz Ib f Unit roll When unit roll is switched on the w
10. p Me CAD T 3 882 Over Under Checkweighing during subtractive weighing Assistance through the coloured background and the graphical weighing aid is also possible during subtractive weighing and subtractive counting Procedure 1 Specify target values as described above The target value is indicated with a negative sign Place a full container on the weighing platform and tare it Remove as much from the weighing sample as required for the display to change to the state good factory setting green Tare the unit again The scale is ready for the next removal GE B Order number 22021161A 12 10 Display for subtractive Over Under Checkweighing With default settings for display type and colours the following is displayed for foo light UB good too heavy T 6 15kg p 1 B8kg T B 18kg AE T 8 887 1 B8kg T 5 887 B les T 8PCS p 5BPCS T SPCS 3 4 Over Under Checkweighing with Quick start If default values for the tolerances are used with tolerance types Relative or Percent Over Under Checkweighing can be started by pressing just one key Prerequisites Y The setting On is selected in the menu under Application Over under gt Default Values Activate v Tolerance values are defined under Application gt Over under gt Default Values gt Relative weight Percent weight Relative pieces v The selected tolerance type matches the entered
11. Input via barcode is recognised as target value Average PW Input via barcode is recognised as average piece weight Recall Input via barcode is recognised as record number Communication COM x Parameter Communication parameters Baud Selecting baud rate Possible settings 300 600 1200 2400 4800 9600 19200 38400 57600 115200 Parity Selecting parity Possible settings 7 none 8 none 7 odd 8 odd 7 even 8 even Handshake Selecting handshake Possible settings Off Xon Xoff RS Type Selecting type of the optional RS422 RS485 interface RS422 or RS485 Net Address Assigning network address O 31 only for RS485 Checksum Activating deactivating checksum byte STX Activating deactivating STX If STX is enabled the STX signal 0x02 is sent at the beginning of each output string that is sent via the interface Print G This functionality can only be enabled if one of the Demand mode templates is selected If it is enabled the gross weight is marked with G Examples Print G enabled no fare _ _2 001_kg_G Print G disabled no tare _ _2 001_kg Print G enabled tare active 2 025_kg_G 2 000 kg T 0 025 kg NET Print G disabled tare active 2 025 kg 2 000 kg T 0 025 kg NET Load resistor Only for the optional RS422 RS485 interface To avoid reflexions on a network we recommend to make a defined termination To this purpose the load resistor within
12. WAININGS rossian 79 Order number 22021161A METTLER TOLEDO User manual ICS669 check 97 To protect your METTLER TOLEDO product s future METTLER TOLEDO Service XXL assures ihe quality measuring accuracy and preservation of value of all METTLER TOLEDO products for years to come Please send for full details about our attractive terms of service Thank you www mt com service For more information Mettler geg Ee dt GmbH D 72458 A Tel da ee 140 Fax 49 7431 14 232 Subject to technical changes al eee Toledo Albstadt GmbH 22021161A Printed in Germany Order number 22021161A
13. 30 meter TOLEDO User manual ICS669 check Order number 22021161A 12 10 12 10 2 12 2 General cleaning recommendations Risk of electric shock A A Before cleaning unplug the power plug in order to disconnect the terminal from the power supply Cover open connectors with cap plugs Clean the protective cover separately The protective hood is dishwasher safe Replace the protective hood regularly Take off the load plate and remove any dirt and foreign substances which may have collected underneath Do not use any hard objects to do so Do not disassemble the weighing device Remove any possibly remaining detergent by rinsing with clear water To prolong the lifetime of the load cell dry it with a soft lint free cloth immediately after cleaning Observe all the existing regulations on cleaning intervals and permissible cleaning agents Cleaning of different weighing platforms as described in this User manual gt Order number 22021161A Make sure fo observe the cleaning instructions for the connected weighing platform The weighing platform may not be designed for wet environments and the cleaning procedures described above METTLER TOLEDO User manual ICS669 check 3 32 2 13 METTLER TOLEDO User manual ICS669 check Verification test The weighing instrument is verified if the accuracy class is displayed in the metrological line the securing seal is not tampered with it bears an official verification
14. 18 diagonal 34 mm 1 34 METTLER TOLEDO User manual ICS669 check 87 ICS669a check f Dim mm mm mm mm mm mm mm mm d 418 16 46 485 19 09 581 22 87 681 26 81 772 30 39 407 16 02 489 19225 640 25 10 ICS669a check c i b E a BB BC cc QA QB ac Dim mm mm mm mm mm mm mm mm d 337 13 27 404 15 91 500 19 69 600 23 62 691 27 2 326 12 83 408 16 06 559 22 01 b 412 mm 16 22 c 34 mm 1 34 88 METTLER TOLEDO User manual ICS669 check Order number 22021161A 11 10 11 10 7 3 Accessories Printers Order no GA46 printer RS232 incl 8 pin M12 plug 2 5 m cable 22 019 925 0 4 m cable 22 019 926 Retrofitable interfaces conversion kits Order no Kit must be fitted by a METTLER TOLEDO service technician RS232 conversion kit terminal version 22012 112 C Version 22012117 RS422 RS485 conversion kit terminal version 22012 113 C Version 22012 118 Ethernet conversion kit terminal version 22012114 C Version 22012119 USB Device conversion kit terminal version 22012115 C Version 22012 120 Digital UO conver
15. 24 26 23 54 29 25 23 54 41 59 32 21 26 45 31 56 24 57 04 62 09 15 29 25 34 21 25 26 59 32 64 55 18 31 56 36 41 17 19 62 09 67 57 19 34 21 38 58 20 64 55 71 21 18 36 41 41 12 15 67 57 75 24 15 38 58 43 26 19 71 21 80 56 24 26 41 12 45 38 26 75 24 90 00 18 factory setting Disposal In conformance with the European Directive 2002 96 EC on Waste Electrical and Electronic Equipment WEEE this device may not be disposed of with domestic waste This also applies to countries outside the EU according to their specific requirements Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment If you have any questions please contact the responsible authority or the distributor from which you purchased this device Should this device be passed on to other parties for private or professional use the content of this regulation must also be related Thank you for your contribution to environmental protection Order number 22021161A 12 10 8 5 Straight weighing Protocol printouts GA46 printouts in English 300000000000000000000000 Gross 1 19 kg Net 0 37 kg Tore 0 82 kg Starline Gross net tare weights 1 New line Printout with header
16. Cancel editing e Exit menu C Clear Clear tare Clear value clear digit e Leave info page EUM Switch e Switch over weight unit Re edit A e Switch over from numerics to upper lower case letters gt 0 lt Zero e Set scale to zero clear tare gt T lt Tare e Tare scale clear tare Quick select e Opening Quick Select menu e g for menu access viewing log files or logout 4 Info e Activate info screen I e Proceed to next info line info page e Freeze and release startup screen E 5 Transfer e Transfer data to a printer or computer e Enter menu item e Confirm entry selection eL e VY Cursor keys e Navigating e Navigating 12 10 Order number 22021161A METTLER TOLEDO User manual ICS669 check 13 14 1 2 6 Page 1 Page 2 Page 3 METTLER TOLEDO User manual ICS669 check Soft keys To meet your specific application requirements ICS669 check offers 15 soft keys which can be configured in the terminal menu The soft keys are divided into three lines pages Default setting gt gt Set target gt Load target ID1 Store target Identification 1 Reference n Operating soft keys gt Press the key below the desired function Changing soft key line gt Press the cursor keys lt or gt to switch from line to line Possible soft key settings Average piece weight Symbol Menu setting Function
17. User access Backlight User access Weight hold Stealth mode Colours Colour mode Configuring the display window For details see Introduction Selecting the presentation of the weight value Possible settings Default 3 Line mode Color mode Selecting the contents of the auxiliary display lines 1 Select auxiliary line number 2 Assign contents to the selected line Setting the contrast of the display Possible settings 1 10 Setting the brightness of the display Possible settings 1 10 Setting whether and after which time the background lighting will be switched off Devices with a storage battery switch the background lighting off automatically by default when no action takes place at the device for approx b seconds Possible settings Off no background lighting 5 sec 10 sec 30 sec 1 min On background lighting always on approximate values Setting how long the weighing result is frozen in the display after the transfer key gt has been pressed or auto print was generated Possible settings O s 10 s With stealth mode on there is no weight display only the coloured display for too light good and too heavy Not available if the scale is approved Configuring colour setting for default MinWeigh and Over Under Checkweighing The device offers 10 predefined colours which can be assigned to the different conditions 1 Select status 2 Assign
18. colour to the selected status Setting the way of changing display colour e Continuous Change in colour with stable or dynamic weight values e Stable Change in colour only when the weight value is stable For default colour settings see below IDs Giving a name to the IDs D1 ID2 ID3 To each ID the following can be assigned e Name Enter a description of the ID for printout datatransfer e g Lot e Value Enter a default value for the ID e g 123 12 10 Order number 22021156A METTLER TOLEDO User manual ICS649 check 59 60 Default colour settings Condition Default Optional MinWeigh Dark blue Tolerance Red Purple lt SP Tolerance Light blue Orange Good range Green Light green Pink gt Tolerance Yellow Default colour White Keyboard Switching keys on off and setting info key Key lock Selecting keys to lock unlock Possible keys Power Clear C Unit switch KC Info C Transfer E gt Quick Select E keyboard Info key Configuring up to 27 items to be displayed using the info key 1 1 Select the item to be configured Item 1 Item 27 2 Assign contents Soft keys Configuring the soft keys Unless a soft key is configured there is no Over Under Checkweighing or counting functionality The soft keys are organised in two pages lines of 5 function keys each 1 Select the page to be configured Page 1 Page 2 Page 3 2 Selec
19. displayed 2 Press OK The menu changes are saved and the terminal returns to the weighing mode or gt Press ESC for further menu settings or gt Press NO to discard changes and return to the weighing mode Order number 22021156A 12 10 Application Terminal Communication Maintenance Calibration Display Units Zero Selecting and setting parameters in the menu Example Setting of the Chain tare function Calibration Display Units Zero Tare Restart Filter Minlleigh Reset Auto tare Chain tare Clear tare Scale Tare Auto tare Chain tare amp Clear tare 12 10 Order number 22021156A In the menu start screen press to switch to the right side The first submenu Calibration is highlighted Use v to select Tare The current Tare settings are displayed on the right side Press gt to open the selected highlighted menu item Tare The Tare submenus are displayed on the left side Use v a fo select Chain tare The current Chain tare setting is displayed on the right side Press to open the selected highlighted menu item Chain tare All possible Chain tare settings are displayed on the right side the current setting is highlighted Use v a to change the Chain tare setting Confirm the setting with LP METTLER TOLEDO User manual ICS649 check 45 4 2 Scale menu block analog scales Factory settings are printed in bold in the following overview
20. e Off No automatic clearing of the tare weight e 9d The tare weight is automatically cleared if the gross weight is within 9 display steps Analog Scale Restart Automatic saving of zero point and tare value Restart When the restart function is activated the last zero point and the tare value are saved After switching off on or after a power interruption the device continues to work with the saved zero point and tare value Analog Scale Filter Adaptation of the ambient conditions and the weighing type Vibration Adaptation to ambient conditions Low e Very steady and stable environment The scale works very rapidly but is very sensitive to external influences Medium e Normal environment The scale operates at medium speed High e Restless environment The scale works more slowly but is insensitive to external influences Process Adaptation to the weighing process Universal e Universal setting for all weighing samples and normal weighing goods Dosing e Dispensing liquid or powdery weighing samples Stability Adjusting the stability detector Fast e The scale operates very fast Standard e The scale operates at medium speed Precise e The scale operates with the greatest possible reproducibility The slower the scale works the greater the reproducibility of the weighing results METTLER TOLEDO User manual ICS649 check Order number 22021156A 12 10 Analog Scale gt MinWeigh
21. fixed to 0 5 d As standard the zero point of the scale is automatically corrected when the scale is unloaded e The zero function is only available within a limited weighing range e After zeroing the scale the whole weighing range is still available e A successful Zeroing will always delete a tare weight Simple weighing 1 Place weighing sample on the scale 2 Wait until the stability monitor O goes out 3 Read the weighing result Weighing with tare Taring gt Place the empty container on the scale and press gt T lt The zero display and the symbol NET appear The tare weight remains stored until it is cleared Clearing the tare Press C The symbol NET goes out the gross weight appears in the display If the symbol is displayed i e the A Clear Tare function is activated in the menu under Scale gt Tare the tare weight is automatically cleared as soon as the scale is unloaded Order number 22021161A 12 10 12 10 2 4 3 Automatic clearing of the tare A tare weight is automatically cleared when the scale is unloaded Prerequisite v The symbol is displayed i e the tare function A Clear Tare is activated in the menu under Scale gt Tare 2 4 4 Automatic taring If you place a weight on an empty scale the scale tares automatically and the symbol NET is displayed Prerequisite v The symbol TJ is displayed i e the tare function Auto Tare is activated in the menu under Scale gt
22. gt Target Enter Over Under Checkweighing parameters Recall Recall Over Under Checkweighing parameters out of the database Store Store Over Under Checkweighing parameters in the database TA Take away Activate Deactivate Take away mode lo To zero Over Under Checkweighing to zero ID1 ID2 ID3 DI ID2 ID3 Enter identification 1 identification 2 identification 3 Reference n Determine the average piece weight Average PW Enter the average piece weight Switch scale Switch over weighing platform Order number 22021161A 12 10 12 10 1 2 7 Order number 22021161A Alphanumeric input When an alphanumeric input is requested one of the following symbols is displayed in the right top edge of the display 1239 for numeric input and special characters ABC for input in upper case letters abc for input in lower case letters 2 To switch between numerics and upper lower case letters press SA Text entries work like e g on a mobile phone Up to four characters are assigned to the keys of the numeric keyboard Entries must be confirmed with LP Example Enter ICS6x9 Oo 5M 0 OT de CO Paz Make sure that ABC amp is displayed To enter letter I press key 4 three times To enter letter C press key 2 three times To enter letter S press key 7 four times Press KCA twice to change to numerics 123 Enter number 6 Press A twice to change to lower c
23. mark e g the green M sticker OIML the validity is not expired The weighing instrument is also verified if the metrological line shows Approved scale labels with the metrological data are placed near the weight display the securing seal is not tampered with it bears an official verification mark e g the green M sticker OIML the validity is not expired The period of validity is country specific It is in the responsibility of the owner to renew verification in due time Terminal and platform combinations Combinations of a weighing terminal and an analog weighing platform use a Geo Code to compensate for gravitational influence The manufacturer of the weighing instrument uses a defined Geo Code value for verification Please check if the Geo Code in the instrument corresponds with the Geo Code value defined for your location The Geo Code value is displayed when you switch on the instrument The Geo Code for your location is shown in the Appendix Call the METTLER TOLEDO service technician if the Geo Code values do not match Order number 22021161A 12 10 12 10 Over Under Checkweighing The device offers Over Under Checkweighing functions The respective settings in the menu are described in the application menu section The corresponding coloured background lighting allows rapid detection of the states too light factory setting red good factory setting green and too heavy factory setting yell
24. must be carried out manually To leave Take away mode press soft key TA again The symbol TA disappears You can select the Good print feature to generate an automatic printout when the weight is within tolerance You can select Auto tare and Auto clear tare to continue working after the full or empty container needs to be replaced without touching one key Order number 22021161A 12 10 12 10 3 6 Over Under Checkweighing to zero The weight value or the number of pieces can also be represented as the difference to the target weight Prerequisites v For Over Under Checkweighing to zero tolerance types Relative or Percent are selected For Over Under Checkcounting to zero tolerance type Relative is selected v The soft key lo to zero is activated v Display layout Colour mode or 3 line mode selected Procedure 1 Specify the target values as described above 2 Press the soft key lo The target is displayed with a minus sign 3 Place the checkweighing material on the scale Depending on the applied weight or the applied amount the colour of the background lighting changes The display value is displayed in accordance with the tolerance type setting The target value is O kg or PCS or 0 00 96 Terminating Over Under Checkweighing to zero gt Press soft key lo again The symbol 0 in the info line disappears the net weight is displayed Order number 22021161A METTLER TOLEDO User manual ICS669 check 41
25. printout for totalising Possible settings Off Template 1 Template 5 Total Template Selecting protocol printout for printing the total Possible settings Off Template 1 Template 5 ASCII Format Line format Line length Separator Expanded Add line feed Selecting formats for the protocol printout Selecting line format from the following e Multiple multiple lines e Single single line Fixed Records output in single lines every record includes the number of characters that was defined under Line length Setting line length Possible settings O to 100 characters Factory setting 24 characters This item is only displayed for the line formats Multiple and Fixed Selecting the separator Possible settings X and space This item is only displayed for the line format Single Printout with bigger font size on METTLER TOLEDO printers Adding line feeds Possible settings 0 9 66 uerreR TOLEDO User manual ICS649 check Order number 22021156A 12 10 Communication gt COM x Destination Destination for barcode input None Input destination is not predefined The input will be shown on the display you can decide what to do with the input Tare preset Input via barcode is recognised as tare preset ID1 ID2 ID3 Input via barcode is recognised as ID1 ID2 resp ID3 User ID Input via barcode is recognised as User ID Target
26. 22 RS485 menu blocks COM 1 COM 2 COM 3 Level 1 Level 2 Level 3 Level 4 Mode Print Auto print Instant print Continuous Dialog External input Demand mode Demand m auto Cont Weight Cont count Printer Type ASCII printer Label printer GA46 printer Print Standard Template Template 5 Template Plus Off Template Template 5 Template Total Off Template Template 5 Template ACII Format Line format Multiple Single Fixed Line length 1 24 100 Separator E oj _ Space Expanded On Off Add line 0 9 feed Destination Off Tare preset ID1 ID2 Target Average PW Recall Parameter Baud 300 600 57600 115200 baud Parity 7 none 8 none 7 odd 8 odd 7 even 8 even Handshake Off Xon Xoff RS Type RS422 RS485 Net Address 0 31 Checksum On Off STX On Off Print G On Off Load On Off resistor Reset Perform Reset METTLER TOLEDO User manual ICS649 check Order number 22021156A 12 10 4 6 3 Description of the RS232 RS422 RS485 menu blocks COM 1 COM 2 Communication gt COM x gt Mode Operating mode of the serial interface Print Manual data output to the printer with gt Auto print Instant print Automatic output of stable results fo the printer e g for series weighing operations Immediate manual data output to the printer with gt not verifiabl
27. 22021156A 12 10 COM x Digital 1 0 Output Mode Behaviour of the digital outputs Continuous Digital outputs are updated continuously Stable Digital ouputs are updated only when the weight is stable COM x Digital 1 0 Reset COM 3 Resetting communication settings to factory settings Perform reset Confirmation inquiry e Reset the Digital I O settings to factory settings with YES e Do not reset Digital I O settings with NO 12 10 4 6 5 4 6 6 Ethernet menu block IP address Hem Reference Mode Printer See RS232 RS422 RS485 menu blocks Destination Parameter DHCP If DHCP is set to On the device will receive the IP address automatically Then IP address Subnet mask and Gateway are read only fields Enter display IP address Subnet mask Enter display Subnet address Gateway Enter display Gateway address Reset See RS232 RS422 RS485 menu blocks Ethernet USB menu block Item Reference Mode Destination Checksum See RS232 RS422 RS485 menu blocks STX Reset USB Order number 22021156A METTLER TOLEDO User manual ICS649 check 69 70 4 6 7 WLAN menu block Level 1 Level 2 Level 3 Mode Printer See RS232 RS422 R8485 menu blocks Destination Parameter IP address Enter IP address Subnet mask
28. 26 5 58 4 cc 35 0 71 2 QA 4 1 9 0 QB 7 8 17 2 ac 13 1 28 9 Length of load cell cable for ICS669a check t Weighing range Length in m Length in ft up to 30 kg 50 Ib 1 5 5 60 kg 100 Ib and higher 2 5 8 Order number 22021161A 11 10 Dimensional drawings The size of the weighing platform A BB B BC CC QB is indicated at the end of the product name e g ICS669a check A6 Weighing platform Front view Side v iew nd a gt b BB BC cc QA QB ac Dim mm mm mm mm mm Imm fmm mm a 175 6 89 235 9 25 335 13 19 435 17 13 503 19 80 170 6 69 233 917 392 15 43 b 240 9 45 300 11 81 400 15 75 500 19 69 600 23 62 229 9 02 305 12 01 457 17 99 c 56 2 20 57 224 57 224 70 276 79 3 11 56 2 20 57 2 24 60 2 36 min height d max height d 15 mm d 0 59 11 10 Order number 22021161A d 95 3 74 97 3 82 100 3 94 108 4 25 130 5 12 95 3 74 108 4 25 100 3 94 e 235 9 25 335 13 19 435 17 13 587 23 11 724 28 50 170 6 69 245 9 65 397 15 63 f 300 11 81 400 15 75 500 19 69 650 25 59 800 31 50 229 9 02 305 12 01 457 17 99 g 22 0 87 15 0 59 15 059 15 0 59 21 083 22 087 15 059 15 0 59 h Circle diameter 30 mm 1
29. 3 EN 8 2 8 3 8 4 8 5 8 6 METTLER TOLEDO User manual ICS669 check Quick Select menu Ouere 73 Entering main MEM oeseri aane codon edet eti e aba conta nddeaasbayimascandovndbaats 73 Keeper 73 Performing el 74 Calling up routine test log file m 75 Calling up alibi log filg 5 irr ierit rer ro n roro e trt e 76 Calling up calibration log file 77 Event and error messages eeeeeeeeeeeenennn enne nnne nnn nnn nnn nnn 78 Enor COMGITIONS ici egene seht te et et ta da riva eee un aa Rua 78 Errors and WOININGS scu pto b e er exer parete red es e tren Md YR RENE 79 Smart weighing counter Spanner icon 8l Technical data and ACCeSsoriesS s ssssssseussnnnnunnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn nn 82 Technical data weighing terminal sse ms 82 Technical data weighing platforms sssssss ereere 85 ACGeSSOTIOS E A T EE 89 LIT EE 92 Tests for utilisation in hygienically sensitive areas sss 92 Notice for verified instruments in EC countries erer 92 Tables of Geo Code volues mene 93 DIE e Ee T ET 94 Protocol priDtOUlS s c ac eoe hp el eee t ad Peas tte e e p nata 95 ly oM 96 Order number 22021161A 12 10 12 10 ull A A A A A A A A gt gt GP A Order number 22021161A Introduction Safety instructions General Do not use the device in a hazardous environment Special de
30. 6 hours If work is continued during the charging process the charging time is extended The storage battery is protected against overcharging The storage battery has a service life of approx 2 years or 500 to 1 000 charging discharging cycles The storage battery is also suitable for permanent mains operation CAUTION Danger of soiling because the charger for the storage battery is not protected to IP69K A A Do not charge the device in humid or dusty rooms After the storage battery has been charged close the cover cap of the charging socket at the device CAUTION No success in charging the storage battery due to low temperatures A Do not charge the battery if the battery temperature is below O C 32 F Charging is not possible in this temperature range A Do not operate the battery charger outside its temperature range of O C to 40 C Order number 22021161A 32 F to 104 F METTLER TOLEDO User manual ICS669 check 19 20 1 5 METTLER TOLEDO User manual ICS669 check Recommended use of the built in storage battery The characteristics mentioned above are only valid if the following recommendations are observed Connect the device to the battery charger as soon as the warning message Low battery appears and the battery symbol starts flashing When the message appeares you still have enough time at least 10 minutes to complete your current task Keep the battery charger connected until the cha
31. D Activate On Off Delete all Del Profile User No 12 10 Order number 22021156A METTLER TOLEDO User manual ICS649 check 57 Level 1 Level 2 Level 3 Level 4 Level 5 Level 6 Access Supervisor Password Reset Perform reset 4 5 2 Description of the Terminal menu block Terminal Device General device settings Language Selecting the language of the operator interface Possible languages English German French Spanish Italian Chinese We will expand the available languages continuously Sleep This menu item only appears on devices in mains operation User access When Sleep is activated the device switches off display and backlighting after the time period set when not in use and gross weight is O Display and backlighting are Switched on again by pressing a key or if the weight changes Possible settings Off 1 min 3 min 5 min 15 min 30 min approximate values Power Off This menu item only appears on devices in battery operation User access When Power Off is activated the device switches itself off automatically after the time period set when not in use After this it must be switched on again using Possible settings Off 1 min 3 min 5 min 15 min 30 min approximate values 58 merer TOLEDO User manual ICS649 check Order number 22021156A 12 10 Display Layout Auxiliary line 1 Auxiliary line 3 Contrast User access Brightness
32. Enter Subnet address Gateway Enter Gateway address Wireless SSID Enter SSID SettIngs Encryption Off WEP WPA WEP Settings 64 Bit 128 Bit WEP key Key 1 Key 4 WPA Settings WPA TKIP WPA2 AES Password Enter password Status Display the current status e g connection status signal strength Reset COM See RS232 RS422 RS485 menu blocks 4 6 8 Templates menu block Level 1 Level 2 Level 3 Template 1 Line 1 Not used Header Date Time Gross Net Tare gos i High resolution ID1 ID2 ID3 Dev identifier Template 5 Line 25 Dev location SNR Terminal SNR Scale 1 Target Tolerance Tolerance Tolerance type Descr field Deviation Weight position Average PW Reference count Quantity Record number Article name Article number Total net n User name Star line New line Form feed Configuring templates Select a template 2 Select the line to be configured 3 Assign the line contents The header can be specified via SICS command I31 see Reference manual MT SICS for ICS6xx METTLER TOLEDO User manual ICS649 check Order number 22021156A 12 10 4 7 Maintenance menu block Test Scale Testing the scale Scales with an analog interface will offer the test procedure described below Scales with an IDNet interface and an internal calibration weight perform an automatic calibration check
33. ICS669 check Weighing terminals Terminal and platform combinations METTLER 996 5 669 ICS669a check CS ICS669d check ee p Sei ICS669a check c CS669a check f S 1 XXL Congratulations on choosing the quality and precision of METTLER TOLEDO Proper use ervi ce according to these instructions and regular calibration and maintenance by our factory F trained service team ensure dependable and accurate operation to protect your investment Tailored Services Contact us about a ServiceXXL agreement tailored to your needs and budget We invite you fo register your product at www mt com productregistration so we can contact you about enhancements updates and important notifications concern ing your METTLER TOLEDO product 2 merrier TOLEDO User manual ICS669 check Order number 22021161A 12 10 Contents 1 tg HL BEE 5 11 RISIKEN 5 1 2 Nee 6 13 Tracing and tracking features essei rrer oi a mm 16 14 elle ll soc veo te Cete sa dete ag a Pisa t e T Rega Re data 18 15 Use in hygienically sensitive oe 20 2 lg m M 21 2 1 owitehing or and offo aeda e EEN 21 2 2 Zeroing Zero point Correction ceannann aaa n a GN 22 2 3 SIMPLE WEIGHING sei aiio uit te tao ee c rd eth itr idet Fe i ad P s 22 24 Weighing with tare emm eene 22 2 5 eine E de 24 2 6 Pribitinig Tes lls Aa Sterne a t RR e P e RE ERR EE P e HERR 24 2 7 VRAS RE 25 2 8 A
34. TOLEDO User manual ICS669 check Order number 22021161A 12 10 ICS669a check c 1 2 3 Options The following options are available for the ICS669 check Hygienic optimal version Weighing terminal and column are seamlessly welded together Easy to clean cables run inside the column ICS669a check ICS669d check ICS669a check f ICS669a check t ICS669a check c Built in storage battery v v Optional scale interface SCALE 2 e analog e digital Optional communication interface e RS232 e RS232 COM 2 e RS422 RS485 e RS422 RS485 e Ethernet e Ethernet e WLAN e WLAN e USB e Digital I O Optional communication interface e RS232 COM 3 e RS422 RS485 e USB e Digital 1 0 Desk mounting plate v 12 10 Order number 22021161A METTLER TOLEDO User manual ICS669 check 8 METTLER TOLEDO User manual ICS669 check Terminal and platform combinations Load cells for more challenging environments Standard hermetically sealed stainless steel load cell identical to PBA430 Option loadcell with KS coating identical to PBA430 with option loadcell with KS coating Other resolutions availability depending on region weighing unit and Weights and Measures approval Verification OIML Class Ill 1 x 3 000 e Verification OIML Class Ill 2 x 3 000 e MR 6 000 d non approvable 10 000 d non approvable 15 000 d non approvable 30 000 d non approvable
35. Tare The weight to be tared automatically e g packaging material must be heavier than 9 display steps of the scale 2 4 5 Chain tare With this function it is possible to tare several times if e g cardboard is placed between individual layers in a container Prerequisite Y The tare function Chain tare is activated in the menu under Scale Tare 1 Place the first container or packaging material on the scale and press T The packaging weight is automatically saved as the tare weight the zero display and the symbol NET appear 2 Weigh the weighing sample and read print out the result 3 Place the second container or packaging material on the scale and press gt T lt again The total weight on the scale is saved as the new tare weight The zero display appears Weigh the weighing sample in the second container and read print the result 5 Repeat steps 3 and 4 for other containers Order number 22021161A METTLER TOLEDO User manual ICS669 check 23 24 2 4 6 2 5 2 6 METTLER TOLEDO User manual ICS669 check Tare preset If you know the weight of your containers you can enter the tare weight numerically So you do not have to tare the empty container Enter the known tare weight and press gt T lt to confirm The weight display shows the negative tare weight and the symbol NET appears 2 Place the full container on the weighing platform The net weight is displayed The entered tare weight is valid until a new
36. al net and n are updated Unload the scale Repeat steps 3 and 4 for further items To finish totalising press the soft key x A safety prompt is displayed Press soffkey YES to clear the total o Press soffkey NO to continue totalising Piece counting results and OverUnder Checkweighing results can be totalised the same way Order number 22021161A 12 10 12 10 2 8 With manual start With automatic start 2 9 2 9 1 Order number 22021161A Average dynamic weighing With the average weighing function it is possible to weigh restless weighing samples such as living animals If this function is activated GC is displayed in the info line With average weighing the scale calculates the mean value from 56 weighing operations within 4 seconds Y Average Manual is selected in the menu v Weighing sample heavier than 9 scale divisions 1 Place the weighing sample on the scale 2 Press LZ to start average weighing 3 During average weighing stars appear in the display and the average result will be displayed with the symbol x 4 Unload the scale to be able to start a new average weighing operation Y Average Auto is selected in the menu v Weighing sample heavier than 9 scale divisions 1 Place the weighing sample on the scale Average weighing starts automatically During average weighing stars appear in the display and the average result will be displayed with the symbol x 2 Unload the sca
37. and Percent e The upper tolerance value has to be greater than the lower one High gt Low or respectively the target weight has to be greater than the lower tolerance value Tol gt Target gt Tol Order number 22021161A 12 10 3 1 3 Specifying target number of pieces Prerequisite n v At least one of the counting soft keys or is activated Determining the piece weight Apply the reference parts factory setting 10 pieces and press A The number of reference pieces is displayed e For alternate procedures to determine the piece weight refer to the counting section e f you use the unit PCS the tolerance type Percent ist not available Weighing in the target number of pieces gt Proceed as described in section Weighing in target values The display unit is PCS 3 1 4 Using stored target values Storing target records A traget record contains the following Tolerance type target weight and tolerances If available tare weight average piece weight and description field can be stored too Up to 500 target values can be stored 1 Enter the target values as described in the sections above 2 Press soft key gt The display changes to green to indicate a free record The record number of the first free memory and status Free are displayed 3 If necessary change the desired record number 4 Press soft key OK Record stored briefly appears in the display The target record is stored under the selec
38. arameter settings for the standard RS232 interface COM 1 COM 2 Parameter settings for the optional interface COM 2 COM 3 Parameter settings for the optional interface COM 3 not for c version The interfaces identify themselves Therefore only those menu settings appear which are relevant for the individual interface If no optional interface is installed the COM 2 and COM 3 menus will not appear Templates Define templates to be selected via COM x gt Printer gt Template Order number 22021156A 12 10 12 10 4 6 1 Standard settings COMI COM2 COM3 RS232 RS232 RS422 Ethernet USB WLAN RS485 Print v v v v v Auto print v v v v v Instant print v v v v v Continuous v v v v v v Dialog Factory Factory Factory Factory Factory Factory setting setting setting setting setting setting External input v v v v v v Demand mode v v v v v Demand m auto v v v v v 2 Cont weight Cont count Printer Destination Parameter Baud Parity Handshake RS Type Net Address Checksum STX Print G Load resistor D for more information see Reference manual MT SICS for ICS4xx 2 for more information see Reference manual MI Demand and Continuous not recommended for new installations Order number 22021156A METTLER TOLEDO User manual ICS649 check 63 64 4 6 2 Overview RS232 RS4
39. article number To search for the contents of the description field or for article names and numbers the entered data must be unique Application Count Setting reference Reference Presetting the number of reference pieces for counting Factory setting 1O pieces Note To activate the counting application you need to activate the corresponding soft keys too see Terminal menu block METTLER TOLEDO User manual ICS649 check Order number 22021156A 12 10 Application Log files Setting up logfiles Alibi Log Activate Item 7 Item 10 Delete all records Routine Test Log Days External test Internal test Delete all records Setting up Alibi log file If set to ON all weighings are saved in a log file You can add additional information to that required by W amp M regulations 1 Select item number 2 Assign the item contents A safety prompt is displayed before deleting all stored alibi records Setting up routine calibration test for your scale Enter interval of routine test e Test weight Enter the test weight value Weight name Enter weight name Tolerance Enter tolerance for passing the routine calibration test Yes for weighing platforms with an internal calibration weight e No for weighing platforms without an internal calibration weight A safety prompt is displayed before deleting all stored routine test records Applicatio
40. ase letters abc To enter letter x press key 9 twice Press XA to change to numerics 123 10 Enter number 9 11 Confirm entry with EP METTLER TOLEDO User manual ICS669 check 15 1 2 8 Connections Weighing terminal only ICS669a check f ICS669a check t 2 3 pe Jr T Cie Pe Optional interface COM3 Optional interface COM2 Standard interface COM1 RS232 Weighing platform connection SCALE 1 Pressure compensation AC power supply or battery charging Verification securing seal Optional weighing platform connection SCALE 2 o d OD OO P Ga MM ICS669a check c 1 Optional interface COM2 9 2 Weighing platform connection 9 4 3 Pressure compensation qe 4 AC power supply or battery charging 3 5 Standard interface COM1 RS232 Note The verification securing seal is applied on the weighing terminal as described in the section above 1 3 Tracing and tracking features ICS669 check offers some advanced features for tracking and tracing operation on your weighing terminal e User management e Alibi log file e Routine test and routine test log file e Calibration weight management 16 mertier roLeDo User manual ICS669 check Order number 22021161A 12 10 1 3 1 User management The user management of the ICS669 check allows you to manage up to 20 users by user name e user profile operator or supv
41. average weighing results Sign For negative weight values o Stability monitor For unstable weight values 1 2343 kg Non approved last digit For approved scales only with e gt d The example shows the weight value for a scale with e 1 g and d 0 1 g The last smaller digit is not approved Order number 22021161A METTLER TOLEDO User manual ICS669 check 1l 12 METTLER TOLEDO User manual ICS669 check Symbols and info line In the symbols and info line the following information can be displayed Symbol Information Remark l 51 1 Weighing range For multi range or multi interval scales only nA Scale number A Weight below MinWeigh must be activated in the menu minimum weight a Average weighing Average must be activated in the menu Automatic taring Auto Tare must be activated in the menu Automatic clearing of the tare weight A Clear Tare must be activated in the menu Over Under Check weighing to zero To zero must be assigned to a soft key in the menu TA Take away mode Take away must be assigned to a soft key in the menu gt 0 lt Center of zero Availability depending on local Weights and indication Measures regulations Order number 22021161A 12 10 1 2 5 Keyboard METTLER TOLEDO Key Name Function in the operating mode Function in the menu l Power e Switching on and off e Cancel editing O e
42. chnician 78 wermesmoepg User manual ICS669 check Order number 22021161A 12 10 6 2 Errors and warnings 6 2 1 Error messages Error messages contain the following information Warning Information Screen Out of range _ 5 Enter value within the allowed range g 4 Press C Key to confirm SWO3 3 Ge ae 1 Error message 2 Warning symbol 3 Message identifier A How to clear the message 5 Remedy 6 2 2 Warnings Warnings are displayed briefly and then disappear automatically Example Warning Information Screen Key not allowed 1 Warning 2 Warning symbol 3 Warning identifier 12 10 Order number 22021161A METTLER TOLEDO User manual ICS669 check 79 6 2 3 Information Information screens are displayed briefly and then disappear automatically Example Warning Information Screen Cleared EIO 7 3 1 Info message 2 Info symbol 3 Info identifier 80 merrier TOLEDO User manual ICS669 check Order number 22021161A 12 10 6 3 Smart weighing counter spanner icon This weighing instrument features several control functions to monitor the condition of the device The METTLER TOLEDO service technician can setup and enable these functions This helps the user and the METTLER TOLEDO service technician to detemine how the device is treated and what measures are needed to keep it in a good shape If the co
43. cian has to determine and enter a minimum weight value Function Switching minimum weight function on off If the weight on the scale drops below the stored minimum weight B appears in the symbols and info line IDNet Scale Reset Resetting scale settings to factory settings Perform reset Confirmation inquiry e Reset the IDNet scale settings to factory settings with YES Do not reset the IDNet scale settings with NO 12 10 Order number 22021156A METTLER TOLEDO User manual ICS649 check 5 4 4 Application menu block Factory settings are printed in bold in the following overview 4 4 1 Overview Level 1 Level 2 Level 3 Level 4 Average Off Auto Manual Over Under Tolerance type Off Absolute Relative Percent Default values Activate Off On Relative weight Tolerance Tolerance Percent weight Tolerance Tolerance Relative pieces Tolerance Tolerance Output Setpoint tol Good print Off On Data storage Descr field Activate Off On Article name Activate Off On Article number Activate Off On Delete all Sure Delete record Count Reference n Log files Alibi Log Activate Item 7 D1 D2 ID3 Dev m identifier Dev location Item 10 SNR Terminal SNR Scale 1 SNR Scale 2 User name Delete all records Routine Test Log Days External test Test we
44. d press the soft key again If activated the total net and the number of items are increased Unload the scale Repeat steps 3 and 4 for further items To finish totalising press the soft key A safety prompt is displayed Press softkey YES to clear the total o Press softkey NO to continue totalising Piece counting results and Over Under Checkweighing results can be totalised the same way METTLER TOLEDO User manual ICS669 check 25 26 2 7 2 METTLER TOLEDO User manual ICS669 check Automatic totalising Prerequisites Y Application gt Totalise is activated in the menu Y Good print is activated in OverUnder Checkweighing Auto print is activated for other weighing applications v For clearing the total the soft key 3 Total must be activated in the menu under Terminal Device Keyboard Softkeys v If you want to follow the totalisation on the display Total net and or n must be activated for the auxiliary lines and or the info key Terminal Device gt Display gt Auxiliary lines Of Terminal gt Device gt Keyboard Info key Totalising l Weigh the first sample The result is printed out automatically and total net and n are updated If activated the total net and the number of items are displayed Unload the scale Load the next sample If activated the total net and the number of items are increased The result is printed out automatically and tot
45. default values Procedure gt Place the target weight or target amount on the scale and press soft key 29 The applied weight or the applied amount is stored as the target weight or target amount respectively The display changes to the state good factory setting green Over Under Checkweighing is activated 12 10 Order number 22021161A METTLER TOLEDO User manual ICS669 check 39 A0 3 5 METTLER TOLEDO User manual ICS669 check Over Under Checkweighing in Take away mode In this mode the same ifems can be weighed into a container or weighed out of a container without a key having to be pressed between the actions Prerequisite v The soft key TA is activated Procedure l Specify target values as described above For subtractive weighing the target value has to be negative For weighing in place the empty container on the scale for weighing out of the container place the full container on the scale Press soft key TA Take away active is displayed briefly The symbol TA appears in the info line and the container is tared When weighing in place the checkweighing material into the container When weighing out negative target value remove weighing material from the container If the applied removed weight or the applied removed amount lies within the tolerance values taring is carried out automatically The next item can be weighed in removed To use an item that is too light or too heavy taring
46. displayed indicating target values The example shows the default colour for a sample below the lower tolerance Tolerances and target weight are displayed in 3 lines The example shows the default colour for a sample above the upper tolerance The example shows the default colour for a good sample Order number 22021161A 12 10 12 10 Metrological data line In the metrological data line the following information is displayed Symbol Information Remark TO Accuracy classes Displayed only if the scale is approved mi according to the Weights and Measures guidelines uI EA Ce Weighing range For multi range devices only information dsplayed only if the scale is approved according to the Weights and Measures guidelines RES ET Maximum capacity DI Minimum capacity Displayed only if the scale is approved according to the OIML Weights and Measures guidelines je Approved resolution OIML Displayed only if the scale is approved NTEP Displayed only if the scale is approved and d is different from e ER Display resolution OIML Displayed only if the scale is not approved or if d is different from e NTEP Displayed always Approved weighing Metrology display disabled device Weights and Measures data must be indicated on a label near the weight display Weight value The weight value can be marked with the following symbols Symbol Information Remark Calculated weight value E g for
47. e Continuous Ongoing output of all weight values via the interface Dialog Bi directional communication via MT SICS commands control of the device via PC External input Input other than via terminal keypad What the input is used for is defined in the Destination menu block Demand mode Manual data transmission with E gt Demand m auto Automatic transmission of stable results e g for series weighing operations Cont Weight TOLEDO Continuous mode Cont count TOLEDO Continuous mode with counting results Note Printing conditions for Auto print and Demand m auto The weight must be heavier than 9 display increments A weight change of at least 9 display increments is required to initiate the next printout 12 10 Order number 22021156A METTLER TOLEDO User manual ICS649 check 65 Communication gt COM x gt Printer Settings for protocol printout Type Print Template Selecting printer type from the following ASCII printer Label printer GA46 printer Note If Label printer is selected the transmitted data does not include the name of the variable e g date gross ID1 but the value and if apropriate the unit as a separate line This allows the label printer to fill its template with the required data Selecting protocol printout for straight weighing Possible settings Standard Template 1 Template 5 Plus Template Selecting protocol
48. eight value can be displayed in all available units with K 3 Notes e n case of verified scales individual sub items of the Display menu item may not be available or only to a limited extent depending on the respective country e On multi range multi interval scales the symbol I 1 with number indicates the current range or interval IDNet Scale Zero Automatic zero update AZM On verified scales this menu item does not appear Switching on off automatic zero update The effective range of the zero update mode 0 5 d 1 d 2 d 3 d can only be set by service technician IDNet Scale gt Tare Tare function Auto Care Switching on off automatic taring Chain tare Switching on off chain tare A Clear tare Switching on off automatic clearing of the tare weight when the load is removed from the scale e On The tare weight is automatically cleared if the gross weight is or below zero e Off No automatic clearing of the tare weight e 9d The tare weight is automatically cleared if the gross weight is within 9 display steps IDNet Scale Restart Automatic saving of zero point and tare value Restart When the Restart function is activated the last zero point and the tare value are saved After switching off on or after a power interruption the device continues to work with the saved zero point and tare value 5O merer TOLEDO User manual ICS649 c
49. ement Operating life with storage battery The operating life during storage battery operation differs depending on the intensity of use the configuration and the connected scale The following approximate values apply with standard RS232 interface and the brightness set to 5 Weighing platform Conditions Duration With 1 strain gauge weighing cell 10 operation 90 power off mode 150 h e g ICS669a check A15 Continuous operation 15h With 4 strain gauge weighing 10 96 operation 90 power off mode 120 h cells e g a floor scale Continuous operation 12h K line weighing platforms 10 operation 90 96 power off mode 60h Continuous operation 6h METTLER TOLEDO User manual ICS669 check 83 84 METTLER TOLEDO User manual ICS669 check Dimensional drawing METTLER TOLEDO m C Dimension mm I d 260 10 24 b 170 6 70 C 114 4 49 Order number 22021161A 60 11 10 7 2 The size of the weighing platform A BB B BC CC QB QC is indicated at the end Technical data weighing platforms of the product name e g ICS669a check A6 Other combinations of weighing range and readability can be adjusted by the METTLER TOLEDO service technician on site The table below indicates the factory settings of weighing range and readability Weighing ranges and readability factory setting
50. enu is highlighted 2 Press CP Enter code is displayed 3 Enter the password The first menu item Scale is highlighted e When the device is first delivered the supervisor password is set to 423 Set your individual password in the Terminal menu e f a password has still not been entered after a few seconds the scale returns to the weighing mode e f a password has been issued for Supervisor access to the menu and you have forgotten it please contact the METTLER TOLEDO service 12 10 Order number 22021156A METTLER TOLEDO User manual ICS649 check 43 A4 4 1 3 METTLER TOLEDO User manual ICS649 check Calling up menu when user management is active If user management is active password entry is required when logging in Press H The Quick Select menu opens Menu is highlighted 2 Press LP The menu start screen is displayed depending on the user profile Display presentation in the menu Menu items are displayed together with their context The following example shows the menu start screen 2 Calibration Display Units Application Zero Tare Terminal Restart Filter Communication Minkleigh Reset Maintenance 2 e Use navigation keys to navigate through the menu Menu info line i e menu path of the current menu item Menu items the selected menu item is highlighted Navigation info line Sub menu items amp 0N Exiting the menu 1 Press The last menu item End appears Save settings is
51. erminal User management Edit User User 1 User 20 Configuring user profiles Enter user name max xx characters Assign profile User or Supervisor Define password max xx characters Assign user language Define user ID e g personnel number User Name e Profile e Password e Language e User ID Activate Activating deactivating user management Delete all Delete all user profiles Delete Profile Selecting a single user profile to be deleted Supervisor Password Retype code Note Terminal Access Password for Supervisor menu access Password entry for Supervisor menu access Request to enter password gt Enter password Request to repeat the password entry gt Enter password again The password can consist of up to 6 characters 12 10 Order number 22021156A METTLER TOLEDO User manual ICS649 check 62 Terminal Reset Resetting terminal settings to factory settings Perform reset Confirmation inquiry e Reset the terminal settings to factory settings with YES Do not reset terminal settings with NO 4 6 METTLER TOLEDO User manual ICS649 check Communication menu block For detailed information on interface protocols and commands refer to the following documents e SICS Reference manual e MT continuous Reference manual The Communication menu block consists of the following subblocks COM 1 P
52. ervisor e user password e user language e user ID When user management is activated any access to the terminal is protected by password To enter the menu a password is no longer necessary e User management can be configured in the menu under Terminal gt User Management e Login logout with user management is described in chapter 2 1 3 2 Alibi log file If requested by national regulations you can activate an alibi log file to trace all weighing activities on the scale The alibi log file stores all weighings with the mandatory data In addition you can store up fo 4 more items such as identifications serial numbers and user name e The alibi log file can be configured in the menu under Application gt Log files e Viewing printing transferring of the alibi log file is described in the Quick Select menu 1 3 3 Routine test and routine test log file For optimum weighing results the device supports routine calibration tests You can configure routine tests by e Interval days e External test for analog scales e internal test for scales with internal calibration weight For the external test for analog scales you can specify the following e Test weight value e Weight name to make sure you always use the same weight e Tolerance e The routine test can be configured inthe menu under Application Log files e Performing the routine test and viewing printing transferring of the routine test log file is described in
53. fo screen is displayed Press I again With one info screen only the weight display appears With several info screens the next info screen is displayed With several info screens press C to leave the info screens An info screen is displayed until iis pressed again or Cis pressed Order number 22021161A METTLER TOLEDO User manual ICS669 check 29 2 12 Environment and cleaning 2 12 1 Overview The devices are designed to be used in a wet environment Depending on the environment and the cleaning procedures we suggest weighing platforms with different types of load cells The following table gives you a detailed overview about the recommended environment and suitable cleaning procedures Terminal Weighing platform ICS669a check Standard version Option ICS669d check hermetically sealed load cell with stainless steel load cell KS coating IP rating IP68 IP69k IP68 IPG9k IP68 IPG9k Environment Short time wet 30 min day v v v Part time wet 120 min day v v v Permanent wet v v v Cleaning procedure Wet wipe down v v v Light hose down v v v 5 min 20 kPa Light wash down v v v 12 51 min 30 kPa Heavy wash down v v v high pressure water and steam jet up to 10000 kPa Cleaning detergents Mild detergents v v v Other detergents in v v v accordance with the manufacturer s specifications and instructions Acids lyes solvents v
54. h SP Tolerance Red Purple Green Tolerance Orange Light green Good range Pink White Default gt Tolerancet colour Colour mode Continuous Stable IDs D1 ID2 ID3 56 merer TOLEDO User manual ICS649 check Order number 22021156A 12 10 Level 1 Level 2 Level 3 Level 4 Level 5 Level 6 Device Keyboard Key lock Power Lock Unlock Clear UnitSwitch Info Transfer Quick Select Keypad Info key Item 1 Not used Date amp Time Gross Net Tare HighRes amp Net Item 27 D1 ID2 ID3 Dev identifier Dev location Dev name SNR Terminal SNR Scale 1 Firmware Vers Target Tolerance Tolerancet Tolerance type Descr field Deviation Average PW Reference count Quantity Record number Article name Article number Total net n Soft keys Page 1 Soft key Not used Target Page 2 x 1 x 5 Recall Store Page 3 Take away To zero D1 ID2 ID3 Reference n Average PW Switch scale Plus Total Date amp Time Format EU US Date dd mm yyyy EU mm dd yyyy US Time hh mm ss Meridian AM PM Calender On Off Beeper On Off Information Identification Location User Edit user User 1 User name Profile Password Language management User 20 User I
55. heck Order number 22021156A 12 10 IDNet Scale Filter Adaptation to the ambient conditions and the weighing type Vibration Adaptation to the ambient conditions Stable e Very steady and stable environment The scale works very rapidly but is very sensitive to external influences Normal e Normal environment The scale operates at medium speed Unstable e Restless environment The scale works more slowly but is insensitive to external influences Process Adaptation to the weighing process Finefill e Dispensing of liquid or powdered weighing samples Universal e Universal setting for all weighing modes and normal weighing goods Absolut e For solid bodies under extreme conditions e g strong vibrations Stability Adjusting stability monitoring ASD 0 ASD A ASD 0 Stability monitoring switched off Only possible for non verified scales ASD 1 Rapid display Good reproducibility ASD 2 ft y ASD 3 f y ASD 4 Slow display Excellent reproducitility IDNet Scale gt Update Setting the display speed of the weight display This menu item is only displayed if the UPDATE function is supported by the connected scale xx UPS Selecting the number of updates per second UPS Note The possible settings depend on the connected scale IDNet Scale gt MinWeigh Minimum weighing in quantity Before you can use this function the METTLER TOLEDO service techni
56. heck c 3 6 kg 7 9 Ib weight of the weighing platform Mains Direct connection to power supply supply voltage fluctuation not connection exceeding 10 of the rated voltage Rated voltage 100 240 VAC 50 60 Hz 300 mA Supply of device 12 V 2 5 A operation e f the supply voltage is interrupted the device automatically Switches over to storage battery operation Battery charger e Ambient conditions O 40 C 32 104 F dry environment Ambient e Application indoor use only conditions e Altitude up to 2 000 m e Temperature range Class Ill 10 40 C 14 104 F e Temperature range Class Il 0 40 C 32 104 F e Overvoltage category I e Pollution degree 2 e Humidity Max rel humidity 80 for temperatures up to 31 C decreasing linearly to 50 96 rel humidity at 40 C Interfaces e interface RS232 integrated e ICS669a check c additional optional communication interface possible e CS669a check d ICS669a check ICS669a check t 2 additional optional communication interfaces possible additional scale interface possible W amp M e OIML Class Il Ill III approvals e NTEP Class Il Ill Order number 22021161A 11 10 11 10 Order number 22021161A Applications e Weighing e Over Under Checkweighing e Piece counting e Average weighing e Totalising e Albi log file e Routine test function e Calibration log file e User manag
57. ighing platforms ICS669d check with digital scale interface fo connect METTLER TOLEDO weighing platforms with IDNet interface On the rear the weighing terminal is equipped with a swivel bracket for mounting the terminal on the wall or to a METTLER TOLEDO column As an accessory a table stand for setting up the terminal on the table is available 1 2 2 Terminal and platform combinations The complete name of a terminal and platform combination also indicates the type size and capacity of the connected analog weighing platform E g ICS669a check A6 c stands for ICS669a check type of weighing terminal and type of weighing interface A design and size of the weighing platform 6 weighing platform capacity in kg C mechanical design As default the weighing platforms are equipped with a hermetically sealed stainless steel load cell and a readability setting of 2 x 3000 or 5000 divisions non approved ICS669a check f The weighing terminal is fixed mounted in front of the weighing platform Terminal and platform can be handled as one unit easy to install and to change location The perfect solution if a stand or a bracket would hinder an effective working process ICS669a check t Weighing terminal and weighing platform are connected by cable Suitable for wall mount operation and desk operation with an additional desk mounting plate see Options The combination can be upgraded with a stand see Accessories 6 mertier
58. ight Weight name Tolerance Internal test Delete all records Totalize Activate Off On Reset Perform reset 52 METTLER TOLEDO User manual ICS649 check Order number 22021156A 12 10 4 4 2 Description Application Average Determining the average weight for an unstable load dynamic weighing Off Calculating average weight switched off Auto Calculating average weight with automatic start of the weighing cycle Manual Calculating average weight with manual start of the weighing cycle via gt Application gt Over Under Setting Over Under Checkweighing parameters Tolerance type Off Absolute Relative Percent Specify which parameters have fo be entered for Over Under Checkweighing No tolerance type predefined it can be set individually when entering Over Under Checkweighing parameters A low and a high weight value must be entered These weights and all weights within this range are treated as being within tolerance The target weight has to be entered as an absolute weight upper and lower tolerances as deviations in weight from the target weight The target weight has to be entered as an absolute weight upper and lower tolerances as deviations in percent from the target weight Not possible for counting Default values Activate Relative weight Percent weight Relative pieces If you always use the same tolerances for Over Under Checkweigh
59. ing you can store these tolerances and thus avoid entering tolerances all the time Activating usage of default tolerance values e Off Default tolerance values not used e On Default tolerance values used Enter the weight values for Tolerance and Tolerance Enter the percentages for Tolerance and Tolerance Enter Tolerance and Tolerance in pieces Output Setpoint tol Good print Setting display and printing options When Setpoint tol is reached the coloured display will change from the Setpoint fol colour to the Tolerance colour This setpoint is available on the optional digital I O interface as well This feature can be used to show the Tolerance colour when you are already near the target or if you need an additional setpoint for I O control Possible settings 0 100 96 of the Tolerance value Automatic printout e Off No automatic printout e On Automatic printout when a stable weight value within the tolerance values exists 12 10 Order number 22021156A METTLER TOLEDO User manual ICS649 check 53 54 Application Data storage Setting data base parameters Descr field Activate Descriptor The description field can be used to expand each target record by one auxiliary field This field can be used for an article description article number or any other information on the record It will be stored as part of the target record it can be disp
60. isplayed Scale not calibrated Call the METTLER TOLEDO service technician to calibrate the scale e When user management is active you are asked to select your name and enter the corresponding password Press Before the display goes out OFF appears briefly Login logout When user management is active a login logout procedure is required The login screen is displayed after switching on or logging out 1 Select your name using the cursor keys v and confirm with gt 2 Enter your password using the displayed keyboard A welcome screen and then the weight display appear Press E to open the Quick Select menu 2 Select Logout using the cursor keys v and confirm with La A safety prompt is displayed 3 Press L gt The login screen is displayed the current user is logged out Always log out when leaving the terminal in order to prevent unauthorised persons from working on it Order number 22021161A METTLER TOLEDO User manual ICS669 check 22 2 2 Manual Automatic 2 3 2 4 2 4 1 2 4 2 METTLER TOLEDO User manual ICS669 check Zeroing Zero point correction Zeroing corrects the influence of slight changes on the load plate or minor deviations from the zero point 1 Unload scale 2 Press gt 0 Zero appears in the display In case of non verified scales the automatic zero point correction can be deactivated in the menu or the zero range can be changed Approved scales are set
61. isplayed The device operates in straight weighing mode Working with identifications Weighing series can be assigned 3 identification numbers ID1 ID2 and ID3 with up to 40 characters that are also printed out in the protocols If for example a customer number and an article number are assigned it can be clearly seen in the protocol which article was weighed for which customer Direct entry v At least one of the soft keys ID1 ID2 or ID3 is activated 1 Press the desired soft key ID1 ID2 or ID3 The last entered ID is displayed 2 Enter the ID and confirm with L gt The entered ID is assigned to the following weighings until the ID is changed Barcode use for one identification only v D1 ID2 or ID3 is selected as destination for external input v To display the identification ID1 ID2 or ID3 is activated in the auxiliary line Using SICS command set up to three identifications v To display the identification s the corresponding IDs are activated in the auxiliary line Order number 22021161A 12 10 12 10 2 11 Displaying information Up to 27 different values for display can be configured in the menu for the info key Depending on the configuration in the menu Terminal gt Device gt Keyboard 3 Info key the following data can be assigned in a free order e g Date amp Time Weight values Identifications Device information Serial numbers and software versions User name Press The first in
62. ith a tolerance type selected in the menu this step does not appear 3 Load requested weight and press soft key OK The next parameter is highlighted 4 Repeat step 3 until New target set is displayed The coloured Over Under Checkweighing display appears the scale is ready for Over Under Checkweighing e If tolerance default values have been set in the menu only the target has to be specified with tolerance types Relative and Percent e The upper tolerance value has to be greater than or equal to the lower one High gt Low or respectively the target weight has to be greater than or equal to the lower tolerance value and smaller than or equal the upper tolerance Tol Target Tol Entering known target values The following section describes the course of the factory setting for the Over Under Checkweighing application 1 Press 2 The current Over Under Checkweighing parameters are displayed 2 Check tolerance type Use soft key to change the tolerance type and soft key OK to proceed to the first weight With a tolerance type selected in the menu this step does not appear 3 Enfer the requested weight value and confirm with D gt The next parameter is highlighted 4 Repeat step 3 until New target set is displayed The coloured checkweighing display appears the scale is ready for checkweighing e f tolerance values have been set in the menu only the target has to be specified with tolerance types Relative
63. ities Please contact your METTLER TOLEDO organisation The first Step of the verification has been carried out at the manufacturing plant If national regulations in individual countries limit the period of validity of the verification the operator of such a weighing instrument is himself responsible for its timely re verification Order number 22021161A 12 10 8 3 Tables of Geo Code values For weighing instruments verified at the manufacturer s the Geo Code value indicates the country or geographical zone for which the instrument is verified The Geo Code value set in the instrument e g Geo 18 appears briefly after switching on Table Geo Code values 3000e shows the Geo Code values for European countries Table Geo Code values 60006 7 500e shows the Geo Code values for different gravitation zones 8 3 1 Geo Code values 3000e OIML Class III European Countries Country Geographical latitude Geo Code value Country Geographical latitude Geo Code value Austria 46 22 49 01 18 Liechtenstein 47 03 47 14 18 Belgium 49 30 51 30 21 Lithuania 53 54 56 24 22 Bulgaria 41 41 44 13 16 Luxemburg 49 27 50 11 20 Croatia 42 24 46 32 18 Netherlands 50 46 53 32 21 Czechia 48 34 51 03 20 Norway 57 57 64 00 24 Denmark 54 34 57 45 23 64 00 71 11 26 Est
64. layed printed or transferred The description field has a maximum length of 40 characters Activating description field for identifying sets of Over Under Checkweighing parameters Off No possibility to enter a value in the description field during target record definition The content of already existing description fields is not affected e On During target record entry you are asked to enter a value in the description field Enter a name for the description field Factory setting Item Article name Activate With each target record an article name can be stored e Off No possibility to enter an article name during target record definition The content of already existing article names is not affected On During target record entry you are asked to enter an article name Article number With each target record an article number can be stored Activate Off No possibility to enter an article number during target record definition The content of already existing article numbers is not affected e On During target record entry you are asked to enter an article number Delete all Delete all records of the database Sure A safety prompt is displayed before deleting the database records Delete record XX Delete a single record in the database Enter number of the record to be deleted Notes There is a SICS command ovailable to write a target record including description field article name and
65. le to be able to perform a new average weighing operation Piece counting Piece counting with a fixed number of reference pieces Prerequisite v The corresponding soft key is activated in the terminal menu 1 Apply the reference parts 2 If itis the defautl number of reference parts press EA Otherwise enter the number of reference parts and press ES The number of reference parts is displayed 3 Load the parts to be counted The number of pieces is displayed e The average piece weight remains stored until a new average piece weight is de termined e The auxiliary line and the info page can be configured to show the average piece weight METTLER TOLEDO User manual ICS669 check 27 28 2 9 2 2 9 3 2 10 METTLER TOLEDO User manual ICS669 check Piece counting with a known piece weight Prerequisite v The corresponding soft key l is activated in the terminal menu Press E Input of the average piece weight APW is requested 2 Enter the average piece weight The weight unit changes to PCS 3 Load the parts to be counted The number of pieces is displayed e The average piece weight remains stored until a new average piece weight is de termined e You can enter the average piece weight via barcode if the external input destination is configured to Average PW Terminating piece counting Press C Cleared appears in the display The average piece weight is cleared and the weight value is d
66. n Totalise Activating deactivating totalisation Activate Activate deactivate totalisation Totalisation is possible for piece counting and OverUnder Checkweighing as well Application Reset Resetting application settings to factory settings Perform reset Confirmation inquiry e Reset the application settings to factory settings with YES Do not reset application settings with NO 12 10 Order number 22021156A METTLER TOLEDO User manual ICS649 check 55 4 5 Terminal menu block Factory settings are printed in bold in the following overview 4 5 1 Overview Level 1 Level 2 Level 3 Level 4 Level 5 Level 6 Device Language English German French Spanish Italian Sleep Off 1 minute 3 minutes 5 minutes 15 minutes Power off 30 minutes Display Layout Default 3 Line mode Color mode Auxiliary Not used Date amp Time Gross Net line 1 3 Tare High Resolution ID1 ID2 ID3 Dev identifier Dev location Target Descr field Deviation Average PW Reference count Quantity Record number Article name Article number Total net n User name Contrast 1 10 Brightness 1 10 Backlight off 5 seconds 10 seconds 30 seconds 1 minutes On Weight hold 0 s 10 s Stealth On Off mode Colours Default color Yellow Light blue MinWeigh Dark blue lt MinWeig
67. nals have been assessed by the EHEDG European Hygienic Engineering and Design Group and the NSF National Sanitation Foundation Both institutes certify the fulfilment of the hygienic requirements for easy cleaning Hygienic Design Criteria The EHEDG is an association of device manufacturers firms in the foodstuff industry research institutes and health autohorities It was founded in 1989 with the aim of promoting the hygienically faultless manufacturing and packaging of foodstuffs A positive expertise of the device by the EHEDG has taken place A corresponding report is available on the internet under www mt com NSF is an independent NGO founded in 1944 in the USA Corresponding regulations were published for the use of devices in the foodstuff industry The device fulfils the NSF cfriteria C 2 Special Equipment and or Devices for use in the foodstuff industry The products are listed on the NSF site www nsf org Notice for verified instruments in EC countries Weighing instruments verified at the place of manufacture bear the preceding mark on the packing label and a green M sticker on the descriptive plate They may be set to work immediately Weighing instruments which are verified in two steps have no green M on the descriptive plate and bear the preceding identification mark on the packing label The second step of the verification must be carried out by the approved METTLER TOLEDO service or Weights and Measures author
68. nes 78 Error conditions 78 Error messages 79 METTLER TOLEDO User manual ICS669 check G Geo value 21 32 93 l Identifications 28 Info dE 29 Information sesoses 80 Interface settings 62 Digital WO 68 Eiliernel uestes 69 TEA KE 64 RS422 R8485 64 UN js EN 69 WAN iue mitten 70 K Keyboard sssess 13 Key Joch 60 L Log files Alibi log file 76 Routine test log ie 75 76 Login logout se 21 M Menu ACCOSS iaceret ideas 61 Application 52 Communication 62 Maintenance 71 Operation 43 Password 48 61 Quick Select 73 Scale analog 46 Scale IDNet 49 Terminal seenen 56 Messages sii 78 MinWeigh 49 51 Mode Demand mode 65 Dialog mode ssssesssriiss 65 Print mode 65 MT continous 62 0 OptIONS n 7 86 Output mode 69 Over Under Checkweighing 10 33 Checkcounting 35 YS 0 0 36 Quick start 39 Stored target values 35 Subtractive weighing 38 Take away mode 40 Tolerance types
69. ng the cursor keys v and confirm with E 3 If Print records by number or Print records by date is selected you are asked to enter start and end number respectively start and end date The selected alibi record s is are printed Order number 22021161A 12 10 Searching alibi records 1 When an alibi record is displayed press soft key C 2 Inthe next screen use X Ain select the search criterion either Search by date or Search by rec no record number and confirm with E 3 Enter record number resp date of the alibi record s you are looking for and confirm with gt The desired alibi record s is are displayed Deleting alibi log files Deleting alibi log files is carried out in the menu under Application gt Log files gt Alibi Log 5 7 Calling up calibration log file For analog scales the results of calibration procedures are stored in the calibration log file Viewing calibration log file 1 Select Calibration Log in the Quick Select menu using the cursor keys v and confirm with L gt The calibration record of the last calibration is displayed Calibration passed 2 To view other calibration records use the cursor keys v Printing calibration records 1 When a calibration record is displayed press D gt 2 In the next screen select either Print selected record to print a single record or Print whole memory fo print all records 3 Confirm selection with LP The calibration record s is are prin
70. nstructions gt Call the METTLER TOLEDO service technician or carry out commissioning in accordance with the installation instructions Order number 22021161A 12 10 12 10 1 4 4 Power supply connection CAUTION Risk of electric shock A A A Before connecting the power supply check whether the voltage value printed on the rating plate corresponds to your local system voltage Do not under any circumstances connect the device if the voltage value on the rating plate deviates from the local system voltage Make sure the weighing platform has reached room temperature before switching on the power supply Plug the power plug into the power socket After it has been connected the device runs a self test The device is ready to operate when zero appears on the display 1 4 5 Handling of the built in storage battery Note the following when operating a device with a built in storage battery The operating life depends on the intensity of use the configuration and the connected scale For details see the technical data The battery symbol shows the current state of charge of the storage battery One segment corresponds with approx 25 capacity If the symbol flashes the storage battery has to be charged A message is displayed too During charging the segments are running until the battery is fully charged and all segments light up continuously The charging time of the storage battery amounts fo approx
71. ntrol functions triggers an alert a message is shown You can confirm the message and continue to work with the weighing instrument The spanner icon a lights up Warning Information Screen Service required Service Next service due Press C Key to confirm SW17 In case of an alert we strongly recommend calling the METTLER TOLEDO service technician e to replace parts which are at the end of lifetime e to correct wrong settings e to educate operators about proper handling e to perform routine service work e to reset the alert The control functions monitor the following conditions e number of weighings e number of overloads e maximum weight e zero commands and zero failures e battery charging cycles e power on time e date for the next service inspection 12 10 Order number 22021161A METTLER TOLEDO User manual ICS669 check 81 82 7 7 1 METTLER TOLEDO User manual ICS669 check Technical data and accessories Technical data weighing terminal Housing Stainless steel 1 4301 or AISI 304 Display e Color TFT graphical display with backlighting e Size 115 x 85 mm 320 x 240 pixels Keyboard e Tactile touch membrane keypad Scratch resistant labelling Storage battery Protection type e Terminal IP68 IPG9k e Standard weighing platform IP68 IP69k e Weighing platform with option load cell with KS coating IP68 IP69k Net weight e Terminal 2 3 kg 5 1 Ib e CS669a c
72. onia 57 30 59 40 24 Poland 49 00 54 30 21 Finland 59 48 64 00 25 Portugal 36 58 42 10 15 64 00 70 05 26 Romania 43 37 48 15 18 France 41 20 45 00 17 Slovakia 47 44 49 46 19 45 00 51 00 19 Slovenia 45 26 46 35 18 Germany 47 00 55 00 20 Spain 36 00 43 47 15 Greece 34 48 41 45 15 Sweden 55 20 62 00 24 Hungary 45 45 48 35 19 62 00 69 04 26 Iceland 63 17 67 09 26 Switzerland 45 49 47 49 18 Ireland 51 05 55 05 22 Turkey 35 51 42 06 16 Italy 35 47 47 05 17 United Kingdom 49 00 55 00 21 Latvia 55 30 58 04 23 55 00 62 00 23 factory setting 12 10 Order number 22021161A METTLER TOLEDO User manual ICS669 check 93 94 8 3 2 x METTLER TOLEDO User manual ICS669 check Geo Code values 6000e 75000e OIML Class III Height lt 1000 m Geographical latitude Geo Code value Geographical latitude Geo Code value 00 00 12 44 18 43 26 47 51 18 05 46 17 10 2 45 38 50 06 22 12 44 20 45 16 47 51 52 22 20 17 10 23 54 18 50 06 54 41 21 20 45 26 45 20 52 22 57 04
73. ow The colours can be modified in the menu Prerequisite v The soft keys for Over Under Checkweighing are activated in the terminal menu section 3 1 Specifying target values Different entries are required at the beginning of Over Under Checkweighing or Over Under Checkcounting depending on the tolerance type setting Tolerance type Absolute A low and a high weight value must be entered These weights and all weights within this range are treated as being within tolerance Tolerance type Relative Target weight Target as well as lower tolerance Tol and upper tolerance Tol have to be specified The tolerances are displayed as relative deviations from the target weight Tolerance type Percent Target weight Target as well as lower tolerance Tol and upper tolerance Tol have to be specified At Over Under Checkweighing the weight value is represented as a percentage of the target weight The target weight 100 is striven for or O 96 at Over Under Checkweighing to zero Order number 22021161A METTLER TOLEDO User manual ICS669 check 33 34 3 1 1 METTLER TOLEDO User manual ICS669 check Weighing in target values The following section describes the course of the factory setting for the Over Under Checkweighing application 1 Press 7 9 The current Over Under Checkweighing parameters are displayed 2 Check tolerance type Use soft key to change the tolerance type and soft key OK to proceed to the first weight W
74. rging process is completed i e all segments of the battery symbol light up continuously For optimum battery performance operate the device with built in storage battery at an ambient temperature in the range of 10 C to 30 C 50 F to 86 F This applies to discharging as well as charging the battery If you plan to put the scale out of operation for a longer period charge the battery completely Even if you do not use the instrument charge the battery at least every 3 months to avoid deep discharge Use in hygienically sensitive areas The device is easy to clean and is designed to be used in the food industry Features Suitability of the materials for contact with foodstuffs Construction made of stainless steel No open threads No screws with recesses Smooth non porous and flat surfaces that are easy to clean Reduced horizontal surfaces Continuous welding seams For further information please refer to the appendix Order number 22021161A 12 10 12 10 2 1 2 1 1 Switching on Switching off Login Logout Operation Switching on and off Switching on off Press For a few seconds the device shows a start up screen with device name software version serial number of the weighing terminal and the Geo value only if an analog weighing device is connected e You can freeze the start up screen by pressing i e When you start a weighing system with an analog scale the first time a message is d
75. rompt is displayed 2 Press gt again to start printing Reset A11 Reset all settings to factory settings Perform reset Confirmation inquiry e Reset all settings to factory settings with YES e Do not reset settings with NO 72 wertuer roLEDo User manual ICS649 check Order number 22021156A 12 10 12 10 5 1 5 2 5 3 Quick Select menu Quick Select menu overview The Quick Select menu offers access to logout routine test and several log files depending on your configuration gt Press H The following menu is displayed Quick Select Log out Routine test Routine Test Log Alibi Log Calibration log The example shows the Quick Select menu with the maximum of configurable items Entering main menu gt In weighing mode press E and then L gt The main menu is displayed without a long key press Logout Prerequisite v User management is activated under Terminal gt User Management e Logout is described in the Operation chapter e Always log out when leaving the terminal in order to prevent unauthorised persons from working on it Order number 22021161A METTLER TOLEDO User manual ICS669 check 73 5 4 Performing routine test By performing a routine test you can check the calibration of your scale regularly Prerequisite v Routine test parameters are set under Application gt Log files If an interval for the routine test is defined Days gt 0 the device au
76. s Order no RS232 adapter 8 pin M12 plug lt gt 8 pin Binder receptacle 0 2 m 22 021 094 RS485 adapter 6 pin M12 plug lt gt 6 pin Binder receptacle 0 2 m 22 021 095 Ethernet adapter 4 pin Coding D M12 plug lt gt 16 pin Binder receptacle 0 2 m 22 021 096 USB adapter 4 pin Coding A M12 plug lt gt 16 pin Binder receptacle 0 2 m 22 021 097 Digital I O adapter 12 pin M12 plug lt gt 19 pin Binder receptacle 0 2 m 22 021 098 Use already installed cables plugs with our new ICS6x9 M12 plug Order number 22021161A 11 10 11 10 Mechanical parts Order no Protective cover for terminals ICS6x9 set of 3 pieces 22 021 110 Stand ICS6x9 height 50 mm 22 018 057 Stand ICS6x9 for PBA430 weighing platform height 330 mm 22 013 964 height 660 mm 22 013 965 Stand ICS6x9 for KA KB MA MB and DB platforms height 330 mm 22 014 836 Bench stand ICS6x9 for scale bench 00503632 or 00504854 height 500 mm 22 014 835 Floor stand ICS6x9 height 1000 mm 22 014 834 Standbase for floor stand 22 011 982 Wall bracket ICS6x9 inclinable and swivelling 22 014 833 Desk mounting plate for teminal and t version only 22021 111 Order number 22021161A METTLER TOLEDO User manual ICS669 check 91 92 8 1 EHEDG METTLER TOLEDO User manual ICS669 check NSF Appendix Tests for utilisation in hygienically sensitive areas ICS669 check weighing termi
77. sion kit 4 outputs and 4 inputs terminal version 22012116 C Version 22 012 121 WLAN conversion kit terminal version 22 012 126 Cc version 22012 127 Plugs Order no RS232 counter plug 8 pin M12 22021 105 RS485 counter plug 6 pin M12 22021 106 Ethernet counter plug 4 pin Coding D M12 22 021 107 USB counter plug 4 pin Coding A M12 22 021 108 Order number 22021161A METTLER TOLEDO User manual ICS669 check 89 METTLER TOLEDO User manual ICS669 check Cables always delivered with 90 angled M12 plug Order no RS232 cable for SICS scale 8 pin M12 lt gt 9 pin sub D plug 3 m 22 021 088 RS232 cable for PC 8 pin M12 lt gt 9 pin sub D receptacle 3 m 22 021 087 RS422 RS485 cable 6 pin M12 lt gt open ends 3 m 22 021 089 Ethernet 10 100 Base T twisted pair cable 4 pin M12 Coding D lt gt RJ45 5m 22 021 090 20m 22 021 091 USB adapter cable 4 pin M12 Coding A lt gt USB series A receptacle 0 2 m 22 021 122 5m 22 021 123 USB cable connection to PC 4 pin M12 Coding A lt gt USB series A plug 3 m 22 021 092 USB cable connection to USB devices 4 pin M12 Coding A lt gt USB series B plug 3 m 22 021 124 Cable to connect Digital 1 0 option with Relay box 12 pin M12 lt gt open ends 10 m 22 021 093 1 0 accessories Order no Relay box for Digital I O option 22 011 967 Power supply for Relay box 4 110 230 VAC 00 505 544 Adapter
78. t protocols use the cursor keys v Order number 22021161A METTLER TOLEDO User manual ICS669 check 75 76 5 6 METTLER TOLEDO User manual ICS669 check Printing routine test log file 1 When a routine test record is displayed press D gt 2 In the next screen select either Print selected record to print a single record or Print whole memory fo print all records 3 Confirm selection with E gt The routine test log record s is are printed Deleting routine test log file Deleting routine test log files is carried out in the menu under Application gt Log files Routine Test Log Calling up alibi log file Prerequisite Y Alibi Log is activated under Application gt Log files Viewing alibi records 1 Select Alibi Log in the Quick Select menu using the cursor keys v and confirm with L gt The alibi record of the last weighing operation is displayed Alibi log HBB DD Rec no BAB I HBB BAS Date amp Time ZAC 13 27 41 HEH BIB Net 9 48 kg 888 411 Tare HDH kg ABB 81 Gross 9 48 kg ABA AIS Scale number ABA Ald ABA 815 888 B16 ABA Al 2 To view other alibi records use the cursor keys v Printing alibi log file 1 When an alibi record is displayed press gt In the next screen you are given the following choice Print selected record Print whole memory Print today s records Print records by number Print records by date 2 Select the desired printing mode usi
79. t the position Soft key x 1 Soft key x 5 3 Assign contents For possible soft keys refer to the Introduction Note e f you want to lock the tare key 9 T and or the zero key gt 0 ask the METTLER TOLEDO service technician e Locked keys cannot be activated by the user but the supervisor can still activate these keys by entering his password Date amp Time Format Date Time Meridian Calender Setting date and time Selecting date format Possible settings EU US Setting date in the selected format dd mm yyyy EU or mm dd yyyy US Setting time in the following format hh mm ss For US format only Setting AM PM Display a calender sheet in the right bottom edge of the display METTLER TOLEDO User manual ICS649 check Order number 22021156A 12 10 Beeper Each keystroke is confirmed by a short beep Switching beeper on off Information Identification Location Entering device information to identify the device according to your company s naming conventions Entering device identification Entering device location Note This device information can be used as follows e fo be displayed in the auxiliary lines of the display e to be displayed via 1 e to be printed transferred together with the weight value In addition Device name provides the complete type information already entered in the factory e g ICS649a check A1 5 t T
80. tare weight is entered or the tare weight is Cleared Switching scales Prerequisites v For manual totalisation the soft key A A Switch scale is activated in the menu under Terminal Devic gt Keyboard Softkeys gt Press the soft key AA to switch from scale 1 to scale 2 or vice versa The current active scale is displayed in the symbol and info line on the top of the display Printing results If a printer or computer is connected weighing results and other information can be printed out or transferred to a compuler Press E gt The defined data is printed out or transferred to the computer The printout content can be defined inthe Templates menu Order number 22021161A 12 10 12 10 2 7 Order number 22021161A Totalising Totalising manually Prerequisites Y Application gt Totalise is activated in the menu v For manual totalisation the soft keys Plus and X Total must be activated in the menu under Terminal gt Device gt Keyboard gt Softkeys v f you want to follow the fotalisation on the display activate Total net and or n for the auxiliary lines and or the info key Terminal Device Display gt Auxiliary lines Of Terminal gt Device gt Keyboard gt Info key Totalising l Weigh the first sample and press the soft key If activated the total net and the number of items are displayed Unload the scale Load the next sample an
81. ted 12 10 Order number 22021161A METTLER TOLEDO User manual ICS669 check 77 Event and error messages 6 1 Error conditions Error Cause Remedy Display dark e Backlighting set too dark gt Set backlighting brighter e No mains voltage Check mains e Unit switched off gt Switch on unit e Mains cable not plugged in Plug in mains cable e Brief fault gt Switch device off and on again Weight display unstable e Restless installation location gt Adjust vibration adapter e Draft Avoid draft e Restless weighing sample gt Dynamic weighing e Contact between weighing pan and or Remedy contact weighing sample and surroundings e Mains fault Check mains Incorrect weight display e Incorrect zeroing gt Unload scale set to zero and repeat weighing operation Incorrect tare value gt Clear tare e Contact between weighing pan and or Remedy contact weighing sample and surroundings e Weighing platform tilted gt Level weighing platform e Load plate not on the scale gt Place load plate on the scale UTE e Weighing range not reached gt Set to zero e Weighing range exceeded gt Unload scale gt Reduce preload pe ren pe e Result not yet stable gt f necessary adjust vibration adapter Attention Approval invalid e Approval was tampered with gt Call METTLER TOLEDO service alternating with metrological data te
82. ted number If the selected record is already occupied the background lighting changes to red and Occupied is displayed 1 In order to overwrite the record with the new values press soft key OK The message Overwrite record appears 2 Press soft key YES to overwrite the record or soft key NO to cancel 12 10 Order number 22021161A METTLER TOLEDO User manual ICS669 check 35 Loading the target records l Press soft key gt The first occupied record is shown 2 Select the desired record number 3 Press soft key OK Record loaded briefly appears in the display The Over Under Checkweighing display appears the scale is ready for Over Under Checkweighing Searching for target records Using the soft key OQ you can search for target records by record number article name and article number or item if specified in the target set 1 Press soft key gt The first occupied record is shown 2 Press soft key ou Search by is displayed 3 Using soft key 2 you can select the search criterion i e record no article no article name and item 4 Confirm search criterion with L2 5 Enter record data to be searched for If existent the searched record is displayed 6 Press soft key gt to load the target record 3 2 Over Under Checkweighing procedure The device facilitates Over Under Checkweighing and Over Under Checkcounting through different coloured background lighting for the states too light factor
83. the Quick Select menu 12 10 Order number 22021161A METTLER TOLEDO User manual ICS669 check 17 1 3 4 1 4 2 1 4 3 18 wetter roLEDo User manual ICS669 check Calibration weight management For verified scales the calibration results are stored in the calibration log Viewing printing transferring of the calibration log file is described in the Quick Select menu Commissioning Selecting the weighing platform location The correct location is crucial to the accuracy of the weighing results gt Select a stable vibration free and if possible a horizontal location for the weighing platform The ground must be able to safely bear the weight of the fully loaded weighing platform gt Observe the following environmental conditions No direct sunlight No strong drafts No excessive temperature fluctuations Levelling the weighing platform Only weighing platforms that have been levelled precisely horizontally provide accurate weighing results Weights and Measures approved weighing platforms have a spirit level to simplify levelling 1 Tum the adjustable feet of the weighing platform until the spirit level s air bubble is inside the inner circle 2 Tighten the lock nuts of the adjustable feet Weighing platform connection and interface commissioning The weighing platform connection to the weighing terminal as well as the comminssioning of the interfaces are described in the ICS4x9 installation i
84. the terminal can be used When set to On a resistor of approx 100 Q between the signal lines is enabled 12 10 Order number 22021156A METTLER TOLEDO User manual ICS649 check 67 Communication gt COM x gt Reset COM x Resetting communication settings to factory settings Perform reset Confirmation inquiry e Reset the communication settings to factory settings with YES Do not reset communication settings with NO 4 6 4 Digital 1 0 menu blocks Level 1 Level 2 Level 3 Input Input pin 1 Input pin 4 Off Zero Tare Transfer Switch Clear Info Target Take away Output Ready Stable Tare Zero Off Output Pin 1 Output Pin 4 MinWeigh MinWeigh Underload Overload Setpoint 1 Setpoint 1 lt Setpoint 2 gt Setpoint 2 Star SP Tolerance lt Tolerance Good range Target gt Tolerance Setpoints Setpoint 1 Setpoint 2 Output mode Continuous Stable Reset COM 2 Perform reset COM x Digital 1 0 gt Input Output Configuring inputs outputs Configuring inputs 1 Select an input pin 2 Assign an input signal to the selected input pin Configuring outputs 1 Select an output signal 2 Assign an output pin COM x Digital 1 0 Setpoints Entering values Setpoint 1 Setpoint 2 Enter value for setpoint 1 Enter value for setpoint 2 68 merer TOLEDO User manual ICS649 check Order number
85. tion process can be aborted using e The calibration protocol is stored in the calibration log file Analog Scale Display Units Weighing unit and display accuracy Unit 1 Select weighing unit 1 g kg oz Ib Ib oz t Unit 2 Select weighing unit 2 g kg oz Ib Ib oz t Resolution Select readability resolution the possible settings depend on the connected scale Unit roll When unit roll is switched on the weight value can be displayed in all available units with E 3 Notes e n case of verified scales individual sub items of the Display menu item may not be available or only to a limited extent depending on the respective country e On dual range dual interval scales resolutions marked with I lt gt I 1 2 are divided up into 2 weighing ranges intervals e g 2 x 3000 d 12 10 Order number 22021156A METTLER TOLEDO User manual ICS649 check 47 48 Analog Scale Zero Automatic zero update AZM On verified scales this menu item does not appear Switching on off automatic zero update and selecting zeroing range Possible settings Off 0 5 d 1 d 2d 5 d 10d Analog Scale Tare Tare function Auto tare Switching on off automatic taring Chain tare Switching on off chain tare A Clear tare Switching on off automatic clearing of the tare weight when the load is removed from the scale e On The tare weight is automatically cleared if the gross weight is or below zero
86. tomatically asks you to perform the routine test With external weight 1 Unload the scale 2 Select Routine Test in the Quick Select menu using the cursor keys v and confirm with E gt You are asked to put the indicated weight on the platform 3 When the required weight is put on the platform press D gt The routine test is carried out and the following test protocol is displayed for a short time Routine test 74 meter TOLEDO User manual ICS669 check Order number 22021161A 12 10 12 10 With internal weight 1 Unload scale 2 Select Routine Test inthe Quick Select menu using the cursor keys v and confirm with E gt The routine test is carried out and the following test protocol is displayed for a short time Routine test e The results of the routine test are stored in the routine test log file If the determined weight is not within the tolerance test protocol is in red Call the METTLER TOLEDO service technician e ifan external test weight is defined an external routine test is performed directly after the internal routine test EI 5 5 Calling up routine test log file Prerequisite v Routine test parameters are set under Application gt Log files Viewing routine test log file 1 Select Routine Test Log inthe Quick Select menu using the cursor keys v and confirm with La The routine test protocol of the last routine test is displayed Routine test passed 2 To view other routine fes
87. verage dynamic weidhinmg mmm 27 29 Piece COUNTING et coat tn de eem vete e od teas 27 2 10 Working Zu ie uer le EE 28 241 Displaying information sss mmm 29 2 12 Environment and cleaning SE cep Ute e teet hb e ena eee da 30 2 13 Verification TOSH ier eis IE 32 3 Over Under Checkweighing csssscccssssesccessseeeeseeesseeeseesessnaeeeesesnaeeeseeeaes 33 3 1 Specifying target values mene 33 3 2 Over Under Checkweighing Drocedume m 36 3 3 Over Under Checkweighing during subtractive welghing reee 38 3 4 Over Under Checkweighing with Quick ao 39 3 5 Over Under Checkweighing in Take away mode 40 3 6 Over Under Checkweighing To zem 4 37 Terminating Over Under Checkweighing ssse mm 42 4 Settings in the menu eeeseseeeseeseeeeeenn eene enne nennen nnne nnnm nnn nnn 43 4 Operating the men 43 4 2 Scale menu block analog scales ssssse ttt tr rrtt tt rrn tte er rnrn 46 4 3 Scale menu block IDNet scales ssssss Hmmm 49 44 Application MENU DIOCK s cuia re ree Diae de oh rta ftne M N 52 4 5 Terminal MOTI DIOCK E s os ses er endete e rer d ehe ERR et EN cus Apud 56 4 6 COMMUNICATION MENU DOCK oc cn Et e e i rr eR Ha eee Pe a 62 4 7 Maintenance Men BOCK eeii t ete t E e e e as 71 12 10 Order number 22021161A METTLER TOLEDO User manual ICS669 check 3 A 5 1 5 2 5 3 5 4 5 5 5 6 5 7 6 1 6 2 6 3 7 72 7
88. vices are available in our range of products for hazardous environments The safety of the device cannot be ensured if it is not operated in accordance with these operating instructions Only authorised personnel may open the device Devices with built in power supply unit Ensure that the power socket outlet for the device is earthed and easily accessible so that it can be de energised rapidly in emergencies Ensure that the supply voltage at the installation site lies within the range of 100 V to 240 V Check the power cable regularly for damage If it is damaged immediately disconnect the device from the power supply Ensure that there is a space of at least 3 cm 1 25 at the rear in order to prevent the power cable from being bent too strongly Devices with built in storage battery Do not use the battery charger in humid or dusty rooms or below 0 C 32 F ambient temperature After the built in storage battery has been charged the cover cap of the charging Socket at the device must be closed Terminal and platform combinations The maximum static safe load must never be exceeded Observe the operation limits See Technical data Avoid falling loads shock loads as well as impacts from the side METTLER TOLEDO User manual ICS669 check 5 1 2 Presentation Weighing terminals There are two versions of the ICS669 check weighing terminal ICS669a check with analog scale interface to connect analog METTLER TOLEDO we
89. y setting red good factory setting green and too heavy factory setting yellow 1 Specify the target values as described above 2 Place the checkweighing material on the scale Depending on the applied weight the colour of the background lighting changes Weight information is displayed in accordance with the display setting and the Over Under Checkweighing settings 36 merrier TOLEDO User manual ICS669 check Order number 22021161A 12 10 Display for Over Under Checkweighing and tolerance type Absolute With default settings for display type and colours the following is displayed for too light good too heavy EA LOW 1 64kg HIGH 2 38kq d Hi LOW 1 64kg HIGH 2 38kg dd NET LOW 1 64kg HIGH 2 38kq Display for Over Under Checkweighing and tolerance type Relative With default settings for display type and colours the following is displayed for too light good too heavy E HL T 28kg 2 10k ECKER c IB T 0 20kq EAR c bbs T 0 18kg T B 28kg CA T 6 18kg 12 10 Order number 22021161A METTLER TOLEDO User manual ICS669 check 37 38 3 3 METTLER TOLEDO User manual ICS669 check Display for Over Under Checkweighing and tolerance type Percent With default settings for display type and colours the following is displayed for too light good too heavy HUI T 5 882 2 18kg T 3 882 d H kg as T 5 88Z 2 1 Bkg T 3 882

Download Pdf Manuals

image

Related Search

Related Contents

SWIS User`s Manual 4.4  PRESSOSTATI A MEMBRANA  Philips Kala 200 User's Manual  E.ON SolarManager Funktionsstecker Bedienungsanleitung (PDF  AL25D-Series  MODE D`EMPLOI  NFC Lock User Manual (Version 2.0)  Panasonic NN-SN651W microwave  

Copyright © All rights reserved.