1. Figure 3 Vacuum Filtration Alloy 2 5 2 Required Apparatus for Vacuum Filtration r Description Unit Cat No Filter Discs glass fiber 47 mm 100 pkg 2530 00 Filter Holder membrane 47 mm each 13529 00 Flask filtering 500 mL each 546 49 Select one of the following Pump vacuum hand operated each 28248 00 Pump vacuum portable 115 VAC each 14697 00 Pump vacuum portable 230 VAC each 28248 01 Tupbing vacuum 20741 45 Tweezers each 14282 00 2 5 3 Gravity Filtration Many chemical procedures use gravity filtration This requires only filter paper a conical funnel and a receiving flask see Figure 4 Gravity filtration is better for retaining fine particles The rate of filtration increases as the volume increases in the filter cone but never fill the cone more than three quarters full Note Pretreatment using acid and heat is often necessary for metals testing Filter paper will not withstand such an environment therefore vacuum filtration with glass fiber filter discs is recommended Also glass filter discs do not retain colored species like the paper filters do To filter using gravity 1 Place a folded filter paper into the funnel Dampen the filter paper with deionized water so it adheres to the funnel 2 3 Place the funnel into an Erlenmeyer flask or graduated cylinder 4 Pour the sample into the funnel 12 Laboratory Practices Figure 4 Gravity Filtration 2 5 4 Requi
2. Section 2 Laboratory Practices 2 1 Temperature Most methods perform most accurately when the sample temperature is between 20 C and 25 C 68 to 77 F A note in the individual procedure will state any special temperature requirements 2 2 Mixing Swirling is recommended when mixing samples in a graduated cylinder or a titration flask Swirling is the gentlest method of mixing and offers the least chance for atmospheric contamination when testing for carbon dioxide and other gases 1 Grip the cylinder or flask firmly with the tips of the thumb and first two fingers Figure 1 2 Hold the cylinder at a 45 degree angle and make a circling motion from the wrist 3 Move the cylinder in an approximately 12 inch circle creating enough rotation to complete the mixing in a few turns Mixing sample in a square sample cell 1 Grasp the neck of the cell with the thumb and index finger of one hand Rest the concave bottom of the cell on the tip of the index finger of other hand 2 Mix by rotating the cell quickly one way and then in the reverse direction Inverting allows for thorough mixing in a capped sample cell or a mixing cylinder 1 Hold the cell or cylinder in a vertical position with the cap on top 2 Invert so that the cap is on the bottom Return the cell to its original position Figure 1 Repeat as needed Figure 1 Swirling a Cylinder and Inverting a Sample Cell Laboratory Practices 2 3 Digestion 2 4 Di
3. e Cells Although sample cells shipped with the spectrophotometer instrument are distortion free nicks and scratches from handling may cause an optical mismatch between two sample cells and introduce error into the test results This type of error may be avoided by optically matching the sample cells as follows Note Refer to the photometer user manual for specific steps required to select wavelengths and zero the instrument 1 Turn the instrument on Make sure Display Lock is off or the Reading mode is set to Continuous 2 Select a wavelength of 510 nm or the wavelength to be used for the test 3 Pour atleast 10 mL 25 mL for 25 mL cells of deionized water into each of two sample cells 4 Place one sample cell into the cell holder with the fill mark toward the front facing the user 5 Zero the instrument 6 Place the other sample cell into the cell holder with the fill line toward the front 7 Wait for the value to stabilize and read the absorbance Record the resulting absorbance 17 Laboratory Practices 8 Turn the cell 180 and repeat step 6 Try to achieve an absorbance value within 0 002 Abs of the first cell Note the orientation of the cell If the sample cells cannot be matched within 0 002 Abs they may still be used by compensating for the difference For example if the second cell reads 0 003 absorbance units higher than the first cell correct future readings when using these two cells by subtrac
4. e determination that uses the same reagent lot Spectrophotometer software allows the reagent blank value to 13 Laboratory Practices be stored and subtracted automatically from each sample value The reagent blank needs to be determined only at first use when a new lot of reagent has been opened or if contamination is suspected In most tests the reagent blank is so small the instrument may be zeroed on either an untreated portion of the original water sample or on deionized water This will not result in a significant loss of accuracy unless the test is for very low levels of the species of interest It is then best to use a reagent blank prepared as above 2 7 Sample Dilution 14 Most colorimetric tests use volumes of 10 and 25 mL However in some tests the color developed in the sample may be too intense to be measured due to high levels of analyte or unexpected colors may develop due to an interference In either case dilute the sample to produce a measurable endpoint or to determine if interfering substances are present To dilute the sample 1 Pipet the chosen sample portion into a clean graduated cylinder or volumetric flask for more accurate work 2 Fill the cylinder or flask to the desired volume with deionized water 3 Mix well Use the diluted sample when running the test Table 4 shows relative quantities and the multiplication factors to use with a 25 mL graduated cylinder The concentration of the samp
5. e pipet CAUTION Never pipet chemical reagent solutions or samples by mouth When filling a pipet keep the tip of the pipet below the surface of the sample as the sample is drawn into the pipet Serological pipets have marks that indicate the volume of liquid delivered by the pipet The marks may extend to the tip of the pipet or may be only on the straight portion of the tube If the marks are only on the straight part of the tube fill serological pipets to the zero mark and discharge the sample by draining the sample until the meniscus is level with the desired mark If the serological pipet has marks extending to the tip of the pipet fill the Laboratory Practices pipet to the desired volume and drain all the sample from the pipet Then blow the sample out of the pipet tip with the pipet filler for accurate measurements Volumetric transfer pipets have a bulb in the middle and a single ring above the bulb to indicate the volume of liquid when it is filled to the mark To discharge a volumetric pipet hold the tip of the pipet at a slight angle against the container wall and drain Do not discharge the solution remaining in the tip of the pipet after draining Volumetric pipets are designed to retain a small amount of sample in the pipet tip If droplets of sample cling to the walls of the pipet the pipet is dirty and is not delivering the correct amount of sample Wash the pipet thoroughly with a laboratory detergent or cleaning solut
6. ion and rinse several times with deionized water 2 12 2 The Pour Thru Cell The Pour Thru Cell is an optional accessory that improves accuracy and makes measuring more convenient for rapid liquid methods Methods that use 25 mL samples and sample cells can use the Pour Thru Cell if specified in the procedure The Pour Thru Cell cannot be used with 10 mL sample sizes and reagents Note See the photometer user manual for installation and operation instructions e Pour the solution into the funnel of the installed Pour Thru Cell Module Avoid spilling solution onto the instrument e The funnel height and orientation may be adjusted The funnel height determines the speed of sample flow through the cell The higher the funnel the faster the flow e To minimize air bubbles adjust the funnel so that it drains completely with the final level of liquid in the tube about 5 cm 2 inches below the tip of the funnel e Take instrument readings after the solution has stopped flowing through the cell e Always rinse the cell thoroughly with deionized water after each series of tests or as often as specified in the procedure Occasionally remove the Pour Thru Cell to check for any accumulation of film on the windows If the windows appear hazy soak the cell in a detergent bath and rinse thoroughly with deionized water 19 Visit us at www hach com
7. le into Snapper point first until the tip touches the ramp at the bottom of the Snapper 3 Hold the Snapper between the index and middle finger like a syringe With the ampule tip down lower the Snapper into the sample until the ampule shoulder is wet 4 Press on the flat end of ampule with the thumb as if depressing the plunger on a syringe until the tip snaps Allow the ampule to fill before removing from sample 5 Rinse the wet end of the Snapper and ampule with clean water if desired Remove the ampule from Snapper 6 Knock the ampule tip retained in the Snapper into a suitable waste receptacle 2 9 PermaChem Pillows Permachem pillows utilize powdered reagents to minimize deterioration and the risk of reagent spills To use PermaChem pillows Figure 6 1 Tap the pillow on a hard surface to collect the powdered reagent in the bottom 2 Locate the tear notch and tear or cut across holding the pillow away from the face 3 Using two hands push both sides toward each other to form a spout 4 Pour the pillow contents into the sample cell and continue the procedure according to the instructions Figure 6 Opening PermaChem Pillows re 2 Tear 3 Push 4 Pour 16 Laboratory Practices 2 10 Sample Cells The same solution in both cells will give the same absorbance within 0 002 Abs for properly matched cells For more information see Matching Sample Cells on page 17 For accurate results use only the
8. le is equal to the diluted sample result multiplied by the multiplication factor Table 4 Sample Dilution Volumes Sample mL Deionized Water p Used to Bring the Volume Multiplication Factor Volume mL to 25 mL 25 0 0 0 12 5 12 5 2 10 01 15 0 2 5 5 01 20 0 5 2 51 22 5 10 1 01 24 0 25 0 2501 24 75 100 1 For sample sizes of 10 mL or less use a pipet to measure the sample into the graduated cylinder or volumetric flask Laboratory Practices More accurate dilutions can be done with a pipet and a 100 mL volumetric flask Table 5 1 Pipet the sample and dilute to volume with deionized water 2 Stopper and invert to mix Table 5 Multiplication Factors for Diluting to 100 mL Sample Volume mL Multiplication Factor 1 100 50 20 10 10 25 4 50 2 2 7 1 Sample Dilution with Interfering Substances Sample dilution may influence the level at which a substance interferes The effect of the interferences decreases as the dilution increases In other words higher levels of an interfering substance can be tolerated in the original sample if it is diluted before analysis An Example Copper does not interfere at or below 100 mg L for a 25 mL sample in a procedure If the sample volume is diluted with an equal volume of water what is the level at which copper will not interfere Total volume Sample volume 25 3 12 5 Dilution factor Interference Le
9. nitrogen ammonia nitrogen phenols selenium and volatile acids The General Purpose Heater and Support Apparatus provides efficient heating and anchoring of the glassware Laboratory Practices Figure 2 General Purpose Distillation Apparatus 2 5 Filtration Filtration separates particulates from an aqueous sample It uses a porous medium that retains particulates but allows liquids to pass through It is useful for removing turbidity which may interfere in colorimetric analyses from water samples Two methods of filtration are most frequently used vacuum filtration and gravity filtration 2 5 1 Vacuum Filtration Vacuum filtration uses both suction and gravity to draw the liquid through the filter An aspirator or vacuum pump is used to create suction see Figure 3 Vacuum filtration is faster than gravity filtration alone To filter using a vacuum 1 2 Use tweezers to place a filter paper into the filter holder Place the filter holder assembly in the filtering flask Dampen the filter with deionized water to ensure adhesion to the holder Position the funnel housing on the filter holder assembly While applying a vacuum to the filtering flask transfer the sample to the filtering apparatus When the filtration is complete slowly release the vacuum from the filtering flask and transfer the solution from the filter flask to another container 11 Laboratory Practices
10. red Apparatus for Gravity Filtration Description Unit Cat No Cylinder graduated 100 mL each 508 42 Funnel poly 65 mm each 1083 67 Filter Paper 12 5 cm pleated 100 pkg 1894 57 Flask erlenmeyer 125 mL each 505 43 2 6 Reagents 2 6 1 Reagent and Standard Stability It is always good laboratory practice to date chemicals upon receipt and to rotate supplies so the older supplies are used first Absorption of moisture carbon dioxide or other gases from the atmosphere bacterial action high temperatures or light with photosensitive compounds may affect the reagent shelf life In some cases reaction with the storage container or interaction of reagent components may occur 2 6 2 Reagent Blank In several tests the contribution of the reagent s to the final reading is of such a magnitude that it must be compensated for whenever the test is performed This produces a positive error in the test results Every effort is made to produce reagents with the lowest possible blank and for most reagents it is less than 0 009 absorbance units However it is sometimes impossible or impractical to produce reagents with such a low blank When using such reagents it is best to determine the reagent blank by performing the procedure using high quality water deionized distilled etc to zero the instrument The resulting value is then expressed in the concentration units of the test and is subtracted from each sampl
11. sample cells specified in each procedure Due to differences in cell pathlengths sample cell substitution will introduce bias in test results For example 1 inch square cells have a pathlength approximately 8 longer than 1 inch round cells substitution of round cells for square cells will introduce a bias in the reading 2 10 1 Orientation of Sample Cells To minimize measurement variability when using a particular cell always orient the cell in the same manner before placing it into the cell holder The fill marks on the cells can be used as orientation guides for positioning the cells 2 10 2 Maintaining Sample Cells When not in use store the sample cells in their boxes to protect them from scratches and breakage After use empty and clean the cells avoid leaving colored solutions in the cells for extended periods of time 2 10 3 Cleaning Sample Cells Most laboratory detergents can be used at recommended concentrations Neutral detergents such as Liquinox are safer when regular cleaning is required Cleaning times are reduced by increasing the temperature or by using an ultrasonic bath Finish by rinsing a few times with deionized water and allow to air dry Cells may also be cleaned with acid followed by rinsing thoroughly with deionized water Individual procedures may require special cleaning methods When using a brush to clean cells take extra care to avoid scratching the inner surfaces of the cells 2 10 4 Matching Sampl
12. stillation 10 Several procedures require sample digestion Digestion uses chemicals and heat to break down a substance into components that can be analyzed This section briefly describes three different digestion procedures The Digesdahl system is a process that yields a digest suitable for the determination of metals total phosphorus and total Kjeldahl nitrogen TKN It is rapid convenient and is very effective at destroying interfering organic materials For USEPA reporting purposes USEPA approved digestions are required USEPA presents two digestions mild and vigorous for metals analysis Other digestion procedures are required for mercury arsenic phosphorus and TKN See Section 3 for more information on sample digestion Distillation is an effective easy and safe method of separating some chemical components for analysis The following equipment is recommended for distillation e General Purpose Distillation Apparatus Cat No 22653 00 shown in Figure 2 e Arsenic Distillation Apparatus Set Cat No 22654 00 e Cyanide Distillation Apparatus Set Cat No 22658 00 e General Purpose Heater and Support Apparatus Cat No 22744 00 115 VAC 60 Hz e General Purpose Heater and Support Apparatus Cat No 22744 02 230 VAC 50 Hz The Distillation Apparatus is suitable for water and wastewater that requires sample pretreatment by distillation Applications for the General Purpose Apparatus include fluoride alouminoid
13. ting 0 003 absorbance units or the equivalent concentration from the reading Likewise if the second cell reads 0 003 absorbance units add 0 003 absorbance units to the reading 2 11 Other Apparatus 2 11 1 Boiling Aids Boiling is required in some procedures Under some conditions bumping may occur causing sample loss or injury Bumping is caused by the sudden almost explosive conversion of water to steam as it is heated Use of a boiling aid such as Boiling Chips Cat No 14835 31 reduces bumping Make sure the boiling aids will not contaminate the sample Do not use boiling aids except glass beads Cat No 2596 00 more than once Use a large enough flask or beaker to allow significant headspace above the solution Loosely covering the sample during boiling will prevent splashing reduce the chances of contamination and minimize sample loss Individual procedures will recommend the specific boiling aid to use 2 12 Achieving Accuracy in Measurement 2 12 1 Pipets and Graduated Cylinders Figure 7 18 When smaller sample quantities are used the accuracy of measurements becomes increasingly important Figure 7 illustrates the proper way to read the sample level using the meniscus formed when the liquid wets the cylinder or pipet walls Reading the Meniscus Before filling rinse the pipet or cylinder two or three times with the sample to be tested Use a pipet filler or pipet bulb to draw the sample into th
14. vel x Dilution Factor Interference level in sample 100 x 2 200 The level at which copper will not interfere in the diluted sample is at or below 200 mg L 2 8 AccuVac Ampuls AccuVac Ampuls contain pre measured powder or liquid vacuum packed in optical quality glass ampules To use AccuVac Ampuls 1 Collect the sample in a beaker or other open container 2 Use one of the following methods to break the tip off the ampule e Use the optional AccuVac Snapper Cat No 24052 00 See Using the AccuVac Snapper for instructions or e Place the ampule tip well below the sample surface and break the tip off Figure 5 against the beaker wall The break must be far enough below the surface to prevent air from being drawn in as the level of the sample drops 3 Secure an ampule cap over the tip of the ampule Invert the ampule several times to dissolve the reagent The cap protects from broken glass and provides a grip for 15 Laboratory Practices inserting and removing the ampul from the cell holder Wipe the ampule with a lint free cloth to remove fingerprints Note Without the cap the liquid will stay in the ampule when inverted DO NOT place fingers over broken glass 4 Insert the ampule into the sample cell holder and read the results directly Figure 5 Using AccuVac Ampuls t 7 ia 2 8 1 Using the AccuVac Snapper 3 1 Hold the Snapper with the open end up 2 Gently slip the ampu