ppm pDS Manual - Myron L Company
Contents
1. BATTERY REPLACEMENT To replace the battery detach the battery connector Pull on the vinyl strap to remove the battery Replace with fresh zinc carbon or alkaline 9 volt battery Reinsert the vinyl strap to secure battery CELL CUP Self conditioning of the built in electrodes occurs each time the button is pressed with a sample in the cell cup This ensures consistent results each time With some samples a small downward swing of the pointer is a result of this conditioning action This action is powerful and removes normal films of oil and dirt However if very dirty samples particularly scaling types are allowed to dry in the cell cup a film will build up This film reduces accuracy When there are visible films of oil dirt or scale in the cell cup or on the electrodes scrub them lightly with a small brush and household cleanser Rinse out the cleanser and the meter is ready for accurate measurements pH SENSOR The unique pH electrode in your pDS meter is a non refillable combination type which features a porous Teflon liquid junction covered by U S Patent No 4128468 It should not be allowed to dry out see pH MEASUREMENT If it does the sensor can sometimes be renewed by soaking in a saturated potassium chloride KCI solution for several days Drifting can be caused by a film on the sensor bulb Use a liquid cleaner such as Windex or Fantastic to clean it The sensor bulb is very thin and delicate Excessive pre
2. To order accessories contact your nearest stocking distributor or the Myron L Company WARRANTY amp SERVICE Myron L pDS meters excluding the pH sensor have a two year warranty The pH sensor has a 6 month limited warranty If an instrument fails to operate properly check the batteries calibrations and cleanliness of the pH sensor bulb If it still fails to properly function return it prepaid to the Myron L Company If in the opinion of the factory failure was due to materials or workmanship repair or replacement will be made without charge A reasonable service charge will be made for diagnosis or repairs due to normal wear abuse or tampering This warranty is limited to the repair or replacement of the Myron L pDS meter only The Myron L Company assumes no other responsibility or liability 2450 Impala Drive Carlsbad CA 92010 7226 USA Tel 760 438 2021 Fax 800 869 7668 760 931 9189 www myronl com Myron L Company 2001 W
3. CE below 4 Replace the bottom cover RANGE DOUBLING 1 Using procedures described above adjust the INTERNAL STANDARD to the value shown on the bottom label 2 Test sample Multiply reading by two For example Indicated value of 3200 ppm would actually be 6400 ppm 3200 x 2 S position Adjust Master Calibration to 12 Internal Standard value Model T2 pH shown NOTE After completing tests requiring range doubling reset the Master Calibration Control to its INTERNAL STANDARD value and replace bottom cover pH STANDARD BUFFER SOLUTIONS All pH sensors drift and decrease in strength as they age pH meters have two Calibration Controls to compensate for those effects ZERO adjust corrects for drift of the pH sensor s response to neutral pH solutlons GAIN adjust amplifies the signal from the pH sensor for accurate readings of other pH values Use pH standard buffer solutions to calibrate all pDS meters The recommended values are 4 7 and 10 For greatest accuracy adjust with the buffer closest in pH to the samples to be tested pH ZERO ADJUST Frequency weekly Test by rinsing and refilling the cell with 7 buffer and pressing the red button The meter should indicate 7 on the red scale If it is necessary to calibrate Zero perform the following steps 1 Remove the bottom cover using fingernails or a small screwdriver to loosen the front or rear edge 2 Refill cell with 7 buffer see pH MEASUREMENT for p
4. H 6 Compare the meter reading to the INTERNAL STANDARD value on the meter s bottom label If they are noticeably different mark the new value in place of the old one Conductivity PH Sensor Master Calibration Control X1 000 NS Battery Strap X10 Trimmer A A A AL T N lt pH Gain X100 Trimmer E y aiia l pH Zero NOTE The feature described below is intended as a quick field calibration check or for using the range doubling technique It is not a replacement for calibration with Standard Solution INTERNAL STANDARD An INTERNAL STANDARD value for each meter is on the label on the bottom cover of each meter Use it between normal calibrations as a field check of your meter s accuracy To verify your meter s calibration 1 Turn the Range Switch to S and press the black button T2 pH or press the green button T6 pH 2 Ifthe reading matches the INTERNAL STANDARD value on the bottom label your meter is in calibration If not see TDS CALIBRATION above FIELD CALIBRATION MAINTENANCE 1 Turn the range switch to S T2 pH only 2 Remove the bottom cover using fingernails or a small screwdriver to loosen the front or rear edge 3 Press the black button T2 pH or green button T6 pH and adjust the Master Calibration Control until the meter reading is the same as the INTERNAL STANDARD value NOTE If the meter will not adjust to the INTERNAL STANDARD value the battery should be replaced See MAINTENAN
5. LITY Both Models 1 By using an RE 10 Range Extender see ACCESSORIES the maximum TDS range of each meter may be increased ten times The pictures in this manual show the major operating parts of your Myron LpDS Meter Handle your pDS Meter and identify these parts to become familiar with it Model T2 pH shown Polyethylene Cell Cup N DH Compartment pper Conductivity Electrode Meter Zero Adjust Pointer Conductivity Range Selector Switch X ee Black Conductivity Red pH Button Button USING YOUR METER TDS MEASUREMENT Model T2 pH 1 Turn the Range Switch to the desired range If you don t know which range to use set it to 1000 2 Rinse the cell cup three times with the sample you want to test For very hot or very cold samples see TEMPERATURE COMPENSATION below NEVER FILL CELL BY DIPPING METER INTO WATER 3 Fill the cell with another sample to at least s 6mm above the upper electrode 4 Press the black button 5 Read the dial s black scale value indicated by the pointer If the pointer is very low on the left or goes off the scale to the right try the next lower or higher range 6 Multiply the dial s black scale value by the range setting to determine TDS in parts per million 7 Ifthe pointer goes off the scale to the right on the x1000 range try RANGE DOUBLING see FIELD CALIBRATION or use an RE 10 range extender see ACCESSORIES TDS MEASUREMENT Model T6 pH 1 Rin
6. PPM pDS METERS Operation Instructions User Manual for Models T2 pH 16 pH 2450 Impala Drive Carlsbad CA 92010 7226 USA Tel 760 438 2021 MYRON L Fax 800 869 7668 760 931 9189 COMPANY pH Conductivity Instrumentation WWW myronl Com Accuracy Reliability Simplicity DESCRIPTION This manual describes your Myron L pDS meter tells you how to use it and how to keep it working accurately for many years Myron L pDS Meters are compact light weight self contained conductivity oH meters They contain a built in sample cell with a replaceable pH sensor Models T2 pH and T6 pH convert conductivity values directly to Total Dissolved Solids TDS and have dials calibrated in parts per million and in pH units They will quickly determine the TDS and pH of almost any solution Both are 3 4 x 4 5 x 4 0 85 x 129 x 126mm and weigh less than one pound 45 kg pDS Meters are completely self contained The built in cell is automatically temperature compensated from 50 to 160 F 10 to 71 C They are powered by a 9 volt transistor battery good for at least 2000 tests or one year shelf life MODEL PARTS PER MILLION PPM pH RANGE RANGES T2 pH 0 50 0 500 0 5000 ppm TDS 2 12 pH T6 pH 0 5000 ppm TDS 2 12 pH ACCURACY Both Models TDS 2 of full scale ppm NaCl standard only Ground water 442 standard values are approximate because of varying sample TDS compositions see ACCESSORIES pH 0 2 pH units REPEATABI
7. e concentrations of the reference solutions are calculated from data in the International Critical Tables Vol 6 pH STANDARD BUFFER SOLUTIONS Ready to use Myron L Company pH buffer solutions are mold inhibited and accurate to within 0 06 pH units 25 C They are traceable to National Bureau of Standards certified pH references and are color coded for instant identification RECOMMENDED STANDARD SOLUTIONS MODEL RANGE SOLUTION NUMBER T2 pH TOTAL DISSOLVED SOLIDS X 10 442 30 X 100 442 300 X 1000 442 3000 with Re 10 442 30 000 pH ZERO 7 buffer pH GAI N 4 buffer 10 buffer T6 pH TOTAL DISSOLVED SOLIDS 442 3000 with RE 10 442 30 000 pH ZERO 7 buffer pH GAIN 4 buffer 10 buffer RANGE EXTENDER The RE 10 Range Extender is a useful accessory for testing high conductivity solutions beyond the normal range of your pDS meter Inserting the RE 10 into a sample filled cell cup increases the maximum range ten times Use the Range Extender whenever the reading is off the scale USING THE RE 10 1 Fill the cell cup three times to rinse it but each time insert the Range Extender to rinse it also Fill the cell cup with your sample Push the Range Extender into the cell cup seating the O ring seal Use and read the pDS Meter in the normal manner Multiply the reading by 10 For best accuracy repeat the complete test with a fresh sample When finished remove the Range Extender Thoroughly rinse the cell cup and Extender with clean wat
8. er preferably distilled or deionized to eliminate dried salts build up This is extremely important when the instrument will be used to test high purity water MYRON L COMPANY Printed in U S A ae O N pH Conductivity Instrumentation Accuracy Reliability e Simplicity Specifications subject to change without notice RE 10 Range Extender nara ere THE RE 10 Calibrate your meter without RE 10 using Standard Solution number 442 3000 Fill cell cup with 442 30 000 Standard Solution Insert RE 10and pressthe button Multiply the reading by 10 and compare it with the value on the Standard Solution label If they are not the same the RE 10 must be recalibrated NOTE DO NOT adjust the Conductivity Calibration Control 4 Adjust the white insert of the Extender as follows If the reading is too high push or tap inward If the reading is too low twist or pull outward with pliers NOTE The Range Extender is factory calibrated to a particular meter It should be recalibrated if it is to be used with another meter w N PORTA PAK Carrying Case for use with all Myron L DS and pDS meters is foam lined and molded of sturdy ABS plastic REPLACEMENT pH SENSORS Order Model RPY for either T2 pH or T6 pH meters When ordering be sure to include the model and serial number of your meter to ensure receiving the proper type Complete installation instructions are provided with each replacement sensor ORDERING
9. nds then immediately fill the cell with the sample you want to test step 3 above This allows the automatic temperature compensation feature time to work properly CAUTION ALWAYS store your pH sensor wet and with the cap on ALWAYS rinse the cell cup with pure water immediately after use or cleaning This will prevent the build up of deposits and possible contamination of the sample DO NOT test solutions which could damage the pH sensor See MAINTENANCE DO NOT use with samples hotter than 160 F 71 C The readings WILL NOT be accurate DO NOT splash solvents such as lacquer thinner acetone benzene or chlorinated solvents on the plastic case DO NOT fix or modify the meter That will void your warranty See SERVICE for details or consult Myron L DO NOT DIP THE INSTRUMENT INTO WATER If water does get inside the instrument see MAINTENANCE for instructions on drying it TDS CALIBRATION TDS STANDARD SOLUTIONS A Standard Solution has a known conductivity and ppm TDS Your meter was calibrated at the factory using Standard Solutions You can keep your meter accurate by using these same Standard Solutions 442 Standard Solutions approximate the conductivity characteristics of most natural waters To pick the right Standard Solution for your meter see ACCESSORIES How often you calibrate your meter depends on how much you use it For once a day use recalibrate it every three months CHECKING TDS CALIBRATION 1 Turn Range Swi
10. roper procedure 3 Press the red button and turn the pH Zero control until the needle indicates 7 4 Recap the pH compartment and replace the bottom cover pH GAIN SLOPE ADJUST Frequency bi monthly The Gain should be calibrated approximately once every eight weeks The Gain control is used with 4 and 10 buffers NOTE Never adjust the Gain control without first adjusting Zero Test by rinsing and refilling the cell with 4 or 10 buffer and pressing the red button The meter should indicate the buffer value on the red scale If it is necessary to calibrate the Gain perform the following steps 1 Remove the bottom cover 2 Rinse and fill the pH compartment with 4 buffer and press the red button Adjust the Gain control until the needle indicates 4 3 Rinse three times and test with 10 buffer If adjustment is necessary adjust to only half way between the initial reading and 10 If after this the 4 and 10 buffer readings are not less than one division from 4 and 10 the sensor is deteriorating and should be replaced The sensor should also be replaced if the Gain control cannot make the pointer reach either 4 or 10 MAINTENANCE BATTERYCHECK Turn the Range Switch to S and press the black button T2 pH or press the green button T6 pH If the pointer does not reach the INTERNAL STANDARD value readjust the S see TDS CALIBRATION If the pointer will not reach the INTERNAL STANDARD value the battery should be replaced
11. ry with an absorbent cloth or tissue Avoid pushing any water into the Calibration Controls or the switches 3 Air dry the meter in a warm area with the bottom cover off Allow several hours for thorough drying lf the water entered through a leak in the case or cell or if the instrument shows erratic readings or other unusual behavior return it to the Myron L Company for servicing ACCESSORIES STANDARD SOLUTIONS TDS STANDARD SOLUTIONS Your pDS Meter has been factory calibrated with the appropriate Standard Solution All Myron L conductivity Standard Solution bottle labels show three values ppm 442 pom sodium chloride and conductivity in micromhos 442 Unless otherwise specified the calibration used for all DS Meters is the 442 standard The 442 Standard Solutions consist of the following salt ratios 40 sodium sulfate 40 sodium bicarbonate and 20 sodium chloride This salt ratio has conductivity characteristics approximating natural waters and was developed by the Myron L Company over two decades ago Sodium Chloride For every ppm 442 Standard Solution there is a ppm sodium chloride NaCl solution which wili have the same conductivity The parts per million of the equivalent NaCl solution is on each Standard Solution label Instruments calibrated to NaCl standards are set using equivalent NaCl values Conductivity All Myron LCompany Conductivity Standard Solutions are within 1 0 of Potassium Chloride reference solutions Th
12. se and fill the cell cup as in steps 2 and 3 above 2 Press the black button 3 Read the TDS value in parts per million indicated by the pointer on the dial s black scale 4 Ifthe pointer goes off the scale to the right try RANGE DOUBLING see FIELD CALIBRATION or use an RE10 range extender see ACCESSORIES NOTE When finished RINSE THE CELL CUP with clean water preferably distilled or deionized pH MEASUREMENT 1 Obtain sample to be tested if necessary heat or cool it to between 50 and 160 F 10 71 C 2 Remove the pH compartment cap and sharply shake out any liquid 3 Pour or squirt the sample into the built in cell cup especially the small pH compartment Shake out and repeat three times with the sample to be tested These may be the same rinsings used to condition the cell for the TDS measurement 4 Press the red button The pointer immediately indicates the OH of the sample on the red scale If the pointer drifts wait approximately 30 seconds and press the button again to obtain a stable reading of pH If drifting continues the electrode may be coated with a film Use a liquid cleaner such as Windex or Fantastic clean the pH sensor bulb see pH SENSOR 5 Rinse both pH and conductivity parts of the cell with clean water LEAVE SOME LIOUID IN AND RECAP THE pH COMPARTMENT TEMPERATURE COMPENSATION For very hot or very cold solutions let the three rinse samples each remain in the cell for several seco
13. ssure during cleaning may break it Leaving high pH alkaline solutions in contact with the pH sensor for long periods of time can damage it Rinsing such liquids from the pH compartment and moistening it with 4 buffer or tap water will extend its useful life Samples containing chlorine sulphur or ammonia can poison any pH electrode If it is necessary to measure the pH of any such sample thoroughly rinse the pH sensor with clean water immediately after taking the measurement Any sample element which will reduce add an electron to silver such as cyanide will attack the reference electrode Replacement sensors are available only from the Myron L Company or our authorized distributors see ACCESSORIES DuPont Company WATER INSIDE THE METER Your Myron L meter is a rugged instrument and will withstand water exposure around its cell meter movement and switches However care should be taken to keep water from leaking in around the bottom cover It is not sealed to prevent condensation from forming lf the water is relatively clean i e tap water or better and there are only a few drops inside the meter dry it as described below Large amounts of water or corrosive or very dirty solutions will almost certainly damage the meter movement or electronics Such meters should be returned to the Myron L Company for repair To dry your meter 1 Shake excess water out of the inside of the meter 2 Dab the exposed surfaces d
14. tch to the x1000 range T2 pH only 2 Test a sample of appropriate Standard Solution CAUTION Throw away the Standard Solution as you use it Don t put the used samples back in the bottle 3 If the pDS Meter does not indicate the same value as is on the Standard Solution bottle s label first clean the cell For directions on how to do this see CELL CUP inside Rinse the cell thoroughly and test the Standard Solution again If the pDS Meter still does not indicate the correct value recalibrate it as described below TO RECALIBRATE TDS 1 Remove the bottom cover using fingernails or a small screwdriver to loosen the front or rear edge Identify the TDS Calibration Control s so you can find them by touch while calibrating Model T2 pH has three see photo below 2 Test another sample of the Standard Solution be careful to not splash solution inside the meter 3 Adjust the Master Calibration Control until the meter indicates the value on the Standard Solution label NOTE T2 pH and T6 pH Meters may be calibrated to read either 442 or NaCl parts per million but not micromhos 4 The Master Calibration Control on model T2 pH adjusts ail ranges For best accuracy in the lower ranges get the recommended Standard Solutions for the x100 and x10 ranges Use the x100 and x10 trimmers to recalibrate those ranges as in steps 2 and 3 above 5 Turn the Range Switch to S and press the black button T2 pH or press the green button T6 p
Download Pdf Manuals
Related Search
Related Contents
Sandrine Audry - Saint-Quentin-en SAシリーズ パンプレットPDFは、 こちら Mounting RAID box Q3.F22 Light Manual backup - F-22 Lightning 3 Online Community TSSBTR926B & TSSBTR926W USER`S MANUAL Multiquip MQ-D306HA Automobile Parts User Manual Rosco Diagnostica - Medica-Tec 2C Frame IT Cinema Flocked Frame Screen Installation Manual Copyright © All rights reserved.