Home
GX10/GX20 Paperless Recorder (Panel mount type)
Contents
1. While memory is stopped Immediately change the time DST The date time for switching between standard time and DST can be specified Time zone Sets the time difference from GMT Date format Select YYYY MM DD MM DD YYYY DD MM YYYY or DD MM YYYY MM expression can be selected from the numeric character or ellipsis Ex January 01 or Jan The delimiter can be selected from Ethernet Communication Functions Electrical specifications Conforms to IEEE 802 3 Connection Ethernet 10BASE T 100BASE TX Max segment length 100 m Max connecting configuration Cascade Max 4 level 1OBASE T Max 2 level 100OBASE TX Connector RJ 45 Protocols TCP UDP IP ICMP ARP DHCP HTTP FTP SMTP SNTP Modbus and dedicated protocols E mail client Automatically send e mail at specified times E mail is sent by events as below Alarm occurring alarm canceling Recover from power failure Report data generating Storage medium error FTP client function error Specified time period POP before SMTP and SMTP authentication PLAIN and CRAM MD85 is available FTP client Automatically transfer data files to the FTP server Applicable files Display data event data screen image data report data etc FTP Server Transfer files delete files manipulate directories and output file lists of the GX Number of the simultaneous connection Max 4 Web server GX real time monitoring and setting changes
2. General GX10 6x20 SMARTTVG Specifications Paperless Recorder Panel mount type GS 04L51B01 01EN m OVERVIEW The GX10 GX20 are paperless recorders that display real time measured data on a touch screen and save data on an external storage medium SD card For the input output module specification please see GX90XA GX90XD GX90YD GX90WD Input Output Module General Specification GS 04L53B01 01EN e The number of analog input is a maximum of 100 channels in GX10 with expandable I O anda maximum of 450 channels in GX20 with expandable I O Max number of input output channels is 500 channels e The GX10 GX20 have the large internal memory GX10 GX20 standard type 500MB GX20 large memory type 1 2 GB and prolonged record and GX10 GX20 preservation are possible e As the input signal a DC voltage thermocouple resistance temperature detector DI DCV input e The measuring accuracies noted in the general TTL contact signal or mA DC current can be specifications have a margin of error that takes set to each channel into account the product s components and the e Input and output have module structure and it can equipment used for adjustment and testing extend them easily GX10 max 3 GX20 max However the actual values calculated from the 10 accuracy testing data upon shipment of the In case of using expandable I O GX10 can instrument from the factory are as follows extend two units GX20 can exte
3. 12 Large memory type can be specified only for the GX20 13 To connect an expandable I O you will need one expansion module for the GX 14 IMC option must be separately specified when the WT communication is selected Analog input module Digital I O module sold separately MODEL and SUFFIX Code GX90XA MODEL and SUFFIX Code GX90XD MODEL and SUFFIX Code GX90YD MODEL and SUFFIX Code GX90WD Please see GX90XA GX90XD GX90YD GX9YOWD Input Output Module General Specification GS 04L53B01 01EN When connecting the Expandable I O Please see GX60 Expandable I O GX90EX Expansion Module General Specification GS 04L53B00 01EN E Standard Accessories E Optional Accessories Sold Separately Product opto Shunt resister for M3 terminal 10 Q 0 1 Mounting bracket for GX10 or GX20 SD memory card 1GB Stylus pen touch pen Tag sheet for GX10 or GX20 Sheet paper for GX10 or GX20 Shunt resister for M3 terminal 100 Q 0 1 X010 100 3 Shunt resister for M3 terminal 250 Q 0 1 X010 250 3 Shunt resister for Clamp terminal 10 Q 0 1 438922 Shunt resister for Clamp terminal 100 Q 0 1 438921 Shunt resister for Clamp terminal 250 Q 0 1 438920 Validation Documents For AS option 1 773230 a Provision of Validation Documents Alicense sheet containing the license key required for installation is provided Download the validation document from the following URL http www s
4. UL 61010 2 030 CSA NRTL C CE EMC directive EN61326 1 compliance Class A Table 2 EN61000 3 2 compliance EN61000 3 3 compliance EN55011 Class A Group 1 Low voltage directive EN61010 1 EN 61010 2 030 Installation category II Pollution degree 2 Measurement category e EMC Regulatory Arrangement in Australia and New Zealand RCM EN55011 compliance Class A Group 1 e KC marking Electromagnetic wave interference prevention standard electromagnetic wave protection standard compliance 4 Installation category overvoltage category II Describes a number which defines a transient overvoltage condition Implies the regulation for impulse withstand voltage IP applies to electrical equipment which is supplied from the fixed installation like a distribution board 2 Pollution degree 2 Describes the degree to which a solid liquid or gas which deteriorates dielectric strength or surface resistivity is adhering 2 applies to normal indoor atmosphere Normally only non conductive pollution occurs 3 Measurement category Depends on the specification of each modules All Rights Reserved Copyright 2012 2014 Yokogawa Electric Corporation Category Measurement Description category CAT II Available in the testing Appliances and measuring circuits portable directly connected equipment to a usage location etc receptacle or the like of a low voltage main power supply facility CAT III A
5. and alarm point marks can be displayed on the scale Moving scale Scale can be moved on any waveform A bitmap image scale can be attached Others Grid Auto 4 to 12 Trip line Message Zone display Partial expanded display Historical trend display T Y display Redisplays the display data or event data in the internal memory or external storage medium Time axis operation The time axis can be reduced or expanded Data search Waveforms from the internal memory can be displayed through the specification of a date and time calendar each summary Moving scale A bitmap image scale can be attached All historical trends can be displayed in one screen Bar graph display Direction Vertical or horizontal Scale Display a scale for each channel Color scale band and alarm point marks can be displayed on the scale Digital display Displays measured values numerically A DI input state can be displayed as an arbitrary character string O Off 1 On etc Update rate 0 5s Overview display Display format All channels each groups Displays the measured values of all channels and the alarm statuses Alarm summary display Displays a log of up to 1000 alarms Specify an alarm with the cursor and jump to the corresponding section on the historical trend display Message summary display Time and content of up to 500 messages including 50 add messages Specify a message with the cursor and jump to the corresponding section on the histori
6. 04L51B01 01EN May 7 2014 00 e Client function Supported clients FTP client and SMTP client Port number 443 when encryption is used Trusted certificate Trusted certificates a total of up to 100 KB can be saved in the internal memory m HARDWARE SPECIFICATIONS MAIN UNIT Display Display unit ee F GX10 5 7 inch TFT color LCD 640 x 480 dots Electronic Signature Function GX20 12 1 inch TFT color LCD 800 x 600 dots Electronic signatures can be added to report files A small number of missing or steady on LCD created in PDF format using the PDF form creation function An electronic signature is provided each time a report file is created Certificate for electronic signature Certificates for electronic signatures created by users can be saved in the internal memory Other Functions All Rights Reserved Copyright 2012 2014 Yokogawa Electric Corporation Buzzer GX makes a buzzer sound at touch screen operation or when alarm occurs Backlight saver function Dim or turn off the LCD backlight if there is no key operation for a specified time Favorite display Register frequently used displays to the Favorite and show them through simple operation The main alarm is indicated using the MENU key LED No alarm Blue same condition as power on Alarm condition Red User function feature A button user function key to which the user can assign a desired function is provided It can be assigned to an
7. 6 1 O module slot 5 ma T 1 a ABs MTR MOT OT El 7 1 O module slot 4 l 1 O module slot 3 i I O module slot 2 Hi I O module slot 1 ste 1 O module slot 0 co Bice 2 amp Eile lall w Power supply terminal RS 422A 485 terminal C3 or RS232 terminal C2 USB port UH Ethernet port GS 04L51B01 01EN VGA output terminal D5 FAIL STATUS terminal FL May 7 2014 00 m SPECIFICATIONS OF OPTIONAL FUNCTIONS ADVANCED SECURITY FUNCTION AS Security and electronic record signature functions have been added that are compliant with the USA s FDA title 21 CFR Part 11 e Enabling disabling the advanced security function You can enable or disable the advanced security function The set values and data stored in the internal memory are initialized each time the function is enabled or disabled e Data anti tamper function Settings and measured data are saved as encrypted binary files e Data type Only for display or event Trigger mode is not possible with event data e Login function Using the login function described below you can enter security settings on the instrument e User name Password User ID depend on setting e User level and number of users System administrator all can be operated General user With user restrictions you can set restrictions on performing operations and sign authority Monitor user monitori
8. Corporation 10 Scalable range e Log input 1 00E 15 to 9 99E 15 15decade maximum Scale_L lt Scale_U If the lower limit mantissa is 1 00 the difference between the exponents must be 1 or more If the lower limit mantissa is a value other than 1 00 the difference between the exponents must be 2 or more e Pseudo Log Input Log linear input 1 00E 15 to 9 99E 15 15decade maximum The higher limit mantissa is the same as the lower limit mantissa If the lower limit mantissa is 1 00 the value must be between 1 00E 15 and 1 00E 15 the difference between the exponents must be 1 or more and the maximum decades is 15 If the lower limit mantissa is a value other than 1 00 the value must be between 1 01E 15 and 9 99E 15 the difference between the exponents must be 1 or more and the maximum decades is 15 Alarm type High limit low limit delay high limit and delay low limit Alarm setting range The range converted into the LOG scale corresponding to 5 to 105 of the span width Alarm hysteresis Fixed to 0 Green band setting range The lower limit to the upper limit of the scale However the lower limit of the display position must be smaller than the upper limit Position of the decimal point 1 to 2 Misc Nonlinear input is possible by correcting the input value MATHEMATICAL FUNCTIONS WITH REPORT FUNCTION MT Mathematical Function Number of math channels GX10 50 GX20 100 Operation General arith
9. FUNCTIONS MC e Number of communication channels GX10 standard type 300 C001 to C300 GX20 large memory type 500 C001 to C500 Required to record data from Modbus devices or PCs on the GX 24 V DC AC POWER SUPPLY P1 e Rated power supply 24 V DC 24 V AC 50 60 Hz e Allowable power supply voltage range 21 6 to 26 4 V DC AC e Insulation resistance 20 MQ or greater at 500 V DC between power terminal and earth e Withstand voltage 500 V AC at 50 60 Hz for one minute between power terminal and earth e Rated power supply frequency for AC 50 60 Hz e Alowwable power supply frequency for AC 50 Hz 2 60 Hz 2 e Influence of power supply voltage variation Variations of the measured value must be within 1 digit in the range from 21 6 to 26 4 VAC VDC e Influence of power supply frequency variation for AC Variations of the measured value must be within 0 1 of rdg 1 digit at a rated frequency of 2 Hz e Power consumption Supply LCD Normal voltage backlight off operation 24VDC GX10 7VA GX10 9 VA GX10 24 VA GX20 13 VA GX20 16VA GX20 48 VA 24VAC GX10 14 VA GX10 17 VA GX10 42 VA GX20 25 VA GX20 29 VA GX20 76 VA The following combinations are assumed for LCD backlight off and normal operation GX10 1 Al module 1 DO module 1 DI module GX20 5 Al modules 4 DO modules 1 DI module e Module power supply voltage The total allowable power consumption of respective
10. If you want to use more than 50 channels please purchase the I O modules individually To connect expandable I O GX60 to GX10 or GX20 you must purchase one unit of expansion module GX90EX for GX10 GX20 in addition to the expandable I O Trademarks The TCP IP software used in this product and the document for that TCP IP software are based in part on BSD networking software Release 1 licensed from The Regents of the University of California e SMARTDAC is a trademark of Yokogawa Electric Corporation e Microsoft MS and Windows are registered trademarks of Microsoft Corporation USA e Pentium are registered trademarks of Intel Corporation e Ethernet is a registered trademark of XEROX Corporation e Modbus is a registered trademark of AEG Schneider e Kerberos is a trademark of MIT e Java and Java related trademarks are trademarks or registered trademarks of Sun Microsystems Inc in the United States and or other countries e Other company and or product names are registered trade mark of their manufactures All Rights Reserved Copyright 2012 2014 Yokogawa Electric Corporation GS 04L51B01 01EN May 7 2014 00
11. characters Tag numbers can be enabled or disabled Tag Number of characters Up to 32 Displayable characters Alphanumeric characters Message e Write messages to the trend display e Number of messages 100 e Number of characters Up to 32 e Write method Write a preset message or write an arbitrary message on the spot e Write destination Select only the displayed group or all groups e Auto message Write a message when the GX recovers from a power failure while memory sampling is in progress Write a message when the trend interval is switched during memory sampling Add message e Write messages to the past data positions e Message The same as the Message item above Number of writable messages per file 50 messages including 10 freehand messages Freehand message e Can be written by dedicated pen Number of writable messages per file 50 messages including 10 Add messages Data Saving Functions Internal memory e Temporarily saves various types of data e Medium Flash memory e File storage capacity GX10 GX20 standard type 500 MB GX20 large memory type 1 2 GB External storage medium e Medium SD card SD SDHC up to 32 GB e Format FAT32 or FAT16 All Rights Reserved Copyright 2012 2014 Yokogawa Electric Corporation Data type e Display data Event data Alarm summary data Manual sampled data Screen image data Setup data and Report data Display data e Target Measurement
12. input output module math communication channels alarm summary message summary Description Maximum or minimum value per recording interval e Recording intervals Determined by the trend interval recording data type display data display data event data GX20 large memory type For GX20 standard type For GX20 large memory type Number of channels Display Display data data Event data 30 s or longer 1000 1000 Note that the maximum number of channels is fixed at 100 in the GX10 e Data size Analog input data 12 bytes ch Digital I O data 4 bytes ch Math channel data 12 bytes ch Communication channel data 12 bytes ch e File size Up to 18 MB Number of files for GX10 GX20 standard type Up to 500 including event data for GX20 large memory type Up to 1000 including event data Operation in the internal memory FIFO First In First Out e Data format Binary or text e Recording Records data at all times e Display data file sample time Measurement channel 30 Math Channel 0 Recording intenal s feos Event data e Target Measurement input output module math communication channels alarm summary message summary operation log Description Instantaneous value per recording interval e Recording intervals Determined by the sample rate recording data type display data display data event data GX20 large memory type GS 04L51B01 01EN May 7 2014 00 4 For GX20 standard t
13. modules is up to 6 W in the GX10 and up to 20 W in the GX20 USB INTERFACE UH e USB port Complies with USB 2 0 and host function e Number of ports 2 one each on the front panel and rear panel All Rights Reserved Copyright 2012 2014 Yokogawa Electric Corporation 11 Connectable devices Only connect the devices listed below to prevent damage to the devices Keyboard Complies with HID Class Ver 1 1 104 89 keyboard US and 109 89 keyboard Japanese Mouse Complies with HID Class Ver 1 1 External medium USB flash memory Does not guarantee the operation of all USB flash memories External medium such as a hard disk ZIP MO and optical discs are not supported Barcode reader USB HID Class Ver 1 1 compatible English U S standard USB keyboard compatible Execution of a communication command using a bar code The entered text can be executed as a communication command Power supply 5 V 10 500 mA 1 4 Devices which need more than 500 mA total bus power for 2 ports cannot be connected at the same time For low powered devices bus power lt 100 mA 5V 5 For high powered devices bus power lt 500 mA 5V 10 GS 04L51B01 01EN May 7 2014 00 m INPUT OUTPUT MODULE SPECIFICATIONS ANALOG INPUT MODULE ee GX90XA or Option Uxx0 DIGITAL INPUT MODULE Model GX90XD or Options CRx1 DIGITAL OUTPUT MODULE Model GX90YD or Options CR1x CR2x ICR4x Please see GX90XA GX90XD GX90YD GX90WD Inpu
14. span for each alarm level Alarm output Output to the internal switch Internal switch operation AND OR operation selectable Display Displays the status on the respective operation screen and an alarm icon on the status display section when an alarm occurs Display operation Hold or not hold the display until the alarm acknowledge operation All Rights Reserved Copyright 2012 2014 Yokogawa Electric Corporation e Alarm hide function alarm no logging function Not display alarms nor record to the alarm summary for each channel Alarm information Displays a log of alarm occurrences on the alarm summary e Reflash The duration for which the reflash relays are deactivated can be set to 500 ms 1 s or 2s e Individual alarm ACK function Alarm display and relay output can be cancelled on individual alarms Event Action Functions e Event action Execute a specified operation when a given event occurs Number of settings 50 Events Remote control input etc Timer Number of timers 4 Match time timer Number of timers 4 Action Specify memory start stop alarm ACK etc Security Functions e Operation lock function Limitations to touch operation access to the external storage medium and various operations e Login function Only registered users can operate the GX It can be set to each of touch operation and communication access System administrators and Users 50 totally Number of Authority of u
15. 20 16 e Communication cycle 500 ms 1 s 2 s 5 s 10 s 15 s 20 s 30 s e Types of data that can be obtained Voltage current power power factor phase electrical energy high frequency wave etc e Number of data allocations GX20 GP20 300 GX10 GP10 50 FAIL OUTPUT FL Contact C contact 1 point e FAIL output The relay contact output on the rear panel indicates the occurrence of CPU failure Relay operation Energized during normal operation and de energized on CPU failure e Status output The relay contact which is de energized in normal output state is energized upon the occurrence of a memory media error measurement error communication error recording stop or alarm e Rated power supply voltage 24 V DC or 250 V AC or less e Rated load current 3A DC 3A AC resistance load e Min load current 100 mA e Recommended replacement periods of contact Electrical 30 000 more ON OFF operations Mechanical 5 000 000 more ON OFF operations Log SCALE LG A logarithmic voltage that has been converted from a physical value is applied to the GX and then the GX s Log scale logarithmic scale is used to display and record the physical value e Input type Log input logarithmic input Pseudo log input An input that supports pseudo logs Log linear input Input that is linear on a logarithmic scale e Range 20mV 60mV 200mV 1V 2V 6V 20V 50V All Rights Reserved Copyright 2012 2014 Yokogawa Electric
16. 24 VA GX20 38 VA GX10 30 VA GX20 45VA GX20 110 VA The following combinations are assumed for LCD backlight off and normal operation GX10 1 Al module 1 DO module 1 DI module GX20 5 Al modules 4 DO modules 1 DI module Module power supply voltage The total allowable power consumption of respective modules is up to 6 W in the GX10 and up to 20 W in the Gx20 Allowable interruption time Less than 1 cycle of the power supply frequency GS 04L51B01 01EN May 7 2014 00 Isolation e Insulation resistance Between the Ethernet RS 422 485 and each insulation terminals and earth 20 MQ or greater at 500 VDC e Withstand voltage Between the power terminal and earth 3000 V AC at 50 60 Hz for one minute Between the contact output terminal and earth 3000 V AC at 50 60 Hz for one minute Between the input output modules and earth Depends on the specification of I O module e Grounding Be sure to set a low grounding resistance e Isolation FAIL output terminal Ethernet port RS 422 485 terminal Input and output module terminal Power terminal Input and output module internal circuit Earth PE terminal RS 232 terminal SD card slot USB port Internal circuit The circuits divided by lines are insulated mutually Safety and EMC Standards e CSA CSA22 2 No 61010 1 installation category ll pollution degree 2 and CSA C22 2 NO 61010 2 030 12 e UL UL61010 1
17. 8 or higher and that can show 65 536 colors 16 bit high color or more Universal Viewer software The universal viewer can display the following data generated by the recorder on the screen and print it out on the printer e Display data file e Event data file e Report data file Including Hour Day Week Month e Manual sample data file 200 2S ess SEM AA BEN oH OBOK SY owe wt Sette t tina e Viewer function Waveform display digital display circular display list display report display operation log display etc e Data conversion File conversion to ASCII or MS Excel format Web application Hardware configurator Online setting on Web browser IE e Offline setting on Web browser IE Both settings can be made using browsers such as Internet Explorer 8 9 10 and 11 8 0 8 8 4 8 E c a aj J E c a EJI a a pi gt a E GS 04L51B01 01EN May 7 2014 00 13 MODEL AND SUFFIX CODES GX10 Model Suffix Code Optional code Description Paperless recorder Panel mount type Small display t3 Paperless recorder Panel mount type Large display 13 Standard Max measurement channels 100 ch Optional features Large memory Max measurement channels 500 ch 2 Display language E English degF DST summer winter time IAS Ad
18. 89 Rear view I O module slot 2 I O module slot 1 I O module slot 0 Power supply terminal RS 422A 485 terminal C3 or RS232 terminal C2 USB port UH Ethernet port FAIL STATUS terminal FL All Rights Reserved Copyright 2012 2014 Yokogawa Electric Corporation GX20 280 2 11 03 295 2 11 62 288 11 34 144 5 67 leg ee ale ler LM olis co Unit mm approx inch Unless otherwise specified tolerance is 3 however tolerance is 0 3 mm when below 10 mm Panel cut dimensions Dimension before fitting attachment MAX 219 2 8 63 1 28 2 168 8 6 65 2 1 11 152 6 6 01 2 2 0 08 to 26 1 02 Thickness of attachment pane O gt i a ii 22 g m SS Q BE S 32 3 1 27 et 247 4 9 74 Ss ite N Dimension after fitting attachment 1 with modules 2 without moduels 360 min 14 17 T FT F Sjo i 1 E N rE i 1 ole oO aak ae aee oer Om oH H H NS i 1 281 6 11 06 Rear view W S 0 module slot 9 I O module slot 8 I O module slot 7 1 O module slot
19. 9200 38400 57600 or 115200 bps Data length 7 or 8 bits Start bit 1 bit Stop bit 1 bit or 2 bit Parity ODD EVEN or NONE Handshaking Off Off XON XON XON RS and CS RS e Communication distance RS 422 485 1200 m RS 232 15 m e Modbus RTU communication Reading or writing of measurement data on other instruments is available by Modbus protocol Communication channel function option MC is needed to read measurement data from other instruments e Operation modes Master or slave e Execution of a communication command using a bar code The entered text can be executed as a communication command CUSTOM DISPLAY CG Using DXA170 DAQStudio screen creation software a custom screen can be constructed and displayed in which display components such as trend digital and bar graphs are freely placed The screen data is transferred from DAQStudio to the internal memory via communication or loaded from an external medium to the internal memory and displayed Number of screens 30 internal memory e Display components e Normal components digital value bar graph tag No tag comment simple digital value simple bar graph alarm mark unit alarm indicator lower limit span value upper limit span value group name system icon memory sample bar date time view batch name and user name e Trend components trend group display with scale board e List components alarm list view and message list
20. ated commands Range Scale unit Alarm Time Moving average e Operation related commands Reset alarm Reset timer Start MATH calculation Rebuild system Initialize Input communication Output communication DO Write message Batch Function e Function Data management using batch names Enter text fields and batch comments in the data file e Batch name Added to the file name of the display data and event data Structure Batch number up to 32 characters lot number up to 8 digits Use not use selectable for lot number on off selectable for auto increment function e Text field Adds text to the display data and event data There are 24 available text fields Up to 20 title characters and 30 other characters can be entered per field e Batch comment Adds text to the display data and event data 3 comments max 50 characters are available Printer Output Function Snapshot Data can be printed out with any LAN connected printer supporting the HP PCL5c language SSL Communication Functions Communication that sends and receives information encrypted by the SSL Secure Socket Layer protocol is possible e Server function Supported servers HTTP server and FTP server Port number 443 when encryption is used Private key Created in GX and saved in the internal memory Server certificate Server certificates created by users can be saved in the internal memory Self signed certificates can be created in GX GS
21. cal trend display Memory summary display Displays the information up to 500 of the data in the memory Specify a file with the cursor and jump to the corresponding section on the historical trend display Report display Displays the report data residing in the internal memory For more details see MATHEMATICAL FUNCTIONS WITH REPORT FUNCTION MT Log display Displays the event log error log communication log FTP log Web log e mail log SNTP log DHCP log and Modbus log Multi panel display Divides the screen into two to six sections and displays some different display formats GS 04L51B01 01EN May 7 2014 00 e Internal switch relay state display Displays the internal switch and ON OFF state of DO Operates the internal switch and ON OFF state e Other displays Network information display System information display System configuration display Auto scroll ON OFF The displayed groups can be automatically switched at a specified interval The display switches in ascending group order Names of channels XXX X __ Channel number Analog input 01 to 10 Digital input 01 to 16 Digital output 01 to 06 Digital input output DI 01 to 08 DO 09 to 14 Slot number GX10 0 to 2 GX20 0 to 9 Expandable I O 0 to 5 Unit number Main unit 0 Expandable I O 1 to 6 Tags Tag and Tag numbers can be displayed Tag number Number of characters Up to 16 Displayable characters Alphanumeric
22. dule Output 6 Input 0 ICR11 With digital I O module Output 6 Input 16 78 ICR20 With digital I O module Output 12 Input 0 ICR21 With digital I O module Output 12 Input 16 ICR40 With digital I O module Output 24 Input 0 ICR41 With digital I O module Output 24 Input 16 1 C2 and C3 cannot be specified together 2 D5can be specified only for the GX20 3 Only one option can be specified 4 Only one option can be specified 5 UC40 UC50 US40 and US50 cannot be specified for the GX10 6 CR20 CR21 CR40 and CR41 cannot be specified for the GX10 7 If UC20 or US20 is specified CR11 cannot be specified for the GX10 8 If UC30 or US30 is specified CRO1 CR10 and CR11 cannot be specified for the GX10 9 Digital input module and digital output module have M3 screw terminals All Rights Reserved Copyright 2012 2014 Yokogawa Electric Corporation GS 04L51B01 01EN May 7 2014 00 14 10 The Display language is selectable from English German French Russian Korean Chinese Japanese As of Mar 2013 To confirm the current available languages please visit the following website URL http www yokogawa com ns language 11 Solid state relay scanner type Type Suffix Code U2 If you need the electromagnetic relay scanner type current scanner type L model DCV TC DI scanner type purchase it separately
23. event triggered by the event action function Firmware update function The Web application the IO module or the expansion module firmware can be updated by operating GX pixels and minor variations in brightness uniformity is a normal display characteristic and not a malfunction Touch screen 4 wire resistive touch screen Construction Mounting Flush panel mounting on a vertical plane Mounting angle Inclined backward up to 30 degrees from a horizontal plane Left and right horizontal Panel thickness 2 to 26 mm Material Case Metal plate Bezel and display cover Polycarbonate Color Case Smoke blue Bezel Charcoal grey light Front panel Water and dust proof Complies with IEC529 IP65 and NEMA No 250 TYPE 4 except External Icing Test except for side by side mounting External dimensions When installing modules GX10 144 W x 144 H x 255 D mm GX20 288 W x 288 H x 220 D mm When uninstalling modules GX10 144 W x 144 H x 174 D mm GX20 288 W x 288 H x 169 D mm D depth from the panel mounting plane Weight GX10 Approx 2 1 kg GX20 Approx 6 0 kg excluding modules Power Supply e Rated supply voltage 100 to 240 VAC Allowable power supply voltage range 90 to 132 180 to 264 VAC Rated power supply frequency 50 60 Hz Power consumption Supply LCD backlight Normal voltage off operation 100 VAC GX10 16 VA GX10 20 VA GX10 48 VA GX20 28 VA GX20 34 VA GX20 90 VA 240VAC GX10
24. martdacplus com software en All Rights Reserved Copyright 2012 2014 Yokogawa Electric Corporation GS 04L51B01 01EN May 7 2014 00 15 Calibration certificate sold separately When ordering the GX10 GX20 with options analog input the calibration certificate for the modules is included in and shipped with the calibration certificate of the main unit When ordering an analog input module separately each module gets its own calibration certificate one certificate per module Test certificate QIC sold separately When ordering the GX10 GX20 with options analog digial I O the QIC for each module is included in and shipped with the QIC of the main unit When ordering analog input modules and digital I O modules separately each module gets its own QIC one QIC per module User s Manual Product user s manuals can be downloaded or viewed at the following URL To view the user s manual you need to use Adobe Reader 7 or later by Adobe Systems URL www smartdacplus com manual en Product Purchase Specifications The GX10 GX20 is composed of the main unit I O modules Expandable I O and Expansion module There are two ways to purchase I O modules One way is to purchase them individually by specifying models GX90XA GX90XD and GX90YD The other way is to purchase them as an option UCxx or USxx Purchasing them as an option is convenient but this places limitations on the number of analog inputs that you can obtain
25. metic operations Four arithmetic operations square root absolute common logarithm natural logarithm exponential and power Relational operations lt lt gt 2 and Logic operations AND OR NOT and XOR Statistical operations TLOG or CLOG Special operations PRE HOLD RESET and CARRY Conditional operation a b c Bit operation BIT Integer extracting operation INT Remainder extracting operation MOD Trigonometric functions SIN COS Computation accuracy Double precision floating point Data that can be used Channel data Measurement channels GX10 0001 to 6516 GX20 0001 to 6516 mathematical channels GX10 A001 to A050 GX20 A001 to A100 Communication channels GX10 standard type C001 to C300 GX20 large GS 04L51B01 01EN May 7 2014 00 memory type C001 to C500 Constants 100 K001 to K100 Internal switch 100 S001 to 100 Flag 20 F01 to F20 Recording state RECO1 Report function e Number of report channels GX10 50 GX20 60 e Report types Hourly daily daily weekly daily monthly Batch Day custom e Computation types Average maximum minimum sum or instantaneous value e Unit of sum OFF s min hour day e Report templates Office Open XML spreadsheet files which can be displayed with Microsoft Office Excel or PDF files can be output or printed out with any LAN connected printer supporting the HP PCL5c language COMMUNICATION CHANNEL
26. nd 9 units e A module type is four types an analog input a Measuring accuracy 1 typical value 2 digital input a digital output and a digital input DCV 0 01 of reading 5 pV output a e Amaximum of six units of expandable I O can be 6V 1 5V 0 01 of reading 2 mV connected to GX10 GX20 Each expandable I O can incorporate a maximum of six modules With expandable I O connected the large memory type GX10 and GX20 can use multiple channels Furthermore the I O ports can be installed in Pt100 0 02 of reading 0 2 C Pt100 0 02 of reading 0 16 C high resolution separate locations from where GX10 GX20 is 1 General operating conditions 2342 C 55 10 RH supply voltage 90 132 180 250 VAC supply frequency within 50 60 Hz 1 warm up of 30 minutes or more no vibrations or other located helping you reduce wiring requirements and build a decentralized configuration e The intuitive operation by flick pinch in pinch out hindrances to performance and swipe are possible 2 For the measuring accuracy guaranteed e The past trend under recording can be seamlessly see the module s general specifications displayed on a trend screen GS04L53B01 01EN e Moreover the measurement data of the time specified on the calendar screen can be searched MAIN UNIT SPECIFICATIONS and displayed e Various functions such as a freehand message FUNCTIONAL SPECIFICATION a PDF Excel output of a rep
27. ng only totally 100 users e User restrictions setting 10 kinds for general users e Password expiration time select form Off 1month 3 month 6month e Password control function Logins are verified by a Kerberos v5 authentication server only user name and password Encryption method AES128 CTS HMAC SHA1 96 AES256 CTS HMAC SHA1 96 ARCFOUR HMAC MD5 Pre Auth function use The function has confirmed compatibility with Windows Server2003 SP2 Windows Server2008 SP2 Windows Server2012 ActiveDirectory e Signature function After checking data that has finished being recorded you can add three levels of electronic signature select a pass fail and enter comments 50 characters maximum e Audit trail function The operation log the settings change log and the settings file when the change was made are saved SERIAL COMMUNICATION INTERFACE C2 C3 Connection EIA RS 232 C2 or EIA RS 422 485 C3 e Protocol Dedicated protocol or Modbus protocol e Setting measurement server function Operation setting or output of measurement data are available by dedicated protocol e Synchronization Start stop synchronization e Transmission mode RS 422 485 RS 422 Four wire half duplex multi drop All Rights Reserved Copyright 2012 2014 Yokogawa Electric Corporation connection 1 n n 1 to 31 RS485 Two wire half duplex multi drop connection 1 n n 1 to 31 e Baud rate 1200 2400 4800 9600 1
28. nsation e Vibration 10 to 60 Hz 4 9 m s maximum e Shock 392 m s maximum in packaged condition GS 04L51B01 01EN May 7 2014 00 m EXTERNAL DIMENSIONS AND PANEL CUTOUT DIMENSIONS GX10 Unit mm approx inch Unless otherwise specified i tolerance is 3 however tolerance is 0 3 mm when o below 10 mm Dimension before fitting attachment MAX 224 4 8 83 1 173 1 6 81 2 151 5 5 96 22 8 144 5 67 0 90 _157 7 6 21 3 40 9 1 61 2 0 08 to 26 1 02 Thickness of attachment panel ie uo ocd Sey l Sy 26 6 1 05 28 Dimension after fitting attachment 1 with modules 2 without moduels 40 9 1 61 103 3 4 07 Panel cut dimensions Single Unit Side by Side Mounting Mounting horizontally 137 5 39 j l a 175 MIN 1137 q 6 89 6 39 1372 5 39 L Ls A Side by Side Mounting vertically max 3 units 137 L4 639 2 282 11 10 3 426 16 77 4 570 22 44 5 714 28 11 L 2 6 858 33 78 G 7 1002 39 45 8 1146 45 12 9 1290 50 79 10 1434 56 46 n 144xn 6 175 MIN 6
29. operations can be performed with the Web browser The screen layout can be determined independently of the screen of the GX main unit Number of the simultaneous connection Max 4 SNTP client Inquires the time to the SNTP server and sets the GX SNTP server Outputs the GX time Time resolution 5 ms DHCP client Automatically obtain the network address settings from the DHCP server Modbus client Reads data from another device and writes to the registers All Rights Reserved Copyright 2012 2014 Yokogawa Electric Corporation Number of connectable sever GX10 GX20 standard type Max 16 GX20 large memory type Max 32 e Modbus server Loads measurement and math channel data Loads and writes communication channel data Some control commands such as memory start Modbus client register access limitations Required MC option Number of the simultaneous connection Max 4 Setting Measurement server Operate set and output data of the GX using a dedicated protocol e DARWIN compatible communication server Supports some DARWIN commands Communication with GX is possible using DARWIN communication commands Output related commands Output measurement IO channel data Output calculation channel data Output relay status Output the position of the decimal point for the measurement IO channel Output the position of the decimal point for the calculation channel Output the information on system configuration e Setup rel
30. ort file a direct output to a network printer a scale movement of a trend ede ae display and a buzzer sound are equipped Input Specifications e tcan be hooked up to network via Ethernet which Please see GX90XA GX90XD GX90YD GX90WD enables to inform by Email and to monitor on Web I O Module General Specifications site as well as to transfer files by using FTP Also Gener it can communicate with Modbus RTU or Modbus kalen eel TCP i e A setup of GX can be performed on line from the GS 04L53B01 01EN web browser on PC A setup by off line is also ossible 7 avera Viewer software allows a PC to display waveforms on its screen and to print out waveforms Yokogawa Electric Corporation GS 04L51B01 01EN YO KOGAWA 2 9 32 Nakacho Musashino shi Tokyo 180 8750 Japan Copyright November 2012 9th Edition May 7 2014 Measuring Functions The number of installable modules and I O channels total for GX and expandable 1 0 GX10 GX20 standard type Number of input output module Max 100 GX20 large memory type GX20 large memory type Max 500 or max 450 for Al only Restrictions of module connection Amaximum of 10 modules can be installed as a total for GX90YD digital output modules and GX90WD digital I O modules e A maximum of one GX90WD digital I O module can be installed in GX and in each expandable O e Expandable I O connection Number of connectable units Max 6 Caution If the maxim
31. ser 10 level Manual Sampled Data e Item Instantaneous value at an arbitrary time Target Measurement input output module math communication channels Number of recording channels GX10 GX20 standard tyep Max 50 GX20 large memory type Max 100 e Maximum number of data values that the internal memory can store 400 Data format Text Report Data e Item Report at each scheduled time of report Target Measurement input output module math communication channels e Maximum number of reports that the internal memory can store 800 e Data format Text Snapshot Data e Item Displayed screen image data e Data format PNG e Output destination External medium or communication output Setup Data e Item GX setup data e Data format Text e Output read destination for saving loading External medium Clock Functions e Clock With a calendar function e Accuracy 5 ppm 0 to 50 C excluding a delay of 1 second maximum caused each time the GS 04L51B01 01EN May 7 2014 00 power is turned on Time setting Using touch operation communication command event action function or SNTP client function Time adjustment method Limit in which the time is gradually adjusted Select from the available settings between 5 s and 15 s Whether to change an out of limit operation immediately or report it as an error can be selected While memory sampling Corrects the time by 1 ms for each second
32. t Output Module General Specification GS 04L53B01 01EN m APPLICATION SOFTWARE SMARTDAC STANDARD e Universal viewer e Web application Hardware configurator Download the latest version of the software from the following URL www smartdacplus com software en Operating environment Cie eee Windows Vista Home Premium SP2 Except for 64 bits Editions Business SP2 Except for 64 bits Editions SP2 Business SP2 Except for 64 bits Editions for 64 bits Editions Windows 7 Home Premium SP1 32 bit and 64 bit a CERURA Professional SP1 Professional SP1 32 bit and 64 bit Editions bit and 64 bit Editions Processor and main memory o Processor and main memory Windows Vista Intel Pentium 4 3GHz or faster x64 or x86 2GB or more Windows 7 32 bit edition Intel Pentium 4 3GHz or faster Windows 8 1 x64 or x86 2GB or more 64 bit edition Intel x64 processor that is equivalent to Intel Pentium 4 3 GHz or faster 2GB or more Browser Browser Version Windows Internet Explorer Internet Explorer 8 Internet Java Runtime Environment Explorer 9 Internet Explorer 10 1 7 0 xx Version 7 Update or Internet Explorer 11 Xx xx is more than 51 Hard disk e 100MB or more of free space All Rights Reserved Copyright 2012 2014 Yokogawa Electric Corporation 12 Display e Avideo card that is recommended for the OS and a display that is supported by the OS has a resolution of 1024 x 76
33. um number of I O channels are assigned and the last channel is assigned to an intermediate channel of a module that module and subsequent modules will not be recognized Module 1 Module n 1 81 91 Maximum number of I O channels 10 90 O O x Display Functions Display groups Number of groups GX10 30 GX20 standard type 50 GX20 large memory type 60 Number of channels that can be assigned to each group GX10 10 GX20 20 Scan interval 10012 2002 500 ms 1 2 5s 1 Cannot be specified if an electromagnetic relay scanner type Type Suffix Code T1 analog input module is in use 2 Cannot be specified for L model DCV TC DI scanner type Type Suffix Code L1 Display color Trend Bar graph Digital display Channel Select from 24 colors A desired display color can be selected freely using its RGB value Background Select from white or black Display type e Trend display T Y Display method Direction Horizontal vertical Trend interval 5 12 10 15 30 s div 1 2 5 10 15 20 30 min div 1 2 4 10 h div 1 Cannot be specified if an electromagnetic relay scanner type analog input module is in use 2 Cannot be specified for L model DCV TC DI scanner type Type Suffix Code L1 Trend line width Thick normal thin All Rights Reserved Copyright 2012 2014 Yokogawa Electric Corporation Scale GX10 Max 6 GX20 Max 10 Current value bar graph color scale band
34. vailable in the testing Distribution and measuring circuits board circuit connected to a power breaker etc distribution portion of a low voltage main power supply facility IV CAT IV Available in the testing verhead and measuring circuits wire cable connected to a power systems etc source of a low voltage main power supply facility WEEE Directive Compliant Normal Operating Conditions Power supply voltage 100 to 240 VAC 10 Power supply frequency 50 60 Hz 2 Ambient temperature 0 to 50 C Ambient humidity 20 to 80 RH at 5 to 40 C no condensation e Magnetic field 400 A m or less DC and 50 60 Hz e Vibration 5 lt f lt 8 4 Hz amplitude 3 5 mm peak 8 4 lt f lt 160 Hz acceleration 9 8 m s e Shock IEC 60068 2 27 Non energization 500 m s or less approximate 10 ms 6 directions X Y Z 3 times in each direction e Mounting position Can be inclined up to 30 degrees backward Left and right horizontal when installing the panel mount and wall mount e Altitude 2000 m or less e Installation location Indoors e Warm up time At least 30 minutes after power on Other Specifications e Memory backup A built in lithium battery backs up the settings and runs the clock e Recommended replacement periods of Battery Approximately 10 years at room temperature Transport and Storage Conditions e Ambient temperature 25 to 60 C e Ambient humidity 5 to 95 RH no conde
35. vanced security function Part 11 Black cover RS 232 RS 422 485 Custom display IDS VGA output EtherNet IP communication WT communication 4 Fail output 1 point ILG Log scale IMT Mathematical function with report function Communication channel function 24 V DC AC power supply IUH USB interface Host 2 ports Analog input module Digital I O module Please add the following suffix codes to the main unit model and specification codes Option Optional code Description Optional features Analog input 31 UC10 With analog input module 10ch Clamp terminal UC20 With analog input module 20ch Clamp terminal 7 UC30 With analog input module 30ch Clamp terminal 8 UC40 With analog input module 40ch Clamp terminal UC50 With analog input module 50ch Clamp terminal US10 With analog input module 10ch M3 screw terminal US20 With analog input module 20ch M3 screw terminal 7 US30 With analog input module 30ch M3 screw terminal 8 US40 With analog input module 40ch M3 screw terminal US50 With analog input module 50ch M3 screw terminal Optional features Digital 1 0 ICRO1 With digital I O module Output 0 Input 16 8 ICR10 With digital I O mo
36. view e Operation components DO DO operable internal switch internal switch operable numeric value operations viewing data of and writing data to communication channels and button operations writing numeric values operating bits switching screens and executing communication commands e Text components labels e Figure components line view rectangle view and circle view e Figure components line view rectangle view and circle view e Configuring screens Screen creation software Creation using DXA170 DAQStudio GX does not have a creation function e Saving loading screen data A specified screen or all the screens is are loaded from an external storage medium to the internal memory or a specified screen or all the screens in the internal memory is are saved on an external storage medium GS 04L51B01 01EN May 7 2014 00 VGA VIDEO OUTPUT D5 Only for GX20 e External display Resolution 800 x 600 dots VGA Connector 15 pin D Sub EtheNet IP COMMUNICATION E1 Can be joined to an Ethernet IP network as an adapter or a server e Loading data from the I O channel or calculation channel e Loading and writing data from to the communication input channel WT COMMUNICATION E2 Collects data by connecting to WT equipment manufactured by Yokogawa Meters amp Instruments Corp via Ethernet communication e Supported models WT1800 WT500 and WT300 e Number of connectable units GX10 8 GX
37. ype Sampling rate Number of channels For GX20 large memory type Number of channels Display Display data data Event data Trend interval div Note that the maximum number of channels is fixed at 100 in the GX10 Data size Analog input data 6 bytes ch Digital I O data 2 bytes ch Math channel data 6 bytes ch Communication channel data 6 bytes ch File size Up to 18 MB Number of files for GX10 GX20 standard type Up to 500 including display data for GX20 large memory type Up to 1000 including display data Operation in the internal memory FIFO First In First Out Data format Binary or text Mode Free Records data at all times Trigger Starts recording data when a certain event occurs and records for the specified interval Repetition trigger Repeat Trigger mode Event data file sample time Measurement channel 30 Math Channel 0 Internal Memory 500 MB Recording interval s Total sample time Approx 1 month Alarm Functions Number of alarms Up to four alarms level for each measurement channels Alarm type High limit low limit difference high limit difference low limit high limit on rate of change alarm low limit on rate of change alarm delay high limit and delay low limit Alarm delay time 1 s to 24 hours for each channel Rate of change calculation interval of rate of change alarms 1 to 32 times the scan interval common to all channels Hysteresis 0 0 to 5 0 of the
Download Pdf Manuals