ANALOX 101D2 Portable - Oxygen Monitor User Manual

image

Contents

1. User Manual Warning These instructions should be read and understood by all individuals who will Be responsible for the operation of this analyser The actions taken as a result of the measured Oxygen readings must be in strict accordance with the Company or Country regulations Document ref 1D2 800 01 January 2009 Page 13 ZINALOX looking after the air YOU breathe
2. AN ALOX looking after the air YOU breathe 0 0196 19 9996 at Atmospheric Pressure 0 196 100 0 at Atmospheric Pressure 0 0196 O2 t 196 of Readout 0 196 of Readout per C 2 years in Air 20 9 4 x MN1500 Alkaline Cells Approx 1 month of continuous use reducing with Alarm Activation External Supply 0 30v DC Relays on Hi Lo Alarms Alternative Sensor I P Depth Overall 205mm Height Overall 68mm Width Overall 148mm Weight 700gms Incl Battery Document ref 1D2 800 01 January 2009 Page 11 ANALOX 101D2 Portable Oxygen Monitor User Manual 10 0 ANALOX OXYGEN COMPENSATION CHART penuesaH SIY IH PIMPHOM liy u6nouqselppiiN ejsexois sed JBlISNpu Kersexois NOD seuojsureM G p SUN ON NOLLO31OHd IWLNSWNOUIANS OTDNV HSILLOOS 9661 1ubu doo XOHLIN U oeJnooe ZO G 0 uleyUIeW o are Alp UyIM JO jana ZO Heyo eui oi IAL eyo eu jo Wed siy ui JaowW sixe HY pue einjejeduie 3U jI S9 Lv vt St 8l 1 80 i HM 96001 1 OZH 1N3233d N39AXO 2DIH3HdSOIWLV ALIQIINQH 3ALVT34 8E 43 LZ LZ 9l Ol v 0 2 dW3L 001 06 08 OL 09 0S Ov c 3 dW3L ALIGINAH 3ALIVT38 ONY 3401V3dWa3l OL NOILV13Y NI LN3233d N39AXO JYAHdSOWLY Ju3HdSOWIV 3H1 NI 33401SIOW 303 LUYVHD NOIIVSN3dWOO N39AXO 3n680011 D GNIWISYIN IVD N NOU VAONAE ABojouypay 10sues XOT NY Document ref 1D2 800 01 January 2009 Page 12 looking after the air YOU breathe ANALOX 101D2 Portable Oxygen Monitor
3. ANALOX looking after the air YOU breathe ANALOX 101D2 Portable Oxygen Monitor User Manual Analox Sensor Technology Ltd 15 Ellerbeck Court Stokesley Business Park North Yorkshire TS9 5PT T 44 0 1642 711400 F 44 0 1642 713900 W www analox net E info analox net ANALOX 101D2 Portable Oxygen Monitor User Manual CONTENTS 1 0 AMP OGUCTIONS s coswcssnnuanesssngnneanstakanssnnasvvantiemnessantntearteatnerts Page 3 20 COMMONS M Page 3 3 0 Battery Check CC Page 4 EHE orioneu E EEEE E et Page 4 5 0 Alarm Operations esee Page 5 6 0 Alarmi Seti pur EM eR RRNU Mn aac S Uu E Page 5 7 0 Sensor Replacement seeueessssss Page 6 8 0 Trouble Shooting em c vv Page 7 9 0 Warranty Statement esses Page 10 10 0 Specifications cm Page 11 11 0 Compensation Chart assecuti tr reboot Sod e thon ds Page 12 Document ref 1D2 800 01 January 2009 Page 2 ZINALOX looking after the air YOU breathe 1 0 1 3 ANALOX 101D2 Portable Oxygen Monitor User Manual INTRODUCTION The ANALOX 101D2 Oxygen Analyser is a battery powered portable instrument which provides a continuous digital display of the Oxygen concentration in the gas sample presented to the front surface of the sensor unit The instrument has two ranges 0 01 19 99 and 0 1 100 0 The sensor assembly incorp
4. User Manual SENSOR REPLACEMENT When the instrument can no longer be calibrated by adjustment of the CAL control it is possible that the Oxygen Sensor unit is exhausted and should be replaced The ANALOX 9212 4 Sensor Assembly has a life of approximately 3 years in normal air but this will be reduced if it is constantly exposed to higher concentrations of Oxygen Conversely if the sensor is normally exposed to lower concentrations of Oxygen its life may be extended The sensor must be replaced with an Analox Oxygen sensor type 9212 4 as this sensor is especially manufactured for this instrument There are no serviceable parts in the sensor and cable assembly the entire unit must be replaced The sensor in the 101 D2 is an electrochemical device and contains a caustic electrolyte Always check to make sure that it is not leaking and do not allow it onto any part of your body or clothing In the event that you do come into contact with the electrolyte wash the contaminated part with copious amounts of water see safety information WARNING If after handling the sensor you fingers or other parts of your body feels slippery or stings wash with a lot of water If stinging persists get medical attention Document ref 1D2 800 01 January 2009 Page 6 AANALOX looking after the air YOU breathe ANALOX 101D2 Portable Oxygen Monitor User Manual 8 0 TROUBLE SHOOTING SYMPTOM REASON SO
5. checks the instrument zero and is the same as the theoretical zero voltage output of a sensor when it is subjected to as containing no Oxygen Alternatively Pure Nitrogen or Helium or other Oxygen free gas may be passed across the sensor at a flow rate of approx 60 Litres Hour for about 3 minutes The Instrument reading should not be greater than about 0 03 800ppm depending on the purity of the gas being used and the sensor performance Span Adjustment this can be carried out in two ways Select the O 100 range and expose the sensor to normal ambient atmosphere When the reading on the display is steady adjust the CAL control on the Instrument front panel until the display reads 20 9 Turning the control clockwise will increase the reading This completes the calibration process NOTE High humidity decreases Oxygen levels in the atmosphere please refer to the humidity chart in 10 0 calibrate using procedure 4 2 2 Using a known accurate Oxygen Nitrogen gas mixture select the range most appropriate to the gas concentration Fit a flow adaptor to the sensor and pass the gas at a flow rate of about 60 Litres Hour for about 5 minutes When the Reading on the display is steady adjust the CAL control on the front panel until the reading on the display agrees with the Oxygen concentration of the gas being used Clockwise rotation of the control increases the reading This completes the calibration process If the des
6. the four securing screws in the base of the instrument and fit the rubber feet in the rectangular holes Document ref 1D2 800 01 January 2009 Page 8 AANALOX looking after the air YOU breathe ANALOX 101D2 Portable Oxygen Monitor User Manual 8 1 4 OVERLAY DRAWING Ox oO OX O a Z Q Q Lu O Q a lt Z LLI Z O cc z Z LL Q O Z lt I e E O Q v 9 L I O Q a O O Document ref 1D2 800 01 January 2009 Page 9 ANALOX looking after the air YOU breathe ANALOX 101D2 Portable Oxygen Monitor User Manual WARRANTY INFORMATION We provide the following Warranties for the Analox 101D2 A 2 year graded sensor warranty A 1 year electronics warranty In both cases the Warranty period runs from the date of our Invoice We warrant that the equipment will be free from defects in workmanship and materials The Warranty does not extend to and we will not be liable for defects caused by the effects of normal wear and tear erosion corrosion fire explosion misuse use in any context or application for which the equipment is not designed or recommended or unauthorised modification Following a valid Warranty claim in accordance with the above the equipment upon return to us would be repaired or replaced without cost or charge but in our discretion we may elect instead to provide to you which ever is the lesser of the cost of replacement or a refund of net pu
7. 22 796 ALARM SETTING Before any adjustments are carried out to the SET ALARM controls the user should release the locks on the control knobs This is done by moving the small lever located at the edge of the control Downward until the knob moves freely After adjustment the lock should be reset to prevent accidental movement The SET HI SET LO toggle switch is normally spring biassed to its central position and in this position displays measured Oxygen concentration The Hi alarm trip point may be reset by moving and holding the switch Upward and adjusting the SET HI ALARM control until the desired setpoint is displayed on the LCD The Lo alarm trip point may be reset by moving and holding the switch Downward and adjusting the SET LO ALARM control until the desired setpoint is displayed on the LCD Alarm trip points may only be set to a resolution of 0 196 O2 If the instrument is being used on the 0 19 99 range then during the adjustment procedure the display will be automatically adjusted to read the correct resolution If the user only requires to check the present alarm set points then this may be done by just pressing the SET HI SET LO switch to the appropriate position and reading the value on the LCD Document ref 1D2 800 01 January 2009 Page 5 AANALOX looking after the air YOU breathe 7 0 7 1 7 2 7 3 7 4 7 5 ANALOX 101D2 Portable Oxygen Monitor
8. LUTION Battery Symbol Low Battery Change battery LED does not read Requires offset adjust See instructions in 8 1 000 0 02 Switch on No display Switched off Check display Bad connection connection Check battery connection Zero reading Sensor disconnected Sensor expired Check connection Change sensor Reading erratic Reading does not change When calibration pot is turned Pressure on sensor Radio transmission Sensor old or faulty Condensation on sensor Faulty connections Sensor failure Check Flow Move unit away Change sensor Dry sensor face Check connections Change sensors Display segments missing Display faulty Return to dealer Will not calibrate Sensor faulty Sensor not in air High altitude Change sensor Check flow adapter Calculate percent Equivalent 20 996 xbar Reading drifts Rapid temperature change Do not move analyser from one temperature to another immediately before use AANALOX Document ref 1D2 800 01 January 2009 looking after the air YOU breathe Page 7 ANALOX 101D2 Portable Oxygen Monitor User Manual 8 1 TROUBLE SHOOTING continued 8 1 1 In order to access the interior of the instrument the top cover requires removal To remove the top cover ensure the instrument is switched off and remove the four screws located in the recesses in the base and the four rectangular rubber feet Th
9. e top cover is retained by four internal catches which can be seen at the bottom of the rectangular holes which normally hold the rubber feet These catches can be released by using the green tool provided with the instrument or a suitable screwdriver Lay the instrument on its side and insert the tool in one of the REAR holes keeping the bevelled edge toward the rear of the instrument Whilst holding the instrument firmly push the tool toward the top over and gently ease the rear corner of the top away from the body Repeat the process on the other REAR CATCH Release the front catches in a similar manner and the top cover may then be removed WARNING UNDER NO CIRCUMSTANCES SHOULD ANY ADJUSTMENTS OTHER THAN THOSE OUTLINED IN THIS PROCEDURE BE MADE TO OTHER CONTROLS IN THE INSTRUMENT ANY ALTERATION OF OTHER CONTROLS WILL AFFECT THE OPERATION AND ACCURACY OF THE INSTRUMENT AND THE SUPPLIER CANNOT BE HELD RESPONSIBLE ADJUSTMENT PROCEDURE The potentiometer that requires adjustment is VR2 This is located on the main printed circuit board SA004B 1 Adjacent to the front panel is a line of five potentiometers VR2 is the second potentiometer from the end nearest the front panel see drawing 8 1 4 RE ASSEMBLY OF THE BOX Place the top cover loosely in position easing the front edge groove of the cover into the tongue on the instrument body Press the top cover down and forward until the retaining latches click into position Replace
10. ired readings cannot be obtained then it is likely that the sensor has come to the end of its life and should be replaced NOTE The Instrument does not need to be calibrated on both ranges Document ref 1D2 800 01 January 2009 Page 4 AANALOX looking after the air YOU breathe 5 0 5 1 5 2 5 3 5 4 6 0 6 1 6 2 6 3 ANALOX 101D2 Portable Oxygen Monitor User Manual ALARM OPERATION If an alarm condition occurs the internal audible buzzer will sound intermittently and the appropriate LED will flash The audible alarm may then be silenced by pressing the MUTE button on the front panel If the reading is still in the alarm condition the LED will continue to flash until the Oxygen concentration returns within the normal band The LED will then turn off If an alarm condition occurs and the Oxygen concentration returns to normal before the MUTE button is pressed then the audible and visual alarms will remain active until the MUTE button is pressed This facility allows the user to be aware of an alarm occurring whilst the instrument was unattended The alarms have a built in Hysteresis of approximately 0 396 O2 to overcome nuisance alarm triggering when measuring near either set point This means that if a high alarm occurs with a set point of 23 096 then having acknowledged the alarm by pressing MUTE button the alarm state will not clear until the Oxygen level drops to below
11. orates advanced circuit design which provides automatic temperature compensation over the range 10 C to 40 C The instrument is easy to calibrate requiring only a single adjustment and user settable High and Low audible and visual alarms are fitted as standard Optional extra facilities can be specified at time of ordering These include an analogue output voltage relays which operate in alarm conditions a 4 20mA Sensor Input to accommodate a remote loop powered Sensor Transmitter Optional extras require the fitting of the 0 30vDC external power facility 2 0 CONTROLS 21 4 ANALOX 101D2 OXYGEN ANALYSER OFF SET H ON SET 10 SENSOR SCOTTISH ANGLO ENVIRONMENTAL PROTECTION LTD Document ref 1D2 800 01 January 2009 Page 3 ANALOX looking after the air YOU breathe 3 0 3 1 4 0 4 1 4 2 4 2 1 4 2 2 ANALOX 101D2 Portable Oxygen Monitor User Manual BATTERY CHECK If the internal battery voltage falls below a pre set threshold approximately 5 volts then the LCD display on the instrument front panel will alternate between its normal reading and 1888 The Instrument should not be used when this occurs since the indicated value of Oxygen concentration may not be accurate CALIBRATION Zero Check Select the 0 19 9996 range using the switch on the front panel and remove the sensor jack plug from the socket on the front panel The LED should read 0 00 0 02 This
12. rchase price paid as per our Invoice on initial purchase from us We shall have no liability for losses damages costs or delays whatsoever We shall have no liability for any incidental or consequential losses or damages All express or implied warranties as to satisfactory or merchantable quality fitness for a particular or general purpose or otherwise are excluded and no such Warranties are made or provided save as set out in this Clause 7 In order to effectively notify a Warranty claim the claim with all relevant information and documentation should be sent in writing to Analox Sensor Technology Limited 15 Ellerbeck Court Stokesley Business Park Stokesley North Yorkshire TS9 5PT Or by e mail to info analox net Or by Fax to 44 1642 713900 We reserve the right to require from you proof of dispatch to us of the notification of Warranty claim by any of the above alternative means The equipment should not be sent to us without our prior written authority All shipping and Insurance costs of returned equipment are to be born by you and at your risk All returned items must be properly and sufficiently packed Document ref 1D2 800 01 January 2009 Page 10 AANALOX looking after the air YOU breathe ANALOX 101D2 Portable Oxygen Monitor User Manual 9 0 SPECIFICATIONS Ranges Minimum Reading Accuracy Temperature Effect Sensor Life Battery Battery Life Optional Extras Dimensions

Download Pdf Manuals

image

Related Search

Related Contents

UFED Logical Analyzer  1 Digital Document Access (DDA)  ThinkPad® X Series Service and Troubleshooting Guide  nouvelles-acquis 0910  Dell PowerConnect J-EX4200 Quick Start Manual  2004 forester service manual quick reference index body section  Le Moniteur d`Issoire 1904-03  Le contrat de génération - AGEFOS PME Ile-de  TLE-12  ZyXEL Communications 645R-A Series Network Router User Manual  