ORTHOPHOS XG 3 - Sirona Support
Contents
1. 5 Attach all remaining unit housing covers IMPORTANT In order to attach the floor stand covers properly the four spring clamps E must be placed on the base plate in sucha way that the two covers F and G remain assembled after they are attached see detail drawing above 60 51 416 D 3352 D 3352 031 03 17 02 43 3 Installation Panoramic X ray unit Sirona Dental Systems GmbH 3 7 Attaching the covers Installation Instructions ORTHOPHOS XG 3 60 51 416 D 3352 44 D 3352 031 03 17 02 The Dental Company 4 Electrical connection ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 17 02 45 4 Electrical connection Sirona Dental Systems GmbH 4 1 Connecting the control cables PAN Installation Instructions ORTHOPHOS XG 3 4 1 Connecting the control cables PAN L108 X103 L2 T _ a aK p a Duplex patch cable 1 1 connection SC SC 50 125 um Media Konverter X101 r 3 B E T n A P re Coiled cable Switch Media convert pare Se LN ies pa Ras E S d Wagoklemme O 1 Connect the personal computer to socket SC SC of the For the installation version without remote control panoramic X ray unit via the media converter
2. 10 Insert screw R through washer S and then through the e Tighten adjusting nuts Q see page 33 and screw R support and into the stand from behind firmly 11 Fit profile clamp T onto screw R from the other front side and screw nut U onto screw R 60 51 416 D 3352 34 D 3352 031 03 17 02 Sirona Dental Systems GmbH 3 Installation Panoramic X ray unit Installation Instructions ORTHOPHOS XG 3 3 4 Installing the floor stand option 2 3 Q 5 Q e Setup the unit including the center styrofoam part IMPORTANT Please observe the required movement range of the X ray unit during installation see section 1 6 12 Remove the styrofoam packaging and rotate the X ray tube assembly counterclockwise to the front side of the unit 60 51 416 D 3352 D 3352 031 03 17 02 35 3 Installation Panoramic X ray unit Sirona Dental Systems GmbH 3 4 Installing the floor stand option 2 Installation Instructions ORTHOPHOS XG 3 VO KATA Q e ase e Loosen screw R and adjusting nuts Q again slightly IMPORTANT 13 Level the unit in both directions by turning adjusting Be sure to tighten all adjusting nuts equally to the same nuts Q while measuring with the spirit level torque after leveling Align the stand first by placing the spirit level against the side and rear of the stand 14 Tighten screws R again firmly Then align the r
3. 13 Drill through the recesses of the stand into the floor In sert wall plug M and check again that the stand is aligned correctly see step 10 Screw the stand to the floor with the two wood screws N and the washers O 14 Attach the covers of the wall holder s 60 51 416 D 3352 D 3352 031 03 17 02 29 3 Installation Panoramic X ray unit 3 4 Installing the floor stand option 2 3 4 Installing the floor stand Must be dis mantled ly ing on its 4A CAUTION For installation with the floor stand the unit remains lying on the pallet until the floor stand has been completely assem bled Only then may the unit be installed For enhanced rep resentation some of the following drawings are shown in the standing state 1 Remove the surrounding packaging the two lateral sty rofoam parts and profile cover A NOTICE The center styrofoam part should remain attached to the unit for protection 2 Loosen the nuts B with washers on both sides of the pallet or the wooden support Take off the pallet or the wooden support Remove the threaded bolt C 30 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 option 2 C i p A i ir n i eee OB Te a IMPORTANT The nuts D on the unit may remain inside the unit when the threaded rods are removed Remove the nuts 3 R
4. Please move the sensor by 6 Adjust the sensor IMPORTANT Positive sign Moves the sensor to the right Negative sign Moves the sensor to the left To do this proceed as follows Mark the current position of the sensor with a per manent marker Loosen the three screws A approx 2 turns do not unscrew fully Move the sensor to the left or right by the displayed value Lx Tighten screws A again 117 Sirona Dental Systems GmbH 11 Appendix Installation Instructions ORTHOPHOS XG 3 11 2 Adjusting the panoramic X ray unit l Pan adjustment Sensor adjustment Please move the sensor by 4 mm in SIDEXIS on the Multipad 60 3 0 60 Fo llow and 0 oO 12 7 Confirm the displacement of the sensor Click OK The exposure dialog box showing the exposure status appears in Sidexis The initialization status is visualized by a progress indi cator on the Multipad The initialization procedure is completed when the ex posure parameters of service routine 010 1 60 kV 3 mA 0 6 s are displayed and the progress in dicator disappears 60 51 416 D 3352 118 D 3352 031 03 17 02 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 V7 ay EZ vu VV OO Dada CE 10 Diaphragm system adjustment kammer3 i x Pan Sensor adjustment Pan Fixed diaphragm Pan Symmetry Pan Reset adjustment gt
5. A1 88 6 mm 1 mm l A2 44 3 0 5 mm lt lt 7g A2 44 3 0 5mm Adjustment ok length measurement with SIDEXIS 7 Evaluate the X ray image The shadow of the center needle the needle AERAN image and the auxiliary line must be coincident and located behind each other A tolerance offset of needle from the central If the X ray image still does not correspond to the ideal image it is possible that one or more criteria have not auxiliary line of 0 75 mm is admissible A Ver peen IMUA Distance A1 must be 88 6 1 mm A Case 1 Center needle not coincident with auxiliary line Distances A2 must be identical each being 44 3 If the shadow of the center needle the needle image and 0 5 mm A the auxiliary line do not yet coincide then repeat the adjust A white border surrounding the image on all ment procedure starting with step 5 sides must be visible B Case 2 Symmetry not OK IMPORTANT If the shadow of the center needle the needle image and the auxiliary line coincide but the symmetry or the distance between the two outer needles is not yet correct then re peat the adjustment procedure starting with step 5 Always measure exactly from pin center to pin center IMPORTANT If you do not reach your goal via automatic adjustment re peat the adjustment procedure with manually determined adjustment values see page 132 60 51 416 D 3352 130 D 3352 031 03 17 02
6. 1585 0 Dy 116 A2 0 By 0 Da 955 Eik 174 BF 0 Werte zurlicksetzen Werte zuruicksetzen Werte speichern Abbrechen 11 2 8Resetting the adjustment NOTICE Important Make sure to note down the values dis played in the text boxes before modifying them This will enable you to reset the adjustment values to the factory settings if necessary Contact the SIRONA Customer Service Center for more information or to enable the menu Phone 0 62 51 16 16 70 The DIAPHRAGM SYSTEM ADJUSTMENT menu offers you the possibility of resetting or manually modifying any or all of the pan adjustment settings you have made in very exceptional cases e To reset the pan settings open the PAN RESET AD JUSTMENT menu 1 IMPORTANT If the adjustment values have been reset the unit must be readjusted 60 51 416 D 3352 134 D 3352 031 03 17 02 Sirona Dental Systems GmbH 11 Appendix Installation Instructions ORTHOPHOS XG 3 11 2 Adjusting the panoramic X ray unit Blenden Geratejustage kammer3 Aufnahmemodus Ray C Dunkelstrom Testumlaut Blende offen I Quickshot Bildaufnahme Abbrechen 11 2 9Sirona Service Sheet This menu is intended only for internal use and is not required for installation 60 51 416 D 3352 D 3352 031 03 17 02 135 11 Appendix Sirona Dental Systems GmbH 11 3 Demo mode Installation Instructions ORTHOPHOS XG 3 11 3Demo mode
7. Manual adjustment of the PAN sensor The manual adjustment procedure is similar to the one for automatic adjustment The only difference is that the default adjustment values automatically determined by SIDEXIS are overwritten by manually determined adjustment values in the PAN SENSOR ADJUSTMENT submenu 1 Measure distances 1 S2 and S3 with the SIDEXIS measuring ruler IMPORTANT To measure S1 S2 and S3 estimate the horizontal center position of the displayed needles Measure in the lower area of the needles if possible since they may have been bent Slightly after repeated use Tip To facilitate the measuring procedure you can color the image in SIDEXIS see also SIDEXIS Operator s Manu al 120 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 Diaphragm system adjustment kammer3 xj stent Pan Fixed diaphragm Pan Symmetry Pan Reset adjustment 4 gt IV Precision adjustment Image acquisition i i l i gt Ste gt 52 e gt 53 Save values 4 gt Cancel 2 Overwrite the default values for S1 S2 and S3 with the measured values in the text boxes of the PAN SENSOR ADJUSTMENT submenu IMPORTANT For information on the direction of displacement input of sign in the menu see page 110 Use points as decimal separators e Proceed with step 5 of the adjustment procedure on page 117 60 51 416 D 3352 D 3352 031 03 17 02 Si
8. Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 3 Installation Panoramic X ray unit 3 4 Installing the floor stand option 2 e Level the unit again in both directions with the help of the spirit level see step 12 and tighten the screws on the wall holder firmly 60 51 416 D 3352 D 3352 031 03 17 02 5x 10 x 160 17 Mount the unit onto the floor Drill the fastening holes in the floor through the holes in the base plate Remove the drilling dust with a vacuum cleaner Slide wall plugs X through the base plate and into the drilled holes Use the five screws Y and washers to screw the base plate firmly onto the floor 39 3 Installation Panoramic X ray unit Sirona Dental Systems GmbH 3 5 Removing the transport safety device Installation Instructions ORTHOPHOS XG 3 device A he Un h 4 7 a pe te a a eF gr a a Ey E 7 qh apa ea salle a Sa Ng 8 1 Remove the transport safety devices B prior to unit startup IMPORTANT Keep transport safety devices B in a safe place You will need them should the unit be moved again 60 51 416 D 3352 40 D 3352 031 03 17 02 Sirona Dental Systems GmbH 3 Installation
9. 1 supported by SIDEXIS function required only if not performed by an additional server 2 Declaration acc to Medical Devices Directive MDD 93 42 EEC and IEC 60 601 1 1 item 3 201 4 o The undersigned herewith certifies that the above system consisting of CE marked medical devices has been assembled in accordance with the present manufacturer s instructions n that it has been tested for its intended use and that all relevant user information including pertinent information provided by the corresponding manufacturers has been provided for placing it on the market and that this declaration shall be kept on hand for the competent authorities for a period of five years 2 z Place date Name of system integrator in block letters Signature 5 a 0 D 3352 128 01 04 02 12 2010 Nr 113 429 59 92 545 D 3352 2 Complete the Declaration of Conformity The system integrator certifies the CE conformity of the installed SIDEXIS system by means of the compatibility list The list is designed so that only the white fields must be filled in The original list remains in the practice the co py is given to the system integrator Only for Germany Perform the acceptance test required according to Art 16 of the X Ray Ordinance now Use the test phantom plug in plate contrast elements needle phantom and relevant instructions included in delivery for this purpose In Germany File in the X ray System Logbook 60 51 416 D 3352 D 33
10. 230 V gV e Remove front cover for details see Service Manual 230 240V 7 5 V 2 Select 300 VAC line voltage range on multimeter _ Connect measuring leads to terminal K1 L and N aoe cy 3 Connect power and switch unit ON Record reading Wait 1 min for self adjustment of the unit Press key R to return X ray tube head into the initial position e Turn unit OFF e Remove meter leads and refit cover 4 Establish exposure readiness via SIDEXIS IMPORTANT If the voltage drop is not within the specified range advise the customer that an adequate power supply must be installed Refer to Pre Installation Instructions Disconnect unit and do not release for use 5 Select highest exposure level e g 90 kV 12 mA A DANGER RADIATION Depress the exposure key until meter reading is obtained 60 51 416 D 3352 D 3352 031 03 17 02 15 8 Startup for USA Canada only Sirona Dental Systems GmbH 8 3 Tube Current Verification Installation Instructions ORTHOPHOS XG 3 8 3 Tube Current Verification e WARNING 1 Remove cover for details see Service Manual The electronics of the X ray tube assembly are always 2 Loosen the four screws A and remove cover plate B of connected to line voltage the electronics box Always switch the X ray unit off and wait until V203 is no longer illuminated before contacting the test leads Z WARNING The test leads and measuring instruments used must have a dielectric stre
11. Panoramic X ray unit Installation Instructions ORTHOPHOS XG 3 3 6 Installing the release button holder 3 6 Installing the release button holder IMPORTANT Only install the holder to the unit if you do not intend to use a remote control If you use a remote control in combina tion with the release button attach the holder to the remote control see page 49 1 Unscrew and remove cover A Attach holder B and screw the cover back onto the unit 60 51 416 D 3352 D 3352 031 03 17 02 41 3 Installation Panoramic X ray unit Sirona Dental Systems GmbH 3 7 Attaching the covers Installation Instructions ORTHOPHOS XG 3 3 7 Attaching the covers 4 Use the 4 screws A to refasten the assembled housing IMPORTANT cover to the unit These covers must be attached only if the unit is installed with the floor stand and does not stand against a wall 1 Loosen the 4 screws A on the upper housing cover B and remove this housing cover toward the rear 2 Insert the cover C included in the floor stand accesso ries in the outer cover B 3 Fasten the cover C to the cover B with the 3 spring steel clamps D 60 51 416 D 3352 42 D 3352 031 03 17 02 Sirona Dental Systems GmbH 3 Installation Panoramic X ray unit Installation Instructions ORTHOPHOS XG 3 3 7 Attaching the covers y Ni
12. 3352 D 3352 031 03 17 02 Sirona Dental Systems GmbH 6 Safety checks Installation Instructions ORTHOPHOS XG 3 6 1 Checking the protective ground wires u Power cable ORTHOPHOS XG Second protective ground wire C Measuring points B and C GNYE power connection and 2nd ground to the ORTHOPHOS XG to the ORTHOPHOS XG Power cable Second Protective ground wire Power cable Protective ground wire 60 51 416 D 3352 D 3352 031 03 17 02 61 6 Safety checks Sirona Dental Systems GmbH 6 1 Checking the protective ground wires Installation Instructions ORTHOPHOS XG 3 LL A a os ot NN K eK SA Z D j LN YA SS Ue X ray tube assembly Measuring points D and E IMPORTANT If the resistance exceeds the value specified in the table check whether the protective ground wire is fastened ac cording to specifications Check whether the flat washer toothed lock washer and cable lug are fitted on the protective ground wire in the right order see page 61 and whether the nuts of the ground wire connections are tightened securely If the ground wire is not fastened according to specifica tions fasten the ground wire properly see page 61 Do not connect the power cable and the second ground wire to the building installation yet but perform a mea surement of the device leakage current f
13. 60 51 416 D 3352 126 D 3352 031 03 17 02 Sirona Dental Systems GmbH 11 Appendix Installation Instructions ORTHOPHOS XG 3 11 2 Adjusting the panoramic X ray unit ZOOM 1 1 B 0 75 mm gt a A1 88 6 mm 1 mm A2 44 3 0 5 mm q _ gt _ pl np a2 44 3 0 5 mm Adjustment ok length measurement with SIDEXIS Adjustment not ok 60 51 416 D 3352 D 3352 031 03 17 02 127 11 Appendix 11 2 Adjusting the panoramic X ray unit 4 Evaluate the X ray image see page 127 The shadow of the center needle the needle image and the auxiliary line must be coincident and located behind each other A tolerance offset of needle from the central auxiliary line of 0 75 mm is admissible A Distance A1 must be 88 6 1 mm A Distances A2 must be identical each being 44 3 0 5 mm A A white border surrounding the image on all sides must be visible B IMPORTANT If one of these criteria is not fulfilled C the pan symmetry must be adjusted 128 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 17 02 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 11 Appendix 11 2 Adjusting the panoramic X ray unit J Diaphragm system adjustment kammer3 Pan Sensor adjustment Pan Fixed diaphragm Pan Symmetry Pan Reset adjustment 4 S1 2 10 mm s2 200mm s3 1 50mm
14. Before you begin Sirona Dental Systems GmbH 1 6 Installation versions Installation Instructions ORTHOPHOS XG 3 60 51 416 D 3352 12 D 3352 031 03 17 02 The Dental Company 2 Pe and transport ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 17 02 13 2 Delivery and transport 2 1 Delivery 2 1 Deliver Dimensions cm 199 x 69 x 122 inches 78 3 8 x 30 3 4 x 28 3 4 Weight 177 kg 390 Ibs NOTICE Possible transport damage If the shipment was damaged during transport document all damage carefully and contact the responsible carrying agent immediately All SIRONA equipment is carefully checked and packed prior to shipment Please carry out an incoming inspection of the equipment in order to make sure that it was not dam aged during transport e Check the packaging and the equipment for visible signs of damage e Check the shipment for completeness based on the at tached scope of supply checklist IMPORTANT Disposal Return the packaging materials to SIRONA or dispose of them in compliance with the legal regulations ap plicable in your country 14 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 IMPORTANT The packaging of the X ray unit is designed both for protec tion during transport and as an installation aid Therefore please remove only the surrounding packaging prior to installation Please leave the styrofoam packaging and transport pallet attached to
15. D 3352 031 03 17 02 71 7 Initial startup Sirona Dental Systems GmbH 7 4 Checking the data paths Installation Instructions ORTHOPHOS XG 3 60 51 416 D 3352 72 D 3352 031 03 17 02 Kr The Dental Company 8 slats for USA Canada only ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 17 02 73 8 Startup for USA Canada only 8 1 Startup measurements and controls Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 8 1 Startup measurements and controls Required measuring instruments 1 Digital multimeter Fluke 87 III or equivalent Accuracy DC voltage 0 1 of reading plus 0 02 of range DC current 0 4 of reading plus 0 1 of range 2 Adose measurement device e g Mult O Meter type 510L is required for dosimetry e Radiation protection Observe the radiation protection guidelines as outlined in the Operating Instructions manual X radiation is emitted as long as the exposure key on the Multitimer is depressed The X ray indicator must light up on the Multitimer during radiation An acoustic signal must also be heard Power supply adequacy To assure that the ORTHOPHOS XG system performance is in accordance with Sirona specifications an adequate power supply for permanent installation is essential The Federal Performance Standard for Diagnostic X ray Units Code of Federal Regulations Title 21 CFR Subchapter J mandates an adequate power supply 74 Duty cy
16. For connection proceed as follows NOTICE Use a 3 wire cable 1 5 mm to connect the X ray warning lamp A maximum load of 50 W is permissible and no additional circuit may be connected 60 51 416 D 3352 D 3352 031 03 17 02 Insert cable A for connection of the X ray warning lamp in the chassis concealed or surface installation and attach it to the unused strain relief Connect the current carrying cables and the ground wire to terminal X105 board DX42 as shown in the connection diagram above Secure the current carrying cables with a cable tie B to prevent them from slipping out Run cable A parallel to control cable L117 before clos ing the housing 57 5 Installation Remote control Sirona Dental Systems GmbH 5 6 Final work Installation Instructions ORTHOPHOS XG 3 5 6 Final work 1 DHHS label WARNING 1 Reassemble the remote control and place the release button if installed in the holder NOTICE Make sure that no cables are pulled off when you clip the cover on IMPORTANT For operation with a remote control the unit must be config ured accordingly After initial startup of the unit check the configuration with the help of service routine S017 test step 6 see section 11 1 6 2 Forthe USA Canada only Attach the DHHS and warn ing labels 60 51 416 D 3352 58 D 3352 031 03 17 02 Kr The Dental Company 6 SEL checks ORTHOPHOS XG 3 60 51 4
17. Primary 4 gt S1 1 34 mm 4 l 52 1 18 mm 53 1 04 mm IV Precision adjustment Image acquisition l i j i gt i Sim gt j 52 gt 3 je Save values gt Cancel 2 ama V NI 9a AN A oo i OOOO Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 in SIDEXIS Preview i for exposure E sirona 9 1 4 Checking the PAN symmetry A WARNING When performing the following tests be sure to ob serve the radiation protection regulations applicable in your country see Operating Instructions e Insert the needle phantom in the bite block holder of the panoramic X ray unit see page 85 IMPORTANT The sensor must be plugged into the sensor slot on the pan oramic X ray unit 1 Goto the PAN SYMMETRY submenu 2 To make SIDEXIS XG ready for exposure Click IMAGE ACQUISITION 88 The exposure dialog box showing the exposure status appears in Sidexis The initialization status is visualized by a progress indi cator on the Multipad The initialization procedure is completed when the ex posure parameters of service routine 010 2 60 kV 3 mA 14 1 s are displayed and the progress indicator disappears Take an exposure 60 kV 3 mA Press the R key on the Multipad to move the unit back to the starting position Press the release button Hold down the release button until image acqui
18. Protective ground wire e Use the emergency shutdown button that is integrated into the power cable to switch on the unit 60 51 416 D 3352 48 D 3352 031 03 17 02 The Dental Company 5 Installation Remote control ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 17 02 49 5 Installation Remote control Sirona Dental Systems GmbH 5 1 Installation material tools Installation Instructions ORTHOPHOS XG 3 5 1 Installation material tools 1 Installation material for remote control 2 Required tools Wood screw 4x30 3 16x1 1 4 3 pc Masonry drill amp ta 6 mm 1 4 Plastic wall plug S6 3 pc Impact drill or percussion drill Blade screwdriver small and medium sized Torx offset screwdriver TX20 Hammer Center punch awl 60 51 416 D 3352 50 D 3352 031 03 17 02 Sirona Dental Systems GmbH 5 Installation Remote control Installation Instructions ORTHOPHOS XG 3 5 2 Mechanical installation 5 2 Mechanical installation 1 Carefully press the blade screwdriver in groove A from below do not pry and remove the chassis to the rear IMPORTANT For concealed installation the control cable is drawn into Hold the chassis against the wall in its mounting posi the chassis from the rear For surface installation it is drawn tion and mark the positions for the three drill holes with in from underneath an awl For concealed installation rii h a a a ue 2 D
19. Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 1 Before you begin 1 4 Dimensions Space requirements 1 4 Dimensions Space requirements min 1411 55 1 2 1371 54 o min 1280 50 3 8 h Sry J Ts d i i 2249 71 1 2 1950 76 3 4 402 5 15 7 8 NOTICE The minimum ceiling height should be 2 10 m 82 11 16 If the ceiling height is lower than 2 27 m 89 3 8 max travel height of 2 25 m 88 1 2 the travel height of the unit must be adjusted or limited prior to startup of the unit see section 11 1 7 60 51 416 D 3352 D 3352 031 03 17 02 1970 77 1 2 NOTICE The dimensions specified here apply to installation of the X ray unit without the floor stand Installation with the floor stand results in an additional 30 mm 1 3 16 increase of all height dimensions 1 Before you begin 1 5 Mounting options Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 1 5 Mounting options 1 2 Standard version MUN CON Standard version see Section 3 3 1 Wall mounted installation with 1 wall holder and floor fastening if both wal
20. Sirona Dental Systems GmbH 11 Appendix Installation Instructions ORTHOPHOS XG 3 11 2 Adjusting the panoramic X ray unit 8 Diaphragm system adjustment kammer3 Pan Sensor adjustment Pan Fixed diaphragm Pan Symmetry Pan Reset adiustment 4 gt S1 0 00 mm m S2 0 00 mm 53 0 00 mm E Er Image acquisition St i Ii Hl 3 1 2 X Save values S a ad Cancel 8 Ifall criteria are fulfilled and the current image is identi cal to the ideal image A B then save the values Click SAVE VALUES IMPORTANT In SIDEXIS versions V01 50 and higher the values for S1 S2 and S3 in the PAN SYMMETRY submenu are set equal to zero with a correct adjustment i e if they are within the permissible tolerance For versions lt V01 50 the values in the text boxes may deviate slightly from zero even with a correct adjustment e The adjustment of the panoramic X ray unit is now completed 60 51 416 D 3352 D 3352 031 03 17 02 131 11 Appendix 11 2 Adjusting the panoramic X ray unit Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 Manual adjustment of the PAN symmetry The manual adjustment procedure is similar to the one for automatic adjustment The only difference is that the default adjustment values automatically determined by SIDEXIS are overwritten by manually determined adjustment values in the PAN SYMMETRY submenu 1 Measure distances 1 S2 and S3 with t
21. aE 45 4 1 Connecting the control cables PAN cc ssscceccsseeeeeseeseeeeeseeeeeeeaeees 46 4 2 Connecting the line Voltage ce ecececcseeseeeeeeeeeaneeeeeeeeeeaaeeeeeeeeeaaaaes 47 Installation Remote control ccceccececsesececeecececuecececueeececueeececeeneeecueaececeeuececeeasceseeueness 49 5 1 Installation material tools ccccccccseeeeeeseeeeeeceeeeeeeeseeeeeeeeaaeeeeesaeseeeeaas 50 5 2 Mechanical INStAIATON x ssiccssacecewenstcetnevecsddesvneceahaestuapentcesieetervensieonesteein 51 5 38 Connecting the control cables REMOTE cc sseeeeeeseeeeeeeeeeeeeeens 53 5 4 Connecting the door Contact SWItCN ccccccsseeeeeeeeeeeeeeeeeeeeeeeseeeeeeens 56 5 5 Connecting the X ray warning lAMP cc eeeeceeeeeeeeeeeeeeeeeeeeeeeeeeeeeneaaees 57 5 6 Gc le 0 Gree teenirre rma reen ters S 58 SLORY CNECKS corna a a A ta uaateenmetaae 59 6 1 Checking the protective ground WireS s ssessesssrrrsresssrrrrrresrrrrrreee 60 6 2 Checking the device leakage CUITeNL ccceeeeeeeeeeeeeeeeeeeeeeaeeeeeeeeeaeas 63 1 PENNE Leal Startup ess cele ce cacan ds eee soe tied ees acca wtewnes a Aap Ean 65 7 1 Inserting the forehead and temple supports ssseeeeeesseeeeeeeeeeeeeees 66 7 2 Inserting the Sensor oxicis ects aicanetc ecdenc ade cushateediitde ane dstaapeweedaeeacs Wieacss 67 7 3 Switching the units ON cccccccssseeeeeceeeeeeeeceseeeceeeeeee
22. before laying it down 60 51 416 D 3352 D 3352 031 03 17 02 11 Appendix 11 3 Demo mode 139 We reserve the right to make any alterations which may be required due to technical improvements Sprache englisch Printed in Germany D 3352 031 03 17 02 07 2013 A Nr 117540 Imprim en Allemagne Sirona Dental Systems GmbH in the USA FabrikstraBe 31 Sirona Dental Systems LLC D 64625 Bensheim 4835 Sirona Drive Suite 100 Order No 60 51 416 D 3352 Germany Charlotte NC 28273 www sirona de USA
23. directions indicated by the plus and minus signs located below and next to the pictograph refer to shifting of the exposed image area in the direction of the stationary auxiliary lines see the following example Sx q_ _ gt In the example the exposed image area is offset to the left by the value Sx and upward by the value Sy In order to shift the image area so that it comes to lie inside the auxiliary lines you must enter e Sx shift to the right with a positive sign e Sy shift downward with a negative sign in the text boxes of the submenu IMPORTANT The pan diaphragm cannot be adjusted via the software i e the pan diaphragm must be adjusted by moving it mechani cally see section 11 2 6 110 Generally speaking the exposed image area must always be shifted toward the auxiliary lines Shift to the right or upward Enter the value measured offset from the auxiliary line with a positive sign or mechanically shift the diaphragm to to the right or upward Shift to the left or downward Enter the value measured offset from the auxiliary line with a negative sign or mechanically shift the diaphragm to the left or downward Displays on the Multipad During the adjustment procedure different service routines are started from Sidexis they are displayed one after the other on the Multipad If there are any error messages displayed during adjust ment please follow the instructi
24. fasten the release button hold er B to the keyboard Plug the connector of the coiled cable into socket X101 on board DX42 and screw the connector down tight Hang the coiled cable in strain relief E of the assembled chassis 60 51 416 D 3352 D 3352 031 03 17 02 Sirona Dental Systems GmbH 5 Installation Remote control Installation Instructions ORTHOPHOS XG 3 5 3 Connecting the control cables REMOTE Connecting the control cable e Connect control cable L117 as described in section 5 3 on page 53 60 51 416 D 3352 D 3352 031 03 17 02 55 5 Installation Remote control Sirona Dental Systems GmbH 5 4 Connecting the door contact switch Installation Instructions ORTHOPHOS XG 3 5 4 Connecting the door contact switch Door contact switch Whitegreen Green Brown Whitebrown BU WHOG Blue Whiteorange WHBU OG Whiteblue Orange 1 Connect the door contact switch between terminal X 108 pin 1 board DX 42 and control cable L117 pin 1 WHGN 60 51 416 D 3352 56 D 3352 031 03 17 02 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 5 Installation Remote control 5 5 Connecting the X ray warning lamp 5 5 Connecting the X ray warning lamo OF RN X RAY Alarm It is possible to activate an X ray warning lamp via the remote control
25. recessed drill holes of the adjustment plate must point downward 60 51 416 D 3352 32 D 3352 031 03 17 02 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 SS 3 Installation Panoramic X ray unit 3 4 Installing the floor stand option 2 Q 3xM10 SA 3x ota 10 9 Position the base plate with the threaded bolts includ ing the mounted support on the adjustment plate and attach the base plate loosely with the 3 adjustment nuts Q and washers NOTICE Make sure that the cables are fed through the support cor rectly and are not crushed 60 51 416 D 3352 D 3352 031 03 17 02 33 3 Installation Panoramic X ray unit Sirona Dental Systems GmbH 3 4 Installing the floor stand option 2 Installation Instructions ORTHOPHOS XG 3 10 11 N z ee D UH eA n S NEIKE lt Q er L il mil D gt o Sa e A O S yeu eS NUNES ites T 1x Profile clamp U Nik Li ri
26. see the Operating Instructions supplied with the media convert 2 Connect the coiled cable of the release button to socket er X103 of the panoramic X ray unit e Install the media converter at a suitable location using fastening screws or the Velcro strap supplied for this purpose For the installation version with remote control 3 Connect the remote control to socket X103 of the pan oramic X ray unit with cable L117 NOTICE Secure the connector on socket X103 Do not fasten the media converter to the X ray unit not included in the scope of supply 60 51 416 D 3352 46 D 3352 031 03 17 02 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 A DANGER Fixed connection The installation of a power plug instead of the pre scribed fixed hard wired connection violates interna tional medical regulations and is prohibited In case of a fault you would thus endanger the life and limb of the patient the operator or other persons A DANGER Danger of electrical shock Be sure to switch off the line power supply before con necting the line voltage 1 Power cable ORTHOPHOS XG Second protective ground wire ORTHOPHOS XG 3 e The unit is suitable for connection to networks of 200 240 V and 50 60 Hz 10 60 51 416 D 3352 D 3352 031 03 17 02 4 2 Connecting the line voltage 4 Electrical connection 4 2 Connecting the line voltage WARNING Be sure to connect the second p
27. the restart the previously set minimum travel The current height position is displayed in selection height is cancelled field 1 The LED above the Memory key lights up 2 To cancel the limitation for the minimum travel height first press the Memory key LED above the R key lights up and then the R key 60 51 416 D 3352 108 D 3352 031 03 17 02 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 11 Appendix 11 2 Adjusting the panoramic X ray unit 11 2Adjusting the panoramic X ray unit Diaphragm system adjustment kammer3 Pan Sensor adjustment Pan Fixed diaphragm Pan Symmetry Pan Reset adjustment 4 gt S1 m S2 0 00 mm 53 0 00 mm IV Precision adjustment Image acquisition Save values Cancel gt j Ste gt j 82 e gt j S3 je 4 AN AN a SERVICE 11 2 1Diaphragm system adjustment menu The DIAPHRAGM SYSTEM ADJUSTMENT menu guides you through the adjustment of the panoramic unit This ser vice routine is started from SIDEXIS XG UTILITIES gt CONSTANCY TEST gt XCXP gt SELECT X RAY DEVICE gt SERVICE EXPOSURE gt PASSWORD INPUT gt SELECT X RAY COMPONENT gt DIAPHRAGM TEST EXPOSURE IMPORTANT The DIAPHRAGM SYSTEM ADJUSTMENT menu is pass word protected As password enter the first four digits of the current system date PC in reverse order Example On 05 30 2004 the service password is 5003 IMPORTANT Th
28. the unit Save one of the lateral styrofoam packaging parts for later use as an installation aid A Scope of supply e Panoramic X ray unit e Profile cover D e Sensor B e Accessories and hygienic protective covers see pp 16 ff B e Installation material see section 3 1 B e Safety strap B e Remote control optional C 60 51 416 D 3352 D 3352 031 03 17 02 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 2 Delivery and transport 2 1 Delivery Two shock indicators A are attached to the side of the packaging to indicate whether the unit was exposed to a shock during transport e White indicator No shock e Red indicator Shock A tilt indicator B that enables you to recognize whether the unit was transported improperly is also attached to the packaging e Red indicator Improper transport The display of improper transport doesn t necessarily mean that the unit is damaged Make a note on the delivery slip that the indicator is acti vated Have this confirmed on the delivery note by the driver of the transport company Fax the delivery slip to the Sirona Customer Service Center CSC Enter the state of the indicators in the startup report in the case of warranty claims 60 51 416 D 3352 D 3352 031 03 17 02 15 2 Delivery and transport 2 1 Delivery Accessories Panoramic X ray unit 1 Forehead 1x and temple supports 2x 2 Buttons 2x 3 Temporoma
29. was not inad vertently deadjusted during transport The TEST EXPOSURE menu is started from SIDEXIS XG UTILITIES gt CONSTANCY TEST gt XCXP gt SELECT X RAY DEVICE gt SERVICE EXPOSURE gt SELECT X RAY COMPONENT gt DIAPHRAGM TEST EXPOSURE IMPORTANT The SELECT X RAY DEVICE and SELECT X RAY COMPO NENT dialogs appear only if several x ray devices or x ray components are available for selection IMPORTANT When you open the TEST EXPOSURE menu the unit switches from the user mode to the PC service mode logged by the PC The service mode is indicated by the SERVICE display on the Multipad 2 In the PC service mode the control options that are available on the Multipad are determined by SI DEXIS and the currently selected service routine General control of the unit by means of the Multipad as in the user mode is not possible in this mode The test exposures are used to check the adjustment They can be started without a password 84 You can switch between the PAN DIAPHRAGM and PAN SYMMETRY menus by clicking the corresponding tab with the mouse To quit the TEST EXPOSURE menu click CANCEL Displays on the Multipad During the adjustment procedure different service routines are started from SIDEXIS they are displayed one after the other on the Multipad If there are any error messages displayed during adjust ment please follow the instructions provided in the Service Manual 60 51 416 D 3352 D 33
30. 16 D 3352 D 3352 031 03 17 02 59 6 Safety checks 6 1 Checking the protective ground wires Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 6 1 Checking the protective ground wires Measuring setup for protective ground wire test A DANGER Shock hazard It is essential that you switch the X ray unit OFF before replacing any components If you have not done it already Switch OFF the line voltage at the main switch of the building installation Disconnect the power cable and the second protective ground wire from the building installation Remove the following cover parts Profile cover Tube assembly cover front Tube assembly cover rear 60 Ammeter Measuring points B E Measuring point Power source Central ground wire Voltmeter moO W Check whether the ground wire resistance complies with the specifications Protective ground wire test between Aand B GNYE wire 0 10 Aand C 2 Protective 0 10 ground wire Aand D Housing DX32 0 20 Aand E Tube assembly 0 20 housing A power source with a current of at least 0 2 A and a no load voltage of 24 V max and 4 V min is required Connect the power source between the measuring points specified in the table for at least 5 s and mea sure the voltage drop with the voltmeter the current with the ammeter and calculate the resistance using the formula R U I 60 51 416 D
31. 51 0 00 mm s2 omm 53 0 00 mm IV Precision adjustment l l Image acquisition l i Ji Sije i 82 je i S3 e Save values gt Cancel 8 Take an exposure 60 kV 3 mA Press the release button Hold down the release button until image acquisition is completed and the preview image appears in the exposure dialog box 9 Evaluate the X ray image The three needle images must lie in the center of the exposed areas and inside the auxiliary lines A IMPORTANT If these criteria are not yet fulfilled repeat the adjustment procedure starting with step 5 NOTICE If you do not reach your goal via automatic adjustment re peat the adjustment procedure with manually determined adjustment values see page 120 60 51 416 D 3352 D 3352 031 03 17 02 11 Appendix 11 2 Adjusting the panoramic X ray unit Adjustment ok 10 If the image is identical to the ideal image A save the values Click SAVE VALUES IMPORTANT In SIDEXIS versions V01 50 and higher the values for S1 S3 in the PAN SENSOR ADJUSTMENT submenu are set equal to zero with a correct adjustment i e if they are within the permissible tolerance For versions lt V01 50 the values in the text boxes may deviate slightly from zero even with a correct adjustment e Go onto the next adjustment step e Reattach the fixed diaphragm see page 113 119 11 Appendix 11 2 Adjusting the panoramic X ray unit
32. 52 031 03 17 02 Sirona Dental Systems GmbH 10 Final work Installation Instructions ORTHOPHOS XG 3 10 4 Unit handover 10 4Unit handover 1 Give the customer the technical documentation as well as all of the patient positioning aids test phantoms and special tools included in delivery including packaging These are important components of the unit and must be stored carefully 60 51 416 D 3352 D 3352 031 03 17 02 95 10 Final work 10 4 Unit handover 96 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 17 02 Kr The Dental Company 11 Ap pendix ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 17 02 97 11 Appendix 11 1 Service routines for installation Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 11 1Service routines for installation Service routine Designation Described in this Appendix Page 17 2 Configuring the hardware version Configuring the hardware version 103 17 6 Configuring the remote control Enable disable the remote control 104 18 2 Setting the maximum Travel height Setting the maximum travel height 105 for the unit 18 3 Undoing the maximum travel height Undo the maximum travel height set 106 setting ting 18 5 Setting the minimum Travel height Setting the minimum travel height 107 for the unit 18 6 Canceling the limitation of minimum Canceling the li
33. 52 031 03 17 02 Sirona Dental Systems GmbH 9 Checking the system adjustment Installation Instructions ORTHOPHOS XG 3 9 1 Test exposures Needles on Top side 9 1 2 Needle phantom In order to check the pan symmetry see section 9 1 4 you must insert needle phantom A in the bite block holder of the panoramic X ray unit IMPORTANT When fitting the needle phantom make sure that it is cor rectly oriented For the adjustment of the X ray unit the phantom must be fitted in such a way that the needles point upward 2 60 51 416 D 3352 D 3352 031 03 17 02 85 9 Checking the system adjustment Sirona Dental Systems GmbH 9 1 Test exposures Installation Instructions ORTHOPHOS XG 3 1 in SIDEXIS Preview Exposure Preview p for exposure E sirona 9 1 3 Checking the diaphragm adjustment A a a a a The exposure dialog box showing the exposure status WARNING appears in Sidexis When performing the following tests be sure to ob serve the radiation protection regulations applicable in your country see Operating Instructions The initialization status is visualized by a progress indi cator on the Multipad The initialization procedure is completed when the ex e Plug the sensor into the sensor slot on the panoramic posure parameters of service routine S030 2 X ray unit 60 kV 3 mA 0 20 s are displayed and the progress indicator disappears IMPORTANT The needle phantom must
34. Acceptance test according T to DIN V 6868 151 for all ype countries Acquisition gen eration with image storage No and display successful Manuf Sirona Dental Systems CE marking in compliance with see Installation instructions For the FRG Q gt Council Directive 93 42 EEC Acceptance test according um a to DIN V 6868 151 for all g ype countries Acquisition gen O wu eration with image storage xQ No and display successful A Manuf Sirona Dental Systems CE marking in compliance with see Installation instructions For the FRG Q z Council Directive 93 42 EEC Acceptance test according Of T to DIN V 6868 151 for all o ype countries Acquisition gen E eration with image storage o No and display successful Manuf Sirona Dental Systems CE marking in compliance with see Installation instructions For the FRG Q Council Directive 93 42 EEC Acceptance test according TEB to DIN V 6868 151 for all Yp countries Acquisition gen eration with image storage No and display successful o IEC 60 950 1 and CE marking Device installed according 2 manui goid R Council Directive tomangmoiers institices Q 89 336 EE 23 Type z 3 q No gt IEC 60 950 1 and CE marking Device installed according 2 Manut according to Council Directive Q to manufacturer s instruc Q S Elm k tions B amp Yp O 3 fi No So IEC 60 950 1 and CE markin Device installed accordin 3 Manut according to Council Directive Q em ecan iste Q Type eas e No
35. Dental Company 3 Installation Panoramic HS unit ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 17 02 19 3 Installation Panoramic X ray unit Sirona Dental Systems GmbH 3 1 Installation material Installation Instructions ORTHOPHOS XG 3 3 1 Installation material J Sirona for installation on wooden stud for installation a frame on wooden stud i sib i Standard version Option 1 with second wall holder see Section 3 3 see Section 3 3 1 Wall floor mounting 2 Additional wall holder for bottom wall mounting Hexagon wood screws 8x80 5 16x3 4 pc Wall holder 1 pc Plastic wall plug S10 2 pc Wood screws 8x80 5 16x3 2 pc Screw M8x30 2 pc Washer dia 8 4 3 pc Washer sia 8 4 6 pc Hexagon head screw M8x50 1 pc Nut M8 2 pc Nut M8 1 pc Screw M4x10 3 pc Mounting kit dia 10 SXR 2 pc Washer ota 4 3 2 pc Profile clamp 1 pc Mounting kit dua 10 SXR 2 pc Cover for wall holder 2 pc Torx offset screwdrivers TX10 TX20 TX25 resp 1 pc Offset Allen key size 6 1 pc 20 Wall holder 1 pc Cover for wall holder 2 pc Interm
36. F Image acquisition Save values e a Cancel sirona Sule V JY o o o A gY C0 HOOD IMPORTANT The default values for S1 S2 and S3 were automatically de termined by SIDEXIS based on the exposure and entered in the text boxes of the menu For manual adjustment the values displayed at this position in the text boxes of the menu can be overwritten First continue with the automatic adjustment Manual deter mination of the adjustment values is required only if you fail to reach your goal via automatic adjustment see page 132 60 51 416 D 3352 D 3352 031 03 17 02 in SIDEXIS Preview p for exposure a sirona Make SIDEXIS XG ready for exposure Click IMAGE ACQUISITION The exposure dialog box showing the exposure status appears in Sidexis The initialization status is visualized by a progress indi cator on the Multipad The initialization procedure is completed when the ex posure parameters of service routine S010 2 60 kV 3 mA 14 1 s are displayed and the progress indicator disappears Take an exposure 60 kV 3 mA Press the release button Hold down the release button until image acquisition is completed and the preview image appears in the exposure dialog box 129 11 Appendix Sirona Dental Systems GmbH 11 2 Adjusting the panoramic X ray unit Installation Instructions ORTHOPHOS XG 3 0 75 mm lt _ lt _
37. Indicateurs de basculement intact oul Indicador de vuelco correcto si A Ersch tterungssensor a b ok E J Vibration sensor a b ok YES Capteur de vibration a b intact oul Sensor de sacudidas a b correcto si Software Hardware Logiciel _ Mat riel ORTHOPHOS XG ee SIDEXIS ion e Inbetriebnahme gem bauseitiger Installati Commissioning performed error free according Mise en service ex cut e conform ment aux Puesta en funcionamiento seg n las instruccio onsanleitung Montageschablone und Instal to the pre installation instructions mountin instructions d installation et aux instruc nes de instalaci n y de montaje plantilla de lationsanleitung sowie bedarfsweise l nder template and installation instructions as well as tions de pr installation calibre de montage montaje asi como si es necesario de haber spezifische sicherheitstechnische country specific safety controls where required et ventuels contr les de s curit sp cifi efectuado controles t cnicos de seguridad es ontrollen fehlerfrei durchgef hrt The unit including accessories and all docu gues au pays ex cut s sans d tection de pecificos del pa s exentos de defectos Das Ger t incl des Zubeh rs und aller Doku ments in accordance with the Operating Instruc La unidad incluyendo los accesorios y todos los mente gem Gebrauchsanweisung wurde _ tions was turned over to the th
38. MPORTANT If these criteria have not yet been fulfilled repeat the me chanical adjustment of the diaphragm see page 123 and then continue with step 5 60 51 416 D 3352 D 3352 031 03 17 02 Reattach the covers Go on to the next adjustment step 125 11 Appendix Sirona Dental Systems GmbH 11 2 Adjusting the panoramic X ray unit Installation Instructions ORTHOPHOS XG 3 in SIDEXIS Exposure Pan Sensor adjustment Pan Fixed diaphragm Pan Symmetry Pan Reset adjustment 4 a S1 E S2 0 00 mm E 53 0 00 mm E Image acquisition Save values 11 2 7Adjusting the PAN symmetry e Insert the needle phantom in the bite block holder of the 3 Take an exposure 60 kV 3 mA panoramic X ray unit see page 112 Press the R key on the Multipad to move the unit 1 Go to the PAN SYMMETRY submenu back to the starting position Press the release button Hold down the release button until image acquisition is completed and the preview image appears in the exposure dialog box 2 To make SIDEXIS XG ready for exposure Click IMAGE ACQUISITION The exposure dialog box showing the exposure status appears in Sidexis The initialization status is visualized by a progress indi cator on the Multipad The initialization procedure is completed when the ex posure parameters of service routine 010 2 60 kV 3 mA 14 1 s are displayed and the progress indicator disappears
39. TENDED DETAILS menu 13216015160 Subnet mask Eam Slonebard jarva aens e Changed Syitenparameter alee cae TS The following systemparamecers have been changed compared ta the factory S2t ings after replacing an B lt L1 Pce the listed values must be reconfigured Changed system conf igir ation S tt ings Serting of syotem version configured in 4 47 23 wa values il Sat tting oF Tanguage configured in S1F a actual value oedod erting of language set index cenfiqured in 27 1 actual value da0003 Tert tting of remore contra activation configured in 317 6 ual value g tive Set cing oF pai configuration configured in i7 9 eTual values mount 1 Ser ttir ng oF inage f ane sonf igur ed in SL 11 4 value enabled 01 cae ring oF prev Ua l iay ation seonfigured in 317 17 actu value disabled 00 zer tings af we came ser enn Sorfiguravton icon gured dn 417 14 ota ed 00 Et ings of CENH BO USUAENE stor i by the adjustment rout inet actual value galpha 7774 be by ache Sent ings of C amp PH ad ud inert guicks shot Stered by the adjudteernt raut ined pr 40060 attual value Galpha 303 D Set erin ng of ring Type Configured in 53 actual value 1 Settings of network configure at fan actu value fe 198 168 15 176 Subnet 255 255 295 0 gateway 192 165 13 1 Changed Wier Preference Sot tings Zrt ting of default pati entsymbol PAH Kconfigured
40. With SIDEXIS Version V02 20 and higher a message win dow indicates when a coarse adjustment is necessary Only if you cannot achieve your goal with precision adjust ment e g if the exposed area is completely outside the im age field should you perform a coarse and then a precision adjustment If a coarse adjustment proves necessary deactivate the PRECISION ADJUSTMENT check box and follow the adjust ment steps described in the present chapter to perform a coarse adjustment The steps and correction procedure required for coarse ad justment are identical to those for precision adjustment The only difference between the two modes is the size of the im age section considered Furthermore there are fewer auxil lary lines in the coarse adjustment mode Default values in the Diaphragm system adjustment menu in SIDEXIS XG During the adjustment the default adjustment values are displayed in the text boxes of the Diaphragm system adjust ment menu First perform the adjustment with these default values If you do not attain the desired result via this automatic adjustment you should determine the adjustment values manually by measuring the exposure with the SIDEXIS measuring ruler and then overwrite the default values in the menu This procedure is described in the following sections 111 11 Appendix Sirona Dental Systems GmbH 11 2 Adjusting the panoramic X ray unit Installation Instructions ORTHOPHOS XG 3 Needles on To
41. ann ENS Ne Omer CUCTEMCOCTI MINL SYSTER SOE TII Prog V 1 2 3 Oooo S through the list until 60 51 416 D 3352 136 D 3352 031 03 17 02 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 11 Appendix 11 3 Demo mode 11 3 1 Switching the demo mode ON When operated in demo mode the unit must not release any radiation For this reason you must take the following safety measures 1 gt 2 N Remove the cover of the tube assembly Loosen screws A and remove cover plate B Set DIP switch S2 DX6 to position 2 Pull cable L5 X RAY off of connector J6 DX6 Radiation release is now no longer possible 5 Check the mode using the Info menu Demo mode ON means that The demo mode is switched ON Radiation release is not possible Demo mode OFF means The demo mode is switched OFF Radiography X ray radiation are possible Reattach cover plate B and the tube assembly covers by following the dismantling procedure in reverse order 11 3 2Switching the demo mode OFF gt 2 oN Remove the cover of the tube assembly Loosen screws A and remove cover plate B Set DIP switch S2 DX6 to position 1 Connect cable L5 X RAY to connector J6 DX6 Radiation release is now once again possible 60 51 416 D 3352 D 3352 031 03 17 02 Check the mode using the Info menu Demo mode ON means that The demo mode is switched ON Radiation release is not p
42. arnings ccceeeeeeeeeeeeeeeeeeaeeeeeeeeeeesaaaceeeeeeesaaneeeeeeees 6 ke oale Ccgrazsenacactacdacs cc tsccascacacuauncnssoot A E 7 1 3 System and sensor VESSIONS ccccccsseeeecceceeeeeecaueeeeeesauseeeeeeeseeeeseaeeeeees 8 1 4 Dimensions Space requireMents ccccssescccccseeeeeeeceeeeeecsseeeeeesneseeeess 9 TS lt IWIOUINTING ODUONS terne nesr cp eontesseaboeciccaaienee 10 1 6 iS tallatlomeverslOnsiiviseitaciccancusad eran a a A 11 D liv ry and transpor sssini E ceveotsnenamcenaneeu ace eantehcwgenws 13 2 1 DEINEM treie E a a nes 14 2 2 Transport to the installation Site eecccsseeeeeeeeeeeeeeeeeseeeeeneeeeeseeeens 18 Installation Panoramic X ray unit cccccceeseeeeeeeeeeeeeeeeeeeeeeneeeaseeeneeeaseeeaseeeaeeeeseneneeeanennaes 19 3 1 Installation Maternal assess won cied aitencitieceduatecesteeesmeehevesuaaedhedvantearh evhanats 20 3 2 FRO CUMCC Tool Se arrape a a eatneateactareseantowes 22 3 3 Wall mounting Standard Option 1 ccccccccsseeeeeeseeeeeesseeeeeeeseeeeeeeenes 23 3 4 Installing the floor Stand option 2 cccsscecceseeeeeeeeeeeeeeeeeeeeeessaeeeeeens 30 3 5 Removing the transport Safety device ccccseeeeceeeseeeeeeeeeeeeeneeeeeeeees 40 3 6 Installing the release button Noldel cece ceeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeaes 41 S7 Attaching ING COV CIS teer a e teince cn apes ncn tn reesei caeeaeen ees 42 Electrical COMMECUHON enea a EE
43. ce routines for installation ccccccsseeeeeceeeeeeeesseeeeeeeseseeeeeeaaees 98 11 2 Adjusting the panoramic X ray unit ccceeccceeeeeeeeeeeeeeeeeeeseeeeeeeeeeess 109 Mi Demo ModE essare a a ASAE 136 60 51 416 D 3352 D 3352 031 03 17 02 Kr The Dental Company 1 Before rel peg in ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 17 02 5 1 Before you begin 1 1 Identification of warnings 1 1 Identification of warnings Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 Warning and safety information To prevent personal injury and material damage please observe the warning and safety information provided in the present operating instructions The structure appearance and use of warning and safety information in Sirona documents are based on the ANSI Z535 standard The following warnings may be used in this document A DANGER An imminent danger that could result in serious bodily injury or death 2 WARNING A possibly dangerous situation that can result in seri ous bodily injury or death A CAUTION A possibly dangerous situation that can result in slight bodi ly injury NOTICE A possibly harmful situation which can lead to damage of the product or an object in its environment Instructions for use The following application information may be used in this document IMPORTANT Application instructions and other important information Tip Informat
44. cle Between exposures maintain at least a cool off time auto matic exposure blockage see Operating Instructions man ual Operating instructions During measurements and controls it is necessary to ener gize or de energize the unit For all operating steps please refer to the Operating Instructions manual CAUTION with PC boards All PC boards are fitted with electronic components sensi tive to electrostatic discharge ESD In an environment of moving people electrostatic charges are unavoidable due to friction caused by clothing carpeting etc NOTICE To prevent damage to electronic chips do not touch same Always handle circuit boards by their edges y AteN NOTICE Electrical shock hazard Always turn unit OFF before connecting and disconnecting the test leads to the test points an OFF 60 51 416 D 3352 D 3352 031 03 17 02 Sirona Dental Systems GmbH 8 Startup for USA Canada only Installation Instructions ORTHOPHOS XG 3 8 2 Power supply adequacy 300 VAC e To determine power supply adequacy the line volt Line voltage no load Max permissible age drop during exposure must be measured line voltage drop 1 Be sure power is disconnected at the central distribu 180 208 V 9 V tion panel 208
45. e SELECT X RAY DEVICE and SELECT X RAY COMPO NENT dialogs appear only if several x ray devices or x ray components are available for selection 60 51 416 D 3352 D 3352 031 03 17 02 IMPORTANT When you open the DIAPHRAGM SYSTEM ADJUSTMENT menu the unit switches from the user mode to the PC ser vice mode logged by the PC The service mode is indicated by the SERVICE display on the Multipad 2 In the PC service mode the control options that are avail able on the Multipad are determined by SIDEXIS and the currently selected service routine General control of the unit by means of the Multipad as in the user mode is not possible in this mode The DIAPHRAGM SYSTEM ADJUSTMENT menu has 5 submenus e PAN Sensor adjustment e PAN Diaphragm e PAN Symmetry e PAN Reset adjustment e Sirona Service Sheet for internal use only You can change between the individual submenus by click ing the menu tabs with the mouse To quit the DIA PHRAGM SYSTEM ADJUSTMENT menu click CANCEL 109 11 Appendix 11 2 Adjusting the panoramic X ray unit Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 Direction of displacement of the exposed image area Information on the pictographs in the system adjustment menu The Pan Sensor adjustment Pan Filter and Pan Sym metry submenus each contain a pictographic representa tion of the expected adjustment image to help you perform the adjustment The shifting
46. e are fewer aux iliary lines in the coarse adjustment mode IMPORTANT With SIDEXIS SW V01 45 and higher a message window indicates whether a coarse adjustment is required on com pletion of the precision exposure 60 51 416 D 3352 116 D 3352 031 03 17 02 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 9 Diaphragm system adjustment kammer3 51 4 43 mm Lx 0 so 118mm m s3 104mm IV Precision adjustment Image acquisition Save values Cancel t l 4 Pi StH i 2 je gt S3 je gt System version 1 Fixed sensor holder IMPORTANT The default values for S1 S2 and S3 were automatically de termined by SIDEXIS based on the exposure and entered in the text boxes of the menu For manual adjustment the values displayed at this position in the text boxes of the menu can be overwritten First continue with the automatic adjustment Manual deter mination of the adjustment values is required only if you fail to reach your goal via automatic adjustment see page 120 5 Make SIDEXIS XG ready for exposure Click IMAGE ACQUISITION The PAN SENSOR ADJUSTMENT dialog box appears in Sidexis The dialog box suggests a value Lx for the mechanical adjustment of the sensor If the suggested value is greater than 0 5 mm then perform a mechanical adjustment of the sensor 60 51 416 D 3352 D 3352 031 03 17 02 11 Appendix 11 2 Adjusting the panoramic X ray unit
47. e customer in documentos en correspondencia con las ins in ordnungsgem en und funktionstiichti poper and functioning condition _ trucciones de uso ha sido entregada en el debi gen Zustand bergeben ISA only The customer received his copy of do estado y en condiciones de funcionamiento Der Unterzeichnende best tigt die f r die In the Federal Report of Assembly FD 2579 arriba indicados L appareil y compris les accessoires et tous les documents conform ment a la notice d utilisation ont t remis dans l tat correct et op rationnel mentionn ci dessus alah z A Le sous sign confirme disposer des con A PER Eii betriebnahme und Einweisung erforderliche The person signing confirms having the re naissances techniques requises pour la mi E firmante confirma los conocimientos t cnicos Sachkenntnis und die Erf llung und Einhal quired technical knowledge for performing com se en service et la prise en main et remplir et necesarios para la puesta en funcionamiento y tung der zuvor aufgef hrten Angaben misioning and training and fulfilling and comply respecter les indications nonc es ci des la instrucci n inicial y el cumplimiento y obser ing with the requirements outline above sus vaci n de las indicaciones arriba mencionadas Name Name Nom Apellido Kunde Customer Unterwiesene Person Trained person Verk ufer Sales rep Techniker Technician Client C
48. ecision adjustment Image acquisition l i i l mH py SS ie Save values 4 2 00 0000 o o JO Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 in SIDEXIS Preview j for exposure E sirona 11 2 5Adjusting the PAN sensor e Dismantle the user fixed diaphragm see page 113 e Plug the sensor into the sensor slot on the panoramic X ray unit e Insert the needle phantom in the bite block holder of the panoramic X ray unit see page 112 1 Open the PAN SENSOR ADJUSTMENT submenu see section 11 2 1 IMPORTANT The menu provides a precision adjustment and a coarse ad justment Always try to use precision adjustment first when adjusting the unit In most cases it is not necessary to per form a coarse adjustment prior to precision adjustment 2 To make SIDEXIS XG ready for exposure Click IMAGE ACQUISITION 114 The exposure dialog box showing the exposure status appears in Sidexis The initialization status is visualized by a progress indi cator on the Multipad The initialization procedure is completed when the ex posure parameters of service routine 010 1 60 kV 3 mA 0 6 s are displayed and the progress in dicator disappears Take an exposure 60 kV 3 mA Press the R key on the Multipad to move the unit back to the starting position Press the release button Hold down the release b
49. ediate piece 1 pc Release button holder 1 pc Handle 1 pc Screw for handle M6x25 1 pc 60 51 416 D 3352 D 3352 031 03 17 02 Sirona Dental Systems GmbH 3 Installation Panoramic X ray unit Installation Instructions ORTHOPHOS XG 3 3 1 Installation material 3 Option 2 CAL ANN TTT MN TTT Mounting hardware Tin A Floor stand Covers Cover Floor stand Slide 0 Z Option 2 Floor stand installation Nut M8 2 pc see Section 3 4 Screw M10x50 1 pc Profile clamp 1 pc Screw M5x12 1 pc Washer a 10 5 13 pc Nut M10 4 pc Spring steel clamp 8 pc Screw M10x25 4 pc Wood screw M10x80 3 8x3 5 pc Mounting kit dua 10 SXR 2 pc 3 Floor stand installation Floor stand Floor stand covers Wood screws 10x160 3 8x6 5 pc Plastic wall plug S12 5 pc Screw EM8x60 2 pc Screw M8x80 2 pc Washer a 8 4 2 pc 60 51 416 D 3352 D 3352 031 03 17 02 21 3 Installation Panoramic X ray unit Sirona Dental Systems GmbH 3 2 Required tools Installation Instruct
50. eight currently set current height at center of bite block is displayed as a numerical value e Select service routine 018 2 IMPORTANT The current travel height minus 10 mm is displayed and saved 60 51 416 D 3352 D 3352 031 03 17 02 2 11 Appendix 11 1 Service routines for installation Confirm the travel height by pressing the save values key and the Service key Setting the mechanical limit stop on the unit 3 Loosen the nut A and slide the mechanical limit stop B for the limit switch to the new maximum position until it switches Retighten the nut firmly Quit service routine S018 see page 101 Switch the unit OFF and restart it After the restart the new maximum travel height is ac tive 105 11 Appendix 11 1 Service routines for installation Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 11 1 8Canceling the maximum travel height setting e Select service routine 018 3 IMPORTANT The current travel height minus 10 mm is displayed and saved 1 To cancel the maximum travel height setting first press the Memory key and then the R key 106 Quit service routine S018 see page 101 Switch the unit OFF and restart it After the restart the previously set maximum travel height is undone 60 51 416 D 3352 D 3352 031 03 17 02 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 11 Appendix 11 1 Service routines for i
51. el lights up This LED 7 Initiate the radiation Hold the release button pressed remains lit as long as the unit is on until the set radiation time has expired CAUTION Activating the release button triggers The start screen appears on the touchscreen of X rays g gg the Easypad and the system s self adjustment routine starts running for approx 1 minute The rotating element rotates briefly clockwise and counterclockwise The diaphragm moves into position The forehead and temple supports on the panoramic unit open and close and then stop moving in fully opened position On completion of the self adjustment routine the main menu appears on the touchscreen Help message H301 prompts you to move the unit into the starting position 3 Press the R key The unit moves to its starting position 60 51 416 D 3352 D 3352 031 03 17 02 79 8 Startup for USA Canada only Sirona Dental Systems GmbH 8 4 kV verification Exposure Time Verification Installation Instructions ORTHOPHOS XG 3 kV verification Analyzing measurements gt Read the voltage values on the Mult O Meter The value for the voltage displayed on the Mult O Meter must correspond to the 62 kV selected in the service routine The permissible tolerance is 10 Ifthe measured values do not fall within the permissible tolerance range replace the tube assembly see Service Manual Ifthe measured values fall within the permissible tolerance range
52. emove the carrying handles E on the lower side of the pallet from the holders F 4 Loosen screws G and remove the holders F 60 51 416 D 3352 D 3352 031 03 17 02 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 3 Installation Panoramic X ray unit 3 4 Installing the floor stand option 2 ies 5 Carefully push the unit toward the base just far enough so that the center styrofoam part nudges lower supporting block H NOTICE Make absolutely sure that the interfaces do not have firm contact with supporting block H and are not damaged when you push the unit 6 Remove the two screws J from the bottom of the stand 60 51 416 D 3352 D 3352 031 03 17 02 31 3 Installation Panoramic X ray unit Sirona Dental Systems GmbH 3 4 Installing the floor stand option 2 Installation Instructions ORTHOPHOS XG 3 a E ee O O 4x M 10x25 4x Sta 10 5 2x M 8x80 7 Screw adjustment plate K onto the stand firmly 8 Screw the support N firmly onto the base plate P using _ the 4 screws O and washers IMPORTANT To do this use screws J for the two front holes and secure them with the corresponding nuts L and washers from the installation material Use two new screws M for the rear holes The
53. ered in the image field as well as b en gt a a tear inside the superimposed auxiliary lines A loc laa A white border surrounding the image on all sides must be visible The maximum density NOTICE must lie in the center of the diaphragm area A The adjustment of the pan diaphragm must be performed mechanically i e this cannot be done via SIDEXIS 60 51 416 D 3352 122 D 3352 031 03 17 02 Sirona Dental Systems GmbH 11 Appendix Installation Instructions ORTHOPHOS XG 3 11 2 Adjusting the panoramic X ray unit v i o not remo e F D Z iT il cH 6 ALD KOJ f j il o S A Ey Y7 J mU af i 5 hj h ifl g AO S EA dal E ldda Hi j Sa im s i T oe ia a a A E S Fe oS Se a A E a ee OS E a Manual adjustment of the pan diaphragm e Remove the covers see also page 113 Tube assembly front Tube assembly rear e Loosen both screws A slightly approx 1 turn e Adjust the diaphragm position via adjusting screws B horizontal adjustment and C vertical adjustment B movement of the diaphragm to the right or left C movement of the diaphragm upward or downward e Tighten the screws A again 60 51 416 D 3352 D 3352 031 03 17 02 123 11 Appendix Sirona Dental Systems GmbH 11 2 Adju
54. erer 07 2073 m The Dental Company ORTHOPHOS XG 3 Installation Instructions isirona aa OO palit ADY Prog s W mA YV VY 0000 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 General information About this document This document describes the installation of the ORTHO PHOS XG XG3 panoramic X ray unit 07 2013 New as of Changes since the last version 04 2013 Chapter or section page 2 A DEINEN ecscaetatcestieh cutee had eaten erent cnateetatones 14 2 2 Transport to the installation site ceeeeeeee 18 3 3 Wall mounting Standard option 1 ccseeeee 23 For installation please refer also to the following docu ments e Installation drawings e Installation Requirements Separate document e Operating Instructions e Service Manual e Installation Report and Warranty Passport e SIDEXIS XG Digital Radiography Installation instructions Our Customer Service Center can provide the technical documentation in paper form free of charge on request pro vided that the respective order numbers are specified cor rectly The latest documentation can always be downloaded from the Sirona homepage www sirona com HOME Service Technical Documentation 60 51 416 D 3352 D 3352 031 03 17 02 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 B fore VOU CCN icicles luca a e at ac e a ciate ee eee 5 1 1 Identification Of w
55. eshooting according to the service manual Now reconnect the unit to the building installation 63 6 Safety checks Sirona Dental Systems GmbH 6 2 Checking the device leakage current Installation Instructions ORTHOPHOS XG 3 60 51 416 D 3352 64 D 3352 031 03 17 02 Kr The Dental Company F Initial aats ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 17 02 65 7 Initial startup Sirona Dental Systems GmbH 7 1 Inserting the forehead and temple supports Installation Instructions ORTHOPHOS XG 3 7 1 Inserting the forehead and temple supports 1 Insert the forehead and temple supports until they lock in place 60 51 416 D 3352 66 D 3352 031 03 17 02 Sirona Dental Systems GmbH 7 Initial startup Installation Instructions ORTHOPHOS XG 3 7 2 Inserting the sensor 7 2 Inserting the sensor NOTICE The sensor is a sensitive component and therefore must be handled with special care 1 Carefully plug sensor A upward into the holder until it audibly locks in place 60 51 416 D 3352 D 3352 031 03 17 02 67 7 Initial startup 7 3 Switching the units ON 7 3 Switching the units ON WARNING When performing the following test be sure to observe the radiation protection regulations applicable in your country see operating instructions It is prohibited for any person to be positioned in the unit when it is switched on NOTICE If the room height is less than 2 27 m 89 3 8 2 30 m 90 1 2 w
56. esponds to a tube current of 1MA per missible tolerance 20 IMPORTANT If specified values cannot be obtained see Service Man ual chapter Tube Current Verification 77 8 Startup for USA Canada only Sirona Dental Systems GmbH 8 3 Tube Current Verification Installation Instructions ORTHOPHOS XG 3 ZRH A TET mm TIL th e f specified value is obtained switch unit OFF 8 Remove upper cover and meter leads e Replace jumper 9 Screw the cover plate back on to the electronics box 10 Reattach the housing covers 60 51 416 D 3352 78 D 3352 031 03 17 02 Sirona Dental Systems GmbH 8 Startup for USA Canada only Installation Instructions ORTHOPHOS XG 3 8 4 kV verification Exposure Time Verification 8 4 kV verification Exposure Time Verification A X RAY 60kV 8mA gt oD Preparing the measurement 4 Call the Service menu and the Service routine S002 3 see Service Manual 1 Attach the Mult O Meter sensor in the middle of the du al sensor 5 Use the arrow keys A in selection field 1 to select the kV mA level 62kV 8mA 2 Switch the unit on via the switch A see also Operating Instructions The X Ray radiation indicator B lights up briefly 6 Use the arrow keys A in selection field 2 to select the radiation time 0 5 s After approx 2 seconds the green LED C in the Performing measurements upper part of the control pan
57. finalize the measurement Exposure Time Verification Analyzing measurements gt Read the radiation time on the Mult O Meter The value for the radiation time displayed on the Mult O Meter must correspond to the radiation time of 0 5s selected in the service routine The permissible tolerance is 10 Ifthe measured radiation time does not fall within the permissible tolerance replace the tube assembly see Service Manual Ifthe measured radiation time falls within the permissible tolerance finalize the measurement Finalizing the measurement gt Exit the service routine gt Switch off the unit see Operating Instructions 60 51 416 D 3352 80 D 3352 031 03 17 02 Sirona Dental Systems GmbH 8 Startup for USA Canada only Installation Instructions ORTHOPHOS XG 3 8 5 Checking the laser for USA Canada only 8 5 Checking the laser for USA Canada only of beens lf 1 Insert the forehead and temple supports Label 2 Check the laser e Fasten a piece of white cardboard between the temple LASER KLASSE 1 supports CLASS 1 LASER PRODUCT an mirror by pressing the left recess A on the TB APPAREIL A LASER DE CLASSE 1 ouch bar e Switch on the light localizer with key B on the Easy i pad The Lasers generate a red line Date of manufacture August 2007 A WARNING Class 1 radiation is emitted during installation Complies with 21 CFR 1040 10 and 1040 11 Always keep eyes a mi
58. he SIDEXIS measuring ruler IMPORTANT Measure in the lower area of the needles if possible since they may have been bent slightly after repeated use Always measure exactly from pin center to pin center IMPORTANT Tip To facilitate the measuring procedure you can color the image in SIDEXIS see also SIDEXIS Operator s Manu al 132 60 51 416 D 3352 D 3352 031 03 17 02 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 11 Appendix 11 2 Adjusting the panoramic X ray unit 2 Diaphragm system adjustment kammer3 xj Pan Sensor adiustment Pan Fixed diaphragm Pan Symmetry Pan Reset adiustment 4 gt 4 2 10 mm 2 00 mm 1 50 mm Image acquisition gt et Kar R 1 52 Save values se a Cancel 2 Overwrite the default values for S1 S2 and S3 with the measured values in the text boxes of the PAN SYMME TRY submenu IMPORTANT For information on the direction of displacement input of sign in the menu see page 110 Use points as decimal separators e Proceed with step 5 of the adjustment procedure on page 129 60 51 416 D 3352 D 3352 031 03 17 02 133 11 Appendix Sirona Dental Systems GmbH 11 2 Adjusting the panoramic X ray unit Installation Instructions ORTHOPHOS XG 3 Blenden Geradtejustage kammer3 x Pan Sensorjustage Pan Festblende Pan Symmetrie Pan Justage zuruecksetzen f gt Blendenwerte R
59. in srlecet starting settings Setting of default patientaymbe CEPA Leoni gureed in select starting settings l IMPORTANT You can scroll down further in the file using the scroll bar The Changed system parameters i e the system param eters that were modified in relation to the factory setting are displayed there This is especially interesting after a module change The parameter settings can thus be easily traced 93 10 Final work 10 3 Installation Report Warranty Passport Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 10 3Installation Report Warranty Passport XGPlus DS Ceph re ORTHOPHOS xG5 Ceph a va EI Installationsprotokoll Installation Report Protocole d installation Protocolo de montaje Garantiepass Warranty Passport fiche de garantie Pasaporte de garantia Kunde Customer Client Cliente H ndler Dealer Depositaire Depositante Adresse Tel Adresse Tel Address tel Address tel Adresse tel Adresse tel Direcci n tel Direcci n tel Kundennummer Customer No Auftragsnummer Order No No du client No de cliente LILILILILILI No de commande No de pedido 11lIl _JLJtJl Seriennummern Serial numbers Num ros de s rie Num ros de serie Pos Serial No XG Ceph Schockindikator ok JA Shock watch ok YES Indicateurs de chocs intact oul Indicador de golpes correcto sl Kippindikator ok JA Tilt watch ok YES
60. iness dialog box appears on the screen and shows the status of the exposure 70 3 E Select Service exposure x Factory Service Diaphragm test exposure Diaphragmsystem adjustment Quality test exposure Cancel v Cag SEIE IMPORTANT During operation in the service mode the unit switches from the user mode to the PC service mode logged by the PC The service mode is indicated by the SERVICE display on the Multipad see page 84 The unit switches back to the user mode as soon as the ex posure is completed 60 51 416 D 3352 D 3352 031 03 17 02 Sirona Dental Systems GmbH 7 Initial startup Installation Instructions ORTHOPHOS XG 3 7 4 Checking the data paths de a e EH J 6 Take an exposure 7 Test pattern for panoramic unit Press the R key on the Multipad to move the unit back to the starting position Check the images according to the following criteria Press the release button Hold down the button Tile structure of the existing segments until the exposure has been completed Tip To facilitate checking you can adjust the image con trast and brightness in SIDEXIS IMPORTANT Linear grayscale gradient A service message box indicates whether the generated test image is correct Acknowledge this message with OK The test image is displayed on the screen Intensity steps clearly recognizable No image artifacts detectable 60 51 416 D 3352
61. ing packaging at the tabs pro vided for that purpose A pull out the carrying handles B and transport the unit to the installa tion site A WARNING When in transport position the unit has a very high center of gravity Take care that the unit does not tip over during transport 2 Transport without pallet exception IMPORTANT If the pallet is too wide for transport to the installation site you may unscrew the pallet from wooden support C and transport the unit by means of the wooden supports without the pallet To do this proceed as follows Remove the surrounding packaging the two card board boxes as well as the two lateral styrofoam parts Loosen the four screws D 18 NOTICE The center styrofoam part should remain attached to the unit for protection If this is not possible SIRONA recom mends securing the tube assembly in its position with the Supplied strap prior to any further transport see the label on the styrofoam packaging Tighten the strap only loosely Do not stretch Pull the carrying handles B out of their holders and insert them through the drillings of the wooden sup port C from the back Insert screws D through the drillings E into the drill ings of the carrying handles to attach them firmly Long or short IMPORTANT The carrying handles have rims which prevent them from slipping out of the holes 60 51 416 D 3352 D 3352 031 03 17 02 The
62. ing with the spirit levelin both directions by placing the spirit level on the ring 60 51 416 D 3352 36 D 3352 031 03 17 02 Sirona Dental Systems GmbH 3 Installation Panoramic X ray unit Installation Instructions ORTHOPHOS XG 3 3 4 Installing the floor stand option 2 15 Mount the upper wall holder CAUTION In case of mounting on weight bearing wood struc tures Use the enclosed wood screws and washers from the mounting kit for mounting the unit on weight bearing wood structures 2x 8x80 2X dia 8 4 IMPORTANT 1980 mm The floor stand version is 78 Z o 30 mm 1 3 16 higher than IMPORTANT the standard version Even when the unit is installed using the floor stand it must CO be secured with the upper wall holder _ 60 51 416 D 3352 D 3352 031 03 17 02 37 3 Installation Panoramic X ray unit Sirona Dental Systems GmbH 3 4 Installing the floor stand option 2 Installation Instructions ORTHOPHOS XG 3 2x M8 2X a 8 4 2x M8x30 e Slide the unit with the assembled floor stand up to the wall NOTICE Please observe the required movement range of the unit during positioning see section 1 2 16 Mount the unit loosely on the upper wall holder Insert the screws V into the groove Screw the unit onto the wall holder loosely using nuts W and washers Do not tighten the screws 60 51 416 D 3352 38 D 3352 031 03 17 02
63. ion on making work easier 60 51 416 D 3352 D 3352 031 03 17 02 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 1 Before you begin 1 2 Safety A DANGER Fixed connection The installation of a power plug instead of the pre scribed fixed hard wired connection violates interna tional medical regulations and is prohibited In case of a fault you would thus endanger the life and limb of the patient the operator or other persons WARNING Installation and startup must be carried out in accor dance with the requirements stated in our Installation Instructions 2 WARNING Installation and startup may be carried out only by per sonnel specifically authorized by SIRONA WARNING Any person who assembles or modifies a medical elec trical system complying with the standard IEC 60 601 1 1 safety requirements for medical electrical equipment by combining it with other equipment is responsible for ensuring that the requirements of this regulation are met to their full extent for the safety of the patients the operators and the environment If any equipment not approved by SIRONA is connect ed it must comply with the applicable standards IEC 60950 1 for information technology equipment and IEC 60 601 1 for medical electrical equipment See also On site installation dimensions technical data and the Compatibility list Declaration of conformity issued by the system integra
64. ions ORTHOPHOS XG 3 4 7 _ 8 9 Se 6 9 l 1 Masonry drill Ava 10mm 3 8 3 2 Required tools 1 Required measuring instruments 2 Impact drill or percussion drill e Multimeter or ammeter battery operated e Test unit for device leakage current measurement 3 Torx offset screwdriver e g Bender tester or line frequency high resistance TX10 measurement voltage source isolation transformer TX20 and measuring circuit MD that meets the require ments of IEC 60 601 1 TX25 l e Power source for protective ground wire test 4 Open end wrench Technical data 13 mm A F No load voltage of at least 4V maximally 24V 5 Socket wrench Short circuit current at least 0 2A Wrench insert 13 Hammer Center punch Spirit level O S N p Additional requirements for installation with floor stand Masonry drill 6ta 12mm 1 2 Allen key 6 mm Socket wrench and extension Wrench insert 17 included in the scope of supply 60 51 416 D 3352 22 D 3352 031 03 17 02 Sirona Dental Systems GmbH 3 Installation Panoramic X ray unit Installation Instructions ORTHOPHOS XG 3 3 3 Wall mounting standard option 1 60 51 416 D 3352 D 3352 031 03 17 02 23 3 Installation Panoramic X ray unit Sirona Dental Systems GmbH 3 3 Wall mounting standard option 1 Installation Instructions ORTHOPHOS XG 3 e Remove profile cover A 1 Position the two i
65. irst see sec tion 60 51 416 D 3352 62 D 3352 031 03 17 02 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 6 Safety checks 6 2 Checking the device leakage current 6 2 Checking the device leakage current Power cable to the ORTHOPHOS XG A Measuring setup for testing the device leakage current A DANGER Shock hazard It is essential that you switch the X ray unit OFF before replacing any components IMPORTANT A high resistance measuring voltage source at line frequen cy and a measuring circuit compliant with the requirements of IEC 60 601 are required Complete test units e g the Bender tester fulfill these re quirements If you have not done it already Switch OFF the line voltage at the main switch of the building installation Disconnect the power cable and the second protective ground wire from the building installation Remove the following cover parts Profile cover 60 51 416 D 3352 D 3352 031 03 17 02 Voltage source 208 V 230 V 10 Voltmeter 1 Check whether the power switch of the unit is turned ON 2 Connect a high resistance measuring voltage source between the short circuited power cable B and ground wire A 3 Measure the voltage drop across MD IMPORTANT The measured value must not exceed 5 mA If the leakage current is not OK perform troubl
66. ith floor stand you must limit the maximum travel height of the unit see section 11 1 7 68 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 Switch the unit ON All LEDs and the LED display on the Multipad light up briefly The green LED at the top of the Multipad is lit permanently as long as the unit is ON The initialization status is visualized by a progress indicator while the unit executes a self adjustment routine approx 2 min At the same time the rotating element rotates briefly clockwise and coun terclockwise and the diaphragm is positioned The forehead and temple supports on the panoramic unit open and close and then stop moving in fully opened position Once the self adjustment routine is completed help message H301 prompts you to move the unit to the starting position Press the R key to move the unit back to the starting position Switch the PC ON and start SIDEXIS XG 60 51 416 D 3352 D 3352 031 03 17 02 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 7 Initial startup 7 4 Checking the data paths Relocate Plug ins Macros Configure system 7 4 1 Generating pan ceph test images Checking the pan ceph PC data path IMPORTANT SIDEXIS XG must be installed and configured before you begin checking the data paths see SIDEXIS Operators Manual 60 51 416 D 3352 D 3352 031 03 17 02 1 In SIDEXIS XG select the con
67. l Systems GmbH 5 Installation Remote control Installation Instructions ORTHOPHOS XG 3 5 3 Connecting the control cables REMOTE 5 3 Connecting the control cables 590 Shorten if necessary Whitegreen Green Brown Whitebrown BU WHOG Blue Whiteorange WHBU OG Whiteblue Orange 5 3 1 Installation version 1 without release button and coiled cable 1 Attach cable L117 at shield clamp A Unscrew clamp A from the board MUP ORT AINE p l For information on the connection of a door contact switch Place the cable in the clamp so that the turned up see section 5 4 cable shield is completely enclosed Re attach the clamp to the board 2 Connect control cable L117 to terminal X108 board DX42 as shown in the connection diagram above 3 Roll cable L117 into a loop and stow it away at the bot tom edge of the chassis before closing the housing 60 51 416 D 3352 D 3352 031 03 17 02 53 5 Installation Remote control 5 3 Connecting the control cables REMOTE Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 5 3 2 Installation version 2 with release button and coiled cable NOTICE Operation of the remote control via the membrane keyboard is prohibited when installing the remote control with release button Installing the release button holder 1 Use an awl to puncture the membrane keyboard at the prepared points C from the rear 54 Using the two screws D
68. l and floor installation are possible on site Option 1 with second wall holder see Section 3 3 2 Wall mounted installation with 2 wall holders and no floor fastening if only wall installation is possible on site 10 Option 1 Option 2 CONN Option 2 with floor stand see section 3 4 3 Installation with floor stand for free installation any where in the room or if wall mounted installation is not possible on site e g with lightweight walls IMPORTANT If the unit is installed freely with a floor stand the quality of the resulting X ray exposures may be impaired depending on the floor or surface conditions Sirona there fore recommends additional fastening of the unit with an up per wall holder also when installing it with a floor stand 60 51 416 D 3352 D 3352 031 03 17 02 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 1 6 Installation versions 1 Standard installation ORTHOPHOS XG 3 without remote control with re lease button on the coiled cable in the treatment room 60 51 416 D 3352 D 3352 031 03 17 02 1 Before you begin 1 6 Installation versions Installation version 1 see section 5 3 1 ORTHOPHOS XG 3 with remote control outside the X ray room without release button on the coiled ca ble Installation version 2 see section 5 3 2 ORTHOPHOS XG 3 with remote control outside the X ray room with release button on the coiled cable 11 1
69. lection and confirmation the service routine is activated 60 51 416 D 3352 102 D 3352 031 03 17 02 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 11 Appendix 11 1 Service routines for installation 0081 ORTHOPHOS XG 3 Sleeve X306 Sleeve X309 Sleeve X503 x318 DX61 x304x320 DX9 1x308 Ox Cr Cx Cr Jumper inside LJ lJ CAN LJ J I I CAN CEPH Sensor Jumper inside 11 1 5Configuring the system version S017 2 Select service routine 017 2 with security access see section 11 1 3 and 11 1 4 Following selection the ID of the currently selected sys tem version appears in selection field 1 Select the desired system version 0081 using thearrow keys of selection field 1 and confirm your selection by pressing the Service key 0081 ORTHOPHOS XG 3 fixed diapraghm After the selection of the system version the LED above the Memory key lights up 60 51 416 D 3352 D 3352 031 03 17 02 To save the selected system version first press the Me mory key the LED above the R key lights up and then the R key IMPORTANT The set value is permanently saved as the relevant system version 2 Check the jumper settings of sockets 306 309 and 503 on board DX1 0081 Jumpers in i
70. liente Personne initi e Persona instruida Vendeur Vendedor Technicien Tecnico Unterschrift Signature Signature Firma Datum Date Datum Date Datum Date Datum Date Date Fecha Date Fecha Date Fecha Date Fecha Wartungsiberblick Maintenance logbook Journal de bord des interventions de maintenance Libro de registro de mantenimiento Datum Date Handler Dealer Techniker Technician Unterschrift Signature Date Fecha Depositaire Depositante Technicien Tecnico Signature Firma Fortsetzung siehe R ckseite Continued see rear Suite voir verso Continuaci n ver dorso FUR DEN KUNDEN FOR THE CUSTOMER POUR LE CLIENT PARA EL CLIENTE Sirona Dental Systems GmbH FabrikstraBe 31 64625 Bensheim Germany A Nr 113 429 59 38 357 D 3352 077 01 05 09 12 2010 1 Enter the serial numbers and software version on the attached document Installation Report and Warranty Passport Complete the remaining points together with the customer Hand over the copy marked FOR THE CUSTOMER to the customer The copy marked FOR THE DEALER is for the customer records kept by the dealer Please fax the FOR THE DEALER copy back to SIRONA immediately after the installation NOTICE Please note that any warranty claims can be fulfilled only if the completely and legibly completed Installation Report Warra
71. mbH 10 1 Attaching the profile covers Installation Instructions ORTHOPHOS XG 3 10 1Attaching the profile covers 1 Attach the profile cover Screw holder A onto the column with screw B Clip on profile C Insert intermediate piece D in the gap IMPORTANT Make sure that intermediate piece D lies flush on the stand surface e The unit is now ready for operation 60 51 416 D 3352 92 D 3352 031 03 17 02 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 10 Final work 10 2 Selecting More details 10 2Selecting More details 1 Opening SIXABCON exe Serialnumber H N System Software version 04 n Permanent 01110111 11 36 53 2006 Mar 23 129804024 o s 2005 Sep 22 1 MPCESSTZPS0s 38 2006 Mar 17 130 39 Oxff Oxf Oxff Oxff Oxff Oxff Oxff Oxff 1 Open the SIXABCON utility from the SIDEXIS XG pro gram folder Click the SIXABCON exe file See screenshot or use the pull down menus PROGRAM gt SIDEXIS gt CONFIG URATION OF THE X RAY COMPONENTS 2 Open the MORE DETAILS menu Click the PROPERTIES tab then click MORE DETAILS The current parameters are retrieved from the unit and stored as an XML file in the PDATA P2K_Config folder using the unit s network name This step may take up to 30 seconds e After the parameters are read an editor displaying the XML file is opened automatically 60 51 416 D 3352 D 3352 031 03 17 02 2 Opening the EX
72. mitation of the mini 108 travel height setting mum travel height setting The service routines listed above may be required for instal lation of the ORTHOPHOS XG X ray unit and are described in this Appendix A complete list and comprehensive description of all ser vice routines can be found in the Service Manual 98 60 51 416 D 3352 D 3352 031 03 17 02 Sirona Dental Systems GmbH 11 Appendix Installation Instructions ORTHOPHOS XG 3 11 1 Service routines for installation 11 1 1Selecting the Service menu 1 Press the Service key for approx 2 seconds 2 Then press the patient symbol keys in the following or der within 4s b d a After you have entered the key combination correctly the Service menu appears 3 To quit the Service menu and return to the main menu press the up arrow key AN at the very right 60 51 416 D 3352 D 3352 031 03 17 02 99 11 Appendix 11 1 Service routines for installation O O O Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 11 1 2Displays and symbols on the Multipad From the Multipad you can start all available service routines and perform important system settings and tests as well as calibrations Depending on the procedure step different hints error messages and parameters are displayed in a context sen sitive way on the Multipad Display Error messages Ex yy zz see Service Manual Service routines test ste
73. ndibular joint supports 1 1x and 2 1x 4 Ear holders 4x 5 Buttons TMJ 2x 6 Chin rest accessories 1x Bite block 5x Bite block fixation 1x Bar 1x Chin pad 1x Chin rest 1x 16 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 Bite block Bar Bite block fixation Chin pad Chin rest 7 Contact segment standard yellow 1x 8 Bite block standard yellow 1x 60 51 416 D 3352 D 3352 031 03 17 02 Sirona Dental Systems GmbH 2 Delivery and transport Installation Instructions ORTHOPHOS XG 3 2 1 Delivery Bottom side Top side Hygienic protection Adjustment set Panoramic X ray unit Hygienic protective sleeves for 14 Panoramic needle phantom 9 Forehead and temple supports 500x 15 Set of Torx offset screwdrivers 10 Bite block 500x 11 Chin rest and bar 100x 12 Bite blocks and contact segments 500x 13 XG hygienic handle 100x 60 51 416 D 3352 D 3352 031 03 17 02 17 2 Delivery and transport 2 2 Transport to the installation site Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 2 2 Transport to the installation site NOTICE If possible leave the packaging attached to the unit during transport in order to protect it against damage 1 Transport with packaging attached normal case Open the surround
74. ngth of at least 1000V Be sure to use a battery powered measuring instru ment with shock hazard protected sockets Use only test leads with shock protection NOTICE The X ray tube assembly moves during the measurement Therefore be sure to use test leads of sufficient length and place the measuring instrument in a location where it is firm ly seated so that it doesn t fall down 60 51 416 D 3352 76 D 3352 031 03 17 02 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 3 20mADC p 49E 6 ME set 2 e li PC board DX6 j WARNING Be sure to switch the X ray unit off before removing the jumper for the mA measuring jack 3 Remove jumper A from MA MA test points on PC board DX6 Connect digital ammeter to MA and MA and select range 20 mADC 4 Reattach cover C and tighten screw D securely 5 Switch unit ON Wait 1 min for self adjustment of the unit Press key R to return X ray tube head into the initial position 6 Select 66kV 8mA Establish exposure readiness via SIDEXIS 7 f P1 program and 66kV 8mA are selected The unit must be ready for radiation 60 51 416 D 3352 D 3352 031 03 17 02 8 Startup for USA Canada only 8 3 Tube Current Verification Measurement A DANGER RADIATION Depress the exposure key and hold depressed until meter reading is obtained The ammeter shall indicate 8mA 1 6mA Record reading IMPORTANT Readings 1mA corr
75. nimum distance of 100 mm away _ except for deviations pursant to from the laser Do not stare into the beam Laser Notice No 50 dated June 24 2007 Sirona Dental Systems GmbH Fabrikstr 31 SAG laser vertical line i D 64625 Bensheim Germany The light beam must strike the center of the head sup Manufactured in Bensheim port in a vertical direction FH laser horizontal line The light beam must strike the FH line of the sheet of cardboard attached to the head support It must be pos sible to displace the light beam vertically using slide C Unwanted movement of the FH light localizer is not per mitted IMPORTANT To adjust the lasers see Maintenance Instructions No controls are available to adjust the laser power 60 51 416 D 3352 D 3352 031 03 17 02 81 8 Startup for USA Canada only Sirona Dental Systems GmbH 8 5 Checking the laser for USA Canada only Installation Instructions ORTHOPHOS XG 3 60 51 416 D 3352 82 D 3352 031 03 17 02 The Dental Company 9 sansa the Se SIT to ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 17 02 83 9 Checking the system adjustment 9 1 Test exposures 9 1 Test exposures Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 1 Exposure Preview fthe exposure here 9 1 1 Test exposure menu The panoramic X ray unit is already completely adjusted on delivery Check the adjustment using the TEST EXPOSURE menu to make sure that the unit
76. nner position Quit service routine S017 see page 101 Switch the unit OFF and restart it After the restart the selected system version is active 103 11 Appendix Sirona Dental Systems GmbH 11 1 Service routines for installation Installation Instructions ORTHOPHOS XG 3 11 1 6Configuring the remote control S017 6 e Select service routine 017 6 with security access see sections 11 1 3 and 11 1 4 Following selection the ID of the currently selected sys tem configuration appears in selection field 1 1 Select the desired system configuration using the ar row keys of selection field 1 and confirm your selection by pressing the Service key 00 Remote control disabled 01 Remote control enabled After the selection of the system version the LED above the Memory key lights up To save the selected system version first press the Me mory key the LED above the R key lights up and then the R key IMPORTANT The set value is permanently saved as the relevant system configuration e Quit service routine S017 Switch the unit OFF and re start it e After the restart the new system configuration is active 60 51 416 D 3352 104 D 3352 031 03 17 02 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 11 1 7Setting the maximum travel height S018 2 1 Move the unit to the required maximum travel height by pressing the Up Down keys in the user mode IMPORTANT The travel h
77. not be fitted on the bite block holder on the panoramic X ray unit 3 Take an exposure 60 kV 3 mA Press the R key on the Multipad to move the unit back to the starting position e Open the TEST EXPOSURE menu Press the release button Hold down the release button until image acquisition is completed and the see section 9 1 1 preview image appears in the exposure dialog box 1 Select the PAN DIAPHRAGM submenu 2 To make SIDEXIS XG ready for exposure Click IMAGE ACQUISITION 60 51 416 D 3352 86 D 3352 031 03 17 02 Sirona Dental Systems GmbH 9 Checking the system adjustment Installation Instructions ORTHOPHOS XG 3 9 1 Test exposures Adjustment ok Adjustment not ok 4 Evaluate the X ray image IMPORTANT If these criteria are not fulfilled B the pan diaphragm must be adjusted The procedure for adjusting the unit is de scribed in the Appendix see section 11 2 and in the OR A white border surrounding the image on all THOPHOS Service Manual sides must be visible The maximum density must lie in the center of the diaphragm area A The exposed diaphragm area must lie horizon tally centered in the image field as well as inside the superimposed auxiliary lines A 60 51 416 D 3352 D 3352 031 03 17 02 87 9 Checking the system adjustment 9 1 Test exposures 1 Diaphragm system adjustment Kammer4 Xx Pan Sensor adjustment Pan Diaphragm Pan Filter Pan Symmetry Ceph
78. nstallation 11 1 9Setting the minimum travel height S018 5 1 Move the unit to the required minimum travel height by pressing the Up Down keys in the user mode IMPORTANT Programming the minimum travel height is possible only for a system height that is below the lower correction switch lev el lt position value of 1500 2 Select test step 2 in selection field 2 with the arrow keys and confirm your selection by pressing the Ser vice key The current height position is displayed in selection field 1 The LED above the Memory key is lit 3 To save the minimum travel height first press the Mem ory key the LED above the R key lights up and then the R key 60 51 416 D 3352 D 3352 031 03 17 02 IMPORTANT The limitation of the minimum travel height is purely soft ware based The lower limit switch is not adapted to the new minimum travel height e Quit service routine S018 see page 101 e Switch the unit OFF and restart it After the restart the new minimum travel height is ac tive 107 11 Appendix Sirona Dental Systems GmbH 11 1 Service routines for installation Installation Instructions ORTHOPHOS XG 3 11 1 10Canceling the minimum travel height setting S018 6 1 Select test step 6 in selection field 2 with the arrow e Quit service routine S018 see page 101 keys and confirm your selection by pressing the Ser e Switch the unit OFF and restart it vice key o After
79. nstallation aids B at the foot C of the device and secure their position with adhesive tape CAUTION The installation aids must be placed on top of each other in such a way that their openings lie on top of each other 2 Setup the unit To do this tilt the transport pallet up right IMPORTANT If you have transported the unit on the wooden support with out a pallet set the unit upright with the wooden support You can also use the lateral styrofoam packaging as a sup port with this variation 3 Loosen the nuts B with washers on both sides of the pallet or the wooden support Take off the pallet or the wooden support Remove the threaded bolts C A CAUTION Remove the lower bolts first followed by the upper bolts IMPORTANT The nuts D on the unit may remain inside the unit when the threaded rods are removed Remove the upper nuts The lower nuts may remain in the unit 60 51 416 D 3352 24 D 3352 031 03 17 02 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 CAUTION If the setup site for the unit is carpeted the carpeting must be removed Wall plugs Each wall plug must withstand an extraction force of 700 N The wall construction must be suitable for installation of the unit see On site installation dimensions technical data In case of mounting on weight bearing wooden con structions Use the enclosed wood screws and washers from the mounting kit for mounting the unit
80. nty Passport document is available at SIRONA im mediately after the installation 94 ORTHOPHOS XG 5 Ceph ene ORTHOPHOS XG 3 ie beapey ESETA Compatibility list Declaration of conformity by system integrator for digital radiography with SIDEXIS System integrator system installer Company address Name address IEC 60 950 1 and CE marking IBM compatible PC with For the Federal Republic of according to Council Directive DualCore 1 GHz Germany Manuf 89 336 EEC Q RaM 1GBwithWindows2000 On site installation accord Q XP 7 Professional 7 Ultimate ing to DIN VDE 0100 710 gt 5 GB free hard disk space Graphics card z 21024 x 768 gt 16 7 Mio colors Type True Color gt 128 MB Q Oo Refresh rate gt 70Hz Operating system according to the op systems No approved in the installation Q Q instructions Removable disk drive CD DVD ROM Q Q Dual Layer Manuf IEC 60 950 1 and CE marking Min screen diagonal 15 with Display quality tested with z aut according to Council Directive flatscreen Q SMPTE test image accord Q O 89 336 EEC 17 with CRT Q ing to the enclosed instruc EYRE Min resolution 1280 x 1024 tions a Refresh rate gt 70Hz CRT Dot mask min 0 28mm No flat screen monitor Pixel pitch 0 30 mm x 0 30 mm Manuf Sirona Dental Systems CE marking in compliance with see Installation instructions For the FRG Q Council Directive 93 42 EEC
81. on Panoramic X ray unit 3 3 Wall mounting standard option 1 H 1x a 8 4 1x M 8x50 Insert screw G through washer H and then through the wall holder and into the stand from the rear Fit profile clamp J onto screw G from the other front side and screw nut K onto the screw Tighten nut K firmly 27 3 Installation Panoramic X ray unit Sirona Dental Systems GmbH 3 3 Wall mounting standard option 1 Installation Instructions ORTHOPHOS XG 3 9 Remove the transport safety device A prior to unit 12 Level the unit by moving the unit base in both directions startup while measuring with the spirit level Align the stand first by placing the spirit level 10 Pull the wooden board L out of the styrofoam packag l l against the side and rear of the stand ing and remove all styrofoam packaging Then align the ring with the spirit level in both 11 Rotate the X ray tube assembly counterclockwise to directions by placing the spirit level on the ring the front side of the unit 60 51 416 D 3352 28 D 3352 031 03 17 02 Sirona Dental Systems GmbH 3 Installation Panoramic X ray unit Installation Instructions ORTHOPHOS XG 3 3 3 Wall mounting standard option 1 walk 2x S 10 O 2x dia 8 4 2x 8x80
82. on weight bearing wood structures i 2x 8x80 2X dia 8 4 60 51 416 D 3352 D 3352 031 03 17 02 3 Installation Panoramic X ray unit 3 3 Wall mounting standard option 1 4 Mount the upper wall holder Aia 10 mm Ava 3 8 Se 25 3 Installation Panoramic X ray unit Sirona Dental Systems GmbH 3 3 Wall mounting standard option 1 Installation Instructions ORTHOPHOS XG 3 Only with second wall holder option 1 5 Mount the lower wall holder 60 51 416 D 3352 26 D 3352 031 03 17 02 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 F 2xM8 2x dia 8 4 2x M 8x30 e Move the panoramic X ray unit into its installation posi tion at the wall Hold the unit laterally at the styrofoam packaging to do this NOTICE SIRONA recommends leaving the styrofoam packaging on the unit during the entire installation procedure If due to on site conditions it is unavoidable to remove the styrofoam packaging already at this point you may move the unit by carefully grasping the bite block bar and the stand 6 Fasten the panoramic X ray unit to the upper wall hold er Insert the screws E into the groove Screw the panoramic X ray unit firmly onto the wall holder using the washers and nuts F IMPORTANT The wall holder must be flush with the upper edge of the unit 60 51 416 D 3352 D 3352 031 03 17 02 3 Installati
83. ons provided in the Service Manual For adjustment please proceed as described in the follow ing sections 60 51 416 D 3352 D 3352 031 03 17 02 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 11 Appendix 11 2 Adjusting the panoramic X ray unit 11 2 2Iimportant information concerning adjustment WARNING When performing the following tests be sure to ob serve the radiation protection regulations applicable in your country see Operating Instructions A JX WARNING Radiation is signaled by an X ray LED and a beep IMPORTANT Be sure to take screenshots of the PAN RESET ADJUST MENT menu before and after the adjustment see section 11 2 8 and save them to the C SIDEXIS XGRAW directory along with the time and date IMPORTANT Before starting the service routines for system adjustment make sure that no unit movements are active especially di aphragm travels Otherwise the system may become inop erable in rare cases 60 51 416 D 3352 D 3352 031 03 17 02 Coarse and precision adjustment z using the Diaphragm system sa f 000mm H adjustment menu in SIDEXIS XG IV Precision adjustment Image ac IMPORTANT The Pan Sensor adjustment submenu contains a coarse adjustment and a precision adjustment precision adjust ment is preset Always try to use precision adjustment first when adjusting the sensor In most cases previous coarse adjustment is not necessary
84. ossible Demo mode OFF means The demo mode is switched OFF Radiography X ray radiation are possible Reattach cover plate B and the tube assembly covers by following the dismantling procedure in reverse order 137 11 Appendix Sirona Dental Systems GmbH 11 3 Demo mode Installation Instructions ORTHOPHOS XG 3 ORTHOPHOS XG 702 27 1 2 965 38 11 3 3lmportant information for repacking and transport 1 Switch the panoramic X ray unit on and move it to its e To uninstall and pack the unit follow the installation pro packing height by pressing the UP DOWN keys on the cedure in reverse order Multipad Bite block height 965 mm displayed as height on the Multipad Bottom edge of slide cover 702 mm IMPORTANT The bottom edge of the slide cover must be at the same height as the A markings in the column 60 51 416 D 3352 138 D 3352 031 03 17 02 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 For units installed with a floor stand NOTICE Before laying the unit down on the pallet swivel the X ray tube assembly inward and make sure to attach the center styrofoam part To ensure that the unit is in a position suitable for disman tling the floor stand after it is laid down on the pallet check to make sure that the distance between the pallet and the back of the unit is approx 49 cm
85. p side 11 2 3PAN needle phantom In order to perform the PAN sensor adjustment and the symmetry adjustment you must insert needle phantom A in the bite block holder of the panoramic X ray unit The needle phantom must be removed from the bite block holder for the pan diaphragm adjustment IMPORTANT When fitting the needle phantom make sure that it is cor rectly oriented For the adjustment of the X ray unit the phantom must be fitted in such a way that the needles point upward 2 60 51 416 D 3352 112 D 3352 031 03 17 02 Sirona Dental Systems GmbH 11 Appendix Installation Instructions ORTHOPHOS XG 3 11 2 Adjusting the panoramic X ray unit D y Oy gt SS V n 11 2 4Dismantling the fixed diaphragm The fixed diaphragm must be dismantled in order to per form the sensor adjustment of the ORTHOPHOS XG3 X ray unit To do this proceed as follows 1 Remove the covers Tube assembly front Tube assembly rear 2 Unscrew and remove the two screws B and take off the fixed diaphragm e Assemble by following the same procedure in reverse order 60 51 416 D 3352 D 3352 031 03 17 02 113 11 Appendix 11 2 Adjusting the panoramic X ray unit 1 Diaphragm system adjustment kammer3 Pan Sensor adjustment Pan Fixed diaphragm Pan Symmetry Pan Reset adjustment 4 gt S1 S2 0 00 mm S3 0 00 mm Aa V Pr
86. ps values codes etc in selection fields 1 3 1 2 3 4 digits 8 digits 4 digits 100 Keys Patient O O O O different functions symbol keysa d a b c d depending on the service routine Memory key For saving an input Service key different functions formerly AES key depending on service routine but in most cases for confirming a selection or jumping to the next test step For starting a test For moving the unit to the starting position or confirming a save operation Test key Return key Light localizer key For switching on the light localizers Up Down keys a For moving the unit up and down W Forehead support For moving the forehead sup key p port Temple support A amp i For moving the temple support key 60 51 416 D 3352 D 3352 031 03 17 02 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 A ADY J 11 1 3Selecting service routines e Select the Service menu see page 99 Selecting a service routine 1 Select the desired service routine using the arrow keys of selection field 1 and confirm your selection by press ing the Service key IMPORTANT If the selected service routine comprises several test steps the first selectable test step is displayed in selection field 2 60 51 416 D 3352 D 3352 031 03 17 02 11 Appendix 11 1 Service routines for installation oO O Selecting a test step 2 Select
87. raw the control cable into the chassis from the wall through rear opening B and fasten the chassis securely to the wall with the three screws C For concealed installation the cable length between the wall exit and the stripped wire ends should be 250 mm 60 51 416 D 3352 D 3352 031 03 17 02 51 5 Installation Remote control 5 2 Mechanical installation Concealed installation For surface installation 3 Fasten the chassis firmly to the wall with the three screws C Shortening cable L117 4 If necessary shorten cable L117 to the required length and prepare it for reconnection Expose the cable shield see condition on delivery The length of the wires should be approx 60 mm 2 3 8 Place the shortened shielding over the insulation and wrap the shielding with 3 layers of self adhe sive copper foil 52 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 Cable shield lt i approx gt 250 mm from wall exit surface installation Strip the wire ends 5 mm Crimp on the end sleeves Insert the orange white blue blue and white orange wires in separate end sleeves Attaching the strain relief 5 Attach cable L117 at the strain relief in the chassis concealed or surface installation The cable length between the strain relief and the stripped wire ends should be 200 mm 8 60 51 416 D 3352 D 3352 031 03 17 02 Sirona Denta
88. rona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 11 Appendix 11 2 Adjusting the panoramic X ray unit in SIDEXIS Preview RIEN mas 11 2 6Adjusting the PAN diaphragm IMPORTANT The fixed diaphragm must be reattached for this adjustment step see page 113 Remove the needle phantom from the bite block holder of the panoramic X ray unit Go to the PAN DIAPHRAGM submenu To make SIDEXIS XG ready for exposure Click IMAGE ACQUISITION The exposure dialog box showing the exposure status appears in Sidexis 60 51 416 D 3352 D 3352 031 03 17 02 The initialization status is visualized by a progress indi cator on the Multipad The initialization procedure is completed when the ex posure parameters of service routine 030 2 60 kV 3 mA 0 2 s are displayed and the progress in dicator disappears Take an exposure 60 kV 3 mA Press the R key on the Multipad to move the unit back to the starting position Press the release but ton Hold down the release button until image acquisition is completed and the preview image appears in the exposure dialog box 121 11 Appendix Sirona Dental Systems GmbH 11 2 Adjusting the panoramic X ray unit Installation Instructions ORTHOPHOS XG 3 Adjustment ok Adjustment not ok 4 Evaluate the X ray image IMPORTANT The exposed diaphragm area must lie horizon If th teri t fulfilled B th daoi i tally cent
89. rotective ground wire to ground Z CAUTION Incorrectly connected units can pose a risk to patients Building installation with 3x1 5 mm or 3x2 5 mm 16 AWG or 14 AWG and a 16 A 20 A overcurrent circuit breaker Connect only units which do not pose any risk to patients when the automatic circuit breaker is triggered Do not connect any EDP units See also Installation requirements Power cable e to the ORTHOPHOS X Protective ground wir to the ORTHOPHOS XG Protective ground wir Power cable 1 Check the protective ground wires and the device leak age current according to IEC 62353 2007 See sec tions 6 1 Checking the protective ground wires Page 60 and 6 2 Checking the device leakage current Page 63 2 Record the measured values in chapter 3 of the docu ment Inspection maintenance and safety related check 3 Carry out the power supply connection as shown above first 47 4 Electrical connection Sirona Dental Systems GmbH 4 2 Connecting the line voltage Installation Instructions ORTHOPHOS XG 3 Media converter A A WARNING When operating the unit at exhibitions and fairs you must observe the information provided in section 11 3 Demo mode e Plug the connector of the media converter s power sup ply unit into the electric outlet EMERGENCY STOP installation if legally prescribed Emergency Shutdown btton Power cable to the ORTHOPHOS XG Second
90. seeaeeeeenenseeeeees 68 7 4 Checking the data pathS cccccccseceeccessseeceeeeeeeeceeseeeeseeaeeessnseeeeeeess 69 60 51 416 D 3352 D 3352 031 03 17 02 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 Startup for USA Canada only ccccceeeseeceeeseeseesseeeenseeseenseeeeenseesoesseneenseesoenseesonseessoass 73 8 1 Startup Measurements and controls ccccceseeeceeceeeeeeeeeeeseeeeeseeeeeees 74 8 2 Power supply adequacy ccccesscecseececeeececeeseeeseeeeseeeeseeeeseeeesseaess 75 8 3 Fube Current Verification serieren wii nae 76 8 4 kV verification Exposure Time VerifiCation ccccccccsseeeeeeseeeeeeeees 79 8 5 Checking the laser for USA Canada Onlly cccccseeeeeeceeeeeeeeneeeeeeeenees 81 Checking the system adjustment cceeseceeeeeeeeneeeeeeeeenseeeeseseeeneeeaneeoeeseeeeneeoaseeeenenes 83 9 1 We ST CXR OSUINC S ane a a nge tesa enatlarnee 84 MAL WOUEK sees eens cictacestnctiansenntidecionced E a eet 91 10 1 Attaching the profile COVEMS ccceeeeeeeeeeceeeeneeeeeeeeeaaneeeeeessaneeeeees 92 10 2 Selecting Mote detallSnicncancndiiinuGtashaAnwaieed nee 93 10 3 Installation Report Warranty Passport cccccseeeeeeeceeeeeeeeeeeeeeeeeaeees 94 LOA UIE MENGOVECN sescsaisis staat ican tain Reet A 95 ADDENGD cara ce seganteceeseweetatca son nee seca tocar cana a weaaaecn wet ciganeate scan sneumesvecassesbaesoaseas 97 11 1 Servi
91. sition is completed and the preview image appears in the exposure dialog box 60 51 416 D 3352 D 3352 031 03 17 02 Sirona Dental Systems GmbH 9 Checking the system adjustment Installation Instructions ORTHOPHOS XG 3 9 1 Test exposures ZOOM 1 1 B 0 75 mm 0 75 mm 1 gt a A1 88 6 mm 1mm A2 44 3 0 5 mm jq _ gt ple ps a2 44 3 0 5 mm Adjustment ok length measurement with SIDEXIS Adjustment not ok 60 51 416 D 3352 D 3352 031 03 17 02 89 9 Checking the system adjustment Sirona Dental Systems GmbH 9 1 Test exposures Installation Instructions ORTHOPHOS XG 3 4 Evaluate the X ray image The shadow of the center needle the needle image and the auxiliary line must be coincident and located behind each other A tolerance offset of needle from the central auxiliary line of 0 75 mm is admissible A Distance A1 must be 88 6 1 mm A Distances A2 must be identical each being 44 3 0 5 mm A A white border surrounding the image on all sides must be visible B IMPORTANT If one of these criteria is not fulfilled C the pan symmetry must be adjusted The procedure for adjusting the unit is de scribed in the Appendix see section 11 2 and in the OR THOPHOS XG Service Manual 60 51 416 D 3352 90 D 3352 031 03 17 02 Kr The Dental Company 10 Final work 60 51 416 D 3352 D 3352 031 03 17 02 91 10 Final work Sirona Dental Systems G
92. stancy test UTILITIES gt CONSTANCY TEST The classical SIDEXIS user interface is started Con stancy test is already preset 2 Start the exposure mode Click XCXP The dialog box for selecting the X ray device opens IMPORTANT If no X ray device has been configured yet in SIDEXIS XG the password input window appears instead of the dialog box for selecting the X ray device For creating a new X ray device see documentation SIDEXIS XG Installation of Stand Alone Systems 69 7 Initial startup 7 4 Checking the data paths Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 3 Select X ray device OX Existing devices P2k_F ammer4 l E Cancel 4 Acceptance test Constancy test Cancel 3 Select confirm the X ray device Select e g ROOM 3 and click OK IMPORTANT The SELECT X RAY DEVICE dialog appears only if several X ray devices are installed The dialog box for selecting the test type appears on the screen 4 Select confirm the test type Click SERVICE EXPOSURE The dialog box for selecting the service exposure ap pears on the screen 5 Select confirm the service exposure Click DIGITAL TEST PATTERN f several different X ray components are available a dialog box for selecting the X ray component appears on the screen In this case select confirm the required component lf only one X ray component is available the expo sure read
93. sting the panoramic X ray unit Installation Instructions ORTHOPHOS XG 3 9 in SIDEXIS Pan Sensor adjustment Pan Fixed diaphragm Pan Symmetry Pan Reset adjustment 4 gt Preview A ana A ap 9 0000 S KW mA vV Prog v 5 Make SIDEXIS XG ready for exposure 6 Take an exposure 60 kV 3 mA Click IMAGE ACQUISITION Press the release button Hold down the release The exposure dialog box showing the exposure status button until image acquisition is completed and the appears in Sidexis preview image appears in the exposure dialog box The initialization status is visualized by a progress indi cator on the Multipad The initialization procedure is completed when the ex posure parameters of service routine S030 2 60 kV 3 mA 0 2 s are displayed and the progress in dicator disappears 60 51 416 D 3352 124 D 3352 031 03 17 02 Sirona Dental Systems GmbH Installation Instructions ORTHOPHOS XG 3 11 Appendix 11 2 Adjusting the panoramic X ray unit 7 Evaluate the X ray image If the image is identical to the ideal image A then the The exposed diaphragm area must lie horizon adjustment of the diaphragm is completed tally centered in the image field as well as inside the superimposed auxiliary lines A A white border surrounding the image on all sides must be visible The maximum density must lie in the center of the diaphragm area A I
94. the desired test step using the arrow keys of se lection field 2 and confirm your selection by pressing the Service key Quitting the service routine To return to the service routine selection menu press the Service key or the right up arrow key 4 To quit the Service menu and return to the main menu press the up arrow key 4 at the very right 101 11 Appendix Sirona Dental Systems GmbH 11 1 Service routines for installation Installation Instructions ORTHOPHOS XG 3 11 1 4Service routines with security access IMPORTANT A security code is required for accessing service routines involving functions such as radiation release or editing of configuration data or stored values This procedure prevents the inadvertent selection or activa tion of these service routines To select a service routine or test step with security access proceed as follows A WARNING Be sure to observe the radiation protection regulations applicable in your country Selecting the service routine test step 1 Select the service routine or the test step and confirm your selection with the Service key After you have confirmed your selection a 0 appears in selection field 1 Confirming the security access 2 Confirm the security access by once again selecting the number of the main routine in our example 2 in se lection field 2 with the arrow keys Confirm your input by pressing the Service key Following this double se
95. tor In case of doubt contact the manufacturer of the sys tem components WARNING Wireless phone interference with medical electrical equipment To ensure safe operation of medical electrical equip ment the use of mobile wireless phones in practice or hospital environments is prohibited 60 51 416 D 3352 D 3352 031 03 17 02 Z CAUTION The unit contains class 1 lasers Keep a distance of at least 4 10 cm between eye and la ser Do not stare into the beam Do not use the system with any other lasers and do not make any changes to settings or processes that are not de scribed in these operating instructions This may lead to a dangerous exposure to radiation NOTICE Use an ESD wrist band during installation Connect it to the protective ground wire UMMM X NOTICE When opening the unit Please observe the usual precautionary bh measures for handling PCBs ESD Touch a ground point to discharge static elec t A tricity before handling any components NOTICE Extreme fluctuations of temperature may cause condensa tion inside the unit Do not switch the unit on before it has reached normal room temperature 1 Before you begin Sirona Dental Systems GmbH 1 3 System and sensor versions Installation Instructions ORTHOPHOS XG 3 1 3 System and sensor versions 1 ORTHOPHOS XG 3 Digital unit 2 XG PAN sensor Sensor for panoramic PAN X ray 60 51 416 D 3352 8 D 3352 031 03 17 02
96. utton until image acquisition is completed and the preview image appears in the exposure dialog box 60 51 416 D 3352 D 3352 031 03 17 02 Sirona Dental Systems GmbH 11 Appendix Installation Instructions ORTHOPHOS XG 3 11 2 Adjusting the panoramic X ray unit Adjustment ok Adjustment not ok 4 Evaluate the X ray image The three needle images must lie in the center of the exposed areas and inside the auxiliary lines A IMPORTANT If these criteria are not fulfilled B the PAN sensor must be adjusted 60 51 416 D 3352 D 3352 031 03 17 02 115 11 Appendix Sirona Dental Systems GmbH 11 2 Adjusting the panoramic X ray unit Installation Instructions ORTHOPHOS XG 3 Image X ray image with unadjusted pan sensor Coarse or precision adjustment Alternatively with precision adjustment setting B and coarse adjustment setting A sensor adjustment can usually be performed directly via precision adjustment Only in exceptional cases e g if one or more needles are completely outside of the image field A in an image acquired with the PRECISION ADJUSTMENT presetting is it necessary to perform a coarse adjustment prior to precision adjustment B To do this deactivate the PRECISION ADJUSTMENT check box see page 111 and then perform a coarse adjustment proceeding in the same way as for precision adjustment The only difference between coarse and precision adjustment is the size of the image area considered Furthermore ther