Home
JAWAHARLAL INSTITUTEOF POST GRADUATE MEDICAL
Contents
1. 4 pairs of outlet terminator Four gel chambers should be connected simultaneously 1 time Safety device No load detection Leakage detection Sudden load change detection Over temperature protection Over load detection Shrouded plug and sockets Program Storage 30 programmed files Accessories to be provided at no extra cost power pack adaptor replacement fuse 2 no Standards Safety and Training e Should be FDA CE UL or BIS approved product e Electrical safety conforms to standards for electrical safety IEC 60601 IS 13450 e The manufacturer supplier has to provide adequate training in operation and maintenance of the equipment at the site of installation Warranty e The supplied equipment and all accessories should be under warranty for a period of 3 years after successful commissioning e All spare parts PC boards and service manuals should be available with the local service centre during the warranty period and steps should be taken immediately for servicing when required to minimize down time Annual Maintenance Contract e Annual maintenance contract rates for a period of 5 years after the end of warranty period should be quoted separately and this would be taken into consideration during price comparisons e Annual maintenance contract should include preventive maintenance as well as breakdown calls A copy of the service manual should be available with the local service centre
2. Installation Commissioning Testing Maintenance and After Sales Service e The equipment and all accessories should be installed tested and commissioned at the Clinical Immunology Laboratory Department of Clinical Immunology JIPMER free of cost e All spare parts and consumables should be available with the supplier or principals for a period of ten years after commissioning JAWAHARLAL INSTITUTEOF POST GRADUATE MEDICAL EDUCATION AND RESEARCH PUDUCHERRY 605006 SPECIFICATION FOR TABLE TOP CENTRIFUGE Operational Requirements Microprocessor based tabletop centrifuge that use brushless induction motor with frequency drive enabling the user to pre set speed amp time with a high degree of accuracy Display of set parameters like speed time etc make the unit an ideal choice for repetitive sample analysis Features e Brushless Induction motor with frequency drive e Stable speed output even under unstable voltage conditions e 7segment LED display of speed R 8C BL e Digital countdown timer e Safety lid interlock to prevent lid opening during centrifugation e Dynamic brake for quick deceleration e Imbalance amp Inverter fault detection with auto shutdown e Recall of last set parameters Useful for repetitive analysis e Automatic door opening through gas hinges Technical specifications e Max Speed 5250 rpm e Max RCF 3600 g e WxDxM 365 x 415 x 350 mm e Type of Head swing out e Maximum capacity 400 ml e Swin
3. lt 30min 20 C to 10 C Standard blocks must be provided as part of equipment not as optional A BLOCK 96 well Plate B BLOCK 0 5ml x 54 C BLOCK 1 5ml x 35 D BLOCK 2ml x 35 E BLOCK 50 ml Power supply 230V 50 60 Hz Accessories to be provided at no extra cost Block extraction tool must be supplied allowing blocks to be removed easily Standards Safety and Training e Should be FDA CE UL or BIS approved product e Electrical safety conforms to standards for electrical safety IEC 60601 IS 13450 e The manufacturer supplier has to provide adequate training in operation and maintenance of the equipment at the site of installation Warranty 1 The supplied equipment and all accessories should be under warranty for a period of 3 years after successful commissioning 2 All spare parts PC boards and service manuals should be available with the local service centre during the warranty period and steps should be taken immediately for servicing when required to minimize down time Annual Maintenance Contract 1 Annual maintenance contract rates for a period of 5 years after the end of warranty period should be quoted separately and this would be taken into consideration during price comparisons 2 Annual maintenance contract should include preventive maintenance as well as breakdown calls A copy of the service manual should be available with the local service centre Installation Commiss
4. C Temperature control Electronic microprocessor based Temperature Sensitivity 0 5 1 C Temperature Sensor Audio visual alarm if temperature deviates from preset temperature Body Inner body SS and outer body MS with powder coated Water circulation Inside the tank through integrated pump to maintain precise temperature uniformity through the liquid medium Tank dimension 30cm x 25cm x 15cm approximately Input voltage 230 V 10 50Hz 15 amps A C Supply Cleaning Should be easy to clean the tank Standards Safety and Training Should be FDA CE UL or BIS approved product Electrical safety conforms to standards for electrical safety IEC 60601 IS 13450 BIS standards The manufacturer supplier has to provide adequate training in operation and maintenance of the equipment at the site of installation WARRANTY The equipment and all accessories should be under WARRANTY for a period of THREE YEARS after successful commissioning All essential spare parts PC boards and service manuals should be available with the local services during WARRANTY period and all steps should be taken for immediate servicing to prevent the down time Annual Maintenance Contract The Annual maintenance Contract rates for a period of 3 years after the warranty period should be quoted separately and this would be taken into consideration in comparing the price bids Annual Maintenance Contract should include preventive maintenance and br
5. Sales Service e The equipment and all accessories should be installed tested and commissioned at JIPMER Pondicherry free of cost e All spare parts and consumables should be available with the supplier or principals for a period of 10 years
6. DC Current 1 400 mA Power 1 100 W Time 00 01 500 h Volt hour 1 500 000 Vh Output resolution Voltage 1 V Current 100 uA 0 39 9 mA 1 mA 40 400 mA Power 1W Programming resolution Voltage 1 V Current 1 mA Power 1W Time 1 min 00 01 99 59 1h 100 500 h Volt hour 1 Vh 1 9999 Vh 100 Vh 10 0 99 9 kVh 1 kVh 100 500 kV Accuracy Voltage 1 3V Current 1 1 mA Power 2 1 W Timer 0 1 1 min 00 01 99 59 h 0 1 10 min 100 500 h Line regulation lt 0 2 Load regulation lt 1 at load change 10 90 of maximum load Ripple lt 1 at 600 V Short term stability lt 0 2 10 h after warm up Long term stability lt 1 year Start current check Resistance not greater than 0 2 MQat 40 V current less than 200 uA Can be disabled Ground leakage check Leakage not greater than 500 uA Output protection Fully protected against any overload conditions Timers 1 999 mins with alarm continuous Recovery after power failure Duration lt 8 s The program must continue automatically Duration gt 8 s The program must continue after manual restart Ambient operating 4 40 C temperature Ambient operating 0 95 humidity Ambient operating pressure 68 106 kPa maximum altitude of 2000 m Mains requirements 100 120 220 240 V 50 60 Hz Power consumption Max 140 W Terminator pairs
7. JAWAHARLAL INSTITUTEOF POST GRADUATE MEDICAL EDUCATION AND RESEARCH PUDUCHERRY 605006 SPECIFICATIONS FOR HEAT BLOCK Operational Requirement The heat block unit must be ideal for clinical laboratory applications like incubation boiling inactivation wet washing enzyme analysis and other clinical uses Technical Specifications Should be a micro computer controlled product withan advanced thermoelectric technique with a LCD display Must be easy to setup and use Must have a robust construction for long term durability and reliability Small light and compact foot print Should possess a simultaneous display of set and actual time temperature Should have a solid anodized aluminum construction ensuring maximum heat transfer between the heater base and tube contents Should have a reliable design with inner extra temperature protection Change of blocks must be easy and convenient Should have a wide range of controlling temperature and temperature deflection for differing purposes Standard blocks given in the table below as a part of the unit for versatile applications Temperature setting Range 5 105 C Temperature control Range 10 100 C Timing Range 1 min 99h59min Temperature control Accuracy lt 0 5 C Temperature setting Push button Display Accuracy 0 1 C Temperature uniformity across the block lt 0 5 C Heating Rate lt 25min 20 C to 100 C Cooling Rate
8. ation WARRANTY e The equipment and all accessories should be under WARRANTY for a period of THREE YEARS after successful commissioning e All essential spare parts PC boards and service manuals should be available with the local services during WARRANTY period and all steps should be taken for immediate servicing to prevent the down time Annual Maintenance Contract e The Annual maintenance Contract rates for a period of 3 years after the warranty period should be quoted separately and this would be taken into consideration in comparing the price bids e Annual Maintenance Contract should include preventive maintenance and breakdown calls A copy of service manual should be available with local service centre Installations Commissioning Testing Maintenance and After Sales Service e The equipment and all accessories should be installed tested and commissioned at JIPMER Pondicherry free of cost e All spare parts and consumables should be available with the supplier or principals for a period of 10 years JAWAHARLAL INSTITUTEOF POST GRADUATE MEDICAL EDUCATION AND RESEARCH PONDICHERRY 605006 SPECIFICATIONS FOR WATER BATH WITH TEMPERATURE CONTROL 37 C and 56 C 2 Numbers Operational requirement 1 Water bath for use in Clinical Immunology Laboratory 56 C 2 Water bath for use in Immunotherapy Cytatherapy unit 37 C Technical specification Maximum capacity 10 12 Litres Temperature range 25 C to 100
9. d allow processing of up to 24 X 0 5ml tubes or 6 X 50ml tubes at a time 3 Construction e Stainless steel basket designed specifically to generate maximum ultrasonic activity prevent items resting on the tank and prevent operators coming into contact with chemical solutions e Plastic lid to reduce noise and minimise potential of aerosol escape 4 Operation at 200 watts ultrasonic frequency 24kHz 5 Automatic frequency tuning system for accurate frequency control 6 Amplitude adjustable from 20 to 100 7 Pulse adjustable from 0 to 100 8 Built in timer 9 Easy to use single touch LCD control panel with user settable parameters 10 Drain valve for convenient emptying 11 Minimal noise levels homogeneous ultrasonic activity throughout the tank minimal dead spots and standing waves 12 Accurate process control of time temperature ultrasonic activity degas and power 13 One bottle of ultrasonic solution to be included as standard Ambient conditions for operation The unit should be capable of withstanding continuous storage and operation at ambient temperatures of 0 50 C and relative humidity of 15 90 Power Requirements Suitable for input power of 220 240V AC 50 Hz fitted with suitable plug for Indian conditions Should have built in safety measures like resettable over current breaker Standards Safety and Training e Should be FDA CE UL or BIS approved product e Electrical safety conforms to standards for elec
10. eakdown calls A copy of service manual should be available with local service centre Installations Commissioning Testing Maintenance and After Sales Service The equipment and all accessories should be installed tested and commissioned at the JIPMER Pondicherry free of cost All spare parts and consumables should be available with the supplier or principals for a period of 10 years JAWAHARLAL INSTITUTEOF POST GRADUATE MEDICAL EDUCATION AND RESEARCH PUDUCHERRY 605006 SPECIFICATIONS FOR PROGRAMMABLE HIGH VOLTAGE POWERPACKS Operational Requirement Programmable high voltage power packs suitable for molecular biology applications Technical Specifications e Should be an ultra compact sized unit providing power to run horizontal agarose or vertical polyacrylamide electrophoresis e Should be of a high quality high precision and must ensure safe power supply for electrophoresis applications that require a maximum of 600 volts and 400 mA e The power pack must be ideal for the following applications like SDS PAGE Polyacrylamide Gel Electrophoresis Native PAGE Agarose electrophoresis Electroblotting DNA pulsed field electrophoresis DNA fragment separation DNA submarine electrophoresis Regulation Maximum voltage current and power with automatic cross over at preset limits Control Microprocessor controlled Output range Voltage 0 600 V DC Current 0 400 mA Power 0 100 W Programming range Voltage 6 600 V
11. g out Rotor heads for running 8 x 15 ml tubes with Maximum speed of 4200 rpm and Maximun RCF of 3000 g e Swing out Rotor heads for running 6 x 50 ml tubes with Maximum speed of 4000 rpm and e Maximun RCF of 2800 g Power Requirements e Suitable for input power of 220 240V AC 50 Hz fitted with suitable plug for Indian conditions e Resettable over current breaker Standards Safety and Training e Should be FDA CE UL or BIS approved product e Electrical safety conforms to standards for electrical safety IEC 60601 IS 13450 e The manufacturer supplier has to provide adequate training in operation and maintenance of the equipment at the site of installation WARRANTY e The equipment and all accessories should be under WARRANTY for a period of THREE YEARS after successful commissioning e All essential spare parts PC boards and service manuals should be available with the local services during WARRANTY period and all steps should be taken for immediate servicing to prevent the down time Annual Maintenance Contract e The Annual maintenance Contract rates for a period of 3 years after the warranty period should be quoted separately and this would be taken into consideration in comparing the price bids e Annual Maintenance Contract should include preventive maintenance and breakdown calls A copy of service manual should be available with local service centre Installations Commissioning Testing Maintenance and After
12. ioning Testing Maintenance and After Sales Service 1 The equipment and all accessories should be installed tested and commissioned at the Clinical Immunology Laboratory Department of Clinical Immunology JIPMER free of cost 2 All spare parts and consumables should be available with the supplier or principals for a period of ten years after commissioning JAWAHARLAL INSTITUTEOF POST GRADUATE MEDICAL EDUCATION AND RESEARCH PUDUCHERRY 605006 SPECIFICATIONS FOR VACUUM MANIFOLD AND PUMP 1 No Operational Requirement Vacuum Manifold with suitable pump for processing of 96 well 384 well filter plates for use with multiplex bioassay analyzer Luminex system Technical Specifications 1 Should be compatible with 96 well filter plates and 384 well filter plates Should be compatible with all vacuum filtration applications Manifold components should be resistant to solvents Should be suitable for use with automated or manual protocols Should be easily configured for both flow to waste and analyte collection Plate and manifold designs should combine to eliminate gaps between the flow directors and receiver wells 7 Should provide a plate on plate processing option for assays where filtrate collection is required 8 Effortless vacuum initiation 9 Should be possible to do complete filtration assay cycles with no manual intervention 10 Should enable crosstalk free filtrate collection 11 Configurations for deep well or standard receiver plate
13. late Shaker required for stirring very low volumes of multiple samples on 96 well plates Technical Specifications Capacity Should allow processing of four 96 well micro plates at a time Construction stainless steel Powder coated heavy gauge MS body Accurate speed control at 20 to 600 rpm Minimal noise levels Electronic DC motor controlled Minimal footprint to volume ratio Should be fitted with digital electronic display of speed and timer MORN Ambient conditions for operation The unit should be capable of withstanding continuous storage and operation at ambient temperatures of 0 50 C and relative humidity of 15 90 Power Requirements Suitable for input power of 220 240V AC 50 Hz fitted with suitable plug for Indian conditions Should have built in safety measures like resettable over current breaker Standards Safety and Training e Should be FDA CE UL or BIS approved product e Electrical safety conforms to standards for electrical safety IEC 60601 IS 13450 e The manufacturer supplier has to provide adequate training in operation and maintenance of the equipment at the site of installation Warranty e The supplied equipment and all accessories should be under warranty for a period of 3 years after successful commissioning e All spare parts PC boards and service manuals should be available with the local service centre during the warranty period and steps should be taken immediately for servicing when required to
14. minimise down time Annual Maintenance Contract e Annual maintenance contract rates for a period of 5 years after the end of warranty period should be quoted separately and this would be taken into consideration during price comparisons e Annual maintenance contract should include preventive maintenance as well as breakdown calls A copy of the service manual should be available with the local service centre Installation Commissioning Testing Maintenance and After Sales Service e The equipment and all accessories should be installed tested and commissioned at the Clinical Immunology Laboratory Department of Clinical Immunology JIPMER free of cost e All spare parts and consumables should be available with the supplier or principals for a period of ten years after commissioning JAWAHARLAL INSTITUTEOF POST GRADUATE MEDICAL EDUCATION AND RESEARCH PONDICHERRY 605006 SPECIFICATIONS FOR VORTEX MIXER e Speed 100 3200 rpm e Operation Touch continuous e Working condition 4 60 C e Accessory 10 15 tube insert attachment for different tube size e Vortexing motion Circular orbit e Input voltage 230 V 10 50Hz Single phase Standards Safety and Training e Should be FDA CE UL or BIS approved product e Electrical safety conforms to standards for electrical safety IEC 60601 IS 13450 e The manufacturer supplier has to provide adequate training in operation and maintenance of the equipment at the site of install
15. nt and all accessories should be under warranty for a period of 3 years after successful commissioning 4 All spare parts PC boards and service manuals should be available with the local service centre during the warranty period and steps should be taken immediately for servicing when required to minimise down time Annual Maintenance Contract 3 Annual maintenance contract rates for a period of 5 years after the end of warranty period should be quoted separately and this would be taken into consideration during price comparisons 4 Annual maintenance contract should include preventive maintenance as well as breakdown calls A copy of the service manual should be available with the local service centre Installation Commissioning Testing Maintenance and After Sales Service 3 The equipment and all accessories should be installed tested and commissioned at the Clinical Immunology Laboratory Department of Clinical Immunology JIPMER free of cost 4 All spare parts and consumables should be available with the supplier or principals for a period of ten years after commissioning JAWAHARLAL INSTITUTEOF POST GRADUATE MEDICAL EDUCATION AND RESEARCH PUDUCHERRY 605006 SPECIFICATIONS FOR BATH SONICATOR Operational Requirement Required for processing of microbeads for multiplexing applications for use with multiplex bioassay analyser Luminex system Technical Specifications 1 Ambient heating 5 C to 70 C 2 Capacity Shoul
16. s 12 Footprint should allow for easy robotic deck integration D UPW O VACUUM PUMP 1 Piston driven high output Vacuum Pump with 70 cm of 1 4 tubing and a 50mm filter for in line moisture protection Vacuum max 921 mbar 27 2 in Hg Pressure max 5 4 bar 80 psig Flow rate max 34 L min Construction Cast aluminium Weight 5 3kg Dimensions 8 x 9 x 10 in Connections Stepped hose barb Should come with a 2L vacuum filter flask with cork and 1 4 solvent resistant tubings 0 00 OBS SYRINGE FILTER 50 mm bi directional syringe filter units with hydrophobic membrane for vacuum line protection Material Polypropylene Filtration Area 19 6 cm 60 psi 4 1 bar maximum inlet pressure 5 Operating temperature up to 121 C Ambient conditions for operation 1 The unit should be capable of withstanding continuous storage and operation at ambient temperatures of 0 50 C and relative humidity of 15 90 Power Requirements 1 Suitable for input power of 220 240V AC 50 Hz fitted with suitable plug for Indian conditions 2 Resettable over current breaker iw IS Standards Safety and Training e Should be FDA CE UL or BIS approved product e Electrical safety conforms to standards for electrical safety IEC 60601 IS 13450 e The manufacturer supplier has to provide adequate training in operation and maintenance of the equipment at the site of installation Warranty 3 The supplied equipme
17. trical safety IEC 60601 IS 13450 e The manufacturer supplier has to provide adequate training in operation and maintenance of the equipment at the site of installation Warranty e The supplied equipment and all accessories should be under warranty for a period of 3 years after successful commissioning e All spare parts PC boards and service manuals should be available with the local service centre during the warranty period and steps should be taken immediately for servicing when required to minimise down time Annual Maintenance Contract e Annual maintenance contract rates for a period of 5 years after the end of warranty period should be quoted separately and this would be taken into consideration during price comparisons e Annual maintenance contract should include preventive maintenance as well as breakdown calls A copy of the service manual should be available with the local service centre Installation Commissioning Testing Maintenance and After Sales Service e The equipment and all accessories should be installed tested and commissioned at the Clinical Immunology Laboratory Department of Clinical Immunology JIPMER free of cost e All spare parts and consumables should be available with the supplier or principals for a period of ten years after commissioning JAWAHARLAL INSTITUTEOF POST GRADUATE MEDICAL EDUCATION AND RESEARCH PUDUCHERRY 605006 SPECIFICATIONS FOR MICROPLATE ROCKER SHAKER Operational Requirement Micro P