Home
Vitalab Flexor E Service Manual
Contents
1. Photo meter board System board start stop CLK from KIO 115 10V C4 f IC7a MAM f 2 100mV reference 4 8 16 bits counter Vx IC13 Input A Amplifier meas IC2 ref Figure 3 10 Lin Log converter Conversion is achieved by utilizing the logarithmic discharge curve of capacitor C4 The voltage over this capacitor is compared to the linear voltage Vx and the time from the beginning of the discharge until both voltages are equal correspond to the logarithm of the measured signal The discharge time is measured by starting a counter at the moment that the actual discharge is activated The counter will be stopped the moment both voltages are equal measuring and capacitor voltage The lin log converter circuit consists of e Precision voltage source which supplies a very accurate 10V IC5 Logarithmic curve generator IC2 and C4 Comparator 1 4 FET switches IC8 Digital parts that consist of a Counter Time Control CTC which is situated on the system board VITAL SCIENTIFIC 3 15 3 SYSTEM ORGANIZATION SERVICE MANUAL At the start of the logarithmic conversion capacitor C4 will be charged to 10 Volts derived from IC5 bv means of closing the FET switch IC8d After a certain time 1 msec the counters in the CTC will be enabled At the same moment the selection of the input signa
2. 395 OT3yRGA 5 17 31 S34NY33101 431 15 ISIMUIHLO 1068 1009 HILIMS WANJVA 08 08 SJM WA SMOTB8 68L 1009 LW VITAL SCIENTIFIC 8 ELECTRICAL DIAGRAMS SERVICE MANUAL WNLVOHJAJI 3LWO A83AI130 SONYTHJHLIN DIALINAIOS 6 Vivi sme 200173135 NUM 38 NOR diti3530 DNNZINVHIO INSA VARA QwiNvY ALUNYNO 19NVAM3A SDV 3003 CUM SSINHSNOW 3509 08 5 A w 9 4 aas jun dund 21009 28 1009 a x 03V08 IW LV 158 1009 9 2 8004 aas SINDA UnnoDA 9 2 8004 aas SaA DA SMO 3q 04 08 WALSAS IV 6511009 VITAL SCIENTIFIC 8 ELECTRICAL DIAGRAMS SERVICE MANUAL NIHIMNV AS 3F31N3DS WIM VYLIITIS 9 SNWYNIS NOLLAWISAQ MINLVONZAZI Y avd 2002 LINWND Maiv DIHMAMW 3o JONVANJA sean OT3WAJA SH3ONY FIZNAL S31NV33101 Q31V1S 3SIMH3H10 0808 318v 001 009 EX 1 sl 9 ib L
3. gt gt gt gt SA Figure 2 5 Syringe drive assembly AGENT SYRINGE S MPLE SYRINGE STEPPER STEPPER MOTOR MOTOR The lower opto is also used for home initialisation during the reset The four opto s are mounted together on a separate PCB The syringes are made of glass and the plungers are teflon tipped The tubes that are connected to the valves are also made of teflon 2 12 VITAL SCIENTIFIC SERVICE MANUAL 2 FUNCTIONAL DESIGN 2 3 7 ISE ARM Optional The ISE arm is integrated in the mixer assembiv Figure 2 6 shows the complete assembiv The communication with the external ISE measuring unit is done bv a serial RS232 inter face The serial interface connector is mounted at the left side of the instrument Mixer ISE needle Guide Opto coupler To ISE unit Figure 2 6 ISE arm An electro magnet that moves the needle into its aspirate position drives the ISE needie An opto switch detects the lower position of the needle The electronic driver circuits for magnet and mixer are located on the ISE driver board The svstem board does the control of these functions VITAL SCIENTIFIC 2 13 2 FUNCTIONAL DESIGN SERVICE MANUAL 2 3 8 MOTOR ENCODERS Some of the stepper motors are equipped with an encoder disc on the shaft In conjunction with two opto switches A and B they check the number of steps an
4. SAMPLE NEEDLE RINSING PLACE RINSING PLACE lt 2 2 REAGENT NEEDLE SAMPLE NEEDLE Y v v AY y 9 ISE waste 522 V5 1 1 a a gt WATER BOTTLE DILUTED CONCENTRATED WASTE BOTTLE WASTE BOTTLE gt v9 v v6 15 22 222 12222 4 42 1 42 ag ag Us Ze ue 3 3 3 3 3 3 Y JU gt 252221255 223 2 ki CUVETTE ROTOR 4 4 4 A 2 RINSING PLACE STEPPER MOTOR 1 BELLOWS PUMP a ASCO VALVE 3 way FAS VALVE 2 wayl PORT 1 PORT 1 PORT 3 DISPENSER n E PORT 2 PORT 2 OFF PORT 1 3 OPEN OFF PORT 1 2 CLOSED SAMPLE NEEDLE REAGENT NEEDLE ON PORT 2 3 OPEN ON PORT 1 2 OPEN Valve no 1 4 5 6 7 Valve no 2 3 13 14 8 9 10 12 Figure 2 12 Fluid and vacuum system Optional the instrument is equipped with two separated waste circuits for diluted and concentrated waste The concentrated waste consists of the waste of the first needle from the wash arm the reaction mixture and the waste from the optional ISE unit The normal diluted waste is coming from the rest of the system like the other needles from the wash arm and the wash positions for reagent and sample needle The waste for both systems is collected into two separate waste containers 2 20 VITAL SCIENTIFIC SERVICE MANUAL 2 FUNCTIONAL DESIGN 2 6 1 FILLING OF THE C
5. VITAL SCIENTIFIC SERVICE MANUAL 8 ELECTRICAL DIAGRAMS ISE SIPPER STIRRER DRIVER BOARD 6001 773 PART NUMBER 3302 180 3302 317 3302 401 3302 465 3315 009 3320 100 3320 150 3320 232 3320 312 3326 521 3326 554 3333 545 3353 057 3355 002 3376 550 3381 052 3820 115 DESCRIPTION RES METAL 33 OHM 0 6W RES METAL 150 OHM RES METAL 1K OHM RES METAL 4K7 OHM RES N W 4X4K7 CAP CER 1NF 40V CAP CER 4 7NF 40V CAP CER 0 022UF 50V CAP CER 220nF 63V CAP ELC 15uF 20V CAP ELC 47uF 63V L 6202 DMOS FULLBRIDGE DR 1 5A CONN 12POL 826470 6 KEYING PLUG 926498 1 ISE SIP STIR DRIVER BRD OPT SWITCH H22B1 STICKER LOCATION R4 R 3 5 R1 R2 RN 1 C1 3 8 9 15 16 18 C2 C6 7 13 14 C5 12 C10 11 C4 IC1 2 PL1 PL1 OC1 Comp ident comp side VITAL SCIENTIFIC SERVICE MANUAL 8 ELECTRICAL DIAGRAMS FA SA STIRRER A 5 1 0 2 A ci 16282 1n 22n SP2 RA STIRRER SP3 SIPP c 22n F3 ISE SIPPERARM Eo SPS 0 65 A C6 22n c2 4n SP6 TOLERANCES UNLESS OTHERUISE STRTED DISPLACES DI ENSIONS ANGLE ROUGHNESS Ra ESEJ a NO 6001 773 fa SHEET 171 439 BA DESCRIPTION ISE SIPPERNSTIRRER DRIVER BOARD DATE 39 03 94 CHECK J Ribberink DATE 09 03 95 VITAL SCIENTIFIC DIEREN THE NETHERLANDS SCIENTIFIC VITAL SCIENT
6. Cause This error will occur in combination with E140 when there is no communication with the ISE Solution See E140 E144 ISE ARM ERROR Cause This error will be generated when the ISE arm is in the lower position when its not allowed probablv touched bv hand Solution Press Check Again and continue E145 ISENOT READV Cause This error will be generated when the ISE is executing a command and another command is sent or when the ISE is in the warming up procedure Solution Press Check Again and continue E146 ISE STANDARD A NOT DETECTED Cause This error will occur when during calibration the liquid is not detected within in certain window Solution See 7 2 of ISE service manual STANDARD NOT DETECTED E147 ISE STANDARD B NOT DETECTED Cause This error will occur when during calibration the liquid is not detected within in certain window Solution See section 7 2 of ISE service manual STANDARD A B NOT DETECTED E150 ISE CHECK TEMPERATURE Cause The temperature is outside its limits 10 C 50 C Solution See section 7 2 of ISE service manual CHECK TEMPERATURE E151 ISE NOT CALIBRATED Cause This error occurs when the ISE detects that the calibration is not correct or when the cover was open for a longer period of time Solution Perform a new calibration E152 ISE SANPLE DETECTION ERROR Cause This error will occur when no sample is detected at the wrong moment Solution See se
7. PART NUMBER DESCRIPTION LOCATION 3002 025 SCREW ST ST M3X30 1 3004 214 SCREW ST ST M3X12 2 3022 010 WASHER ST ST 3 2 3 3029 049 O RING 36X2 4 3064 017 BELT TIMING 132T T2 5 5 6001 464 INNER STATOR ASSY 6 6001 465 ROTOR ASSY 7 6001 820 STATOR ASSY 8 6001 782 INSULATION ASSY 9 INNER STATOR ASSEMBLY 6001 464 PART NUMBER DESCRIPTION LOCATION 3002 005 SCREW ST ST M2X12 10 3008 040 SCREW ST ST M4X5 11 3022 005 WASHER ST ST 2 2 12 3060 053 LENS 08 F13 5 13 3750 007 SPRING 14 4100 452 INNER STATOR 15 4105 193 BUSH 16 4501 466 BLOCK 17 6001 447 PHOTO DIODE UNIT 18 ROTOR ASSEMBLY 6001 465 PART NUMBER DESCRIPTION LOCATION 3053 036 BALL BEARING 19 3130 080 SPROCKET 20 4509 150 ROTOR 21 STATOR ASSEMBLY 6001 820 PART NUMBER DESCRIPTION LOCATION 3002 015 SCREW ST ST M3X4 22 3029 040 O RING 23 3060 053 LENS 08 F13 5 24 4509 151 STATOR 25 4802 286 PLATE 26 6001 822 THERMISTOR ASSEMBLY 27 6001 821 ROTOR RESET BOARD 28 INSULATION ASSEMBLY 6001 782 PART NUMBER DESCRIPTION LOCATION 3002 002 SCREW ST ST M2X5 29 3002 018 SCREW ST ST M3X8 30 3022 005 WASHER ST ST 2 2 31 3022 010 WASHER ST ST 3 2 32 3023 006 WASH LOCK ST ST M3 33 4509 170 INSULATION BLOCK 34 4802 256 PLATE 35 6001 469 REFERENCE PHOTO DIODE UNIT 36 ENCODER STEPPER MOTOR UNIT 6001 430 PART NUMBER DESCRIPTION LOCATION 3004 204 SCREW ST ST M2X6 37 3008 040 SCREW ST M4X5 38 3130 076 SPROCKET T2 5 15 39 3363 505 STEPPER MOTOR 40 4700 543 ENCODER DISC 41 ENC
8. 0 30 0 40 Bar VACUUM 6 Also check at which level the detector is switching from NO VACUUM to VACUUM This must be at approx 0 40 bar VITAL SCIENTIFIC 5 15 5 MECHANICAL ADJUSTMENTS SERVICE MANUAL 5 10 SYRINGE TIP ADJUSTMENT 100 uL ONLY The syringe tip adjustment has to be done when a complete new syringe or a new plunger is placed The adjustment of the sample syringe is as follows 1 2 5 16 Remove the complete syringe from the dispenser unit Use the CHANGE SYRINGE menu Loosen the set screw of the plunger stop and adjust the plunger in sucha way that the plunger stop is against the bottom of the glass barrel and the top of the teflon tip is in line with the top of the syringe TOP OF TEFLON TIP epu PLUNGER STOP SET SCREW Figure 5 12 Sample syringe tip adjustment Fix the adjust screw again Place the syringe in the dispenser unit VITAL SCIENTIFIC SERVICE MANUAL 5 MECHANICAL ADJUSTMENTS 5 11 OPTO ENCODER ADJUSTMENT The opto encoder checks the strobe disk position The position of the two opto s is adjusted at the factory and only has to be re adjusted when the two mounting screws have been loosened The strobe disk is used to check all horizontal rotations measurement disk sample disk reagent disk sample arm horizontal reagent arm horizontal Check opto encoder position 1 Select drive to be checked in the service menu 2 Reset drive 3 Disable d
9. gt r EL U ol LI WATER RUNNING OUT after a time out of 25 sec BI INSUFFICIENT WATER 1 1 1 OUT Figure 2 9 Water level detection The svstem board controls the water pump When the upper float switch detects no water the water pump will be switched on for 25 seconds When after this time the float switch is still detecting no water the error message WATER RUNNING OUT will be displaved on the screen and the pump will be switched on again for 25 seconds When the lower float switch detects no water the error INSUFFICIENT WATER will be displaved When this message appears the instrument will not dispense sample or reagent anv more however itwill finish all the measurements in the cuvette rotor VITAL SCIENTIFIC 2 17 2 FUNCTIONAL DESIGN SERVICE MANUAL 2 4 4 OVERFLOW FOR CUVETTE The cuvette rotor is equipped with an overflow detection to detect if the liquid level in becomes too high The principle of the detection is based on conductivitv The principle is shown in figure 2 10 The resistance between the liquid sensor and the needles is measured as soon as the wash arm is going down If the liquid sensor touches the water in the low position of the wash arm the resistance is low and the svstem will report a WATER OVERFLOW MEASUREMENT DISC error
10. heat exchanger analyser compressor cooling unit The cooling liquid is based on Glycol The temperature of the cooling unit is set at the cooling unit itself and is not checked by the analyser Note that the temperature of the reagent is always higher than the set temperature The temperature setting of the cooling unit is protected so the user can not change the temperature Below the procedure for changing this temperature is described 2 22 VITAL SCIENTIFIC SERVICE MANUAL 2 FUNCTIONAL DESIGN When changing this temperature the situation in the lab should be taken into consideration For example a setting of 3 degrees might work well in an air conditioned lab but it could cause excessive condensation on tubes and heat exchanger in a warm lab with a high humiditv Changing of the temperature setting push and at a time for seconds now parameter mode select by means of the and U keys parameter P19 protection push set and keep it depressed and push or to set P19 to 0 protection is now off push and at a time for seconds return to normal operating mode fatina and keep it depressed and set the required temperature by means and push at a time for seconds now parameter mode select by means of the and keys parameter P19 protection push set and keep it depressed and push or to set P19 to 1 prote
11. 2 10 VITAL SCIENTIFIC SERVICE MANUAL 2 FUNCTIONAL DESIGN During the reset also the presence of the cuvette rotor is detected The wash arm will go downwards until it touches the bottom of the cuvette At this position opto 1 the vane is still positioned in opto 1 When no cuvette is present the is in opto 1 and error will be displayed Wash arm opto readings at the various wash arm positions High Up Down Detection of cuvette during reset Opto 0 The bellows pump consists of six separate bellows mounted together in one unit A stepper motor drives the bellows The bellows pump has two defined positions a zero position and a predefined full stroke volume position 500 ul is equal to 833 steps The detection of both positions is done bv opto switches VITAL SCIENTIFIC 2 11 2 FUNCTIONAL DESIGN SERVICE MANUAL 2 3 6 PIPETTOR The pipettor is a precise diluter with motor driven valves and svringes The pipettor is controlled bv one of the Motion Control Boards The liquid displacement is controlled bv two stepper motor driven svringes the sample svringe 100 ul and the reagent svringe 1000 ul Each svringe is connected to a special connection block that contains the valve unit The valves are driven bv small DC motors via a worm wheel construction The svringe drive assembly is shown in figure 2 5 The top and bottom position of the plunger is detected bv two opto switches
12. Chemical The operator is responsible for taking all necessary precautions against hazards associated with the use of clinical laboratory chemicals Specific recommendations for each reagent used with the analyser are normally found on the manufacturer s package inserts or on product information sheets for each chemical Wipe away any reagent spillage on the instrument immediately Biohazardous Materials As with all in vitro diagnostic equipment patient samples and serum based quality control QC products that are assayed on this system as well as all waste from the waste container should be treated as potentially biohazardous All materials and mechanical components associated with the sampling and waste system should be handled according to your facility s biohazard procedure Use the personal protective equipment recommended by your facility when handling any of these components Detailed recommendations Samples Treat all samples as potentially biohazardous and infectious If any sample is spilt on the instrument utilise the correct personal protective equipment PPE gloves lab coat etc wipe it away immediately and clean the contaminated surface with a disinfectant Fibron 1 Manual Vital Scientific NV Service manual Rev 1 0 Page 3 77 vital scientific Waste Solution and Solid Wastes Avoid direct contact with waste solution and or solid waste Both should be handled as potentially biohazardous Dispose of waste solut
13. REF 01 4 3333 518 ULN2013A IC 11 3342 311 SWITCH SLIDE 0 161390 1 SW 1 2 3348 191 FUSE 3 15A SLOW F 1 3 3350 006 CONN FAST 2 8X0 5 1 16 3350 137 CONN 2POL 280609 2 3350 146 8136 650 2 42 3351 216 CONN 280610 2 PL2 3351 235 CONN 4POL 826467 4 PL4 3352 011 CONN 28061 1 2 PL 1 3352 014 CONN 6 POLE 826467 6 PL 5 7 3355 002 KEYING PLUG 926498 1 PL 4 7 3359 002 POWER UNIT DC CONVERTOR IC 3 3362 622 COIL ZKB 422 061 L 1 2 3366 006 FUSE HOLDER PCB MOUNT FOR TR5 F 1 3 3370 109 MOUNT SET TO 220 T 1 5 3376 452 TEMPERATURE BOARD 3381 071 OPTOCOUPLER MCT 6 OC 1 3 3820 115 STICKER 4502 353 BEAM A3 5 x gt 3 5 CJ E ta H M min 2 4 s VITAL SCIENTIFIC 8 ELECTRICAL DIAGRAMS SERVICE MANUAL 3100 48301730 eb 9IJILIN3IOS 18110 sita 4 1244 1889 108L1NOJ dHIL 1141 5530 SSINHDNOM 1N39U34 SNAS 4 3 2944403 n a iN39834 ten 1N3JIBHU bena D 5 On 99 DI 514 2 2 34405 1081802 35010834031 3510 135430 0 1N34313 NOI1UN0d0n3 8311134 820 6 53713 Wanod dit 0 vati 0 VITAL SCIENTIFIC 2 5 342053114 3114 vee Garda b
14. SONGUT83H1JN Et 2EL N3831 13395 2131193125 Waira UNIBIBRI C 652 1009 3403 Nd9 pue uaiseu TE TE on NOLL4182530 m SSINHONOM SNOISN3UIO FI NE 031015 553780 32NU37101 BAQB 10 424 UNAS 924 725099102529 05182724 48518 624 KBINASILId KBAQN LM 02d 201 22 B 29317312 01 22 981 413 VITAL SCIENTIFIC SERVICE MANUAL 8 ELECTRICAL DIAGRAMS SHEET 3 4 130 6001 Mastercard SIO module VITAL SGIENTIFILS Racon 5 10178 4 VITAL SCIENTIFIC SERVICE MANUAL 8 ELECTRICAL DIAGRAMS jj D D O N MODIFICATION TOLERANCES UNLESS OTHERUISE STATED DIMENSIONS DELIVERY DATE NO 6001 759 DESCRIPTION Mastercard PIO module VITAL SCIENTIFIC DIEREN THE NETHERLANDS UITRL SCIENTIFIC VITAL SCIENTIFIC SERVICE MANUAL PHOTOMETER BOARD 6001 761 PART NUMBER DESCRIPTION 3002 016 SCREW ST ST M3X6 3019 009 NUT M3 HEIGHT 15 mm 3023 006 LOCKWASHER J 3 2 3302 301 RES METAL 100 OHM 3302 305 RES METAL 110 OHM 3302 365 RES METAL 470 OHM 3302 417 RES METAL 1K5 OHM 3302 433 RES METAL 2K2 OHM 3302 445 RES METAL 3K OHM 3302 501 RES METAL 10K OHM 3302 517 RES METAL 15K OHM 3302 521 RES METAL 16K OHM 3302 525 RES METAL 18K OHM 3302 541 RES METAL 27K OHM 3302 565 RES METAL 47K OHM 3302 633 RES METAL 220K OHM 3305 249 RES
15. VITAL SCIENTIFIC SERVICE MANUAL 9 MECHANICAL DRAWINGS MAIN ELECTRICAL UNITS VITAL SCIENTIFIC 9 19 9 MECHANICAL DRAWINGS MECHANICAL ASSEMBLY 6001 752 sheet 1 PART NUMBER DESCRIPTION LOCATION 3002 016 SCREW ST ST M3X6 1 3002 333 SCREW ST ST M4X12 8 3002 336 SCREW ST ST M4X16 9 3002 388 SCREW RVS M6X25 10 3002 621 SCREW ST ST M3X16 11 3004 211 SCREW ST ST M3X8 12 3008 024 SCREW ST ST M3X5 13 3008 040 SCREW ST ST M4X5 14 3008 043 SCREW ST ST M4x12 15 3009 016 SCREW M8x50 16 3011 011 NUT ST ST M3 17 3011 019 NUT M4X27 5MM 18 3019 021 NUT M4 19 3022 015 WASHER ST ST 4 3 21 3022 025 WASHER ST ST 6 4 22 3023 010 WASH LOCK ST ST M4 25 3023 042 SHIMRING 26 3023 035 SPRING WASHER 27 3029 017 O RING 12x1 5 28 3029 051 V SEAL RING 29 3030 045 DOWEL PIN 6M6X20 30 3030 051 DOWEL PIN 04M6X16 31 3053 001 BEARINGBALL 6X19X6 32 3064 017 TIMING BELT 132T T2 5 33 3070 038 TUBE CLAMP 34 3105 005 ADJ RING 06 8 12 8 35 3105 006 ADJUSTING RING 36 3105 007 ADJUSTING RING 37 3106 113 BEARINGBUSH 38 3350 050 CONN IP HDP 20PIN 39 3350 107 CONN 1PM 40 3350 149 FASTON TAB 41 3352 019 CONN HOUSING 6P 42 3370 444 TIE ANCHOR MOUNT NYLON 44 3381 053 OPTO INTERRUPTOR H21B1 45 4102 319 SHAFT 46 3019 049 STUD 47 4102 329 SHAFT 48 4102 331 SHAFT 49 4102 336 STUD 50 4105 186 BUSH 51 4105 187 BUSH 52 4105 196 BUSH 53 4105 197 BUSH 54 4106 227 RING 55 4106 268 RING 56 4501 461 BLOCK 58 4502 357 STRIP 59 4502 358 STRIP 60
16. R 14 28 R 36 37 39 R 32 R 6 9 22 23 R 34 R 2 13 18 26 R 3 4 8 19 20 21 R5 R15 16 30 31 1 12 17 25 RN 1 4 C 29 37 42 46 47 54 C9 C 7 13 23 28 C 5 11 14 16 18 26 31 33 C 1 4 6 8 10 12 15 17 19 20 22 30 32 33 45 48 53 5 27 C 24 25 C 36 35 C 34 1 4 9 16 21 24 25 26 1 4 IC 1 7 9 13 IC 3 IC 8 IC 11 IC 4 IC 15 IC 5 IC 2 10 IC 12 IC 14 SW 1 SW 2 SW 4 SW 3 F 1 3 TAB 1 8 PL 1 2 PL3 PL 4 5 PL 6 8 IC 5 6 IC 4 15 IC 12 PL 1 2 4 8 F 1 3 D 5 8 17 20 X1 A3 8 ELECTRICAL DIAGRAMS SERVICE MANUAL A Ud EU JA UU AUY UI 0 SSS JU LENJJMPUJJJE D3A3852J 5140 NY WO VON3A11 3LVO E d SONVIBGHLIN JHL DHLINADS 6 Vivi Ee 05 Diva BINOJ 03 08 1081402 ume 000 00790 0 n ta 80 13 da BCH E O Cal SES 20 t m 90 002 19 Ua DaSJy 05 110 ua 8 50 29 19 epfiz VITAL SCIENTIFIC 8 ELECTRICAL DIAGRAMS SERVICE MANUAL 13 21 4 1 25 18110 a 95 2158 eet v 4335 1838814 C P al 8S2 R II TOYLNOD NOILOW 114182530 OR esa mm 31088 5 015 10 021915 3SIN IHLO SSJINN 32N0N3701 za HIS 2 882
17. SERVICE MANUAL 2 FUNCTIONAL DESIGN The table below shows the timing in the dual mode for each individual cuvette Action Time after Time after sample addition Reagent 2 addition Washing ofthe 00 00 0024 K 10 KO C i 21 Endpoint measurement Bichromatic VITAL SCIENTIFIC 2 7 2 FUNCTIONAL DESIGN SERVICE MANUAL 2 3 MECHANICAL DESIGN 2 3 11 REAGENT ARM The reagent arm is responsible for the moving of the reagent needle to the correct position The construction of the arm is such that the driving for the movements is decoupled separate stepper motors drive the up down movement and the sideward rotating movement The reagent arm has in fact seven discrete horizontal positions in which the arm normally is positioned A reset opto and a subsequent number of stepper motor steps determine these positions Horizontal Cuvette position Wash position 5 Reagent Rotor position There are three discrete vertical positions and one variable vertical position Vertical Up or reset position Cuvette position Wash position Reagent variable value depending on the liquid levelin the reagent bottle For the discrete positions a value will be assigned with respect to the reset position which will comply with the number of steps the stepper motor needs for the arm to reach its destination The variable position for the reagent level indicates the arm to lower itself just below the reagent
18. This error occurs when the opto reading of the horizontal movement of the sample arm is not in accordance with the current theoretical position This normally happens when the sample arm is obstructed by something or when the safety switch inside the sample needle is activated Perform a selective reset of the sample arm Check when this error is occurring frequently the operation of the sample needle switch and clean if necessary the contacts of the switch only systems with mechanical needle switch Check the operation of the opto couplers and replace if necessary VITAL SCIENTIFIC SERVICE MANUAL 7 ERROR HANDLING E63 SAMPLE ARM VERTICAL ERROR Cause Solution This error occurs when the opto reading of the vertical movement of the sample arm is not in accordance with the current theoretical position This normallv happens when the sample arm is obstructed bv something or when the safetv switch inside the sample needle is activated Perform a selective reset of the sample arm Check when this error is occurring frequently the operation of the sample needle switch and clean if necessarv the contacts of the switch only mechanical needle switch Check the operation of the opto couplers and replace if necessarv E72 SAMPLE ARM INIT FAILED see E75 E75 SAMPLE ARM HORIZ RESET FAILED Cause Solution This error will be generated when the motion control board does not report the horizontal reset position after the
19. 3002 034 SCREW ST CAD M4X45 5 3004 210 SCREW ST ST M3X6 6 3004 230 SCREW ST ST M4X20 7 3008 040 SCREW ST M4X5 8 3011 004 NUT ST ST M2 9 3020 017 CIRCLIP 4 10 3022 005 WASHER ST ST 2 2 11 3022 010 WASHER ST ST 3 2 12 3022 015 WASHER ST ST 4 3 13 3066 088 SOCKET AD1 14 3070 038 TUBE CLAMP 15 3130 082 GEAR 1 Z 60 16 3350 004 CONN HDP 20 CONTACT 17 3350 008 CONN SOLDER PIN CONTACT 18 3350 050 CONN IP HDP 20 PIN 19 3353 183 CONN 37P HDP 20 PLUG H 20 3363 505 STEPPER MOTOR 21 3366 120 SOLDERTAG 22 3370 004 SLEEVE CABLE 23 3370 006 SLEEVE CABLE 24 3370 401 CLIP CABLE NXO 25 3370 410 CLIP CABLE NX3 26 3370 430 CLIPCABLE TIE 27 3370 525 SCREWLOCKS FEMALE 28 3372 003 WIRE 0 15 GREY 29 3372 007 WIRE 0 15 VIOLET 30 3390 013 PELTIER ELEMENT 31 3750 036 SPRING 32 3820 255 STICKER EXCLAMATION MARK 33 4100 461 WASTE CUP SAMPLE 34 4100 557 WASTE CUP REAG 35 4102 330 SHAFT 36 4104 085 PIPE 37 4104 089 PIPE 38 4105 225 BUSH 39 4106 226 RING 40 4107 072 SCREW 41 4107 085 SCREW 42 4509 171 PLATE 43 4509 172 PLATE 44 4509 176 HEATSINK BLOCK 45 4801 100 PLATE 46 4803 381 BRACKET 47 6001 430 ENCODER STEPMOTOR UNIT 48 6001 757 MAIN ASSEMBLV 49 6001 817 MEASURING OPTO UNIT CABLE 50 6001 930 ENCODER OPTO UNIT ROTORS 51 9 34 SERVICE MANUAL VITAL SCIENTIFIC SERVICE MANUAL 9 MECHANICAL DRAWINGS MEASURING UNIT VITAL SCIENTIFIC 9 35 9 MECHANICAL DRAWINGS MAIN ASSEMBLV MEASURING UNIT 6001 757
20. 3066 072 pipettor valve Note that the maintenance requirements depend on the workload of the instrument and the environmental conditions as temperature and humidity Individual instruments might need maintenance more or less often than stated here 4 8 VITAL SCIENTIFIC SERVICE MANUAL 4 INSTALLATION AND MAINTENANCE 4 5 2 Maintenance steps Time required approximatelv 8 hours excluding customer acceptance tests The scheduled maintenance includes the following steps and must be performed in the order as stated here and in accordance with the parts list as stated before Svstem cleaning procedure Replacement of reagent and sample probe tubing Replacement of ISE sample tubing oniv for instruments with an ISE unit connected Revision of vacuum pump Revision of water pump Replacement of mixer belts Checking the pipettor replacing pipettor valves Replacement of svringes Replacement of water filter 10 Replacement of wash block 11 Checking the cuvette rotor adjustment 12 Checking the lamp adjustment replacing the lamp 13 Checking the photometer adjustment 14 Checking the reagent needle and cuvette temperature 15 Checking the wash arm adjustment 16 Checking the bellows aspirate volume 17 Checking reagent and sample arm adjustment 18 Checking the needle switch 19 Checking the mixer position 20 Checking reagent and sample rotor adjustment 21 Checking filter wheel adjustment 22 Checking the opto encoder adjus
21. 5 85965 A 6 N AN N IERI Al LS 8 Qu an 3 cei 819 01 409 BABJS 9 UONEJIUNWWOS elles drivers for bellows valves vac valves pumpunit Figure 3 3 Block diagram Svstem Board VITAL SCIENTIFIC 3 4 SERVICE MANUAL 3 3 MOTION CONTROL BOARD In figure 3 4 is the block diagram given of the motion control board sio PIO XTAL CLOCK 280 MREQ 3 SYSTEM ORGANIZATION CTC baudrate 810 RS 232 ADRESS DECODER PORT ADRESS DECODER gt RD gt WR gt IOR gt TOW PROM 32K RAM 8K reset half full A stepclk B motor A CONTR c 4 enable D home PIO1 input input reset half full A stepclk B motor CONTR A B enable D home input input output output output output input input input input PIO 2 MOTION CONTROL BOARD Service mode switch Figure 3 4 Motion Control Board
22. Check if all voltages are present the power supplv and repair or exchange if necessarv E121 WASHARM BELLOWS PUMP 30V FAILED Cause This error will be generated when the circuit see sec 8 of this manual motion control board sheet 1 4 diode 026 does not detect the correct voltage Solution Check the fuses and voltage on the motion control board Check connector PL3 and PI4 Check if all voltages are present on the power supplv and repair or exchange if necessarv Check wash unit stepper motor and bellows pump stepper motor E122 WASH ARM ERROR Cause This error will be generated when the opto reading is not in accordance with calculated software position The cause can be that the motor is not operating due to mechanical obstruction electrical malfunctioning bad connection or because of malfunctioning of the opto s Solution Check if the washarm is fixed properly Check if the wash arm can move smoothly and repair if necessary Check if the electrical connections are correct and repair if necessary Check if the opto coupler functions are working and repair if necessary E123 BELLOWS PUMP ERROR Cause This error will be generated when the opto reading is not in accordance with calculated software position The cause can be that the motor is not operating due to mechanical obstruction electrical malfunctioning bad connection or because of malfunctioning of the opto s Note that this error can also be generated
23. F3 Valve V of the ISE unit will be activated for 2 seconds F4 Valve S of the ISE unit will be activated for 2 seconds F5 peristaltic pump will be switched on for 2 seconds F6 ISE unit will send the arm down command the ISE arm control status will change The arm itself will not go down F7 The measured voltage of the ISE electrodes for Na K and CI will be displayed for standard A When pressing Shift F7 the same will be displayed for standard B F8 The measured voltage of the sample sensor will be displayed For really sending the ISE arm down you must use the arrow down key The arm will go down for 5 seconds For more detailed ISE test facilities we refer to the ISE service manual where the internal test diagnostics are explained VITAL SCIENTIFIC 4 25 4 INSTALLATION AND MAINTENANCE SERVICE MANUAL 4 6 8 Pipettor Analyser status Inactive Req tray buffer 1 18 Run mode DUAL External 88 232 ON LINE Display mode Funct Check Adj Adjust Lamp Reagent Reagent valve Mater p Sample Arw Reagent syringe 600 Reagent Disc Sample valve Needle p Sample Disc Sample syringe 8 Measurement Disc Filter Wash Arm Reagent valve opto ISE Arm ISE Unit Reagent valve opto 1 Pipettor Reagent sur res opto Vacuum system Reagent syr down apta Water system Sample valve opto B Optical electronics Sample valve opto 1 Electronics Sample sur res opto Barcode scanner Sampl
24. FUNCTIONAL DESIGN As a result of this cvcle after 48 cvcles all cuvettes have been washed pipetted and measured In the table vou will find the order of actions for the individual cuvettes Action Time sec min sec Washing of the cuvette 120 to 70 s Reagent dispensing espe dee s BJ HEN a pr j L I EB Endpoint measurement Bichromatic 1699 11 30 VITAL SCIENTIFIC 2 5 2 FUNCTIONAL DESIGN SERVICE MANUAL 2 2 2 DUAL MODE CVCLE The cycle time of the Dual Mode is 27 seconds therefore the theoretical throughput of the instrument is 133 tests hour In the figure below the sequence of the 27 seconds cycle is shown Ef E2 Endpoint dichromatic after 4 5 minutes E4 Endpoint dichromatic R1 R2 Reagent Blank dichromatic Figure 2 3 The 27 second cycle Because in the Dual Mode use can be made of a second or starter reagent in this cycle there is time for the second reagent to be dispensed and mixed Instead of dispensing a second reagent these extra 7 seconds can also be used for predilution of a sample In case of predilution in one cycle the sample is added to the diluent and in the next cycle in the extra 7 seconds the diluted sample is picked up by the sample needle and dispensed in the next cuvette which already contains the appropriate reagent 2 6 VITAL SCIENTIFIC
25. LIQUID SENSOR CUVETTE ROTOR WASH ARM 1 Figure 2 10 Overflow detection 2 18 VITAL SCIENTIFIC SERVICE MANUAL 2 FUNCTIONAL DESIGN 2 5 OPTICAL SYSTEM Figure 2 11 shows the optical system of the analyser Cuvet Beamspliffer Interference Filter Diaphragm Lens Lamp 4 Detector Lens Filferwheel Reference Detector Figure 2 11 Optical system The light source consists of a long life quartz iodine lamp 12V 20W 2000 hour life time and the detector is made of a UV enhanced silicon photocell Monochromatic light is achieved by using single IFL InterFerence Line filters The filters are mounted in a stepper motor driven 8 position filter wheel Depending on the programmed wavelength the microprocessor will automatically select the correct filter and the filter wheel will move to the correct position The used IFL filters have an integrated coloured glass filter depending on the wavelength of the IFL filter two types are used For the low wavelength range a filter is used to reduce the effects from strav light bv blocking light of higher wavelength For the high wavelength range a filter is used to level difference in energv at the high ends of the spectral range This will be necessarv because the spectral emission of the quartz iodin
26. SERVICE MANUAL 3 SYSTEM ORGANIZATION 3 5 2 1 INPUT AMPLIFIER The input amplifier circuit is built around which is a voltage OP AMP 7 The purpose of this amplifier is to amplifv the difference at the inputs to an adequate level The amplification is approximatelv 70x The non inverting input is derived from the output 0 408 V at 37 To extend the dvnamic range of the amplifier the inverting input of IC7 is derived from a reference voltage of 408 8 mV 3 5 2 2 DIFFERENTIAL AMPLIFIER The second stage of the temperature control consists of differential amplifier IC6c The output of IC7 that corresponds to the actual temperature is present at the inverting input of IC6c while the other input is connected to the reference voltage This reference voltage can be adjusted and is derived from the 10 volts reference source IC6c will amplify the difference between the inputs 3 5 2 3 DRIVING CIRCUIT The output signal from IC10a is going via a amplification circuit IC6d to the switched power supply circuit The System board can monitor if the temperature is within 0 1 C by means of the READY signal RDY If the temperature is not reached within 2 5 minutes the microprocessor will switch off the heating element and will generate the REAGENT NEEDLE TEMPERATURE ERROR 3 5 3 LAMP VOLTAGE DRIVER The lamp stabilizer circuit is built around IC5 a voltage regulator One of the lamp leads is always connected to the v
27. TZ 1 4 6 TZ 2 3 C 1 5 7 9 13 16 C 6 14 C3 11 C2 10 4 8 12 16 1 2 IC 2 4 IC 1 3 1 3 PL 2 5 PL 1 3 PL 4 PL 2 4 5 Comp ident comp side 14 aC 1 VITAL SCIENTIFIC Motion Ctrl EJ TAB3 TZ6 e c 8 ELECTRICAL DIAGRAMS SERVICE MANUAL SONGTN3MI3N DEL 13345 Jj I 4 I N a I x S 78 fl I 3141443125 Wain 55 26 62 zwa AXNIN388IN C ar ik 94 in uo 32832ap anal 114 t 6 8e cz n 87390 ast Vasi NZnee 114182530 7N A A i Eld 3140 43 1730 pug 43003 duo per U 9 J RE man D severi i 031915 JS1053M10 55370 S33N553201 834 4 330 lL ese jeu s co P438 J U0J u BA 82 5 1 ZE 2 7 570 sen 4 D 3o do aidees 5 do z o D O aaan 231003 BS D 2 weer vers ws ma 4 wen a 2222 Z tu B E 57330 1 x 2 T SERVICE MANUAL 8 ELECTRICAL DIAGRAMS VITAL SCIENTIFIC SERVICE MANUAL VALVE BOARD 6001 768 PART NUMBER 3002 016 3011 011 3302 120 3302 180 3315 037 3320 100 3320 150 3320 232 3320 311 3320 312 3326 521 3330 045 3331 416 3333 331 3333 545 3335 004 3348 191 3350 136 3351 217 3352 012 3353 057 3355 002 3366 006 3376 432 DESCRIPTION SCRE
28. The reagent arm must be able to move in two directions and is therefore driven by two stepper motors The reagent arm is equipped with a level detector to check if sufficient reagent is present It is also equipped with a controlled heating element to preheat the cooled reagent 2 1 6 ISE ARM Optionally the instrument is equipped with an ISE arm to connect the instrument to an optional external ISE lon Selective Electrode measuring unit This external ISE unit is controlled by the instrument and the ISE arm aspirates sample from the cuvette rotor The ISE arm is integrated in the mixer unit 2 1 7 PIPETTORS The two Hamilton svringes a 1000 ul and a 100 ul tvpe are used in combination with two valves for reagent and sample pipetting The pipetting svstem is water filled with air bubble separation A stepper motor drives each svringe and the detection of the svringe up and down positions is done bv opto switches 2 2 VITAL SCIENTIFIC SERVICE MANUAL 2 FUNCTIONAL DESIGN 2 1 8 CUVETTE ROTOR The cuvette rotor contains 48 cuvettes which are thermostatted at 37 C The path length of a cuvette is about 7mm but absorbance values are re calculated to 10 mm path length A stepper motor drives the cuvette rotor The heating of the cuvette rotor is done by means of two controlled Peltier elements The cover of the cuvette rotor is equipped with a heating element to avoid condensation at the inside of this cover 2 1 9 WASHING UN
29. the wiring the SIO on the system board the KIO on the motion control board or the main PCB fuse on the motion control board Replace the board s or replace the defective components Check if there is a mechanical obstruction of the reagent arm and repair if necessary Check if the opto couplers are working correctly and repair if necessary E96 REAGENT ARM VERTICAL RESET FAILED Cause Solution This error will be generated when the motion control board does not report the vertical reset position after the reset command Check if the communication between system board and motion control board is present When there is no communication this can either be caused by the wiring the SIO on the system board the KIO on the motion control board or the main PCB fuse on the motion control board Replace the board s or replace the defective components Check if there is a mechanical obstruction of the reagent arm and repair if necessary Check if the opto couplers are working correctly and repair if necessary VITAL SCIENTIFIC 7 15 7 ERROR HANDLING SERVICE MANUAL E97 REAGENT ARM COMMUNICATION ERROR Cause Solution This error will be generated when the communication to motion control board for the reagent arm stalls or if an unknown character is echoed Normally this error will occur when the communication is disrupted distorted Check if the communication between svstem board and motion control board is present W
30. 2 X1 8 ELECTRICAL DIAGRAMS SERVICE MANUAL 65 009 HWOLYDHIAJ ILVO AYIANIO SONVINIHLIN 16 6 ivi ES 0308 WILSAS ONtvN38 14823530 ONNZONV 38 WINVW 19Nv AM 532 1634 E 2 GT3WAJA SH30NY FIZNIL SILNVUJIOL QILVLS JSIMHIHLO 5531 SIINVYIFNOL I 928096 870 0701 915 bid La P M LY 53 19 Oe mu m 15 en d m kud sg J BJ DI 82 m O CN Qa a Ca sha 50 ZSC 9 CC VITAL SCIENTIFIC 8 ELECTRICAL DIAGRAMS SERVICE MANUAL D Santen unnoen mm o je SNO log zp p 8v2 eb dO SONBINIHLIN N38310 3141183125 Wita L US z ess z es 266 9801 d 096 ner oe ve U S nat 9 L 22 2 u ui I oi uL bia ut ur ui vy 692 Sv2 8802 aneis D yh 562 662 HIS y 6S n ma 5 xnu i 5e9 varen 9 pum 2 REP J 5 af pl OZN T t 330 330 962 2 93 BI 492 bi ET 3 9u03 44135 229815 351 yiun 3SI 3 VITAL SCIENTIFIC 8 ELECTRICAL DIAGRAMS SERVICE MANUAL 3190 A 3N1730
31. 20V 3326 554 CAP ELC 47uF 63V 3330 304 DIODE ZENER ZPD 3 6V 3333 532 MA 232 CPE 5V 3335 004 IC 74 04 3335 008 IC 74 HCT 08 3335 032 IC 74 HCT 32 3335 074 IC 74 HCT 74 3335 138 IC 74 HCT 138 3335 161 IC 74 161 3336 008 284 00 280 CPU 3336 009 284 20 4 PIO 3336 010 284 C30 3336 051 32K X 8 STATIC CMOS RAM 3336 060 IC 280 3336 063 MICRO MONITOR CHIP DS1232 3336 074 280 SER IN OUTPUT CONTR 3342 217 SWITCH PUSHBUTTON 3342 311 SWITCH SLIDE 0 161390 1 3348 191 FUSE 3 15A SLOW 3350 006 CONN FAST 2 8X0 5 3351 235 CONN 4POL 826467 4 3352 011 CONN 28061 1 2 3352 014 CONN 6 POLE 826467 6 3353 056 CONN 2X6 POL 826469 6 3353 127 CONN 16PM 1800 826469 8 3353 340 CONN PCB DIL 28PF 3353 342 CONN 32PF IC SOCKET 3353 344 SOLDER TAIL SOCKET 44P 3353 345 CONN PCB DIL 40 3353 384 CONN 84P PLCC 0 0821573 1 3355 002 KEYING PLUG 926498 1 3366 006 FUSE HOLDER PCB MOUNT FOR TR5 3376 352 SYSTEM PC BOARD 3390 027 CRYSTAL 7 3728 Mc HC 18 3820 115 STICKER VITAL SCIENTIFIC 8 ELECTRICAL DIAGRAMS LOCATION R3 R 9 11 R5 R 2 4 R 7 8 R1 RN 1 3 C 6 16 31 32 35 62 C 1 3 5 7 13 17 20 23 33 C 14 15 C2 C 21 22 24 25 29 34 C 30 D 1 IC 18 19 IC 10 IC 2 IC 9 12 IC 1 IC 3 11 IC 8 IC 5 IC 14 15 IC 4 IC 6 IC 16 IC 20 IC 13 17 21 SW 1 SW 2 F 1 2 TAB 1 7 PL 6 7 PLS PL 1 8 11 PL9 10 12 13 PL 2 4 IC 4 6 IC 7 IC 13 17 21 IC 5 14 15 IC 16 PL 1 4 6 13 F 1
32. 4802 331 PLATE 61 4802 245 PLATE 62 4803 381 BRACKET 63 4803 383 BRACKET 64 4803 457 TOPPLATE 65 4803 458 BASEPLATE 66 6001 888 ARM RESET BOARD 67 6001 882 WASH ARM BOARD 69 6001 420 ADJUST BLOCK ASSY 70 6001 430 ENCODER STEPMOTOR UNIT 71 6001 756 MEASURING UNIT 72 R amp S STIRRING ISE UNIT 73 6001 445 ROTOR UNIT 74 6001 930 ENCODER OPTO UNIT ROTORS 75 6001 448 CARRIAGE ASSY 76 6001 449 CLAMP ASSY 77 6001 931 ENCODER OPTO UNITS ARM 78 6001 455 UP DOWN STEPMOTOR UNIT 79 6001 456 SWING UNIT 80 6001 811 MOVEMENT ASSY 81 9 20 SERVICE MANUAL VITAL SCIENTIFIC SERVICE MANUAL 9 MECHANICAL DRAWINGS MECHANICAL ASSEMBLY SHEET 1 VITAL SCIENTIFIC 9 21 9 MECHANICAL DRAWINGS MECHANICAL ASSEMBLY 6001 752 sheet 2 PART NUMBER DESCRIPTION LOCATION 3002 016 SCREW ST ST M3X6 1 3002 030 SCREW ST ST M4X6 6 3002 333 SCREW ST ST M4X12 8 3002 336 SCREW ST ST M4X16 9 3002 388 SCREW RVS M6X25 10 3002 621 SCREW ST ST M3X16 11 3004 211 SCREW ST ST M3X8 12 3008 024 SCREW ST ST M3X5 13 3008 040 SCREW ST ST M4X5 14 3008 043 SCREW ST ST M4x12 15 3009 016 SCREW M8x50 16 3011 011 NUT ST ST M3 17 3011 019 NUT M4X27 5MM 18 3019 021 NUT M4 19 3022 015 WASHER ST ST 4 3 21 3022 032 SHIM RING PS8X10X0 15 23 3023 006 LOCKWASHER J 3 2 24 3029 017 O RING 12x1 5 28 3029 051 V SEAL RING 29 3030 045 DOWEL PIN 6M6X20 30 3030 051 DOWEL PIN 04M6X16 31 3053 001 BEARINGBALL 6X19X6 32 3064 017 TIMING BELT 132 2 5 33 3070 038 TUBE CLAMP
33. 81802 4315 02 4315 nace 220 pca 3 ave S sza cca 2 373 c k 123 153 553 8 2 6 9 252 862 8 330 HIS 2 2 852 3 42 6 445 L 2 9 4 9 1 14 33a 0 fi idns sancg neero VITAL SCIENTIFIC SERVICE MANUAL 8 ELECTRICAL DIAGRAMS VITAL SCIENTIFIC SERVICE MANUAL 8 ELECTRICAL DIAGRAMS iii VITAL SCIENTIFIC 8 ELECTRICAL DIAGRAMS SERVICE MANUAL SON6 838138 N38310 21411N3125 181101 LAW gt 330 b 2 Kiso A 66628 72 as2 Sta Kosa INAS gt 3u0H 13535 amp 11047 30H 22 amp n22 n3 VITAL SCIENTIFIC 8 ELECTRICAL DIAGRAMS SERVICE MANUAL SONBINIHLIN N38310 2131193195 Wir lt 250 INAS gt 3u0H lt 13535 K 11047 4068 13 amp n227n2 VITAL SCIENTIFIC SERVICE MANUAL SVSTEM BOARD 6001 759 PART NUMBER DESCRIPTION 3302 301 RES METAL 100 OHM 3302 465 RES METAL 4K7 OHM 3302 469 RES METAL 5K1 OHM 3302 501 RES METAL 10K OHM 3302 549 RES METAL 33K OHM 3302 601 RES METAL 100K OHM 3315 009 RES N W 4X4K7 3320 110 CAP CER AXIAL 1 nF 50V 3320 310 CAP CER AXIAL 100nF 50V 3320 942 CAP CERAMIC 33pF 100V 3326 410 CAP ECLEC TANTAL 1uF 35V 3326 521 CAP ELC 15uF
34. 9 i 5 9 4 L 8 2 8 L 6 2 5 3 08 7041802 NOLLOW 8V a 85 1009 2510 31dHVS VITAL SCIENTIFIC 8 ELECTRICAL DIAGRAMS SERVICE MANUAL SONVINGHLIN 6 Dr 04 409 ZVYLIITIS 08 08 HILSASAY 65L 1009 WALVONIAJI WWHOS ABJANIO ANSHA1V3HL 3002 ww WINYY Wied va ocu ALLENVIND ON 93050 180 LWA Pu SSINKDNOW Sve IZN31 S3UNVHTIOL 15 3SIMY3HLO 5531 SIINVUJIOL 708 08 7081802 NOLLOW AW 708 08 104102 NOLLOR CV Way HSVM 954 4009 9 2 PQA PJDOQ JAJDA SA0188 VITAL SCIENTIFIC 8 ELECTRICAL DIAGRAMS SERVICE MANUAL 4103135 di 530 3330 NIHIMNV AS FILNIOS_IVLIA 9N 30NvH38 3 53387438 3003 Wil 553008 113 483 SH30NV FIZN31 531 3101 031 15 3SIMHIHLO ie 95 4009 VITAL SCIENTIFIC 8 ELECTRICAL DIAGRAMS SERVICE MANU
35. CUVETTE ROTOR In the active state the vacuum pump is always running When all valves are closed de activated this will cause a vacuum in the collecting cask In order to empty the cuvettes valve V2 will open and the wash arm needle will aspirate the cuvette contents into the collecting cask Optionally a separate parallel concentrated waste circuit will aspirate the contents of the first cuvette into a concentrated waste collecting cask 2 6 3 EMPTYING THE WASH POSITIONS Valve V2 is closed and valve V4 and V5 are activated When V2 will be opened the waste of the reagent wash position will be sucked into the collecting cask When V5 is de activated the waste of the sample wash position is sucked into the collecting cask and the collecting cask will be emptied 2 6 4 EMPTYING THE COLLECTING CASK Valve V2 is closed and valve V3 is opened Pressure will be put on the collecting cask by activating V1 and V12 and the collecting cask will be emptied and collected into the waste container Valve V1 V3 and V12 are de activated again VITAL SCIENTIFIC 2 21 2 FUNCTIONAL DESIGN SERVICE MANUAL 2 7 REAGENT COOLING Reagent is cooled by means of an external cooling unit A cooling liquid is cooled down to a temperature set at the cooling unit This cooling liquid is then pumped through a heat exchanger in the analyser from where cooled air is pumped through the reagent compartment NER liquid
36. Each motion control board controls two stepper motors They receive the signals from reset opto s decoder discs and needle switches to check the position Note that there are six identical motion control boards in the instrument with each a different function therefore the EPROM version and the place in the instrument determines the function of the board VITAL SCIENTIFIC 3 5 3 SYSTEM ORGANIZATION SERVICE MANUAL 3 3 1 STEPPER MOTOR CIRCUIT Figure 3 5 shows the block diagram with the stepper motor circuit data Stepper motor B 2 address PIO 14 gt mea 280 KIO L297 _ Micro Stepper Drivers processor motor controller Figure 3 5 Stepper motor circuit The two stepper motors and their control circuits are identical The first two blocks are the microprocessor and the PIO that are part of the I O array In principle there are only four lines necessary to control the stepper motor device L297 This device has 8 output lines that control the stepper motor drivers L6203 On the output lines of the drivers fast ramping rectifiers can be found to protect the drivers against high induction peaks There are also LED s to indicate if the output of the driver is activated When the stepper motors are normally running it seems that all LED s are on because the switching time i
37. F4 Fo F7 13 Reset Disable ashArm Drive Fi F2 r3 First press reset F1 to activate the wash arm function and the wash arm position can be controlled with the cursor keys The following active states are displayed Arm opto switch high position Arm opto switch up position Arm opto switch down position Overflow detection of the cuvette rotor No cuvette rotor detection With function key F3 can the wash arm motor be disabled Refer to paragraph 2 3 5 for the function of the various opto s 4 24 VITAL SCIENTIFIC SERVICE MANUAL 4 6 7 ISE arm ISE unit Analyser status Inactive Run mode DUAL Adjust Lamp Reagent Arm Sample Arw Reagent Disc Sample Disc Measurement Disc Filter Mash Arm TSE Avm ISE Unit Pipettor Vacuum system Water system Optical electronics Electronics Barcode scanner Taba 4 INSTALLATION AND MAINTENANCE Req tray buffer 1 18 External 88 232 ON LINE Display mode Funct Check Adj Up Note ISE arm can stay max Doun l 5 seconds down ISE arm opto down ISE arm control up alve AMValve Pump 29 0penfi2s Openfi2s Openfils On BZ FI F2 M T5 Fo Service On Sensor WZA F8 9 The following test functions are selectable Fl Valve A of the ISE unit will be activated for 2 seconds F2 Valve B of the ISE unit will be activated for 2 seconds
38. MECHANICAL ADJUSTMENTS SERVICE MANUAL A o 2 Figure 5 7 Bellows pump After adjustment of the pump volume vou have to check if the aspirated volume is correct The best wav is to check this during a wash cvcle When the wash arm is down and the cuvette is filled the shortest needle C must be approximatelv 1 mm below the water level When checking this be sure the wash unit adjustment is OKI 3 Figure 5 8 Checking the pump volume 5 8 VITAL SCIENTIFIC SERVICE MANUAL 5 MECHANICAL ADJUSTMENTS 5 6 SAMPLE ARM AND REAGENT ARM ADJUSTMENT For all arm adjustments the covers of the arms must be removed Small horizontal arm adjustment 1 In the service menu reset the arm and move the arm to the wash position horizontally and vertically 2 Loosen the larger conic screw on top of the arm and loosen the smaller screw 3 Adjust the arm in a way that the needle is in the centre of the wash position 4 Fasten the smaller screw well 5 Fasten the larger conic screw well COMPLETE ARM ADJUSTMENT This adjustment is only necessary when a repair on this part of the arms has been carried out If only a small horizontal adjustment is necessary for example because the needle has bent a little bit then carry out the small horizontal arm adjustment The height of the arms is determined by ring G see figure 5 6 the horizontal position is determined by the cla
39. METAL 30M OHM 3306 701 RES METAL 1M OHM 3312 112 POTM ADJ 1K MT 3312 117 POTM ADJ 2K MT 3312 132 POTM ADJ 10K MT 3312 162 POTM ADJ 50K MT 3320 210 CAP CER 10NF 40V 3320 311 CAP CER 0 1uF 50V 3322 020 CAP STYROFLEX 22pF 160V 3323 410 CAP P E 1UF 63V 3325 162 CAP PP 6 8NF 100V 3325 262 CAP PP 0 068UF 250V 3326 521 CAP ELC 15uF 20V 3333 052 IC 111AM 3333 061 IC OPA 121KP 3333 323 IC REF 01 HP 3333 508 IC CMP 01 EP 3333 514 IC DG 212 CJ 3342 311 SWITCH SLIDE 0 161390 1 3346 002 RELAY REED V23100 V4005 A010 3348 191 FUSE 3 15A SLOW 3350 006 CONN FAST 2 8X0 5 3350 030 CONN 2 POL SUBCLIC 3350 137 CONN 2POL 280609 2 3353 127 CONN 16PM 180o 826469 8 3355 002 KEYING PLUG 926498 1 3359 002 POWER UNIT DC CONVERTOR 3366 006 FUSE HOLDER PCB MOUNT FOR TR5 3370 361 MOUNTINGPAD TO 5 3376 341 PHOTOMETER PC BOARD 3820 115 STICKER 4801 060 COVER VITAL SCIENTIFIC 8 ELECTRICAL DIAGRAMS LOCATION 2 3 10 27 6 R4 R 11 24 R1 R 12 23 R5 R 17 18 R 14 21 R 13 22 R7 R 16 19 R 15 20 R 9 26 R 8 25 P4 P 2 3 1 5 P 6 7 C 5 6 C 8 11 15 17 7 12 14 16 C4 C 1 3 13 IC 10 11 IC 2 IC 5 IC 3 4 IC 6 9 SW 1 REL 1 2 F1 TAB 1 11 PL 3 4 PL2 PL 1 PL 1 IC 1 F1 IC 10 11 A4 COVER 1 8 ELECTRICAL DIAGRAMS SERVICE MANUAL 91 1009 WNLVONJAJI 3LWO AN3ANIO JHLUNADS 6 VIN 8 08 ILIWOLOHA DNIHYN3G 82530 SONY Ti N333l0 GMI3N N3HJNN
40. PART NUMBER DESCRIPTION LOCATION 3002 016 SCREW ST ST M3X6 1 3002 307 SCREW ST ST M3X8 2 3002 318 SCREW M3X25 3 3002 336 SCREW ST ST M4X16 4 3004 212 SCREW ST ST M3X10 5 3008 034 SCREW ST M3X10 6 3011 011 NUT ST M3 7 3022 010 WASHER ST ST 3 2 8 3022 015 WASHER ST ST 4 3 9 3030 023 DOWEL PIN 10 3053 002 BEARINGBALL 11 3064 019 TIMING BELT 12 3106 109 BEARING 13 3366 122 SOLDERTAG 14 3370 401 CLIP CABLE 15 4105 184 BUSH 16 4105 185 BUSH 17 4501 459 BLOCK 18 4501 460 BLOCK 19 4502 290 BEAM 20 4753 010 CLIP 21 4802 242 PLATE 22 4802 243 VANE 23 UP DOWN STEPPER MOTOR 6001 455 PART NUMBER DESCRIPTION LOCATION 3008 024 SCREW ST M3X5 24 3130 075 SPROCKET T2 5 15 25 3363 505 STEPPER MOTOR 26 ADJUST BLOCK ASSEMBLY 6001 420 PART NUMBER DESCRIPTION LOCATION 3008 038 SCREW ST M3X16 27 3019 009 NUT M3 28 3022 030 WASHER ST ST 3 2 29 3023 055 SCR SPR 30 3030 045 DOWEL PIN 31 3053 001 BEARINGBALL 32 4501 501 BLOCK 33 4501 502 BLOCK 34 WASH ARM 6001 961 PART NUMBER DESCRIPTION LOCATION 6001 961 WASH ARM 35 6001 962 BLOCK ASSY 36 6001 963 FILTERBLOCK ASSY 37 3004 211 SCREW M3X8 38 3002 007 SCREW 39 9 32 VITAL SCIENTIFIC SERVICE MANUAL 9 MECHANICAL DRAWINGS WASH ARM VITAL SCIENTIFIC 9 33 9 MECHANICAL DRAWINGS MEASURING UNIT 6001 756 PART NUMBER DESCRIPTION LOCATION 1513 009 PVC TUBE 1 5X3 5 MM 1 3002 015 SCREW ST ST M3X4 2 3002 016 SCREW ST ST M3X6 3 3002 028 SCREW ST ST M4X12 4
41. SAMPLE ADJUSTMENT See section 5 4 and 5 5 this manual 4 5 2 16 CHECKING THE REAGENT AND SAMPLE ROTOR ADJUSTMENT See section 5 7 and 5 8 of this manual 4 5 2 17 THE FILTER WHEEL ADJUSTMENT See section 5 11 of this manual 4 5 2 18 CHECKING THE OPTO ENCODER ADJUSTMENT See section 5 12 of this manual VITAL SCIENTIFIC 4 15 4 INSTALLATION AND MAINTENANCE SERVICE MANUAL 4 5 2 19 PERFORM THE DICHROMATE END TEST To check the sample and reagent svringe reproducibilitv and the stabilitv of the photometer vou have to perform a Dichromate test A special liquid is used Potassium Dichromate which has an absorbance of approximatelv 8 Abs for the 340nm filter and used as sample As a reagent distilled water is used Description of the SR3 SR30 test Sample Potassium Dichromate 8 Abs Reagent Distilled water SR3 8 Abs sample dilution 1 100 gt 80 mAbs SR30 8 Abs sample dilution 1 10 gt 800 mAbs 1 Program the following two tests TESTPARAMETERS Name SR3 SR30 Abs Decimals No No No N Set all other absorbance and concentration limits to 0 to prevent error messages 2 Position in the program reagent positions menu the reagents for both tests Place a bottle of distilled water on these positions Put in the sample rotor on position 1 to 24 tubes with 8 Abs dichromat Run for all 24 samples both the SR3 and the SR30 test Wait until the results are
42. TUBE ASSY 4 3066 050 T CONNECTOR 5 3070 338 COVER 6 3344 004 VACUUM DETECTOR 7 3366 920 ASCO VALVE 3 WAY 8 6001 861 TUBING SET FOR WATER VACUUM 9 6001 754 PIPETTOR UNIT 10 6001 826 DILUTED WASTE CONTAINER 11 6001 405 TUBING ASSY REAGENT PROBE 12 6001 788 COLLECTING CASK 14 6001 423 VACUUM TUBE 15 6001 753 EXTERNAL PUMP UNIT 16 6001 428 TUBING SET FOR SAMPLE 17 6001 431 TUBING ASSY PIPETTOR UNIT 18 6001 756 MEASURING UNIT 19 6001 801 SAMPLE HEAD ASSY 20 6001 797 HEAT EXCHANGER REAGENT ROTOR 21 6001 839 BELLOWS PUMP 22 6002 172 CONNECTION BLOCK ASSEMBLY 24 6001 787 CONNECTOR PLATE ASSEMBLY 25 6001 961 WASH ARM 26 6001 872 TUBE ASSEMBLY 28 6001 860 WATER CONTAINER 31 6001 786 WATER CASK 32 1513 052 TUBE PVC 1 5X3 5 34 1513 053 TUBE PVC 1 5X2 5 35 6001 827 CONCENTRATED WASTE CONTAINER 36 Remark is only available in length of 5 mtr TUBE FROM TO CODE NO LENGTH A VALVE SAMPLE DISPENSER T CONNECTOR 3066 050 1513 001 260mm B VALVE REAGENT DISPENSER T CONNECTOR 3066 050 T CONNECTOR 3066 050 T CONNECTOR 3066 050 1513 001 550mm D VALVE14 GATE2 SOCKET 3066 088 E VALVE 5 GATE 1 WASHPLACE REAGENT 1513 001 1100mm G VALVE 4 GATE 3 CONNECTION BLOCK ASSY 1513 001 450mm 3066 050 3066 050 SOCKET 3066 087 J 3066 050 VACUUM TUBE UPPER 1513 001 120mm VACUUM TUBE LOW
43. Y 0888 090 9 ELBUL 8 kl ec2 eo zea 0550853158 40 90 so ro co zo 0 31 330 a 911233 1801 SERVICE MANUAL 8 ELECTRICAL DIAGRAMS U gt 2 0892 O c17 c LAMP gt 8 DELIVERV DATE DESCRIPTION NO TEMP CONTROL BRD 6BB1 771 3 SHEET 2 4 130 84 A DATE 05 10 93 CHECK J Ribberink DATE 09 03 95 UITRL SCIENTIFIC DIEREN THE NETHERLANDS VITAL SCIENTIFIC 8 ELECTRICAL DIAGRAMS SERVICE MANUAL SOMBIBINLIN n EE 13385 23141153125 18 10 S6 E0 50 X4u J9QQ 3 B 122 1003 71051409 dial 56 9 2 80114182530 ON M 3100 A31730 553 004 533814510 L TTWBAZ TZVBNZ 9 801 n duo 90730 lt zz I 1 823 oa si n ja Sito ELIS Karen ara sin CO sisi 9 B wa ssa v Kars C135440 n i VITAL SCIENTIFIC 8 ELECTRICAL DIAGRAMS SERVICE MANUAL N38310 214113125 Wain 55 59 60 iu 7038 1031N03 nae 91 50 0114182530 1242 13245 44 5 18110 Gv 92EE 3003 SSINHDNOM SNOISN3UIQ 532874510 a TINA 1004 Si n 5 Sinn 1 erq f U ry CS z 4 9124 RZE U b N 8921 ecu
44. a serial RS 232 line The System board is connected to six slave controllers the so called Motion Control Boards The Motion Control Boards control all the mechanical movements Each board contains a slave processor and can handle two stepper motors with the corresponding in and outputs The instrument is equipped with 12 stepper motors with the following subdivision Board Motor 1 Motor 2 Reagent rotor Sample rotor PCB PCB PCB PCB PCB PCB PCB PCB PCB 6001 759 6001 758 6001 758 6001 758 6001 771 6001 758 6001 758 6001 758 6001 761 EPROM EPROM EPROM EPROM EPROM EPROM EPROM Figure 3 2 PCB configuration 3 2 VITAL SCIENTIFIC SERVICE MANUAL 3 SYSTEM ORGANIZATION The Svstem Board is also connected to the Photometer Board where the actual photometric measurement is done The measured values are transmitted to the Svstem Board and the external computer 3 2 SYSTEM BOARD The overall functional control of the instrument is done by the System board and is exercised by a Z80 microprocessor The program for this board is stored an EPROM The svstem board receives requests for tests from the external c
45. and replace if necessary Check if the V sealing is sufficient greased and grease if necessary VITAL SCIENTIFIC SERVICE MANUAL 7 ERROR HANDLING E103 SAMPLE DISC ERROR Cause Solution This error occurs when the opto reading of the sample disc is not in accordance with the current theoretical position This happens normally when the disc is obstructed by something Perform a selective reset of the sample disc Check the operation of the opto couplers and replace if necessarv E112 DISCS INIT FAILED see E115 E116 E115 REAGENT DISC RESET FAILED Cause Solution This error will be generated when the motion control board does not report the reset position of the reagent disc after the reset command Check if the communication between system board and motion control board is present When there is no communication this can either be caused by the wiring the SIO on the system board the KIO on the motion control board or the main PCB fuse on the motion control board Replace the board s or replace the defective components Check if there is a mechanical obstruction of the reagent disc and repair if necessary Check if the opto couplers are working correctly and repair if necessary E116 SAMPLE DISC RESET FAILED Cause Solution This error will be generated when the motion control board does not report the reset position of the sample disc after the reset command Check if the communication between system boar
46. board Wash arm board lavout Arm reset board Encoder opto board Cooling unit diagrams VITAL SCIENTIFIC 8 ELECTRICAL DIAGRAMS 6001 772 6001 772 6001 772 6001 955 6001 773 6001 773 6001 773 6001 858 6001 858 6001 858 6001 882 6001 882 6001 882 6001 888 6001 900 6001 956 6001 959 8 ELECTRICAL DIAGRAMS SERVICE MANUAL SONY THIHLIN HL 26 Vivi SES 63193135 WWADVIG DNIZIM 31945 2530 k 31033N JAWA 314445 WNIVONIAZI HVO YILVM JAWA SN39v38 WiNYV WILY PI T JSNVA JA Sav Nase OTWAJA SH30NV FIZNIL SILNVUJIOL 031 15 3SIMH3HLO SSIINN SIINVUJIOL ONFAZJIM NOLI VITTO 31033N JAWA SN39v38 EE E x 5 D rome Warps e 000000000000 LX 202 GX 33 9X 0000900000000 a 8X 861009 40000000000900000 VITAL SCIENTIFIC 8 ELECTRICAL DIAGRAMS SERVICE MANUAL WNLVON3AJI ILVO AJMO 103135 INWVNIS NOU AYISI SONVINGHAZN N383I0 Nas3sNVdS 2818305 WIA 84056 TTG 914 214 tV IN3841Y381 JBNVA JA 3002
47. cycle the following adjustment can be done e Measure the temperature in the cuvette rotor with a very small temperature probe When the probe has too much mass it will cool down the cuvette and a good indication of the temperature is not possible anymore e Place the temperature probe in the cuvette that is in the lightpath e Adjust the temperature to 37 5 C 0 2 by means of P3 which is mounted behind the front plate Turning P3 360 clockwise means as an indication an increase of temperature of 0 4 C After an adjustment of the temperature repeat the wash and fill cycle and check the new set temperature 6 5 ADJUSTMENT OF THE PLL LIQUID SENSE This procedure describes the adjustment of the PLL circuit for liquid detection The potentiometers for this adjustment can be found just behind the tube connectors for the sample and the reagent arm if you remove the top cover next to the printer The procedure is the same for both arms Most important is to prevent that the PLL circuit is too sensitive and thinks it aspirates liquid while in fact it does not By means of surrounding the needle by full cups bottles the capacitive coupling between needle and rotor is made very high The tube and the liquid act as dielectricum Even then the needle may not sense liquid erroneously VITAL SCIENTIFIC 6 3 6 ELECTRICAL ADJUSTMENTS SERVICE MANUAL Liq uid sense sensitivitv adjustment PLL circuit SAMPLE ROTOR Turn P2 to th
48. have designed two different forms e Installation report e Malfunctioning report The Installation report should be filled out when the instrument is installed at the customers site The malfunctioning report should be filled out each time when a service action is performed on the instrument We request you to return the filled out forms to us on a regular base for example once a month by mail or fax whatever is convenient VITAL SCIENTIFIC 1 3 The contents of this manual with all figures tables and graphics are intellectual propertv of Vital Scientific Unauthorized commercial or non commercial excerption or copving of contents and use of this manual in total or in parts are strictly forbidden unless the editor gives written permission for it This manual was written and produced with the utmost care However errors cannot be fullv excluded Vital Scientific does not take anv responsibilitv and accepts no liabilities for incidents of anv kind that occur because of errors in the manual All product names that are mentioned in this manual are registered trademarks The manual describes the analvser svstem Fibron I and its software version 1 0 November 2002 The Fibron I is conceptualised manufactured and tested in accordance with the declaration of conformitv The declaration is supplied with each device in a separate file Please call vour local distributor if vou need advice or vou have anv questions Article No
49. in the service menu This will show the following screen Display wode Funct Check Add Adjust Lamp tt Reagent Arm Sample Arm LAMP Reagent Disc REF Abs Sample Disc Measurement Disc Filter Wash firm ISE Arm ISE Unit Pipettor Vacuum system Water sustem ni i Min Optical electronics Electronics Barcode scanner in 12345678 Dery ice By selecting this function the wavelength is automatically set to 340 nm VITAL SCIENTIFIC 5 1 5 MECHANICAL ADJUSTMENTS SERVICE MANUAL The lamp must be adjusted in such a way that the two bars LAMP REF and CUVET reach a maximum These values must be at least as high as the displaved minimum The two left bars represent the intensity of light measured by the reference detector LAMP and the measurement detector CUVET Procedure e Loosen screw Turn screw B to the extreme right position Turn screw C to lowest position Adjust screw B and C such that the bars CUVET and REF reach their maximum Fasten screw A Note Normally both the CUVET and REF signals reach their maximum at the same lamp position If this is not the case adjust for maximum REF signal and execute the adjustment described below 5 2 VITAL SCIENTIFIC SERVICE MANUAL 5 MECHANICAL ADJUSTMENTS Sometimes it is possible to gain some lamp energy by readjusting the cuvette photo sensor Normallv this is only the case when the inner rotor assembly has been remov
50. meter 472 digit a vacuum meter 0 1 bar vacuum and a temperature meter with a tolerance of 0 1 C 4 7 1 TWO ABSORBANCE TEST BOX The Two Absorbance test box is a test box that contains a simple circuit consisting of a number of precision resistors a capacitor and a push button switch By means of this box the photometer board can be calibrated see section 6 3 15V 10K 910 30M 110 6 8 Figure 7 1 Test box circuit The operation of the circuit is as follows A negative voltage 15V is connected to the test box presenting the input signal to the input amplifier of the photometer board When the push button is pressed 1 100 of the original voltage is presented to the input amplifier via the resistors This will result in an absorbance of 2 000 mAbs on the display The Two Absorbance test box can be ordered from Vital Scientific under code number 6001 283 4 7 2 MECHANICAL ADJUSTMENT TOOL There is a special tool available to easily perform mechanical adjustment of the sample and reagent arm see section 5 8 and 5 9 The tool consists of a metal block of 97mm which supports the reagent or sample arm during these adjustments Please note that it is not strictly necessary to have this tool but it will make the adjustment easier The tool can be ordered from Vital Scientific under code number 4999 001 4 32 VITAL SCIENTIFIC SERVICE MANUAL 4 INSTA
51. of separate motion control boards is started VITAL SCIENTIFIC 3 19 3 SYSTEM ORGANIZATION SERVICE MANUAL 3 11 SLAVES MOTION CONTROL BOARD INITIALISATION The sequence of initialization of the six motion control boards is the same for all boards except for the pipettor motion control board A2 this board will perform besides the below mentioned initializations a ROM test 1 KIO initialisation CTC initialisation PIO initialisation ROM test only for board A2 RAM test SIO initialisation DARON After this initialization the hardware reset procedure is started and all functions are checked The sequence of the reset procedure is fixed and is as follows 1 wash arm reagent and sample arm initially up reset sample arm reset reagent arm reset measuring rotor reset diluter reset filterwheel reset sample and reagent rotor reset bellows pump reset wash arm move arms to wash position fill bellows pump OO DX p OT PO IO When one the functions mentioned above is found not reset position not found or otherwise the reset procedure will stop at that step An error message will be displaved on the screen for the failing step and possiblv also for the remaining not performed steps 3 20 VITAL SCIENTIFIC SERVICE MANUAL 4 INSTALLATION AND MAINTENANCE 4 INSTALLATION AND MAINTENANCE 4 1 INTRODUCTION This section contains information for unpacking the instrument and the procedure for ins
52. policv 1 5 SPARE PARTS Refer to the section General Conditions of Sale in the Price List for the general spare parts policv 1 2 VITAL SCIENTIFIC SERVICE MANUAL 1 GENERAL 1 6 TECHNICAL ASSISTANCE Should vou encounter a problem that requires technical assistance vou mav call fax or email the Service Department of Vital Scientific Please be prepared to give a clear and complete description of the problem Also be prepared to give an answer to the following questions tvpe of instrument serial number software version when applicable revision level in case of a PCB 1 7 SERVICE TRAINING Service training courses are organised on a regular base Information is available from the Sales department of Vital Scientific In principal courses will be held at Vital Scientific but in consultation it is possible to organise a course at the customers site Participants are expected to have knowledge of analog digital electronics microprocessor techniques basic photometric techniques the English language Please contact the Service Department of Vital Scientific for information about service trainings 1 8 FIELD SERVICE INFORMATION The purpose of the Field Service Information svstem is to provide detailed information on the frequency and nature of failures that occur during field use This information enables us to take corrective actions and to improve on the design and manufacturing methods For this purpose we
53. power supplv Note that the power supplv accepts input voltages from 95 to 264V with a frequencv from 47 to 63 Hz The specifications of the switched power supplv are as follows Output Maximum current se These outputs supply the different voltages to the boards fan s etceteras Refer to the DC wiring diagram of section 8 and the table below 5V 12V 14V 30V GND Black X II Discs Board 6001 758 X Photometer Board 6001 761 X TT X Valve Boards 60090 X For the power supply 3359 039 exchangeable parts are available 3 18 VITAL SCIENTIFIC SERVICE MANUAL 3 SYSTEM ORGANIZATION 3 9 FUSES The table below illustrates the fuses used in the instrument Description Type Location Cable Filter Board 3 15 A Slow 3 10 SYSTEM INITIALISATION 3 10 1 1 PIO initialisation The microprocessor will first initialise the PIO s and set the lines that are designated as inputs and outputs 3 10 1 2 RAM test 1 bits and O bits are shifted through the memory to check the correct working 3 10 1 3 ROM test During this test the contents of the program memory EPROM is tested by means of a checksum 3 10 1 4 CTC initialisation At this time the counter registers are cleared and the interrupt frequency is set 3 10 1 5 SIO initialisation At this time the six serial interfaces to the motion control boards are initialized Then the initialisation
54. printed and calculate the CV from SR3 and SR30 The CV of the SR30 test should be lt 1 The CV from the SR3 test should be lt 2 3 4 5 4 16 VITAL SCIENTIFIC SERVICE MANUAL 4 INSTALLATION AND MAINTENANCE 4 6 TEST AND DIAGNOSTIC PROCEDURES In order to support the service technician a number of test and adjustment routines are implemented in the service menu of the instrument This section supplies information about the use and the functions of these menus The service menu can be displayed by selecting Special Functions and Service menu Then the following menu will be displayed Analyser status Stand hy Req tray buffer 1 18 Run mode DUAL External 8 232 ON LINE Display mode Service Functional Check fidjustments Needle rinse history Last needle rinse date s ISE clean history 29 SEP ISE deproteinise history 15 0CT Error History 31 007 16 N0U SEES Eval Request Load Funct KLD IEE IES Fi F2 F3 M 5 Fo F8 Fg In this menu vou have the possibilitv to select Functional Check Adjustments Needle rinse historv ISE clean historv ISE deproteinise historv or Error historv Needle rinse history In this menu you can see the last dates of the needle rinse procedures ISE clean history In this menu you can see the last dates of the ISE clean procedure ISE deproteinise history In this menu you can see the last dates of the ISE deprot
55. reagent level detection The detailed diagrams are shown in section 8 of this manual An oscillator signal of 13 kHz 10V top top is made by the oscillator circuit built around opamp 2 IC4 This signal is connected to the mechanical assembly of the rotor This oscillator signal will be returned to this circuitry via the cup bottle needle the FET transistor of the arm board inside the arm and will be recognized by the PLL Phase Locked Loop circuit IC1 IC3 The output of IC1 IC3 willbecome high and will stop the vertical motor via the PIO and KIO of the Motion Control Board A3 A7 for the sample arm control The coupling between the metal rotor and the metal needle is capacitive where the tube and the liquid act as di electricum The sensitivity of the liquid detection can be adjusted by means of one of the two potentiometers mounted on top of the PLL board Needle gt SAMPLE Sample Arm E e MOTION CONTROL Board LIQU ID BOARD SAMPLE ARM DETECTION Oscilator Needle REAGENT Reagent Arm MOTION CONTROL Board L Q U D BOARD REAGENT ARM DETECTION Oscilator Figure 3 11 Block diagram PLL Board VITAL SCIENTIFIC 3 17 3 SYSTEM ORGANIZATION SERVICE MANUAL 3 8 POWER SUPPLV The DC power supply of the instrument is provided by a switched power supply The primarv voltage is routed via the connection terminal to the switched
56. repair if necessary Check if the opto couplers are working correctly and repair if necessary E137 WASHARM BELLOWS PUMP COMM ERROR Cause This error will be generated when the communication to motion control board for the washarm bellows pump stalls or if an unknown character is echoed Normally this error will occur when the communication is disrupted or distorted Solution Check if the communication between system board and motion control board is present When there is no communication this can either be caused by the wiring the SIO on the system board the KIO on the motion control board or the main PCB fuse on the motion control board Replace the board s or replace the defective components 7 1 10 ISE ERRORS E140 ISE NOT PRESENT Cause This error will be generated when the instrument detects that the ISE unit is not well connected A signal has been send to the ISE and no return signal is detected Solution Check is the communication cable is connected properly and if the ISE unit is switched on E141 ISE SYNCHRONISATION ERROR Cause A request is sent and the ISE busy signal is active Solution Check communication between ISE and analyser cable connections switched on fuses F3 or F4 on SBC card of ISE E142 ISE COMMUNICATION ERROR Cause Failure detected in communication between analyser and ISE Solution See E141 7 20 VITAL SCIENTIFIC SERVICE MANUAL 7 ERROR HANDLING W143 ISE RESET
57. rotor the vane see figure 5 5 will be in the fully down position opto 2 and not in opto 1 any more Before checking the cuvette detection the wash arm adjustment must be checked 12 Select wash arm service menu 13 Reset the wash arm F1 5 6 VITAL SCIENTIFIC SERVICE MANUAL 5 MECHANICAL ADJUSTMENTS Check if cuvettes is 0 if not 13 1 Disable the wash arm 13 2 Push the wash arm downward until the needles touch the cuvette bottom 13 3 Check if opto 1 is 0 if not bend opto 1 carefully downwards in such a way that it just switches to 0 14 Remove cuvette 15 Reset wash arm F1 Check if cuvettes is 1 If not check if opto 1 is in right position and reset the wash arm again 5 5 BELLOWS PUMP ADJUSTMENT The bellows pump consists of 6 bellows one not used which aspirate or dispense simultaneously One bellow on the motor side controls the wash water of the sample needle Four bellows control the wash water in the 4 cuvette wash positions The optimum cycle volume for the pump is 400 ul and can be adjusted by means of shifting beam A towards respectively away from the bellows To decrease the volume loosen screw 1 and 2 move beam A away from bellows tighten screw 1 and 2 To increase the volume loosen screw 1 and 2 move beam A towards bellows tighten screw 1 and 2 The screws are accessible through a slit in the baseplate VITAL SCIENTIFIC 5 7 5
58. the Valve board in the pump unit Check all the wiring Replace KIO on the svstem board Replace svstem board E12 WATER RUNNING OUT Cause When the highest float switch in the water cask is not detecting water for more than 25 seconds Solution Check the water in the water container Make sure the pump is pumping water bv holding the pump unit and the water container at a higher level for a short while Check the water filter Check level switches reed relavs Check if the water tubing is leaking or blocked If signal is on PL7 of system board replace KIO E13 LAMP FAILURE Cause When during a measurement a counter overrange signal is detected The signal counter overrange signal is generated when the photocell signal is too low Solution Check lamp adjustment and replace lamp if necessarv Check if the lamp voltage is present and well adjusted Check if the right filter is selected When also the blank data values are not correct the problem can be caused bv awrong gain setting of the input amplifier or a defective input amplifier on the Photometer Board Check the photometer board adjustment Replace photometer board 7 6 VITAL SCIENTIFIC SERVICE MANUAL 7 ERROR HANDLING E14 CUVETTES TEMP ERROR Cause When the temperature control circuit is reporting the temperature not OK signal for 5 minutes The circuit will be switched off to prevent hazardous situations cuvet therm control signal pin 5 PL7 Soluti
59. the cuvette bottom must be 0 5mm if this is according the procedure given in section 5 3 Check the functioning of valve V2 and V4 7 24 VITAL SCIENTIFIC SERVICE MANUAL 7 ERROR HANDLING 7 2 2 3 Low measuring values Svmptom The values of the controls are too low Probable cause The temperature of the cuvette rotor is too low Solution Check the temperature of the cuvette rotor bv following the adjustment procedure of section 6 5 Blank error Svmptom The SD value of the cuvette blank is higher then 0 0150 Probable cause The rotor is damaged or polluted Solution Replace the rotor bv a new one and perform a new blank procedure Probable cause When the reagent needle is dispensing water which contains air bubbles the cuvette rotor caused bv leakage of the liquid svstem or leakage the teflon svringe tips or svringe valve Solution Check the tubing on leakage Check the condition of the teflon tips and valve and replace if necessary 7 2 3 General trouble shooting 7 2 3 1 The instrument does not start up Check the fuses in the mains cable connection Open slot A2 and check the power supply voltages with respect to tab6 tab 2 5 1 V tab 8 14 V tab 1 30 V The 12 V has no connection to this board but the fans run on 12V 7 2 3 2 Communication check system board lt gt motion control boards If a reset fails see paragraph 3 11 for the power on sequence the instrument will stop the reset procedure and wil
60. the pallet 4 bolts and place it in its operating position Figure 4 2 Rear panel 6 Remove the accessories packing and the packing of the cooling unit 7 Remove the small packing foam between cuvette cover and large foam piece Remove the large piece of foam under the arms Remove the tubes around the arm shafts 8 Connect the cable from the pump unit to the connector marked PUMPUNIT at the rear of the instrument see figure 4 2 VITAL SCIENTIFIC 4 3 4 INSTALLATION AND MAINTENANCE SERVICE MANUAL 4 4 ANALVSER COOLING 1 2 3 4 8 9 10 IN OUT 1 2 3 4 OUT IN PUMP UNIT COOLING 5 6 7 UNIT WATER Figure 4 3 Tubing diagram Connect all tubing according to figure 4 3 Connect the waste detector plug of the diluted waste container to the connector marked waste Fill the water container with system liquid and bidest 25 ml on a full water container Connect tubing between the cooling unit and analyser as shown in figure 4 3 Place the Cooling unit at the final operating position Remove the filling lid from the cooling unit and fill the cooling unit with 2 litres of the cooling liquid Add demineralised or distilled water until the liquid level almost reaches the top of the container Close the lid again The filling lid has an air tap that should be open so that it is possible for the liquid to expand because of tempera
61. the photometer is ignored There are two operating modes the Mono Mode for single reagent use and the Dual Mode where a second reagent can be used The user can switch over from the Mono Mode to the Dual Mode but not during testing After switching from one mode to the other the instrument will reset 2 2 1 MONO MODE CYCLE The Mono Mode has a cycle time of 20 seconds so the theoretical throughput of the instrument is 180 tests hour 3x60 In practice the throughput is a little less due to possible sample blanks and the start up time which is 13 minutes In the figure below you find a graphic representation of the 20 second cycle The line represents the cuvette that starts at the sample dispensing position T 25 cuvette position 30 measuring kinetic points 22 35 RY L 40 4 45 jil 4 4 disp sample disp 1 48 g reagentdisp measuring endpoint reagent disp bichromatic T hin 10 washing 15 1 reagents 12 20 blank NG WO Ae 1 2 3 4 5 6 9 10 11 12 13 14 15 16 17 18 19 20 1 2 3 4 5 6 7 9 10 11 12 13 14 15 time Figure 2 2 Graphic representation Mono Note that after every cycle the cuvette rotor shifts one position 2 4 VITAL SCIENTIFIC SERVICE MANUAL 2
62. the spindle and drive nuts still mounted in the drive assembiv with a solvent e g refined petrol and a brush 4 12 VITAL SCIENTIFIC SERVICE MANUAL 4 INSTALLATION AND MAINTENANCE pa CIRCLIP DRIVE NUT vi za E DRIVE ASSY L4 il LH HI 2 BUSH HH A _ CH DRIVE PIN LL GUIDE SHAFT LEAD SCREW BALL BEARING U SPROCKET Figure 4 4 Pipettor drive assembly Drive nut mounting alignment This alignment is only necessary when the drive nuts are removed out of the drive assembly 1 Turn both drive nuts over the spindle measure the distance between the legs of the drive assembly and keep the same distance on the spindle 2 Insert the spindle with drive nuts through the drive assembly while pushing the drive nuts radial towards the spindle Slide the spindle with drive nuts until both drive nuts are in the middle of a drive assembly leg and at both ends you can see the edge of the drive nut 3 Tighten the screws which keep the drive nuts in place but do not overtight them Check to see that the spindle turns well in the drive assembly and has some axial play in the drive assembly VITAL SCIENTIFIC 4 13 4 INSTALLATION AND MAINTENANCE SERVICE MANUAL Spindle mounting 1 Insert the spindle and drive assy on its place Install the upper ball bearing with circlip on the spindle and the lower ball bearing in the lower beam Slide the spindle with drive assembl
63. to be trained in the use of the instrument bv the manufacturer or their representative A service technician is also expected to be familiar with the normal and special operations of the instrument as described in this User s Manual Description of Svmbols A This means that the labelled item could lead to personal injurv and or damage to the analvser Note Notes contain additional information corresponding to the text Fibron 1 Manual Vital Scientific NV Service manual Rev 1 0 Page 2 77 vital scientific Electrical Please pav attention to the following items e To prevent the risk of electrical shock and or damage to the instrument Operators should not open the top cover of the instrument Only authorised personnel for example service technicians may open the instrument to perform maintenance or repair e Don t touch the main board when the power is on as this may cause severe injury or death Mechanical There is no risk presented by the mechanical parts of the instrument when the covers are closed If the covers of the instrument are removed mechanical parts could cause personal injury if touched or become damaged Therefore the following advice must be followed DO NOT wear loose garments or jewelry that could catch in mechanisms DO NOT put your fingers hands into the path of any part while the analyzer is in operation DO NOT attempt to make mechanical repairs unless the instrument is not in operation or turned OFF
64. when a cuvette error occurs Usually the error disappears when the cuvette error has been solved Solution Check if the pump can move smoothly and repair if necessary Check if the electrical connections are correct and repair if necessary Check the opto coupler functions are working and repair if necessary 7 18 VITAL SCIENTIFIC SERVICE MANUAL 7 ERROR HANDLING E124 WATER OVERFLOW MEASUREMENT DISC Cause Solution This error will be generated when the overflow contact signal detects a low impedance to ground water level in the cuvette rotor is too high The overflow of the cuvette rotor can be caused by not sufficient vacuum badiv adjusted wash arm needle against bottom cuvette rotor or an electrical short circuit of the detection wire to frame Check visuallv if there is reallv a water overflow can also be caused bv bubbles If this is the case clean the washarm and check the vacuum and the valves V2 V4 and V14 If there is no real water overflow clean the underside of the washarm For older washarms the resistance of the isolator can become too low replace washarm E125 NO CUVETTES Cause Solution This error will be generated when at resetting the wash arm can reach a too low position according to the opto read out Check if a cuvette rotor is present and place one if necessarv Check if the washarm is fixed properly and fix if necessary Check if the wash arm adjustment is according the procedure giv
65. 0 4 3002 029 SCREW ST ST M4X16 5 3002 333 SCREW ST ST M4X12 8 3002 336 SCREW ST ST M4X16 9 3002 388 SCREW RVS M6X25 10 3002 621 SCREW ST ST M3X16 11 3004 211 SCREW ST ST M3X8 12 3008 024 SCREW ST ST M3X5 13 3008 040 SCREW ST ST M4X5 14 3008 043 SCREW ST ST M4x12 15 3009 016 SCREW M8x50 16 3011 011 NUT ST ST M3 17 3011 019 NUT M4X27 5MM 18 3019 021 NUT 4 19 3022 010 WASHER ST ST 3 2 20 3022 015 WASHER ST ST 4 3 21 3023 042 SHIMRING 26 3023 035 SPRING WASHER 27 3029 017 O RING 12x1 5 28 3029 051 V SEAL RING 29 3105 005 ADJ RING 06H8X12X8 35 3105 006 ADJUSTING RING 36 3105 007 ADJUSTING RING 37 3106 113 BEARINGBUSH 38 3350 050 CONN IP HDP 20PIN 39 3350 107 CONN 1PM 40 3350 149 FASTON TAB 41 3352 019 CONN HOUSING 6P 42 3370 444 TIE ANCHOR MOUNT NYLON 44 3381 053 INTERRUPTOR H21B1 45 4102 319 SHAFT 46 3019 049 STUD 47 4102 329 SHAFT 48 4102 331 SHAFT 49 4102 336 STUD 50 4105 186 BUSH 51 4105 187 BUSH 52 4105 196 BUSH 53 4105 197 BUSH 54 4106 227 RING 55 4106 268 RING 56 4501 461 BLOCK 58 4502 357 STRIP 59 4502 358 STRIP 60 4802 331 PLATE 61 4802 245 PLATE 62 4803 381 BRACKET 63 4803 383 BRACKET 64 4803 457 TOPPLATE 65 4803 458 BASEPLATE 66 6001 888 ARM RESET BOARD 67 6001 882 WASH ARM BOARD 69 6001 420 ADJUST BLOCK ASSY 70 6001 430 ENCODER STEPMOTOR UNIT 71 6001 756 MEASURING UNIT 72 R amp S STIRRING ISE UNIT 73 drawing not like reall 6001 445 ROTOR UNIT 74 6001 930 ENCODER UNIT ROTORS 75
66. 066 090 CONNECTOR 20 3066 092 ELBOW 21 3066 111 CONNECTOR NYLON WHITE 22 3105 028 BUSH 23 6001 965 VALVE 2 WAY INCL CONNECTOR 26 4501 516 PLATE 29 4501 517 BLOCK 30 4509 174 PLATE 31 6001 733 CONNECTOR ASSY 32 In case of a standard instrument without concentrated waste the collecting cask contains one single chamber Part No 6002 172 9 8 VITAL SCIENTIFIC SERVICE MANUAL 9 MECHANICAL DRAWINGS CONNECTOR PLATE ASSY COLLECTING CASK ASSY Note This drawing is valid for the instrument version with the concentrated waste option installed VITAL SCIENTIFIC 9 9 9 MECHANICAL DRAWINGS SERVICE MANUAL ASSEMBLY OF MAIN UNITS FOR WATER AND VACUUM SYSTEM PART NUMBER DESCRIPTION LOCATION 3002 018 SCREW M3X8 1 3002 019 SCREW M3X10 2 3002 027 SCREW M4X10 3 3004 210 SCREW M3X6 4 3002 313 SCREW M3X16 5 3002 336 SCREW M4X16 6 3004 211 SCREW M3X8 7 3004 218 SCREW M3X30 9 3011 016 NUT M3 10 3019 021 STUD M4 11 3066 052 CONNECTOR 12 3070 338 COVER 14 3344 004 VACUUM DETECTOR 15 6001 754 PIPETTOR UNIT 17 6001 826 DILUTED WASTE CONTAINER 18 6001 756 MEASURING UNIT 19 6001 860 WATER CONTAINER 20 6001 788 COLLECTING CASK ASSY 21 6001 423 VACUUM TUBE 22 6001 753 EXTERNAL PUMP UNIT 23 6001 797 HEAT EXCHANGER REAGENT ROTOR 27 6001 839 BELLOWS PUMP 28 6002 172 CONNECTION BLOCK ASSEMBLY 30 6001 787 CONNECTION PLATE ASSEMBLY 31 6001 961 WASH ARM 32 6001 786 WATER CASK 36 6001 827 CONCENTRATED WASTE CONTAINER 37 9 10 VITAL SCI
67. 14 If this does not solve the problem replace the main board pn 6002 511 Fibron 1 Manual Vital Scientific NV Service manual Rev 1 0 Page 6 77 vital scientific If the instrument does not stat up When the instrument does not start up check the fuses at the line entrv if this is ok check the voltage on the orange connector on the main board This should be 4 85 to 5 15 V If this is not ok remove the cover over the power supply and check the fuse and if necessary adjust the voltage using the potentiometer SVR1 in the left front corner If this fuse is ok but the power is not adjustable replace the power supplv pn 3359 045 If this does not solve the problem replace the main board pn 6002 511 In all case of replacing one of the boards alwavs check the settings in the hidden menu and set them as follow Light level 41000 Gain PT threshold 37 APTT threshold 25 Open threshold 37 Auto start threshold 512 Number Count 4 When setting these numbers always move to the next screen by pressing ENTER until you get to the screen Reset Fibron Then switch off and on the instrument by using the main switch Only when entering all system settings these will be stored Fibron 1 Manual Vital Scientific NV Service manual Rev 1 0 Page 7 4 7 vital scientific Parts overview 3374 100 6002 505 3374 124 3381 033 6002 503 3359 045 6002 511 Fibron 1 Manual Vital Scientific NV Se
68. 3 2 11 3022 015 WASHER ST ST 4 3 12 3022 030 WASHER ST ST 3 2 13 3029 039 O RING 24X2 36624 14 3030 028 DOWEL PIN 3m6X16 15 3053 002 BEARINGBALL 3X10X4 16 3053 025 BEARINGBALL 6X15X5 17 3064 018 TIMING BELT T 2 5 160 18 3106 016 BEARING BUSH 19 3130 081 SPROCKET 20 3381 053 OPTO INTERRUPTOR H21B1 21 4100 347 BELLOWS 22 4100 532 BUSH 23 4102 246 PLUG 24 4102 352 SHAFT 25 4102 346 SHAFT 26 4104 085 PIPE 28 4105 192 BUSH 29 4501 464 BLOCK 30 4501 465 BLOCK 31 4502 291 BEAM 32 4509 122 CAP 33 4802 251 PLATE 34 4802 252 PLATE 35 4802 253 PLATE 36 4802 254 PLATE 37 6001 831 BELLOWS MOTOR UNIT 38 9 14 SERVICE MANUAL VITAL SCIENTIFIC SERVICE MANUAL 9 MECHANICAL DRAWINGS BELLOWS PUMP VITAL SCIENTIFIC 9 15 9 MECHANICAL DRAWINGS EXTERNAL PUMP UNIT 6001 753 SERVICE MANUAL PART NUMBER DESCRIPTION LOCATION 1513 001 PVC TUBER 2 X4 CL 1 1513 002 PVC TUBE 4 6 CLEAR 2 3002 016 SCREW ST ST M3X6 3 3002 018 SCREW ST ST M3X8 4 3002 026 SCREW ST ST M4X8 5 3002 027 SCREW ST ST M4X10 6 3002 030 SCREW ST ST M4X6 7 3002 033 SCREW ST ST M4X40 8 3002 322 SCREW ST ST M3X40 9 3004 213 SCREW ST ST M3X8 10 3011 011 NUT ST ST M3 11 3019 013 NUT INSERT ST 1 8mm 12 E 3019 014 NUT INSERT ST GAL M4 1 8 mm 13 3022 015 WASHER ST ST 4 3 14 3066 052 CONN 10 32 UNF 15 3066 087 SOCKET 16 3066 090 CONNECTOR 17 3066 095 SOCKET 2 5 18 3066 096 SOCKET 3 4 19 30
69. 3 620 MOTOR 12V ESCAP 37 4102 243 WORM 38 4501 518 BLOCK 39 4802 329 PLATE MOTOR U D MOV 40 9 42 SERVICE MANUAL VITAL SCIENTIFIC SERVICE MANUAL 9 MECHANICAL DRAWINGS VALVE DRIVE UNIT VITAL SCIENTIFIC 9 43 9 MECHANICAL DRAWINGS HEAT EXCHANGER REAGENT ROTOR 6001 797 PART NUMBER DESCRIPTION LOCATION 3002 029 SCREW ST ST M4X16 1 3002 021 SCREW ST ST M3X16 1A 3002 025 SCREW ST ST M3X30 2 3002 031 SCREW ST ST M4X20 3 3002 034 SCREW ST CAD M4X45 4 3002 322 SCREW ST ST M3X40 5 3004 216 SCREW ST ST M3X20 6 3009 019 SCREW M8x30 POL 7 3011 028 NUT M8 POLVAM 8 3066 081 TUBE PILLAR M10x1keg 9 4501 497 BLOCK 10 4501 499 BLOCK 11 4509 158 BLOCK 12 4801 105 PLATE 13 4801 106 PLATE 14 4801 107 PLATE 15 4801 108 PLATE 16 4801 109 PLATE 17 4802 298 PLATE 18 4802 299 PLATE 19 4802 300 INSULATION PLATE 20 4802 301 PACKING 21 6001 393 BLOCK ASSV 22 6001 796 FAN COOLUNIT 23 6001 396 BLOCK ASSV 24 6001 798 SINK ASSV 25 BLOCK ASSY 6001 393 PART DESCRIPTION LOCATION 3004 215 SCREW ST ST 26 3029 050 V SEAL RING 27 3066 073 PVC KNEE 28 4105 223 BUSH 29 4105 224 BUSH 30 4501 490 BLOCK 31 4501 498 BLOCK 32 BLOCK ASSY 6001 396 PART NUMBER DESCRIPTION LOCATION 3029 050 V SEAL RING 33 3066 073 PVC KNEE 34 4105 223 BUSH 35 4105 224 BUSH 36 4501 463 BLOCK 37 9 44 SERVICE MANUAL VITAL SCIENTIFIC SERVICE MANUAL 9 MECHANICAL DRAWINGS HEAT REAGENT ROTOR BLOCK ASSV B
70. 34 3105 005 ADJ RING 06 8 12 8 35 3105 006 ADJUSTING RING 36 3105 007 ADJUSTING RING 37 3106 113 BEARINGBUSH 38 3350 050 CONN IP HDP 20PIN 39 3350 107 CONN 1PM 40 3350 149 FASTON TAB 41 3352 019 CONN HOUSING 6P 42 3370 444 TIE ANCHOR MOUNT NYLON 44 3381 053 OPTO INTERRUPTOR H21B1 45 4102 319 SHAFT 46 3019 049 STUD 47 4102 329 SHAFT 48 4102 331 SHAFT 49 4102 336 STUD 50 4105 186 BUSH 51 4105 187 BUSH 52 4105 196 BUSH 53 4105 197 BUSH 54 4106 227 RING 55 4106 268 RING 56 4501 461 BLOCK 58 4502 357 STRIP 59 4502 358 STRIP 60 4802 331 PLATE 61 4802 245 PLATE 62 4803 381 BRACKET 63 4803 383 BRACKET 64 4803 457 TOPPLATE 65 4803 458 BASEPLATE 66 6001 888 ARM RESET BOARD 67 6001 882 WASH ARM BOARD 69 6001 420 ADJUST BLOCK ASSY 70 6001 430 ENCODER STEPMOTOR UNIT 71 6001 756 MEASURING UNIT 72 R amp S STIRRING ISE UNIT 73 6001 445 ROTOR UNIT 74 6001 930 ENCODER OPTO UNIT ROTORS 75 6001 448 CARRIAGE ASSY 76 6001 449 CLAMP ASSY 77 6001 931 ENCODER OPTO UNITS ARM 78 6001 455 UP DOWN STEPMOTOR UNIT 79 6001 456 SWING UNIT 80 6001 811 MOVEMENT ASSY 81 9 22 SERVICE MANUAL VITAL SCIENTIFIC SERVICE MANUAL 9 MECHANICAL DRAWINGS MECHANICAL ASSEMBLY SHEET 2 ri ME 04 f VITAL SCIENTIFIC 9 23 9 MECHANICAL DRAWINGS MECHANICAL ASSEMBLY 6001 752 sheet 3 PART NUMBER DESCRIPTION LOCATION 3002 016 SCREW ST ST M3X6 1 3002 028 SCREW ST ST M4X12 2 3002 026 SCREW ST ST M4X8 3 3002 027 SCREW ST ST M4X1
71. 4129 019 PULLEY 38 4129 020 PULLEY 39 4129 021 PULLEY 40 4501 511 LEVER 41 4501 512 BLOCK 42 4509 162 STRIP 43 4509 163 NEEDLE HOLDER 44 4509 165 PLATE 45 4509 167 PLATE 46 4802 330 PLATE 47 4803 450 BRACKET 48 6001 773 STIRRER DRIVER BOARD 49 Parts specific for the ISE arm are optional and not present in all instruments Maintenance parts for stirrer unit PART NUMBER DESCRIPTION LOCATION 6001 982 Bearing assembly sample mixer 19 37 6001 983 Bearing assembly reagent mixer 21 40 9 28 VITAL SCIENTIFIC SERVICE MANUAL 9 MECHANICAL DRAWINGS STIRRING ISE UNIT VITAL SCIENTIFIC 9 29 9 MECHANICAL DRAWINGS SERVICE MANUAL SAMPLE REAGENT ROTOR BASE UNIT 6001 445 PART DESCRIPTION LOCATION 3002 015 SCREW 5 5 M3X4 1 3002 016 SCREW 5 5 M3X6 2 3002 030 SCREW ST ST M4X6 3 3002 336 SCREW ST ST M4X16 4 3022 010 WASHER ST ST 3 2 5 3022 015 WASHER ST ST 4 3 6 3053 001 BEARING BALL 7 3053 035 BALLBEARING 8 3130 077 SPROCKET 9 4100 447 HOLDER 10 4102 355 SHAFT 11 4102 324 SHAFT 12 4802 250 VANE 14 ENCODER UNIT FOR ROTOR 6001 930 PART NUMBER DESCRIPTION LOCATION 3002 016 SCREW ST ST M3X6 15 3019 011 NUT INSERT ST M3 16 4803 472 BRACKET 17 6001 900 ENCODER OPTO BOARD 18 9 30 VITAL SCIENTIFIC SERVICE MANUAL 9 MECHANICAL DRAWINGS ROTOR UNIT REAGENT SAMPLE ROTOR VITAL SCIENTIFIC 9 31 9 MECHANICAL DRAWINGS SERVICE MANUAL WASH ARM ASSEMBLV MOVEMENT ASSEMBLY 6001 811
72. 6001 448 CARRIAGE ASSV 76 6001 449 CLAMP ASSV 77 6001 931 ENCODER OPTO UNITS ARM 78 6001 455 UP DOWN STEPMOTOR UNIT 79 6001 456 SWING UNIT 80 6001 811 MOVEMENT ASSY 81 9 24 SERVICE MANUAL VITAL SCIENTIFIC SERVICE MANUAL 9 MECHANICAL DRAWINGS MECHANICAL ASSEMBLV SHEET 3 VITAL SCIENTIFIC 9 25 9 MECHANICAL DRAWINGS SWING UNIT 6001 456 PART NUMBER DESCRIPTION LOCATION 3002 016 SCREW ST ST M3X6 1 3008 040 SCREW ST ST 4 5 2 3053 034 BALLBEARING 3 3106 114 BEARING BUSH 4 3130 074 SPROCKET 5 4102 319 SHAFT 6 4501 456 BLOCK 7 4501 457 BLOCK 8 4802 239 VANE 9 CLAMP ASSEMBLY 6001 449 PART NUMBER DESCRIPTION LOCATION 3002 333 SCREW ST ST M4X12 10 3022 010 WASHER ST ST 4 12 11 3022 015 WASHER ST ST 3 2 12 3030 023 DOWEL PIN 13 3053 002 BEARINGBALL 14 4509 118 CLAMP 15 CARRIAGE ASSEMBLY 6001 448 PART NUMBER DESCRIPTION LOCATION 3002 306 SCREW ST ST M3X6 16 3022 010 WASHER ST ST 3 2 17 3030 013 DOWEL PIN 18 3064 019 TIMING BELT T2 5 19 3106 113 BEARING BUSH 20 4105 183 BUSH 21 4501 458 CARRIAGE 22 4753 009 CLIP 23 4753 009 VANE 24 ENCODER OPTO UNIT 6001 930 PART NUMBER DESCRIPTION LOCATION 3002 016 SCREW ST ST M3X6 25 3019 011 NUT INSERT ST M3 26 4803 472 BRACKET 27 6001 900 ENCODER OPTO BOARD 28 ENCODER STEPPER MOTOR UNIT 6001 430 PART NUMBER DESCRIPTION LOCATION 3004 204 SCREW ST ST M2X6 29 3008 040 SCREW ST M4X5 30 3130 076 SPROCKET T2 5 15 31 3363 505 STEPPER MOTOR 32 4700 543 EN
73. 6002 500 450 iy vital scientific Van Rensselaerweg 4 6956 AV Spankeren The Netherlands Phone 31 3134 30500 Fax 31 3134 27807 email service vital nl 7 vital scientific TABLE OF CONTENTS SAFETV PRECAUTIONS AND POTENTIAL HAZARDS 2 GENERAL 2 OPERATOR QUALIFICATIONS 2 SERVICE TECHNICIAN QUALIFICATIONS 2 DESCRIPTION OF SVMBOLS 2 ELECTRICAL 3 MECHANICAL 3 CHEMICAL 3 BIOHAZARDOUS MATERIALS 3 SAMPLES 3 WASTE SOLUTION AND SOLID WASTES 4 BIOHAZARDOUS PARTS 4 ADDITIONAL PRECAUTIONS 4 FLAMMABLES 4 ACCURACY PRECISION OF MEASURED RESULTS 4 APPLICATION 4 INSTALLATION REQUIREMENTS 5 INSTRUMENT UNUSED FOR A LONG PERIOD OF TIME 5 ERROR HANDLING 6 PARTS OVERVIEW 8 PARTS LIST 10 Fibron 1 Manual Vital Scientific NV Service manual Rev 1 0 Page 1 77 vital scientific Safetv precautions and Potential hazards General Before vou start installing and working with the analvser vou should read the safetv precautions and regulations detailed in this chapter Safety comes first Operator Qualifications Operation should be conducted under the management of a technician operator who has undergone training at the facilitv specified bv the sales agent For clinical tests the instrument should be used under the management of a doctor or clinical inspector Service Technician Qualifications To install maintain and repair the instrument a service technician has
74. 66 097 NUT 20 3066 098 CODE RING WHITE 21 3066 100 CODE RING RED 22 3066 101 CODE RING GREEN 23 3066 102 CODE RING BLUE 24 277 BUE CER EVO TRANSP Revision set Waterpump 6001 516 3351 027 3P PANEL MOUNT CONN 29 3365 014 PUMP VACUUM 12VDC 30 3370 042 SLEEVE CODE 1 VELLOW 32 3370 043 SLEEVE CODE 2 VELLOW 33 3370 053 FEED THROUGH 1440015 4 34 CP 3370 430 CLIPCABLE TIE INSULOK T18R W 35 3370 525 SCREWLOCKS FEMALE 36 4105 130 BUSH 37 4502 360 BEAM 38 4800 181 PLATE 39 4800 190 PLATE 40 4802 295 PLATE 41 4802 296 PLATE 42 4802 333 PLATE 43 4803 404 BRACKET 44 4803 459 BRACKET 45 6001 770 CABELFILTER BOARD 46 6001 791 PUMP UNIT BOARD 47 6001 847 FAN UNIT 48 6001 881 WATER PUMP 49 Membraneset Waterpump 6002 056 3913 049 Membrane 6002 041 Plunger kit see drawing 6002 042 Chamber kit chamber incl membrane see drawing 6001 516 Revisionset Waterpump see drawing 6002 056 Membraneset Waterpump see drawing Plunger kit 9 16 0 60 Chamber kit VITAL SCIENTIFIC 9 MECHANICAL DRAWINGS SERVICE MANUAL EXTERNAL PUMP UNIT 9 17 VITAL SCIENTIFIC 9 MECHANICAL DRAWINGS MAIN ELECTRICAL UNITS 6001 753 PART NUMBER DESCRIPTION LOCATION 6001 758 MOTION CONTROL BOARD 2 6001 759 SYSTEM BOARD 3 4102 354 SHAFT 4 6001 768 VALVE BOARD 5 6001 766 POWER UNIT 6 6001 761 PHOTOMETER BOARD 7 6001 771 TEMPERATURE BOARD 9 6001 858 TEMPERATURE COMPENSATION BOARD 12 9 18 SERVICE MANUAL
75. AL 4 195 852 1089 WUBDUIG DNIAIM 20 NOLLALBIS3O 1N3U10381 3192 45301730 3003 S53NHONOW 031015 351082410 SSIINN SJIJNU IJOL 110213100 5 015 3419 NS See W FS TN PS KA BIONUKIXI caeca I 2180212113 wave wanted VITAL SCIENTIFIC 8 ELECTRICAL DIAGRAMS SERVICE MANUAL 20168 L 9 59 L 00 aAA N3d0 7 1 180d NO 34 Z 1504 NO 3500 Z 1804 480 E L 10d 5 n 1010 6 avo 1 10 00 1 180d ADK Z JAWA SVJ ADM OJSV DONIS TV NSN TOM D ALVINOLLIINNOI BIZA TVNV VITAL SCIENTIFIC SERVICE MANUAL 8 ELECTRICAL DIAGRAMS VITAL SCIENTIFIC SERVICE MANUAL MOTION CONTROL BOARD 6001 758 PART NUMBER DESCRIPTION 3002 018 SCREW ST ST M3X8 3023 006 LOCKWASHER J 3 2 3302 301 RES METAL 100 OHM 3302 329 RES METAL 200 OHM 3302 341 RES METAL 270 OHM 3302 353 RES METAL 360 OHM 3302 373 RES METAL 560 OHM A 270 3302 429 RES METAL 2K OHM 3302 465 RES METAL 4K7 OHM 3302 469 RES METAL 5K1 OHM 3302 481 RES METAL 6K8 OHM 3302 501 RES
76. CODER DISC 33 9 26 SERVICE MANUAL VITAL SCIENTIFIC SERVICE MANUAL 9 MECHANICAL DRAWINGS SWING UNIT CLAMP ASSY CARRIAGE ASSY ENCODER OPTO ENCODER STEPPER MOTOR UNIT VITAL SCIENTIFIC 9 27 9 MECHANICAL DRAWINGS SERVICE MANUAL STIRRING UNIT with ISE NEEDLE 6001 783 optional STIRRING UNIT 6002 023 PART NUMBER DESCRIPTION LOCATION 1573 002 SIL TUBE 0 7 X 2 7 MM 1 3002 001 SCREW ST ST M2X4 2 3002 002 SCREW ST ST M2X5 3 3002 016 SCREW ST ST M3X6 4 3002 122 SCREW ST ST M1 6X3 5 3004 007 SCREW ST ST M2X6 6 3004 203 SCREW ST ST M2X5 7 3008 020 SCREW ST ST M3X3 8 3008 040 SCREW ST ST M4X5 9 3008 125 SCREW ST ST M6X5 10 3009 025 SCREW 11 3019 011 NUT INSERT ST 0 9mm 12 3019 013 NUT INSERT ST M3 1 8mm 13 3022 010 WASHER ST ST 3 2 14 3023 003 LOCKWASHER J 2 2 15 3023 006 LOCKWASHER J 3 2 16 3029 001 O RING 2 9 X 1 78 BUNA 17 3053 005 BALL 5MM BALLBEARINGSTEEL 18 3053 020 BEARINGBALL 8X16X5 ADR WX8ZZ 19 3053 037 BALLBEARING 02 5X06X2 6 20 3053 038 BALL BEARING 07X014X5 21 3053 204 BALL SLIDE ASSEMBLY AD 2 22 3064 041 BELT DRIVING 23 3363 625 MOTOR BALLBEARING 18VDC 24 3366 301 ROTARY SOLENOID 30VDC 25 3370 401 CLIP CABLE NXO 26 3370 430 CLIPCABLE TIE INSULOK T18R W 27 3372 010 WIRE 0 15 BLACK 28 3372 024 WIRE 0 15 WHITE ORANGE 29 3372 025 WIRE 0 15 WHITE GREEN 30 3750 043 SPRING 31 3751 024 SPRING 32 4100 650 PEN 33 4104 083 PIPE 34 4105 238 BUSH 35 4129 017 PULLEY 36 4129 018 PULLEY 37
77. ELLOWS PUMP 30V FAILED WASHARM ERROR BELLOWS PUMP ERROR WATER OVERFLOW MEASUREMENT DISC NO CUVETTES WASHARM BELLOWS PUMP INIT FAILED WASHARM RESET FAILED BELLOWS PUMP RESET FAILED WASHARM BELLOWS PUMP COMM ERROR VITAL SCIENTIFIC 7 3 7 ERROR HANDLING SERVICE MANUAL 7 4 E140 W141 E142 W143 E144 E145 E146 E147 E150 E151 E152 E153 E154 E155 E156 E157 E160 E161 E162 E163 E164 E165 E166 ISE NOT PRESENT ISE SYNCHRONISATION ERROR ISE COMMUNICATION ERROR ISE RESET ACTIVE ISE ARM ERROR ISE NOT READY ISE STANDARD A NOT DETECTED ISE STANDARD B NOT DETECTED ISE CHECK TEMPERATURE ISE NOT CALIBRATED ISE SAMPLE DETECTION ERROR ISE SAMPLE DETECTION ERROR ISE VALVES ERROR ISE ELECTRODES OR FLUIDS ERROR ISE INSUF CLEANER OR CONDIT ISE POWER FAIL Z80 MASTER CARD RS 232 ERROR TOO MUCH ETRANSMISSIONS TO z80 Z80 MASTER CARD RS 232 ERROR TOO MUCH DOUBLE ACKS SENT TO z80 Z80 MASTER CARD RS 232 ERROR CTS TOO LONG INACTIVE Z80 MASTER CARD RS 232 ERROR CTS TOO LONG ACTIVE MEMORY CHECKSUM CORRUPTED MEMORY CHANGED WATCHDOG EXPECTED IN 5 SECONDS VITAL SCIENTIFIC SERVICE MANUAL 7 ERROR HANDLING 7 1 3 280 ERRORS E00 280 MASTER CARD RAM CHECK ERROR Cause One or more bits are not readable writable E01 280 MASTER CARD ROM CHECK ERROR Cause Checksum is not equal to zero The cause of this error is most likely that the RAM or ROM on the system board are malfunctioning Solution Replace
78. ENT DISC ERROR Cause Solution 7 10 This error will be generated when e the reset position is detected but according to the software this is not possible e the direction encoders do not detect any movement of the disc e the direction encoders detect a movement which is not in accordance with the software co ordinates This can either be caused by loss of steps because of a mechanical malfunctioning friction blocking or electrical malfunctioning opto wiring control circuit Check if the measuring disc is moving smoothly and if there is any blockage Also check if the timing belt of the stepper motor is adjusted well not too loose or too tight Repair or adjust if necessary Check if the opto couplers are working and repair if necessary When there is no communication at all exchange the PIO on the motion control board for the measuring disc VITAL SCIENTIFIC SERVICE MANUAL 7 ERROR HANDLING E43 FILTER ERROR Cause This error will be generated when the reset position is detected but according to the software this is not possible This can either be caused by loss of steps because of a mechanical malfunctioning friction blocking or electrical malfunctioning opto wiring control circuit Solution Check if the filter wheel is moving smoothiv and if there is anv blockage Repair or adjust if necessarv Check if the opto coupler is working and repair if necessarv When there is no communication at all
79. ENTIFIC 9 MECHANICAL DRAWINGS SERVICE MANUAL ASSEMBLY OF MAIN UNITS FOR WATER AND VACUUM SYSTEM 9 11 VITAL SCIENTIFIC 9 MECHANICAL DRAWINGS WATER CASK 6001 786 PART NUMBER DESCRIPTION LOCATION 3002 013 SCREW ST ST M2 5X10 1 3004 226 SCREW ST ST M4X8 2 3029 020 O RING 036 36624 3 3066 052 CONN 10 32 UNF 4 3344 015 SWITGH LIQUID LEVEL 5 3350 107 CONN 1PM 6 3352 019 CONN HOUSING 6P 7 4100 492 PROP 8 4100 527 PIPE 9 4100 531 BUSH 10 4100 558 PROP 11 WASTE TUBES CONNECTION BLOCK ASSEMBLY 6002 172 PART NUMBER DESCRIPTION LOCATION 3066 052 CONNECTOR 12 3066 089 CONNECTOR 13 4501 567 BLOCK 14 VACUUM TUBE 6001 423 PART NUMBER DESCRIPTION LOCATION 3002 019 SCREW ST ST M3X10 36 3019 013 NUT INSERT ST M3 37 3029 010 RING 38 3066 052 CONNECTOR 39 4100 478 PLUG 40 4100 479 PLUG 41 4104 091 PIPE 42 4803 398 BRACKET 43 9 12 SERVICE MANUAL VITAL SCIENTIFIC SERVICE MANUAL 9 MECHANICAL DRAWINGS WATERCASK CONNECTION BLOCK ASSV VACUUM TUBE VITAL SCIENTIFIC 9 13 9 MECHANICAL DRAWINGS BELLOWS PUMP 6001 839 PART NUMBER DESCRIPTION LOCATION 1513 009 PVC TUBE 1 5X3 5 MM 1 3002 001 SCREW ST ST M2X4 2 3002 018 SCREW ST ST M3X8 3 3002 019 SCREW ST ST M3X10 4 3002 020 SCREW ST ST M3X12 5 3002 026 SCREW ST ST M4X8 6 3002 028 SCREW ST ST M4X12 7 3004 212 SCREW ST ST M4X12 7 3008 040 SCREW ST ST 4 5 8 3020 020 CIRCLIP 9 3022 005 WASHER ST ST 2 2 10 3022 010 WASHER ST ST
80. ER CONN PLATE ASSV CONN 2 L VALVE 5 GATE 2 VALVE 4 GATE 1 1513 001 200mm N VALVE 14 GATE 1 COLLECTING CASK 1513 001 360mm O VALVE1 GATE2 j COLLECTING CASK j 1513 001 140mm P VALVE 2 GATE 1 COLLECTING CASK 1513 001 240mm QO VALVE1 GATE1 CONN PLATE ASSY CONN 3 R VACUUMSWITCH ELBOW 3066 092 1513 002 40mm ELBOW 3066 092 SOCKET 3066 087 T CONNECTOR 3066 050 WATER CASK LOWER U WATER CASK UPPER 2X CONN PLATE ASSV CONN 1 AND 4 1513 001 160mm W SOCKET 3066 088 PIPE NO 1 6001 961 1513 010 580mm 580mm 5X Y BELLOWS PUMP PIPE 1513 010 700mm 4X 9 6 VITAL SCIENTIFIC 9 MECHANICAL DRAWINGS SERVICE MANUAL TUBING DIAGRAM 9 7 VITAL SCIENTIFIC 9 MECHANICAL DRAWINGS SERVICE MANUAL TUBE CONNECTOR PLATE ASSEMBLV 6001 787 PART DESCRIPTION LOCATION 3004 211 SCREW ST ST M3X8 1 3019 013 NUT INSERT ST 1 8 2 3066 095 SOCKET 2 5 3 3066 096 SOCKET 3 4 4 3066 097 NUT 5 3066 098 CODE RING WHITE 6 3066 100 CODE RING RED 7 3066 101 CODE RING GREEN 8 3066 102 CODE RING BLUE 9 3370 525 SCREWLOCKS FEMALE 10 4802 332 PLATE 11 6001 770 CABELFILTER BOARD 12 6001 788 COLLECTING CASK ASSY 13 COLLECTING CASK ASSY 6001 788 PART NUMBER DESCRIPTION LOCATION 1513 002 PVC TUBE R 4X6CLEAR 14 3002 020 SCREW 5 5 M3X12 15 3004 216 SCREW ST ST M3X20 16 3019 013 NUT INSERT ST M3 1 8mm 17 3029 058 O RING 16X2 FPM VITON 18 3029 060 O RING 7X1 19 3
81. ETAK 13K OHM RES METAL 15K OHM RES METAL 20K OHM RES METAL 30K OHM RES METAL 47K OHM RES METAL 100K OHM RES METAL 470K OHM RES METAL 2M2 OHM RES WW 0 22 OHM RES 6K98 OHM RES 10K1 OHM RES 215K 25PPM POTM ADJ 200 OHM POTM ADJ 500 OHM POTM ADJ 2K MT POTM ADJ 10K MT CAP CER AXIAL 1 nF 50V CAP CER AXIAL 22nF 50V CAP CER AXIAL 100nF 50V CAP POL 0 15 uF 63VDC 20 CAP ELC 2 2 uF CAP ELC 15uF 20V CAP ELC 47UF 20V CAP ELC 100 uF CAP ELEC 100 uF 63V DIODE IN 4148 DIODE MUR 1610 CT DIODE ZENER ZPD 2 7 DIODE ZENER ZPD 5 1 DIDOE ZENER 8 2V TRANS 549 B TRANS IRFZ 20 IC OP 07EP IC OP400 QUAD OP AMPLIFIER LOCATION 41 R 26 R 36 R 33 76 78 R 24 R 18 32 34 R 66 R 44 R 1 19 R 16 25 52 R 20 R 27 55 65 72 R 45 R 40 R 69 70 61 R 9 11 15 17 R 30 39 43 49 R 50 60 62 67 R 73 74 35 57 R 29 R 21 63 64 4 5 58 68 28 42 38 47 53 3 7 8 10 48 31 22 23 75 79 54 71 2 6 37 R 59 R 46 51 56 P2 1 P4 C 6 30 32 2 7 24 25 22 C4 10 C 1 3 5 8 C 14 16 23 C 18 21 26 29 C 12 13 C 9 11 17 D 4 6 8 10 D 12 15 1 2 D 11 D3 D7 T 6 11 T 1 5 IC 7 10 IC 9 VITAL SCIENTIFIC SERVICE MANUAL 8 ELECTRICAL DIAGRAMS 3333 028 IC LM 324N IC 6 8 VITAL SCIENTIFIC SERVICE MANUAL 8 ELECTRICAL DIAGRAMS PART NUMBER DESCRIPTION LOCATION 3333 311 IC 723PC IC 5 3333 316 IC L296 1 2 3333 323
82. IENT WATER E20 SAMPLE SYR POS ERROR E21 SAMPLE SYR POS ERROR 7 2 VITAL SCIENTIFIC SERVICE MANUAL 7 ERROR HANDLING E22 E23 E24 E25 E30 E31 E32 E35 E36 E37 E40 41 42 E43 E52 E55 E56 E57 E60 E61 E62 E63 E72 E75 E76 E77 E80 E81 E82 E83 E92 E95 E96 E97 E100 E101 E102 E103 E112 E115 E116 E117 E120 E121 E122 E123 E124 E125 E132 E135 E136 E137 SAMPLE VALVE ERROR REAG SYR POS ERROR REAG SYR POS ERROR REAG VALVE ERROR PIPETTOR 14V FAILED PIPETTOR 30V FAILED PIPETTOR INIT FAILED SAMPLE SYR RESET FAILED REAGENT SYR RESET FAILED PIPETTOR COMMUNICATION ERROR MEAS DISC 14V FAILED MEAS DISC 30V FAILED MEASUREMENT DISC ERROR FILTER ERROR MEAS DISC INIT FAILED MEAS DISC RESET FAILED FILTER RESET FAILED MEAS DISC COMMUNICATION ERROR SAMPLE ARM 14V FAILED SAMPLE ARM 30V FAILED SAMPLE ARM HORIZONTAL ERROR SAMPLE ARM VERTICAL ERROR SAMPLE ARM INIT FAILED SAMPLE ARM RESET FAILED SAMPLE ARM RESET FAILED SAMPLE ARM COMMUNICATION ERROR REAGENT ARM 14V FAILED REAGENT ARM 30V FAILED REAGENT ARM HORIZONTAL ERROR REAGENT ARM VERTICAL ERROR REAGENT INIT FAILED REAGENT ARM RESET FAILED REAGENT ARM RESET FAILED REAGENT ARM COMMUNICATION ERROR DISCS 14V FAILED DISCS 30V FAILED REAGENT DISC ERROR SAMPLE DISC ERROR DISCS INIT FAILED REAGENT DISC RESET FAILED SAMPLE DISC RESET FAILED REAG SAMP DISCS COMMUN ERROR WASHARM BELLOWS PUMP 14V FAILED WASHARM B
83. IFIC SERVICE MANUAL TEMP COMP BOARD ASSEMBLV A3 6001 858 PART NUMBER DESCRIPTION 3002 016 SCREW ST ST M3X6 3312 112 POTM ADJ 1K MT 3333 431 IC LM 3342 3350 039 CONN 167301 4 3351 236 CONN HOUSING 4PF 926475 4 3355 004 RETENTION PLUG 926477 1 3370 008 SLEEVE CABLE H30X20BL 3370 430 CLIPCABLE TIE INSULOK T18R W 3372 010 WIRE 0 15 BLACK 3372 645 CABLE SCR 4 X 0 14 3376 292 TEMP COMP PC BOARD 4501 495 BLOCK comp zijde 8 ELECTRICAL DIAGRAMS LOCATION 1 IC 1 A4 A4 VITAL SCIENTIFIC SERVICE MANUAL 8 ELECTRICAL DIAGRAMS WASH ARM BOARD 6001 882 PART NUMBER DESCRIPTION 3302 501 RES METAL 10K OHM 3302 601 RES METAL 100K OHM 3315 267 TRANS VOLTAGE SUPP 14 5V 3320 100 CAP CER 1NF 40V 3320 232 CAP CER 0 022UF 50V 3331 214 TRANS BC 559B 3351 235 CONN 4POL 826467 4 3353 056 CONN 2X6 POL 826469 6 3355 002 KEYING PLUG 926498 1 3376 590 WASH ARM PC BOARD 3381 052 OPT SWITCH H22B1 3820 115 STICKER N lt 0 0 O U1 oD N gt aaa LOCATION R1 R2 TZ 1 C1 C 2 6 T 1 PL 2 PL 1 A4 OC 1 3 VITAL SCIENTIFIC SERVICE MANUAL 8 ELECTRICAL DIAGRAMS Cooling unit diagrams Control unit cooling liquid container cooling liquid pumpi ME VITAL SCIENTIFIC SERVICE MANUAL 9 MECHANICAL DRAWINGS 9 MECHANICAL DRAWINGS 9 1 INTRODUCTION This section co
84. IT The washing unit washes the cuvettes after usage At the first 4 positions of the washing unit a long needle aspirates the cuvette contents and a short needle fills the cuvette with water At the fifth position the water is aspirated at the sixth position a drying block dries the cuvette With the option concentrated waste installed the liquid aspirated at the first cleaning action i e the reaction mixture is lead to the concentrated waste deposit The washing unit is equipped with liquid sensors to detect flooding of the cuvette 2 1 10 COOLING UNIT Cooling of the reagents is done by means of an external cooling unit This unit keeps the cooling liquid in the container at a fixed temperature which must be set at the cooling unit On its turn the cooling liquid is pumped through a heat exchanger in the analyser where cool air is fed through a closed reagent compartment VITAL SCIENTIFIC 2 3 2 FUNCTIONAL DESIGN SERVICE MANUAL 2 2 TIMING PRINCIPLE The instrument always works with a fixed cycle time In this fixed cycle all the necessary dispensing mixing washing and diluting take place and also a fixed number of measurements are made regardless if they are necessary or not This fixed cycle assures that all the necessary measurements for a test always take place at the right moment Even if no measurements have to be taken for example during washing of the first cuvettes the cycle takes place In that case the reading of
85. LLATION AND MAINTENANCE 4 8 PRINTED CIRCUIT BOARD REPAIR The detailed information supplied bv this manual provides the opportunitv for trouble shooting at component level however at an economical and qualitv point of view vou can decide to exchange the complete PCB The electronic design of the printed circuit boards 1 to 9 is such that most electronic parts used are commoniv available in everv well equipped workshop However do not use anv components with lower specifications then the original items because this can affect the performance of the instrument A list of components is included with everv circuit diagram at section 8 of this manual The electronic boards of the computer assembly are designed in such a way that repair is virtually impossible The reason is the small and compact design the use of SMD components and the price of the boards Therefore the components used on these boards cannot be ordered from Vital Scientific only the complete PCB s are available NOTE More and more components are verv sensitive for static electricitv Please take the necessarv precautions to avoid damage of the PC boards VITAL SCIENTIFIC 4 33 SERVICE MANUAL 5 MECHANICAL ADJUSTMENTS 5 MECHANICAL ADJUSTMENTS 5 1 INTRODUCTION The following paragraphs describe the mechanical adjustment procedures for the main mechanical units 5 2 ADJUST LAMP The lamp adjustment can be carried out with the help of the ADJUST LAMP function
86. LOCATION 6001 833 COVER CUVETTE ROTOR 10 3002 231 SCREW PLASTITE M4X12 17 3002 232 SCREW PLASTITE M4X16 18 3002 016 SCREW M3X6 19 3002 313 SCREW M3X16 20 3008 040 SCREW M4X5 21 3002 233 SCREW PLASTITE M4X30 22 3002 022 WASHER 23 3002 029 SCREW M4X16 24 3002 228 SCREW PLASTITE M3X16 25 3055 008 MAGNET STRIP 27 4802 269 PLATE 28 3002 224 SCREW PLASTITE M3X8 29 3002 227 SCREW PLASTITE M3X12 30 3022 010 WASHER 3 2 31 3070 347 HINGE 32 3004 210 SCREW M3X6 33 3002 026 SCREW M3X6 38 4803 318 TUBE PLATE 39 3070 343 SAMPLE COVER 40 3070 351 REAGENT COVER 41 6001 957 SAMPLE ROTOR 42 6001 784 REAGENT ROTOR 43 4107 051 FIXING SCREW 4 4100 476 BUSH 44 4100 210 BUSH 45 3002 318 SCREW M3X25 48 3002 329 SCREW M4X8 49 3019 011 NUT INSERT 50 3019 027 NUT M6X075 51 3022 025 WASHER 52 3370 525 SCREWLOCK FEMALE 53 4100 656 54 4100 657 55 4104 102 56 4501 523 PANEL 57 4803 452 58 6001 767 ISE INTERFACE BOARD 59 4100 555 ADAPTOR 60 ONLY WHEN ISE OPTION INSTALLED 9 4 SERVICE MANUAL VITAL SCIENTIFIC SERVICE MANUAL 9 MECHANICAL DRAWINGS ASSEMBLV OF COVERS VITAL SCIENTIFIC 9 5 9 MECHANICAL DRAWINGS SERVICE MANUAL TUBING DIAGRAM Length and position of tubing mav be different from the drawing because of changes in the design PART NUMBER DESCRIPTION LOCATION 1513 051 TUBE PVC 4X6 1 1513 050 TUBE PVC 2X4 2 6001 877 TUBE ASSY 3 6001 877
87. LOCK ASSV VITAL SCIENTIFIC 9 45 9 MECHANICAL DRAWINGS SERVICE MANUAL DILUTED WASTE CONTAINER 6001 826 PART NUMBER DESCRIPTION LOCATION 1513 002 PVC TUBE 4 6 CLEAR 1 3004 211 SCREW ST ST M3X8 2 3066 094 3 3066 105 NUT RED 4 3344 015 SWITGH LIQUID LEVEL 5 3351 028 CABLE CONN 6 3370 727 SLEEVE CODE 7 VELLOW 7 3370 728 SLEEVE CODE 8 VELLOW 8 3820 261 STICKER 9 3820 265 STICKER 10 4100 497 10 L CONTAINER 11 4100 668 PROP 12 4100 669 CAP 13 4104 104 PIPE 14 CONCENTRATED WASTE CONTAINER 6001 827 OPTIONAL PART NUMBER DESCRIPTION LOCATION 1513 002 PVC TUBE R 4X6 CLEAR 15 3004 211 SCREW ST ST M3X8 16 3066 094 CONNECTOR 17 3066 105 NUT RED 18 3344 015 SWITCH LIQUID LEVEL 19 3351 028 3P CABLE CONN 20 3370 720 SLEEVE CODE 0 YELLOW 21 3370 721 SLEEVE CODE 1 YELLOW 22 3370 729 SLEEVE CODE 9 YELLOW 23 3820 260 STICKER 24 3820 264 STICKER 25 4100 668 PROP 26 4100 669 CAP 27 4100 670 CONC WASTE CONTAINER 28 4104 104 PIPE 29 WATER CONTAINER 6001 860 PART NUMBER DESCRIPTION LOCATION 1513 001 PVC TUBE R2X4 30 3066 093 CONNECTOR 34 3066 107 NUT BLUE 35 3370 736 CODEMARKER 5 36 3370 737 CODEMARKER 6 37 3820 262 LOGO 38 3820 266 STICKER 39 4100 703 PROP 40 4100 497 CONTAINER 41 4104 080 PIPE 44 6001 620 WATERFILTER 45 9 46 VITAL SCIENTIFIC SERVICE MANUAL 9 MECHANICAL DRAWINGS DILUTED WASTE CONTAINER CONCENTRATED WASTE CONTAINER OPTIONAL WATER CONTAI
88. METAL 10K OHM 3302 533 RES METAL 22K OHM 3302 565 RES METAL 47K OHM 3304 105 RES 0 47 OHM 1W 3315 009 RES N W 4X4K7 A 4 x 1KO 3320 110 CAP CER AXIAL 1 nF 50V 3320 150 CAP CER 4 7NF 40V 3320 151 CAP AXIAL 4 7nF 50V 3320 233 CAP CER AXIAL 22nF 50V 3320 310 CAP CER AXIAL 100nF 50V 3320 942 CAP CERAMIC 33pF 100V 3326 521 CAP ELC 15uF 20V 3326 551 CAP ELC 47UF 20V 3326 554 CAP ELC 47uF 63V 3330 045 DIODE BYV 28 100 100V 3330 304 DIODE ZENER ZPD 3 6V 3331 004 TRANS BC 549 B 3333 544 L 6203 DMOS FULL BRIDGE DR 3335 004 IC 74 HCT 04 3335 032 IC 74 HCT 32 3335 193 IC 74 HCT 193 3336 008 284 00 280 CPU 3336 009 Z84 C20 4 PIO 3336 022 IC TC 5564 PL 15 3336 050 L 297 STEPPER MOTOR CONTR 3336 060 IC Z80 KIO 3336 063 MICRO MONITOR CHIP DS1232 3342 217 SWITCH PUSHBUTTON 3342 229 SWITCH PUSH BUTTON 3342 311 SWITCH SLIDE 0 161390 1 3342 516 SWITCH ROTARV HEX CODED 3348 191 FUSE 3 15A SLOW 3350 006 CONN FAST 2 8X0 5 3351 235 CONN 4POL 826467 4 3352 011 CONN 280611 2 3352 014 CONN 6 POLE 826467 6 3353 056 CONN 2X6 POL 826469 6 3353 340 CONN PCB DIL 28PF 3353 345 CONN PCB DIL 40 3353 384 CONN 84P PLCC 0 0821573 1 3355 002 KEYING PLUG 926498 1 3366 006 FUSE HOLDER PCB MOUNT FOR TR5 3370 116 RING 3376 362 MOTION CONTROL PC BOARD 3380 330 LED RED 3390 027 CRYSTAL 7 3728 Mc HC 18 3820 115 STICKER 4502 353 BEAM VITAL SCIENTIFIC 8 ELECTRICAL DIAGRAMS LOCATION R 7 29 38 R 10 24 R 11 27
89. NER VITAL SCIENTIFIC 9 47 9 MECHANICAL DRAWINGS 9 2 SPARE PARTS This chapter contains the spare parts that can not be found in the exploded views Power supply The power supply can only be exchanged as a whole There are no separate spare parts available for the power supplv 3359 039 Fan 6001 847 Cooling unit 2206 007 3913 034 3913 035 3913 045 3913 046 3913 036 3913 037 3913 100 3913 101 3913 102 3913 103 3913 104 power supply fan incl connector cooling liquid pump 230V 50Hz pump 115V 60Hz pump 220V 60Hz pump 110V 50Hz impeller 50 Hz pump impeller 60 Hz pump temperature control unit transformer 230 V 12 V transformer 115 V 12 V hosecoupling 90 hosecoupling 180 Optional parts See product guide 9 48 SERVICE MANUAL VITAL SCIENTIFIC SERVICE MANUAL 9 MECHANICAL DRAWINGS Reagent arm reagent needle urit 6002 008 ballbearing 3053 037 Sample arm bearing 3053 037 VITAL SCIENTIFIC 9 49
90. ODER OPTO UNIT ROTORS 6001 930 PARTNUMBER DESCRIPTION LOCATION 3002 016 SCREW ST ST M3X6 42 3019 011 NUT INSERT ST M3 43 4803 472 BRACKET 44 6001 900 ENCODER OPTO BOARD 45 9 36 SERVICE MANUAL VITAL SCIENTIFIC SERVICE MANUAL 9 MECHANICAL DRAWINGS MAIN ASSEMBLV MEASURING UNIT VITAL SCIENTIFIC 9 37 9 MECHANICAL DRAWINGS LAMP HOLDER MOUNTING PLATE ASSEMBLY 6001 884 PART NUMBER DESCRIPTION LOCATION 3002 015 SCREW ST ST M3X4 1 3002 016 SCREW ST ST M3X6 2 3002 027 SCREW ST ST M4X10 3 3004 226 SCREW ST ST M4X8 4 3022 015 WASHER ST ST 4 3 5 3023 006 LOCKWASHER 6 3060 053 LENS 08 7 3750 035 SPRING 8 4102 239 EXCENTER 9 4105 195 BUSH 10 4502 293 BEAM 11 4802 257 PLATE 12 4802 343 PLATE 13 LAMP HOLDER UNIT 6001 461 PART NUMBER DESCRIPTION LOCATION 3002 016 SCREW ST ST M3X6 14 3002 027 SCREW 5 5 M4X10 15 3023 006 LOCKWASHER 16 3023 010 WASH LOCK ST ST M4 17 3350 008 CONN SOLDER PIN CONTACT 18 3350 156 CONNECTOR LAMPSOCKET 19 3372 525 WIRE 7X0 4 WHITE 20 3380 018 Q I LAMP 12V 20W 21 4502 292 BEAM 22 4803 441 BRACKET 23 FILTER WHEEL WITH FILTERS 6001 460 PART NUMBER DESCRIPTION LOCATION 3029 041 O RING 26 3067 340 IFL FILTER 340 nm 27 3067 376 IFL FILTER 376 nm 28 3067 405 IFL FILTER 405 nm 29 3067 436 IFL FILTER 436 nm 30 3067 505 IFL FILTER 505 nm 31 3067 546 IFL FILTER 546 nm 32 3067 578 IFL FILTER 578 nm 33 3067 620 IFL FILTER 620 nm 34 3106 031 BEARING 5x8x10 35 4106 214 RING F
91. OR FILTERWHEEL 36 4700 540 FILTERWHEEL 37 BEAM SPLITTER BOARD 6001 818 PART NUMBER DESCRIPTION LOCATION 3002 016 SCREW ST ST M3X6 38 3004 226 SCREW ST ST M4X8 39 3059 015 WINDOW 40 4105 194 BUSH 41 4501 533 BLOCK 42 4509 124 BS BLOCK 43 4752 044 LEAF SPRING 44 6001 819 FILTER RESET BOARD 45 9 38 SERVICE MANUAL VITAL SCIENTIFIC SERVICE MANUAL 9 MECHANICAL DRAWINGS MOUNTING PLATE ASSY LAMP UNIT FILTER WHEEL WITH FILTERS BEAM SPLITTER BOARD VITAL SCIENTIFIC 9 39 9 MECHANICAL DRAWINGS PIPETTOR UNIT 6001 754 PART NUMBER DESCRIPTION LOCATION 3002 016 SCREW ST ST M3X6 1 3002 018 SCREW ST ST M3X8 2 3002 026 SCREW ST ST M4X8 3 3002 029 SCREW ST ST M4X16 4 3002 032 SCREW ST ST M4X35 5 3002 333 SCREW ST ST M4X12 6 3002 336 SCREW ST ST M4X16 7 3004 211 SCREW ST ST M3X8 8 3008 040 SCREW ST ST M4X5 9 3008 048 SCREW RVS 5 40 10 3011 016 NUT ST ST 4 11 3019 048 NUT M3X45 12 3020 017 CIRCLIP 4 13 3022 015 WASHER ST ST 4 3 14 3023 010 WASH LOCK ST ST M4 15 3053 001 BEARINGBALL 6X19X6 16 3064 016 BELT TIMING 80T T2 5 17 3066 070 SYRINGE 1ml 18 3066 071 SYRINGE 100ul 19 3066 072 VALVE DISPENSER UNIT 20 3130 069 SPROCKET T2 5 24 20 3130 070 SPROCKET T2 5 25 22 3750 033 SPRING 23 4100 410 BUSH FOR GUIDE SHAFT 24 4100 411 DRIVE NUT 25 4102 304 GUIDE SHAFT 26 4102 305 LEAD SCREW 27 4102 306 DRIVE PIN 28 4102 344 SHAFT 29 4105 215 BUSH 30 4107 065 KNOB FOR VALVE BLOCK 31 4107 066 KNOB FOR SYRINGE PL
92. RAM and or ROM of the System board slot 1 E02 SYSTEM EMERGENCY HALT Cause During rotation of the measurement rotor is detected that either the ISE arm the reagent arm the sample arm or the wash arm is not at the high position Probably touched The detection for the ISE arm and the wash arm is done by an opto coupler Solution Check wash arm opto 0 If occurs together with E144 ISE arm error see E144 E05 NO CLEAN CUVETTE Cause When according to the instrument there is no empty clean cuvette The bellows pump has stopped there was a system reset and the cuvettes were not emptied yet Solution Take out the cuvette rotor Empty and clean the cuvette rotor Reset the system and continue the measurements E07 SYSTEM RESET INCOMPLETE Cause When the system reset is not completed properly This error is always in combination with one or more other errors Solution Solve other errors and reset the system E10 NO VACUUM Cause Vacuum is below 0 4 bar for a period longer than 2 5 seconds Solution Check the vacuum pump and membranes Check Valve board in pump unit Check the vacuum tubing for leakage and clamping Check the valves on leakage Check and adjust if necessary the vacuum sensor VITAL SCIENTIFIC 7 5 7 ERROR HANDLING SERVICE MANUAL E11 E16 WASTE FULL CONCENTRATED WASTE FULL Cause When the WASTE FULL signal is high for at least 1 second Solution Check the floating switches reed relavs Check
93. TION The instrument has been designed as a fullv selective Patient Orientated Analvser svstem The chemistries include kinetic two point and endpoint determinations with both linear and non linear calculation modes The instrument has two modes mono mode for running mono reagents and dual mode for dual reagents The mono reagent mode has a maximum throughput of 180 tests hour The dual reagent mode has a maximum throughput of 133 tests hour 1 22 SCOPE AND ORGANIZATION This Service Manual is part of a two volume set The Instruction Manual is the first part of this set The Instruction manual provides detailed operating instructions technical specifications and some brief trouble shooting instructions The Service manual has been prepared for service technicians and other technical personnel directiv involved in maintaining and repairing the instrument The reference information is provided in 10 separate sections Section 1 General This section provides general information about the service and repair policv warrantv spare parts technical assistance and service trainings Section 2 Functional design This section provides information about the functional design of the analvser This includes a short description of the functional units like sample rotor washing unit monitor etc a description of the mechanical and optical design and the fluid system Section 3 System organisation This section describes the system o
94. TION AND MAINTENANCE Req tray buffer External 18 232 ON Display mode Valve Valve Valve Valve Valve 5 Valve 12 Valve 13 Valve 14 Vacuum pump Vacuum sensor V 18 LINE Funct Check Adj off off off off off off off Vacuum checked The current state of the valves pump and vacuum detector is displayed Valve V13 and V14 are for the concentrated waste option VITAL SCIENTIFIC 4 27 4 INSTALLATION AND MAINTENANCE SERVICE MANUAL 4 6 10 Water Svstem Analyser status Inactive Req tray buffer 1 18 Run mode DUAL External 88 232 ON LINE Display mode Funct Check Adj Adjust Lamp Reagent Arn Valve off Sample Arw Valve off Reagent Disc Valve Sample Disc Valve 9 off Measurement Disc Filter Valve 18 off Mash Arm Valve il off ISE Arm ISE Unit Bellows pump unknown Pipettor Mater pump Vacuum sustem Pump up if Pump opto dom 8 Optical electronics Haste sensor not full Electronics Conc waste sensor not full Barcode scanner Mater running out no Insuff water no Taba alve UCZE Reset Water AT Pump Pump Fi F2 B 4 hi Fo FT 9 In this menu all the functional parts of the water svstem can controlled and checked The active state of the valves bellows pump with opto s water pump waste sensor concentrated waste sensor and the level
95. TRL SCIENTIFIC VITAL SCIENTIFIC SERVICE MANUAL CABLE FILTER BOARD 6001 770 PART NUMBER 3348 193 3351 216 3353 056 3353 401 3355 002 3366 006 3376 470 3820 115 DESCRIPTION FUSE 4A SLOW CONN 280610 2 CONN 2X6 POL 826469 6 CONN 25PF AMPLIMITE HD 20 KEYING PLUG 926498 1 FUSE HOLDER PCB MOUNT FOR TR5 CABLEFILTER BOARD STICKER 8 ELECTRICAL DIAGRAMS LOCATION F1 PL 1 PL2 1 VITAL SCIENTIFIC SERVICE MANUAL 8 ELECTRICAL DIAGRAMS TEMPERATURE BOARD 6001 771 PART NUMBER 3002 018 3302 285 3302 317 3302 341 3302 349 3302 381 3302 401 3302 405 3302 409 3302 433 3302 445 3302 449 3302 465 3302 469 3302 473 3302 481 3302 493 3302 501 3302 505 3302 509 3302 513 3302 517 3302 529 3302 545 3302 565 3302 601 3302 665 3302 733 3304 032 3306 474 3306 504 3306 622 3312 106 3312 108 3312 117 3312 132 3320 110 3320 233 3320 310 3323 321 3326 431 3326 521 3326 551 3326 613 3326 615 3330 010 3330 041 3330 303 3330 308 3330 318 3331 004 3331 415 3333 018 3333 019 DESCRIPTION SCREW ST ST M3X8 RES METAL 75 OHM RES METAL 150 OHM RES METAL 270 OHM RES METAL 330 OHM RES METAL 680 OHM RES METAL 1K OHM RES METAL 1K1 OHM RES METAL 1K2 OHM RES METAL 2K2 OHM RES METAL 3K OHM RES METAL 3K3 OHM RES METAL 4K7 OHM RES METAL 5K1 OHM RES METAL 5K6 OHM RES METAL 6K8 OHM RES METAL 9K1 OHM RES METAL 10K OHM RES METAL 11K OHM RES METAL 12K OHM RES M
96. UNGER 32 4107 079 SCREW 33 4501 424 CARRIAGE 34 4501 487 BLOCK 35 4502 334 BEAM 36 4502 335 BEAM 37 4502 356 STRIP 38 4509 132 PLATE 39 4509 169 PLATE 40 4509 180 PLATE 41 4802 227 VANE ON CARRIAGE 42 6001 459 DISPENSER OPTO BOARD 43 6001 769 DISPENSER DRIVER BOARD 44 6001 780 VALVE DRIVE UNIT 45 6001 812 VALVE MOTOR CABLE 46 6001 813 STEPPERMOTOR DISPENSER 47 4105 214 BUSH DILUTER VALVE 48 3916 030 RING 49 9 40 SERVICE MANUAL VITAL SCIENTIFIC SERVICE MANUAL 9 MECHANICAL DRAWINGS PIPETTOR UNIT VITAL SCIENTIFIC 9 41 9 MECHANICAL DRAWINGS VALVE DRIVE UNIT 6001 780 PART NUMBER DESCRIPTION LOCATION 3002 019 SCREW ST ST M3X10 2 3002 023 SCREW ST ST M3X20 3 3004 205 SCREW ST ST M2X8 4 3008 040 SCREW ST ST M4X5 5 3008 041 SCREW ST ST M4X5 6 3020 017 CIRCLIP 4 8 3022 010 WASHER ST ST 3 2 9 3022 018 SHIM RING 10 3022 023 SHIM RING 11 3053 001 BEARINGBALL 6X19X6 12 3053 002 BEARINGBALL 3X10X4 13 3053 019 BEARINGBALL 5X1 1X4 14 3130 062 GEAR 15 4100 408 BUSH 17 4100 409 BUSH 18 4102 269 SHAFT 19 4102 302 SHAFT 20 4102 303 SHAFT 21 4106 203 DISC 22 4501 422 BLOCK 23 4501 519 BLOCK 24 4803 297 STRIP 25 6001 781 MOTOR UNIT U D MOVEMENT 27 1513 001 PVC TUBE R 2X4 CL 28 3002 001 SCREW ST ST M2X4 29 3002 003 SCREW ST ST M2X6 30 3002 016 SCREW ST ST M3X6 31 3002 307 SCREW ST ST M3X8 32 3011 011 NUT ST ST M3 33 3020 01 1 CIRCLIP 2 3 34 3022 005 WASHER ST ST 2 2 35 3053 002 BEARINGBALL 3X10X4 36 336
97. UVETTE ROTOR AND WASH POSITIONS The filling of the cuvette rotor is done bv means of a bellows pump that is driven bv a stepper motor The volume is 4 x 500 ul per rotation cvcle In total 4 cuvettes are washed filled and emptied at the same time Also the sample needle is washed with 500 ul water at the wash position When the water pump is switched on water will be pumped from the water bottle to the water cask The water cask is equipped with a svstem for detecting the water level see section 2 4 3 When the water level is too low the water pump will be switched on and when the level is high enough the pump will be switched off When the switch off circuit is defective the water will flow back into the water container via the overflow connection The filling of the cuvette rotor and sample wash position is performed as follows Valve V6 to V10 are not activated the bellows goes downwards and is aspirating water Then V6 to V10 are activated and the bellows goes upwards and is dispensing water in the cuvette rotor and sample wash position and the valves V6 to V10 are de activated again The reagent svringe fills the reagent wash position The reagent valve will move to the aspirate position and the syringe plunger will go downwards and the syringe will be filled with water Then the valve will move to the dispense position and the plunger will go upwards and water will be dispensed into the reagent wash position 2 6 2 EMPTYING THE
98. VdS 219193025 WAA 102116 ITE 3 bib WYKIS JWIS Winvy 23030 ON 19 SIJVIdIU 3002 Wisil O33 imn ENS QIINYIA SYIONV l IZN31 S31NY23101 031 15 JSIMUIHLO 102000 928066 916 VITAL SCIENTIFIC SERVICE MANUAL 8 ELECTRICAL DIAGRAMS 130 A3 SPLACES 6001 1 1 C SWEET PHOTOMETER BOARD O Lie z LU 4 l c gt VITAL SCIENTIFIC SERVICE MANUAL PLL BOARD 6001 763 PART NUMBER 3302 401 3302 501 3302 509 3302 529 3312 261 3315 267 3315 270 3320 310 3323 214 3323 321 3323 350 3326 521 3331 004 3333 023 3333 513 3350 006 3352 014 3352 082 3353 056 3355 002 3376 333 3820 115 otion Ctrl Brd DESCRIPTION RES METAL 1K OHM RES METAL 10K OHM RES METAL 12K OHM RES METAL 20K OHM POTM ADJ 1M 0 5W TRANS VOLTAGE SUPP 14 5V TRANSIENT SUPPRES 33 3V CAP CER AXIAL 100nF 50V CAP 0 01uF 63V POLY CAP POL 0 15 uF 63VDC 20 CAP P E 0 47UF 63V CAP ELC 15uF 20V TRANS BC 549 B IC AD741LN IC MONOLITHIC TONE DECODER CONN FAST 2 8X0 5 CONN 6 POLE 826467 6 CONN 9P HD 20 WITH FILTER CONN 2X6 POL 826469 6 KEYING PLUG 926498 1 PLL PC C BOARD STICKER 8 ELECTRICAL DIAGRAMS LOCATION 4 12 1 3 5 9 11 13 7 15 6 8 14 16 1 2
99. Vitalab Flexor E Service Manual Vitalab Flexor Service Manual 6002 190 450 date of release 1 Februarv 2000 version 1 VITAL SCIENTIFIC N V Service Manual Vitalab Flexor E Service Manual Preface The contents of this manual with all figures tables and graphic representations are intellectual propertv of Vital Scientific N V Unauthorised commercial or non commercial excerption or copving of contents and use of this manual ion total or in parts are strictly forbidden unless the editor gives written permission for it This manual was written and reproduced with the utmost care However errors cannot be fullv excluded Vital Scientific N V does not take anv responsibilities and accepts no liabilities for incidents of anv kind that mav occur because of errors in the manual All products names that are mentioned in this manual are registered trademarks The manual describes the analvser svstem Vitalab Flexor E The Vitalab Flexor E was conceptualised manufactured and tested in accordance with the declaration of conformitv The declaration of conformitv is supplied with each device in a separate file Vital Scientific N V P O Box 100 6950 AC Dieren Van Rensselaerweg 4 6956 AV Spankeren The Netherlands 31 313 430 500 This manual was conceptualized written and lavouted Vital Scientific Service Manual VITAL SCIENTIFIC N V SERVICE MANUAL 1 GENERAL 1 GENERAL 1 1 INTRODUC
100. W ST ST M3X6 NUT ST ST M3 RES METAL 4 7 OHM RES METAL 33 OHM 0 6W RES NETWORK 10 KOHM 10 X CAP CER 1NF 40V CAP CER 4 7NF 40V CAP CER 0 022UF 50V CAP CER 0 1uF 50V CAP CER 220nF 63V CAP ELC 15uF 20V DIODE BYV 28 100 100V PROTECTED POWER MOS FET VOLTAGE REGULATOR 5V 100mA L 6202 DMOS FULLBRIDGE DR 1 5A IC 74 04 FUSE 3 15A SLOW CONN 2PM MOD 1 RIGHTANGLE CONN 4 POLE 280616 2 CONN 280617 2 CONN 12POL 826470 6 KEYING PLUG 926498 1 FUSE HOLDER PCB MOUNT FOR TR5 VALVE BOARD VITAL SCIENTIFIC 8 ELECTRICAL DIAGRAMS LOCATION R1 R 2 7 RN 1 C 1 7 C 17 22 C 5 9 10 12 13 15 16 3 4 C2 8 11 14 C6 D 1 T 1 IC 1 IC 3 5 IC 2 F1 PL 2 5 10 PL 1 4 11 12 11 12 1 SERVICE MANUAL 8 ELECTRICAL DIAGRAMS Comp ident como side Tid ass R2 ULTI e e 819914 LC R3 9 ld Hi ptt z 61d ER 162 015 T K lt Ri Z SI FI 1 EM 3376 433 P12 lt EE SAW PLZ VITAL SCIENTIFIC 8 ELECTRICAL DIAGRAMS SERVICE MANUAL SONUISINLIN N33310 231411N3125 Turin zza u 69 Ez leso vey rz esci su 066 eusense s s9 p Cza 6 L sa 8697 n D a
101. ally when the reagent arm is obstructed by something or when the safety switch inside the reagent needle is activated Perform a selective reset of the reagent arm Check when this error is occurring frequently the operation of the reagent needle switch and clean if necessary the contacts of the switch only mechanical safety switch Check the operation of the opto couplers and replace if necessary VITAL SCIENTIFIC SERVICE MANUAL 7 ERROR HANDLING E83 REAGENT ARM VERTICAL ERROR Cause Solution This error occurs when the opto reading of the vertical movement of the reagent arm is not in accordance with the current theoretical position This happens normallv when the reagent arm is obstructed bv something or when the safety switch inside the reagent needle is activated Perform a selective reset of the reagent arm Check when this error is occurring frequently the operation of the reagent needle switch and clean if necessary the contacts of the switch only mechanical safety switch Check the operation of the opto couplers and replace if necessary E92 REAGENT INIT FAILED see E95 E95 REAGENT ARM HORIZ RESET FAILED Cause Solution This error will be generated when the motion control board does not report the horizontal reset position after the reset command Check if the communication between system board and motion control board is present When there is no communication this can either be caused by
102. avana z vs 814 8221 8221 sau u 2 8 sl 13 z ra T p T SZ b u s9 uzz ELI Tia ZNP 11 813 m uzz pe p 223 63 914 kal uw d n 2 LLI Y G E za 3231 o u 9 a vez Ned sere O ELI 2 5 vezz J Et 4 22 23 813 330 ino m lt 4431 T T 3 2 SERVICE MANUAL DISPENSER DRIVER BOARD 6001 769 PART NUMBER 3002 016 3011 011 3302 120 3302 373 3302 501 3320 232 3320 312 3333 545 3351 217 3351 234 3353 130 3355 002 3376 511 3381 053 3820 115 DESCRIPTION SCREW ST ST M3X6 NUT ST ST M3 RES METAL 4 7 OHM RES METAL 560 OHM RES METAL 10K OHM CAP CER 0 022UF 50 220nF 63V L 6202 DMOS FULLBRIDGE DR 1 5A CONN 4 POLE 280616 2 CONN PLUG 4POL 90 826468 4 CONN 22PM MOD2 RIGHTANGLE KEYING PLUG 926498 1 DISPENSER OPTO BOARD OPTO INTERRUPTOR H21B1 STICKER Comp ident comp side 8 ELECTRICAL DIAGRAMS LOCATION R2 3 R 1 6 R 4 5 C 1 4 IC 1 PL 1 PL 2 4 PL 3 OC 1 4 VITAL SCIENTIFIC SERVICE MANUAL 8 ELECTRICAL DIAGRAMS 2 Oucc 3538 oare 23 11 94 TOLERANCES UNLESS OTHERUISE ai E DISPLACES DIMENSIONS ANGLE ROUGHNESS Ra M SEL NO DESCRIPTION DISPENSER OPTO BOARD VITAL SCIENTIFIC DIEREN THE NETHERLANDS 6001 769 VI
103. ction 7 2 of ISE service manual NO SAMPLE E153 ISE SANPLE DETECTION ERROR Cause This error will occur when sample is detected at the wrong moment Solution See section 7 2 of ISE service manual SAMPLE DETECTION VITAL SCIENTIFIC 7 21 7 ERROR HANDLING SERVICE MANUAL E154 ISE VALVES ERROR Cause This error will a occur when one of the valves is not working correctiv Solution See section 7 2 of ISE service manual CHECK VALVES E155 ISE ELECTRODES OR FLUIDS ERROR Cause This error will occur after a calibration when one or more electrodes cannot be calibrated Solution See section 7 2 of ISE service manual CHECK ELECTRODES or NA K CLNOT CALIBRATED E157 ISE POWER FAIL Cause A power fail of shorter then 30 seconds is detected Solution Check the power and continue measurements 7 1 11 PC ERRORS E160 Z80 MASTER CARD RS 232 ERROR Cause This error will be generated when there are too much retransmissions more than 12 a message sent by the PC to the system board If a message sent bv the PC to the svstem board is not acknowledged bv the Z80 within 5 seconds it is retransmitted Solution Communication line is not ok 280 main board is Replace the board or replace the defective components Check the wiring E161 Z80 MASTER CARD RS 232 ERROR Cause This error will be generated if a message from system board to is not acknowledged the PC according to the svstem board or
104. ction is now on push and at a time for 3 seconds return to normal operating mode Vital Scientific will take no responsibilitv for damaged cooling unit or analvser because of too low temperature setting or because of the use of an inadequate cooling liquid VITAL SCIENTIFIC 2 23 SERVICE MANUAL 3 SYSTEM ORGANIZATION 3 SYSTEM ORGANISATION 3 1 INTRODUCTION A general svstem overview is shown in figure 3 1 EXTERNAL COMPUTER RS232 RS232 RS232 COM1 driver driver driver EXTERNAL CONNECTIONS 1 SYSTEM BOARD PL8 PL10 PL9 PL2 PL3 PL9 211 PLS PHOTO MOTION MOTION MOTION CONTROL A2 CONTROL CONTROL 4 METER TEMP CONTROL BOARD BOARD A9 A5 PIPETTOR SAMPLE WASH UNIT MOTORS BELLOW PUMP MOTORS 7 8 TA qe EN MOTION MOTION MOTION CONTROL A6 CONTROL CONTROL MEASURING REAGENT SAMPLE ROTOR REAGENT FILTER MOTORS ROTOR WHEEL 9 10 11 12 Figure 3 1 System overview VITAL SCIENTIFIC 3 1 3 SYSTEM ORGANIZATION SERVICE MANUAL The communication between the external computer and the Svstem board is realized by means of
105. d the direction of the stepper motor This wav loss of steps can be detected These are the stepper motors that are equipped with an encoder disc Sample Rotor Reagent Rotor Cuvette Rotor Sample Arm horizontal movement e Reagent Arm horizontal movement 1 A 0 1 gt SIER 0 Em 0 9 1 1 1 0 1 0 0 1 0 Figure 2 7 Motor encoders The phase relation between the two encoder signals should be around 90 degrees otherwise the software cannot detect all transitions and loss of steps will be reported even if no actual loss of steps has occurred 2 14 VITAL SCIENTIFIC SERVICE MANUAL 2 FUNCTIONAL DESIGN 2 4 DETECTION SYSTEMS The instrument is equipped with the following detection systems e sample level detection reagent level detection water buffer level detection overflow detection for cuvette waste full detection 2 4 1 SAMPLE LEVEL DETECTION The sample arm is equipped with a detector that will detect the liquid level of the sample and will given an error when not enough sample is present in the cup SAMPLE SAMPLE 10V top top 3 f 13 KHz OSCILLATOR SIGNAL Figure 2 8 Sample level detection VITAL SCIENTIFIC 2 15 2 FUNCTIONAL DESIGN SERVICE MANUAL The pri
106. d and motion control board is present When there is no communication this can either be caused by the wiring the SIO on the system board the KIO on the motion control board or the main PCB fuse on the motion control board Replace the board s or replace the defective components Check if there is a mechanical obstruction of the sample disc and repair if necessary Check if the opto couplers are working correctly and repair if necessary E117 REAG SAMP DISCS COMMUNICATION ERR Cause Solution This error will be generated when the communication to motion control board for the reagent or sample disc stalls or if an unknown character is echoed Normally this error will occur when the communication is disrupted or distorted Check if the communication between system board and motion control board is present When there is no communication this can either be caused by the wiring the SIO on the system board the KIO on the motion control board or the main PCB fuse on the motion control board Replace the board s or replace the defective components VITAL SCIENTIFIC 7 17 7 ERROR HANDLING SERVICE MANUAL 7 1 9 WASH ARM ERRORS E120 WASHARM BELLOWS PUMP 14V FAILED Cause This error will be generated when the circuit see sec 8 of this manual motion control board sheet 1 4 diode 025 does not detect the correct voltage Solution Check the fuses and voltage on the motion control board Check connectors PL3 PL4
107. e and spectral response of the detector are much higher in the visible and near infra red range than in the near ultra violet range The filament of the lamp is projected in the lens in front of the filterwheel After the filterwheel a beam splitter is mounted which splits the light beam 90 of the signal is going through the second lens and cuvette and 10 of the signal is going into the direction of the reference detector The function of the reference detector is to detect and to compensate for the fluctuations of the lamp The optimal alignment and lamp intensitv can be reached bv adjusting the lamp VITAL SCIENTIFIC 2 19 2 FUNCTIONAL DESIGN SERVICE MANUAL 2 6 FLUID AND VACUUM SYSTEM Figure 2 12 shows an overview of the main parts of the fluid and vacuum svstem VACUUM DETECTOR SELECTRA CONNECTION PLATE PUMPUNIT 4522 VACUUM CASK IP 25 4 3 VACUUM 2 M EM PUMP 2 3 2 lt lt gt gt gt 8 1 2 b p lt 2 4 2 DRY BLOCK WATER gt amp gt PUMP COLLECTING 4 Lp CASK C W COLLECTING 2 3 7 4 1 6 4 5 CASK 1 2 3 n 5 6 CUVETTE ROTOR gt WATER CASK 1 1
108. e errors can be divided into two groups e flagerrors e hardware errors 7 1 1 FLAG ERRORS The flag errors are printed behind the result of a certain test An explanation of the flag is given in the EVALUATE SAMPLES menu The following flag errors can occur GENERAL HARDWARE ERROR INSUFFICIENT REAGENT INSUFFICIENT SAMPLE LAMP UNDERRANGE LAMP OVERRANGE LAMP REF CHANNEL UNDERRANGE LAMP REF CHANNEL OVERRANGE MEAS UNDERRANGE MEAS OVERRANGE MEAS REF UNDERRANGE MEAS REF OVERRANGE CUVET TEMPERATURE ERROR RERUN LINEARITY ERROR ABSORBANCE LIMIT ERROR REAGENT ABSORBANCE DEVIATION ERROR REAGENT BLANK LIMIT ERROR CALIBRATION LIMIT ERROR CONTROL LIMIT ERROR REFERENCE LIMIT ERROR REAGENT ABORBANCE ERROR The errors marked with are also displaved the screen a window can be recognised bv an audible alarm repeating four short beeps Pressing anv kev can stop the audio alarm These errors are normally caused by a hardware error and are explained hereafter The other errors are normally caused by an another error programming or chemical error and are explained in the instruction manual VITAL SCIENTIFIC 7 1 7 ERROR HANDLING SERVICE MANUAL GENERAL HARDWARE ERROR This error will be displaved when one of the hardware errors mentioned in the next section has occurred and the user ignores this error The best way to find this error is to make a pri
109. e extreme left CCW position very insensitive Put a dry and empty Sample tube in the Rotor on position 2 surrounded by emptv paediatric adapters on position s 1 3 E3 4 W B and S1 Select in the service menu sample arm and reset the sample arm F1 Turn the sample arm into the Sample tube on position 2 5 5 4 The liquid detection has to be NO Turn P2 to the right CW position until the liquid detection changes to a YES Now turn P2 to the left CCW until the bit changes again to a Turn P2 360 one complete turn to the left CCW REAGENT ROTOR Turn P1 to the extreme left CCW position very insensitive Put an empty 5ml bottle on position 29 surrounded by water filled bottles on position s 15 24 25 26 27 29 30 31 32 1 and 16 Select in the service menu reagent arm and reset the reagent arm F1 Turn the reagent arm into the empty bottle on position 28 1151 14 11 The liquid detection has to be NO Turn P1 to the right CW position until the liquid detection changes to YES Now turn P1 to the left CCW until the bit changes again to a Turn P1 360 one complete turn to the left CCW Check also for some other positions if the liquid sense is working well 6 4 VITAL SCIENTIFIC SERVICE MANUAL 7 ERROR HANDLING 7 ERRORHANDLING 7 1 ERROR CODE LIST In the analvser a large number of error messages are present Th
110. e syr down opto Taba Reset 1 1 Service Pipett 11 1 FI r2 F5 Fo F7 Fa 9 First press reset F1 to activate the pipettor functions and the following function kevs are active Switch the reagent valve to the water position F4 Switch the reagent valve to the reagent needle position F5 Switch the sample valve to the water position F6 Switch the sample valve to the sample needle position F7 Program the reagent syringe position in motor steps 0 to 1000 steps F8 Program the sample syringe position in motor steps 0 to 1000 steps The following states are displayed The current positions of the reagent and sample valve The current positions of the reagent and sample syringe The state of the reagent and sample valve opto switches The state of the syringe reset opto s The state of the syringe down position opto s 4 26 VITAL SCIENTIFIC SERVICE MANUAL 4 6 9 Vacuum System Analyser status Run mode Taba a R M Pi Pu l FI Pump Fi F2 H T5 Fo F7 FR rg In this menu all the functional parts of the vacuum svstem can be controlled and Inactive DUAL Adjust Lamp Reagent Arn Sample Arm Reagent Disc Sample Disc Measurement Disc Filter Mash Arm ISE Arm ISE Unit Pipettor Water system Optical electronics Electronics Barcode scanner 4 INSTALLA
111. easiest wav to have access to the tubing of the ISE pick up arm is bv removing the power unit 2 Mount the new tubing in the same way as the old tubing was mounted 4 10 VITAL SCIENTIFIC SERVICE MANUAL 4 INSTALLATION AND MAINTENANCE 4 5 2 4 REVISION OF VACUUM PUMP 1 Loosen the eight socket nuts at the pressure and suction side of the pump 2 Remove the pumphead and gaskets Washer 3 Remove the old membranes 4 Mount the new membrane as shown in the figure left Vou have to lock the nut bv means of using Locktite 243 5 Mount the pumphead by tighten the socket nuts very equally like the nuts of cylinder head of a car 6 Check the vacuum level and vacuum Black side detector bv following the adjustment procedure described in section 5 10 In case of wear the plungers you can order a separate plunger kit which includes the ball bearing and a mounting description part nr 6002 041 In case of wear of the chambers on the pump head a separate chamber kit is available part nr 6002 042 Please note that this and the former part should alwavs be replaced in pairs 4 5 2 5 REVISION OF WATER PUMP 1 Loosen the four screws of the pump 2 Remove the old pump head valves and gasket and unscrew the old membrane of the plunger 3 Place the new revision set including membrane rubber mini valves and housing 4 Tighten the four screws again 4 5 2 6 REPLACEMENT OF MIXER BELTS Remove t
112. ect the Two Absorbance testbox to PL3 Connect and the power cable to TAB3 15V and push BLANK F2 Read the LAMP MEAS value on the screen and write down Press the button on the Two Absorbance testbox and keep it depressed Read the LAMP MEAS value again and write it down The difference must be 20000 100 if not adjust with 1 Connect the Two Absorbance testbox to PL4 Connect the power cable to 15V Select BLANK F2 Read the cuvette measurement value on the screen and write it down Press the button on the Two Absorbance testbox and keep it depressed the difference for the cuvette measurement value must again be 20000 100 Mount the metal shielding Connect PL3 and PL4 again Close slot A9 VITAL SCIENTIFIC SERVICE MANUAL 6 ELECTRICAL ADJUSTMENTS 6 4 TEMPERATURE ADJUSTMENT CUVETTE ROTOR Select in the service menu ELECTRONICS page 3 In this menu the complete description for the adjustment is given Temperature adjustment of cuvette rotor First check if the cuvette detection is correct in the service menu The status of the analyser should be INACTIVE because when the status is HALTED the Peltiers of the cuvette rotor are not activated The screen must show CUVETTE TEMPERATURE OK This adjustment must be done in the main menu You have to run a wash and fill cycle During the run all the covers must be mounted and the cuvette cover must placed After the wash and fill
113. ectronics Liquid sense Electronics Needle switch Barcode scanner down bit Submerge depth Taba Reset ELIE Start Edit Service SampArn 11980011 01910017 501110 Fi F2 F3 M T5 Fo T9 First press reset F1 to activate the sample arm function and the sample arm can be moved with the cursor kevs The active state of the following functions are displaved Horizontal reset opto switch Encoder opto switches Vertical reset opto switch Liquid sensor Needle switch Max down bit Pressing F3 disables the horizontal stepper motor and the vertical motor can be disabled with F4 The stirrer motor can be started and stopped with respectivelv F5 and F6 4 20 VITAL SCIENTIFIC SERVICE MANUAL 4 INSTALLATION AND MAINTENANCE 4 6 3 Reagent Disc Analyser status Inactive Req tray buffer 1 18 Run mode DUAL External 88 232 ON LINE Display mode Funct Check Adj Adjust Lamp Reagent Arm Sample Arw Disc position Sample Disc Measurement Disc Filter Mash Arm ISE Arm ISE Unit Pipettor Reset opto Vacuum system Encoder opto Water system Encoder opto B Optical electronics Electronics Barcode scanner Taba Reset Edit Disable Service Reagbscm BS Drive Fi F2 r3 4 F5 Fo F 9 First press reset F1 to activate the reagent disc function and then reagent disc position 32 positions can be edited with F2 The active state o
114. ed This adjustment is as follows 1 Loosen screws D at top of rotor 2 Turn the inner rotor assembly slowly clockwise or counter clockwise and fix housing on position with the highest reading on CUVET 3 Re adjust lamp position for maximum signal on CUVET D c lE 040 0 NE Pa 7 gt Z 7 gt 2 AL AA Figure 5 2 Photo sensor adjustment Note So far when adjusting the lamp and or sensor position for maximum signal at 340 nm it does not matter whether there is a cuvette rotor placed or not or if this cuvette rotor is empty or filled with water For the next check a water filled cuvette rotor has to be placed The condition of the filters can be checked by pressing F1 Check Cuv Abs and the absorbance is measured for each wavelength 340 to 620 nm The absorbance values are graphicallv displaved on the right side of the screen The absorbance must be at least as high as the minimum level The actual absorbance and counter values are not displaved in this menu but are shown in a special menu called OPTICAL ELECTRONICS VITAL SCIENTIFIC 5 3 5 MECHANICAL ADJUSTMENTS SERVICE MANUAL 5 3 CUVETTE ROTOR ADJUSTMENT In principle this adjustment has been done at the factorv of Vital Scientific Only when the measurement rotor has been disassembled it is necessary to check the lightpath and in some cases to re adjust it It has no sense to re adjust unless both adjust
115. ed the config sys file For further information check your DOS manual 4 6 VITAL SCIENTIFIC SERVICE MANUAL 4 INSTALLATION AND MAINTENANCE 4 5 PREVENTIVE MAINTENANCE The analvser is designed to operate with a minimum of maintenance Repair equipment is kept to a minimum and common tools used in most cases A list of special tools and other support equipment can be found in the price list To assure a high MTBF Mean Time Between Failure and an optimal operation a program of scheduled preventive maintenance is required The schedule is detailed in the following paragraphs Note that the maintenance that can and should be performed by the operator is detailed in the Instruction Manual 4 5 1 SCHEDULED MAINTENANCE The maintenance described in this paragraph must be performed every 6 months Required equipment 2 Abs test box 4 digit Digital voltage meter Vacuum meter 0 1 bar vacuum Temperature meter max tolerance of 0 1 C VITAL SCIENTIFIC 4 7 4 INSTALLATION AND MAINTENANCE SERVICE MANUAL Required parts for preventive maintenance Everv 6 months PARTNUMBER DESCRIPTION 3013049 3064 041 mixing belt 3066 071 100 ul syringe 3066 070 1 ml syringe 6001 405 tubing set for reagent probe 6001 428 tubing set for sample probe 3066 083 filter water tank 6001 963 wash block Every 12 months PARTNUMBER DESCRIPTION Every 24 months PARTNUMBER DESCRIPTION 6002 042 chamber kit vacuum pump
116. einise procedure VITAL SCIENTIFIC 4 17 4 INSTALLATION AND MAINTENANCE SERVICE MANUAL Error historv In this menu vou can inspect all the errors that occurred error code error description date amp time of occurrence operator action after the error If vou select Functional Check the following menu will be displaved Analyser status Stand by Req tray buffer 1 18 Run mode DUAL External 88 232 ON LINE Display mode Funct Check Adj Adjust Lamp Reagent Arn Sample Arw Reagent Disc Sample Disc Measurement Disc Filter Mash Arm ISE Arm ISE Unit Pipettor Vacuum system Water system Optical electronics Electronics Barcode scanner Taba Special Funct Fi F2 r3 F4 hi Fo FT T3 In this menu a number of functional checks can be selected like adjust lamp reagent arm sample arm etceteras The different checks can be selected bv moving the cursor with the cursor kevs and press Enter For the ADJUST LAMP procedure see chapter 5 4 18 VITAL SCIENTIFIC SERVICE MANUAL 4 6 1 Reagent Arm Analyser status Inactive Run mode Taba DUAL Adjust Lamp Sample Arw Reagent Disc Sample Disc Measurement Disc Filter Mash Arm ISE Arm ISE Unit Pipettor Vacuum system Water system Optical electronics Electronics Barcode scanner 4 INSTALLATION AND MAINTENANCE Req tray buffer 1 18 External 28 232 ON LINE Display mode Funct Check Adj Horizontal
117. en in section 5 3 of this manual Check if the opto couplers are malfunctioning and repair if necessarv E135 WASH ARM RESET FAILED Cause Solution This error will be generated when the wash arm motion control board does not report the correct wash arm position within 5 seconds after the reset command Check if the communication between svstem board and motion control board is present When there is no communication this can either be caused by the wiring the SIO on the system board the KIO on the motion control board or the main PCB fuse on the motion control board Replace the board s or replace the defective components Check if there is a mechanical obstruction of the wash arm and repair if necessarv Check if the opto couplers are working correctiv and repair if necessarv VITAL SCIENTIFIC 7 19 7 ERROR HANDLING SERVICE MANUAL E136 BELLOWS PUMP RESET FAILED Cause This error will be generated when the motion control board does not report the correct bellow position within 25 seconds after the reset command Solution Check if the communication between svstem board and motion control board is present When there is no communication this can either be caused by the wiring the SIO on the system board the KIO on the motion control board or the main fuse on the motion control board Replace the board s or replace the defective components Check if there is a mechanical obstruction of the bellow pump and
118. ent 5 10 Reset the arm F1 Loosen screw C of the clamp assembly A and apply some new Locktite In the service menu move the arm to the wash position only horizontally Disable vertical drive F4 Push the arm down by hand just above the wash position and adjust the arm so that the needle is in the centre of the wash position Firmly push ring G block E and clamp assembly A against each other and fix screw C very well VITAL SCIENTIFIC SERVICE MANUAL 5 MECHANICAL ADJUSTMENTS 12 Reset the arm and check the needle in every position cuvette wash position and cups bottles Pressure release screw for guiding bearings By means of the screw in the clamp assembly you can release the pressure of the guiding bearings on the shaft After fastening the clamp assembly always check this pressure e Disable the vertical drive of the arm e Check the pressure of the bearings on the shaft by moving the arm up and down by hand e 5 movement is difficult release the pressure by turning the little screw clockwise and check again e Do not release the pressure too much this causes play CHECK THE MIXER POSITION 1 Reset reagent arm F1 and move reagent arm to the mixer position 2 Start the mixer F5 and check if the needle touches the sides of the cuvettes If it touches noisy e Loosen the 2 screws on top of the mechanical assembly with which the mixer assembly is mounted in the mechanical assembly e Shift t
119. et it right e Exchange the lamp Check lamp voltage 12 V e Check sensor voltages 2100 mV Exchange Optometer board e Exchange the cuvette rotor e lf the results are bad only for one filter exchange this filter e lf the results are bad for certain cuvettes at all wavelengths e Take out the cuvette rotor and put it back 90 degrees turned After running a new cuvette blank you can see if the bad cuvettes shifted 12 places or not If not e Check lightpath and lenses on dirt and hairs e Check that the light beam goes through the centre of the cuvettes with adhesive transparent tape sticked to the cuvette wavelength 546 nm If not change the setting of the hexacode switch on MCB A6 7 26 VITAL SCIENTIFIC SERVICE MANUAL 7 ERROR HANDLING 7 2 4 Check filters 111 UVE REF Min 12345578 LAMP REF indicates the intensitv of light on the reference sensor CUVET indicates the intensitv of light on the normal sensor CUVET Abs indicates the absorbance of the cuvette for the various filters The relative length of the bars that indicate the absorbances should be as in the figure i e first bar longest then everv bar shorter If a bar is much longer or shorter than expected this indicates bad filter The 340 nm 1 and the 546 nm 6 filters are usually the first to deteriorate After adjustment of the lamp no infinite signs may appear on top of the bars LAMPREF and CUVET 7 2 4 1 Water or vacu
120. exchange the PIO on the motion control board for the filter wheel E52 MEAS DISC INIT FAILED see E55 E55 MEAS DISC RESET FAILED Cause This error will be generated when the measuring disc motion control did not report the correct disc position within 8 seconds after the reset command Solution Check if the communication between svstem board and motion control board is present When there is no communication this can either be caused by the wiring the SIO on the system board the KIO on the motion control board or the main PCB fuse on the motion control board Replace the board s or replace the defective components Check if there is a mechanical obstruction of the measuring disc and repair if necessary E56 FILTER RESET FAILED Cause This error will be generated when the filter wneel motion control board did not report the correct filter reset position within 8 seconds after the reset command Solution Check if the communication between system board and motion control board is present When there is no communication this can either be caused by the wiring the SIO on the system board the KIO on the motion control board or the main PCB fuse on the motion control board Replace the board s or replace the defective components Check if there is a mechanical obstruction of the filter wheel and repair if necessary VITAL SCIENTIFIC 7 11 7 ERROR HANDLING SERVICE MANUAL E57 MEAS DISC COMMUNICATION ERROR Cause S
121. f the following functions are displaved Disc position Reset opto switch Encoder opto switches The reagent disc motor can be disabled with function kev F3 VITAL SCIENTIFIC 4 21 4 INSTALLATION AND MAINTENANCE SERVICE MANUAL 4 6 4 Sample Disc Analyser status Inactive Req tray buffer 1 18 Run mode DUAL External 88 232 ON LINE Display mode Funct Check Adj Adjust Lamp Reagent Arn Sample Arw Disc position Reagent Disc Measurement Disc Filter Wash Arm ISE Arm ISE Unit Pipettor Reset opto Vacuum system Encoder opto Water system Encoder opto B Optical electronics Electronics Barcode scanner Taba Reset H Edit Disable Service Drive Fi 2 r3 4 F5 Fo F 9 First press reset F1 to activate the sample disc function and then the sample disc position can be edited with F2 The active state of the following functions are displaved Disc position Reset opto switch Encoder opto switches The sample disc motor can be disabled with function kev F3 4 22 VITAL SCIENTIFIC SERVICE MANUAL 4 INSTALLATION AND MAINTENANCE 4 6 5 Measuring disc Filter Analyser status Inactive Req tray buffer 1 18 Run mode DUAL External 88 232 ON LINE Display mode Funct Check Adj Adjust Lamp Reagent Arn Sample Arw Disc position Reagent Disc Filter position Sample Disc Mash Arm ISE Arm ISE Unit Pipettor Disc reset opto Vacuum system Disc encode
122. he analvzer INTRODUCTION Wiring diagram Motion control board parts list Motion control board 4 sheets Motion control board lavout Svstem board parts list Svstem board 4 sheets Svstem board lavout Photometer board parts list Photometer board Photometer board lav out PLL board parts list PLL board PLL board lav out ISE interface board parts list ISE interface board ISE interface board lavout Valve board parts list Valve board Valve board lavout Dispenser driver board parts list Dispenser driver board Dispenser driver board lavout Cable filter parts list Cable filter board Cable filter lavout Temperature board parts list Temperature board 4 sheets Temperature board lavout VITAL SCIENTIFIC 8 ELECTRICAL DIAGRAMS 6001 750 6001 758 6001 758 6001 758 6001 759 6001 759 6001 759 6001 761 6001 761 6001 761 6001 763 6001 763 6001 763 6001 767 6001 767 6001 767 6001 768 6001 768 6001 768 6001 769 6001 769 6001 769 6001 770 6001 770 6001 770 6001 771 6001 771 6001 771 SERVICE MANUAL Reagent Sample arm board parts list Reagent Sample arm board Reagent Sample arm board lavout Reagent Sample arm board ISE Sipper stirrer driver board parts list ISE Sipper stirrer driver board ISE Sipper stirrer driver board lavout Temperature compensation board parts list Temperature compensation board Temperature compensation board lavout Wash arm board parts list Wash arm
123. he differential amplifier is connected to the inputs of the dual followers IC8a and b Depending whether heating or cooling is required the output of the differential amplifier is positive or negative The two dual driver outputs are connected together via diodes and connected to the switched power supply This signal that is connected to the power supply varies between OV and 15V When the input signal is OV the output voltage of the power supply is also OV Depending which output of the followers is positive one half of the bridge driver is active This means that the two outputs are controlling the current direction through the Peltier elements and therefore selecting heating or cooling 3 5 1 3 PELTIER DRIVING CIRCUIT The Peltier elements driver circuit is built around the four FET transistors T2 T3 T4 and T5 The two Peltiers are connected in series between point 3 and 4 of PL2 If the measured temperature is lower then 37 C the output voltage of IC9d is positive The output voltage of IC8a will also be positive and will drive T3 T7 will take away the gate voltage of T2 The current through the Peltier elements is going via T5 the Peltier elements and T2 to ground If the measured temperature is higher then 37 C the output voltage of IC5b will be positive and will drive T4 The current through the Peltier elements is then going via the Peltier elements and T4 to ground The System board can monitor the temperature with
124. he mixer assembly such that the noise of the needle touching the cuvette disappears e Fasten the 2 screws again Reset the sample arm F1 and move sample arm to the mixer position Repeat step 2 VITAL SCIENTIFIC 5 11 5 MECHANICAL ADJUSTMENTS SERVICE MANUAL NEEDLE SAFETV SWITCH ADJUSTMENT The instrument is provided with safetv switches optical on the reagent arm and on the sample arm to detect obstruction especiallv when moving down Optical needle switch adjustment 1 2 3 4 5 Select Sample Arm or Reagent Arm in the Service Menu Reset and disable the horizontal and vertical movement of the arm Move the arm so that the needle tip just touches the surface of the working table Now move down the arm very carefully in a way that you can feel the steps of the stepper motor After the 2nd or the 3rd step of the stepper motor the screen should say needle switch 1 If this is not the case you must adjust the vane of the switch see figure below in such a way that the switching point becomes within the range of 2 or 3 steps 5 12 vane 4 setscrew o 21 spring 4 VITAL SCIENTIFIC SERVICE MANUAL 5 MECHANICAL ADJUSTMENTS 5 7 REAGENT ROTOR ADJUSTMENT 1 Place a reagent bottle in position 1 of the reagent rotor 2 Select in the service menu reagent disc and reset the roto
125. he old belt and place a new one 4 5 2 7 CHECKING THE REPLACING PIPETTOR VALVES Replacing pipettor valves It is recommended to replace the pipettor valves every 2 years but in case of intensive use and depending on the temperature variations in the lab a higher replacement frequency might be necessary 1 Remove the syringe of the valve that needs replacement 2 Unscrew the valve from the frame and mount the new valve Make sure the shaft that drives the valve is in the right position 3 Mount the syringe again VITAL SCIENTIFIC 4 11 4 INSTALLATION AND MAINTENANCE SERVICE MANUAL Spindle drive removal 1 The two spindles are connected between two upper and lower beams It is not recommended to remove the beams from the plate otherwise a new alignment of the upper and lower beam will be necessarv 2 Loosen the two sprocket set screws from the spindle Push the spindle with the drive assembiv a little up until the sprocket comes loose from the spindle and the spindle comes loose from the lower ball bearing 3 Remove the guide shaft bv loosening the two set screws at each end and slide it out of the beams see figure 4 4 Be careful with the spring to jump awav Remove the circlip upper ball bearing and lower ball bearing and slide the complete spindle with drive assembly out Cleaning the spindle 1 Turn the spindle out of the drive assembly 2 Handle the spindle with care scratches etc Clean
126. hen there is no communication this can either be caused by the wiring the SIO on the system board the KIO on the motion control board or the main PCB fuse on the motion control board Replace the board s or replace the defective components 7 1 8 REAGENT SAMPLE DISC ERRORS E100 DISCS 14V FAILED Cause Solution This error will be generated when the circuit see sec 8 of this manual motion control board sheet 1 4 diode 025 does not detect the correct voltage Check the fuses and voltage on the motion control board Check connectors PL3 and PL4 Check if all voltages are present on the power supply and repair or exchange if necessary E101 DISCS 30V FAILED Cause Solution This error will be generated when the circuit see sec 8 of this manual motion control board sheet 1 4 diode 026 does not detect the correct voltage Check the fuses and voltage on the motion control board Check connectors PL3 and PL4 Check if all voltages are present on the power supply and repair or exchange if necessary Check sample rotor and reagent rotor stepper motors E102 REAGENT DISC ERROR Cause Solution 7 16 This error occurs when the opto reading of the reagent disc is not in accordance with the current theoretical position This happens normally when the disc is obstructed by something and the stepper motor is loosing steps Perform a selective reset of the reagent disc Check the operation of the opto couplers
127. his error will occur when the communication is disrupted distorted Check if the communication between svstem board and motion control board is present When there is no communication this can either be caused by the wiring the SIO on the system board the KIO on the motion control board or the main PCB fuse on the motion control board Replace the board s or replace the defective components 7 1 7 REAGENT ARM ERRORS E80 REAGENT ARM 14V FAILED Cause Solution This error will be generated when the circuit see sec 8 of this manual motion control board sheet 1 4 diode 025 does not detect the correct voltage Check the fuses and voltage on the motion control board Check connectors PL3 and PL4 Check if all voltages are present on the power supply and repair or exchange if necessary E81 REAGENT ARM 30V FAILED Cause Solution This error will be generated when the circuit see sec 8 of this manual motion control board sheet 1 4 diode 026 does not detect the correct voltage Check the fuses and voltage on the motion control board Check connector PL3 and PL4 Check if all voltages are present on the power supply and repair or exchange if necessary Check reagent arm stepper motors E82 REAGENT ARM HORIZONTAL ERROR Cause Solution 7 14 This error occurs when the opto reading of the horizontal movement of the reagent arm is not in accordance with the current theoretical position This happens norm
128. his stage of the signal processing is selected the amount of clock pulses stored in the counters after conversion corresponds to the logarithm of this signal To compensate for temperature and component variations measuring and reference signals are measured alternately As already mentioned before the reference value is equal to 100 mVolts which in turn corresponds to 20 000 counts and 2 000 mAbs after conversion When due to changes in the characteristics of the circuit the reference value is no longer equal to 20 000 counts it needs no argument that the measuring value will be affected proportionally Therefore measuring values are compensated for variations in the log conversion circuit by means of the measured reference signal It will be obvious that this kind of compensation has its limits Therefore a range of plus and minus 1000 counts has been established and as long as the reference value is within the specified window of 19 000 to 21 000 counts the circuit is assumed to function correctly and compensation will be carried out as described before When a reference value is measured which is outside the specified range the instrument will give an error 3 16 VITAL SCIENTIFIC SERVICE MANUAL 3 SYSTEM ORGANIZATION 3 7 PLLBOARD The liquid detection circuits of the sample needle and the reagent needle are located on the PLL board and are built around IC1 and 2 for the sample level detection and IC3 and IC4 for the
129. history screen is shown below Analyser status Stand hy Req tray buffer 1 18 Run mode DUAL External 88 232 ON LINE Display mode Error history L0 SYSTEM RESET INCOMPLETE 12 DEG 1994 10 58 E35 SAMPLE SYR RESET FAILED 12 DEC 1994 11 82 E36 REAGENT SYR RESET FAILED 12 DEG 1994 11 07 PIPETTOR COMMUNICATION ERROR 12 DEG 1994 11 07 12 DEC 1994 11 11 12 DEG 1994 11 11 12 DEC 1994 11 12 12 DEC 1994 11 47 Operator command RESET SYSTEM Print Print 1 11100 Eval Request Load Main All Funct Control IES Fi F2 M T5 Fo FR Fg The error historv shows the date time when error was generated When a certain date is selected the hardware error message is displaved for example the screen above four errors were displayed E07 5 E36 and E37 The operator action after the error message will also be displaved in this case the operator has pressed RESET SYSTEM A maximum of 100 errors can be stored in the error history VITAL SCIENTIFIC 4 31 4 INSTALLATION AND MAINTENANCE SERVICE MANUAL 47 SUPPORT TOOLS The instrument is designed in such a way that a minimum of repair equipment and special tools are required In most cases the common standard tool set will be quite sufficient Beside this tool set and the special tools described in section 7 3 1 and 7 3 2 the service technician should have a Digital Multi
130. in 0 1 C by means of the READY signal RDY If the temperature is not reached within 5 minutes the microprocessor will switch off the Peltier elements and will generate the CUVETTE TEMPERATURE ERROR Note that if the cuvette temperature error is acknowledged the error will not re appear any more until the instrument is reset The heating circuit is switched off VITAL SCIENTIFIC 3 9 3 SYSTEM ORGANIZATION SERVICE MANUAL 3 5 2 TEMPERATURE CONTROL REAGENT NEEDLE The temperature of the reagent needle is controlled bv a heating element inside it The temperature is measured bv a Negative Temperature Coefficient resistor The NTC with a resistance 22kQ at 25 is connected to a 1V precision voltage supplv and is wired to produce a voltage output that increases with the temperature The temperature control circuit can be divided into a number of separate circuits with the following functions e input amplifier e differential amplifier e driving circuit In figure 3 7 a schematic diagram is given for the temperature control A complete schematic diagram is given in section 8 of this manual 1V 408 8mV Input t Amplifier Differential Amplifier EE Switched Heating Power Element 10V Supply mm 10V at 0 1 Figure 3 7 Block diagram temperature control 3 10 VITAL SCIENTIFIC
131. ion and or solid waste according to the relevant governmental regulations Consult the reagent manufacturer for information on the concentrations of heavy metals and other toxic constituents in each reagent Biohazardous Parts Avoid direct contact with all parts of the sample area Treat these areas as potentially biohazardous and or infectious Avoid direct body contact with reagents Direct body contact may result in irritation or damage to your skin Refer to the manufacturer s reagent kit box and package inserts or product information sheets for specific instructions Avoid direct body contact with cleaning solutions Direct body contact may result in skin irritation or damage Refer to the manufacturers kit box and package inserts or product information sheets for specific instructions Additional precautions Flammables Avoid using dangerous flammable material around the instrument Fire or explosion may be caused by ignition Accuracy Precision of Measured Results Ap For proper use of the instrument measure control samples and monitor the instrument during operation An incorrectly measured result may lead to an error in diagnosis thereby posing a danger to the patient Treat all reagents according the manufacturer s recommendations Refer to the reagent kit box and package inserts or product information sheets for specific instructions plication The Fibron 1 is a single channel coagulometer that works on the p
132. it in the down position and the cuvette bottom is 0 5mm 0 1 The checking procedure is as follows 1 Place a cuvette in the measuring rotor 2 Select wash arm in the service menu 3 Reset the wash arm 1 4 Move wash arm to the down position in the service menu 5 Measure distance d d 6 Disable wash arm F3 7 Push the wash arm downward until the needles touch the cuvette bottom 8 Measure distance d again The difference between d5 and d8 must be 0 5 mm 0 1mm if not adjust wash arm The adjustment procedure is as follows 9 Loosen the locking screw 10 Adjust the setscrew in such a way that the distance between d5 and d8 becomes 0 5 mm 0 1 mm 360 cw 0 5 mm 360 ccw 0 5 mm You have to check every adjustment by repeating step 3 through to 8 Make sure that when turning the set screw counterclockwise that the wash arm board is really going down and not the screw coming up This can be done by pushing on the screw with the screwdriver 11 Tighten the locking screw VITAL SCIENTIFIC 5 5 5 MECHANICAL ADJUSTMENTS SERVICE MANUAL WASH ARM ADJUSTMENT SET SCREW LOCKING SCREW WASH ARM BOARD Figure 5 5 Wash arm adjustment Checking the cuvette detection Every time the wash arm is reset it goes down completely to check if the cuvette rotor is present If there is no
133. l drives used in vour There are three possibilities If your has only a drive C add the next two lines to the config svs file in the directorv devicehigh c windows ramdrive sys 4 e devicehigh c windows ramdrive sys 1024 e This creates TWO ramdrives called D and E size 4kB and 1024kB Ramdrive D is a dummy drive but is needed to be able to install a ramdrive on E If your PC has already drives and D harddisk and or CDROM add the next line to the config sys file in the C directory devicehigh c windows ramdrive sys 1024 e This creates a ramdrive called E with size 1024kB If your PC has already drives C D and E harddisk and or CDROM move the drive on E to another location F through Z and proceed with the steps mentioned above 4 4 2 DOS memory maximisation After the program is started the program uses the memory left for the test results This memory should be as large as possible You can check the total free memory with the mem command in a DOS box It should be as close to 600k as possible To get as much memory for the DOS application as possible check if the extended memory drivers are installed The config sys file should contain the next three lines device c windows himem sys device c windows emm386 exe ram dos high umb Depending on the Windows version the second line should be omitted Try this and check if this gives more memory Always reboot your PC when you have chang
134. l for conversion takes place photocell signal or reference signal and the FET switch IC8d is switched off causing the logarithmic discharge curve to appear at the output of IC2 This signal is compared to either a reference voltage via FET switch IC8a or to the measuring voltage via FET switch IC8b The reference voltage for the comparison measurement is also obtained from the 10 Volts precision reference source and has a nominal value of 100 mV During the time the logarithmic decreasing signal is higher than the measuring or the reference voltage Depending which one has been selected for conversion the counters in the CTC are fed with the high frequency clock 1 8432Mhz derived from the main clock signal KIO When the logarithmic decreasing signal is higher than the measuring signal as described above the output of comparator IC13 is high This signal is referred as start stop signal PH STRT and the time this signal is high is depending of the voltage of the measured signal This signal together with the high frequency clock are connected to an AND port The output of this AND port IC6b is connected to the CLK Trigger input of the 16 bits counter of the CTC When the logarithmic signal becomes less than the measuring reference signal the output of comparator IC4 is dropped low and the high frequency clock will not pass the AND port anymore In case the measuring signal which is in fact the relative transmission in t
135. l generate an error for the part that fails but probably also for parts that have not been reset yet If the reset fails one can check if this is caused bv a failing communication between the svstem board and the motion control boards in the following wav select INSTALLATION MENU select RELEASE and press lt Enter gt The system board now communicates with the motion control boards that send back the version of the EPROM they contain If this communication fails retry a few times you might try to replace the KIO that usually is responsible for this failure If the communication suddeniv fails during normal operation the instrument generates error Exxx COMMUNICATION ERROR VITAL SCIENTIFIC 7 25 7 ERROR HANDLING SERVICE MANUAL 7 2 3 3 Motion Control Board exchanges As all motion control boards are the same vou can exchange two boards in order to know if a problem is caused by the board or by something else The EPROMs determine the function of the board and they must stay with the slot Note that if you exchange board 6 measurement rotor that you copy the position of the hexacode switch from this board to the new board on this position Replacing the KIO of the MCB sometimes helps if it turns out that a sensor signal is present on the board but not recognised by the system 7 2 3 4 cuvette blank results If the results of the cuvette blank are bad low C V values e Check the lamp adjustment if you can not g
136. l scientific Maintenance There is no periodic maintenance The customer should run their internal qualitv control on a regular basis Error handling Error Inadequate light Only when this error appears with every sample the following steps should be followed Check the red LED If the LED is off and the heating and display are working replace the LED board pn 6002 513 If the LED is on check the light level by using the hidden menu and check if the parameters are ok The light level should be close to 41000 when there is no cuvette in the lightpath The gain setting can be adjusted by using the lt and gt key s To get into the hidden menu Switch on the instrument When the first screen appears Vital Scientific press first lt and immediately gt Always go to the next screen by using the ENTER key until the instrument tells you to restart the Fibron By switching off and on again The following numbers should be in there Light level 41000 Gain PT threshold 37 APTT threshold 25 Open threshold 37 Auto start threshold 512 Number Count 4 If the lightlevel can not be adjusted replace the detector board pn 6002 514 Error excessive light Only when this error appears with every sample the following steps should be followed Check the light level in the hidden menu See above for the numbers If the lightlevel can not be adjusted replace the detector board pn 6002 5
137. level in the bottle enough to aspirate 400 ul This position is determined by a level detection and the subsequential lowering with a predefined number of steps 2 8 VITAL SCIENTIFIC SERVICE MANUAL 2 FUNCTIONAL DESIGN 2 3 2 SAMPLE ARM The sample arm is responsible for moving the sample needle in the correct position The construction of the arm is such that the driving for the movements is decoupled separate stepper motors drive the up down movement and the sideward rotating movement The sample arm has five discrete horizontal positions which are represented by five stepper motor co ordinates Horizontal Cuvette position Wash position Sample rotor outer position Sample rotor middle position Sample rotor inner position In the vertical direction there are three discrete and one variable position Vertical Up or reset position Cuvette position Wash position Sample position variable position depending on the sample volume 2 3 3 REAGENT AND SANPLE ROTOR The reagent rotor has 32 discrete positions that correspond with the 32 reagent bottle positions although optional different reagent rotors are possible The sample rotor has 72 discrete positions that correspond with the 72 sample cup positions 2 3 4 MEASURING UNIT The measuring unit consists of the following items mounted together as one unit e measuring rotor e filter wheel with lamp unit The measuring rotor is stepper motor driven and the reset position i
138. low the instructions on the screen Page 3 Temperature check cuvette rotor The O K on the screen only indicates that the instrument reaches its set temperature It might take a few minutes before the instrument reaches this temperature To check if the temperature really is O K see paragraph 6 4 Page 4 Reagent arm thermostatting Wait for the indication on the screen to change to O K Page 5 Input amplifier offset adjustment Follow the instructions on the screen Page 6 Discharge buffer offset adjustment Follow the instructions on the screen VITAL SCIENTIFIC 6 1 6 ELECTRICAL ADJUSTMENTS SERVICE MANUAL Page 7 Output comparator offset adjustment Follow the instructions on the screen It is not necessary that you adjust to exactiv 1 5V The potentiometers must be adjusted in such a way that the outputs of the opamps switch over from about OV to about 5V On this switching point vou will find that the reading of the meter is unstable but this is normal behaviour Page 8 Reference counter adjustment Follow the instructions on the screen Switching on the sound F9 will make the adjustment easier Vou will hear a beep while doing the adjustment The lower the tone the closer vou are to exactiv 20 000 Adjust until the beep does not sound anv more 6 3 PHOTOMETER BOARD ADJUSTMENT Two Absorbance reference adjustment 6 2 Open slot A9 and remove the metal shielding Select in the service menu OPTICAL ELECTRONIGS Conn
139. ments of paragraph 5 2 have been executed first Checking the lightspot Remove the cuvette rotor Select in the service menu measurement disc filter and reset the cuvette rotor F1 and filter wheel F2 e Put some transparent tape Scotch on the detector side of the cuvette rotor Select 546nm F6 and switch on the lamp F8 There will be a green lightspot visible on the tape The lightspot must be in the middle of the hole behind the tape if not perform the cuvette rotor adjustment Cuvette rotor adjust UM 000 y X A P d 2 Figure 5 3 Cuvette rotor adjustment e Open slot 6 measurement disc filter wheel motion control board e When the lightspot is on the left side Turn hexaswitch 3 on board A6 clockwise one position higher e When the lightspot is on the right side Turn hexaswitch 3 on board 6 counterclockwise one position lower Note After the adjustment of hexaswitch 3 a reset F1 of the measurement disk has to be done in order to make the selected position become active If the Motion Control Board is being exchanged remember to copy the position of the old to the new board The off factory position can also be found on the sticker on the plastic panel next to the lamp 5 4 VITAL SCIENTIFIC SERVICE MANUAL 5 MECHANICAL ADJUSTMENTS 5 4 WASH UNIT ADJUSTMENT The wash unit must be adjusted in such a way that the distance between the needle un
140. mp assembly B However in order to readjust the height the clamp assembly also has to be loosened This means that after performing the vertical adjustment the horizontal adjustment must be performed too Initial Check 1 Reset the sample arm F1 2 Check if the free space between upper side of clamp assembly B and bottom side of swing unit A is 0 5 mm If not shift the reset opto a little up or downwards by means of the two fixing screws Reset the sample arm and check again if the free space is sufficient 3 Check if the distance between the top of the mechanical assembly and the bottom of the arm without plastic cover is 97 mm If not the vertical adjustment must be carried out VITAL SCIENTIFIC 5 9 5 MECHANICAL ADJUSTMENTS SERVICE MANUAL 97 17 FREE __ w RESET OPTO m 5 Figure 5 6 Arm adjustment Vertical adjustment 4 5 Loosen and remove screw C of the clamp assembly apply some Locktite 243 and mount again Do not fasten the screw Loosen screw F while holding ring G with your hands and adjust the shaft of the arm so that the distance is 97 mm Press ring G firmly upwards against block E and fasten screw F firmly After this adjustment the horizontal adjustment must be performed Horizontal adjustm
141. n experienced service engineer vou will appreciate that considerable part of problems with instruments in general result from external causes Environment problems like heavy line voltage fluctuations peaks etc inadequate earth ground connections dust especiallv in combination with high ambient humiditv corrosive vapours acid dirt etc can all affect the performance of an instrument For sensitive photometric equipment high intensitv light direct sunlight or from artificial sources might affect the operation as well In addition extreme high or low ambient temperatures will adverselv effect these instruments As for everything prevention is far better than repair therefore we strongly advise to follow the preventive maintenance instructions closelv Finallv the customer maintenance will be of utmost importance as well It is verv important that the cleaning instructions for the fluid svstem as described in the Instruction Manual are followed closelv 7 2 2 LIST OF PROBABLE FAILURES 7 2 21 Inconsistent measuring results Inconsistent measuring results can be defined as follows drop outs and too low measuring results 7 2 2 2 Drop outs Symptom Occasional drop outs in the measurement values too high or too low Probable cause Remaining water in the cuvette rotor because the wash arm is not emptying sufficiently Solution Check the adjustment of the wash arm the distance between the needle unit in the down position and
142. n of the reference measurement is to compensate variations of the lamp Because these circuits are identical only the operation of the actual input amplifier is explained The principle of the input amplifier is illustrated in figure 3 9 The figure shows the actual amplifier IC11 with one of the photo diodes as input source and four switches controlled by the microprocessor of the system board By means of these four switches sixteen different possibilities for the gain settings can be programmed 3 12 VITAL SCIENTIFIC SERVICE MANUAL 3 SYSTEM ORGANIZATION E IC 11 V b1 i b2 G1 Gain control Photodiode G2 G3 Da L b ase 1 5 Figure 3 9 Input amplifier principle diagram Referring to the electrical diagram of the photometer board of section 8 the gain setting is performed by four signals which are derived from the PIO of the System Board P G0 to P G3 These signals directly control the FET switches of IC9 Three of these FET switches perform in combination with the relav contact from REL2 the actual amplification setting By switching resistors R19 20 21 22 23 and 24 into the circuit whilst the remaining resistor R26 is switched bv means of the relav contact The relav is switched bv means of a FET switch IC8c and is controlled bv the fourth signal P G3 The same fourth sig
143. nal that controls the relav controls also FET switch and has a compensation function This FET switch will switch resistor R18 into the circuit from amplification step 8 and onwards Note that the functions are reverse from amplification step 0 to 7 P G3 is active high this means that REL2 is activated and FET switch IC14c also and will short the resistor R24 The gain factor per step equals the power of 1 5 and can be programmed in 16 steps from 1 5 to the power 0 up to 1 5 to the power 15 which will result in an amplification of 1 and 438 respectivelv Changing the amplification bv 1 step will cause an absorbance change of approximatelv 255 mAbs log 1 5 x 10 6 88 0 255 Abs According the law of Lambert Beer the absorbance is proportional with the lightpath The log 1 5 value has to be recalculated because the amplification steps are based on a standard lightpath of 10mm Therefore the absolute range is 15 x 0 255 3 82 Abs VITAL SCIENTIFIC 3 13 3 SYSTEM ORGANIZATION SERVICE MANUAL During the Blank procedure a series of actions is started to produce the optimal gain for the programmable input amplifier Because the gain of this amplifier is controlled in steps only It needs no argument that it will be virtually impossible to control the amplifier to produce a certain voltage corresponding to 100 transmission 0 Abs exactiv Therefore utilising a method of successive approximation commoniv used in A D co
144. nciple of this detection is based on capacitv An oscillator signal of 10V 13 kHz is connected to the metal sample rotor which is isolated from the rest of the frame The sample cup and sample itself serve as dielectric so the signal can reach the sample needle When the sample needle touches the sample this is immediately detected by the PLL circuit Phase Locked Loop after which the sample needle will go down a few more steps in order to be able to aspirate enough sample When the sample cup is empty no capacity is detected Message INSUFFICIENT SAMPLE FOR XXX will appear in the results Note that the dead volume depends on the type of the sample rotor 2 4 2 REAGENT LEVEL DETECTION The principle of the reagent level detection is exactly the same as that of the sample level detection Note that when no reagent is detected the error message NO REAGENT ON XXX is immediately displayed on the screen 2 16 VITAL SCIENTIFIC SERVICE MANUAL 2 FUNCTIONAL DESIGN 2 4 3 WATER BUFFER LEVEL DETECTION The water cask is provided with a svstem to detect the water level and to switch the water pump on and off Two float switches based on reed relavs which are mounted in the water cask see figure 2 9 detect the level To the Svstem board Ei U NA
145. nd detects reset opto not in accordance with current co ordinate This can have a mechanical cause bad functioning of the opto switches or stepper motor Note that the upper opto switch is the DOWN OPTO and the lower opto switch is the RESET Solution Check if the syringe motors are operating by means of the service menu Check if the timing belts of the two syringe motors are adjusted well not too loose or too tight and if the spindle is moving smoothly Clean grease and adjust the spindle if necessary Check opto switches E22 SAMPLE VALVE ERROR Cause When the opto reading is not in accordance with the current theoretical position Solution Check the opto switches of the valves Check the DC motor The resistance of older motors can get low causing the multifuse on the Motion Control Board to act Check the dispenser driver board E23 24 REAGENT SYRINGE POSITIONING ERROR see E21 E22 E25 REAG VALVE ERROR see E22 E30 PIPETTOR 14V FAILED Cause This error will be generated when the circuit see sec 8 of this manual motion control board sheet 4 diode 025 does not detect the correct voltage Solution Check the fuses and voltage on the motion control board Check connectors PL3 and PL4 Check if all voltages are present on the power supply and repair or exchange if necessary 7 8 VITAL SCIENTIFIC SERVICE MANUAL 7 ERROR HANDLING E31 PIPETTOR 30V FAILED Cause Solution This error will be ge
146. nerated when the circuit see sec 8 of this manual motion control board sheet 1 4 diode 026 does not detect the correct voltage Check the fuses and voltage on the motion control board Check connectors PL3 and PL4 Check if all voltages are present on the power supply and repair or exchange if necessary Check pipettor stepper motors E35 SAMPLE SYR RESET FAILED Cause Solution This error will be generated when the motion control board does not report the sample syringe reset after the reset command Check if the communication between the system board and motion control board is present When there is no communication this can either be causer by the SIO on the system board the KIO on the motion control board or the PCB fuses on the motion control board Replace the board s or replace the defective components E36 REAGENT SYR RESET FAILED Cause Solution This error will be generated when the motion control board does not report the reagent syringe reset after the reset command Check if the communication between the system board and motion control board is present When there is no communication this can either be causer by the SIO on the system board the KIO on the motion control board or the PCB fuses on the motion control board Replace the board s or replace the defective components E37 PIPETTOR COMMUNICATION ERROR Cause Solution This error will be generated when the communication to the m
147. ntains all the relevant exploded view drawings of the analyser Page Unit 2 Jmainunits 0000 4 assembly of covers 6 tubing diagram tube connector plate assemblv Er collecting casks i0 2 1 assembly of main units for water and vacuum system 12 water cask waste tubes connection block assembly ul vacuum tube 16 Jexternalpumpunit 1 main electrical units 4 0 6 8 0 6 8 0 2 mechanical assemblv 20 sample reagent arm swing unit clamp assy carriage assy encoder opto unit encoder stepper motor 28 30 wash arm movement assembly up down stepper motor wash arm adjust block assembly 34 Measuring unit optical block main assembly encoder stepper motor unit encoder opto unit lamp holder mounting plate assembly lamp holder unit filterwheel with filters beam splitter 46 water container waste containers 2 2 2 stirrer unit ise arm option 3 3 SPARE PARTS Page Unit 48 Power supplv Cooling unit Sample reagent arm VITAL SCIENTIFIC 9 1 9 MECHANICAL DRAWINGS MAIN UNITS PART NUMBER DESCRIPTION LOCATION 6001 956 COOLING UNIT 220V 50Hz 1 6001 959 COOLING UNIT 110V 60Hz 6001 753 EXTERNAL PUMP UNIT 2 3374 059 PUMPUNIT CABLE 3 9 2 SERVICE MANUAL VITAL SCIENTIFIC SERVICE MANUAL 9 MECHANICAL DRAWINGS MAIN UNITS VITAL SCIENTIFIC 9 3 9 MECHANICAL DRAWINGS ASSEMBLV OF COVERS PART NUMBER DESCRIPTION
148. ntout of the error historv of the instrument INSUFFICIENT REAGENT This error will be displaved when the safetv switch of the reagent needle is activated bottle emptv or when liquid detection of the reagent needle is not detecting anv liquid no bottle present This error can also occur when the liquid detection is not working INSUFFICIENT SAMPLE This error will be displayed when the safety switch of the sample needle is activated sample cup empty or when liquid detection of the sample needle is not detecting any liquid after aspirating the sample detection of air bubble LAMP ERRORS Reference detector Cuvette detector 100 mV signal 2 21 000 counts Lamp ref underrange ref underrange 100 mV signal lt 19 000 counts Lamp ref overrange Lamp signal gt 29 000 counts Lamp underrange Lamp signal lt 10 counts 7 1 2 HARDWARE ERRORS These errors are displaved on the screen in a window and can be recognised bv an audio alarm continuous beeps at three different pitches Pressing the space bar can stop the audio alarm Here is a list of all the hardware errors E00 Z80 MASTER CARD RAM CHECK ERROR E01 Z80 MASTER CARD ROM CHECK ERROR E02 SYSTEM EMERGENCY HALT W03 CYCLE TIMING PROBLEM E05 NO CLEAN CUVETTE E07 SYSTEM RESET INCOMPLETE E10 NO VACUUM E11 WASTE FULL E12 WATER RUNNING OUT E13 LAMP FAILURE E14 CUVETTES TEMP ERROR E15 REAGENT NEEDLE TEMP ERROR E16 CONCENTRATED WASTE FULL E17 INSUFFIC
149. nverters the gain is set in such way that it drives the output of the amplifier to fit within a specific window This window determines the maximum and minimum negative absorbance limits of the instrument The lower limit is fixed at 145 mAbs and the upper limit of the blanking window is 436 mAbs The theoretical range of the photometer is 0 436 3 82 4 256 Abs but at this maximum amplification the noise and drift effects are also high Therefore a maximum amplification step of 12 is chosen At amplification step 12 the maximum absorbance value is 3 06 Abs The lowest level that still can be measured is therefore 3 496 Abs 3 06 0 436 The table below illustrates how the FET switches must be set to produce the required gain Note that if during the blank procedure the instrument is not able to produce a gain resulting that the absorbance value is within the blank range an error will be displayed Right after blanking the output voltage of the input amplifier will be between 5V and 7 94 V unless overrange or under range has occurred G3 G2 Gi GO RELAY AMPLIFICATION ABSORBANCE O 0 fo coed o zi 0 0 1 closed 15 005 0 1 JO closed 225 055 0 closed 56 102 O closed 114 15 O open 297 36 0 1 1946 3945 0 open 92 35 3 14 VITAL SCIENTIFIC SERVICE MANUAL 3 SYSTEM ORGANIZATION 3 6 2 Lin log converter
150. ol 1V 408 8mV Input Amplifier Differential Amplifier om Temperature Compensation Circuit Switched Dual Power Followers Supply heat Bridge driver cool Temp off UE TEMP OK 40V 3 at 0 1 C Figure 3 6 Temperature control cuvette rotor A complete schematic diagram is given in section 8 of this manual 3 8 VITAL SCIENTIFIC Peltier Elements SERVICE MANUAL 3 SYSTEM ORGANIZATION 3 5 INPUT AMPLIFIER The input amplifier circuit is built around IC10 which is a voltage OP AMP OPO7 The purpose of this amplifier is to amplifv the output of the thermistor 5 76 to an adequate level The amplification is approximately 70x therefore the output voltage of IC10 is approx 400 To extend the dvnamic range of the amplifier the inverting input of IC12 is derived from a reference voltage of 408 8mV 3 5 1 1 DIFFERENTIAL AMPLIFIER The second stage of the temperature control consists of a differential amplifier IC9 and a compensation circuit board which is mounted at the front of the analvser The compensation board consist of an adjustable current source that is used as a remote sense temperature application IC4 will amplify the difference between the inputs 3 5 1 2 DUAL FOLLOWERS The output signal from t
151. oltage 14 V The other lead is connected to ground via FET transistor T1 When the power is applied the gate of the FET is controlled via resistor R34 and diode D10 The lamp voltage is measured via a resistor network and compared against the reference voltage The regulator controls the gate of the FET transistor in such a way that the measured lamp voltage is the same as the reference voltage The lamp voltage can be adjusted with P1 The lamp can be switched on and off by activating or deactivating opto coupler OC3B The lamp will be activated by the microprocessor on the System board The overcurrent through the FET transistor T1 is protected by means of transistor T6 and R6 VITAL SCIENTIFIC 3 11 3 SYSTEM ORGANIZATION SERVICE MANUAL 3 6 PHOTOMETER BOARD On the Photometer Board is the photometer electrical circuit In the figure below the block diagram of this board shown 17 PHOTO PHOTO MEASURING DIODE SIGNALS M ET E R TO SYSTEM PHOTO BOARD aye DODE CIRCUIT _ REF Figure 3 8 Block diagram Photometer Board The complete electrical diagram is given in section 8 of this manual 3 6 1 PHOTOMETER CIRCUIT The photometer circuit can be divided into the following two circuits e programmable input amplifier e lin log converter On the photometer board are two identical circuits present one for the actual measurement and one for the reference measurement LAMP REF The functio
152. olution This error will be generated when the communication to motion control board for the measuring disc stalls or if an unknown character is echoed Normallv this error will occur when the communication is disrupted or distorted Check if the communication between svstem board and motion control board is present When there is no communication this can either be caused by the wiring the SIO on the system board the KIO on the motion control board or the main PCB fuse on the motion control board Replace the board s or replace the defective components 7 1 6 SAMPLE ARM ERRORS E60 SAMPLE ARM 14V FAILED Cause Solution This error will be generated when the circuit see sec 8 of this manual motion control board sheet 1 4 diode 025 does not detect the correct voltage Check the fuses and voltage on the motion control board Check connectors PL3 and PL4 Check if all voltages are present on the power supply and repair or exchange if necessary E61 SAMPLE ARM 30V FAILED Cause Solution This error will be generated when the circuit see sec 8 of this manual motion control board sheet 1 4 diode 026 does not detect the correct voltage Check the fuses and voltage on the motion control board Check connectors PL3 and PL4 Check if all voltages are present on the power supply and repair or exchange if necessary Check sample arm stepper motors E62 SAMPLE ARM HORIZONTAL ERROR Cause Solution 7 12
153. omputer Bv means of serial communication ports it sends commands to the Motion Control Boards for all the movements and bv means of digital outputs it drives via driver boards the valves and the pumps Besides it switches on and off via digital outputs the various heating elements In combination with the photometer board the svstem board measures the extinction The intermediate results are stored in the svstem data memorv RAM that consists of a 32 Kbvtes CMOS RAM The serial interface between the svstem board and the motion control boards is controlled via the three Serial Input Output Z80 SIO devices The Killer Input Output Z80 KIO is an integrated device which contains a Counter Timer Control a PIO Parallel Input Output and a SIO with the following functions KIO CTC conversion of the photometer signal and timing control KIO PIO controls the valves V1 to V14 and the signals for the vacuum switch waste switch pumps etcetera KIO SIO serial connections with external computer and optional ISE interface The schematics and the wiring diagram of the svstem board is shown in section 8 of this manual The block diagram of the svstem board is shown in fig 3 3 VITAL SCIENTIFIC 3 3 SERVICE MANUAL 3 SYSTEM ORGANIZATION a 6 28586
154. on Check Cuvet Temp OK signal on A5 PL7 pin 4 hoog Check the resistance of the Peltier elements A5 PL2 Pin 3 4 2 to 10 ohms Check the temperature sensor thermistor in the cuvette block A5 PL5 Pin 1 3 36KOhms at room temperature Check fuses and voltages on temperature board Exchange A5 Temperature board check that the switch is in the on position For repairs check the Powerbox 15V tab8 and 15V tab7 generator E15 REAGENT NEEDLE TEMP ERROR Cause When the temperature control circuit is reporting the temp not OK signal for 2 5 minutes The circuit will be switched off to avoid overheating of the reagent needle Solution Check temperature sensor NTC of the needle red and blue wires in the arm 25KOhms Check heating element of the needle purple wires 50 Ohms Check wiring to the Temperature board Exchange temperature board For repairs check the Powerbox 15V tab8 and 15V tab7 generator E16 CONCENTRATED WASTE FULL see E11 Waste Full E17 INSUFFICIENT WATER Cause This error will be generated when the lowest float switch in the water cask is not detecting water This error must be preceded by a WATER RUNNING OUT error Solution See E12 WATER RUNNING OUT VITAL SCIENTIFIC 7 7 7 ERROR HANDLING SERVICE MANUAL 7 1 4 PIPETTOR ERRORS E20 21 SAMPLE SVR POS ERROR Cause When the svringe is going down and detects down opto or when svringe is going up a
155. otion control board for the pipettors stalls or if an unknown character is echoed Normally this error will occur when the communication is disrupted or distorted Check if the communication between system board and motion control board is present When there is no communication this can either be caused by the wiring the SIO on the system board the KIO on the motion control board or the main PCB fuse on the motion control board Replace the board s or replace the defective components VITAL SCIENTIFIC 7 9 7 ERROR HANDLING SERVICE MANUAL 7 1 5 MEASUREMENT DISC FILTER WHEEL ERRORS E40 MEAS DISC 14 FAILED Cause Solution This error will be generated when the circuit see sec 8 of this manual motion control board sheet 1 4 diode 025 does not detect the correct voltage Check the fuses and voltage on the motion control board Check connectors PL3 and PL4 Check if all voltages are present on the power supply and repair or exchange if necessarv E41 MEAS DISC 30V FAILED Cause Solution This error will be generated when the circuit see sec 8 of this manual motion control board sheet 1 4 diode 026 does not detect the correct voltage Check the fuses and voltage on the motion control board Check connectors PL3 and PL4 Check if all voltages are present on the power supply and repair or exchange if necessarv Check measuring rotor stepper motor and filter wheel stepper motor E42 MEASUREM
156. owever the vacuum level varies continuously due to opening and closing valves while the vacuum cask works as a buffer The vacuum sensor will become 0 when the vacuum level becomes too low The actual indication of the sensor can be checked in the service menu VACUUM SYSTEM The adjustment of the vacuum sensor has to be done with a vacuum meter The procedure is as follows 1 Select in the service menu vacuum system 2 Connect the vacuum meter between tube connector 2 at the rear of the instrument green vacuum and the tube going to the vacuum sensor 3 Select in the service menu OFF for all valves F1 to F8 and switch on the vacuum pump The vacuum level must be 2 0 7 bar After switching off the vacuum level it mav not reduce more then 0 1 Bar in 90 seconds When this is the case there is a leakage in the vacuum pump or in the tubing 4 Switch on the vacuum pump again and cover the rinsing place of the sample needle with your finger 5 While keeping vour finger on the rinsing place switch on valve 2 and valve 4 Let the vacuum escape little bv little bv means of lifting vour finger a little bit and check that the vacuum sensor switches from VACUUM to NO VACUUM at a level of about 0 3 bar If not adjust the switch level of the vacuum sensor by turning the setscrew accessible from the rear of the instrument remove back cover Turning clockwise increasing switch level VACUUM SENSOR VACUUM NO VACUUM
157. packing and installing the analvser The instrument is shipped in two boxes that contain e The analyser itself pump unit accessories e Cooling unit The boxes should be carefully inspected for damage If there is any major damage or when parts are missing this should be reported by Wood frome Accessory means of filling out the Installation report Figure 4 1 Shipping box top view The following procedure describes the recommended method for unpacking and installing the instrument 1 Place the pallet with the instrument on the floor and open the top side of the box 4 2 VITAL SCIENTIFIC SERVICE MANUAL 4 INSTALLATION AND MAINTENANCE 2 A wooden frame is visible that protects the arms together with 2 accessorv boxes 3 Unscrew the frame from the wooden box and remove it Carefully remove the 2 accessory boxes 4 Remove the cardboard packing material unscrew the big cardboard box from the pallet and lift it over the instrument Take the remaining accessorv box from the pallet MAINS PUMPUNIT voltage frequency power tuses 10 240 Vac Vitalab SERIAL NO 50 60 320VA AF WARNING agansi risk of spec 1 2 Manufactured by gt 1 0 VITAL QTY SCENTIAC DIEREN THE NETHERLANDS 5 Unscrew the instrument from
158. r 1 3 Select in the service menu reagent arm and reset the reagent arm Turn the reagent arm to position 1 above bottle 1 Disable the vertical drive 4 Push the reagent needle into the bottle and check if the distance d is the same on each side see drawing below Take radial play of bottle in rotor into account Figure 5 8 Reagent rotor adjustment 6 When this is not the case adjust the rotor bv means of the four clamping screws A see figure 5 9 Figure 5 9 Reagent rotor adjustment VITAL SCIENTIFIC 5 13 5 MECHANICAL ADJUSTMENTS SERVICE MANUAL 5 14 SAMPLE ROTOR ADJUSTMENT Select in the service menu sample arm and reset the sample arm F1 Turn the sample arm to position 1 above sample cup 1 gt gt Disable the vertical drive F4 Push the sample arm downwards and check if the needle housing is in the middle of the sample cup Take radial play of cups in rotor into account When this is not the case adjust the rotor by means of the four clamping screws A see drawing below 4 SZ J 22 9 e E Z lt lt gt M M 3 ES 222 Figure 5 10 Sample rotor adjustment VITAL SCIENTIFIC SERVICE MANUAL 5 MECHANICAL ADJUSTMENTS 5 9 VACUUM SENSOR ADJUSTMENT When the instrument is in the active state the vacuum pump is always running H
159. r opto A Water system Disc encoder opto B Optical electronics Filter reset opto Electronics Barcode scanner This 118 Reset BEJG VERNE Edit Edit Cont Service 102 180 On H F2 r3 4 hi Fo F F8 9 First press 1 to activate the cuvette rotor function and press F2 to reset the filter wheel and the following functions are active F3 F4 F5 F6 F7 F8 Disable the cuvette rotor motor Disable the filter wheel Edit the cuvette rotor position cuvette position from 1 to 48 can be entered Edit the position of the filter wheel filter number 1 to 8 can be entered Continuous run of the cuvette rotor Switch lamp on off The active state of the following functions is displayed Disc position Filter position Cuvette rotor reset opto switch Filter wheel reset opto switch Encoder opto switches VITAL SCIENTIFIC 4 23 4 INSTALLATION AND MAINTENANCE 4 6 6 Wash Arm Inactive DUAL Analyser status Run mode Adjust Lamp Reagent Arm Sample Arm Reagent Disc Sample Disc Measurement Disc Filter ISE Arm ISE Unit Pipettor Vacuum system Water system Optical electronics Electronics Barcode scanner Taba SERVICE MANUAL Req tray buffer External 15 232 Display mode 1 18 ON LINE Funct Check Adj High l Up Down Arm opto 8 Arm opto 1 Arm opto 2 Overflow detection No cuvettes 8 n R m 8 M
160. reset command Check if the communication between svstem board and motion control board is present When there is no communication this can either be caused the wiring the SIO on the svstem board the KIO on the motion control board or the main PCB fuse on the motion control board Replace the board s or replace the defective components Check if there is a mechanical obstruction of the sample arm and repair if necessarv Check if the opto couplers are working correctiv and repair if necessarv E76 SAMPLE ARM VERTICAL RESET FAILED Cause Solution This error will be generated when the motion control board does not report the vertical reset position after the reset command Check if the communication between svstem board and motion control board is present When there is no communication this can either be caused by the wiring the SIO on the system board the KIO on the motion control board or the main PCB fuse on the motion control board Replace the board s or replace the defective components Check if there is a mechanical obstruction of the sample arm and repair if necessary Check if the opto couplers are working correctly and repair if necessary VITAL SCIENTIFIC 7 13 7 ERROR HANDLING SERVICE MANUAL E77 SAMPLE ARM COMMUNICATION ERROR Cause Solution This error will be generated when the communication to motion control board for the sample arm stalls or if an unknown character is echoed Normally t
161. reset opto Encoder opto A B 0 8 Vertical reset opto Liquid sense Needle switch ax down bit Submerge depth Reset 0 1 Start Service Reaghrm 31910017 50111301 Fi F2 M hh Fo Fg First press reset F1 to activate the reagent arm function and the arm can be moved with the cursor kevs The active state of the following functions are displaved Horizontal reset opto switch Encoder opto switches Vertical reset opto switch Liquid sensor detected Needle switch Max down bit 0 yes liquid detected 1 no no liquid 1 arm is maximal down The horizontal steppermotor can be disabled by pressing F3 and the vertical motor can be disabled with F4 The stirrer motor can be started and stopped with respectively F5 and F6 Note that for all opto s opto 1 means opto is detecting light so there is nothing in the light path VITAL SCIENTIFIC 4 19 4 INSTALLATION AND MAINTENANCE SERVICE MANUAL 4 6 2 Sample Arm Analyser status Inactive Req tray buffer 1 18 Run mode DUAL External 88 232 ON LINE Display mode Funct Check Adj Adjust Lamp Cuv Mh Out Mid Inn Reagent fm Saupe Am 1 11 Reagent Disc Sample Disc Down l l l l Measurement Disc Filter Mash Arm ISE Arm ISE Unit Pipettor Horizontal reset opto Vacuum system Encoder opto B Water system Vertical reset opto Optical el
162. rganisation of the instrument that includes the theory of operation and the design of the different electronic circuits Section 4 Installation amp Maintenance This section provides instructions for installation and maintenance of the instrument and test diagnostics VITAL SCIENTIFIC 1 1 1 GENERAL SERVICE MANUAL Section 5 Mechanical adjustments This section describes the procedures for mechanical adjustments The assembly disassembly instructions are included as well Section 6 Electrical adjustments This section describes the electrical adjustment procedures Section 7 Error handling This section contains the error handling and a comprehensive procedure for trouble shooting A consolidated list of necessarv support equipment can also be found in this chapter Section 8 Electrical diagrams This section provides all the relevant electrical diagrams PCB assemblies electrical component part lists and wiring diagrams Section 9 Mechanical drawings This section provides the mechanical drawings part lists and facilitates the identification and ordering of service parts Section 10 This section is initiallv left blank but is meant for service information regarding the analvser 1 3 REPAIR POLICV Refer to the section General conditions of sale in the Price List for the general repair policv 1 4 WARRANTV Refer to the section General Conditions of Sale in the Price List for the general warrantv
163. rinciple of photometric clot detection The detection of the coagulation end point is achieved by measuring the change in optical density of a plasma reagent sample Please note that other types of analyses may not be applicable to this instrument Fibron 1 Manual Vital Scientific NV Service manual Rev 1 0 Page 4 7 vital scientific e Operation and Maintenance of the analyser should be performed in accordance with the instructions in this manual e During operation and maintenance of the instrument proceed according to the instructions and do not touch anv parts of the instrument other than those specified e Always clean the instrument after a batch of measurements e Ensure that both the covers are closed when the instrument is in operation e Avoid touching the instrument inside while the instrument is operating This may cause operation to stopped or damage the instrument Installation Requirements Installation should only be performed by a representative of the instrument s manufacturer The customer is responsible for providing the necessary facilities as detailed in chapter Technical Data in this User s Manual Instrument Unused for a Long Period of Time If the instrument is not due to be used for a long period of time ensure that it is thoroughly cleaned Place the dustcover over the analyzer to prevent dust from entering the analyzer Fibron 1 Manual Vital Scientific NV Service manual Rev 1 0 Page 5 77 vita
164. rive 4 Turn the strobe disk bv hand step bv step and check the reading on the screen for the following Everv two steps of the strobe disk there must be one change of the opto s Note Everv time it takes about a second before the screen displavs the actual situation Therefore after everv step wait at least 1 second for the screen to update If the code of the opto s is not changing properly it is necessary to reposition the complete opto encoder assembiv The distance between the opto s must be 1 6 mm and is fixed glued VITAL SCIENTIFIC 5 17 SERVICE MANUAL 6 ELECTRICAL ADJUSTMENTS 6 ELECTRICAL ADJUSTMENTS 6 1 INTRODUCTION The following paragraphs describe the electrical adjustment procedures The adjustments are carried out according to the text on the help screens of the service menu submenu ELECTRONIOS The description of the service menu is given in section 4 4 of this manual To the electrical adjustments the following equipment is required Digital voltmeter 4 digits 2 Absorbance test box Temperature meter accuracy of 0 1 C 6 2 CHECK POWER SUPPLV VOLTAGES Select in the service menu ELECTRONICS page 1 In this menu the complete description for the adjustment is given Perform all adjustments as indicated in the 8 pages The lt Page Down gt key steps to the next page Page 1 Power supply voltage check Follow the instructions on the screen Page 2 Lamp voltage adjustment Fol
165. rvice manual Rev 1 0 Page 8 7 vital scientific 3390 044 6002 512 6002 514 6002 513 Fibron 1 Manual Vital Scientific NV Service manual Rev 1 0 Page 9 Parts List 3315 276 Line filter 250 110 3342 857 3359 045 Power supply 3374 100 Printer cable 3374 123 LCD flat cable 3374 124 Detector cable 3381 033 LCD module 3390 044 Power resistor 1 5 heater 6002 503 Incubator unit 6002 505 Printer unit 6002 511 Main board 6002 512 Temperature board 6002 513 LED board 6002 514 Detector board 77 vital scientific Fibron 1 Manual Vital Scientific NV Service manual Rev 1 0 Page 10 7 vital scientific Fibron 1 Manual Vital Scientific NV Service manual Rev 1 0 Page 11 SERVICE MANUAL 2 FUNCTIONAL DESIGN 2 1 FUNCTIONAL UNITS 2 1 1 INTRODUCTION In the figure below vou find an overview of the main functional units of the analvser Reagent arm _ Sample arm Reagent Sample Rotor Rotor LLL H L 2 TT TT Cuvette rotor Wash arm ISE arm Figure 2 1 Overview of the analyser 2 1 2 SAMPLE ROTOR The various instruments can have different sample rotors Refer to the special pages for the variou
166. s Overrange reference counter Underrange reference counter CTC Overrange error signal Underrange error signal Gain bit gain For example err_ga 20 means after blanking an overrange signal is detected with a gain of 0 The following functions can be performed F1 With this function the next filter can be selected Each time when vou press this button the wavelength will be increased starting with a wavelength of 340 nm When vou press the SHIFT F1 a filter of a lower wavelength will be selected F2 After pressing this button the instrument will search for a gain high enough to reach a low absorbance value F3 After pressing this button the gain of the lamp circuit LAMP will be increased When pressing SHIFT F3 the gain will be decreased F4 After pressing this button the gain of the cuvette circuit CUVET will increased When pressing SHIFT F4 the gain will be decreased 4 30 VITAL SCIENTIFIC SERVICE MANUAL 4 INSTALLATION AND MAINTENANCE 4 6 12 Electronics In the Electronics menu the adjustment procedure is described for power supply photometer board lamp voltage temperature cuvette rotor and reagent needle A detailed description of these adjustments is given in section 6 of this manual 4 6 13 Error historv The analyser is equipped with a error history storage This menu can be entered by selecting in the main menu Special Functions and Service menu An example of an error
167. s detected bv an opto switch The measuring rotor has 48 discrete positions that correspond with the 48 cuvettes which must be positioned in the lightpath Since a full turn of the stepper motor is 2400 steps the distance between two cuvettes is 50 steps Two Peltier elements are mounted under the measuring rotor to heat the cuvettes The selection of the wavelength is done bv means of an 8 position filter wheel The standard mounted filters are 340 376 405 436 505 546 578 and 620 nm The filter wheel has 8 discrete positions that are 48 steps awav from each other The reset position is detected bv an opto switch The lamp unit consists of a lamp holder with a mechanism for adjusting the lamp VITAL SCIENTIFIC 2 9 2 FUNCTIONAL DESIGN SERVICE MANUAL 2 3 5 WASH UNIT AND BELLOWS PUMP The wash unit is built up with a stepper motor that is driving via a timing belt the wash arm WASH ARM ADJUSTMENT SET SCREW LOCKING SCREW WASH ARM BOARD d CUVETTE ROTOR Figure 2 4 Wash unit The positions of the wash arm are detected by three opto switches The opto switches are mounted on a PCB the so called wash arm board The wash unit normally has three vertical positions During normal operation two of them are accessed up and down Opto 1 is for detecting the up position and opto 2 is for the down position Opto 0 is for detecting the high position of the wash arm for changing of the cuvette rotor
168. s machines for the right rotor A stepper motor drives the rotor 2 1 3 REAGENT ROTOR There are 24 positions for 25 ml bottles and 8 positions for 7 ml bottles in the reagent rotor Some openings for 25 ml bottles can be combined to fit 50 ml bottles Adapters for a placing a 7ml bottle on a 25 ml position are also available All positions can be assigned as Reagent 1 or 2 start reagent The reagent disk compartment can be cooled by an external cooling unit A stepper motor drives the reagent rotor Some instruments have an optional special reagent rotor VITAL SCIENTIFIC 2 1 2 FUNCTIONAL DESIGN SERVICE MANUAL 2 1 4 SAMPLE ARM The sample needle aspirates a programmed volume and will dispense it into a cuvette of the cuvette rotor After pipetting the sample needle mixes the reaction mixture After sampling and mixing the needle will be washed inside and outside Because the sample arm has to move in two directions two stepper motors horizontal and vertical drive the sample arm The sample arm is equipped with a level detector that will detect the liquid level of the sample cup and will generate an error message when not enough sample is present in the cup 2 1 5 REAGENT ARM The reagent needle will aspirate a programmed reagent volume and will dispense it into a cuvette of the cuvette rotor After pipetting reagent 2 the reagent needle will mix the mixture in the cuvette Then the needle will be washed inside and outside
169. s much faster than the human eye can detect 3 6 VITAL SCIENTIFIC SERVICE MANUAL 3 SYSTEM ORGANIZATION 3 4 TEMPERATURE CONTROL BOARD The temperature control board has the following functions temperature control of the cuvette rotor temperature control of the reagent needle lamp voltage control temperature driver for cuvette cover temperature driver for plate heat exchanger The temperature control of the cuvette rotor is alwavs working unless the instrument is in the halted state or if a cuvette temperature error occurs The reagent needle is only heated when the instrument is in the active state On the service menu the needle temperature can be checked 3 4 1 TEMPERATURE CONTROL CUVETTE ROTOR The cuvette rotor is heated by two Peltier elements and the temperature is measured by a linear thermistor network This network consists of two precision resistors which are mounted on the board The thermistor is mounted inside the cuvette rotor assembly The thermistor is connected to a 1V precision voltage supply and is wired to produce a voltage output that increases linearly with the temperature The temperature control circuit can be divided into a number of separate circuits with the following functions e input amplifier e differential amplifier e dual followers e Peltier element driving circuit VITAL SCIENTIFIC 3 7 3 SYSTEM ORGANIZATION SERVICE MANUAL Figure 3 6 shows the block diagram of the temperature contr
170. s of the water detection svstem water running out and insufficient water is displaved 4 28 VITAL SCIENTIFIC SERVICE MANUAL 4 INSTALLATION AND MAINTENANCE 4 6 11 Optical electronics Analyser status Run mode Taba Inactive DUAL Display mode Adjust Lamp Reagent Arn Sample firm Reagent Disc Sample Disc Measurement Disc Filter Abs Mash Arm meas ISE Arm ISE Unit ref Pipettor err ga BB Vacuum system Mater sustem Electronics Barcode scanner LAMP ku Req tray buffer 4 1262 28838 19974 V a Funct Check Adj Wavelength 348 nu CUVET Gain Abs 1 6984 meas 7469 pef 19997 err ga BB M gt Fo Next Blank Hode ea F2 By means of this menu the optical functions can be showed Normally you will use this menu together with the lamp adjust menu In this screen the following actual values of the reference detector REF and the cuvette detector CUVET are displayed gain 0 15 Abs Absorbance value meas measured counter value ref measured value for 100 mV reference signal err ga error gain signal The err ga error gain signal is a read out of the bit map for the measurement and lamp signal that is generated after blanking VITAL SCIENTIFIC 4 29 4 INSTALLATION AND MAINTENANCE SERVICE MANUAL The bit map is build up as follows error bit error code XXXX No error
171. tallation Please note that a qualified technician must do the installation and initial checkout of the instrument Important remarks for installation e Install the analyser in such a way that it is not exposed to high intensity direct sunlight and vibration i e centrifuges e Avoid extreme high or low temperature and humidity Ambient temperature 15 to 32 maximum altitude of 3000 meter and maximum relative humidity 8096 at 32 non condensing e Note that dust corrosive vapour dirt etc may affect the performance of the instrument e The line voltage preferably a reliable connection should be selected Additional connections of instruments with high surge on peak currents to the same line should be avoided Also make sure that the instrument is well grounded e Do not remove from or connect interface cables to the system when the power is on Cooling unit The separate cooling unit accepts a line voltage of 115 or 230V Please note that the line voltage of the cooling unit is fixed and cannot be changed the line voltage is mentioned at the type plate of the unit The power consumption of the cooling unit is 350 VA max The power cords supplied with the instrument can either be the 220V non polarised or the 110V type depending on the type you have ordered Two fuses of 10 A slow must be placed VITAL SCIENTIFIC 4 1 4 INSTALLATION AND MAINTENANCE SERVICE MANUAL 42 UNPACKING AND INSTALLATION Un
172. the 280 does not receive the acknowledge the message is sent again to the PC If this happens 12 times for the same message this error is displaved Solution The communication line is not ok see E160 E162 Z80 MASTER CARD RS 232 ERROR Cause This error will be generated if the handshake line RTS of the svstem board is too long inactive more than 60 seconds because the PC is not cleared to sent a message to the 280 Solution Communication line is not ok or svstem board is not ok Replace the board or replace the defective components or replace wiring E166 WATCHDOG EXPECTED IN 5 SECONDS Cause This error will be generated when the software is crashed or there is a hardware problem and the program is stucked After one minute the 7 22 VITAL SCIENTIFIC SERVICE MANUAL 7 ERROR HANDLING watchdog goes off and restarts the system It will continue where it stopped Solution The error can be found in the error historv VITAL SCIENTIFIC 7 23 7 ERROR HANDLING SERVICE MANUAL 7 2 TROUBLE SHOOTING 7 2 1 INTRODUCTION The svstem trouble shooting information in this section provides a means of isolating faults possible causes and remedies Note that the information in this section does not provide an overall step bv step trouble shooting guide but merely is intended to supply the general guide lines on how to proceed in cases of a failure and in addition some symptoms probable causes and remedies are provided Being a
173. the 1 ml syringe there is an O ring inside the tip which you have to use with the new tip 2 Remove the plunger from the sample syringe and install a new syringe tip by means of the replacement tool 3 Check if the plunger tips are adjusted according the procedure in section 5 11 4 The glass barrels of the syringes are available under separate part numbers 3066 112 Glass barrel for 0 1 ml pipettor 3066 114 Glass barrel for 1 ml pipettor 4 5 2 9 REPLACEMENT OF WATERFILTER Replace the filter of the water tank by a new one 4 5 2 10 REPLACEMENT OF DRYING BLOCK Replace the drying block 6001 963 of the wash arm by a new one 4 14 VITAL SCIENTIFIC SERVICE MANUAL 4 INSTALLATION AND MAINTENANCE Before fixing the screw again lower the wash arm in the service menu with the new drving block alreadv mounted This assures that the drving block is entered in the position Fix the screw of the drving block 4 5 2 11 CHECKING THE CUVETTE ROTOR ADJUSTMENT The position of the cuvette rotor has to be checked by following the adjustment procedure of section 5 7 of this manual 4 5 2 12 THE PHOTOMETER BOARD ADJUSTMENT See section 6 3 of this manual 4 5 2 13 CHECKING REAGENT NEEDLE AND TEMPERATURE See section 6 5 and 6 6 ofthis manual 4 5 2 14 CHECKING WASH ARM ADJUSTMENT AND BELLOWS VOLUME See section 5 2 and 5 3 of this manual Alwavs perform wash arm adjustment firstl 4 5 2 15 CHECKING THE REAGENT AND
174. tment 23 Perform the dichromate endtest VITAL SCIENTIFIC 4 9 4 INSTALLATION AND MAINTENANCE SERVICE MANUAL 4 5 2 1 SYSTEM CLEANING This cleaning procedure cleans the complete tubing of the instrument The automatic clean system can be started by selecting F5 Special Functions F1 Rotor System and CLEAN SYSTEM The operating instructions are displayed on the screen Please note that the clean system procedure will take about 2 5 hours and after this procedure you have to run all programmed tests on a control serum When the results are out of limits the Empty and Fill System procedure must be repeated to remove the remaining cleaning solution To prevent contamination clean the water tank with cleaning solution Rinse three times with water afterwards During the procedure the operator will have to put the water filter in a bottle with cleaning liquid The machine will take about 500 ml of it 4 5 2 2 REPLACEMENT OF REAGENT AND SANPLE PROBE TUBING Remove the old reagent and sample tubing and mount the new tubing 4 5 2 3 REPLACEMENT OF ISE SAMPLE TUBING This is only relevant when the instrument is equipped with a separate ISE unit The length and tvpe of this sample tubing has direct effect to the ISE measurement therefore it is verv important to use the correct tubing code number 1573 002 with a length of 650 2 mm 1 Remove the ISE needle unit and disconnect the old tubing The
175. ture differences without building a pressure or vacuum Take care that you do not switch on the cooling unit without liquid because the pump will be damaged almost immediately e Switch ON the cooling unit by means of the two switches in front of the unit marked cooling and pump e See chapter 2 7 for changing temperature setting of the cooling unit Place a cuvette rotor into the measurement rotor Make sure not to touch the sides of the cuvettes Plug the power cord in the instrument and the other end into a grounded power outlet and turn power on VITAL SCIENTIFIC SERVICE MANUAL 4 INSTALLATION AND MAINTENANCE 43 SUPPRESSION CONCENTRATED WASTE ALARM Standard the instruments are delivered with the Concentrated Waste option NOT installed In this case the alarm indicating that the Concentrated Waste container is full has to be suppressed This is done by installing a special connector part number 3351 028 Part number 3351 028 The connector has to be placed in the slot of the pump unit where the fluid level detector for the Concentrated Waste container is connected VITAL SCIENTIFIC 4 5 4 INSTALLATION AND MAINTENANCE SERVICE MANUAL 44 SOFTWARE INSTALLATION 4 4 1 Ramdrive installation The software needs a ramdrive This must be drive E size 1024 If your PC doesn t have such a ramdrive it has to be installed The ramdrive will alwavs be installed after the loca
176. um system If there are problems in the water or vacuum system water overflow no vacuum e select the appropriate menu and check both pumps and all the valves you can hear them switch If pumps or valves do not operate e Connect a spare valve or exchange two connectors to see whether the component or the electronics is defective e Change valve board check fuse first e Change system board Check the complete water and vacuum system systematically with the help of the service menus and the Vacuum diagram see figure 2 12 VITAL SCIENTIFIC 7 27 7 ERROR HANDLING SERVICE MANUAL 7 2 4 2 Liquid detection problems If there are problems with the liquid detection first make sure that the shaft and the bearing of the needle are completely dry Also note that the minimum sample volume is 300 microliter for normal samples 200 microliter for controls and standards and 100 microliter for paediatrics Check the PLL adjustment if necessarv see paragraph 6 5 If this is not enough e Ifthe instrument gives the INFO INSUFFICIENT SAMPLE although there is clearly enough sample turn the corresponding potentiometer 360 degrees more clockwise more sensitive e lf the instrument thinks it picked up sample that was not there turn the potentiometer 360 degrees counterclockwise less sensitive 7 28 VITAL SCIENTIFIC SERVICE MANUAL 8 ELECTRICAL DIAGRAMS 8 1 This section contains all relevant electrical diagrams and parts lists of t
177. y in its place while holding the sprocket with timing belt in its final position 2 Pull the drive assembly and the sprocket towards each other to remove all play and tighten the two set screws note for the flat surface on the spindle 3 Slide the guide shaft on its place together with the two bushings and spring Note that the bushings must fit freely 4 Lock the set screws at both ends Apply a little grease Molykote PG75 to the bushings so they will slide smoothly over the guide shaft Grease the spindle over the full length with Molykote PG75 by applying it with a brush 5 Check the belt tension and adjust if necessary by loosening the stepper motor screws sliding the motor in the play of the four mounting holes until the correct belt tension is reached and the tightening the mounting screws Now let the drive assembly run up and down by the service menu to spread the grease and remove eventually overspill of grease Check if the drive assembly is running smoothly without irregularities 4 5 2 8 REPLACE SYRINGES For the procedure of replacement of the syringes see the instruction manual Note that the life time of the teflon tips sealings can be increased if a very little bit of silicon grease is applied to it If only the sealings are replaced the following procedure can be followed 1 Remove the plunger from the reagent syringe and install a new syringe tip by means of the replacement tool 4100 533 Please note that in
Download Pdf Manuals
Related Search
Related Contents
Samsung Leitor de MP3 YP-Q1JCW manual de utilizador 045F - Lamber TE150 Manual - Wireless Weather Station with Weather Boy Sony SS-XB5A User's Manual American Sensor SA379 User's Manual カタログ Osram Eco Pro Classic P Directive 94/35 Edulcorants Fleurines &Vous 酸素濃縮器(LiteTEC Copyright © All rights reserved.
Failed to retrieve file