8J4071 CS 7600 User Guide, Ed. 01

image

Contents

1. 41 Storing the Images 41 Acquisition Related Procedures 41 Clearing Scanner Memory 41 Re recording imaging plates 42 Erasing the Imaging Plates 42 Imaging Plate Ss ee ee seep enairen res 43 Handling and Storage of Imaging Plates 43 Cleaning the Imaging Plate 44 Cleaning Materials 44 Cleaning Method 44 Disinfecting Imaging Plates 44 Frequency of Disinfection 44 Disinfection Solutions 2144444 224 eo 44 Disinfection Method 45 Disinfecting the Hygienic Sheath 45 Frequency of Disinfection 45 Disinfection Solutions 45 Disinfection Method 46 Disposing of the Imaging Plate 46 Seamer CIE da mas ied Ook drame EEE WE OA ae 47 Cleaning the Scanner Body 47 Cleaning and Disinfecting the Detachable Insertion PAM es sidi ss SO REe Keene SD ed aed de 47 Cleaning the Scanner Rollers 49 Scan amp Go Device Care 2023 02dhi625i 0400 eden nnn 52 Cleaning the Scan amp Go 52 Replacing the Scan amp Go Battery 52 Quick HoubleshOOunGi s 0iracceeeeiwieesehGoaiws aeae Gas 55 CS
2. 18 Scanner Interface ICONS 19 Starting the Imaging Software 21 Starting the KODAK Dental Imaging Software 21 Creating a Patient Record 21 Accessing the Acquisition Interface 22 Starting the CS Imaging Software 23 Creating a Patient Card 23 Accessing the Acquisition Interface 23 PEON LOU DRE cena ans Seve ae ee a ete ok 25 Preparing for Acquisition 24 26 4eiuntoie ee as enaa 25 Preparing the Imaging Plates 26 Performing the X rays 232 raser attentions 27 Scanning the Imaging Plates 28 Processing the MAS 4 402262 esas dem resss rss 30 Storing the MAS 4e des hu eress 30 Acquisition Related Procedures 30 Clearing Scanner Memory 30 CS 7600_User Guide 8J4071 Ed01 Chapter 6 Acquiring Images using Scan amp Go Chapter 7 Maintenance Chapter 8 Troubleshooting Appendix Consumables amp Accessories Retrieving No ID Images 32 Erasing the Imaging Plates 33 ROOMS rt ee oe en i eed we ed es oes 35 Preparing for Acquisition 36 Preparing the Imaging Plates 36 Performing ba arn eae Gee hoes ce ee car can ears 38 Scanning the Imaging Plates 39 Processing the Images
3. Cleaning and Disinfecting the Detachable Insertion Panels The CS 7600 makes use of two insertion panel sizes according to the inserted plates size 0 3 and 4 The maintenance procedure is identical for both insertion panels WARNING It is the user s responsibility to disinfect the detachable insertion panel s daily to prevent cross contamination CAUTION Do not autoclave autoclaved detachable insertion panel s must be discarded To clean and disinfect the insertion panels follow these steps 1 Remove the installed insertion panel by inserting a finger in the designated removal depression at the bottom of the insertion panel A and lifting out the insertion panel B 2 Use disinfection wipes for medical equipment to clean the detachable insertion panels thoroughly from all sides including the slot CS 7600_User Guide 8J4071 Ed01 47 3 The scanner displays a prompt to put in place an insertion panel either a size 0 3 left figure in prompt or a size 4 right figure in prompt insertion panel Detachable Insertion Panel Insert panel in place 4 Install the cleaned and disinfected insertion panel by positioning the insertion panel at a 45 degree angle to the right side of the insertion panel s housing A and pressing the insertion slot panel in from right to left B until it s completely inserted C 5 The scanner displays an insertion confirmation Detachable Insertion Panel
4. Important The imaging plate has a marker attached to its inactive inactive side contrary to its intended use The image orientation can be corrected using the mirroring tool in the acquisition software If a diagnosis is not possible in the area of the marker the X ray should be repeated To process the scanned images from your workstation follow these steps 1 Images retrieved from recorded imaging plates populate the correct frame in the layout regardless of the scanning sequence If the image is a new image retake of a previous image that already populates a frame the new image will populate the correct frame and the old image is moved to the Image Gallery panel 2 Use the Acquisition interface processing options to optimize the scanned image refer to CS 7600 Acquisition Interface Overview on page 11 3 Click the X icon top right corner to exit the Acquisition interface and return to the imaging application Images are transferred from the CS 7600 Acquisition interface to the Imaging Software automatically Storing the Images Image management storage and retrieval is performed through the Imaging Software Refer to the Imaging Software documentation Acquisition Related Procedures Clearing Scanner Memory When the CS 7600 Acquisition interface is open at the designated workstation the scanned imaging plate image is automatically routed to the correct exam acquisition If the CS 7600 Acquisition interfa
5. 7600 Consumables amp Accessories 57 1 Conventions in this Guide Conventions in this Guide The following special messages emphasize information or indicate potential risk to personnel or equipment WARNING Warns you to avoid injury to yourself or others by following the safety instructions precisely CAUTION Alerts you to a condition that might cause serious damage Important Alerts you to a condition that might cause problems Note Emphasizes important information Tip Provides extra information and hints CS 7600_User Guide 8J4071 Ed01 1 2 Chapter 1 Conventions in this Guide 2 CS 7600 System Overview Introduction to have training to use the CS 7600 Note The CS 7600 is intended for dental professionals You are required CS 7600 is a scanner system for intraoral X ray exposed imaging plates The CS 7600 is intended for digital dental radiography using an imaging plate storage phosphor screen for radiographic diagnostic intraoral images Use the CS 7600 system to scan and review intraoral dental X ray images When scanning the X ray exposed imaging plate a digital image is previewed on the scanner s LCD and saved to the scanner s internal memory After scanning the scanner erases the imaging plate and ejects it The imaging plate is ready for re use The exam acquisition s scanned image is sent over the network to the workstation s Acquisition interface Using the Acqu
6. Go device to Chapter 6 Acquiring Images using Scan amp Go Gi Note If your configuration includes the optional Scan amp Go device refer A typical workflow consists of the following steps e Registering the patient choosing the tooth to X ray e Inserting the imaging plate into a hygienic sheath and sealing it e Placing the imaging plate in the patient s mouth behind the selected tooth or teeth e Exposing the imaging plate to X ray e Removing the imaging plate from the patient s mouth e Disinfecting the hygienic sheath e Scanning the imaging plate in the CS 7600 scanner imaging plate is scanned without its hygienic sheath e Reviewing the scanned image on the Acquisition interface e Storing the image in the Kodak Dental Imaging Software Preparing for Acquisition Before starting the acquisition procedure make sure you have e Selected the patient record e Accessed the imaging window e Accessed the Acquisition interface The above procedures are detailed in the previous chapter Note Keep the Acquisition interface open throughout your exam acquisition session close it only after you have completed scanning and optimizing all images CS 7600_User Guide 8J4071 _Ed01 25 Preparing the Imaging Plates To prepare the intraoral imaging plate follow these steps 1 Choose an imaging plate of the right size for your examination Important Use only CS 7600 intraoral imaging plates from Carestr
7. Press the middle function button to confirm and return to the main screen 48 Chapter 7 Maintenance Cleaning the Scanner Rollers Note Before you perform this procedure make sure you have the rollers cleaning plate The scanner rollers should be cleaned in the following cases e When the scanner prompts to clean the rollers e When there are dirt related artifacts on the scanned image To clean the scanner rollers follow these steps 1 Cleaning the rollers requires using the size 4 insertion slot If it is not installed replace the installed insertion slot with the correct size See page 47 2 Press the scanner s left function button to access the Settings menu 3 Select Rollers Cleaning _Settings Exit Brightness Control IP Setup Rollers Cleaning About Information Diagnostics 4 Press the middle function button to confirm the Rollers Cleaning due message CS 7600_User Guide 8J4071 _Ed01 49 5 Peel the adhesive protective sheet from the rollers cleaning plate ZA CS 7600 Rollers Cleaning Plate ers Za CLEANING SIDE P N AS000674 6 Follow the scanner prompt Rollers Cleaning Insert Cleaning plate and insert gently the rollers cleaning plate into the size 4 insertion slot with the adhesive side first and the black stripe facing downwards until the rollers draw the rollers cleaning plate in 50 Chapter 7 Maintenance 7 When the cleaning procedure is completed the scanner will
8. its hygienic sheath to minimize exposure to light degradation of image quality Do not remove the imaging plate from Important If the imaging plate is exposed to light it could result in A N its hygienic sheath until you insert it in the scanner s insertion slot 2 At the scanner location tear open the hygienic sheath A and insert the exposed imaging plate into the scanner by positioning the open side of the sheath in the insertion panel slot B with the foam side to the right Squeeze out the tip of the imaging plate from its hygienic sheath C The CS 7600_User Guide 8J4071 _Ed01 39 imaging plate without its protective sheath will be drawn into the insertion slot by the scanner and scanning starts automatically The scanner preview shows the scan progress and the scanned patient s details ad SE 3 Dispose of the used protective hygienic sheath D in previous figure 4 Collect the ejected imaging plate from the lower tray after it is scanned including its recorded information before ejecting it The imaging plate is ready for re use i Note After scanning the image the scanner erases the imaging plate 40 Chapter 6 Acquiring Images using Scan amp Go Processing the Images side It will indicate if the imaging plate has been exposed from the wrong side contrary to its intended use If the shadow of the marker is visible in the X ray image the imaging plate was exposed from the
9. sizes according to the inserted imaging plates size 0 3 or 4 7 Front cover 8 Back covers 4 Chapter 2 CS 7600 System Overview Figure 2 CS 7600 Rear View o M ia 1 Service door 2 Base 3 Cables compartment for power supply connection and interface ports RJ 45 Ethernet port 4 Back cover Figure 3 Cables Compartment in detail 1 Power supply inlet Zi RJ 45 Ethernet port Ethernet e 2 activity LEDs yellow indicates L speed green indicates link status Service ONLY ports 4 Compartment cover CS 7600_User Guide 8J4071 _Ed01 5 Smart Imaging Plates Overview The CS 7600 works with five imaging plate sizes O to 4 The scanned image file size depends on the imaging plate size and the scanning resolution set by the user The smart function of the imaging plates is utilized when using the Scan amp Go device Table 1 Intraoral imaging plates sizes Size Illustration Application O Periapical exam Pediatric dentistry 1 Bitewing exam Pediatric dentistry 2 Periapical exam Adult 3 Bitewing exam Adult 4 Occlusal exam 6 Chapter 2 CS 7600 System Overview The imaging plate has two sides the active side for capturing the X ray image and the inactive side The active side of the imaging plate includes an orientation mark c to facilitate accurate positioning within the patients mouth The inactive side has a printed dot that corresponds to the location of the orientati
10. CS 7600 User Guide Notice CARESTREAM Systems are designed to meet international safety and performance standards Personnel operating the unit must have a thorough understanding of the proper operation of the system This guide has been prepared to aid dental professionals to understand and operate the system Do not operate the system before reading this guide and gaining a clear understanding of the operation of the system If any part of this guide is not clear please contact your Carestream Health representative for clarification Specifications User Guide and make sure you observe all the WARNING Read the Safety Regulatory and the Technical A N Warnings and Cautions before using the CS 7600 The CS 7600 is intended for digital dental radiography using an imaging plate storage phosphor screen for radiographic diagnostic intraoral images The CS 7600 in addition optionally provides the Scan amp Go technology to enable dental practitioners prior to acquisition to record the dental imaging plates with exam acquisition data No part of this guide may be reproduced without the express permission of Carestream Health Inc U S Federal law restricts this device to sale by or on the order of a dentist or physician This document is originally written in English Guide Name CS 7600 User Guide Part Number 8J4071 Revision Number 01 Print Date 2011 08 In this guide all trademarks and registered trademarks are the propert
11. Insert the imaging plate in the scanner s insertion slot without using a hygienic sheath CS 7600_User Guide 8J4071 Ed01 33 The Erasing screen is displayed Erase plate Workstation D Destination1 OREX IDANM Collect the imaging plate while taking care to maintain hygienic protocol and minimal exposure to light 34 Chapter 5 Acquiring Images Acquiring Images using Scan amp Go Introduction This chapter describes the acquisition procedure for configurations that include the optional Scan amp Go device Using the Scan amp Go option streamlines and shortens the scanning process The scanned image reaches its destination automatically no need to select the destination workstation when scanning the imaging plate In FMS mode the image automatically populates its designated frame in the template The Scan amp Go option allows multiple users to scan images from multiple patients simultaneously automatically and securely device refer to Chapter 5 Acquiring Images Gi Note If your configuration does not include the optional Scan amp Go A typical workflow consists of the following steps e Registering the patient choosing the tooth to X ray e Recording the imaging plate with the exam acquisition information using the Scan amp Go device e Inserting the imaging plate into a hygienic sheath and sealing it e Placing the imaging plate in the patient s mouth behind the selected t
12. Perform steps 3 and 4 for each imaging plate the imaging plate on top of the Scan amp Go device and check the Imaging plate Scan amp Go information in the Settings Imaging plate information tab Q Tip To display the imaging plate s recorded information at any time place To record exam acquisition data on an imaging plate that was already recorded refer to Re recording imaging plates on page 42 CS 7600_User Guide 8J4071 Ed01 37 Performing X rays Perform the required X rays according to your clinical procedure It is recommended to continue using X ray positioning techniques and tools to ensure the resulting image is complete and includes all the information required for diagnosis It is recommended to use the Rinn Suresoft holder or equivalent products available on your market To facilitate matching the orientation of the image in the software to the clinical reality we recommend to position the imaging plate in the mouth of the patient with the orientation mark always towards the bottom Important Make sure the active side of the imaging plate is facing the patient s teeth 38 Chapter 6 Acquiring Images using Scan amp Go Scanning the Imaging Plates 1 Remove the imaging plate from the patient s mouth Do not remove the imaging plate from its hygienic sheath Wipe and disinfect the exposed imaging plate s hygienic sheath see Disinfecting the Hygienic Sheath on page 45 Keep the imaging plate in
13. Social Security Number fields are required 3 Click OK to save The patient record is automatically assigned a 7 digit number starting with a letter for example MO000001 CS 7600_User Guide 8J4071 Ed01 21 fs ms 4 Click to access the Imaging window Alternatively double click the selected patient name Accessing the Acquisition Interface To access the Acquisition interface follow these steps 1 In the Imaging window from the toolbar click the Acquisition interface access icon on the top toolbar e To start the Acquisition software in CS 7600 FMS mode Template mode click the icon s expand triangle and click e Tostart the Acquisition software in CS 7600 Acquisition mode image acquisition mode click the icon s expand triangle and click ay acquisition session close it only after you have completed scanning and optimizing all images i Note Keep the Acquisition interface open throughout your exam 2 For non Scan amp Go configurations If you need to change the scanning resolution for this exam acquisition do this through the scanner resolution setting see page 29 3 For Scan amp Go configurations only Verify that the Scan amp Go device is connected see CS 7600 Acquisition Interface Overview on page 11 4 Ifyou are not using a Scan amp Go device see Chapter 5 Acquiring Images to launch an acquisition If you are using a Scan amp Go device see Chapter 6 Acquiring Imag
14. ce is closed the scanned imaging plate image is stored in the scanner s memory The scanner s storage capacity is 25 images To clear the scanner s memory follow steps 1 through 5 on page 31 Clearing Scanner Memory CS 7600_User Guide 8J4071 _Ed01 41 Re recording imaging plates To record an imaging plate that was already recorded with exam acquisition information re recording follow these steps 1 Place the imaging plate upon the Scan amp Go device The following prompt is displayed This imaging plate was already tagged for fame 1 Do you wantto override the imaging plate recorded data Cancel 2 Click OK to re record this imaging plate The imaging plate is recorded with the current exam acquisition and the current active frame information If you do not want to re record the imaging plate either remove the imaging plate from the Scan amp Go surface or click Cancel in the overwrite prompt The current recorded data as displayed in this prompt remains unchanged 3 Wait until the exam acquisition information recording is completed as indicated by an audible beep successive short flashes of the blue LED and a change of color in the frame border and number Erasing the Imaging Plates The imaging plates are automatically erased and ejected after their information is scanned Exposure to light and so forth requires erasing manually the imaging plates before use To erase the imaging plate manually follow s
15. configuration e Check the computers s firewall settings and disable the firewall e Exit the Acquisition interface and No connection to scanner relaunch the application after launching Acquisition e Check the scanner s network interface configuration by checking in the scanner s Settings IP Setup screen if DHCP is enabled or not and consult with your system administrator what the scanner configuration should be e Replace the network cable e Ifthe scanner icon frame does not change to green within 40 s restart the scanner by turning it off and then on again Scanner Error 071 Scanner memory full ee Uleaiing Canne Memon on page 30 e Press the middle function button to eject the imaging plate and then Scanner Error 073 Imaging plate loading failure reinsert the imaging plate again e Replace imaging plate if reinsertion does not work Imaging plate is not ejected Scanner Error 001 from the scanner into the imaging plates tray Remove the imaging plate manually from the scanners insertion slot CS 7600_User Guide 8J4071 Ed01 55 Table 1 Quick Troubleshooting Continued Malfunction Possible Cause Action e Make sure you are using Carestream imaging plates e Make sure you follow the preparations and scanning instructions in the Acquiring Images procedures to Either imaging plate inserted prevent incorrect insertion of the incorrectly or the imaging imaging plates Scanner Error 008 pla
16. cquisition destination address 90000000 Series 00000000 Scan resolution Manufacturing date 20110724 Exam acquisition ID 00000000 Unique ID E004010053D8E9C6 CS 7600 Scanner Interface Overview The CS 7600 scanner interface is used primarily for scanning the exposed imaging plates and provides easy to understand prompts The scanner interface also provides the option of changing the scanner default settings as described in the CS 7600 installation guide All scanner functions are applied using the three buttons beneath the LCD The functional buttons are used for navigating the LCD screens menus and used to apply your selections 18 Chapter 3 Software Overview Scanner Interface Icons The following table lists the icons that are part of the various LCD screens F R Lu E 6 N Scan resolution Stored images counter and available memory status Blue background indicates available scanner memory Stored images counter and available memory status Red background indicates scanner memory is either nearing its full capacity or is already used up Network connection state connected Network connection state disconnected Network connection state limited connection Scan amp Go state indicator Blue indicates the scanner supports Scan amp Go smart scanning When scanning recorded imaging plates this icon blinks Scan amp Go state indicator White indicates the scanner does not suppo
17. d Procedures Clearing Scanner Memory The scanner s storage capacity is 25 images Images are stored in the scanner s memory in the following cases e When No ID is selected in the scanner s Select patient screen See Retrieving No ID Images on page 32 e When the destination workstation is not available due to network problems and so forth 30 Chapter 5 Acquiring Images When the scanner s memory nears Its limit the scan number icon changes to red Insert Plate 0 1 2 3 When memory is full error 071 Memory full Retrieve images from workstation is displayed Error 071 gq Memory full Retrieve images from workstation To clear the scanner s memory follow these steps 1 Start the CS 7600 Acquisition interface 2 Open the exam acquisition to which the images belong or alternatively create a new temporary exam acquisition in which you will save these images 3 Select the scanner it is indicated by the following full memory error state icon 4 Inthe Unclaimed panel select the images to clear CTRL left click to multi select and drag them to their correct place in the template if applicable or to the Image Gallery panel 5 If images are not needed delete them CS 7600_User Guide 8J4071 Ed01 31 Retrieving No ID Images No ID is a term for scanned images that are not associated with any exam acquisition When you scan a plate and select the No ID option in the scanner
18. display a message that the roller cleaning Is done Rollers Cleaning Cleaning progress done Please take Cleaning plate Remove the roller cleaning plate from the insertion slot Tip The cleaning procedure is to be repeated 3 times to ensure the rollers are thoroughly cleaned CS 7600_User Guide 8J4071 Ed01 51 Scan amp Go Device Care Cleaning the Scan amp Go Use disinfection wipes for medical equipment to clean the Scan amp Go device thoroughly from all sides Replacing the Scan amp Go Battery f Important Replace only with a Carestream genuine BL 5C battery To replace the Scan amp Go battery follow these steps 1 Press the On Off button for 5 seconds to turn off the Scan amp Go device A The blue LED will flash quickly 3 times B and then the device is powered off Disconnect the Scan amp Go device s USB cable C 52 Chapter 7 Maintenance 2 Open the battery compartment by carefully pressing the clip outwards and lifting out the cover A B 3 Extract the battery as shown in the figure above C D using the removal depression to lift the battery out 4 Insert the new battery by first pressing its contacts against the battery housing terminals A and then place it in the housing B 5 Press the cover firmly to close it C D in above figure 6 Connect the USB cable to the computer CS 7600_User Guide 8J4071 Ed01 53 7 Charge for 3 hours 8 V
19. eam and hygienic sheaths from Carestream The use of plates or hygienic sheaths from a third party supplier may cause a malfunction of the system and void the warranty Tip If the imaging plate has not been used for a while proceed to erase it first See Erasing the Imaging Plates on page 33 Important If you select a size 4 imaging plate make sure the scanner has the size 4 insertion slot panel installed Instructions on replacing the insertion panel are included in Cleaning and Disinfecting the Detachable Insertion Panels on page 47 There is no need to power off the system when replacing insertion panels 2 Insert the disinfected imaging plate see Disinfecting Imaging Plates on page 44 into its protective hygienic sheath with the inactive side facing the adhesive strip so the imaging plate s size number is seen through the transparent side of the hygienic sheath and the orientation mark visible in the bottom corner A Peel the adhesive strip s protective paper to seal the hygienic sheath B Handle the imaging plate by holding the hygienic sheath s empty edge C where the foam is er A em 26 Chapter 5 Acquiring Images Performing the X rays Perform the required X rays according to your clinical procedure It is recommended to continue using X ray positioning techniques and tools to ensure the resulting image is complete and includes all the information required for diagnosis It is recommend
20. ed to use the Rinn Suresoft holder or equivalent products available on your market To facilitate matching the orientation of the image in the software to the clinical reality we recommend to position the imaging plate in the mouth of the patient with the orientation mark always towards the bottom Important Make sure the active side of the imaging plate is facing the patient s teeth CS 7600_User Guide 8J4071 Ed01 27 Scanning the Imaging Plates Remove the imaging plate from the patient s mouth Do not remove the imaging plate from its hygienic sheath Wipe and disinfect the exposed imaging plate s hygienic sheath see Disinfecting the Hygienic Sheath on page 45 Keep the imaging plate in its hygienic sheath to minimize exposure to light YY ME C a LC i E N S Important If the imaging plate is exposed to light it could result in degradation of image quality Do not remove the imaging plate from its hygienic sheath until you insert it in the scanner s insertion slot At the scanner location tear open the hygienic sheath A and insert the exposed imaging plate into the scanner by positioning the open side of the sheath in the insertion panel slot B with the foam side to the right Squeeze out the tip of the imaging plate from its hygienic sheath C The imaging plate without its protective sheath will be drawn into the insertion slot by the scanner The scanner preview shows the scan progres
21. eres to the following detailed specifications and restrictions e Diluted bleach solution dilute one part 5 25 sodium hypochlorite with ten parts of water e A commercially prepared equivalent solution of diluted bleach that does not contain any ingredients Isopropyl alcohol hydrogen peroxide and other peroxides citrus based cleaners 44 Chapter 7 Maintenance hand lotions and water less hand sanitizers as well as surfactants and lubricants that can cause damage to the intraoral imaging plates Disinfection Method bleach is used follow the manufacturer s specifications and warnings A CAUTION If a commercially prepared equivalent solution of diluted To disinfect the imaging plates follow these steps 1 Moisten thoroughly but not dripping a non abrasive lint free cloth with the prepared diluted bleach and wipe the imaging plate clean 2 After 1 minute wipe off the bleach residue with a new soft lint free cloth dampened with water 3 Wipe dry with a non abrasive lint free cloth Disinfecting the Hygienic Sheath CAUTION Do not soak the hygienic sheath in any cleaning or disinfecting solutions Do not autoclave autoclaved hygienic sheaths must be discarded Frequency of Disinfection After the hygienic sheath has been removed from a patient s mouth and before extracting the imaging plate from within it Disinfection Solutions e Diluted bleach solution dilute one part 5 25 sodium hypochlorite w
22. erify the device operates press the On Off button for 1 second and verify that the LEDs function as described in the Scan amp Go Overview on page 9 9 Dispose of the old battery in accordance with all applicable federal state and local regulations 54 Chapter 7 Maintenance Troubleshooting Quick Troubleshooting The quick troubleshooting guides you through actions you need to take to correct malfunctions you may encounter and that do not require authorized technical support For errors and malfunctions not detailed in this section contact technical support The following table lists the malfunctions that are displayed on the scanner s LCD and in the Acquisition interface and details the corrective action Table 1 Quick Troubleshooting Malfunction Possible Cause Action mp e Check the network cable is connected to its RJ45 port in both the scanner and or on scanner display workstation computer Check the local network connectivity Check the computers s firewall settings and disable the firewall Replace the network cable e Check the scanner s network configuration Limited or no connection to scanner physical connection is Intact but the IP address in the Acquisition is not correct interface and or e Check that the scanner s power supply is connected e Check the network cable is connected to its RJ45 port in both the scanner and workstation computer e Check the local network
23. es using Scan amp Go to launch an acquisition 22 Chapter 4 Getting Started Starting the CS Imaging Software To start the CS Imaging Software follow these steps 1 On your desktop double click The Patient Browser tab is displayed 2 You must now create a patient card Creating a Patient Card To create a patient card follow these steps 1 Click in the Patient Browser tab The Patient Card window is displayed Patient Card Date of birth Gender Male C Female other Pregnancy status Administrative Social Security CO Patient ID G4eltade Setb 4atba7a5 1c2aaredes DPMS ID Patent Browser Contact Address f Town Zip Postal code Home phane CS 3 ea 2 Enter the patient information in the Patient Card window Note The First Name and Last Name fields are mandatory All other fields are optional Shaded fields are filled automatically and are not editable 3 When you have finished entering data click OK The Patient Card window closes and the patient now appears in the patient list 4 You can now access the Acquisition interface Accessing the Acquisition Interface To access the Acquisition interface see Accessing the Acquisition Interface on page 22 CS 7600_User Guide 8J4071 _Ed01 23 24 Chapter 4 Getting Started Acquiring Images Introduction This chapter describes the acquisition procedure for configurations that do not include the optional Scan amp
24. ging plate make sure the scanner has the size 4 insertion slot panel installed Instructions on replacing the insertion panel are included in Cleaning and Disinfecting the Detachable Insertion Panels on page 47 There is no need to power off the system when replacing insertion panels 2 Insert the disinfected imaging plate see Disinfecting Imaging Plates on page 44 into its protective hygienic sheath with the inactive side facing the adhesive strip so the imaging plate s size number is seen through the transparent side of the hygienic sheath and the orientation mark 36 Chapter 6 Acquiring Images using Scan amp Go visible in the bottom corner A Peel the adhesive strip s protective paper to seal the hygienic sheath B Handle the imaging plate by holding the hygienic sheath s empty edge C where the foam is a a 5 oN wy pe 3 Inthe Acquisition interface select the frame to which the image exposed on this imaging plate should be routed to If no frame is selected the scanned image will be routed to the first available frame in the selected template 4 Place the imaging plate on the Scan amp Go device to record the exam acquisition data An audible beep and two successive blinks of the blue LED indicate the exam acquisition information was recorded Additionally the CS 7600 Acquisition interface marks the corresponding frame as recorded frame color changes to indicate recorded status 5
25. imaging plates where they can become damaged by liquid or chemical spills e Do not autoclave CS 7600_User Guide 8J4071 _ Ed01 43 Cleaning the Imaging Plate Sheets MSDS for Kodak Screen Cleaner A CAUTION Read and follow the instructions in Material Safety Data Clean the imaging plate to remove dirt dust particles finger prints and so forth that could affect image quality Cleaning Materials e Clean dry non abrasive lint free wipes or cloths e Kodak Screen Cleaner CAT No 845 4977 Cleaning Method To clean the imaging plates follow these steps 1 Fold a non abrasive lint free wipe or cloth and dampen with a small amount of the solution Be careful not to pour the solution directly on the imaging plate Excessive amounts of the screen cleaner may damage the imaging plate 2 Wipe the imaging plate thoroughly dry with a clean dry non abrasive lint free cloth to remove residual cleaner DO NOT LEAVE THE imaging plate TO AIR DRY Apply pressure to remove persistent dirt if necessary Disinfecting Imaging Plates disinfecting solutions x CAUTION Do not soak the imaging plate in any cleaning or Frequency of Disinfection Disinfect the imaging plate after every use after the imaging plate is scanned before it is inserted in the hygienic sheath for the next use Disinfection Solutions To disinfect the imaging plates use either a self prepared solution or a commercially prepared product that adh
26. inactive side contrary to its intended use The image orientation can be corrected using the mirroring tool in the acquisition software If a diagnosis is not possible in the area of the marker the X ray should be repeated To process the scanned images from your workstation follow these steps The scanned image populates the first available frame If there is no such frame all frames are occupied applicable only when working with FMS mode the image is routed to the Image Gallery panel In image acquisition mode a new frame is added to the images layout For manual reordering of the frames drag and drop the image to the desired frame If the frame is populated with an image the former image in the frame will be moved automatically to the Image Gallery panel when the new image is dropped within it Note When exiting the Acquisition interface images stored in the Gallery panel are sent to the Imaging Software database Use the Acquisition interface processing options to optimize the image See CS 7600 Acquisition Interface Overview on page 11 Click the X icon top right corner to exit the Acquisition interface and return to the Imaging Software Images are transferred from the CS 7600 Acquisition interface to the Imaging Software automatically Storing the Images Image management storage and retrieval is performed through the Imaging Software Refer to the Imaging Software documentation Acquisition Relate
27. ion for identifying imaging plates with the Scan amp Go device Note that this scan resolution does not change the scanner s default resolution which is set via the scanner s settings By default turn the image 90 clockwise CW or 90 counter clockwise CCW to ensure correct orientation The scanned image save file format Scanned images are always saved as DICOM files and there is an option to save the scanned image also in JPEG format If needed change the workstation name by entering the new name 6 Image processing Default filter and preset Set the default filter to be applied to the acquired image The installation default is Endo Perio Periodontal filter Designed to assist in detecting disease or inflammation in the interface between the gum gingival part and the bone Endo Endodontic filter Provides overall good contrast and look of the image and high quality view of the root canals It visualizes any diseases or injuries that affect the root tips or nerves in the teeth Dej dentine to enamel junction filter Designed to improve the contrast resolution between the dentine and the enamel It enables the dentist to detect small cracks and evaluate the inspected region Preset enables setting a default image processing parameters preset that is applied after the image Is acquired e Sharp Sharpening filter for the sharpest look contrast transition emphasis e Soft Soft look smooth transition fro
28. is the common imaging platform for all our digital systems for dentistry The Imaging Software has the following features e Patient management e Image management CS 7600 Acquisition Interface Overview The CS 7600 Acquisition interface is a user friendly application designed and developed specifically for the CS 7600 intraoral imaging system It provides a rich array of image processing options in addition to the acquisition functions The Acquisition interface provides two operation modes e Full Mouth Series predefined templates for full mouth X ray images e Image Acquisition automatic dynamic addition of frames according to the number of images acquired CS 7600_User Guide 8J4071 Ed01 11 Main Window Q Tip Selected and active options are highlighted in blue PTT THATTTTTTTTT REN 12 Chapter 3 Software Overview 5 Exit Settings Scan amp Go status Scanner area Image enhancement buttons Patient information Tooth selection Click z to exit the CS 7600 Acquisition interface Click to customize the default settings See Settings on page 16 The Scan amp Go s active connection type white color indicates connection red indicates no connection USB connected Allows browsing all connected scanners if there are multiple scanners in the same network Click to list all scanners R Green frame indicates a connected scanner Red frame indicates a scanner in er
29. isition interface the image is processed and reviewed The Acquisition interface sends the images to the Kodak Dental Imaging Software which archives the acquired images The scanner connects directly to the network or is connected peer to peer to the workstation The scanner can be optionally mounted on a wall or placed on a desk counter top using a special shelf adapter The system is composed of the following elements e CS 7600 scanning device page 4 e Smart imaging plates page 6 e CS 7600 Acquisition interface page 11 e Scan amp Go device optional page 45 Note The scanner is supplied with a separate viewing and archiving software package Kodak Dental Imaging Software approved by Carestream Health Note The CS 7600 User Guide is also provided as a PDF file on the supplied CD ROM CS 7600_User Guide 8J4071 _Ed01 3 CS 7600 Components Overview Scanner Overview The CS 7600 scanner includes the functions necessary for scanning and acquiring images from the imaging plates Figure 1 CS 7600 Front View 1 3 5 QVGA color LCD 320 X 240 pixels landscape matrix 2 Function buttons for menus 3 On Off button power indication LED On Off button functions e Turn scanner On press for 1 second e Shut down close application and files and then turn off press for 2 seconds 4 Base 5 Intraoral imaging plates tray 6 Detachable insertion panel The CS 7600 makes use of two insertion panel
30. ith ten parts of water e Acommercially prepared equivalent solution of diluted bleach that does not contain any ingredients Isopropyl alcohol hydrogen peroxide and other peroxides citrus based cleaners hand lotions and water less hand sanitizers as well as surfactants and lubricants that can cause damage to the hygienic sheaths CS 7600_User Guide 8J4071 _Ed01 45 Disinfection Method bleach is used follow the manufacturer s specifications and warnings x CAUTION If a commercially prepared equivalent solution of diluted To disinfect the hygienic sheaths follow these steps 1 Moisten thoroughly but not dripping a non abrasive lint free cloth with the prepared diluted bleach and wipe the hygienic sheath clean 2 Wipe off the bleach residue with a new soft lint free cloth dampened with water 3 Wipe dry with a non abrasive lint free cloth Disposing of the Imaging Plate Important The imaging plate contains Barium and should be considered hazardous or special waste in specific conditions at the end of its useful service life For disposal or recycling information contact your local authorities 46 Chapter 7 Maintenance Scanner Care Cleaning the Scanner Body Use disinfection wipes for medical equipment to clean the scanner body thoroughly from all sides including the tray tray daily to prevent cross contamination T WARNING Itis the user s responsibility to disinfect the imaging plates
31. m high to low density e Standard Intermediate processing between sharp to soft e None Raw image with no processing filters applied Adjustment mode Imaging processing adjustment mode enables setting the optimal filter and preset combination for each filter type 7 Default exposure values Enter the default X ray exposure values to add to all scanned images attributes DICOM tags 8 DICOM Character Set Select the applicable DICOM character set encoding for the site to ensure the information is rendered correctly on DICOM viewers regardless of the locale 9 Default template selection Set the template used when opening a new exam acquisition or to change the current template applicable to a newly opened acquisition session ONLY before images have been retrieved from the scanner 10 Demo mode This option allows putting the Acquisition interface in demonstration mode for demonstration purposes Click Apply to apply your settings Click Cancel to exit to the main screen without saving changes Click OK to save the changes and exit to the main screen CS 7600_User Guide 8J4071 _ Ed01 17 Imaging plate information tab The Imaging plate Scan amp Go information fields are populated with the imaging plate s identifying information when it is placed upon the Scan amp Go device Imaging plate Scan amp Go information Serial number ZD00001 Imaging plate size amp amp type 20110724 Activation date Scan count Exam a
32. n amp Go device driver which you should do When the icon color changes from red to white connection is established 56 Chapter 8 Troubleshooting Appendix Consumables amp Accessories CS 7600 Consumables amp Accessories 1014364 1014372 1014380 1014398 1014406 1014429 1014430 1014448 1014455 1014463 1014315 1014331 1014513 1015098 1023704 1023712 1014497 1018894 1020338 1023720 P N Description CS 7600 hyg sheaths size O CS 7600 hyg sheaths size 1 CS 7600 hyg sheaths size 2 CS 7600 hyg sheaths size 3 CS 7600 hyg sheaths size 4 CS 7600 smart plate size O CS 7600 smart plate size 1 CS 7600 smart plate size 2 CS 7600 smart plate size 3 CS 7600 smart plate size 4 CS 7600 Scan amp Go pack 1 license CS 7600 Scan amp Go pack 3 licenses CS 7600 imaging plate transport box amp pad CS 7600 imaging plate pad CS 7600 slot panel size O CS 7600 slot panel size 4 CS 7600 wall mount adapter CS 7600 cleaning plate CS 7600 battery for Scan amp Go device CS 7600 quality control tool CS 7600_User Guide 8J4071 Ed01 57 Quantity 200 pcs 200 pcs 200 pcs 200 pcs 200 pcs 4 pcs 4 pcs 4 pcs 4 pcs 4 pcs J 58 Chapter Appendix Consumables amp Accessories
33. nsider lower exposure values for better image quality and diagnostics Click to delete ALL images from the Image Gallery panel J Click to delete the selected image from the Image Gallery panel w Use the brightness slider to manually adjust the image s brightness Use the contrast slider to manually adjust the image s contrast o Click to apply an automatic optimization of brightness and contrast CS 7600_User Guide 8J4071 _Ed01 15 Settings General settings tab Tooth numbering system QO European American Default scan resolution O High speed High resolution Super high resolution Default orientation 90 cw 90 CCW Output files format 4 Workstation name Q Workstation D 1 Tooth numbering system 2 Default scan resolution 3 Default orientation 4 Output files format 5 Workstation name 16 Chapter 3 Software Overview Image processing 6 Default filter Preset Default template selection Image Acquisition Mode Perio Standard FMS 10 CAM tnt Endo Sharp FMS1 tint De Soft FMS2 fint FMS3 fmt Adjustment mode FMiS4 fmt FMSS frrit Default exposure values Demo node_______ M Acquisition standalone demo mode DICOM Character Set Default Set the tooth numbering system European or American When using individual tooth selection the tooth number is stored in the image s DICOM tags Set the default scan resolut
34. on mark on the active side This dot is visible through the transparent hygienic sheath and is helpful too in correct positioning of the imaging plate in the patient s mouth Active side Inactive side Important Use only Carestream imaging plates and hygienic sheaths CS 7600_User Guide 8J4071 _Ed01 7 Hygienic Sheaths For each imaging plate size there is a corresponding one time use disposable hygienic sheath for preventing cross contamination The transparent side allows viewing the orientation mark for positioning the imaging plate correctly in the patient s mouth while the opaque side protects the active side of the inserted imaging plate from light damaging effects The foam strip is a stopper that prevents inserting the hygienic sheath by mistake into the scanner as long as the insertion direction is correct Figure 4 Hygienic Sheath front view 1 Adhesive strip for sealing the hygienic sheath 2 Foam stopper 3 Sheath transparent and light protection barrier 8 Chapter 2 CS 7600 System Overview Scan amp Go Overview Scan amp Go is an optional device that enables you to link the imaging plate to a specific exam acquisition e g patient prior to acquisition This information is retrieved by the scanner and the imaging plate is scanned automatically according to this information resolution target workstation etc Scanned images are then routed automatically to the correct workstation associated to
35. only If using a 3rd party imaging software refer to the 3rd party software documentation Gi Note This chapter is applicable to CS Imaging Software and Kodak Starting the Imaging Software Depending on your particular configuration you may have either Kodak Dental Imaging Software or CS Imaging Software installed e For KODAK Dental Imaging Software see Starting the KODAK Dental Imaging Software e For CS Imaging Software see Starting the CS Imaging Software Starting the KODAK Dental Imaging Software To start the imaging software follow these steps 1 On your desktop double click K OR From your computer click Start gt All Programs gt Kodak gt Kodak Dental Imaging Software 3 Calculator fan Kodak EasyShare printer dock gt 3 Internet Explorer K FMS Editor G Outlook Express Kodak Dental Software i Show Desktop C2 Windows Explorer Radiological Log ars Dental sortware C Windows Media Player Specific Panoramic Installation All Programs K Trophy Link g Undock Computer A Log Off QC Turn Off Computer ig Start Ep C amp A blank Patient Window is displayed 2 Create or open an existing patient record Creating a Patient Record To create a patient record follow these steps 1 Inthe Patient Window from the toolbar click OR From the menu bar select Patient gt New 2 Enter the required patient information The Last name the First name the Date of birth and the
36. ooth or teeth e Exposing the imaging plate to X ray e Removing the imaging plate from the patient s mouth e Disinfecting the hygienic sheath e Scanning the imaging plate in the CS 7600 scanner imaging plate is scanned without its hygienic sheath e Reviewing the scanned image on the Acquisition interface e Storing the image in the Kodak Dental Imaging Software CS 7600_User Guide 8J4071 Ed01 35 Preparing for Acquisition Before starting the acquisition procedure make sure you have Verified that the Scan amp Go device is connected Selected the patient record Accessed the imaging window Accessed the Acquisition interface The above procedures are detailed in Getting Started on page 21 acquisition session close it only after you have completed scanning and optimizing all images Note Keep the Acquisition interface open throughout your exam Preparing the Imaging Plates To prepare the intraoral imaging plate follow these steps 1 Choose an imaging plate of the right size for your examination Important Use only CS 7600 intraoral imaging plates from Carestream and hygienic sheaths from Carestream The use of plates or hygienic sheaths from a third party supplier may cause a malfunction of the system and void the warranty Tip If the imaging plate has not been used for a while proceed to erase it first See Erasing the Imaging Plates on page 33 Important If you select a size 4 ima
37. or for discarding an image panel when necessary Orientation b Click to rotate the selected image 90 degrees clockwise uttons s gS S Click to rotate the selected image 180 degrees Preview panel Displays the active frame s image if frame is populated Apply the required processing and comments to the preview image Full window Click to display the image in a separate full screen window dark room we mode Exposure DICOM tags of the image and can be viewed later using the Imaging Comments amp ___ y Click to add a comment to the image The comments will be added to the Values y Software or DICOM viewer inserted in the image s DICOM header Otherwise the default exposure values as defined in the General settings tab are inserted in the DICOM Click to enter custom X ray exposure values This information will be ff header 14 Chapter 3 Software Overview 16 17 18 19 Selected Image Resolution X ray exposure indication Delete Image brightness and contrast Indicates the selected image s resolution level preview image panel High speed HS High resolution HR Super high resolution SHR Partly red indicates underexposed image Review the X ray technique and consider increasing the exposure values for better image quality and diagnosis Green indicates good exposure values All red indicates the image is over exposed Review the X ray technique and co
38. ror state click the scanner icon to display the error tool tip Click to display the images in the memory of the selected scanner Click to enhance a different zone of interest 7 Click to optimize the display of periodontal tissues Perio Click to optimize the contrast over the entire image Endo i A Click to optimize the contrast at the dentin to enamel junction DEJ Displays the patient information 4A 3 NSNMAMANG j PINN or acquisition respectively When using an FMS template the tooth associated with the selected frame is highlighted green in the panoramic arch icon reflects the selected template FMS or image When using image acquisition clicking the arrow allows selecting individual teeth in either child or adult EI dentitions To clear the selection click the tooth again CS 7600_User Guide 8J4071 _Ed01 13 10 11 12 13 14 15 Imaging Imaging Software details Software Details Scan The scan resolution to apply when recording information on the imaging plates with resolution the Scan amp Go device To change the scan resolution for the current exam acquisition click the required icon before you record the imaging plate f High speed HS High resolution HR z Super high resolution SHR Frames area This area is populated by frames in a layout determined by the selected template Image gallery Useful for storing images when rearranging the layout
39. rt Scan amp Go smart scanning or that the scanned imaging plate does not contain Scan amp Go recorded information The image scan resolution High speed HS High resolution HR Super high resolution SHR Press the scanner s left function button to access the Settings menu For related Settings procedures refer to the Maintenance chapter and the CS 7600 installation guide Press the scanner s middle function button to manually erase the imaging plate See Erasing the Imaging Plates on page 33 Press the scanner s middle function button to return to the Insert Plate screen CS 7600_User Guide 8J4071 _ Ed01 19 Press the scanner s right function button to access scanner information such as scanner name IP address and total scans performed by the system and the system s set date and time Press the scanner s left function button to scroll down forwards Press the scanner s right function button to scroll up backwards Press the scanner s middle function button to confirm your selection To change the scan resolution for the current exam acquisition when acquiring images without Scan amp Go device press the right scanner function button and toggle to the required scan resolution after inserting the imaging plate If the current exam acquisition is changed the scanner reverts to its default resolution 20 Chapter 3 Software Overview 4 Getting Started Dental Imaging Software
40. s g D 4 ad SE 28 Chapter 5 Acquiring Images 3 Dispose of the used protective hygienic sheath D in previous figure 4 Select the patient from the list If the required patient does not appear in the list select the No ID option See Retrieving No ID Images on page 32 5 Verify the scanning resolution is as desired High Speed High Resolution or Super High Resolution A Select Patient No Id OWorkstationD Doe John Destination HE Bi OREX IDANM_ Imaging_Quality HS High Speed HR High Resolution SHR Super High Resolution If it is not you can change the scanning resolution for this specific scan by pressing the scanner s right function button depicted in B to toggle between the resolution levels The resolution displayed is the scan resolution 6 Press the middle function button to scan the imaging plate 7 Collect the ejected imaging plate from the lower tray after it is scanned Note After scanning the image the scanner erases the imaging plate before ejecting it The imaging plate is ready for re use CS 7600_User Guide 8J4071 _Ed01 29 Processing the Images Important The imaging plate has a marker attached to its inactive side It will indicate if the imaging plate has been exposed from the wrong side contrary to its intended use If the shadow of the marker is visible in the X ray image the imaging plate was exposed from the
41. s Select Patient screen the scanned image is stored as a No ID image in the scanners memory No ID is usually selected when the required exam acquisition is not available such as when the Acquisition interface is closed on the workstation These images are routed to the scanner s Unclaimed list for further operations To retrieve a No ID image follow these steps 1 Access the CS 7600 Acquisition interface from the patient file 2 Inthe Scanner list select the applicable scanner The list of No ID images for this scanner is displayed 3 Locate the image Each image displays its scan time and date and can be previewed to aid in the identification process Important Make sure you have correctly identified the image before retrieving it to a specific patient file 4 Click to select the correct image from the No ID list and drag it to a frame of your choice The exam acquisition s data is now associated to the image and the image is no longer No ID 32 Chapter 5 Acquiring Images Erasing the Imaging Plates The imaging plates are automatically erased and ejected after they are scanned If the imaging plate is exposed to light for an extended period of time the plate will need to be manually erased before use To erase the imaging plate manually follow these steps 1 Press the scanner s middle function button Rocm 13 Room 09 Room 135 Room 17 2 The LCD will display the Erase plate screen Erase plate 3
42. te hygienic sheath is e Reinsert the imaging plate in the right inserted in the wrong direction with the active side facing direction outwards e Turn the hygienic sheath around to switch between the open side of the sheath with the foam to the sealed side then position it correctly into the insertion slot Use the Acquisition interface to check the Scanned images do not populate Les ca ere selected scanner s unclaimed images and frame nie retrieve the required images See Retrieving y No ID Images on page 32 Imaging plate is blocked when inserting an imaging plate into the Incompatible insertion panel Check if the insertion panel size is insertion panel size compatible to the imaging plate size Artifacts black dots or smeared Scanner rollers have See Cleaning the Scanner Rollers on lines on scanned image accumulated grime page 49 e Press the Scan amp Go device On Off button for 1 5 s If the icon color changes from red to white connection is established e Disconnect the USB cable from the For Scan amp Go configurations only Scan amp Go device and reconnect tt Acquisition interface displays If the icon color changes from red to No USB connection to Scan white connection is established amp Go device e Insert the Scan amp Go device in an alternative USB port in your workstation If this is the first time the Scan amp Go device is detected you may be prompted to reinstall the Sca
43. teps 1 through 4 on page 33 Erasing the Imaging Plates Note Erasing the imaging plate manually does not erase the recorded exam acquisition information 42 Chapter 6 Acquiring Images using Scan amp Go Maintenance Imaging Plate Care If used with care Carestream Health imaging plates can be used repeatedly However the imaging plates will show wear over time due to continuous use Inspect the imaging plates regularly and replace them if they are scratched and or show signs of excessive wear CAUTION To prevent damage to the imaging plates and the possibility of image artifacts avoid contact between the imaging plates and the following materials solutions solvents Isopropyl alcohol hydrogen peroxide and other peroxides citrus based cleaners hand lotions and water less hand sanitizers as well as surfactants and lubricants Handling and Storage of Imaging Plates e Imaging plates should be stored in their original packing or storage box when they are not in use Always store imaging plates in a dark and dry place e Do not expose the imaging plates to light for long periods as this can have a degrading effect e Do not store imaging plates in hot or moist conditions e Do not fold crease or bend the imaging plates e Avoid touching the active side of the imaging plates and be careful not to drag the active side of the imaging plate across any surface as this will damage the imaging plate e Do not leave
44. the correct exam acquisition and placed in the correct frame in the CS 7600 Acquisition interface 1 USB 2 0 port 2 On Off button 3 On Off Recording status indicator LED blue Turning On Off 3 short flashes Standby intermittent flashing Recording data to the imaging plate successive short flashes 4 Battery status indicator LED amber Charging static light Low battery successive short flashes C light Off battery charged 5 Recording surface for imaging plate 6 Mounting adapter CS 7600_User Guide 8J4071 Ed01 9 Shelf The CS 7600 can be optionally mounted on the wall using an adapter 10 Chapter 2 CS 7600 System Overview 3 Software Overview Computer System Requirements Important It is MANDATORY to check that the computer system configuration is compatible with the computer minimum system requirements specified in the Safety Regulatory and the Technical Specifications User Guide If necessary you must update your computer system configuration General Software Overview The CS 7600 intraoral imaging system operates with the following software e Imaging Software e CS 7600 Acquisition interface e CS 7600 Scanner interface In addition the system can interface with patient data management software applications Imaging Software Overview The Imaging Software is a user friendly working interface that was designed and developed specifically for radiological diagnosis It
45. y of their respective holders The brand names and logos reproduced in this guide are copyright CS 7600 complies with Directive 93 42 CEE relating to medical equipment CE 0086 Manufacturer Carestream Health Inc aal 150 Verona Street Rochester NY 14 608 Authorized Representative in the European Community EC REP Carestream Health France COPERNIC Il Immeuble Le Neptune 1 Rue Galil e 93192 Noisy Le Grand FRANCE Contents Chapter 1 Conventions in this Guide Chapter 2 CS 7600 System Overview Chapter 3 Software Overview Chapter 4 Getting Started Chapter 5 Acquiring Images Conventions in this Guide 1 eoe le e sy 5 2 OENE bar hace os cma es et ae oes 2 3 CS 7600 Components Overview 4 Scanner N E 1122 dar ds drecoodesaleteot s 4 Smart Imaging Plates Overview 6 Hygienic SES 2 0 8 hi es ss doisesesesses 8 Scan amp Go OveMiEW a aides ss de ein aw sree ee uew 9 SR ns ce dc dm a 10 Computer System Requirements 14 General Software Overview 11 Imaging Software Overview 11 CS 7600 Acquisition Interface Overview 11 MeN LL Te 0 ctacwadacmddnenea ee ee aeiede donee hee 12 Re ee di E 16 General settings tab 16 Imaging plate information tab 18 CS 7600 Scanner Interface Overview