Centrifan PE Personal Evaporator User Manual

image

Contents

1. 3 2 Cold trap frozen water bottle 3 3 Configuration 1 4 D Drying rates Typical solvents 3 2 E Electrical Requirements 2 2 Environmental Conditions 2 1 Evaporation flow path 1 3 G Gas Flows 1 2 Inert gas fittings 2 6 Inspection 4 2 Installation 2 3 Introduction 1 1 L Loading samples 3 4 Location of the System 2 1 M Maintenance 1 4 4 1 N NOTE vii O Operating Symbols vii Centrifan PE Personal Evaporator Operation 3 1 S Safety Precautions iii Safety Symbols vii Shipping kit 2 2 Space Requirements 2 1 Spare Parts 1 Specifications 1 Stainless liner Cleaning 4 1 T Temperature Setting 3 6 Temperature Range 2 1 Timer Setting 3 6 Troubleshooting 4 1 U Unpacking 2 2 W Warning Symbols vii Warnings iii Warranty v D 1 Index D 2 Centrifan PE Personal Evaporator
2. be placed near a window air conditioning duct etc or any device in which significant temperature changes may occur e g an oven The relative humidity of the area should meet the ASHRE standard for human comfort e lf flammable solvents are to be used at any stage of the operation of the system suitable ventilation should be provided and open flames must be prohibited in the laboratory e Corrosive vapors solvents and dust should not be allowed in the laboratory as these can affect system performance The general guidelines provided above apply to typical installations In certain special applications such as drying proteins and peptides it may be desirable to place the unit in a refrigerator to maintain near 0 C conditions to minimize degradation The unit has been tested at the aqueous freezing point and the heater has sufficient power to raise the temperature about 35 C above ambient In other words to achieve a peptide evaporation environment of 9 C the Centrifan PE could be placed in a 4 C refrigerator and the temperature controller set for 9 C 2 3 Space Requirements The dimensions of the Centrifan PE Personal Evaporator are 12 30 5 cm W x 18 45 7 cm H x 8 20 3 cm D Some clearance to the front of the unit should be provided to set the temperature control and to operate the lid latch for the loading and unloading of sample vials Centrifan PE Personal Evaporator 2 1 Chapter 2 2 4 Electrica
3. it continues to circulate b Gently lower a frozen water bottle into the top of the cold trap and then pour a small amount of solvent denatured ethanol or ethyleneglycol into the space between the trap and the bottle to maximize thermal conduction IES Note Do not use water between the trap and the bottle because it could freeze and cause damage to the unit Centrifan PE Personal Evaporator 3 3 Chapter 3 3 3 Loading Samples in the Centrifan PE A view of the splitter separated from the rotor in the Centrifan PE is presented in Figure 3 1 Splitter Tabs Inserted into Vial Openings ya Thumbscrew Attaches Rotor to top of Spindle Figure 3 1 Internal Components of the Centrifan PE Shown Centrifan PE Model 78 0071 78 1071 3 4 Centrifan PE Personal Evaporator Operation To load samples in the Centrifan PE a Open the lid and remove the splitter from the openings of the vials sitting in the pockets of the rotor b If used vials are present in the rotor remove them and replace with new vials containing samples that are to be dried c Note Several different rotors are available for the system If the rotor is to be replaced because of different size containers remove the present rotor by unscrewing the fastener at the top of the rotor and lifting the rotor straight up out of the fan Lower the new rotor in position and turn the rotor within the fan until it drops and locks itself re
4. lid 3 5 3 Switching between Manual and Timer Modes The user can switch between the two timer module modes manual and timer by following the steps below Press and hold the SET button for about 8 seconds at which time the display will show the characters Ut Use the A or Y buttons to cycle through the four parameters Ut toP bc and MAn Press the buttons until MAn is displayed Press SET to view the current value of the MAn parameter ONzmanual mode OFF timer mode Use the A or Y buttons to toggle between ON and OFF Press SET to lock in the parameter Press the SET and Y buttons simultaneously to quit the programming mode or wait one minute and it will guit automatically Centrifan PE Personal Evaporator 3 7 Chapter 3 3 8 Centrifan PE Personal Evaporator Chapter 4 Maintenance and Troubleshooting 4 1 Maintenance This section describes a series of activities that should be performed on a routine basis to optimize the performance and ensure long term safe operation of the Centrifan PE Personal Evaporator The suggested freguency for each activity indicated below is based on the typical operation of the system and is dependent on a variety of factors such as the nature of the solvent and of the sample Over time it is likely that the user will determine an appropriate schedule that meets the specific application of system 4 1 1 Cleaning the System The exterior surfaces of
5. sound Spin the rotor again and latch the lid to insure there is no contact between the rotor components and the heatsink on the underside of the lid I Note Ensure that the system is configured for the voltage supplied by the mains The voltage configuration in indicated on the rear of the unit d Check that the power switch at the back of the system is set to off and plug the unit into an appropriate AC power source The line cord receptacle is located at the rear of the unit Open the lid again reach behind the unit just to the outside of the power cord and turn on the main power switch The temperature controller display will illuminate red indicating that the power is on Latch the lid again which will cause the rotor to begin spinning Allow the rotor about 5 seconds to achieve full speed and be aware of any unusual noise or vibration The instrument is extremely guiet and smooth running One way to know that it is running properly is to listen for the distinctive rushing sound of the trapped gas volume circulating within the enclosure When it is clear that the unit is running properly lift the lid The motor brake will stop the rotor in a few seconds The unit is now ready to install the sample containers the splitter disk and to dry samples WARNING Warning Let the centrifugal fan and rotor come to a complete stop without using your hand or any other object as a brake to stop the rotor assembly more guickly Some rotor co
6. that the centrifugal fan turns freely Centrifan PE Personal Evaporator 2 3 Chapter 2 c Gently withdraw each tab attached to the polypropylene splitter disk from the sample vials Verify that the knurled retaining nut at the center of the rotor is snugly tightened to retain the rotor within the fan Remove the vials from the rotor and the rotor is now ready to accept your samples Give the rotor assembly a good spin tentatively close the lid again and lock the latch There should be no visual or audible indication that there is any contact between the rotor and any internal parts of the enclosure Locate the large glassware clamp and slide it down on the top of the bar a few inches and loosely tighten the retaining screw Feed the cold trap down through the retaining ring and rest the curved bottom cold trap glassware in the U shaped cradle of the bracket as shown in Figure 2 3 Rotate the cold trap into its final position so that the tubing connection fittings are directed to the rear of the Centrifan PE unit Figure 2 3 Attaching the Cold Trap Assembly and Associated Tubing f Locate the three pieces of 72 ID tubing and make the connections on the back of the Centrifan PE unit as shown in Figure 2 3 Twist the plastic fittings as required to Centrifan PE Personal Evaporator Installation make the connections from the cold trap The 72 T fitting is used to attach a 20mL scintillation vial at the bo
7. the system should be cleaned with a cloth using water with a mild detergent on a periodic basis and whenever mobile phase or sample is accidentally spilled on the system If a mild detergent is not able to remove the foreign material an organic solvent such as methanol or isopropyl alcohol can be used Do not get solvent on the temperature controller or it may be damaged The interior components that are exposed to solvent vapor are composed of stainless steel and other materials that can be cleaned with a mild detergent If the solvent is not compatible with water isopropanol or a similar solvent should be employed 4 1 2 Removing the Stainless Liner for Cleaning To remove the liner for cleaning a Make sure the power switch is turned off and unplug the line cord b Remove the splitter from the vial openings on the top of the rotor c Unscrew the thumbscrew from the center of the rotor and carefully lift the rotor out of the center of the fan using care not to apply excessive force d Reach down inside the fan and turn the black nut on the driveshaft counterclockwise to loosen the fan If the driveshaft turns with the nut and slips in the fan hub press lightly on the top rim of the fan in the front left corner while unscrewing the nut The uneven force on the fan will keep the driveshaft from turning Remove the nut and then lift the fan assembly straight upwards off of the driveshaft e At the rear of the unit pull off from each fittin
8. In others the condensate may be extraction solvent suitable for recycling I Note If you are evaporating different solvents in consecutive runs rinse out the flask to ensure that the previous solvent is not present if there is the possibility of chemical interactions between the solvents me Note The Centrifan is designed so that the solvent is retained in the vials and vapor is rapidly swept from the system The components inside the Centrifan are constructed from polypropylene stainless steel and other materials with solvent resistant coatings so that the possibility of corrosion due to the evaporated solvent is minimized In the unlikely event of liquid spilling into the Stainless liner mop up the liquid as soon as possible and rinse with water or a miscible solvent such as isopropanol Setting the Temperature When the system is powered up the temperature controller will indicate the ambient temperature To change the temperature a Press the SET button 2 two times to display the present set point b Use the A or Wbuttons to display the desired setpoint temperature c Press the SET button again to accept the indicated temperature as the set point c Note If you do not press the SET button within one minute the display will revert to the ambient temperature and the edited setpoint change will be lost Setting the Timer The Centrifan PE T version of the Centrifan product includes a timer module designed to shu
9. cation interfaces tape drives and or all similar devices is in all cases limited to ninety 90 days from the date of shipment to Buyer Replaced products assume the ongoing warranty of the originally purchased instrument with the above exceptions Centrifan PE Personal Evaporator Safety Operating Symbols The following symbols appearing in the manual are defined as follows The WARNING statement used throughout the manual presents dangers that could result in personal injury WARNING The CAUTION statement used throughout the manual presents hazards on conditions that could cause damage to the instrument or the reporting of m erroneous results I The NOTE statement used throughout the manual highlights important information about the instrument and its use Failure to follow these statements may invalidate the warranty Centrifan PE Personal Evaporator vii Warranty viii Centrifan PE Personal Evaporator Table of Contents Warnings and Safety Precautions ccceccceeeeeeeeeeeeeee eee ceeeeeeeeeeeeeeaaeeeeeseeeeeeeseesaesseaeeaeeeseeeeseseeeeenees iii curl nian eect MAA SPIKA MASWA MNA NIA KAA KUSAKA AWA aie pd inten NA KUA Sa SARA SAMANI ANA KESS AAA V Safety Operating Symbols eese eene nennen ennt nnn KUKAAA nnne annu nnne n innen nnne nnne nnn n n nnns vii Table of Contents IMMER ix Chapter 1 Introduction iere rhet ee rerit ignes 1 1 1 4 SOVervVieWrs en sees otl sese e
10. dent This warranty extends only to the original consumer purchaser IN NO EVENT SHALL KD SCIENTIFIC BE LIABLE FOR INCIDENTAL OR CONSEGUENTIAL DAMAGES Some states do not allow the exclusion or limitation of incidental or conseguential damages so the above limitation or exclusion may not apply to you THERE ARE NO IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE OR OF ANY OTHER NATURE Some states do not allow this limitation on an implied warranty so the above limitation may not apply to you If a defect arises within the warranty period promptly contact KD Scientific 84 October Hill Road Holliston Massachusetts 01746 at 1 508 429 6809 Email address is info kdscientific com Goods will not be accepted for return unless an RMA returned materials authorization number has been issued by our customer service department The customer is responsible for shipping charges for non warranty repairs Please allow a reasonable period of time for completion of repairs or replacement If the unit is replaced the replacement unit is covered only for the remainder of the original warranty period dating from the purchase of the original device This warranty gives you specific rights and you may also have other rights which vary from state to state Repair Facilities and Parts KD Scientific stocks replacement and repair parts When ordering please describe parts as completely as possible preferably using a part number obtained
11. e Acetone Isopropyl Alcohol DMF Ethyl Acetate Methylene Chloride Acetonitrile AcN H20 70 30 Rotor Key Rotor Total Volume in Rotor 6 X 10mL 6 X 5mL 6 X 10mL 6 X 10mL 6 X 10mL 6 X 1mL 6 X 10mL 6 X 10mL rrrrryrrYr ey 8 X 1mL 8 X 1mL 8 X 1mL 8 X 1mL 8 X 1mL 8 X 1mL 8 X 1mL 8 X 1mL 8 X 1mL U V UU WWW WD DW Temp Time to Dry C 40 40 40 40 40 40 40 40 40 40 40 40 40 40 40 40 40 6 x 20mL Scintillation Vial 8 x 1 6mL Microcentrifuge Tube 8 x 16mm x 100mm Test Tube Rotor 60 min 3 hrs 15 min 35 min 70 min 12 hrs 25 min 3 hrs 65 min 15 min 20 min 50 min 4 hrs 25 min 18 min 35 min 3 hrs Centrifan PE Personal Evaporator Operation 3 2 2 Frozen Water Bottle Convenience with Higher Boiling Solvents a Ifthe solvent to be condensed has a higher boiling point such as Water IPA DMF or DMSO frozen water is probably sufficiently cold to condense the vapor from the recirculating gas and give a good evaporation rate Freezing an appropriately sized water bottle is a convenient no mess way of condensing most solvents in the trap Note The amount of vapor removed from the recirculating gas as it passes throught the trap and around the internal cold finger is dependent upon the temperature of the trap surfaces and the residence time of the vapor The dryer the gas which is returned to the Centrifan the more vapor it will be able to carry to the trap as
12. e Any Part Of The Instrument Which Requires A Tool There are no user serviceable parts inside the base of the instrument enclosure or under the heater assembly in the lid It is not necessary to remove any enclosure screws to perform the cleaning procedures envisaged for the user of the instrument Wear Protective Eyewear Reagents used in the system can damage your eyes Protective eyeware should be worn by the operator The system should be used in proximity to an eyewash sink If any solvents or chemicals splash on the skin or eyes immediately rinse the affected parts in the sink Obtain medical help as needed Centrifan PE Personal Evaporator iii Warnings and Safety Precautions 10 11 12 Use the System In A Proper Manner Do not use the instrument and or its accessories in a manner not specified by KD Scientific If you do so the protection provided by safety interlocks may be impaired Do Not Attempt To Bypass Any Safety Interlocks The safety interlocks are provided to comply with safety reguirements of various regulatory agencies and must be employed to protect the operator The lid should be fully down and the latch engaged when the instrument is in use Do not use your hands or any implement to slow the rotor at the completion of an evaporation process Allow the rotor to come to rest without interference The motor is eguipped with a braking function which is engaged when the latch is disengaged and will stop th
13. e rotor in just a few seconds Wear Gloves When Handling Dry Ice Dry ice can cause serious burns to the skin Casually dropping dry ice nuggets into the trap can splash the ethanol liguid out of the internal cold finger Use care when adding dry ice to not cause a spill Ensure That Circulation Exists Between The Cold Trap Ports The function of the Centrifan PE is to accelerate evaporation of sample solvent The solvent must be condensed out of the recirculating gas volume within the instrument or continued evaporation will not occur The outlet port at the back left of the instrument and the inlet port at the back right should be connected in series with a dry ice trap or a mechanical cold trap to obtain maximum green solvent recovery benefit Avoid use of acidic solvents The standard system is not designed to handle acid vapors Extended use of acidic reagents can cause degradation of internal parts i e rusting will occur on stainless steal Liner Use of acidic solvents will void warranty Contact the factory for information on ordering an acid proof system Centrifan PE Personal Evaporator Warranty KD Scientific warranties this instrument for a period of three years from date of purchase At its option KD Scientific will repair or replace the unit if it is found to be defective as to workmanship or materials This warranty does not extend to damage resulting from misuse neglect or abuse normal wear and tear or acci
14. elative to the tube opening DAT ETL This will prevent the cap from potentially slowing the drying rate of the adjacent tube to the right by blocking the tube opening from the gas flow This cap location will also assure that the heater fins will not contact the tubes when the lid of the Centrifan is closed Centrifan PE Personal Evaporator 3 Instructions for cutting splitter tabs to accommodate Microcentrifuge tube liguid volumes Please note that the gasflow splitter for the Microcentrifuge rotor is diecut with long tabs to maximize the drying rate for the last few microliters in the tip of a tube However it is not effective to use the long tabs when there is more liquid present because if the tab contacts the liquid then the evaporative gas flow is effectively blocked In these instances where more liquid is present in the tubes a second splitter has been provided with the Microcentrifuge rotor so that the tabs may be cut shorter to suit the specific liquid volume requirements The two illustrations show the original splitter with the long tabs above and a second modified splitter which shows the tabs shortened for use with tubes containing volumes of 1 mL below Additional splitters long tabs may be ordered using part number 78 8410 B 4 Centrifan PE Personal Evaporator Appendix D Index A Accessories 1 4 1 Assembly 2 3 C Cleaning 4 1 Cold trap assembly 2 4 Cold trap dry ice
15. et EE Y e iab atk dal Gone 1 1 1 2 Mode of iii mide bt baee a es bet CURE 1 2 13 Benefits of the Centrifan M PE 1 4 1 4 System Configuration and Accessories ssssssssee een 1 4 1 5 Contents of this Manual sss errem eene 1 4 1 6 ForFurtherInformation outre es tarot tete sg YR GYD car dere 1 5 Chapter 2 Installation niece reete rec ce rea eo KA EEEH n rin arte iden On dn Fan ens 2 1 24 OV rVIew su oett ete rode out odas e obtu t ESI cocco d 2 1 2 2 LOCATION ofthe System asset e pedes aa 2 1 2 3 Space Requirements iet taeda eee de pU Ie bec dtes egeta 2 1 2 4 Electrical Require Ments issin eea aena na ta a eoe a e Raa 2 2 2 5 Unpacking the Systems e a a e 2 2 2 6 Assembling the System a yasin na E 2 3 2 7 Connections for Evaporation with Inert Gas 2 ccceceeceeeeeeeeeeeeeeeeeneecaeeeeeeeeess 2 6 le IETICTERNSDIITLCMTgI WR Pr 3 1 3 1 Preparing the Centrifan for Operatiol sei eb ee Sheet he SON 3 1 9 2 SeteUp the Cold TEBpzs suot et AA dud seb o Mod Ld al 3 2 not included in Centrifan PE Lite Mense 3 2 3 2 1 LOW Boiling SolVentSu uii repere ee Pee Poe da eid 3 2 3 2 2 Frozen Water Bottle Convenience with Higher Boiling Solvents 3 3 3 3 Loading Samples in the Centrifan PE niei iet iron iate paid 3 4 3 4 Setting the Temperature ee trier pene Leti tee obrero diee addens 3 6 9 9 oett
16. from our Customer Service department If practical enclose a sample part or sketch We offer a complete reconditioning service Serial Numbers All inquiries concerning our product should refer to the serial number of the unit Centrifan PE Personal Evaporator Warranty vi Limitations Products are warranted to be free from defects in materials and workmanship for parts and labor to the original purchaser with the exceptions indicated below e Any components of the system that are in direct contact with liquids are warranted to be free from defects in materials and workmanship at time of delivery but cannot be further warranted due to the unknown nature of the use of the product Materials of construction should be compatible with the compounds and solvents being processed e Consumable items such as Tubing Fittings Test Tubes and Glassware are excluded from this warranty The rotor insert which holds the solvent containers and the rotor cover cannot be warranted due to their fragile nature e Loss damage or defects resulting from transportation to the Buyer s facility improper use software or interfaces supplied by the buyer unauthorized modification or operation outside the environmental specifications for the instrument use by unauthorized or untrained personnel or improper site maintenance or preparation e The sole and exclusive warranty applicable to software and firmware products provided by KD Scientific for use
17. g the connected tubing so that the fittings are free to be unscrewed f Unscrew the two right angle fittings on the back of the unit which actually pass through the space between the outer wall of the Centrifan PE and thread into the stainless liner to form a sealed volume for the recirculating gas g Lift the liner from the outer enclosure and wash or sterilize as desired h When the liner has been cleaned replace the liner by repeating the above procedure in the reverse order Centrifan PE Personal Evaporator 4 1 Chapter 4 Figure 4 1 Remove Nut Fan and Two Fittings to Free Liner for Cleaning 4 1 3 Routine Inspection On a periodic basis the following activities should be performed a Inspect the tubing between the evaporator module and the cold trap If the tubing is cracked deformed or solid material is deposited inside the tubing replace the tubing b Inspect the inert gas supply tubing or the jumper tube If the tubing is cracked deformed or solid material is deposited inside the tubing replace the tubing c Inspect the rotor components for signs of damage Replace as required d Replace the solvent in the cold trap 4 2 Troubleshooting es Note There are no components inside the system that require any user interaction If the system does not operate satisfactorily please contact the manufacturer or your local agent AZ Centrifan PE Personal Evaporator Appendix A Specifications R
18. ifan PE 8 x 4mL 1 Dram Vials 78 8403 Rotor Centrifan PE 10 x 1 5mL HPLC Vials 78 8406 Splitter Centrifan PE 4 x 40mL Scintillation Vials 78 8415 Splitter Centrifan PE 6 x 20mL Scintillation Vials 78 8409 Splitter Centrifan PE 6 x 16mm x 100mm Test Tube 78 8411 Splitter Centrifan PE 8 x AmL 1 Dram Vials 78 8412 Splitter Centrifan PE 6 x 30mL Scintillation Vials 78 8414 Splitter Centrifan PE 8 x 1 6mL Eppendorf Tubes 78 8410 Splitter Centrifan PE 10 x 1 5mL HPLC Vials 78 8416 Cold Trap Kit Centrifan PE to add cold trap to PE L Lite version 78 8417 Cold Trap Centrifan PE 78 8418 Circulator Coil 78 8419 Flask Round 500mL 35 20 Socket Joint fits Centrifan PE Cold Trap 78 8421 Flask Round 100mL 35 20 Socket Joint fits Centrifan PE Cold Trap 78 8422 Adapter Kit Cold Trap Socket to 20mL Scint Vial low loss distillate collection 78 8423 Clamp Pinch Locking 35 20 Socket Joint 78 8424 Manual Centrifan PE 78 8425 Centrifan PE Personal Evaporator B 1 B 2 Centrifan PE Personal Evaporator Appendix C Operator s Note Centrifan PE 8 x 1 6mL Microcentrifuge Rotor Instructions for positioning tube caps when loading the Microcentrifuge tube rotor The photo below shows the proper orientation of the caps of the Microcentrifuge tubes after they have been loaded into the Centrifan PE rotor Rotate each cap clockwise until it reaches the two o clock position r
19. ingdheTllnela aiii iei digid nats e de tele ten Hee Pad sgen ie dte caben 3 6 3 5 1 Operating in Manual Mode sssss emere 3 6 3 5 2 Operating in Timer Mode ia ka emen 3 6 3 5 3 Switching between Manual and Timer Modes eeeeeese 3 7 Chapter 4 Maintenance and Troubleshooting eeueeseeseesesiesseeseeseeen enne nnne 4 1 Ad Maintenances sooo a T eate bbc o pute T YY MA LEL 4 1 41 4 Cleaning the System e eerte hd ee E Peer tt Paneg 4 1 4 1 2 Removing the Stainless Liner for Cleaning seeeeeees 4 1 A 1 3 Routine Inspection uu Nee ie Sieb ede KA dtes YND 4 2 4 2 Droubleshooting eee ent Reti when dnr Maa Ee eR ea 4 2 AppendbcA Specifications 5 iunc eiit feet nete Hoe ea Eae ER Y GUDD CR DE CE ER ua ao Niais etina 1 Appendix B Spare Parts and Accessories eeeececeeeeeeeeeeeeeaeeeeeeeeeeeeeseeeeeeaseaeeeseeeeeeeeeeeeeseeeneaees 1 Appendix D Index eere E tenax en rer rx uen scLure ees aa 1 Centrifan PE Personal Evaporator ix Table of Contents x Centrifan PE Personal Evaporator Chapter 1 Introduction 1 4 Overview The Centrifan PE Personal Evaporator Figure 1 1 employs an innovative recirculating evaporation technology to evaporate both agueous and organic solvents to allow for the isolation of non volatile solutes from liquid samples It replaces conventional blow down systems and ensure
20. kd Scientific Centrifan PE Personal Evaporator User Manual Shown Centrifan PE T Model 78 0070 with Circulator Coil Accessory KD Scientific 84 October Hill Road Holliston MA 01746 Phone 508 429 6809 Fax 508 893 0160 Part 78 8425 Manual www kdscientific com Notices This system is covered by a limited lifetime warranty A copy of the warranty is included with this manual The operator is not reguired to perform routine maintenance but procedures for disassembly and cleaning are as described herein For cleaning procedures refer to Chapter 4 All information in this manual is subject to change without notice and does not represent a commitment on the part of KD Scientific The system and various components in the system are the subject of various pending patents Modular SFC was acquired by Harvard Bioscience in May 2012 KD Scientific is a division of Harvard Bioscience No part of this manual may be reproduced or transmitted in any form or by any means without the written permission of KD Scientific All trademarks of other companies are acknowledged to the property of the respective owners Centrifan is a trademark of KD Scientific Printed in the United States of America Warnings and Safety Precautions The following precautions should be followed to minimize the possibility of personal injury and or damage to property while using the Centrifan PE 1 Maintain a Well Venti
21. l Reguirements The Centrifan PE Personal Evaporator is provided in two versions for either 115VAC or 230VAC Both versions operate with 50 60Hz and 200 watts A label on the rear panel indicates the version and the user should ensure that the power reguirements match the power provided by the mains The AC voltage connection to the system must include an earth ground for operator safety The system should not be connected to a power line that is susceptible to significant changes in power reguirements such as a power line that is connected to a compressor oven or other source of heat generation If it is possible that significant changes in the power may occur a constant voltage transformer may be reguired The system should not be installed near eguipment that generates strong electromagnetic fields If the power line is noisy it may be necessary to install a noise filter These general electrical guidelines are far more applicable to a microprocessor based instrument with analog data processing capability the Centrifan PE with only a temperature controller module performing any analog control functions should be immune to all but the most severe power conditions 2 5 Unpacking the System The Centrifan PE Personal Evaporator and its accessories are shipped in the same container Unpack components carefully making sure all items in the list below have been included and are in good condition Save the container and packing mate
22. lated Laboratory Volatile organic solvents may be used in the system Ensure that the laboratory is well ventilated so that a buildup of vaporized solvent cannot occur Ensure That Solvent Vapor Is Not Allowed To Escape Into The Laboratory Any organic solvent vapor must be vented to an external venting system so that it cannot collect in the laboratory Avoid Open Flames And Sparks Volatile organic solvents may be used in the system Do not use an open flame in the laboratory and do not install any eguipment that can cause sparks or other ignition sources in proximity to the instrument The System Must be Plugged Into A Grounded Power Receptacle Ensure that all parts of the system are properly grounded Do not attempt to bypass the earth ground connection A serious shock hazard could result Treat All Samples And Solvents As If They May Contain Hazardous Substances Or Transmit Disease The samples and or solvents that are handled by the system may contain compounds that could present a hazard to the operator of the instrument Take all precautions to ensure that they do not come into contact with the skin or eyes A sink and eyewash should be located in the laboratory In the event of an accident wash the affected part with copious quantities of water and seek medical assistance If you are handling biological clinical samples treat them in accordance with the infectious disease control program of your institution Do Not Disassembl
23. lative to the fan When the rotor and fan can only turn together reinstall the screw without overtightening WARNING Warning Make sure to rotate the rotor within the fan until the rotor drops the last half inch onto the top of the driveshaft The flat at the top of the driveshaft must align properly with matching flat or dowel pin inside the hole in the bottom of the rotor If the rotor is not properly seated on the driveshaft the rotor components will not be sitting low enough inside the fan assembly and they could strike the heat sink fins on the underside of the lid when it is closed This could cause significant damage to the rotor and splitter C Insert the vials into the positions in the rotor located inside the centrifugal fan The rotor should be balanced within a three or four milliliters by locating vials containing similar amounts of liquid on opposite sides of the rotor If there are less than the maximum number of samples to run insert empty vials into the unused locations in order to make sure the splitter fits properly on top of the rotor and to preserve drying gas flow patterns Caution For safe operation the system should be balanced Ensure that there is a vial with the equivalent amount of material at a position 180 from each vial placed in the rotor In addition it is recommended that all unused positions on the rotor are filled with empty vials d Insert the tabs of the splitter into each vial e Ensure that
24. lvent vapor will be sent into the atmosphere If an organic solvent is employed this may lead to environmental and or regulatory issues even if the solvent is passed into a fume hood The Centrifan PE condenses the solvent vapor in a cold trap and allows for safe disposal or recycling e A vacuum pump is not required reducing both procurement and operating costs and eliminating a major source of downtime in vacuum concentration systems Prefreezing of samples required by lyophilization is not required when using the Centrifan PE e The very small portable size of the Centrifan PE makes it suitable for flexible use in virtually any one square foot space 1 4 System Configuration and Accessories The system can be configured with a variety of rotors to accommodate 78 8400 6x20 mL Scintillation Vials 78 8404 6x30 mL Scintillation Vials 78 8405 4x40 mL Scintillation Vials 78 8401 8 x 1 6 mL Microcentrifuge Tubes 78 8403 8x4mL 1 dram Vials 78 8402 6x 16mm x 100mm Test Tubes 78 8406 10x 1 5mL HPLC Vials The optional cold trap cover is no longer available although in our experience Parafilm over the top of the trap with a small perforation may be just as effective at slowing the evaporation rate of the alcohol or acetone being used in the trap 1 5 Contents of this Manual This Operator s Manual contains e Installation Chapter 2 describes the facilities required for the system and how to make the inert gas purge connecti
25. nfigurations contain plastic and are relatively fragile and can be damaged if they contact a solid object sharply Replacement of carrier components is not covered under the warranty Centrifan PE Personal Evaporator 3 1 Chapter 3 3 2 Set Up the Cold Trap not included in Centrifan PE Lite version 3 2 1 Low Boiling Solvents a Fill cold trap with approximately 1L of the desired solvent into the top of the dry ice trap until the level is about 15cm deep In most cases denatured ethanol is an appropriate solvent c Note The solvent that should be used in the cold trap is dependent on the solvent that is being removed from the solvent If the temperature of the cold trap is too low it is possible that the evaporated solvent may freeze in the trap or tubing and cause a blockage thereby restricting the flow of recirculating gas in the system 3 2 b Gently drop several dry ice pellets into the trap to begin the chilling process of the cold finger at the center of the trap Add enough dry ice to maintain the cold temperature of the trap for the period of time required for the drying operation Replace the cover of the trap to minimize the loss of the solvent through evaporation Typical drying rates of some common solvents are presented in Table 3 1 Table 3 1 Drying Rates for Typical Solvents Solvent Methanol Water Hexane Acetone Isopropyl Alcohol DMSO Methylene Chloride AcN H20 70 30 Methanol Hexan
26. ngth of 1 8 ID tubing which will serve as the means of venting an egual gas volume to that being introduced through the other fitting This arrangement maintains approximate atmospheric pressure within the enclosure at all times d To visualize and help set the low volume of gas introduced to the system it may be helpful to dip the end of the vent tube in a beaker of water and regulate the minimal inert gas bleed by monitoring the trickle of bubbles within the beaker Outlet for Inert Gas Figure 2 4 Inlet left and Vent right Connections for Inert Gas Evaporation 2 6 Centrifan PE Personal Evaporator Chapter 3 Operation 3 4 Preparing the Centrifan for Operation To prepare the system for initial operation I Note The Centrifan PE and the attached cold trap should be placed in a convenient location for loading and removing sample vials and for the occasional addition of dry ice to the cold trap a b Insure that all hoses are routed correctly and securely pressed onto their fittings Unlatch the lid and open the instrument for rotor access Remove the splitter disk and any vials Make sure the thumbscrew in the center of the rotor is reasonably but not overly tight Turn the centrifugal fan containing the rotor and look inside the Centrifan PE enclosure to make sure there is no foreign material in the unit which could cause damage to the eguipment or your samples The fan should turn freely and make no
27. ons if reguired e Operation Chapter 3 describes how to install and secure sample vials prior to running e Maintenance and Troubleshooting Chapter 4 describes how to remove the stainless liner for cleaning 1 4 Centrifan PE Personal Evaporator Introduction e Specifications Appendix A describes general system specifications e Spare Parts Appendix B lists consumable items and replacement parts for the system 1 6 For Further Information For further information please contact KD Scientific 84 October Hill Road Holliston MA 01746 Phone 508 429 6809 Fax 508 893 0160 www kdscientific com or your local KD Scientific Authorized Distributor Centrifan PE Personal Evaporator 1 5 Chapter 1 1 6 Centrifan PE Personal Evaporator Chapter 2 Installation 2 1 Overview c Note The Centrifan PE Personal Evaporator is designed to be installed by the user This chapter describes general information about the location and the physical reguirements of the system 2 2 Location of the System The Centrifan PE Personal Evaporator should be installed in a facility with the following environmental conditions e The temperature range of the laboratory should be between 15 C and 25 C 59 F 77 F The system should be installed in an area in which the temperature range is fairly constant the suggested temperature variation is 2 C hr 5 F hr e The system should not
28. otor Capacities 6 x 20mL Scintillation Vials 6 x 30mL Scintillation Vials 4 x 40mL Scintillation Vials 8 x 4mL 1 Dram Vials 8 x 1 6mL Microcentrifuge Tubes 6 x 16mm x 100mm Test Tubes 10 x 1 5mL HPLC Vials Temp Controller Range Ambient to 55 C Ambient Temperature 2 C to 40 C Cold Finger Capacity 1 L non freezing liquid plus dry ice nuggets Power Reguirements 115 230VAC 50 60Hz 200 watts Fuse Ratings 3A 115V and 230V fast acting 5mm x 20mm Certifications Dimensions 12 30 5 cm w x 8 20 3 cm d x 18 45 7 cm h Weight 12 Ibs 5 5 kg Centrifan PE Personal Evaporator A 2 Centrifan PE Personal Evaporator Appendix B Spare Parts and Accessories Assembly or Part Description Part Number Centrifan PE T Includes Trap amp Timer 115V 230V version is 78 1070 78 0070 Centrifan PE Includes Trap 115 V 230V version is 78 1071 78 0071 Centrifan PE L Lite version No Trap 115V 230V version is 78 1072 78 0072 Rotor Centrifan PE 6 x 20mL Scintillation Vials 78 8400 Rotor Centrifan PE 8 x 1 6mL Microcentrifuge Tubes 78 8401 Rotor Centrifan PE 6 x 16mm x 100mm Test Tube 78 8402 Rotor Centrifan PE 4 x 40mL Scintillation Vials 78 8405 Rotor Centrifan PE 6 x 30mL Scintillation Vials 78 8404 Rotor Centr
29. rial for convenient and secure relocation of the instrument The Centrifan PE Personal Evaporator is shipped with the following items e Centrifan PE unit e Rotor including splitter for installation instructions see page 3 5 e Cold trap glassware e 500mL round collection flask e Cold trap holder e Clamp for ball joint e Line cord e Manual I Note Not included with Centrifan PE L Lite versions models 78 0072 and 78 1072 2 2 Centrifan PE Personal Evaporator Installation The Centrifan PE Personal Evaporator has been carefully packed and shipped to ensure that it is received in proper condition Any damage to the container or its contents should be reported immediately to your local distributor or to KD Scientific Please refer to the warranty page v Warranty Returns and Repairs for more information If replacement parts are needed please refer to Appendix B Replacement Parts for part numbers 2 6 Assembling the System The interior of the Centrifan PE Personal Evaporator is shown In Figure 2 1 Splitter Tabs Inserted into Vial Openings Thumbscrew Attaches Rotor to top of Spindle Sample Vials Placed in Rotor Figure 2 1 Internal Components To install the system a Remove the Centrifan PE unit from its carton set it on a suitable work surface and open the lid by lifting the spring loaded latch Remove any packing material present and confirm
30. s safe rapid evaporation of solvent while ensuring sample yield and purity and optimizing solvent recovery Figure 1 1 The Centrifan PE Personal Evaporator Shown Centrifan PE T Model 78 0070 with Circulator Coil Accessory Centrifan PE Personal Evaporator 1 1 Chapter 1 1 2 Mode of Operation The operation of the Centrifan PE involves spinning of the liquid samples via a rotor and directing a gas flow from a fan over the solvent surface to accelerate evaporation The vaporized solvent is swept from the sample area and directed to the cold trap where the vapors are condensed The solvent free gas is then returned to the Centrifan PE enclosure to pick up more solvent vapor The two circulation paths are traced in a basic functional diagram of the instrument depicted in Figure 1 2 Dried Gas s 5 Recirculating Gas amp Vapor Figure 1 2 Evaporation and Condensation Gas Flows The actual circulation paths are traced through a cross section of the instrument depicted in Figure 1 3 1 2 Centrifan PE Personal Evaporator Introduction SECONDARY RECIRCULATING CONDENSATION GAS HEATER o T INSIDE LID CENTRIFAN y CENTRIFAN DRIVES GAS FLOW THROUGH EVAPORATION LOOP Figure 1 3 Evaporation Path in Green Condensation Path in Black There are two gas flows in the system a A primary or evaporation high flow path of recirculating gas removes solvent molec
31. t off the unit and stop the evaporation process at a pre determined time It also has a manual mode which allows the timer module to function like an on off switch for the unit Operating in Manual Mode In the manual mode the display will show the status of the switched output of the timer as either ON or OFF a Press the SET button to toggle the output between the ON and OFF states Operating in Timer Mode If the display is showing OFF and pressing the SET button causes the number 0 to appear then the module is in the timer mode To confirm that the module is in the timer mode no lighted dot will be visible at the lower left corner of the display a Use the A or Y buttons to set the desired amount of time in minutes that the Centrifan is to evaporate the samples b When the display shows the desired run time for the samples press the SET button again to turn the module ON the display will momentarily show RUN and then display the time remaining until the module shuts off the Centrifan While the timer Centrifan PE Personal Evaporator Operation is counting down the minutes a lighted dot will blink in the top left corner of the display To interrupt the evaporation the timer module can be deactivated by holding down the SET button for 1 second The unit will beep and display End and OFF showing that the time interval has been deleted I Note To immediately shut off and stop the Centrifan just open the
32. the power switch at the rear panel AC line cord connector is set to on f Ifan inert gas is used to displace oxygen from the unit during the drying operation make certain that the gas is turned on and that the vent tube is located in a hood Dipping the end of the vent tube in a beaker of water will show the slow trickle of gas flowing from the enclosure and confirm that excess inert gas is not being consumed unnecessarily The spinning fan generates the blow down gas flow not the gas introduced into the unit through this oxygen displacement tubing g Close and latch the lid to start the fan and rotor assembly and the drying process Check the temperature controller to be sure the desired temperature has been entered see Section 3 4 for operation of the temperature controller Set the desired temperature using the buttons on the temperature display The temperature can be set to a maximum of 55 C If the controller is set to a temperature below ambient the enclosure will drop several degrees below ambient due to evaporative cooling effects and the decreased temperature of the dried gas returning from the cold trap Centrifan PE Personal Evaporator 3 5 Chapter 3 3 4 3 5 3 5 1 3 5 2 3 6 h On a periodic basis empty the round bottom flask The waste solvent should be disposed of in a manner that is consistent with local environmental regulations In some applications the round bottom flask may contain the desired product
33. ttom of the cold trap return tubing This vial helps prevent any condensate which forms in the cold trap exit tube from re entering the enclosure and slowing the evaporation process g If your application will not require drying in an oxygen free environment leave the jumper piece of 1 8 ID tubing between the two small barbed fittings as shown in Figure 2 3 c Note If you require evaporation with an oxygen free environment see the instructions for adding the tubing designed to purge the enclosure with inert gas in Section 2 7 Centrifan PE Personal Evaporator 2 5 Chapter 2 2 7 Connections for Evaporation with Inert Gas If you desire to purge the drying enclosure of air by bleeding dry nitrogen or another gas into the process as it runs then procure a length of 1 8 ID tubing It should be long enough to connect from an inert gas of choice to the back of the Centrifan PE unit and also for providing a vent connection to any convenient exhaust system such as a fume hood or elephant trunk a Locate the Centrifan PE a convenient distance from both the supply of inert gas and the hood or exhaust trunk which will be used to vent the system b The 1 8 barb fitting screwed into the larger 77 fitting closest to the cold trap should be connected to a nitrogen argon or other inert gas supply at a very low pressure c The 1 8 barb fitting located on the fitting above the power connector should be connected to a le
34. ules from the liquid surface of the samples which are contained in a rotor right yellow pathway of Figure 1 3 b A secondary or condensation low flow loop continuously transfers gas with the solvent vapor to a cold trap for condensation and collection After the solvent is removed the gas is returned to the evaporation chamber left green pathway of Figure 1 3 A heater in the lid maintains the desired gas temperature the evaporation of solvent will cool the sample to below ambient unless thermal energy is added If desired the user can bleed in an inert gas such as Ar or N to displace the air in the captured gas volume and protect samples from oxygen degradation Centrifan PE Personal Evaporator 1 3 Chapter 1 1 3 Benefits of the Centrifan PE The Centrifan PE system provides a number of benefits compared to commonly employed methods of solvent operation including e The centrifugal force obtained by rotation of the samples forces the solute to the bottom of the collection vessel In contrast when a stream of gas is passed over the sample it is possible that solute will be distributed into the atmosphere When the Centrifan PE is employed the dried sample remains in the bottom of the vial e f a blow down apparatus is used a significant amount of compressed air or an inert gas is required The Centrifan PE generates its own gas flow via the internal centrifugal fan e lf a blow down apparatus is used so
35. with a processor internal or external to the Product will be as follows KD Scientific warrants that such software and firmware will conform to KD Scientific s program manuals or other publicly available documentation made available by KD Scientific s current at the time of shipment to Buyer when properly installed on that processor provided however that KD Scientific does not warrant the operation of the processor or software or firmware will be uninterrupted or error free e Products that have been altered or repaired by individuals other than KD Scientific personnel or its duly authorized representatives unless the alteration or repair has been performed by an authorized factory trained service technician in accordance with written procedures supplied by KD Scientific e Products that have been subject to misuse neglect accident or improper installation The warranty herein applies only to Products within the country of original delivery Products transferred outside the country of original delivery either by KD Scientific at the direction of the Buyer or by Buyer s actions subsequent to delivery may be subject to additional charges prior to warranty repair or replacement of such Products based on the actual location of such Products and KD Scientific s warranty and or service surcharges for such location s The warranty period for data processing equipment including data storage devices processors printers terminals communi