CliChem 240 Auto Chemistry Analyzer User's Manual
Contents
1. Date Detail Add Delete Delete All gt Start Pre Next Search 54 Figure 9 2 Maintenance Service record Add a service record Press Add button enter the service record needed to add in the opened new form and then press Save Delete a service record Select the record needed to delete from the list and then press Delete button If needing to delete all the service records press directly Delete all button Note Only the administrator can add and delete service records 2009 10 26 16 12 27 Help Maintenance Auto Rerun No Auto Send No Admin Sample request Status Results Parameter Reagent Setup Statistics 8 Exit A 2009 10 22 11 59 28 Logging in Admin B 2009 10 22 13 43 05 Logging in Admin 2009 10 22 13 44 11 Logging in Admin E 2009 10 22 13 46 51 Logging in Admin A 2009 10 22 14 41 38 Logging in Admin i 2009 10 22 14 42 29 Logging in Admin 2009 10 22 15 08 43 Logging in Admin 2009 10 22 17 05 00 Logging in Admin 2009 10 23 17 01 08 Logging in Admin 2009 10 26 09 33 20 Logging in Admin 2009 10 26 09 41 03 Logging in Admin 2009 10 26 09 42 33 Logging in Admin I 2009 10 26 10 55 39 Logging in Admin 2009 10 26 11 00 51 Logging in Admin 2009 10 26 11 10 00 Logging in Admin k 2009 10 26 11 10 03 Logging in Admin gt a Clear Log b Start H Pau Pre gt Next Search Figure 9 3 Maintenance Lo
2. User Daily Testing Calibration Se Result Edit Result Check Maintenance Admin NES NES NES Yes NES E Jack Chen Yes Yes Yes Yes Yes Yes No User Access Setup M Daily Testing Parameter Setup F Calibration Setup l Control Setup f Result Edit Result Check f Maintenance New Save Delete Password b Stat HH g Pre 3 Next X Search Figure 5 4 User setup Add a user Press New button and then enter the user name in the User entry box Select to set the authority and then press Save button Modify a user s authority Select a user from the list select again the authority setup and then press Save button Delete a user Select a user from the list and then press Delete button Note Only the administrator user name Admin can add and delete users and modify a user s authority Modify a password Select a user from the list and then press Password button to open a page as illustrated in Figure 5 6 Revise paszword User Password y New Password A Input Again A Figure 5 5 Password modification Enter user password and new password enter the new password again to confirm and then press 25 Revise button 5 5 Unit Setup Function description Set commonly used units of the system so as to select them from the list when setting item parameters Select Unit Setup tab to open a page as illustrated in the following fi
3. gt Start i Pause W Stop Pre Next Search Figure 8 3 Workload statistics Designate a time period and then press Statistics button The system will perform the workload statistics to all the sent by doctors tested by doctors and checked by doctors within the time period 50 Chapter 9 Service In order to ensure that the instrument can function accurately and reliably as well as to prolong the service life of parts and components users need to perform proper routine maintenance according to the requirements This chapter will introduce the instrumental routine maintenance suggestions on locating troubles calibration and replacement of commonly used components etc Note There is potential biological contamination on the component surface of the instrument therefore proper safety protection measures should be taken during operation and reparation Improper maintenance may damage the instrument Operators must perform the maintenance according to the manual If the troubles or problems in your instrument are not listed in this manual please contact BGT BioGenTechnologies GmbH s Customer Service Department The specialized personnel approved by BGT BioGenTechnologies GmbH will give you the maintenance suggestions The parts and components provided by BGT BioGenTechnologies GmbH must be used when performing maintenance If you have any questions please contact BGT BioGenTechnologies GmbH s Customer Ser
4. Auto Rerun No Auto Send No Admin BGT BioGenTechnologies CliChem 240 Auto Chemisty Analyzer All Rights Reserved http www biogentechnologies de b Start 1 l Sto t Pre Hext Search Figure 3 3 Main menu Note After each re powering it takes 30 minutes for light source to be stable If tests are performed before light source stabilization the result accuracy will be affected 3 2 Exit Click Exit on the main menu to perform logout or exit operations See Figure 3 4 Logout and Exit C Fast Exit Figure 3 4 Logout and Exit Select Logout to log out and log in again Select Exit to close the software system after system cleaning Select Fast Exit to directly close the software system without system cleaning Chapter 4 Parameter Setup 4 1 Test Parameter Setup Function description Setting test and control parameters of all test items Click Parameter on the main menu to open the Test Parameter Setup Page See Figure 4 1 Auto Rerun No Auto Send No lAdmin 2009 10 26 11 42 00 Help Sample request Status Results Parameter Reagent Setup Statistics Maintenance a Exit Test Parameter Setup Calibration Setup Quality Control Setup Profile Setup Calculation Setup External Resi 4 ALT FullName Unit U L Decimallxx Ref Range Volume ul Rea Method ss Normal Rerun Reac mode Kinetic Sample 110 M Prediluted Bl
5. Sample request Status Results Parameter Reagent Setup Statistics Maintenance a Exit Sample Disk History Results Workload Search Date 2009 10 26 2009 10 26 Sex Po Age Year Year Test TP m Print Sample No Age Test Remar Sample No Sex Age Test Result TP 0 0910261001 Unknown 10910261002 Unknown OY TP 18 59 Totus 10910261003 Unknown oY TP 4844 High 0 10910261004 Unknown OY TP 78 29 10910261005 Unknown OY TP 72 01 Low 0 Abnormality ratio 0 0 Mean value 43 466 SD 30 200 CV 69 5 gt Stat Pause W Stop Pre 3 Next 3 Search Figure 8 2 Statistics of History Results Designate searching conditions date sex age and Test as required and then press Statistics button The system will display the test records meeting the conditions in the list and perform the result statistics automatically The statistics include total amount higher amount lower amount abnormality rate mean value SD and CV 8 3 Workload Statistics Select Workload page as illustrated in Figure 8 3 49 Lamp Stabilizing 37 1C Auto Rerun No Auto Send No Admin 2009 10 26 11 23 36 Help Sample request Status Results Parameter Reagent Setup Maintenance Exit M Date 2009 10 26 7 2009 10 26 y oK Sent by Tested by Checked by Department Sentby Workl Testedby Workload _ Checked by Workload Admin 5 gt 5
6. Note e 11 sE expresses all items e If the special cleaning is not necessary between two reagents select Distilled water otherwise select Cleaning Solution and the position number of the special cleaning solution should be designated e Performing special cleaning will consume the normal testing cycle and slow the instrumental testing speed Therefore please set special cleaning only if necessary Item mask setup Press Mask Test button to open a page as illustrated in Figure 4 7 Masked Test Select Masked Test ALB ALT CR TP Mask All Cancel Mask Back Figure 4 7 Item mask setup Select the item needed to mask And the system will automatically skip the item Press the Mask All button to mask all items press Cancel Mask button to cancel the item mask Press Save button to save the modifications otherwise press the Back button not to save them 4 2 Calibrator Setup Function description Set the parameters of calibrators used in the system Select Calibration Setup to open a page as illustrated in Figure 4 8 16 Auto Rerun No Auto Send No lAdmin 2009 10 26 11 42 48 Help Sample request Status Results Parameter Reagent Setup Statistics Maintenance o Exit Test Parameter Setup Calibration Setup Quality Control Setup Profile Setup Calculation Setup External Res Calibration Test rest Reference unt TP 30 mmol L New Delete New Delete
7. Pre 3 Next Search Figure 7 9 Reagent blank searching The left list displays all items tested reagent blank Click the item needed to search and the right list will display the history reagent blank record of the item Press Delete button to delete the selected reagent blank record in the list Press Delete all button to delete all reagent blank records in the list Press Print button to print all reagent blank records in the list Note The blank results beyond reagent blank range will be marked with OR 47 Chapter 8 Statistics 8 1 Worksheet Statistics Click Statistics in the main menu to open the worksheet statistics page as illustrated in Figure 8 1 Lamp Stabilizing 37 0C Auto Rerun No AutoSend No Admin 2009 10 26 11 20 51 Help Sample request Status Results Parameter Reagent Setup Statistics Maintenance Gg Exit 53 069 65 7 24 353 33 05 12 714 56 6 28 687 55 85 P Start Fi Pause W Stop tPre Nex Search Figure 8 1 Worksheet statistics Select the worksheet needed statistics from the pull down list and the statistical results of the worksheet will be displayed beneath the list The statistical results include total amount completed amount higher amount lower amount and abnormality rate 8 2 Statistics of History Results Select History Results page as illustrated in Figure 8 2 48 Lamp Stabilizing 37 0C ETA AN Send No Admin 2009 10 26 11 21 43 Help
8. Position Sample No Sample Type Test Result Unit Remark jStatus 4 ALT Start at09 57 10 7 0910261002 Serum 107 892 UL Completed 8 0910261003 Serum ALT 49 042 UL Completed 3 0910261004 Serum ALT 49 246 UL Completed 10 0910261005 Serum ALT 96 653 U L Completed 6 0910261001 Serum CR 33 388 ummol L Completed 7 0910261002 Serum CR 32 571 ummol L Completed 8 0810261003 Serum CR 32 571 ummol L Completed 9 0910261004 Serum CR 6 857 ummol L Completed 10 0910261005 Serum CR 6 94 ummol L Completed 1 tp1 Serum TP 30 mmol L Completed 2 tp2 Serum TP 55 6 mmol L Completed 3 tp3 Serum TP 82 mmol L Completed 4 tp4 Serum TP 92 8 mmol L Completed 5 zk101 Serum TP 36 267 mmol L Completed 6 0910261001 Serum TP 0 mmol L Completed 7 0910261002 Serum TP 18 594 mmol L Completed 8 0910261003 Serum TP M1 3 0910261004 Serum M 111 10 0910261005 Serum TP M1 v in gt gt i Curve Retest Print Total 26 Clear li Pause 8 Stop Pre 3 Next Search Figure 6 13 Test list e Select records to be displayed Select All Completed or In process to display test records respectively according to different requirements e Display reaction curves Select a test from the list and press Reaction curve button to view the corresponding reaction curve of the test e Retest Select the result needed to retest from the list and then press Retest to mark the selected record as Retest If the
9. Save b Start Pre Next Search Figure 5 7 Setup of sample information testing SI test parameters When performing SI tests dilute the sample according to the set dilution parameter and then analyze the diluted sample to obtain the sample information The test parameters include sample volume reagent volume diluent volume and reagent position diluent position Calculation parameters The basic information of serum includes turbidity L hemolysis H and icterus 1 A B C D E and F are calculation parameters Markers Set the correlation of information and markers in the report SI test switch Selecting Turbidity Examination means need to examine the turbidity of patients samples eo Select Hemolysis Examination means need to examine the hemolysis of patients samples Select Icterus Examination means need to examine the icterus of patients samples Note If the system has selected any item of sample information examinations the SI item tests will be automatically performed before testing the patient s sample 5 7 Reagent Setup Function description Designate reagent positions and other reagent information for test items Click Reagent in main menu to open the reagent setup page as illustrated in Figure 5 8 21 Lamp Stabilizing T ERROR Auto Rerun No Auto Send No Admin 2009 10 22 13 59 41 Help Sample request Status Results Parameter l Reagent Setup
10. MR CT Mi AROS 46 O Unknown 5 B Eror lt T Stat H Pause B Stop Pre Next Search Figure 6 11 Testing status Reaction disk e Replacing cuvettes All cuvettes can be replaced by pressing the Replace cuvettes button e View reaction curve Select the reaction cuvette from the list or illustrations of reaction plate and then press Reaction curve to view the reaction curve of this test See the figure 37 Reaction curve Cuvette No 11 Blank 55356 M With mark Sample 10 TestTP Result55 6 mmol L 1 2 3 4 5 6 78 8 1011 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 ame Prim m Sec Prim Sec Last ll Figure 6 12 Reaction curve Press Last and Next to view the reaction curves of reaction cuvettes at each position in order Press Print button to print the reaction curve of the reaction cuvette at current position Press Back button to exit the viewing of reaction curves Test list View all tests completed tests and uncompleted tests of the current worksheet respectively See the figure Testing Auto Rerun No Auto Send No Admin 2009 10 26 10 27 43 Help Sample request Status Results Parameter Reagent Setup Statistics Maintenance a Exit Sample disk No 1 Test mode Single Dual Reagent Total 26 Remaining time 2 min Reagent Sample Disk Reaction Disk Test List All C Completed C n Process Note
11. Next Search Figure 6 3 Quality control request Step 1 Designate Position Step 2 Select repeat times Step 3 Select control no and the QC items contained in the quality control product will be displayed in the right list Step 4 Select QC items from the list Step 5 Press Save button Step 6 A QC request is completed 6 1 4 Calibration Request Select Calibration Request tab to open a page as illustrated in Figure 6 4 31 Lamp Stabilizing 37 1C Auto Rerun No lAuto Send No Admin 2009 10 26 11 30 01 Help Sample request Status Results Parameter Reagent Setup Sample Disk No 1 Statistics Maintenance 8 Exit Active Sample Request Quality Control Request Calibration Request List Test Position 1 Small Tube i Repeat Mo Calibration Calibration Noltp1 11 Delete Delete All Input position and press Delete eg 15 8 means all gt Start d Pause M Stop Pre Next Search Figure 6 4 Calibration request Step 1 Designate position Step 2 Select repeat times Step 3 Select a calibration no and the calibration items contained in the calibrator will be displayed in the right list Step 4 Select calibration items from the list Step 5 Press Save button Step 6 A calibration request is completed List Select List page as illustrated in Figure 6 5 32 Lamp Stabilizing KA Auto Rerun No Auto Send No Admin 2
12. Password OK Cancel Figure 3 1 Login Select a user name and enter the password and then press OK button to perform user login otherwise press Cancel to exit the software Note The user name of system administrator is Admin and the initial password is 888888 Users can freely change the password Please remember once you have changed the password After login the system automatically performs self inspection and initialization See the following figure Initializing Checking reagentisample disk Figure 3 2 Initialization Self checking and initialization contains the following contents e Connecting the front end Test if the communication of control software and intermediary device software is normal e Module connection Test if the statuses of all modules are normal e Parameter delivery Read system parameters from the intermediary device e System resetting Reset the whole system and perfuse the liquid line so as to be ready for testing Note If the system uses disposable cuvettes please replace new cuvettes during each start up After self inspection and initialization the system will wait for system light source stabilization at this time the words Lamp Stabilizing will display on the left corner of the screen After light source stabilization the word Ready will display on the left corner of the screen only at this time can tests be performed See the following figure
13. 3 25 5 consecutive values on the same side of target value 5 5 T consecutive values on the same side of target value 7 T Figure 4 10 QC rule setup Select a QC warning rule for this item from the seven rules and then press Save button to save 4 4 Profile Setup Function description Integrate multiple test items supporting a diagnosis into a set Select Profile Setup page as illustrated in Figure 4 11 Auto Rerun No Auto Send No Admin 2009 10 26 11 43 39 Help Sample request Status Results Parameter Reagent Setup Statistics Maintenance a Exit Test Parameter Setup Calibration Setup Quality Control Setup Profile Setup Calculation Setup External Res Profile Profile2 New Delete Save gt Stat i Pause M Stop Pre 3 Next X Search Figure 4 11 Integration setup Add a profile 19 Press New button enter the name of new profile and then press OK button Select the items contained in the profile and then press Save button Modify a profile Select the profile needed to modify from the profile list modify the items contained in the profile and then press Save button Delete a profile Select the profile needed to delete from the profile list and then press Delete button 4 5 Calculation Setup Function description Set the items which results are obtained by calculating test items Select Calculation Setup page as illustrated
14. Statistics Maintenance a Exit Test R1 Pos R1 Vol R1 Size R2 Pos R2 Vol aj T iL R1 Position a Volume ml Large Size 40ml Small Size 18ml 39 DILUENT 40 ATER 4 gt Save Delete Delete All Check b Start 3 Pre Next Search Temperatue sensor error Figure 5 8 Reagent setup Modify reagent information Select an item set the positions of Reagent 1 and Reagent 2 surplus volume and model of reagent bottle respectively and then press Save button If the fixed bar code scanning module is available please set the corresponding reagent bar code Note The Reagent Positions 39 and 40 are respectively used to place diluent and deionized water other items cannot use them Delete reagent information Select an item and press Delete button to delete the reagent information of this item Press Delete All button to delete the reagent information of all items Reagent checking Press Check button and the system begins to test the reagent surplus volume in the designated position 28 Chapter 6 Routine Testing 6 1 Sample Request Function description Edit worksheets sample disks and set test items for samples Click Sample request in main menu to open the sample request page as illustrated in Figure 6 1 Lamp Stabilizing 37 0C Auto Rerun No Auto Send No Admin 2009 10 26 11 29 09 Help Sample request Status Results Parameter Reagent Set
15. Step 6 Press Save button Step 7 A sample request is completed e Edit multiple sample in batch Step 1 Select repeat times and dilution ratio The default repeat times is 1 and dilution ratio is None Step 2 Input the initial sample no Step 3 Select test items and test profile Step 4 Press Batch Save button to open a page as illustrated in Figure 6 2 Batch 5ave Figure 6 2 In batch saving Step 5 Input the initial and end sample positions and then press the OK button Step 6 A batch sample request is completed Note When using the Batch Save the system will automatically generate sample no the generation rule is date sample disk number sample position such as 0906161001 After request users can modify it manually If small sample tubes are used please select Small Tube 6 1 3 Quality Control Request Select Quality Control Request tab to open a page as illustrated in Figure 6 3 30 Lamp Stabilizing 37 0C Auto Rerun No Auto Send No Admin 2009 10 26 11 29 41 Help Sample request Status Results Parameter Reagent Setup Statistics Maintenance a Exit Sample Disk No 1 Active Sample Request Quality Control Request Calibration Request List Test Position 5 T Small Tube Repeat M a Quality Control Control No zk101 5 Delete Delete All Input position and press Delete eg 15 8 means all b Start Pause M Stop Pre
16. b Start Pre Next Search Figure 4 8 Calibration setup Add a calibrator Press New button under calibrator list enter the calibration no and lot in the opened page and click OK to save Then add all calibration items included in the calibrator one by one Delete a calibrator Select the calibrator needed to delete from the calibrator list and press Delete button under calibrator list Add or modify calibration items Select a calibrator and then press New button under test list Select calibration item needed to modify in the opened page enter reference values and then press the Save button Delete calibration items Select a calibrator from the calibrator list and its contained calibration items will be displayed in the left list Select the calibration item needed to delete and then press Delete button Note If an existent calibrator is deleted the corresponding calibration results of this calibrator will also be deleted at the same time If an existent calibration item is edited or deleted the corresponding calibration results of this calibration item will also be deleted at the same time 4 3 QC Setup Function description Set the parameters of quality controls used in the system Select Quality Control Setup page as illustrated in Figure 4 9 17 Auto Rerun No Auto Send No lAdmin 2009 10 26 11 43 08 Help Sample request Status Results Parameter Reagent Setup Statistics Maintena
17. be displayed beneath the list Edit it as required and then press Save button Edit patient test results Click a patient record and the corresponding test results of the patient will be displayed in the right list Users can add modify or delete the test results Add patient records Press New button and then directly enter a patient s basic information beneath the list After that press Save button Note e The retested manually added or manually modified test results are marked with the symbol e The symbols t and respectively express that the result is too high and too low e The new added patient sample no cannot be as same as the existing ones otherwise the system will update the existing patient information 41 Delete patient records Select the records needed to delete from the patient record list or press Select All and then press Delete button The system will delete the designated patient records and all the corresponding test records Check patient results Select the patient records from the patient record list or press Select All and then press Check button The checked by doctor of the selected patients will be automatically updated as current operator Indirect calculation Select the patient record needed to perform indirect calculation from the patient record list or press Select all and then press Calculate button The system will automatically calculate the calcu
18. cause physical injuries Warnings and Safety Hints This instrument is only used for in vitro diagnosis Please read the following warnings carefully before use The warnings must be strictly complied Warning Please carefully read the following precautions before using this instrument e f abnormal odor or smoke occurs cut off the power supply immediately and pull the power plug out of the power socket At this time present an inspection application to your distributor or an agent of our company If continuing to use the instrument at this time fire electric shock or personnel injuries or death may occur Liquid should not drip into the instrument metal objects such as staples and pins should not fall into the instrument either Otherwise it is likely to cause short circuit or fire and smoke Operators shall not contact the electronic circuit within the instrument in particular contacting the circuit with wet hands has a higher risk of electric shock Rubber gloves must be worn and prescribed tools and components should be used when servicing and inspecting the instrument After work please wash your hands with disinfectant solution Otherwise the skin contacting the liquid may be burned or infected e When treating samples please be very careful and wear rubber gloves otherwise it is possible to cause infections If the sample enters the eyes or wounds rinse well with clean water immediately and visit
19. exceeds the limit content prescribed in the Standard SJ T11363 2006 2 Marker description Environmental Protection Service Life Marker 9 Marker meaning This electronic information product contains some toxic and harmful substances Its environmental protection service life is 20 years Within the environmental protection service life users can use it securely beyond the environmental protection service life it should enter the recycling system 63
20. left list and press Print button 7 4 QC Search Function description Search the test results of designated QC item within the designated time period Select Quality Control page as illustrated in Figure 7 7 Lamp Stabilizing 37 1C Auto Rerun No Auto Send No Admin 2009 10 26 11 19 21 Help Sample request Status Results Parameter Reagent Setup Statistics Maintenance a Exit Patient Test Calibration Quality Control Reagent Blank Quality Control No Test ALB Date 2009 10 01 l 2009 10 26 TP v 2009 10 05 10 26 20 37 00mmol L v 2009 10 26 10 26 20 36 27mmol L v i Select All Graph Delete Print elect All b Start Fi Pai 5 Pre 3 Next X Search Figure 7 7 QC search View QC data Select the QC item needed to view from the left list and designate the searching date range Then all the test results meeting the conditions will be displayed in the right data list View QC graph Select the data points needed to draw from the data list and then press QC Graph to display the QC graph 45 QC Graph Display Type All C Daily Cumulative mmol L zk101 TP 38 Target 36 500 Mean 36 390 SD 0 500 SD 0 582 CV 1 6 Figure 7 8 QC graph This page displays the target value and SD value of the QC item as well as the statistical results of the selected data mean value SD and CV The system can automatically convert the QC graph according to
21. the external item results tested by other devices into this system please first set external item parameters needed to enter Select External Result Setup page as illustrated in Figure 4 13 Auto Rerun No Auto Send No Admin 2009 10 26 11 48 23 Help Sample request Status Results Parameter Reagent Setup Statistics Maintenance o Exit Calibration Setup Quality Control Setup Profile Setup Calculation Setup External Result Setup Test FullName QuatijQuan low High RBC Quan 0 000 0 000 X XXX PLT Quan 0 000 0 000 X XXX HbeAsg Quali X v gt Test Full Name Quantitive Qualitative Reference 0 0 Unit Decimal x xxx b Stat i Pre Next Search Figure 4 13 Setting non biochemical items Add external items Press New button enter one by one the item name full name qualitative or quantitative values and reference values of the new non biochemical item and then press Save button Modify external items Select the external item needed to modify from the list edit its full name qualitative or quantitative values and reference values and then press Save button Delete external items Select the external item needed to delete from the list and then press Delete button 21 Chapter 5 System Setup 5 1 System Control Parameters Function description Including active sample disk automatic rerun switch automat
22. to select Reagent 2 and then enter Reagent 2 and select the mixing intensity after adding sample Note If the dilution ratio of patient sample is designated all the tests of this patient sample should be pre diluted according to the designated dilution ratio Otherwise tests should be performed according to the designated pre dilution ratio in item setup parameters Note Only the test result is beyond the instrumental linear range and performed an automatic or manual rerun the test is performed according to the set Rerun volume Otherwise tests are performed according to the set Normal volume 6 Wavelength Set the wavelength required according to the reagent instructions For single wavelength tests set only Wavelength 1 and select None for Wavelength 2 However in order to eliminate external interference it is recommended to use double wavelength tests 7 Reac mode Select End Point Fixed time or Kinetic 8 E type Select None R Blank S Blank Pre R Blank or Pre S Blank Note None Meaning that it is not needed to subtract the blank value R Blank Meaning Reagent blank he determination method is to test by converting the sample into distilled water according to the reagent and sample volumes during normal testing S Blank Meaning Sample blank The determination method is to test by converting the reagent into distilled water or saline according to the reagent and sample volumes durin
23. 009 10 26 11 30 21 Help Sample request Status Results Parameter Reagent Setup Statistics Maintenance a Exit Sample Disk No 1 Active Sample Request Quality Control Request Calibration Request List Positior Smal Tube Sample No Test fome TestType Repeat Diluti 1 No Itp1 ITP 9 7 tp2 Cali tp3 Cali tp4 Call zk101 QC 0910261001 ALB ALT CR TP Routine 0910261002 ALB ALT CR TP Routine 0910261003 ALB ALT CR TP Routine A 0910261004 ALB ALT CR TP Routine i 0910261005 ALB ALT CR TP Routine 1 Delete Delete All Input position and press Delete eg 15 8 means all b Start Pre Next Search Figure 6 5 List The list displays all the applied sample tests QC tests and calibration tests of the worksheet It can be used check and confirm applied test contents in the worksheet Note After users complete the entire sample request please select List to review and confirm so as to avoid errors or omissions Delete samples Select the sample from the list or enter the sample position needed to delete in the entry box at the left bottom of the page entry format such as 1 3 5 8 12 and then press Delete button Press Delete all button to delete all samples in the worksheet 6 2 Sample Testing Function description Start to perform the applied tests in the current worksheet Press Start button at the bottom of the screen and then a pop
24. 1 Product Name Auto Chemistry Analyzer isses essen enean nana nnns 3 VALL Model GlIChem a E Idus 3 fais CBROQUOLI YITICIDIB a NN a 3 1 14 Product Application Scope ti A an 3 1 2 Product Structure and COMPDOSINON sitas aane A a a aA 3 Em Sucre ODA idas 3 1 2 2 Input and Output Interfaces on Analyzer Main Part eese 4 LR COMPUTO interfaces sait esposas de cde esM 4 1 3 Instrumental Technical Parameters cooccoccoccconconcococonconoonnconcnnnononnnconnnnnnnnnnnnnncnnnonennnnaneneos 5 Chapter 2 Installation and Calibration 6 2 1 Instrument UNS Cad O gs eb o 6 2 2 ANstument Installatlon caida 6 2 3 Turning On Off the InstruMent ccoocccccnccccccccccocncccnoconcnncnononononoconnonnnonnncnnnnnnnnonaninnnnoncnnnnninons 6 2 4 Instrument Calibration Requirements cooccccccocncocnconcocnoncncococonocononononcnonnnnononcnonronononaninonnnnos 7 Chapter 3 Login and Shutdown ccsececeeeeeeneecenseeenneeeeseeeneeeenseeenees 8 3 1 LOO Renee near art a RN ML E T PE 8 32 a RR Me eR Or nae ra Oe ee ee ee 9 Chapter 4 Parameter Setup cccccssecessecceseeessecenseecesesenneeceseeeeneeeaes 10 a Jest arameler SM eli Io paribus Da p UN NI UM eias Cice Medusa i acd nae DO e 10 42 Calibrator oetUD sindical 16 q 39 QC Sal aii 17 AeA SSA A O A O RO eencenied a beaaboviasie 19 45 O ONY SOU esse tee estate ap ae iudi
25. 51 9 1 2 PE HOGIC MANTECA m 51 9 2 Common Troubles and Messul6es o tarada das eo eed 52 9 3 AYES TU Wee IRIS ERU an nO Cae aE AE A 52 Chapter 10 Using Precautions Leere rere nene 59 Appendix A Calculation Methods eese 60 A 1 Ho aos ee nsec E pete tale SERERE A UR 60 PEZ AKO C TIS VIC TO Gis RE dico 60 7 6 AMO IMIS TAO GC ONERE ld e a 61 Appendix B Name and Content of Toxic and Harmful Substances or Elements In PFOGUCL riii 25i exon a FRU xVFE VER EREME T a EENEPrVxkgE CEEVEK REDI aUE 63 Chapter 1 Instrument Introduction 1 1 Product Introduction 1 1 1 Product Name Auto Chemistry Analyzer 1 1 2 Model CliChem 240 1 1 3 Product Principle CliChem 240 is an analyzer for clinical biochemical tests which is researched and developed by photoelectricity automation and computer technologies The analyzer tests samples by the transmission of specific turbidity testing method and calculates sample concentration according to the Lambert Beer Law 1 1 4 Product Application Scope It is used for routine clinical biochemical assays and other absorbance determination 1 2 Product Structure and Composition It mainly includes an optical part mechanical motion part and computer control part 1 2 1 Structure Graph ca vl E AAA Figure 1 1 CliChem 240 vertical view 1 Sampling needle 2 Reagent sample plate 3 Mixing arm o ICO sa
26. 7096 without dew condensation Its power supply must be a c 220V 50Hz 60Hz and well grounded If the voltage variation gt 10 it is recommended to install an external voltage stabilizer with the power of above 1500W This analyzer is a precision instrument fully and automatically controlled by computer therefore an above 3000W UPS uninterruptible power supply must be configured Note e The AC power supply must be well grounded zero ground voltage 5V e The AC power supply must be stable Do not share the power supply with high powered electrical appliances e When pulling out the power cable you must grab the plug itself not the power cable e If there is smoke peculiar smell or strange sound from the instrument please turn off the power supply immediately and contact your distributor Connecting control computer e Connect mouse and keyboard to the corresponding interfaces at the back of the control computer e insert one end of display cable into the signal interface of display and the other end into the displaying interface at the back of control computer Then connect the display to AC power supply with power cable e Insert one end of print cable parallel interface into the signal interface of printer and lock up with wire thread and then insert the other end of cable into the parallel printer port at the back of control computer Then connect the printer to AC power supply with the power cable accompan
27. Admin 2009 10 26 11 27 26 Help Sample request Status Results Parameter Reagent Setup Statistics Maintenance a Exit System Setup Report Setup Department Setup User Setup Unit Setup SI Setup Department Doctor Jack Dept2 Smith ee 6 New Save Delete New Save Delete b Stat i pa Pre 3 Next Search Figure 5 3 Department setup Add a department Press New button below the department list enter the department name and then press Save button Delete a department Select the department needed to delete from the department list and press Delete button Add a doctor Select a department and all doctors in this department will be displayed in the doctor list Below the doctor list press New button enter the doctor name and press Save button Delete a doctor Select a department and all doctors in this department will be displayed in the doctor list Select the doctor needed to delete from the doctor list and press the Delete button 5 4 User Setup Function description Set a user s authority and password Select User Setup tab to open a page as illustrated in Figure 5 4 24 Lamp Stabilizing 37 0C Auto Rerun No Auto Send No Admin 2009 10 26 11 27 46 Help Sample request Status Results Parameter Reagent Setup Statistics Maintenance a Exit System Setup Report Setup Department Setup User Setup Unit Setup SI Setup
28. CliChem 240 Auto Chemistry Analyzer User s Manual BGT BioGenTechnologies LAA mm o s n9 o p BGT BioGenTechnologies BGT BioGenTechnologies GmbH Von Langen Weg 10 D 48565 Steinfurt Tel 02551 4090 Fax 02551 1298 Web www biogentechnologies de E Mail info biogentechnologies de Version 1 0e How to Use This Manual Thank you very much for becoming a user of CliChem 240 Auto Chemistry Analyzer In order to get the optimal effects you must become familiar with our instrument and its performance before using it to perform clinically diagnostic tests This User Manual discusses operating instructions of BGT BioGenTechnologies GmbH CliChem 240 Auto Chemistry Analyzer Its contents include instrument installation routine operations maintenance and service and etc If you have any questions please contact your distributor Please keep all the packaging for storage shipping and to send back reparation in the future Instruments with different versions or configurations have slightly different functions Due to version upgrading the manual contents may slightly change without notice Note Hints counsels and suggestions Write in italics to show a distinction Warning Warnings must be strictly complied so as to ensure the normal functioning of instrument as well as the validity and truth of the test results Write in bold
29. Rerun No Auto Send No Admin 2009 10 22 14 07 54 Help Sample request Status Results Parameter Reagent Setup Statistics Maintenance 9 Exit Patient Test Calibration Quality Control ReagentBlank 0909101001 Unknown 0Y 0909101002 Unknown 0Y 0909101003 Unknown 0Y 0909101004 Unknown 0Y 0909101005 Unknown 0Y 0909101006 Unknown 0Y 0909101007 Unknown 0Y 0909101008 Unknown OY 0909101009 Unknown 0Y 0909101010 Unknown 0Y 0909101011 Unknown 0Y 0909101012 Unknown 0Y 0909101013 Unknown 0Y 0909101014 Unknown 0Y lt Sample No 1090900 Profile All M Clinic No G123456 Sex Male Department ABCDefgh New Test Name Jack Age 24 Yea Sent by Leo Chen Save Result Find Delete Print Send Calculate Check Select All Save Delete b Stat Fi Pa E Pre Next Search Temperatue sensor error Figure 7 1 Patient record Search patient records Press Find button to open a page as illustrated in Figure 7 2 40 Find Patient sample No Clinic No Name Sex Age Department send by Tested by Checked by v Date From 2009 10 22 To 2009 10 22 OK Back Figure 7 2 Searching conditions of patient records Enter the searching conditions and then press OK button The list will display all the patient records meeting the conditions Edit patient information Select a patient from the list and the corresponding basic information will
30. Select All and then press Print button View reaction curve Click the test records needed to review and press Reaction curve button The system will display the reaction curve of the test as illustrated in Figure 7 5 Reaction curve 12 3 4 5 6 amp 6 7 8 9 1011 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 Prim SEC Prim Sec Print Back Figure 7 5 Reaction curve Press Print button to print the curve 7 3 Calibration Search Function description Search the calibration results of designated items Select Calibration page as illustrated in Figure 7 6 Lamp Stabilizing 37 1C Auto Rerun No AutoSend No Admin 2009 10 26 11 18 48 Help Sample request Status Results Parameter Reagent Setup Statistics Maintenance a Exit Patient Test Calibration Quality Control Reagent Blank ALB Calculation Log logit E Calibration No Reference mmol L OD TP calibration curve Log logit 123 136 1681 2282 y 0 1392 2784 4176 5568 696 8352 9744 111 36 s x cong mmol L Ln R K Ln C B K 0 468 B 7 763 Delete Print b Stat dH Pause 3 Stop Pre 3 Next 4 Search Figure 7 6 Standard Search 44 Delete calibration results Select the calibration results needed to delete from the calibration data list and press Delete button Print calibration results Select the item needed to print its calibration results from the
31. Working power supply Power Fuse End point method Fixed time method Kinetic method Factor single point standard multiple point standard linear regression non linear regression etc 8 pieces of optical filters 340 405 450 510 546 578 630 670 nm 40 positions 40 positions 12 V 20 W long life halogen lamp The absorbance should not be less than 4 04 LCD display Peripheral printer 10 C 30 C relative humiditys70 without dew condensation atmospheric pressure 86kPa 106kPa 0 C 40 C relative humiditys80 stored in a well ventilated room without corrosive gases a c 220V 50Hz 60Hz 1500VA T5 0AL250V 05x20 Chapter 2 Installation and Calibration 2 1 Instrument Unsealing Open the instrument packaging remove shipping material and keep them appropriately stored so as to repack the instrument in the future Take out the instrument from the box Check if articles in the packing box correspond to those on the packing list Note If any component is damaged or inconsistent with the packing list please contact your distributor 2 2 Instrument Installation The CliChem 240 Auto Chemistry Analyzer must be installed by specialized personnel The instrument must be installed in a clean and dust free room oscillation moisture strong electromagnetic field and direct sunshine must be avoided The requirements for the working environment are as follows temperature 10 C 30 C relative humiditys
32. a doctor for examination e Used instrument consumables and other waste should be considered as medical waste or infectious waste and should be properly disposed of according to the related national regulations Power voltage connection and grounding Do not insert the power plug into a non 220V AC power socket Otherwise it is possible to cause fire or electric shock e When installing the instrument the three core power cable accompanying the instrument must be used and the grounding must be good Otherwise it is possible to cause fire or electric shock Do not damage the isolation protection cover of power cables Do not strongly draw the power cable or hang heavy objects on it Otherwise it is possible to cause short circuit or broken circuit thereby inducing electric shock or fire Before connecting peripheral devices power supply must be first cut off Otherwise it is possible to cause electric shock or failure The Drug Administration Law prescribes that it is prohibited to modify medical instruments Contents How to Use This Manual 11 eese ieeeleeree nnne nenne nnn anna n nnt I Warnings and Safety Hints 1 ecce Lees eL eese inrer nennen nnns lil COMENS gs scarce cee cece daemeveanpeceee seas E D i E 1 Chapter 1 Instrument Introduction 3 1 1 PLOCUCE Imiirerelden o ER D o E Lm 3 1 1
33. ank type No v l E Test time 90 z 180 a Sec Calculation Factor 2572 Limited Range 1 111 Reag 1 180 Mixing 1 High mixir Linearity range H o o VIL T Reag 2 0 Mixing2 High mix Reagentblanko 0 Wavelength nm Linearity limit 20 Substrate exhaust Oo Wavelength 1 340 Wavelength 2 No K H B o o New Save Delete Print Test Sequence Special Cleaning MaskTest b Start H Paus i Stoj Pre Next Search Figure 4 1 Test Parameter Setup Add an item Press New button and enter name of the new item then enter all parameters of the new item one by one after that press Save button Modify an item Select the item needed to modify from the list and then its corresponding parameters will display on the right Move the cursor to the parameters needed to modify one by one and modify them then press Save button Delete an item Select the item needed to delete from the list and then press Delete button Print item parameters Select the item needed to print from the list and then press Print button 10 Item parameter description 1 Full name Enter the full name or descriptive information of the item It only appears in patient reports 2 Unit Enter or select the unit of item test result 3 Decimal Select the decimal point of item result for printing a summary report 4 Ref range Designate the division standard of reference range and the corresponding valu
34. c Gain Dark AD 430 Read Save Read Save Front Version 101010101010 b Start Pre Next Search Figure 9 6 Maintenance Configuration Limit parameters Blank range is the standard for judging if the reaction cuvette can be used If the blank of the reaction cuvette is within the range it is judged as valid otherwise it is judged as invalid The system will automatically skip the invalid reaction cuvettes Press Read parameters button to obtain the current blank limit parameters of the system Enter the blank range and then press Update Parameters button to save 57 Note The blank range must be obtained by actual testing If the range is not rationally set the system cannot test normally When replacing different batches of reaction cuvettes please open Blank page first and view the cuvette blank range of this batch then set a proper limit parameter according to the test results of all wavelengths Filter parameter Wavelength displays the corresponding wavelengths of the eight filters configured by the system one by one The Column Gain displays in turn the corresponding gain coefficients of the eight wavelengths If needing to adjust the gain coefficient corresponding to a certain wavelength users can directly modify it by hand or press Automatically Configure Gain to automatically obtain a proper gain coefficient and then press Update Parameters to save the obtained parameter If th
35. current worksheet can be tested If needing to modify active sample disk refer to 5 1 System Control Parameters 36 Total Indicating the total test amount of current worksheet Remaining time Indicating the remainder time to finish the tests of current worksheet menge mp Disk View the testing status of designated sample Click Sample List page and select Position from the list and all the corresponding tests will be displayed in the list Select a test from the list and press Curve button to view the corresponding reaction curve of the test or press Retest for retesting request e View reagent status Click Reagent List page and the reagent information designed in the worksheet will be displayed in list Reaction Disk Click the Reaction Disk tab in the main menu to open a page as illustrated in the figure 2009 10 26 10 24 45 Help Testing Auto Rerun No Auto Send No Admin Sample request Status Results Parameter Reagent Setup Statistics Maintenance o Exit Sample disk No 1 Test mode Single Dual Reagent Total 26 Remaining time 5 min Reagent Sample Disk Reaction Disk Test List O Predilution Reaction disk info B Testing B Finished Completed 107 892 Completed 49 042 Completed 49 245 Completed 96 553 Completed 33 388 Completed 32 571 Completed 32 571 Completed 6 857 Completed 6 940 Completed 30 000 Completed 55 600 M1 M1 M1 M1 M1 M1
36. cuvettes in the reaction plate one by one users can press Stop button to stop cleaning at any time during the cleaning process Applicable to the system equipped with auto cleaning modules Replace cuvettes Press Replace cuvettes and then the system turns the nine set cuvettes to the manual replacing window one by one and remind users to replace Press stop button to stop replacing the rest cuvettes Test cuvette blank Press Check Blank button and then the system begins to automatically test the cuvette blank of the 81 cuvettes at each wavelength after completion the system will automatically display the corresponding cuvette blank value of designated wavelength Blank screening Set the screening range of the corresponding cuvette blank of designated wavelength and then press Refresh button The system will automatically judge the validity of each set of cuvette blank according to the set range and mark with colors Print Press Print button and the system will automatically print the cuvette blank value displayed in the list Auto Rerun No Auto Send No Admin 2009 10 26 16 13 28 Help Sample request Status Results Parameter Reagent Setup Statistics Maintenance o Exit Backup Service Log Debug Blank Configure Filter Setup Wavelength Gain Configure Blank Range 50000 60000 4 mo Automatic Gain 3 200 Automatic Gain 5 200 Automatic Gain 7 0 foo Automatic Gain 8 200 Automati
37. e cuvette blank energy of a certain wavelength is too low increase the corresponding gain coefficient conversely decrease the gain coefficient Note Common users shall not modify the parameters in Configuration page otherwise the accuracy of system testing may be affected 58 Chapter 10 Using Precautions Precautions Cause The instrument must be kept far away from all oscillation 1 sources and electromagnetic interference so as not to affect Avoid affecting signal acquisition the signal acquisition 2 Please check water supply and drainage tubes before use Do Avoid affecting sample adding not bend and fold them and avoid causing liquid to spill Do not let any object enter the working area when the l e S nep P Avoid collision instrument is functioning so as to avoid collision Please do not replace cleaning solution and distilled water 4 during the testing process otherwise the liquid seal cannot Avoid affecting sample adding form in the liquid lines thereby affecting the injector sampling If the system always reminds of invalid blank please confirm if 5 the cuvettes are contaminated or worn if the gain coefficient is Affect the test accuracy proper and if the light source is aging Please do not start testing before the light source and Affect the test accuracy temperature are stable 7 The opening of reaction plate must be always closed during the Affect the test accuracy testing process Affect the te
38. e factor values can be manually input Note In speed method the factor symbols of test items reflect the curve variation direction For a downward reaction the factor is a negative number otherwise it is a positive number 11 Linearity range It means the testable range of the instrument or reagents If a test result is beyond the range it is unreliable in that case a diluted retest or concentrated retest should be performed 12 Reagent blank It means the valid range of reagent blank If the reagent blank is beyond this range the system will treat the reagent as invalid The unit is 1 10 000 absorbance Linearity limit Only applicable to tests using speed method The system automatically calculates the linearity during test period if the linearity of reaction curve is beyond the set range the result will be marked accordingly The permissible setup range of linearity is 0 300 and the system default value is 20 The calculation formula of linearity limit is as follows 13 ma Test points gt Nine points Linearity Variation rate of the first six points Variation rate of the last six points Variation rate of all points Four points lt Test points lt Eight points Linearity Variation rate of the first three points Variation rate of the last three points Variation rate of all points Substrate exhaustion limit Only applicable to tests using two point method and speed method Some high concentration ac
39. ed on e Check if the control computer and analyzer are normally connected by the control series interface cable e Check if the control computer uses the port designated by system software The cuvette blank is invalid e Check if replacing clean cuvettes e Check if the light source energy is normal e Check if the cuvette blank is invalid e Check if the corresponding gain coefficient of each wavelength is rational The sampling volume is inaccurate e Check if the sampling lines are well sealed Check if the liquid lines are well perfused Check if the corrective coefficient of liquid volume is normal Check if the injector status is normal Check if the deionized water is exhausted The sampling needle cannot accurately localize e Check if the sampling needle is vertically downward e Check if standard reagent bottles and sample tubes are used There is residual liquid on the surface of the sampling needle Check if there is dirt on the outer wall of sampling needle Check if the tube of sampling needle is worn Check if the fountain height in the cleaning pool is proper when cleaning the sampling needle Check if the sampling lines are well sealed Check if the liquid lines are well perfused Check if the deionized water is exhausted The mixing arm cannot accurately localize e Check if the sampling needle is vertically downward e Check if standard reagent bottles and sample tubes are used 9 3 Mainte
40. ere Era chu d rd rica ue Rr wget o core EORR Vince oiacsot a UE Pere draft 20 40 External RESUS CGD ici ter eth a Md osa 21 5 1 System Control Parameters esris chta e iaa E E E E EAE aaa E REE 22 S MEME O TER E A A a OT on ee ene 23 959 Department Setup uiro rec d oer De ke d ecc s d ei v Pa a urbt t D c 23 SC US SO EET 24 IG SOI cM 26 S WEE Ec nn ao o a rege enn er en a eee 26 Oar dJseadent oe lc iain OUR oa feet mad cu Mese await ad gees area cheats 2f Chapter 6 Routine Testing eee eee eee rere nennen nnn 29 OF Sample REGUE tua A o 29 Sn 29 BA SAM EQS ST hadi ett ial ease Um 29 COEM MEMN QU CONTO i o V cR 30 6 14 GCalibrallon ROQUES eris adde oe eut A ade eae uar dua 31 627 Sample Teslng sso e kie dis 33 0 3 TESI dl cee ee E O ee ee RSE S ERI OA 36 Chapter 7 PESUIE SC aN CI iacts gti se vacteacsivessumecenaueenaductwsaantecsanasaianestnaceaiacut 40 71 Searching Patient Records iris ertet aa A ii 40 EZ gt Searcnind Test Rie io RE t EUER 42 E S eiie ne y A TDEx 44 VES ME BS o A A RM RN 45 75 Searching Reagent Blessed 46 Chapter 8 Stats RETE TETTE TTD T D D 48 8 1 Worksheet SEtatistleS 2 do tos Sere dee Ed da 48 0 2 SlaliStiGs OF AISO RESUS ite too O niue Lad 48 S3 WVONRKIOAG SAUS el oe DER EE E dI 49 Chapter 9 Sn 51 9 1 Mitzila sin zi Ie UI o e de seni 51 9 1 1 Inspeccion Detore ANALYSIS uui tastes testes psa tes bec ene teehee ad
41. es according to the reagent description See Figure 4 2 iv By Age Male Fed B 116 Year 0 lt 35 Year 0 a lt 60 Year 0 Above 0 Figure 4 2 Reference range 5 Volume e Setting the sample volume for normal test Enter sample volume the applicable setup range is 3 45 ul and the precision is O 1pl If this item is needed to pre dilute samples before test check Prediluted option and enter the sample volume pre diluted and diluent volume at one time in the box below the option See the following figure Normal Rerun Sample 10 M Prediluted Figure 4 3 Normal test sample volume e Setting the sample volume for rerun test If the test result is beyond the linearity range a rerun test can be performed The designated sample volume should be re pipetted according to higher or lower test results when rerunning test Enter the sample volumes of higher retest and lower retest one by one If it is necessary to pre dilute sample before rerunning check Prediluted option and enter the sample volume pre diluted and diluent volume in the box below the option See the following figure 11 Normal Rerun Sample Higher 3 v fPredil ted sample 30 Diluent 279 Sample Lower 5 Figure 4 4 Second test sample volume Reagent volume The precision of reagent volume is 1 ul Enter the volume of Reagent 1 and select the mixing intensity after adding sample For double reagent testing it is also
42. g The list displays all the system log records If needing to delete all log records press Clear Log button Note Only the administrator can remove logs 55 Auto Rerun No Auto Send No Admin 2009 10 26 16 12 37 Help Sample request Status Results Parameter Reagent Setup Statistics Maintenance a Exit RATA Temperature 1 37 1C Home System Initializing Temperature 2 37 1C Home Reaction Disk Perfusion Temperature 3 37 1C Home Reagent Disk Wash Cuvettes Dark AD 0 Home Sampling System Wash Wipper Home Mixing System Wash Sampling System Home Auto washing Wash Mixing System gt Start i Pause M Stop Pre Next Search Figure 9 4 Maintenance Debug In Status testing page the status testing of all components can be performed Auto Rerun No Auto Send No Admin 2009 10 26 16 13 15 Help Sample request Status Results Parameter Reagent Setup Statistics Exit Backup Service Log Debug Blank Configure Blank Rule Range 50000 60000 O oK O Abnormal O Unknown Refresh Result Wavelength 1 340 Wavelength 2 No Wash Cuvettes Replace Cuvettes Check Blank Print gt Stat H Pause M Stop Pre Next Search Figure 9 5 Maintenance Cuvette blank 56 Wash cuvettes Option function standard instrument does have this function Press Wash cuvettes and then the system begins to clean the 81
43. g normal testing Pre R Blank Meaning that it is needed to subtract the absorbance of R1 Pre S Blank Meaning that it is needed to subtract the mixture s absorbance of R1 and sample e Using end point method for reagent blank test means subtract the absorbance of the reagent blank test Using fixed time method and kinetic method for reagent blank test means subtract the variation per minute of the reagent blank lest 12 e Only end point method test can select Pre R Blank and only end point method for double reagent test can select Pre S Blank 9 Testtime Set the measure beginning time and end time respectively The measure time range for double reagent items is 20 300 sec and that for single reagent items is 20 600 sec 13 Note Generally for tests using the end point method only one measure point is needed to designate if multiple measure points are designated the mean value of the multiple points is used as the calculation value The measure time of fixed time method cannot be less than 30 sec In order to ensure the accuracy of test results obtained from speed method it is recommended that the measure points should be less than six points i e the measure time is longer than 90 sec Moreover for items using fixed time method and kinetic method set the measure time as long as possible 10 Calculation Select the corresponding calculation method according to actual requirements If the Factor is selected th
44. gure Lamp Stabilizing 37 1C Auto Rerun No Auto Send No Admin 2009 10 26 11 28 12 Help Sample request Status Results Parameter Reagent Setup Statistics Maintenance 9 Exit System Setup Report Setup Department Setup User Setup Unit Setup SI Setup Unit E New Save Delete gt Start 1 Pre 3Next Search Figure 5 6 Unit setup Unit list Displaying all units defined by users Add a unit Press New button enter a unit in the entry box and then press Save button Delete a unit Select the unit needed to delete from the list and press Delete button 5 6 SI Setup Function description If testing turbidity hemolysis or icterus of patients samples please first set the test and calculating parameters of sample information before test Select SI Setup tab to open a page as illustrated in Figure 5 2 26 Lamp Stabilizing 37 2C Auto Rerun No AutoSend No Admin 2009 10 26 11 28 29 Help Sample request Status Results Parameter Reagent Setup Statistics Maintenance a Exit System Setup Report Setup Department Setup User Setup Unit Setup SI Setup SI Parameter Sample Volume ul 45 Reagent Volume ul 180 Reagent Position 40 Turbidity Check I Hemolysis Check M Icterus Check Range Note Range Note Range Note Ro p lt p po lt F e B M RF e F a c m 7 e e o
45. ic send switch system temperature and testing mode Click Setup in the main menu to open the system setup page as illustrated in Figure 5 1 Lamp Stabilizing 37 1C Auto Rerun No Auto Send No Admin 2009 10 26 11 25 59 Help Sample request Status Results Parameter Reagent Setup Statistics Maintenance a Exit System Setup Report Setup Department Setup User Setup Unit Setup SI Setup Active Sample Disk o 1 l Linearity range exceeded auto rerun Auto send results to server Temperature Ambient 30C 37C Test Mode Single Dual Reagent C Single Reagent Save Port Setup gt Start Pre Next Search Figure 5 1 Setting system control parameters Active sample disk In total the system has five worksheets and can be virtually divided into five sample disks The system always tests active disk Users can designate the disk needed to test by modifying active sample disk Setting control switch e Selecting Option 1 means if the test result is beyond the linear range the system will automatically retest the result after completing the normal test e Selecting Option 2 means that the test result will send to the system designated server in a real time way The coordinated receiving software should be installed in the server System temperature Set the system target temperature Testing mode e Single Dual Reagent mode Performing single reagent and double reagent co
46. in Figure 4 12 Auto Rerun No Auto Send No Admin 2009 10 26 11 47 36 Help Sample request Status Results Parameter Reagent Setup Statistics Maintenance a Exit Test Parameter Setup Calibration Setup Quality Control Setup Profile Setup Calculation Setup External Res 4 Test Ful Name Unit Decimal Formula JALB GLB mmol L x xx TP ALB Ib AST ALT x x AST ALT oT CR P lt gt Test Full Name Unit Decimal X x Formula KASTHALT Delete Ref Range Input expression including and Method gt Stat Pre Next Search Figure 4 12 Calculation setup Add calculation items Press New button enter one by one the item name full name unit decimal point and calculation formula of the new calculation item and then press Save button After that enter the reference range of the new calculation item Modify calculation items Select a calculation item from the list Modify its full name unit decimal point and calculation formula Then press Save button Or modify the reference range of this item according to the Reference Range Delete calculation items Select a calculation item from the list and press Delete button 20 Note Calculation formulae can only be composed of number and item name Otherwise the calculation formula is invalid 4 6 External Result Setup Function description Before entering
47. it Figure 6 9 Reminding of lower cuvette blank energy At this time users should locate the trouble according to the reminder for example whether the gain setup is normal and clean cuvettes have been placed If the trouble is caused by aging of light source please contact product service personnel to replace the light source Note If the system is equipped with the auto cleaning module cuvettes should be cleaned after startup and replacing cuvettes Otherwise the accuracy of test results Will be affected It takes about 20 minutes to clean all cuvettes 6 3 Testing Status Function description View the test results and status of the current worksheet Click Status in main menu to open a page as illustrated in the following figure Testing Auto Rerun No Auto Send No Admin 2009 10 26 09 57 20 Help Sample request Status Results Parameter Reagent Setup Statistics Maintenance a Exit Sample disk No 1 Test mode Single Dual Reagent Total 26 Remaining time 31 min Reagent Sample Disk Reaction Disk TestList Routine sample Reagent List Sample List e Stat sample Need ml Control A MTV U S eo Standard Empty O Untested O Testing O Finished O Abnormal Hj Pause m Stop Pre 3 Next Search Figure 6 10 Testing status reagent sample disk Sample Disk Current worksheet It indicates the worksheet that the system is testing The system has a total of five worksheets but only the
48. lation item results of the selected patients according to the system setup parameters Send patient test results Select the patient record needed to deliver from the patient record list or press Select all and then press Send button The system will send test results of the selected patient to the designated receiving software Print patient report Select the patient records needed to print a report from the patient record list or press Select all and then press Print button The system will automatically print the patient report 7 2 Searching Test Records Function description Search the test records meeting designated conditions Select the Test page as illustrated in Figure 7 3 42 Lamp Stabilizing T ERROR Auto Rerun No Auto Send No Admin 2009 10 22 14 08 41 Help Sample request Status Results Parameter Reagent Setup Statistics Maintenance a Exit Patient Test Calibration Quality Control Reagent Blank A 2009 09 09 090909001 ALB Quan di mmol L 1 2009 09 09 090909001 ALT Quan 23 U L E 2009 09 09 090909001 CR Quan 126 ummol L E 2009 09 09 090909001 GLB Quan 33 05 mmol L 2009 09 09 090909001 TP Quan 65 25 mmol L 2009 09 10 0909101001 ALB Quan 0 mmol L x 1 2009 09 10 0909101002 ALB Quan 0 mmol L f 2009 09 10 0909101003 ALB Quan 0 mmol L f 2009 09 10 0909101004 ALB Quan 0 mmol L t 2009 09 10 0909101005 ALB Quan 0
49. mmol L x f 2009 09 10 0909101006 ALB Quan 0 mmol L 1 2009 09 10 0909101007 ALB Quan 0 mmol L 1 2009 09 10 0909101008 ALB Quan 0 mmol L 1 2009 09 10 0909101009 ALB Quan 0 mmol L 1 2009 09 10 0909101010 ALB Quan 0 mmol L 1 2009 09 10 0909101011 ALB Quan 0 mmol L 1 2009 09 10 0909101012 ALB Quan 0 mmol L f 2009 09 10 0909101013 ALB Quan 0 mmol L f 2009 09 10 0909101014 ALB Quan 0 mmol L 1 2009 09 10 0909101015 ALB Quan 0 mmol L f 2009 09 10 0909101016 ALB Quan 0 mmol L x 1 lt E Delete Print Curve Modify f Select All b Start 8 Pause Search test records Press Find button to open a page as illustrated in Figure 7 4 Search Pre 3 Next Temperatue sensor error Figure 7 3 Test records Find history result CUE Test Sample No v Date From 2009 10 22 2009 10 22 To OK Figure 7 4 Searching conditions of test records Enter the searching conditions and then press OK button The list will display all the test records meeting the conditions Note The retested manually added or manually modified test results will be marked with the symbol behind them Delete test records Select the records needed to delete from the test record list or press Select All and then press Delete button 43 Print test records Select the record needed to print from the test record list or press
50. mping 5 Sampling position 6 Needle cleaning position position 7 Opening of reaction plate Figure 1 2 CliChem 240 front view 1 Sampling needle 2 Injector 3 Mixing arm 1 2 2 Input and Output Interfaces on Analyzer Main Part RS 232 The series interface for analyzer main part and computer communication AC INPUT Deionized water connection Water inlet of the analysis part the water supply tube entering the instrument from here Deionized water sensor There is a liquid level sensor in deionized water leading out cables into the instrument Waste connection Waster outlet the waste tube leading out from here to the waste barrel Waste sensor Waste liquid level sensor leading out cables into the instrument Main part reset button Use for reset operations if the main part abnormally goes down Grounding terminal Protective grounding o pes qo WO Ol 1 2 3 Computer interfaces RS 232 The series interface for computer and analyzer main part communications Network interface The interface connecting inner and outer networks Parallel interface The interface connecting the computer and peripheral printer USB The interface connecting computer and handhold bar code scanner o a 1 3 Instrumental Technical Parameters Test method Calculation method Spectral mode Reagent position Sample position Light source Stray light Display Printing Working conditions Storage conditions
51. n or By Sample button or drag the tests in the list with mouse After readjusting press OK button to start tests otherwise press Cancel to abandon tests Note The Column Status in the list indicates if a test is completed If the test is completed it displays Completed otherwise it displays none The Column Remark indicates if the test is masked if the test is masked it displays Masked otherwise it displays none For the first test after startup or replacing cuvettes the system will automatically test the cuvette blank before sampling The possible test results are as follows Result 1 All cuvette blanks are valid and the system starts to test Result 2 Some cuvette blanks are invalid and the system reminds users to select if replacing the cuvettes See the figure Ho 1 trip of cuvettes are contaminated it may depress testing speed Do you want to change them zem E Figure 6 8 Reminding to replace cuvettes Select Yes to cancel this test After that users can replace the cuvettes Select No to continue this test And the system will automatically skip the cuvettes that cannot be used 35 Result 3 If the cuvette blank energy is too low all the cuvette blanks at 340 nm are lower than 30 000 the system will automatically display the following information Cuvettes blank too largelPozzible reason illezal cuvettes setting wrong gain setting or lamp 5 power too low Do you want to ignore
52. nance Click Maintenance in the main menu of software to open the maintenance page 92 Auto Rerun No Auto Send No Admin 2009 10 26 16 11 58 Help Sample request Status Results Parameter Reagent Setup Statistics Maintenance o Exit Backup Service Log Debug Blank Configure Configure QC results Back up Recover Date 2009 10 26 2009 10 26 Export Import Database Clear Export Import History results QC results Test parameters f Calibration Parameters es Export Import gt Start Pre Next amp Search Figure 9 1 Maintenance data maintenance Backup and restoration of configuration parameters Configuration parameters are the optimal corrective parameters that can adapt the hardware system and are obtained by debugging during the instrument production process These parameters include localization parameter motor control parameter temperature corrective parameter gain parameter cuvette blank threshold value and filter configuration parameter The functions of each parameter are as follows The localization parameter is used for correcting the mechanical localization parameter of the system such as the descending height of sampling needle the positions of liquid pipetting and releasing the position of photoelectric acquisition etc The motor control parameter is used for controlling the working mode of each motor The temperature c
53. nce a Exit Test Parameter Setup Calibration Setup Quality Control Setup Profile Setup Calculation Setup External Res gt Quality Control Quality Control No 12101 1090909 gt ALB 30 0 5 mmol L TP 36 5 0 5 mmol L New Delete New Delete Control Rule P Start Fi g Pre Nex amp Search Figure 4 9 Quality control setup Add a QC Press New button enter the control no and lot in the opened page and click OK to save Then add the QC items contained in the quality control product one by one Delete a QC Select the QC needed to delete from the QC list and press Delete button Add or modify QC items Select a QC and then press New button Select the QC item needed to modify in the opened page enter target value and SD and then press the Save button Delete a QC item Select a QC from the QC list and its contained QC items will be displayed on the left list Select the QC item needed to delete and then press Delete Items button Set QC rules Select a QC from the QC list and its contained QC items will be displayed in the left list Select the item needed to set QC rules and then press Control Rule button to open a page as illustrated in the following figure 18 Control rule setting zk101 ALB w 1 value exceed 45D 1 45 iw 1 value exceed 3SD 1 35 l 1 value exceed 25D 1 25 w 2 consecutive values exceed 25D 2 25 3consecutive values exceed 25D
54. niformly this causes the absorbance of test solution at a certain wavelength to decrease or increase uniformly the decrease or increase speed A min is directly proportional to the activity or concentration of test material Speed method e generally called dynamic method is also named as continuous monitoring method it is mainly used for the determination of enzyme activity In fact the substrate concentration is impossible to be large enough therefore during the reaction course when the substrate is consumed to a certain degree the reaction speed is not directly proportional to enzyme activity So the speed method is applicable to a specific time period 61 Absorbance Time l Delay Time A Measure Time Kinetic method 62 Appendix B Name and Content of Toxic and Harmful Substances or Elements in Product 1 Name and content of toxic and harmful substances or elements Toxic and harmful substances or elements Component Polybromi Polybromi name Lead Mercury Cadmium ere nated Nia Pb Hg Cd Cr VI Biphenyl Internal circuit board Photoelectric components Internal electronic wires Attachments o Expressing that the contents of the toxic and harmful substance in all uniform materials of this component are lower than the limit content prescribed in the Standard SJ T11363 2006 x Expressing that the content of the toxic and harmful substance at least in a uniform material of this component
55. nstant speed tests single reagent items and double reagent items can be tested at the same time e Single Reagent mode Only testing single reagent items 22 5 2 Print Setup Function description Set the title annotations and print format of a summary report Select Report Setup tab to open a page as illustrated in Figure 5 2 Lamp Stabilizing 37 1C Auto Rerun No Auto Send No Admin 2009 10 26 11 27 02 Help Sample request Status Results Parameter Reagent Setup Statistics Maintenance o Exit System Setup Report Setup Department Setup User Setup Unit Setup SI Setup Report format A4 With Full Name y Report title xx Hospital Report Report annotations Note xxxxxxxxxx 0 000000000000 j Print Sequence b Start y Pre Next Search Figure 5 2 Print setup Report format Several report formats are available to select from Report title Set the title of a patient s report for example Test Report of xxxxHospital Report annotations Set the annotation information of a patient s report for example Note This test result is only responsible for this sample Print sequence Set the print order for items in a summary report 5 3 Department Setup Function description Set the departments and doctors relevant to the patient information entry Select Department Setup tab to open a page as illustrated in Figure 5 3 23 Lamp Stabilizing 37 1C Auto Rerun No Auto Send No
56. ntered item setup parameter calibrator setup parameter QC setup parameter integration setup parameter calculation item setup parameter non biochemical item setup parameter reagent setup parameter system setup parameter worksheet setup and test data e Overall data Including all the data above Export and import of QC results Export all the QC test results within designated time period and the exported data will be saved to the designated path If required the QC results within designated time period can be imported from the file Export and import of history data Export the patient sample test results within designated time period and the exported data will be saved to the designated path If required the patient sample test results within designated time period can be imported from the file Export and import of item setup parameters Export the current item setup parameter of the system and the exported data will be saved to the designated path If required the item setup parameters can be imported from the file Note Only the administrator can perform the operations of configuration parameter restoration database restoration QC data import history data import importing of item setup parameters and data removal Auto Rerun No Auto Send No Admin 2009 10 26 16 12 16 Help Sample request Status Results Parameter Reagent Setup Statistics Maintenance o Exit Backup Service Log Debug Blank Configure
57. orrective parameter is used for correcting the temperature control accuracy of the reaction plate and preheating The gain parameter is used for controlling the corresponding signal gain coefficient of each wavelength The cuvette blank threshold value is used for setting the cuvette blank warning limit If a cuvette blank is beyond the set range the system considers it as invalidity and will not use this cuvette The filter configuration parameter is used for recording the corresponding wavelength of each filter configured by the system Damages or loss of configuration parameters can cause abnormal functioning of the instrument In order to prevent it please back up the configuration parameters under normal status so as to restore them to the backup status by using the restoration function if necessary Backup and restoration of database The database is used for storing data documents on setting parameters and history test results Please periodically back it up so as to restore it to the backup status if the data are damaged or lost 53 Data removal It is used fro removing the data with designated types in database The data types can be selected according to the actual requirements the removed data cannot be restored e History data Meaning patients test results and patients basic information records e QC data Meaning QC test results e Calibration data Meaning calibration test results e Set parameters Meaning the user e
58. s the Save button In order to restore the default test order press Home button and then press Save button Note The default test order is according to the letter order of item names Set special cleaning Press Special Cleaning button to open a page as illustrated in Figure 4 6 Special Cleaning Contaminant test Contaminated test R1 gt R1 R1 gt R2 Contaminanttest CR y CR P Contaminated test R1 gt R1 Routine cleaning Cleanser R1 R2 Routine cleaning Cleanser R2 gt R1 Routine cleaning Cleanser 40 R2 gt R2 C Routine cleaning Cleanser 40 39 DILUENT 40 ATER Delete Delete All Figure 4 6 Special cleaning setting There is cross contamination in different reagents of some items thereby affecting the test results It is necessary to perform special cleaning to reduce or eliminate this effect At this time users can designate the system to perform a specific cleaning operation for the according case by the button Special Cleaning For example CR can contaminate other items the sampling needle after sampling CR Reagent 2 15 should be performed special cleaning with the liquid in Reagent Position 40 before sampling other reagents At this time set according to the figure above and then press Save button To delete the set special cleaning item select a set item from the list and press the Delete button To delete all the set items press Delete All button
59. st accuracy be shaken e The cuvettes must be placed at a proper location and cannot 59 Appendix A Calculation Methods A 1 End Point Method After a period of time the test material is fully converted into product during the reaction course At this time the absorbance of the reaction solution does not increase or decrease any more i e reaching the reaction end point Calculate the concentration of test material according to the absorbance value at the reaction end point This method is named as end point method Absorbance Time Delay Time Measure Time End point method A 2 Fixed Time Method The fixed time method means that within a certain reaction time reaction speed is directly proportional to reactant concentration The reactant continues to consume therefore the overall reaction speed continues to be reduced i e the speed of absorbance increase or decrease continues to lower The measure must be performed within a specific time period and within the time period the absorbance increase or decrease is directly proportional to the concentration of test material 60 Absorbance Time Delay Time Measure Time Fixed time method A 3 Kinetic Method It means that reaction speed is directly proportional to the zero power of test material concentration i e irrelevant to the concentration of test material Therefore during the entire reaction course the reactant can generate a certain product u
60. system is testing the record will be automatically superadded and marked as a Retest test otherwise the system will automatically perform a retest to the test marked as Retest 38 Reminder information Displaying the reminder information during the process of worksheet testing including beginning and end time of the test information of deficient reagents samples and etc Press Clear button to remove the information in the list Remarks The symbols A and A gt express that test results are beyond linear range The symbol expresses that the retest result is still beyond linear range Status PREP indicates that testing will begin soon R1 indicates adding Reagent 1 S indicates adding test samples M1 indicates performing the first time of mixing R2 indicates adding Reagent 2 M2 indicates performing the second time of mixing Completed indicates that the testing is completed Unit SE indicates that the substrate is exhausted OL indicates that the result is beyond linear range 39 Chapter 7 Result Search 7 1 Searching Patient Records Function description Search the patient records meeting designated conditions Users can edit and view the selected basic patient information and test results one by one Click the Result in the main menu to open the patient record page as illustrated in Figure 7 1 Lamp Stabilizing T ERROR Auto
61. the user designated drawing mode Press Print button to print a QC graph as illustrated in the figure eo Real time QC graph All data are drawn and each QC data corresponds to a different abscissa of the graph e Daily QC graph Average the data obtained in the same day Only the corresponding points of daily mean values are drawn and the daily mean values correspond to different abscissa e Accumulation graph All data are drawn and the data obtained in the same day correspond to the same abscissa Delete QC data Select the QC data needed to delete from the QC data list and press Delete button Print QC data Select the QC data needed to print from the QC data list and press Print button 7 5 Searching Reagent Blank Function description Search the reagent blank history records of designated items Select the Search Reagent Blank page as illustrated in Figure 7 9 46 Lamp Stabilizing 37 0C Auto Rerun No AutoSend No Admin 2009 10 26 11 20 18 Help Sample request Status Results Parameter Reagent Setup Statistics Maintenance a Exit Patient Test Calibration Quality Control Reagent Blank Reagent Blank Remark 2009 10 12 10 00 00 8015 OR 2009 10 12 10 00 00 8015 OR 2009 10 13 10 12 00 8065 OR 2009 10 14 11 00 00 8034 OR 2009 10 20 10 00 00 8051 OR 2009 10 25 10 00 00 8023 OR 2009 10 26 10 00 00 1235 OR v lt gt Delete Delete All Print b Start H Paus W Sto
62. tivity samples exhaust the substrate and the dynamic reaction cannot continuously proceed In order to correctly reflect test results it is needed to set the substrate exhaustion limit at a specific absorbance This absorbance limit should be just the critical point of median area and non linear area zero grade dynamic method in reaction curve or the critical point of one grade reaction area and multiple grade reaction area fixed time method it is the minimal downwardness of reaction curve or maximal upwardness of reaction curve absorbance value when the substrate is not exhausted during the reaction time period The substrate exhaustion limit of an item closely correlates with the reagent kit used The unit is 1 10 000 absorbance The set value 0 means not judge Corrective coefficient lt is the linear corrective coefficient to system test results result measure value K B Generally it is not necessary to correct and at this time K 1 B 0 But for test items using factor method this coefficient can be used to correct errors caused by the instrument 14 a 15 a Set test sequence Press Test Sequence button to open a page as illustrated in Figure 4 5 14 Test Sequence Up Down Home Save Back Figure 4 5 Test sequence setup Select the item which test order is needed to adjust and then press Up and Down buttons or drag it with mouse to the proper position After all items have been adjusted pres
63. to show a distinction Statement BGT BioGenTechnologies GmbH owns the final interpretation right of this manual The illustrations provided in this manual can only be used as examples and they may be not fully accordant to the actual product Therefore please take the actual product as the standard The illustrations should not be used for other purposes Without BGT BioGenTechnologies s written permission any person or organization cannot reproduce revise or translate the contents of this manual BGT BioGenTechnologies GmbH is responsible for product safety reliability and performance if all the following requirements are met i e Assembly re debugging expansion modification and reparation should all be performed by personnel of BGT BioGenTechnologies s approval e Product operations are performed according to this manual e Relevant electric devices accord with national standards Note This instrument must be used by laboratory medicine professionals or trained doctors nurses or laboratory technicians Warning e fthe user fails to perform the maintenance service scheme satisfactorily abnormal instrument failure may occur and it may endanger physical health Please ensure that the analyzer is used under the conditions prescribed in the manual If beyond the conditions the analyzer may not function well in that case the test results are not reliable and it is also possible to damage instrument components and
64. up Statistics Maintenance a Exit Sample Disk No 1 Active Sample Request Quality Control Request Calibration Request List Test Position e M Small Tube i Profile Profile2 Repeat Mo Dilution ratio No Routine C STAT Patient Sample Sample No 0910261001 Sample Type Serum Edit Batch Save Save 6 Delete Delete All Input position and press Delete eg 15 8 means all b Stat Fi y Pre Next Search Figure 6 1 Sample request 6 1 1 Select a sample disk The system provides five editable worksheets in each worksheet 40 test samples can be entered in total 200 test samples can be entered Users can get five virtual sample disks from the five worksheets The system only tests active sample disk therefore users can realize the test to designated disk by setting different disks as active sample disk Please refer to Chapter 5 System Setup for setting active sample disk 6 1 2 Sample Request e Edit single sample Step 1 Designate position Input the sample position and press Enter button then the stored information of the sample will be displayed Step 2 Select repeat times and dilution ratio The default repeat times is 1 and dilution ratio is None Step 3 Select sample type as Routine or STAT 29 otep 4 Input sample no Press Edit button to input the basic patient information of the entered sample Step 5 Select test items and test profile
65. up box as illustrated in Figure 6 6 appears 33 Confirm Test mode SinglelDual Reagent sample disk No 1 Normal C Rerun Clear all results Cancel Figure 6 6 Worksheet confirmation Test mode It indicates the current selected test mode Refer to the description in 5 1 System Control Parameters Note The Single Dual Reagent mode can test both double reagent items and single reagent items but the Single Reagent mode can only test single reagent items Normal test It indicates to perform the normal test to the uncompleted or retested tests in the current worksheet Rerun test It indicates to perform a rerun test to the completed test with the result beyond the linearity range Clear all results It means clear the results of completed tests in current worksheet and restart all the tests Before confirming to start tests please confirm the information above if needing to modify press Cancel button Otherwise press OK button and then the test sorting page appears as illustrated in the following figure 34 Test sorting Normal Remark ALB 0909141007 ALB Routine 0909141001 ALT Routine 0909141002 CR Routine 3 11 Total 6 By Test By Sample OK Cancel Figure 6 7 Test sorting The system default ordering method is the designated order according to Test Sequence refer to 4 1 Test Parameter Setup If users need to readjust the order press By Test butto
66. vice Department 9 1 Maintenance Guide 9 1 1 Inspection before Analysis 1 The analyzer main part has no obvious physical damages There is no accumulated dust and liquid on the tabletop of the main part worktable The lines connect well The connection of the instrument to computer printer and etc is normal 2 The power supply is well connected and well grounded 3 The waste barrel is empty and the cleaning solution is enough 9 1 2 Periodic Maintenance 1 Daily maintenance Clean the dust on the worktable with a duster cloth that has soaked in neutral cleaning solution 2 Weekly maintenance 1 Softly clean the sampling needle and mixing arm with a duster cloth dipped in alcohol 2 Make sure that the flow rate of cleaning water flow in the two cleaning pools is normal 3 Monthly maintenance 1 Cleaning the external cleaning solution water tank 2 Cleaning the water supply and drainage systems 4 Non periodic maintenance as required 1 Check and replace the halogen lamp once per 1000 hours 2 Checkthe injector piston once per 150 hours 3 Check and replace the diaphragm pump once per 1000 hours 4 Check and replace the mixing arm once a year Note The replacement of diaphragm pump and mixing arm must be performed under the permission of BGT BioGenTechnologies GmbH s engineers 91 9 2 Common Troubles and Measures The software cannot function normally e Check if the analyzer main part is turn
67. ying with the printer e Connect the control computer to AC power supply with power cable Connecting instrument with control computer Insert one end of series interface cable into the series interface of instrument and the other end into the series interface COM1 of control computer or change the port setting to use the series interface COM2 Connecting instrument to power supply Connect the instrument to AC power supply with power cable 2 3 Turning On Off the Instrument Turn on the Instrument e Turn on the instrument power switch e Turn on the computer and start the instrument control software Turn off the Instrument First close the instrument control software and then turn off the instrument power switch 2 4 Instrument Calibration Requirements Calibrators can be used for calibrating the instrument It is not necessary to calibrate the instrument before each test however at least one calibration test should be performed for items needing to be calibrated Changes of system environment may bring some influences on testing therefore it is recommended to perform a calibration test after each instrument start up so as to ensure the accuracy of test results Refer to Chapter 6 Routine Testing for calibration test methods Chapter 3 Login and Shutdown 3 1 Login After start up run the CliChem 240 Software and the system opens the user log in interface as illustrated in Figure 3 1 Logging in User Admin
Download Pdf Manuals
Related Search
Related Contents
Edge Products 30104 Performance Module Installation PaseoBebe.com GDM-400PS2015-05-08 21:584.4 MB - Wiki Karat Instruction manual 取扱説明書及び安全ガイドを必ずご参照下さい。 **2014 年 3 月 19 日 Montageanleitung 2.7 - Bender T-4030 - Onkyo Instruction manual Manuel d'utilisation Bedienungsanleitung Copyright © All rights reserved.