1. August 2001 sss Seventh Edition part number Sixth Edition part number First Edition part number 16193 90000 Second Edition Third Edition 16193 90020 16193 90020 16193 90030 16193 90040 Safety Summary The following general safety precautions must be observed during all phases of operation service and repair of this instrument Failure to comply with these precautions or with specific WARNINGS given elsewhere in this manual violates safety standards of design manufacture and intended use of the instrument The Agilent Technologies assumes no liability for the customer s failure to comply with these requirements DO NOT Operate In An Explosive Atmosphere Do not operate the instrument in the presence of flammable gasses or fumes Operation of any electrical instrument in such an environment constitutes a safety hazard Keep Away From Live Circuits Operating personnel must not remove instrument covers Component replacement and internal adjustments must be made by qualified maintenance personnel Do not replace components with the power cable connected Under certain conditions dangerous voltages may exist even with the power cable removed To avoid injuries always disconnect power and discharge circuits before touching them DO NOT Service Or Adjust Alone Do not attempt internal service or adjustment unless another person capable of rendering first aid and resuscitation is present DO NO
2. 1 Assembly Replacement 4 8 Disassembly and Assembly 4 8 Pressure Arm Assembly 4 8 Pressure Arm Assembly Disassemble 4 8 Replaceable Parts ll 4 le s 4 8 Troubleshooting 4 9 Open Impedance Check ooa a a 4 9 Short Impedance Check 1 2 2 a 4 10 Short Impedance Repeatability Check 4 10 Contents 1 Figures Applicable DUT Size 1 3 Simulation Setup 1 5 16193A Product Overview 2 2 Electrodes configuration and DUT Size 2 2 2 2 2 8 Connecting the Test Fixture with APC 7 Connector 4291B 2 4 16193A Test Fixture Overview D 3 2 1 Replaceable Parts for Pressure Arm Assembly 2 2 e 4 5 Replaceable Parts for Test Stage Assembly llle 4 7 Obsolete parts manufactured earlier than 1996 4 8 Shorting Device Position 4 9 Tables Furnished Accessories E 1 2 Supplied Shorting Device Residual Inductance Typical 1 4 Option 010 Shorting Device Residual Inductance Typical 1 4 16198 Contents ll ll 4l s 2 2 Measurement Error Correcting Functions for Each Instrument 2 6 1 16193A Product Overview 4 2 Replaceable Parts for M
3. 2 6 Preparation for Use Operation Introduction This chapter describes how to use the test fixture and the measurement error correcting techniques for the test fixture Operating Flow Before performing a measurement you have to compensate for the residual error of the test fixture by using the measurement error correcting functions Perform the following steps when measuring the DUT The 16193A has inherent stray capacitance residual inductance and residual resistance that affect the measurement To compensate for these residuals and thus minimize their effect on measurement accuracy the measurement instrument s OPEN and SHORT compensation capabilities must be used gt indicates procedures in the following pages Select the DUT size Calibrate your analyzer Described in Chapter 2 Connect the Test Fixture Sel the electrical length The 16193A s electrical length is 14 0 mm typical value y To set the electrical length see your analyzer s Operation Manual Perform fixture To set the fixture and devices for fixture compensation see compensation gt How to set the SHORT condition and the OPEN condition SHORT compensation i LOAD compensation These procedures show how to set the test fixture for the OPEN SHORT LOAD compensation procedure OPEN compensation For how to measure OPEN SHORT LOAD see the Operation Manual supplied with your analyzer Y
4. Device 2 4x2 4x3 2 mm 8 16191 29004 1 Shorting Device 2 4x2 4x4 5 mm 9 1540 0692 1 Case for shorting devices 1 1 16193 90040 Operation and Service Manual 1 Not shown in the figure 4 2 Service Assembly Replacement This section includes Disassembly and Assembly Procedures and Replacement Parts for the Main Assembly Disassembly and Assembly Procedure This section contains the information required to disassemble and assemble the Main Assembly Pressure Arm Assembly Removal 1 Rotate the Locking Knob counter clockwise and remove it see Table 4 2 2 Remove the Pressure Arm Assembly including Spring Bushing and Washer from the Test Stage Assembly Pressure Arm Assembly Disassemble 1 Remove two screws marked with 8 and 12 in Figure 4 1 2 Remove bottom cover marked with 1 in Figure 4 1 3 Remove other screws and other parts Replaceable Parts Table 4 2 through Table 4 4 show and list the replaceable parts for the Main Assembly The parts listed can be ordered from your nearest Agilent Technologies Office Ordering information must include the Agilent part number and the quantity required Service 4 3 Table 4 2 Replaceable Parts for Main Assembly Reference Agilent Part Designator Number Qty Description 1 16193 23004 1 Locking Knob 2 1460 2368 1 Spring 3 16198 25008 1 Bushing 4 not assigned 1 Pressure Arm Assembly 5 16193 87101 1 Label
5. H mm P N 16191 29008 I ences 3 2 L x 1 6 W x 0 8 H mm Electrical length when the length between electrodes is 2 mm 14 0 mm Additional Error 1 5 x f Repeatability for inductive component ss 30 250 x f mQ impedance of 30 mQ 40 pH o 2 30 x f uS admittance of 2 pS 5 fF f frequency GHz H 3mm W 3mm 0 5mm lt L lt 3 2mm electrodes AC201001 Figure 1 1 Applicable DUT Size General Information 1 3 Residual Inductance of the Shorting Bar The usual method to compensate the test fixture s residual inductance is to let SHORT OH In this method the measurement result is the relative value of the measured impedance to the shorting bar s impedance The short bar s residual inductance as a result of its size and shape is not estimated On the other hand there is a definition method to let SHORT x H In this method the measurement result is the absolute value of the device s impedance The short bar s residual inductance as a result of its size and shape is estimated under specific conditions and is used as a reference value This method is useful for devices with values which are close to the short conditions of the measurement system The reference inductance values presented Table 1 2 and Table 1 3 were simulated as the relative difference to a disk type t
6. is not shown in Figure 2 1 2 Depending on your order configuration 3 The magnifying lens is packed separately from the 16193A body Connect it as shown in Figure 2 1 4 Not contained if you ordered the 16193A option 001 2 2 Preparation for Use Adjust the Configuration to SMD Size The 16193A s electrodes length are 0 2 mm 0 5 mm 1 mm and 2 mm change the electrodes configuration to match the SMD as follows d electrode length 1 Loose the o screws Tighten the screws E Ol no L N 2 Change the position AC202 002 Figure 2 2 Electrodes configuration and DUT Size Preparation for Use 2 3 Connecting the Test Fixture for Use Note Some instruments require an ada
7. shorting device on the electrode 5 D Loosen the knob 2 Position the DUT holder so that the short bar contacts both electrodes Tighten the knob to fix the DUT holder cA Shorting Device 06103046 06103047 6 D Reposition the positioner so that the nose of the 7 D Push the arm lifter so that the arm latch is pressure arm is just above the shorting device released Let down the pressure arm to press down Tighten the locking knob the shorting device 3 4 Operation Setting the OPEN condition C610 3051 06103048 1 O Push the arm lifter Push the arm latch 2 Remove the shorting device button Release the arm lifter while pushing the arm latch button so that the pressure arm is held How to connect the DUT DUT 06103042 C6103049 1 D Loosen the locking knob Rotate the 2 Place the DUT on the electrode positioner until the nose of the pressure arm is located outside the frame Operation 3 5 C6103044 06103046 3 D Loosen the knob Position the DUT holder so Reposition the positioner so that the nose of the that the D
8. the 16193A meets the following specifications Maximum DC Bias Voltage I ee 40V Operating Temperature ooo 55 to 85 C Operating Humidity wet bulb temperature lt 40 C AN 15 to 95 RH Non operating Temperature o 55 to 85 C Non operating Humidity wet bulb temperature lt 65 C IEEE Up to 90 RH Weight 500 g Dimension ooo 150 W x 70 x 80 D mm 1 2 General Information Supplemental Performance Characteristics This section lists supplemental performance characteristics Supplemental performance characteristics are not specifications but are typical characteristics included as additional information for the operator Supplemental performance characteristics are not guaranteed Frequency Range e eaa DC to 2 GHz Applicable DUT size see Figure 1 1 sss sss 0 5 mm to 3 2 mm Supplied shorting device size P N 16191 29001 ssssssssssssssss II e 1x1 x 2 4 mm P N 16191 29002 ssssssssssssssse e e e aes 1 6 x 2 4 x 2 mm P N 16191 29008 I ehe n 2 4 x 2 4 x 3 2 mm P N 16191 29004 iiissssssssssssssss I eme n 2 4 x 2 4 x 4 5 mm Option 010 shorting device size P N 16191 29005 I 1 0 L x 0 5 W x 0 5 H mm P N 16191 29006 ssssssssssssss I e 1 6 L x 0 8 W x 0 8 H mm P N 16191 29007 I 2 0 L x 1 2 W x 0 8
9. 04 1 Ground Plate 10 16193 60011 1 Contact Assembly 11 0515 0952 3 Screw Flat Head M2L4 12 0515 0976 2 Screw Pan Head M2L6 13 16193 24005 2 Spacer 14 1250 0907 1 Contact RF Connector Caution If your 16193A test fixture was manufactured earlier than 1996 and you need to replace either the plate Figure 4 3 9 or the contact assembly Figure 4 3 10 it will be necessary to buy replacements for both parts Figure 4 2 9 and 10 since they were modified at the same time Neither of the new parts Figure 4 2 9 and 10 can be fitted to the other obsolete parts Figure 4 3 9 and 10 AC204008 Figure 4 3 Obsolete parts manufactured earlier than 1996 4 8 Service Troubleshooting This section includes the functional check procedure used to troubleshoot the 16193A Small Side Electrode SMD Test Fixture Open Impedance Check 1 Set the Size Selection Frame to the position for small size DUT 2 Adjust the Device Holder for connecting the 1x1x2 4 mm Shorting Device PN 16191 29001 Refer to Figure 4 4 Shorting Device PN M NEN 2 4 mm AC204005 Figure 4 4 Shorting Device Position 3 Place the fixture on the calibrated APC 70 terminal of the 4291B RF Impedance Material Analyzer 4 Set the 4291B RF Impedance Material Analyzer as follows Measurement Parameter Cp parallel capacitance Start Frequency 100 MHz Stop Fre
10. 6 16193 25004 1 Washer 7 not assigned 1 Test Stage Assembly 4 4 Service Figure 4 1 Replaceable Parts for Pressure Arm Assembly Service 4 5 Table 4 3 Replaceable Parts for Pressure Arm Assembly Reference Agilent Part Designator Number Qty Description 1 16193 04002 1 Bottom Cover 2 1460 2367 1 Spring 3 16193 23005 1 Shaft 4 2950 0054 1 Nut 5 2190 0054 1 Washer 6 16 0515 0914 2 Screw Flat Head M3L6 7 16193 25006 1 Arm Latch 8 12 0515 0914 2 Screw Flat Head M3L6 9 16193 04001 1 Top Cover 10 16193 24008 1 Screw 11 0515 1052 1 Screw Socket Head M3L12 13 16193 25007 1 Screw 14 3050 1139 1 Flat Washer Teflon 15 1460 2316 1 Spring 17 16193 25008 1 Spacer 18 16193 01201 1 Angle 19 16193 25005 2 Bushing 20 0515 0976 1 Screw Pan Head M2L6 21 3050 1066 1 Flat Washer M2 22 16193 24007 2 Nut 23 3030 0285 1 Set Screw 6 32 24 16193 23001 1 Rod 25 16193 23002 1 Holder 4 6 Service AC204007 Figure 4 2 Replaceable Parts for Test Stage Assembly Service 4 7 Table 4 4 Replaceable Parts for Test Stage Assembly Reference Agilent Part Designator Number Qty Description 1 16193 24011 1 Device Holder Knob 2 16193 25009 1 Device Holder 3 0515 1050 2 Screw Socket Head M3L8 4 16193 24004 1 Size Selection Frame 5 16193 00613 1 Plate 6 16193 23003 1 Flange 7 0515 0914 6 Screw Flat Head M3L6 8 16193 00201 1 Table 9 16198 006
11. Agilent 16193A Small Side Electrode SMD Test Fixture Operation and Service Manual Agilent 16193A Small Side Electrode SMD Test Fixture Operation and Service Manual RE Agilent Technologies Agilent Part No 16193 90040 Printed in JAPAN August 2001 Seventh Edition Notice The information contained in this document is subject to change without notice This document contains proprietary information that is protected by copyright All rights are reserved No part of this document may be photocopied reproduced or translated to another language without the prior written consent of the Agilent Technologies Agilent Technologies Japan Ltd Component Test PGU Kobe 1 3 2 Murotani Nishi ku Kobe shi Hyogo 651 2241 Japan Copyright Agilent Technologies Japan Ltd 1993 1996 1999 2000 2001 Manual Printing History The manual printing date and part number indicate its current edition The printing date changes when a new edition is printed Minor corrections and updates that are incorporated at reprint do not cause the date to change The manual part number changes when extensive technical changes are incorporated December 1993 June 1996 8 eee eh e e eee October 1996 isses eee August 1999 Fourth Edition part number Fifth Edition part number December 1999 September 2000
12. Ltd Call Center 9 1 Takakura Cho Hachioji Shi Tokyo 192 8510 Japan tel 81 426 56 7832 fax 81 426 56 7840 Latin America Agilent Technologies Latin American Region Headquarters 5200 Blue Lagoon Drive Suite 950 Miami Florida 33126 U S A tel 305 267 4245 fax 305 267 4286 Australia New Zealand Agilent Technologies Australia Pty Ltd 347 Burwood Highway Forest Hill Victoria 3131 tel 1 800 629 485 Australia 16193 90040 O Copyright Agilent Technologies Inc 1993 1996 1999 2000 2001 Printed in Japan 8 2001 Reorder No 16193 90040
13. T Substitute Parts Or Modify Instrument Because of the danger of introducing additional hazards do not substitute parts or perform unauthorized modifications to the instrument Return the instrument to a Agilent Technologies Sales and Service Office for service and repair to ensure the safety features are maintained Dangerous Procedure Warnings Warnings such as the example below precede potentially dangerous procedures throughout this manual Instructions contained in the warnings must be followed Warning Dangerous voltages capable of causing death are present in this instrument Use extreme caution when handling testing adjusting this instrument The voltage levels found in this test fixture when used with the intended instruments do not warrant more than normal safety precautions for operator safety Operating Precaution Do not exceed the operating input power voltage and current level and signal type appropriate for the instrument being used refer to your instrument s operation manual Caution Electrostatic discharge EspD can damage the highly sensitive microcircuits in your instrument ESD damage is most likely to occur as the test fixtures are i being connected or disconnected Protect them from ESD damage by wearing a grounding strap that provides a high resistance path to ground Alternatively ground yourself to discharge any static charge built up by touching the outer shell of any grounded i
14. UT contacts both electrodes Tighten the pressure arm is just above the DUT Tighten the knob locking knob C6103050 3 D Push the arm lifter so that the arm latch is released Let down the pressure arm to press down the DUT 3 6 Operation Service Introduction This chapter covers the following subjects m General Information m Assembly Replacement Disassembly Procedure for the Main Assembly Replaceable Parts m Troubleshooting Warning These servicing instructions are for use by qualified personnel only Do NOT perform any servicing other than that contained the operating section unless you qualified to do so Caution When you repair the 16193A put on lint free groves to avoid contaminating inner parts of the 16193A Agilent Technologies supplies lint free groves i Agilent part number 9300 0163 General Information Table 4 1 shows all items included with the 16193A Small Side Electrode SMD Test Fixture Service 4 1 Table 4 1 16193A Product Overview Lr M l 16191 29021 Styrofoam Holder Reference Agilent Part Designator Number Qty Description 1 not assigned 1 Main Assembly 2 16193 60002 1 Magnifying Glass 3 8710 1181 1 Wrench 4 8710 2081 1 Tweezers 5 16191 29001 1 Shorting Device 1x 1x2 4 mm 6 16191 29002 1 Shorting Device 1 6x2 4x2 mm 7 16191 29008 1 Shorting
15. ain Assembly 2 2 2 2 2 25 4 4 Replaceable Parts for Pressure Arm Assembly 4 6 Replaceable Parts for Test Stage Assembly lll 4 8 Open Impedance Value Guideline MM D 4 10 Short Impedance Value Guideline MM D 4 10 Short Impedance Repeatability Guideline MM D 4 10 Contents 2 General Information Introduction The purpose of this manual is to enable you to use your 16193A Small Side Electrode SMD Test Fixture efficiently and confidently Manual Summary This manual contains the following m The specifications of the 16193A see this chapter m Installing the 16193A see Chapter 2 m Operating the 16193A see Chapter 3 m Ordering replaceable parts for the 16193A see Chapter 4 Product Description The 16193A is used to measure a small side electrodes surface mount devise SMD with high repeatability Applicable Instruments The 16193A has been designed to operate specifically with the following LCR meters and impedance analyzers You can directly connect the 16193A to 4191A 4291B E4991 A You can use the 16193A with 43961A 4395A option 010 4396A B option 010 You can use the 16193A with 41951A 4195A You can use the 16193A with RF adapter 42942A 4294A You can use the 16193A with RF adapter 16085B 4284A1 You can use the 16193A with RF adapter 16099A 4194A with 41941A 1 The 42844 is a typical in
16. echnologies warrants that its software and firmware designated by Agilent Technologies for use with an instrument will execute its programming instruction when property installed on that instrument Agilent Technologies does not warrant that the operation of the instrument or software or firmware will be uninterrupted or error free Limitation Of Warranty The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer Buyer supplied software or interfacing unauthorized modification or misuse Operation outside of the environmental specifications for the product or improper site preparation or maintenance No other warranty is expressed or implied Agilent Technologies specifically disclaims the implied warranties of merchantability and fitness for a particular purpose Exclusive Remedies The remedies provided herein are buyer s sole and exclusive remedies Agilent Technologies shall not be liable for any direct indirect special incidental or consequential damages whether based on contract tort or any other legal theory Assistance Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products For any assistance contact your nearest Agilent Technologies Sales and Service Office Addresses are provided at the back of this manual vi Safety Symbols General definitions of safety symbols used on equipment or i
17. ermination on either the Y mm or the 3 5 mm connector The measurement of these short bars under other conditions than shown below cannot reproduce the reference inductance values Table 1 2 Supplied Shorting Device Residual Inductance Typical Shorting Bar l mm d mm h mm Offset mm Connector Inductance Typical P N 16191 29001 1 2 4 1 0 75 3 5 mm 0 2 nH P N 16191 29002 1 6 2 4 2 0 45 3 5 mm 0 2 nH P N 16191 29002 2 2 4 1 6 0 25 3 5 mm 0 2 nH P N 16191 29002 2 4 2 1 6 1 3 7mm 0 7 nH P N 16191 29003 3 2 2 4 2 4 0 9 7mm 0 6 nH P N 16191 29004 4 5 2 4 2 4 0 7mm 0 6 nH Table 1 3 Option 010 Shorting Device Residual Inductance Typical Shorting Bar l mm d mm h mm Offset mm Connector Inductance Typical P N 16191 29005 1 0 0 5 0 5 0 75 3 5 mm 0 5 nH P N 16191 29006 1 6 0 8 0 8 0 45 3 5 mm 0 4 nH P N 16191 29007 2 0 1 2 0 8 1 5 7 mm 0 9 nH P N 16191 29008 3 2 1 6 0 8 0 9 7mm 0 8 nH 1 4 General Information Short Bar Center Electrode Outer Electrode 7 mm 3 5 mm Connector Figure 1 2 Simulation Setup General Information 1 5 Preparation for Use Introduction This chapter explains how to install the 16193A The topics include the following m Initial inspection m Connecting the test fixture for use m Repackaging the test fixture for shipment m Measurement error correcting function for each instrument I
18. esult damage to or destruction of part or all of the product Note denotes important information It calls attention to a procedure practice condition or the like which is essential to highlight vii APC 7 is a U S registered trademark of the Bunker Ramo Corporation viii Contents 1 General Information Introduction 2 1 1 Manual Summary 2 2 ee a L 1 1 Product Description MEL 1 1 Applicable Instruments VV 1 1 Options Available MM D 1 2 Accessories Supplied 6 2 2 1 2 Specifications 2 1 2 Supplemental Performance Characteristics 0 0 2 00 1 3 Residual Inductance of the Shorting Bar aoa a a a a a a a 1 4 2 Preparation for Use Introduction 2 2 1 Initial Inspection 4 ee 2 1 Adjust the Configuration to SMD Size 2 3 Connecting the Test Fixture for 2 4 Repackaging the Test Fixture For Shipment 2 5 Measurement Error Correcting Function for Each Instrument 2 6 3 Operation Introduction 9 1 Operating Flow 9 1 Overview 9 2 How to set the SHORT condition and the OPEN condition 3 3 Setting the SHORT condition 3 3 Setting the OPEN condition ooo a a 35 How to connect the 3 5 4 Service Introduction 2 a 4 1 General Information 1 2 4
19. n manuals Warning Y Caution Y Note Instruction manual symbol the product is marked with this symbol when it is necessary for the user to refer to the instruction manual in order to protect against damage to the instrument Indicates dangerous voltage terminals fed from the interior by voltage exceeding 1000 volts must be so marked Protective conductor terminal For protection against electrical shock in case of a fault Used with wiring terminals to indicate the terminal which must be connected to ground before operating equipment Low noise or noiseless clean ground earth terminal Used for a signal common as well as providing protection against electrical shock in case of fault A terminal marked with this symbol must be connected to ground in the manner described in the installation Operation manual and before operating the equipment Frame or chassis terminal A connection to the frame chassis of the equipment which normally includes all exposed metal structures Alternating current power line Direct current power line Alternating or direct current power line Warning denotes a hazard It calls attention to a procedure practice condition or the like which if not correctly performed or adhered to could result in injury or death to personnel Caution sign denotes a hazard It calls attention to a procedure practice condition or the like which if not correctly performed or adhered to could r
20. nd check the value is within the Short Impedance Value Guideline described in Table 4 6 Calculate the difference between the first Ls measurement value and second Ls measurement value at each frequency The guideline for the short impedance repeatability is as follows Table 4 7 Short Impedance Repeatability Guideline Parameter Frequency Guideline difference Ls 100 MHz 45 pH Ls GHz 20 pH 4 10 Service REGIONAL SALES AND SUPPORT OFFICES For more information about Agilent Technologies test and measurement products applications services and for a current sales office listing visit our web site http www agilent com find tmdir You can also contact one of the following centers and ask for a test and measurement sales representative 11 29 99 United States fax 61 3 9272 0749 Agilent Technologies tel 0 800 738 378 New Zealand Test and Measurement Call Center fax 64 4 802 6881 P O Box 4026 Englewood CO 80155 4026 Asia Pacific tel 1 800 452 4844 Agilent Technologies 24 F Cityplaza One 1111 King s Road Canada Taikoo Shing Hong Kong Agilent Technologies Canada Inc tel 852 3197 7777 5150 Spectrum Way fax 852 2506 9284 Mississauga Ontario L4W 5GI tel 1 877 894 4414 Europe Agilent Technologies Test amp Measurement European Marketing Organization P O Box 999 1180 AZ Amstelveen The Netherlands tel 31 20 547 9999 Japan Agilent Technologies Japan
21. nitial Inspection The small side electrode SMD test fixture has been carefully inspected before being shipped from the factory It should be in perfect physical condition no scratches dents or the like It should also be in perfect electrical condition Verify this by carefully performing an incoming inspection to check the small side electrode SMD test fixture set for signs of physical damage and missing contents If any discrepancy is found notify the carrier and Agilent Technologies Your Agilent Technologies sales office will arrange for repair and replacement without waiting for the claim to be settled m Inspect the shipping container for damage Keep the shipping materials until the inspection is completed m Verify that the shipping container contains everything shown in Figure 2 1 and listed in Table 2 1 m Inspect the exterior of the 16193A for any signs of damage Preparation for Use 2 1 Figure 2 1 16193A Product Overview Table 2 1 16193A Contents Description Agilent Part Quantity Number D Test fixture 16193A 1 0 Operation and Service Manual P N 16193 90040 1 Shorting device 1x1x2 4 mm P N 16191 29001 1 1 6x2 4x2 mm P N 16191 29002 1 2 4x2 4x3 2 mm P N 16191 29003 1 2 4x2 4x4 5 mm P N 16191 29004 1 Case for shorting device P N 1540 0692 1 5 Magnifying lens P N 16193 60002 1 Tweezers P N 8710 2081 1 D Wrench P N 8710 1181 1 1 Operation and Service Manual
22. nstrument chassis before touching the test port connectors Never touch the test clip contacts Use a work station equipped with an anti static work surface Certification Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology to the extent allowed by the Institution s calibration facility or to the calibration facilities of other International Standards Organization members Warranty This Agilent Technologies instrument product is warranted against defects in material and workmanship for a period of one year from the date of shipment except that in the case of certain components listed in Instrument Specifications of this manual the warranty shall be for the specified period During the warranty period Agilent Technologies will at its option either repair or replace products which prove to be defective For warranty service or repair this product must be returned to a service facility designated by Agilent Technologies Buyer shall prepay shipping charges to Agilent Technologies and Agilent Technologies shall pay shipping charges to return the product to Buyer However Buyer shall pay all shipping charges duties and taxes for products returned to Agilent Technologies from another country Agilent T
23. ou can omit LOAD compensation Connect the device under test DUT N ly Measurement To connect the DUT see gt How to connect the DUT The compensating functions are different for each instrument See Table 2 2 and perform the listed functions Operation 3 1 Overview Pressure adjustment Arm lifer Positioner Locking knob Pressure arm Device holder knob Device holder Arm latch AC203002 Arm lifter Positioner Locking knob Pressure arm Device holder knob Device holder Size selection frame Electrodes Arm latch Pressure adjustment Shorting device agnifying lens Size selection frame Electrodes Figure 3 1 16193A Test Fixture Overview Tilts the pressure arm up Slides to set the pressure arm atop the DUT Fixes the positioner Fixes the DUT Fixes the holder Holds the DUT Fixes the electrode These are connected to the device Latches the pressure arm tilted up Push the arm latch tilt the arm up by pushing the arm lifter To set the latch to off push the arm lifter Adjusts the pressure load on the device Not shown in Figure 3 1 Used for short compensation Four types of shorting devices are supplied with the 16193A Use one that has dimensions similar to the DUT When making a shorting device of same dimensions as your DUT very low residual impedance and a high conductivity metal that is not easily co
24. packing material Use sufficient shock absorbing material on all sides of the 16193A to provide a thick firm cushion and to prevent movement Seal the shipping container securely and mark it FRAGILE Preparation for Use 2 5 Measurement Error Correcting Function for Each Instrument Each LCR meter or analyzer has measurement error correcting functions Table 2 2 shows the functions of the instrument that can be used for calibration setting the electrical length and Open Short Load compensation Table 2 2 Measurement Error Correcting Functions for Each Instrument Instrument Calibration Electrical length OPEN SHORT LOAD Compensation Compensation Compensation 4291B Calibration Selecting Test Fixture OPEN SHORT LOAD OPEN Port Extension Compensation Compensation Compensation SHORT LOAD LOW LOSS Capacitor 41914 Calibration Electrical Length No capability No capability No capability 08 00 50 Q 4194A with No capability No capability OPEN Correction SHORT Correction LOAD Correction 16085B 4195A with Calibration OPEN Port Extension 0 S OFFSET 0 Q OFFSET No capability 41951A SHORT Port 1 or Port 2 LOAD 4194A with Calibration No capability Compensation Compensation No capability 41941A B 08 ZERO OPEN ZERO SHORT 16099A 00 50 Q 4194A with Calibration No capability Compensation Compensation No capability 16085B 08 ZERO OPEN ZERO SHORT 00 50 Q 1 According to demand for precise measurement
25. pter to be connected to the 16193A see i Applicable Instruments in Chapter 1 Y 1 Calibrate your analyzer at the APC 7 connector plane before connecting test fixture See Table 2 2 for the applicable error correcting function After calibration disconnect any standards from the APC 7 connector 2 Place the fixture on the APC 7 connector 3 Tighten turn counterclockwise the coupling nut of the APC 7 connector uen TEST Fiume 2 Place the fixture on APC 7 connector 3 Tighten the coupling nut of the APC 7 connector AG202 001 Figure 2 3 Connecting the Test Fixture with APC 7 Connector 4291B 2 4 Preparation for Use Repackaging the Test Fixture For Shipment If shipment to an Agilent Technologies service center is required each test fixture should be repackaged using the original factory packaging materials If this material is not available comparable packaging materials may be used Wrap the small side electrode SMD test fixture in heavy paper and pack in anti static plastic
26. quency 1 GHz OSC Level 0 12 V Number of Point 2 Point Averaging Factor 16 Point Averaging ON 5 Perform a single sweep measurement Service 4 9 6 Read Cp value at 100 MHz and 1 GHz The guideline for the open impedance value is as follows Table 4 5 Open Impedance Value Guideline Parameter Frequency Guideline absolute value Cp 100MHz 750 fF 400 fF Cp 1 GHz 750 fF 400 fF Short Impedance Check l After the Open Impedance Check is completed put the 1x1x2 4 mm Shorting Device PN 16191 29001 between the center electrode and the ground electrode refer to Figure 4 4 Set the 4291B RF Impedance Material Analyzer as follows Measurement Parameter Ls series inductance Start Frequency 100 MHz Stop Frequency 1 GHz OSC Level 0 12 V Number of Point 2 Point Averaging Factor 16 Point Averaging ON Perform a single sweep measurement Read Ls value at 100 MHz and 1 GHz The guideline for the short impedance value is as follows Table 4 6 Short Impedance Value Guideline Parameter Frequency Guideline absolute value Ls 100 MHz 2nH 1 nH Ls 1 GHz 2nH 1 nH Short Impedance Repeatability Check 1 After the Short Impedance Check is completed remove the Shorting Device and put it back on the same place again refer to Figure 4 4 Perform a single sweep measurement with the same measurement conditions Read Ls value at 100 MHz and 1 GHz a
27. rroded must be used to construct the shorting device It must also be clean Magnifying lens Not shown in Figure 3 1 Used to set a small device within the electrodes Warning DO NOT look at the sun through the magnifying glass DO NOT operate or leave the magnifying glass under direct sunlight 3 2 Operation How to set the SHORT condition and the OPEN condition Setting the SHORT condition Arm Lifter Positioner C6103040 06103041 The names of each part are listed above 1 D Push the arm lifter 2 Push the arm latch button Release the arm lifter while pushing the arm latch button so that the pressure arm is held 06103042 Ol 06103043 Electrode Gap 0 2mm 0 5mm imm 2mm es 2 D Loosen the locking knob Rotate the positioner until the nose of the pressure arm is located outside the stage 3 Position the size selection frame depending on the size of the DUT This procedure shows the case for D as an example Operation 3 3 TO Shorting Device C6103045 C6103044 4 Place the
28. strument Other 4 terminal pair instruments that can be used with the 16085B are also applicable 2 The 4194A with 419414 is a typical instrument Other instruments that can be used with the 16099A are also applicable General Information 1 1 Options Available The following option is supplied for the 16193A Option 001 Delete the magnifying lens and tweezers Option 010 Add industry standard size short bar set Accessories Supplied The following accessories are supplied with the 16193A Table 1 1 Furnished Accessories Description Part Number Quantity Operation and Service Manual P N 16193 90040 1 Shorting device 4 types P N 16191 29001 1 P N 16191 29002 1 P N 16191 29003 1 P N 16191 29004 1 Case for shorting devices P N 1540 0692 1 Option 010 P N 16191 29005 1 P N 16191 29006 1 P N 16191 29007 1 P N 16191 29008 1 Magnifying lens P N 16193 60002 1 Tweezers P N 8710 2081 1 Wrench P N 8710 1181 1 1 Option 010 sizes are the same as industry standard EIA EIAJ SMD sizes This short bar set has the following SMD sizes included 1005 mm 0402 inch 1608 mm 0603 inch 2012 mm 0805 inch 3216 mm 1206 inch Order option 010 if the SMD that is to be measured has the same size as the EIA EIAJ sizes Specifications This section lists the complete 16193A specifications These specifications are the performance standards and limits against which the 16193A is tested When shipped from the factory