traceable® jumbo refrigerator/freezer thermometer instructions
Contents
1. To change between ambient IN and probe OUT temperature display modes press the IN OUT button located on the back of the unit To change between displaying the memory MIN MAX and the alarm limits HI LO press the MODE button DISPLAYING F OR C To display the temperature readings in Fahrenheit or Celsius press the C F button located on the back of the unit MINIMUM AND MAXIMUM MEMORY There are four points that are automatically recorded into memory Minimum MIN Ambient Temperature Achieved Maximum MAX Ambient Temperature Achieved Minimum MIN Probe Temperature Achieved Maximum MAX Probe Temperature Achieved All four memories are independent of each other Minimum and maximum temperature memories are NOT programmable The minimum temperature recorded into memory is the minimum temperature achieved since the last time the memory was cleared The maximum temperature recorded into memory is the maximum temperature achieved since the last time the memory was cleared The minimum and maximum temperature memories are maintained over the period since the memory was cleared Minimum and maximum memories are recorded for both ambient and probe temperatures regardless of the display mode CLEARING THE PROBE OUT TEMPERATURE MEMORY 1 Select the Probe Temperature Display Mode by pressing the IN OUT button located on the back of the unit until OUT appears on the display 2 Press the MODE button until MIN
2. 2 Once the above items have been manually recorded clear both the minimum and maximum temperature memory See the Clearing the Probe OUT Temperature Memory section By clearing the memory each day the minimum and maximum temperature memory will provide a record of the minimum and maximum temperature that has been achieved inside the refrigerator over the past 24 hour monitoring period The memory will also allow the user to see if the temperature inside the refrigerator went outside of the acceptable range when there was no one available to hear the alarm overnight ALL OPERATIONAL DIFFICULTIES If this unit does not function properly for any reason replace the battery with a new high quality battery see the Battery Replacement section Low battery power can occasionally cause any number of apparent operational difficulties Replacing the battery with a new fresh battery will solve most difficulties BATTERY REPLACEMENT Erratic readings a faint display no display or a battery symbol appearing on the display are all indications that the battery must be replaced Open the battery compartment by depressing the tab located at the top of the battery cover and lifting it out The battery cover is hinged at the bottom so that the battery compartment may be opened without completely removing the battery cover Remove the exhausted battery and replace with a AAA alkaline battery Make certain to insert the new battery with
3. Set a screw into the wall at the desired location Do not set the screw flush to the wall the head of the screw will need to slip into the receptacle on the back of the unit Once the screw has been properly set hang the unit in place by sliding the receptacle on the back of the unit over the head of the screw USING THE ALARM AND MEMORY TO MONITOR A REFRIGERATOR FREEZER Following is an example of how to use the alarm and memory to monitor the temperature inside a refrigerator or freezer This example is provided only as a helpful guide and is not intended to replace existing facility requirements or procedures In this example the refrigerator temperature must be monitored and logged for each 24 hour period and certain actions must be taken if temperature falls below 0 C or rises above 5 C at any time during the 24 hour period Unit Setup Example Plug the probe into the unit Install the battery Place the probe sensor inside the refrigerator Place the display outside the refrigerator Select the Probe Temperature Display mode See the Temperature Display Modes section OV EO Ne At this point if using a bottle probe allow sufficient time for the bottle probe to reach equilibrium with the true current temperature inside the refrigerator 6 Set the probe temperature low alarm limit to 0 C See the Setting Probe OUT Temperature Alarm Limits section 7 Set the probe temperature high alarm limit to 5 C See
4. and MAX appears on the display 3 Press the MIN RESET button to clear the minimum temperature memory The display will show 88 8 under MIN and will then display the new minimum temperature memory 4 Press the MAX RESET button to clear the maximum temperature memory The display will show 88 8 under MAX and will then display the new maximum temperature memory CLEARING THE AMBIENT IN TEMPERATURE MEMORY 1 Select the Ambient Temperature Display Mode by pressing the IN OUT button located on the back of the unit until IN appears on the display 2 Press the MODE button until MIN and MAX appears on the display 3 Press the MIN RESET button to clear the minimum temperature memory The display will show 88 8 under MIN and will then display the new minimum temperature memory 4 Press the MAX RESET button to clear the maximum temperature memory The display will show 88 8 under MAX and will then display the new maximum temperature memory ALARMS Both ambient IN and probe OUT temperature alarm limits may be set in 1 increments With the alarm enabled A the unit will sound an alarm when the temperature being displayed is outside the alarm limits that have been set for that display mode equal to or lower than the low alarm set point or equal to or greater than the high alarm set point Example If the unit is displaying the ambient temperature IN and the probe temperature OUT is o
5. the proper polarity as indicated in the illustration in the battery compartment Close the battery cover Replacing the battery has the same affect as pressing the RESET button See the RESET BUTTON section Warranty Service or Recalibration For warranty service or recalibration contact CONTROL COMPANY 4455 Rex Road Friendswood Texas 7 546 USA Ph 281 482 1714 Fax 281 482 9448 E mail sales control3 com www control3 com Control Company is ISO 9001 Quality Certified by DNV and ISO 17025 accredited as a Calibration Laboratory by A2LA Control Company Cat No 92 4148 00 Traceable is a registered trademark of Control Company 2008 Control Company
6. TRACEABLE JUMBO REFRIGERATOR FREEZER THERMOMETER INSTRUCTIONS SPECIFICATIONS Range 58 0 to 158 0 F 50 0 to 70 0 C Resolution 0 1 Update Rate 10 seconds Battery AAA 1 5V Probe Supplied Model 4148 Supplied with a standard probe sensor with cable Both sensor and cable may be immersed in liquid Probe sensor may be mounted using the supplied sensor holder and mounting screw Model 4548 amp 4648 Supplied with a bottle sensor and cable Bottle sensor is filled with a patented nontoxic glycol solution that is GRAS Generally Recognized As Safe by the FDA Food and Drug Administration eliminating concerns about incidental contact with food or drinking water The solution filled bottle simulates the temperature of other stored liquids Velcro and a magnetic strip are provided to mount the bottle to the inside of a refrigerator freezer and to mount the display on the outside The micro thin probe cable permits refrigerator freezer doors to close on it TEMPERATURE DISPLAY MODES Ambient Temperature Display Mode IN Indicated by IN appearing on the display displays the current ambient temperature and either the ambient temperature memory MIN MAX or the ambient temperature alarm limits LO HI Probe Temperature Display Mode OUT Indicated by OUT appearing on the display displays the current probe temperature and either the probe temperature memory MIN MAX or the probe temperature alarm limits LO HI
7. he probe temperature OUT is outside the probe temperature alarm limits no alarm will sound RESET BUTTON The RESET button located on the back of the unit may be used to simultaneously reset the memory and alarm limits for both the ambient IN and probe OUT sensor To reset the unit press the RESET button When the reset button in pressed the following will occur The display will show all segments for approximately 5 seconds The ambient IN and probe OUT minimum maximum memory will be cleared The ambient IN and probe OUT alarm limits will be returned to the factory default value of 10 C LO and 30 C HI The unit will display the ambient IN temperature in C ON OFF BUTTON The ON OFF button located on the back of the unit may be used to turn the unit off when the thermometer is not being used The minimum maximum memory and alarm limits are not cleared when the unit is turned off To turn the unit off press the ON OFF button Each successive press of the ON OFF button will turn the unit on or off No alarms will sound and no minimum maximum memory will be recorded when the unit is off BENCH STAND The unit is supplied with a bench stand that is a part of the back of the unit To use the bench stand locate the small openings on either side of the bottom of the unit Place your fingernail into one of the openings and flip the stand out To close the stand simply snap it shut WALL MOUNTING THE UNIT
8. le the alarm to sound when the temperature being displayed is outside the alarm limits press the button located on the back of the unit until A appears on the display Each successive press of the A button will enable or disable the alarm ALARM SOUNDING With the alarm enabled 4A the unit will sound an alarm when the temperature being displayed is outside the alarm limits that have been set for that display mode equal to or lower than the low alarm set point or equal to or greater than the high alarm set point If the alarm is sounding based on the low alarm limit LO will flash on the display If the alarm is sounding based on the high alarm limit HI will flash on the d isplay 2 3 The unit will continue to alarm and flash the corresponding LO or HI until the temperature being displayed is within the alarm limits Once the temperature being displayed is within the alarm limits the alarm will stop sounding and the corresponding LO or HI will no longer flash To silence an alarm manually press any button When manually silenced the corresponding LO or HI will continue to flash on the display until the temperature is within the alarm limits When enabled A the alarm will only sound if the temperature being displayed is outside the alarm limits that have been set for that display mode IN or OUT Example If the unit is displaying ambient temperature IN and t
9. the Setting Probe OUT Temperature Alarm Limits section 8 Enable the alarms See the Enable Disable Alarms section 9 Clear the minimum and maximum memory See the Clearing the Probe OUT Temperature Memory section 10 Select the probe temperature with Min Max memory display mode See the Temperature Display Modes section The alarm limits have been set and the alarm has been enabled The display has been set to show the current temperature inside the refrigerator along with the minimum and maximum temperature that has been achieved inside the refrigerator If the temperature inside the refrigerator goes outside the alarm limits equal to or lower than the low alarm set point or equal to or greater than the high alarm set point the alarm will sound If no one is available to hear the alarm sounding overnight and the temperature returns to an in range condition the memory will provide a record of the single lowest and highest temperature achieved When the temperatures are recorded into the manual log the memory will allow the user to see that the temperature inside the refrigerator had gone outside of the acceptable range Monitoring Procedure Example Keep a notebook or spreadsheet as a manual log 1 Atthe same time every day record the following into the manual log Current Date and Time Current Temperature Reading Minimum Temperature Reading MIN Maximum Temperature Reading MAX
10. utside the probe temperature alarm limits no alarm will sound SETTING PROBE OUT TEMPERATURE ALARM LIMITS 1 Select the Probe Temperature Display Mode by pressing the IN OUT button located on the back of the unit until OUT appears on the display 2 Press MODE button until LO and HI appears 3 To set the low temperature alarm limit LO press the LO ALM SET button to advance the display until the desired low temperature is displayed Press and hold the LO ALM SET button to rapidly advance the display 4 To set the high temperature alarm limit HI press the HI ALM SET button to advance the display until the desired high temperature is displayed Press and hold the HI ALM SET button to rapidly advance the display SETTING AMBIENT IN TEMPERATURE ALARM LIMITS 1 Select the Ambient Temperature Display Mode by pressing the IN OUT button located on the back of the unit until IN appears on the display 2 Press MODE button until LO and HI appears 3 To set the low temperature alarm limit LO press the LO ALM SET button to advance the display until the desired low temperature is displayed Press and hold the LO ALM SET button to rapidly advance the display 4 To set the high temperature alarm limit HI press the HI ALM SET button to advance the display until the desired high temperature is displayed Press and hold the HI ALM SET button to rapidly advance the display ENABLE DISABLE ALARMS To enab
Download Pdf Manuals
Related Search
Related Contents
N°4 - Le Canard en Chainon パットウォール取扱説明書 TDSHーBA 保管用 Philips Soft-shell case DLM6304 Mounting and Operating Instructions EB 8330 EN Honda Power Equipment BF130A Boat User Manual LDRCheck-D Explorer® séries 8000 et 8300, ENP et ENP HD - Affaires Manuel d``utilisation Hannspree ST19KMAW 19" HD-Ready Red LCD TV Copyright © All rights reserved.