User`s Manual - Oak Tree Health
Contents
1. The HF and thermometer icon appears Temperature is too high outside the required range of 50 F 104 F 10 C 40 C This alerts users that an incorrect result may occur if the test continues The LF and thermometer icon appears Temperature is too low outside the required range of 50 F 104 F 10 C 40 C This alerts users that an incorrect result may occur if the test continues Relocate the meter to a location with temperature between 50 F 104 F 10 C 40 C Relocate the meter to a location with temperature between 50 F 104 F 10 C 40 C 42 Chapter 6 Technical Information Specifications Brand name Range Response time Memory sets Operating condition Relative Humidity Storage and transportation condition Relative Humidity Blood sample Hematocrit Hct Power Battery life Display dimension Device dimension H x Wx D Weight Principles Software via USB EasyMax NG Blood Glucose Meter 20 600 mg dL 5 seconds 480 test results 50 F 104 F 10 C 40 C R H lt 90 4F 122 F 20 C 50 C R H lt 90 0 6 uL Fresh blood from fingertip 20 6096 Alkaline Battery 2 ct Over 2000 tests 1 4 x 1 7 inches 35 0 x 43 0 mm 3 7 x 2 0 x 0 8 inches 94 x 50 x 19 5 mm 1 68 oz without batteries 47 7 1 gram Electrochemical biosensor technology GlucoManager H Limitations The test strips are used for fresh capillary whole bl
2. 14 Setting the Time and Date Setting the current time and date on your meter is important if you use the meter memory Eum HN CE M A Nal INI 1 Press d Power Button 3 The last 2 digits of the year flashes 4 Repeat step 3 to set the date and hold the button until at the top of the display Press and time The flashing field is the meter turns on Left button or gt Right button to the one you are currently 2 Press gt Right button for adjust the year and press 0 setting 2 seconds to get into Power Button to confirm the setting mode setting 15 Using EasyMax 15 Blood Glucose Test Strips Use only with EasyMax NG or EasyMax LTC Blood Glucose Meters EasyMax NG Self Monitoring Blood Glucose System is a no code system and does not require test strip calibration EasyMax NG Self Monitoring Blood Glucose System utilizes glucose dehydrogenase chemistry to avoid incorrect readings on patients receiving Oxygen therapy Run a control solution test every time you open a new box of strips See Chapter 2 Control Solution Testing After you take a test strip out of the vial tightly close the vial immediately to keep the test strips dry Use the test strip within three minutes after taking it out of the vial The strip is for single use only Do not reuse it When you open a new test strip vial check the expiration date The test strips are good for six months from the initial date the vial is opened
3. Cleaning Frequency 9 cycles day X 365 days year X 4 year product life 13 140 cycles of cleaning Disinfection Frequency 1 cycle week X 52 weeks year X 4 year product life 208 cycles of disinfections 20 000 validated cycles of cleaning and disinfection for the meter 14 600 validated cycles of cleaning and disinfection for the lancing device 4 year product life is for properly cleaning and disinfection After 4 years the meter must be replaced with a new meter 37 Maintenance and Testing EAsYMax NG n n LU DD DO D D D Your meter needs little or no maintenance with normal use It automatically tests its own systems every time you turn it on and lets you know if something is wrong See Screen Messages and what to do about them To make sure the display is working properly turn off the meter Press and hold power button to see the complete display All the indicators should be clear and look exactly like the picture to the left If not please contact Customer Service at 866 994 3345 38 Screen Messages And Troubleshooting Never make treatment decisions based on an error message If you have any concerns call the Customer Care Service toll free at 866 994 3345 Eastern Standard Time Mon Fri 8 00AM 6 00PM Message What it means What to do Humidified Used strips The meter has detected a problem with the test strip Low power The meter batteries do not have enough power to perform
4. a test Repeat the test with a new strip and puncture site Refer to pages 25 27 for information on sample application Replace with new batteries 39 Message What it means What to do System error Replace the batteries first There may be a problem with the meter Refer to pages 14 and 32 If this error message appears again please contact Customer Service Memory Error Replace the batteries first If ERROR 005 appears again please contact Customer Service No result in memory You can still perform a blood glucose non The test proceeds incompletely The test and get an accurate test result meter was unable to recall this result o g gt D A0 Message What it means What to do Insufficient Volume error The volume of blood or control solution is NOT enough Test result is higher than 600 mg dL The test result is lower than 20 mg dL Repeat the test with a new strip puncture site and enough blood control solution Refer to pages 25 27 for information on sample application Re check your glucose level If the glucose result is HI again obtain and follow instructions from your healthcare professional without delay This may require immediate treatment according to your healthcare professional s recommendations Although this message could be due to a test error it is safer to treat first and then do another test 41 Message What it means What to do
5. again after you replace the battery 1 5V AAA x 2 batteries are available at most stores You may take the old battery with you for replacement Remove the battery when you will not be using the meter for one month or more 32 Cleaning and Disinfecting the EasyMax NG Meter in LTC Settings Choosing a disinfectant According to Guideline for Disinfection and Sterilization in Healthcare Facilities in 2008 you should clean your glucose meter after every use to prevent any possibility of cross infection Wipes used should be EPA registered as a disinfectant Many EPA registered disinfectant wipes have 2 minute claims If your wipes have higher or lower kill times you should adjust your technique accordingly Cleaning and Disinfection Instructions Please keep the meter free of dirt dust bloodstain and water stains After every use follow both the cleaning and disinfection instructions below using the EPA registered wipes Meter cleaning area Possibility of Code Name contact Uih biodi A Strip Slot High B Front Case Medium C Side Case High D Front Buttons High E Back Case Low l F Battery Cover Medium 33 Step 1 Cleaning Instruction Clean first to remove any heavy soil left on the surface of the meter All blood and bodily fluids must be thoroughly cleaned from surfaces and objects before disinfection with a germicidal wipe Open unfold and use the first germicidal wipe to remove heavy soi
6. meter will also void the warranty Always keep the meter clean and store it in a safe place Protect the meter from direct sunlight to ensure a longer lifespan You should not store the meter and test strips in a car a bathroom or a refrigerator Keep the meter test strips and lancets away from children and pets You should not test critically ill patients with home use blood glucose meters Critically ill patients should not be tested with blood glucose meters See page 9 for more information m Warning for potential biohazard Healthcare professionals using this system on multiple patients should be aware that all products or objects that come in contact with human blood even after cleaning should be handled as if capable of transmitting viral disease Note Do not touch the strips with wet hands m Do not use expired strips the expiration date is shown on the vial m Do not bend cut or twist the strips m Altitude up to 10 000 feet above sea level has no effect on readings Health Related Information m Ifthe patient is experiencing dehydration frequent urination low blood pressure shock or hyperosmolar hyperglycemic nonketotic coma HHNKC you may get a test result that is lower than the actual blood glucose level We recommend confirming the result on such patients via a clinical laboratory m If you have followed the steps in the user s manual but still have symptoms that don t seem to match your test results or if
7. or until the expiration date whichever comes first Record the date you first open the test strip vial in the designated spot on the vial Store the test strip vial and your meter in a cool dry place between 36 F 86 F 2 C 30 C Do not freeze Insert the test strip into the meter before applying blood or control solution Do not touch the test strip with wet hands Do not bend cut or twist the test strips 16 Chapter 2 Control Solution Testing When to Run a Control Solution Test Running a EasyMax 15 Control Solution Test will let you know that your meter and test strips are working properly You should run control solution tests If you are using a new EasyMax NG Blood Glucose Meter When you open a new vial of EasyMax 15 test strips If you think the meter or test strips may be working incorrectly If the meter has been dropped If the results are lower or higher than expected and you have repeated the test with similar results When you practice the test procedure 17 About Control Solution Use only with EasyMax 15 test strips and EasyMax NG or EasyMax LTC meters Write the date you first open the Control Solution bottle on the label The control solution is good for three months from the date the bottle is opened or until the expiration date on the bottle whichever comes first Do not use a control solution that is past the expiration date The control solution can stain clothing If you spill it wash yo
8. seconds and will show the result 20 Understanding Control Solution Results The label on your test strip vial shows the acceptable ranges for the different control solution levels The result you get should be inside the acceptable range for the appropriate control solution level Make sure you compare the result to the correct level of control When the control result is inside the range on the test strip vial your test strips and your meter are working properly If your control solution result is not within the acceptable range printed on your test strip vial look at the troubleshooting checks on next page Note In the Control Solution mode control Solution values will NOT be included in the average memory 21 Troubleshooting Checks Was the test strip exposed to open air for a long period of time Action Repeat the control test with properly stored strips Was the test strip vial capped tightly If it wasn t the strips could be humidified Replace the test strips Is the meter functioning well You can use control solution to verify the meter s functions Chapter 2 Is the control solution expired or contaminated Replace with new EasyMax 15 control solution to check the performance of your EasyMax NG meter Were test strips and control solutions stored in cool dry places Repeat the control test with properly stored strips and or control solutions Did you follow the testing st
9. 2 Put a test strip into the meter in the 4 Please wait at least 5 seconds warm water Rinse and dry direction of the arrow The meter until the intended area is dry thoroughly You can also use an turns on and the icon shows and clean before using the alcohol prep pad to clean the site automatically lancing device If you are testing others a new 3 When the icons of and 5 Position the lancing device pair of clean gloves should be flashes on the display the testing and press it firmly against the worn before testing each patient mode of blood drop is starting puncture site 25 6 Gently squeeze and or massage your fingertip until a round drop of blood forms on your fingertip O Apply blood to the edge of the test strip f O Do not apply blood on top of the test strip y 778 Be sure to get enough blood on strip to make it to the confirmation window 7 Touch the blood drop at the tip of the transparent window of the test strip Do not put blood on top of the strip Be sure to get enough blood on the strip s reaction zone Otherwise an error code will appear 26 9 Push the Strip Ejector to eject the test strip Dispose of both strip and lancet properly 10 After discarding wash hands thoroughly with soap and water Rinse and dry thoroughly 11 Clean and disinfect the meter following the instruction in Chapter 5 8 The meter starts to count down from 5 seconds and then displays th
10. Control Solution Test 17 About Control Solution EE 18 Running a Control Solution Test 19 Understanding Control Solution Test Results sse nemen 21 Chapter 3 Testing Your Blood Glucose emen enne nnne nennen 23 Instructions for Using Single use lancets with a Lancing Device 29 Inserting a Lancet Into The Lancing Device ee eene nnne 23 Running a Blood Glucose Test with Blood from Your Fingertip se 25 Understanding Your Test Results ect coe dic 28 Unusual Test Results ooccccccccccccococonnconnnonnnnnnonononnnnnccnnnannnnononnnonnnrnnrnnnnnnnnnnonnnnnnrnnnrnnnenanonenennnonnnenarineraninns 28 Chapter 4 Meter Memory Setup c cccccceeeeecceceeeeceeeaeeeeeeceaeceeeesecaeceseescaaeeeeesecaeeeeeeseeaeeeeseeenseeeess 29 Memory Storing Test Result itis canicas 29 Viewing and Deleting Test Results itte tret eet ier iter nr trei iren fne 30 Using With Your Computer 2 2 ood e ias 31 Chapter 5 Maintenance and Troubleshooting e emen 32 Battery Information Zieser er 32 Cleaning and Disinfecting the EasyMax NG Meter in LTC Gettings 33 Cleaning And Disinfection Freouencn enm ener nene ennemi enne nnne enne 37 Maintenance and Testing aiheen aea aai pias 38 Screen Messages and Troubleshooting A 39 Chapter 6 Technical Information oooooooniccccnnnonocccccncononononcccncnnnnnnnno ene cn cnn cnn rn nn caen nn cnc naar rn nn nnne
11. EASYMax NG Next Generation NG Self Monitoring Blood Glucose System User s Manual Please read this User s Manual thoroughly before using your blood glucose meter Introduction Welcome to the EasyMax Next Generation NG Self Monitoring Blood Glucose SMBG System This system is dependable amp easy to use utilizing a compact lightweight and portable meter that will aide you in monitoring you or your patients blood glucose levels Please read this manual thoroughly before you begin testing It provides you and your diabetes care team important information and step by step directions for use of the EasyMax NG Self Monitoring Blood Glucose System Intended Use The EasyMax NG Self Monitoring Blood Glucose Test System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertip palm or forearm Testing is done outside the body In Vitro diagnostic use The EasyMax NG Blood Glucose System can be used by a single patient or for multiple patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control When used in Long Term Care Settings this system must be used with safety lancets The system is not to be used on neonates nor for the diagnosis of or screening for diabetes mellitus Important Safety Instructions All parts of the kit are considered biohazardous and may transmit infection even if you have performed cleaning and dis
12. call the Customer Care Service toll free at 866 994 3345 Eastern Standard Time Mon Fri 8 00AM 6 00PM P N 718063742 Rev A01 46
13. e test result Note Used test strips and lancets are considered bio hazardous waste in accordance with U S amp local regulations and should be handled as if capable of transmitting infection Users may discuss proper methods for disposing of used test strips and lancets with their doctor 27 Understanding Your Test Results Expected Values The EasyMax 15 Blood Glucose test strips are plasma referenced and calibrated for easier comparison to lab results The American Diabetes Association recommends a post meal glucose level of less than 140 mg dL and a pre meal glucose of less than 100 mg dL American Diabetes Association Diagnosis and Classification of Diabetes Mellitus Position Statement Diabetes Care 34 Supp 1 S66 2011 Unusual Test Results If the results are inconsistent with you or your patient s symptoms please follow these steps 1 Run a control solution test Chapter 2 Control Solution Testing 2 Repeat a blood glucose test Chapter 3 Testing Your Blood Glucose 3 After running the quality control and retesting your blood glucose if there are still concerns about the performance of the meter call EasyMax Customer Care at 866 994 3345 for further troubleshooting or if it is an emergency contact your healthcare professional 4 For accuracy and precision data and for important information on limitations see the instructions that come with your test strips Note 1 Extremely high humidity may affect the t
14. eps properly Read Chapter 2 Control Solution Testing and test again Stop using the meter if you continue to obtain inaccurate results Please contact the EasyMax Customer Care Service line at 866 994 3345 for further troubleshooting 22 Chapter 3 Testing Your Blood Glucose Instructions for using single use lancets with a Lancing Device You must first load the lancet into the lancing device to get it ready for use 1 Unscrew the Cap 2 Insert the lancet into the lancing 3 Put the front cap back on device firmly then twist off the and turn clockwise to protective cover secure Note Lancets are for single use only and a new sterile lancet should be used each time you perform a test 23 5 Pull on the sliding barrel of the 6 Set the lancing device 4 Select the desired penetration aside until later in the test lancing device until it clicks and depth then release Now the lancing device is ready Do not prick your finger until your meter and strip are prepared Note 1 Select 1 2 for soft or thin skin 3 4 for average and 5 6 for thick or calloused skin 2 Lancing device and lancets are not to be shared between users Sharing lancing devices and lancets may transmit bloodborne pathogens such as viral hepatitis 24 Running a Blood Glucose Test with Blood from Your Fingertip D I KA FED D I DD D C DXX D I 1 Wash your hands with soap and
15. est results A relative humidity above 90 may cause inaccurate results 2 Hematocrit below 20 may cause higher results Hematocrit above 60 may cause lower results 28 Chapter 4 Meter Memory Setup Memory Storing Test Results Your meter stores a maximum of 480 test results with the time and date of the test You can review them at any time When the memory is full the oldest result is dropped as the newest is added so it is very important to have the correct time and date set in the meter The memory function is necessary in order to transmit data to a central location separate software amp instructions are available Note Us 2 o Do not change your therapy based on one individual result in memory The memory is not lost when you replace the battery You do need to check that the time and date are still correct See Section Setting the Time and Date in Chapter 1 Once 480 results are in memory adding a new result causes the oldest one to be deleted In the Control Solution mode control solution values will NOT be included in the average memory 29 Viewing and Deleting Test Results Each review requires the user to go back to the main screen testing mode by pressing the power button DM Ee Vi t 7t ut al Is TO t 3 To delete a test result press Left button for more than 2 seconds and display shows dEL press to 1 Press the Left button or gt 2 F
16. infection Wash hands thoroughly with soap and water after handling the meter Follow Standard Precautions when using the meter and obtaining a sample A new pair of gloves should be worn by the user before testing each patient The meter should be disinfected after each use The EasyMax NG Self Monitoring Blood Glucose Test System may be used for testing multiple patients when Standard Precautions and the disinfection procedures included in this manual are followed EasyMax NG Blood Glucose Meter Accessories EasyMax 15 Blood Glucose Test Strips EasyMax 15 Level 1 Low Control Solution EasyMax 15 Level 2 Normal Control Solution User s Manual Logbook 2 AAA batteries Carrying case For further information please see Biosafety in Microbiological and Biomedical Laboatories BMBL http www cdc gov biosafety publications bmbl5 Protection of Laboratory Workers From Occupationally Acquired Infections Approved Guideline Third Edition Clinical and Laboratory Standards Institute CLSI M29 A3 Guideline for Isolation Precautions Preventing Transmission of Infectious Agents in Healthcare Settings 2007 http www cdc gov hicpac 2007ip 2007isolationprecautions html FDA Public Health Notification Use of Fingerstick Devices on More than One Person Poses Risk for Transmitting Bloodborne Pathogens Initial Communication 2010 http www fda gov MedicalDevices Safety AlertsandNotices ucm224025 htm CDC Cl
17. inical Reminder Use of Fingerstick Devices on More than One Person Poses Risk for Transmitting Bloodborne Pathogens 2010 http www cdc gov injectionsafety Fingerstick DevicesBGM html Why is it so important to test blood glucose regularly Testing blood glucose regularly can make a big difference in how you manage you or your patients day to day diabetes care We ve made EasyMax NG Self Monitoring Blood Glucose System as simple as possible to help you use it regularly Do you need help If you have questions or need assistance please call the Customer Care Service toll free at 866 994 3345 Eastern Standard Time Mon Fri 8 00AM 6 00PM Important Information about your new EasyMax NG Meter The EasyMax NG Blood Glucose Meter is designed and approved for testing fresh capillary whole blood samples from the fingertip palm and forearm The meter is for in vitro use ONLY for testing outside the body It should not be used to diagnose diabetes Remove batteries if the meter will not be used for one month or more The EasyMax NG Blood Glucose Meter can only be used with EasyMax 15 Blood Glucose Test Strips Other test strips will give inaccurate results EasyMax 15 Blood Glucose Test Strips feature a unique and accurate glucose dehydrogenase chemistry system Testing is not valid for neonatal blood specimens Do not disassemble the meter as this may cause damage to the components resulting in incorrect readings Disassembling the
18. l Step 2 Disinfection Instruction After cleaning and before using the meter again unfold a new wipe and thoroughly wet all meter surfaces including the strip port and the connection port Treated area must remain visibly wet for a full 2 minutes Use additional wipe s if needed to ensure a continuous 2 minute wet contact time Let the device air dry for 30 seconds Do W Make sure the meter is turned off during cleaning and disinfection W Keep test strip vial s tightly closed when performing cleaning and disinfection procedures the fumes from the disinfectant may affect the performance of the strips W After cleaning and disinfection please perform a physical appearance and performance check of devices W Always wear gloves when cleaning and testing Note Do Not Get any moisture into the test strip slot Spray any cleaning solution directly onto the meter Put the meter under water or liquid Pour liquid onto the meter After disinfection users gloves should be removed and hands should be thoroughly washed with soap and water before proceeding to the next patient 34 Performance check of the meter Check item After each cleaning Does the meter turn on Does the meter beep Does the meter have stored results Is the display clear Can you read the result After each disinfection 1 Insert test strip Does the meter turn on 1 Press and hold power button for 3 sec 2 Press and release left or
19. nennen nnns 43 Ree iia eee e die uiae eie 44 iure cct PP 44 Device Informatlon ET 46 lucu AM 46 Chapter 1 Understanding Your Meter EasyMax NG Blood Glucose Meter Test Strip Slot Insert test strip here Display Shows results messages and results stored in memory Power Button Press to confirm the setting Left Button Press to enter memory adjust setting and scroll through results Strip Ejector Push the Strip Ejector to remove the strip Battery Door Flip open the battery door by pushing the tab in the direction of the arrow and pulling the door up Right gt Button Press to enter memories adjust setting and scroll through results 12 The EasyMax NG Accessories EasyMax 15 Blood Glucose Test Strip Electrode Em Strip Insert Direction Hand Hold Area Reaction area m Blood collection area N Test Strip Bottle amp Control Solution Bottle A EASYMax 25 Blood Glucose Test pdfacturer EPS Bio Tee Jf No 706400008 Rev AOI a Expiration Date 13 Inserting Batteries A 1 Open the battery door on the back of the meter by pushing the tab in the direction of the arrow and pulling the door up 2 Insert two AAA batteries The meter will beep to confirm the batteries are inserted correctly 3 Put the battery door back in place and snap it closed
20. ood samples 1 2 3 4 5 DO NOT use neonate blood sample Extreme humidity may affect the results A relative humidity greater than 9096 may cause incorrect results The system should be used at temperatures between 50 F and 104 F 10 C and 40 C Outside this range the system may get incorrect results DO NOT reuse test strips The test strips are for single use only Hematocrit The hematocrit between 20 and 60 will not affect the results Hematocrit below 20 may cause higher results Hematocrit above 60 may cause lower results 6 Altitude up to 10 000 feet above sea level has no effect on readings Healthcare Professionals Please note these additional Limitations 1 If the patient has the following conditions the result may fail Severe dehydration Severe hypotension low blood pressure Shock Astate of hyperglycemic hyperosmolar state with or without ketosis Lipemic samples Cholesterol level up to 500 mg dL and triglycerides up to 3 000 mg dL do not affect the results Grossly lipemic patient samples have not been tested and are not recommended for testing with the EasyMax NG Blood Glucose Meter Critically ill patients should not be tested with home use blood glucose meters DO NOT use during xylose absorption testing Xylose in the blood will interfere with the Self Monitoring Blood Glucose System Interfering Substances depend on the concentration EPS EasyMax NG mfg has tested ab
21. ove the therapeutic levels and at or above the suggested testing levels of the EPA for many common substances 44 Based on this chart many of these substances do not have an effect on EasyMax NG blood glucose results Blood Concentration mg dL Suggestion tested Result EPS Exogenous Interferent Therapeutic Conc conc by EPA tested cane 7 4 Acetaminophen 172 __053 J 009 GentisicAcid___ 18 J A 18 ___ Glipzide 02 0 2 EDopa HBO 0 1 0 Tolbutamide 5310 PA B mbimuneonugaed 12 w Cholesterol 300 2 950 Creatinine 15 X 1550 330 Triglycerides 190 150250 1000 UricAcid 7 14 39 J 5119 Fructose 230 9 Galactose 10 10 J 20 9 Device Information EasyMax NG SMBG System EasyMax 15 Level 1 Low Control Solution EasyMax NG Blood Glucose Meter EasyMax 15 Level 2 Normal Control Solution EasyMax 15 Blood Glucose Test Strip Manufacturer EPS BIO TECHNOLOGY CORP No 8 R amp D RD IIl Hsinchu Science Park Hsinchu Taiwan 30077 Oak Tree Health warranty During the warranty period if your EasyMax NG Self Monitoring Blood Glucose System does not work for any reason other than obvious abuse Oak Tree Health will replace it with a new system or an equivalent product free of charge Please
22. right button Does the meter turn off Does the meter turn on button for 3 sec Does the meter beep Does the meter have stored results right button Is the display clear Can you read the result Action If any of the results are No the user should call the Customer Care Service toll free at 866 994 3345 Eastern Standard Time Mon Fri 8 00AM 6 00PM for assistance 2 Run control solution test Any reading Is the reading within control range 35 Physical Appearance check of the meter after each cleaning or disinfection Check item Accept Result Is the display clear to read Is the screen clear Yes Is the strip slot and other parts free of corrosion Yes Action If any of the results are No the user should call the Customer Care Service toll free at 866 994 3345 Eastern Standard Time Mon Fri 8 00AM 6 00PM for assistance Meter Life when Cleaned and Disinfected The meter has been tested through 20 000 validated cycles 10 000 cycles of cleaning and 10 000 cycles of disinfection If the meter is used for 9 cycles of cleaning and 9 cycles of disinfection per day it will last in professional use for 3 years 36 Cleaning And Disinfection Frequency At least 9 cycles of cleaning per day and 1 cycle of disinfection per week will allow the user to use the device for 4 years The meter can sustain 20 000 cycles of cleaning and disinfection and the lancing device can sustain 14 600 cycles of cleaning and disinfection
23. rom the main screen press Right button to view averages Left button for more than 2 of 7 14 30 90 day test results first seconds to view the control solution Ge and then all results records in order from 30 to 1 4 Press Left button or Press back to the testing Press back to the testing mode Right button to keep mode reviewing the results 30 Using With Your Computer To Transfer data the meter can be turned on or turned off The USB mini port can be found on the bottom of your meter below the power left and right buttons EASYMax NG EASYMax NG EASYMax NG 1 Use the USB cable to connect 2 The meter starts to transfer 3 When all data is uploaded from your meter and computer The data to the GlucoManager the meter to the PC the meter display shows PC software shows OK and beeps In the software select ndividual Management 31 Chapter 5 Maintenance and Troubleshooting Battery Information The meter uses two alkaline 1 5V AAA batteries Batteries will normally last for more than 2000 tests Other types of 1 5V AAA batteries are also acceptable but the capacity of test times may differ Install the batteries when you first use the meter or replace with new batteries when the LP low power message and the low battery symbol appear on the display Note 1 2 3 The meter won t delete earlier records after you replace the battery You should reset the time and date
24. ur clothes with soap and water Close the bottle tightly after every use Left over control solution should not be returned to the control solution bottle Store the bottle of control solution at room temperature between 36 F 86 F 2 C 30 C Do not freeze 18 Running a Control Solution Test You need the meter a test strip and control solution EASYMax NG 3 Place the meter on a flat surface like a table 4 Remove the control solution bottle cap and wipe the tip of j the bottle with a tissue 1 Put a test strip into 2 The meter turns on automatically and the strip i 5 Squeeze the bottle until a the meter in the icon appears tiny drop forms at the tip of direction of the Press Left button to change the testing mode to the bottle and place the arrow Control Solution and the E icon will flash solution on top of bottle cap Note Setting the meter in control solution mode will prevent the control solution reading from being added to your memory and averages 19 6 Touch the drop to the blood 7 Do not remove the test strip until you 8 Push the strip ejector to collection area at the tip of the confirm that the reading falls within eject the test strip into a test strip Do not put control the range printed on the test strip vial proper trash receptacle solution on the top of test Be sure to match the correct control strip The meter starts to count solution range to your test result down from 5
25. you have questions please call the Customer Care Service toll free at 866 994 3345 Eastern Standard Time Mon Fri 8 00AM 6 00PM m Healthcare personnel should wash hands amp change gloves between patients even if patient dedicated testing devices and single use lancing devices are used Note Warning for potential biohazard Healthcare professionals using this system on multiple patients should handle all products or objects in contact with human blood carefully to avoid transmitting viral disease even after cleaning Explanation of Meter Symbols Date ru rut Luce cuc on the left side M Memory Time nni Record Average on the right side NJ JM display Result Apply Blood Apply Control Solution Temperature Application of blood or control mg dL Unit solution Insert a strip ER Battery Table of Contents Chapter 1 Understanding Your Meter 12 EasyMax NG Blood Glucose Meter 12 The EasyMax NG Accessories eee d ed Lens ttes lias oed dece Nene dde d eode d audeat e a eee iced 13 Inserting Batteries ccoo M O NN 14 Setting the Time and Date iii aen eee te eben ie ite qu e E din eek gun e ene Re pela Se A BER Yee 15 Using EasyMax 15 Blood Glucose Test Strips oooocococcococccccncncnonnnnnnnno cnn nnnn conan r rr nnnr em emen nen 16 Chapter 2 Control Solution Test ener ESAE EEEn nen nn nen nnns nennen sse n enne nns 17 When to Run a
Download Pdf Manuals
Related Search
Related Contents
小型で高利得です。 Texto de la Etiqueta del Producto Samsung Samsung S6102 Manual de utilizare SYSTEM VII Access Control Software: User Manual DEWALT DWE4887N Use and Care Manual Manuale di installazione, uso e manutenzione I nstallation Avaya IP Telephony BCM Rls 6.0 User's Manual Guia_QUÍMICA NA AGRICULTURA AZTEC On-Line Communications – Getting Started Guide Philips HD3620 Home draft system Copyright © All rights reserved.